GUIDANCE FOR THE APPLICATION OF EQUIPMENT CALIBRATION AND TRACEABILITY OF MEASUREMENTS ON INSPECTION BODIES

Similar documents
Supersedes: The copy of this document located on Measurement Canada s website is considered to be the controlled copy.

EN ISO/IEC Technical Requirements

tradition. knowledge. responsibility. CALIBRATION OF ELECTRICAL MEASUREMENT EQUIPMENT CALIBRATION LABORATORY

General Accreditation Criteria Reference Material Producers. ISO Standard Application Document

CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE CO 2 PRESSURE. (+34)

Service ISO17025 explained

TG GUIDELINES CONCERNING CALIBRATION INTERVALS AND RECALIBRATION

Stability and traceability of calibration gases for EN 14181

Traceable calibration of automatic weighing instruments in dynamic operation

Accreditation for In-Service Inspection of Transportable Pressure Receptacles (TPRs)

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Date of Shipment: Xxxxx 00, 20xx 3402c National Institute of Advanced Industrial Science and Technology National Metrology Institute of

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience -

Keeping It Simple. Gas Mixtures for Stack Emissions Monitoring.

MET-101 Introduction to Metrology

Certificate of Accreditation

Bi-lateral Comparison (APMP.L-K ) Calibration of Step Gauge

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Certified Reference Materials FOR REFRACTOMETERS AND POLARIMETERS

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

GM/GN2646. Guidance on Axle Bearing Maintenance. Issue One: March 2011 Rail Industry Guidance Note for GM/RT2004 Issue Four

Recommended Practices for Calibration Laboratories

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 ANSI/NCSL Z

ISO Vacuum gauges Calibration by direct comparison with a reference gauge

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Procedure: Pressure equipment safety

CERTIFICATE OF ACCREDITATION ISO/IEC 17025:2005 ANSI/NCSL Z (R2002)

Light Lab. Reliable Measurements through Regular Calibration. DAkkS

Requirements for the certification of movements and mechanical watches resistant to magnetic fields of 1.5 T ( G)

Gas mixtures. Individual solutions specifically for your application

Accreditation for In-Service Inspection of Pressure Systems/Equipment

Interlaboratory comparison in the pressure range from 0 to 2 MPa for accredited calibration laboratories

Gas Metrology. SI-traceable: Primary Reference Materials Certified Reference Materials Calibrated Gas Mixtures Interlaboratory Comparisons

CALIBRATION CERTIFICATE # 500

GMC-I Service GmbH Calibration Center

INSPECTORATE FOR DIVING EQUIPMENT SERVICING & TESTING CODE OF PRACTICE CP11:2011

SFHGEN77 - SQA Code HD48 04 Perform first line calibration on clinical equipment to ensure it is fit for use

Balancing and Settlement Code. Code of Practice Four

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

KENYA ACCREDITATION SERVICE

INSPECTION FOCUS ON:

BRML INM. National Metrology Institute. COOMET Comparison COOMET.L-S20 676/MD/15. Calibration of gauge blocks by comparison. Technical protocol

CALIBRATION BOOK DOC

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

CALIBRATION SYSTEM REQUIREMENTS. ESCC Basic Specification No

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. PRECISION MEASURING CORP Groesbeck Hwy. Fraser, MI Robert Meyer Phone:

Maximum precision for quantifiable success.

AUSTRALIAN SAILING POLICY SPO National Officiating Program. Australian Sailing Race Officer Accreditation. 1 Executive Summary...

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Research and Development. Money in the Pipeline

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

REFERENCE SPECIFICATION FOR END

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z TROXLER ELECTRONICS LABORATORIES, INC. James Pratt Phone:

COOMET TECHNICAL COMMITTEE 1.8 PHYSICAL CHEMISTRY

Systems of Accounting for and Control of Nuclear Material

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z CALIBRATION

Accreditation of laboratories for the performance evaluation of CMMs according to the ISO

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

Title: INSTRUMENT AND EQUIPMENT DOCUMENTATION AND RECORDS

UKAS calibration laboratory 8809 Temperature, pressure and electrical signals calibration guide

Recalibration of gas meters

Any use of this material except in accordance with a written agreement with Western Power is prohibited.

Transferpette -8/-12

Level 3 Diploma in Moving Loads in Construction. Qualification Specification

Speciality Gases Practice 4. Our quality, your safety. Accredited Test and Calibration Laboratory Speciality Gases Centre in Hörstel.

Primary Standardisations and Traceability for Radioactivity supporting Nuclear Medicine John Keightley

GAS TRANSPORTER S REQUIREMENTS FOR GAS MEASUREMENT SYSTEMS CONNECTED TO THE NATIONAL GRID GAS NETWORK

General Accreditation Guidance. User checks and maintenance of laboratory balances

Control of Legionella

2014 Programme of Training Courses in Measurement & Calibration

Effective Date: May 2017 Void After: May Product: Press and Nail Connections File No. M-8332

PIQCS HACCP Minimum Certification Standards

AUSTRALIAN SAILING POLICY SPO National Officiating Program. Australian Sailing Judge Accreditation. 1 Executive Summary...

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

INSPECTION, TESTING AND CERTIFICATION OF GASFITTING WORK DONE UNDER SUPERVISION

Pressure Equipment Directive PED 2014/68/EU Commission's Working Group "Pressure"

COOMET TECHNICAL COMMITTEE 1.8 PHYSICAL CHEMISTRY. CООМЕТ project 576/RU/12

NZQA registered unit standard version 3 Page 1 of 7. Plan electrical installations for explosive atmospheres

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

Improvement of reliability in pressure measurements and international mutual recognition

NZQA registered unit standard version 1 Page 1 of 8. Conduct detailed inspection of electrical installations for explosive atmospheres

Gas Safety Management Policy

The IEC61508 Inspection and QA Engineer s hymn sheet

SOUTH AFRICAN NATIONAL STANDARD

Suction fittings for use in swimming pools, wading pools, spas, hot tubs, and other aquatic facilities.

Significant Change to Dairy Heat Treatment Equipment and Systems

RJ Kendall Pty Ltd. Building & Engineering Consultant. 5 th May Peter Fitzpatrick Director of Properties Diocese of Wagga Wagga

CRITERIA FOR LABORATORY ACCREDITATION IN MASS METROLOGY

SUP 15 Health & Safety Management Pressure Systems. Unified procedures for use within NHS Scotland

oliet gaso danmark Norsk olje &qass Guidelines for Mutual Recognition of Specialised Safety and Emergency Response Training for North Sea Operations

Certificate of Accreditation

Pressure Metrology in Mexico

Standard Pneumatic Test Procedure Requirements for Piping Systems

Proficiency Testing for Surface Roughness Standard and Groove Depth Standard

INTERNATIONAL STANDARD

NZQA registered unit standard version 2 Page 1 of 7. Conduct a conformity assessment of explosion-protected apparatus

How we calibrate the Wave Height and Period Measurements from the Gravitational Acceleration Wave Buoys in RMIC/AP

Safety assessments for Aerodromes (Chapter 3 of the PANS-Aerodromes, 1 st ed)

1995 Metric CSJ SPECIAL PROVISION ITEM 7. Legal Relations And Responsibilities To The Public

Transcription:

GUIDANCE FOR THE APPLICATION OF EQUIPMENT CALIBRATION AND TRACEABILITY OF MEASUREMENTS ON INSPECTION BODIES ESYD G-DIAKRIB/01/01/26-03-2015 1/6

ESYD G-DIAKRIB Issue: 01 Revision: 01 Issue Date: 28-11-2007 Revision Date: 26-03-2015 Prepared by: The Quality Manager of ESYD Approved by: The Head of ESYD Quality Manager of ESYD Head of ESYD ESYD G-DIAKRIB/01/01/26-03-2015 2/6

An Inspection Body (IB), where appropriate, shall perform calibration of its measurement equipment, having a significant influence on the results of the inspections performed by the IB, before being put into service, and thereafter calibrate it according to an established programme (ELOT EN ISO/IEC 17020:2012, 6.2.6). The calibration may be performed with one of the ways presented below: A. Internal calibration (in house), thus the following must apply: 1. IBs that perform internal calibrations of their measurement equipment must fulfill the requirements listed below, according to the standards ELOT EN ISO/IEC 17025, ELOT EN ISO/IEC 17020, the guide ILAC P15:06/2014 ( 6.2.7a) and the guidance ILAC-P10:01/2013, for the performance of these calibrations: a) The IBs must have and apply documented procedures for the calibration or/and check of the equipment, as well as records of calibration method validation, in case of using non-standard calibration methods, b) The IBs must have a proper procedure for the calculation of uncertainty of measurements, taking into account all components, c) The personnel that perform the internal calibrations must have the proper training, d) The IBs must have appropriate equipment and environmental conditions, e) The IBs must have measurement reference standards calibrated by: - National Metrology Laboratory, which provides services appropriate for intended use, covered by mutual recognition agreement CIPM MRA or - Accredited calibration laboratory, having Scope of Accreditation (SoA) appropriate for intended use, by ESYD or other Accreditation Body, signatory to ILAC mutual recognition agreement or local agreements recognized by ILAC, f) have internal and external (participation in programmes of interlaboratory comparisons) quality control results. 2. The IB must inform ESYD, either with its Application Form or relevant letter, that it performs internal calibrations of its equipment in specific quantities. 3. The IB that performs internal calibrations of its equipment must submit to ESYD the relevant procedures, including the procedures for the calculation of uncertainty of measurements. 4. The assessment of the IB s competence at performing internal calibrations is mandatory in the initial assessment, for the granting of accreditation. 5. During the initial assessment of an IB, there must be included in the assessment team either an ESYD Bodies Assessor, properly trained on the requirements of ELOT EN ISO/IEC 17025, or an ESYD Calibrations Assessor as a Technical Expert. ESYD G-DIAKRIB/01/01/26-03-2015 3/6

6. In the special case that the IB performs internal calibrations, which have not been previously stated and therefore the assessment team is not technically competent to assess them, the Lead Assessor informs ESYD to modify the assessment team by adding an appropriate Assessor, who performs a complementary assessment. 7. Any modifications in the calibration procedures must be communicated to ESYD and can be assessed on-site, in the frame of the next surveillance visit, on the basis of a relevant ESYD decision. 8. The assessment of the IB s competence at performing internal calibrations, in the frame of surveillance and reassessment visits, must be conducted either by ESYD Assessors for the standard ELOT EN ISO/IEC 17020, on condition that they have been previously trained on the requirements of ELOT EN ISO/IEC 17025, or by an ESYD Calibrations Assessor as a Technical Expert. 9. Regarding already accredited IBs performing internal calibrations, an assessment of their competence at performing internal calibrations must be conducted in the frame of the next surveillance visit, either by an ESYD Bodies Assessor, properly trained on the requirements of ELOT EN ISO/IEC 17025, or by an ESYD Calibrations Assessor as a Technical Expert. B. External calibration, thus: External suppliers providing calibration services are used. There are four (4) separate cases, according to the guide ILAC P15:06/2014 ( 6.2.7b) and the guidance ILAC-P10:01/2013: 1. The supplier used is National Metrology Laboratory, which provides services appropriate for intended use, covered by mutual recognition agreement CIPM MRA. 2. The supplier used is accredited according to ELOT EN ISO/IEC 17025, in the specific technical field where the equipment calibration service is requested, by ESYD or other Accreditation Body, signatory to ILAC mutual recognition agreement or local agreements recognized by ILAC. Accreditation is a sufficient condition for co-operation between the IB and the provider of calibration services, which assures quality of services, provided that copies of the supplier s scope of accreditation are maintained in the IB s records. 3. The supplier used is National Metrology Laboratory, which provides services appropriate for intended use, but not covered by mutual recognition agreement CIPM MRA. ESYD G-DIAKRIB/01/01/26-03-2015 4/6

4. The supplier used is not accredited according to ELOT EN ISO/IEC 17025, by ESYD or other Accreditation Body, signatory to ILAC mutual recognition agreement or local agreements recognized by ILAC, in the specific technical field where the equipment calibration service is requested. The cases 3 and 4 are selected only if cases 1 and 2 are not available, either in Greece or abroad. Moreover, if selecting cases 3 and 4, the IB shall document the compliance of the supplier / calibration laboratory with ESYD traceability requirements. Appropriate evidence of compliance includes at least the following (numbers in the parentheses refer to clauses in ELOT EN ISO/IEC 17025:2005): - Calibration of the standards of the supplier / laboratory by (5.6.1): - National Metrology Laboratory, which provides services appropriate for intended use, covered by mutual recognition agreement CIPM MRA or - Accredited calibration laboratory, having SoA appropriate for intended use, by ESYD or other Accreditation Body, signatory to ILAC mutual recognition agreement or local agreements recognized by ILAC, - Results of participation in programmes of interlaboratory comparisons (5.9.1) - Appropriate equipment and environmental conditions (5.5.1 & 5.3) - Standard calibration methods or records of non-standard calibration methods validation (5.4.2 & 5.4.5) - Procedures for estimation of uncertainty (5.4.6) - Documentation for assuring the quality of calibration results (5.9) - Documentation for competence of staff (5.2) - Audits of the calibration laboratory (4.6.4 & 4.14) - The contents of the calibration certificate (5.10) For the documentation of all the previously mentioned, it may be necessary for the IB to perform a practical assessment of the supplier / laboratory providing calibration services (second party assessment), similar to that which would be undertaken by an Accreditation Body, to evaluate the technical competence of the supplier / calibration laboratory. Especially for aforementioned cases 3 and 4, an assessment of the competence of the IB to evaluate its supplier / calibration laboratory will take place, by an ESYD Bodies Assessor, properly trained on the requirements of ELOT EN ISO/IEC 17025, or an ESYD Calibrations Assessor. ESYD G-DIAKRIB/01/01/26-03-2015 5/6

Relevant findings must be assessed and, if necessary, the evaluation conducted by the IB to its supplier / calibration laboratory must be witnessed on-site. For this reason, during the initial assessment of an IB, there must be included in the assessment team either an ESYD Bodies Assessor, properly trained on the requirements of ELOT EN ISO/IEC 17025, or an ESYD Calibrations Assessor as a Technical Expert. In the frame of surveillance and reassessment visits of IBs, the evaluation by the IB of the provider of calibration services must be subject to assessment either by ESYD Bodies Assessors, on condition that they have been previously trained on the requirements of ELOT EN ISO/IEC 17025, or by an ESYD Calibrations Assessor as a Technical Expert. Regarding already accredited IBs, an assessment of their competence at evaluating their suppliers of calibration services must be conducted in the frame of the next surveillance visit, either by an ESYD Bodies Assessor, properly trained on the requirements of ELOT EN ISO/IEC 17025, or by an ESYD Calibrations Assessor as a Technical Expert. ESYD G-DIAKRIB/01/01/26-03-2015 6/6