Micro GC Fusion Pre-Installation Checlist
|
|
- Garey Patterson
- 5 years ago
- Views:
Transcription
1 Site Preparation Proper site preparation prior to installing is very important. Contact INFICON ( with any questions regarding site preparation. Checkboxes ( ) identify hardware and actions that must be addressed when installing. List of Required Tools 1 1/8" wrench recommended, or adjustable large wrench (for attaching regulator to gas tank) Teflon tape for regulator connections The following tools are included in the optional Install Kit (PN G1). Open-ended wrench, 1/4 in. and 5/16 in. Open-ended wrench, 7/16 in. and 9/16 in. Open-ended wrench, 3/8 in. and 7/16 in. Hex driver, 2.5 mm ball-point long shank Tubing cutter Carrier Gas....Carrier gas (e.g., helium, argon) purity of %, or minimum % when the appropriate carrier gas trap is installed. Contact INFICON for carrier gas trap assistance. A pressurized [>16,547 kpa (>2,400 psi)] 50 liter tank of carrier gas will last approximately one year or more.....dual Stage Regulator for pressurized carrier gas tank with outlet pressure regulated to kpa (58-62 psi). A Swagelok 1/8 in. fitting is preferred on the regulator outlet to connect the carrier gas feed tube to. In North America, use the CGA 580 inlet connection (PN P1) for helium, nitrogen and argon carrier gas. Use the CGA 350 inlet connection (PN P1) for hydrogen carrier gas. For installations outside of North America, use the appropriate inlet connections as standard for the region. Exceeding 427 kpa (62 psi) carrier gas pressure may result in instrument damage. 1 of 5
2 Calibration Gas....Moisture/hydrocarbon trap (PN G2870A-01) required for the carrier gas when using Molsieve or Alumina columns. To protect and to enhance performance, carrier gas traps should be used to purify the carrier gas before introduction into. A hydrocarbon/moisture trap (PN G2870A-01) is usually sufficient.....copper or stainless steel tubing (pre-cleaned) 1/8 in. OD for carrier gas connections between the tank regulator output and carrier gas inlet. has up to two 1/8 in. Swagelok male carrier gas inlet connections. 3 m (10 ft.) of pre-cleaned GC grade copper tubing is included with. Additional copper tubing is available in the installation kit (PN G1), or can be ordered in 15 m (50 ft.) lengths (PN ). Do not clean copper or stainless steel tubing with solvent. Do not use Teflon or Tygon tubing for carrier gas.....assortment of male and female Swagelok fittings for regulator, tubing, etc., including nuts and front and back ferrules. Typical size is 1/8 in. but adapters for 1/4 in. to 1/8 in. may be needed if 1/4 in. tubing is used.....1/8 in. tee fitting and 1/8 in. ball valve, depending on connection requirements.....supply of calibration gas specific to the analysis from a commercial gas supplier. Calibration gas must be available at the time of installation. Calibration gas must closely resemble the composition of the actual process gas.....a regulator, if necessary, to reduce calibration gas pressure below 69 kpa (10 psi) at the inlets. The regulator must be clean, grease-free and non-venting. Checkout gas mixtures are only used for instrument installation and start up. Checkout gas mixtures supplied by INFICON are not calibration gases. 2 of 5
3 Sample Gas Pressure Range Sample gas pressure greater than 172 kpa (25 psi) will damage.....maximum pressure for direct sample injection into either the front or rear sample inlet is 172 kpa (25 psi). To prevent damage to, limit input sample pressure to 69 kpa (10 psi) or lower.....when the Integrated Sample Conditioner is installed, sample pressures up to 6,895 kpa (1,000 psi) can be introduced into either the front or rear sample inlet. Sample pressures higher than 6895 kpa (1,000 psi) must be reduced.....when a Genie Filter Assembly (PN G1) is installed, sample pressures up to 3447 kpa (500 psi) can be introduced into either the front or rear sample inlet. Sample pressures higher than 3447 kpa (500 psi) must be reduced. Installation Space and Venting Module instrument Dimension (L x W x H): 43.3 x 20 x 26.5 cm (17 x 7.9 x 10.4 in.) Weight: 6.8 kg (11 lb.)....install in a location with easy access to the instrument and associated computer.....for optimum performance and to dissipate heat, do not restrict the airflow around. WARNING Safely vent sample streams potentially toxic, noxious, or flammable gases outside the instrument and away from the operating area. Vent toxic gases to a fume hood, chemical trap, or reaction medium.....avoid venting into drafty areas, such as in front of a heating/cooling vent. 3 of 5
4 Power Requirement....Standard V (ac) electric outlet.....maximum power for 2-Module instrument: 260 W....Uninterruptible power supply with surge protection is recommended. Computer Hardware Requirement Minimum hardware configuration required to run software. Tablet....RAM Gb....Memory Gb....CPU GHz....Screen Size in.....resolution x 768 pixels....network Connectivity Wi-Fi wireless Computer....RAM Gb....Hard Drive Gb....CPU GHz....Screen Size in.....resolution x 768 pixels....network Connectivity Wi-Fi wireless or RJ-45 wired 4 of 5
5 Network Preparation....If will be connected to a local area network (LAN) using a static IP address, the hostname and static IP address must be assigned by the LAN administrator.....it is not necessary for a LAN administrator to set up the hostname and static IP address when Dynamic Host Configuration Protocol (DHCP) is used to assign the IP address. Web Browser Requirement.... supports the following tablet web browsers: Android 4+ stock browser Google Chrome or Firefox for Android Safari on ios 6 for ipad Symbian is not supported..... supports the following computer web browsers: Google Chrome for Mac OS and Windows XP or above Firefox for Mac OS and Windows XP or above Internet Explorer version 10 and above for Windows 7 or above 2014 INFICON reachus@inficon.com Due to our continuing program of product improvements, specifications are subject to change without notice. All trademarks are the property of their respective owners. 5 of 5
6890 GC Site Preparation
6890 GC Site Preparation (a16011) This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this material. The information
More information490 Micro GC Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationSite Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)
Purpose of Procedure Your site must meet this specification or set of requirements to assure a successful and timely installation of your 7820MSD Series Mass Selective Detector (MSD). This checklist is
More informationSite Preparation Specification for GCMS system. Dimensions and Weight. Agilent 5975C MSD Series G3170A, G3171A, G3172A, G3174A
Purpose of Procedure Your site must meet this specification or set of requirements to assure a successful and timely installation of your 5975C Series Mass Selective Detector (MSD). This checklist is designed
More informationDetector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane
Gas requirements Gases for packed columns The carrier gas you use depends upon the type of detector and the performance requirements. Table 520-1 lists gas recommendations for packed column use. In general,
More informationFor additional information about our solutions, please visit our web site at US/Pages/HomePage.
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationAgilent 7697A Headspace Sampler
Agilent 7697A Headspace Sampler Site Preparation Agilent Technologies Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by any means (including electronic
More informationCustomer Responsibilities. Important Customer Information GCMS QQQ Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationAgilent 7820A GC. Site Preparation Guide. Agilent Technologies
Agilent 7820A GC Site Preparation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2009 No part of this manual may be reproduced in any form or by any means (including electronic storage and
More informationCustomer Responsibilities
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationCustomer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist
Thank you for purchasing an Agilent system. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is
More information7820A GC Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationAgilent 7667A. Mini Thermal Desorber. Site Preparation
Agilent 7667A Mini Thermal Desorber Site Preparation Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval
More information490 Micro GC Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationManual Solvent Purification System. Innovative Technology, Inc.
Innovative Technology, Inc. 2 New Pasture Road, Newburyport, MA 01950-4054 USA Email: Info@gloveboxes.com TEL: +1 978 462 4415 FAX: +1 978 462 3338 Manual Solvent Purification System Copyright Innovative
More informationAgilent 7890 / 220 Ion Trap GCMS Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationInstalling a Probe. Typical Installation
for hazardous locations. Prior to any installation in a Classified Hazardous Location, verify installation methods by the Control Drawing referenced on the product s name tag 2. The Feed Through Assemblies
More informationMarkes Thermal Desorption Systems Site Preparation Page 1 of 8
Thermal Desorption Systems Site Preparation Purpose of site preparation Your site must meet this specification to assure a successful and timely installation of your Agilent thermal desorption equipment.
More informationTA Instruments Installation Requirements for the Thermal Discovery Series
Notice TA Instruments Installation Requirements for the Thermal Discovery Series Thank you for ordering a Discovery Series thermal analysis system from TA Instruments. To ensure that installation of your
More informationbe present throughout these services; otherwise, they will miss important operational, maintenance and safety information.
Thank you for purchasing an Agilent Instrument. To get you started and to assure a successful and timely installation of your 9000 GC, please refer to this site prep checklist. Correct site preparation
More informationUltraLo 1800 Alpha Particle Counter
XIA LLC UltraLo 1800 Alpha Particle Counter Site Requirements & Planning Version: 0.4 Friday, February 15, 2013 Table of Contents I. Site Preparation Tasks... 3 A. Instrument Overview... 3 B. Installation
More informationCustomer Responsibilities. Important Customer Information. Agilent Infinity II Preparative HPLC System Site Preparation Checklist
Agilent Site Preparation Infinity II Checklist Preparative HPLC System Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer
More information28 Site Preparation. Supplying valve actuator air
28 Site Preparation Temperature and humidity ranges Ventilation requirements Venting oven exhaust Venting toxic or noxious gases Benchtop space requirements Electrical requirements Grounding Line voltage
More informationPlease refer to the Hydrogen Safety Guide which is shipped with the Instrument. 5977B Series GC/MS Systems Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationVarian 240-MS GC/MS Ion Trap Mass Spectrometer Pre-installation Instructions
Varian 240-MS GC/MS Ion Trap Mass Spectrometer Pre-installation Instructions Checklist NOTE: Do not unpack the boxes. Place a check in the box after satisfying each requirement as described in the instructions.
More informationDiscovery HP-TGA 75/750. Site Preparation Guide
Discovery HP-TGA 75/750 Site Preparation Guide Revision A Issued August 2018 Table of Contents Table of Contents... 2 Ideal Setup... 3 System Components... 4 Instrument Measurements... 5 Utility Requirements...
More informationSuper-Clean Gas Filters from Restek
Super-Clean Gas Filters from Restek Chromatography Products www.restek.com 800-356-688 84-353-300 regulator instrument-grade tubing on/off valve GC carrier gas moisture trap hydrocarbon trap high-capacity
More informationPegas 4000 MF Gas Mixer InstructionManual Columbus Instruments
Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...
More informationGeneral Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!
The portable Gas Chromatograph is now more versatile than ever. DPS Instruments is pleased to present the newest Portable Companion 2 Gas Chromatograph with room for 2 Detectors. The Companion 2 GC was
More informationCustomer Responsibilities
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationStorage Systems and Accessories
Storage Systems and Accessories High Pressure Air and Inert Gases 4500-6000 PSIG DOT and ASME Cylinders High Pressure Storage Systems If an application requires storage, either in the form of a single
More informationMASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS
MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS Using state-of-the-art measuring and microprocessor technologies, Advanced has assembled a series of systems which can measure mass
More informationEnvironmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!
After years of development and testing, DPS Instruments is pleased to present the newest, most expandable and versatile Gas Chromatography Systems in history. The DPS 600 Series GC systems are the world
More informationGas Management Supplies for GC Labs
Restek GC Accessories Gas Management Supplies for GC Labs Gas Generators Pressure Regulators Tubing and Fittings Gas Purifiers www.restek.com Gas Management for Your Lab A clean gas stream is critical
More informationCustomer Responsibilities. Important Customer Information. Agilent 6400 Series Triple Quad LC/MS Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationMATHESON s LaserPro Gases and Gas Delivery Equipment
MATHESON s LaserPro Gases and Gas Delivery Equipment MATHESON s LaserPro Support Gases and Services MATHESON delivers the LaserPro line of gases specifically designed to meet the stringent purity requirements
More informationCustomer Responsibilities. Important Customer Information. Cary 4000/5000/6000i UV-Vis spectrophotometer Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationAgilent 7820A GC. Site Preparation Guide. Agilent Technologies
Agilent 7820A GC Site Preparation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2010-2015 No part of this manual may be reproduced in any form or by any means (including electronic storage
More informationCryogenic Liquid Ultimate Liquid Cylinder Hose
Cryogenic Liquid Ultimate Liquid Cylinder Hose Flexible Components brand liquid cylinder hose is considered the highest quality liquid cylinder hose on the market. It effectively transfers cryogenic liquids
More informationFeatures. Main Voltage: Vdc. Input Control Signal: 0-5 Vdc or 4-20 madc (Sourcing) Monitor Output Voltage: 0-5 Vdc or 4-20 madc (Sourcing)
1 X-Flow Mass Flow Controller Flow Range from 0.8 ml/min to 20 l/min Target Markets Laboratory and Process Instrumentation Pharma / Bio-Pharma Equipment Air Quality Monitoring Systems Furnace and Coatings
More informationPressure Dump Valve Service Kit for Series 3000 Units
Instruction Sheet Pressure Dump Valve Service Kit for Series 000 Units. Overview The Nordson pressure dump valve is used to relieve hydraulic pressure instantly in Series 00, 400, 500, and 700 applicator
More informationPreparing Your Laboratory for the Optima 7300 V and 8300 ICP-OES Spectrometers
LABORATORY PREPARATION Preparing Your Laboratory for the Optima 7300 V and 8300 ICP-OES Spectrometers TABLE OF CONTENTS Section Subject 1 Suitable Working Area 2 Exhaust Vent 3 Vent Positions 4 Handling
More informationTA Instruments Installation Requirements for Thermal Analysis Systems
TA Instruments Installation Requirements for Thermal Analysis Systems Thank you for ordering a Thermal Analysis system from TA Instruments. To ensure that installation of that system goes as smoothly as
More informationAPPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT
APPLICATION NOTE Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT The INFICON 3000 Micro GC provides two fast and accurate solutions for the analysis of coal mine gas components.
More informationDiscovery SDT 650. Site Preparation Guide. Revision G Issued December Page 1
Discovery SDT 650 Site Preparation Guide Revision G Issued December 2018 Page 1 Table of Contents Table of Contents... 2 Ideal Setup... 3 System Components... 4 With Discovery Mass Spectrometer... 4 With
More informationHow to order guide. Spectronus FTIR Trace Gas Analyser. Version: 1.0
How to order guide Spectronus FTIR Trace Gas Analyser Version: 1.0 www.ecotech.com How to order guide Spectronus FTIR Tr ace Gas Analyser 1.0 This page is intentionally blank Page 2 Table of Contents Table
More informationWHAT ARE THE SPECIFICATIONS FOR THE ARGON GAS OPTION USED IN JOULE THOMSON REFRIGERATORS?
MMR Technologies, Inc. WHAT ARE THE SPECIFICATIONS FOR THE ARGON GAS OPTION USED IN JOULE THOMSON REFRIGERATORS? Joule Thomson refrigerators operate using high pressure gas to cool a sample stage with
More informationThermo Scientific FOCUS GC. Gas Chromatograph Site Preparation and Installation Manual. PN , Revision May 2010
Thermo Scientific FOCUS GC Gas Chromatograph Site Preparation and Installation Manual PN 317 094 11, Revision May 2010 FOCUS GC Site Preparation and Installation Manual May 2010 Edition Part Number 317
More informationTM8 On-Line Gas Chromatography DGA Monitor Site Preparation Guide Model TMx Series
TMx Site Preparation Guide 810-1647-03 Rev B TM8 On-Line Gas Chromatography DGA Monitor Site Preparation Guide Model TMx Series June, 2017 Document ID: 810-1647-03 Rev B www.serveron.com Legal Information
More informationAvio 500 ICP-OES. ICP-Optical Emission Spectroscopy PREPARING YOUR LAB. Preparation Checklist. Suitable Working Area
PREPARING YOUR LAB Avio 500 ICP-OES ICP-Optical Emission Spectroscopy Preparation Checklist Suitable Working Area Exhaust Vent Vent Positions Handling of Gas Cylinders and Other Suggested Safety Practices
More informationThermo Scientific. TRACE GC Ultra. Gas Chromatograph Site Preparation and Installation Manual. PN , Revision May 2010
Thermo Scientific TRACE GC Ultra Gas Chromatograph Site Preparation and Installation Manual PN 31709190, Revision May 2010 TRACE GC Ultra Site Preparation and Installation Manual May 2010 Edition Part
More informationPre-installation Checklist
Pre-installation Checklist Varian 1200/1200L Mass Spectrometer Use this checklist as a guide while preparing your site for the installation of the Varian 1200 or 1200L MS system. Make sure that your site
More informationAgilent 5975T LTM GC/MSD
Agilent 5975T LTM GC/MSD Site Preparation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2010 No part of this manual may be reproduced in any form or by any means (including electronic storage
More informationCustomer Responsibilities
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationVarian 325-MS LC/MS Quadrupole Mass Spectrometer Pre-installation Instructions
Varian 325-MS LC/MS Quadrupole Mass Spectrometer Pre-installation Instructions Checklist NOTE: Do not unpack the shipping cartons. Check off each checklist box after satisfying each requirement as described
More informationReview of the Hall B Gas System Hardware. George Jacobs
of the Hardware George Jacobs DSG Staff 2 Hall B Gas Utilities for detectors Drift Chamber (DC) Low Threshold Cherenkov Counter (LTCC) Micromegas Vertex Tracker (MVT) Forward Tagger (FT) Ring Imaging Cherenkov
More informationAgilent Intuvo 9000 Gas Chromatograph
Agilent Intuvo 9000 Gas Chromatograph Installation and First Startup Agilent Technologies Notices Agilent Technologies, Inc. 2016 No part of this manual may be reproduced in any form or by any means (including
More informationGas delivery systems. How to ensure uninterrupted high quality gas delivery from the source to the apparatus? Dr. Gastons Vereskuns Trakai,Lithuania
Gas delivery systems. How to ensure uninterrupted high quality gas delivery from the source to the apparatus? Dr. Gastons Vereskuns Trakai,Lithuania May 7, 2014 What it is gas delivery system? It is a
More information21700 Nordhoff St. Chatsworth, CA Phone
21700 Nordhoff St. Chatsworth, CA. 91311 Phone 818.882.2331 DILUTION SYSTEM DILUTION SYSTEM ~ 1 ~ 09 21 07 TABLE OF CONTENTS DEFINITIONS.. 3 INTRODUCTION.. 4 INSTALLATION & SETUP 5 CONVERSION TO PRESSURE
More informationAir Operated Hydraulic Pumping Systems to 50,000 psi
High Pressure Equipment Air Operated Hydraulic Pumping Systems to 50,000 psi PS-10: 10,000 psi PS-20: 20,000 psi PS-30: 30,000 psi PS-40: 40,000 psi PS-50: 50,000 psi PS-90: 90,000 psi High Pressure air
More informationGasTrak Pro TM Delivery/Control Systems
GasTrak Pro TM Delivery/Control Systems Regulator, Purifier and Outlet Sections Introduction GasTrak Pro TM Systems are a unique approach to controlling and delivering high-purity gases distributed in
More informationPressure Dump Valve Service Kit for Series 2300 Units
Instruction Sheet Pressure Dump Valve Service Kit for Series 00 Units. Overview The Nordson pressure dump valve is used to relieve hydraulic pressure instantly in Series 00 applicator tanks when the unit
More informationNonrefillable Cylinders
Nonrefillable Cylinders A DIVISION OF NORCO, INC. Norlab, the Specialty Gas Division of Norco, Inc., has been producing and distributing calibration mixtures and pure specialty gases in refillable and
More information780S Ultra High Purity Series
Smart Ultra High Purity Thermal Gas Mass Flow Meter Features Measures mass flow directly, no seperate temperature or pressure inputs required Field adjustment of critical flow meter settings via on-board
More informationAC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW
AC 2010-2145: MEASUREMENT OF HYDROGEN IN HELIUM FLOW Randy Buchanan, University of Southern Mississippi Christopher Winstead, University of Southern Mississippi Anton Netchaev, University of Southern Mississippi
More informationTITAN Oxygen Regulators
EMS OXYGEN REGULATORS TITAN Oxygen Regulators EMS TITAN SERIES OXYGEN EMS AND FIRE SERVICE APPLICATIONS CLICK-STYLE REGULATOR Whether its EMS, ambulance, fire rescue or tactical medical use, Titan Oxygen
More informationNexION 1000/2000 ICP-MS
PREPARING YOUR LAB NexION 1000/2000 ICP-MS ICP - Mass Spectrometry Preparation Checklist Environmental conditions Electrical requirements Space requirements Exhaust ventilation Coolant requirements Argon
More informationPreinstallation Checklist and Instructions for the TriStar These Preinstallation Checklist and Instructions were reviewed and approved by:
Title: Preinstallation Checklist and Instructions for TriStar 3000 Date: 08/04 300/42870/01 Revision ( A ) MICROMERITICS Preinstallation Checklist and Instructions for the TriStar 3000 These Preinstallation
More informationATIS OPERATION MANUAL
ATIS OPERATION MANUAL Adsorbent Tube Injector System T702019A 1 Introduction The Adsorbent Tube Injector System (ATIS) is a sample preparation device that provides a quick and efficient way to transfer
More information6200E Installation Requirements
Requirements When purchasing a standalone 6200E unit with the calibrator, the analyzer will be supplied as three 19 rack mount units. Analyzer Converter Calibrator Included with the unit are two lengths
More informationMB-SP Series Solvent Purification System
Operation Manual MB-SP Series Solvent Purification System Table of contents 1. System overview 2 2. Column activation.. 5 3. System purging.. 6 4. Glove box integration.... 7 5. Solvent reservoir installation....
More informationInstallation, Operation, and Maintenance Manual
Installation, Operation, and Maintenance Manual Welker Probe Instrument Regulator Model The information in this manual has been carefully checked for accuracy and is intended to be used as a guide for
More informationLUMAZOTE T ECHNICAL BRO CHU RE
by Chemiluminescence Standards AST D4629 / ASTM D6069 / ASTM D 5176 / ASTM D 5762 / NF EN 12260 / NF M 07-058 Licence TOTAL LUMAZOTE T ECHNICAL BRO CHU RE ERALY & Associés Phone: +33 (0) 1 77 04 80 97
More informationMass Flowmeters. Mass Flowmeters for Gases
Mass Flowmeters Mass Flowmeters for Gases Fast, Accurate, Low Pressure Drop for critical measurement applications! Flow (Std l/min) Response to a step change in flowrate 300 200 100 4 Milliseconds 63%
More informationDiscovery TGA 55, TGA 550, TGA Site Preparation Guide
Discovery TGA 55, TGA 550, TGA 5500 Site Preparation Guide Revision E Issued March 2018 Table of Contents Table of Contents... 2 Ideal Setup... 3 System Components... 4 With Discovery Mass Spectrometer...
More informationSpecialty Gases & Equipment ISO 9001: 2008 CERTIFIED. A Division of Norco, Inc.
Specialty Gases & Equipment ISO 9001: 2008 CERTIFIED A Division of Norco, Inc. Located in Boise Idaho, Norlab has been producing and distributing calibration gas mixtures and specialty gases in refillable
More informationProcess Micro-chromatograph PGC 490 ATEX explosion-proof rapid gas analyzer General user manual
Process Micro-chromatograph PGC 490 ATEX explosion-proof rapid gas analyzer General user manual p. 1 / 26 Table of contents 1. INTRODUCTION 4 2. SAFETY INSTRUCTIONS 4 2.1 Important general information
More informationTA Instruments Installation Requirements for Thermal Analysis Systems
TA Instruments Installation Requirements for Thermal Analysis Systems Notice The material contained in this manual, and in the online help for the software used to support TA Instruments products, is believed
More informationVOLUMATIC WATER REGULATOR
VOLUMATIC WATER REGULATOR For Cage Systems - PDS and Manual Adjustment Part Number 00-, 00-, 00- Introduction The Chore-Time VOLUMATIC Water Regulator is designed to regulate the water pressure for the
More informationGenerating Calibration Gas Standards
Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and
More informationBaffles and Traps Cryotrap
Baffles and Traps 362-6 Cryotrap Varian Low-Profile Water-Cooled Baffles combine 100% optical density with high conductance and unusually low overall height. They are especially useful in applications
More informationGAS DETECTION: CALIBRATION GAS
ENVIRONMENTAL SURVEY/PURGING MONITORS GAS DETECTION: CALIBRATION GAS Gas Mixtures Non-Reactive Single & Two-Gas Mixtures 98 Reactive Single & Two-Gas Mixtures 98 Non-Reactive Multi Mixtures 99 Reactive
More informationEnergy Management solutions for compressed air and industrial gases. vpinstruments.com
Energy Management solutions for compressed air and industrial gases vpinstruments.com 6 Introduction VPINSTRUMENTS Measure, Discover and Save! Compressed air is a very expensive energy source. In fact,
More informationInstructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems
Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the
More informationAmmonia Treatment Instructions Page 1 of 5
Page 1 of 5 Original version by Douglas Call Updated January 21, 2009 by David Jones Department: Chemical Engineering Location: Fenske 102 (access thru room 103) Contact: BP Matthew Holbrook 161 Fenske
More informationAgilent 490 Micro GC Natural Gas Analyzer
Agilent 490 Micro GC Natural Gas Analyzer User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by any means (including electronic
More informationNGN Series nitrogen generator
NGN Series 8-64 nitrogen generator Applications how it works RK series 5-40 F Blanketing of Chemicals and Pharmaceuticals Gas Assisted Injection Moulding (GAIM) Laser Cutting Food Heat Treament of Ferrous
More informationAgilent GC, GC/MS, and ALS
Agilent GC, GC/MS, and ALS Site Preparation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2017 No part of this manual may be reproduced in any form or by any means (including electronic
More informationBGA244 Binary Gas Analyzer
Quick Start Guide Revision 1.0 Certification Warranty Service certifies that this product met its published specification at the time of shipment. This product is warranted against defects in materials
More informationGas Absorption Draft Standard Operating Procedure
Gas Absorption Draft Standard Operating Procedure R.C. 9-14 Scope Page 1 Equipment Overview Page 1 Safety Page 2 Operating Procedure Page 4 Equipment Description Page 16 Gas Absorption Experiment Draft
More informationIRwin Methane Leak Detector. Innovation makes the difference
IRwin Methane Leak Detector Innovation makes the difference Innovative methane detector system for easy survey of gas pipelines IRwin is a portable methane leak detector for fast and easy survey of gas
More informationTechnical Information Conducal CLY421
TI00496C/07/EN/01.12 71142150 Products Solutions Services Technical Information Conducal CLY421 Conductivity calibration set for ultrapure water applications Factory calibrated measurement box Certified
More informationAgilent G3180B Two-Way Splitter Kit
Agilent G3180B Two-Way Splitter Kit With Makeup Gas Installation and Operation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2006 No part of this manual may be reproduced in any form or
More informationInstruction Manual For Model Equilibrium Flash Separator P/N: D01 Revision E February 2016
Instruction Manual For Model 2353-805 Equilibrium Flash Separator P/N: 2353-1D01 Revision E February 2016 S/N: 2001 N. Indianwood Ave. Broken Arrow, Oklahoma 74012 U.S.A. Telephone: 918-250-7200 Fax: 918-459-0165
More informationSPECIFICATIONS. Approximate Weight: 10 oz. Surface Finish: Ra micro inch or less MATERIALS OF CONSTRUCTION
DIFFUSION-RESISTANT, DIAPHRAGM SEAL, EXCESS FLOW SHUT-OFF VALVES (FS SERIES) FS Series Excess Flow Shut-Off Valves are designed to automatically shut-off the delivery of gas in a line if the flow exceeds
More informationTECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1
Dry 131h 1. DESCRIPTION The Viking Model G-1 Maintenance Air Compressor is an electric motor-driven, aircooled, single-stage, oil-less compressor. The unit is equipped with a check valve and provides a
More informationRunning HPLC Methods on an ACQUITY UPLC System
Running HPLC Methods on an ACQUITY UPLC System Note: This document is an addendum to the ACQUITY UPLC System Operator s Guide. Contents Topic Page System considerations 1-1 Choosing fittings 1-3 System
More informationMEDICAL EQUIPMENT CATALOG
THE MOST TRUSTED MEDICAL EQUIPMENT IN NORTH AMERICA MEDICAL EQUIPMENT CATALOG Total Components Including Bulk Oxygen Supply, Back up and Delivery Equipment Designs for High Pressure, Liquid or Bulk Applications
More informationSAFETY MANUAL FOR FLAMMABLE PRODUCT TRANSFER
SAFETY MANUAL FOR FLAMMABLE PRODUCT TRANSFER SUPPLIMENT TO eom IMPORTANT READ THIS MANUAL BEFORE PRODUCT INSTALLATION, OPERATION, INSPECTION & MAINTENANCE Tougher and more rigid guidelines are being established
More informationNHR Powder Feed Hopper
Instruction Sheet P/N 0804A0 NHR-8-50 Powder Feed Hopper Description This instruction sheet provides specifications, installation and operation instructions, and parts lists for the NHR-8-50 powder feed
More information