Complex Care Hub Manual: Pressure Ventilation on the LTV 1200
|
|
- Audra Jennings
- 5 years ago
- Views:
Transcription
1 Complex Care Hub Manual: Pressure Ventilation on the LTV 1200 Table of Contents 1 Mechanical Ventilation Important information for support worker Commonly used terms Inspiration Expiration Breath rate (machine set breaths) Initiate Sensitivity Triggered breaths Peak Inspiratory Pressure (PIP) Positive End Expiratory Pressure (PEEP) Pressure Pressure Control breaths Pressure support breaths Continuous Positive Airway Pressure (CPAP) Synchronised Intermittent Mandatory Ventilation (SIMV) Monitored Information (monitored data) Ventilator controls Silence Reset Scroll button On/Standby Power Oxygen The breathing circuit Humidifier (mechanical) Inspiratory limb Heater wire Temperature probe Expiratory limb Water trap (only in home use wet circuits) Exhalation valve Expiration drive line Wye Tracheostomy connector Sensor lines Oxygen inlet Dry circuit (no mechanical humidification) Troubleshooting common alarms How to silence and turn off the flashing alarm message Page 1 of 17
2 10 Care of the ventilated child Start of shift equipment checks Standard checks of a ventilated child Hourly checks Changing and cleaning the circuit Cleaning the home circuit To clean the exhalation valve Cleaning the inlet filter Version 4.0 Date revised: 2017 Next due: 2019 This core section only includes basic information on the LTV 1200 ventilator features that are relevant to the usual daily care provided by a support worker to a child at home. It does not include all features found on the ventilator. Further information can be obtained from the manufacturers User Manual or by contacting the Clinical Technologists at The RCH. Page 2 of 17
3 1 Mechanical Ventilation Mechanical ventilation describes the use of a machine (ventilator) to give breaths to someone who can t breathe at all, or to support the breathing of someone who can t breathe well enough on their own. 2 Important information for support worker The support worker will not be required to change any of the settings on the ventilator. This is only to be done by trained medical professionals or appropriately trained family members of the child and always in consultation with the child s medical team. The only buttons the support worker will need to use are: Silence/Reset, Scroll button and On/Standby. These buttons will be further explained in your training. 3 Commonly used terms 3.1 Inspiration Describes the act of breathing in or taking a breath 3.2 Expiration Describes the act of breathing out. 3.3 Breath rate (machine set breaths) The minimum amount of breaths given in one minute as set on the ventilator. Nurses may describe these as mandatory breaths or machine set breaths. The child may be able to take more breaths than the mandatory breaths, but if the child isn t breathing on their own then the machine will deliver the set breaths. 3.4 Initiate Initiate (start) describes what causes a breath to be given by the ventilator. Breaths may be initiated by the child or by the ventilator according to the set breath rate 3.5 Sensitivity The machine has a special function that recognises the child s effort when taking a breath. When the child tries to take a breath for themselves, the air or gas in the ventilator tubing moves. For the ventilator to recognise when the child is trying to take a breath we need to make it sensitive to this gas flow in the ventilator tubing. When a large enough gas flow is detected the ventilator will respond with a breath. The ventilator can be made very sensitive or not at all sensitive by changing the sensitivity dial. We like the ventilator to be sensitive to the child s effort to breathe, but if it detects interference (e.g. water in the tubing or a larger leak around the tracheostomy tube) it may not recognize the child s breathing efforts or deliver too many breaths. An example: 1 = very sensitive this is usually set for smaller patients 4= less sensitive- this is usually set for older children or young adults Page 3 of 17
4 3.6 Triggered breaths A triggered breath is when the ventilator senses the child is trying to breathe and responds by pushing gas into the lungs to deliver a breath. A triggered breath can also be called an assisted breath. 3.7 Peak Inspiratory Pressure (PIP) During inspiration the ventilator delivers a set pressure to inflate the lungs. For every machine breath this will be the highest pressure that is reached during inspiration. This highest pressure is called PIP. It is important that the child receives the correct PIP as they can become unwell if the PIP is too high or too low. 3.8 Positive End Expiratory Pressure (PEEP) PEEP is the pressure that is held in the airways at the end of expiration. Everyone has a natural PEEP when they breathe out, this pressure helps to stop the small air sacs in the lungs from completely collapsing. By setting a PEEP on the ventilator we guarantee the child s lungs are always kept slightly open at the end of expiration, this makes it easier for the child to take each breath. 3.9 Pressure The amount (pressure) of gas delivered to the child from the ventilator to help the lungs inflate 3.10 Pressure Control breaths Pressure control describes the pressure that is delivered to the child for the set breath rate (3.3). The pressure delivered (PIP) is a combination of the number seen on the pressure control dial and the number seen on the PEEP dial. For example Pressure control is set at 15 PEEP is set at 5 Which means that PIP is 20 ( = 20) 3.11 Pressure support breaths These are the breaths that the child triggers that are above the set breath rate (3.3) on the LTV All pressure supported breathes are triggered by the child. Usually the pressure delivered to the child in a support breath is less than the pressure delivered during a pressure control breath. This means that the child has a smaller breaths. The LTV 1200 can be programed to deliver only pressure supported breaths Continuous Positive Airway Pressure (CPAP) This is a mode of ventilation on the LTV In this mode the ventilator will deliver a continuous pressure which means the child s lungs/ airways are always kept slightly open. Keeping some pressure in the lungs continuously makes breathing easier for the child. Imagine the lungs are like a balloon, it is harder to blow up a brand new balloon but it gets easier once there is some air in the balloon. It is the same with the lungs if there is no air in the lungs the child has to work hard to get enough air into them, but if there is a little air in the lungs then the child doesn t have to work as hard to get enough air into them. CPAP can be delivered/given by mask or tracheostomy. CPAP can only be used by patients who are able to breathe on their own. Page 4 of 17
5 3.13 Synchronised Intermittent Mandatory Ventilation (SIMV) This is a mode of ventilation on the LTV In SIMV mode the ventilator gives a set number of breaths and can also recognise when a child is trying to take a breath. It is able to synchronise, or give a breath at the same time as the child initiates a breath. If the child does not take a breath within a set time frame then the ventilator will give a mandatory breath. Although the child always receives a set number of mandatory breaths, they are able to take extra breaths. In order to support the child who takes extra breaths SIMV is often used in conjunction with Pressure Support. 4 Monitored Information (monitored data) This table contains data that is displayed on the ventilator. The support worker does not have to record this data and it is included in the manual for information only. Display Monitored Data Units of measurement Explanation PIP Peak Inspiratory Pressure cmh20 The PIP displays the greatest pressure measured during the inspiratory phase MAP Mean Airway Pressure cmh20 The MAP displays a running average of the airway pressure for the last 60 seconds. PEEP Positive End Expiratory Pressure cmh20 The PEEP monitor displays the pressure in the circuit at the end of expiration. F Total breath rate Breaths per minute Displays the breaths per minute based on the last 8 breaths, and includes all breath types Vte Exhaled Tidal Volume millilitres Tidal volume indicates how much air is exhaled in one breath. VE Minute Volume Litres The minute Volume displays how much air is exhaled over the last 60 seconds as calculated from the last 8 breaths I:E I:E ratio Smaller unit normalised to 1 The I:E Ratio displays the unit-less ratio between measured inspiratory time and measured expiration time Page 5 of 17
6 5 Ventilator controls The only buttons the support workers will need to use are: 5.1 Silence Reset Press once to silence and then press once again to reset the alarm when the issue/problem has been resolved. 5.2 Scroll button Located on the top left hand corner of the ventilator. This takes you through the messages in the display window 5.3 On/Standby It is to be used when transferring the child between ventilators e.g. night ventilator and day ventilator. The support worker should only touch this button when the child is not attached to the ventilator. The support worker may be required to turn off the ventilator if it is not in use. To do this press the control lock to off and press and hold the On/Standby button. Once the ventilator has turned off a constant alarm will sound and can be silenced by pressing and holding the silence/reset button 6. Power Plug the ventilator into mains power whenever possible to ensure the internal battery is fully charged at all times Page 6 of 17
7 The ventilator has an internal battery that will power the ventilator for up to 1 hour when fully charged. It is recharged when the ventilator is plugged in to any external power source (including the external battery) In emergencies, or when the child is being transported, the ventilator can be plugged directly into a 12 volt DC supply such as a car cigarette lighter (you will need correct attachment). You must always ensure the car motor is running and the car is in a well ventilated place before plugging in the ventilator or running the ventilator in the car The external battery will run the ventilator for up to 5 hours when fully charged (i.e. each battery is equal to 2.5hours). It is important to remember that battery times can vary. In the event of a power failure the ventilator will automatically change to either the external or internal battery (which ever has the most charge remaining). If the external battery is flat the ventilator will audibly and visually alarm and automatically transfer to the internal battery. You must find a source of external power as soon as possible (and not longer than 1 hour) or the battery will run out of charge. The External Power indicator shows the level of the external power while the ventilator is operating from an external power source When the ventilator is running from the internal battery, the External Power indicator is off External power may be provided by connecting the ventilator to an external battery or LTV AC Power Adapter. When running from external power the Charge Status indicator will show the following levels: LED Colour Green Yellow Red Power Level External Power level is acceptable External Power level is low or checking power level External power is about to run out 7. Oxygen From time to time the child may require oxygen to be delivered through the ventilator circuit. This can be done from a low flow source such as an oxygen concentrator or oxygen cylinder. Oxygen from either source is mixed in the ventilator circuit with air from the ventilator. Oxygen is a medication and the amount of oxygen is prescribed by the child s medical team In the home, the Low Pressure O2 source button on the ventilator should be lit up at all times. This ensures that the ventilator can deliver Page 7 of 17
8 room air (21% Oxygen) with or without low flow home oxygen attached to the circuit. 8. The breathing circuit Most ventilator s will have a standard set up, though the tubing may look different depending on what has been given to the child. Picture of full wet circuit and the connection from ventilator to humidifier 8.1 Humidifier (mechanical) The humidifier actively warms and moistens the air. It is connected to the ventilator and the child by the inspiratory limb. This is called a wet circuit. Water is held in the humidifier chamber and heated to create water vapour. The humidifier is ideally positioned below the child. There are two types of chambers: A reusable chamber that is refilled by disconnecting the circuit tubing (without the heating wires) and pouring water into the chamber through the opening. A single patient use, automatically filling chamber with a reusable water bag. In order for water to flow from the bag to the chamber, the chamber must be level and the bag must be hung at least 50cm above the chamber. Water bags can be used for two weeks before being replaced. In a dry circuit the mechanical humidifier is removed and a special attachment called a Humid-vent must be added to provide humidification. Page 8 of 17
9 MR850 humidifier displayed, other types of humidifiers are also in use in the community, see specific instructions for the child s humidifier. 8.2 Inspiratory limb This tubing delivers the air to the child from the ventilator. It is plugged into the ventilator and then connects with the humidifier chamber. A separate piece of tubing goes from the humidifier to the child. 8.3 Heater wire The tubing that goes from the humidifier to the child will usually have a heating wire in it. The heater wire maintains the temperature of the air as it travels from the humidifier to the child. This prevents the air from cooling and reduces rain out (formation of condensation) within the tubing. The heater wire is connected to the humidifier by a heater wire adaptor (not disposable). The heater wire adaptor for the reusable circuits is different to that of the single patient use circuits. 8.4 Temperature probe This plugs into the inspiratory limb both at the patient end and the humidifier end of the tubing. This senses the temperature of the air at each end of the tubing so that the humidifier can safely maintain the correct temperature. The temperature probe at the patient end must be positioned in an upright position to stop it from getting wet. 8.5 Expiratory limb This tubing goes from the child to the expiration valve. It allows the expired air to be removed from the child. Page 9 of 17
10 Expiratory limb, water trap, expiration valve, sensor lines, wye 8.6 Water trap (only in home use wet circuits) This is a small removable jar with a self-sealing valve in the expiratory limb of the ventilator circuit. It is used to collect the condensed water (rain out) that gathers in the unheated expiratory limb and should be emptied regularly. Single patient use circuits do not require a water trap because the expiratory section of the circuit is heated by the humidifier to reduce rain out. 8.7 Exhalation valve This controls the flow of air during inspiration and expiration. It is also used to control the pressure left in the lungs when the child breathes out (PEEP). It is connected to the expiratory limb and the expiration drive line. 8.8 Expiration drive line This line is used to control the operation of the expiration valve. Shutting it during inspiration and then opening it to control the pressure left in the lungs when the child breathes out (PEEP). The line connects the expiration valve to the ventilator. At the ventilator end this connects to the port marked Exhl Valve. 8.9 Wye The inspiratory and expiratory limbs connect to this; the Wye is connected to the patient. It measures flow and pressure using the sensor lines Tracheostomy connector This connector has swivels in two directions to minimise the effect of the circuit pulling on the tracheostomy tube. The patient end of the connector should be visibly clean at all times. The tracheostomy should be cleaned in warm soapy water, rinsed and dried at least once per week or more often if required. Wye with sensor lines, top of inspiratory limb (with temperature probe) top of expiratory limb. Tracheostomy connector Sensor lines Sensor lines monitor the flow of air to and from the patient. They go from the ventilator to the Wye (please do not try to remove them from the Wye). They must always point upwards to stop them from getting wet. At the ventilator end they connect to the ports marked Flow Xdcer. The sensor lines cannot connect to the wrong ports. Page 10 of 17
11 Top of picture shows the three Sensor lines connecting to the ventilator. The bottom connector is the inspiratory limb connected to the ventilator Oxygen inlet This is used to add in extra oxygen to the circuit. It can be added to the circuit on the inspiratory side between the ventilator and humidifier Dry circuit (no mechanical humidification) In a dry circuit, mechanical humidification is replaced with a humidivent which is a special filter to moisten and warm the air. The humidivent is inserted between the top of the wye and the tracheostomy connector. The humidivent is single use and should be replaced daily or if soiled with secretions. Dry circuits are usually used when taking the child out of the house (e.g. shopping), or when the child is up and about in their wheelchair. See the child specific care page for the times when the child uses a dry circuit. 9. Troubleshooting common alarms The ventilator has alarms that are set to alert the support worker that there is a problem. When an alarm goes off you must always check the child before you take any further action, including silencing the alarm. When an alarm occurs you will See a flashing alarm message in the display window e.g. Disc/Sense Hear an alarm tone Page 11 of 17
12 See the control that is connected to the alarm flashing e.g. if the High Pressure alarm goes off then the High Pressure control on the ventilator will flash When the cause of the alarm is fixed The alarm tone stops The alarm message will continue to flash until you press the reset button The control button will continue to flash until you press the reset button 9.1 Possible alarms The table lists the alarm displays that you will see on the machine and what you may need to do (after checking the child) to fix the cause of the alarm. A complete list of alarms is contained in the LTV 1200 manual. Alarm message What it might mean What to do (if required) Disc/sense High PRES (high pressure) Low Pressure alarm Tubing has come off the tracheostomy. A connection along the tubing has come apart Sensor lines loose or disconnected Child is coughing Child is upset and crying Kink in tubing Partial/complete blockage of tracheostomy or humidivent Water in tubing Tubing has come off the tracheostomy A connection along the tubing has come apart Hole in the circuit Large leak around the tracheostomy Accidental decannulation Put tube back on to tracheostomy Check tubing for broken connection and fix Monitor child & suction if required. If tracheostomy completely blocked, change tracheostomy Comfort child if upset. Remove kinks (may need to replace circuit). Replace humidivent Remove any water that may be in tubing Reconnect the tubing Reconnect the tubing or do a circuit change Find the leak and repair/replace tubing Check the ventilator circuit and tighten any loose connections Check the child; ensure the tracheostomy ties are tight. Page 12 of 17
13 Bat Low The ventilator has been running on internal battery and it is running out of power Plug the ventilator into mains power within 5 mins Apnoea The child has stopped breathing Commence DRSABC If the child is ok and still breathing, troubleshoot the circuit and check tracheostomy is clear. Bat Empty LMV (Low Minute Volume) Locked The internal battery level is dangerously low The child is breathing out (exhaling) less volume than the machine has been programmed for (low minute volume alarm) Tracheostomy leak/disconnection Increased secretions The LTV is in locked mode and the settings can t be changed You must plug the ventilator into mains power immediately. If you don t the child will require bag and mask ventilation Check position of tubing and tracheostomy. Try to stop the pulling on the tubing. Check that the child s chest is rising and falling equally on the left and right side Keep airway clear with suctioning Inform the child s parents if alarm continues despite troubleshooting measures. Do not unlock the machine. Support workers are not allowed to change the settings on the ventilator How to silence and turn off the flashing alarm message Once you have seen the child is safe you may press the silence/reset button once. When you have fixed the cause of alarm you can press the silence/reset button again to stop the flashing alarm message. Repeating Alarms If the child has multiple events of the same alarm with no obvious cause this may indicate an illness with the child or a problem with the ventilator. Please tell the parents if the machine is repeatedly showing the same alarm and the support worker is not able to fix the problem with the usual trouble shooting methods outlined above. It is helpful to make note of what the child is doing during the repeated alarms as this can help medical staff troubleshoot the problem. 10 Care of the ventilated child 10.1 Start of shift equipment checks Ventilator Page 13 of 17
14 Make sure ventilator circuit is attached to child (via tracheostomy or mask) Check that the settings and alarm limits are correct Starting from the ventilator check the circuit for leaks and ensure connections are tight Check for water in the circuit and drain back to the humidifier or water trap Empty the water trap Make sure the ventilator is plugged into the mains power Check that all the batteries are plugged into the ventilator and that they are fully charged Humidifier Check that humidifier is on the right setting (is it mask or tube ventilation?) Check the water level and add cooled boiled water if needed Check the temperature probe on the wye is facing up Is the humidifier at the right temperature? Oxygen If oxygen is required check if there are enough oxygen equipment (O2 cylinders, concentrators, tubing, connectors) for your shift and the next shift. Suction Check that suction is available, working and fully charged or charging. Resuscitation equipment Make sure the resuscitation equipment is available and working Make sure the emergency tracheostomy kit is available and fully stocked (if required) If any of the equipment used for ventilating the child is not working, or needs servicing or repairs, please inform the parents. The parents will need to arrange to return the equipment to the ICU clinical technologists Standard checks of a ventilated child Along with equipment checks, there are a number of other checks that the support worker should attend to at the beginning of the shift. The support worker should get a handover from the family or support worker in attendance and together you should check the following: Is the child breathing normally (for them)? o o o o Is the breathing faster or slower than normal? Are they working harder to breathe than normal? Is the breathing softer or shallower than normal Can you hear normal breath sounds? Page 14 of 17
15 Do the settings on the ventilator match the ventilation order for the child? Does the child need suctioning? Is the tracheostomy secure? Is the ventilator circuit properly connected to the tracheostomy and lower than the tracheostomy? Is the ventilator plugged into mains power? If not then how long has the ventilator been on battery power? Is there enough water in the humidifier? Is the ventilation tubing free of water? ( rain out or condensation). If not, drain the water out, always keeping the circuit lower than the child and their tracheostomy (so water doesn t enter the tracheostomy). You may need to disconnect the circuit from the child to do this and they will need manual bagging while you drain the water out Hourly checks During the shift the support worker needs to complete standard checks of the child every hour (or more if required). Check the child looks/acts like their usual self Does the child need suctioning? Is the ventilator circuit properly connected to the tracheostomy? Check the pressures on the ventilator Is there enough water in the humidifier? Is the ventilation tubing free of water? ( rain out or condensation) Check the child specific care pages for any other checks required. During the support worker s shift they may be required to record information from the ventilator. The child specific care page will tell you what you need to record and when to record it. 11 Changing and cleaning the circuit A disposable circuit is changed every week. The temperature probe is not disposable but the rest of the circuit can be thrown out. These are used in hospital and generally not used at home. A home use circuit needs to be changed and washed weekly (or more frequently if required) Cleaning the home circuit Make sure circuit is removed from the ventilator remove heater wire from the inspiratory limb Wash the inspiratory and expiratory limbs in warm soapy water. Rinse well and allow to dry completely in room air Wipe temperature probe Put heater wire back in the inspiratory limb when it is dry Page 15 of 17
16 Page 16 of 17
17 11.4 To clean the exhalation valve Disassemble and remove the diaphragm. Wash gently in warm soapy water Rinse well and leave to dry Store the clean dry reassembled circuit in a clean, dry pillowcase Cleaning the inlet filter Remove the inlet filter by gently pinching the foam filter and pulling it out Inspect the filter for damage and replace if necessary Hand wash the filter using warm soapy water (a mild dish liquid detergent) Rinse the filter thoroughly in warm water to remove all detergent Allow filter to air dry before replacing it into the ventilator Spare filters can be obtained from the Clinical Technologists at The RCH Caution: do not install a wet or damp filter into the ventilator, as this could result in damage to the ventilator. Page 17 of 17
Manual: Biphasic Positive Airway Pressure (BiPAP) Ventilation
RCH@Home Manual: Biphasic Positive Airway Pressure (BiPAP) Ventilation 1. Commonly used terms... 2 1.1 Inspiration... 2 1.2 Expiration... 2 1.3 Breath rate (bpm)... 2 1.4 Ventilation... 2 1.5 Biphasic
More informationpatient education program 4156 South 52 nd Street, Omaha, NE ChildrensOmaha.org/HomeHealthcare
patient education program 4156 South 52 nd Street, Omaha, NE 68117 800-747-7334 402-734-6741 ChildrensOmaha.org/HomeHealthcare We know children CHILDREN S HOME HEALTHCARE Trilogy Ventilator Instructions
More informationVentilator Training Module
Ventilator Training Module LTV (Lap Top Ventilator) Teaching Script LTV User guides: http://www.carefusion.com/documents/guides/user-guides/rc_ltv-1000_ug_en.pdf http://www.carefusion.com/documents/guides/user-guides/rc_ltv-1100_ug_en.pdf
More informationASSISTANCE. Cardinal Health Pulmonetic Systems Medina Rd., Suite 100 Minneapolis, Minnesota
ASSISTANCE Cardinal Health Pulmonetic Systems 17400 Medina Rd., Suite 100 Minneapolis, Minnesota 55447-1341 Customer Care: (800) 754-1914 (763) 398-8500 Fax: (763) 398-8403 Website: www.cardinalhealth.com/viasys
More informationAUTOVENT 4000 VENTILATOR
OVERVIEW AUTOVENT 4000 Only properly trained and approved Escambia County Bureau of Public Safety Paramedics are to use the AutoVent 4000 ventilator manufactured by LSP to transport patients already on
More informationOperation of Oxylog 3000 ventilator
Operation of Oxylog 3000 ventilator Overview Suitable for patients from approximately 7Kg body weight upwards (smallest tidal volume 50mls). Entirely power dependant- will not ventilate if no available
More informationOperation and Maintenance of the EPV200 Portable Ventilator
Operation and Maintenance of the EPV200 Portable Ventilator 1 Applications of the EPV200 The EPV200 Portable Ventilator is a gas powered electronically controlled mechanical ventilator, designed to provide
More informationMedical Instruments in the Developing World
2.2 Ventilators 2.2.1 Clinical Use and Principles of Operation Many patients in an intensive care and the operating room require the mechanical ventilation of their lungs. All thoracic surgery patients,
More informationVirginia Beach EMS. Oxylator EMX. Debra H. Brennaman, RN, MPA, NREMT-P
Virginia Beach EMS Oxylator EMX Debra H. Brennaman, RN, MPA, NREMT-P Oxylator EMX Overview Patient responsive oxygen powered resuscitation / ventilation device intended to provide emergency ventilatory
More informationHallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting
Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting This is a quick reference for setting up, using, and troubleshooting the Hallowell EMC 2000 and 2002 Small Animal Ventilator.
More informationThe Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)
The Crossvent 2i+ 1. How is this ventilator classified 2. Ventilator Concept (brief theory of operation and features) -Your Two Choices with this Ventilator 3. An overview of the device (an in-service)
More informationComplex Care Hub Manual: Oxygen Therapy
Complex Care Hub Manual: Oxygen Therapy Table of Contents RCH@Home Manual: Oxygen Therapy... 1 1. What is oxygen therapy?... 2 2. Symptoms of low oxygen... 2 3. Giving oxygen at home and in the community...
More informationAutomatic Transport Ventilator
Automatic Transport Ventilator David M. Landsberg, MD, FACP, FCCP, EMT-P Luke J. Gasowski, RRT, NPS, ACCS, CCP-C, FP-C Christopher J. Fullagar, MD, FACEP, EMT-P Stan Goettel, MS, EMT-P Author credits /
More informationLP6 Plus Volume Ventilator -And- LP10 Volume Ventilator With Pressure Limit
LP6 Plus Volume Ventilator -And- LP10 Volume Ventilator With Pressure Limit Clinician s Manual Important Note: Read this manual in its entirety before using the ventilator. Keep this manual for future
More informationInstructions for Use. OptiLife Nasal Lab Mask
Instructions for Use OptiLife Nasal Lab Mask Intended Use The OptiLife Nasal Mask is intended to provide an interface for application of CPAP or bi-level therapy to patients. The mask is for multi-patient
More informationCPR MEDICAL DEVICES INC. OPERATING MANUAL FOR RESUSCITATOR
CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR OXYLATOR FR-300 RESUSCITATOR Symbols used in the Operating Manual for the Oxylator FR-300 and/or on the Oxylator FR-300 unit: WARNING! Do not use grease or
More informationAutomatic Transport Ventilators. ICU Quality Ventilation on the Street.
Automatic Transport Ventilators ICU Quality Ventilation on the Street. Kevin Bowden, March 20 th 2014 Ventilator Definition A ventilator is an automatic mechanical device designed to provide all or part
More informationPERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:
PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: 1. **Identify and name the filters on the 7200ae. 2. **Explain how each filter is sterilized. 3. **Trace the gas flow through the ventilator
More informationTracheostomy and Ventilator Education Program Module 11: Emergency Preparedness
Tracheostomy and Ventilator Education Program Module 11: Emergency Preparedness Disclaimer This material is intended for use by trained family members and caregivers of children with tracheostomies who
More informationAchieva. Clinician s Manual. Ve nt i l at or with Flow Acceleration Limit and Apnoea Backup
Achieva Clinician s Manual Ve nt i l at or with Flow Acceleration Limit and Apnoea Backup Nellcor Puritan Bennett Inc. is an affiliate of Tyco Healthcare. Achieva and Puritan Bennett are registered trademarks
More informationBasics of Mechanical Ventilation. Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity
Basics of Mechanical Ventilation Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity Overview of topics 1. Goals 2. Settings 3. Modes 4. Advantages and disadvantages
More informationTechnical Data and Specifications
Technical Data and Specifications INTENDED USE Ventilator designed to provide Invasive and Non-invasive ventilation for the critical care management of adult, pediatric and neonate-infant (including premature)
More informationS/5 Aespire. Quick Reference Guide
S/5 Aespire Quick Reference Guide Anesthesia system controls 1 Turn on the system 2 7100 controls and monitored data 3 Menu map 4 Set the ventilation mode 5 Set ventilator controls 6 Vaporizer controls
More informationHONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS
RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS 1. INTRODUCTION An anaesthesia delivery system includes any machine, equipment or apparatus which supplies gases, vapours, local anaesthesia and/or
More informationCPR MEDICAL DEVICES INC. OPERATING MANUAL
CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR OXYLATOR EMX RESUSCITATION AND INHALATION SYSTEM Symbols used in the Operating Manual for the OXYLATOR EMX and/or on the OXYLATOR EMX unit: WARNING! Do not
More informationHow to use the SAVe II Ventilator
TRAININGGROUNDS How to use the SAVe II Ventilator Designed to elevate the standard of care and improve triage capabilities simply select the height of the patient to deliver a targeted tidal volume. This
More informationSimplified Automated Ventilator Training Slides
Simplified Automated Ventilator Training Slides Export Department: J.D. Honigberg International, Inc. 155 N. Pfingsten Road, Suite 150 Deerfield, IL 60015 USA Tel: 847-412-0200 Email: medical@jdhintl.com
More informationMy BiPAP ventilator. My helpful guide
My BiPAP ventilator My helpful guide Inhale. Exhale. And everything in between. How does it work? Brief Overview The lungs are vital organs that bring fresh oxygen into the body and remove carbon dioxide
More informationUNDERSTANDING NEONATAL WAVEFORM GRAPHICS. Brandon Kuehne, MBA, RRT-NPS, RPFT Director- Neonatal Respiratory Services
UNDERSTANDING NEONATAL WAVEFORM GRAPHICS Brandon Kuehne, MBA, RRT-NPS, RPFT Director- Neonatal Respiratory Services Disclosures Purpose: To enhance bedside staff s knowledge of ventilation and oxygenation
More informationOperation and Maintenance of the MCV100 and MCV100-B Portable Ventilator
Operation and Maintenance of the MCV100 and MCV100-B Portable Ventilator 1 Applications of the MCV100(B) The MCV100 and MCV100-B Mass Casualty Ventilators are electronically controlled portable ventilators,
More informationO-Two Self-Study Guide. e600 Transport Ventilator Ventilation Modes
O-Two Self-Study Guide e600 Transport Ventilator Ventilation Modes e600 VENTILATION MODES The e600 ventilator has 13 ventilation modes: Manual Ventilation, Controlled Mandatory Ventilation (CMV), Assist
More informationWelcome to the Specialized Medical Services Respiratory training webinar series!
Welcome to the Specialized Medical Services Respiratory training webinar series! SMS is your Long Term Care (LTC) facility single source for oxygen, medical equipment, respiratory care services and supplies
More informationMy CoughAssist. A patient guide to CoughAssist T70. Please visit
Philips Healthcare is part of Royal Philips Electronics Europe, Middle East, Africa +49 7031 463 2254 How to reach us www.philips.com/healthcare healthcare@philips.com Latin America +55 11 2125 0744 Asia
More informationRecommendations on Checking Anaesthesia Delivery Systems
Page 1 of 11 Recommendations on Checking Anaesthesia Delivery Version Effective Date 1 Oct 1992 (reviewed Feb 07, Feb 02) 2 2004 3 Nov 2011 4 Dec 2016 Document No. HKCA T1 v4 Prepared by College Guidelines
More informationA few seconds after turning on, the screen changes to Set Up Screen Before adjusting the pressure, calibrate the O 2 sensors
SiPAP Set-Up Plug into mains, air & O 2, then setup the circuit as shown below Set-up & test before spiking water bag A few seconds after turning on, the screen changes to Set Up Screen Before adjusting
More informationHallowell EMC Anesthesia WorkStation Operating Manual
Hallowell EMC Anesthesia WorkStation Operating Manual PN 000a2770 Anesthesia WorkStation A compact (9" W x 9" D x 15" H), convenient system for use on animals under 7 kg (15 lb.) Hallowell Engineering
More informationSATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products
Saturn Evo SATURN EVO COLOR A N A E S T H E S I A S Y S T E M i n t e l l i g e n t s i m p l i c i t y Manufacturer of high quality anaesthesia products SATURN EVO COLOR PATIENT SAFETY AND USER PATIENT
More informationC-1 C-7. C-6 Short air tubing Upper Velcro straps. C-a C-5 C-2 C-4 C-3 B-1 B-9 B-10 B-2 B-7 B-8 B-3 B-6 D-1 D-2 D-13 D-3 D-12 D-10 D-11 D-4 D-5 D-8
A Forehead support Forehead pads C Mask Frame Mask Elbow Cushion clip Mask Cushion C-7 C-1 Vent cover Ports cap Cuff ActiveCell C-6 Short air tubing Upper Velcro straps Swivel Lower Velcro straps Headgear
More informationAcornVac Vacuum Plumbing Systems - Trouble Shooting Guide
AcornVac Vacuum Plumbing Systems - Trouble Shooting Guide 1. Accumulators Problem Accumulator is overflowing If Accumulator continues to overflow Correction 1. Push and hold the manual activation button
More informationInstruction Manual for Configura Cushionair Portable Pump
Instruction Manual for Configura Cushionair Portable Pump Fitted with battery powered pump, suitable for Configura Portable chairs V E R S I O N O N E M A Y 2 0 1 6 Contents Introduction 3 Set up of Cushionair
More informationUpdate to RS-232 commands. Changing patient from NIV to INVASIVE Vent Type. SNDF command
Changing patient from NIV to INVASIVE Vent Type Table 9 shows automatic settings changes that occur when changing the same patient from NIV to INVASIVE Vent Type. Table 9: Automatic settings changes NIV
More informationAirox Supportair Ventilator
Airox Supportair Ventilator A versatile hospital ventilator with a titration platform that enables a smooth transition to the patient s home care ventilator Titration Platform The Supportair ventilator
More informationBunnell LifePulse HFV Quick Reference Guide # Bunnell Incorporated
Bunnell Incorporated n www.bunl.com n 800-800-4358 (HFJV) n info@bunl.com 436 Lawndale Drive n Salt Lake City, Utah 84115 n intl 801-467-0800 n f 801-467-0867 Bunnell LifePulse HFV Quick Reference Guide
More information2015 Hamilton Medical AG. All rights reserved. Printed in Switzerland.
HAMILTON-T1 Quick Guide This Quick Guide is intended as a useful reference for ventilation of adult and pediatric patients. It does not replace the clinical judgment of a physician or the content of the
More information2 PRE USE FUNCTION CHECK 3 OPERATING PROCEDURE 4 SERVICING. 5 CAREvent ALS ACCESSORIES 11
THE RESUSCITATOR MUST BE THOROUGHLY CLEANED AFTER EACH PATIENT USE. 1. Operate CAREvent ALS Handheld Resuscitator to blow out any contaminant from the patient valve. 2. Ensure CAREvent ALS Handheld Resuscitator
More informationOxylog 3000 plus Emergency & Transport Ventilation
Oxylog 3000 plus Emergency & Transport Ventilation Offering high ventilation performance with features such as AutoFlow, integrated capnography and non-invasive Ventilation, the compact and robust Oxylog
More informationOxygen Dialflow Meter. Instructions for Use
Oxygen Dialflow Meter Instructions for Use 702-0031.12 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient,
More informationSTANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA
STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA SOP#: 1157.1 Date Issued: 05/14 Date Revised: 5/15 Page 1 of 6 TITLE: SCOPE: RESPONSIBILITY: PURPOSE: SurgiVet
More informationQUICK REFERENCE GUIDE
cm H O 2 cm H O 2 cm HO 2 PSI cm H O 2 ON OFF UPPER LIMIT LOWER LIMIT UPPER LIMIT LOWER LIMIT LIFE PULSE HIGH-FREQUENCY VENTILATOR QUICK REFERENCE GUIDE 01388-08.11 MONITOR PIP JET VALVE ALARMS READY SILENCE
More informationUnderstanding Tracheostomy Care
Understanding Tracheostomy Care Inside this guide: This guide will help you learn how to take care of your tracheostomy (trach). It is important to ask questions. You will be given time to learn. Working
More informationSLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters
SLE4000 Infant Ventilator with touch-screen operation When the smallest thing matters SLE4000 - The Total Solution for Conventional Infant Ventilation SLE is a world leader in the design and manufacture
More informationHealth Professional Info
Health Professional Info Mouthpiece Ventilation (MPV) What is MPV? MPV is a less intrusive form of noninvasive ventilation that uses a portable home mechanical ventilator (HMV) with a single-limb open-circuit
More informationElectronics Respironics Lifecare PLV-102 / PLV-102 Ventilator To t a l S o l u t i o n This lightweight ventilator is well suited for transport and portable applications. Controlled by a microprocessor,
More informationHome Oxygen Therapy Patient Handbook
Home Oxygen Therapy Patient Handbook Office Hours 8:30 a.m. to 5:00 p.m. Monday-Friday Emergency Answering Service 24 hours a day, 7 days a week Birmingham Office: (205) 981-2333 5025 Highway 280 Suite
More informationFlow meter. bellow vaporizer. APL valve. Scavenging system
Introductory Lecture Series: The Anesthesia Machine PORNSIRI WANNADILOK Objectives Anesthesia Machine Ventilators Scavenging Systems System Checkout 1 Flow meter ventilator bellow vaporizer Corrugated
More informationCompleted downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo
Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo Link full download: http://testbankcollection.com/download/pilbeams-mechanicalventilation-physiological-and-clinical-applications-5th-edition-test-bank-cairo
More informationLiquid Oxygen- Helios & Helios Marathon A Guide for Patients in the Home
Liquid Oxygen- Helios & Helios Marathon A Guide for Patients in the Home Contact Name: Phone number: Table of Contents Contents Receiving Liquid Oxygen in the Home... 2 Instructions for Liquid Oxygen-Helios
More informationTest your PICU/ NICU/ CHSU Knowledge
Test your PICU/ NICU/ CHSU Knowledge 1. On the LTV 1150 you have a 12-year-old trach patient. You are adding in 2 liters of 100% Oxygen through the flow inlet. The patient s spontaneous respiratory rate
More informationR409 Plus Veterinary Ventilator Machine
R409 Plus Veterinary Ventilator Machine User Manual v1.2 CONTENTS 1 User/Owner Responsibility... 1 2 Warnings and Cautions... 1 3 Warranty... 2 4. Introduction... 2 5 Unpacking... 4 6 Schematic diagram
More informationCATALOGUE. Adult and Infant Respiratory Care Catalogue
CATALOGUE Adult and Infant Respiratory Care Catalogue TABLE OF CONTENTS F&P 850 System Adult Care Adult Invasive Ventilation Adult Noninvasive Ventilation Adult Optiflow and Oxygen Therapy 6 F&P 850 System
More informationTafonius Junior A USERS GUIDE. v10mar2011
Tafonius Junior A USERS GUIDE J2 v10mar2011 Notes: Printed: 28/Mar/11 Page 2 of 19 DOCA4962 Table of Contents An Overview...5 Junior sans gas...5 Junior with gas...7 Controller Boot Up Sequence...9 INITIALISING
More informationSLE5000 Neonatal Ventilator with High Frequency Oscillation
SLE5000 Neonatal Ventilator with High Frequency Oscillation The SLE5000 is superb: the best High Frequency Ventilator I ve used When the smallest thing matters SLE5000 - The Total Solution for Infant Ventilation
More informationUnderstanding Tracheostomy Care for your Child
Understanding Tracheostomy Care for your Child 2 Inside this guide: This guide will help you learn how to take care of your child s tracheostomy (trach). It is important to ask questions. You will be given
More information200 PSI HIGH-FLOW AIR SOURCE KIT
200 PSI HIGH-FLOW AIR SOURCE KIT 50% Duty Compressor on 2.0 Gallon Air Tank PART NO. 20008 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of
More information200 PSI FAST-FILL AIR SOURCE KIT
200 PSI FAST-FILL AIR SOURCE KIT 55% Duty Compressor on 2.0 Gallon Air Tank PART NO. 20007 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of
More informationContents Warnings, Cautions, and Notes Contents of the Package Introduction to the Harmony S/T A Tour of the Harmony S/T...
S/T Patient Guide 1004221 JR 3/15/01 Contents 1 Warnings, Cautions, and Notes... 3 Warnings... 3 Cautions... 4 Temperatures... 4 Intended Use/Indications for Use... 5 Contraindications... 5 1 Contents
More informationEnglish. ivolve Full Face Mask. User Manual
ivolve Full Face Mask User Manual 0123 Table of Contents ivolve Full Face Mask... 1 Intended Use... 1 Before Using the Mask... 2 Parts of the ivolve Full Face Mask... 4 Getting the Right Cushion Size...
More informationTITLE: Pulse Oximetry COMPETENCY #: Resp #1 NEW COMPETENCY REVISION DATE: 10/18/12 EMPLOYEE NAME. DATE INITIAL RE-EVALUATION ANNUAL (if required) PRN
: Pulse Oximetry COMPETENCY #: Resp #1 NEW COMPETENCY REVISION : 10/18/12 INITIAL RE-EVALUATION ANNUAL (if required) PRN 1. Verifies physician order. 2. Gathers equipment and supplies. 3. Knocks on door
More informationYour patients, our primary focus
Your patients, our primary focus Siesta i Whispa anesthesia machine Functional, fail-safe and flexible to your needs Take a good look, and you ll be surprised by the many forward-thinking features of the
More informationAstral in AirView: Improving patient care through connectivity. ResMed.com
Astral in AirView: Improving patient care through connectivity ResMed.com Using Astral in AirView via the ResMed Connectivity Module (RCM) Astral is ResMed s portable, invasive and non-invasive life support
More informationUnit 15 Manual Resuscitators
15-1 Unit 15 Manual Resuscitators GOAL On completion of this unit, the student should comprehend the proper operation of self-inflating resuscitation bags, flow-inflating resuscitation bags and gas-powered
More informationGE Healthcare. Centiva/5 Critical Care Ventilator. Meet a new level of expectations
GE Healthcare Centiva/5 Critical Care Ventilator Meet a new level of expectations Intergrating performance and value Marry form and function In an environment where time and space are tight, the design
More informationAstral in AirView: Improving patient care through connectivity. ResMed.com
Astral in AirView: Improving patient care through connectivity ResMed.com This guide will assist you with: Setting up the ResMed Connectivity Module for Astral 2 Troubleshooting the ResMed Connectivity
More informationNEONATAL/INFANT VENTILATOR
n a!! i le ion b a a i l ve r s v w a lone o N d-a n sta FEATURES: MDES F VENTILATIN PRESSURE LIMITED/ CNTINUUS FLW FLW TRIGGERED VLUME VENTILATIN NENATAL/INFANT VENTILATR The Crossvent 2i+ ventilator
More informationInstructions/Procedure for Use of the Speaking Valve
Instructions/Procedure for Use of the Speaking Valve Notes on the use of the speaking valve for ventilator-dependent and non-ventilator dependent patients The speaking valve is a one-way valve that allows
More informationCOMPANY PROFILE. Engler manufactures the Sonus V ultrasonic dental unit for the human market
REV. F 10 / 21 / 2014 COMPANY PROFILE Engler Engineering Corporation has been in business since 1964 and occupies an 8000 square foot facility in Hialeah, Florida (USA). We manufacture ultrasonic dental
More informatione700 Automatic Transport Ventilator User Manual e700 User Manual (15PL Rev.14 28_8_2017)
e700 Automatic Transport Ventilator User Manual Table of Contents Chapter 1 Safety... 1 Chapter 2 Intended Use.. 4 Chapter 3 Overview.... 5 3.1 Control and Display Layout...... 5 3.2 Function Keys......
More informationSLE5000 Infant Ventilator with HFO
SLE5000 Infant Ventilator with HFO When the smallest thing matters SLE5000 - The Total Solution for Infant Ventilation Shown on optional roll stand. Ventilator may be mounted either way round on stand.
More informationMicro Dial-Flowmeter. Instructions for Use
Micro Dial-Flowmeter Instructions for Use 702-0082.8 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user
More informationVERTICAL AIR COMPRESSORS
VERTICAL AIR COMPRESSORS MODEL NO: VE15C150, VE18C150, VE25C150 PART NO: 2226010, 2226020, 2226025 OPERATION & MAINTENANCE INSTRUCTIONS LS0715 INTRODUCTION Thank you for purchasing this CLARKE Vertical
More informatione600 Automatic Transport Ventilator User Manual e600 User Manual (15PL Rev.7 28_8_2017)
e600 Automatic Transport Ventilator User Manual Table of Contents Chapter 1 Safety. 1 Chapter 2 Intended Use.. 4 Chapter 3 Overview. 5 3.1 Control and Display Layout........ 5 3.2 Function Keys......
More informationOxygen Dialflow Meter. Instructions for Use
Oxygen Dialflow Meter Instructions for Use 702-0031.9 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user
More informationpro pro 2 AND User Manual
pro 2 AND This BiPAP system is covered by one or more of the following patents: US Patent Nos. 5,148,802; 5,313,937; 5,433,193; 5,632,269; 5,803,065; 6,029,664; 6,305,374; 6,539,940; 5,239,995; Re 35,295;
More informationR E D I C O N T R O L S
R E D I C O N T R O L S Operation & Maintenance Manual Portable Service Purger for Low Pressure Chillers Model: PSP-LP-1B For Refrigerants R-11, R-113, R-114 & R-123 & Other Similar Refrigerants File Literature
More informationMirage Vista Nasal Mask
English 60 50 40 30 20 10 0 4 6 10 12 14 16 1 20 ENGLISH Mirage Vista Nasal Mask Rx Only The Mirage Vista Nasal Mask is an accessory to a non-continuous ventilator (respirator) intended for multipatient
More informationCPAP Devices - Masks - Accessories for respiratory assistance in CPAP
Devices - Masks - Accessories for respiratory assistance in HAROL is one of the Italian companies leader in developing and selling systems, determining the most appropriate solutions for all your needs.
More informationMODEL NUMBER: PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank
IMPORTANT SAFETY INSTRUCTIONS CAUTION - To reduce risk of electrical shock or Electrocution: MODEL NUMBER: 20008 200 PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank IMPORTANT: It is
More informationNegative Pressure Wound Therapy
Guide to Continuous Therapy DeRoyal Pro Series easy use guide to continuous therapy To turn unit On: Press the ON or OK button. To reach option mode to input settings: Hold the up arrow button and the
More informationSiesta i Whispa Integrated technology
Siesta i Whispa Integrated technology Functional, fail-safe and flexible to your needs Take a good look, and you ll be surprised by the many forward-thinking features of the Siesta i Whispa anaesthesia
More informationHome Oxygen Therapy in NHS Scotland. Level 1 Education for Patients & Carers
Home Oxygen Therapy in NHS Scotland Level 1 Education for Patients & Carers National Home Oxygen Service Guidance/Best Practice Recommendations Changes to the National Oxygen Service in Scotland in 2012-2013
More informationVENTILATION SERVO-s EASY AND RELIABLE PATIENT CARE
VENTILATION SERVO-s EASY AND RELIABLE PATIENT CARE Critical Care SERVO-s 3 SERVO-s SIMPLY MAKES SENSE MAQUET THE GOLD STANDARD Leading the way: MAQUET is a premier international provider of medical products
More informationTafonius. Auxiliary Mode Operating Manual A USERS GUIDE
Tafonius Auxiliary Mode Operating Manual A USERS GUIDE v09feb2011 Notes: Printed: 28/Feb/11 Page 2 of 15 DOCA4956 Table of Contents Running Tafonius with the Auxiliary Control System...5 Turning Tafonius
More information1 of 9 10/20/2010 4:45 PM
1 of 9 10/20/2010 4:45 PM This is the ventilator you've been looking for - compact, lightweight, unsurpassed reliability, and self contained. Its comprehensive, versatile features meet the demands of almost
More informationMODEL NUMBER: M20005 AIR SOURCE KIT. 30% Duty Compressor on. 2.0 Gallon Air Tank SAVE THIS MANUAL FOR FUTURE REFERENCE
MODEL NUMBER: M20005 AIR SOURCE KIT 30% Duty Compressor on 2.0 Gallon Air Tank SAVE THIS MANUAL FOR FUTURE REFERENCE USER MANUAL IMPORTANT SAFETY INSTRUCTIONS CAUTION - To reduce risk of electrical shock
More informationSelecting the Ventilator and the Mode. Chapter 6
Selecting the Ventilator and the Mode Chapter 6 Criteria for Ventilator Selection Why does the patient need ventilatory support? Does the ventilation problem require a special mode? What therapeutic goals
More information42045 Heavy Duty ADA Base Model Kit: 85/105 PSI (ADA Compressor Only) Heavy Duty ADA Base Model Kit: 110/145 PSI (ADA Compressor Only)
42045 Heavy Duty ADA Base Model Kit: 85/105 PSI (ADA Compressor Only) 42047 Heavy Duty ADA Base Model Kit: 110/145 PSI (ADA Compressor Only) 45052 Constant Duty ADA Base Model Kit: 85/105 PSI (ADA Compressor
More informationRAM 4021 Operation Manual
RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For your safety...3 Description...3 Set-up mode...4 Annunciator lights/alarms...4 Operation...5 Calibration...6
More informationVENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV
VENTIlogic LS VENTIlogic plus 100 % Mobility and Reliability in IV and NIV VENTIlogic LS VENTIlogic plus Your requirements for reliability and mobility are our benchmark. VENTIlogic LS and VENTIlogic plus
More informationINSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y
INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model 50 Systems Outfitted with ABM-725 Monitor M A R T E C H S E R V I C E S C O M P A N Y OFFICE: (507) 843-4700 P.O. BOX 7079 Toll
More informationFLIGHT 50 Ventilator
FLIGHT MEDICAL INNOVATIONS LTD. FLIGHT 50 Ventilator Operator's Manual LIT-0002 Rev. A00 Safety Instructions General Warnings Legal Notice Disclaimer FLIGHT MEDICAL INNOVATIONS LTD. (FLIGHT MEDICAL) provides
More information