1 1 of 7 7/11/2014 3:39 PM May 1, 2010 By John Heim,Doug Staples Anabolic steroid screening analysis in urine is complex and labor intensive requiring sensitive instrumentation and optimized chromatographic separations. This research presents the practical application of comprehensive two-dimensional gas chromatography time-of-flight-mass spectrometry (GC GC TOF-MS) for the identification and quantification of five androgenic anabolic steroids in urine. Conventional methods for steroid analysis in urine rely heavily on one-dimensional GC separations and selected ion monitoring (SIM) MS methods. This study utilizes GC GC to increase peak capacity and resolution in combination with time-of-flight mass spectrometry (TOF-MS) detection followed by data processing with deconvolution software algorithms for positive confirmation of anabolic steroids in urine. A steroid mixture containing stanozolol (Winstrol), 4-hydroxystanozolol, boldenone, 19-norandrosterone, 17 -methylandrostan diol, and 3-hydroxystanozolol was prepared from commercial standards. Experimental results for the identification of 3-hydroxystanozolol at the 2-ppb level are presented. The illegal anabolic steroid stanozolol and metabolites are known to be particularly difficult to detect and separate chromatographically. Methyltestosterone was used as an internal standard (ISTD). Stanozolol was not included in the calibration curve development due to poor chromatographic response below 10 ppb. Although urine sampling is simple and easy to obtain, it poses a variety of challenges for the laboratory analyst. Sample preparation is labor intensive, complex, and difficult to reproduce consistently. Urine is a complicated biochemical mixture. The matrix effects from urine often can obscure detection of trace-level steroids and their metabolites. Instrumentation must be able to provide absolute confirmation of the analytical results. Sample preparation followed a well established extraction and derivatization procedure for antidoping control. Results from this study show significant improvements in chromatographic resolution and peak capacity, as well as the enhanced detectability that GC GC TOF-MS provides for antidoping control screening. Successful trace level identifications of the five steroid standards mixture will be shown at the 2-ng/mL (2-ppb) level. This exploratory research investigation demonstrates favorable and practical applicability of GC GC TOF-MS for the positive identification of anabolic steroids at the lowest allowable concentration limits which meet the strict guidelines set by the World Anti-Doping Agency (WADA). The increased peak capacity and enhanced chromatographic resolution of GC GC coupled with a fast acquisition TOF-MS, up to 500 Hz, is essential for the successful acquisition and analysis of the data density needed to characterize low levels of steroids fully in complex sample matrices such as urine. These data-rich files are processed with deconvolution algorithms, which deliver qualitative identification as well as a multiple compound quantification in a single run. The results show limitof-detection values at or below 2 ppb for five anabolic steroids with a calibration linearity of greater than 99.9%. Experimental Initial studies focused on the development of the sample acid hydrolysis and derivatization procedure, in addition to chromatographic (GC GC) and MS (TOF-MS) method optimization. Sample preparation was conducted with established protocols for steroid analysis in urine (1). Steroid reference standards were purchased from (Cerilliant,

2 2 of 7 7/11/2014 3:39 PM Round Rock, Texas) and (Alltech-Applied Science Labs, Deerfield, Illinois). A five-component steroid standards mixture was prepared in methanol. Serial dilutions were made from the steroid standards mixture and then spiked into 2-mL urine aliquots to create a working five point calibration at 2, 10, 20, 50, and 100 ng/ml. Methyltestosterone was spiked into the urine matrix at 200 ng/ml as an internal standard. Acid hydrolysis with beta-glucuronidase was carried out before solvent extraction with methyl- tert-butyl-ether. Extracts were dried completely using nitrogen evaporation. Trimethylsilyl-derivatization was conducted with a commercially available MSTFA-ammonium-iodide-ethanethiol mixture from (Sigma-Aldrich, Saint Louis, Missouri). A 3-µL pulsed splitless injection was used to analyze each sample immediately after derivatization. GC GC TOF-MS results were generated with a Pegasus 4D time-of-flight mass spectrometer (LECO, St. Joseph, Michigan). The Pegasus 4D GC GC TOF-MS instrument was equipped with a model 7890 gas chromatograph (Agilent Technologies, Santa Clara, California) featuring a two-stage cryogenic modulator and a secondary oven (LECO). ChromaTOF software (LECO) was used for all acquisition control, data processing, and calibration curve development. A 30 m 0.25 mm, 0.25-µm d f Rxi-5ms, (Restek Corp., Bellefonte, Pennsylvania) GC capillary column was used as the primary column for the GC GC TOF-MS analysis. In the GC GC configuration, a second column (1.2 m 0.18 mm, 0.20-µm d f BPX-50, (SGE Analytical Science, Austin, Texas) was placed inside the secondary GC oven after the thermal modulator. The helium carrier gas flow rate was set to 1.5 ml/min at a corrected constant flow via pressure ramps. The primary column was programmed with an initial temperature of 140 C for 0.20 min then ramped at 20 C/min to 170 C, next ramped at 5 C/min to 260 C for 2.0 min, with a final ramp at 10 C/min to 315 C for 12.0 min. The secondary column temperature program was set to an initial temperature of 145 C for 0.20 min then ramped at 20 C/min to 175 C, next ramped at 5 C/min to 265 C for 2.0 min, with a final ramp at 10 C/min to 320 C for 12.0 min. The thermal modulator was set to +20 C relative to the primary oven and a modulation time of 4 s was used. The GC GC TOF-MS analysis total run time was 39.2 min. The MS mass range was m/z with an acquisition rate of 100 spectra/s. The ion source chamber was set to 230 C and the detector voltage was 1950 V with electron energy of 70 ev. GC GC TOF-MS analysis was conducted immediately after sample derivatization. Subsequently, calibration curves were developed and loaded into the software for automated data processing and quantification purposes. Method development achieved an optimized GC GC TOF-MS analysis capable of effectively separating the five anabolic steroids in complex urine matrix at or below 2-ppb detection limits. Results and Discussion Figure 1 Anabolic Steroid Identification: Successful trace level identifications of the five steroid standards mixture was achieved at 2 ng/ml with library search matches of greater than 60%. Calibration curve linearity of 99.9% or greater was attained. Over 5000 peaks per sample were detected for this analysis. Figure 1 displays the two-dimensional contour plot chromatogram of the 20-ng/mL standard GC GC TOF-MS analysis. The highlighted peaks show the labeled spiked steroids in urine throughout the chromatographic plane. The two-dimensional contour plot chromatogram clearly shows the ability of GC GC to maximize separation and resolution of anabolic steroids from the complex urine matrix in both the first and second dimensions. The three-dimensional surface plot chromatogram in Figure 2 shows a zoomed-in portion of the 20-ng/mL steroid GC GC TOF-MS analysis where stanozolol, 3-hydroxystanozolol, and 4-hydroxystanozolol are eluted. This figure is an excellent example of the increased peak capacity available with comprehensive two-dimensional chromatography. The stanozolol and hydroxystanozolols derivatize with one, two, or three trimethylsilyl groups. Each of the compounds, stanozolol, 3-hydroxystanozolol, and 4-hydroxystanozolol are eluted in the same first-dimension retention time modulation periods shown by the arrows in Figure 2. However, the same stanozolol and stanozolol-metabolites that are derivatized with fewer tri-methylsilyl groups are separated completely in the second dimension designated by the stars in Figure 2. This example illustrates how comprehensive two-dimensional GC GC fully resolves analytes that would otherwise be coeluted by one-dimensional chromatography. Figure 2 Meeting WADA Minimum Required Performance Levels for Detection of Prohibited Substances Table I shows the five steroid standards and the internal standard results of the low-level standards at 2 ng/ml, which is the WADA cut-off limit (2). The column labeled (QUANT S/N) lists the quantitative signal-to-noise ratio (S/N) for each steroid. S/N values for the five steroids ranged from 85 to 1353 at the 2 ppb concentration. The low-level cut-off

3 3 of 7 7/11/2014 3:39 PM Table I: Peak table for five steroids and steroid metabolites for the low level WADA cut-off limits at 2 ppb Table II: Calibration table used to develop calibration curves in ChromaTOF software guidelines are achieved and exceed the limits set by WADA. Limits of detection at a 10:1 S/N were calculated from the 2 ng/ml standard quant S/N results shown in Table I. The calculated limits of detection extend from 15 to 235 pg/ml in the part per trillion range. Calibration Curve Development A five point calibration curve was developed in the software in the concentration range of ng/ml for each of the five derivatized steroid standards. Figure 3 illustrates the calibration curve developed in the software for 3-hydroxystanozolol (3TMS) with a first order linear correlation coefficient value of Calibration curve development for all GC GC results for the 2-, 10-, 20-, 50-, and 100-ng/mL steroid standards was achieved with greater than 99.9% linearity. The generation of quantitative calibration curves for the five steroid standards used in this study demonstrates the practical application of utilizing GC GC TOF-MS for anti-doping control screening. Mass Spectral Deconvolution Figure 3 The column headings in Table II show the steroid analyte name, absolute first and second dimension retention time, minimum and maximum concentration range for the calibration, the analyte type, curve fit, the first-order equation, and the calculated correlation coefficient for each standard. The calibration table developed for this research produced excellent linearity for all five steroids. Results for the calibration curve development yielded linearity of greater than 99.9% for the five TMS-derivatized steroid standards. Figure 4 displays a deconvolution example for the trace level concentration of the derivatized anabolic steroid (19-norandrosterone) at 2 ppb. It is coeluted and buried under the endogenous anabolic steroid peak, 5 -androst-16-en-3 -OL. The chromatogram in Figure 4 shows the extracted ion traces for the masses at m/z 315, 241, 57, and 117. The peak markers for the unique masses m/z 315 and 241 are separated by approximately 20 ms. The mass spectrum labeled A is the Caliper or (total ion mass spectra) before deconvolution. The mass spectra labeled B is the Peak True deconvoluted mass spectrum. The mass spectra labeled C is the library search match for the deconvoluted mass spectrum. A library match similarity hit of 749 out of 1000 was made for the 19-norandrosterone 2 ppb derivatized standard Figure 4 peak. This illustration clearly shows the deconvolution of a trace anabolic steroid even when it is buried in and masked under heavy sample matrix. Conclusion This challenging work accomplished the goals set for this research. Calibration linearity for quantitation was attained for all analytes at greater than 99.9%. Limits of detection were calculated from acquired data that achieved part per trillion sensitivity. This research demonstrated the ability of GC GC TOF-MS to provide improved and enhanced analyte detection, deconvolution of trace level concentrations of steroids from heavy sample matrices, as well as exceptional quantitation capabilities for the analysis of illegal anabolic steroids in complex urine sample matrix. The experimental results of this work established that significantly increased analytical performance was accomplished by utilizing comprehensive GC GC TOF-MS for anti-doping screening control. In conclusion the research outcome shows favorable and practical applicability of GC GC TOF-MS for the positive identification of anabolic steroids in urine at or below the lowest allowable concentration limits in compliance with the guidelines set by WADA. John Heim and Doug Staples are with LECO Corporation, St. Joseph, Michigan. References (1) Recent Advances in Doping Analysis (12), W. Schanzer, H. Geyer, and A. Gotzmann, U. Mareck, Eds. (Sport und Buch Straub, Koln, 2004), pp

4 4 of 7 7/11/2014 3:39 PM (2) World Anti-Doping Agency (WADA) Technical Document TD2009MRPL The World Anti-Doping Code: The 2009 Prohibited List; International Standard.

5 5 of 7 7/11/2014 3:39 PM

6 6 of 7 7/11/2014 3:39 PM

7 7 of 7 7/11/2014 3:39 PM Figure 1 Figure 2 Table I: Peak table for five steroids and steroid metabolites for the low level WADA cut-off limits at 2 ppb Figure 3 Table II: Calibration table used to develop calibration curves in ChromaTOF software Figure Advanstar Communications Inc.. Permission granted for up to 5 copies. All rights reserved. You may forward this article or get additional permissions by typing into any web browser. Advanstar Communications Inc. and Chromatography Online logos are registered trademarks of Advanstar Communications Inc. The icopyright logo is a registered trademark of icopyright, Inc.

Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites

Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites Application Note Doping Control Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites Steroid profiling, screening, and confirmatory analysis using the

More information

Longterm Detection of Anabolic Steroid Metabolites in Urine

Longterm Detection of Anabolic Steroid Metabolites in Urine Longterm Detection of Anabolic Steroid Metabolites in Urine Application Note Forensics/Doping Control Author Stephan Baumann Agilent Technologies, Inc. 5301 Stevens Creek Blvd Santa Clara, CA 95051 USA

More information

WADA Technical Document TD2014EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting

WADA Technical Document TD2014EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting Endogenous Anabolic Androgenic Steroids Measurement and Reporting 1.0 Introduction The purpose of this Technical is to harmonize the approaches to the measurement and reporting of endogenous anabolic androgenic

More information

Comparative analysis of mass spectral matching for confident compound identification using the Advanced Electron Ionization source for GC-MS

Comparative analysis of mass spectral matching for confident compound identification using the Advanced Electron Ionization source for GC-MS TECHNICAL NOTE 1598 Comparative analysis of mass spectral matching for confident compound identification using the Advanced Electron Ionization source for GC-MS Authors Amit Gujar, 1 Tim Anderson, 1 Daniela

More information

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS UCT Part Numbers: CUQAX22Z Clean-Up C8+QAX, 2mg/1mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQID21-3UM

More information

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion: Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: The RSK-175 standard operating procedure was developed in order to determine the amount of dissolved gas in water. Due to the expansion

More information

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS Abstract Highly sensitive analysis of metanephrines in plasma is critical in the diagnosis and treatment of pheochromocytoma and paraganglioma. Here, a HILIC LC-MS/MS method was developed using a Raptor

More information

Sample purification procedure for confirmation of 3 -hydroxy stanozolol by gas chromatography/high resolution mass spectrometry

Sample purification procedure for confirmation of 3 -hydroxy stanozolol by gas chromatography/high resolution mass spectrometry Author proof done Research Article Sample purification procedure for confirmation of 3 -hydroxy stanozolol by gas chromatography/high resolution mass spectrometry S. Ahi, I. M. Reddy, A. Beotra, R. Lal,

More information

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE. Gases&Technology FEATURE Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. B Y M A T T H E W M O N A G L E Abstract Bulk gases are often delivered to

More information

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note Gas Chromatography Author Matthew S. Klee Agilent Technologies 2850 Centerville Road Wilmington, DE 19808-1610 USA Abstract

More information

GCMSD-Headspace Analysis SOP

GCMSD-Headspace Analysis SOP Before you start GCMSD-Headspace Analysis SOP Method and Sequences names are restricted to 50 characters (MSD program will crash otherwise) The GC oven and Injection Ports need to be cooled to 50 oc to

More information

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS Application Note: Analysis of and Cyanuric Acid in Food Matrices by LC-MS/MS PeterVarelis, National Center for Food Safety and Technology, Illinois Institute of Technology. Jonathan Beck, Kefei Wang, and

More information

WADA Technical Document TD2016EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting

WADA Technical Document TD2016EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting Endogenous Anabolic Androgenic Steroids Measurement and Reporting Introduction The purpose of this Technical Document (TD) is to harmonize the approaches to the measurement and reporting of Endogenous

More information

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation Using Nitrogen instead of Helium as carrier gas, while maintaining exact the same separation efficiency, retention times, peak elution order, without changing the oven temperature conditions. Jaap de Zeeuw

More information

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES 1. Applicability and Principle 1.1 Applicability. This method applies to the analysis of carbon dioxide

More information

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT APPLICATION NOTE Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT The INFICON 3000 Micro GC provides two fast and accurate solutions for the analysis of coal mine gas components.

More information

Rapid and Reliable Detection of Dissolved Gases in Water

Rapid and Reliable Detection of Dissolved Gases in Water Rapid and Reliable Detection of Dissolved Gases in Water Andrea Caruso and Massimo Santoro Thermo Fisher Scientific, Milan, Italy Application Note 005 Key Words Chromeleon CDS, Environmental, Fracking,

More information

Methylprednisolone detection in urine following local and oral administrations

Methylprednisolone detection in urine following local and oral administrations Susana MS Simões, Marta Calçada, Luís Horta, Xavier de la Torre Methylprednisolone detection in urine following local and oral administrations Laboratório de Análises e Dopagem. Instituto do Desporto de

More information

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS) Toxicology Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose - This procedure specifies the required elements for the calibration and use of the Agilent Gas Chromatograph interfaced to the Agilent

More information

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer application Note Gas Chromatography/ Mass Spectrometry Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer Overview A mass spectrometer

More information

Development of a Clinical Research Method for the Measurement of Testosterone. Dihydrotestosterone

Development of a Clinical Research Method for the Measurement of Testosterone. Dihydrotestosterone Development of a Clinical Research Method for the Measurement of Testosterone and Dihydrotestosterone Martijn van Faassen, Maarten van den Noort and Ido Kima University Medical Center, Groningen, Netherlands

More information

Automated Determination of Dissolved Gases in Water Anne Jurek

Automated Determination of Dissolved Gases in Water Anne Jurek Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: Tapping the natural gas reservoirs throughout the United States has long been a viable solution for energy independence; however

More information

High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis

High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis PO-CON695E High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis MSACL 6 EU Neil Loftus, Stephane Moreau, Mikaël Levi 3, Anja Grüning

More information

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI Low-Pressure Retention Locking with the 789A GC Application HPI Authors Courtney Milner and Russell Kinghorn BST International 41 Greenaway Street Bulleen, VIC 315 Australia Matthew S. Klee Agilent Technologies,

More information

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution. Name: 1. You are asked to measure the percent iron in a steel sample you believe to be about 50% Fe by mass. You dissolve four replicate samples in nitric acid and dilute them by mass to a concentration

More information

Now it is. easy to switch. CHS TM Steroid Profiling kit and software. Brochure not for distribution in the USA and Canada

Now it is. easy to switch. CHS TM Steroid Profiling kit and software. Brochure not for distribution in the USA and Canada Now it is easy to switch to Clinical MSMS CHS TM Steroid Profiling kit and software Brochure not for distribution in the USA and Canada 2 Meeting your needs for steroid assays PerkinElmer s new CHS MSMS

More information

SPECIFIC DETECTION OF ANDROST-4-ENE-3,17-DIONE

SPECIFIC DETECTION OF ANDROST-4-ENE-3,17-DIONE 7 SPECIFIC DETECTION OF ANDROST-4-ENE-3,17-DIONE ADMINISTRATION BY PROFILING OXYGENATED METABOLITES Abstract For the selective detection of androst-4-ene-3,17-dione ( 4-AEDIONE) administration, the analysis

More information

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS Peter Alden and Michael Jones Waters Corporation, Milford, MA, U.S. APPLICATION BENEFITS UPLC combines

More information

Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS)

Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS) Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS) 1.0 Purpose - This procedure specifies the required elements for the calibration and use of the Agilent 6890 GC interfaced

More information

Experiment GC : Analysis of BTEX by GC-FID

Experiment GC : Analysis of BTEX by GC-FID 46 Experiment GC : Analysis of BTEX by GC-FID Learning Goals: Familiarity with gas chromatography Gain experience in temperature programming and method development Correctly use an internal standard, and

More information

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (16) 17 R QUALIFICATION OF EQUIPMENT ANNEX 2: QUALIFICATION OF GC EQUIPMENT Full document title and reference Document type Legislative

More information

WADA Technical Document TD2017NA

WADA Technical Document TD2017NA HARMONIZATION OF ANALYSIS AND REPORTING OF 19-NORSTEROIDS RELATED TO NANDROLONE 1.0 Introduction This document has been established to harmonize the Confirmation Procedure for the analysis and reporting

More information

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek Abstract: Due to the expansion of natural gas drilling through horizontal fracturing, there has been

More information

CORESTA RECOMMENDED METHOD Nº 67

CORESTA RECOMMENDED METHOD Nº 67 CORESTA RECOMMENDED METHOD Nº 67 DETERMINATION OF WATER IN THE MAINSTREAM SMOKE OF CIGARS BY GAS CHROMATOGRAPHIC ANALYSIS (November 2005) 1. FIELD OF APPLICATION The method is applicable to the particulate

More information

Retention Time Locking: Concepts and Applications. Application

Retention Time Locking: Concepts and Applications. Application Retention Time Locking: Concepts and Applications Application Gas Chromatography December 1997 Authors Vince Giarrocco Bruce Quimby Matthew Klee Agilent Technologies, Inc. 2850 Centerville Road Wilmington,

More information

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries. Informal Commentary Limit of Diethylene Glycol (DEG) and Ethylene Glycol (EG) in Sorbitol Solution, Sorbitol Sorbitan Solution and Noncrystallizing Sorbitol Solution December 2009 Monograph/Section(s):

More information

APPLICATION NOTE. GC Integrated Permeation Device

APPLICATION NOTE. GC Integrated Permeation Device GC Integrated Permeation Device GC-integrated Design PPB to PPM Level Calibration No Cylinders or Regulators Required Option for Dual Ovens Cost Effective, Safe, Clean, Flexible, NIST Traceable Keywords:

More information

WADA Technical Document TD2010EAAS. Endogenous Anabolic Androgenic Steroids Testing, Reporting and Interpretive Guidance

WADA Technical Document TD2010EAAS. Endogenous Anabolic Androgenic Steroids Testing, Reporting and Interpretive Guidance Endogenous Anabolic Androgenic Steroids Testing, Reporting and Interpretive Guidance 1.0 Introduction The purpose of this Technical Document is to further harmonize the approaches to the detection and

More information

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Quantitative Analysis of Hydrocarbons by Gas Chromatography Quantitative Analysis of Hydrocarbons by Gas Chromatography Introduction Gas-liquid chromatography (GLC) accomplishes a separation by partitioning solutes between a mobile gas phase and a stationary liquid

More information

The instrument should have a mass range from 2 to 2000 m/z or better Interface

The instrument should have a mass range from 2 to 2000 m/z or better Interface Revised Specifications for Liquid Chromatography Mass Spectrometry (LC-MS/MS ) Triple Quadrupole Specification for Ultra-Fast Liquid Chromatography system Solvent Delivery 1) A quaternary Pump system for

More information

The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood

The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood UCT Part Numbers: ECQUUS1015CT- Enviro-Clean 15 ml centrifuge tube with 400 mg MgSO 4 and 100 mg NaCl CUMPS2CT - Enviro-Clean

More information

SOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15

SOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15 Chemicals needed: Precipitate solution, Methanol Derivatization Solution, Amplifex Keto Reagent Kit Mobile phases, H 2 0 with 0.1% Formic Acid LC/MS grade and Acetonitrile with 0.1% Formic Acid LC/MS grade

More information

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER Michael T Tang, Ph.D. Grace Feng Greg Merideth Rui Huang Matheson Gas Applied Lab

More information

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS Application Note: 424 Analysis of and Cyanuric Acid in Food Matrices by LC-MS/MS PeterVarelis, National Center for Food Safety and Technology, Illinois Institute of Technology. Jonathan Beck, Kefei Wang,

More information

GC-MS(/MS) investigations on long-term metabolites of 17-methyl steroids

GC-MS(/MS) investigations on long-term metabolites of 17-methyl steroids M.K. Parr 1), G. Fußhöller 1), M. Gütschow 2), C. Hess 1,2), W. Schänzer 1) GC-MS(/MS) investigations on long-term metabolites of 17-methyl steroids 1) Institute of Biochemistry, German Sport University,

More information

Introduction. Experimental

Introduction. Experimental PO-CON162E Development of Automated Screening and Quantitation Approach on Novel On-Line SFE-SFC-MS/MS Platform (I) For 23 Restricted Perflurocompounds in Textiles ASMS 2016 MP 283 Jie Xing, Jun Xiang

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results. Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions

More information

LCMSMS Solutions For Steroid Analysis

LCMSMS Solutions For Steroid Analysis LCMSMS Solutions For Steroid Analysis Dan Blake Manager, Clinical Applications For Research Use Only. Not for use in diagnostic procedures. RUO-MKT-18-1166-A Agenda Special considerations for steroid analysis

More information

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist ASTM WK 43267 Standard Test Method for Dissolved Gases Anne Jurek Applications Chemist Rationale Hydraulic Fracturing is becoming more and more common. Nearby water well can be affected during the fracking

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310 www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Varian 310 310 LC/MS Manual System INCLUDES: 310-MS LC/MS triple quadrople

More information

WADA Technical Document TD2018MRPL

WADA Technical Document TD2018MRPL MINIMUM REQUIRED PERFORMANCE LEVELS FOR DETECTION AND IDENTIFICATION OF NON-THRESHOLD SUBSTANCES In order to ensure that all WADA-accredited Laboratories can report the presence of Prohibited Substances,

More information

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection This user guide does not have the intention to introduce you to any of the theory and/or principles behind

More information

Determination of Selected ASTM Perfluorinated Compounds (PFCs) Using the Shimadzu LCMS No. LCMS-083. No. SSI-LCMS-083

Determination of Selected ASTM Perfluorinated Compounds (PFCs) Using the Shimadzu LCMS No. LCMS-083. No. SSI-LCMS-083 Liquid Chromatography Mass Spectrometry No. LCMS-083 Determination of Selected ASTM Perfluorinated Compounds (PFCs) Using the Shimadzu LCMS-8060 Summary ASTM methods D7968-14 and D7979-15 are used to determine

More information

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler Optimizing Vial Pressurization Parameters for the Analysis of Residual Solvents Using the 7697A Headspace Sampler Application Note Pharmaceuticals Author Roger L Firor Agilent Technologies, Inc.

More information

Agilent Dimension Software for ELSD User Manual

Agilent Dimension Software for ELSD User Manual Agilent Dimension Software for ELSD User Manual Agilent Dimension Software for ELSD User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2011 No part of this manual may be reproduced in

More information

WADA Technical Document TD2018EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting

WADA Technical Document TD2018EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting Endogenous Anabolic Androgenic Steroids Measurement and Reporting 1.0 Introduction The purpose of this Technical Document (TD) is to harmonize the approaches to the measurement and reporting of Endogenous

More information

GUIDE TO ULTRA-FAST HPLC. Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations

GUIDE TO ULTRA-FAST HPLC. Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations GUIDE TO ULTRA-FAST HPLC Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations GUIDE TO ULTRA-FAST HPLC Quick Tips for Converting Conventional Reversed-Phase

More information

1. INTRODUCTION CH 3 CH3

1. INTRODUCTION CH 3 CH3 1. ITRODUCTIO TITLE: Analysis of stanozolol and metabolites S.O.P. : ARO/446 in urine Page : 1 of 12 This method describes the analysis of stanozolol, 4ß-hydroxy-stanozolol and 16ß-hydroxy-stanozolol in

More information

Extended Application Note

Extended Application Note Extended Application Note Harmful Substances in Dietary Supplements HPLC Columns with APP A-337 www.mtc-usa.com 1-732-578-1777 INTRODUCTION A dietary supplement is a substance to be consumed for the purpose

More information

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY Nengbing Xu, Hongmei Ying and Libo Zhu Ningbo Environmental Monitoring

More information

During today's webinar we're going to look at the following topics:

During today's webinar we're going to look at the following topics: Alternative Carrier Gas for GC and GC-MS Presentation Transcript Hello and welcome to the Peak scientific webinar which today will concern alternative carrier gas for GC and GC-MS. My name is Ed Connor,

More information

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Oneida Research Services, Inc. 8282 Halsey Road Whitesboro, NY 13492 Phone: (315) 736-5480 FAX: (315) 736-9321 1.0 Purpose The

More information

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC Evaluation of linear and step gradient performance and retention time precision of the Agilent 12 Series Rapid Resolution LC This document is believed to be accurate and up-to-date. However, Agilent Technologies,

More information

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor Anwar Sutan, Metco Services Ltd. Charles Johnson, Metco Services Ltd. Jason Laidlaw,

More information

Carrier Gases in Capillary GC

Carrier Gases in Capillary GC Carrier Gases in Capillary GC LC Columns and Consumables Mark Sinnott Application Engineer January 15, 2009 CARRIER GAS Mobile Phase Carries the solutes down the column Selection and velocity influences

More information

CHEMTRACE Fremont Blvd. Fremont, CA 94538, Tel: (510) Fax: (510)

CHEMTRACE Fremont Blvd. Fremont, CA 94538, Tel: (510) Fax: (510) Report Date: 2/12/04 Page 1 of 5 Analysis Requested: Organic Outgassing Analysis by GC-MS at 100 C for 30 minutes. Sample Information: Sample ID: Wafer Carrier Material Samples were received wrapped in

More information

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction A Column-Flow Independent Configuration for QuickSwap Application Authors Matthew S. Klee and Bruce Quimby Agilent Technologies, Inc. 80 Centerville Road Wilmington, DE 08 USA Abstract As can be seen from

More information

Chemical Analysis for Methyltestosterone

Chemical Analysis for Methyltestosterone APPLIED CONSUMER SERVICES, INC. Chemical Analysis for Methyltestosterone Report #: 27584 Report to: Mr. Mitch Ross 24 Dockside Lane # 131 Key Largo, Florida 33017 2016 1 1 8 9 0 N W 8 7 t h C o u r t,

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs. Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography

More information

RECENT ADVANCES IN DOPING ANALYSIS (2)

RECENT ADVANCES IN DOPING ANALYSIS (2) M. DONIKE H. GEYER A. GOTZMANN U. MARECK-ENGELKE (EDITORS) RECENT ADVANCES IN DOPING ANALYSIS (2) Proceedings of the 12th Cologne Workshop on Dope Analysis 10th to 15th April 1994 SPORT und BUCH StrauB

More information

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module TECHNICAL NOTE High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module TN42256 Dr. Liliana Krotz, Dr. Francesco Leone, Walter Galotta and Dr. Guido Giazzi

More information

Gas chromatographic determination of 1,4-dioxane at the

Gas chromatographic determination of 1,4-dioxane at the J. Soc. Cosmet. Chem., 42, 97-104 (March/April 1991) Gas chromatographic determination of 1,4-dioxane at the parts-per-million level in consumer shampoo products MARK P. ITALIA and MATHEWS A. NUNES, Johnson

More information

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial Supporting Information for Micro-Collection of Gases in a Capillary Tube: Preservation of Spatial and Temporal Resolution AUTHOR INFORMATION Kristin A. Herrmann Favela, 1 * Pieter Tans, 2 Thomas Jaeckle,

More information

no peaks missing peaks too small peaks increasing retention decreasing retention

no peaks missing peaks too small peaks increasing retention decreasing retention GC Troubleshooting We can help you. 1. Observation: no peaks no peaks missing peaks too small peaks increasing retention decreasing retention declining baseline rising baseline bleeding plateaus interfering

More information

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay Summary This assay measures Human Granulocyte Colony Stimulating Factor (G-CSF) in a 96-well

More information

Impact of Air Leaks on the Productivity of GC and GC/MS Systems

Impact of Air Leaks on the Productivity of GC and GC/MS Systems Impact of Air Leaks on the Productivity of GC and GC/MS Systems Application Note Environmental Author Ken Lynam Agilent Technologies, Inc. Abstract Agilent 7890A GC dual-channel FID, Agilent 7890B GC,

More information

Weiyi Zheng, Xiaoxia Dong, Evan Rogers, Jimmie C. Oxley, and James L. Smith*

Weiyi Zheng, Xiaoxia Dong, Evan Rogers, Jimmie C. Oxley, and James L. Smith* Improvements in the Determination of Decomposition Gases from 1,3,3-Trinitroazetidine and 5-Nitro-2,4-dihydro-3H-1,2,4,-trizol-3-one using Capillary Gas Chromatography-Mass Spectrometry Weiyi Zheng, Xiaoxia

More information

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems Inorganic Analysis Chromatography Molecular Spectroscopy Thermal/Elemental Analysis Informatics Gas Chromatography MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass

More information

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD) Procedure No: SOP-026 Revision No: 1.0 January 24, 2011 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to meet

More information

System Overview. TCD Detector temperature setpoint Regulator

System Overview. TCD Detector temperature setpoint Regulator System Overview POPULAR CONFIGURATION GCs Your educational TCD GC is configured on the compact 310 chassis. It is equipped with a TCD Detector, a temperature programmable Column Oven, a 3 Silica Gel packed

More information

Gas Chromatographic Applications of the Dielectric Barrier Discharge Detector

Gas Chromatographic Applications of the Dielectric Barrier Discharge Detector Gas Chromatographic Applications of the Dielectric Barrier Discharge Detector Matthew Monagle, Advanced Industrial Chemistry Corp*. Ronda Gras, Jim Luong, Dow Chemical Canada 28th International Symposium

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 6974-4 First edition 2000-04-01 Natural gas Determination of composition with defined uncertainty by gas chromatography Part 4: Determination of nitrogen, carbon dioxide and

More information

CHM Gas Chromatography: Our Instrument Charles Taylor (r10)

CHM Gas Chromatography: Our Instrument Charles Taylor (r10) 1/8 Introduction You should already be familiar with the basic concepts of gas chromatography. The purpose of this note pack is to familiarize you with the specific instrument that we will use in our labs.

More information

TRACE LEVEL DETERMINATION OF PER- AND POLYFLUOROALKYL SUBSTANCES (PFASs) IN WATER USING THE AGILENT 6460 LC/MS/MS

TRACE LEVEL DETERMINATION OF PER- AND POLYFLUOROALKYL SUBSTANCES (PFASs) IN WATER USING THE AGILENT 6460 LC/MS/MS ENVIRONMENTAL MARKET TRACE LEVEL DETERMINATION OF PER- AND POLYFLUOROALKYL SUBSTANCES (PFASs) IN WATER USING THE AGILENT 6460 LC/MS/MS A method for the determination of per- and polyfluoroalkyl substances

More information

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Your local gas generation partner. Precision series Modular gas generation solution for GC. Your local gas generation partner Precision series Modular gas generation solution for GC www.peakscientific.com Perform with Precision Specifically designed and engineered for GC laboratory applications,

More information

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic. 458 LCGC NORTH AMERICA VOLUME 24 NUMBER 5 MAY 2006 www.chromatographyonline.com LC TROUBLESHOOTING Dwell Volume Revisited Gradient transfer does not have to be problematic. John W. Dolan LC Troubleshooting

More information

TESTOSTERONE AND ESTERS* Latest Revision: June 23, 2005

TESTOSTERONE AND ESTERS* Latest Revision: June 23, 2005 TESTSTERNE AND ESTERS* Latest Revision: June 23, 2005 *ther esters of testosterone have been synthesized and identified; however, only the following are treated in this monograph: C H 3 H CH 3 C H 3 CH

More information

Hiden QIC Series Gas Analysers Real-time gas analysis systems

Hiden QIC Series Gas Analysers Real-time gas analysis systems Hiden QIC Series Gas Analysers Real-time gas analysis systems vacuum analysis surface science plasma diagnostics gas analysis Application Focussed Options Multistream valves 20, 40 and 80 way versions

More information

The analysis of complex multicomponent

The analysis of complex multicomponent The analysis of complex multicomponent gas mixtures for emission monitoring 31 July 2017 Test and Measurement Prelly Marebane Fit-for-purpose reference materials for air pollution monitoring in SA and

More information

Advancements in Gas Chromatography Analyzers - Keeping up with New Technology. Chuck Runkle Gas Phase Product Specialist ASTS June 2013

Advancements in Gas Chromatography Analyzers - Keeping up with New Technology. Chuck Runkle Gas Phase Product Specialist ASTS June 2013 Advancements in Gas Chromatography Analyzers - Keeping up with New Technology Chuck Runkle Gas Phase Product Specialist ASTS June 0 Capillary Flow Technology -- solves difficult application problems easily

More information

Renzo BAGNATI Mario Negri Institute of Pharmacological Research- IRCCS Milano - Italy

Renzo BAGNATI Mario Negri Institute of Pharmacological Research- IRCCS Milano - Italy STEROIDS IN WASTE WATERS AS MARKERS OF ABUSE Renzo BAGNATI Mario Negri Institute of Pharmacological Research- IRCCS Milano - Italy A first attempttoapplythe sewage epidemiologyapproachin doping Compare

More information

M/s. Thermo fisher Scientific

M/s. Thermo fisher Scientific CLARIFICATIONS AND RESPONSES IN THE PRE BID MEETING HELD ON 19 TH AUGUST 2013 IN OUR COMMITTEE ROOM FOR THE PROCUREMENT OF DIOXIN ANALYSER (GC-MS/MS) File No : PUR/IMP/178/14 Description of Item : DIOXIN

More information

SUPPORTING INFORMATION

SUPPORTING INFORMATION SUPPORTING INFORMATION Aging of Organochlorine Pesticides and Polychlorinated Biphenyls in Muck Soil: Volatilization, Bioaccessibility and Degradation Fiona Wong, Terry F. Bidleman *, Centre for Atmospheric

More information

Measuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability

Measuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability Measuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability R. C. Kullberg*, D. J. Rossiter** Oneida Research Services, Inc., 8282 Halsey Road, Whitesboro, NY13492 Abstract Many

More information

Gas chromatographic and mass spectrometric analysis of conjugated steroids in urine

Gas chromatographic and mass spectrometric analysis of conjugated steroids in urine Gas chromatographic and mass spectrometric analysis of conjugated steroids in urine JONG MAN YOON* and KYUNG HO LEE Department of Marine Biomedical Science, College of Ocean Science and Technology, Kunsan

More information

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Jamie Brown R&D Scientist Supelco 595 North Harrison RD. Bellefonte, PA 16823 National

More information

Agilent 6890N, Option 652 Oral Fluids Analyzer

Agilent 6890N, Option 652 Oral Fluids Analyzer Agilent 6890N, Option 652 Oral Fluids Analyzer Installation and Operation Agilent Technologies Notices Agilent Technologies, Inc. 2006 No part of this manual may be reproduced in any form or by any means

More information

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1 Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost HERMAN HOLM SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 4C CONFERENCE APRIL 2018 SAN ANTONIO, TX SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC

More information

Agilent 7800 ICP-MS. Specifications and Typical Performance. Fast track metals analysis with Solution-Ready ICP-MS

Agilent 7800 ICP-MS. Specifications and Typical Performance. Fast track metals analysis with Solution-Ready ICP-MS Agilent 7800 ICP-MS Specifications and Typical Performance Fast track metals analysis with Solution-Ready ICP-MS The Solution-Ready Agilent 7800 Quadrupole ICP-MS combines proven, robust hardware, auto-optimization

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 16017-1 First edition 2000-11-15 Indoor, ambient and workplace air Sampling and analysis of volatile organic compounds by sorbent tube/thermal desorption/capillary gas chromatography

More information