Integration of amino acid, acylcarnitine and steroids analysis in single FIA/LC-MS/MS platform
|
|
- Meredith Payne
- 5 years ago
- Views:
Transcription
1 POCON3E Integration of amino acid, acylcarnitine and steroids analysis in single FIA/LCMS/MS platform ASMS 2 ThP2 Tetsuo Tanigawa, Toshikazu Minohata Shimadzu Corporation, Kyoto, Japan
2 analysis in single FIA/LCMS/MS platform Introduction Analysis of amino acids (AA) and acylcarnitines (AC) in dried blood spot (DBS) sample collection method by flow injection analysis (FIA) is now widely used. On the other hand, traditionally, analysis of steroid such as 7hydroxyprogesterone is done by immunoassays but LC/MS/MS will be an attractive analytical alternative because it can also screen for other related steroids. The use of LC/MS/MS results in a reduction of false positives and a more accurate quantitative performance. The requirements against steroid analysis by LC/MS/MS are getting more stringent issues. In this study, we present a strategy for performing both AA/AC and steroids analysis within a single LC/MS/MS platform. Methods and Materials The experimental setup designed to combine a FIA measurement covers 8 AAs and 7 ACs and a LCMS measurement for steroids includes cortisol, 2deoxycortisol (2DOF), deoxycortisol (DOF), androstenedione (4AD) and 7hydroxyprogesterone (7OHP) with 2 position 6 port high pressure valve. AAs, ACs Guard LC3AD SIL3AC F2AH LCMS (LCMS) F2AH Steroids Guard Guard (SIL3AC) SIL3AC LC3AD F2AH LCMS Figure Flow Diagram of FIA/LCMS/MS system 2
3 analysis in single FIA/LCMS/MS platform The isotopically labeled internal standards for amino acids, acylcarnitines, and steroids were purchased from Cambridge Isotope Laboratories, Inc. Quality control materials were obtained from the Newborn Screening Quality Assurance Program at the Centers for Disease Control and Prevention (CDC). Analytical Conditions HPLC Mobile Phase A :. Formic acid water Mobile Phase B : Methanol temperature : 4 ºC [for Amino Acids and Acylcarnitines] Guard : GL Sciences Cartridge Guard E ( mml x. mm I.D.) Gradient Program : Time B conc. (%) Flow rate (ml min) Injection Volume : µl [for Steroids] : Phenomenex Kinetex 2.6u XBC8 ( mml x 2. mm I.D., 2.6µm) Flow Rate :.3 ml/min Gradient Program : Time B conc. (%) Injection Volume : µl 3
4 analysis in single FIA/LCMS/MS platform Mass (LCMS8 triple quadrupole mass spectrometry) Ionization : heated ESI Nebulizing Gas Flow : 3 L / min Heating gas flow : L/min BH Temperature : ºC MRM parameter : Drying Gas Pressure : L / min DL Temperature : 2 ºC Interface Temperature : 4 ºC Q>Q3 IS Q>Q3 Q>Q3 IS Q>Q3 Phe 66.>2. Phe IS 72.>26. C 246.2>8. C IS 2.2>8. Leu 32.>86. Leu IS 3.>89. CDC 276.>8. CDC IS 279.>8. Met.>4. Met IS 3.>7. COH 262.2>8. COH IS 26.2>8. Tyr Val 82.>36. 8.>72. Tyr IS Val IS 88.> >8. C6 C8 26.2> >8. C8 IS 29.2>8. Cit Arg 76.>3. 7.>7. Cit IS Arg IS 78.>. 8.>7. C C2 36.2> >8. C2 IS 347.3>8. Ala 9.>44. Ala IS 94.>48. C >8. C4 IS 38.3>8. C C2 62.>3. 24.>8. C IS C2 IS 7.>3. 27.>8. C6 C6OH 4.3> >8. C6 IS 43.3>8. C3 C4 C4OH 28.> > >8. C3 IS C4 IS 22.> >8. C8 C8OH 428.4> >8. C8 IS 43.4>8. R.T(min) Q>Q3 CE (V) R.T(min) Q>Q3 CE (V) Cortisol. 363.> > CortisolD2. 36.> > DOF > > DOFD > > DOF > > DOFD > > AD > > AD3C >. 29.> OHP >9. 33.> OHPD > >
5 analysis in single FIA/LCMS/MS platform DBS samples (d = mm) were placed in 96well plates, and AAs, ACs and steroids were extracted with 8 μl of acetonitrilewater solution consists of the known concentrations of stable isotope labeled standards of each compounds. The extraction were performed in an ultrasonic bath for 3 min. Samples were measured using a Nexera UHPLC system coupled to LCMS8 triple quadrupole mass spectrometer (Shimadzu Corporation, Japan). Result DBS provides a number of advantages, for examples, a less invasive and much simpler sample collection method rather than venipuncture technique. Furthermore, it provides you simpler storage and transportation as well as it can lower the infection risk of various pathogens, and requires a smaller blood volume. To date, DBSLCMS/MS has emerged as an important method for quantitative analysis of small molecules. Previously we developed an innovative AAs and ACs screening method makes it possible to inject just µl of sample and successfully reduce analytical run time as fast as 74 seconds (conventional method >2 sec.) using the combination of Nexera MP and LCMS84 (Shimadzu Corporation, Japan). In addition to that, we independently developed a method for steroids in DBS. Steroids were separated on a Phenomenex kinetex XBC8 (x2mm, 2.6μm) at a column temperature of 4 C for min. In this study, we present a strategy for performing AAs, ACs and steroids analysis within a single LC/MS/MS platform. AAs, ACs and steroids were extracted from only one dried blood spot. This system enables to automatically analyze 7 min in all target analytes in 2 injections. B conc. (%). 9 9 Injection.7 for steroids. Injection.3 for AAs, ACs min min 4 min min Analysis of AAs and ACs wash valve switching Flow rate (ml/min) equilibration Analysis of steroids wash valve switching Figure 2 Time Program of amino acids, acylcarnitines and steroids analysis
6 6 Integration of amino acid, acylcarnitine and steroids analysis in single FIA/LCMS/MS platform Table Data Summary of 8 amino acids, 7 acylcarnitines and steroids No.32 No Phe Leu Met Tyr Val Cit 4.33 Arg Ala C C C3.. C4 No.36.. C4. 7%. C4OH.. C.2 CDC.72.6 COH C6.2 3 C8.2 3 C. 2 C C6. C6OH.7.6 C8 C8OH No.322 No %.9 No.323 No No % % No % No % % No.324 No.364 target % % % 4.4
7 analysis in single FIA/LCMS/MS platform Conc. Area %RSD S/N LOD LOQ Conc. Area %RSD S/N LOD LOQ, , , , Cortisol ,97 322, AD 7 2,247,236 2,87, , ,46, ,37, ,43, , , , , DOF ,22 7, OHP 7 2 4,27 874, , , ,, ,694, , , DOF ,44 964, ,29, ,872, Conclusions This platform is an effective tool for an initial screening and able to minimize sample consumption down to ul. However, the use of a high performance LC/MS/MS system is highly recommended in order to achieve the appropriate level of sensitivity for the steroids especially. Using high performance LC/MS/MS (LCMS8, Shimadzu Corporation, Japan), the observed limit of detection (LOD) for the analysis of 7hydroxyprogesterone was.9 ng/ml. First Edition: May, 2 For Research Use Only. Not for use in diagnostic procedures. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. Shimadzu Corporation, 2
High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis
PO-CON695E High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis MSACL 6 EU Neil Loftus, Stephane Moreau, Mikaël Levi 3, Anja Grüning
More informationDetermination of Selected ASTM Perfluorinated Compounds (PFCs) Using the Shimadzu LCMS No. LCMS-083. No. SSI-LCMS-083
Liquid Chromatography Mass Spectrometry No. LCMS-083 Determination of Selected ASTM Perfluorinated Compounds (PFCs) Using the Shimadzu LCMS-8060 Summary ASTM methods D7968-14 and D7979-15 are used to determine
More informationIntroduction. Experimental
PO-CON162E Development of Automated Screening and Quantitation Approach on Novel On-Line SFE-SFC-MS/MS Platform (I) For 23 Restricted Perflurocompounds in Textiles ASMS 2016 MP 283 Jie Xing, Jun Xiang
More informationDevelopment of a Clinical Research Method for the Measurement of Testosterone. Dihydrotestosterone
Development of a Clinical Research Method for the Measurement of Testosterone and Dihydrotestosterone Martijn van Faassen, Maarten van den Noort and Ido Kima University Medical Center, Groningen, Netherlands
More informationTrace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS
Abstract Highly sensitive analysis of metanephrines in plasma is critical in the diagnosis and treatment of pheochromocytoma and paraganglioma. Here, a HILIC LC-MS/MS method was developed using a Raptor
More informationSimultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS
Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS UCT Part Numbers: CUQAX22Z Clean-Up C8+QAX, 2mg/1mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQID21-3UM
More informationThe instrument should have a mass range from 2 to 2000 m/z or better Interface
Revised Specifications for Liquid Chromatography Mass Spectrometry (LC-MS/MS ) Triple Quadrupole Specification for Ultra-Fast Liquid Chromatography system Solvent Delivery 1) A quaternary Pump system for
More informationThe Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood
The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood UCT Part Numbers: ECQUUS1015CT- Enviro-Clean 15 ml centrifuge tube with 400 mg MgSO 4 and 100 mg NaCl CUMPS2CT - Enviro-Clean
More informationLCMSMS Solutions For Steroid Analysis
LCMSMS Solutions For Steroid Analysis Dan Blake Manager, Clinical Applications For Research Use Only. Not for use in diagnostic procedures. RUO-MKT-18-1166-A Agenda Special considerations for steroid analysis
More informationNow it is. easy to switch. CHS TM Steroid Profiling kit and software. Brochure not for distribution in the USA and Canada
Now it is easy to switch to Clinical MSMS CHS TM Steroid Profiling kit and software Brochure not for distribution in the USA and Canada 2 Meeting your needs for steroid assays PerkinElmer s new CHS MSMS
More informationSetting a New Standard in Flash Chromatography Performance
Setting a New Standard in Flash Chromatography Performance What s New Faster Flow rates, up to 300 ml/min Higher pressures, up to 300 psi (20 bar) Bigger touchscreen, 12 or 15 inches Wider Dynamic Range
More informationProudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Varian 310 310 LC/MS Manual System INCLUDES: 310-MS LC/MS triple quadrople
More informationN Korea Japan China Thailand Taiwan Philipines EP-5 EP km. Source: Google Maps and
OCCURRENCES AND TRANSPORT OF PERFLUORINATED COMPOUNDS IN EAST CHINA SEA Kunacheva C 1, Fujii S 1, Tanaka S 1, Shimizu J 2 and Shivakoti B R 1 1 Graduate School of Global Environmental Studies, Kyoto University,
More informationAutomated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:
Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: The RSK-175 standard operating procedure was developed in order to determine the amount of dissolved gas in water. Due to the expansion
More informationSOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15
Chemicals needed: Precipitate solution, Methanol Derivatization Solution, Amplifex Keto Reagent Kit Mobile phases, H 2 0 with 0.1% Formic Acid LC/MS grade and Acetonitrile with 0.1% Formic Acid LC/MS grade
More informationTRACE LEVEL DETERMINATION OF PER- AND POLYFLUOROALKYL SUBSTANCES (PFASs) IN WATER USING THE AGILENT 6460 LC/MS/MS
ENVIRONMENTAL MARKET TRACE LEVEL DETERMINATION OF PER- AND POLYFLUOROALKYL SUBSTANCES (PFASs) IN WATER USING THE AGILENT 6460 LC/MS/MS A method for the determination of per- and polyfluoroalkyl substances
More informationMethylprednisolone detection in urine following local and oral administrations
Susana MS Simões, Marta Calçada, Luís Horta, Xavier de la Torre Methylprednisolone detection in urine following local and oral administrations Laboratório de Análises e Dopagem. Instituto do Desporto de
More informationExtend the time window for the detection of low level anabolic androgenic steroids and their metabolites
Application Note Doping Control Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites Steroid profiling, screening, and confirmatory analysis using the
More informationAnalysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS
Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS Peter Alden and Michael Jones Waters Corporation, Milford, MA, U.S. APPLICATION BENEFITS UPLC combines
More informationValidation of the MRM pesticides from the Working Document SANCO/12745/2013 of November 2017 rev. 9(1)
Validation of the MRM pesticides from the Working Document SANCO/12745/2013 of 21-22 November 2017 rev. 9(1) CONTENTS 1. Aim and scope... 2 2. Short description... 2 3. Apparatus and consumables... 2 4.
More informationFast method for Ginseng Analyses using Agilent Poroshell 120 Columns Scaled from a Traditional Method
Fast method for Ginseng Analyses using Agilent Poroshell Columns Scaled from a Traditional Method Application Note Pharmaceutical Author Rongjie Fu Agilent Technologies (Shanghai) Co. Ltd. 4 Ying Lun Road
More informationUser guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection
User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection This user guide does not have the intention to introduce you to any of the theory and/or principles behind
More informationExtended Application Note
Extended Application Note Harmful Substances in Dietary Supplements HPLC Columns with APP A-337 www.mtc-usa.com 1-732-578-1777 INTRODUCTION A dietary supplement is a substance to be consumed for the purpose
More informationAutomated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek
Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek Abstract: Due to the expansion of natural gas drilling through horizontal fracturing, there has been
More informationM/s. Thermo fisher Scientific
CLARIFICATIONS AND RESPONSES IN THE PRE BID MEETING HELD ON 19 TH AUGUST 2013 IN OUR COMMITTEE ROOM FOR THE PROCUREMENT OF DIOXIN ANALYSER (GC-MS/MS) File No : PUR/IMP/178/14 Description of Item : DIOXIN
More informationBETTER SAMPLE CARE. Spark Holland introduces. SPH1299 UHPLC Gradient Pump
BETTER SAMPLE CARE Spark Holland introduces SPH1299 UHPLC Gradient Pump SPH1299 Spark Holland introduces SPH1299 UHPLC Gradient Pump Robust, Precise and Easy to use Innovative pumping technology created
More informationEvaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC
Evaluation of linear and step gradient performance and retention time precision of the Agilent 12 Series Rapid Resolution LC This document is believed to be accurate and up-to-date. However, Agilent Technologies,
More informationSupporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial
Supporting Information for Micro-Collection of Gases in a Capillary Tube: Preservation of Spatial and Temporal Resolution AUTHOR INFORMATION Kristin A. Herrmann Favela, 1 * Pieter Tans, 2 Thomas Jaeckle,
More informationThe instrument will process the samples one by one till all the samples have been cleaned up.
Table of Contents SPE-01 Cleanup Station... 1 LC-04SP Valve System... 5 LC-04SP Standalone Valve... 10 LC-05 Auto Injector... 12 LC-06 Dosing Pump... 16 LC-08 Column Selector... 17 SPE-01 Cleanup Station
More informationAutomated Determination of Dissolved Gases in Water Anne Jurek
Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: Tapping the natural gas reservoirs throughout the United States has long been a viable solution for energy independence; however
More informationGases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.
Gases&Technology FEATURE Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. B Y M A T T H E W M O N A G L E Abstract Bulk gases are often delivered to
More informationMay 1, By John Heim,Doug Staples
1 of 7 7/11/2014 3:39 PM May 1, 2010 By John Heim,Doug Staples Anabolic steroid screening analysis in urine is complex and labor intensive requiring sensitive instrumentation and optimized chromatographic
More informationHumiSys HF High Flow RH Generator
HumiSys HF High Flow RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The HumiSys HF is a high flow version of the previously
More informationUWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column
Before you connect your trap and column You need to provide your own trap and column (Note our fittings accommodate trap and column made with fused silica 360 µm OD). Check our guide for packing traps
More informationASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist
ASTM WK 43267 Standard Test Method for Dissolved Gases Anne Jurek Applications Chemist Rationale Hydraulic Fracturing is becoming more and more common. Nearby water well can be affected during the fracking
More informationMore Speed and Resolution for Current Methods
More Speed and Resolution for Current Methods A Simple How to Guide for Easier Method Transfer in the Current Lab Environment Edward Kim Application Engineer June 15, 2010 Page 1 Key Points for Successful
More informationApplication Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction
Application Note Rapid performance verification of AZURA systems with refractive index detector Method Keywords ID HPLC Quality control, system verification, refractive index, AZURA Analytical HPLC Plus
More informationGCMSD-Headspace Analysis SOP
Before you start GCMSD-Headspace Analysis SOP Method and Sequences names are restricted to 50 characters (MSD program will crash otherwise) The GC oven and Injection Ports need to be cooled to 50 oc to
More informationRenzo BAGNATI Mario Negri Institute of Pharmacological Research- IRCCS Milano - Italy
STEROIDS IN WASTE WATERS AS MARKERS OF ABUSE Renzo BAGNATI Mario Negri Institute of Pharmacological Research- IRCCS Milano - Italy A first attempttoapplythe sewage epidemiologyapproachin doping Compare
More informationVideo Notes. LC Troubleshooting Series Retention Time Shifts
Video Notes LC Troubleshooting Series Retention Time Shifts Introduction John Henderson is an Applications Scientist for Agilent Chemistries and Supplies Division. Maureen Joseph is the HPLC Columns Product
More informationAgilent 1290 Infinity LC System Applications requiring the Agilent Ultra-Low Dispersion Kit
Agilent 19 Infinity LC System Applications requiring the Agilent Ultra-Low Dispersion Kit Technical Overview Author Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany mau 15 1 1 3 1 All components
More information(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.
Name: 1. You are asked to measure the percent iron in a steel sample you believe to be about 50% Fe by mass. You dissolve four replicate samples in nitric acid and dilute them by mass to a concentration
More informationTROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.
458 LCGC NORTH AMERICA VOLUME 24 NUMBER 5 MAY 2006 www.chromatographyonline.com LC TROUBLESHOOTING Dwell Volume Revisited Gradient transfer does not have to be problematic. John W. Dolan LC Troubleshooting
More informationWADA Technical Document TD2014EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting
Endogenous Anabolic Androgenic Steroids Measurement and Reporting 1.0 Introduction The purpose of this Technical is to harmonize the approaches to the measurement and reporting of endogenous anabolic androgenic
More informationANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION
ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION Standard OQ Test Suite This document describes the test program for qualifying LC analytical-scale Agilent systems; the following table lists
More informationOptimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler
Optimizing Vial Pressurization Parameters for the Analysis of Residual Solvents Using the 7697A Headspace Sampler Application Note Pharmaceuticals Author Roger L Firor Agilent Technologies, Inc.
More informationGUIDE TO ULTRA-FAST HPLC. Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations
GUIDE TO ULTRA-FAST HPLC Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations GUIDE TO ULTRA-FAST HPLC Quick Tips for Converting Conventional Reversed-Phase
More informationPicoChip. Nanospray System. Column Care & Use Manual
PicoChip Nanospray System Column DOCUMENT: 34414 RELEASE DATE: December 22, 2014 Read This First Read This First This document is provided to customers who have purchased New Objective equipment to use
More informationLongterm Detection of Anabolic Steroid Metabolites in Urine
Longterm Detection of Anabolic Steroid Metabolites in Urine Application Note Forensics/Doping Control Author Stephan Baumann Agilent Technologies, Inc. 5301 Stevens Creek Blvd Santa Clara, CA 95051 USA
More informationAnalysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS
Application Note: Analysis of and Cyanuric Acid in Food Matrices by LC-MS/MS PeterVarelis, National Center for Food Safety and Technology, Illinois Institute of Technology. Jonathan Beck, Kefei Wang, and
More informationOMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT
OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (16) 17 R QUALIFICATION OF EQUIPMENT ANNEX 2: QUALIFICATION OF GC EQUIPMENT Full document title and reference Document type Legislative
More informationApplications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples
Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples Outline of Presentation Introduction Process Gas Analysis Magnetic Sector Mass Spectrometer Standard
More informationextraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.
Informal Commentary Limit of Diethylene Glycol (DEG) and Ethylene Glycol (EG) in Sorbitol Solution, Sorbitol Sorbitan Solution and Noncrystallizing Sorbitol Solution December 2009 Monograph/Section(s):
More informationPrep LC Training Manual
Prep LC Training Manual (11-6-2013) Location: 1238 Hach Hall Contacts: kamel Harrata and Steve Veysey, 1234 Hach Hall Safety: Introduction: The Agilent PrepStar SD-1 preparative HPLC is a student operated
More informationPrep LC Training Manual S.V.
1234 Hach Hall 515-294-5805 www.cif.iastate.edu Prep LC Training Manual 11-6-2013 S.V. Location: Contact: 1238 Hach Hall Kamel Harrata, 1236 Hach Hall and Steve Veysey, 1234 Hach Hall Safety: All researchers
More informationDEBUBBLERS & DEGASSERS ACTIVE DEBUBBLERS APPLICATION NOTE WHY DEGAS YOUR MOBILE PHASE? PAGE 177 STAND ALONE DEGASSERS PAGE 175 PAGE 178
174 DEBUBBLERS & DEGASSERS ACTIVE DEBUBBLERS PAGE 175 APPLICATION NOTE WHY DEGAS YOUR MOBILE PHASE? PAGE 177 STAND ALONE DEGASSERS PAGE 178 SEE STARRED PRODUCTS FOR TOP SELLERS! Series DEBUBBLERS & DEGASSERS
More informationModel GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272)
Manual Addendum Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272) P/N M07273 DATE 06/11/13 TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments 16830 Chestnut Street City of Industry,
More informationComparative analysis of mass spectral matching for confident compound identification using the Advanced Electron Ionization source for GC-MS
TECHNICAL NOTE 1598 Comparative analysis of mass spectral matching for confident compound identification using the Advanced Electron Ionization source for GC-MS Authors Amit Gujar, 1 Tim Anderson, 1 Daniela
More informationMV5: Automated concentration system
MV5: Automated concentration system Abstract LabTech MV5 is the latest automatic nitrogen evaporator developed by our R&D Team able to process up to 54 samples simultaneously. This advanced instrument
More informationAutomated Low Background Solid Phase Extraction of Perfluorinated Compounds in Water. Ruud Addink Fluid Management Systems Watertown MA
Automated Low Background Solid Phase Extraction of Perfluorinated Compounds in Water Ruud Addink Fluid Management Systems Watertown MA Introduction Perfluoralkylated compounds contain a perfluorinated
More informationMSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay
MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay Summary This assay measures Human Granulocyte Colony Stimulating Factor (G-CSF) in a 96-well
More informationNew Generation System M, leading the World in the Non-Invasive Measurement of Critical Real-Time Parameters.
New Generation System M, leading the World in the Non-Invasive Measurement of Critical Real-Time Parameters. System M Spectrum Medicals total commitment to continuous product improvement is demonstrated
More informationGas Sampling in the DST
UCRL-ID-129732 Gas Sampling in the DST L. DeLoach M. Chiarappa R. Martinelli B. Glassley January 12, 1998 Lawrence Livermore National Laboratory This is an informal report intended primarily for internal
More informationUSER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve
USER'S GUIDE Nanoflow Probe 3-4mm Red Stripe ptfe Sleeve To Injector ZU1XC Fused Silica 20µm 90µm TTF115 ptfe Sleeve Fused Silica 25µm 375µm Micromass UK Limited Floats Road Wythenshawe M23 9LZ Tel: +44
More informationImproving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer
Improving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer Ben D. Gardner, Phillip M. Erwin, Wai Tak Lee, Amber M. Tissandier, Souzan M. Thoresen Hamilton Sundstrand
More informationRapid and Reliable Detection of Dissolved Gases in Water
Rapid and Reliable Detection of Dissolved Gases in Water Andrea Caruso and Massimo Santoro Thermo Fisher Scientific, Milan, Italy Application Note 005 Key Words Chromeleon CDS, Environmental, Fracking,
More informationDOS in Food Science & Nutrition, Manasagangotri, Mysore
Sl No UNIVERSITY OF MYSORE DOS in Food Manasagangotri, Mysore 570 006. TENDERS FOR THE SUPPLY OF EQUIPMENTS, NETWORKING & COMPUTATIONAL FACILITIES FOR UNIVERSITY OF MYSORE (UNDER UGC XI SCHEME) INVITATION
More informationReal-time PCB Gas Monitoring System (MOHMS-21GP)
Real-time PCB Gas Monitoring System (MOHMS-21GP) SHINSAKU DOBASHI* 1 YOSHIHIRO DEGUCHI* 1 MASAHARU KIRA* 2 RYUICHIRO TANAKA* 3 YOSHINORI IZAWA* 3 This report describes the technical characteristics and
More informationUNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008
UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method
More informationDemystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.
Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Jamie Brown R&D Scientist Supelco 595 North Harrison RD. Bellefonte, PA 16823 National
More informationAcclaim RSLC. (Rapid Separation Liquid Chromatography)
Acclaim RSLC (Rapid Separation Liquid Chromatography) Product Manual for Acclaim RSLC Columns Page 1 of 9 Product Manual for Acclaim RSLC Columns (Rapid Separation Liquid Chromatography) Analytical Columns
More informationTechnical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS)
Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS) 1.0 Purpose - This procedure specifies the required elements for the calibration and use of the Agilent 6890 GC interfaced
More informationHigh Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module
TECHNICAL NOTE High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module TN42256 Dr. Liliana Krotz, Dr. Francesco Leone, Walter Galotta and Dr. Guido Giazzi
More informationSensitive Analysis of Genotoxins by HPLC-ECD Using Boron-Doped Diamond Electrochemical Cell
Sensitive Analysis of Genotoxins by HPLC-ECD Using Boron-Doped Diamond Electrochemical Cell Bruce Bailey, Marc Plante, Christopher Crafts, Paul Gamache, John Waraska, and Ian Acworth ESA A Dionex Company,
More informationToxicology Gas Chromatography-Mass Spectrometry (GC-MS)
Toxicology Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose - This procedure specifies the required elements for the calibration and use of the Agilent Gas Chromatograph interfaced to the Agilent
More informationUsing the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer
application Note Gas Chromatography/ Mass Spectrometry Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer Overview A mass spectrometer
More informationZebron Gas Management
Zebron www.phenomenex.com/gasmanagement What are the key benefits of Zebron filters and traps? The high adsorption capacity of the Zebron Filters removes contaminants from mobile phase gases, reducing
More informationSOP: Buck Scientific BLC-20P HPLC Operation
Page 1 of 11 Approvals Preparer: John Buford Reviewer: Tim Kull Reviewer: Dr. Margaret Bryans Date: 16OCT13 Date: 30OCT13 Date: 31OCT13 1. Purpose 1.1. Basic operation of the Buck Scientific BLC-20P isocratic
More informationAKTA pure 25 New Owner s Intro
AKTA pure 25 New Owner s Intro The exercise below will give a quick demonstration of how easy and intuitive the AKTA pure 25 will be for you in demonstrating downstream processing to your students. Steps
More informationChemical Analysis for Methyltestosterone
APPLIED CONSUMER SERVICES, INC. Chemical Analysis for Methyltestosterone Report #: 27584 Report to: Mr. Mitch Ross 24 Dockside Lane # 131 Key Largo, Florida 33017 2016 1 1 8 9 0 N W 8 7 t h C o u r t,
More informationAgilent 7800 ICP-MS. Specifications and Typical Performance. Fast track metals analysis with Solution-Ready ICP-MS
Agilent 7800 ICP-MS Specifications and Typical Performance Fast track metals analysis with Solution-Ready ICP-MS The Solution-Ready Agilent 7800 Quadrupole ICP-MS combines proven, robust hardware, auto-optimization
More informationA NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS
A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS B.Lakshmi,Kallam Haranadhareddy Institute of Technology, Guntur Prof T.V.Reddy,Prof Mallareddy College of Engineering, Secunderabad
More informationAgilent 1260 Infinity Purification Solution
Agilent 1260 Infinity Purification Solution System User Guide Agilent Technologies Notices Agilent Technologies, Inc. 2014-2015, 2016 No part of this manual may be reproduced in any form or by any means
More informationLaboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.
Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions
More informationDegassing ERC KK Saitama, Japan
Degassing Degassing Solutions IDEX Health & Science degassers improve fluidic instrument precision and reliability by removing dissolved gases from fluids before they outgas and form problem-causing bubbles
More informationGradients in ion chromatography. Optimal separation with shorter analysis time
Gradients in ion chromatography Optimal separation with shorter analysis time Gradients yield precise results faster 02 The goal of ion chromatography in a nutshell reads: optimum separation within an
More informationMouse and Rat Tail Cuff Blood Pressure Systems from 1-24 Animals. Mouse rat tail cuff blood pressure system components
Mouse and Rat Tail Cuff Blood Pressure Systems from 1-24 Animals Mouse rat tail cuff blood pressure system components Tail Cuff Method Sensors Rodent Restrainers Software Chart Recorder Amplifier with
More informationID-Cube. Installation and Operations Manual For ID-Cube Source and Power Supply. Revision 2 Document #:
ID-Cube Installation and Operations Manual For ID-Cube Source and Power Supply Revision 2 Document #: 7.1.002 IonSense Inc. 999 Broadway Suite 404 Saugus, MA 01906 Copyright 2014 by IonSense Inc. All rights
More informationTrace analysis using ambient ionization on miniature mass spectrometers
Trace analysis using ambient ionization on miniature mass spectrometers Zheng Ouyang, R. Graham Cooks, Sandilya Garimella, He Wang and Fatkhulla Tadjimukhamedov Purdue University Miniaturization of Analysis
More informationGeneral Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!
The portable Gas Chromatograph is now more versatile than ever. DPS Instruments is pleased to present the newest Portable Companion 2 Gas Chromatograph with room for 2 Detectors. The Companion 2 GC was
More informationUP-F Delrin TM Winged Nut. UP-F PEEK TM Nut. with F-142 PEEK TM Ferrule
One Piece or Two? A one-piece fitting is more convenient and less cumbersome, since the ferrule cannot stick in a receiving port and the fitting is more easily found if dropped. With two-piece Fingertight,
More informationAnalysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS
Application Note: 424 Analysis of and Cyanuric Acid in Food Matrices by LC-MS/MS PeterVarelis, National Center for Food Safety and Technology, Illinois Institute of Technology. Jonathan Beck, Kefei Wang,
More informationDevelopment of DMS and Acetonitrile, and Formaldehyde gas standards. Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin
Development of DMS and Acetonitrile, and Formaldehyde gas standards Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin The amount of chemical substances in cylinder container Purity uncertainty Gravimetry
More informationIJPAR Vol.6 Issue 1 Jan - Mar Journal Home page:
IJPAR Vol.6 Issue 1 Jan - Mar -2017 Journal Home page: ISSN:2320-2831 Research article Open Access Optimization of stability indicating RP-HPLC method for the estimation of an anti-cancer drug Sorafenib
More informationOperation of the Perkin Elmer TGA-GC/MS
Operation of the Perkin Elmer TGA-GC/MS Summary of the TGA-GC/MS: The TGA-GC/MS allows a user to decompose a sample by heating, measure its loss of mass, and simultaneously analyze the chemical composition
More informationPortable Oxygen Concentrator. Introduction
Portable Oxygen Concentrator Introduction 1 Oxus Facilities Oxus America, Inc. Auburn Hills, MI USA Incorporated 4/2010 Product Design Manufacturing Regulatory approvals FDA/GMP CE / ISO13485 Responsible
More informationPrecisely HPLC. Hints, Tips, Guidelines and Important Considerations. Presenters: Wilhad Reuter / Tom Breslin
HPLC Hints, Tips, Guidelines and Important Considerations Presenters: Wilhad Reuter / Tom Breslin Important HPLC Solvent Characteristics: ALL injected compounds/diluents must be soluble High purity and
More informationPFC Analysis Kit for ACQUITY UPLC System Guide
PFC Analysis Kit for ACQUITY UPLC System Guide 71500183002/Revision A Copyright Waters Corporation 2009 All rights reserved Copyright notice 2009 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA
More informationA simple and rapid ESI-LC-MS/MS method for simultaneous screening of doping agents in urine samples
Research Article A simple and rapid ESI-LC-MS/MS method for simultaneous screening of doping agents in urine samples I. Madhusudhana Reddy, Alka Beotra, S. Jain, S. Ahi ABSTRACT National Dope Testing Laboratory,
More informationSpecifications...Easy CLIA Classification: Sample Type: Blood gas analyzer operation has never been simpler Sample Size: Measured Parameters PCO2
medicacorp.com easy inside and out With EasyBloodGas, Medica has redefined blood gas analyzer design The sophistication of traditional blood gas analyzers has been packaged in a new compact format with
More information