Precisely HPLC. Hints, Tips, Guidelines and Important Considerations. Presenters: Wilhad Reuter / Tom Breslin

Size: px
Start display at page:

Download "Precisely HPLC. Hints, Tips, Guidelines and Important Considerations. Presenters: Wilhad Reuter / Tom Breslin"

Transcription

1 HPLC Hints, Tips, Guidelines and Important Considerations Presenters: Wilhad Reuter / Tom Breslin

2 Important HPLC Solvent Characteristics: ALL injected compounds/diluents must be soluble High purity and degassed Noncorrosive Low viscosity UV transparency Solvents should not alter component s activity in sample Commercial bottled solvents should not be re-filtered Add 1% MeOH or 0.5% ACN to water. Not recommended using pure water > 2 days old Do not use halogenated acids; e.g., HCl Typically excludes IPA For UV detection, solvents must not absorb close to wavelength(s) of interest; above UV cutoff Especially for pharmaceuticals and biomolecules Select such that components elute between 2 to 20 column volumes ( k )

3 Solvent Degassing Mobile phase degassing is important for accurate pump blending and to avoid bubble formation in the detector flowcell. Techniques: BOILING Hmm!, not advised VACUUM Useful for only hours; Must combine w/ sonication HELIUM (at least High Purity) SPARGE CONTINUOUS SPARGE SPARGE AND PRESSURIZE ON-LINE VACUUM { Uses too much helium Best choice, when available Sparge for 5 min; then let pressurize Good choice when helium unavailable; 80% as effective as helium degassing

4 Effect of (Not) Degassing on Retention Time Pump mixed mobile phase at 30 O C; Injections every 6 min and, after Helium stopped, every 4 hours Solvent A = 1% HOAc in MeOH Solvent B = 1% HOAc in H 2 O Mobile Phase = 34% A 0.22% CCA Chlorthalidone 1.6 Retention Time (min) % (RSD) Not Degassed Not Pressurized Degassed and Pressurized 1.5% Stop Helium Pressurization % % 1.3% min 16 Hours Injection Number

5 Effect of temperature Bottom line: Avoid ambient temperatures Premixed solvent; starting at 30 Degrees 3.4% CCA Chlorthalidone Retention Time (min) % (RSD) o 30 C w/ Oven Pressurized and Degassed Room Temperature o C Pressurized and Degassed 2.5% 0.27% o 30 C w/ Oven Pressurized and Degassed % 0.30% Hours Injection Number

6 Common Mobile Phase Modifiers Buffers To stabilize phs of mobile phase under reverse phase (RPC) or ionexchange (IEC) chromatography (phosphate, acetate, citrate); ~5-20 mm Triethylamine (TEA) To reduce tailing of basic analytes RPC; ~0.1% Phosphoric, acetic acid or formic acids Ion suppressors To suppress ionization of acidic analytes under RPC; ~0.1% Ion-pairing reagents For separation of ionic compounds using RPC column (C18) Anions: Quaternary ammonium salts (e.g., Tetrabutylammonium hydroxide); ~2-20 mm Cations: Alkyl sulfonates (e.g., Hexane sulfonate); ~2-20 mm

7 Series 200 Analytical Pump: Low Pressure Mixing Solvent toward column Solvents Solvent into metering chamber 100 µl 107 µl Solvent mixing delay volume: ~ uL, depending on pressure Advantages: Great solvent mixing and built-in solvent compressibility factor

8 Low-Pressure Mixing Pump Blending Performance

9 Series 200 Micro-Pump: Binary High-Pressure Mixing To Injector Purge Valve Purge Valve Tee Pump A Pump B Solvent mixing delay volume: ~10-400uL with static mixer, depending on flow rate Advantage: Very low solvent mixing delay; good for high speed gradients and LC/MS

10 Method Transferability Concerns Low-Pressure Mixing: 1.0mL Delay Volume = High-Pressure Mixing: 0.1mL Delay Volume = 2 ml/min

11 Using Injection Delay Time (for Method Transferability) Low-Pressure Mixing: 1.0mL Delay Volume = High-Pressure Mixing: 0.1mL Delay Volume = Low-Pressure mixing with 0.45 min Injection Delay Time 2 ml/min

12 HPLC: Preferred Materials for Wettable Parts STAINLESS STEEL Industry Standard SS 316; May rarely effect biological activity or allow adverse chealation to occur TITANIUM (alternative to SS) PEEK (Poly-Ether Ether-Ether-Ketone) TEFLON / VESPEL / TEFZEL POLYAMIDE / POLYIMIDE QUARTZ-SILICA GLASS Good biocompatibility. Somewhat more brittle than SS Good biocompatibility and often used for basic compounds. Becomes brittle w/ >20% THF or MeCl 2 Used for rotor seals and low- pressure tubing. Note: : Use PEEK for basic compounds Do not use with MeCl 2 Commonly used for flow cells RUBY ` Used for balls in pump check valves SAPPHIRE Used for pump pistons

13 Pressure Problems PRESSURE TOO HIGH Dirty column Blockages e.g., system filter, guard column or column inlet frit, tubing, injector needle or valve, detector flow cell. >> Check by isolating pressure point(s); like a plumber Check flow rate Different solvent mixture? Are solvent lines switched?

14 Pressure Problems PRESSURE TOO LOW Leaks Pump malfunction e.g., piston seal, connections, rotor seal e.g., vapor lock, sticking check valves or solenoid valve, broken piston Plugged solvent filter Check flow rate Different solvent mixture? Lines switched?

15 Autosampler: Basic Injection Cycle Based on Rheodyne injector valve Poor peak shape may result if either the connections to injector ports 2 and 3 are reversed or one fills sample loop volume over 50% (in partial-loop mode). Sample from Syringe

16 Precision Problems Possibly Related to Autosampler Overfilling the sample vial Recommended: No more than 2/3 full Tightly sealed vials not vented Turn venting option on Excessive sampling speed Especially for viscous and low volume samples (<5µL) L): : Reduce speed to slow Bubbles in the autosampler sampling lines or syringe Degas flush solvent Worn rotor seal in the injection valve Results in poor RSD: Replace seal

17 Detectors for Any Application PerkinElmer Series 200 Detectors UV/Vis PDA UV/VIS considerations: Can detect predominant compounds being analyzed by HPLC Useful for any compounds that absorb between 190 and 700 nm Very good sensitivity (ppb). Typically, better than RI but not as good as Fluor. Exceptional baseline stability Wavelength programmable during run PDA considerations: Approaches UV/VIS sensitivity (ppb) Dual lamps (Deuterium and Tungsten) for full UV/Vis wavelength range Spectral data useful for confirmation and identification of components Capability of peak purity assessment Network-Direct connection for direct transfer of data and spectra to data directories on the network

18 Detectors for Any Application PerkinElmer Series 200 Detectors FL RI Fluorescence considerations: Very selective (must fluoresce) Exceptional sensitivity (ppt) May have to derivatize if compounds do not fluoresce in native state. Refractive Index considerations: Universal detection Good sensitivity (ppm) Sensitive to temperature changes (RI must have built-in Peltier temp. control; 35 C usually good for most labs) Can not be used for gradient method

19 Detectors: What to Consider Lamp Energy: Lamp Hours: Wavelength Accuracy: Reference Wavelength (PDA): Flow Cell Condition: RI Reference Purge Valve: Column Connections: Check periodically Lamp lifetime: typically ~1500 hours for UV/Vis, ~1000 for PDA and 500 for FL detectors Check periodically Choose a wavelength transparent to both sample and solvent On PDA, remove flow cell periodically; hold it toward light and check clarity Shine light into UV/Vis flow cell Check position Check that column is connected to Inlet tubing on detector

20 Detector Pointers Baseline problems related to detectors may be: Mechanical Leaky flow cell, trapped bubble Optical Aging lamp, dirty flow cell Bad mobile phase Not detector Electrical Call vendor Don't wash columns while they are connected to the detector. When possible, the use of a 50psi back pressure device is recommended after the detector (suppresses outgassing)

21 Suggested Periodic Performance Testing LC PUMPS Flow, composition calibration and pressure Leak test AUTOSAMPLER Precision, carryover, linearity testing DETECTORS Wavelength check Noise and drift testing Linearity testing

22 Suggested Spare Parts and Replacement Period LC PUMPS High/low pressure piston seals and backup rings kit (~ once a year); Part# N Might want to replace high pressure piston every other year AUTOSAMPLER Injector rotor seal (~ once a year) Injector needle (if properly maintained, should hardly ever need replacing) Might want to replace flush and/or sampling syringe periodically (inspect first) DETECTORS Lamp (order new one only after 1000 hours on current UV/Vis or 500 hours on PDA - they have a shelf life)

23 Chromatographic Diagnostics NO PEAKS: Lamp off/bad, check wavelength Injector problem Check solvent selection / strength No sample in vial CARRYOVER: Materials incompatibility eg., Teflon vs PEEK Insufficient injector flushes or flush speed Column overloaded Some columns exhibit component carryover For that application, switch to another column vendor

24 Chromatographic Diagnostics Normal noise Expanded; always a good practice before starting a run

25 Chromatographic Diagnostics 5 min Air bubbles in solvent Degas solvent with in-line degasser or 5min helium sparge then pressurize Aging/bad detector lamp Replace lamp (typically, order new lamp after 500 hours on current one) Damaged pulse dampener on pump Replace

26 Chromatographic Diagnostics 5 min Contamination in system (solvent, degasser, pump, AS, column, detector) Remove column(s)/guard column and flush any buffers out of system; Rinse all solvent lines with HPLC water (may require co-solvent flush); Place all solvent lines into common bottle of 10% Ammonium Hydroxide; Flush for 20 min at 3mL/min. then switch to HPLC water for 2 minutes; Replace solvent bottle with 10% Nitric Acid and do another 20 min flush; Replace solvent bottle with HPLC water and do another 10 min flush; May also want to perform 20 min./1ml/min Isopropanol flush (usually not necessary); otherwise, move back to original solvents with clean bottles

27 Chromatographic Diagnostics Noise from detector, surroundings or poor analog cable grounding Monitor lines and recommend adding line conditioner Aging/bad detector lamp Replace lamp Air bubbles in solvent or dirty flow cell Degas solvent / rinse flow cell with strong co-solvent; The latter may require base/acid rinse (previous slide) May arise from fines from column > flush thoroughly without detector or change vendor Especially observed on RI and electrochemical detectors

28 Chromatographic Diagnostics A B Void volume; solvent front Unretained peak volume/time Elution volume k = 1 (capacity factor) Refractive index effect (typically from flow cell) Replace flow cell or retire detector

29 Chromatographic Diagnostics Column voids/channeling Bad column: Discard

30 Chromatographic Diagnostics Sample overload Inject less sample, dilute sample or change diluent to match mobile phase Forgot to add anti-tailing modifier to mobile phase Add Triethylamine (TEA) or similar Deteriorating column Change column Components may be too basic or acidic If using ion exchange column, switch to C18 and ion pairing mode

31 Example of Fronting Fronting Peak Caused by sample overload or component miscibility problem Inject less sample, dilute sample or change diluent to match mobile phase Check sample miscibility

32 Chromatographic Diagnostics All could be caused by sample overload Inject less sample, less concentrated sample or change diluent to match mobile phase Change to another column with higher sample capacity

33 Chromatographic Diagnostics Proportionally increasing retention times Low flow rate If same peak height, solvent leak before injector Column temperature may be lower; use column oven Possible mobile phase change (modifier, ph, molarity); however, not always proportional Possible column deterioration; however, usually not proportional

34 Chromatographic Diagnostics Disproportional retention time change Mobile phase change (modifier, ph, molarity); Column deterioration; replace column Change in column temperature; use column oven

35 Chromatographic Diagnostics Proportional drop in peak height Air in autosampler flush line Check lamp energy: lamp energy may be getting significantly lower. Time to change lamp Check wavelength (not always proportional) Sample absorbing to column or other material; change all teflon contact points in autosampler to PEEK material (rotor seal, etc.) or change to other vendor s column Autosampler: leaky rotor seal or plugged needle Volatile components: use capped vials (not always proportional)

36 Thank you for attending!

37 Chromatographic Diagnostics Component interconversion or isomerization Check ph, especially relative to pka(s) of components Check Temperature

Acclaim RSLC. (Rapid Separation Liquid Chromatography)

Acclaim RSLC. (Rapid Separation Liquid Chromatography) Acclaim RSLC (Rapid Separation Liquid Chromatography) Product Manual for Acclaim RSLC Columns Page 1 of 9 Product Manual for Acclaim RSLC Columns (Rapid Separation Liquid Chromatography) Analytical Columns

More information

HPLC systems. Overview

HPLC systems. Overview HPLC systems Overview An hplc system is just a way to get a sample onto the end of a column, to pump solvents through the column until the analytes drop out the end, and to detect them as they appear.

More information

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION Standard OQ Test Suite This document describes the test program for qualifying LC analytical-scale Agilent systems; the following table lists

More information

Troubleshooting. Select the Table of Contents here Rheodyne All rights reserved

Troubleshooting. Select the Table of Contents here Rheodyne All rights reserved nmlkji Literature nmlkj Site Troubleshooting This section contains our complete Troubleshooting Guide. It presents detailed discussions of the most common problems encountered when using sample injection

More information

AKTA pure 25 New Owner s Intro

AKTA pure 25 New Owner s Intro AKTA pure 25 New Owner s Intro The exercise below will give a quick demonstration of how easy and intuitive the AKTA pure 25 will be for you in demonstrating downstream processing to your students. Steps

More information

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager Ghost Peaks: How to Fix a Haunting Problem Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager Outline of Topics What is a ghost peak? How to identify ghost peaks and their source How to correct

More information

HPLC troubleshooting guide

HPLC troubleshooting guide CASE STUDY 70303 HPLC troubleshooting guide A free app that will save time and frustration Introduction During the development of a new HPLC method, there were unexpected results all peaks, except for

More information

no peaks missing peaks too small peaks increasing retention decreasing retention

no peaks missing peaks too small peaks increasing retention decreasing retention GC Troubleshooting We can help you. 1. Observation: no peaks no peaks missing peaks too small peaks increasing retention decreasing retention declining baseline rising baseline bleeding plateaus interfering

More information

HPLC. Solvent Delivery Module Systems

HPLC. Solvent Delivery Module Systems HPLC Solvent Delivery Module Systems Who is GBC Scientific Equipment? GBC Scientific Equipment was founded in 1978 and is now the world's third largest and fastest growing manufacturer of atomic absorption

More information

The instrument should have a mass range from 2 to 2000 m/z or better Interface

The instrument should have a mass range from 2 to 2000 m/z or better Interface Revised Specifications for Liquid Chromatography Mass Spectrometry (LC-MS/MS ) Triple Quadrupole Specification for Ultra-Fast Liquid Chromatography system Solvent Delivery 1) A quaternary Pump system for

More information

DOS in Food Science & Nutrition, Manasagangotri, Mysore

DOS in Food Science & Nutrition, Manasagangotri, Mysore Sl No UNIVERSITY OF MYSORE DOS in Food Manasagangotri, Mysore 570 006. TENDERS FOR THE SUPPLY OF EQUIPMENTS, NETWORKING & COMPUTATIONAL FACILITIES FOR UNIVERSITY OF MYSORE (UNDER UGC XI SCHEME) INVITATION

More information

GCMSD-Headspace Analysis SOP

GCMSD-Headspace Analysis SOP Before you start GCMSD-Headspace Analysis SOP Method and Sequences names are restricted to 50 characters (MSD program will crash otherwise) The GC oven and Injection Ports need to be cooled to 50 oc to

More information

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC Evaluation of linear and step gradient performance and retention time precision of the Agilent 12 Series Rapid Resolution LC This document is believed to be accurate and up-to-date. However, Agilent Technologies,

More information

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion: Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: The RSK-175 standard operating procedure was developed in order to determine the amount of dissolved gas in water. Due to the expansion

More information

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE. Gases&Technology FEATURE Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. B Y M A T T H E W M O N A G L E Abstract Bulk gases are often delivered to

More information

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (16) 17 R QUALIFICATION OF EQUIPMENT ANNEX 2: QUALIFICATION OF GC EQUIPMENT Full document title and reference Document type Legislative

More information

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction Application Note Rapid performance verification of AZURA systems with refractive index detector Method Keywords ID HPLC Quality control, system verification, refractive index, AZURA Analytical HPLC Plus

More information

BETTER SAMPLE CARE. Spark Holland introduces. SPH1299 UHPLC Gradient Pump

BETTER SAMPLE CARE. Spark Holland introduces. SPH1299 UHPLC Gradient Pump BETTER SAMPLE CARE Spark Holland introduces SPH1299 UHPLC Gradient Pump SPH1299 Spark Holland introduces SPH1299 UHPLC Gradient Pump Robust, Precise and Easy to use Innovative pumping technology created

More information

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON'

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON' DISSOLUTION ASSAY Release & Stability Studies Ed. 04. General Index: 1. PRODUCT SPECIFICATIONS (USP MONOGRAPH) 2. IDENTIFICATION BY HPLC 3. ASSAY - HPLC SETUP 4. STANDARD PREPARATION 5. SYSTEM SUITABILITY

More information

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries. Informal Commentary Limit of Diethylene Glycol (DEG) and Ethylene Glycol (EG) in Sorbitol Solution, Sorbitol Sorbitan Solution and Noncrystallizing Sorbitol Solution December 2009 Monograph/Section(s):

More information

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note Gas Chromatography Author Matthew S. Klee Agilent Technologies 2850 Centerville Road Wilmington, DE 19808-1610 USA Abstract

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs. Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography

More information

Setting a New Standard in Flash Chromatography Performance

Setting a New Standard in Flash Chromatography Performance Setting a New Standard in Flash Chromatography Performance What s New Faster Flow rates, up to 300 ml/min Higher pressures, up to 300 psi (20 bar) Bigger touchscreen, 12 or 15 inches Wider Dynamic Range

More information

Automated Determination of Dissolved Gases in Water Anne Jurek

Automated Determination of Dissolved Gases in Water Anne Jurek Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: Tapping the natural gas reservoirs throughout the United States has long been a viable solution for energy independence; however

More information

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD) Procedure No: SOP-026 Revision No: 1.0 January 24, 2011 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to meet

More information

Indian Institute of Technology Kanpur Environmental Engineering and Management Department of Civil Engineering

Indian Institute of Technology Kanpur Environmental Engineering and Management Department of Civil Engineering Indian Institute of Technology Kanpur Environmental Engineering and Management Department of Civil Engineering Professor Vinod Tare Phones:+91-0512-2597791,7792 Email: vinod@iitk.ac.in Post Office: IIT

More information

Figure 1: Plug in ion inlet of mass spec

Figure 1: Plug in ion inlet of mass spec Starting procedure: by Vicky - is the vacuum pump running (should run all time, otherwise it ll need 4-5 hours to have a adequate vacuum for the MS) otherwise something has gone wrong- F.T - Remove plug

More information

Agilent 1260 Infinity Purification Solution

Agilent 1260 Infinity Purification Solution Agilent 1260 Infinity Purification Solution System User Guide Agilent Technologies Notices Agilent Technologies, Inc. 2014-2015, 2016 No part of this manual may be reproduced in any form or by any means

More information

Valves FLUIDICS VALVE OVERVIEW & FUNCTIONS ROTARY SHEAR VALVES STAND ALONE VALVE PRODUCTS MANUAL VALVES SPARE PARTS VALVE ACCESSORIES

Valves FLUIDICS VALVE OVERVIEW & FUNCTIONS ROTARY SHEAR VALVES STAND ALONE VALVE PRODUCTS MANUAL VALVES SPARE PARTS VALVE ACCESSORIES 114 Valves FLUIDICS Our valves are an integral part of advanced fluid-handling solutions for a wide range of analytical instrumentation and clinical diagnostic systems. Our valve options include manual

More information

For that reason VICI Jour developed:

For that reason VICI Jour developed: VICI Jour Safety VICI Safety Products VICI Safety Products NEW VICI strives to provide customers with the best solutions for their problems to make the work better and/or easier, with respect to environment

More information

UP-F Delrin TM Winged Nut. UP-F PEEK TM Nut. with F-142 PEEK TM Ferrule

UP-F Delrin TM Winged Nut. UP-F PEEK TM Nut. with F-142 PEEK TM Ferrule One Piece or Two? A one-piece fitting is more convenient and less cumbersome, since the ferrule cannot stick in a receiving port and the fitting is more easily found if dropped. With two-piece Fingertight,

More information

DEBUBBLERS & DEGASSERS ACTIVE DEBUBBLERS APPLICATION NOTE WHY DEGAS YOUR MOBILE PHASE? PAGE 177 STAND ALONE DEGASSERS PAGE 175 PAGE 178

DEBUBBLERS & DEGASSERS ACTIVE DEBUBBLERS APPLICATION NOTE WHY DEGAS YOUR MOBILE PHASE? PAGE 177 STAND ALONE DEGASSERS PAGE 175 PAGE 178 174 DEBUBBLERS & DEGASSERS ACTIVE DEBUBBLERS PAGE 175 APPLICATION NOTE WHY DEGAS YOUR MOBILE PHASE? PAGE 177 STAND ALONE DEGASSERS PAGE 178 SEE STARRED PRODUCTS FOR TOP SELLERS! Series DEBUBBLERS & DEGASSERS

More information

ACQUITY UPLC Sample Manager - Fixed Loop Operator s Overview and Maintenance Information

ACQUITY UPLC Sample Manager - Fixed Loop Operator s Overview and Maintenance Information ACQUITY UPLC Sample Manager - Fixed Loop Operator s Overview and Maintenance Information Revision A Copyright Waters Corporation 2010 2011 All rights reserved Copyright notice 2010 2011 WATERS CORPORATION.

More information

Prep LC Training Manual S.V.

Prep LC Training Manual S.V. 1234 Hach Hall 515-294-5805 www.cif.iastate.edu Prep LC Training Manual 11-6-2013 S.V. Location: Contact: 1238 Hach Hall Kamel Harrata, 1236 Hach Hall and Steve Veysey, 1234 Hach Hall Safety: All researchers

More information

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port So, Why Do I Have To Do Maintenance? 1. Things wear out septa, syringes, nuts, ferrules, o-rings, etc.

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results. Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions

More information

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek Abstract: Due to the expansion of natural gas drilling through horizontal fracturing, there has been

More information

SOP: Buck Scientific BLC-20P HPLC Operation

SOP: Buck Scientific BLC-20P HPLC Operation Page 1 of 11 Approvals Preparer: John Buford Reviewer: Tim Kull Reviewer: Dr. Margaret Bryans Date: 16OCT13 Date: 30OCT13 Date: 31OCT13 1. Purpose 1.1. Basic operation of the Buck Scientific BLC-20P isocratic

More information

Autosampler problems!

Autosampler problems! Laserchrom HPLC Laboratories Ltd Technical Bulletin Autosampler problems! Air in the syringe Bent or blocked needle Valve problems Unreproducible injections Sample carryover Wash solution problems Vial,

More information

Running HPLC Methods on an ACQUITY UPLC System

Running HPLC Methods on an ACQUITY UPLC System Running HPLC Methods on an ACQUITY UPLC System Note: This document is an addendum to the ACQUITY UPLC System Operator s Guide. Contents Topic Page System considerations 1-1 Choosing fittings 1-3 System

More information

Gradients in ion chromatography. Optimal separation with shorter analysis time

Gradients in ion chromatography. Optimal separation with shorter analysis time Gradients in ion chromatography Optimal separation with shorter analysis time Gradients yield precise results faster 02 The goal of ion chromatography in a nutshell reads: optimum separation within an

More information

The instrument will process the samples one by one till all the samples have been cleaned up.

The instrument will process the samples one by one till all the samples have been cleaned up. Table of Contents SPE-01 Cleanup Station... 1 LC-04SP Valve System... 5 LC-04SP Standalone Valve... 10 LC-05 Auto Injector... 12 LC-06 Dosing Pump... 16 LC-08 Column Selector... 17 SPE-01 Cleanup Station

More information

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column Before you connect your trap and column You need to provide your own trap and column (Note our fittings accommodate trap and column made with fused silica 360 µm OD). Check our guide for packing traps

More information

Generating Calibration Gas Standards

Generating Calibration Gas Standards Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and

More information

OPERATORS MANUAL G U A R A N T E E D T R Y C H E M APPROVED

OPERATORS MANUAL G U A R A N T E E D T R Y C H E M APPROVED OPERATORS MANUAL G U A R A N T E E D C H E M I S T R Y APPROVED Operator Manual Post-Column Derivatization Instrument 1280 Space Park Way Mountain View CA 94043 USA 800-654-3330 650-694-6700 G U A R A

More information

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Jamie Brown R&D Scientist Supelco 595 North Harrison RD. Bellefonte, PA 16823 National

More information

Dionex IonPac AS19-4µm

Dionex IonPac AS19-4µm Total Thermo Scientific Dionex IonPac AS19-4µm Column Product Manual P/N: March 2015 Part of Thermo Fisher Scientific Product Manual for Dionex IonPac AS19-4µm Column Dionex IonPac AG19-4µm Capillary Guard

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 9 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 3.1 Canister Receipt 3.2 Canister Storage 4.0 INTERFERENCE AND POTENTIAL PROBLEMS

More information

FROG-4000 TM Chemical Analysis System

FROG-4000 TM Chemical Analysis System FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies, Inc. 2015 Vol.5 This page is intentionally left blank. FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies,

More information

LC, RSLC, Bio RS and Dual Gradient Systems

LC, RSLC, Bio RS and Dual Gradient Systems Thermo Scientific Dionex UltiMate 3000 LC, RSLC, Bio RS and Dual Gradient Systems P/N 4823.0205 Revision B December 2014 Automated Method Scouting Considerations All information in this manual is subject

More information

Extended Application Note

Extended Application Note Extended Application Note Harmful Substances in Dietary Supplements HPLC Columns with APP A-337 www.mtc-usa.com 1-732-578-1777 INTRODUCTION A dietary supplement is a substance to be consumed for the purpose

More information

Prep LC Training Manual

Prep LC Training Manual Prep LC Training Manual (11-6-2013) Location: 1238 Hach Hall Contacts: kamel Harrata and Steve Veysey, 1234 Hach Hall Safety: Introduction: The Agilent PrepStar SD-1 preparative HPLC is a student operated

More information

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation CHE 4115 Chemical Processes Laboratory 2 Experiment 1 Batch Distillation BACKGROUND Distillation is one of the most commonly used unit operations in chemical engineering. In general, a distillation operation

More information

Best Practice for Identifying Leaks in GC and GC/MS Systems

Best Practice for Identifying Leaks in GC and GC/MS Systems Best Practice for Identifying Leaks in GC and GC/MS Systems Technical Overview Maintaining a leak-free GC or GC/MS system is critical for obtaining optimal system performance with reliable, reproducible,

More information

Multiple Gas#5 GC configuration Jan 2016

Multiple Gas#5 GC configuration Jan 2016 History: Unfortunately there is no single column that can separate: Hydrogen Oxygen Nitrogen Methane CO CO2 Ethane Water Propane Butane Pentane Over the years SRI Instruments has devised several solutions

More information

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more! After years of development and testing, DPS Instruments is pleased to present the newest, most expandable and versatile Gas Chromatography Systems in history. The DPS 600 Series GC systems are the world

More information

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation User s Booklet for the Wyatt minidawn Light Scattering Instrumentation The Wyatt minidawn Light Scattering instrument is able to measure the weight average molecular weight of a synthetic polymer or a

More information

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS UCT Part Numbers: CUQAX22Z Clean-Up C8+QAX, 2mg/1mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQID21-3UM

More information

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane Gas requirements Gases for packed columns The carrier gas you use depends upon the type of detector and the performance requirements. Table 520-1 lists gas recommendations for packed column use. In general,

More information

Video Notes. LC Troubleshooting Series Retention Time Shifts

Video Notes. LC Troubleshooting Series Retention Time Shifts Video Notes LC Troubleshooting Series Retention Time Shifts Introduction John Henderson is an Applications Scientist for Agilent Chemistries and Supplies Division. Maureen Joseph is the HPLC Columns Product

More information

Fiber Optic SMA Z-Flow Cell Manual Version 2.1

Fiber Optic SMA Z-Flow Cell Manual Version 2.1 Fiber Optic SMA Z-Flow Cell Manual Version 2.1 The line of fiber optic SMA Z-Cells is the top choice for many absorbance measurements. SMA-Z-Cell path lengths are available between 2.5mm - 100mm and can

More information

Rev. CO Packed Columns for Normal Phase Chromatography. TSKgel NH m INSTRUCTION MANUAL

Rev. CO Packed Columns for Normal Phase Chromatography. TSKgel NH m INSTRUCTION MANUAL Rev. CO067000 Packed Columns for Normal Phase Chromatography TSKgel NH2-100 3 m INSTRUCTION MANUAL Safety Precautions To help protect you and/or your property from potential damage, please read this manual

More information

Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION

Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION Install your Blue River Technologies Port-A-Poly mixing system in a clean dry area.

More information

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment FAST SHIPPING AND DELIVERY TENS OF THOUSANDS OF IN-STOCK ITEMS EQUIPMENT DEMOS HUNDREDS OF MANUFACTURERS SUPPORTED

More information

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS Abstract Highly sensitive analysis of metanephrines in plasma is critical in the diagnosis and treatment of pheochromocytoma and paraganglioma. Here, a HILIC LC-MS/MS method was developed using a Raptor

More information

Peak Biotech - Company & Product Presentation

Peak Biotech - Company & Product Presentation Peak Biotech - Company & Product Presentation July 2005 Peak Biotech Company & Product Presentation Peak Biotech Company Profile Location Jacob Nøhr Peak Biotech A/S was founded Kvistgaard, Denmark QA

More information

Rapid and Reliable Detection of Dissolved Gases in Water

Rapid and Reliable Detection of Dissolved Gases in Water Rapid and Reliable Detection of Dissolved Gases in Water Andrea Caruso and Massimo Santoro Thermo Fisher Scientific, Milan, Italy Application Note 005 Key Words Chromeleon CDS, Environmental, Fracking,

More information

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial Supporting Information for Micro-Collection of Gases in a Capillary Tube: Preservation of Spatial and Temporal Resolution AUTHOR INFORMATION Kristin A. Herrmann Favela, 1 * Pieter Tans, 2 Thomas Jaeckle,

More information

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany Manual Degasser DDG-75 April 2007 1 Manual Online Degasser DDG-75 DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE-68766 Hockenheim Germany Tel. +49-6205-9450-0 Fax. +49-6205-9450-33 email info@duratec.com

More information

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment FAST SHIPPING AND DELIVERY TENS OF THOUSANDS OF IN-STOCK ITEMS EQUIPMENT DEMOS HUNDREDS OF MANUFACTURERS SUPPORTED

More information

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions Varian 3300 Gas Chromatograph Abbreviated Operating Instructions 1. Introduction and Startup 2 2. Basic Configurations for the GC 3 3. Sample Preparation 3 4. Building a Method 3 5. Split/Splitless Injection

More information

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI NanoSight NS300 Operation instructions Injection/flushing brief overview: 1. Do not exceed flow of 1 ml per 20 seconds. 2. Inject two 1 ml syringes with nano-pure or DI water. 3. If the water does not

More information

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic. 458 LCGC NORTH AMERICA VOLUME 24 NUMBER 5 MAY 2006 www.chromatographyonline.com LC TROUBLESHOOTING Dwell Volume Revisited Gradient transfer does not have to be problematic. John W. Dolan LC Troubleshooting

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310 www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Varian 310 310 LC/MS Manual System INCLUDES: 310-MS LC/MS triple quadrople

More information

FROG-4000 TM Chemical Analysis System

FROG-4000 TM Chemical Analysis System FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies, Inc. 2014 Vol.4 This page is intentionally left blank. FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies,

More information

NMR Service. How to Prepare Samples for NMR

NMR Service. How to Prepare Samples for NMR NMR Service How to Prepare Samples for NMR In NMR, unlike other types of spectroscopy, the quality of the sample has a profound effect on the quality of the resulting spectrum. If you follow a few simple

More information

Operating Instructions Multi-Channel Vacuum Degasser

Operating Instructions Multi-Channel Vacuum Degasser Operating Instructions Multi-Channel Vacuum Degasser Introduction The Vacuum Degasser is a high-efficiency in-line module that removes dissolved gasses from HPLC solvents. Its unique design assures reliable

More information

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS Peter Alden and Michael Jones Waters Corporation, Milford, MA, U.S. APPLICATION BENEFITS UPLC combines

More information

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES 1. Applicability and Principle 1.1 Applicability. This method applies to the analysis of carbon dioxide

More information

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection This user guide does not have the intention to introduce you to any of the theory and/or principles behind

More information

MV5: Automated concentration system

MV5: Automated concentration system MV5: Automated concentration system Abstract LabTech MV5 is the latest automatic nitrogen evaporator developed by our R&D Team able to process up to 54 samples simultaneously. This advanced instrument

More information

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY Nengbing Xu, Hongmei Ying and Libo Zhu Ningbo Environmental Monitoring

More information

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013 The SRI ASTM 3612/TOGA/ Dissolved Gas GC configuration ( TOGA GC ) permits measurement of gasses dissolved in water, oil or other liquids, and is based on the requirements of ASTM method 3612, option C,

More information

Zebron Gas Management

Zebron Gas Management Zebron www.phenomenex.com/gasmanagement What are the key benefits of Zebron filters and traps? The high adsorption capacity of the Zebron Filters removes contaminants from mobile phase gases, reducing

More information

HICHROM. HPLC Accessories

HICHROM. HPLC Accessories HICHROM HPLC Accessories HICHROM Index PEEK Tubing SLIPFREE HPLC Column Connectors 1 Stainless Steel Tubing 7 PEEK Unions, Tees and Crosses 8 2 Micro Flow Fittings Tubing Cutters 9 3 Fingertight Fittings

More information

Buck Scientific, Inc.

Buck Scientific, Inc. Buck Scientific, Inc. 910/310 Gas Chromatographs The Buck 910/310 Gas Chromatographs In the lab, the 910-series GCs perform routine industrial analyses and complicated research separations easily, generating

More information

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator Procedure No: SOP-034 Revision No: 1.0 Revised Dec. 29, 2010 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE This procedure is intended to describe the operations of the Dasibi model 5008 calibrator. The Dasibi

More information

Capillary Conversion System CCS-4/NC & CCS-05/NC. Installation and Operating Instructions. Operating Instructions

Capillary Conversion System CCS-4/NC & CCS-05/NC. Installation and Operating Instructions. Operating Instructions Operating Instructions Check the detector for the correct operating conditions. If make-up gas is being used a flow of 30ml/min. of carrier plus make-up is normally recommended. This can most effectively

More information

Advances in Capillary Column Backflushing

Advances in Capillary Column Backflushing Advances in Capillary Column Backflushing Matthew S. Klee, Ph.D. ISCCE 2010 1 Capillary Flow Technology Traits Ease of use - ruggedness Inertness Usability to > 350 oc Excellent exclusion of air No unpurged

More information

FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS

FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS Analysis Conducted BY CC RESEARCH (P) Ltd., For Nanoceutical Solutions Inc. USA 2950 Thousand Oaks drive, Suite 3, San Antonio, Texas EXECUTIVE SUMMARY

More information

Columbus Instruments

Columbus Instruments 0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011

More information

Organic Elemental Analysis Series II CHNS/O Elemental Analyzer. proven performance. reliable results

Organic Elemental Analysis Series II CHNS/O Elemental Analyzer. proven performance. reliable results Organic Elemental Analysis 2400 Series II CHNS/O Elemental Analyzer proven performance reliable results 2400 Series II CHNS/O Elemental Analyzer raising the standard Q U I C K G L A N C E One analyzer

More information

System Overview. TCD Detector temperature setpoint Regulator

System Overview. TCD Detector temperature setpoint Regulator System Overview POPULAR CONFIGURATION GCs Your educational TCD GC is configured on the compact 310 chassis. It is equipped with a TCD Detector, a temperature programmable Column Oven, a 3 Silica Gel packed

More information

Svensk läkemedelsstandard

Svensk läkemedelsstandard Svensk läkemedelsstandard 2019.0 Carbon dioxide Postadress/Postal address: P.O. Box 26, SE-751 03 Uppsala, SWEDEN Besöksadress/Visiting address: Dag Hammarskjölds väg 42, Uppsala Telefon/Phone: +46 (0)18

More information

Preparative Valve-Based Fraction Collector

Preparative Valve-Based Fraction Collector Agilent 1260 Infinity II Preparative Valve-Based Fraction Collector User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2016-2018 No part of this manual may be reproduced in any form or

More information

Standard Operational Procedure for Capillary Electrophoresis

Standard Operational Procedure for Capillary Electrophoresis Page 1 of 14 Standard Operational Procedure for Capillary Electrophoresis *Please ask an experienced operator for training before you use the CE if you can. A demo is most probably more efficient than

More information

Oxygen Contamination

Oxygen Contamination MPMS Application Note 1014-210 Oxygen Contamination This application note describes potential sources for oxygen contamination in the sample chamber and discusses its possible effects. Molecular oxygen,

More information

POLYMER TEST KIT (CLAPPER METHOD) OFI PART No

POLYMER TEST KIT (CLAPPER METHOD) OFI PART No OFI Testing Equipment Polymer Test Kit (Clapper Method) Part # 295-00 Instructions Page 1 of 4 POLYMER TEST KIT (CLAPPER METHOD) OFI PART No. 295-00 This test determines the Partially Hydrolyzed Polyacrylamide

More information

INSTALLATION AND OPERATING INSTRUCTIONS

INSTALLATION AND OPERATING INSTRUCTIONS INSTALLATION AND OPERATING INSTRUCTIONS For 3D Instruments Direct Drive Pressure Gauges 3D Instruments Inc., 15542 Chemical Lane, Huntington Beach, CA, 92649, U.S.A. Tel. (714) 894-5351, Fax. (714) 895-4309

More information

Agilent 1290 Infinity Pump Head Maintenance

Agilent 1290 Infinity Pump Head Maintenance Agilent 1290 Infinity Pump Head Maintenance Technical Note Agilent Technologies Notices Agilent Technologies, Inc. 2012-2015, 2016 No part of this manual may be reproduced in any form or by any means (including

More information