1 The analysis of complex multicomponent gas mixtures for emission monitoring 31 July 2017 Test and Measurement Prelly Marebane
2 Fit-for-purpose reference materials for air pollution monitoring in SA and SADC region Have you Ever wondered who is responsible for traceability of measurements in South Africa? Measurements affect our daily lives, from the food we eat, the quality of drinking water, the monitoring of air pollution, health care and even for law enforcement. NMISA provides the confidence that measurements in South Africa are accurate, comparable and internationally accepted. In air pollution monitoring, the provision of reference materials is important to ensure accuracy of results reported for compliance to national air quality ambient air standards as set out by the air quality act of 2004, health and safety limits and global climate change programmes. This accuracy often depends on; o Continuous development of reference methods o Reduction of uncertainty o Matrix reference materials o Cutting edge technology This presentation focuses on the effects of differences in matrix gases in the analysis of stack gas by the NMISA and the implications to industry. It also reports on the findings of another important emission industry; automotive.
3 Production Measurement Improvements Significance and methodology The significance of this study is to better the measurement capability of the laboratory for emission reference gases in an effort to ultimately provide suitable reference gas mixtures representative of emission samples. For consideration; o Matrix effect (78% calibrated MFCs etc. ) o Reactivity of gases o Gravimetric preparation and purity analysis o Cross interferences non dispersive infrared spectroscopy o Simultaneous determination Preparation of compressed gas mixtures by gravimetry - Six components for stack mixtures -five for automotive mixtures Confirm or reject the hypothesis of synergistic effects resulting from differences in matrix gas of sample and calibration mixture Check if matrix matching will result in better measurement of myriad complex mixtures - Sort out other calibration methods Target mass, purity analysis
4 High pure Sulphur dioxide, nitric oxide, propane, carbon monoxide, carbon dioxide, oxygen and nitrogen Methodology and experimental m i = x ip f V cyl M i RTZ f Assume Z=1 1. ISO : Gas Analysis - Preparation of calibration gas mixtures Gravimetric method, (2015) 2. Hodgkinson J., Smith R., Ho W.O., Safell J.R., Tatam R.P., Non-dispersive infrared (NDIR) measurement of carbon dioxide at 4.2 μm in a compact and optically efficient sensor. Sensors and Actuators B (2013)
5 Methodology and experimental Analytical parameters for the analysis of propane in stack gas mixtures using gas chromatography coupled with flame ionisation detector (FID) Same conditions may be used for carbon dioxide analysis by gas chromatography with thermal conductivity detector at 175 deg. Celsius 20 µl sample volume 7 injections Matrix match only
6 Methodology and experimental continued. Calibration techniques offer the ability to compare methods as well in efforts to measure emission gases better. Calibration curve γ = mx + c Single point calibration Matrix match and same mole fraction of gas component Monitor drift Time efficient Residual gas analyser (simultaneous analysis by multi-dimensional chromatography)
7 Results stack gas ppm CO, 10-16% CO 2, ppm NO, 100 ppm SO 2, 1-10 ppm C 3 H 8 in N 2 The accuracy of the measurement for matrix matching = 0.14% and at % Relative expanded uncertainty of 0.5%
8 07 December December December December December 2016 Gravimetric concentration, µmol/mol Results stack gas C 3 H 8 - D Analysis The accuracy of the measurement for matrix matching = 0.4% and at % Relative expanded uncertainty of 0.3% The accuracy of the measurement for matrix matching = 0.4% and at % Relative expanded uncertainty of 1.8%
9 Results stack gas Source of Sulphur dioxide in stack gas -Elemental Sulphur impurities in coal, Hancox et al., (2014). -combustion of conventional fuels Non-dispersive infrared spectroscopy Synergistic effects observed when other stack mixtures where used for calibration Defied the hypothesis SO 2 /N 2 calibration gave better results The accuracy of the measurement for matrix matching = 0.09% and at % Relative expanded uncertainty of 1.5% Therefore, the results show that when calibrating a measuring instrument for stack gas, traditional methods calibration methods may be used. However, this has not been investigated on the use of chromatography (Sulphur chemiluminescent detector)
10 Results automotive Method for analysis Column: Molecular sieve 13X for oxygen, nitrogen and carbon monoxide separation Shin Carbon St for carbon dioxide PLOT column for C 3 H 8 Oven; 30 ᵒC Temp program: initial 30 ᵒC for 2 min, and 85 ᵒC for 3 min 30 ᵒC/min Sample flow: 35 ml/min Injections: 7 Carrier gases: N 2 and H 2 FID and TCD: both at 175 ᵒC Run time: 6 min
11 Results Automotive Traditionally binary reference gas mixtures (one component and nitrogen balance) and multipoint calibration technique have been used to determine unknown mole fractions of samples. However, this method has been found to be associated with synergistic effects. New system exploits the linearity of gas chromatographs to provide a simultaneous and quicker analysis Matrix matching exploits the ability to measure without interferences from matrix effect by matching not only the matrix but mole fraction as well, Component %difference (max) %REU O CO CO C 3 H The capability of the method is illustrated in the table above Target: Emphasis on not only accuracy but the reduction of uncertainty Goal: 0.5% Relative expanded uncertainty, long-term
12 Conclusion 1 -Need for representative matrix reference materials -sort out and develop new methods to control challenges posed by complex mixtures -Ongoing development of stack and automotive reference materials for emission monitoring 2 CO 2 stack good, both types of matrices C 3 H 8 stability problems but good, only matrix matching evaluated NO matrix matching results in better capability SO 2 matrix matching doesn t result in better measurement CO needs improvement 3 The new method used for automotive gas mixtures is fit for purpose
13 Acknowledgements The NMISA Dr. J Tshilongo Gas Analysis section head (NMISA) Prof H Tutu Supervisor (University of the Witwatersrand) Gas team: David Mogale, Napo Ntsasa, Portia Seemane, Tshepiso Mphamo, Kennedy Ramahala, Lincon Maluleke, Mudalo Jozela, Nompumelelo Leshabane, Silindile Lushozi, Gumani Mphaphuli and Goitsemang Lekoto. National Research Foundation
Traceability of greenhouse gases and volatile organic compounds for air monitoring in South Africa
Traceability of greenhouse gases and volatile organic compounds for air monitoring in South Africa James Tshilongo *1, Silindile L. Lushozi 1,2, Napo Ntsasa 1, Mudalo Jozela 1,2 Luke Chimuka 2 and Goitsemang
More informationGravimetric preparation of NO 2 primary reference gas mixtures at the NMISA. Tshepiso Mphamo 03 September 2012
Gravimetric preparation of NO 2 primary reference gas mixtures at the NMISA Tshepiso Mphamo 03 September 2012 NMISA 2012 Contents Introduction Experimental Verification Uncertainty CCQM- Pilot study results
More informationEN ISO/IEC Technical Requirements
EN ISO/IEC 17025 - Technical Requirements Wolfgang Garche Environmental Protection Agency Department Air Quality Monitoring, Information, Assessment 1 What means accreditation in accordance to EN ISO/IEC
More informationKeeping It Simple. Gas Mixtures for Stack Emissions Monitoring.
Keeping It Simple Gas Mixtures for Stack Emissions Monitoring www.uk.airliquide.com Stack Emissions Monitoring The Stack Emissions Monitoring market has undergone substantial legislative and quality changes
More informationDevelopment of DMS and Acetonitrile, and Formaldehyde gas standards. Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin
Development of DMS and Acetonitrile, and Formaldehyde gas standards Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin The amount of chemical substances in cylinder container Purity uncertainty Gravimetry
More informationWhite Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:
Title: Comparison of Chemical Sensor and NDIR Technologies TSN Number: 25 File:\\MII- SRV1\Metron\Bridge_Analyzers\Customer_Service_Documentation\White_Papers\25_G en EGA NDIR vs Chemical Sensor.docx Created
More informationComparison APMP.QM-S2.1 Oxygen in nitrogen at atmospheric level
APMP.QM-S2.1 report Comparison APMP.QM-S2.1 Oxygen in nitrogen at atmospheric level Final Report ByungMoon Kim 1, Kwangsub Kim 1, Jinsang Jung 1,*, Sanghyub Oh 1, Liu Hui 2, Hou Li 2, Teo Beng Keat 2,
More informationHiQ laboratory gas generators.
Product information HiQ laboratory gas generators. The convenient alternative for laboratory environments. 02 HiQ laboratory gas generators HiQ laboratory gas generators. Designed with safety and convenience
More informationMETROLOGICAL ASSURENCE OF O3, CO2, CH4 AND CO CONTROL IN ATMOSPHERE
METROLOGICAL ASSURENCE OF O3, CO2, CH4 AND CO CONTROL IN ATMOSPHERE Leonid A. Konopelko Yuri A. Kustikov Vitaly V. Beloborodov Dmitry N. Selukov D.I.Mendeleyev Institute for Metrology (VNIIM), 19 Moskovsky
More informationLINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER
LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER Michael T Tang, Ph.D. Grace Feng Greg Merideth Rui Huang Matheson Gas Applied Lab
More informationDate of Shipment: Xxxxx 00, 20xx 3402c National Institute of Advanced Industrial Science and Technology National Metrology Institute of
Date of Shipment: Xxxxx 00, 20xx 3402c01151112-170831 National Metrology Institute of Japan Reference Material Certificate NMIJ CRM 3402-c01 Sulfur Dioxide This certified reference material (CRM) was produced
More informationGases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.
Gases&Technology FEATURE Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. B Y M A T T H E W M O N A G L E Abstract Bulk gases are often delivered to
More informationInternational Comparison CCQM-K111.1 Propane in nitrogen
International Comparison CCQM-K111.1 Propane in nitrogen Adriaan M.H. van der Veen 1, J. Wouter van der Hout 1, Paul R. Ziel 1, Mudalo Jozela, James Tshilongo, Napo G. Ntsasa, Angelique Botha 1 Van Swinden
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
code Tarapur Qatar Location code Customers sites Dove House Dove Fields Staffordshire ST14 8HU Contact: Dr Gavin Squire Tel: +44 (0)1889 569229 Fax: +44 (0)1889 569220 E-Mail: gavin.squire@effectech.co.uk
More informationMETHOD 3A - DETERMINATION OF OXYGEN AND CARBON DIOXIDE CONCENTRATIONS IN EMISSIONS FROM STATIONARY SOURCES (INSTRUMENTAL ANALYZER PROCEDURE)
METHOD 3A - DETERMINATION OF OXYGEN AND CARBON DIOXIDE CONCENTRATIONS IN EMISSIONS FROM STATIONARY SOURCES (INSTRUMENTAL ANALYZER PROCEDURE) 1.0 Scope and Application What is Method 3A? Method 3A is a
More informationENVIRONMENTAL AND POLLUTANTS GAS ANALYZERS
XIX IMEKO World Congress Fundamental and Applied Metrology September 6, 9, Lisbon, Portugal ENVIRONMENTAL AND POLLUTANTS GAS ANALYZERS Ana Madeira, Florbela A. Dias and Eduarda Filipe 3 Instituto Português
More informationAPPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT
APPLICATION NOTE Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT The INFICON 3000 Micro GC provides two fast and accurate solutions for the analysis of coal mine gas components.
More informationVehicle and Engine Emissions Testing
Vehicle and Engine Emissions Testing Calibration gases and equipment for compliance Benefits and Features Guaranteed analytical accuracy facilitates precise instrument calibration to ensure Part 1065,
More informationFinal Report of APMP.QM-K46 Ammonia in Nitrogen at 30 µmol/mol Level
Final Report of APMP.QM-K46 Ammonia in Nitrogen at 30 µmol/mol Level Shinji Uehara 1), HAN Qiao 2), Takuya Shimosaka 3)* 1) Chemicals Evaluation and Research Institute, Japan (CERI) 2) National Institute
More informationFlare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1
Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost HERMAN HOLM SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 4C CONFERENCE APRIL 2018 SAN ANTONIO, TX SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC
More informationThere is no protection without detection. HiQ specialty gases, equipment and services for environmental monitoring.
There is no protection without detection. HiQ specialty gases, equipment and services for environmental monitoring. 02 HiQ specialty gases, equipment and services for environmental monitoring. HiQ pure
More informationThere is no protection without detection. HiQ Specialty Gases, Equipment and Services for Environmental Monitoring.
There is no protection without detection. HiQ Specialty Gases, Equipment and Services for Environmental Monitoring. 02 HiQ Specialty Gases, Equipment and Services for Environmental Monitoring HiQ Specialty
More informationCOOMET TECHNICAL COMMITTEE 1.8 PHYSICAL CHEMISTRY
COOMET TECHNICAL COMMITTEE 1.8 PHYSICAL CHEMISTRY CООМЕТ PROJECT 664/RU/14 SUPPLEMENTARY COMPARISON OF STANDARD GAS MIXTURES: «Atmospheric air pollutants: CO in Nitrogen, 5 µmol/mol - Technical Protocol,
More informationLaboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.
Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions
More informationMaximum precision for quantifiable success.
Accredited calibration gas mixtures Maximum precision for quantifiable success. Reference materials and other accredited calibration gas mixtures. 02 Accredited calibration gas mixtures It s now official:
More informationMETHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES
METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES 1. Applicability and Principle 1.1 Applicability. This method applies to the analysis of carbon dioxide
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 6974-4 First edition 2000-04-01 Natural gas Determination of composition with defined uncertainty by gas chromatography Part 4: Determination of nitrogen, carbon dioxide and
More informationLaboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.
Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography
More informationAPPLICATION NOTE. GC Integrated Permeation Device
GC Integrated Permeation Device GC-integrated Design PPB to PPM Level Calibration No Cylinders or Regulators Required Option for Dual Ovens Cost Effective, Safe, Clean, Flexible, NIST Traceable Keywords:
More informationReport on Co-production of CH 4 /air within ACRM framework
Report on Co-production of CH 4 /air within ACRM framework 8 th APMP/TCQM Gas CRM workshop NMIJ/AIST, Tsukuba, 9~11 June 2010 By Hai Wu, NIM 1 Brief Introduction to ACRM ACRM Concept Asian Collaboration
More informationNitrogen subtraction on reported CO 2 emission using ultrasonic flare gas meter
Nitrogen subtraction on reported CO 2 emission using ultrasonic flare gas meter Kjell-Eivind Frøysa, Christian Michelsen Research AS, Bergen, Norway Henning Ekerhovd, StatoilHydro ASA, Kollsnes, Norway
More informationSvensk läkemedelsstandard
Svensk läkemedelsstandard 2019.0 Carbon dioxide Postadress/Postal address: P.O. Box 26, SE-751 03 Uppsala, SWEDEN Besöksadress/Visiting address: Dag Hammarskjölds väg 42, Uppsala Telefon/Phone: +46 (0)18
More informationAutomated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:
Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: The RSK-175 standard operating procedure was developed in order to determine the amount of dissolved gas in water. Due to the expansion
More informationGas Metrology. SI-traceable: Primary Reference Materials Certified Reference Materials Calibrated Gas Mixtures Interlaboratory Comparisons
Gas Metrology SI-traceable: Primary Reference Materials Certified Reference Materials Calibrated Gas Mixtures Interlaboratory Comparisons T +31 15 2691500 E vsl@vsl.nl www.vsl.nl 1 This guide provides
More informationHigh Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module
TECHNICAL NOTE High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module TN42256 Dr. Liliana Krotz, Dr. Francesco Leone, Walter Galotta and Dr. Guido Giazzi
More informationSvensk läkemedelsstandard
Svensk läkemedelsstandard 2017.2 Nitrous oxide Postadress/Postal address: P.O. Box 26, SE-751 03 Uppsala, SWEDEN Besöksadress/Visiting address: Dag Hammarskjölds väg 42, Uppsala Telefon/Phone: +46 (0)18
More informationHiQ. Specialty Gases & Specialty Equipment. Precision matters in everything we do.
HiQ. Specialty Gases & Specialty Equipment. Precision matters in everything we do. Content 1 HiQ Overview Page 3-6 2 HiQ Customers Page 7-13 3 HiQ Products Page 14-30 4 HiQ Service Page 31-39 HiQ. Global
More informationSpeciality Gases Practice 4. Our quality, your safety. Accredited Test and Calibration Laboratory Speciality Gases Centre in Hörstel.
Speciality Gases Practice 4 Our quality, your safety. Accredited Test and Calibration Laboratory Speciality Gases Centre in Hörstel. Accredited Westfalen quality. Speciality Gases Centre in Hörstel. They
More information3 HiQ Specialty Gases and Specialty Equipment. Precision matters in everything we do
3 HiQ Specialty Gases and Specialty Equipment Precision matters in everything we do 02 Introduction Introduction 03 Contents. 4-5 What we can do for you 6-7 Specialty gases products 8-9 How we measure
More informationYour local gas generation partner. Precision series Modular gas generation solution for GC.
Your local gas generation partner Precision series Modular gas generation solution for GC www.peakscientific.com Perform with Precision Specifically designed and engineered for GC laboratory applications,
More informationCALIBRATION. empanelled laboratories following standard procedures and using certified reference standards.
CALIBRATION A. Calibration of Air Analysers Gaseous Parameter a. The instruments/ analysers for real time monitoring of gaseous emissions shall be calibrated with respect to their functioning, drift, linearity
More informationPharmaceutical Grade Nitrogen & Gardner Denver / CompAir PSA Nitrogen Gas Generation Systems
Pharmaceutical Grade Nitrogen & Gardner Denver / CompAir PSA Nitrogen Gas Generation Systems The purpose of this document is to clarify the specification for pharmaceutical grade nitrogen gas and confirm
More informationYour local gas generation partner. Precision series Modular gas generation solution for GC.
Your local gas generation partner Precision series Modular gas generation solution for GC www.peakscientific.com Perform with Precision Specifically designed and engineered for GC laboratory applications,
More informationYour local gas generation partner. Precision series Modular gas generation solution for GC.
Your local gas generation partner Precision series Modular gas generation solution for GC www.peakscientific.com Perform with Precision Specifically designed and engineered for GC laboratory applications,
More informationThree Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor
Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor Anwar Sutan, Metco Services Ltd. Charles Johnson, Metco Services Ltd. Jason Laidlaw,
More informationHiQ Specialty Gases and Specialty Equipment. Precision matters. In everything we do.
HiQ Specialty Gases and Specialty Equipment Precision matters. In everything we do. 02 Introduction Petrochemical refinery complex Our industry is highly complex. But our vision is simple. We provide precisely
More informationImproved Reliability in Natural Gas Energy Measurement with the Revised ISO Standard. Gerard Nieuwenkamp - VSL
Improved Reliability in Natural Gas Energy Measurement with the Revised ISO 10723 Standard Gerard Nieuwenkamp - VSL Improved reliability in natural gas energy measurement with the revised ISO 10723 standard
More informationHYDROGEN - METHANE MIXTURES : DISPERSION AND STRATIFICATION STUDIES
HYDROGEN - METHANE MIXTURES : DISPERSION AND STRATIFICATION STUDIES A. Marangon, M.N. Carcassi Department of Mechanical, Nuclear and Production Engineering, University of Pisa, Via Largo Lucio Lazzarino,
More informationGas mixtures. Individual solutions specifically for your application
Gas mixtures Individual solutions specifically for your application For many applications besides pure gases also defined gas mixtures are required. For many routine applications, from operating gases
More informationSvensk läkemedelsstandard
Svensk läkemedelsstandard 2018.2 Nitrogen Postadress/Postal address: P.O. Box 26, SE-751 03 Uppsala, SWEDEN Besöksadress/Visiting address: Dag Hammarskjölds väg 42, Uppsala Telefon/Phone: +46 (0)18 17
More informationRare Gas Applications
Rare Gas Applications Specializing in EPA, Electronic, Medical, and NMR & MRI GLOBAL REACH, LOCAL SERVICE 1.800.649.6704 WWW. MIDDLESEXGASES.COM The Middlesex specialty gas advantage Middlesex Gases &
More informationHIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION
HIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION SCENTROID is the global leader in sensory technology providing intelligent environmental solutions across a wide range of industries
More informationThe. «Modular Alliance» Gas Generators. ALLIANCE, the partnership that keeps the gas flowing. One step ahead!
The «Modular Alliance» Gas Generators One step ahead! ALLIANCE, the partnership that keeps the gas flowing. F-DGSi EXPERTISE IS DEFINED BY ALWAYS BEING ONE STEP AHEAD! WE HAVE DESIGNED NEW MODULAR SOLUTION
More informationCORESTA RECOMMENDED METHOD Nº 67
CORESTA RECOMMENDED METHOD Nº 67 DETERMINATION OF WATER IN THE MAINSTREAM SMOKE OF CIGARS BY GAS CHROMATOGRAPHIC ANALYSIS (November 2005) 1. FIELD OF APPLICATION The method is applicable to the particulate
More informationDetector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane
Gas requirements Gases for packed columns The carrier gas you use depends upon the type of detector and the performance requirements. Table 520-1 lists gas recommendations for packed column use. In general,
More informationCOAL MINES TECHNICAL SERVICES, GAS DETECTION RELATED SERVICES FOR MINES RESCUE
COAL MINES TECHNICAL SERVICES, GAS DETECTION RELATED SERVICES FOR MINES RESCUE Peter Mason Manager, Coal Mines Technical Services, a division of Mines Rescue Pty Ltd, Australia ABSTRACT The NSW Mines Rescue
More informationCOOMET TECHNICAL COMMITTEE 1.8 PHYSICAL CHEMISTRY. CООМЕТ project 576/RU/12
COOMET TECHNICAL COMMITTEE 1.8 PHYSICAL CHEMISTRY CООМЕТ project 576/RU/12 SUPPLEMENTARY COMPARISON STANDARDS IN THE FIELD OF ANALYSIS OF GAS MIXTURES CONTAINING СО 2, СО, С 3 Н 8 IN NITROGEN ( AUTOMOTIVE
More informationTitle: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)
Procedure No: SOP-026 Revision No: 1.0 January 24, 2011 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to meet
More informationGas monitoring using GASMET FTIR
Gas monitoring using GASMET FTIR Dominic Duggan Quantitech dd@quantitech.co.uk Andrew Hobson Quantitech Antti Heikkila Gasmet Outline of the workshop: FTIR for emissions measurement Applications in ambient
More informationGAS CHROMATOGRAPHY MAINTENANCE USING UNCERTAINTY BASED CBM
GAS CHROMATOGRAPHY MAINTENANCE USING UNCERTAINTY BASED CBM Anwar Sutan, i-vigilant Technologies Paul Daniel, i-vigilant Technologies 1 INTRODUCTION On-line gas chromatography is frequently used within
More informationResearch and Development. Money in the Pipeline
On-line gas chromatography Gerard Nieuwenkamp Research and Development Money in the Pipeline Delft, 2 November 2010 Natural gas metering 2" 2" 65 m3/h usm 6" 400 m3/h M etering runs 2500 m3/h Gas Oil Piston
More informationProtocol Gas Verification Program (PGVP)
Protocol Gas Verification Program (PGVP) Michael Hayes ICAC/ MARAMA Workshop Advances in Control Technologies July 9-10, 2008 Exton, PA PGVP Protocol Gases from an historic perspective Historical Audit
More informationResearch in the waste area
Prof. ing. Marco Ragazzi Department of Civil and Environmental Engineering, University of Trento Municipal Solid Waste as is MSW residual of selective collection under-sieve and over-sieve from MSW mechanical
More informationInternational comparison CCQM-K82: Methane in Air at Ambient level ( ) nmol/mol. (Final report)
Version 1 7 November 2014 International comparison CCQM-K82: Methane in Air at Ambient level (1800-2200) nmol/mol (Final report) Edgar Flores *1, Joële Viallon 1, Tiphaine Choteau 1, Philippe Moussay 1,
More informationDemystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.
Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Jamie Brown R&D Scientist Supelco 595 North Harrison RD. Bellefonte, PA 16823 National
More informationCertification of AMS acc. EN 15267, Part 3 - Overview and First Experience -
Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience - Dr. Wolfgang Jockel, Martin Schneider, TÜV Rheinland Group, D-51105 Cologne / Germany 1. Introduction A new basis for the certification
More informationEasyLine Continuous Gas Analyzers. So smart, they re simple
EasyLine Continuous Gas Analyzers So smart, they re simple Reliable Economic Powerful EasyLine is both a powerful and affordable line of instruments for the monitoring of gas concentrations in numerous
More informationGases and welding distributors continue
Specialty Gas Instrumentation By Ray Borzio, Mike Lee, and Scott Varian Gases and welding distributors continue to have a significant future in the production, delivery, and related services for specialty
More informationWeiyi Zheng, Xiaoxia Dong, Evan Rogers, Jimmie C. Oxley, and James L. Smith*
Improvements in the Determination of Decomposition Gases from 1,3,3-Trinitroazetidine and 5-Nitro-2,4-dihydro-3H-1,2,4,-trizol-3-one using Capillary Gas Chromatography-Mass Spectrometry Weiyi Zheng, Xiaoxia
More informationINSTRUMENT INSTRUMENT YL INSTRUMENT
INSTRUMENT INSTRUMENT INSTRUMENT Nitrogen Generator Whisper-N2 Series Features Membrane Technology Compressed air is forced through a hollow fiber membrane, with selective permeation of the different components
More informationCommercial Practice Test Method Internal Vapor Analysis of Hermetic Devices
Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Oneida Research Services, Inc. 8282 Halsey Road Whitesboro, NY 13492 Phone: (315) 736-5480 FAX: (315) 736-9321 1.0 Purpose The
More informationGas-Detection Instruments
Gas-Detection Instruments Regulators Regulators are used to ensure you receive the required flow of calibration gas for your application. MSA regulators save customers money by delivering calibration test
More informationSuperior Result from Analytical Instruments The constant purity of the nitrogen improves system stability and ensures reproducible results
Nitrogen Generator Whisper-N2 Series Features Membrane Technology Superior Result from Analytical Instruments The constant purity of the nitrogen improves system stability and ensures reproducible results
More informationAPPENDIX C. 40 CFR PART 60 App. B SPEC. 1-4, 6
APPENDIX C 40 CFR PART 60 App. B SPEC. 1-4, 6 Part 60, Appendix B - Performance Specification 2 Specifications and Test Procedures for SO2 and NOx Continuous Emission Monitoring Systems in Stationary Sources
More informationCALCULATING THE SPEED OF SOUND IN NATURAL GAS USING AGA REPORT NO Walnut Lake Rd th Street Houston TX Garner, IA 50438
CALCULATING THE SPEED OF SOUND IN NATURAL GAS USING AGA REPORT NO. 10 Jerry Paul Smith Joel Clancy JPS Measurement Consultants, Inc Colorado Engineering Experiment Station, Inc (CEESI) 13002 Walnut Lake
More informationRapid and Reliable Detection of Dissolved Gases in Water
Rapid and Reliable Detection of Dissolved Gases in Water Andrea Caruso and Massimo Santoro Thermo Fisher Scientific, Milan, Italy Application Note 005 Key Words Chromeleon CDS, Environmental, Fracking,
More informationOrganic Elemental Analysis Series II Nitrogen Analyzer. proven performance. reliable results
Organic Elemental Analysis 2410 Series II Nitrogen Analyzer proven performance reliable results PerkinElmer 2410 Series II Nitrogen Analyzer Advanced Combustion Method Q U I C K G L A N C E Advanced combustion
More informationAutomated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek
Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek Abstract: Due to the expansion of natural gas drilling through horizontal fracturing, there has been
More informationAND EQUIPMENT FOR ENVIRONMENTAL APPLICATIONS (800) B Polk Street Houston, TX
CALIBRATION GASES AND EQUIPMENT FOR ENVIRONMENTAL AND INDUSTRIAL HYGIENE APPLICATIONS (800) 548-2268 6717-B Polk Street Houston, TX 77011 www.portagas.com IS DEDICATED TO PROVIDING INNOVATIVE CALIBRATION
More informationUltimate Performance in Gas Chromatography
Ultimate Performance in Gas Chromatography by Randall Bramston-Cook Lotus Consulting 5781 Campo Walk Long Beach, California 90803 310/569-0128 Fax 714/898-7461 Email ebramstoncook@msn.com August 7, 2007
More informationASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist
ASTM WK 43267 Standard Test Method for Dissolved Gases Anne Jurek Applications Chemist Rationale Hydraulic Fracturing is becoming more and more common. Nearby water well can be affected during the fracking
More informationGas Chromatographic Applications of the Dielectric Barrier Discharge Detector
Gas Chromatographic Applications of the Dielectric Barrier Discharge Detector Matthew Monagle, Advanced Industrial Chemistry Corp*. Ronda Gras, Jim Luong, Dow Chemical Canada 28th International Symposium
More informationOxygen Analyser And Monitor Viamed
Oxygen Analyser And Monitor Viamed 1 / 7 2 / 7 3 / 7 Oxygen Analyser And Monitor Viamed Viamed Ltd, established in 1977, has built a strong reputation for the design, manufacture and distribution of a
More informationLFE OEM TCD - Thermal Conductivity Detector
LFE OEM TCD - Thermal Conductivity Detector Key features Quick response - T90 3 sec Extremely suppressed ranges High corrosion resistance High temperature capability up to 180 C Infallible containment
More informationHot extractive multi-channel gas analyser for continuous process and emissions monitoring.
Product Data Sheet CODEL Continuous Emission Monitoring GCEM4100 Extractive Gas Analyser CO, CO2, NO, NO2, SO2, HCL, CH4 & H2O Hot extractive multi-channel gas analyser for continuous process and emissions
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 16017-1 First edition 2000-11-15 Indoor, ambient and workplace air Sampling and analysis of volatile organic compounds by sorbent tube/thermal desorption/capillary gas chromatography
More informationLUMAZOTE T ECHNICAL BRO CHU RE
by Chemiluminescence Standards AST D4629 / ASTM D6069 / ASTM D 5176 / ASTM D 5762 / NF EN 12260 / NF M 07-058 Licence TOTAL LUMAZOTE T ECHNICAL BRO CHU RE ERALY & Associés Phone: +33 (0) 1 77 04 80 97
More informationInternational Comparison CCQM-K111 Propane in nitrogen
International Comparison CCQM-K111 Propane in nitrogen Adriaan M.H. van der Veen 1, J. Wouter van der Hout 1, Paul R. Ziel 1, Rutger J. Oudwater, Andreia L. Fioravante, Cristiane R. Augusto, Mariana Coutinho
More informationSystem Overview. TCD Detector temperature setpoint Regulator
System Overview POPULAR CONFIGURATION GCs Your educational TCD GC is configured on the compact 310 chassis. It is equipped with a TCD Detector, a temperature programmable Column Oven, a 3 Silica Gel packed
More informationGas Mixture Two Components. Gas Mixtures Two Components
Gas Mixture Two Components Two Components Contents Two-Component Mixtures Ammonia...M4 Argon...M5 Benzene...M6 n-butane...m7 Carbon Dioxide...M8 Carbon Monoxide...M10 Chlorine...M12 Ethane...M13 Ethylene
More informationMeasurement of Representative Landfill Gas Migration Samples at Landfill Perimeters: A Case Study
Measurement of Representative Landfill Migration Samples at Landfill Perimeters: A Case Study Breda M. Kiernan, PhD. National Centre for Sensor Research, Dublin City University Glasnevin Dublin 9 Ireland
More informationSPECIFICATION. date of manufacture. Intermittent: -20 o C to +50 o C. Pressure Range:
The ECO-Sure (2e) is a high quality, cost effective 2-electrode electromechanical cell designed for the detection of Carbon Monoxide CO in a range of applications but particularly for domestic/residential
More informationPractical approach and problems in in-situ RGA calibration
Practical approach and problems in in-situ RGA calibration Oleg Malyshev and Keith Middleman Vacuum Science Group, ASTeC Accelerator Science and Technology Centre STFC Daresbury Laboratory UK Workshop
More informationThis test shall be carried out on all vehicles equipped with open type traction batteries.
5.4. Determination of hydrogen emissions page 1 RESS-6-15 5.4.1. This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4.2. The test shall be conducted following
More informationEuroFID Total Hydrocarbon Analyzer. Precise Determination of Total Hydrocarbons in Air for Corrosive as well as Condensing Gases
P r o d u c t i n f o r m at i o n EuroFID Total Hydrocarbon Analyzer Precise Determination of Total Hydrocarbons in Air for Corrosive as well as Condensing Gases Proven Analyzer Technology Measuring total
More informationRetention Time Locking: Concepts and Applications. Application
Retention Time Locking: Concepts and Applications Application Gas Chromatography December 1997 Authors Vince Giarrocco Bruce Quimby Matthew Klee Agilent Technologies, Inc. 2850 Centerville Road Wilmington,
More informationHot extractive multi-channel gas analyser for continuous process and emissions monitoring.
Product Data Sheet CODEL Continuous Emission Monitoring GCEM4100 Extractive Gas Analyser CO, CO2, NO, NO2, SO2, HCL, CH4 & H2O Hot extractive multi-channel gas analyser for continuous process and emissions
More informationThe Role of Gas Monitoring in the Prevention and Treatment of Mine Fires
University of Wollongong Research Online Coal Operators' Conference Faculty of Engineering and Information Sciences 28 The Role of Gas Monitoring in the Prevention and Treatment of Mine Fires D. Brady
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares
More informationBackground Statement for SEMI Draft document 4657B NEW STANDARD: SPECIFICATIONS FOR TUNGSTEN HEXAFLUORIDE (WF 6 )
Background Statement for SEMI Draft document 4657B NEW STANDARD: SPECIFICATIONS FOR TUNGSTEN HEXAFLUORIDE (WF 6 ) Note: This background statement is not part of the balloted item. It is provided solely
More information