The desflurane (Tec 6) vaporizer: design, design considerations and performance evaluation

Size: px
Start display at page:

Download "The desflurane (Tec 6) vaporizer: design, design considerations and performance evaluation"

Transcription

1 British Journal of Anaesthesia 1994; 72: 474-^479 The desflurane (Tec 6) vaporizer: design, design considerations and performance evaluation R. B. WEISKOPF, D. SAMPSON AND M. A. MOORE SUMMARY We have described the design and design considerations of the desflurane Tec 6 "vaporizer" and have tested its performance characteristics. The vaporizer differs from previous vaporizers designed for anaesthesia in that electromechanical rather than mechanical controls accommodate the different physical characteristics of desflurane. This design, while offering perhaps an increased risk of failure (owing to sophisticated electronic components and circuitry), on the other hand offers the decreased likelihood of accidental delivery of very large concentrations of liquid anaesthetic resulting from tilting or overfilling and alarms and warnings not previously incorporated into the design of anaesthetic vaporizers. The output characteristics of the vaporizer are as expected, based on the design: desflurane concentration output in oxygen has accuracy (±75%) which is similar to that of the mechanical vaporizers; output decreases when nitrous oxide is added owing to the lower viscosity, but remains within 20% of the dial setting or 0.5% absolute. (Br. J. Anaesth. 1994; 72: ) KEY WORDS Anaesthetics, volatile: desflurane. Equipment: vaporizers. The new anaesthetic, desflurane (1,2,2,2-tetrafluorethyl-difluorethyl ether), has a relatively large vapour pressure of 669 mm Hg at 20 C and 757 mm Hg (boils) at 22.8 C. Current temperature-compensated, agent-specific vaporizers were not designed to accommodate anaesthetics with a boiling point or a very steep vapour pressure vs temperature curve within its expected temperature of operation (fig. 1) [1]. This report describes the design, considerations that went into the design of a new "vaporizer" created specifically to deliver desflurane and an evaluation of the accuracy of its output. DESIGN AND CONSIDERATIONS Compatibility It was thought advisable that a new vaporizer should have external design and performance characteristics similar to those of current vaporizers. The vaporizer would have to be able to be mounted, in a plug-on fashion, on the "back bar" of current types of anaesthesia machines. These considerations dictated vaporizer height and width, but allowed modest flexibility in depth. Also dictated was the incorporation of a "safety interlock", which prevents the simultaneous use of more than one vaporizer. Perhaps most important was the need to provide a dial indicating the concentration of desflurane output of the vaporizer. Concentration had to be affected little by changes in background flow, room temperature and background gas composition. The MAC of desflurane (7.25% in oxygen, in humans aged yr) suggested the incorporation of a relatively large sump, while the very low blood-gas solubility (0.42) of desflurane indicated a need to be able to fill the vaporizer during the delivery of an anaesthetic, without having to shut the output of the vaporizer (to avoid substantial decrease of depth of anaesthesia). Size limitations suggested the latter approach. Temperature control The options are either to cool or to heat the liquid anaesthetic to some constant temperature to produce a constant vapour pressure. Failure to do so would result in large variations in temperature of the liquid anaesthetic. If the anaesthetic were cooled it would have been possible to use the same stream-splitting technology now used in the variable bypass vaporizers. However, this was not practical because few commercially available methods for cooling are compatible with the size constraints for the vaporizer, and those that are available either consume much power or are unreliable, or both. In addition, failure of cooling could result in high vaporizer temperatures, with the potential for excessive anaesthetic output, requiring additional potentially cumbersome systems to safeguard against this possibility. Consequently, the alternative approach was taken: to heat desflurane to a constant temperature of approximately 39 C, thus providing a constant vapour pressure of approximately 1460 mm Hg, and therefore delivery of desflurane as a gas. This pressure was chosen as sufficient to overcome the resistances of the internal components of the vaporizer (the "regulator", the rotary valve and passage RICHARD B. WEISKOPF, M.D. Department of Anesthesia, Cardiovascular Research Institute, University of California, San Francisco, CA , U.S.A. DAVE SAMPSON, B.SC., Research and Development, Ohmeda, Steeton, West Yorkshire, U.K. MARK A. MOORE, B.A., University of California, San Francisco, CA, U.S.A. Accepted for Publication: October 21, Correspondence to R.B.W.

2 DESFLURANE VAPORIZER 475 failure, the vaporizer ceases to deliver anaesthetic and the battery provides power exclusively for the warning and status lights for approximately 20 h (see below). The vaporizer draws a maximum current of 2.5 A. Thus if plugged into a multiple socket outlet available on some anaesthesia machines, total consumed current may exceed the combined socket rating if other electrical devices are in operation. 800-, Pb at sea level ex ^ 6003 o FEATURES Q Temperature ( C) 21 I FIG. 1. Vapour pressure of desflurane at temperatures of C. ways within the vaporizer; see fig. 2) and of the breathing system, and also provide an extra margin of safety in the event of partial occlusion of the breathing system. To provide the necessary heat, two elements heat the sump containing the anaesthetic and two more heat the vapour in the outflow path (see below) to prevent condensation. Liquid anaesthetic temperature is monitored by a resistive thermometer device (RTD). The electronics determine the amount of power to be supplied to the heaters. A second RTD monitors die temperature of the heater block within the sump and shuts off power to the heaters if the temperature exceeds 57 C. The vaporizer is powered by 110 V or 220 V a.c. It contains a 9-V battery which is of insufficient power to supply the electricity required for heating, which is variable, but is approximately 50 W during steadystate conditions and 200 W maximum to permit addition of room temperature anaesthetic, without affecting output. Thus in the event of main power The functional aspects of the vaporizer may be thought of as being composed of two portions: (1) components constituting a regulator function (shown in the shaded area of fig. 2); and (2) the other components handling the gases: inlet, fixed restrictor, sump, shut-off valve, manual variable control valve and outlet. The gases comprising the "background flow" follow a path through a fixed resistor, and unlike current Tec 3, 4, and 5 vaporizers, leave the vaporizer without ever having entered the sump containing the anaesthetic. The sump is heated (and, as a result, pressurized). Its output of gaseous desflurane is controlled by a series of transducers, valves and resistors. The "sump shut-off" valve opens when the vaporizer is enabled and the anaesthetist-controlled dial ("manually variable restrictor ") is rotated to any position other than " 0 % " desflurane (as detected by an optical sensor). The sump shut-off valve is closed by any of several conditions (see below), when cessation of delivery of desflurane is desired or mandated. The anaesthetic vapour then passes through the regulator (shown in shaded area of fig. 2) and then through the manually controlled variable restrictor, after which it joins the background gases and leaves the vaporizer. The manually controlled variable restrictor is controlled by the anaesthetist. It comprises a groove Fixed restrictor R1 Fixed restrictor R2 Carrier gas inlet Alarm control electronics Pressure Temperature level Battery Mains system Control electronics LJ I Differential pressure tri sducer - Output Jilt Dial I 1 Interlock L_J solenoid Liquid ll level X, Sump shut-off value Heated sump Variable I restrictor Regulator Manually variable restrictor select Temperature control F I G. 2. Schematic d r a w i n g of the vaporizer courtesy of O h m e d a, I n c..

3 476 BRITISH JOURNAL OF ANAESTHESIA cut on the underside of a plate of the external dial mounted on top of the vaporizer, on which the calibrated scale of concentrations of desflurane appears. Only desflurane vapour passes though this groove. As the dial is turned, the portion of the groove which is exposed to desflurane vapour has an increasingly deep cut, allowing for greater flow. THE REGULATOR The purpose of the regulator is to reduce the pressure of desflurane vapour (1460 mm Hg) leaving the sump (fig. 2, point D) and make the resulting pressure equal to the pressure of the background gas. This is accomplished by an electromechanical system, which was chosen rather than a simple mechanical device for several reasons: (1) the ease of achieving the range of vapour flow (between 2 and 2200 ml min" 1 ); (2) the linear pressure-flow characteristic; (3) size constraints; and (4) electronic control permitted warning, alarm ("operational", "no output", "warm-up" and "alarm battery low") and control functions not otherwise achievable. The differential pressure transducer senses pressure at point B in the background gas flow, upstream of the fixed restrictor and at point E in the desflurane vapour flow. When the pressures are not equal, an electrical signal is sent to the variable restrictor of the regulator to correct the discrepancy. If the pressure in the desflurane flow is less than that in the background flow, the valve opening is increased to allow for greaterflowof desflurane; conversely, if the pressure in the desflurane flow exceeds that of the background gas, the opening of the variable restrictor is decreased, to decrease the flow of desflurane. Because of the fixed, laminar flow restrictor (R2), an increase in background gas flow increases the pressure in the proximal portion of the background flow channel TABLE I. Viscosity of gases commonly used in anaesthesia (from [2]) E (O Viscc Gas Oxygen Nitrous oxide Nitrogen Carbon dioxide Helium 280 -, l_l or ^ 206 x 10-8 at 19.1 C 152 x 10" 8 at 26.9 C 182 x 10-8 at 27.4 C 153 x 10" 8 at 19 C 198 x 10" 8 at 20 C ^ Viscosity (kg s" 1 m" 2 ) i j i 1 ' 1 ' 1 ' Temperature (K) FIG. 3. Viscosity of oxygen ( ) and nitrous oxide (O)- (point B), creating an imbalance of pressures between the background gas channel and the desflurane channel (point E). This is sensed by the differential pressure transducer, and an electrical signal causes the variable restrictor to open further, increasing the flow and pressure of desflurane in the channel. Conversely, decreasing the background gas flow results in an opposite imbalance at the pressure transducer and a resultant electrical signal which decreases theflowand pressure of desflurane at point E. For safety, there is a second differential pressure transducer (not shown in fig. 2) which senses pressures at the same locations. If the two transducers do not agree, an electrical signal is sent to the sump shut-off solenoid, which stops desflurane output from the sump. This redundancy in the system also safeguards against zero drift in the transducers. Should the variable restrictor (R2) in the desflurane channel be open fully, but the pressure at point E in the desflurane channel still be less than that in the background gas channel (point B), additional power is supplied to the sump heaters. Ordinarily this is required only when the sump is filled with additional anaesthetic while the vaporizer is delivering large desflurane concentrations with large flows of background gas. For the above to function satisfactorily, the fixed restrictor (R2) in the background gas channel and the rotary variable restrictor ("manually variable restrictor select", fig. 2) must have laminar flow so that any change in pressure produces a linear change in flow. Within the working limits specified ( litre min" 1 ) the flow through these devices is essentially laminar. However, the flow through the fixed restrictor, as is true for all laminarflowdevices, is affected by the viscosity of the gas. The vaporizer is calibrated at the factory with a backgroundflowof 100% oxygen. At any pressure change across the resistor (R2), as the viscosity of the gas increases, the flow of gas decreases. For example, because nitrous oxide has a lower viscosity than oxygen (table I, fig. 3), as it is added to the background gas, the viscosity of the background gas decreases and thus for a given gas pressure at point B (proximal to the fixed restrictor, R2), background gasflowis greater than if the background gas flow were oxygen alone. As the regulator balances pressures (and not flows) in the two gas channels, at this pressure (but greaterflowin the background gas channel) the amount of desflurane allowed to flow though the rotatory variable control valve is unchanged, resulting in a decreased concentration of desflurane leaving the vaporizer. The change in pressure decrease and flow is directly proportional to the change in viscosity. Fortunately the viscosities of the other gases used commonly in anaesthesia are sufficiently similar to that of oxygen [2] (table I) that the resultant desflurane concentration emanating from the vaporizer when nitrous oxide or nitrogen is added, should remain within 20 % of the value when oxygen alone comprises the background flow (see below). The background gas channel contains one additional laminar flow, fixed resistor (Rl) in line before the gas enters the regulator component of the vaporizer. The purpose of this resistor is to provide

4 DESFLURANE VAPORIZER 477 a small positive pressure (19-22 cm H 2 O at an oxygenflowof 10 litre min"')j on the side of the plate within the anaesthetist-controlled dial (manually variable restrictor) of the vaporizer, in order to transmit the pressure at point A to the plate. Desflurane vapour passes through the groove on the opposite side of this plate. The positive pressure helps to prevent the plate from lifting. If there were no additional resistor, and if the plate were to lift, then the groove through which the desflurane passes would have a larger effective cross-sectional area, allowing a greater than desired amount of desflurane to pass. With this additional restrictor any lift of the rotary valve results in a net inflow of fresh gas into the control groove area, causing a dilution of desflurane output. SAFETY FEATURES Several safety features are incorporated into the Tec 6 desflurane vaporizer. A second differential pressure transducer (not shown in fig. 2) measures the difference in pressures at the same place as the first transducer. The two must read identically, or the shut-off solenoid is activated, and output of desflurane ceases. This guards against zero drift of both transducers. Both resistive thermometer devices (one for the sump, and one for the upper manifolds and rotary valve) have back-ups (not shown in fig. 2). As for all pressurized devices, a "relief" valve is desirable; with the Tec 6 this comprises the use of "thermal fuses", one in each limb of the main power supply to the sump heaters (not shown in fig. 2). If the temperature exceeds 57 C, the power to the heater is interrupted, thus preventing build up of excessive pressure. To ensure that the vaporizer does not deliver desflurane when the anaesthetist controlled dial is turned to 0% (stand-by), not only is the resistor (groove) in the dial closed, but in addition, a positive interlock mechanism shuts the sump shut-off valve which controls output from the sump. Before the control dial valve can be rotated to a position other than 0%, the dial release button must be depressed. When the dial release button is pushed in, an optical sensor sends a signal to open the solenoid controlling the sump output. An interlock prevents the release button from being depressed unless the vaporizer is in operational mode. A second button on the anaesthetist-controlled dial must be depressed to deliver desflurane concentrations in excess of 12%. If tilted, some conventional, mechanical vaporizers are subject to excessive output, owing to liquid anaesthetic entering the tubes (designed to carry vapour only) leaving the sump. Because the sump shut-off solenoid opens only when the vaporizer is operational and thus electrically powered, and placed correctly on the back bar of an anaesthesia machine, the likelihood of liquid desflurane leaving the sump when the vaporizer is tilted is reduced greatly. Nevertheless, the Tec 6 also contains a "tilt switch" which shuts the sump shut-off valve (should it be open) when the angular displacement of the vaporizer exceeds approximately 15 from vertical. The Tec 6 desflurane vaporizer incorporates electronic sensors and display of the hquid level of the sump. This method was chosen, rather than the direct visual " sight glass " of conventional vaporizers because of the possibility of potential for leaks caused by the heated pressurized liquid anaesthetic. The liquid level sensor is a ceramic covered electrode within a stainless steel tube, which is part of the sump. The capacitance of this sensor, which forms one arm of a comparator, varies with the level of liquid between the ceramic covered electrode and the stainless steel tubes. The electrical signal through the electronics drives the 20-bar liquid crystal linear display on the front of the vaporizer. When the sump is full, all 20 bars are illuminated; when the sump contains less than 60 ml of hquid, no bar is illuminated. A second device switches off the heaters when the sump is almost empty. This results in a no output alarm, which changes to a low agent alarm when the rotary dial is returned to a standby position. Overfilling the sump of a standard mechanical vaporizer may result in liquid anaesthetic entering the collecting tubing exiting the sump. In conventional vaporizers, because background gas flows through the sump, this liquid is carried out of the vaporizer rapidly, either as liquid or a very large concentration of vaporized anaesthetic. This has resulted in several reports of anaesthetic overdose. Because no background gas flows through the sump of the Tec 6 vaporizer, this risk is lessened. However, it is important to avoid overfilling of the sump. There are two mechanisms designed to prevent overfilling of the sump. First the filler aperture of the sump is placed so that when the level of desflurane in the sump is above the aperture, thus occluding it, the sump is full, but not overfilled. Second, the sump outlet is higher than the liquid level of a full bottle of desflurane in itsfillingposition. Thus under ordinary circumstance, the sump fluid level would not reach this point. However, it should be noted that the design of both of these mechanisms depends upon the vaporizer being filled in its normal upright position. Filling the vaporizer while it is in another attitude could bypass this design and result in overfilling. However, even if the vaporizer were overfilled while tilted, the solenoid shut-off valve would prevent liquid from leaving the sump. When the vaporizer was subsequently turned on, a small amount of liquid might leave the sump. However, the resultant vapour would cause an increase in pressure in the desflurane gas channel, resulting in closure of the shut-off solenoid. EVALUATION OF OUTPUT PERFORMANCE We have (1) compared the desflurane concentration in the gas leaving the vaporizer with the dial setting; (2) measured desflurane output as a function of background gas flow rates; and (3) measured desflurane output as a function of the background gas composition. Methods We tested six desflurane Tec 6 vaporizers at desflurane dial settings of 1, 4, 7 and 10% with

5 478 BRITISH JOURNAL OF ANAESTHESIA a 10 CO T H Q. g g a 9 8 -t-4 i A i 4 S- S 5 T i a " i i x -a 9 r Gas flow (litre min" 1 ) -9 fr A i i 10% 7%» 4% FIG. 4. Desflurane output of six vaporizers (each represented by a different symbol) at vaporizer settings of 1, 4, 7 and 10% desflurane with various carrier gas flow rates of 100% oxygen. c 5= O 8-6-.H 2 - Q I 8 S B 8 10% g i 7% i i n 8 4 % Gas flow (litre min" 1 ) FIG. 5. Desflurane output of six vaporizers (each represented by a different symbol) at vaporizer settings of 1, 4, 7 and 10% desflurane with various carrier gasflowrates of 70 % nitrous oxide in oxygen. I 25-, -M I «* ft ft H 5- o «o Gas flow (litre min" 1 ) FIG. 6. Difference (relative; %) in vaporizer output at dial settings of 1 (O), 4 ( )» 7 (O) and 10% (A) desflurane, between carrier gasflowsof 100 % oxygen and 70 % nitrous oxide in oxygen. Data are mean of six vaporizers. background flow rates of litre min" 1 of 100 % oxygen and 70% nitrous oxide in oxygen. To determine desflurane concentration in the gas leaving the vaporizer, we used the analogue output, to the nearest 0.01 % desflurane, of a clinical infra-red spectrophotomer (Datex Ultima, Helsinki), which 10 1% 10 had been calibrated against gas samples, the concentration of which had been measured by gas chromatography against primary gravimetric standards. In addition, spot samples of vaporizer output were analysed by gas chromatography. Results With carrier flow of oxygen litre min" 1, the desflurane concentration in the outflow gas was within ±15% relative or ±0.5% absolute, of the dial setting for all vaporizers, except one, which was 17 % low at the single condition of 4 % desflurane at 0.5 litre min" 1 (fig. 4). At carrier flow rates less than 2 litre min" 1 of oxygen, desflurane concentration was generally smaller (by approximately 8%) than at larger carrier gas flow rates, but was still within the above limits. With carrier gas flow rates of 70% nitrous oxide in oxygen, the concentration of desflurane was less (average 16% smaller) than with similar carrier flow rates of oxygen (figs 5 and 6). The desflurane concentration in the output gas was within these limits for all vaporizers at a dial setting of 1 % desflurane. At larger desflurane concentrations, and larger carrier flow rates of 70 % nitrous oxide in oxygen, the desflurane output of the vaporizers remained within these tolerances; however as the carrier gas flow rate decreased, so did the relative output of the vaporizers. Only at a dial setting of 1 % desflurane did output error exceed 20 % (range 0-32 %; mean error (SD) : 13.6 (6.5) %); however, output remained within ±0.5% desflurane, absolute. Discussion The accuracy of the Tec 6 vaporizer output is dependent upon the desflurane concentration, and carrier gas flow rate and composition. The vaporizer is calibrated at the factory with oxygen as the carrier gas. Thus it is not surprising that the vaporizer output closely matched the dial settings when oxygen was the carrier gas. Because of the lower viscosity of nitrous oxide and the vaporizer design, a smaller desflurane concentration is to be expected when nitrous oxide is part of the carrier gas than when 100 % oxygen is the carrier gas. At C, the viscosity of nitrous oxide is 71 % that of oxygen [2]. Thus the viscosity of 70 % nitrous oxide in oxygen should be approximately 80% that of 100% oxygen. This should result in an approximate 20% decrease in the concentration of desflurane leaving the vaporizer when 70 % nitrous oxide in oxygen is the carrier. At a dial setting of 1 % desflurane, we found that the desflurane concentration with 70% nitrous oxide in oxygen as the carrier gas was 20% less than the desflurane concentration with 100% oxygen as the carrier gas. As the dial setting was increased the difference in desflurane concentration between the two carrier gases lessened to % at 7-10 % desflurane. The small difference in desflurane output between analysed and theoretical values when nitrous oxide was added to the carrier gas, at any dial setting, suggests that although flow within the fresh gas restrictors is essentially laminar, it is not perfect. Fortunately, the error is in the direction of safety: as the second

6 DESFLURANE VAPORIZER 479 anaesthetic (nitrous oxide) is added, the vaporizer output is less than indicated by the dial setting. REFERENCES 1 Andrews JJ, Johnston RV jr, Kramer GO Consequences of A /~vwnurr irrw-cuxexrr misfilling contemporary vaporizers with desflurane. Canadian Journal of Anaesthesia 1993; 40: The authors thank E. I. Eger n, M.D., for his review of this 2 Hodgman, Weast RC, Selby SM, eds. Handbook of manuscript. This work was supported by the Anesthesia Chemistry and Physics. 39th Edn. Cleveland: Chemical Research Foundation. Rubber Publishing Co, 1957;

Flow meter. bellow vaporizer. APL valve. Scavenging system

Flow meter. bellow vaporizer. APL valve. Scavenging system Introductory Lecture Series: The Anesthesia Machine PORNSIRI WANNADILOK Objectives Anesthesia Machine Ventilators Scavenging Systems System Checkout 1 Flow meter ventilator bellow vaporizer Corrugated

More information

Understanding vaporizers

Understanding vaporizers Stephen Boumphrey MB BS FRCA Nick Marshall MB BS FRCA Matrix reference 1A03 Key points Vaporizers allow a known and reproducible concentration of anaesthetic vapour to be delivered in a safe and reliable

More information

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS 1. INTRODUCTION An anaesthesia delivery system includes any machine, equipment or apparatus which supplies gases, vapours, local anaesthesia and/or

More information

Recommendations on Checking Anaesthesia Delivery Systems

Recommendations on Checking Anaesthesia Delivery Systems Page 1 of 11 Recommendations on Checking Anaesthesia Delivery Version Effective Date 1 Oct 1992 (reviewed Feb 07, Feb 02) 2 2004 3 Nov 2011 4 Dec 2016 Document No. HKCA T1 v4 Prepared by College Guidelines

More information

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet: OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle

More information

Lesson 6: Flow Control Valves

Lesson 6: Flow Control Valves : Flow Control Valves Basic Hydraulic Systems Hydraulic Fluids Hydraulic Tank Hydraulic Pumps and Motors Pressure Control Valves Directional Control Valves Flow Control Valves Cylinders : Flow Control

More information

Anes Machine Use and Pre-use Check - CA1 focus

Anes Machine Use and Pre-use Check - CA1 focus Anes Machine Use and Pre-use Check - CA1 focus BWH Anesthesia Grand Rounds August 31, 2005 James H. Philip, MEE, MD, CCE Copyright 2000-2005, James H Philip, all rights reserved BWH Anesthesia Workspace

More information

OIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply

OIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply OIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply 4.1.1 General The primary function of a system for handling fuel oil is to transfer oil from the storage tank to the oil burner at specified conditions

More information

VAPOMATIC ANESTHETIC VAPORIZER

VAPOMATIC ANESTHETIC VAPORIZER VAPOMATIC ANESTHETIC VAPORIZER INSTRUCTION MANUAL SOLD FOR VETERINARY USE ONLY BY: A.M. BICKFORD, INC. 12318 BIG TREE ROAD WALES CENTER, NY 14169 1-800-795-3062 Fax: 716-652-2046 www.ambickford.com sales@ambickford.com

More information

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS Using state-of-the-art measuring and microprocessor technologies, Advanced has assembled a series of systems which can measure mass

More information

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS Annex 3, page 2 ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS The text of existing chapter 15 is replaced by the following: "1 Application This

More information

Vehicle-mounted meters, pump supplied

Vehicle-mounted meters, pump supplied Purpose This inspection procedure outline (IPO) defines the minimum tests which must be performed to ensure that basic volumetric measuring devices comply with the legislation. Application Vehicle-mounted

More information

Vehicle- or rack-mounted liquefied gas meters, pump supplied

Vehicle- or rack-mounted liquefied gas meters, pump supplied Purpose This inspection procedure outline (IPO) defines the minimum tests which must be performed to ensure that basic volumetric measuring devices comply with the legislation. Application Rack/fixed metering

More information

Mass Flow Controller (MFC) for Gases

Mass Flow Controller (MFC) for Gases Mass Flow Controller (MFC) for Gases Type 8713 can be combined with... Direct flow measurement by MEMS- Technology for nominal flow rates from 1 ml N /min to 8 l N /min (N 2 ) High accuracy and repeatability

More information

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions L 100 Bubble-Tube Level System Installation, Operation and Maintenance Instructions Figure 1 Contents Section Description Page 1.0 Introduction 2 2.0 Specifications 3 3.0 Installation 3 4.0 Warranty 6

More information

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE TELEDYNE HASTINGS TECHNICAL PAPERS INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE Proceedings of FEDSM 98 1998 ASME Fluids Engineering

More information

Anesthesia Machine Intro

Anesthesia Machine Intro Anesthesia Machine Intro Brigham and Women s Hospital New Residents July 9, 2010 Copyright 2000-2010, James H Philip, all rights reserved ADS1 Ready JHP Anesthesia Machine Intro James H. Philip, MEE, MD,

More information

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features ANALOG MASS FLOW CONTROLLERS Model AF mass fl ow controllers are designed to indicate fl ow rates and control set fl ow rates of gases. Each of these units incorporates an advanced straight tube sensor

More information

Ventam 85 Installation & Commissioning Instructions

Ventam 85 Installation & Commissioning Instructions Ventam Systems Ltd Unit D4 Seedbed Business Centre Vanguard Way Shoeburyness Essex SS3 9QY Phone 01702 382 307 Fax 01702 382 340 Ventam 85 Installation & Commissioning Instructions 1 General The Ventam

More information

IT is frequently desirable to be able to determine the

IT is frequently desirable to be able to determine the NEW INVENTIONS A PORTABLE OXYGEN ANALYSER By A. BRACKEN Research and Development Department, The British Oxygen Company, Limited IT is frequently desirable to be able to determine the oxygen concentration

More information

Datum 'L' Vaporizer. Blease. L Series. Anaesthesia Vaporizer. User Manual. Blease. User Manual

Datum 'L' Vaporizer. Blease. L Series. Anaesthesia Vaporizer. User Manual. Blease. User Manual Blease R Anaesthesia Vaporizer L Series Blease 1 2 Datum 'L' Vaporizer 'L' Series Datum are smaller and lighter than other Datums. The serial number, found on the back of the vaporizer, ends with the letter

More information

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111 INTRODUCTION Water vapor in natural gas has more than a substantial effect on the quality of the gas stream. Without quality measurement of water vapor the gas is basically not saleable. Contracts are

More information

Working version subject to Departmental approval. Drafted by Dr. Olympio and reviewed by clinical members of The Capital Equipment Committee

Working version subject to Departmental approval. Drafted by Dr. Olympio and reviewed by clinical members of The Capital Equipment Committee Wake Forest University Baptist Medical Center Department of Anesthesiology TECHNICIAN ABBREVIATED CHECKLIST OF RESPONSIBILITIES For Anesthesia Apparatus Checkout Specific to the Draeger APOLLO anesthesia

More information

Mass Flow Meter (MFM) for Gases

Mass Flow Meter (MFM) for Gases 873 Mass Flow Meter (MFM) for Gases Direct flow measurement by MEMS- Technology for nominal flow rates from 1 ml N /min to 8 l N /min (N 2 ) High accuracy Short response time Compact design and digital

More information

Aespire 7100 Essential performance Compact design

Aespire 7100 Essential performance Compact design GE Healthcare Aespire 7100 Essential performance Compact design Features Enhanced monitor integration capabilities with our Datex-Ohmeda Anesthesia Monitor and Compact Anesthesia monitor Lightweight and

More information

Design Features. General Description

Design Features. General Description Model thermal Mass Flow Controllers are designed to indicate and control set flow rates of gases. The combines the characteristics and accuracy of conventional mass flow devices into a unique compact design

More information

DPC-30 DPC-100. Reference Manual

DPC-30 DPC-100. Reference Manual DPC-30 DPC-100 Reference Manual 1. Introduction 1.1 Description The Martel DPC Digital Pneumatic Calibrator improves upon traditional dial gauge pneumatic calibrators. The Martel DPC improves accuracy,

More information

The Univentor 1250 Anaesthesia Unit

The Univentor 1250 Anaesthesia Unit THE UNIVENTOR 1200/1250 ANAESTHESIA UNIT The Univentor 1250 Anaesthesia Unit TABLE OF CONTENTS EDITION 1 Section 1 - WARRANTY & SERVICE 1.1. WARRANTY 2 1.2. DAMAGED SHIPMENTS 2 1.3. SERVICE 2 Section 2

More information

This test shall be carried out on all vehicles equipped with open type traction batteries.

This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4. Determination of hydrogen emissions page 1 RESS-6-15 5.4.1. This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4.2. The test shall be conducted following

More information

BUBBLER CONTROL SYSTEM

BUBBLER CONTROL SYSTEM BUBBLER CONTROL SYSTEM Description: The HDBCS is a fully automatic bubbler system, which does liquid level measurements in water and wastewater applications. It is a dual air compressor system with, air

More information

RTG 25. Introduction

RTG 25. Introduction Introduction RTG 25 pressure regulator is included within the direct acting and balanced valve regulator class. These regulators have a large range of applications both in industrial and domestic installations.

More information

Selecting the right pressure sensor for your application

Selecting the right pressure sensor for your application Selecting the right pressure sensor for your application This guide from Acal BFi details the key factors that you must consider when selecting a pressure sensor for your application. Selecting the best

More information

DURING the course of certain investigations it became

DURING the course of certain investigations it became VOLUMETRIC DETERMINATION OF ETHER OR CYCLOPROPANE, CARBON DIOXIDE, NITROUS OXIDE AND OXYGEN IN ANESTHETIC MIXTURES By F. J. PRIME DURING the course of certain investigations it became necessary to be able

More information

Columbus Instruments

Columbus Instruments 0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011

More information

Bubble Tube Installations

Bubble Tube Installations Instruction MI 020-328 September 2013 Bubble Tube Installations For Liquid Level, Density, and Interface Level Measurements 2 Contents Introduction... 5 Abbreviations... 5 Principle of Operation... 5 Alternative

More information

Qi Litre Ultrasonic Cleaner

Qi Litre Ultrasonic Cleaner Qi-300 17Litre Ultrasonic Cleaner Ultrawave s Qi-Series are robust and reliable, delivering outstanding cleaning performance in a wide range of industries. From precision, pharmaceutical and medical device

More information

AUTOMATIC GAS MANIFOLDS

AUTOMATIC GAS MANIFOLDS AUTOMATIC GAS MANIFOLDS Phoenix Pipeline Products Limited. Unit 8, McKenzie Industrial Park, Tel No.: 44 (0) 161 428 7200 Bird Hall Lane, Fax No.: 44 (0) 161 428 7010 Stockport, Email: info@p3-phoenix.com

More information

(12) Patent Application Publication (10) Pub. No.: US 2013/ A1

(12) Patent Application Publication (10) Pub. No.: US 2013/ A1 US 2013 0186486A1 (19) United States (12) Patent Application Publication (10) Pub. No.: US 2013/0186486A1 Ding (43) Pub. Date: Jul. 25, 2013 (54) SYSTEM FOR AND METHOD OF (52) U.S. Cl. MONITORING FLOW

More information

The Anesthesia Machine (con t) The primary cause of machine malfunction is failure to check Never start without the American Express items What is the

The Anesthesia Machine (con t) The primary cause of machine malfunction is failure to check Never start without the American Express items What is the Anesthesia Machine The Anesthesia Machine You are the master of the machine You are responsible for checking the machine prior to each case 1 The Anesthesia Machine (con t) The primary cause of machine

More information

Model 130M Pneumatic Controller

Model 130M Pneumatic Controller Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate

More information

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW AC 2010-2145: MEASUREMENT OF HYDROGEN IN HELIUM FLOW Randy Buchanan, University of Southern Mississippi Christopher Winstead, University of Southern Mississippi Anton Netchaev, University of Southern Mississippi

More information

THE GE AISYS ANAESTHETIC MACHINE

THE GE AISYS ANAESTHETIC MACHINE THE GE AISYS ANAESTHETIC MACHINE All essential components of the anaesthetic machine are shown in the diagram below. Each major component will be explained in detail. The Gas Inlet Module In the diagram

More information

Point level switches for safety systems

Point level switches for safety systems Point level switches for safety systems By: Bill Sholette Level Products Business Manager Northeast US - Endress+Hauser Point level switches are often used in applications designed to prevent accidents.

More information

AMS 6916 Board mount pressure sensor with ratiometric analog output

AMS 6916 Board mount pressure sensor with ratiometric analog output FEATURES Piezoresistive pressure sensor with amplified analog output Calibrated and temperature compensated Ratiometric voltage output, 0.5 4.5 V Digital signal conditioning, 12 bit output resolution Differential,

More information

AMT-Ex Dewpoint Transmitter

AMT-Ex Dewpoint Transmitter AMT-Ex Dewpoint Transmitter Instruction Manual Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Tel: +44 1274 733100 Fax: +44 1274 733200 email: mail@amsytems.co.uk web: www.amsystems.co.uk

More information

Operation manual Oxygen sensor

Operation manual Oxygen sensor Using oxygen sensors made by SENSORE, oxygen can be measured in gaseous media within a concentration range from 10 ppm O 2 up to 96 % O 2 depending on the sensor type. The sensor is thus characterized

More information

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE Surgical Care at the District Hospital 1 15 Anesthetic Infrastructure and Supplies Key Points 2 15.1 Equipment & Supplies for Different Level Hospitals However well trained you are as an anaesthetist,

More information

Prof AH Basson, Pr Eng March Departement Meganiese en Megatroniese Ingenieurswese. Department of Mechanical and Mechatronic Engineering

Prof AH Basson, Pr Eng March Departement Meganiese en Megatroniese Ingenieurswese. Department of Mechanical and Mechatronic Engineering An evaluation of the effectiveness of the Sinapi Chest Drain's flutter valve in comparison to a water seal as found in an UWD to resolve pneumo- or hemothorax Prof AH Basson, Pr Eng March 2010 Departement

More information

LPG DRIVER ATTENDED TRANSPORT LOADING

LPG DRIVER ATTENDED TRANSPORT LOADING November 1993 Prepared By: Ken A. Steward. P.E. Linco-Electromatic, Inc. 4580 West Wall Street Midland, Texas 78703 LPG DRIVER ATTENDED TRANSPORT LOADING INTRODUCTION The safest and most accurate method

More information

OIL AND GAS INDUSTRY

OIL AND GAS INDUSTRY This case study discusses the sizing of a coalescer filter and demonstrates its fouling life cycle analysis using a Flownex model which implements two new pressure loss components: - A rated pressure loss

More information

with O 2 Controller Instruction Manual

with O 2 Controller Instruction Manual 14500 Coy Drive, Grass Lake, Michigan 49240 734-475-2200 E-mail: sales@coylab.com www.coylab.com Hypoxic Cabinets (In-Vivo / In-Vitro) with O 2 Controller Instruction Manual Index Page Warranty 2 Warnings

More information

Measurement of Low GAS & Liquid Flows

Measurement of Low GAS & Liquid Flows Measurement of Low GAS & Liquid Flows The Measurement Measurement with a float in a tapered borosilicate glass tube. Series 2100 with a 100 mm flow tube Series 2150 with a 150 mm flow tube Series 2300

More information

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008 UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method

More information

Section 14: Equipment failure

Section 14: Equipment failure Risks associated with your anaesthetic Section 14: Equipment failure Summary This leaflet describes the types of equipment used by anaesthetists. It gives information about the ways in which equipment

More information

BUBBLER CONTROL SYSTEM

BUBBLER CONTROL SYSTEM BUBBLER CONTROL SYSTEM Description: The LDBCS is a fully automatic bubbler system, which does liquid level measurements in water and wastewater applications. It is a dual air compressor system with, air

More information

Mitos Fluika Pressure and Vacuum Pumps Datasheet

Mitos Fluika Pressure and Vacuum Pumps Datasheet Unit 1, Anglian Business Park, Orchard Road, Royston, Hertfordshire, SG8 5TW, UK T: +44 (0)1763 242491 F: +44 (0)1763 246125 E: sales@dolomite-microfluidics.com W: www.dolomite-microfluidics.com Dolomite

More information

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler. Exercise 5-2 Bubblers EXERCISE OBJECTIVE Learn to measure the level in a vessel using a bubbler. DISCUSSION OUTLINE The Discussion of this exercise covers the following points: Bubblers How to measure

More information

Warner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

Warner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax DH-40i, Rev. 04.03.02 Micro-Incubation Platform For RC-40 Chambers Model DH-40i 1125 Dixwell Avenue, Hamden, CT 06514 (800) 599-4203 / (203) 776-0664 (203) 776-1278 - fax DH-40i, Rev. 04.03.02 Table of

More information

ACV-10 Automatic Control Valve

ACV-10 Automatic Control Valve ACV-10 Automatic Control Valve Installation, Operation & Maintenance General: The Archer Instruments ACV-10 is a precision automatic feed rate control valve for use in vacuum systems feeding Chlorine,

More information

Oxidation Stability of Gasoline and Aviation Fuels

Oxidation Stability of Gasoline and Aviation Fuels Oxidation Stability of Gasoline (Induction Period Method) Oxidation Stability of Aviation Fuels (Potential Residue Method) Test Method Provides an indication of the tendency of gasoline and aviation fuels

More information

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson A Journal of Practical and Useful Vacuum Technology From By Phil Danielson Thermal Conductivity Gauges Thermal conductivity pressure gauges are extremely common in vacuum technology, but an understanding

More information

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products Saturn Evo SATURN EVO COLOR A N A E S T H E S I A S Y S T E M i n t e l l i g e n t s i m p l i c i t y Manufacturer of high quality anaesthesia products SATURN EVO COLOR PATIENT SAFETY AND USER PATIENT

More information

Powers Controls RC 195 Multiple Input Receiver- Controller

Powers Controls RC 195 Multiple Input Receiver- Controller Powers Controls RC 195 Multiple Input Receiver- Controller Document No. 155-036P25 RC 195-1 Description Features Application The RC 195 Multiple lnput Receiver-Controller is a pneumatic instrument which

More information

CORESTA RECOMMENDED METHOD N 6

CORESTA RECOMMENDED METHOD N 6 CORESTA RECOMMENDED METHOD N 6 DETERMINATION OF VENTILATION DEFINITIONS AND MEASUREMENT PRINCIPLES (2015 Revision September 2016) 1. SCOPE This CORESTA Recommended Method specifies a method for the determination

More information

MACH ONE MASS FLOW CONTROLLER. MACH ONE SERIES flow control. MASS FLOW CONTROLLERS at the speed of sound.

MACH ONE MASS FLOW CONTROLLER. MACH ONE SERIES flow control. MASS FLOW CONTROLLERS at the speed of sound. MACH ONE MASS FLOW CONTROLLER MACH ONE SERIES flow control MASS FLOW CONTROLLERS at the speed of sound. MACH ONE SERIES MASS FLOW CONTROLLER FLOW CONTROL AT THE SPEED OF SOUND The Mach One revolutionary

More information

Operating Manual. Check and Go Disposable Medical Depot, Inc Port Washington N.Y

Operating Manual. Check and Go Disposable Medical Depot, Inc Port Washington N.Y Check and Go Disposable Operating Manual 2005 Medical Depot, Inc Port Washington N.Y. 11050 visit us at www. drivemedical.com Check and Go Disposable Part# 18580 2005 Medical Depot, Inc Port Washington

More information

Manual for continuous distillation

Manual for continuous distillation Manual for continuous distillation 1. Week 1: Objectives: Run the column at total reflux. When steady state is reached, take the sample from the top and bottom of the column in order to determine the overall

More information

SINGLE VALVE WITH LOW-FLOW BYPASS

SINGLE VALVE WITH LOW-FLOW BYPASS CONTROL VALVES Pressure Reducing Valve Sizing Guide Sizing pilot operated reducing valves is not a complicated process. It starts with determining requirements and following these guidelines in valve size

More information

Anaesthetic infrastructure and supplies

Anaesthetic infrastructure and supplies 15 Anaesthetic infrastructure and supplies Key Points 15.1 EQUIPMENT AND SUPPLIES FOR DIFFERENT LEVEL HOSPITALS However well trained you are as an anaesthetist, your ability to provide safe anaesthesia

More information

2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic. Pressure Measurement Engineered solutions for all applications

2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic. Pressure Measurement Engineered solutions for all applications Application Description AG/266PILD-EN Rev. C 2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic Pressure Measurement Engineered solutions for all applications Increase plant productivity

More information

Technical Data Sheet TI-F50 Locking Units series KFH

Technical Data Sheet TI-F50 Locking Units series KFH English translation of German original Locking Units series KF Further important practical advice is given in Operating Manual BA-F50., Rod diameter 18 mm 50 mm øz 8 L 2 6 x 6 0 min. 4x30 KF 18 to KF 32,

More information

CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS

CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS Brit. J. Anaesth. (1956), 28, 196 CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS BY RUSSELL M. DAVIES, I. R. VERNER Queen Victoria Hospital, East Grinstead AND A. BRACKEN Research and Development Centre,

More information

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series Introduction Powerex manifolds are cleaned, tested and prepared for the indicated gas service and are built in accordance with the Compressed Gas Association guidelines. The manifold consists of a regulator

More information

Keeping anaesthetic pollution to minimum in theatre environment

Keeping anaesthetic pollution to minimum in theatre environment Vet Times The website for the veterinary profession https://www.vettimes.co.uk Keeping anaesthetic pollution to minimum in theatre environment Author : Gianluca Bini Categories : General, Vets Date : September

More information

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1250 METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1.0 Scope and Application. 1.1 Analytes. Analyte CAS No. Sensitivity Total Organic Compounds N/A

More information

Chapter 10: Bidirectional Flow Controls

Chapter 10: Bidirectional Flow Controls Chapter 10: Bidirectional Flow Controls Objectives Learn about the patented, bidirectional flow control valves. Understand how the flow force affects the performance of the ZL70-36. Learn why there is

More information

The Ins and Outs of I/P Transducers

The Ins and Outs of I/P Transducers The Ins and Outs of I/P Transducers By Mark B. Levine, ControlAir Inc. General description I/P transducers are versatile instruments that use an electrical control signal to proportionally regulate gas

More information

How to specify a product. Process Sensors and Mechanical Instruments

How to specify a product. Process Sensors and Mechanical Instruments How to specify a product Process Sensors and Mechanical Instruments Keep the overview. Here is some guideline information on how to specify our products. Intended as supplementary help to specification

More information

Self-operated Pressure Regulators for special applications

Self-operated Pressure Regulators for special applications Self-operated Pressure Regulators for special applications Type 2357-31 Pressure Build-up Regulator with safety function and integrated excess pressure valve Application Pressure regulators for cryogenic

More information

Simple, Robust, Reliable Measures ph Doses either acid (lower) or alkali (raise) Displays ph, Total Dose Count (DCT) and Dose Count per hour (DC/h)

Simple, Robust, Reliable Measures ph Doses either acid (lower) or alkali (raise) Displays ph, Total Dose Count (DCT) and Dose Count per hour (DC/h) Simple, Robust, Reliable Measures ph Doses either acid (lower) or alkali (raise) Displays ph, Total Dose Count (DCT) and Dose Count per hour (DC/h) Integrated peristaltic dosing pump When using automatic

More information

Validation of Custody Transfer Metering Skid at Site After Laboratory Proving

Validation of Custody Transfer Metering Skid at Site After Laboratory Proving flotek.g 2017- Innovative Solutions in Flow Measurement and Control - Oil, Water and Gas August 28-30, 2017, FCRI, Palakkad, Kerala, India Validation of Custody Transfer Metering Skid at Site After Laboratory

More information

H 2. Sensor Series Take no risk! Install a Hydrogen Sensor in your Gaschromatograph

H 2. Sensor Series Take no risk! Install a Hydrogen Sensor in your Gaschromatograph Sensor Series 9000 Take no risk! Install a Hydrogen Sensor in your Gaschromatograph Carrier Gases for GC Probably more than 90% of the present GC instruments run with helium as carrier gas. Some people

More information

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES PROCESS ANALYSERS DF-550E The DF-550E Nano Trace oxygen analyser is a Coulometric sensor based analyser designed to measure oxygen as a contaminant at ultra trace levels in ultra high purity electronic

More information

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance Introduction Expired Gas Analysis or indirect calorimetry, can be used to measure ventilation and the fractions of oxygen and carbon dioxide in expired air. From these measurements, the body's oxygen consumption

More information

Dival 500 Pressure Regulators

Dival 500 Pressure Regulators Dival 500 Pressure Regulators Dival 500 Classification and Range of use The DIVAL 500 is a downstream direct-acting pressure regulator with balanced plug, for low, medium and high pressures. Suitable for

More information

Air Eliminators and Combination Air Eliminators Strainers

Air Eliminators and Combination Air Eliminators Strainers Description Air Eliminators and Combination Air Eliminator Strainers are designed to provide separation, elimination and prevention of air in piping systems for a variety of installations and conditions.

More information

Sampling Considerations for Equilibrium Dissolved Oxygen [DO] Sensors

Sampling Considerations for Equilibrium Dissolved Oxygen [DO] Sensors Sampling Considerations for Equilibrium Dissolved Oxygen [DO] Sensors Abstract Boiler water in the low parts per billion dissolved oxygen concentration range can be significantly contaminated by oxygen

More information

Operation manual Level sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50

Operation manual Level sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50 Rev.02 (2014.06) OL Page 1 of 8 Table of contents I. STRUCTURE OF THE MANUAL / CLARIFICATION... 3 II. SAFETY AND HEALTH CONCERNS... 4 1 INTRODUCTION... 5 1.1 Use

More information

We Warm Animals. User Manual For. Darvall Heated Smooth-Wall Circuits. Darvall Heated ZDS Qubes

We Warm Animals. User Manual For. Darvall Heated Smooth-Wall Circuits. Darvall Heated ZDS Qubes We Warm Animals User Manual For Warm Inspired Air Control Unit to use with Darvall Heated Smooth-Wall Circuits or Darvall Heated ZDS Qubes For Veterinary Use Only Warm Inspired Air Control Unit Use for

More information

Powers Controls RC 195 Single Input Receiver-Controller

Powers Controls RC 195 Single Input Receiver-Controller Document No. 155-119P25 RC 195-4 Powers Controls RC 195 Single Input Receiver-Controller Description The RC 195 Single lnput Receiver-Controller is a pneumatic instrument which receives one pneumatic input,

More information

Drilling Efficiency Utilizing Coriolis Flow Technology

Drilling Efficiency Utilizing Coriolis Flow Technology Session 12: Drilling Efficiency Utilizing Coriolis Flow Technology Clement Cabanayan Emerson Process Management Abstract Continuous, accurate and reliable measurement of drilling fluid volumes and densities

More information

PRESSURE SENSOR - ABSOLUTE (0 TO 700 kpa)

PRESSURE SENSOR - ABSOLUTE (0 TO 700 kpa) Instruction Sheet for the PASCO Model CI-6532A PRESSURE SENSOR - ABSOLUTE (0 TO 700 kpa) 012-06859B 10/98 $1.00 polyurethane tubing syringe cable with DIN s to computer interface quick release s (4) pressure

More information

Radar, Ultrasonic and RF Level Transmitters

Radar, Ultrasonic and RF Level Transmitters Radar, Ultrasonic and RF Level Transmitters Both measures the time it takes the wave to travel between the transmitter and that reflected wave off the surface of the material to reach the transmitter again.

More information

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel TANK MANAGER FOR TWO TANKS OPERATING MANUAL 10/31/11 C-More T6C L color touch panel 1 TABLE OF CONTENTS GENERAL...3 INSTALLATION...4 STONE TEST PROCEDURE...7 OPERATIONAL SUMMARY...7 AUTO CARBONATION...10

More information

Instruction Manual Updated 7/26/2011 Ver. 2.2

Instruction Manual Updated 7/26/2011 Ver. 2.2 4-Unit Model MB HTHP Filter Press #171-50-4: 115-Volt #171-51-4: 230-Volt Instruction Manual Updated 7/26/2011 Ver. 2.2 OFI Testing Equipment, Inc. 11302 Steeplecrest Dr. Houston, Texas 77065 U.S.A. Tele:

More information

Exhibit 4. Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N)

Exhibit 4. Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N) 1 APPLICABILITY Exhibit 4 Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N) Definitions common to all certification and test procedures

More information

Typical Single Stage Design SR 450 Series Shown. Two stage regulators add the following in addition to the features listed above:

Typical Single Stage Design SR 450 Series Shown. Two stage regulators add the following in addition to the features listed above: Single Stage : Recommended for applications where slight delivery pressure increases (due to decreasing cylinder pressures) would not affect the performance characteristics of the work or test results.

More information

Level MEASUREMENT 1/2016

Level MEASUREMENT 1/2016 Level MEASUREMENT 1/2016 AGENDA 2 A. Introduction B. Float method C. Displacer method D. Hydrostatic pressure method E. Capacitance method G. Ultrasonic method H. Radar method I. Laser method J. Level

More information

Digital Vacuum Regulator

Digital Vacuum Regulator Temperature Control for Research and Industry Digital Vacuum Regulator User s Manual Model 300 INDEX SECTION PAGE 1. QUICK OPERATING INSTRUCTIONS........................... 3 Safety Notices.................................................

More information

A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers

A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers Slemeyer, A. University of Applied Sciences of Giessen, Dept. of Electrical Engineering 1 Wiesenstr. 14, D-35390

More information