Autosampler problems!

Size: px
Start display at page:

Download "Autosampler problems!"

Transcription

1 Laserchrom HPLC Laboratories Ltd Technical Bulletin Autosampler problems! Air in the syringe Bent or blocked needle Valve problems Unreproducible injections Sample carryover Wash solution problems Vial, cap and septum problems Run time errors X-Y calibration Units B16-18, Laser Quay, Medway City Estate, Rochester, Kent. ME2 4HU

2 Introduction An autosampler has more moving parts than the rest of the system put together. Hence it is hardly surprising that it is the source of a number of problems. They tend to fall into three categories; those resulting from lack of care or maintenance, those resulting from human error, and normal wear and tear. Here s a list of the common problems and some suggestions for how to avoid them, and how to fix them once they occur! Air in the syringe The syringe is used to draw sample from the vials, and to pump the wash solution. So we already know where the air must have come from! If the sample level is too low for the amount required for the injection(s) that have been programmed, air will be drawn in once the level goes below the tip of the needle. This can also occur if the needle depth is set too high. Air will also be drawn in if the wash solution runs out, or if it is not degassed prior to use. Run a few wash cycles, and the air should be eliminated automatically into the autosampler waste bottle. Bent Needle A needle has to take considerable force along its axis, as it is forced into through the septum into a vial. Hence in time it becomes weakened, and will eventually bend under the pressure. A clue to this beginning to happen is that the hole in the vial caps starts to move away from the centre. As soon as this is observed, the needle should be checked for straightness, and if necessary, replaced. Needles for autosamplers are inordinately expensive, so the cost must be weighed against the cost of an overnight run being wasted if the needle lets go at 2am! Try to resist the temptation to straighten a bent needle. It sometimes works, but like a car which has been straightened after a crash, the next time it will bend really easily. The other reason for finding a bent needle is that the X-Y calibration is out. As a consequence, the vial positions are not quite where the autosampler thinks they are, and the needle hits the cap instead of the septum. X-Y calibration does go out slowly with time, and is one of the things that should be checked when an autosampler is serviced. If the holes in the vial caps are drifting away from the centre but the needle is straight, this is almost certainly the problem. This is a job only for a service engineer or a very competent user. Replacing a needle is relatively straightforward, although sometimes it is necessary to replace some tubing with the needle (eg Agilent 1050 etc) If there is a needle guide, ensure that this is straight when the new needle is fitted. Autosamplers tend to use reasonable force with the needle, so if the needle is bent, the needle guide could be out of position too. Reconnect the tubing to the needle, and bear in mind that sometimes this forms part of the flow path after injection, so the connection to the needle may need to be tight.

3 Blocked Needle A needle has a fairly fine bore, and can easily be blocked. The blockage normally comes from a piece of rubber septum which has broken off as the needle passes through. This is known as coring. There are two things which can be done to prevent or minimise this. The first is to use a PTFE septum. Essentially this is just a plastic disc, which splits as the needle passes through. The only drawback is that if the run time is long, or the sample solvent is volatile, the septum does not re-seal, and so evaporation can occur, giving a higher concentration for the duplicate injection. An alternative is to use a rubber septum which has a layer of PTFE on one or both sides. There are two other causes of a blocked needle. One is particles in the sample. This is clearly not an acceptable situation for lots of reasons, but assuming that the sample was correctly made up with no undissolved material, particles can arise because of precipitation when the sample is chilled in the autosampler tray, or because the sample is not sufficiently soluble in the eluent. In this latter case, precipitation occurs at the moment of injection, and subsequent samples are then not drawn up because of the blockage. The second cause is particles in the wash solution. The wash solution should ideally be the same as the eluent at the point of injection, but without the buffer. This ensures that it is totally miscible, does not disturb the dynamic equilibrium in an isocratic system, and has no content which can dry in the needle causing a blockage. Valve problems The injection valve is never made by the autosampler manufacturer, but sourced from a valve specialist such as Rheodyne or Valco. The valve manufacturer supplies the valve, its actuator (the motor which turns it), a control box, and a cable between the control box and the actuator. So if there is a problem and the valve won t turn, assuming the autosampler is sending the control box the signal, the problem must lie with one of these components, and generally it is not a user-serviceable problem. However it is worth removing the valve from the actuator and trying again. If the actuator turns, the problem lies with the valve, which is usually quite easy to remove and replace without paying 100 an hour for a service engineer to queue round the M25! What causes a valve to jam? Either leaving it unused for ages so it dries out, or not flushing it out regularly. If a sample dries to leave a sticky material, then this can easily dry in a valve overnight and cause it to stick. It tends to build up, until the motor just can t make it turn. The danger is that by then the motor itself can burn out, or pull so much current that the cable fails. If it all goes, it will be expensive (possibly 2K) to fix. The secret is to keep the valve flushed out, not to use buffer in the wash solution, and to be very careful when using an autosampler for the first time after a period of prolonged disuse. If in doubt, please call us on the number below for free advice.

4 Needle Depth too high or too low If the needle depth is set too high, the last part of the sample cannot be drawn up and the needle may draw in air. If the needle depth is set too low, it drives a hole through the bottom of the vials. It is necessary that the needle depth should be adjustable to accommodate different types of vials. So care must be taken when using a new type of vial to establish an appropriate needle depth. This is usually done using the autosampler in service mode, and driving the needle down one step at a time until the vial moves, indicating that the needle has made contact with the bottom. The depth used should be several steps higher than this to allow for variations in the vial base thickness. Wash solution a stronger eluent than the mobile phase In a fixed volume autosampler, at least three times the loop volume of sample is injected, flushing the loop and making sure it contains exactly one loop volume of sample. Fixed volume autosamplers therefore use more sample than variable volume models and are less flexible, but they give much more reproducible injections and this problem does not apply to fixed volume autosamplers. A Variable volume autosampler uses a much larger loop, and meters in a small amount of sample with the syringe. The rest of the loop is filled with wash solution. So if the wash solution is a strong eluent, the same problem occurs as does if the sample solvent is a strong eluent, only much more so, because there is much more wash solution present than sample solvent! The problem is that the peaks become very broad, because the strong eluent disperses the sample into the column before it is itself washed away. The solution is that the wash solution must be no stronger than the eluent, and preferably the eluent itself only without any buffer. If running a gradient, the wash solution should be the eluent at the point of injection. Other wash solution issues 1. The wash solution must not be allowed to run out. Its obvious, but especially if there is a wash cycle after each injection, quite a large volume of wash solvent will be required. 2. The wash solvent must be degassed. Otherwise air can be sucked up, which may be in the syringe for a few injections before finally clearing itself. Hence the injection volume will be suspect. 3. Waste bottle for the wash solution must have enough capacity for all the wash solution used.

5 Sample carryover Sample carryover occurs if a trace of the previous sample remains on or in the needle when it is inserted into the next sample vial.. Some autosamplers minimise this by leaving the needle in the flow path (making it a part of the loop) so that in the inject position, the needle is flushed with eluent throughout the run. Others have various wash cycles which can be run before or after each sample or each injection. To test for sample carryover, it is necessary to follow an injection with an injection from a vial of clean solvent. Extreme care must therefore be taken to ensure that the vial and solvent is clean! If doing a gradient run, ghost peaks may appear which are nothing to do with the sample, and these will appear if a blank gradient run is performed without making any injection at all. Carryover must also involve peaks which were present in the previous injection. If the current wash routine is proving inadequate to prevent carryover, check the procedure. You can change the wash solvent, perform two or perform more procedures between each run. Also watch out for contamination arising from sample on the outside of the needle. Some autosamplers have the option of washing the needle in a wash port, and then airdrying the outside of the needle. The Sykam model does this and we would be pleased to advise if required. Program Errors - Runtime and Vial Range errors You have to put your hand up for these! If the autosampler is sending the start signals and therefore effectively controlling the system, it must have a longer run time than you give the data system or the gradient controller etc. Otherwise it will make an injection before other devices are ready, and if they are not looking for a start signal, the sample will be injected but nothing will take any notice. Vial 1 will be processed correctly, vial 2 will be injected but ignored, vial 3 will be injected but processed as vial 2 etc. If the system is controlled totally from a PC, which looks for a ready signal from all devices before giving the green light for an injection, then this does not apply. But otherwise, allow a longer run time on the autosampler. Remember for a gradient run, the gradient time will be much longer than the data system run time, because it must allow for returning to the starting conditions and re-equilibrating. The autosampler must wait till this is complete. Injection before the system is re-equilibrated will give variable retention times, and hence dodgy peak identification and quantitation. Vial range errors arise if the samples are not placed in the vial range which is programmed into the autosampler. Its an easy mistake, but it fools the autosampler. If an autosampler reaches a blank space with no vial it will usually stop and report an error. If the cap is not fiited (I know its crazy, but people do!) sometimes the autosampler will notice this and stop anyway, reporting a vial error.

6 Unreproducible injections If we are asking an autosampler to inject the same sample several times to check reproducibility, we would expect that an automated machine which just runs programs should be able to do this fairly easily. However sometimes this is not the case. Before blaming the autosampler, do check that its not an air bubble in the pump, or integration problems. But assuming the problem lies within the autosampler, here s a few places to look: 1. Not enough sample. The needle barely dips into the sample, so air is drawn in. 2. Needle depth too high. Same as above. 3. Blocked needle. It causes a resistance to flow, and varying amounts of sample get in each time. 4. Sample too viscous, or syringe speed too fast for the sample viscosity. The syringe generates a vacuum and the sample follows more slowly. Slow down the syringe speed, or use a less viscous sample solvent. 5. Worn syringe. If the syringe cannot suck hard enough, and resealing vial septa are being used, there is a reduction in pressure in the vial after each injection, making it progressively harder for the autosampler to withdraw sample. Replace the syringe, or use PTFE septa. 6. Air in the syringe. It can only meter sample accurately in the total absence of air. See the earlier section on dealing with air in the syringe. 7. Wash routine. Sometimes it is found that the first injection gives a lower result than subsequent injections, and the reason is that for the first injection, the syringe starts wet with wash solution, but for subsequent injections it is wet with sample. If this is a problem, the solution is to wash before each injection, not just before each sample. 8. Sample solvent evaporating. This occurs if the sample solvent is volatile, and PTFE septa are used (they do not reseal) allowing solvent evaporation between injections. Hence the concentration of sample is gradually increasing, so the results will increase with each injection (unlike the previous case where only the first injection is low). 9. Sample not down to temperature. If working at (eg) 4 o C, the samples must be allowed to reach temperature before the run is started. Reducing temperature reduces the volume, and hence increases the apparent concentration in a 20ul injection. It can take at least 30 minutes for sample temperature to stabilise. 10. No air gap left between sample and eluent. When a sample is withdrawn from the vial, it is kept separate from the eluent by a minute (5ul) air gap. Failure to do this allows it to diffuse into the eluent that is pushing it around the autosampler, and this causes a variation in peak height from injection to injection.

7 Vial, cap and septum problems It is important to choose appropriate vials, caps and septa for your autosampler, and for your analytical method. Here s a few check points: 1. Vial Size. Most autosampler use the standard 1.8ml vials. However these come in several flavours. They can be clear or amber, with or without a label, straight sided or with a narrow tip to enable the last drop to be removed, with screw top, snap on top or crimp top etc. Other size vials also offer similar options. The vial size must be big enough to contain enough sample for the injection volume x the number of injections and a little to spare, but also be small enough to allow the needle to withdraw the required amount. If the vial is too big, evaporation causes changes in concentration, and there is a real risk that the needle will suck air. Amber vials are for lightsensitive samples. The label makes sample identification easier, and the top style is to suit you. Crimping is quicker if you have a lot of vials, but the tool costs well over 100! 2. Septa. These can be made of PTFE, rubber or silicone, and the latter two options can come with a layer of PTFE on one or both sides. PTFE is cheaper, but does not reseal after piercing. Hence evaporation can occur between duplicate injections if a volatile sample solvent is used. Be careful to use only one PTFE disc! They are punched out several hundred layers thick and tend to stick together. If you use two at a time, the resistance might be enough to bend a tired needle. Silicone is fine for aqueous samples, but rubber is most commonly used. It reseals, but can leave a slug of rubber in the needle (known as coring). If this happens it can sometimes be cleared, but spoils the run. To prevent this, use a septum with a layer of PTFE on top. Rubber is also soluble in some organic solvents. In this situation it is recommended to use a layer of PTFE on the bottom surface of the septum. Both problems at once? Us a rubber septum with PTFE on both sides! 3. Caps. Caps must clearly be appropriate for the vials. Note that not all screw top vials use the same screw thread, and sometimes the screw will not stay tight. Mixing and matching vials and caps may work, but may not so try a sample before buying. 4. Quality. As with everything this century, there are ultra-cheap versions arriving every day. There are deals to be had, but be careful to test thoroughly before switching. The worst problems are vials which break as you tighten the cap, and vials which have a different thickness of base from batch to batch! The cost saving on the vials might be needed to replace the needle!

Ace Glass offers through VWR full line of

Ace Glass offers through VWR full line of Septa Catalog Septa 3 Septa Ace Glass offers through full line of caps, closures, seals and septa. This new brochure combines all the choices in an easy-to-read and understandable format. Our extensive

More information

Solvent System Walkthrough

Solvent System Walkthrough v2.0 1 Solvent System Walkthrough The Contour Glass solvent system is designed to give you air-free anhydrous solvent with minimal effort. However, the system is only as good as its users so please read

More information

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager Ghost Peaks: How to Fix a Haunting Problem Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager Outline of Topics What is a ghost peak? How to identify ghost peaks and their source How to correct

More information

AKTA pure 25 New Owner s Intro

AKTA pure 25 New Owner s Intro AKTA pure 25 New Owner s Intro The exercise below will give a quick demonstration of how easy and intuitive the AKTA pure 25 will be for you in demonstrating downstream processing to your students. Steps

More information

Video Notes. LC Troubleshooting Series Retention Time Shifts

Video Notes. LC Troubleshooting Series Retention Time Shifts Video Notes LC Troubleshooting Series Retention Time Shifts Introduction John Henderson is an Applications Scientist for Agilent Chemistries and Supplies Division. Maureen Joseph is the HPLC Columns Product

More information

Experiment 18 Properties of Gases

Experiment 18 Properties of Gases Experiment 18 Properties of Gases E18-1 E18-2 The Task In this experiment you will investigate some of the properties of gases, i.e. how gases flow, their phase changes and chemical reactivity. Skills

More information

Valve INV-907. Instructions. 1 Introduction. 2 Installation. Safety. Unpacking. Installing the valve

Valve INV-907. Instructions. 1 Introduction. 2 Installation. Safety. Unpacking. Installing the valve Instructions 1 Introduction is a motorised rotary 7 port valve. The valve is used in ÄKTA design chromatography systems. It is powered from the ÄKTAdesign P-900 series system pump, and is controlled from

More information

NMR Service. How to Prepare Samples for NMR

NMR Service. How to Prepare Samples for NMR NMR Service How to Prepare Samples for NMR In NMR, unlike other types of spectroscopy, the quality of the sample has a profound effect on the quality of the resulting spectrum. If you follow a few simple

More information

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction Application Note Rapid performance verification of AZURA systems with refractive index detector Method Keywords ID HPLC Quality control, system verification, refractive index, AZURA Analytical HPLC Plus

More information

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008 UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method

More information

The instrument will process the samples one by one till all the samples have been cleaned up.

The instrument will process the samples one by one till all the samples have been cleaned up. Table of Contents SPE-01 Cleanup Station... 1 LC-04SP Valve System... 5 LC-04SP Standalone Valve... 10 LC-05 Auto Injector... 12 LC-06 Dosing Pump... 16 LC-08 Column Selector... 17 SPE-01 Cleanup Station

More information

no peaks missing peaks too small peaks increasing retention decreasing retention

no peaks missing peaks too small peaks increasing retention decreasing retention GC Troubleshooting We can help you. 1. Observation: no peaks no peaks missing peaks too small peaks increasing retention decreasing retention declining baseline rising baseline bleeding plateaus interfering

More information

Troubleshooting. Select the Table of Contents here Rheodyne All rights reserved

Troubleshooting. Select the Table of Contents here Rheodyne All rights reserved nmlkji Literature nmlkj Site Troubleshooting This section contains our complete Troubleshooting Guide. It presents detailed discussions of the most common problems encountered when using sample injection

More information

Thermo Scientific Chromatography Well Plates

Thermo Scientific Chromatography Well Plates Thermo Scientific Chromatography Well Plates High-throughput sample handling solutions Well Plates for Chromatography A full selection of high quality sample well plates for your UHPLC, LC and GC applications.

More information

Thermo Scientific Chromatography Well Plates

Thermo Scientific Chromatography Well Plates Thermo Scientific Chromatography Well Plates High-throughput sample handling solutions Well Plates for Chromatography A full selection of high quality sample well plates for your UHPLC, LC and GC applications.

More information

Automated Liquid Handling Station

Automated Liquid Handling Station Automated Liquid Handling Station Software User s Guide Manual Part Number 32-0442-048 Rev 0 2018 Teledyne Technologies Incorporated. All rights reserved. Printed in USA. Installation See the SimPrep Quick

More information

6900 Maintenance Instruction System Flush

6900 Maintenance Instruction System Flush Equipment Required FA74005 Damper Drain Tube FA16005 Cover Removal Tool FA900005 Beaker 0.25 Litre FA900003 Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves Latex

More information

CORESTA RECOMMENDED METHOD Nº 67

CORESTA RECOMMENDED METHOD Nº 67 CORESTA RECOMMENDED METHOD Nº 67 DETERMINATION OF WATER IN THE MAINSTREAM SMOKE OF CIGARS BY GAS CHROMATOGRAPHIC ANALYSIS (November 2005) 1. FIELD OF APPLICATION The method is applicable to the particulate

More information

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013 The SRI ASTM 3612/TOGA/ Dissolved Gas GC configuration ( TOGA GC ) permits measurement of gasses dissolved in water, oil or other liquids, and is based on the requirements of ASTM method 3612, option C,

More information

Pipettor. User Manual

Pipettor. User Manual Pipettor User Manual Product Code Description Single Chanel Pipettor Variable 550.002.055 Volume 0.5 to 10ul 550.002.060 2 to 20ul 550.002.065 10 to 100ul 550.002.070 20 to 200ul 550.002.075 100 to 100ul

More information

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column Before you connect your trap and column You need to provide your own trap and column (Note our fittings accommodate trap and column made with fused silica 360 µm OD). Check our guide for packing traps

More information

DURRIDGE STAINLESS STEEL SOIL GAS PROBE

DURRIDGE STAINLESS STEEL SOIL GAS PROBE DURRIDGE STAINLESS STEEL SOIL GAS PROBE CONTENTS 1. Introduction... 3 2. Placement... 3 3. Connection... 4 4 Measurement... 4 4.1 GRAB Protocol... 5 4.2 Continuous Monitoring... 6 4.3 Thoron Protocol...

More information

An Overview of Types of Isolation Valves

An Overview of Types of Isolation Valves Isolation Valves 101 An Overview of Types of Isolation Valves To control flow without contaminating media such as blood, pharmaceuticals or reagents, a wide variety of medical devices and analytical equipment

More information

7.9. Flash Column Chromatography Guide

7.9. Flash Column Chromatography Guide WARNING NOTICE: The experiments described in these materials are potentially hazardous and require a high level ofsafety training, special facilities and equipment, and supervision by appropriate individuals.

More information

UsER manual for Watersens ph -REDOX

UsER manual for Watersens ph -REDOX UsER manual for Watersens -REDOX Cl 8 1 2 6 3 3 7 7 4 4 4 4 Parts List 1 Redox Probe 1 x 2 PH Probe 1 x 5 Tube Weight 2 x 6 Connection Valve 1 x chlorine 3 Chlorine and Pumps 2 x 7 Dosing Valve 2 x 5 5

More information

GCMSD-Headspace Analysis SOP

GCMSD-Headspace Analysis SOP Before you start GCMSD-Headspace Analysis SOP Method and Sequences names are restricted to 50 characters (MSD program will crash otherwise) The GC oven and Injection Ports need to be cooled to 50 oc to

More information

POWER FLUSH KIT INSTRUCTIONS

POWER FLUSH KIT INSTRUCTIONS 1 GENERAL INFORMATION POWER FLUSH KIT INSTRUCTIONS ACE Power Flush is an ozone-safe solvent engineered specifically for flushing refrigeration and air conditioning systems. Used exactly like the old R-11

More information

6800 Maintenance Instruction System Flush Procedure

6800 Maintenance Instruction System Flush Procedure Equipment Required FA74005 FA65318 FA900005 FA900003 Damper Drain Tube 6800 Cover Removal Tool Beaker 0.25 Litre Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves

More information

BS ml Bottle Sampling Unit

BS ml Bottle Sampling Unit BS500 500ml Bottle Sampling Unit User Guide www.mpfiltri.co.uk 200.307-EN Covers Model Numbers BS500 SAFETY WARNING Hydraulic systems contain dangerous fluids at high pressures and temperatures. Installation,

More information

Using the Akta Prime plus October 22, 2012

Using the Akta Prime plus October 22, 2012 Some starting precautions: 1. Vacuum filter all buffers. Removes any large particles/debris that may clog your column De-gases the buffers 2. Clarify lysates first by centrifugation and then filtration

More information

mechanical APPLICATION

mechanical APPLICATION mechanical APPLICATION Many SWAROVSKI ELEMENTS, such as Rivets, Jeans Buttons, Snap Fasteners, Decorative Buttons and Magnet Fasteners, can be applied manually or mechanically, using either semi- or fully

More information

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany Manual Degasser DDG-75 April 2007 1 Manual Online Degasser DDG-75 DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE-68766 Hockenheim Germany Tel. +49-6205-9450-0 Fax. +49-6205-9450-33 email info@duratec.com

More information

Technical Bulletin AL-134

Technical Bulletin AL-134 Technical Bulletin AL-134 Handling Air-Sensitive Reagents The Aldrich Sure/Seal system Anhydrous solvents and air-sensitive reagents from Aldrich are packaged in our exclusive Sure/Seal bottles which provide

More information

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment FAST SHIPPING AND DELIVERY TENS OF THOUSANDS OF IN-STOCK ITEMS EQUIPMENT DEMOS HUNDREDS OF MANUFACTURERS SUPPORTED

More information

How to use Solvent Purification System A brief manual

How to use Solvent Purification System A brief manual 1 How to use Solvent Purification System A brief manual By V. Azev. SPS (Figure 1) can be used to collect dry oxygen free solvents into two different types of containers. One of them is a syringe. Syringes

More information

Discussion and guidance on the definition and qualification of porous loads

Discussion and guidance on the definition and qualification of porous loads SUMMARY Porous Loads requiring Sterilisation are often a Critical Control Point for the manufacture of Sterile products. Particularly Aseptically filled products. Regulatory guidance in this area is not

More information

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port So, Why Do I Have To Do Maintenance? 1. Things wear out septa, syringes, nuts, ferrules, o-rings, etc.

More information

COLD WELD COPPER TUBE SAMPLING EQUIPMENT AND PROCEDURES

COLD WELD COPPER TUBE SAMPLING EQUIPMENT AND PROCEDURES COLD WELD COPPER TUBE SAMPLING EQUIPMENT AND PROCEDURES THE COLD WELDING EQUIPMENT The equipment needed to take ~ 45 gram water samples in copper tubes is shown in the figures below. Figure 1 shows the

More information

CRP INSTALLATION, OPERATING AND MAINTENANCE INFORMATION FOR INLINE SAMPLING VALVES

CRP INSTALLATION, OPERATING AND MAINTENANCE INFORMATION FOR INLINE SAMPLING VALVES CRP INSTALLATION, OPERATING AND MAINTENANCE INFORMATION FOR INLINE SAMPLING VALVES Sampling Valves Installation Commissioning and Operating Instructions SD IL 300 & SD IL 400 Inline Sampling Valve This

More information

Patients Guide for Self Administering Intravenous Antibiotics General points

Patients Guide for Self Administering Intravenous Antibiotics General points Patients Guide for Self Administering Intravenous Antibiotics General points An intravenous line is an open route for infection therefore you need to prevent this by washing your hands before any procedure

More information

Navigator 600 Silica analyzers

Navigator 600 Silica analyzers ABB MEASUREMENT & ANALYTICS INFORMATION INF09/020 REV. B Navigator 600 Silica analyzers Troubleshooting Measurement made easy Navigator 600 silica analyzers 1 Introduction This publication details troubleshooting

More information

Pump out insert (condenser and still lines) overnight with turbo pump

Pump out insert (condenser and still lines) overnight with turbo pump Running the Dilution Fridge Before starting a run: Pump out insert (condenser and still lines) overnight with turbo pump Make sure LN and LHe traps are clean pump out with He4 rotary pump Start He3 rotary

More information

CHM250 Calibration and Measurement Lab. Balance Calibration

CHM250 Calibration and Measurement Lab. Balance Calibration CHM250 Calibration and Measurement Lab Green Profile Balance Calibration Introduction: Balances that are properly operated, calibrated and maintained are crucial for laboratory operations. The accuracy

More information

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111 INTRODUCTION Water vapor in natural gas has more than a substantial effect on the quality of the gas stream. Without quality measurement of water vapor the gas is basically not saleable. Contracts are

More information

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02 ! Read this first Zetasizer nano series Self installation and Quick start guide I N S T R U M E N T S MRK825-02 Zetasizer Nano series Self installation and Quick start guide MAN0383 Issue 1.1 July 2007

More information

Model : LMW-9620 Please read this manual carefully before using the instrument Labnics Equipment

Model : LMW-9620 Please read this manual carefully before using the instrument Labnics Equipment MICROPLATE WASHER User Manual Model : LMW-9620 Please read this manual carefully before using the instrument Labnics Equipment Table of Content Sr.No. Contents Pg. No. 1 INTRODUCTION 1 1.1 Working Conditions

More information

Setting up and running a column

Setting up and running a column Setting up and running a column PART 1: What you need and theory PART 2: Using a bellow and packing the column PART 3: Loading sample to silica PART 4: Running column PART 5: TLC analysis and combining

More information

OPERATION. Estimated kerf width compensation. HPR260 Manual Gas Instruction Manual 4-9

OPERATION. Estimated kerf width compensation. HPR260 Manual Gas Instruction Manual 4-9 Estimated kerf width compensation The widths in the chart below are for reference. Differences between installations and material composition may cause the specific user results to vary from those shown

More information

The second stage cannot begin until the first stage is complete.

The second stage cannot begin until the first stage is complete. Addendum to Pneumatic Corker (August 12, 2002) Please read the Operating Instructions before using the pneumatic corker. The purpose of this addendum is three-fold. 1) Describe in more detail how the corker

More information

Microfluidic Demonstration

Microfluidic Demonstration Microfluidic Demonstration Co-flow of liquids across a microfluidic chip Create a microfluidic controller to flow liquids across a microfluidic chip. 1. Assemble fluidic circuits and attach to breadboard.

More information

Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems

Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems The bucket brigade air sampler. For the square Tedlar bags with the push/pull valves: Put the valve of a tedlar

More information

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve USER'S GUIDE Nanoflow Probe 3-4mm Red Stripe ptfe Sleeve To Injector ZU1XC Fused Silica 20µm 90µm TTF115 ptfe Sleeve Fused Silica 25µm 375µm Micromass UK Limited Floats Road Wythenshawe M23 9LZ Tel: +44

More information

User's Manual. MixRite TF 10. Edition 05.08

User's Manual. MixRite TF 10. Edition 05.08 User's Manual MixRite TF 10 Edition 05.08 1 Tefen MixRite TF 10 fertilizer and chemicals Injector Congratulations on your purchase of one of Tefen s high quality products. To get the best results from

More information

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT ATTENTION SYRINGE REPLACEMENT Please read through the document completely before starting any repairs. Refer to the proper section in the service manual for complete removal and replacement procedures.

More information

Filtration FILTRATION PRODUCT TIPS

Filtration FILTRATION PRODUCT TIPS Filtration 195 18 Proved and tested, durable, optimally resistant against acids, caustic solutions and other aggressive chemicals: our solutions for efficient and safe filtration. PRODUCT TIPS BOLA Filtration

More information

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION Standard OQ Test Suite This document describes the test program for qualifying LC analytical-scale Agilent systems; the following table lists

More information

UltiMate 3000 Viper Capillary Kits for Intelligent LC (LCi) System Solutions

UltiMate 3000 Viper Capillary Kits for Intelligent LC (LCi) System Solutions UltiMate 3000 Viper Capillary Kits for Intelligent LC (LCi) System Solutions January 2010 Rev. 02 Table of Contents 1. Introduction... 1 1.1 How to Use this Manual... 1 1.2 Definition of LCi... 2 1.3

More information

Basic Flows in a Microfluidic Device

Basic Flows in a Microfluidic Device Basic Flows in a Microfluidic Device Vishal Tandon 1 Introduction Microfluidic devices are used to transport, mix, and separate analytes in small sample volumes for chemical or biological analysis. Understanding

More information

BioAerosol Nebulizing Generator. Operation and Maintenance User Manual

BioAerosol Nebulizing Generator. Operation and Maintenance User Manual BioAerosol Nebulizing Generator Operation and Maintenance User Manual INTRODUCTION The BANG or BioAerosol Nebulizing Generator is a unique nebulizer for the generation of aqueous aerosols at a low air

More information

1. Introduction Why use the Flow-Through Cell? Component Identification... 1

1. Introduction Why use the Flow-Through Cell? Component Identification... 1 Contents 1. Introduction... 1 2. Why use the Flow-Through Cell?... 1 3. Component Identification... 1 3.1. Standard Components... 2 3.2. Optional Components... 2 4. Assembly and Probe Insertion... 3 4.1.

More information

SOP: Buck Scientific BLC-20P HPLC Operation

SOP: Buck Scientific BLC-20P HPLC Operation Page 1 of 11 Approvals Preparer: John Buford Reviewer: Tim Kull Reviewer: Dr. Margaret Bryans Date: 16OCT13 Date: 30OCT13 Date: 31OCT13 1. Purpose 1.1. Basic operation of the Buck Scientific BLC-20P isocratic

More information

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI NanoSight NS300 Operation instructions Injection/flushing brief overview: 1. Do not exceed flow of 1 ml per 20 seconds. 2. Inject two 1 ml syringes with nano-pure or DI water. 3. If the water does not

More information

IMPROVING QUALITY THROUGH INNOVATION: A FULLY AUTOMATED SOLUBLE SALT METER

IMPROVING QUALITY THROUGH INNOVATION: A FULLY AUTOMATED SOLUBLE SALT METER ARP INSTRUMENTS, INC. SSM MODEL # RPCT-07-001 IMPROVING QUALITY THROUGH INNOVATION: A FULLY AUTOMATED SOLUBLE SALT METER For more information contact: ARP Instruments, Inc. www.arpinstruments.biz arp.instruments@gmail.com

More information

Students measure the change in pressure by varying the volume of trapped air in a syringe while:

Students measure the change in pressure by varying the volume of trapped air in a syringe while: How Does a Trapped Gas Behave? Teacher Information Objective Students investigate the effect of changes in the volume of a confined gas on pressure at constant temperature. Using the pressure sensor, students

More information

For that reason VICI Jour developed:

For that reason VICI Jour developed: VICI Jour Safety VICI Safety Products VICI Safety Products NEW VICI strives to provide customers with the best solutions for their problems to make the work better and/or easier, with respect to environment

More information

Running HPLC Methods on an ACQUITY UPLC System

Running HPLC Methods on an ACQUITY UPLC System Running HPLC Methods on an ACQUITY UPLC System Note: This document is an addendum to the ACQUITY UPLC System Operator s Guide. Contents Topic Page System considerations 1-1 Choosing fittings 1-3 System

More information

HPLC systems. Overview

HPLC systems. Overview HPLC systems Overview An hplc system is just a way to get a sample onto the end of a column, to pump solvents through the column until the analytes drop out the end, and to detect them as they appear.

More information

ACQUITY UPLC Sample Manager - Fixed Loop Operator s Overview and Maintenance Information

ACQUITY UPLC Sample Manager - Fixed Loop Operator s Overview and Maintenance Information ACQUITY UPLC Sample Manager - Fixed Loop Operator s Overview and Maintenance Information Revision A Copyright Waters Corporation 2010 2011 All rights reserved Copyright notice 2010 2011 WATERS CORPORATION.

More information

Parallel Reactor Standard Operating Procedure THIS SOP IS NOT MEANT TO REPLACE HANDS-ON TRAINING FROM QUALIFIED PERSONNEL!

Parallel Reactor Standard Operating Procedure THIS SOP IS NOT MEANT TO REPLACE HANDS-ON TRAINING FROM QUALIFIED PERSONNEL! Parallel Reactor Standard Operating Procedure THIS SOP IS NOT MEANT TO REPLACE HANDS-ON TRAINING FROM QUALIFIED PERSONNEL! This SOP describes the basic procedure for using the HEL Cat-24 parallel reactor,

More information

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC Evaluation of linear and step gradient performance and retention time precision of the Agilent 12 Series Rapid Resolution LC This document is believed to be accurate and up-to-date. However, Agilent Technologies,

More information

LC, RSLC, Bio RS and Dual Gradient Systems

LC, RSLC, Bio RS and Dual Gradient Systems Thermo Scientific Dionex UltiMate 3000 LC, RSLC, Bio RS and Dual Gradient Systems P/N 4823.0205 Revision B December 2014 Automated Method Scouting Considerations All information in this manual is subject

More information

INSTALLATION & MAINTENANCE INSTRUCTIONS - ORIGINAL VERSION AVK KNIFE GATE VALVES

INSTALLATION & MAINTENANCE INSTRUCTIONS - ORIGINAL VERSION AVK KNIFE GATE VALVES 1. INTRODUCTION The AVK knife gate valve is a unique, patented, bi-directional on-off valve that ensures a bubble-tight shut off. The valve is available with body of ductile iron and seat of nitrile, viton,

More information

UP-F Delrin TM Winged Nut. UP-F PEEK TM Nut. with F-142 PEEK TM Ferrule

UP-F Delrin TM Winged Nut. UP-F PEEK TM Nut. with F-142 PEEK TM Ferrule One Piece or Two? A one-piece fitting is more convenient and less cumbersome, since the ferrule cannot stick in a receiving port and the fitting is more easily found if dropped. With two-piece Fingertight,

More information

Agilent 1220 Infinity II LC Mobile Upgrade Kit

Agilent 1220 Infinity II LC Mobile Upgrade Kit Agilent 1220 Infinity II LC Mobile Upgrade Kit Note 1220 Infinity II LC Mobile Upgrade Kit This note describes the procedures to install an Mobile Upgrade kit to an existing Agilent 1220 Infinity II LC.

More information

Aseptic Techniques. Techniques for Sterile Compounding. Pharmacy Technician Training Systems Passassured, LLC

Aseptic Techniques. Techniques for Sterile Compounding. Pharmacy Technician Training Systems Passassured, LLC Aseptic Techniques Techniques for Sterile Compounding Pharmacy Technician Training Systems Passassured, LLC Aseptic Techniques, Tech for Sterile Compounding PassAssured's Pharmacy Technician Training Program

More information

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions Varian 3300 Gas Chromatograph Abbreviated Operating Instructions 1. Introduction and Startup 2 2. Basic Configurations for the GC 3 3. Sample Preparation 3 4. Building a Method 3 5. Split/Splitless Injection

More information

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE GENERAL INFORMATION This family of dissolved oxygen electrodes is of the steam sterilizable polarographic type, designed to be interchangeable with NBS

More information

Mass Spec will not pump down

Mass Spec will not pump down Mass Spec will not pump down Applies to 5973A/N MSD There are several situations that could cause the 5973A/N to have problems pumping down. The following scenarios should help in isolating problems related

More information

RB70 Automatic Diluent Valve Maintenance Manual. Version 1.1 November 2006 Written by Tino de Rijk. Page 1 of 23

RB70 Automatic Diluent Valve Maintenance Manual. Version 1.1 November 2006 Written by Tino de Rijk. Page 1 of 23 RB70 Automatic Diluent Valve Maintenance Manual Version 1.1 November 2006 Written by Tino de Rijk Page 1 of 23 Table of Contents 1. Introduction... 3 2. ADV diagram and parts list (Pre June 2006)... 4

More information

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion: Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: The RSK-175 standard operating procedure was developed in order to determine the amount of dissolved gas in water. Due to the expansion

More information

130 SSP Soluble Salt Profiler

130 SSP Soluble Salt Profiler 130 SSP Soluble Salt Profiler Conforms to: SSPC Guide 15, Bresle Equivalent ISO 8502-9 Test Method Page 1 of 6 The Elcometer 130 Soluble Salt Profiler provides fast and accurate measurement of the level

More information

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents AQUARIUS Operating Instructions for the Type 70 Electrolytic Gas Soldering / Welding Unit Table of Contents 1. Important General Remarks 2 2. Safety Instructions 2 3. Description of the Soldering Unit

More information

Versatility. Accuracy. Compact. Versa. Automated Headspace Vial Sampler

Versatility. Accuracy. Compact. Versa. Automated Headspace Vial Sampler Versatility Accuracy Compact Versa Automated Headspace Vial Sampler Small Size. Big Value. Automated Static Headspace System Static headspace is one of the most popular techniques due to its versatility

More information

THERMALLING TECHNIQUES. Preface

THERMALLING TECHNIQUES. Preface DRAFT THERMALLING TECHNIQUES Preface The following thermalling techniques document is provided to assist Instructors, Coaches and Students as a training aid in the development of good soaring skills. Instructors

More information

Unblocking the MicroSprayer Aerosolizer Model IA-1C With the High Pressure Reverse Cleaning Adapter. Instructions for Use

Unblocking the MicroSprayer Aerosolizer Model IA-1C With the High Pressure Reverse Cleaning Adapter. Instructions for Use Penn-Century, Inc. 7670 Queen Street Suite 200 Wyndmoor, PA 19038-8031 USA 215-753-6540 Tel 215-220-3040 Fax www.penncentury.com Unblocking the MicroSprayer Aerosolizer Model IA-1C With the High Pressure

More information

SPIROVENT VACUUM DEGASSERS

SPIROVENT VACUUM DEGASSERS SPIROVENT VACUUM DEGASSERS S P I R O V E N T A I R S U P E R I O R FLUID AND GAS There are always a certain amount of gasses in a fluid, dependent on pressure and temperature. The British physicist William

More information

Extraction systems. How do you extract your solvents? Trust the market leader: S.C.A.T. - SafetyCaps are THE safety solution!

Extraction systems. How do you extract your solvents? Trust the market leader: S.C.A.T. - SafetyCaps are THE safety solution! Extraction systems How do you extract your solvents? The first and foremost concern is chromatography and secondly, safe handling of solvents is taken into consideration. It is remarkable how little attention

More information

CP-8400 AutoSampler and CP-8410 AutoInjector

CP-8400 AutoSampler and CP-8410 AutoInjector Varian, Inc. 2700 Mitchell Drive Walnut Creek, CA 94598-1675/USA CP-8400 AutoSampler and CP-8410 AutoInjector Operator's Manual Varian, Inc. 2003 Printed in the Netherlands 03-914852-00:Rev. 4.1 Table

More information

Equipment and Accessories. ROMIL LiQuiSure. Designed for High Purity Containment & Dispensing PURE CHEMISTRY

Equipment and Accessories. ROMIL LiQuiSure. Designed for High Purity Containment & Dispensing PURE CHEMISTRY Designed for High Purity Containment & Dispensing To maintain the high purity of ROMIL products during use and handling, we have developed a special range of equipment and accessories under the LiQuiSure

More information

Agilent Auxiliary Gas Module AGM 2. User s Guide

Agilent Auxiliary Gas Module AGM 2. User s Guide Agilent Auxiliary Gas Module AGM 2 User s Guide Notices Agilent Technologies, Inc. 1992, 1996, 1999, 2010 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

Sampling Considerations for Equilibrium Dissolved Oxygen [DO] Sensors

Sampling Considerations for Equilibrium Dissolved Oxygen [DO] Sensors Sampling Considerations for Equilibrium Dissolved Oxygen [DO] Sensors Abstract Boiler water in the low parts per billion dissolved oxygen concentration range can be significantly contaminated by oxygen

More information

Best Practice for Identifying Leaks in GC and GC/MS Systems

Best Practice for Identifying Leaks in GC and GC/MS Systems Best Practice for Identifying Leaks in GC and GC/MS Systems Technical Overview Maintaining a leak-free GC or GC/MS system is critical for obtaining optimal system performance with reliable, reproducible,

More information

4 CONTROL-CHART EXAMPLES

4 CONTROL-CHART EXAMPLES 4 CONTROL-CHART EXAMPLES The following charts are examples of how various conditions will affect your control plots. They are intended only as a guide; your plot may not look exactly like these examples.

More information

Mass Spec will not Autotune

Mass Spec will not Autotune Mass Spec will not Autotune Applies to 5973A/N MSD What could be the problem? There could be several things that would cause your Mass Spec not to Autotune. The most common, easily corrected Autotune problems

More information

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic. 458 LCGC NORTH AMERICA VOLUME 24 NUMBER 5 MAY 2006 www.chromatographyonline.com LC TROUBLESHOOTING Dwell Volume Revisited Gradient transfer does not have to be problematic. John W. Dolan LC Troubleshooting

More information

Technical Procedure for General Laboratory Equipment

Technical Procedure for General Laboratory Equipment Technical Procedure for General Laboratory Equipment 1.0 Purpose - This procedure specifies the required elements for the use of general laboratory equipment. 2.0 Scope - This procedure applies to all

More information

Setting a New Standard in Flash Chromatography Performance

Setting a New Standard in Flash Chromatography Performance Setting a New Standard in Flash Chromatography Performance What s New Faster Flow rates, up to 300 ml/min Higher pressures, up to 300 psi (20 bar) Bigger touchscreen, 12 or 15 inches Wider Dynamic Range

More information

OPERATING INSTRUCTIONS Sherlock Chamber Testing System

OPERATING INSTRUCTIONS Sherlock Chamber Testing System OPERATING INSTRUCTIONS Sherlock Chamber Testing System Arrangement Equipment Kit FL90 & FL100 Corbel Test Plate with Gasket Seal S8, S14, S15 & S16 Sump Test Plate with Gasket Seal S7 & S5 Chamber Test

More information

Agilent 1290 Infinity Pump Head Maintenance

Agilent 1290 Infinity Pump Head Maintenance Agilent 1290 Infinity Pump Head Maintenance Technical Note Agilent Technologies Notices Agilent Technologies, Inc. 2012-2015, 2016 No part of this manual may be reproduced in any form or by any means (including

More information

Procedure for Field Calibration Check of Digital Pressure Gauges (Using a DG-700 or DG-500 as the Reference Standard)

Procedure for Field Calibration Check of Digital Pressure Gauges (Using a DG-700 or DG-500 as the Reference Standard) Procedure for Field Calibration Check of Digital s (Using a or DG-500 as the Reference Standard) 1. Testing Apparatus Set up the (or DG-500) Standard and the gauge to be tested as shown in the diagrams

More information