Patient Manual. Oxygen Concentrator Oxygen Monitor, Air Outlet, Dual Flow, and Pediatric Flow Options

Size: px
Start display at page:

Download "Patient Manual. Oxygen Concentrator Oxygen Monitor, Air Outlet, Dual Flow, and Pediatric Flow Options"

Transcription

1 Patient Manual Oxygen Concentrator Includes Oxygen Monitor, Air Outlet, Dual Flow, and Pediatric Flow Options

2

3 DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL. IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS AND INSTRUCTIONS, CONTACT YOUR EQUIPMENT PROVIDER BEFORE ATTEMPTING TO USE THIS EQUIPMENT; OTHERWISE, INJURY OR DAMAGE MAY RESULT. Smoking while using oxygen is the number one cause of fire, injuries and related deaths. You must follow these safety warnings: Do not allow smoking, candles, or open flames in the same room with the device or within 5 feet (1.52 meters) of the oxygen-carrying accessories. Smoking while wearing an oxygen cannula may cause facial burns and possibly death. Removing the cannula and putting it on bedding, sofas, or other cushion material will cause a flash fire when exposed to a cigarette, heat source, or flame. If you smoke, these 3 steps may save your life: turn off the oxygen concentrator, take off the cannula, and leave the room where this device is located. No Smoking Oxygen in Use signs must be prominently displayed in the home, or where the oxygen concentrator is in use. Patients and their caregivers must be informed about the dangers of smoking in the presence of, or while using, medical oxygen. English: A multilingual version of the manual is available through your equipment provider. Español: Una versión multilingüe del manual está disponible a través de su proveedor de equipo. Français: Une version multilingue du manuel est disponible par l'intermédiaire de votre fournisseur de matériel. Deutsche: Eine mehrsprachige Version des Handbuchs ist in Ihrer Geräte-Anbieter.

4

5 TABLE OF CONTENTS AirSep NewLife Elite Oxygen Concentrator AirSep NewLife Elite Oxygen Concentrator 1 Important Safety Rules 1 Why Your Physician Prescribed Oxygen 6 What is an Oxygen Concentrator? 7 How to Operate Your Oxygen Concentrator 8 Filters 12 Oxygen Without Humidifier 12 Oxygen With Humidifier 13 Nasal Cannula 14 Proper Setting of Oxygen Flowmeter 15 Cleaning, Care, and Proper Maintenance 16 Cabinet 16 Filters 17 Reserve Oxygen Supply 17

6 TABLE OF CONTENTS AirSep NewLife Elite Oxygen Concentrator Troubleshooting 18 Product Specifications 20 Classification 21 Oxygen Monitor Option 22 Function of the Oxygen Monitor 22 Alarm Signal 22 Air Outlet Option 23 Operating the NewLife Elite Air Outlet 23 Dual Flow and Pediatric/ Low Flow Options 30 Dual Flow Application 31 Pediatric/Low Flow Application 31 Setting the Pediatric Flowmeter 32 Symbols/Abbreviations 33

7 [Read the Important Safety Rules section before operating this equipment.] AirSep NewLife Elite Oxygen Concentrator This Patient Manual will acquaint you with AirSep s NewLife Elite Oxygen Concentrator. Make sure you read and understand all the information contained in this manual before you operate your unit. Should you have any questions, your Equipment Provider will be happy to answer them for you. Important Safety Rules Carefully review and familiarize yourself with the following important safety information about the NewLife Elite Oxygen Concentrator. No Smoking Oxygen in Use signs must be prominently displayed in the home, or where the oxygen concentrator is in use. Patients and their caregivers must be informed about the dangers of smoking in the presence of, or while using, medical oxygen. This device supplies high-concentration oxygen that promotes rapid burning. Do not allow smoking or open flames within the same room of (1) this device, or (2) any oxygen-carrying accessory. Failure to observe this warning can result in severe fire, property damage and / or cause physical injury or death. This unit is not to be used for life support. Geriatric, pediatric, or any other patients unable to communicate discomfort while using this machine may require additional monitoring. Patients with hearing and/or sight impairment(s) may need assistance with monitoring alarms. If you feel discomfort or are experiencing a medical emergency, seek medical assistance immediately Use no oil, grease, or petroleum-based or other flammable products on or near nasal end of cannula or the NewLife Elite unit. Oxygen accelerates the combustion of flammable substances. 1

8 Electrical shock hazard. Disconnect the power cord from the electric outlet before you clean the unit to prevent accidental electrical shock and burn hazard. Only your Equipment Provider or a qualified service technician should remove the covers or service the unit. Do not leave a nasal cannula under bed coverings or chair cushions. If the unit is turned on but not in use, the oxygen will make the material flammable. Set the I/0 power switch to the 0 (off) position when the NewLife Elite unit is not in use. Do not use extension cords with this unit or connect too many plugs into the same electrical outlet. The use of extension cords could adversely affect the performance of the device. Too many plugs in one outlet can result in an overload to the electrical panel causing the breaker/fuse to activate or fire if the breaker or fuse fails to operate. Use only voltage specified on back panel label. Do not use your oxygen concentrator in the presence of flammable gases. This can result in rapid burning causing property damage, bodily injuries or death. Care should be taken to prevent NewLife Elite from getting wet or allowing water to enter the unit. This can cause the unit to malfunction or shut down, and cause an increased risk for electrical shock or burns. Do not use liquid directly on the unit. A list of undesirable chemical agents includes but is not limited to the following: alcohol and alcohol-based products, concentrated chlorine-based products (ethylene chloride), and oil-based products (Pine-Sol, Lestoil ). These are NOT to be used to clean the plastic housing on NewLife Elite, as they can damage the unit s plastic. 2

9 Clean the cabinet, control panel, and power cord only with a mild household cleaner applied with a damp cloth or sponge, and then wipe all surfaces dry. Do not allow any liquids to enter the concentrator. Ensure concentrator is operated in an upright position. AirSep recommends an alternate source of supplemental oxygen in the event of a power outage, alarm condition, or mechanical failure. Consult your physician or Equipment Provider for the type of reserve system required. Federal (USA) law restricts this device to sale or rental by order of a physician or other licensed health care provider. It is very important to follow the prescribed level of oxygen flow. Do not increase or decrease the flow until you first consult your physician. In the event of an alarm or you observe the NewLife Elite is not working properly; consult the troubleshooting section of this manual. If you cannot resolve the problem, consult your Equipment Provider. Do not operate this unit in a restricted or confined space where ventilation can be limited. This can cause the device to overheat and affect performance. Storing the device outside of its temperature specifications may affect performance (refer to the specification section of the manual). 3

10 If the alarm is weak or does not sound at all, consult your Equipment Provider immediately. Position the unit away from curtains or drapes, hot air registers or heaters. Be certain to place the unit on a flat surface and make sure all sides are at least 1 foot (30 cm) away from a wall or other obstruction. Do not place the unit in a confined area. Choose a dust and smoke free-location away from direct sunlight. Do not operate the unit outdoors unless the unit is plugged into a Ground Fault Circuit Interrupter (GFCI ) protected outlet. If the humidifier bottle tubing is not properly connected to the humidifier bottle fitting or the oxygen outlet, an oxygen leak can occur. Normally, you should not need to adjust the flowmeter on your unit. If you turn the flowmeter adjustment knob clockwise, you will decrease and can shut off the flow of oxygen from your unit. For your convenience, the flowmeter is marked in ½ lpm increments from 1 to 5 lpm. For units with the pediatric flowmeter option, the flowmeter is marked in 1/8 lpm increments for flow settings up to 2 lpm. To prevent a voided AirSep warranty, follow all manufacturers instructions. Replace the disposable cannula periodically following normal usage according to your Equipment Providers recommendations. AirSep does not recommend the sterilization of this equipment. Do not attempt any maintenance other than the possible solutions listed below. 4

11 Always follow the cannula manufacturer s instructions for proper use. Consult your licensed health care provider to determine how often the cannula should be replaced. The standard NewLife Elite Oxygen Concentrator accommodates prescriptions from 1 to 5 lpm. For prescriptions levels below 1 lpm, you must operate the NewLife Elite unit for at least 5 minutes at 2 lpm. The use of some oxygen administration accessories not specified for use with this oxygen concentrator may impair its performance. The use of certain humidifiers not specified for use with this oxygen concentrator may impair its performance. To Equipment Provider: The following humidifier bottle is recommended for use with the NewLife Elite Oxygen Concentrator: AirSep Part No. HU003-1 To Equipment Provider: The following oxygen administration accessories are recommended for use with the NewLife Elite Oxygen Concentrator: Nasal Cannula with 7 feet (2.1 m) of tubing: AirSep Part No. CU Oxygen Outlet Adapter: AirSep Part No. F

12 [Read the Important Safety Rules section before operating this equipment.] Why Your Physician Prescribed Oxygen Many people suffer from a variety of heart, lung, and other respiratory diseases. A significant number of these patients can benefit from supplemental oxygen therapy at home, in the hospital, or at a medical facility. Oxygen is a gas that makes up 21% of the room air we breathe. Our bodies depend on a steady supply to function properly. Your physician prescribed a flow or setting to address your particular respiratory condition. Although Oxygen is a non-additive drug, unauthorized oxygen therapy can be dangerous. You must seek medical advice before you use this oxygen concentrator. The Equipment Provider who supplies your oxygen equipment will demonstrate how to set the prescribed flow rate. It is very important to select the prescribed level of oxygen flow. Do not increase or decrease the flow until you first consult your physician. 6

13 [Read the Important Safety Rules section before operating this equipment.] What is an Oxygen Concentrator? Oxygen concentrators were introduced in the mid-1970 s and have become the most convenient, reliable source of supplemental oxygen available today. Oxygen concentrators are the most cost-effective, efficient, and safest alternative to using high-pressure oxygen cylinders or liquid oxygen. An oxygen concentrator provides all the oxygen you need with no cylinder or bottle deliveries required. The air we breathe contains approximately 21% oxygen, 78% nitrogen, and 1% other gases. In the NewLife Elite unit, room air passes through a regenerative, adsorbent material called molecular sieve. This material separates the oxygen from the nitrogen. The result is a flow of high-concentration oxygen delivered to the patient. There is never a danger of depleting the oxygen in a room when you use your NewLife Elite unit. 7

14 [Read the Important Safety Rules section before operating this equipment.] How to Operate Your Oxygen Concentrator E F First, become familiar with the important parts of your NewLife Elite Oxygen Concentrator (Figures 1a and 1b). A. On/Off (I/0) Power Switch: Starts and stops the operation of the unit. B. Circuit Breaker Reset Button: Resets the unit after electrical overload shutdown C. Digital Hour Meter: Records the unit s total hours of operation. D. Flowmeter/Adjustment Knob: Controls and indicates the oxygen flow rate in liters per minute (lpm). E. Oxygen Outlet : Provides connections for a humidifier (if required) or nasal cannula. F. Top and Side Handles: Figure 1a Enables convenience in carrying the unit. G. Operating Instructions: Explains procedures to operate the unit. C B A D G F 8

15 [Read the Important Safety Rules section before operating this equipment.] I. Air Intake Gross Particle Filter: Prevents dust and other airborne particles from entering the unit. J. Power Cord: Allows connection of unit into electrical outlet. I Do not use extension cords with this unit or connect too many plugs into the same outlet. Too many plugs in one outlet can result in an overload to the electrical panel causing the breaker/fuse to activate or fire if the breaker or fuse fails to operate. J 1. Locate the unit near an electrical outlet in the room where you spend most of your time. 2. Position the unit away from curtains or drapes, hot air registers, heaters, and fireplaces. Be certain to place the unit so all sides are at least 12 inches (30.5 cm) away from a wall or other obstruction. Do not place the unit in a Figure 1b confined area. 3. Turn the unit so that the operating controls are within easy reach and the air intake on the back of the unit is not obstructed. 4. Connect oxygen accessories such as a humidifier (if required), nasal cannula, and/or extension tubing to the oxygen outlet. 5. Completely unwrap the power cord (Figure 1b). 9

16 [Read the Important Safety Rules section before operating this equipment.] 6. Insert power cord into the electrical outlet. 7. Locate the power switch on the front of the unit, and switch it to the I position (on). (Figure 2.) A battery-operated audible alarm must sound for a 5-second test to indicate a good battery and alarm. On/Off Power Switch If the alarm is weak or does not sound at all, consult your Equipment Provider immediately. The standard NewLife Elite Oxygen Concentrator accommodates prescriptions from 1 to 5 lpm. Figure 2 8. Dual Flow and High Flow Applications: Set the flowmeter adjustment knob(s) to the prescribed lpm, in any combination of flows up to a total of 5 lpm for Dual 5 units (Figure 3). or Pediatric/Low Flow Applications: Set the flowmeter adjustable knob(s) to the prescribed lpm up to 2 lpm in 1/8 liter (125 ccm) increments. The concentrator is now ready for use. Figure 3 10

17 [Read the Important Safety Rules section before operating this equipment.] 9. To turn the concentrator off, press the I/0 switch to the 0 position. Ensure concentrator is operated in an upright position For prescriptions levels below 1 lpm, you must operate the NewLife Elite unit for at least 5 minutes at 2 lpm. 10. If the NewLife Elite unit fails to operate properly, refer to the Troubleshooting section for a list of probable causes and solutions. 11

18 [Read the Important Safety Rules section before operating this equipment.] Filters Air enters the NewLife Elite unit through an air intake gross particle filter located on the back off the oxygen concentrator. This filter removes dust particles and other large particles from the air. Before you operate the NewLife Elite unit, make sure this filter is clean and positioned correctly (Figure 4). The supplemental oxygen produced by the NewLife Elite unit receives additional filtration from a product filter located within the oxygen concentrator. Your Equipment Provider performs maintenance on the product filter in addition to other maintenance on the unit. Figure 4 The use of some oxygen administration accessories not specified for use with this oxygen concentrator may impair its performance. Oxygen Without Humidifier 1. If your physician did not prescribe a humidifier, connect the oxygen tubing directly to the unit s Oxygen outlet. A separate outlet fitting is supplied for this type of connection (Figure 5). Figure 5 12

19 [Read the Important Safety Rules section before operating this equipment.] Operating With Humidifier Following these steps if your physician prescribed an oxygen humidifier as part of your therapy: 1. Remove or unscrew the reservoir bottle from the humidifier (If you have a pre-filled unit, do not perform this step. Proceed directly to step 4.) 2. Fill the reservoir with cool or cold water (distilled water is preferred) to the fill line indicated on the bottle. DO NOT OVERFILL. 3. Screw the reservoir bottle back together. 4. On the top of the humidifier, turn the thread nut counterclockwise while you connect the humidifier to the oxygen outlet, and tighten securely (Figure 6). Figure 6 5. Connect oxygen tubing from the nasal cannula to the humidifier outlet fitting (Figure 7). The use of certain humidifiers not specified for use with this oxygen concentrator may impair its performance. To Equipment Provider: The following humidifier bottle is recommended for use with the NewLife Elite Oxygen Concentrator: AirSep Part No. HU003-1 Figure 7 13

20 [Read the Important Safety Rules section before operating this equipment.] Nasal Cannula Your physician has prescribed either a nasal cannula, face mask, or other accessories (Figure 8). In most cases the manufacturer has already connected the oxygen supply tubing to the nasal cannula, face mask, or other accessory. If not, follow the manufacturer s instructions for proper connection. Connect the oxygen tubing to the oxygen outlet adapter or humidifier. To Equipment Provider: The following oxygen administration accessories are recommended for use with the NewLife Elite Oxygen Concentrator: Nasal Cannula with 7 feet (2.1 m) of tubing: AirSep Part No. CU Oxygen Outlet Adapter: AirSep Part No. F Figure 8 14

21 [Read the Important Safety Rules section before operating this equipment.] Proper Setting of Oxygen Flowmeter To set the proper flow of supplemental oxygen, turn the flowmeter adjustment knob left or right until the ball inside the flowmeter centers on the flow line number prescribed by your physician (Figure 9). To view the flowmeter at the proper angle, note that the back line and the front numbered line must give the appearance of just one line. It is very important to follow the prescribed level of oxygen. Do not increase or decrease the flow until you first consult your physician. Normally, you should not need to adjust the flowmeter on your unit. If you turn the flowmeter adjustment knob clockwise, you will decrease and can shut off the flow of oxygen from your unit. For your convenience, the flowmeter is marked in ½ lpm increments from 1 to 5 lpm. For units with the pediatric flowmeter option, the flowmeter is marked in 1/8 lpm increments for flow settings up to 2 lpm. Figure 9 15

22 [Read the Important Safety Rules section before operating this equipment.] Cleaning, Care, and Proper Maintenance Cabinet Electrical shock hazard. Disconnect the power cord from the electric outlet before you clean the unit to prevent accidental electrical shock and burn hazard. Only your Equipment Provider or a qualified service technician should remove the covers or service the unit. Do not use liquid directly on the NewLife Elite unit to clean it. A list of undesirable chemical agents includes but is not limited to, the following, according to the plastics manufacturer: alcohol and alcoholbased products, concentrated chlorine-based products (ethylene chloride), and oil-based products (Pine- Sol, Lestoil ). These are NOT to be used to clean the plastic housing on NewLife Elite, as they can damage the unit s plastic. Clean the cabinet, control panel, and power cord only with a mild household cleaner applied with a damp cloth or sponge, and then wipe all surfaces dry. Do not allow any liquids to enter the concentrator. To prevent a voided AirSep warranty, follow all manufacturers instructions. AirSep does not recommend the sterilization of this equipment. 16

23 [Read the Important Safety Rules section before operating this equipment.] Filters Do not operate the unit without the air intake gross particle filter in place. At least one time each week, wash the air intake gross particle filter, which is located in the back of the unit. Your Equipment Provider may advise you to clean it more often, depending upon your operating conditions. Follow these steps to properly clean the air intake filter: 1. Remove the filter and wash it in a warm solution of soap and water. 2. Rinse the filter thoroughly, and remove excess water with a soft, adsorbent towel. Ensure that the filter is dry before replacing it. 3. Replace the dry filter. Reserve Oxygen Supply Your Equipment Provider may recommend another source of supplemental oxygen therapy in case there is a mechanical failure or a power outage. 17

24 Troubleshooting If your NewLife Elite Oxygen Concentrator fails to operate properly, refer to the chart on the following pages for possible causes and solutions and, if needed, consult your Equipment Provider. If you cannot get the unit to operate, connect your nasal cannula, face mask, or other accessories to the reserve supplemental oxygen supply (if provided). Do not attempt any maintenance other than the possible solutions listed below. Problem Probable Cause Solution Unit does not operate. Power failure condition causes a continuous alarm to sound. Power cord not connected into electrical outlet. No power at electrical outlet. Oxygen concentrator circuit breaker is activated. Check power cord plug at the electrical outlet for a proper connection. Check power source, wall switch, fuse, or circuit breaker in-house. Contact your Equipment Provider for service. Limited oxygen flow. Dirty or obstructed humidifier bottle. Defective nasal cannula, face mask, catheter, and/or oxygen delivery tube, or other accessory. Other leak or restriction. Remove the humidifier bottle (if used) from the oxygen outlet. If flow is restored, clean or replace with a new humidifier bottle. Remove nasal cannula, face mask, or other accessories from oxygen tubing. If proper flow is restored, replace with new nasal cannula, face mask, or other accessories. Disconnect delivery tubing at oxygen outlet (front of unit). If proper flow is restored, check oxygen tubing for kinks or obstructions. Replace if needed. Contact your Equipment Provider. 18

25 Problem Probable Cause Solution Condensation collects in the oxygen tubing when you use the humidifier bottle. Unit not properly ventilated. Elevated operating temperature. Make sure unit is positioned away from curtains or drapes, hot air registers, heaters, and fireplaces. Be certain to place the unit so all sides are at least 12 inches (30.5 cm) away from a wall or other obstruction. Do not place the unit in a confined area. Allow oxygen tubing to dry out, or replace with new tubing. Refill humidifier bottle with COLD water. DO NOT OVERFILL. Intermittent alarm sounds at one second intervals. Equipment malfunction. Set I/0 power switch to 0 position, use your reserve oxygen supply (if provided), and consult your Equipment Provider immediately. Unit does not alarm, or weak alarm sounds for 5 seconds during start-up. Weak 9-volt battery Call your Equipment Provider to replace 9-volt battery. All other problems. Set I/0 power switch to the 0 position, use your reserve oxygen supply (if provided), and consult your Equipment Provider immediately. 19

26 Product Specifications Oxygen Concentration: * 1-3 lpm: 95.5% % 4 lpm: 92% ± 3% 5 lpm: 90% ± 3% Dimensions: 28.5 in. high x 15.7 in. wide x 14.5 in. deep (72.4 cm high x 40 cm wide x 36.8 cm deep) Weight: 54 lb (24.5 kg) Electrical: 120 VAC, 60 Hz, 4.0 amps 350 watts 5 lpm model Two-prong polarized plug Double insulated cabinet CSA-Approved to IEC Export Models: VAC, 50 Hz, 2.0 amps 220 VAC, 60 Hz, 2.0 amps Double-insulated cabinet Alarms: Storage Temperature: Battery test Power failure High and low pressure Low oxygen concentration (with Oxygen M onitor option) -4 F to 140 F (-20 C to 60 C) * Based on an atmospheric pressure of 14.7 psi (101 kpa) at 70 F (21 C) * Operating unit outside of operational temperature range can affect performance. 20

27 Classification Type of protection against electric shock: Class II Protection from electric shock is achieved by double insulation. Protective earthing or reliance upon installation conditions are not required. Degree of protection against electric shock: Type B Equipment providing a particular degree of protection against electric shock particularly regarding: 1) allowable leakage current; 2) reliability of protective earth connection (if present). Not intended for direct cardiac application. Degree of protection against harmful ingress of water: Drip-proof equipment IPX1. Equipment provided with an enclosure preventing of such an amount of falling liquid as might interfere with the satisfactory and safe operation of the equipment. Method of cleaning and infection control allowed: Please refer to Maintenance section in the NewLife Elite Service Manual. Degree of safety of application in the presence of flammable anesthetic gases: Equipment not suited for such application. Mode of operation: Continuous duty. 21

28 Oxygen Monitor Option [Read the Important Safety Rules section before operating this equipment.] The following information will acquaint you with Oxygen Monitor option of the NewLife Elite Oxygen Concentrator. Make sure you read and understand all the information contained in this manual before you operate your unit. Should you have any questions, your Equipment Provider will be happy to answer them for you. If you feel discomfort or are experiencing a medical emergency, seek medical assistance immediately. This unit is not to be used for life support. Geriatric, pediatric, or any other patients unable to communicate discomfort while using this machine may require additional monitoring. Patients with hearing and/or sight impairment(s) may need assistance with monitoring alarms. Contact your Equipment Provider immediately if the amber OXYGEN MONITOR light remains on for more than 15 minutes. When you turn the unit on, it s normal for the amber OXYGEN MONITOR light to turn on and remain on for up to five minutes. Function of the Oxygen Monitor The oxygen monitor is a small electronic device within the NewLife Elite Oxygen Concentrator that monitors the concentration of oxygen produced by the unit. Alarm Signal If oxygen concentration falls below the acceptable therapeutic level, an amber OXYGEN MONITOR light on the Oxygen Concentrator turns on (Figure 10). If the light remains on for more than 15 minutes, an intermittent alarm sounds. Amber Light Figure 10 22

29 [Read the Important Safety Rules section before operating this equipment] Air Outlet Option The following information will acquaint you with the Air Outlet option for the NewLife Elite Oxygen Concentrator. Make sure you read and understand all the information in this NewLife Elite Patient Manual before you operate your unit. Should you have any questions, your Equipment Provider will be happy to answer them for you. If you feel discomfort or are experiencing a medical emergency, seek medical assistance immediately. This unit is not to be used for life support. Geriatric, pediatric, or any other patients unable to communicate discomfort while using this machine may require additional monitoring. Patients with hearing and/or sight impairment(s) may need assistance with monitoring alarms. Operating the NewLife Elite Air Outlet 1. Read and understand all information contained in the NewLife Elite Patient Manual s How to Operate Your Oxygen Concentrator section before you operate your unit. 23

30 [Read the Important Safety Rules section before operating this equipment.] The NewLife Elite Air Outlet option allows you to connect a hand held nebulizer (Figure 1). You may continue to receive oxygen from the unit while you use the Air Outlet option. Figure 1 Figure 2 Air Outlet option * Shown with hand held nebulizer Air Valve Part No. VA007-1 Air Outlet Barb Fitting Part No. F

31 [Read the Important Safety Rules section before operating this equipment.] 2. Locate the air outlet barb fitting on the front of the unit (Figure 2). If a hand held nebulizer will be used, connect one end of the air supply tubing to the air outlet barb fitting and the other end to the bottom of the nebulizer. Oxygen-enriched air is not delivered at the air outlet. In high humidity environments or during extended period of non-use, open the air valve completely (Figure 3) to purge/flush the system. 3. Fill the nebulizer cup with medication as prescribed by your physician (Refer to the Nebulizer With Medication section for filling instructions). 4. Operate the NewLife Elite unit for at least five minutes, and then open the air valve completely (Figure 3). 5. Begin your treatment (Refer to the Inhaling Medication/Treatment Instruction sections). Nebulizer medication will now be visible as a fine mist.* Figure 3 25

32 [Read the Important Safety Rules section before operating this equipment.] 6. When treatment is complete, turn the air valve to the OFF position. (Figure 4) The NewLife Elite Air Outlet regulator is preset to deliver 6 liters per minute ( lpm) at 12 psig (85 kpa). Figure 4 7. Disconnect the nebulizer and the air supply tubing from the air outlet barb fitting. 8. Clean the nebulizer. (Refer to the Cleaning the Nebulizer section). * If you think that your nebulizer is not operating properly, contact your Equipment Provider. Filling the Nebulizer with Medication 1. Wash your hands thoroughly. 2. Use an eyedropper, syringe, or other measuring device to measure out the proper amount of medication, as prescribed by your physician. Use only the amount of medication and frequency of treatment that your physician prescribed 26

33 [Read the Important Safety Rules section before operating this equipment.] 3. Remove or unscrew the medication cup on the nebulizer, and place your prescribed measured dosage into the medication cup (Figure 5). 4. Connect the medication cup to the nebulizer, and then connect the T piece or mouthpiece to the nebulizer (Figure 6). Figure 5 Figure 6 5. Connect one end of the air supply tubing to the air outlet barb fitting and the other end to the bottom of the nebulizer, and open the air valve completely as shown in Figure Begin your treatment. (Refer to the Inhaling Medication/Treatment Instruction section) 27

34 [Read the Important Safety Rules section before operating this equipment.] Inhaling Medication/Treatment Instructions The following instructions for inhaling medication are often recommended. If your physician or health care professional has given you special instructions, make sure you follow them instead, as prescribed. 1. Close your mouth around the mouthpiece, but do not hold it with your teeth (Figure 7). 2. Take a slow, deep breath, and pause at the end of the inhalation for 1-2 seconds, then exhale slowly and completely. 3. Repeat this procedure until the prescribed amount of medication nebulizes or the Prescribed amount of treatment time elapses (whichever occurs first). 4. If your physician or health care professional instructed you to take short rest periods Figure 7 During your treatment, make sure you turn the air valve to the OFF position, as shown in Figure 4. This will conserve your medication. Prolonged treatment time can indicate a defective nebulizer. Contact your Equipment Provider if this condition exists. 28

35 [Read the Important Safety Rules section before operating this equipment.] Cleaning the Nebulizer Perform steps 1 and 2 below after each treatment to prevent medication from collecting and hardening inside the nebulizer parts. 1. After each treatment, separate the nebulizer and the T piece or mouthpiece assembly. 2. Remove or unscrew the nebulizer cup, and rinse each component thoroughly in warm water. 3. Once a day, clean all nebulizer parts (excluding air supply tubing) with a mild detergent or soap solution in warm water. Rinse thoroughly, and soak all parts in a solution of Figure 8 one (1) part white vinegar and three (3) parts water for 30 minutes to disinfect. 4. Rinse thoroughly in warm water to remove the disinfectant solution. 5. Place all nebulizer parts on a paper towel or soft absorbent material to air dry. DO NOT WIPE DRY. 6. When dry, store the nebulizer parts in a clean container or plastic bag. 7. Repeat the above procedure after each treatment/patient use. Federal (USA) law restricts this device to sale or rental by order of a physician or other licensed health care provider. 29

36 [Read the Important Safety Rules section before operating this equipment.] Dual Flow and Pediatric/Low Flow Options The following information will acquaint you with the 5 liter dual flow and pediatric/low flow options of the NewLife Elite Oxygen Concentrator (see Figure 9). Make sure you read and understand all the information contained in this manual before you operate your unit. Should you have any questions, your Equipment Provider will be happy to answer them for you. No Smoking Oxygen in Use signs must be prominently displayed in the home, or where the oxygen concentrator is in use. Patients and their caregivers must be informed about the dangers of smoking in the presence of, or while using, medical oxygen. If you feel discomfort or are experiencing a medical emergency, seek medical assistance immediately. This unit is not to be used for life support. Geriatric, pediatric, or any other patients unable to communicate discomfort while using this machine may require additional monitoring. Patients with hearing and/or sight impairment(s) may need assistance with monitoring alarms. Use no oil, grease, or petroleum-based or other flammable products on or near nasal end of cannula or the NewLife Elite unit. Oxygen accelerates the combustion of flammable substances. This device supplies high-concentration oxygen that promotes rapid burning. Do not allow smoking or open flames within the same room of (1) this device, or (2) any oxygen-carrying accessory. Failure to observe this warning can result in severe fire, property damage and / or cause physical injury or death. 30

37 [Read the Important Safety Rules section before operating this equipment.] Dual Flow Application The NewLife Elite unit s dual flow option allows a single concentrator to meet the high flow requirements of one patient (Figure 9) or the needs of two patients, in any combination of flows up to the maximum capacity of the concentrator. Excellent for use in the home, extended care facility, hospital, or physician s waiting room. Two Patients/ Dual Flow Applications Pediatric/Low Flow Application Figure 9 The pediatric flowmeter (available for use with the dual flow NewLife Elite unit) meets low flow requirements up to 2 lpm in 1/8 liter (125 ccm) increments Figure 10). Figure 10 31

38 Setting the Pediatric Flowmeter [Read the Important Safety Rules section before operating this equipment.] When using the pediatric flowmeter, the unit will not reach proper concentration at the pediatric setting (less than 2 lpm) until you bleed off a portion of the oxygen by opening the primary flowmeter (on the left side of the unit). Follow the procedure below when using the pediatric flowmeter. 1. Follow the start-up instructions 1-7 as outlined on page 9 & Set the pediatric flowmeter to the prescribed flow. 3. Set the primary flowmeter to 2 lpm to bleed off excess product, and allow the unit to achieve maximum concentration. The NewLife Elite Oxygen Concentrator must be operated for at least five minutes at 2 lpm before using the unit. The NewLife Elite is appropriate for usage by two patients, provided the combined flow is a minimum of 2 lpm and does not exceed the maximum capacity of the concentrator. 32

39 Symbols/Abbreviations Symbols are frequently used on equipment and/or the manual in preference to words with the intention of decreasing the possibility of misunderstanding caused by language differences. Symbols can also permit easier comprehension of a concept within a restricted space. The following table is a list of symbols and definitions that may be used with the NewLife Elite Oxygen Concentrator. Symbol Description Symbol Description ON (power switch on) OFF (power switch off) No smoking Do not disassemble Type BF equipment Consult instructions for use Warning Describes a hazard or unsafe practice that if not avoided can result in severe bodily injury, death or property damage Caution Describes a hazard or unsafe practice that if not avoided can result in minor bodily injury or property damage Class II Device Complies with the 93/42/EEC directive drawn up by the approved organization No Note Provides information important enough to emphasize or repeat Safety agency for CAN/CSA C22.2 No M90 for medical electrical equipment 33

40 Symbol Description Symbol Description Safety agency for CAN/CSA C22.2 No M90 for medical electrical equipment. Oxygen concentration warning LED Keep unit and accessories dry Consult the accompanying documents Use no oil or grease Proper disposal of waste of electrical and electronic equipment required Oxygen outlet connection to the cannula Do not expose to open flames Fragile handle with care Keep in the vertical position Caution: Federal law (USA) restricts this for sale or rental by or on the order of a physician or licensed health care provider. 34

41 For European representative: Gavin Ayling 9 Bungham Lane Penkridge Stafford Staffordshire ST19 5NH England eurorepcontact@ airsep.com

42 For service on your NewLife Elite Oxygen Concentrator, please contact your local Equipment Provider at: Manufactured by: AirSep Corporation Buffalo, NY USA MN078-1_FÍ~

Companion 5 TM. Oxygen Concentrator. User Manual

Companion 5 TM. Oxygen Concentrator. User Manual Companion 5 TM Oxygen Concentrator User Manual User Controls & System Status Indicators Symbol Definition Symbol Definition Read user manual before operation. See user manual for instructions. I/O On/Off

More information

NewLife Family. User Manual (English)

NewLife Family. User Manual (English) NewLife Family User Manual (English) User Controls & System Status Indicators ISO 7000; Graphical symbols for use on equipment Index and synopsis Read user s manual before operation. Reg. # 1641 Operating

More information

Oxygen Concentrator Instruction

Oxygen Concentrator Instruction WARNING-Read instruction before operating this equipment Oxygen Concentrator Instruction K5BW WARNING Oxygen therapy can be hazardous in certain conditions. Seeking medical advice before using an oxygen

More information

U S E R M A N U A L. Model: PM4300 Series SAVE THESE INSTRUCTIONS

U S E R M A N U A L. Model: PM4300 Series SAVE THESE INSTRUCTIONS U S E R M A N U A L Model: PM4300 Series (PM4351 Shown) SAVE THESE INSTRUCTIONS CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090

More information

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website:

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website: USER MANUAL Model: PM4300 Series SAVE THESE INSTRUCTIONS For the most current manual revision, please visit our Website: www.precisionmedical.com 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA

More information

USER MANUAL SAVE THESE INSTRUCTIONS. 300 Held Drive Tel: (+001) Northampton, PA USA Fax: (+001) ISO Certified

USER MANUAL SAVE THESE INSTRUCTIONS. 300 Held Drive Tel: (+001) Northampton, PA USA Fax: (+001) ISO Certified USER MANUAL Model: PM4300 Series (PM4351 Shown) SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090 Northampton,

More information

Companion 5 TM. Oxygen Concentrator. User Manual

Companion 5 TM. Oxygen Concentrator. User Manual Companion 5 TM Oxygen Concentrator User Manual User Controls & System Status Indicators Symbol Definition Symbol Definition Read user manual before operation. See user manual for instructions. I/O On/Off

More information

Home Oxygen Therapy Patient Handbook

Home Oxygen Therapy Patient Handbook Home Oxygen Therapy Patient Handbook Office Hours 8:30 a.m. to 5:00 p.m. Monday-Friday Emergency Answering Service 24 hours a day, 7 days a week Birmingham Office: (205) 981-2333 5025 Highway 280 Suite

More information

Portable oxygen concentrator

Portable oxygen concentrator Table of contents G3 1. Introduction 2 2. Important Safety Rules 3 3. Unpack your Lovego 4 4. User Controls & System Status Indicators 5 5. Quick Start 6 6. How to operate 7 7. How to operate with battery

More information

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator Operating Instructions Table of Contents Resource v.2 Conserving Regulator Safety Information Device Precautions Introduction Product Features Product Specifications Feature Illustrations Set Up Usage

More information

Regalia Oxygen Concentrator

Regalia Oxygen Concentrator Regalia Oxygen Concentrator REGALIA INSTRUCTION MANUAL P/N 5167 Rev B January 2014 Table of Contents Warnings and Cautions... 3 Indications for Use... 4 Introduction... 4 Important Safety Instructions...

More information

MODEL Maverick Heavy Duty Aerosol Compressor Operator s Manual

MODEL Maverick Heavy Duty Aerosol Compressor Operator s Manual MODEL 50012 Maverick Heavy Duty Aerosol Compressor Operator s Manual This page left blank intentionally CONTENTS Page No. INSPECTION... 1 GENERAL INFORMATION... 1 SAFETY PRECAUTIONS... 2 BASIC OPERATING

More information

Oxygen concentrators. A user guide.

Oxygen concentrators. A user guide. Oxygen concentrators. A user guide. Service provided by: 02 Oxygen concentrators. A user guide. Oxygen concentrators. A user guide. 03 Contents. 3 The basics 4 Millennium Important: Please ensure you read

More information

INTEGRA E-Z INSTRUCTION MANUAL

INTEGRA E-Z INSTRUCTION MANUAL Oxygen Concentrator INTEGRA E-Z INSTRUCTION MANUAL Model 6323A-OM-10 Model 6323A-10 P/N 7040 Rev A July 2005 Table of Contents Warnings and Cautions... 3 Indications for Use... 4 Contraindications... 4

More information

Welcome to the Specialized Medical Services Respiratory training webinar series!

Welcome to the Specialized Medical Services Respiratory training webinar series! Welcome to the Specialized Medical Services Respiratory training webinar series! SMS is your Long Term Care (LTC) facility single source for oxygen, medical equipment, respiratory care services and supplies

More information

VisionAire. Oxygen Concentrator Patient Manual

VisionAire. Oxygen Concentrator Patient Manual Oxygen Concentrator Patient Manual VisionAire AirSep Corporation 401 Creekside Drive Buffalo, New York 14228-2085 USA Telephone: (716) 691-0202 24-Hour Fax (716) 691-4141 DO NOT OPERATE THIS EQUIPMENT

More information

Rev. B. Operating Instructions

Rev. B. Operating Instructions 780745-00 Rev. B Operating Instructions 780745-00 Rev. B TABLE OF CONTENTS PAGE Warnings, Cautions, and Notes..............................4-7 Definition of Symbols......................................

More information

Instructions for Cleaning & Maintaining Air Compressor

Instructions for Cleaning & Maintaining Air Compressor Instructions for Cleaning & Maintaining Air Compressor 1. The air compressor should be kept at least 6-8 inches away from the wall for ventilation and to keep the compressor from overheating. 2. The compressor

More information

Patient Manual. Oxygen Concentrator Oxygen Monitor, Air Outlet, Dual Flow, and Pediatric Flow Options

Patient Manual. Oxygen Concentrator Oxygen Monitor, Air Outlet, Dual Flow, and Pediatric Flow Options Patient Manual Oxygen Concentrator Includes: Oxygen Monitor, ir Outlet, Dual Flow, and Pediatric Flow Options DO NOT OPERTE THIS EQUIPMENT WITHOUT FIRST REDING ND UNDERSTNDING THIS MNUL. IF YOU RE UNBLE

More information

OC4000 USER'S GUIDE I O OXYGEN CONCENTRATOR FOR VETERINARY USE ONLY. by Supera Anesthesia Innovations GLOSSARY OF SYMBOLS. : ON (power switched on)

OC4000 USER'S GUIDE I O OXYGEN CONCENTRATOR FOR VETERINARY USE ONLY. by Supera Anesthesia Innovations GLOSSARY OF SYMBOLS. : ON (power switched on) by Supera Anesthesia Innovations USER'S GUIDE OC4000 OXYGEN CONCENTRATOR FOR VETERINARY USE ONLY I O : ON (power switched on) : OFF (power switched off) : Class II protection : Do not expose to open flames

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.9 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

HELiOSTM Plus & Marathon

HELiOSTM Plus & Marathon HELiOSTM Plus & Marathon User Manual (English) HELiOS TM Plus & Marathon Symbols Glossary ISO 7000; Graphical symbols for use on equipment Index and synopsis ADR: European Agreement concerning the International

More information

OC4000 USER'S GUIDE I O OXYGEN CONCENTRATORS OXYGEN CONCENTRATOR FOR VETERINARY USE ONLY GLOSSARY OF SYMBOLS. : ON (power switched on)

OC4000 USER'S GUIDE I O OXYGEN CONCENTRATORS OXYGEN CONCENTRATOR FOR VETERINARY USE ONLY GLOSSARY OF SYMBOLS. : ON (power switched on) OXYGEN CONCENTRATORS USER'S GUIDE OC4000 OXYGEN CONCENTRATOR FOR VETERINARY USE ONLY I O : ON (power switched on) : OFF (power switched off) : Class II protection : Do not expose to open flames GLOSSARY

More information

BROCKTON AREA MULTI-SERVICES, INC. MEDICAL PROCEDURE GUIDE. Date(s) Reviewed/Revised:

BROCKTON AREA MULTI-SERVICES, INC. MEDICAL PROCEDURE GUIDE. Date(s) Reviewed/Revised: Page 1 of 6 DEFINITION: Oxygen is a colorless, odorless, and tasteless gas that makes up about 20% of the air we breathe. Oxygen is essential to human, animal, and plant life. This training is intended

More information

INDUSTRIAL - SOLVENT PARTS CLEANER

INDUSTRIAL - SOLVENT PARTS CLEANER INDUSTRIAL - SOLVENT PARTS CLEANER OPERATION & INSTRUCTION MANUAL I. Introduction A. Manufacturer s Statement CM400 This manual will provide you with important information about the Solvent Parts Cleaners

More information

Pressure Regulator. Instructions for Use

Pressure Regulator. Instructions for Use Pressure Regulator Instructions for Use 702-0083.9 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient, the

More information

AUTOVENT 4000 VENTILATOR

AUTOVENT 4000 VENTILATOR OVERVIEW AUTOVENT 4000 Only properly trained and approved Escambia County Bureau of Public Safety Paramedics are to use the AutoVent 4000 ventilator manufactured by LSP to transport patients already on

More information

SeQual Technologies Inc.

SeQual Technologies Inc. SeQual Technologies Inc. ECLIPSE Model 1000 User Manual PN Rev 01/06 REV. CO No. DESCRIPTION REL. DATE PREP. BY APPR. BY X1 D017 Initial Design Release 11/16/01 SWB PJJ X2 D120 Update manual to remove

More information

Oxygen Concentrators Standard, with. and HomeFill

Oxygen Concentrators Standard, with. and HomeFill User Manual Platinum Series Oxygen Concentrators Standard, with SensO 2 and HomeFill Invacare 10 with SensO 2 Model IRC10LXO2 Invacare 10 Model IRC10LX Dealer: This manual MUST be given to the end user.

More information

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289 3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS Item #31289 The EASTWOOD 3 GALLON, OILLESS PANCAKE COMPRESSOR, with an Integral Air Regulator, efficiently supplies all compressed air requirements for

More information

ACRYLIC FLOW METERS 0-30, 0-70 & LPM

ACRYLIC FLOW METERS 0-30, 0-70 & LPM ACRYLIC FLOW METERS 0-30, 0-70 & 0-120 LPM ASSEMBLY INSTRUCTIONS & INSTRUCTIONS FOR USE R219P86 (0-120LPM) R219P87 (0-70LPM) R219P88 (0-30 LPM) R138P11 Rev. D TABLE OF CONTENTS: 1.0 PRODUCT OVERVIEW...1

More information

Hi Flow Stroller H I F L O W S T R O L L E R

Hi Flow Stroller H I F L O W S T R O L L E R Hi Flow PAT I E N T O P E R AT I N G I N S T R U C T I O N S Stroller H I F L O W S T R O L L E R Warning Information Read this page carefully before using your unit. The unit contains liquid oxygen which

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.12 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient,

More information

Liquid Oxygen- Helios & Helios Marathon A Guide for Patients in the Home

Liquid Oxygen- Helios & Helios Marathon A Guide for Patients in the Home Liquid Oxygen- Helios & Helios Marathon A Guide for Patients in the Home Contact Name: Phone number: Table of Contents Contents Receiving Liquid Oxygen in the Home... 2 Instructions for Liquid Oxygen-Helios

More information

PRODUCT INFORMATION AND INSTRUCTIONS OM-810

PRODUCT INFORMATION AND INSTRUCTIONS OM-810 PRODUCT INFORMATION AND INSTRUCTIONS OM-810 TABLE OF CONTENTS General Information 2 Important Safety Rules and Precautions 3-4 Introduction 5 Description of Parts and Controls 6-7 Assembly and Use Non-Portable

More information

Med Aire Alternating Pressure Pump and Pad System

Med Aire Alternating Pressure Pump and Pad System User Manual Med Aire Alternating Pressure Pump and Pad System 14002E 14001E Symbols & Statements NOTE Indicates some tips or some information users should be aware of. CAUTION Indicates correct operating

More information

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST OIL-LESS TYPE IMPORTANT: PLEASE READ CAREFULLY BEFORE STARTING OPERATIONS. THE CONTENTS ARE FOR GENERAL INFORMATION OF ALL THE SIMILAR MODELS. Record

More information

Children s Home Healthcare s. Oxygen Parent Handbook

Children s Home Healthcare s. Oxygen Parent Handbook Children s Home Healthcare s Information and Instructions For Parents Who Have A Child Who Uses Oxygen Your child s physician prescribed home oxygen therapy for. The current order states that you are to

More information

VENTI-O 2 plus. O 2 switching valve for: VENTImotion from serial number 4000 VENTIlogic from serial number 4000 VENTImotion 2

VENTI-O 2 plus. O 2 switching valve for: VENTImotion from serial number 4000 VENTIlogic from serial number 4000 VENTImotion 2 VENTI-O 2 plus O 2 switching valve for: VENTImotion from serial number 4000 VENTIlogic from serial number 4000 VENTImotion 2 Device description and operating instructions Overview VENTI-O 2 plus with therapy

More information

Dialflow Regulator. Instructions for Use

Dialflow Regulator. Instructions for Use Dialflow Regulator Instructions for Use 702-0030.11 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

In case of an emergency please call 000

In case of an emergency please call 000 Oxygen User Guide 1 The Drive DeVilbiss Story Dr. Allen DeVilbiss invented the first atomizer in the late 1880 s. In order to help his patients breathe more freely, Dr. DeVilbiss became an innovator, dedicated

More information

PRODUCT INFORMATION AND INSTRUCTIONS OM-411A OM-411AF

PRODUCT INFORMATION AND INSTRUCTIONS OM-411A OM-411AF PRODUCT INFORMATION AND INSTRUCTIONS OM-411A OM-411AF TABLE OF CONTENTS 1 General Information.............................................. 2 Important Safety Rules and Precautions..............................

More information

Home Oxygen Therapy in NHS Scotland. Level 1 Education for Patients & Carers

Home Oxygen Therapy in NHS Scotland. Level 1 Education for Patients & Carers Home Oxygen Therapy in NHS Scotland Level 1 Education for Patients & Carers National Home Oxygen Service Guidance/Best Practice Recommendations Changes to the National Oxygen Service in Scotland in 2012-2013

More information

Micro Dial-Flowmeter. Instructions for Use

Micro Dial-Flowmeter. Instructions for Use Micro Dial-Flowmeter Instructions for Use 702-0082.8 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

U S E R M A N U A L AIR FLOWMETER. MODELS: 1MFA2001 (shown) 1MFA9001 CAUTION. ISO Certified

U S E R M A N U A L AIR FLOWMETER. MODELS: 1MFA2001 (shown) 1MFA9001 CAUTION. ISO Certified U S E R M A N U A L AIR FLOWMETER MODELS: 1MFA2001 (shown) 1MFA9001 ISO 13485 Certified Authorized Your EU local Representative: distributor: EMERGO EUROPE, INC. Molenstraat 15 2513866 BH -624 The -3952

More information

Stroller. Sprint S T R O L L E R S P R I N T

Stroller. Sprint S T R O L L E R S P R I N T Stroller S T R O L L E R PAT I E N T O P E R AT I N G I N S T R U C T I O N S Sprint S P R I N T Warning Information Read this page carefully before using your unit. The unit contains liquid oxygen which

More information

Portable Oil Lube Air Compressors

Portable Oil Lube Air Compressors Portable Oil Lube Air Compressors 8003631 8003632 0410149 8018968 8018940 Owner s Manual Read and understand operating instructions before use Safety definitions The information listed below should be

More information

Pressure Regulators. Instructions for Use

Pressure Regulators. Instructions for Use Pressure Regulators Instructions for Use 702-0032.11 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

I. Subject: Compressed gas cylinders, regulators, and portable liquid systems.

I. Subject: Compressed gas cylinders, regulators, and portable liquid systems. I. Subject: Compressed gas cylinders, regulators, and portable liquid systems. II. Objective: To provide safe and effective use, storage, and transport of therapeutic gases. II. Procedure: A. Compressed

More information

VERTICAL AIR COMPRESSORS

VERTICAL AIR COMPRESSORS VERTICAL AIR COMPRESSORS MODEL NO: VE11C150, VE15C150, VE18C150 PART NO: 2226005, 2226000, 2226015 OPERATION & MAINTENANCE INSTRUCTIONS LS0615 INTRODUCTION Thank you for purchasing this CLARKE Vertical

More information

AIR COMPRESSOR. Failure to follow all instructions as listed below may result in electrical shock, fire, and/or serious personal injury.

AIR COMPRESSOR. Failure to follow all instructions as listed below may result in electrical shock, fire, and/or serious personal injury. 2 GALLON AIR COMPRESSOR Model: 7517 DO NOT RETURN TO STORE. Please CALL 800-348-5004 for parts and service. CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or grind materials

More information

Alternating Pressure / Low Air Loss Mattress System

Alternating Pressure / Low Air Loss Mattress System 750000 Alternating Pressure / Low Air Loss Mattress System User Manual Important: Do not operate the Mattress System without first reading and understanding this manual! Save this manual for future use.

More information

MODEL NUMBER: PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank

MODEL NUMBER: PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank IMPORTANT SAFETY INSTRUCTIONS CAUTION - To reduce risk of electrical shock or Electrocution: MODEL NUMBER: 20008 200 PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank IMPORTANT: It is

More information

INSTALLATION GUIDE Gas Rangetops

INSTALLATION GUIDE Gas Rangetops INSTALLATION GUIDE Gas Rangetops Contents Wolf Gas Rangetops........................... 3 Safety Instructions............................ 4 Gas Rangetop Specifications.................... 5 Gas Rangetop

More information

BlenderBuddy 2. Operating Manual & Instructions For Use ENGLISH R233M01 REV. A

BlenderBuddy 2. Operating Manual & Instructions For Use ENGLISH R233M01 REV. A BlenderBuddy 2 Operating Manual & Instructions For Use ENGLISH R233M01 REV. A Maxtec 2305 South 1070 West Salt Lake City, Utah 84119 USA TEL (800) 748.5355 FAX (801) 973.6090 www.maxtec.com E-mail: sales@maxtec.com

More information

MODEL NUMBER: M20005 AIR SOURCE KIT. 30% Duty Compressor on. 2.0 Gallon Air Tank SAVE THIS MANUAL FOR FUTURE REFERENCE

MODEL NUMBER: M20005 AIR SOURCE KIT. 30% Duty Compressor on. 2.0 Gallon Air Tank SAVE THIS MANUAL FOR FUTURE REFERENCE MODEL NUMBER: M20005 AIR SOURCE KIT 30% Duty Compressor on 2.0 Gallon Air Tank SAVE THIS MANUAL FOR FUTURE REFERENCE USER MANUAL IMPORTANT SAFETY INSTRUCTIONS CAUTION - To reduce risk of electrical shock

More information

AirSep Corporation Centrox PSA Oxygen Concentrator

AirSep Corporation Centrox PSA Oxygen Concentrator Centrox PSA Oxygen Concentrator Instruction Manual MN124-1 03/11 Rev. B AirSep Corporation 260 Creekside Drive Buffalo, NY 14228-2075 USA Phone: 1-800-320-0303 Fax: (716) 691-1255 www.airsep.com Ownership

More information

VERTICAL AIR COMPRESSORS

VERTICAL AIR COMPRESSORS VERTICAL AIR COMPRESSORS MODEL NO: VE15C150, VE18C150, VE25C150 PART NO: 2226010, 2226020, 2226025 OPERATION & MAINTENANCE INSTRUCTIONS LS0715 INTRODUCTION Thank you for purchasing this CLARKE Vertical

More information

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician. U S E R M A N U A L Model: 19MFA1001 Series SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090 Northampton,

More information

testo Leakage detector for refrigerants Instruction manual

testo Leakage detector for refrigerants Instruction manual testo 316-3 Leakage detector for refrigerants Instruction manual 2 1 Contents 1 Contents 1 Contents... 3 2 Safety and the environment... 4 2.1. About this document... 4 2.2. Ensure safety... 4 2.3. Protecting

More information

AND OPERATION INSTRUCTIONS

AND OPERATION INSTRUCTIONS ADI 2111-D Certified ISO 9001 MEDICAL FLOWMETER (Back Pressure Compensated Thorpe Tube) INSTALLATION AND OPERATION INSTRUCTIONS Before Installing or Operating, Read and Comply with These Instructions Controls

More information

Oxygen Concentrators without SensO 2. , Oxygen Concentrators with SensO 2 HomeFill Compatible

Oxygen Concentrators without SensO 2. , Oxygen Concentrators with SensO 2 HomeFill Compatible User Manual Perfecto 2 Series Oxygen Concentrators without SensO 2, Oxygen Concentrators with SensO 2 HomeFill Compatible Perfecto 2 with SensO 2 Model IRC5PO2 Model IRC5PO2W Model IRC5PO2V Perfecto 2

More information

VisionAire TM Family. User Manual (English)

VisionAire TM Family. User Manual (English) VisionAire TM Family User Manual (English) User Controls & System Status Indicators ISO 7000; Graphical symbols for use on equipment Index and synopsis Read user s manual before operation. Reg. # 1641

More information

Oxygen Therapy Customer Handbook

Oxygen Therapy Customer Handbook Oxygen Therapy Customer Handbook Sanford Health HealthCare Accessories is committed to providing quality products and services to promote health and healing. Through ongoing customer education, physical

More information

EvolutionTM OM-900 PRODUCT INFORMATION AND INSTRUCTIONS OXYGEN CONSERVER

EvolutionTM OM-900 PRODUCT INFORMATION AND INSTRUCTIONS OXYGEN CONSERVER EvolutionTM OXYGEN CONSERVER PRODUCT INFORMATION AND INSTRUCTIONS OM-900 TABLE OF CONTENTS General Information....3 Important Safety Rules and Precautions Operating the Unit... 4-6 Accessories/Repairs....6

More information

DFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER)

DFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER) DFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER) 0-3, 0-15, 0-30 & 0-70 LPM ASSEMBLY INSTRUCTIONS & INSTRUCTIONS FOR USE R219P87-400 (0-70 LPM) R219P88-400 (0-30 LPM) R219P79-400 (0-15 LPM) R219P99-400

More information

oxyoasis oxygen facial system

oxyoasis oxygen facial system owners manual oxyoasis oxygen facial system #5415 OXYOASIS comfortel.com.au IMPORTANT PLEASE READ THESE SAFETY INSTRUCTIONS GENERAL SAFETY 1. Read carefully and keep these instructions in a safe place.

More information

24L OIL FREE AIR COMPRESSOR MODEL NO: TIGER 7/250 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS10/13

24L OIL FREE AIR COMPRESSOR MODEL NO: TIGER 7/250 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS10/13 24L OIL FREE AIR COMPRESSOR MODEL NO: TIGER 7/250 PART NO: 2244030 OPERATION & MAINTENANCE INSTRUCTIONS LS10/13 INTRODUCTION Thank you for purchasing this product. Before attempting to use this product,

More information

OWNER S MANUAL. Page: 1 of 8

OWNER S MANUAL. Page: 1 of 8 Air Needle Scaler OWNER S MANUAL WARNING: Read carefully and understand all INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety precautions may result in serious personal

More information

Complex Care Hub Manual: Oxygen Therapy

Complex Care Hub Manual: Oxygen Therapy Complex Care Hub Manual: Oxygen Therapy Table of Contents RCH@Home Manual: Oxygen Therapy... 1 1. What is oxygen therapy?... 2 2. Symptoms of low oxygen... 2 3. Giving oxygen at home and in the community...

More information

Before using or working with the Grab n Go Heliox system, please read and understand the usage and safety instructions provided in this booklet.

Before using or working with the Grab n Go Heliox system, please read and understand the usage and safety instructions provided in this booklet. Operating Instructions for the Before using or working with the Grab n Go Heliox system, please read and understand the usage and safety instructions provided in this booklet. Introduction The Praxair

More information

Instrument Description and Operating Instructions OXYMAT 2. Its own oxygen source

Instrument Description and Operating Instructions OXYMAT 2. Its own oxygen source Instrument Description and Operating Instructions OXYMAT 2 Its own oxygen source OXYMAT 2 Fig. 1 5 6 1 15 2 3 4 8 15 7 9 12 10 13 11 14 Legends 1 OXYMAT 2 oxygen concentrator 2 Oxygen outlet 3 Main switch

More information

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. U S E R M A N U A L 1600 SERIES OXYGEN REGULATOR 168715G (Shown) SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090

More information

Operation and Maintenance of the EPV200 Portable Ventilator

Operation and Maintenance of the EPV200 Portable Ventilator Operation and Maintenance of the EPV200 Portable Ventilator 1 Applications of the EPV200 The EPV200 Portable Ventilator is a gas powered electronically controlled mechanical ventilator, designed to provide

More information

HAS2400 Hepa Air Scrubber

HAS2400 Hepa Air Scrubber HAS2400 Hepa Air Scrubber Diamonds Products 15 SW 40th Ave Great Bend, KS 67530 USA Toll Free: (866) 539-1694 Local: (620) 792-4542 Purchase Date: Serial Number: Purchased From: The DIAMOND HAS2400 HEPA

More information

Instruction Manual for Configura Cushionair Portable Pump

Instruction Manual for Configura Cushionair Portable Pump Instruction Manual for Configura Cushionair Portable Pump Fitted with battery powered pump, suitable for Configura Portable chairs V E R S I O N O N E M A Y 2 0 1 6 Contents Introduction 3 Set up of Cushionair

More information

SC505 Sof Care Inflator. Operating Instructions and Service Manual

SC505 Sof Care Inflator. Operating Instructions and Service Manual SC505 Sof Care Inflator Gaymar Industries, Inc. 10 Centre Drive Orchard Park, NY 14127 Toll Free +1 800.828.7341 + 1 716.662.2551 Fax +1 800.993.7890 Outside USA +1 716.662.8636 Outside USA Fax +1 716.662.0730

More information

2 GAL AIR COMPRESSOR OPERATOR S MANUAL 1/3HP/2-GALLON IMPORTANT: READ THIS OPERATOR S MANUAL BEFORE USING

2 GAL AIR COMPRESSOR OPERATOR S MANUAL 1/3HP/2-GALLON IMPORTANT: READ THIS OPERATOR S MANUAL BEFORE USING ITEM# AT01101 SKU# 207-1525 OPERATOR S MANUAL 1/3HP/2-GALLON 2 GAL AIR COMPRESSOR IMPORTANT: READ THIS OPERATOR S MANUAL BEFORE USING Toll Free Helpline: 1-888-899-0146 Versions 01 TABLE OF CONTENTS SAFETY

More information

EasyFit Lite Nasal Mask, Silicone. Nasal Mask. EasyFit Lite Nasal Mask. Device Description and Instructions for Use. Caution:

EasyFit Lite Nasal Mask, Silicone. Nasal Mask. EasyFit Lite Nasal Mask. Device Description and Instructions for Use. Caution: EasyFit Lite Nasal Mask EasyFit Lite Nasal Mask, Silicone Nasal Mask Device Description and Instructions for Use Caution: Federal Law (U.S.A.) restricts this device to sale by or on the order of a physician.

More information

200 PSI FAST-FILL AIR SOURCE KIT

200 PSI FAST-FILL AIR SOURCE KIT 200 PSI FAST-FILL AIR SOURCE KIT 55% Duty Compressor on 2.0 Gallon Air Tank PART NO. 20007 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of

More information

AirSep Corporation Reliant PSA Oxygen Concentrator. Instruction Manual

AirSep Corporation Reliant PSA Oxygen Concentrator. Instruction Manual AirSep Corporation Reliant PSA Oxygen Concentrator Instruction Manual MN021-1 AirSep Corporation 260 Creekside Drive Buffalo, NY 14228-2075 USA Phone: (716) 691-0202 Fax: (716) 691-1255 Ownership Data

More information

User Manual. VisionAire 2 & 3

User Manual. VisionAire 2 & 3 User Manual VisionAire 2 & 3 User Controls & System Status Indicators Symbol Definition Symbol Definition Read user manual before operation. See user manual for instructions. ON (power switch on) No Smoking

More information

SILENTAIRE TECHNOLOGY

SILENTAIRE TECHNOLOGY SILENTAIRE TECHNOLOGY 8614 Veterans Memorial. Houston, Texas 77088 832/327-7452 800/972-7668 Fax: 832/327-0668 E-mail: silentaire@silentaire.com Thank you and congratulations on the purchase of your OILLESS

More information

HOT SHOT OPERATING INSTRUCTIONS AND OWNER S MANUAL. Model SPC-500. (Propane Torch)

HOT SHOT OPERATING INSTRUCTIONS AND OWNER S MANUAL. Model SPC-500. (Propane Torch) OPERATING INSTRUCTIONS AND OWNER S MANUAL Model # SPC-500 READ INSTRUCTIONS CAREFULLY: Read and follow all instructions. Place instructions in a safe place for future reference. Do not allow anyone who

More information

2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL

2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL 2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL Model: 52024 CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or grind materials such as wood, paint, metal, concrete, cement,

More information

AIR/OXYGEN BLENDER INSTRUCTION MANUAL

AIR/OXYGEN BLENDER INSTRUCTION MANUAL BIO-MED DEVICES AIR/OXYGEN BLENDER INSTRUCTION MANUAL CATALOG #2120 REV 042203 BIO-MED DEVICES INC. 1445 BOSTON POST ROAD, GUILFORD, CT 06437 (203) 458-0202 FAX (203) 458-0440 www.biomeddevices.com TABLE

More information

Oxygen Concentrator 7F-5A User s manual

Oxygen Concentrator 7F-5A User s manual TABLE OF CONTENTS SAFETY NOTES.2-5 Oxygen Concentrator 7F-5A User s manual FEATURES..6-7 HANDLING 8 OPERATING & INSTALLATION 9-14 MAINTENANCE 15-20 DISASSEMBLY..21-23 OTHER ATTENTION ITEMS 24-25 DO NOT

More information

Operation and Maintenance of the MCV100 and MCV100-B Portable Ventilator

Operation and Maintenance of the MCV100 and MCV100-B Portable Ventilator Operation and Maintenance of the MCV100 and MCV100-B Portable Ventilator 1 Applications of the MCV100(B) The MCV100 and MCV100-B Mass Casualty Ventilators are electronically controlled portable ventilators,

More information

Oxygen Concentrator. User Manual

Oxygen Concentrator. User Manual Oxygen Concentrator User Manual Model Numbers JB0160-010B, JB0160-010, JB0160-010B-220, JB0160-010-220 and JB0160-015B, JB0160-015, JB0160-015B-220, JB0160-015-220 with O 2 Sensor Important: Do not operate

More information

PremAire INSTRUCTIONS FOR VORTEX TUBE MODE OF OPERATION

PremAire INSTRUCTIONS FOR VORTEX TUBE MODE OF OPERATION DUAL SUPPLY REGULATOR ESCAPE CYLINDER VORTEX MAIN INLET PremAire INSTRUCTIONS FOR VORTEX TUBE MODE OF OPERATION WARNING THIS MANUAL MUST BE READ CAREFULLY BY ALL PERSONS WHO HAVE OR WILL HAVE THE RESPONSIBILITY

More information

DUAL ACTION SANDER MODEL NO: CAT121 OPERATION & MAINTENANCE INSTRUCTIONS PART NO: LS0211

DUAL ACTION SANDER MODEL NO: CAT121 OPERATION & MAINTENANCE INSTRUCTIONS PART NO: LS0211 DUAL ACTION SANDER MODEL NO: CAT121 PART NO: 3110877 OPERATION & MAINTENANCE INSTRUCTIONS LS0211 INTRODUCTION Thank you for purchasing this CLARKE product. Before attempting to use this product, please

More information

200 PSI COMPRESSORS - MODEL NUMBERS

200 PSI COMPRESSORS - MODEL NUMBERS 200 PSI COMPRESSORS - MODEL NUMBERS 380C AIR COMPRESSOR KIT PART NO. 38033 480C AIR COMPRESSOR KIT PART NO. 48043 380C 480C IMPORTANT: It is essential that you and any other operator of this product read

More information

200 PSI HIGH-FLOW AIR SOURCE KIT

200 PSI HIGH-FLOW AIR SOURCE KIT 200 PSI HIGH-FLOW AIR SOURCE KIT 50% Duty Compressor on 2.0 Gallon Air Tank PART NO. 20008 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of

More information

100C Air Compressor Kit

100C Air Compressor Kit 10010 100C Air Compressor (standard mounting bracket, CE Spec) 10014 100C Air Compressor (no leader hose or check valve, CE Spec) 10016 100C Air Compressor (with Omega Bracket, CE Spec) IMPORTANT: It is

More information

User s Guide. VetPro 1000 Mobile Dental Delivery System. For Models:

User s Guide. VetPro 1000 Mobile Dental Delivery System. For Models: VetPro 1000 Mobile Dental Delivery System For Models: 8000-001 8000-002 8000-003 8000-004 8000-005 8000-006 8000-007 8000-008 User s Guide 003-1879-00 Rev L Product Information (The information below is

More information

TABLE OF CONTENTS. AirSep FreeStyle Portable Oxygen Concentrator. AirSep s FreeStyle Portable. Why Your Physician Prescribed Oxygen

TABLE OF CONTENTS. AirSep FreeStyle Portable Oxygen Concentrator. AirSep s FreeStyle Portable. Why Your Physician Prescribed Oxygen TABLE OF CONTENTS AirSep FreeStyle Portable Oxygen Concentrator AirSep s FreeStyle Portable Oxygen Concentrator Symbols Why Your Physician Prescribed Oxygen What is the FreeStyle Portable Oxygen Concentrator?

More information

AMP Oil Free Manual AMP 50-8-TC AMP 50-6-D AMP General User and Maintenance Instructions

AMP Oil Free Manual AMP 50-8-TC AMP 50-6-D AMP General User and Maintenance Instructions AMP Oil Free Manual AMP 50-8-TC AMP 50-6-D AMP 50-24 General User and Maintenance Instructions Silentaire Technology 8614 Veterans Memorial Dr. Houston, TX 77088 800-972-7668 Fax 832-327-0669 www.silentaire.com

More information

May 2016 Med Tips. Helpful Hints and FAQs. Replacing AirSep VisionAire Mixing Tank

May 2016 Med Tips. Helpful Hints and FAQs. Replacing AirSep VisionAire Mixing Tank May 2016 Med Tips Helpful Hints and FAQs Replacing AirSep VisionAire Mixing Tank CAIRE offers a kit designed to allow the replacement of any mixing tank originally installed in the AirSep VisionAire 5

More information

Model 1700 Patient Operating Instructions. Model 1700

Model 1700 Patient Operating Instructions. Model 1700 1700-00C.QXD 10/21/99 8:38 AM Page 1 The Respironics Tranquility Bilevel is a medical device. It has been prescribed by a physician to assist breathing. Compact Compact Bilevel Bilevel System System Model

More information

12 Volt Inflator. Operator s Manual. Model: ICC12F, ICC12FC, ICC12FCN, ICC12FCM, ICC12N

12 Volt Inflator. Operator s Manual. Model: ICC12F, ICC12FC, ICC12FCN, ICC12FCM, ICC12N Operator s Manual 12 Volt Inflator Model: ICC12F, ICC12FC, ICC12FCN, ICC12FCM, ICC12N 873 E. Citation Court Boise, Idaho 83716 USA 800-874-5771 www.bon-aireindustries.com Safety Guidelines Assembly Operation

More information