Final Report of APMP.QM-K46 Ammonia in Nitrogen at 30 µmol/mol Level

Size: px
Start display at page:

Download "Final Report of APMP.QM-K46 Ammonia in Nitrogen at 30 µmol/mol Level"

Transcription

1 Final Report of APMP.QM-K46 Ammonia in Nitrogen at 30 µmol/mol Level Shinji Uehara 1), HAN Qiao 2), Takuya Shimosaka 3)* 1) Chemicals Evaluation and Research Institute, Japan (CERI) 2) National Institute of Metrology (NIM), No.18, Bei-San-Huan Dong Str., Beijing , China 3) National Metrology Institute of Japan (NMIJ), AIST Tsukuba Central 3, 1-1-1, Umezono, Tsukuba, Ibaraki , Japan *) Coordinator of this comparison Field Amount of Substance Subject Comparison of 30 µmol/mol ammonia in nitrogen Participants CERI (Japan), NIM (China) Coordinating Laboratory National Metrology Institute of Japan (NMIJ) Organizing Body APMP 1 / 19

2 Introduction Ammonia is an important compound in the chemical industry. It is widely used and is the basis for producing other compounds containing nitrogen. Ammonia is also very hazardous, and consequently emissions of ammonia need be controlled and monitored. VSL coordinated CCQM-K46 (30 µmol/mol - 50 µmol/mol, NH 3 in nitrogen) in 2006 [1]. CERI participated in the comparison, but NIM did not participate in CCQM-K46. NIM would like to compare the gas standard with CERI through the bilateral comparison. Initially, analytical comparison between CERI and NIM was planned and CERI gravimetrically prepared two primary gas mixtures of NH 3 in N 2 for this comparison. However the cylinder was not able to be transported to NIM. Difficulty of cylinder transport often occurs. Therefore, the protocol of this comparison was changed where NIM will send their cylinder to CERI for analysis. This report describes the results for a bilateral comparison of ammonia in nitrogen gas mixture. The nominal amount-of-substance fraction was 30 µmol/mol. Protocol NIM prepared one gas mixture sample and sent it to CERI for analysis. The amount-of-substance fraction NH 3 in N 2 should be between 30 µmol/mol and 50 µmol/mol. The pressure of the gas mixture should be greater than 10 MPa. A cylinder with an inner volume of 10 dm 3 or larger one should have been used. CERI analyses the sample from NIM against CERI s primary gas mixtures. The sample was returned to NIM after the measurement at CERI. Because CERI participated in CCQM-K46 and the result was consistent with the KCRV, the key comparison reference value, amount-of-substance fraction as obtained from gravimetry at CERI is used as a reference value of this comparison. The participating laboratory is responsible for the calculation of the amount-of-substance fraction ammonia from gravimetric preparation and its uncertainty. For a proper evaluation of the data, it is necessary that purity analysis of raw gases, evaluation of gravimetric preparation process and stability were done properly at each NMI. After transfer of the cylinder to NIM, NIM had to reanalyse the returned cylinder for stability check (pressure at least 5 MPa). The stability data is to be combined into stability data for uncertainty evaluation in the report. Schedule Shipping to CERI : February, / 19

3 Shipping to NIM : November, 2015 Draft A : March, 2016 Draft B :June, 2016 Analysis and s The gas mixture was analysed by an NDIR calibrated by three standard gas mixtures of ammonia in nitrogen prepared by the gravimetric method according to ISO 6142 and ISO 6143 [2, 3]. The NDIR was a specific model of Shimadzu CGT-7000, modified for ammonia analysis. The measurement method of the ammonia fraction in the APMP.QM-K46 sample is the same as the method that CERI did in CCQM-K46. The experimental conditions for this comparison are summarized in Table 1. The amount-of-substance fractions of the standard gas mixtures, RM1, RM2 and RM3, and their uncertainties are summarized in Table 2. Table 1 Experimental conditions of NDIR for determination of NH 3 in N 2. Condition Analyzer NDIR (CGT-7000, Shimadzu) Sample flow rate 1 L/min Table 2 Standard gas mixtures of ammonia in nitrogen of CERI. amount fraction Standard uncertainty RM RM RM The APMP sample and the primary standard gas mixtures were measured in the following order; RM1 RM2 Sample RM3. This sequence was done 3 times within one day. Tables 3 and 4 show the results of this comparison, where x NIM is the NIM s reported value of amount-of-substance fraction of ammonia in the sample. Table 3 Reported value of NIM Cylinder number x NIM U(x NIM ) Coverage Factor CPE / 19

4 Table 4 s of determination of ammonia in the sample Ammonia molar fraction Cylinder number Average Standard uncertainty CPE The results shown in Table 4 are determined by the same method used in CCQM.K46 and the standard gas mixtures of ammonia in nitrogen prepared by the gravimetric method, and therefore the averaged value can be considered as a KCRV of this comparison. A degree of equivalence (D APMP ) is defined as the difference between NIM s reported value and the KCRV. Δx = D APMP = x NIM x KCRV U(D APMP ) 2 = U 2 (x NIM ) + U 2 (x KCRV ) Table 5 and Figure 1 show the result of this comparison. Table 5 Reuslts of this comparison x NIM U(x NIM ) x KCRV U(x KCRV ) D APMP U(D APMP ) (k = 2) NMI NIM Degree of Equivalence NIM CERI Figure 1 Degree of equivalence of APMP.QM-K46. 4 / 19

5 Linking APMP.QM-K46 to CCQM-K46 The results of the CCQM-K46 key comparison are summarized in Table 6. The degree of equivalence and its uncertainty of APMP.QM-K46 are transferred to the CCQM key comparison with the equations given as D i = D APMP + D CCQM, U(D i ) = U(D APMP ), where D CCQM is the degree of equivalence for CERI s result of CCQM-K46. The result of this comparison is added to Table 6, and Figure 2 shows the degree of equivalence of CCQM-K46 and APMP.QM-K46. Table 6 Summary of results of CCQM-K46 and APMP.QM-K46 Laboratory x prep u(x prep ) x lab U(x lab ) D i U(D i ) µmol/ mol µmol/mol µmol/mol VSL NPL VNIIM METAS NIST KRISS CERI x prep u(x prep ) x KCRV U(x KCRV ) D i U(D i ) NIM / 19

6 3 2.5 Degree of Equivalence CCQM-K46 APMP.QM- K46-2 VSL NPL VNIIM METAS NIST KRISS CERI NIM CERI Participants Figure 2 s of CCQM-K46 and APMP.QM-K46. Conclusion This bilateral comparison was carried out to support NIM s capability for the measurement of ammonia in nitrogen at a level of 30µmol/mol. In this key comparison, the result of the participant, NIM is consistent with the KCRV within the uncertainty. Therefore, this key comparison supports the capability of ammonia in nitrogen or air in the range of 30 to 100. Reference [1] Adriaan M.H. van der Veen et al, International comparison CCQM-K46: Ammonia in nitrogen, Metrologia, 47, Tech. Suppl (2010). [2] International Organization for Standardization, ISO 6142:2001 Gas analysis - Preparation of calibration gas mixtures - Gravimetric methods, 2nd edition. [3] International Organization for Standardization, ISO 6143:2001 Gas analysis Comparison methods for determining and checking the composition of calibration gas mixtures. 6 / 19

7 Annex A Report Form APMP.QM-K46 Ammonia in nitrogen Laboratory name: Chemicals Evaluation and Research Institute, Japan Cylinder number: CPE Measurement #1 Component (dd/mm/yy) Standard deviation number of replicates NH3 3/6/ Measurement #2 Component (dd/mm/yy) Standard deviation number of replicates NH 3 4/6/ Measurement #3 Component (dd/mm/yy) Standard deviation number of replicates NH 3 5/6/ Measurement #4 Component (dd/mm/yy) Standard deviation number of replicates NH 3 8/6/ s Component Expanded Uncertainty Coverage factor 1 NH The coverage factor shall be based on approximately 95% confidence. 7 / 19

8 Reference Method: Instruments for NH 3 measurement Principles : NDIR (Type:CGT-7000, Make : Shimadzu corporation) Data collection : integrator Calibration Standards: Preparation : Gravimetric method Purity analysis ; NH 3, N 2 : Impurities in NH 3 and N 2 are determined by analyses and the amount of the major component is conventionally determined by, N X pure = 1 x i i= 1 where: x I = mole fraction of impurity i, determined by analysis N = number of impurities X pure = mole fraction purity of the pure parent gas Instrument Calibration: Component Table 1 Composition of PSMs amount-of-substance fraction ( µmol/mol ) R 1 R 2 R 3 NH Calibration method Analytical scheme was, R 1 R 2 APMP Sample R 3. This scheme was repeated 3-times in a day. These measurements were carried out for 4-days. 8 / 19

9 Uncertainty: Uncertainty source Measurement R 1 R 2 R 3 Round off total Value +/ Method of evaluation (type A or typeb) Assumed probability distribution Divisor A 1 A 1 A 1 A 1 B rectangular 3 Standard uncertainty u(x i ) Sensitivity coefficient c i Contribution u ⅰ (y) Coverage factor: 2 Expanded uncertainty: 0.17 µmol/mol Authorship Shinji UEHARA 9 / 19

10 National Institute of Metrology (NIM), China Transmission of International Comparison s The title of international comparison: Ammonia in nitrogen Serial number for international comparison: APMP.QM-K46 Comparison experiment period: Jan 2015~Aug 2015 Experiment reporter: HAN Qiao Phone number of experiment reporter: hanqiao@nim.ac.cn NIM address: No.18, Bei-San-Huan Dong Str., Beijing , China Phone/Fax number of Department of Metrology Services: Phone/Fax number of Department of R&D and Planning (International Cooperation): yw@nim.ac.cn Transmission date: December 4, / 19

11 Key Comparison APMP.QM-K46 Ammonia in nitrogen 1. General information Institute Address Contact person National Institute of Metrology (NIM), China No.18, Bei-San-Huan Dong Str., Beijing , China Mr. Qiao Han, Mr. Defa Wang Telephone Fax SN of cylinders Cylinder pressure CPE Mpa 2. s Cylinder No. Methane mole fraction x NH 3 /µmol/mol Expanded uncertainty U ( x NH 3) /µmol/mol Coverage factor CPE Purity data of parent gases Table 1. Purity table of parent gas N2, pure, A655029# fraction Std. u Technique Ar(Argon) GC-PDHID CH 4 (Methane) GC-FID-Cat.** CO(Carbon monoxide) GC-FID-Cat.** CO 2 (Carbon dioxide) GC-FID-Cat.** H 2 (Hydrogen) GC-PDHID H 2 O(Water) CRDS O 2 (Oxygen) Oxygen analyzer N 2 (Nitrogen) NH3, pure, MK-20385# fraction Std. u Technique Ar(Argon) GC-PDHID CH 4 (Methane) GC-PDHID H 2 O(Water) Product Spec.* O 2 (Oxygen) GC-PDHID N 2 (Nitrogen) GC-PDHID 11 / 19

12 NH 3 (Ammonia) * Product Spec., data was from the product specification provided by the manufacturer. ** GC-FID-Cat., GC-FID with methanator catalyst. 4. Gravimetric preparation of gas mixtures The standard gas mixtures of ammonia in nitrogen were prepared by using gravimetric method according to ISO6142:2001, and the parent gases were nitrogen, and ammonia of high purity. Figure 1 showed the dilution chart from pure gases to the final gas mixtures of NH 3 /N 2. Pure NH3 Pure N g, u=0.005g CPE NH3= E-03 mol/mol g, u=0.020g g, u=0.005g CPE NH3= E-05 mol/mol g, u=0.020g Figure 1. Dilution scheme from pure gases to NH 3 /N 2 Mass comparator with capacity of 26kg and resolution of 1mg is provided by Mettler Toledo. Temperature and relative humidity in balance room were controlled at 20 C±1 C and 50%RH±10%RH, respectively. Tare cylinder and substitution method were used during weighing of cylinder, by automatic weighing system, in order to cancel buoyancy effect. Fig.1 showed that standard uncertainty to the mass of parent gas added into the target cylinder was estimated as 0.005g when the added gas was around 3g~6g. By 2-step dilutions, around 33ppm ammonia in nitrogen could be achieved from the pure gases. 5. Measurement The new sample cylinder was verified against our PRM cylinder by using a photoacoustic infrared spectra ammonia analyzer. The gas pressure from each sample cylinders was reduced by a regulator, and then the sample gas was introduced into analyzer by using a pump installed inside the instrument. Sample gas flow rate was around 500mL/min. 12 / 19

13 Measurement repeatability is about 0.05ppm. PRMs used in measurement are listed in table 2. These gas mixtures were prepared by gravimetric method in our lab, which was described in 4. Gravimetric preparation of gas mixtures. The certified value and uncertainty of PRMs included the contribution from gravimetric method and short time stability. Table 2. Certified value and uncertainty of PRMs Cylinder Number Certified value µmol/mol Standard uncertainty µmol/mol CPE CPE Single point calibration method was used to calculate the concentration of ammonia in sample cylinder. The sample gas and reference gases were measured in the order of Reference Sample Reference alternatively. The sample gas mixture was prepared in 18/12/2014 and measured against our calibration reference mixtures 4 days after the preparation, for the first time. Half month later the second measurement was carried out. The results are in table 3 and table 4. Table 3. s of first measurement Relative standard Number of repeat (dd/mm/yy) deviation 22/12/ % 6 23/12/ % 6 24/12/ % 6 Average Table 4. s of second measurement Relative standard Number of repeat (dd/mm/yy) deviation 5/1/ % 6 6/1/ % 6 7/1/ % 6 Average The uncertainty of each measurement was evaluated according to the following equation: 2 uu mmmmmm = uu rrrrrr + uu2 aaaaaa Where, u mea is the standard uncertainty of each measurement, u ref is the uncertainty contribution from PRMs, u ana is the uncertainty contribution from analysis method. The uncertainty sources of analysis method are listed in table / 19

14 Table 5. Uncertainty sources of analysis method Uncertainty sources Relative standard uncertainty Distribution Analyzer reading repeatability of 0.16% Normal sample gas mixture Analyzer reading repeatability of reference gas mixture 0.16% Normal Repeatability of measurement 0.12% Normal results The relative standard uncertainty contribution of this analysis method was 0.256%, which led to u ana =0.086 µmol/mol for this sample gas mixture. Table 2 shows u ref =0.088 µmol/mol. So, the standard uncertainty of each measurement was about µmol/mol and the expanded uncertainty was about 0.25 µmol/mol (k=2). 6. Stability In order to study the stability of sample gas mixture, it was re-analyzed when arriving at NIM after the measurement in CERI, the coordinating laboratory. The re-analyzed results were listed in table 6. Table 7 and figure 2 are the summary of stability study, which shows about 0.32 µmol/mol decreasing happened during about 12 months period. Table 6. s of re-analysis Relative standard Number of repeat (dd/mm/yy) deviation 26/11/ % 6 27/11/ % 4 1/12/ % 5 Average Table 7. Summary of stability test Time of test Test result Expanded uncertainty (k=2) December of January of November of / 19

15 Test results µmol/mol Month after preparation Figure2. Stability test during 12 month 7. Certified value and uncertainty of sample gas mixture The certified value of ammonia in sample cylinder CPE was determined according to the following equation: XX = XX 0 + XX Where, X is the certified value, X 0 is the measurement result (=33.62 µmol/mol) after the sample gas mixture preparation in December of 2014 and X is the average variation of ammonia concentration in 12 months. The stability study showed that the variation of ammonia concentration was from µmol/mol to µmol/mol. According to rectangular distribution, X =-0.16 µmol/mol and u( X) was 0.10 µmol/mol. So, the certified value of sample gas mixture was µmol/mol. The standard uncertainty and expanded uncertainty of X were estimated by following equations: uu(xx) = uu 2 (XX 0 ) + uu 2 ( XX) UU(XX) = kk uu(xx), kk = 2 So, the expanded uncertainty of certified value was 0.32 µmol/mol. 15 / 19

16 Annex B APMP.QM-K46 Protocol for Bilateral Comparison on Ammonia in Nitrogen between NIM and CERI coordinated by NMIJ Introduction Ammonia is an important compound in the chemical industry. It is widely used and is the basis for producing other compounds containing nitrogen. Ammonia is also very hazardous, and consequently emissions of ammonia need be controlled and monitored. VSL coordinated CCQM-K46 (30 µmol/mol - 50 µmol/mol, NH 3 in N 2 ) in CERI participated in the comparison, but NIM did not participate in CCQM-K46. NIM would like to compare the gas standard with CERI through the bilateral comparison. Initially, analytical comparison between CERI and NIM was planned and CERI prepared two primary gas mixtures of NH 3 in N 2 for this comparison, but the cylinder was not transported to NIM. Difficulty of cylinder transport often occurs. Therefore, the protocol of this comparison was changed where NIM will send their cylinder to CERI. Participants CERI (Chemicals Evaluation and Research Institute, Japan) and NIM (National Institute of Metrology, China). Method of the Bilateral Comparison NIM will prepare one gas mixture sample and send it to CERI. Concentration of NH 3 in N 2 should be between 30 µmol/mol and 50 µmol/mol. The pressure in the cylinders should be more than 10 MPa. Cylinder of 10 dm 3 nominal or larger one should be used. CERI will measure the cylinder from NIM against CERI s primary gas mixtures. The cylinder will be sent back to NIM after the measurement at CERI. CERI participated in CCQM-K46 and the result was consistent with the KCRV, key comparison reference value. The amount-of-substance fractions as obtained from gravimetry at CERI will be used as reference value. Each cylinder will have its own reference value. Each participating laboratory will be responsible for the calculation of the gravimetric concentration and its uncertainty. For a proper evaluation of the data, it is necessary that purity analysis of raw gases, evaluation of gravimetric preparation process and stability 16 / 19

17 are done properly at each NMI. After the transfer of cylinder to NIM, NIM has to measure the returned cylinder for stability check (pressure at least 5 MPa), the stability data is to be combined into stability data for uncertainty evaluation in the report. NIM will bear cost for all transportations. Transport of the cylinder from NIM to CERI and CERI to NIM will be arranged by NIM. Schedule (original) February 2011 April 2011 June 2011 July 2011 August 2011 September 2011 Preparation and verification of gas mixture in CERI Transport to NIM (8MPa or more) Measurement of gas mixture in NIM Return of cylinder to CERI (5 MPa or more) Verification of gas mixture in CERI Report to NMIJ Draft A Report Draft B Report Coordinator Dr. Kenji Kato National Metrology Institute of Japan (NMIJ) AIST Tsukuba Central 3, 1-1-1, Umezono, Tsukuba, Ibaraki , Japan Phone : Fax : k.kato@aist.go.jp Contact person of CERI Mr. UEHARA Shinji Chemicals Evaluation and Research Institute, Japan (CERI) 1600 Shimotakano, Sugito-machi, Kitakatsushika-gun, Saitama , Japan Phone : Fax : / 19

18 Report Form APMP.QM-K46 Ammonia in nitrogen Laboratory name: Cylinder number: Purity analysis of NH 3 Component Standard deviation number of replicates (dd/mm/yy) (mol/mol) (% relative) Ammonia Purity analysis of N 2 Component Standard deviation number of replicates (dd/mm/yy) (mol/mol) (% relative) Ammonia Gravimetric concentration of NH 3 in N 2 Component Standard deviation number of replicates (dd/mm/yy) (mol/mol) (% relative) Ammonia 18 / 19

19 Assigned concentration and uncertainty of the sample Component Expanded Uncertainty Coverage factor 2 (mol/mol) Purity of NH 3 Purity of N 2 Gravimetry Stability 2 The coverage factor shall be based on approximately 95% confidence. 19 / 19

Comparison APMP.QM-S2.1 Oxygen in nitrogen at atmospheric level

Comparison APMP.QM-S2.1 Oxygen in nitrogen at atmospheric level APMP.QM-S2.1 report Comparison APMP.QM-S2.1 Oxygen in nitrogen at atmospheric level Final Report ByungMoon Kim 1, Kwangsub Kim 1, Jinsang Jung 1,*, Sanghyub Oh 1, Liu Hui 2, Hou Li 2, Teo Beng Keat 2,

More information

COOMET TECHNICAL COMMITTEE 1.8 PHYSICAL CHEMISTRY

COOMET TECHNICAL COMMITTEE 1.8 PHYSICAL CHEMISTRY COOMET TECHNICAL COMMITTEE 1.8 PHYSICAL CHEMISTRY CООМЕТ PROJECT 664/RU/14 SUPPLEMENTARY COMPARISON OF STANDARD GAS MIXTURES: «Atmospheric air pollutants: CO in Nitrogen, 5 µmol/mol - Technical Protocol,

More information

International comparison CCQM-K82: Methane in Air at Ambient level ( ) nmol/mol. (Final report)

International comparison CCQM-K82: Methane in Air at Ambient level ( ) nmol/mol. (Final report) Version 1 7 November 2014 International comparison CCQM-K82: Methane in Air at Ambient level (1800-2200) nmol/mol (Final report) Edgar Flores *1, Joële Viallon 1, Tiphaine Choteau 1, Philippe Moussay 1,

More information

COOMET TECHNICAL COMMITTEE 1.8 PHYSICAL CHEMISTRY. CООМЕТ project 576/RU/12

COOMET TECHNICAL COMMITTEE 1.8 PHYSICAL CHEMISTRY. CООМЕТ project 576/RU/12 COOMET TECHNICAL COMMITTEE 1.8 PHYSICAL CHEMISTRY CООМЕТ project 576/RU/12 SUPPLEMENTARY COMPARISON STANDARDS IN THE FIELD OF ANALYSIS OF GAS MIXTURES CONTAINING СО 2, СО, С 3 Н 8 IN NITROGEN ( AUTOMOTIVE

More information

International Comparison CCQM-K111.1 Propane in nitrogen

International Comparison CCQM-K111.1 Propane in nitrogen International Comparison CCQM-K111.1 Propane in nitrogen Adriaan M.H. van der Veen 1, J. Wouter van der Hout 1, Paul R. Ziel 1, Mudalo Jozela, James Tshilongo, Napo G. Ntsasa, Angelique Botha 1 Van Swinden

More information

Sangil LEE, Mi Eon KIM, Sang Hyub OH, Jin Seog KIM. Center for Gas Analysis Korea Research Institute of Standards and Science (KRISS) Daejeon, Korea

Sangil LEE, Mi Eon KIM, Sang Hyub OH, Jin Seog KIM. Center for Gas Analysis Korea Research Institute of Standards and Science (KRISS) Daejeon, Korea Determination of physical adsorption loss of primary reference gas mixtures in cylinders using cylinder-to-cylinder division Lee et al., 2017 Metrologia 54 L26) Sangil LEE, Mi Eon KIM, Sang Hyub OH, Jin

More information

Report on Co-production of CH 4 /air within ACRM framework

Report on Co-production of CH 4 /air within ACRM framework Report on Co-production of CH 4 /air within ACRM framework 8 th APMP/TCQM Gas CRM workshop NMIJ/AIST, Tsukuba, 9~11 June 2010 By Hai Wu, NIM 1 Brief Introduction to ACRM ACRM Concept Asian Collaboration

More information

Date of Shipment: Xxxxx 00, 20xx 3402c National Institute of Advanced Industrial Science and Technology National Metrology Institute of

Date of Shipment: Xxxxx 00, 20xx 3402c National Institute of Advanced Industrial Science and Technology National Metrology Institute of Date of Shipment: Xxxxx 00, 20xx 3402c01151112-170831 National Metrology Institute of Japan Reference Material Certificate NMIJ CRM 3402-c01 Sulfur Dioxide This certified reference material (CRM) was produced

More information

METROLOGICAL ASSURENCE OF O3, CO2, CH4 AND CO CONTROL IN ATMOSPHERE

METROLOGICAL ASSURENCE OF O3, CO2, CH4 AND CO CONTROL IN ATMOSPHERE METROLOGICAL ASSURENCE OF O3, CO2, CH4 AND CO CONTROL IN ATMOSPHERE Leonid A. Konopelko Yuri A. Kustikov Vitaly V. Beloborodov Dmitry N. Selukov D.I.Mendeleyev Institute for Metrology (VNIIM), 19 Moskovsky

More information

Development of DMS and Acetonitrile, and Formaldehyde gas standards. Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin

Development of DMS and Acetonitrile, and Formaldehyde gas standards. Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin Development of DMS and Acetonitrile, and Formaldehyde gas standards Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin The amount of chemical substances in cylinder container Purity uncertainty Gravimetry

More information

International comparison Refinery gas (CCQM-K77) Final Report

International comparison Refinery gas (CCQM-K77) Final Report Blad 1 van 71 International comparison Refinery gas (CCQM-K77) Final Report Page 2 of 71 VSL Thijsseweg 11 2629 JA Delft Postbus 654 2600 AR Delft Nederland T +31 15 269 15 00 F +31 15 261 29 71 E info@vsl.nl

More information

Final Report. International Key Comparison CCQM-K94

Final Report. International Key Comparison CCQM-K94 Final Report International Key Comparison CCQM-K94 (10 mol/mol Dimethyl sulfide in nitrogen) Sangil Lee 1*, Gwi Suk Heo 1, Yongdoo Kim 1, Sanghyub Oh 1, Qiao Han 2, Hai Wu 2, L.A. Konopelko 3, Y.A. Kustikov

More information

International Comparison CCQM-K111 Propane in nitrogen

International Comparison CCQM-K111 Propane in nitrogen International Comparison CCQM-K111 Propane in nitrogen Adriaan M.H. van der Veen 1, J. Wouter van der Hout 1, Paul R. Ziel 1, Rutger J. Oudwater, Andreia L. Fioravante, Cristiane R. Augusto, Mariana Coutinho

More information

International comparison CCQM-K84. Carbon monoxide in Synthetic air at ambient level

International comparison CCQM-K84. Carbon monoxide in Synthetic air at ambient level CCQM-K Final Report International comparison CCQM-K Carbon monoxide in Synthetic air at ambient level 1 1 1 1 1 1 1 1 0 1 0 Jeongsoon Lee 1, Dongmin Moon 1, Jinbok Lee 1, Jeongsik Lim 1, Brad Hall, Paul

More information

Gravimetric preparation of NO 2 primary reference gas mixtures at the NMISA. Tshepiso Mphamo 03 September 2012

Gravimetric preparation of NO 2 primary reference gas mixtures at the NMISA. Tshepiso Mphamo 03 September 2012 Gravimetric preparation of NO 2 primary reference gas mixtures at the NMISA Tshepiso Mphamo 03 September 2012 NMISA 2012 Contents Introduction Experimental Verification Uncertainty CCQM- Pilot study results

More information

International Comparison CCQM K53 Oxygen in Nitrogen

International Comparison CCQM K53 Oxygen in Nitrogen CCQM-K53 Final Report International Comparison CCQM K53 Oxygen in Nitrogen Jeongsoon Lee 1, Jin Bok Lee 1, Dong Min Moon 1, Jin Seog Kim 1*, Adriaan M.H. van der Veen 2, Laurie Besley 3, Hans-Joachim Heine

More information

Version /12/11. Final Report. Federal Institute for Materials Research and Testing (BAM), Germany.

Version /12/11. Final Report. Federal Institute for Materials Research and Testing (BAM), Germany. International comparison CCQM-K74: Nitrogen dioxide, 10 μmol/mol Final Report Edgar Flores *1, Faraz Idrees 1, Philippe Moussay 1, Joële Viallon 1, Robert Wielgosz 1, Teresa Fernández, Sergio Ramírez,

More information

Improved Reliability in Natural Gas Energy Measurement with the Revised ISO Standard. Gerard Nieuwenkamp - VSL

Improved Reliability in Natural Gas Energy Measurement with the Revised ISO Standard. Gerard Nieuwenkamp - VSL Improved Reliability in Natural Gas Energy Measurement with the Revised ISO 10723 Standard Gerard Nieuwenkamp - VSL Improved reliability in natural gas energy measurement with the revised ISO 10723 standard

More information

Euramet comparison for Ethanol in Nitrogen EURAMET.QM-K4.1

Euramet comparison for Ethanol in Nitrogen EURAMET.QM-K4.1 Blad 1 van 28 Euramet comparison for Ethanol in Nitrogen EURAMET.QM-K4.1 Page 2 of 28 VSL Thijsseweg 11 2629 JA Delft Postbus 654 2600 AR Delft Nederland T +31 15 269 15 00 F +31 15 261 29 71 E info@vsl.nl

More information

Gas Metrology. SI-traceable: Primary Reference Materials Certified Reference Materials Calibrated Gas Mixtures Interlaboratory Comparisons

Gas Metrology. SI-traceable: Primary Reference Materials Certified Reference Materials Calibrated Gas Mixtures Interlaboratory Comparisons Gas Metrology SI-traceable: Primary Reference Materials Certified Reference Materials Calibrated Gas Mixtures Interlaboratory Comparisons T +31 15 2691500 E vsl@vsl.nl www.vsl.nl 1 This guide provides

More information

International Comparison CCQM-K119 Liquefied Petroleum Gas

International Comparison CCQM-K119 Liquefied Petroleum Gas International Comparison CCQM-K119 Liquefied Petroleum Gas M L Downey 1, P J Brewer 1, E Atkins 1, R J C Brown 1, A S Brown 1, E T Zalewska 2, A M H van der Veen 2, D E Smeulders 3, J B McCallum 3, R T

More information

International Comparison CCQM-K76: Sulfur Dioxide in Nitrogen

International Comparison CCQM-K76: Sulfur Dioxide in Nitrogen International Comparison CCQM-K76: Sulfur Dioxide in Nitrogen Franklin R. Guenther 1, Michael E. Kelley 1, Gerald D. Mitchell 1, Manuel de Jesús Avila Salas 2, Jorge Koelliker Delgado 2, Francisco Rangel

More information

CCQM-K90, Formaldehyde in nitrogen, 2 μmol mol -1 Final report

CCQM-K90, Formaldehyde in nitrogen, 2 μmol mol -1 Final report CCQM-K90, Formaldehyde in nitrogen, 2 μmol mol -1 Final report J. Viallon 1, E. Flores 1, F. Idrees 1, P. Moussay 1, R. I. Wielgosz 1, D. Kim 2, Y. D. Kim 2, S. Lee 2, S. Persijn 3, L.A. Konopelko 4, Y.A.

More information

ENVIRONMENTAL AND POLLUTANTS GAS ANALYZERS

ENVIRONMENTAL AND POLLUTANTS GAS ANALYZERS XIX IMEKO World Congress Fundamental and Applied Metrology September 6, 9, Lisbon, Portugal ENVIRONMENTAL AND POLLUTANTS GAS ANALYZERS Ana Madeira, Florbela A. Dias and Eduarda Filipe 3 Instituto Português

More information

Determination of Air Density with Buoyancy Artefacts

Determination of Air Density with Buoyancy Artefacts IMEKO 20 th TC3, 3 rd TC16 and 1 st TC22 International Conference Cultivating metrological knowledge 27 th to 30 th November, 2007. Merida, Mexico. Determination of Air Density with Buoyancy Artefacts

More information

International Key Comparison CCQM-K26b and Pilot Study CCQM P50b (SO 2 )

International Key Comparison CCQM-K26b and Pilot Study CCQM P50b (SO 2 ) International Key Comparison CCQM-K6b and Pilot Study CCQM P50b (SO ) Field - Gas standards Subject - Comparison of primary standards of sulphur dioxide (SO ) in synthetic air. Participants (in CCQM K6b)

More information

Subject - Comparison of primary standards of nitrogen monoxide (NO) in nitrogen

Subject - Comparison of primary standards of nitrogen monoxide (NO) in nitrogen International Key Comparison CCQM-K26a and Pilot Study CCQM-P50a (NO) Final Report Field - Gas standards Subject - Comparison of primary standards of nitrogen monoxide (NO) in nitrogen Participants (in

More information

Bi-lateral Comparison (APMP.L-K ) Calibration of Step Gauge

Bi-lateral Comparison (APMP.L-K ) Calibration of Step Gauge APMP ASIA PACIFIC METROLOGY PROGRAMME Asia-Pacific Metrology Programme Bi-lateral Comparison (APMP.L-K5.2006.1) Calibration of Step Gauge Final Report NMIJ, January 2017 Contents 1 Document control 2 Introduction

More information

A comparison of nitrogen dioxide (NO 2 ) in nitrogen standards at 10 μmol/mol by Fourier Transform Infrared Spectroscopy (FT-IR)

A comparison of nitrogen dioxide (NO 2 ) in nitrogen standards at 10 μmol/mol by Fourier Transform Infrared Spectroscopy (FT-IR) A comparison of nitrogen dioxide (NO 2 ) in nitrogen standards at 10 μmol/mol by Fourier Transform Infrared Spectroscopy (FT-IR) (Final report of the pilot study CCQM-P110-B1) Edgar Flores *1, Faraz Idrees

More information

International Comparison COOMET.QM-K111 Propane in nitrogen

International Comparison COOMET.QM-K111 Propane in nitrogen International Comparison COOMET.QM-K111 Propane in nitrogen Final Report L.A. Konopelko 1, Y.A. Kustikov 1, A.V. Kolobova 1,V.V. Pankratov 1, A.A. Pankov 1, O.V. Efremova 1, M.S. Rozhnov 2, D.M. Melnyk

More information

The analysis of complex multicomponent

The analysis of complex multicomponent The analysis of complex multicomponent gas mixtures for emission monitoring 31 July 2017 Test and Measurement Prelly Marebane Fit-for-purpose reference materials for air pollution monitoring in SA and

More information

International Comparison CCQM-K116: 10 µmol mol -1 water vapour in nitrogen

International Comparison CCQM-K116: 10 µmol mol -1 water vapour in nitrogen International Comparison CCQM-K116: 10 µmol mol -1 water vapour in nitrogen P J Brewer 1, B Gieseking 1, V F Ferracci 1, M Ward 1, J van Wijk 2, A M H van der Veen 2, A A Lima 3, C R Augusto 3, S H Oh

More information

Version /02/2013. (Pilot Study CCQM-P110 Nitrogen dioxide in Nitrogen 10 μmol/mol_protocol B2) Final Report

Version /02/2013. (Pilot Study CCQM-P110 Nitrogen dioxide in Nitrogen 10 μmol/mol_protocol B2) Final Report Study on the accuracy and uncertainty of FT-IR methods calibrated with synthetic spectra for NO 2 concentration measurements (Pilot Study CCQM-P110 Nitrogen dioxide in Nitrogen 10 μmol/mol_protocol B2)

More information

International Comparison CCQM-K41. Final Report. Page 1 of 55

International Comparison CCQM-K41. Final Report. Page 1 of 55 International Comparison CCQM-K41 Final Report Franklin R Guenther 1, Walter R Miller 1, David L. Duewer 1, Gwi Suk Heo 2, Yong-Doo Kim 2, Adriaan M.H. van der Veen 3, Leonid Konopelko 4, Yury Kustikov

More information

Supersedes: The copy of this document located on Measurement Canada s website is considered to be the controlled copy.

Supersedes: The copy of this document located on Measurement Canada s website is considered to be the controlled copy. Bulletin Category: GAS Bulletin: G-23 Page: 1 of 7 The copy of this document located on Measurement Canada s website is considered to be the controlled copy. Recognition of data from specialty gas suppliers

More information

Gas Mixture Two Components. Gas Mixtures Two Components

Gas Mixture Two Components. Gas Mixtures Two Components Gas Mixture Two Components Two Components Contents Two-Component Mixtures Ammonia...M4 Argon...M5 Benzene...M6 n-butane...m7 Carbon Dioxide...M8 Carbon Monoxide...M10 Chlorine...M12 Ethane...M13 Ethylene

More information

AIR PRESSURE CONTROLLED MASS MEASUREMENT SYSTEM

AIR PRESSURE CONTROLLED MASS MEASUREMENT SYSTEM Measurement of Mass, Force and Torque (APMF 2013) International Journal of Modern Physics: Conference Series Vol. 24 (2013) 1360002 (6 pages) The Authors DOI: 10.1142/S2010194513600021 AIR PRESSURE CONTROLLED

More information

All-Russia D.I.Mendeleyev Scientific and Research Institute for Metrology (VNIIM)

All-Russia D.I.Mendeleyev Scientific and Research Institute for Metrology (VNIIM) All-Russia D.I.Mendeleyev Scientific and Research Institute for Metrology (VNIIM) APMP 2015 TCFF Meeting 2 Nov. 2015 Beijing, China Konstantin Popov k.v.popov@vniim.ru Contents Participating in second

More information

A marked improvement in the reliability of the measurement of trace moisture in gases

A marked improvement in the reliability of the measurement of trace moisture in gases Research paper A marked improvement in the reliability of the measurement of trace moisture in gases - Establishment of metrological traceability and a performance evaluation of trace moisture analyzers

More information

Maximum precision for quantifiable success.

Maximum precision for quantifiable success. Accredited calibration gas mixtures Maximum precision for quantifiable success. Reference materials and other accredited calibration gas mixtures. 02 Accredited calibration gas mixtures It s now official:

More information

Research and Development. Money in the Pipeline

Research and Development. Money in the Pipeline On-line gas chromatography Gerard Nieuwenkamp Research and Development Money in the Pipeline Delft, 2 November 2010 Natural gas metering 2" 2" 65 m3/h usm 6" 400 m3/h M etering runs 2500 m3/h Gas Oil Piston

More information

Gas mixtures. Individual solutions specifically for your application

Gas mixtures. Individual solutions specifically for your application Gas mixtures Individual solutions specifically for your application For many applications besides pure gases also defined gas mixtures are required. For many routine applications, from operating gases

More information

Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA)

Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA) Int. J. Metrol. Qual. Eng. 5, 303 (2014) c EDP Sciences 2014 DOI: 10.1051/ijmqe/2014012 Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA) D. Jonker and E.P. Tarnow

More information

International Comparison CCQM K23b Natural gas type II

International Comparison CCQM K23b Natural gas type II International Comparison CCQM K23b Natural gas type II Adriaan M.H. van der Veen 1, Hima Chander 1,Paul R. Ziel 1, Ed W.B. de Leer 1, Damian Smeulders 2, Laurie Besley 2, Valnei Smarçao da Cunha 3, Zei

More information

Keeping It Simple. Gas Mixtures for Stack Emissions Monitoring.

Keeping It Simple. Gas Mixtures for Stack Emissions Monitoring. Keeping It Simple Gas Mixtures for Stack Emissions Monitoring www.uk.airliquide.com Stack Emissions Monitoring The Stack Emissions Monitoring market has undergone substantial legislative and quality changes

More information

Speciality Gases Practice 4. Our quality, your safety. Accredited Test and Calibration Laboratory Speciality Gases Centre in Hörstel.

Speciality Gases Practice 4. Our quality, your safety. Accredited Test and Calibration Laboratory Speciality Gases Centre in Hörstel. Speciality Gases Practice 4 Our quality, your safety. Accredited Test and Calibration Laboratory Speciality Gases Centre in Hörstel. Accredited Westfalen quality. Speciality Gases Centre in Hörstel. They

More information

EN ISO/IEC Technical Requirements

EN ISO/IEC Technical Requirements EN ISO/IEC 17025 - Technical Requirements Wolfgang Garche Environmental Protection Agency Department Air Quality Monitoring, Information, Assessment 1 What means accreditation in accordance to EN ISO/IEC

More information

Preparation of high precision standards (with ± 1 ppm) using a gravimetric method for measuring atmospheric oxygen

Preparation of high precision standards (with ± 1 ppm) using a gravimetric method for measuring atmospheric oxygen Preparation of high precision standards (with ± 1 ppm) using a gravimetric method for measuring atmospheric oxygen Nobuyuki Aoki 1, Takuya Shimosaka 1 Shigeyuki Ishidoya 2, Shohei Murayama 2 1.National

More information

NMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013

NMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013 NMISA NEW GAS FLOW CALIBRATION FACILITY Deona Jonker 7 October 2013 Introduction Flow the total volume of a fluid that flows past a fixed point in a given time Flow measurements used in industry as a means

More information

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES 1. Applicability and Principle 1.1 Applicability. This method applies to the analysis of carbon dioxide

More information

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER Michael T Tang, Ph.D. Grace Feng Greg Merideth Rui Huang Matheson Gas Applied Lab

More information

Development of an Analytical System Based on a Magnetopneumatic Oxygen Analyzer for Atmospheric Oxygen Determination

Development of an Analytical System Based on a Magnetopneumatic Oxygen Analyzer for Atmospheric Oxygen Determination ANALYTICAL SCIENCES APRIL 2018, VOL. 34 487 2018 The Japan Society for Analytical Chemistry Development of an Analytical System Based on a Magnetopneumatic Oxygen Analyzer for Atmospheric Oxygen Determination

More information

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation This intercomparison will be carried out under the lead of KRISS with the cooperation of ARPANSA. This comparison has been approved

More information

THE USE OF Z FACTORS IN GRAVIMETRIC VOLUME MEASUREMENTS

THE USE OF Z FACTORS IN GRAVIMETRIC VOLUME MEASUREMENTS THE USE OF Z FACTORS IN GRAVIMETRIC VOLUME MEASUREMENTS Author: T. Mautjana NMISA Private Bag X 34 Lynwood Ridge; 0040 Email: tmautjana@nmisa.org Phone: 012 841 4798; Fax: 012 841 2131 Abstract Using gravimetric

More information

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology: Title: Comparison of Chemical Sensor and NDIR Technologies TSN Number: 25 File:\\MII- SRV1\Metron\Bridge_Analyzers\Customer_Service_Documentation\White_Papers\25_G en EGA NDIR vs Chemical Sensor.docx Created

More information

Traceability of greenhouse gases and volatile organic compounds for air monitoring in South Africa

Traceability of greenhouse gases and volatile organic compounds for air monitoring in South Africa Traceability of greenhouse gases and volatile organic compounds for air monitoring in South Africa James Tshilongo *1, Silindile L. Lushozi 1,2, Napo Ntsasa 1, Mudalo Jozela 1,2 Luke Chimuka 2 and Goitsemang

More information

APPENDIX C. 40 CFR PART 60 App. B SPEC. 1-4, 6

APPENDIX C. 40 CFR PART 60 App. B SPEC. 1-4, 6 APPENDIX C 40 CFR PART 60 App. B SPEC. 1-4, 6 Part 60, Appendix B - Performance Specification 2 Specifications and Test Procedures for SO2 and NOx Continuous Emission Monitoring Systems in Stationary Sources

More information

Central scientific research automobile and automotive engine institute

Central scientific research automobile and automotive engine institute Central scientific research automobile and automotive engine institute The next mandatory requirements for gas analytical equipment act both in Russian Federation and countries of Customs Union (Belarus,

More information

NMISA New Gas Flow Calibration Facility

NMISA New Gas Flow Calibration Facility NMISA New Gas Flow Calibration Facility Speaker / Author: D. Jonker Co-author: E.P. Tarnow National Metrology Institute of South Africa (NMISA) Private Bag X34, Lynnwood Ridge, Pretoria, 0040, South Africa

More information

Generating Calibration Gas Standards

Generating Calibration Gas Standards Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and

More information

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE. Gases&Technology FEATURE Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. B Y M A T T H E W M O N A G L E Abstract Bulk gases are often delivered to

More information

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1 Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost HERMAN HOLM SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 4C CONFERENCE APRIL 2018 SAN ANTONIO, TX SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC

More information

Vehicle and Engine Emissions Testing

Vehicle and Engine Emissions Testing Vehicle and Engine Emissions Testing Calibration gases and equipment for compliance Benefits and Features Guaranteed analytical accuracy facilitates precise instrument calibration to ensure Part 1065,

More information

Beamex. Calibration White Paper. Weighing scale calibration - How to calibrate weighing instruments

Beamex. Calibration White Paper.  Weighing scale calibration - How to calibrate weighing instruments Beamex Calibration White Paper info@beamex.com Weighing scale calibration - How to calibrate weighing instruments Weighing scale calibration - How to calibrate weighing instruments Weighing scales, weighing

More information

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Oneida Research Services, Inc. 8282 Halsey Road Whitesboro, NY 13492 Phone: (315) 736-5480 FAX: (315) 736-9321 1.0 Purpose The

More information

Field verification of on-line HC dew point measurements

Field verification of on-line HC dew point measurements Field verification of on-line HC dew point measurements Andy Benton, Technical Consultant, Michell Instruments Limited, Ely, UK andy.benton@michell.com www.michell.com Introduction: The Michell Condumax

More information

(NIST Special Publication )

(NIST Special Publication ) (NIST Special Publication 260-126) THE NIST TRACEABLE REFERENCE MATERIAL PROGRAM FOR GAS STANDARDS by Franklin R. Guenther, William D. Dorko, Walter R. Miller and George C. Rhoderick Analytical Chemistry

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 6974-4 First edition 2000-04-01 Natural gas Determination of composition with defined uncertainty by gas chromatography Part 4: Determination of nitrogen, carbon dioxide and

More information

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010 Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600 November 2010 Table of Contents ``Abstract...3 ``Introduction...4 ``Methods & Results...6 ``Data Analysis...9 ``Conclusion...12

More information

ali-q TM Gravimetric Verification Procedure

ali-q TM Gravimetric Verification Procedure ali-q TM Gravimetric Verification Procedure Introduction This document describes the recommended procedure for verifying performance of the ali-q aliquoting pipet controller using the industry standard-

More information

The Gold Standard in Calibration Mixtures

The Gold Standard in Calibration Mixtures The Gold Standard in Calibration Mixtures Calibration standards are used in a wide variety of industries. The need for accurate and precise standards has become increasingly important because of the consequences

More information

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly

More information

APPLICATION NOTE. GC Integrated Permeation Device

APPLICATION NOTE. GC Integrated Permeation Device GC Integrated Permeation Device GC-integrated Design PPB to PPM Level Calibration No Cylinders or Regulators Required Option for Dual Ovens Cost Effective, Safe, Clean, Flexible, NIST Traceable Keywords:

More information

International comparison protocol for the key comparison of 1 kg stainless steel standards.

International comparison protocol for the key comparison of 1 kg stainless steel standards. CCM.M-K4 06/10/ International comparison protocol for the key comparison of 1 kg stainless steel standards. Organized by the Working Group on Mass Standards of the Consultative Committee for Mass and related

More information

CHM 111 Unit 5 Sample Questions

CHM 111 Unit 5 Sample Questions Name: Class: Date: As you work these problems, consider and explain: A. What type of question is it? B. How do you know what type of question it is? C. What information are you looking for? D. What information

More information

CALCULATING THE SPEED OF SOUND IN NATURAL GAS USING AGA REPORT NO Walnut Lake Rd th Street Houston TX Garner, IA 50438

CALCULATING THE SPEED OF SOUND IN NATURAL GAS USING AGA REPORT NO Walnut Lake Rd th Street Houston TX Garner, IA 50438 CALCULATING THE SPEED OF SOUND IN NATURAL GAS USING AGA REPORT NO. 10 Jerry Paul Smith Joel Clancy JPS Measurement Consultants, Inc Colorado Engineering Experiment Station, Inc (CEESI) 13002 Walnut Lake

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 FLUKE CALIBRATION, PHOENIX PRIMARY PRESSURE AND FLOW LABORATORY 4765 East Beautiful Lane Phoenix, AZ 85044 Mike DiGioia Phone: 602 773

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 10156 Third edition 2010-04-01 Gases and gas mixtures Determination of fire potential and oxidizing ability for the selection of cylinder valve outlets Gaz et mélanges de gaz

More information

Figure Vapor-liquid equilibrium for a binary mixture. The dashed lines show the equilibrium compositions.

Figure Vapor-liquid equilibrium for a binary mixture. The dashed lines show the equilibrium compositions. Another way to view this problem is to say that the final volume contains V m 3 of alcohol at 5.93 kpa and 20 C V m 3 of air at 94.07 kpa and 20 C V m 3 of air plus alcohol at 100 kpa and 20 C Thus, the

More information

Protocol Gas Verification Program (PGVP)

Protocol Gas Verification Program (PGVP) Protocol Gas Verification Program (PGVP) Michael Hayes ICAC/ MARAMA Workshop Advances in Control Technologies July 9-10, 2008 Exton, PA PGVP Protocol Gases from an historic perspective Historical Audit

More information

CHEMISTRY - CLUTCH CH.5 - GASES.

CHEMISTRY - CLUTCH CH.5 - GASES. !! www.clutchprep.com CONCEPT: UNITS OF PRESSURE Pressure is defined as the force exerted per unit of surface area. Pressure = Force Area The SI unit for Pressure is the, which has the units of. The SI

More information

Interlaboratory comparison in the pressure range from 0 to 2 MPa for accredited calibration laboratories

Interlaboratory comparison in the pressure range from 0 to 2 MPa for accredited calibration laboratories Int. J. Metrol. Qual. Eng. 6, 307 (2015) c EDP Sciences 2015 DOI: 10.1051/ijmqe/2015021 Interlaboratory comparison in the pressure range from 0 to 2 MPa for accredited calibration laboratories Lovorka

More information

INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION.

INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION. Andrzej Hantz RADWAG BALANCES AND SCALES INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION. The article presents the influence of comparison process automatization

More information

Part 6: Critical flow orifices

Part 6: Critical flow orifices Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 6145-6 Third edition 2017-07 Gas analysis Preparation of calibration gas mixtures using dynamic methods Part 6: Critical flow orifices Analyse

More information

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1250 METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1.0 Scope and Application. 1.1 Analytes. Analyte CAS No. Sensitivity Total Organic Compounds N/A

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 FLUKE CALIBRATION, PHOENIX PRIMARY PRESSURE AND FLOW LABORATORY 4765 East Beautiful Lane Phoenix, AZ 85044 Mike DiGioia Phone: 602 773

More information

Ch. 14 The Behavior of Gases

Ch. 14 The Behavior of Gases Ch. 14 The Behavior of Gases 14.1 PROPERTIES OF GASES Compressibility Compressibility: a measure of how much the volume of matter decreases under pressure Gases are easily compressed because of the spaces

More information

Chapter 5 TEST: Gases

Chapter 5 TEST: Gases Chapter 5 TEST: Gases 1) Gases generally have A) low density B) high density C) closely packed particles D) no increase in volume when temperature is increased E) no decrease in volume when pressure is

More information

3 HiQ Specialty Gases and Specialty Equipment. Precision matters in everything we do

3 HiQ Specialty Gases and Specialty Equipment. Precision matters in everything we do 3 HiQ Specialty Gases and Specialty Equipment Precision matters in everything we do 02 Introduction Introduction 03 Contents. 4-5 What we can do for you 6-7 Specialty gases products 8-9 How we measure

More information

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist ASTM WK 43267 Standard Test Method for Dissolved Gases Anne Jurek Applications Chemist Rationale Hydraulic Fracturing is becoming more and more common. Nearby water well can be affected during the fracking

More information

PERFORMANCE SPECIFICATION PROPELLANT PRESSURIZING AGENT, ARGON

PERFORMANCE SPECIFICATION PROPELLANT PRESSURIZING AGENT, ARGON METRIC 07 August 2013 SUPERSEDING MIL-PRF-27415B 08 February 2007 PERFORMANCE SPECIFICATION PROPELLANT PRESSURIZING AGENT, ARGON Comments, suggestions, or questions on this document should be addressed

More information

EPA Traceability Protocol for Assay and Certification of Gaseous Calibration Standards

EPA Traceability Protocol for Assay and Certification of Gaseous Calibration Standards EPA 600/R-12/531 May 2012 www.epa.gov/ord EPA Traceability Protocol for Assay and Certification of Gaseous Calibration Standards Office of Research and Development National Risk Management Research Laboratory

More information

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV 16-1 INTRODUCTION 16-1.1 Purpose. The purpose of this chapter is to familiarize divers with the techniques used to mix divers breathing gas. 16-1.2 Scope. This chapter

More information

Traceable calibration of automatic weighing instruments in dynamic operation

Traceable calibration of automatic weighing instruments in dynamic operation Traceable calibration of automatic weighing instruments in dynamic operation Matej Grum* Metrology Institute of the Republic of Slovenia, Grudnovo nabrežje 17, 1000 Ljubljana, Slovenia Abstract. The article

More information

Dr. Rogers Chapter 5 Homework Chem 111 Fall 2003

Dr. Rogers Chapter 5 Homework Chem 111 Fall 2003 Dr. Rogers Chapter 5 Homework Chem 111 Fall 2003 From textbook: 7-33 odd, 37-45 odd, 55, 59, 61 1. Which gaseous molecules (choose one species) effuse slowest? A. SO 2 (g) B. Ar(g) C. NO(g) D. Ne(g) E.

More information

Application Note AN-107

Application Note AN-107 SPEC Sensor TM Characterization & Calibration Considerations Scope This document is provided to describe the considerations needed to characterize, calibrate, verify and validate the measurement performance

More information

Interchangeability Criteria.

Interchangeability Criteria. Interchangeability Criteria. A substitute gas should burn satisfactorily with negligible change in burner performance on all types of burners without the need for special adjustment. The important requirements

More information

Phosgene Sensor. Sensoric COCl2 3E 1

Phosgene Sensor. Sensoric COCl2 3E 1 Phosgene Sensor 1 FEATURES Amperometric 3 electrode sensor cell Fixed organic electrolyte High reliability TYPICAL APPLICATIONS Ambient monitoring of TLV levels Chemical Industry, Homeland Security PART

More information

CERTIFICATE OF ACCREDITATION ISO/IEC 17025:2005 ANSI/NCSL Z (R2002)

CERTIFICATE OF ACCREDITATION ISO/IEC 17025:2005 ANSI/NCSL Z (R2002) CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Antibus Scales & Systems, Inc. 1919 Research

More information

VIC offers a variety of Calibrated Gas Leaks

VIC offers a variety of Calibrated Gas Leaks CALIBRATED GAS LEAKS VIC offers a high quality line of calibrated gas leaks traceable to National Institute of Standards Technology (NIST). Calibrated gas leaks are used to provide a controlled delivery

More information