PURCHASE AND DELIVERY OF A HELIUM CLOSED-CYCLE OPTICAL CRYOSTAT. The present tender is only an unofficial translation provided as a courtesy.

Size: px
Start display at page:

Download "PURCHASE AND DELIVERY OF A HELIUM CLOSED-CYCLE OPTICAL CRYOSTAT. The present tender is only an unofficial translation provided as a courtesy."

Transcription

1 ICMCB - CNRS 87 avenue du Docteur Albert Schweitzer Pessac Cedex France PURCHASE AND DELIVERY OF A HELIUM CLOSED-CYCLE OPTICAL CRYOSTAT Contenu 1. General context Project presentation General objectives Technical specifications Brief description Performances and specificications Warranty Technical support On-site delivery Documentation Execution deadline Operations of check and Admission of the instrument General context The present tender is only an unofficial translation provided as a courtesy Project presentation This project deals with the purchase of a He closed-cycle optical cryostat to allow cryogenfree optical and electrical measurements down to 4 K. 1

2 The present specifications give the general description of the scientific instrument and specify performance levels to be reached General objectives This cryostat will integrate the existing diffuse reflectance setup that uses for temperature control the natural temperature gradient existing in a LHe dewar. The new cryostat will allow measurement to be conducted year round without any dependence on the availability of liquid helium, and reduce the ICMCB consumption of this cryogenic fluid. 2. Technical specifications 2.1. Short description This tender deals with the purchase, the delivery, the installation and the commissioning of a He closed-cycle optical cryostat to allow cryogen-free optical and electrical measurements down to 4 K. The present tender will be the object of a contract with an unique lot with possible options Performances and specifications The desired system is composed of an optical cryostat, the stand necessary for autonomous operation and/or operation within the cavity of a standard spectrometer (UV-Vis, FTIR, ), the compressor for He liquefaction, the corresponding tubing and He reservoir, and sample holders for operation of the cryostat (transmission and reflection optical measurements, electrical measurements). A solution for the measurement and the control of the temperature, either provided or adapted from material existing in the laboratory, will be proposed in option. The cryostat should conform to the following specifications: - Base temperature of 4.2 K or less; - Temperature range: at least up to 300 K (if available up to 400 K, to be given as option,) - Temperature stability :<0.1 K. - Cold head cooling power: >0.1 W@4.2 K (50 Hz). - Cool down time: 2h from room temperature to 4.2 K. - Optical access: measurements in transmission and reflection. Possibility to perform measurements in reflection at 45 incidence (2 windows at 90 ). 2

3 - Optical range: nm. - Electrical measurements: at least one 8-pin connector available. - Compatibility with permanent magnetic fields to be specified. - The sample environment is to be specified (vacuum or exchange gas). The stand should conform to the following specifications: - Compatible with metric fixings. - Compatibility of the footprint with commercial spectrometers cavities (about 22 cm width by 25 cm depth) to be specified. The compressor should conform to the following specifications: - Air-cooled - Electrical feed: 230 V@50 Hz - Weight/footprint less than 300 kg/m 2 - Power consumption, and startup/running amperage to be specified - Noise and vibration level to be specified Relative to sample holders, it must be specified: - Number, nature, temperature range for each type of sample holder. The technical offer must include sample holders allowing to work on the whole temperature range, and for all measurements planned for (optical measurements in transmission and reflection, electrical measurements) Warranty Warrant duration : The instrument is guaranteed for 12 months minimum starting from the purchase date. The total duration of the warranty will be defined at the time of the notification of the contract according to the offer made by the candidate in conformance with the possible additional services (price of the guarantee extension). Warranty cover : - All the travels, without limit of frequency. 3

4 - The whole labor, - Unlimited, on-the-spot interventions, - All the spare parts without exception and without value limit, - Spare parts for the parts subject to the wear - Supply of the necessary equipment to insure the operation and the maintenance of the instrument Technical support The contract holder has to propose a technical support including a permanent and personalized technical support. An engineer will be available during the warranty period. The response time of the engineer after reception of the intervention request must be specified in the offer of the candidate Additional services The following additional services are compulsory (n 1) and optional (n 2). They must be quoted separately as options to the basic offer. Compulsory Option n 1 Temperature Controller A solution for measurement and control of the temperature will be proposed as an option. This solution can consist either of a new temperature controller supplied by the holder, or can be adapted by the holder from the material of this type available in the laboratory. This solution will have to provide measurement of the sample temperature with a resolution lower than 0.05 K on all the temperature range. It will have to allow piloting the cryostat on all the temperature range. Computer control will have to be possible for all functions, modalities must be specified in the offer. Facultative Option n 2 Temperature range extension A solution for the extension of the temperature range above ambient temperature (typically 400 K) can be proposed optionally if the cryostat allows it. 4

5 2.6. On-site delivery The holder of the subsequent contract will have to deliver the instrument looked for at the address indicated in page 1 of the present document. The delivery of all the constituent elements will be chargeable to the supplier. The supplier will have to make sure that the configuration of the room in which will be implanted the equipment will allow its perfect functioning with the announced performances Documentation The holder makes a commitment to supply, at the latest in the admission, all the documentation on paper and digital base, without additional charge, necessary for use and for optimal operation of the equipment. This documentation will contain any information useful for the management of common breakdowns and for the maintenance of the equipment, and will explain clearly the conditions and the technical methods (specifications, plans, technical lists, procedures) allowing an adapted and successful operation of the equipment. 3. Execution deadline The deadline for the installation and the commissioning is fixed on 22/07/2016. The holder specifies in his offer the deadline of delivery and installation and commissioning on which he makes a commitment. 4. Operations of check and Admission of the instrument The commissioning (installation and validation) will be insured by the buyer according to a protocol established by the holder and validated by the buyer. The operations of quantitative check are made at the time of the delivery by the Director of ICMCB or his(her) authorized representative. The operations of checking the capacities and the commissioning of the instrument must be finished at the latest 2 weeks after the delivery of the instrument. These operations are realized under the supervision of the people responsible for the purchase at ICMCB. These operations will contain in particular tests demonstrating the technical capacities of the instrument, realized further to the installation in the premises. At the end of the checking operations, if the checking tests are considered as satisfactory with regard to the contract specifications, the commissioning and the admission of the instrument are pronounced upon presentation of a report of definitive admission signed by the Director of ICMCB or by his(her) authorized representative. 5

6 If the compliance check is negative, the CNRS pronounces either the postponement of the services, with a check of the consistency of service during an additional period of 15 days, or the admission with allowance, or the discharge of the services according to the article 25 of the CCAG / FCS. 6

Microscopy Cryogenic WORKSTATION. Performance by Design

Microscopy Cryogenic WORKSTATION. Performance by Design CRYO Microscopy Cryogenic WORKSTATION Optical cryostat for use in microscopy and spectroscopy Performance by Design RC102-CFM Microscopy Cryostat offers fast cooldown, high efficiency, excellent temperature

More information

Technical Specifications (In-Cash Procurement)

Technical Specifications (In-Cash Procurement) IDM UID WCNER4 VERSION CREATED ON / VERSION / STATUS 26 Apr 2018 / 1.1 / Approved EXTERNAL REFERENCE / VERSION Technical Specifications (In-Cash Procurement) 34.00.00 - IOTS - 000009 : Technical summary

More information

Customer Responsibilities

Customer Responsibilities Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Customer Responsibilities

Customer Responsibilities Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Customer Responsibilities

Customer Responsibilities Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

INTRODUCTION. The Quantum Technology system has the following advantages:, as it does not need plastic gas-bags which are volume-consuming,

INTRODUCTION. The Quantum Technology system has the following advantages:, as it does not need plastic gas-bags which are volume-consuming, 1 INTRODUCTION Quantum Technology is a leading scientific equipment supplier. For the last thirty years we served our customers by tailoring our products to each customer s unique requirements. Quantum

More information

Customer Responsibilities

Customer Responsibilities Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Customer Responsibilities. Important Customer Information. Cary 4000/5000/6000i UV-Vis spectrophotometer Site Preparation Checklist

Customer Responsibilities. Important Customer Information. Cary 4000/5000/6000i UV-Vis spectrophotometer Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Discovery SDT 650. Site Preparation Guide. Revision G Issued December Page 1

Discovery SDT 650. Site Preparation Guide. Revision G Issued December Page 1 Discovery SDT 650 Site Preparation Guide Revision G Issued December 2018 Page 1 Table of Contents Table of Contents... 2 Ideal Setup... 3 System Components... 4 With Discovery Mass Spectrometer... 4 With

More information

Specifications for the CMI 2409 Magnetocardiograph

Specifications for the CMI 2409 Magnetocardiograph Specifications for the CMI 2409 Magnetocardiograph Revision date April 14, 2006 1 518-381-1000 (tel) The CMI-2409 system consists of a holder (A), a patient bed (B) and a workstation (not shown in the

More information

ATL. Take Control of your Helium Supply. Advanced Technology Liquefier The next generation of helium recycling

ATL. Take Control of your Helium Supply. Advanced Technology Liquefier The next generation of helium recycling ATL Take Control of your Helium Supply Advanced Technology Liquefier The next generation of helium recycling Helium A Limited and Costly Resource Cryogenic researchers and technicians around the world

More information

DD2 Site Preparation Checklist Hardware Site Preparation Specification

DD2 Site Preparation Checklist Hardware Site Preparation Specification Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

TECHNICAL SPECIFICATION. SPC-4 Stirling Process Cryogenerator

TECHNICAL SPECIFICATION. SPC-4 Stirling Process Cryogenerator TECHNICAL SPECIFICATION SPC-4 Stirling Process Cryogenerator Reference 80 8043 00 Issue Date August 1, 2015 1. INTRODUCTION This document contains detailed technical information and specifications for

More information

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist Thank you for purchasing an Agilent system. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is

More information

Customer Responsibilities. Important Customer Information. Agilent 6400 Series Triple Quad LC/MS Site Preparation Checklist

Customer Responsibilities. Important Customer Information. Agilent 6400 Series Triple Quad LC/MS Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Heliox TM. He systems to suit all applications

Heliox TM. He systems to suit all applications Heliox TM He systems to suit all applications heliox eliox Which Heliox is right for you? When choosing a Heliox system, there are technical parameters that you need to consider in order to achieve the

More information

Rimage Everest 600 Thermal Retransfer Printer

Rimage Everest 600 Thermal Retransfer Printer Rimage Everest 600 Thermal Retransfer Printer Superior Print Technology: Superior Results CONTENTS I. Overview II. Printer Comparison III. Print Modes IV. Advantages of Thermal Retransfer Printing V. Printer

More information

Current Leads for Cryogenic Systems

Current Leads for Cryogenic Systems Current Leads for Cryogenic Systems American Magnetics Inc. s (AMI) Vapor Cooled Current Leads provide reliable and consistent electrical current transfer from room temperature (300K) to a Liquid Helium

More information

SNOW PICKETS` Climbing and Mountaineering Equipment. Foreword

SNOW PICKETS` Climbing and Mountaineering Equipment. Foreword SNOW PICKETS` Climbing and Mountaineering Equipment UIAA 155 Foreword This UIAA Standard is only published in the English language version, which is the master text. For any validations in translation,

More information

Discovery TGA 55, TGA 550, TGA Site Preparation Guide

Discovery TGA 55, TGA 550, TGA Site Preparation Guide Discovery TGA 55, TGA 550, TGA 5500 Site Preparation Guide Revision E Issued March 2018 Table of Contents Table of Contents... 2 Ideal Setup... 3 System Components... 4 With Discovery Mass Spectrometer...

More information

TECHNICAL SPECIFICATION. SPC-1 Stirling Process Cryogenerator

TECHNICAL SPECIFICATION. SPC-1 Stirling Process Cryogenerator TECHNICAL SPECIFICATION SPC-1 Stirling Process Cryogenerator Reference 80 8013 00 Issue Date August 1, 2015 1. INTRODUCTION This document contains detailed technical information and specifications for

More information

Series Environmental Chambers

Series Environmental Chambers 3119-600 Series Environmental Chambers Challenges in Non-Ambient Testing Testing at non-ambient temperatures adds another layer of challenges to your testing laboratory. Ensuring you get accurate and stable

More information

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service? Pipette specifications according to ISO 8655 Chapter 5 5.1 The ISO 8655 standard gives the accuracy and precision limits as both absolute and relative values. Specifications will depend on the technique

More information

RC102 & RC110 Sample in Vacuum Continuous Flow Cryogenic Workstation Cryostats

RC102 & RC110 Sample in Vacuum Continuous Flow Cryogenic Workstation Cryostats RC102 & RC110 Sample in Vacuum Continuous Flow Cryogenic Workstation Cryostats www.cryoindustries.com INDUSTRIES RC102 & RC110 Sample in Vacuum Continuous Flow Cryogenic Workstation Cryostats Designed

More information

Agilent Auxiliary Gas Module AGM 2. User s Guide

Agilent Auxiliary Gas Module AGM 2. User s Guide Agilent Auxiliary Gas Module AGM 2 User s Guide Notices Agilent Technologies, Inc. 1992, 1996, 1999, 2010 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

Customer Responsibilities

Customer Responsibilities Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Fermilab MTF Cryogenic System & Vertical RF Cavity Test Facility. Yuenian Huang, on behalf of Fermilab, TD/TID May 11, 2006

Fermilab MTF Cryogenic System & Vertical RF Cavity Test Facility. Yuenian Huang, on behalf of Fermilab, TD/TID May 11, 2006 Fermilab MTF Cryogenic System & Vertical RF Cavity Test Facility Yuenian Huang, on behalf of Fermilab, TD/TID May 11, 2006 1 Magnet Test Facility (MTF) Tevatron magnet test stands Testing temperature range

More information

Technical Bulletin, Meter Factor Linearization

Technical Bulletin, Meter Factor Linearization Last Updated: 06-April-2009 TB-970803B Technical Bulletin, OMNI FLOW COMPUTERS, INC. 12620 West Airport Boulevard, Suite 100 Sugar Land, Texas 77478 United States of America Phone-281.240.6161 Fax: 281.240.6162

More information

Linde Cryogenics. Helium Recovery Systems.

Linde Cryogenics. Helium Recovery Systems. Linde Cryogenics. Helium Recovery Systems. 02 Contents. 3 Introduction 4 Recovery Compressor Options 6 Installation Recovery Piping and Fittings Kits 6 Model 500A Automatic External Helium Purifier 7 Purity

More information

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc. Advanced Calibration Designs, Inc. Calibration Gas Instrument INSTRUCTION MANUAL Release I www.goacd.com Instruction Manual Gas Generator Release I TABLE OF CONTENTS I. General Description Page 2 II. Start-Up

More information

LC-SPE. Organizer 3.0 User Manual. Innovation with Integrity. Version 001 NMR

LC-SPE. Organizer 3.0 User Manual. Innovation with Integrity. Version 001 NMR LC-SPE Organizer 3.0 User Manual Version 001 Innovation with Integrity NMR Copyright by Bruker Corporation All rights reserved. No part of this publication may be reproduced, stored in a retrieval system,

More information

Figure 1 The LHC cryogenic islands and plants layout

Figure 1 The LHC cryogenic islands and plants layout The LHC cryogenic operation for first collisions and physics run Brodzinski K, Barth K, Benda V, Bremer J, Casas-Cubillos J, Claudet S, Delikaris D, Ferlin G, Fernandez Penacoba G, Perin A, Pirotte O,

More information

LESER Deutschland Standard Functional Tightness Test (Cryo) Content

LESER Deutschland Standard Functional Tightness Test (Cryo) Content Page 1/10 Content 1 Purpose... 1 2 Scope... 1 3 Range of Application... 1 4 Normative Background... 2 5 Terms and Definitions... 2 6 Test Equipment... 3 7 Test Procedure... 5 8 Acceptance Criteria... 8

More information

OPERATION MANUAL FOR VP 710E-3HM SERIES ALLIGATOR TUMBLE STIRRERS *US Patent # 6,176,609

OPERATION MANUAL FOR VP 710E-3HM SERIES ALLIGATOR TUMBLE STIRRERS *US Patent # 6,176,609 TECHNICAL NOTE 299 OPERATION MANUAL FOR VP 710E-3HM SERIES ALLIGATOR TUMBLE STIRRERS *US Patent # 6,176,609 WARNING!!!!! Be advised that the Magnetic Tumble Stirrer has very strong magnetic fields coming

More information

Flexible Metal Hose Products

Flexible Metal Hose Products Flexible Metal Hose Products Specialty Flexible Hose Solutions of Stainless Steel The AEROCOM Advantage Aerocom offers engineered solutions that address specific flexible piping challenges such as vibration,

More information

IMPROVED XENON LOADING EQUIPMENT WITH LOADING CAPACITY UP TO 1200 KG FOR ALPHABUS

IMPROVED XENON LOADING EQUIPMENT WITH LOADING CAPACITY UP TO 1200 KG FOR ALPHABUS IMPROVED XENON LOADING EQUIPMENT WITH LOADING CAPACITY UP TO 1200 KG FOR ALPHABUS Patrick BRAVAIS, Roland SALOME +, Cécile GELAS + * AIR LIQUIDE DTA + CNES Toulouse Abstract Plasmic propulsion systems

More information

CODE OF PRACTICE 42 IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS REVISION 2: 2018

CODE OF PRACTICE 42 IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS REVISION 2: 2018 CODE OF PRACTICE 42 IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS REVISION 2: 2018 CODE OF PRACTICE 42 IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS REVISION 2: 2018 Copyright 2018 by British Compressed

More information

AB AMBER GRID RULES FOR NATURAL GAS TRANSMISSION SYSTEM BALANCING I. GENERAL PROVISIONS

AB AMBER GRID RULES FOR NATURAL GAS TRANSMISSION SYSTEM BALANCING I. GENERAL PROVISIONS AB AMBER GRID RULES FOR NATURAL GAS TRANSMISSION SYSTEM BALANCING I. GENERAL PROVISIONS 1. AB Amber Grid Rules for Natural Gas Transmission System Balancing (hereinafter - the Rules) are designed to identify

More information

IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS

IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS CODE OF PRACTICE 42 IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS REVISION 1 : 2014 CODE OF PRACTICE 42 IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS REVISION 1: 2014 Copyright 2014 by British Compressed

More information

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Your local gas generation partner. Precision series Modular gas generation solution for GC. Your local gas generation partner Precision series Modular gas generation solution for GC www.peakscientific.com Perform with Precision Specifically designed and engineered for GC laboratory applications,

More information

Oxford Instruments Plasma Technology. PlasmaPro NGP80. Installation Data Sheet. Original Instructions.

Oxford Instruments Plasma Technology. PlasmaPro NGP80. Installation Data Sheet. Original Instructions. Oxford Instruments Plasma Technology www.oxford-instruments.com About this data sheet This data sheet provides basic information about the installation of a tool. Floor and wall loadings Table 1 lists

More information

AB AMBER GRID RULES FOR NATURAL GAS TRANSMISSION SYSTEM BALANCING I. GENERAL PROVISIONS

AB AMBER GRID RULES FOR NATURAL GAS TRANSMISSION SYSTEM BALANCING I. GENERAL PROVISIONS AB AMBER GRID RULES FOR NATURAL GAS TRANSMISSION SYSTEM BALANCING I. GENERAL PROVISIONS 1. AB Amber Grid Rules for Natural Gas Transmission System Balancing (hereinafter - the Rules) are designed to identify

More information

Engineering: Measurement Technology Pressure/Level (SCQF level 6)

Engineering: Measurement Technology Pressure/Level (SCQF level 6) National Unit Specification: general information CODE F5KT 12 SUMMARY This Unit can be delivered as part of a National Qualification Group Award but can also be taken as a free-standing Unit. This Unit

More information

Fiji Electricity Authority Transmission Unit 2 Marlow Street, Suva

Fiji Electricity Authority Transmission Unit 2 Marlow Street, Suva TENDER DOCUMENT AND SPECIFICATION MR 333/2017 Preferred Supplier for Double Tail Shock Absorbing Elasticated Webbing Lanyard The Fiji Electricity Authority (FEA) invites bids from reputable suppliers for

More information

All TSOs of the Nordic Capacity Calculation Region Proposal for fallback procedures in accordance with Article 44 of Commission Regulation (EU)

All TSOs of the Nordic Capacity Calculation Region Proposal for fallback procedures in accordance with Article 44 of Commission Regulation (EU) All TSOs of the Nordic Capacity Calculation Region Proposal for fallback procedures in accordance with Article 44 of Commission Regulation (EU) 2015/1222 of 24 July 2015 establishing a guideline on capacity

More information

Baffles and Traps Cryotrap

Baffles and Traps Cryotrap Baffles and Traps 362-6 Cryotrap Varian Low-Profile Water-Cooled Baffles combine 100% optical density with high conductance and unusually low overall height. They are especially useful in applications

More information

Vacuum. Dry Turbo Pumping Stations. Roughing Pumps for 1 K Helium Pots and Lambda Refrigerators. Hermetically Sealed Pumps for He3 Recirculation

Vacuum. Dry Turbo Pumping Stations. Roughing Pumps for 1 K Helium Pots and Lambda Refrigerators. Hermetically Sealed Pumps for He3 Recirculation Cryogenics Vacuum Dry Turbo Pumping Stations Roughing Pumps for 1 K Helium Pots and Lambda Refrigerators Hermetically Sealed Pumps for He3 Recirculation Oils, Filters and Traps Valves, Fittings and Adaptors

More information

Varian 400-MR Magnetic Resonance Spectrometer Pre-installation Instructions

Varian 400-MR Magnetic Resonance Spectrometer Pre-installation Instructions Checklist NOTE: Do not unpack the boxes. Varian 400-MR Magnetic Resonance Spectrometer Pre-installation Instructions Mark the check box after satisfying each requirement as described in the instructions.

More information

RDK-408D2 Cold Head. Technical Manual. SHI-APD Cryogenics Inc Vultee Street Allentown, PA U.S.A. Revision A: September 2005

RDK-408D2 Cold Head. Technical Manual. SHI-APD Cryogenics Inc Vultee Street Allentown, PA U.S.A. Revision A: September 2005 RDK-408D2 Cold Head Technical Manual SHI-APD Cryogenics Inc. 1833 Vultee Street Allentown, PA 18103-4783 U.S.A. Revision A: September 2005 (Reference SHI Manual: December 18, 2003 266404A CD32ZZ-160A)

More information

Varian 325-MS LC/MS Quadrupole Mass Spectrometer Pre-installation Instructions

Varian 325-MS LC/MS Quadrupole Mass Spectrometer Pre-installation Instructions Varian 325-MS LC/MS Quadrupole Mass Spectrometer Pre-installation Instructions Checklist NOTE: Do not unpack the shipping cartons. Check off each checklist box after satisfying each requirement as described

More information

First Experimental Data of the Cryogenic Safety Test Facility PICARD

First Experimental Data of the Cryogenic Safety Test Facility PICARD First Experimental Data of the Cryogenic Safety Test Facility PICARD C. Heidt, A. Henriques, M. Stamm and S. Grohmann 26 th International Cryogenic Engineering Conference 2016 in New Delhi, India KIT,

More information

Digital Melting Point Apparatus

Digital Melting Point Apparatus Digital Melting Point Apparatus Heating Plateau Ramping Start/Stop Plateau set Ramp stop Hold User Guide Version 1.1 Heating Viewing tube Sample Chamber IEC power inlet socket Power on/off Temperature

More information

BPS-2, Rev TABLE OF CONTENTS

BPS-2, Rev TABLE OF CONTENTS BPS-2, Rev. 050404 Bilayer Perfusion System Model BPS-2 125 Dixwell Avenue, Hamden, CT 06514 (800) 599-4203 / (203) 776-0664 (203) 776-1278 fax www.warneronline.com BPS-2, Rev. 050404 TABLE OF CONTENTS

More information

490 Micro GC Site Preparation Checklist

490 Micro GC Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Flow controller Type ED/CCB311 Instruction manual

Flow controller Type ED/CCB311 Instruction manual Flow measurement Flow controller Type ED/CCB311 1. GENERAL DESCRIPTION page 4 2. DETAILED DESCRIPTION page 4 3. PIPE - EXPLOITATION page 5 4. PREVENTIVE MAINTENANCE page 5 5. REPLACEMENT OF THE CONTACT

More information

Dynamic Operation of a 4 K Pulse Tube Cryocooler with Inverter Compressors

Dynamic Operation of a 4 K Pulse Tube Cryocooler with Inverter Compressors Dynamic Operation of a 4 K Pulse Tube Cryocooler with Inverter Compressors C. Wang Cryomech, Inc. Syracuse, NY 13211 ABSTRACT This paper introduces the concept of operating a 4 K pulse tube cryocooler

More information

PRC CO ² -LASER PRESENTATION

PRC CO ² -LASER PRESENTATION Page 1 of 7 PRC CO ² -LASER PRESENTATION GENERAL CHARACTERISTICS - Embedded PC104 Electronics with exchangeable software allowing very easy integration of customer specified functions: such as eg. specific

More information

TESTING BENCHES for watermeters

TESTING BENCHES for watermeters TESTING BENCHES for watermeters +372 6 480 235 Haljas tee 25, 12012 TALLINN Fax +372 6 480 125 Testing benches are designed by Tepso Ltd. for using in calibration laboratories and also in watermeter production

More information

ASSE International Product (Seal) Listing Program

ASSE International Product (Seal) Listing Program ASSE International Product (Seal) Listing Program ASSE 1070-2015 / ASME A112.1070-2015 / CSA B125.70-15 Performance Requirements for Water Temperature Limiting Devices Manufacturer: Contact Person: E-mail:

More information

User s Guide. Pressure Controller for liquid delivery systems

User s Guide. Pressure Controller for liquid delivery systems Flow Control User s Guide Pressure Controller for liquid delivery systems Precise Pressure Control throughout the experiment Flow control Compatible with any perfusion system Ideal for Small Volume Delivery

More information

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures

More information

Gas Vapor Injection on Refrigerant Cycle Using Piston Technology

Gas Vapor Injection on Refrigerant Cycle Using Piston Technology Purdue University Purdue e-pubs International Refrigeration and Air Conditioning Conference School of Mechanical Engineering 2012 Gas Vapor Injection on Refrigerant Cycle Using Piston Technology Sophie

More information

EXPERIENCE WITH A PRE-SERIES SUPERFLUID HELIUM TEST BENCH FOR LHC MAGNETS

EXPERIENCE WITH A PRE-SERIES SUPERFLUID HELIUM TEST BENCH FOR LHC MAGNETS EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH European Laboratory for Particle Physics Large Hadron Collider Project LHC Project Report 388 EXPERIENCE WITH A PRE-SERIES SUPERFLUID HELIUM TEST BENCH FOR LHC

More information

BSP Pressure Sensor Balluff GmbH, BA Marco Zorcic.

BSP Pressure Sensor Balluff GmbH, BA Marco Zorcic. 1 2 Standard Pressure Sensor with display Overview Pressure Range Relative nominal Overload Burst Permitted negative -1..2 bar 2 bar 4 bar 10 bar -1..10 bar 10 bar 20 bar 35 bar 0..2 bar 2 bar 4 bar 10

More information

Changes to the Gas. (Safety and Measurement) Regulations. effective 1 July 2013

Changes to the Gas. (Safety and Measurement) Regulations. effective 1 July 2013 Changes to the Gas (Safety and Measurement) Regulations effective 1 July 2013 www.energysafety.govt.nz www.pgdb.co.nz www.standards.co.nz Changes to the gas certification regime under the Gas (Safety and

More information

GAS LIFT IN NEW FIELDS

GAS LIFT IN NEW FIELDS GAS LIFT IN NEW FIELDS GAS LIFT in NEW FIELDS NEW DESIGN FOR WELLS AND COMPRESSORS RESERVOIR AND PRODUCTION SIMULATION MATCHING REQUIRED PRODUCTION NODAL ANALYSIS SIMULATION USED TO SELECT INJECTION DEPTH

More information

Finding Leak Testing Success Through Calibration and Validation. Paul Chamberlain President, CEO

Finding Leak Testing Success Through Calibration and Validation. Paul Chamberlain President, CEO Finding Leak Testing Success Through Calibration and Validation Paul Chamberlain President, CEO 2 Overview Confidence in Your Leak Test Results Sources of Uncertainty Pressure decay leak testing Helium

More information

Pneumatic spool valve islands

Pneumatic spool valve islands Pneumatic spool valve islands P7.GB.R MEGA SPOOL VALVE ISLANDS MEGA pneumatic spool valve islands offer maximum performance along with flexibility, easy installation and simple operation. Due to their

More information

FLANGED TWO-PIECE BALL VALVES

FLANGED TWO-PIECE BALL VALVES INTRODUCTION This instruction manual includes installation, operation, and maintenance information for FNW flanged split-body ball valves. This manual addresses lever operated ball valves only. Please

More information

User Manual MFCS TM -EZ

User Manual MFCS TM -EZ User Manual MFCS TM -EZ Version 6A_December 2016 1. INTRODUCTION 5 2. GENERAL INFORMATION 6 2.1 General functioning 6 2.2 Different MFCS TM 6 3. PACKAGE CONTENT 8 4. MFCS TM -EZ DESCRIPTION 9 5. CONNECTIONS

More information

Major Design Features

Major Design Features Major Design Features 3Flex Features Design Benefits Discussion 316 SS fittings and pneumatically actuated, hard-seal valves Provides virtually leak-free gas management with the lowest outgassing rate

More information

Discovery HP-TGA 75/750. Site Preparation Guide

Discovery HP-TGA 75/750. Site Preparation Guide Discovery HP-TGA 75/750 Site Preparation Guide Revision A Issued August 2018 Table of Contents Table of Contents... 2 Ideal Setup... 3 System Components... 4 Instrument Measurements... 5 Utility Requirements...

More information

An Investigative and Synoptic Review on Helium Liquefaction Using a Commercial 4 K Gifford- McMahon Cryocooler

An Investigative and Synoptic Review on Helium Liquefaction Using a Commercial 4 K Gifford- McMahon Cryocooler An Investigative and Synoptic Review on Helium Liquefaction Using a Commercial 4 K Gifford- McMahon Cryocooler Mahesh N 1, S A Mohan Krishna 2 1 PG Student, Vidyavardhaka College of Engineering, Mysore,

More information

GENERAL SAFETY INSTRUCTION GSI-M-2 STANDARD PRESSURE EQUIPMENT

GENERAL SAFETY INSTRUCTION GSI-M-2 STANDARD PRESSURE EQUIPMENT GENERAL SAFETY INSTRUCTION GSI-M-2 STANDARD PRESSURE EQUIPMENT Publication date: 09-06-2015 All CERN Safety Rules are available on the web site: https://www.cern.ch/safety-rules Page 1 of 8 1 INTRODUCTION

More information

attoliquid Liquid Research Cryostats

attoliquid Liquid Research Cryostats WITTENSTEIN Group WITTENSTEIN Group 2018, attocube systems AG - Germany. attocube systems and the logo are trademarks of attocube systems AG. Registered and/or otherwise protected in various countries

More information

Title: INSTRUMENT AND EQUIPMENT DOCUMENTATION AND RECORDS

Title: INSTRUMENT AND EQUIPMENT DOCUMENTATION AND RECORDS Page 1 of 7 Sections Included in this Document and Document History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References/(Deleted date) 6. Procedure 7. Definitions 8. Records 9. Supporting

More information

HumiSys HF High Flow RH Generator

HumiSys HF High Flow RH Generator HumiSys HF High Flow RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The HumiSys HF is a high flow version of the previously

More information

Information Sheet. About this information sheet. Floor and wall loadings. Attachment of process module to floor. PlasmaPro Estrelas100

Information Sheet. About this information sheet. Floor and wall loadings. Attachment of process module to floor. PlasmaPro Estrelas100 Information Sheet PlasmaPro Estrelas100 About this information sheet This data sheet provides basic information about the installation of a PlasmaPro Estrelas100 tool. Some of the information in this sheet

More information

Safety. Operating instructions Solenoid valve for gas VG 6 VG 15/10 DANGER. Contents WARNING CAUTION. Changes to edition 09.14

Safety. Operating instructions Solenoid valve for gas VG 6 VG 15/10 DANGER. Contents WARNING CAUTION. Changes to edition 09.14 15 Elster GmbH Edition 7.15 Translation from the German 519 D F NL I E DK S N P GR TR CZ PL RUS H www.docuthek.com Operating instructions Solenoid valve for gas VG VG 15/1 Contents Solenoid valve for gas

More information

WP1- NA1: ACTRIS Management and coordination Deliverable D1.3: Launch for call of TNA

WP1- NA1: ACTRIS Management and coordination Deliverable D1.3: Launch for call of TNA WP1- NA1: ACTRIS Management and coordination Deliverable D1.3: Launch for call of TNA The Coordination Office officially posted the permanent call for Transnational Access to ACTRIS in July 2010 (month

More information

LUMAZOTE T ECHNICAL BRO CHU RE

LUMAZOTE T ECHNICAL BRO CHU RE by Chemiluminescence Standards AST D4629 / ASTM D6069 / ASTM D 5176 / ASTM D 5762 / NF EN 12260 / NF M 07-058 Licence TOTAL LUMAZOTE T ECHNICAL BRO CHU RE ERALY & Associés Phone: +33 (0) 1 77 04 80 97

More information

Sensor for Air Bubble Detection at Liquid Filled Tubes. SONOCHECK Type ABD06.xx. Operating Manual

Sensor for Air Bubble Detection at Liquid Filled Tubes. SONOCHECK Type ABD06.xx. Operating Manual Sensor for Air Bubble Detection at Liquid Filled Tubes SONOCHECK Type ABD06.xx Operating Manual Manufacturer: Model: Type: SONOTEC Ultraschallsensorik Halle GmbH Air Bubble Detector ABD06.xx SONOTEC Ultraschallsensorik

More information

Single and double acting, magnetic, self-centering Bores: ø 16, 20, 25, 32, 40 mm

Single and double acting, magnetic, self-centering Bores: ø 16, 20, 25, 32, 40 mm > Series CGPT self-centering parallel grippers with T-guide Series CGPT self-centering parallel grippers with T-guide Single and double acting, magnetic, self-centering Bores: ø 6, 20, 25, 32, 40 mm New

More information

Siphon vessel / Priming aid

Siphon vessel / Priming aid Manufacturer: sera GmbH sera-straße 1 34376 Immenhausen Germany Tel.: +49 5673 999-00 Fax: +49 5673 999-01 info@sera-web.com Keep the operating manual for future use! Record the exact type and serial number

More information

Deadweight Tester / Pressure Balance High pressure up to 5,000 bar / 70,000 psi CPB5000-HP. Applications. Special features. Description. Rel.

Deadweight Tester / Pressure Balance High pressure up to 5,000 bar / 70,000 psi CPB5000-HP. Applications. Special features. Description. Rel. High up to 5,000 bar / 70,000 psi CPB5000-HP Rel. 20150206 Applications Primary standard for defining the scale in a range up to 5,000 bar Reference instrument for testing, adjusting and calibrating measuring

More information

SAPAG. Safety valves, type 5700 Storage, Use, Operation and Maintenance Instructions. IMPORTANT NOTICE

SAPAG. Safety valves, type 5700 Storage, Use, Operation and Maintenance Instructions.  IMPORTANT NOTICE SAPAG IMPORTANT NOTICE Contents Important notice 1 0 Valve identification 2 1 Storage 2 2 Installation 2 3 Operation 2 4 Maintenance 3 4.1 Dismantling 3 4.2 Inspection 3 4.3 Repair 3 4.4 Assembly 4 4.5

More information

Genesis MX SLM-Series

Genesis MX SLM-Series Single Frequency Visible OEM and End-User OPS Laser Systems Applications like Flow Cytometry, Particle Counting, DNA Sequencing and Microscopy are enable by low noise, visible true CW lasers. The Genesis

More information

Hydraulic hand test pump Models CPP700-H and CPP1000-H

Hydraulic hand test pump Models CPP700-H and CPP1000-H Calibration technology Hydraulic hand test pump Models CPP700-H and CPP1000-H WIKA data sheet CT 91.07 for further approvals see page 2 Applications Simple test pressure generation on site, in the laboratory

More information

A Brief Introduction to Helium Liquefaction At ITER

A Brief Introduction to Helium Liquefaction At ITER ITER Cryogenic System LHe PLANTS A Brief Introduction to Helium Liquefaction At ITER ITER Organization PLANT ENGINEERING DIVISION CRYOGENIC SECTION LHe PLANTS TRO: Eric FAUVE A Brief Introduction to to

More information

SPECIFIC SAFETY INSTRUCTION SSI-M-1-3 NON-FIXED LOAD-LIFTING ACCESSORIES

SPECIFIC SAFETY INSTRUCTION SSI-M-1-3 NON-FIXED LOAD-LIFTING ACCESSORIES SPECIFIC SAFETY INSTRUCTION SSI-M-1-3 NON-FIXED LOAD-LIFTING ACCESSORIES Publication date: 09-06-2015 All CERN Safety Rules are available on the website: https://www.cern.ch/safety-rules Page 1 of 5 1

More information

MIL-STD-883G METHOD

MIL-STD-883G METHOD STEADY-STATE LIFE 1. PURPOSE. The steady-state life test is performed for the purpose of demonstrating the quality or reliability of devices subjected to the specified conditions over an extended time

More information

Variable Temperature Storage Dewar Mount Inserts

Variable Temperature Storage Dewar Mount Inserts Variable Temperature Storage Dewar Mount Inserts All inserts are available for either Liquid Helium or Liquid Nitrogen Dewars! DStat - External cryostat with insert leg DStatMag - Super Conducting Magnet

More information

UltraLo 1800 Alpha Particle Counter

UltraLo 1800 Alpha Particle Counter XIA LLC UltraLo 1800 Alpha Particle Counter Site Requirements & Planning Version: 0.4 Friday, February 15, 2013 Table of Contents I. Site Preparation Tasks... 3 A. Instrument Overview... 3 B. Installation

More information

EXPERIMENT XI. Careful!! Improper handling of the vacuum line may result in the release of SO 2 which is an irritating and suffocating gas.

EXPERIMENT XI. Careful!! Improper handling of the vacuum line may result in the release of SO 2 which is an irritating and suffocating gas. Chem 366-3 Page XI - 1 EXPERIMENT XI INFRARED SPECTRUM OF SO2 (S&G, 5th ed. Expt 36, 6th ed. Expt. 35) 1. Pre-Lab preparation. The description of this experiment has disappeared from the more recent editions

More information

Liquid Helium Bath Cryostat

Liquid Helium Bath Cryostat Liquid Helium Bath Cryostat Operating Manual Model TK1865M CONTENTS First A Word of Warning Lifting and Handling the Cryostat Using Cryogens Introduction The liquid helium cryostat. Type TK1865M Serial

More information

The instrument should have a mass range from 2 to 2000 m/z or better Interface

The instrument should have a mass range from 2 to 2000 m/z or better Interface Revised Specifications for Liquid Chromatography Mass Spectrometry (LC-MS/MS ) Triple Quadrupole Specification for Ultra-Fast Liquid Chromatography system Solvent Delivery 1) A quaternary Pump system for

More information

Improve Process Reliability

Improve Process Reliability Convenient Conductivity Calibration to Meet Pharmacopeia Regulations Pharmacopeias require that conductivity measurement systems be periodically calibrated to ensure their precision. The range of accessories

More information

TECHNICAL SPECIFICATION. StirLNG-4 Stirling Cryogenics gas liquefier for LNG production

TECHNICAL SPECIFICATION. StirLNG-4 Stirling Cryogenics gas liquefier for LNG production TECHNICAL SPECIFICATION StirLNG-4 Stirling Cryogenics gas liquefier for LNG production Reference 80 8414 04 Issue Date November 1, 2016 1. INTRODUCTION Since more than sixty years Stirling Cryogenics has

More information

TLN WRO Document. Back to Back CAS support

TLN WRO Document. Back to Back CAS support TELENET N.V. Page 1 of 8 TLN-WRO_GA_P_O_PDAA TLN WRO Document Back to Back CAS support TELENET N.V. Page 2 of 8 TLN-WRO_GA_P_O_PDAA Document Category and type CAT TYPE DOC ID TV General TLN_WRO_GA_P_O_PDAA

More information

SUP 15 Health & Safety Management Pressure Systems. Unified procedures for use within NHS Scotland

SUP 15 Health & Safety Management Pressure Systems. Unified procedures for use within NHS Scotland SUP 15 Health & Safety Management Pressure Systems Unified procedures for use within NHS Scotland September 2015 Contents Page Acknowledgements... 3 1. Introduction... 4 2. Purpose of this Procedure...

More information