INSTRUCTION MANUAL PARTICLE SENSOR KS-40AF Higashimotomachi, Kokubunji, Tokyo , Japan
|
|
- Ella Fleming
- 5 years ago
- Views:
Transcription
1 INSTRUCTION MANUAL PARTICLE SENSOR KS-40AF Higashimotomachi, Kokubunji, Tokyo , Japan
2
3 Organization of this manual This manual describes the features and operation principles of the light-scattering type liquid-borne particle sensor KS-40AF. In sections related to the configuration of a measurement system comprising the KS-40AF and other equipment, this manual also touches upon operation of the other equipment, but to assure safe and correct operation, always consult the documentation of the other equipment as well. The following pages contain important information on safety. Be sure to read this part. This manual contains the following sections. Outline Provides an outline of the unit and descriptions of measurement system configuration examples. Controls and Features Briefly identifies and explains all parts of the unit. Preparations Describes how to make sampling tube, purge gas, and sample fluid connections and how to set the unit up for use. Measurement Describes the basic steps for measurement, and gives information on error messages. Maintenance Describes steps to perform after measurement, including how to clean the piping and countermeasures to take when the CELL indicator or LIQUID LEAK indicator or LASER indicator is lit in red, or when the LASER indicator flashes in red. Specifi cations Lists the technical specifications of the unit. * All company names and product names mentioned in this manual are trademarks or registered trademarks of their respective owners. i
4 ii
5 FOR SAFETY In this manual, important safety instructions are specially marked as shown below. To prevent the risk of death or injury to persons and severe damage to the unit or peripheral equipment, make sure that all instructions are fully understood and observed. WARNING Disregarding instructions printed here incurs the risk of death or severe injury to persons. Caution Disregarding instructions printed here incurs the risk of injury to persons and/or damage to peripheral equipment. Important Disregarding instructions printed here incurs the risk of damage to the product. Note Mentioned about the tips to use this unit properly. (This messages do not have to do with safety.) iii
6 iv
7 Caution Types of sample fluids The unit is designed for measuring liquid-borne particles. Never use the unit with the liquid types listed below, to prevent false measurement and the possibility of accidents and damage to the unit. - Explosive liquid or mixtures containing explosive fluid - Fluid which can cause corrosion of fluid-contacting parts or mixtures containing such fluid - Liquid with a temperature of less than +15 C or more than +35 C - Liquid with a pressure of more than 300 kpa (gauge pressure) When using different types of sample fluid where mixing can lead to reactions such as thermal build-up, hardening, or particle generation, always perform thorough cleaning to ensure complete replacement. v
8 WARNING Sample fluid danger prevention When toxic sample fluid comes into contact with any part of the human body or when toxic gases generated by the sample fluid are breathed, there is a danger of severe injury or death. Sample fluid also involves the risk of fire, explosion, heat generation, corrosion, deformation and other effects. Observe the following points closely to ensure safe use of sample fluid. Do not pass sample fluids which can cause corrosion of fluid-contacting parts (PFA, sapphire) through the system. Observe the following points when setting up or dismantling a measurement system. - Check the sample fluid for any risks due to toxicity or other harmful properties. - Make sure that sample fluid cannot leak from the system and come into contact with body parts, clothing etc. - When attaching or detaching tubes, connectors etc., always use appropriate protective tools and wear dual-layer gloves. - Double-check all connections before starting the sample fluid flow. - This unit incorporates a leak sensor or other safety device, but to verify safety you should always perform a thorough leak check before starting operation. To ensure that the leak sensor operates properly, the unit must be placed on a horizontal surface (inclination within 2 degrees; see page 2). The pressure of sample fluid passed through the unit may not exceed 300 kpa (gauge pressure). Ensure that the supply system (sampler, sample fluid lines etc.) has sufficient pressure resistance. When using sample fluid which may produce harmful gases, provide adequate ventilation as prescribed by applicable laws and regulations. vi
9 Before and after measurement, clean the entire sample fluid system thoroughly to remove any remnants. Otherwise unwanted effects such as thermal build-up, hardening, particle accumulation etc. can occur. This is especially important to prevent the possibility of serious accidents when using the unit to measure different types of sample fluids. Before starting to prepare the system for storage, complete the cleaning process with pure water and verify neutrality (ph7). Dispose of waste sample fluid only as prescribed by applicable laws and regulations. Never mix waste sample fluid with other substances unless absolute safety has been established. vii
10 viii
11 WARNING The unit uses a laser source This unit is designated as a class 1 laser product according to IEC (2001). The internal particle detector mechanism uses a laser, but it is fully shielded so that the beam cannot exit from the enclosure. The identification label shown at right is affixed to the top of the unit. Laser Precautions The particle detector mechanism in this unit uses a laser classified as a class 3B laser product according to IEC (2001). The laser source can cause blindness if viewed directly, and can cause skin injuries if the skin is exposed directly to the beam. Never open the cover, because this involves the risk of exposure to the laser beam. The warning label shown at right is affixed to the top of the unit. Caution Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure. ix
12 Warning Labels The following types of warning labels are used with this unit. Transport label Instructs the user to purge the sample fluid system with pure water and fill it with alcohol before transporting or shipping the unit for maintenance or servicing. The label is affixed to the rear of the unit. Important Before returning the unit for maintenance or servicing, purge the sample fluid system thoroughly with pure water to remove any sample fluid remnants. Then fill the piping with alcohol (isopropyl alcohol, ethyl alcohol or methyl alcohol). Sample fluid pressure warning label Warns that sample fluid pressure may not exceed 300 kpa (gauge pressure). INLET and OUTLET warning label Warns against subjecting the INLET and OUTLET to strong force or shocks. The label is affixed to the rear of the unit. WARNING Maximum allowable sample fluid pressure is 300 kpa (42 PSI) at gage. Attach and remove tubes carefully and do not exert strong force on the INLET and OUTLET to prevent deformation and fluid leaks. x
13 Installation warning label Instructs the user that the unit must be installed horizontally (inclination angle within 2 degrees), to ensure that the leak sensor operates properly. The label is affixed to the top of the unit. CAUTION This side must face up. The unit must be installed horizontally (maximum inclination 2 degrees). Laser product warning label This unit is designated as a class 1 laser product according to IEC (2001). The internal particle detector mechanism uses a laser, but it is fully shielded so that the beam cannot exit from the enclosure. The label is affixed to the top of the unit. Laser product description label The particle detector mechanism in this unit uses a laser classified as a class 3B laser product according to IEC (2001). The laser source can cause blindness if viewed directly, and can cause skin injuries if the skin is exposed directly to the beam. Never open the cover, because this involves the risk of exposure to the laser beam. The label is affixed to the top of the unit. xi
14 Precautions Operate the unit only as described in this manual. Do not use or store the unit in locations which - may exceed the temperature range specified on page 24, or may be subject to moisture condensation. If the temperature of the sample fluid is lower than the ambient temperature, condensation may occur. Take suitable countermeasures such as using purge gas. - may be subject to temperatures above +50 C or below -10 C, or to relative humidity of more than 90%, or where condensation or freezing inside the piping may occur. - may be subject to splashes of water or to direct sunlight. - may be subject to air with high salt or sulphur content or to gases. - do not provide a level support or may subject the unit and piping to vibrations or shock. Avoid sudden temperature changes which may cause moisture condensation on the optical assembly. Before using the unit, make sure that all cord, cable and tube connections are correctly and safely established. Do not subject the cords, cables and tubes to excessive bending or other force. When disconnecting the cords, cables and tubes, disconnect while holding onto the end of the plug, connector or tube. Excessive electrical noise or external noise can cause the unit to operate abnormally or malfunction. Do not disassemble the unit or attempt internal alterations. If a malfunction should happen to occur, do not try to solve the problem on your own, but rather note the condition of the unit clearly and contact the supplier. The end plugs are required to seal the unit when returning it for maintenance or servicing. Store the plugs in a safe location for later use. xii
15 The sample fluid system of the unit must always be filled with sample fluid (during measurement, storage, and transport). If the sample fluid system dries up, particle detector problems (see page 5) can occur. When the unit is not installed in a clean room environment, purge gas must be used to prevent contamination. Before returning the unit for maintenance or servicing, thoroughly purge the piping with pure water. Then fill the piping with ethyl alcohol and install the end plugs. In addition, use the special packaging box so as to prevent the unit from being subjected to vibrations and shock. xiii
16 xiv
17 Contents FOR SAFETY... iii Outline...1 Light Source Service Life...1 Particle Size Detection...1 Leak detection...2 Measurement System Configuration Examples...2 Controls and Features...4 Front Panel...4 Rear Panel...6 Preparations...7 Attaching the Particle Size Labels...7 Connection to Controller KE Connecting the Sampling Tubes...9 Removing the Sampling Tubes...10 Purge Gas...11 Before Supplying the Sample Fluid...12 Sample Fluid Flow Rate Adjustment...14 Turning the System On...15 Measurement...16 Measurement...16 Error Message...16 Maintenance...18 Steps After Use...18 When CELL indicator is lit in red...19 When LIQUID LEAK indicator is lit in red...19 When LASER indicator is lit in red or flashes in red...19 Cleaning the Piping...20 Periodic maintenance...21 Transport...21 Specifications...22 xv
18 xvi
19 Outline The KS-40AF is a sensor which uses the light-scattering method for measuring the particle number concentration in liquid. The particle count is determined for various sizes. Sample fluid contacting parts are made of sapphire and PFA, allowing direct measurement of hydrofluoric acid. By connecting the KS-40AF to the controller KE-40, a liquid-borne particle counter system with five size ranges can be created. Particles are measured in five size ranges ( 0.1 µm, 0.15 µm, 0.2 µm, 0.3 µm, 0.5 µm), and the rated sample fluid flow is 10 ml per minute. The KS-40AF does not have measurement controls or a display for measurement results. It is designed to be used under control of a separate controller which also supplies power to the KS-40AF. As the KS-40AF does not incorporate a flow control circuit for the sample fluid, the flow rate of the sample fluid must be controlled by external means. Light Source Service Life The life expectancy of a laser diode such as used in the KS-40AF is shorter than that of other semiconductors. When the laser diode nears the end of its service life, the optical power falls below the rated level. There are considerable differences in service life between individual diodes. Some may fail already after several thousand operation hours. The KS-40AF detects when the laser diode no longer provides the required output. In such a case, the LASER indicator flashes in red, which indicates that the laser diode must be replaced. Particle Size Detection The unit is calibrated at the factory, using particles of known diameter and refractive index (monodisperse polystyrene latex spheres with a refractive index of 1.6) in pure water. As the amount of scattered light by such particles is used as reference to determine the particle size, the measured size corresponds to the light-scattering size of polystyrene latex particles. 1
20 PARTICLE MONITOR LIQUID LEAK CELL LASER POWER PARTICLE SENSOR A PARTICLE SENSOR POWER LASER CELL Outline Leak detection The unit incorporates a leak sensor. If a leak is detected, the LIQUID LEAK indicator lights up in red and the LIQUID LEAK ALARM terminals become OPEN. To ensure that the leak sensor operates properly, the unit must be placed on a horizontal surface (inclination within 2 degrees). For proper activation of the leak sensor, the leak sample fluid must be held at the bottom of the unit. PARTICLE SENSOR KS-40 F KS-40AF LIQUID LEAK PARTICLE MONITOR (inclination within 2 degrees) OK Incorrect Incorrect PARTICLE MONITOR LIQUID LEAK CELL LASER POWER KS-40AF Caution If a leak is detected, stop the sample fl uid supply to the unit. Remove any sample fl uid that has accumulated in the unit. Then contact your supplier. Measurement System Configuration Examples With the KS-40AF, measurement control and processing of measurement results are entirely performed by the connected controller KE-40 and a printer KP-06 or computer connected to the KE-40. Two basic measurement configurations are shown page 3. Configure your actual system according to the measurement requirements and types of sample fluid. 2
21 Outline Important The service life of the light source (laser diode) is limited. While no measurement is being carried out, the light source should be turned off as much as possible. After turning on the power, wait at least 10 minutes before starting the measurement. This allows the light source output to stabilize. When using a flow controller to control sample flow Sample fluid line Pressuring sampler or similar Particle sensor Flow controller KS-40AF INLET OUTLET INLET OUTLET CONTROLLER PFAtube(2 4dia.) Connection cable PFA tube (2 4 dia.) SENSOR Sample fluid waste bottle Controller KE-40 SERIAL Printer KP-06 or computer When using the syringe sampler KZ-30W1 to control sample flow PFA tube (2 4 dia.) Syringe sampler Particle sensor KZ-30W1 KS-40AF INLET OUTLET INLET OUTLET CONTROLLER Connection cable PFAtube(2 4dia.) Sample fluid bottle Sample fluid SENSOR Controller KE-40 SERIAL KZ-30W1 Sample fluid waste bottle Printer KP-06 or computer 3
22 Controls and Features Front Panel LIQUID LEAK indicator PARTICLE MONITOR indicator PARTICLE MONITOR LIQUID LEAK CELL LASER POWER PARTICLE SENSOR KS-40AF POWER indicator LASER indicator CELL indicator LIQUID LEAK indicator Lit (green): Unit operates normally. Lit (red): A leak has occurred within the unit. PARTICLE MONITOR indicator Briefly flashes green when particles above the minimum particle size are detected. POWER indicator Lit while power to the unit is on. 4
23 Controls and Features LASER indicator Lit (green): Light source (laser diode) operates normally. Lit (red): Light source temperature is out of range. Flashing (red): Light source output is below normal. Off: Light source is off. CELL indicator Lit (green): Lit (red): Off: Particle detector flow cell operates normally. Particle detector flow cell cannot carry out normal measurement due to contamination, condensation, or similar causes. The indicator also lights up in red when particle number concentration exceeds the measurement limit. Light source is off. 5
24 Controls and Features Rear Panel INLET connector PURGE inlet EXT SELECT switch OUTLET connector EXT connector PURGE WARNING Maximum allowable sample fluid pressure is 300 kpa at gage. Attach and remove tubes carefully and do not exert strong force on the INLET and OUTLET to prevent deformation and fluid leaks. EXT SELECT EXT INLET OUTLET CONTROLLER LIQUID LEAK ALARM CONTROLLER connector LIQUID LEAK ALARM terminals INLET connector Sample fluid inlet PURGE inlet Serves for routing purge gas through the unit (see page 11). OUTLET connector Sample fluid outlet EXT connector, EXT SELECT switch Test output terminal and output selector switch LIQUID LEAK ALARM terminals Serves to drive an external leak alarm. Normally the terminals are closed (short-circuited). When a leak is detected, the terminals become open. The M3 terminal block accepts either electric wire with a 1.25 mm 2 cross section or ring tongue terminals. Maximum load: 30 V DC, 1 A (AC not allowed) CONTROLLER connector Serves for connection of the controller KE-40. 6
25 Preparations Attaching the Particle Size Labels Use the supplied particle size labels to mark settings at the controller KE-40 and the printer KP-06. Attach the labels so that the figures match the CH1 to CH6 indication. Note For information on affixing the particle size labels, please refer to the documentation of the respective equipment. The particle size labels serve only as a reminder and have no influence on operation of the unit. Affixing the labels is not mandatory. The following four label types are supplied Attach to KP Not used Attach to KE Not used 7
26 Preparations Connection to Controller KE Set the POWER switch of the KE-40 to OFF. Important When other components are also used in the system, set the power switch of these components also to OFF. 2. Link the CONTROLLER connector on the KS-40AF with the SEN- SOR connector on the KE-40, using the supplied connection cable. KS-40AF rear panel CONTROLLER connector Controller KE-40 rear panel SENSOR connector Connection to controller 8
27 Preparations Connecting the Sampling Tubes 1. Remove the end plugs from the INLET and OUTLET connectors of the KS-40AF by turning the plugs counterclockwise. 2. Loosen the nut of the sampling tube by turning it counterclockwise, and remove the plug from the sampling tube. Important Store the end plugs and plugs together in a safe place. The end plugs will be needed when the unit is sent to the supplier for servicing and maintenance. The plugs will be needed again when removing the sampling tube for storage. Plug Nut Sampling tube 3. Slide the tip of the sampling tube onto the INLET connector and turn the nut clockwise to fasten the tube. Connect the OUTLET side in the same way. INLET/OUTLET End plug Sampling tube Nut 9
28 Preparations Removing the Sampling Tubes 1. Purge the unit with pure water to remove any sample fluid remnants. If the use of pure water poses the risk of sample fluid reaction and thermal build-up, hardening, or particle accumulation, use a different cleaning agent that is suitable for the sample fluid, and finally purge with clean water. 2. Turn the nut on the INLET connector counterclockwise and pull the sampling tube off. Disconnect the OUTLET side in the same way. INLET/OUTLET Sampling tube Nut End plug 3. Place the end plugs on the INLET and OUTLET connectors and secure the plugs by turning them clockwise (see above illustration). 4. Insert the plug of the sampling tube into the flared section at the tip of the sampling tube and turn the nut clockwise to secure it. Important The fl ared section will contract slightly over time. Always store the sampling tube with the plug in place. Plug Flared section Nut Sampling tube 10
29 Preparations Purge Gas In the cases listed below, the interior of the unit should be cleaned with purge gas supplied to the PURGE inlet, to prevent adverse effects on the optical system and electrical circuitry. - If the temperature of the sample fluid is lower than the ambient temperature, so that moisture condensation may occur in the particle detector. - If there is the possibility that corrosive gases in the vicinity may affect the unit. - If the supplied sample fluid has the possibility that corrosive gases may leak from internal PFA tubes. - To cool the interior of the unit. - In environments with high particle number concentration. Important When the unit is not used in a clean room environment, airborne contamination may adversely affect the measurement. To prevent this, cleaning with purge gas during installation is required. Purge gas requirements are as follows. - Dry clean air or nitrogen gas - Temperature 15 to 35 C - Flow rate 3 to 10 L / min The PURGE inlet is an RC 1/8 type (PT 1/8 female). 11
30 Preparations Before Supplying the Sample Fluid Since the KS-40AF does not have a flow rate adjustment facility, additional equipment is required to adjust the sample fluid flow. Refer to Measurement System Configuration Examples on page 2 and configure a system which can supply sample fluid at a rate of 10 ml per minute. WARNING Do not exceed the sample fl uid pressure rating of the unit. The flow cell and INLET and OUTLET of the KS-40AF are made of sapphire and PFA. Never use sample fluids which can cause corrosion of fluid-contacting parts. Otherwise leaks or seepage may occur. The fluid system of the KS-40AF is designed to withstand sample fluid with a pressure of up to 300 kpa (gauge pressure). If this range is exceeded, leaks or seepage may occur. WARNING Always perform a leak test before supplying sample fl uid. Fill the system with pure water and leave it exposed to a pressure of 300 kpa (gauge pressure) for a while to verify that there are no leaks. WARNING When using different types of sample fl uid, verify that there is no danger of a chemical reaction. 12
31 Preparations Before changing to a different type of sample fluid, always clean the entire sample fluid system first to remove any remnants. Otherwise unwanted effects such as thermal build-up, hardening, particle accumulation etc. can occur. Use the system only after verifying that the cleaning process was successful. Otherwise there is a danger of serious accidents due to chemical reactions. Some sample fluids may react with water, leading to emulsification or thermal build-up. 13
32 Preparations Sample Fluid Flow Rate Adjustment Since this unit does not have a flow rate adjustment facility, additional equipment is required to adjust the sample fluid flow. Adjust the flow to a rate of 10 ml per minute. Pressurized sample fluid The pressure in the sample fluid system moves the sample fluid through the particle sensor. Use a flow controller or similar to adjust the flow rate. For details, please refer to the documentation of the flow controller. Non-pressurized sample fluid When the sample fluid is taken from a sample fluid bottle or similar and there is no pressure, the syringe sampler KZ-30W1 or KZ-30W2 should be used to draw the sample fluid into the KS-40AF. For details regarding flow rate adjustment, please refer to the documentation of the syringe sampler. WARNING Do not exceed the sample fl uid pressure rating of the unit. 14
33 Preparations Turning the System On Important Before turning on the controller KE-40, be sure to fi ll the fl uid system of the sensor with sample fl uid. If the interior of the fl ow cell is dry, the laser beam may cause burn-in of contamination particles, leading to an increase in noise. The KS-40AF does not have a power switch. When power to the controller KE-40 is turned on, the KS-40AF also will be powered. Important After turning on the power, wait at least 10 minutes before starting the measurement. This allows the light source output to stabilize. After turning the system off, wait at least 1 minute before turning it on again. Otherwise the light source may be damaged. 15
34 Measurement Measurement The KS-40AF does not incorporate a flow rate adjustment facility, additional equipment is required to adjust the sample fluid flow. Refer to Measurement system configuration examples on page 2 and configure a system which can supply sample fluid at a rate of 10 ml per minute. The KS-40AF does not have measurement controls or a display for measurement results. It is designed to perform measurement under control of a controller connected to it. For details, please refer to the documentation of the controller. Error Message When there is a problem with the unit, the LIQUID LEAK indicator, CELL indicator, and LASER indicator are used to indicate the condition. The following explanation explains operation and error messages when the Controller KE-40 is connected. For information on countermeasures, please refer to the section Maintenance on page 18. Status of indicators on KS-40AF Status: KS-40AF status Operation: KS-40AF operation and countermeasures Output: KE-40 serial interface output Shows the *** part of the measurement error report message having the format E / ***<EOL>. When LIQUID LEAK indicator is lit in red Status: Leak has occurred inside cabinet Operation: If measurement is in progress, it is stopped immediately and measurement data are discarded. Output: LIQUID_LEAK 16
35 Measurement When CELL indicator is lit in red Status: Measurement may not have been carried out properly, due to excessive particle number concentration in sample fluid, sensor contamination, condensation, or other cause. Operation: Measurement is possible, but may not be correct. Result data should be used with care. Output: NOISE_HIGH When LASER indicator is lit in red Status: Temperature of light source (laser diode) is outside of rated temperature range. Operation: If measurement is in progress, it is stopped and data are discarded. Output: LASER_TEMP. When LASER indicator flashes in red Status: Output of light source (laser diode) has dropped below rated limit. Operation: If measurement is in progress, measurement operation continues, but data should be discarded. Output: LASER_FAIL When LASER indicator and CELL indicator are off Status: Light source (laser diode) is off. Operation: If measurement is in progress, it is stopped and data are discarded. Output: LASER_OFF _ represents a space. 17
36 Maintenance Steps After Use When removing the KS-40AF from the measurement system, observe the following points. 1. Clean the piping with pure water. For more information, please refer to the section Cleaning the Piping on page Set the POWER switch of the controller KE-40 to OFF and disconnect the KS-40AF from the KE Fill the piping with ethyl alcohol to prevent bacteria growth and other contamination. 4. Disconnect the INLET and OUTLET tubes and attach the end plugs. Important The sample fluid system of the unit must always be filled with sample fluid (during measurement, storage, and transport). If the sample fl uid system dries up, particle detector problems (see page 5) can occur. 18
37 Maintenance When CELL indicator is lit in red This indicates that sample fluid with particle number concentration exceeding specifications was passed through the system, or that contamination or condensation has occurred in the particle detector flow cell. Perform purging as described in the section Cleaning the Piping on page 20. If the CELL indicator stays lit in red also after purging of the piping, here may be a problem with the particle detector flow cell or particle detector assembly. Please contact your supplier. When LIQUID LEAK indicator is lit in red A leak has occurred in the cabinet, or there is a problem with the leak sensor. Stop the sample fluid flow to the unit and turn off the Controller KE-40. Then contact your supplier. When LASER indicator is lit in red or flashes in red If these occur frequently, there may be a problem with the light source (laser diode). Please contact your supplier. 19
38 Maintenance Cleaning the Piping Clean the piping of the unit as follows. 1. Pass an ample amount of pure water through the system to replace the sample fluid with pure water. If the use of pure water poses the risk of a chemical reaction, use a different cleaning agent that is suitable for the sample fluid, and then purge with clean water. 2. Clean the KS-40AF by purging with pure water. It is not necessary to control the flow rate at this time. WARNING Keep the pressure (pressurization and suction) during purging within the rated sample fl uid pressure range (300 kpa (gauge pressure) for purging). If the rated pressure is exceeded, there is a risk of accidents. A sample system configuration for circular purging is shown below. Apply this example to your own requirements. Required equipment: External pump or syringe sampler, filter (0.1 µm or less) KS-40AF Filter ( 0.1µm or finer) Always use a filter. External pump or syringe sampler Pure water or cleaning fluid 20
39 Maintenance 3. After sufficient purging, pass pure water through the system at the prescribed flow rate (10 ml / min). 4. Verify that the CELL indicator of the KS-40AF is now lit in green. If the CELL indicator is still lit in red, repeat the cleaning procedure from step 2, but use cleaning fluid instead of pure water. Note Use cleaning fluid that does not contain solid matter, such as optical cleaning fluid or cleaning alcohol. Important If the CELL indicator is still lit in red also after purging with cleaning fl uid, cleaning of the particle detector fl ow cell is required. Please contact your supplier. Periodic maintenance In order to maintain continued measurement precision, it is recommended to have the unit checked and serviced once per year. Contact the supplier regarding periodic maintenance. Transport Before returning the unit for maintenance or servicing, purge the sample Important fluid system thoroughly with pure Before returning the unit for maintenance or servicing, purge the sample fluid system water to remove any sample fluid thoroughly with pure water to remove any remnants. sample fluid remnants. Then fill the piping with alcohol (or Then fill the piping with alcohol (isopropyl ethyl alcohol) and seal it with the end alcohol, ethyl alcohol or methyl alcohol). plugs. Wipe the outside of the unit clean so that there are no remnants of chemical substances on the unit. The label is affixed on the rear of the unit. 21
40 Specifications Optical system 90 sideway light-scattering method Light source Laser diode (rated output 200 mw; wave length 830 nm) Laser product classification Class 1, IEC (2001) Internal particle detection mechanism uses Class 3B laser Light detector PIN type photodiodes Materials of component parts exposed to sample fluid Sapphire, PFA Allowable sample fluid types Fluids which do not corrode the fluid contact materials Calibration Polystyrene latex (PSL) spheres* with refractive index 1.6 in pure water *The PSL spheres are calibrated according to the TEM (transmission electron microscopy) by the supplier, JSR Corporation. Minimum particle size 0.1 µm Measurable particle size 0.1 to 2 µm (with PSL particles of refractive index 1.6 in pure water) Measurement size range Five channels ( 0.1 µm, 0.15 µm, 0.2 µm, 0.3 µm, 0.5 µm) Count efficiency 60 ±15% (measuring PSL particles in the range of 0.3 µm, using count of 0.2 µm and above for comparison with reference unit) Sample flow rate 10 ml / min Maximum particle number concentration 1200 particles / ml (coincidence loss 5% for 0.1 µm particles) 22
41 Specifications Sample fluid temperature range +15 to +35 C (no moisture condensation on flow cell) Allowable sample fluid pressure 300 kpa or less (gauge pressure) Warm-up time 10 minutes Sample fluid connectors INLET Sample fluid inlet, 2 4 dia. flared tube joint OUTLET Sample fluid outlet, 2 4 dia. flared tube joint PURGE Purge gas inlet, Rc 1 / 8 (1 / 8 PT female) Indicators PARTICLE MONITOR Briefly flashes green when particles above minimum particle size are detected. LIQUID LEAK Lit (green) during normal operation Lit (red) when leak was detected CELL Lit (green) during normal operation Lit (red) when contamination was detected or maximum particle number concentration exceeded Off when light source is off LASER Lit (green) during normal operation Lit (red) when light source temperature is out of range Flashing (red) when light source output is below normal Off when light source is off POWER Lit while power to unit is on Input/output connectors CONTROLLER For connection of controller KE-40 LIQUID LEAK ALARM Shorted during normal operation, open when internal leak was detected (M3 terminal, fit in with 1.25 mm 2 electric wire) Maximum load: 30 V DC, 1 A EXT Test output Power requirements Supplied via Controller KE-40 23
42 Specifications in the following graph (no con- Installation inclination angle Max. 2 Ambient conditions for operation Shaded section densation) Temperature ( C) Relative humidity (%) Ambient conditions for storage -10 to +50 C, less than 90% RH (no condensation and no freezing in internal piping) Dimensions 139 (H) 245 (W) 179 (D) mm (maximum) 125 (H) 240 (W) 151 (D) mm (excluding joints and other protruding parts) Weight Approx. 4 kg Supplied Accessories Tube A vacuum pack 1 (2 4 dia., 1.5 m flared PFA tube 2, union joint 1) Particle size labels 1 Instruction manual 1 Instruction sheet for Transport and Installation 1 Liquid-borne particle counter usage precautions 1 Inspection certificate 1 24
43 Specifications Rear view Left-side view Top view Front view Right side view Dimensional Drawings Unit: mm 25
44 No
SPECIFICATIONS PARTICLE SENSOR KS-18F Higashimotomachi, Kokubunji, Tokyo , Japan
SPECIFICATIONS PARTICLE SENSOR KS-18F 3-20-41 Higashimotomachi, Kokubunji, Tokyo 185-8533, Japan No. 05083-5E 18-01 Printed in Japan Outline The KS-18F is a sensor which uses the light scattering method
More informationSPECIFICATIONS PARTICLE SENSOR KS-19F Higashimotomachi, Kokubunji, Tokyo , Japan
SPECIFICATIONS PARTICLE SENSOR KS-19F 3-20-41 Higashimotomachi, Kokubunji, Tokyo 185-8533, Japan No. 13012-3E 17-06 Printed in Japan Outline The KS-19F is a sensor which uses the light scattering method
More informationTHE BP-301 SERIES. Operating and Service Manual. Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv)
THE BP-301 SERIES Operating and Service Manual Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv) Issue B October 2015 1 TABLE OF CONTENTS 1. Description... 3 2. Installation... 3 3. Operation...
More informationColumbus Instruments
0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011
More informationPre-filter for Brushless Blower Instruction Manual
Pre-filter for Brushless Blower 781-70-702 Instruction Manual Minebea Co., Ltd. Industrial Machinery Business Unit 251-8531 1-1-1 Katase,Fujisawa,kanagawa Contact Us Industrial Machinery Product Sales
More informationUser s Guide Temperature Sensor Converter TSC-599
User s Guide Temperature Sensor Converter TSC-599 ILX Lightwave Corporation 31950 Frontage Road Bozeman, MT, U.S.A. 59715 U.S. & Canada: 1-800-459-9459 International Inquiries: 406-556-2481 Fax 406-586-9405
More informationTHE BP-690 SERIES. Operating and Service Manual. Series includes all variants of BP-LF/MF-690/691
THE BP-690 SERIES Operating and Service Manual Series includes all variants of BP-LF/MF-690/691 Issue B April 2015 1 TABLE OF CONTENTS 1. Description... 3 2. Installation... 3 3. Operation... 4 4. Special
More informationTHE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301
THE HF-300 SERIES Operating and Service Manual Series includes all variants of HF-300/301 Issue A July 2015 1 TABLE OF CONTENTS 1. Description... 3 2. Installation... 3 3. Operation... 4 3.1. Spring Loaded...
More informationOPERATION MANUAL NTF-15
OPERATION MANUAL NTF-15 Nitrogen Tire Filling Valve Stem Caps (Qty=200) Order P/N 436075 RTI Technologies, Inc 10 Innovation Drive York, PA 17402 800-468-2321 www.rtitech.com 035-81235-00 (Rev B) TABLE
More informationTHE MF-400 SERIES. Operating and Service Manual. Series includes all variants of MF-400/401
THE MF-400 SERIES Operating and Service Manual Series includes all variants of MF-400/401 Issue A October 2013 1 TABLE OF CONTENTS 1. Description... 3 2. Installation... 3 3. Operation... 4 4. Special
More informationE8AA. Pressure Sensor of Stainless Steel Construction Is Ideal for a Wide Range of Applications. Pressure Sensor (Stainless Steel Diaphragm)
Pressure Sensor (Stainless Steel Diaphragm) CSM DS_E_3_1 Pressure Sensor of Stainless Steel Construction Is Ideal for a Wide Range of Applications Incorporates double diaphragms consisting of SUS316L stainless
More informationUser Manual. Quantos Automated Dosing Liquid Module
User Manual Liquid Module 1 Safety Information 1.1 Definition of warnings and symbols Signal Words WARNING for a hazardous situation with medium risk, possibly resulting in severe injuries or death if
More information6900 Maintenance Instruction System Flush
Equipment Required FA74005 Damper Drain Tube FA16005 Cover Removal Tool FA900005 Beaker 0.25 Litre FA900003 Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves Latex
More informationOperation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER
Operation Manual - PN 0017946001 A MENSOR MODEL 73 SHOP AIR BOOSTER Mensor Model 73 Shop Air Booster System (750 psi Version) April 23, 2012 Trademarks / Copyright Mensor is a registered trademark of Mensor
More informationPRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.
PRS(TC)4,8 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to
More informationP5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.
Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 TestEquipmentDepot.com P5513 Pneumatic Comparison Test Pump Users Manual PN 3963372 November 2010 2010 Fluke Corporation. All
More informationACV-10 Automatic Control Valve
ACV-10 Automatic Control Valve Installation, Operation & Maintenance General: The Archer Instruments ACV-10 is a precision automatic feed rate control valve for use in vacuum systems feeding Chlorine,
More informationLRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator.
LRS(H)4 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to the
More informationOPERATION MANUAL NTF-60 Plus
OPERATION MANUAL NTF-60 Plus Nitrogen Tire Filling Valve Stem Caps (Qty=200) Order P/N 436075 RTI Technologies, Inc 10 Innovation Drive York, PA 17402 800-468-2321 www.rtitech.com 035-81264-00 (Rev A)
More informationSeries IRV1000/2000/3000
CAT.ES60-16 A Regulator Series /2000/3000 Allows adjustment of vacuum line pressure Regulator Series /2000/3000 3 sizes offered in the series Variations have been expanded to three sizes from only one
More informationRS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.
RS(H)10,15 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to
More informationSpiratec ST14, ST16 and ST17 Sensor Chambers and sensors
0862050/1 IM-P086-18 MI Issue 1 Spiratec ST14, ST16 and ST17 Sensor Chambers and sensors Installation and Maintenance Instructions 1. Safety Information 2. General product information 3. Installation 4.
More informationCalibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.
Advanced Calibration Designs, Inc. Calibration Gas Instrument INSTRUCTION MANUAL Release I www.goacd.com Instruction Manual Gas Generator Release I TABLE OF CONTENTS I. General Description Page 2 II. Start-Up
More informationInstruction Manual. CG16K Barometrically Compensated Capsule Dial Gauge. CG16K Capsule Dial Gauge, 0 to 25 mbar
Instruction Manual D356-10-880 Issue G Original CG16K Barometrically Compensated Capsule Dial Gauge Description CG16K Capsule Dial Gauge, 0 to 1040 mbar CG16K Capsule Dial Gauge, 0 to 125 mbar CG16K Capsule
More informationSA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Dial Adjustment)
3820050/4 IM-P382-01 CH Issue 4 SA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Dial Adjustment) Installation and Maintenance Instructions 1. Safety information 2. Use 3.
More informationtesto Leakage detector for refrigerants Instruction manual
testo 316-3 Leakage detector for refrigerants Instruction manual 2 1 Contents 1 Contents 1 Contents... 3 2 Safety and the environment... 4 2.1. About this document... 4 2.2. Ensure safety... 4 2.3. Protecting
More informationSeries SFD. Clean Air Filter. Variations. Made to Order. Type. Cartridge (replaceable element) Disposable (non-replaceable element) Up to 100
Clean Air Filter Variations SFD SFD SFD SFD Made to Order Type Disposable (non-replaceable element) Cartridge (replaceable element) Flow rate L/min (ANR) (at inlet pressure.7 MPa) Up to 6 Up to 8 Up to
More informationOXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:
OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle
More informationTEST BENCH SAFETY VALVES ¼ - 5 DN10 DN125
TEST BENCH SAFETY VALVES ¼ - 5 DN10 DN125 Model: VC-40-VYC Table of contents 1. - Installing the test bench 1.1.1- Connecting the compressed air / nitrogen source 1.1.2- Maximum test pressure according
More informationATD LB PRESSURE BLASTER INSTRUCTION MANUAL
ATD-8402 90LB PRESSURE BLASTER INSTRUCTION MANUAL SAVE THESE INSTRUCTIONS SAFETY INSTRUCTIONS FOR SANDBLASTER 1. Before opening the tank release the air pressure on the sand tank. To do this, turn off
More informationCustomer Responsibilities. Important Customer Information. Cary 4000/5000/6000i UV-Vis spectrophotometer Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationMODEL GT820 OXYGEN SENSOR
INSTRUCTION MANUAL MODEL GT820 OXYGEN SENSOR 70046 The information and technical data disclosed by this document may be used and disseminated only for the purposes and to the extent specifically authorized
More informationPVC Quick Drain Valve
PVC Quick Drain Valve LVW Series Complies to JIS standard for polyvinyl chloride piping (JIS K 6742) Applicable fluids: Deionized water, Chemical liquids Application Example Cleaning Equipment Process
More informationWafer Check Valve. Contents. User s Manual. (1) Be sure to read the following description of our product warranty 1
Serial No. H-V066-E-3 Wafer Check Valve User s Manual Contents (1) Be sure to read the following description of our product warranty 1 (2) General operating instructions 2 (3) General instructions for
More informationCalcimeter Instruction Manual
Hohner (UK - Canada - Texas) Calcimeter Instruction Manual The Hohner Calcimeter is based on industry standard versions, and is used to measure the calcium carbonate and magnesium carbonate in samples.
More informationManual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany
Manual Degasser DDG-75 April 2007 1 Manual Online Degasser DDG-75 DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE-68766 Hockenheim Germany Tel. +49-6205-9450-0 Fax. +49-6205-9450-33 email info@duratec.com
More information3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289
3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS Item #31289 The EASTWOOD 3 GALLON, OILLESS PANCAKE COMPRESSOR, with an Integral Air Regulator, efficiently supplies all compressed air requirements for
More informationOPERATOR S MANUAL Ar-Gone Weld Gas Analyzer
July 2011 OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer WARNING! Before operating this product, read and understand this Operator s Manual. Become familiar with the potential hazards of this unit. Contact
More informationINSTALLATION GUIDE Gas Rangetops
INSTALLATION GUIDE Gas Rangetops Contents Wolf Gas Rangetops........................... 3 Safety Instructions............................ 4 Gas Rangetop Specifications.................... 5 Gas Rangetop
More informationCompressed Gas Cylinders - Guideline
Environmental Health and Safety Compressed Gas Cylinders - Guideline Date of Issuance: 1/10/18 Revision Number: Initial Revision Date: Prepared by: EH&S 1. Purpose Carnegie Mellon University has developed
More informationTranslation Principles of testing and certification for insulating hose assemblies on cable cutting devices
Translation Principles of testing and certification for insulating hose assemblies on cable cutting devices Status as of 2014-01 Principles of testing Insulating hose assemblies on cable cutting devices
More informationOperation Manual Air Saver Unit ASV5000 Series
9IM-V066-b Operation Manual Air Saver Unit ASV5000 Series Thank you for your choice of Kuroda Pneumatics LTDs product on this time. Please read this operation manual carefully and use the product correctly.
More informationOEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE
OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE 1 These instructions generally describe the installation, operation, and maintenance of subject equipment. The manufacturer reserves the right
More informationNon-corrosive gas Liquid. Corrosive gas
997 PRESSURE SENSRS INFRMATIN INTRDUCTIN Principle of operation A sensor converts changes in of a gas or a liquid into an electrical signal by means of a sensing device, and generates an analog output
More informationType 2000 INOX. Quickstart. English Deutsch Français. 2/2-way angle seat valve 2/2-Wege Schrägsitzventil Vanne à siège incliné 2/2 voies
Type 2000 INOX 2/2-way angle seat valve 2/2-Wege Schrägsitzventil Vanne à siège incliné 2/2 voies Quickstart English Deutsch Français Contents 1 QUICKSTART... 2 2 CONTACT ADDRESS... 2 3 INTENDED USE...
More informationCSA Sample Draw Aspirator Adapter Operator s Manual
30-0951-CSA Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0367 Revision: 0 Released: 4/30/15 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating
More informationSA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Knob Adjustment)
3810050/5 IM-P381-01 CH Issue 5 SA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Knob Adjustment) Installation and Maintenance Instructions 1. Safety information 2. Use 3.
More informationINSTALLATION & MAINTENANCE INSTRUCTIONS
DESCRIPTION / IDENTIFICATION The QBX series valve uses Proportion-Air closed loop technology for Pressure control. It gives an output pressure proportional to an electrical command signal input. The QB1X
More informationType C-CUT. Quickstart. English Deutsch Français
Capillary modules for micro and ultrafiltration Kapillarmodule für die Mikro- und Ultrafiltration Modules capillaires pour micro- et ultrafiltration Quickstart English Deutsch Français We reserve the right
More informationINSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES
INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES THIS BOOKLET CONTAINS PROPRIETARY INFORMATION OF ADVANCED SPECIALTY GAS EQUIPMENT CORP. AND IS PROVIDED TO THE PURCHASER SOLELY FOR USE
More information6301 TYPE CAST IRON SAFETY VALVES
Pressure (bar) 6301 TYPE CAST IRON SAFETY VALVES FEATURES The 6301 type safety valve is a device designed to protect installations against possible overpressure. It operates automatically and closes when
More informationAssembly Drawing: W-311B-A01, or as applicable Parts List: W-311B-A01-1, or as applicable Special Tools: , , &
REDQ Regulators Model 411B Barstock Design Powreactor Dome Regulator OPERATION AND MAINTENANCE Contents Scope..............................1 Installation..........................1 General Description....................1
More informationReflow Oven HHL3000 INSTRUCTION MANUAL POHUA - jedyny autoryzowany przedstawiciel w Polsce
POHUA - jedyny autoryzowany przedstawiciel w Polsce www.pohua.pl AOYUE TONGYI ELECTRONIC EQUIPMENT FACTORY Jishui Industrial Zone, Nantou, Zhongshan City, Guangdong Province, P. R. China www.aoyue.com
More informationVS18/26 with PROFINET and EtherNet/IP Interface. ATEX Installation Instructions
VS18/26 with PROFINET and EtherNet/IP Interface ATEX Installation Instructions INDEX 1. INTENDED USAGE 3 2. OPERATING MANUAL ATEX 4 2.1 General conditions 4 2.2 Installation 5 2.3 Operating 5 2.4 Failures
More informationOverview. Front Panel: Keypad and Display
Overview The GA-200B is an analyzer that integrates a gas sampling system with sensors to measure and display the concentrations of oxygen and carbon dioxide in a sample as the percentage of a gas in the
More informationInert Air (N2) Systems Manual
INSTRUCTION MANUAL Inert Air (N2) Systems Manual N2-MANUAL 2.10 READ AND UNDERSTAND THIS MANUAL PRIOR TO OPERATING OR SERVICING THIS PRODUCT. GENERAL INFORMATION Positive pressure nitrogen gas pressurizing
More informationTESCOM 50-4X Series Safety, Installation & Start-Up Procedures
Operations & Service Manual TESCOM 50-4X Series Safety, Installation & Start-Up Procedures Do not attempt to select, install, use or maintain this product until you have read and fully understood this
More informationPegas 4000 MF Gas Mixer InstructionManual Columbus Instruments
Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...
More informationAUTOVENT 4000 VENTILATOR
OVERVIEW AUTOVENT 4000 Only properly trained and approved Escambia County Bureau of Public Safety Paramedics are to use the AutoVent 4000 ventilator manufactured by LSP to transport patients already on
More informationFLOWMETER. FM-PF Series
Integral mold flowmeter available for ultra-pure water/chemicals Caution FLOWMETER Instruction manual for FM-PF Series (Excel type) Before use, please read this Instruction manual with caution for your
More informationSection 3- Repair and Calibration Procedures
Section 3- Repair and Calibration Procedures 1. Refrigerant Removal Procedures 2. Scale/PC Board Recalibration Procedure 3. PC Board Configuration Procedure of AR400 4. Scale Tare Procedure 5. Gauge Calibration
More informationOPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N
OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC. 4075 East Market Street York, PA 17402 Manual P/N 035-80589-02 Table of Contents Components... 2 Test... 4 Recovery/Recycling... 5 Evacuation... 7 Charging...
More informationEnglish. Introduction. Safety Instructions. All Products. Inspection and Maintenance Schedules. Parts Ordering. Specifications WARNING WARNING
Contents All Products... Gb-1 Control Valves... Gb-2 Control Valve Actuators... Gb-3 Regulators... Gb-3 Relief Valves... Gb-4 Instruments, Switches, and Accessories... Gb-4 Products Covered by Battery
More information! Warning, refer to accompanying documents.
About this Manual To the best of our knowledge and at the time written, the information contained in this document is technically correct and the procedures accurate and adequate to operate this instrument
More information36G22, 36G23, 36G24 & 36G52 36J22, 36J23, 36J24 & 36J52 DSI and HSI Single Stage Combination Gas Valve
Operator: Save these instructions for future use! FAILURE TO READ AND FOLLOW ALL INSTRUCTIONS CAREFULLY BEFORE INSTALLING OR OPERATING THIS CONTROL COULD CAUSE PERSONAL INJURY AND/OR PROPERTY DAMAGE. DESCRIPTION
More informationAPC Pressure Reducer with Sensor
OBJ_DOKU-194490-001.fm Seite 1 Donnerstag, 16. Februar 2017 4:22 16 NOTES ON USE USA APC Pressure Reducer with Sensor 20134-006 OBJ_DOKU-194490-001.fm Seite 2 Donnerstag, 16. Februar 2017 4:22 16 Table
More informationOperating Instructions Part No
DIGITAL AUTOMATIC TYRE INFLATOR Operating Instructions Part No. 11.0545 Thank you for selecting this Jamec Pem Automatic Tyre Inflator. Please read this manual before carrying out any installation or service
More informationMaximum 0.85 MPa pressure setting Long-life, high flow perfect for balancer applications
Outstanding performance in extremely low pressure and low pressure ranges from 0.003 to. Realizing high performance, energy saving, and compact size. Realize precise pressure control in a pressure range
More informationTechnical Specifications
No. : PS##-OMG0006 Technical Specifications PSE560 series Contents Safety.P2 Model Indication Method.P6 Specification.P7 Full View with Dimensions.P8 Installation.P9 Examples of Internal Circuit and Wiring.P10-1
More informationQuick Installation Guide Doc nr.: C Date:
Quick Installation Guide Doc nr.: 9.17.093C Date: 13-05-2016 SCOPE OF THIS GUIDE mini CORI-FLOW instruments are highly accurate instruments for measuring and controlling the mass flow rate of liquids and/or
More informationTemperature gauge Installation Guide
Temperature gauge Installation Guide improper functioning, or damage. Additional information can be found at: CAUTION: Read this installation guide carefully before unpacking the temperature gauge. Improper
More informationCP10 Sensor Installation and Maintenance Instructions
4030150/9 IM-P403-26 EMM Issue 9 CP10 Sensor Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4. Maintenance 5. Spare parts Copyright 2017
More informationPRODUCT NAME. Main Line Filter. MODEL / Series / Product Number
PRODUCT NAME Main Line Filter MODEL / Series / Product Number AFF75A-20 / AFF75A-20X13 AFF125A-30 / AFF125A-30X13 AFF150A-40 / AFF150A-40X13 AFF220A-40 / AFF220A-40X13 This operation manual is for model
More informationVACUUM REGULATORS CONTENTS
CAD drawing data catalog is available. ACCESSORIES GENERAL CATALOG AIR TREATMENT, AUXILIARY, VACUUM, AND FLUORORESIN PRODUCTS CONTENTS Small Regulators Features 759 Specifications, Order Codes, Flow Rate
More information* * Data Sheet and Operating Manual. Digital manometer ME01 ## # 87 # HL S#### Dust explosion protection Zone 22.
*09005087* DB_BA_EN_ME01_S Rev.A 12/16 *09005087* d e v e l o p i n g s o l u t i o n s Data Sheet and Operating Manual ME01 Digital manometer Table of Contents ME01 ## # 87 # HL S#### Dust explosion protection
More informationNo.IL02-OM00011-A PRODUCT NAME LOCK UP VALVE. MODEL/ Series IL201 IL211
No.IL02-OM00011-A PRODUCT NAME LOCK UP VALVE MODEL/ Series IL201 IL211 Table of Contents Instructions for your Safety 1~2 1. Summary 3 2. Specification 3 3. Operation Principle 3~4 4. Transportation and
More informationCompact Ball Valves ASAHI AV VALVES. Contents. User s Manual. (Page) (1) Be sure to read the following warranty clauses of our product 1
Serial No. H-V030-E-5 Compact Ball Valves User s Manual Contents (Page) (1) Be sure to read the following warranty clauses of our product 1 (2) General operating instructions 2 (3) General instructions
More informationSDX Submersible Depth Transmitter User Manual
SDX Submersible Depth Transmitter User Manual October 2007 USER INFORMATION Stevens makes no warranty as to the information furnished in these instructions and the reader assumes all risk in the use thereof.
More informationPropane Conversion Kit Instruction
Propane Conversion Kit Instruction Condensing gas boiler Required Input Rates Logamax plus GB62-80 kw 270,000 btu/hr Logamax plus GB62-00 kw 35,000 btu/hr This kit and instructions are for converting the
More informationSupplementary Operator s Manual 42/24-14 EN Rev. 3
Advance Optima Continuous Gas Analyzers AO2000 Series with Uras26, Magnos206, Caldos25, Caldos27 Designed per Category 3G for Measurement of Flammable Gases ( Safety Concept ) and Non-flammable Gases Supplementary
More informationBSS 2 Bottle Sampling System. Instruction manual Version 1.8
BSS 2 Bottle Sampling System Instruction manual Version 1.8 Serial-No.:... Version-/Rev.-Date : 12.06.2012 Contents 1. General... 3 2. Safety Instructions... 4 3. Extent of Delivery... 5 4. Switch Panel...
More informationPH01 Perfusion Cannula Manual
PH01 Perfusion Cannula Manual Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted without the express written permission of Multi
More informationFOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM
IMI CORNELIUS INC One Cornelius Place Anoka, MN 55303-623 Telephone (800) 238-3600 Facsimile (612) 22-326 INSTALLATION INSTRUCTIONS FOR INSTALLING CO 2 BLENDER KIT (P/N 111612000 IN BEER SYSTEM SECONDARY
More informationDPC-30 DPC-100. Reference Manual
DPC-30 DPC-100 Reference Manual 1. Introduction 1.1 Description The Martel DPC Digital Pneumatic Calibrator improves upon traditional dial gauge pneumatic calibrators. The Martel DPC improves accuracy,
More informationInstructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems
Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the
More informationP499 Heavy Duty Pressure Transducer
P499 Heavy Duty Pressure Transducer Product Bulletin Code No. LIT-??? E Issued 11 2006 The P499 Series is a new global Pressure Transducer with an excellent price performance ratio. The P499 exceeds the
More informationwith O 2 Controller Instruction Manual
14500 Coy Drive, Grass Lake, Michigan 49240 734-475-2200 E-mail: sales@coylab.com www.coylab.com Hypoxic Cabinets (In-Vivo / In-Vitro) with O 2 Controller Instruction Manual Index Page Warranty 2 Warnings
More informationTHE HF-210 SERIES. Operating and Service Manual. Series includes all variants of HF-210/211. Issue A October 2014 *Section 6.
THE HF-210 SERIES Operating and Service Manual Series includes all variants of HF-210/211 Issue A October 2014 *Section 6.1 in development 1 TABLE OF CONTENTS 1. Description... 3 2. Installation... 3 3.
More informationLiquid Nitrogen Cooling Accessory CE
109 Lukens Drive New Castle, DE 19720 Thermal Analysis & Rheology A SUBSIDIARY OF WATERS CORPORATION Liquid Nitrogen Cooling Accessory CE Operator's Manual PN 891300.001 Rev. D (Text and Binder) PN 891300.002
More informationXpressFill XF2500 Counter Pressure Filler. Operating Instructions
XpressFill XF2500 Counter Pressure Filler Operating Instructions ii XPRESSFILL Systems LLC lling machine. ank you for choosing our handcrafted bottle filler as the technology to bottle your passion. We
More informationSDX Submersible Depth Transmitter User Manual
SDX Submersible Depth Transmitter User Manual July 2017 USER INFORMATION Stevens makes no warranty as to the information furnished in these instructions and the reader assumes all risk in the use thereof.
More informationAIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST
AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST OIL-LESS TYPE IMPORTANT: PLEASE READ CAREFULLY BEFORE STARTING OPERATIONS. THE CONTENTS ARE FOR GENERAL INFORMATION OF ALL THE SIMILAR MODELS. Record
More informationInstallation Instructions
48011HW NATURAL GAS TO PROPANE CONVERSION KIT 58HDX, 359AAV, PG9YAB, PG9YAA Installation Instructions SAFETY REQUIREMENTS Installing and servicing heating equipment can be hazardous due to gas and electrical
More informationDigital Melting Point Apparatus
Digital Melting Point Apparatus Heating Plateau Ramping Start/Stop Plateau set Ramp stop Hold User Guide Version 1.1 Heating Viewing tube Sample Chamber IEC power inlet socket Power on/off Temperature
More informationV48 Series 3-Way Pressure-Actuated Water-Regulating Valves With Union Fittings
Installation Instructions V48 Issue Date 08/30/01 V48 Series 3-Way Pressure-Actuated Water-Regulating Valves With Union Fittings Application IMPORTANT: The V48 Series 3-Way Pressure-Actuated Water-Regulating
More informationUser Manual. Quantos Automated Dosing Powder Module
User Manual Powder Module 1 Introduction Thank you for purchasing the Quantos dosing system. Quantos is based on the renowned and reliable XPE series balances from METTLER TOLEDO. Besides the highly accurate
More informationSinorix MANOB480, MANOCAB, MANOEOL Installation Maintenance
Sinorix MANOB480, MANOCAB, MANOEOL Installation Maintenance Building Technologies 2017-10-24 Control Products and Systems Legal notice Legal notice The specifications and availability are subject to change
More informationMODEL 154XST INSTALLATION AND SERVICE INSTRUCTIONS FOR MODEL 154XST STROBE LIGHT FOR USE IN HARSH ENVIRONMENTS/ HAZARDOUS LOCATIONS
MODEL 154XST INSTALLATION AND SERVICE INSTRUCTIONS FOR MODEL 154XST STROBE LIGHT FOR USE IN HARSH ENVIRONMENTS/ HAZARDOUS LOCATIONS 2561762B REV. B 210 Printed in U.S.A. INSTALLATION AND SERVICE INSTRUCTIONS
More informationOxford Instruments Plasma Technology. PlasmaPro NGP80. Installation Data Sheet. Original Instructions.
Oxford Instruments Plasma Technology www.oxford-instruments.com About this data sheet This data sheet provides basic information about the installation of a tool. Floor and wall loadings Table 1 lists
More informationJUMO dtrans p30 Pressure Transmitter
609 Fulda, Germany Postal address: 605 Fulda, Germany Phone: +49 661 600-0 Fax: +49 661 600-607 Phone: +44 179 6 55 Data Sheet 40466 Page 1/6 JUMO dtrans p0 Pressure Transmitter General application Pressure
More information