1 HPLC Hints, Tips, Guidelines and Important Considerations Presenters: Wilhad Reuter / Tom Breslin
2 Important HPLC Solvent Characteristics: ALL injected compounds/diluents must be soluble High purity and degassed Noncorrosive Low viscosity UV transparency Solvents should not alter component s activity in sample Commercial bottled solvents should not be re-filtered Add 1% MeOH or 0.5% ACN to water. Not recommended using pure water > 2 days old Do not use halogenated acids; e.g., HCl Typically excludes IPA For UV detection, solvents must not absorb close to wavelength(s) of interest; above UV cutoff Especially for pharmaceuticals and biomolecules Select such that components elute between 2 to 20 column volumes ( k )
3 Solvent Degassing Mobile phase degassing is important for accurate pump blending and to avoid bubble formation in the detector flowcell. Techniques: BOILING Hmm!, not advised VACUUM Useful for only hours; Must combine w/ sonication HELIUM (at least High Purity) SPARGE CONTINUOUS SPARGE SPARGE AND PRESSURIZE ON-LINE VACUUM { Uses too much helium Best choice, when available Sparge for 5 min; then let pressurize Good choice when helium unavailable; 80% as effective as helium degassing
4 Effect of (Not) Degassing on Retention Time Pump mixed mobile phase at 30 O C; Injections every 6 min and, after Helium stopped, every 4 hours Solvent A = 1% HOAc in MeOH Solvent B = 1% HOAc in H 2 O Mobile Phase = 34% A 0.22% CCA Chlorthalidone 1.6 Retention Time (min) % (RSD) Not Degassed Not Pressurized Degassed and Pressurized 1.5% Stop Helium Pressurization % % 1.3% min 16 Hours Injection Number
5 Effect of temperature Bottom line: Avoid ambient temperatures Premixed solvent; starting at 30 Degrees 3.4% CCA Chlorthalidone Retention Time (min) % (RSD) o 30 C w/ Oven Pressurized and Degassed Room Temperature o C Pressurized and Degassed 2.5% 0.27% o 30 C w/ Oven Pressurized and Degassed % 0.30% Hours Injection Number
6 Common Mobile Phase Modifiers Buffers To stabilize phs of mobile phase under reverse phase (RPC) or ionexchange (IEC) chromatography (phosphate, acetate, citrate); ~5-20 mm Triethylamine (TEA) To reduce tailing of basic analytes RPC; ~0.1% Phosphoric, acetic acid or formic acids Ion suppressors To suppress ionization of acidic analytes under RPC; ~0.1% Ion-pairing reagents For separation of ionic compounds using RPC column (C18) Anions: Quaternary ammonium salts (e.g., Tetrabutylammonium hydroxide); ~2-20 mm Cations: Alkyl sulfonates (e.g., Hexane sulfonate); ~2-20 mm
7 Series 200 Analytical Pump: Low Pressure Mixing Solvent toward column Solvents Solvent into metering chamber 100 µl 107 µl Solvent mixing delay volume: ~ uL, depending on pressure Advantages: Great solvent mixing and built-in solvent compressibility factor
8 Low-Pressure Mixing Pump Blending Performance
9 Series 200 Micro-Pump: Binary High-Pressure Mixing To Injector Purge Valve Purge Valve Tee Pump A Pump B Solvent mixing delay volume: ~10-400uL with static mixer, depending on flow rate Advantage: Very low solvent mixing delay; good for high speed gradients and LC/MS
10 Method Transferability Concerns Low-Pressure Mixing: 1.0mL Delay Volume = High-Pressure Mixing: 0.1mL Delay Volume = 2 ml/min
11 Using Injection Delay Time (for Method Transferability) Low-Pressure Mixing: 1.0mL Delay Volume = High-Pressure Mixing: 0.1mL Delay Volume = Low-Pressure mixing with 0.45 min Injection Delay Time 2 ml/min
12 HPLC: Preferred Materials for Wettable Parts STAINLESS STEEL Industry Standard SS 316; May rarely effect biological activity or allow adverse chealation to occur TITANIUM (alternative to SS) PEEK (Poly-Ether Ether-Ether-Ketone) TEFLON / VESPEL / TEFZEL POLYAMIDE / POLYIMIDE QUARTZ-SILICA GLASS Good biocompatibility. Somewhat more brittle than SS Good biocompatibility and often used for basic compounds. Becomes brittle w/ >20% THF or MeCl 2 Used for rotor seals and low- pressure tubing. Note: : Use PEEK for basic compounds Do not use with MeCl 2 Commonly used for flow cells RUBY ` Used for balls in pump check valves SAPPHIRE Used for pump pistons
13 Pressure Problems PRESSURE TOO HIGH Dirty column Blockages e.g., system filter, guard column or column inlet frit, tubing, injector needle or valve, detector flow cell. >> Check by isolating pressure point(s); like a plumber Check flow rate Different solvent mixture? Are solvent lines switched?
14 Pressure Problems PRESSURE TOO LOW Leaks Pump malfunction e.g., piston seal, connections, rotor seal e.g., vapor lock, sticking check valves or solenoid valve, broken piston Plugged solvent filter Check flow rate Different solvent mixture? Lines switched?
15 Autosampler: Basic Injection Cycle Based on Rheodyne injector valve Poor peak shape may result if either the connections to injector ports 2 and 3 are reversed or one fills sample loop volume over 50% (in partial-loop mode). Sample from Syringe
16 Precision Problems Possibly Related to Autosampler Overfilling the sample vial Recommended: No more than 2/3 full Tightly sealed vials not vented Turn venting option on Excessive sampling speed Especially for viscous and low volume samples (<5µL) L): : Reduce speed to slow Bubbles in the autosampler sampling lines or syringe Degas flush solvent Worn rotor seal in the injection valve Results in poor RSD: Replace seal
17 Detectors for Any Application PerkinElmer Series 200 Detectors UV/Vis PDA UV/VIS considerations: Can detect predominant compounds being analyzed by HPLC Useful for any compounds that absorb between 190 and 700 nm Very good sensitivity (ppb). Typically, better than RI but not as good as Fluor. Exceptional baseline stability Wavelength programmable during run PDA considerations: Approaches UV/VIS sensitivity (ppb) Dual lamps (Deuterium and Tungsten) for full UV/Vis wavelength range Spectral data useful for confirmation and identification of components Capability of peak purity assessment Network-Direct connection for direct transfer of data and spectra to data directories on the network
18 Detectors for Any Application PerkinElmer Series 200 Detectors FL RI Fluorescence considerations: Very selective (must fluoresce) Exceptional sensitivity (ppt) May have to derivatize if compounds do not fluoresce in native state. Refractive Index considerations: Universal detection Good sensitivity (ppm) Sensitive to temperature changes (RI must have built-in Peltier temp. control; 35 C usually good for most labs) Can not be used for gradient method
19 Detectors: What to Consider Lamp Energy: Lamp Hours: Wavelength Accuracy: Reference Wavelength (PDA): Flow Cell Condition: RI Reference Purge Valve: Column Connections: Check periodically Lamp lifetime: typically ~1500 hours for UV/Vis, ~1000 for PDA and 500 for FL detectors Check periodically Choose a wavelength transparent to both sample and solvent On PDA, remove flow cell periodically; hold it toward light and check clarity Shine light into UV/Vis flow cell Check position Check that column is connected to Inlet tubing on detector
20 Detector Pointers Baseline problems related to detectors may be: Mechanical Leaky flow cell, trapped bubble Optical Aging lamp, dirty flow cell Bad mobile phase Not detector Electrical Call vendor Don't wash columns while they are connected to the detector. When possible, the use of a 50psi back pressure device is recommended after the detector (suppresses outgassing)
21 Suggested Periodic Performance Testing LC PUMPS Flow, composition calibration and pressure Leak test AUTOSAMPLER Precision, carryover, linearity testing DETECTORS Wavelength check Noise and drift testing Linearity testing
22 Suggested Spare Parts and Replacement Period LC PUMPS High/low pressure piston seals and backup rings kit (~ once a year); Part# N Might want to replace high pressure piston every other year AUTOSAMPLER Injector rotor seal (~ once a year) Injector needle (if properly maintained, should hardly ever need replacing) Might want to replace flush and/or sampling syringe periodically (inspect first) DETECTORS Lamp (order new one only after 1000 hours on current UV/Vis or 500 hours on PDA - they have a shelf life)
23 Chromatographic Diagnostics NO PEAKS: Lamp off/bad, check wavelength Injector problem Check solvent selection / strength No sample in vial CARRYOVER: Materials incompatibility eg., Teflon vs PEEK Insufficient injector flushes or flush speed Column overloaded Some columns exhibit component carryover For that application, switch to another column vendor
24 Chromatographic Diagnostics Normal noise Expanded; always a good practice before starting a run
25 Chromatographic Diagnostics 5 min Air bubbles in solvent Degas solvent with in-line degasser or 5min helium sparge then pressurize Aging/bad detector lamp Replace lamp (typically, order new lamp after 500 hours on current one) Damaged pulse dampener on pump Replace
26 Chromatographic Diagnostics 5 min Contamination in system (solvent, degasser, pump, AS, column, detector) Remove column(s)/guard column and flush any buffers out of system; Rinse all solvent lines with HPLC water (may require co-solvent flush); Place all solvent lines into common bottle of 10% Ammonium Hydroxide; Flush for 20 min at 3mL/min. then switch to HPLC water for 2 minutes; Replace solvent bottle with 10% Nitric Acid and do another 20 min flush; Replace solvent bottle with HPLC water and do another 10 min flush; May also want to perform 20 min./1ml/min Isopropanol flush (usually not necessary); otherwise, move back to original solvents with clean bottles
27 Chromatographic Diagnostics Noise from detector, surroundings or poor analog cable grounding Monitor lines and recommend adding line conditioner Aging/bad detector lamp Replace lamp Air bubbles in solvent or dirty flow cell Degas solvent / rinse flow cell with strong co-solvent; The latter may require base/acid rinse (previous slide) May arise from fines from column > flush thoroughly without detector or change vendor Especially observed on RI and electrochemical detectors
28 Chromatographic Diagnostics A B Void volume; solvent front Unretained peak volume/time Elution volume k = 1 (capacity factor) Refractive index effect (typically from flow cell) Replace flow cell or retire detector
29 Chromatographic Diagnostics Column voids/channeling Bad column: Discard
30 Chromatographic Diagnostics Sample overload Inject less sample, dilute sample or change diluent to match mobile phase Forgot to add anti-tailing modifier to mobile phase Add Triethylamine (TEA) or similar Deteriorating column Change column Components may be too basic or acidic If using ion exchange column, switch to C18 and ion pairing mode
31 Example of Fronting Fronting Peak Caused by sample overload or component miscibility problem Inject less sample, dilute sample or change diluent to match mobile phase Check sample miscibility
32 Chromatographic Diagnostics All could be caused by sample overload Inject less sample, less concentrated sample or change diluent to match mobile phase Change to another column with higher sample capacity
33 Chromatographic Diagnostics Proportionally increasing retention times Low flow rate If same peak height, solvent leak before injector Column temperature may be lower; use column oven Possible mobile phase change (modifier, ph, molarity); however, not always proportional Possible column deterioration; however, usually not proportional
34 Chromatographic Diagnostics Disproportional retention time change Mobile phase change (modifier, ph, molarity); Column deterioration; replace column Change in column temperature; use column oven
35 Chromatographic Diagnostics Proportional drop in peak height Air in autosampler flush line Check lamp energy: lamp energy may be getting significantly lower. Time to change lamp Check wavelength (not always proportional) Sample absorbing to column or other material; change all teflon contact points in autosampler to PEEK material (rotor seal, etc.) or change to other vendor s column Autosampler: leaky rotor seal or plugged needle Volatile components: use capped vials (not always proportional)
36 Thank you for attending!
37 Chromatographic Diagnostics Component interconversion or isomerization Check ph, especially relative to pka(s) of components Check Temperature
Acclaim RSLC. (Rapid Separation Liquid Chromatography)
Acclaim RSLC (Rapid Separation Liquid Chromatography) Product Manual for Acclaim RSLC Columns Page 1 of 9 Product Manual for Acclaim RSLC Columns (Rapid Separation Liquid Chromatography) Analytical Columns
More informationHPLC systems. Overview
HPLC systems Overview An hplc system is just a way to get a sample onto the end of a column, to pump solvents through the column until the analytes drop out the end, and to detect them as they appear.
More informationANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION
ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION Standard OQ Test Suite This document describes the test program for qualifying LC analytical-scale Agilent systems; the following table lists
More informationTroubleshooting. Select the Table of Contents here Rheodyne All rights reserved
nmlkji Literature nmlkj Site Troubleshooting This section contains our complete Troubleshooting Guide. It presents detailed discussions of the most common problems encountered when using sample injection
More informationAKTA pure 25 New Owner s Intro
AKTA pure 25 New Owner s Intro The exercise below will give a quick demonstration of how easy and intuitive the AKTA pure 25 will be for you in demonstrating downstream processing to your students. Steps
More informationGhost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager
Ghost Peaks: How to Fix a Haunting Problem Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager Outline of Topics What is a ghost peak? How to identify ghost peaks and their source How to correct
More informationHPLC troubleshooting guide
CASE STUDY 70303 HPLC troubleshooting guide A free app that will save time and frustration Introduction During the development of a new HPLC method, there were unexpected results all peaks, except for
More informationno peaks missing peaks too small peaks increasing retention decreasing retention
GC Troubleshooting We can help you. 1. Observation: no peaks no peaks missing peaks too small peaks increasing retention decreasing retention declining baseline rising baseline bleeding plateaus interfering
More informationHPLC. Solvent Delivery Module Systems
HPLC Solvent Delivery Module Systems Who is GBC Scientific Equipment? GBC Scientific Equipment was founded in 1978 and is now the world's third largest and fastest growing manufacturer of atomic absorption
More informationThe instrument should have a mass range from 2 to 2000 m/z or better Interface
Revised Specifications for Liquid Chromatography Mass Spectrometry (LC-MS/MS ) Triple Quadrupole Specification for Ultra-Fast Liquid Chromatography system Solvent Delivery 1) A quaternary Pump system for
More informationDOS in Food Science & Nutrition, Manasagangotri, Mysore
Sl No UNIVERSITY OF MYSORE DOS in Food Manasagangotri, Mysore 570 006. TENDERS FOR THE SUPPLY OF EQUIPMENTS, NETWORKING & COMPUTATIONAL FACILITIES FOR UNIVERSITY OF MYSORE (UNDER UGC XI SCHEME) INVITATION
More informationGCMSD-Headspace Analysis SOP
Before you start GCMSD-Headspace Analysis SOP Method and Sequences names are restricted to 50 characters (MSD program will crash otherwise) The GC oven and Injection Ports need to be cooled to 50 oc to
More informationEvaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC
Evaluation of linear and step gradient performance and retention time precision of the Agilent 12 Series Rapid Resolution LC This document is believed to be accurate and up-to-date. However, Agilent Technologies,
More informationAutomated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:
Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: The RSK-175 standard operating procedure was developed in order to determine the amount of dissolved gas in water. Due to the expansion
More informationGases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.
Gases&Technology FEATURE Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. B Y M A T T H E W M O N A G L E Abstract Bulk gases are often delivered to
More informationOMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT
OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (16) 17 R QUALIFICATION OF EQUIPMENT ANNEX 2: QUALIFICATION OF GC EQUIPMENT Full document title and reference Document type Legislative
More informationApplication Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction
Application Note Rapid performance verification of AZURA systems with refractive index detector Method Keywords ID HPLC Quality control, system verification, refractive index, AZURA Analytical HPLC Plus
More informationBETTER SAMPLE CARE. Spark Holland introduces. SPH1299 UHPLC Gradient Pump
BETTER SAMPLE CARE Spark Holland introduces SPH1299 UHPLC Gradient Pump SPH1299 Spark Holland introduces SPH1299 UHPLC Gradient Pump Robust, Precise and Easy to use Innovative pumping technology created
More informationInternational Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON'
DISSOLUTION ASSAY Release & Stability Studies Ed. 04. General Index: 1. PRODUCT SPECIFICATIONS (USP MONOGRAPH) 2. IDENTIFICATION BY HPLC 3. ASSAY - HPLC SETUP 4. STANDARD PREPARATION 5. SYSTEM SUITABILITY
More informationextraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.
Informal Commentary Limit of Diethylene Glycol (DEG) and Ethylene Glycol (EG) in Sorbitol Solution, Sorbitol Sorbitan Solution and Noncrystallizing Sorbitol Solution December 2009 Monograph/Section(s):
More informationSimplified Backflush Using Agilent 6890 GC Post Run Command Application Note
Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note Gas Chromatography Author Matthew S. Klee Agilent Technologies 2850 Centerville Road Wilmington, DE 19808-1610 USA Abstract
More informationLaboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.
Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography
More informationSetting a New Standard in Flash Chromatography Performance
Setting a New Standard in Flash Chromatography Performance What s New Faster Flow rates, up to 300 ml/min Higher pressures, up to 300 psi (20 bar) Bigger touchscreen, 12 or 15 inches Wider Dynamic Range
More informationAutomated Determination of Dissolved Gases in Water Anne Jurek
Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: Tapping the natural gas reservoirs throughout the United States has long been a viable solution for energy independence; however
More informationTitle: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)
Procedure No: SOP-026 Revision No: 1.0 January 24, 2011 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to meet
More informationIndian Institute of Technology Kanpur Environmental Engineering and Management Department of Civil Engineering
Indian Institute of Technology Kanpur Environmental Engineering and Management Department of Civil Engineering Professor Vinod Tare Phones:+91-0512-2597791,7792 Email: vinod@iitk.ac.in Post Office: IIT
More informationFigure 1: Plug in ion inlet of mass spec
Starting procedure: by Vicky - is the vacuum pump running (should run all time, otherwise it ll need 4-5 hours to have a adequate vacuum for the MS) otherwise something has gone wrong- F.T - Remove plug
More informationAgilent 1260 Infinity Purification Solution
Agilent 1260 Infinity Purification Solution System User Guide Agilent Technologies Notices Agilent Technologies, Inc. 2014-2015, 2016 No part of this manual may be reproduced in any form or by any means
More informationValves FLUIDICS VALVE OVERVIEW & FUNCTIONS ROTARY SHEAR VALVES STAND ALONE VALVE PRODUCTS MANUAL VALVES SPARE PARTS VALVE ACCESSORIES
114 Valves FLUIDICS Our valves are an integral part of advanced fluid-handling solutions for a wide range of analytical instrumentation and clinical diagnostic systems. Our valve options include manual
More informationFor that reason VICI Jour developed:
VICI Jour Safety VICI Safety Products VICI Safety Products NEW VICI strives to provide customers with the best solutions for their problems to make the work better and/or easier, with respect to environment
More informationUP-F Delrin TM Winged Nut. UP-F PEEK TM Nut. with F-142 PEEK TM Ferrule
One Piece or Two? A one-piece fitting is more convenient and less cumbersome, since the ferrule cannot stick in a receiving port and the fitting is more easily found if dropped. With two-piece Fingertight,
More informationDEBUBBLERS & DEGASSERS ACTIVE DEBUBBLERS APPLICATION NOTE WHY DEGAS YOUR MOBILE PHASE? PAGE 177 STAND ALONE DEGASSERS PAGE 175 PAGE 178
174 DEBUBBLERS & DEGASSERS ACTIVE DEBUBBLERS PAGE 175 APPLICATION NOTE WHY DEGAS YOUR MOBILE PHASE? PAGE 177 STAND ALONE DEGASSERS PAGE 178 SEE STARRED PRODUCTS FOR TOP SELLERS! Series DEBUBBLERS & DEGASSERS
More informationACQUITY UPLC Sample Manager - Fixed Loop Operator s Overview and Maintenance Information
ACQUITY UPLC Sample Manager - Fixed Loop Operator s Overview and Maintenance Information Revision A Copyright Waters Corporation 2010 2011 All rights reserved Copyright notice 2010 2011 WATERS CORPORATION.
More informationPrep LC Training Manual S.V.
1234 Hach Hall 515-294-5805 www.cif.iastate.edu Prep LC Training Manual 11-6-2013 S.V. Location: Contact: 1238 Hach Hall Kamel Harrata, 1236 Hach Hall and Steve Veysey, 1234 Hach Hall Safety: All researchers
More informationConsiderations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port
Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port So, Why Do I Have To Do Maintenance? 1. Things wear out septa, syringes, nuts, ferrules, o-rings, etc.
More informationLaboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.
Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions
More informationAutomated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek
Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek Abstract: Due to the expansion of natural gas drilling through horizontal fracturing, there has been
More informationSOP: Buck Scientific BLC-20P HPLC Operation
Page 1 of 11 Approvals Preparer: John Buford Reviewer: Tim Kull Reviewer: Dr. Margaret Bryans Date: 16OCT13 Date: 30OCT13 Date: 31OCT13 1. Purpose 1.1. Basic operation of the Buck Scientific BLC-20P isocratic
More informationAutosampler problems!
Laserchrom HPLC Laboratories Ltd Technical Bulletin Autosampler problems! Air in the syringe Bent or blocked needle Valve problems Unreproducible injections Sample carryover Wash solution problems Vial,
More informationRunning HPLC Methods on an ACQUITY UPLC System
Running HPLC Methods on an ACQUITY UPLC System Note: This document is an addendum to the ACQUITY UPLC System Operator s Guide. Contents Topic Page System considerations 1-1 Choosing fittings 1-3 System
More informationGradients in ion chromatography. Optimal separation with shorter analysis time
Gradients in ion chromatography Optimal separation with shorter analysis time Gradients yield precise results faster 02 The goal of ion chromatography in a nutshell reads: optimum separation within an
More informationThe instrument will process the samples one by one till all the samples have been cleaned up.
Table of Contents SPE-01 Cleanup Station... 1 LC-04SP Valve System... 5 LC-04SP Standalone Valve... 10 LC-05 Auto Injector... 12 LC-06 Dosing Pump... 16 LC-08 Column Selector... 17 SPE-01 Cleanup Station
More informationUWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column
Before you connect your trap and column You need to provide your own trap and column (Note our fittings accommodate trap and column made with fused silica 360 µm OD). Check our guide for packing traps
More informationGenerating Calibration Gas Standards
Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and
More informationOPERATORS MANUAL G U A R A N T E E D T R Y C H E M APPROVED
OPERATORS MANUAL G U A R A N T E E D C H E M I S T R Y APPROVED Operator Manual Post-Column Derivatization Instrument 1280 Space Park Way Mountain View CA 94043 USA 800-654-3330 650-694-6700 G U A R A
More informationDemystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.
Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Jamie Brown R&D Scientist Supelco 595 North Harrison RD. Bellefonte, PA 16823 National
More informationDionex IonPac AS19-4µm
Total Thermo Scientific Dionex IonPac AS19-4µm Column Product Manual P/N: March 2015 Part of Thermo Fisher Scientific Product Manual for Dionex IonPac AS19-4µm Column Dionex IonPac AG19-4µm Capillary Guard
More informationSTANDARD OPERATING PROCEDURES
PAGE: 1 of 9 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 3.1 Canister Receipt 3.2 Canister Storage 4.0 INTERFERENCE AND POTENTIAL PROBLEMS
More informationFROG-4000 TM Chemical Analysis System
FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies, Inc. 2015 Vol.5 This page is intentionally left blank. FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies,
More informationLC, RSLC, Bio RS and Dual Gradient Systems
Thermo Scientific Dionex UltiMate 3000 LC, RSLC, Bio RS and Dual Gradient Systems P/N 4823.0205 Revision B December 2014 Automated Method Scouting Considerations All information in this manual is subject
More informationExtended Application Note
Extended Application Note Harmful Substances in Dietary Supplements HPLC Columns with APP A-337 www.mtc-usa.com 1-732-578-1777 INTRODUCTION A dietary supplement is a substance to be consumed for the purpose
More informationPrep LC Training Manual
Prep LC Training Manual (11-6-2013) Location: 1238 Hach Hall Contacts: kamel Harrata and Steve Veysey, 1234 Hach Hall Safety: Introduction: The Agilent PrepStar SD-1 preparative HPLC is a student operated
More informationCHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation
CHE 4115 Chemical Processes Laboratory 2 Experiment 1 Batch Distillation BACKGROUND Distillation is one of the most commonly used unit operations in chemical engineering. In general, a distillation operation
More informationBest Practice for Identifying Leaks in GC and GC/MS Systems
Best Practice for Identifying Leaks in GC and GC/MS Systems Technical Overview Maintaining a leak-free GC or GC/MS system is critical for obtaining optimal system performance with reliable, reproducible,
More informationMultiple Gas#5 GC configuration Jan 2016
History: Unfortunately there is no single column that can separate: Hydrogen Oxygen Nitrogen Methane CO CO2 Ethane Water Propane Butane Pentane Over the years SRI Instruments has devised several solutions
More informationEnvironmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!
After years of development and testing, DPS Instruments is pleased to present the newest, most expandable and versatile Gas Chromatography Systems in history. The DPS 600 Series GC systems are the world
More informationUser s Booklet for the Wyatt minidawn Light Scattering Instrumentation
User s Booklet for the Wyatt minidawn Light Scattering Instrumentation The Wyatt minidawn Light Scattering instrument is able to measure the weight average molecular weight of a synthetic polymer or a
More informationSimultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS
Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS UCT Part Numbers: CUQAX22Z Clean-Up C8+QAX, 2mg/1mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQID21-3UM
More informationDetector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane
Gas requirements Gases for packed columns The carrier gas you use depends upon the type of detector and the performance requirements. Table 520-1 lists gas recommendations for packed column use. In general,
More informationVideo Notes. LC Troubleshooting Series Retention Time Shifts
Video Notes LC Troubleshooting Series Retention Time Shifts Introduction John Henderson is an Applications Scientist for Agilent Chemistries and Supplies Division. Maureen Joseph is the HPLC Columns Product
More informationFiber Optic SMA Z-Flow Cell Manual Version 2.1
Fiber Optic SMA Z-Flow Cell Manual Version 2.1 The line of fiber optic SMA Z-Cells is the top choice for many absorbance measurements. SMA-Z-Cell path lengths are available between 2.5mm - 100mm and can
More informationRev. CO Packed Columns for Normal Phase Chromatography. TSKgel NH m INSTRUCTION MANUAL
Rev. CO067000 Packed Columns for Normal Phase Chromatography TSKgel NH2-100 3 m INSTRUCTION MANUAL Safety Precautions To help protect you and/or your property from potential damage, please read this manual
More informationBlue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION
Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION Install your Blue River Technologies Port-A-Poly mixing system in a clean dry area.
More informationArtisan Technology Group is your source for quality new and certified-used/pre-owned equipment
Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment FAST SHIPPING AND DELIVERY TENS OF THOUSANDS OF IN-STOCK ITEMS EQUIPMENT DEMOS HUNDREDS OF MANUFACTURERS SUPPORTED
More informationTrace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS
Abstract Highly sensitive analysis of metanephrines in plasma is critical in the diagnosis and treatment of pheochromocytoma and paraganglioma. Here, a HILIC LC-MS/MS method was developed using a Raptor
More informationPeak Biotech - Company & Product Presentation
Peak Biotech - Company & Product Presentation July 2005 Peak Biotech Company & Product Presentation Peak Biotech Company Profile Location Jacob Nøhr Peak Biotech A/S was founded Kvistgaard, Denmark QA
More informationRapid and Reliable Detection of Dissolved Gases in Water
Rapid and Reliable Detection of Dissolved Gases in Water Andrea Caruso and Massimo Santoro Thermo Fisher Scientific, Milan, Italy Application Note 005 Key Words Chromeleon CDS, Environmental, Fracking,
More informationSupporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial
Supporting Information for Micro-Collection of Gases in a Capillary Tube: Preservation of Spatial and Temporal Resolution AUTHOR INFORMATION Kristin A. Herrmann Favela, 1 * Pieter Tans, 2 Thomas Jaeckle,
More informationManual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany
Manual Degasser DDG-75 April 2007 1 Manual Online Degasser DDG-75 DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE-68766 Hockenheim Germany Tel. +49-6205-9450-0 Fax. +49-6205-9450-33 email info@duratec.com
More informationArtisan Technology Group is your source for quality new and certified-used/pre-owned equipment
Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment FAST SHIPPING AND DELIVERY TENS OF THOUSANDS OF IN-STOCK ITEMS EQUIPMENT DEMOS HUNDREDS OF MANUFACTURERS SUPPORTED
More informationVarian 3300 Gas Chromatograph Abbreviated Operating Instructions
Varian 3300 Gas Chromatograph Abbreviated Operating Instructions 1. Introduction and Startup 2 2. Basic Configurations for the GC 3 3. Sample Preparation 3 4. Building a Method 3 5. Split/Splitless Injection
More informationNanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI
NanoSight NS300 Operation instructions Injection/flushing brief overview: 1. Do not exceed flow of 1 ml per 20 seconds. 2. Inject two 1 ml syringes with nano-pure or DI water. 3. If the water does not
More informationTROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.
458 LCGC NORTH AMERICA VOLUME 24 NUMBER 5 MAY 2006 www.chromatographyonline.com LC TROUBLESHOOTING Dwell Volume Revisited Gradient transfer does not have to be problematic. John W. Dolan LC Troubleshooting
More informationProudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Varian 310 310 LC/MS Manual System INCLUDES: 310-MS LC/MS triple quadrople
More informationFROG-4000 TM Chemical Analysis System
FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies, Inc. 2014 Vol.4 This page is intentionally left blank. FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies,
More informationNMR Service. How to Prepare Samples for NMR
NMR Service How to Prepare Samples for NMR In NMR, unlike other types of spectroscopy, the quality of the sample has a profound effect on the quality of the resulting spectrum. If you follow a few simple
More informationOperating Instructions Multi-Channel Vacuum Degasser
Operating Instructions Multi-Channel Vacuum Degasser Introduction The Vacuum Degasser is a high-efficiency in-line module that removes dissolved gasses from HPLC solvents. Its unique design assures reliable
More informationAnalysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS
Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS Peter Alden and Michael Jones Waters Corporation, Milford, MA, U.S. APPLICATION BENEFITS UPLC combines
More informationMETHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES
METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES 1. Applicability and Principle 1.1 Applicability. This method applies to the analysis of carbon dioxide
More informationUser guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection
User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection This user guide does not have the intention to introduce you to any of the theory and/or principles behind
More informationMV5: Automated concentration system
MV5: Automated concentration system Abstract LabTech MV5 is the latest automatic nitrogen evaporator developed by our R&D Team able to process up to 54 samples simultaneously. This advanced instrument
More informationDETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY
DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY Nengbing Xu, Hongmei Ying and Libo Zhu Ningbo Environmental Monitoring
More informationASTM 3612/TOGA/Dissolved Gas GC Revised October 2013
The SRI ASTM 3612/TOGA/ Dissolved Gas GC configuration ( TOGA GC ) permits measurement of gasses dissolved in water, oil or other liquids, and is based on the requirements of ASTM method 3612, option C,
More informationZebron Gas Management
Zebron www.phenomenex.com/gasmanagement What are the key benefits of Zebron filters and traps? The high adsorption capacity of the Zebron Filters removes contaminants from mobile phase gases, reducing
More informationHICHROM. HPLC Accessories
HICHROM HPLC Accessories HICHROM Index PEEK Tubing SLIPFREE HPLC Column Connectors 1 Stainless Steel Tubing 7 PEEK Unions, Tees and Crosses 8 2 Micro Flow Fittings Tubing Cutters 9 3 Fingertight Fittings
More informationBuck Scientific, Inc.
Buck Scientific, Inc. 910/310 Gas Chromatographs The Buck 910/310 Gas Chromatographs In the lab, the 910-series GCs perform routine industrial analyses and complicated research separations easily, generating
More informationTitle: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator
Procedure No: SOP-034 Revision No: 1.0 Revised Dec. 29, 2010 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE This procedure is intended to describe the operations of the Dasibi model 5008 calibrator. The Dasibi
More informationCapillary Conversion System CCS-4/NC & CCS-05/NC. Installation and Operating Instructions. Operating Instructions
Operating Instructions Check the detector for the correct operating conditions. If make-up gas is being used a flow of 30ml/min. of carrier plus make-up is normally recommended. This can most effectively
More informationAdvances in Capillary Column Backflushing
Advances in Capillary Column Backflushing Matthew S. Klee, Ph.D. ISCCE 2010 1 Capillary Flow Technology Traits Ease of use - ruggedness Inertness Usability to > 350 oc Excellent exclusion of air No unpurged
More informationFRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS
FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS Analysis Conducted BY CC RESEARCH (P) Ltd., For Nanoceutical Solutions Inc. USA 2950 Thousand Oaks drive, Suite 3, San Antonio, Texas EXECUTIVE SUMMARY
More informationColumbus Instruments
0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011
More informationOrganic Elemental Analysis Series II CHNS/O Elemental Analyzer. proven performance. reliable results
Organic Elemental Analysis 2400 Series II CHNS/O Elemental Analyzer proven performance reliable results 2400 Series II CHNS/O Elemental Analyzer raising the standard Q U I C K G L A N C E One analyzer
More informationSystem Overview. TCD Detector temperature setpoint Regulator
System Overview POPULAR CONFIGURATION GCs Your educational TCD GC is configured on the compact 310 chassis. It is equipped with a TCD Detector, a temperature programmable Column Oven, a 3 Silica Gel packed
More informationSvensk läkemedelsstandard
Svensk läkemedelsstandard 2019.0 Carbon dioxide Postadress/Postal address: P.O. Box 26, SE-751 03 Uppsala, SWEDEN Besöksadress/Visiting address: Dag Hammarskjölds väg 42, Uppsala Telefon/Phone: +46 (0)18
More informationPreparative Valve-Based Fraction Collector
Agilent 1260 Infinity II Preparative Valve-Based Fraction Collector User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2016-2018 No part of this manual may be reproduced in any form or
More informationStandard Operational Procedure for Capillary Electrophoresis
Page 1 of 14 Standard Operational Procedure for Capillary Electrophoresis *Please ask an experienced operator for training before you use the CE if you can. A demo is most probably more efficient than
More informationOxygen Contamination
MPMS Application Note 1014-210 Oxygen Contamination This application note describes potential sources for oxygen contamination in the sample chamber and discusses its possible effects. Molecular oxygen,
More informationPOLYMER TEST KIT (CLAPPER METHOD) OFI PART No
OFI Testing Equipment Polymer Test Kit (Clapper Method) Part # 295-00 Instructions Page 1 of 4 POLYMER TEST KIT (CLAPPER METHOD) OFI PART No. 295-00 This test determines the Partially Hydrolyzed Polyacrylamide
More informationINSTALLATION AND OPERATING INSTRUCTIONS
INSTALLATION AND OPERATING INSTRUCTIONS For 3D Instruments Direct Drive Pressure Gauges 3D Instruments Inc., 15542 Chemical Lane, Huntington Beach, CA, 92649, U.S.A. Tel. (714) 894-5351, Fax. (714) 895-4309
More informationAgilent 1290 Infinity Pump Head Maintenance
Agilent 1290 Infinity Pump Head Maintenance Technical Note Agilent Technologies Notices Agilent Technologies, Inc. 2012-2015, 2016 No part of this manual may be reproduced in any form or by any means (including
More information