Original Article. Simultaneous determination of androstenedione and testosterone in human serum by liquid chromatography-tandem mass spectrometry

Size: px
Start display at page:

Download "Original Article. Simultaneous determination of androstenedione and testosterone in human serum by liquid chromatography-tandem mass spectrometry"

Transcription

1 Original Article Simultaneous determination of androstenedione and testosterone in human serum by liquid chromatography-tandem mass spectrometry L M Gallagher, L J Owen and B G Keevil Abstract Address Department of Clinical Biochemistry, South Manchester University Hospitals NHS Trust, Manchester M23 9LT, UK Correspondence Mrs L J Owen laura.owen@smuht.nwest.nhs.uk Background We aimed to develop a sensitive assay to quantitate serum concentrations of both androstenedione and testosterone within the female range simultaneously, using liquid chromatography-tandem mass spectrometry (LC-MS/ MS) for use in the routine clinical laboratory and to compare this method with immunoassay. Method Samples (2 ml) were prepared by liquid liquid extraction using (1 ml) methyl-tert-butyl-ether. Deuterated androstenedione and testosterone were used as internal standards. Results The standard curve was linear to 5 nmol/l, the lower limit of quantitation was.25 nmol/l, and intra- and inter-assay coefficients of variation were o1% for both androgens over the range.3 35 nmol/l. There was a poor relationship between the LC-MS/MS and the radioimmunoassay methods for androstenedione with the LC-MS/MS generally giving lower results. For testosterone, the LC-MS/MS and immunoassay methods compared well at all concentrations. However, when female samples only were examined, the agreement deteriorated. Conclusions We have developed a sensitive and precise LC-MS/MS method, which gives more accurate results for all androstenedione measurements and low testosterone concentrations than immunoassay. Introduction Testosterone concentration is routinely requested in the investigation of oligo- or amenorrhoea, hirsutism and/or acne in women. The calculation of a free androgen index (FAI) using testosterone and sex-hormone-binding globulin (SHBG) concentrations has been shown to be a sensitive diagnostic marker of polycystic ovarian syndrome (PCOS). 1 The likelihood of a diagnosis of PCOS is increased if the androstenedione level is also raised. 2 Testosterone concentrations often do not correlate well with symptoms, but may help in prompting further investigation. A testosterone concentration of o5 nmol/ L measured by immunoassay is rarely associated with serious pathology, but higher concentrations warrant further investigation for the detection of an androgensecreting neoplasm 3 or congenital adrenal hyperplasia. 4 It has been documented that radioimmunoassays (RIAs) for androstenedione measurement may overestimate steroid concentrations in children and adults. 5,6 Commercial direct immunoassays for testosterone, which have almost completely replaced old extraction assays, can also lead to spuriously high measurements in women and children. 5,7--9 We report a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method which is sensitive, accurate and precise to measure androstenedione and testosterone simultaneously. This method should prove useful in the assessment of female patients presenting with hirsutism, virilization or acne. Although LC-MS/ MS assays for testosterone do exist, 1,11 to our knowledge, none that can measure androstenedione alone or simultaneously has been reported. Materials and methods Androstenedione and testosterone were supplied by Sigma Diagnostics (Poole, Dorset, UK). Deuterated 48 r 27 The Association for Clinical Biochemistry

2 Simultaneous determination of androstenedione and testosterone 49 androstenedione (d 7 A,98% isotopic purity) and deuterated testosterone (d 2 T, 99% isotopic purity) were supplied by CDN Isotopes (Quebec, Canada). Stock androstenedione and testosterone solutions were prepared for standardization by dissolving pure compound in methanol and stored at 21C. A separate stock solution was prepared for quality controls (QCs) for both analytes in agreement with published guidelines. 12 Mixed working standards were prepared by diluting the stock solutions of androstenedione and testosterone in phosphate-bu ered saline (PBS), ph 7.4, containing.1% bovine serum albumin (BSA), to give a range of concentrations (--5 nmol/l) for both androstenedione and testosterone. QCs were prepared in the same manner to give concentrations of nmol/l for both analytes. The use of PBS-based QCs allowed us to calculate the percentage deviation from target concentrations, which would not have been possible had we used commercial QCs. Aliquots of 5 ml were stored in microcentrifuge tubes at 21C for up to six months. Internal standards were prepared by diluting d 7 A and d 2 T to concentrations of.1 and.2 mg/l, respectively, in methanol (high-performance liquid chromatography [HPLC] grade, Fisher, Loughborough, UK) and stored at 21C. Sample preparation Serum samples, standards or QC samples (2 ml) were transferred in singleton into polypropylene microcentrifuge tubes (1.5 ml). Internal standard (1 ml) and 1mL methyl-tert-butyl-ether (HPLC grade, Fisher, Loughborough, UK) were added. Following this, the tubes were stoppered and vortex mixed on a mechanical mixer for 4 min. The supernatant from each tube was then transferred into a glass tube, and the tubes were placed in a 41C heating block. The solvent was evaporated under a gentle air ow for approximately 15 min. The residue was then reconstituted with 1 ml of 5:5 mobile phase (water [A] and methanol [B: HPLC grade, Fisher, Loughborough, UK]), each containing 2 mmol/l ammonium acetate (Sigma, Poole, UK) and.1% (v/v) formic acid (Analar, VWR International, Poole, UK). The tubes were vortex mixed for 1min and extract was then transferred into a 96-well microtitre plate. The plate was sealed and transferred directly to the autosampler for analysis. Liquid chromatography-tandem mass spectrometry Chromatography was performed using a Shimadzu HPLC system comprising a SIL-HT sampler and a degasser with two 1-AD pumps (Shimadzu, Milton Keynes, UK). Extract (5 ml) was directly injected from the microtitre plate onto a Security Guard C18 4mm 2 mm column, connected to a Synergi 4 mm Hydro-RP column (5 mm 3 mm), both from Phenomenex (Maccles eld, UK). The column was maintained at ambient temperature. Following elution from the column, the sample was pumped directly to the electrospray probe of the mass spectrometer, with no splitting or solvent diversion. The mobile phase initially consisted of 7% B for 2 min, allowing isocratic elution. The column was then washed with 95% B for.5 min, then re-equilibrated with starting conditions for 1min. The ow rate was maintained at.6 ml/min. The total run time was 4.3 min injection to injection. A Quattro Micro tandem mass spectrometer (Waters, Manchester, UK) tted with a Z spray ion source was used for all analyses. The instrument was operated in electrospray positive ionization mode and was directly coupled to the HPLC system. System control and data acquisition were performed with MassLynx NT 4. software with automated data processing, using the MassLynx QuanLynx Program provided with the mass spectrometer. Calibration curves were constructed using least-squares regression with 1/x weighting. To tune the mass spectrometer, solutions of androstenedione and d 7 A (1mg/L in methanol) and testosterone and d 2 T (1mg/L in methanol) were infused into the ion source. The cone voltage was optimized to maximize the intensity of the precursor ions for androstenedione, d 7 A, testosterone and d 2 T, seen at m/z 287.3, m/z 294.4, m/z and m/z 291.3, respectively. The collision energy was then adjusted to optimize the signal for the most abundant product ions, seen at m/z 96.8, m/z 99.8, m/z 19. and m/z 98.9, respectively. Tuning conditions for androstenedione and d 7 A were as follows: electrospray capillary voltage.9 kv, sample cone voltage 26 V, and collision energies 22 and 26 ev, respectively, at a collision gas pressure mbar argon. For testosterone and d 2 T, tuning conditions were as follows: electrospray capillary voltage.9 kv, sample cone voltage 28 V and collision energy 28 ev, at a collision gas pressure mbar argon. Desolvation gas ow and temperature were maintained at 6 L/h and 371C, respectively; source temperature was 141C. Transitions were monitored in multiple reaction monitoring (MRM) mode with a dwell time of.2 s. Validation The assay was validated against published acceptance criteria for linearity, precision, recovery and sample stability. 12 Linearity To evaluate linearity of the calibration curves, three curves were prepared and analysed in a single batch.

3 5 Gallagher et al. The ratios of analyte peak height to internal standard peak height were plotted against androgen concentration in nmol/l. Calibration curves were judged linear if the correlation coe cient (r 2 )wasbetterthan.99as calculated by weighted linear regression. Six serum samples were subject to serial dilution with PBS containing.1% BSA to assess assay linearity. Imprecision Imprecision of the method was assessed against a range of concentrations using QC samples. These samples were analysed daily for 15 days to calculate interassay imprecision. To determine intra-assay imprecision, the same samples were analysed 12 times within one batch. Percentage deviation was calculated from the di erence between mean observed and nominal concentrations to assess bias. Limit of quantitation The lower limit of quantitation (LLOQ) was de ned as the concentration for which 1 replicates of PBS-based samples, prepared with low concentrations of androstenedione and testosterone, gave a coe cient of variation (CV) of less than 2% and bias of less than 2%. Recovery The recovery was determined by comparing the concentration of each androgen in patient samples, before and after the addition of a known amount of the androgen. We took six serum samples with androstenedione and testosterone concentrations ranging from.8 to 4.5 nmol/l and from.6 to 18 nmol/l, respectively. The concentrations of androstenedione and testosterone added to each sample were, 7.5, 15 and 3 nmol/l, and,1, 2 and 4 nmol/l, respectively. Sample stability We assessed stability by carrying out: (1) analysis of extract before and after 24 h storage at 41C, (2) analysis of a single extract injected every 8 min over 15 h, (3) analysis of serum samples after one to ve freeze-- thaw cycles. Samples were judged to be stable if the change in response was less than 1%. Ion suppression Ion suppression is a matrix e ect, which occurs when compounds in a sample compete with the analyte for ionization in the source. To investigate this, we infused the following solutions directly into the mass spectrometer to give a constant background signal: 34 nmol/l deuterated testosterone in methanol and 34 nmol/l of deuterated androstenedione in methanol. An extracted serum sample was injected simultaneously via the autosampler. A reduction in the background signal is observed when ion suppression is occurring. The ion suppression is deemed signi cant if a reduction in signal of 41% is observed where the compound of interest elutes. 13 Specificity Solutions of various related steroids, both natural and synthetic, were prepared in 5% methanol/water to a nal concentration of 1 nmol/l and injected directly, without extraction. This excess concentration was chosen to allow easy identi cation of any potentially interfering peaks. Any compounds found to give a signal in the speci c channels for testosterone, androstenedione or their deuterated counterparts were subsequently prepared to the same concentration in PBS containing.1% BSA and extracted along with a full standard curve and QCs to allow quantitation. Steroids tested were levonorgestrel, 19 nortestosterone, estrone, dehydroepiandrosterone, oestradiol, epitestosterone, cyproterone acetate, desogestrel, ethinyloestradiol and norethisterone. Accuracy To assess the accuracy of our method, we compared our assay with an established LC-MS/MS method 1 (n ¼ 28) and three pools provided by the United Kingdom National QualityAssessment Scheme (UKNEQAS) with GC-MS assigned values (n ¼ 4) for testosterone. Comparative assays We compared the results obtained by LC-MS/MS with those of a non-extracted RIA method for androstenedione (DSL-38 Active, Diagnostic Systems Laboratories Inc., TX, USA) and those of the Roche E17 electrochemiluminescence immunoassay (ECLIA) for testosterone (Roche Diagnostics, Lewes, UK). Samples from patients were stored at 31C until analysis, rst by these methods and then by LC-MS/MS. The samples taken for analysis for androstenedione and testosterone were subject to local ethical approval. Androstenedione concentrations were measured in 46 female and 46 male anonymized sera by RIA and LC-MS/MS. Testosterone concentrations were measured in 77 female and 52 male anonymized sera by both ECLIA and LC-MS/MS.We then assessed the accuracy of the ECLIA method by analysing low concentra-

4 Simultaneous determination of androstenedione and testosterone 51 tion (.3 and 1. nmol/l) QC samples 1 times. These QCs were made up in PBS containing.1% BSA to reduce any matrix e ects and possible crossreactivities. Although the LC-MS/MS method is aimed at the routine measurement of female samples, male samples were included in the comparison to allow us to assess the linear range of the assay. Including male samples in the comparison would also allow us to use this method for any male samples that gave anomalous testosterone results by immunoassay. Statistical analysis All statistical analyses were performed using Analyseit software (Analyse-it Software Ltd, Leeds, UK). Results Validation Using the chromatographic conditions above, androstenedione and testosterone co-eluted with their deuterated counterparts. Androstenedione and d 7 A, and testosterone and d 2 T were found to have retention times of 1.8 and 2.3 min, respectively. There were no interfering peaks (Figure 1). Figure 2 shows the result of investigation for ion suppression. The peak for androstenedione is shown in Figure 2b, eluting at 1.8 min. Figure 2a represents the constant infusion of deuterated androstenedione, showing that the main ion suppression occurs at.6 and 3. min. The peak for testosterone (Figure 2d) is shown to elute at 2.3 min. The main ion suppression occurs at min and 3. min (Figure 2c). Both androgens are seen to elute away from the main areas of ion suppression. The standard curve was linear to 5 nmol/l for both androgens, with a correlation coe cient (r 2 )of.999 (n ¼ 3) (Figure 3). Serial dilution of six serum samples demonstrated dilutional linearity (Table 1). The lower limit of quanti cation was.25 nmol/l for both androgens. At this concentration, the %CVand bias of1 replicates were 1.9% and 11.5%, respectively, for androstenedione and 11.9% and 7.5%, respectively, for testosterone. Concentrations below.25 nmol/l result in %CVs greater than the acceptable limit of 2%. Both inter- and intra-assay imprecision were below 1% CV for androstenedione and testosterone at all concentrations, with less than 15% bias for all samples (Table 2). The mean recovery of androstenedione was 99% (range %) and for testosterone 93% (range %). In all experiments for stability, there was a decrease in response of less than 5.5% in measured concentrations for both androgens. There was no loss of sensitivity over repeat injection for 15 h; the extract wasstablefor24hat41c and samples were stable for ve freeze--thaw cycles. None of the steroids tested for speci city gave a signi cant signal in the speci c MRM channels for androstenedione, testosterone or the internal standards. Figure 1 A patient sample showing a typical chromatogram for each androgen and their deuterated form, at concentrations of 2.1 and 1.2 nmol /L for androstenedione and testosterone, respectively. (a) d 7 androstenedione, (b) d 2 testosterone, (c) testosterone and (d) androstenedione

5 52 Gallagher et al. a > e3 b > e4 Time c > e > e3 d Figure 2 Ion suppression. (a) Represents the constant infusion of deuterated androstenedione, while (b) is a serum sample showing a peak where androstenedione elutes. (c) Represents the constant infusion of deuterated testosterone, while (d) is a serum sample showing a peak where testosterone elutes Bland--Altman analysis of female samples analysed for testosterone by this method and an established LC- MS/MS method in routine use in the SAS Centre for Steroid Hormones (Leeds), showed good agreement (Figure 4). Our method demonstrated a small negative bias of.5 nmol/l. Analysis of UKNEQAS samples also demonstrated good agreement for all four replicates for each of the three pools (Table 3). Comparison with other methods Androstenedione When results obtained by LC-MS/MS were compared with those by RIA, Passing and Bablok regression analysis showed LC-MS/MS (nmol/l) ¼.58 RIA (nmol/l).21, r 2 ¼.74. Bland--Altman agreement analysis showed the RIA to have an overall bias of 1.97 nmol/l

6 A Androstenedione/d 7 - androstenedione peak height B Androstenedione concentration (nmol/l) Simultaneous determination of androstenedione and testosterone 53 Table 2 Intra-assay imprecision for androstenedione (a) and testosterone (b), and inter-assay imprecision for androstenedione (c) and testosterone (d) Bias (%) Imprecision CV (%) (a) Androstenedione nominal concentration (nmol/l) (b) Testosterone nominal concentration (nmol/l) Testosterone/d 2 - testosterone peak height Testosterone concentration (nmol/l) Figure 3 Typical standard curves for (a) androstenedione and (b) testosterone Table 1 Testosterone (a) and androstenedione (b) concentrations in serum samples diluted 1 in 2, 1 in 5 and 1 in 1 with PBS containing.1% BSA Sample number Undiluted 1 in 2 1 in 5 1 in 1 (a) Testosterone (nmol/l) (b) Androstenedione (nmol/l) (c) Androstenedione nominal concentration (nmol/l) (d) Testosterone nominal concentration (nmol/l) n=12 LC-MS/MS comparative assay (Figure 5a). However a concentration-dependent bias was observed. At androstenedione concentrations of approximately 1nmol/L, the bias was close to zero and it increased with increasing androstenedione concentrations up to a bias of 6 nmol/l at approximately 1 nmol/l androstenedione. Testosterone Mean of all methods Zero bias Figure 4 Bland Altman analysis comparing this method with an established LC-MS/MS assay for testosterone only When results obtained by LC-MS/MS were compared with those by ECLIA, Passing and Bablok regression analysis showed LC-MS/MS (nmol/l) ¼.98 ECLIA (nmol/l).5, r 2 ¼.97 and Bland--Altman analysis revealed a bias of.22 nmol/l (Figure 5b).

7 54 Gallagher et al. Table 3 Testosterone concentrations (nmol/l) measured by LC-MS/MS of UKNEQAS samples with GC-MS assigned values Pool GC-MS target LC-MS/MS individual values LC-MS/MS mean 31A ;.67;.61; B ;1.14;1.28; C ;2.81;2.82; LC-MS/MS as a percentage of GC-MS target Table 4 Table showing greater variability at low concentrations when measured by ECLIA (Roche) than by LC-MS/MS.3 nmol/l 1. nmol/l Testosterone concentration LC-MS/MS ECLIA LC-MS/MS ECLIA Mean SD CV (%) Bias (%) On examining results from female samples alone (n ¼ 77), we found the correlation (r 2 )was.25.the regression equation was LC-MS/MS (nmol/ L) ¼.74 ECLIA (nmol/l) þ.1. Bland--Altman analysis revealed a bias of.24 nmol/l, i.e., concentrations obtained by ECLIA were higher than those obtained by LC-MS/MS (Figure 5c). One signi cant outlier was observed in which the ECLIA testosterone concentration was almost 6 nmol/l higher than the LC- MS/MS. Repeat analysis of PBS-based samples using the LC-MS/MS and automated immunoassay for testosterone showed that the ECLIA underestimated testosterone considerably in the female range (Table 4). Discussion The chromatograms that are produced with this method are clear with no interfering peaks seen near the relevant retention times. Despite the use of liquid-- liquid extraction, which theoretically should remove many substances that may result in ion suppression, regions of ion suppression are observed. The chromatography has been optimized to ensure that both testosterone and androstenedione elute away from these areas, and therefore negligible ion suppression is seen. The assay s LLOQ and good linearity at high concentrations would make the method, particularly for testosterone, applicable to both female and male samples. The high working range will also cover both the normal and pathological concentrations; therefore the need for sample dilution is extremely rare. Stability studies showed that both androstenedione and testosterone are stable for 24 h when extracted and stored in the fridge. There was no systematic loss in sensitivity upon repeat injection over 15 h, demonstrating that long assays may be performed without a decline in performance. It was also observed that both testosterone and androstenedione were stable for at least ve freeze--thaw cycles. None of the steroids tested showed signi cant interference even at high concentrations. However, the authors acknowledge that only a limited number of steroids were tested. Metabolites of steroids or drugs may possibly cause interfering peaks, but as these are not commercially available, it would be impractical to test these and the vast number of other steroids that are available. By comparing our method with both LC-MS/MS and GC-MS (UKNEQAS samples), we are con dent that our assay gives accurate results for testosterone. Unfortunately, it was not possible to conduct such experiments for androstenedione as GC-MS target pools and routine LC-MS/MS assays are not, to our knowledge, available for this analyte. Although we appreciate that comparing an extracted LC-MS/MS assay to non-extracted immunoassays is not comparing like with like, our intention was to compare the results obtained by our new assay (intended for routine use) with the methods that were in routine use in our laboratory at the time. These comparisons showed that much lower results were obtained by the LC-MS/MS method for androstenedione. This is perhaps due to the speci city of the LC-MS/MS assay because the calibrators from the RIA kit agreed closely with those of the LC-MS/MS assay.

8 Simultaneous determination of androstenedione and testosterone 55 A 1 1 LC-MS/MS RIA Mean of all methods Zero Bias Testosterone results by both methods compared well with a high correlation coe cient and low bias (.22 nmol/l); however, female samples alone showed a much poorer correlation and a slightly larger bias. As we are con dent of the LC-MS/MS method s accuracy and precision at these low concentrations, we can conclude that the automated immunoassay s performance is questionable in this range. The nding of a high bias by the immunoassay from the nominal testosterone concentration in an arti cial matrix supports this. This method has been used routinely for female samples in our laboratory for 13 months, measuring in excess of 12 samples in this time. The method is also often used to measure testosterone in male serum, including samples from other hospitals in the region, when the results of immunoassay are under question. B LC-MS/MS ECLIA Zero bias Summary We have developed a simple, sensitive, robust, accurate and precise LC-MS/MS method for simultaneous determination of androstenedione and testosterone. Using a preparatory liquid--liquid extraction step, combined with adequate chromatography, this method separates androstenedione and testosterone from possible interfering substances, thereby removing inherent problems seen with direct immunoassay. This method will prove useful in clinical practice. C Mean of all methods Acknowledgements We acknowledge Dr H Field, Leeds SAS Centre for Steroid Hormones, for kindly analysing a series of samples for this comparison and DrJ Middle, UKNEQAS, for providing specimens for analysis. 2 References LC-MS/MS Roche Mean of all methods Zero bias 1 Fox R, Corrigan E, Thomas PA, Hull MG. The diagnosis of polycystic ovaries in women with oligo-amenorrhoea: predictive power of endocrine tests. Clin Endocrinol (Oxf) 1991; 34: Robinson S, Rodin DA, Deacon A, Wheeler MJ, Clayton RN. Which hormone tests for the diagnosis of polycystic ovary syndrome? Br J Obstet Gynaecol 1992; 99: O Driscoll JB, Mamtora H, Higginson J, Pollock A, Kane J, Anderson DC. A prospective study of the prevalence of clear-cut Figure 5 Bland Altman difference plots of: (a) Androstenedione concentrations for all male and female samples measured by LC-MS/MS and RIA. (b) Testosterone concentrations measured by LC-MS/MS and ECLIA for all male and female samples. (c) Female testosterone samples alone measured by LC-MS/MS and ECLIA. Dotted lines indicate two standard deviations (2 SD) from the mean

9 56 Gallagher et al. endocrine disorders and polycystic ovaries in 35 patients presenting with hirsutism or androgenic alopecia. Clin Endocrinol 1991; 41: Tietz NW. Fundamentals of Clinical Chemistry. 3rd edn. Philadelphia: WB Saunders, 1996; Wudy SA, Wachter UA, Homoki J, Teller WM. 17 Alphahydoxyprogesterone, 4-androstenedione, and testosterone profiled by routine stable isotope dilution/gas chromatography-mass spectrometry in plasma of children. Pediatr Res 1995; 38: Dorgan JF, Fears TR, McMahon RP, Aronson Friedman L, Patterson BH, Greenhut SF. Measurement of steroid sex hormones in serum: a comparison of radioimmunoassay and mass spectrometry. Steroids 22; 67: Wheeler MJ, Lowy C. Warning on serum testosterone measurement. Lancet 1987; 2: Boots LR, Potter S, Potter D, Azziz R. Measurement of total serum testosterone levels using commercially available kits: high degree of between-kit variability. Fertil Steril 1998; 69: Taieb J, Mathian B, Millot F, et al. Testosterone measured by 1 immunoassays and by isotope dilution gas chromatography mass spectrometry in sera from 116 men, women, and children. Clin Chem 23; 49: Cawood ML, Field HP, Ford CG, et al. Testosterone measurement by isotope-dilution liquid chromatography-tandem mass spectrometry: validation of a method for routine clinical practice. Clin Chem 25; 51: Wang C, Catlin DH, Demers LM, Starcevic B, Swerloff S. Measurement of total serum testosterone in adult men: comparison of current laboratory methods versus liquid chromatography-tandem mass spectrometry. J Clin Endocrinol Metab 24; 89: Health and Human Services, US. Department of Food and Drug Administration, Center for Drug Evaluation and Research (CDER), Center for Veterinary Medicine (CVM). Guidance for Industry Bioanalytical Method Validation. Rockville. MD: FDA/Center for Drug Evaluation and Research, May 21; Annesley TM. Ion suppression in mass spectrometry. Clin Chem 23; 49: Accepted for publication 7 September 26

LCMSMS Solutions For Steroid Analysis

LCMSMS Solutions For Steroid Analysis LCMSMS Solutions For Steroid Analysis Dan Blake Manager, Clinical Applications For Research Use Only. Not for use in diagnostic procedures. RUO-MKT-18-1166-A Agenda Special considerations for steroid analysis

More information

High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis

High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis PO-CON695E High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis MSACL 6 EU Neil Loftus, Stephane Moreau, Mikaël Levi 3, Anja Grüning

More information

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS Abstract Highly sensitive analysis of metanephrines in plasma is critical in the diagnosis and treatment of pheochromocytoma and paraganglioma. Here, a HILIC LC-MS/MS method was developed using a Raptor

More information

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS UCT Part Numbers: CUQAX22Z Clean-Up C8+QAX, 2mg/1mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQID21-3UM

More information

Development of a Clinical Research Method for the Measurement of Testosterone. Dihydrotestosterone

Development of a Clinical Research Method for the Measurement of Testosterone. Dihydrotestosterone Development of a Clinical Research Method for the Measurement of Testosterone and Dihydrotestosterone Martijn van Faassen, Maarten van den Noort and Ido Kima University Medical Center, Groningen, Netherlands

More information

SOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15

SOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15 Chemicals needed: Precipitate solution, Methanol Derivatization Solution, Amplifex Keto Reagent Kit Mobile phases, H 2 0 with 0.1% Formic Acid LC/MS grade and Acetonitrile with 0.1% Formic Acid LC/MS grade

More information

Comparison of 7 Published LC-MS/MS Methods for the Simultaneous Measurement of Testosterone, androstenedione, and dehydroepiandrosterone.

Comparison of 7 Published LC-MS/MS Methods for the Simultaneous Measurement of Testosterone, androstenedione, and dehydroepiandrosterone. Papers in Press. Published October 26, 2015 as doi:10.1373/clinchem.2015.242859 The latest version is at http://hwmaint.clinchem.org/cgi/doi/10.1373/clinchem.2015.242859 Clinical Chemistry 61:12 000 000

More information

Now it is. easy to switch. CHS TM Steroid Profiling kit and software. Brochure not for distribution in the USA and Canada

Now it is. easy to switch. CHS TM Steroid Profiling kit and software. Brochure not for distribution in the USA and Canada Now it is easy to switch to Clinical MSMS CHS TM Steroid Profiling kit and software Brochure not for distribution in the USA and Canada 2 Meeting your needs for steroid assays PerkinElmer s new CHS MSMS

More information

Comparison of 7 Published LC-MS/MS Methods for the Simultaneous Measurement of Testosterone, androstenedione, and dehydroepiandrosterone.

Comparison of 7 Published LC-MS/MS Methods for the Simultaneous Measurement of Testosterone, androstenedione, and dehydroepiandrosterone. Clinical Chemistry 61:12 1475 1483 (2015) Endocrinology and Metabolism Comparison of 7 Published LC-MS/MS Methods for the Simultaneous Measurement of Testosterone, Androstenedione, and Dehydroepiandrosterone

More information

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries. Informal Commentary Limit of Diethylene Glycol (DEG) and Ethylene Glycol (EG) in Sorbitol Solution, Sorbitol Sorbitan Solution and Noncrystallizing Sorbitol Solution December 2009 Monograph/Section(s):

More information

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay Summary This assay measures Human Granulocyte Colony Stimulating Factor (G-CSF) in a 96-well

More information

WADA Technical Document TD2014EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting

WADA Technical Document TD2014EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting Endogenous Anabolic Androgenic Steroids Measurement and Reporting 1.0 Introduction The purpose of this Technical is to harmonize the approaches to the measurement and reporting of endogenous anabolic androgenic

More information

May 1, By John Heim,Doug Staples

May 1, By John Heim,Doug Staples 1 of 7 7/11/2014 3:39 PM May 1, 2010 By John Heim,Doug Staples Anabolic steroid screening analysis in urine is complex and labor intensive requiring sensitive instrumentation and optimized chromatographic

More information

The ICL Insider. Lab Testing: Testosterone. In This Issue. The Debate

The ICL Insider. Lab Testing: Testosterone. In This Issue. The Debate The ICL Insider February 2017 Volume 2, Issue 2 Lab Testing: Testosterone In This Issue Testosterone shashtilak@iclabs.ca Next Issue: Expected Release April 2017 The Debate Aging is accompanied by various

More information

The instrument should have a mass range from 2 to 2000 m/z or better Interface

The instrument should have a mass range from 2 to 2000 m/z or better Interface Revised Specifications for Liquid Chromatography Mass Spectrometry (LC-MS/MS ) Triple Quadrupole Specification for Ultra-Fast Liquid Chromatography system Solvent Delivery 1) A quaternary Pump system for

More information

Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites

Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites Application Note Doping Control Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites Steroid profiling, screening, and confirmatory analysis using the

More information

Retention Time Locking: Concepts and Applications. Application

Retention Time Locking: Concepts and Applications. Application Retention Time Locking: Concepts and Applications Application Gas Chromatography December 1997 Authors Vince Giarrocco Bruce Quimby Matthew Klee Agilent Technologies, Inc. 2850 Centerville Road Wilmington,

More information

CORESTA RECOMMENDED METHOD Nº 67

CORESTA RECOMMENDED METHOD Nº 67 CORESTA RECOMMENDED METHOD Nº 67 DETERMINATION OF WATER IN THE MAINSTREAM SMOKE OF CIGARS BY GAS CHROMATOGRAPHIC ANALYSIS (November 2005) 1. FIELD OF APPLICATION The method is applicable to the particulate

More information

The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood

The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood UCT Part Numbers: ECQUUS1015CT- Enviro-Clean 15 ml centrifuge tube with 400 mg MgSO 4 and 100 mg NaCl CUMPS2CT - Enviro-Clean

More information

Androgen Measurements

Androgen Measurements Chapter 5 / Androgen Measurements 63 Androgen Measurements Methods, Interpretation, and Limitations 5 Frank Z. Stanczyk SUMMARY Androgens are measured primarily in serum by direct and indirect (with one

More information

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS Application Note: Analysis of and Cyanuric Acid in Food Matrices by LC-MS/MS PeterVarelis, National Center for Food Safety and Technology, Illinois Institute of Technology. Jonathan Beck, Kefei Wang, and

More information

Inaccurate First-Generation Testosterone Assays Are Influenced by Sex Hormone Binding Globulin Concentrations

Inaccurate First-Generation Testosterone Assays Are Influenced by Sex Hormone Binding Globulin Concentrations Inaccurate First-Generation Testosterone Assays Are Influenced by Sex Hormone Binding Globulin Concentrations Annemieke C. Heijboer, 1 * Edo Savelkoul, 1 Adrian Kruit, 2 Erik Endert, 3 and Marinus A. Blankenstein

More information

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS Peter Alden and Michael Jones Waters Corporation, Milford, MA, U.S. APPLICATION BENEFITS UPLC combines

More information

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion: Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: The RSK-175 standard operating procedure was developed in order to determine the amount of dissolved gas in water. Due to the expansion

More information

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (16) 17 R QUALIFICATION OF EQUIPMENT ANNEX 2: QUALIFICATION OF GC EQUIPMENT Full document title and reference Document type Legislative

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310 www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Varian 310 310 LC/MS Manual System INCLUDES: 310-MS LC/MS triple quadrople

More information

Extended Application Note

Extended Application Note Extended Application Note Harmful Substances in Dietary Supplements HPLC Columns with APP A-337 www.mtc-usa.com 1-732-578-1777 INTRODUCTION A dietary supplement is a substance to be consumed for the purpose

More information

Quantitative Assay Validation for Oxandrolone in Human Plasma Using LC MS-MS

Quantitative Assay Validation for Oxandrolone in Human Plasma Using LC MS-MS Journal of Analytical Toxicology 2015;1 6 doi:10.1093/jat/bkv056 Journal of Analytical Toxicology Advance Access published May 27, 2015 Article Quantitative Assay Validation for Oxandrolone in Human Plasma

More information

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY Nengbing Xu, Hongmei Ying and Libo Zhu Ningbo Environmental Monitoring

More information

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution. Name: 1. You are asked to measure the percent iron in a steel sample you believe to be about 50% Fe by mass. You dissolve four replicate samples in nitric acid and dilute them by mass to a concentration

More information

GCMSD-Headspace Analysis SOP

GCMSD-Headspace Analysis SOP Before you start GCMSD-Headspace Analysis SOP Method and Sequences names are restricted to 50 characters (MSD program will crash otherwise) The GC oven and Injection Ports need to be cooled to 50 oc to

More information

Rat myeloperoxidase-antineutrophil cytoplasmic antibody(mpo-anca) ELISA Kit

Rat myeloperoxidase-antineutrophil cytoplasmic antibody(mpo-anca) ELISA Kit Rat myeloperoxidase-antineutrophil cytoplasmic antibody(mpo-anca) ELISA Kit Catalog Number. CSB-E08675r For the quantitative determination of rat myeloperoxidase-antineutrophil cytoplasmic antibody(mpo-anca)

More information

Figure 2. RESULTS DATA ANALYSIS

Figure 2. RESULTS DATA ANALYSIS ANDROGEN PARAMETERS IN HIRSUTE AND NORMAL FEMALE PATIENTS: IS THERE A ROLE FOR THE FREE ANDROGEN INDEX (FAI)? Castracane VD 1, Childress E 1, Tawwater B 1, Vankrieken L 2, El Shami AS 2 ( 1 Department

More information

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE. Gases&Technology FEATURE Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. B Y M A T T H E W M O N A G L E Abstract Bulk gases are often delivered to

More information

Rapid and Reliable Detection of Dissolved Gases in Water

Rapid and Reliable Detection of Dissolved Gases in Water Rapid and Reliable Detection of Dissolved Gases in Water Andrea Caruso and Massimo Santoro Thermo Fisher Scientific, Milan, Italy Application Note 005 Key Words Chromeleon CDS, Environmental, Fracking,

More information

FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS

FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS Analysis Conducted BY CC RESEARCH (P) Ltd., For Nanoceutical Solutions Inc. USA 2950 Thousand Oaks drive, Suite 3, San Antonio, Texas EXECUTIVE SUMMARY

More information

Agilent 7800 ICP-MS. Specifications and Typical Performance. Fast track metals analysis with Solution-Ready ICP-MS

Agilent 7800 ICP-MS. Specifications and Typical Performance. Fast track metals analysis with Solution-Ready ICP-MS Agilent 7800 ICP-MS Specifications and Typical Performance Fast track metals analysis with Solution-Ready ICP-MS The Solution-Ready Agilent 7800 Quadrupole ICP-MS combines proven, robust hardware, auto-optimization

More information

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic. 458 LCGC NORTH AMERICA VOLUME 24 NUMBER 5 MAY 2006 www.chromatographyonline.com LC TROUBLESHOOTING Dwell Volume Revisited Gradient transfer does not have to be problematic. John W. Dolan LC Troubleshooting

More information

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection This user guide does not have the intention to introduce you to any of the theory and/or principles behind

More information

CHEMILUMINESCENCE ENZYME IMMUNOASSAY (CLIA) TESTOSTERONE. Testosterone. Cat #

CHEMILUMINESCENCE ENZYME IMMUNOASSAY (CLIA) TESTOSTERONE. Testosterone. Cat # DIAGNOSTIC AUTOMATION, INC. 21250 Califa Street, Suite 102 and 116, Woodland Hills, CA 91367 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com

More information

Introduction. Experimental

Introduction. Experimental PO-CON162E Development of Automated Screening and Quantitation Approach on Novel On-Line SFE-SFC-MS/MS Platform (I) For 23 Restricted Perflurocompounds in Textiles ASMS 2016 MP 283 Jie Xing, Jun Xiang

More information

Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS)

Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS) Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS) 1.0 Purpose - This procedure specifies the required elements for the calibration and use of the Agilent 6890 GC interfaced

More information

MSD 96-Well MULTI-ARRAY CRP Assay

MSD 96-Well MULTI-ARRAY CRP Assay MSD 96-Well MULTI-ARRAY CRP Assay The following assay protocol has been optimized for analysis of C-reactive protein (CRP) in human serum and plasma samples. MSD Materials Storage Read Buffer T (4X), with

More information

Determination of Selected ASTM Perfluorinated Compounds (PFCs) Using the Shimadzu LCMS No. LCMS-083. No. SSI-LCMS-083

Determination of Selected ASTM Perfluorinated Compounds (PFCs) Using the Shimadzu LCMS No. LCMS-083. No. SSI-LCMS-083 Liquid Chromatography Mass Spectrometry No. LCMS-083 Determination of Selected ASTM Perfluorinated Compounds (PFCs) Using the Shimadzu LCMS-8060 Summary ASTM methods D7968-14 and D7979-15 are used to determine

More information

IJPAR Vol.6 Issue 1 Jan - Mar Journal Home page:

IJPAR Vol.6 Issue 1 Jan - Mar Journal Home page: IJPAR Vol.6 Issue 1 Jan - Mar -2017 Journal Home page: ISSN:2320-2831 Research article Open Access Optimization of stability indicating RP-HPLC method for the estimation of an anti-cancer drug Sorafenib

More information

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek Abstract: Due to the expansion of natural gas drilling through horizontal fracturing, there has been

More information

TECHNICAL BULLETIN. TriControls. Control Materials for Liquid QC and Calibration Verification

TECHNICAL BULLETIN. TriControls. Control Materials for Liquid QC and Calibration Verification i-stat TECHNICAL BULLETIN Tris Materials for Liquid QC and OVERVIEW As part of the READi initiative (Responds, Enhances, And Delivers innovations), Abbott Point of Care (APOC) has released a new set of

More information

Methylprednisolone detection in urine following local and oral administrations

Methylprednisolone detection in urine following local and oral administrations Susana MS Simões, Marta Calçada, Luís Horta, Xavier de la Torre Methylprednisolone detection in urine following local and oral administrations Laboratório de Análises e Dopagem. Instituto do Desporto de

More information

Automated Determination of Dissolved Gases in Water Anne Jurek

Automated Determination of Dissolved Gases in Water Anne Jurek Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: Tapping the natural gas reservoirs throughout the United States has long been a viable solution for energy independence; however

More information

Integration of amino acid, acylcarnitine and steroids analysis in single FIA/LC-MS/MS platform

Integration of amino acid, acylcarnitine and steroids analysis in single FIA/LC-MS/MS platform POCON3E Integration of amino acid, acylcarnitine and steroids analysis in single FIA/LCMS/MS platform ASMS 2 ThP2 Tetsuo Tanigawa, Toshikazu Minohata Shimadzu Corporation, Kyoto, Japan analysis in single

More information

A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS

A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS B.Lakshmi,Kallam Haranadhareddy Institute of Technology, Guntur Prof T.V.Reddy,Prof Mallareddy College of Engineering, Secunderabad

More information

Indian Institute of Technology Kanpur Environmental Engineering and Management Department of Civil Engineering

Indian Institute of Technology Kanpur Environmental Engineering and Management Department of Civil Engineering Indian Institute of Technology Kanpur Environmental Engineering and Management Department of Civil Engineering Professor Vinod Tare Phones:+91-0512-2597791,7792 Email: vinod@iitk.ac.in Post Office: IIT

More information

Reproductive DHT Analyte Information

Reproductive DHT Analyte Information Reproductive DHT Analyte Information - 1 - DHT Introduction Dihydrotestosterone (DHT) together with other important steroid hormones such as testosterone, androstenedione (ASD) and dehydroepiandrosterone

More information

Thermo Scientific Chromatography Well Plates

Thermo Scientific Chromatography Well Plates Thermo Scientific Chromatography Well Plates High-throughput sample handling solutions Well Plates for Chromatography A full selection of high quality sample well plates for your UHPLC, LC and GC applications.

More information

Thermo Scientific Chromatography Well Plates

Thermo Scientific Chromatography Well Plates Thermo Scientific Chromatography Well Plates High-throughput sample handling solutions Well Plates for Chromatography A full selection of high quality sample well plates for your UHPLC, LC and GC applications.

More information

Improved Performance in Capillary Electrophoresis using Internal Standards

Improved Performance in Capillary Electrophoresis using Internal Standards Improved Performance in Capillary Electrophoresis using Internal Standards Kevin D. Altria, GlaxoSmithKline R&D, Ware, Hertfordshire, UK. The use of internal standards significantly improves the quantitative

More information

CHEMILUMINESCENCE IMMUNOASSAY (CLIA) Testosterone

CHEMILUMINESCENCE IMMUNOASSAY (CLIA) Testosterone DIAGNOSTIC AUTOMATION, INC. 21250 Califa Street, Suite 102 and 116, Woodland Hills, CA 91367 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com

More information

ab Androstenedione ELISA Kit

ab Androstenedione ELISA Kit ab108672 Androstenedione ELISA Kit Instructions for Use A competitive immunoenzymatic assay for the quantitative measurement of Androstenedione in serum and plasma (citrate). This product is for research

More information

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist ASTM WK 43267 Standard Test Method for Dissolved Gases Anne Jurek Applications Chemist Rationale Hydraulic Fracturing is becoming more and more common. Nearby water well can be affected during the fracking

More information

Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples

Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples Outline of Presentation Introduction Process Gas Analysis Magnetic Sector Mass Spectrometer Standard

More information

Chemical Analysis for Methyltestosterone

Chemical Analysis for Methyltestosterone APPLIED CONSUMER SERVICES, INC. Chemical Analysis for Methyltestosterone Report #: 27584 Report to: Mr. Mitch Ross 24 Dockside Lane # 131 Key Largo, Florida 33017 2016 1 1 8 9 0 N W 8 7 t h C o u r t,

More information

WADA Technical Document TD2016EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting

WADA Technical Document TD2016EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting Endogenous Anabolic Androgenic Steroids Measurement and Reporting Introduction The purpose of this Technical Document (TD) is to harmonize the approaches to the measurement and reporting of Endogenous

More information

Rapid Analysis of Metanephrine and Normetanephrine in Urine by Gas Chromatography Mass Spectrometry

Rapid Analysis of Metanephrine and Normetanephrine in Urine by Gas Chromatography Mass Spectrometry Clinical Chemistry 48:2 332 337 (2002) Endocrinology and Metabolism Rapid Analysis of Metanephrine and Normetanephrine in Urine by Gas Chromatography Mass Spectrometry David K. Crockett, 1* Elizabeth L.

More information

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction Application Note Rapid performance verification of AZURA systems with refractive index detector Method Keywords ID HPLC Quality control, system verification, refractive index, AZURA Analytical HPLC Plus

More information

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010 Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600 November 2010 Table of Contents ``Abstract...3 ``Introduction...4 ``Methods & Results...6 ``Data Analysis...9 ``Conclusion...12

More information

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS) Toxicology Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose - This procedure specifies the required elements for the calibration and use of the Agilent Gas Chromatograph interfaced to the Agilent

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs. Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography

More information

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note Gas Chromatography Author Matthew S. Klee Agilent Technologies 2850 Centerville Road Wilmington, DE 19808-1610 USA Abstract

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results. Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions

More information

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve USER'S GUIDE Nanoflow Probe 3-4mm Red Stripe ptfe Sleeve To Injector ZU1XC Fused Silica 20µm 90µm TTF115 ptfe Sleeve Fused Silica 25µm 375µm Micromass UK Limited Floats Road Wythenshawe M23 9LZ Tel: +44

More information

RayBio Human vwf ELISA Kit

RayBio Human vwf ELISA Kit RayBio Human vwf ELISA Kit Catalog #: ELH-vWF User Manual Last revised April 15, 2016 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite 100 Norcross, GA

More information

Simultaneous Measurement of Serum Testosterone and Dihydrotestosterone. by liquid chromatography tandem mass spectrometry

Simultaneous Measurement of Serum Testosterone and Dihydrotestosterone. by liquid chromatography tandem mass spectrometry Clinical Chemistry 54:11 1855 1863 (2008) Endocrinology and Metabolism Simultaneous Measurement of Serum Testosterone and Dihydrotestosterone by Liquid Chromatography Tandem Mass Spectrometry Steve Shiraishi,

More information

SIMPLE, EFFICIENT AND SIMULTANEOUS DETERMINATION OF ANABOLIC STEROIDS IN HUMAN URINE

SIMPLE, EFFICIENT AND SIMULTANEOUS DETERMINATION OF ANABOLIC STEROIDS IN HUMAN URINE SIMPLE, EFFICIENT AND SIMULTANEOUS DETERMINATION OF ANABOLIC STEROIDS IN HUMAN URINE Pages with reference to book, From 216 To 218 S.J. Khurshid, M. Riaz, S. Ahmed ( Nuclear Chemistry Division, Pakistan

More information

ab IgG1 Human ELISA Kit

ab IgG1 Human ELISA Kit ab100548 IgG1 Human ELISA Kit Instructions for Use For the quantitative measurement of Human IgG1 in serum and plasma This product is for research use only and is not intended for diagnostic use. Version

More information

International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage:

International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage: Research Article CODEN: IJRPJK ISSN: 2319 9563 International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage: www.ijrpns.com SIMULTANEOUS ESTIMATION AND VALIDATION OF TAMSULOSIN

More information

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS Application Note: 424 Analysis of and Cyanuric Acid in Food Matrices by LC-MS/MS PeterVarelis, National Center for Food Safety and Technology, Illinois Institute of Technology. Jonathan Beck, Kefei Wang,

More information

biosensis Mouse Vascular endothelial growth factor A/VEGF-A/VEGF-164/VEGF-1/VEGF- 120/VEGF-2 ELISA Kit Protocol

biosensis Mouse Vascular endothelial growth factor A/VEGF-A/VEGF-164/VEGF-1/VEGF- 120/VEGF-2 ELISA Kit Protocol biosensis Mouse Vascular endothelial growth factor A/VEGF-A/VEGF-164/VEGF-1/VEGF- 120/VEGF-2 ELISA Kit Protocol Catalog No: BEK-2110-1P For quantitative detection of mouse VEGF-A (VEGF164&VEGF120) in mouse

More information

National Institute for Public Health and the Environment Annual CRL workshop 22 October Update on natural Hormone studies

National Institute for Public Health and the Environment Annual CRL workshop 22 October Update on natural Hormone studies 2008 Annual CRL workshop 22 October 2008 Update on natural Hormone studies Natural hormone studies: update Possible approaches - C12/C13 ratio: can result in proof of abuse - Determination of intact esters

More information

GUIDE TO ULTRA-FAST HPLC. Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations

GUIDE TO ULTRA-FAST HPLC. Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations GUIDE TO ULTRA-FAST HPLC Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations GUIDE TO ULTRA-FAST HPLC Quick Tips for Converting Conventional Reversed-Phase

More information

ab Testosterone ELISA Kit

ab Testosterone ELISA Kit Version 10 Last updated 29 January 2018 ab108666 Testosterone ELISA Kit A competitive immunoenzymatic assay for the quantitative measurement of Testosterone in serum, plasma and urine. This product is

More information

MSD 96-Well MULTI-ARRAY Human (6E10) Abeta 42 Ultra-Sensitive Kit

MSD 96-Well MULTI-ARRAY Human (6E10) Abeta 42 Ultra-Sensitive Kit MSD 96-Well MULTI-ARRAY Human (6E10) Abeta 42 Ultra-Sensitive Kit The first protocol has been optimized for quantifying Aβ 1-42 peptide in human cerebrospinal fluid (CSF). An additional general protocol

More information

biosensis Human IGF-II, Insulin-like growth factor II, Somatomedin-A ELISA Kit Protocol

biosensis Human IGF-II, Insulin-like growth factor II, Somatomedin-A ELISA Kit Protocol biosensis Human IGF-II, Insulin-like growth factor II, Somatomedin-A ELISA Kit Protocol Catalog No: BEK-2029-1P For quantitative detection of human IGF-II in cell culture supernatants, cell lysates, tissue

More information

Agilent Dimension Software for ELSD User Manual

Agilent Dimension Software for ELSD User Manual Agilent Dimension Software for ELSD User Manual Agilent Dimension Software for ELSD User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2011 No part of this manual may be reproduced in

More information

APPLICATION NOTE. GC Integrated Permeation Device

APPLICATION NOTE. GC Integrated Permeation Device GC Integrated Permeation Device GC-integrated Design PPB to PPM Level Calibration No Cylinders or Regulators Required Option for Dual Ovens Cost Effective, Safe, Clean, Flexible, NIST Traceable Keywords:

More information

Automated Low Background Solid Phase Extraction of Perfluorinated Compounds in Water. Ruud Addink Fluid Management Systems Watertown MA

Automated Low Background Solid Phase Extraction of Perfluorinated Compounds in Water. Ruud Addink Fluid Management Systems Watertown MA Automated Low Background Solid Phase Extraction of Perfluorinated Compounds in Water Ruud Addink Fluid Management Systems Watertown MA Introduction Perfluoralkylated compounds contain a perfluorinated

More information

Human ABCD1 ELISA KIT

Human ABCD1 ELISA KIT Human ABCD1 ELISA KIT Cat. No.:DEIA8716 Pkg.Size:96T Intended use The Human ABCD1 ELISA KIT is for the quantitative detection of human ABCD1 in serum and plasma. General Description ABCD1, also known as

More information

HUMAN IL6 KITS PROTOCOL

HUMAN IL6 KITS PROTOCOL HUMAN IL6 KITS PROTOCOL Part # 62HIL06PEG & 62HIL06PEH Test size: 500 tests (62HIL06PEG), 10,000 tests (62HIL06PEH) - assay volume: 20 µl Revision: 04 (Jan. 2018) Store at: -60 C or below This product

More information

Validation of the MRM pesticides from the Working Document SANCO/12745/2013 of November 2017 rev. 9(1)

Validation of the MRM pesticides from the Working Document SANCO/12745/2013 of November 2017 rev. 9(1) Validation of the MRM pesticides from the Working Document SANCO/12745/2013 of 21-22 November 2017 rev. 9(1) CONTENTS 1. Aim and scope... 2 2. Short description... 2 3. Apparatus and consumables... 2 4.

More information

Analysis of Pressure Rise During Internal Arc Faults in Switchgear

Analysis of Pressure Rise During Internal Arc Faults in Switchgear Analysis of Pressure Rise During Internal Arc Faults in Switchgear ASANUMA, Gaku ONCHI, Toshiyuki TOYAMA, Kentaro ABSTRACT Switchgear include devices that play an important role in operations such as electric

More information

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC Evaluation of linear and step gradient performance and retention time precision of the Agilent 12 Series Rapid Resolution LC This document is believed to be accurate and up-to-date. However, Agilent Technologies,

More information

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler Optimizing Vial Pressurization Parameters for the Analysis of Residual Solvents Using the 7697A Headspace Sampler Application Note Pharmaceuticals Author Roger L Firor Agilent Technologies, Inc.

More information

biosensis Rat Fibronectin ELISA Kit Protocol

biosensis Rat Fibronectin ELISA Kit Protocol biosensis Rat Fibronectin ELISA Kit Protocol Catalog No: BEK-2017-2P For quantitative detection of rat Fibronectin in cell culture supernatants, serum, and citrate, heparin, or EDTA plasma samples only

More information

biosensis Human TNFα/Cachectin/TNFSF2 ELISA Kit Protocol

biosensis Human TNFα/Cachectin/TNFSF2 ELISA Kit Protocol biosensis Human TNFα/Cachectin/TNFSF2 ELISA Kit Protocol Catalog No: BEK-2100-1P For quantitative detection of human TNFα in cell culture supernatants, serum, and heparin, EDTA or citrate treated plasma

More information

biosensis Human Lipocalin-2/NGAL ELISA Kit Protocol

biosensis Human Lipocalin-2/NGAL ELISA Kit Protocol biosensis Human Lipocalin-2/NGAL ELISA Kit Protocol Catalog No: BEK-2141-2P For quantitative detection of human Lipocalin-2 in cell culture supernatants, serum, and heparin treated plasma, saliva, and

More information

Increasing the productivity of ICP-OES with the Thermo Scientific Qtegra ISDS Software

Increasing the productivity of ICP-OES with the Thermo Scientific Qtegra ISDS Software TECHNICAL NOTE 43225 Increasing the productivity of ICP-OES with the Thermo Scientific Qtegra ISDS Software Authors Dr. Elena Chernetsova, Application Specialist, Thermo Fisher Scientific, Bremen, Germany

More information

Experiment GC : Analysis of BTEX by GC-FID

Experiment GC : Analysis of BTEX by GC-FID 46 Experiment GC : Analysis of BTEX by GC-FID Learning Goals: Familiarity with gas chromatography Gain experience in temperature programming and method development Correctly use an internal standard, and

More information

biosensis Mouse Brain-derived neurotrophic factor (BDNF) ELISA Kit Protocol

biosensis Mouse Brain-derived neurotrophic factor (BDNF) ELISA Kit Protocol biosensis Mouse Brain-derived neurotrophic factor (BDNF) ELISA Kit Protocol Catalog No: BEK-2003-2P For quantitative detection of mouse BDNF in cell culture supernatants, cell lysates, serum, and citrate,

More information

TRACE LEVEL DETERMINATION OF PER- AND POLYFLUOROALKYL SUBSTANCES (PFASs) IN WATER USING THE AGILENT 6460 LC/MS/MS

TRACE LEVEL DETERMINATION OF PER- AND POLYFLUOROALKYL SUBSTANCES (PFASs) IN WATER USING THE AGILENT 6460 LC/MS/MS ENVIRONMENTAL MARKET TRACE LEVEL DETERMINATION OF PER- AND POLYFLUOROALKYL SUBSTANCES (PFASs) IN WATER USING THE AGILENT 6460 LC/MS/MS A method for the determination of per- and polyfluoroalkyl substances

More information

Video Notes. LC Troubleshooting Series Retention Time Shifts

Video Notes. LC Troubleshooting Series Retention Time Shifts Video Notes LC Troubleshooting Series Retention Time Shifts Introduction John Henderson is an Applications Scientist for Agilent Chemistries and Supplies Division. Maureen Joseph is the HPLC Columns Product

More information

CGMP KITS PROTOCOL ASSAY PRINCIPLE

CGMP KITS PROTOCOL ASSAY PRINCIPLE CGMP KITS PROTOCOL Part # 62GM2PEG & 62GM2PEH Test size#: 500 tests (62GM2PEG), 10,000 tests (62GM2PEH) - assay volume: 20 µl Revision: 02-Jan.2018 Store at: 2-8 C (62GM2PEG); 2-8 C (62GM2PEH) For research

More information