1 Anes Machine Use and Pre-use Check - CA1 focus BWH Anesthesia Grand Rounds August 31, 2005 James H. Philip, MEE, MD, CCE Copyright , James H Philip, all rights reserved
2 BWH Anesthesia Workspace
3 The Anesthesia Delivery System Continuous Flow Machine for Gas Mixing and Vaporization Breathing Circuit to provide for breathing and CO2 elimination
4 Continuous Flow Machine for Gas Mixing and Vaporization Provides continuous flow of - Oxygen for Life Air to safely lower F I O 2 Nitrous Oxide for partial anesthesia Agent vapor for complete or partial anesthesia
5 N 2 O N 2 O SERVICE INLET 55 psi SIMPLIFIED ANESTHESIA MACHINE SCHEMATIC DIAGRAM No Oxygen safety features N 2 O 750 psi N 2 O FILTER N 2 O N 2 O 750 psi CHECK VALVE 35 psi PRESSURE REGULATOR S CALIBRATED VAPORIZERS I D WALL NEEDLE VALVE O psi O 2 NEEDLE VALVE O 2 [Anesthetizing gases flow] Fresh Gas Flow (FGF) DISS diameter index safety system CONNEC- TION O psi 35 psi OXYGEN FLUSH VALVE [first] Oxygen Flush Flow (last) COMMON OUTLET FRESH GAS HOSE O 2 O 2 SERVICE INLET 55psi
6 SIMPLIFIED ANESTHESIA MACHINE SCHEMATIC DIAGRAM N 2 O N 2 O SERVICE INLET 55 psi N 2 O 750 psi N 2 O FILTER FAILSAFE VALVE N 2 O N 2 O 750 psi CHECK VALVE 35 psi PRESSURE REGULATOR S CALIBRATED VAPORIZERS I D WALL NEEDLE VALVE O psi O 2 ON/OFF SWITCH NEEDLE VALVE O 2 [Anesthetizing gases flow] Fresh Gas Flow (FGF) DISS diameter index safety system CONNEC- TION O psi 35 psi OXYGEN FLUSH VALVE [first] Oxygen Flush Flow (last) COMMON OUTLET FRESH GAS HOSE O 2 O 2 SERVICE INLET 55psi
7 N 2 O N 2 O SERVICE INLET 55 psi SIMPLIFIED ANESTHESIA MACHINE SCHEMATIC DIAGRAM GEDO (GE-Datex-Ohmeda) Chain Link 25 for Delivered O 2 Modulus 2+, Aestiva N 2 O FILTER FAILSAFE VALVE N 2 O CHECK VALVE 35 psi PRESSURE REGULATOR S CALIBRATED VAPORIZERS I D WALL DISS diameter index safety system CONNEC- TION O psi O PSI Is OK 35 psi ON/OFF SWITCH 3:1 OXYGEN FLUSH VALVE [first] NEEDLE VALVE Mechanical Stop 0, 50 or 250 ml/min O 2 [Anesthetizing gases flow] Oxygen Flush Flow (last) Fresh Gas Flow (FGF) COMMON OUTLET FRESH GAS HOSE O 2 O 2 SERVICE INLET 55psi
8 N 2 O N 2 O SERVICE INLET 55 psi SIMPLIFIED ANESTHESIA MACHINE SCHEMATIC DIAGRAM Draeger ORMC = Oxygen Ratio Monitor/Controller Fabius, Tiro, Narkomed MR N 2 O FILTER Linear Regulator N 2 O CHECK VALVE 35 psi PRESSURE REGULATOR S CALIBRATED VAPORIZERS I D WALL NEEDLE VALVE O 2 O 2 ON/OFF SWITCH [Anesthetizing gases flow] Fresh Gas Flow (FGF) DISS diameter index safety system CONNEC- TION O psi PSI Is OK 35 psi Bypass Flow 0, 50 or 250 ml/min OXYGEN FLUSH VALVE [first] Oxygen Flush Flow (last) COMMON OUTLET FRESH GAS HOSE O 2 O 2 SERVICE INLET 55psi
9 Continuous Flow Gas Machine N O 2 Air O 2 L/m L/m L/m ml/m Isoflurane Sevoflurane Desflurane Vaporizer Common Outlet FGF (Fresh Gas Flow)
10 Ohmeda Modulus 2+
11 GEDO Modulus 2 SE
12 Draeger Narkomed MRI-2
13 Draeger Fabius with Real Needle Valves, Virtual Flowmeters
14 Direct-reading temperature-compensated Vaporizers Sevo Iso Des
15 Performance curve, Tec 4, 5, 7 % isoflurane Dial Setting Flow rate (L/min O 2 )
16 GE-Datex-Ohmeda Aestiva Draeger Fabius GS Two Companies
17 Breathing Circuit & CO 2 Absorber Schematics
18 Breathing Circuit & CO 2 Absorber Pictures
19 GEDO Aestiva CO2 Absorber
20 Machine Check Out Calibrate Oxygen Monitor Check Machine for gas delivery Check Circuit for integrity Check Ventilator for function Check Exhaust for function
21 Calibrate Oxygen Monitor Oxygen Monitor switched on, sensor in air 3 minutes, hole plugged. Calibrate to read 21%. Reconnect Oxygen sensor to inspired limb of breathing circuit This is called Circuit Oxygen
22 Aestiva O2 Sensor in Air for Cal
23 Fabius O2 Sensor in Air for Cal
24 Beware of missing sensor
25 Some have obvious label (good) not Fabius, yet
26 Expose sensor to air
27 Ohmeda CD - Calibrate Oxygen System Cal More
28 Tell machine to calibrate to 21%
29 Datex Capnomac Ultima Similar Calibration
30 Fabius
31 Aestiva
32 Check Machine for Gas Delivery Machine On Oxygen flows smoothly Disconnect oxygen hose Oxygen alarm sounds Switch ON oxygen tank Verify P > 1000 psi and Alarm silences Check Vaporizer is present and not empty *
33 Switch ON oxygen tank (with tank wrench)
34 Check that P > 1000 psi
35 Other ADSs
36 Check Machine for Gas Delivery 2 Tank off with wrench Turn oxygen flow up Alarm sounds on oxygen failure Connect oxygen hose to silence alarm Turn up nitrous oxide and observe alarm and/or control Perform negative pressure leak test now (possibly shown later when it makes sense)
37 Ensure there is no leak - Check circuit for integrity
38 Check Circuit for integrity Attach circuit including reservoir bag Occlude "Y" with thumb Flush and fill bag to P = 30 cmh 2 O
39 Check Circuit for integrity Attach circuit including reservoir bag Occlude "Y" with thumb Flush and fill bag to P = 30 cmh 2 O Ensure there is no leak Observer for 10 seconds Pressure should not fall
40 Check Circuit for integrity Attach circuit including reservoir bag Occlude "Y" with thumb Flush and fill bag to P = 30 cmh 2 O Ensure there is no leak Test pressure alarms Continuing pressure alarm 12 cmh 2 O for > 20 seconds High pressure alarm > 60 cmh 2 O
41 Check Circuit for integrity Attach circuit including reservoir bag Occlude "Y" with thumb Flush and fill bag to P = 30 cmh 2 O Ensure there is no leak Test Pressure Alarms Open Relief (APL, pop off) Valve Verify Reservoir bag empties
42 Check Circuit for integrity Attach circuit including reservoir bag Occlude "Y" with thumb Flush and fill bag to P = 30 cmh 2 O Ensure there is no leak Test Pressure Alarms Open Relief (APL, pop off) Valve Verify Reservoir bag empties Observe waste gas collector (scavenger) bag fills and empties
43 Leak and Compliance tests Automated on Draeger Fabius and Tiro Discussed a bit in next hour
44 Ensure that you can ventilate By hand with reservoir With mechanical ventilator
45 Check Waste Gas Collection System
46 Waste gas collector fills when reservoir bag empties Empties into evacuation system (Scavenger)
47 Check Scavenger Valves + -
48 Newer ADSs have Vacuum Gauge and Vacuum Adjustment
49 We missed one potential leak Low pressure system upstream of breathing circuit Negative Pressure Leak Test
50 Negative Pressure Leak Test Is not necessary if there is no valve Is not possible if no hose connector Perform only with GEDO (Ohmeda) Modulus 2+
51 Check Ventilator for function Ventilate reservoir bag Observe no volume loss
52 Check Suction for Function Rarely used emergently Very important when it is needed
53 Workspace Setup MRSMAID Machine Room Suction Monitors Airway I V Drugs Suction off if: DISS hose unscrewed Wall switch off Hose disconnect from Canister Canister switch off Insert seated incorrectly Any hole open Flap valve closed because unit was shaken or canister is full Flexible hose kinked Flexible hose stepped on
54 GE-Datex-Ohmeda Aestiva Draeger Fabius GS Two Companies
55 Draeger Fabius GS System Test Software, internal hardware Cal Flow, Calibrates the flow sensor Later, compensate for Des high specific heat Cal O2 FIO2 Monitor like all others - expose to air Leak Test Unique to this machine Automatic computation of leak and circuit compliance Press leak test button and follow directions: seal circuit, fill circuit, measure leak&compliance
56 All this is on Anesthesia Department Web Site Navigate to resources All checkouts are there Special thanks to Anes IS Team Jamie Bell, Dan Holtzman, Fred Donovan, Athena Dimas, and new Director, Bob Gimlich
57 Technology Matters! I have done what I can to prepare you for your future It is up to you to learn to use your tools as a true expert The resources are available The rest is up to you
58 Thank you
59
Anesthesia Machine Intro
Anesthesia Machine Intro Brigham and Women s Hospital New Residents July 9, 2010 Copyright 2000-2010, James H Philip, all rights reserved ADS1 Ready JHP Anesthesia Machine Intro James H. Philip, MEE, MD,
More informationDraeger Fabius GS Pre-Use Check
Draeger Fabius GS Pre-Use Check James H Philip MEE, MD, CCE 2011 James H Philip All rights reserved Remember to enter Machine # Fabius GS front view Fabius GS pre-use checkout procedure Fabius GS pre-use
More informationAnesthesia Delivery Systems What Do I Need to Know About My Machine? A Workshop James H. Philip, MEE, MD, CCE
Anesthesia Delivery Systems What Do I Need to Know About My Machine? A Workshop James H. Philip, MEE, MD, CCE Anesthesiologist and Director Bioengineering Department of Anesthesiology, Pain, and Peri-operative
More informationFabius will not compensate for Lung Compliance or Changes in lung compliance
Technology Topics 2007 James H Philip MEE, MD, CCE page 1 Technology Topics 2007 Full Text ANESTHESIA DELIVERY SYSTEMS (ADS) ADS Names and Brands BWH ADSs come from GE-Datex-Ohmeda and Draeger-Siemens
More informationTechnology Topics 2017 James H Philip MEE, MD, CCE page 1
Technology Topics 2017 James H Philip MEE, MD, CCE page 1 Technology Topics 2017 Full Text ANESTHESIA DELIVERY SYSTEMS (ADS) ADS Names and Brands BWH ADSs come from GE-Datex-Ohmeda and Draeger-Siemens
More informationWorking version subject to Departmental approval. Drafted by Dr. Olympio and reviewed by clinical members of The Capital Equipment Committee
Wake Forest University Baptist Medical Center Department of Anesthesiology TECHNICIAN ABBREVIATED CHECKLIST OF RESPONSIBILITIES For Anesthesia Apparatus Checkout Specific to the Draeger APOLLO anesthesia
More informationFlow meter. bellow vaporizer. APL valve. Scavenging system
Introductory Lecture Series: The Anesthesia Machine PORNSIRI WANNADILOK Objectives Anesthesia Machine Ventilators Scavenging Systems System Checkout 1 Flow meter ventilator bellow vaporizer Corrugated
More informationNew Anesthesia Delivery Systems What Do I Need to Know About My New Machine?
New Anesthesia Delivery Systems What Do I Need to Know About My New Machine? James H. Philip, MEE, MD, CCE Anesthesiologist and Director Bioengineering Department of Anesthesiology, Pain, and Peri-operative
More informationVIMA. (Volatile Induction and Maintenance Anesthesia) How and Why. James H. Philip M.E.(E), M.D.
VIMA (Volatile Induction and Maintenance Anesthesia) How and Why James H. Philip M.E.(E), M.D. Copyright 1995-2007, James H Philip, all rights reserved VIMA (Volatile Induction and Maintenance Anesthesia)
More informationAespire 7100 Essential performance Compact design
GE Healthcare Aespire 7100 Essential performance Compact design Features Enhanced monitor integration capabilities with our Datex-Ohmeda Anesthesia Monitor and Compact Anesthesia monitor Lightweight and
More informationAnesthesia Machines. Inhaled Anesthetic Agents
Anesthesia Machines Key Points Anesthesia Machines: - basic components of the generic machine understand how the gasses and agent-mixing part of the machine connects to the patient breathing circuit -
More informationSATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products
Saturn Evo SATURN EVO COLOR A N A E S T H E S I A S Y S T E M i n t e l l i g e n t s i m p l i c i t y Manufacturer of high quality anaesthesia products SATURN EVO COLOR PATIENT SAFETY AND USER PATIENT
More informationFabius Tiro TM. Designed for your environment. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care
Fabius Tiro TM Designed for your environment Emergency Care Perioperative Care Critical Care Perinatal Care Home Care Because you care Traditional hospital ORs have their needs... With more surgical procedures
More informationHallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting
Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting This is a quick reference for setting up, using, and troubleshooting the Hallowell EMC 2000 and 2002 Small Animal Ventilator.
More informationNarkomed MRI-2. A perfect fit for the MRI environment. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care
Narkomed MRI-2 A perfect fit for the MRI environment Emergency Care Perioperative Care Critical Care Perinatal Care Home Care Because you care The complete anesthesia solution for the MRI environment A
More informationRecommendations on Checking Anaesthesia Delivery Systems
Page 1 of 11 Recommendations on Checking Anaesthesia Delivery Version Effective Date 1 Oct 1992 (reviewed Feb 07, Feb 02) 2 2004 3 Nov 2011 4 Dec 2016 Document No. HKCA T1 v4 Prepared by College Guidelines
More informationTHE GE AISYS ANAESTHETIC MACHINE
THE GE AISYS ANAESTHETIC MACHINE All essential components of the anaesthetic machine are shown in the diagram below. Each major component will be explained in detail. The Gas Inlet Module In the diagram
More informationEffective 2/1/17. SC1 Service Call PER Visit $76.00
PSI VT Verify & Test Anesthetic Machine Components $74.00 VOT Verify Vaporizer Output $42.00 PT Pressure Test Machine & Vaporizer $32.00 Total PER Machine $148.00 SC1 Service Call PER Visit $76.00 PSI
More informationKeeping anaesthetic pollution to minimum in theatre environment
Vet Times The website for the veterinary profession https://www.vettimes.co.uk Keeping anaesthetic pollution to minimum in theatre environment Author : Gianluca Bini Categories : General, Vets Date : September
More informationAnesthesiology Research and Scholarship. Follow this and additional works at:
University of New Mexico UNM Digital Repository Anesthesiology Research and Scholarship Anesthesiology 9-1-2013 Breathing Systems Janet Brierley Follow this and additional works at: https://digitalrepository.unm.edu/anesthesiology_pubs
More informationS/5 Aespire. Quick Reference Guide
S/5 Aespire Quick Reference Guide Anesthesia system controls 1 Turn on the system 2 7100 controls and monitored data 3 Menu map 4 Set the ventilation mode 5 Set ventilator controls 6 Vaporizer controls
More informationOperators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200
Operators Manual Supera Floor Stand Anesthesia Machine Models: M1000 Basic Non-rebreathing M1200 Basic M1000 M1200 (All shown with optional equipment) Copyright Supera, LLC 2017 Examination and Preparation
More informationThe Anesthesia Machine (con t) The primary cause of machine malfunction is failure to check Never start without the American Express items What is the
Anesthesia Machine The Anesthesia Machine You are the master of the machine You are responsible for checking the machine prior to each case 1 The Anesthesia Machine (con t) The primary cause of machine
More informationHONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS
RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS 1. INTRODUCTION An anaesthesia delivery system includes any machine, equipment or apparatus which supplies gases, vapours, local anaesthesia and/or
More informationFabius Tiro TM. Designed for your environment. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care
Fabius Tiro TM Designed for your environment Emergency Care Perioperative Care Critical Care Perinatal Care Home Care Traditional hospital ORs have their needs... With more surgical procedures being performed
More informationOperators Manual. Supera Wall Mount Anesthesia Machines. Models: M2300 M2500. (All shown with optional equipment) M2300 M2500
Operators Manual Supera Wall Mount Anesthesia Machines Models: M2300 M2500 (All shown with optional equipment) M2300 M2500 Examination and Preparation for Use Thank you for ordering an Supera Anesthesia
More informationSomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings
SomnoSuite FAQ V1.3 January 2015 Setup 1. How do I connect the SomnoSuite to my oxygen source? 2. Is there a way to speed up the downward movement of the pusher block when setting the empty position? 3.
More informationChecking a Breathing System
Year Group: BVSc3 + Document number: CSL_A04 Equipment list: Equipment for this station: Breathing systems Circle T-Piece Bain (and plunger of a 5ml syringe or red adapter) Parallel Lack Magill Anaesthetic
More informationHallowell EMC Vacuum / Scavenging Interface
Hallowell EMC Vacuum / Scavenging Interface For use in veterinary medical applications only. Operating Manual and Scavenging Primer with Receiving Information, Installation and Maintenance Instructions
More informationTable Top and Convertible Anesthesia Machine Models: M3000 and M3100 (SHOWN WITH OPTIONAL EQUIPMENT)
OPERATOR MANUAL Table Top and Convertible Anesthesia Machine Models: M3000 and M3100 M3000 M3100 (SHOWN WITH OPTIONAL EQUIPMENT) 12330 SE Hwy 212 Clackamas, Oregon 97015 (503) 822-1157 Local and International
More informationHallowell EMC Anesthesia WorkStation Operating Manual
Hallowell EMC Anesthesia WorkStation Operating Manual PN 000a2770 Anesthesia WorkStation A compact (9" W x 9" D x 15" H), convenient system for use on animals under 7 kg (15 lb.) Hallowell Engineering
More informationAnesthesia Workstation
Your healthcare, we care. A9800 Anesthesia Workstation Technical Parameters A9800 Ventilation Modes Alarms Volume Control Pressure Control SIMV with Pressure Support (SIMV + PS) - with Volume or Pressure-type
More informationLandmark. Veterinary Anesthesia Systems. Operator s Manual P/N
Landmark Veterinary Anesthesia Systems Operator s Manual P/N 510-952 Notice This is a Vetland Medical Sales and Services, L.L.C. publication which is protected by copyright. No part of this document may
More informationGE Healthcare. Aestiva/5. More than superior ventilation
GE Healthcare Aestiva/5 More than superior ventilation A balance of superior ventilation and award-winning design SmartVent the shape of superior ventilation At the heart of the Datex-Ohmeda Aestiva /5
More informationA9600 Anesthesia Workstation. Your healthcare, we care.
A9600 Anesthesia Workstation Your healthcare, we care. A9600 is an advanced yet easy to use anesthesia workstation that provides accurate, pneumatically driven and electronically controlled ventilation.
More informationAestiva/5. Operation Manual - Part 1. Software Revision 4.X System Controls, Operation, Checkout
Aestiva/5 Operation Manual - Part 1 Software Revision 4.X System Controls, Operation, Checkout User Responsibility This Product will perform in conformity with the description thereof contained in this
More informationPC7-C Flowmeter O2/N2O Sedation Unit Instruction Manual
PC7-C Flowmeter O2/N2O Sedation Unit Instruction Manual ! WARNING IMPORTANT: READ MANUAL COMPLETELY BEFORE OPERATING THIS DEVICE This manual contains instructions on periodically required checks to be
More informationGE Healthcare. Avance Carestation. Upgrade promotions guide
GE Healthcare Avance Carestation Upgrade promotions guide Contents Display upgrade package 1 for Avance 3 Display upgrade package 2 for Avance 3 Ventilation modes upgrade (PCV-VG, PSVPro, SIMV, Cardiac
More informationA few seconds after turning on, the screen changes to Set Up Screen Before adjusting the pressure, calibrate the O 2 sensors
SiPAP Set-Up Plug into mains, air & O 2, then setup the circuit as shown below Set-up & test before spiking water bag A few seconds after turning on, the screen changes to Set Up Screen Before adjusting
More informationAnesthesia. Midmark Animal Health Products
Anesthesia Midmark Animal Health Products The first and still the best. Safe, simple and controllable inhalation anesthesia Matrx anesthesia machines are the number one choice of veterinary professionals
More informationCarestation 650. Physical Specifications Product Description Carestation 620 A1 ANESTESI CARESTATION 650
Carestation 650 The Carestation 620 is a compact, versatile and easy to use anesthesia system designed to help clinicians deliver reliable anesthesia care to solve today s toughest challenges. Key Features
More informationM5400 Operators Manual
M5400 Operators Manual For Veterinary Use Only Supera, LLC Copyright 2012 Examination and Preparation for Use Thank you for ordering an Supera Anesthesia Innovations anesthesia machine! We are delighted
More informationInternational Journal of Science, Environment and Technology, Vol. 5, No 5, 2016,
International Journal of Science, Environment and Technology, Vol. 5, No 5, 016, 75 764 ISSN 78-3687 (O) 77-663X (P) ANAESTHETIC DELIVERY SYSTEM Rukmani Dewangan* and S.K. Tiwari Department of Veterinary
More informationYour patients, our primary focus. Siesta i Whispa anesthesia machine
Your patients, our primary focus Siesta i Whispa anesthesia machine Functional, and flexible At Dameca we are not just committed to meeting the healthcare needs of today. We also think about tomorrow.
More informationFabius plus. The plus that makes all the difference. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care
Fabius plus The plus that makes all the difference Emergency Care Perioperative Care Critical Care Perinatal Care Home Care Because you care More features, more options add up to more advantages Economical
More informationBreathing Systems. Professor Khalid Bashir
Breathing Systems Professor Khalid Bashir Introduction The devices that connect and deliver anaesthetic gases from anaesthetic machine to patient Spontaneous respiration / IPPV Ports for gas sampling,
More informationPORTER INSTRUMENT COMPANY, INC.
PORTER INSTRUMENT COMPANY, INC. MXR FLOWMETER Instructions and Safety Considerations MXR Models of Porter Conscious Sedation Flowmeters and Bag Tee (Accessory) Wall Mount Models: 2000 and 2000-OS, 3000
More informationAUTOVENT 4000 VENTILATOR
OVERVIEW AUTOVENT 4000 Only properly trained and approved Escambia County Bureau of Public Safety Paramedics are to use the AutoVent 4000 ventilator manufactured by LSP to transport patients already on
More informationService Manual. Anesthesia System ANESTHESIA S ANESTHESIA SYSTEM
Service Manual ANESTHESIA S Anesthesia System ANESTHESIA SYSTEM Service Manual ANESTHESIA S Anesthesia System ANESTHESIA SYSTEM A5 and A3 are U.S. trademarks of Mindray DS USA, Inc. Selectatec is a registered
More informationPrima SP2/3 Anaesthetic Machine Range Service Manual Partnership for Life
Sign-off Copy Prima SP2 USA August 2005 Prima SP2/3 Anaesthetic Machine Range Service Manual Partnership for Life IMPORTANT Servicing and Repairs In order to ensure the full operational life of this anaesthetic
More information2 CYLINDER MOBILE CART USER S MANUAL / INSTRUCTIONS
Parker Hannifin Corporation Precision Fluidics Division Porter Instrument 245 Township Line Road Hatfield, PA 9440 Office 25-723-4000/Fax 25-723-506 www.porterinstrument.com 2 CYLINDER MOBILE CART USER
More informationCOMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS
COMPACT TRIPLE WALL OUTLET STATION MODEL 6255-1 INSTALLATION AND OPERATING INSTRUCTIONS To assure safe operation and conformation to local fire codes, all Porter Outlet Stations are designed to be used
More informationWelcome to the Specialized Medical Services Respiratory training webinar series!
Welcome to the Specialized Medical Services Respiratory training webinar series! SMS is your Long Term Care (LTC) facility single source for oxygen, medical equipment, respiratory care services and supplies
More informationTechnical Data and Specifications
Technical Data and Specifications INTENDED USE Ventilator designed to provide Invasive and Non-invasive ventilation for the critical care management of adult, pediatric and neonate-infant (including premature)
More informationTesting anaesthetic machines, ventilators and related equipment
4. Remove the drum-actuating lever and withdraw the drum. (If the drum is seized or cannot readily be withdrawn, remove the locking ring from the gas outlet and turn the adjusting ring in a clockwise direction
More informationThe Anesthesia Gas Machine
Table of Contents The Anesthesia Gas Machine By Michael P. Dosch CRNA MS ส บค นโดย สาธ ต นฤภ ย ANESTHESIA GAS MACHINE> COMPONENTS & SYSTEMS> SUPPLY OF GASES & ELECTRICITY Gas sources Pipeline Cylinders
More informationAnesthesia Solutions. Prima SP and Prima SP2 Anesthesia Systems. The Future of Anesthesia. Partnership for Life
Anesthesia Solutions The Future of Anesthesia Prima SP and Prima SP2 Anesthesia Systems The Penlon Anesthesia System Range provides the user with their choice of advanced, easy to use anesthesia systems
More informationYour customized anesthesia workstation DRÄGER FABIUS GS PREMIUM. Dräger. Technology for Life
D-9355-2009 Your customized anesthesia workstation DRÄGER FABIUS GS PREMIUM Dräger. Technology for Life 02 DRÄGER FABIUS GS PREMIUM One of the most challenging places D-9354-2009 on earth The modern hospital
More informationInnovating with you Shaping exceptional care
S/5 Avance Innovating with you Shaping exceptional care Shown with S/5 Anesthesia Monitor Features Complete patient monitoring capabilities: respiratory gas, hemodynamic and adequacy of anesthesia Our
More informationState-of-the-art anesthesia technology
State-of-the-art anesthesia technology anesthesia system Get closer to the MRI scanner Superb certified performance in close proximity to strong magnetic fields. Our innovative machine provides benefits
More informationUser Guide and Operating Manual
User Guide and Operating Manual Table Top Laboratory Animal Anesthesia System 901806 & Mobile Laboratory Animal Anesthesia System 901807 / 901809 VetEquip, Inc. - Pleasanton, California USA - 800-466-6463
More informationINTEGRUSPSV ANAESTHESIA WORKSTATION USER MANUAL
INTEGRUSPSV ANAESTHESIA WORKSTATION USER MANUAL This page intentionally left blank 2 Copyright 2005,2006 by Ulco Medical All rights reserved. No part of this publication may be reproduced in any form,
More informationInstallation and commissioning instructions 255 series and 256 series
Installation and commissioning instructions 255 series and 256 series Table of contents 1 General information... 3 1.1 About these instructions... 3 1.2 About this product... 3 1.3 Appropriate usage...
More informationWarning: NEVER ALTER TEST EQUIPMENT
PAGE 1 OF 6 General: Pressure Testing Shop Built Tanks Pressure testing tanks with either positive pressure or vacuum can be a very dangerous exercise. It is important to understand the equipment involved
More informationDialflow Regulator. Instructions for Use
Dialflow Regulator Instructions for Use 702-0030.11 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user
More informationFabius plus. The plus that makes all the difference. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care
Fabius plus The plus that makes all the difference Emergency Care Perioperative Care Critical Care Perinatal Care Home Care More features, more options add up to more advantages Economical To meet today
More informationThe centerpiece of a comprehensive anaesthesia workstation
The centerpiece of a comprehensive anaesthesia workstation IntelliSave AX700 anaesthesia machine A slim design with rich functionality The IntelliSave AX700 is a comprehensive anaesthesia machine from
More informationReclaim Basic Set Up
This purpose of the document is to simplify the set up and understand the Gas Services reclaim system functions. The Gas Services Reclaim Manual is to be used for reference, maintenance, and servicing.
More informationVETROSON. OXY-GEN SYSTEMS (Oxygen Generating Systems) Question & Answer Brochure
VETROSON OXY-GEN SYSTEMS (Oxygen Generating Systems) Question & Answer Brochure Q & A VETROSON OXY-GEN SYSTEMS 1) Q. How does the VETROSON OXY-GEN System work? A: A simplified explanation is that air (21%
More informationHelium-Oxygen Blender
Helium-Oxygen Blender Service Manual Model No. PM5400 Series PM5500 Series (shown) SAVE THESE INSTRUCTIONS 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA 18067 USA Fax: (+001) 610-262-6080 www.precisionmedical.com
More informationNarkomed M Anesthesia System Part No
Technical Service Manual Part No. 4114108 Rev. D Narkomed M Anesthesia System Part No. 4114179 Draeger Medical, Inc. Technical Service Department 3122 Commerce Drive Telford, PA 18969 20 May 2002 2002
More informationNARKOVET ANAESTHETIC UNIT
NARKOVET ANAESTHETIC UNIT MANUAL Art. No. 213062 TELEPHONE +49 7461 96 580 0 www.eickemeyer.com Table of Contents 1 Description...4 2 Overview of the Device...4 3 Startup Procedure...5 4 Operation of
More informationErrors in Monitoring. BWH Clinical Conference 10/06/04. Copyright 2004, James H Philip, all rights reserved.
Errors in Monitoring BWH Clinical Conference 10/06/04 Copyright 2004, James H Philip, all rights reserved. Technology Block 2004 10/06/04 James Philip MD Low flow&closed circuit safety&danger 10/06/04
More informationOxygen Dialflow Meter. Instructions for Use
Oxygen Dialflow Meter Instructions for Use 702-0031.9 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user
More informationYour patients, our primary focus
Your patients, our primary focus Siesta i Whispa anesthesia machine Functional, fail-safe and flexible to your needs Take a good look, and you ll be surprised by the many forward-thinking features of the
More informationStatement. Aeonmed have the rights to replace any content in this manual without notice.
Statement Beijing Aeonmed Co, Ltd. (Aeonmed for short) holds the copyrights to this manual, which is non-public published, and reserves the rights to keep it as a secure document. Refer to this manual
More informationMicro Dial-Flowmeter. Instructions for Use
Micro Dial-Flowmeter Instructions for Use 702-0082.8 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user
More informationO-Two Self-Study Guide. Equinox Relieve
O-Two Self-Study Guide Equinox Relieve PRODUCT DESCRIPTION/PRINCIPLES OF OPERATION The Equinox Relieve provides spontaneously breathing patients with a preset mixture of oxygen and nitrous oxide for analgesia
More information1 Exam Prep NFPA 99 Health Care Facilities Questions and Answers (Plumbing Contractor)
1 Exam Prep NFPA 99 Health Care Facilities Questions and Answers (Plumbing Contractor) 1. According to NFPA 99, SCFM is an acronym for. a) Surface Conditions at Fahrenheit Mercury b) Standard Conditions
More information2011 Equipment Catalog
2011 Equipment Catalog Veterinary Equipment Products Table of Contents Monitoring Equipment Temperature Management Anesthesia Equipment MR Conditional Equipment Ventilators Vaporizers Gas Delivery Management
More informationVT900A + VAPOR. Technical data. One Solution. Auto-detection to Ensure Patient Safety. Key benefits and features
Technical data VT900A + VAPOR The Fluke Biomedical VT900A + VAPOR is designed to efficiently and reliably perform a full anesthesia machine PM, from ventilators to vaporizers. One Solution VT900A + VAPOR
More informationSERVICE KITS AND CALIBRATION NOTES
SERVICE KITS AND CALIBRATION NOTES Copyright 2006 by Ulco Medical All rights reserved. No part of this publication may be reproduced in any form, in an electronic retrieval system or otherwise, without
More informationSIGNET 615 ANAESTHESIA WORKSTATION USER MANUAL
SIGNET 615 ANAESTHESIA WORKSTATION USER MANUAL Copyright 2002, 2005 by Ulco Medical All rights reserved. No part of this publication may be reproduced in any form, in an electronic retrieval system or
More informationIV-ANE. User s Manual. Anesthesia equipment for test animals. Petition
User s Manual IV-ANE Anesthesia equipment for test animals Petition Thank you for your purchase of Anesthesia equipment for test animals at this time. Prior to using these products, read this instruction
More informationYoke Block Instruction Manual
Yoke Block Instruction Manual ! WARNING IMPORTANT: READ MANUAL COMPLETELY BEFORE OPERATING THIS DEVICE This manual contains instructions on periodically required checks to be performed by the user. These
More informationParker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA USA Tel: Ref Rev.
Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA 19440 USA Tel: 215-723-4000 Ref. 10049700 Rev. L Table of Contents Warnings...2 Examination...2 Description and
More informationNew product release. Universal Rebreather Monitor (URBM) Three cell / Independent backup
New product release Closed Circuit Research is pleased to announce the launch of our range of Universal Rebreather Monitors Universal Rebreather Monitor (URBM) Three cell / Independent backup Key features
More informationSTANDARD FLOWMETER USER S MANUAL
STANDARD FLOWMETER USER S MANUAL Instructions and Safety Considerations MXR Models of Porter Conscious Sedation Flowmeters and Bag Tee (Accessory) Flowmeter Models: 1000 1100 Bag Tee Model: P1407A Porter
More informationPressure Regulator. Instructions for Use
Pressure Regulator Instructions for Use 702-0083.9 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient, the
More informationCompact Triple Cabinet Outlet Station Model B Installation and Operating Instructions
PORTER Parker Hannifin Corporation Porter Instrument Division 245 Township Line Rd. P.O. Box 907 Hatfield, PA 19440-0907 USA (215) 723-4000 / fax (215) 723-5106 Compact Triple Cabinet Outlet Station Model
More informationCompact Triple Cabinet Outlet Station Model Installation and Operating Instructions
Compact Triple Cabinet Outlet Station Model 6258-1 Installation and Operating Instructions The Porter Compact Triple Outlet Station (6258-1) provides a quick, safe, and reliable method of connection to
More informationSelecting and Connecting Breathing Systems
Selecting and Connecting Breathing Year Group: BVSc3 + Document number: CSL_A03 Equipment for this station: Equipment list: Pen Paper Calculator T-piece (in CSL a strip of white tape is around this system)
More informationNitrous Oxide / Oxygen FLOOR MOUNTED OUTLET STATION Installation and Instructions
FM-171 Rev. B 8/00 Nitrous Oxide / Oxygen FLOOR MOUNTED OUTLET STATION Installation and Instructions MODEL 6300-1 PORTER INSTRUMENT COMPANY, INC. 245 TOWNSHIP LINE RD. P.O. BOX 907 HATFIELD, PA 19440-0907
More informationComparison Of The Efficacy Of Two Low Fresh Gas Flow Techniques In Low Flow Anaesthesia
ISPUB.COM The Internet Journal of Anesthesiology Volume 7 Number 1 Comparison Of The Efficacy Of Two Low Fresh Gas Flow Techniques In Low Flow Anaesthesia S Kulandayan, L Ng, D Kamarudin Citation S Kulandayan,
More informationAIR/OXYGEN BLENDER INSTRUCTION MANUAL
BIO-MED DEVICES AIR/OXYGEN BLENDER INSTRUCTION MANUAL CATALOG #2120 REV 042203 BIO-MED DEVICES INC. 1445 BOSTON POST ROAD, GUILFORD, CT 06437 (203) 458-0202 FAX (203) 458-0440 www.biomeddevices.com TABLE
More informationSTANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA
STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA SOP#: 1157.1 Date Issued: 05/14 Date Revised: 5/15 Page 1 of 6 TITLE: SCOPE: RESPONSIBILITY: PURPOSE: SurgiVet
More informationR409 Plus Veterinary Ventilator Machine
R409 Plus Veterinary Ventilator Machine User Manual v1.2 CONTENTS 1 User/Owner Responsibility... 1 2 Warnings and Cautions... 1 3 Warranty... 2 4. Introduction... 2 5 Unpacking... 4 6 Schematic diagram
More informationINSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y
INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model 50 Systems Outfitted with ABM-725 Monitor M A R T E C H S E R V I C E S C O M P A N Y OFFICE: (507) 843-4700 P.O. BOX 7079 Toll
More information3:1 High Ratio Oil Pump W. Extn. Kit
3:1 High Ratio Oil Pump W. Extn. Kit OWNER S MANUAL WARNING: Read carefully and understand all INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety precautions may result
More informationKinetics of Inhaled Anesthetic Agents
Kinetics of Inhaled Anesthetic Agents Copyright 1995-2011, James H Philip, all rights reserved. Dr. James Philip has performed funded research on Isoflurane, Sevoflurane, and Desflurane. He is often supported
More information