Government of India DR. RAM MANOHAR LOHIA HOSPITAL, NEW DELHI (PROCUREMENT SECTION)
|
|
- Victoria Holt
- 5 years ago
- Views:
Transcription
1 Tele No D.R.M.L.H (A) Government of India DR. RAM MANOHAR LOHIA HOSPITAL, NEW DELHI (PROCUREMENT SECTION) No /2014-LH(S)/5523 NEW DELHI Dated: C O R R I G E N D U M The following amendment in Tender Enquiry No.11-01/2014-LH(S) Dated are hereby authorized:- Schedule No. Tender Code Number Brief Description of Goods Amendment Date of Opening /2016-LH(S)# ICU Ventilator (04 Nos.) Revised Technical Specification /2016-LH(S)# Coagulation & Platelet Function Analyser No Change (04 Nos.) /2016-LH(S)# ICU Beds (16 Nos.) Revised Technical Specification /2016-LH(S)# Portable Ultrasound System (03 Nos.) Revised Technical Specification /2016-LH(S)# Defibrillator with Monitor & Pacer with Revised Technical AED (10 Nos.) Specification /2016-LH(S)# Fluid Warming Delivery System (08 Nos.) Revised Technical Specification /2016-LH(S)# DVT Pump (20 Nos.) No Change /2016-LH(S)# Cerebral Oximeter (04 Nos.) Revised Technical Specification /2016-LH(S)# Anaesthesia Workstation (02 Nos.) Revised Technical Specification /2016-LH(S)# Multiparameter Monitors for ICU/Recovery (06 Nos.) Revised Technical Specification All other terms and conditions of the subject tender enquiry shall remain unchanged. Sd/- C.M.O. I/c. Procurement For Medical Superintendent
2 Schedule No. 33 Tender Code No /2016-LH(S) REVISED SPECIFICATION FOR ICU VENTILATOR (HIGH END) 1. The ventilator should be microprocessor based with active exhalation valve. 2. Should work on hospital external high pressure line/ compressor. The compressor of the same manufacturer needs to inbuilt. 3. Suitable for use in ICU for Adult and Paediatric patients. 4. Should undergo automatic calibration system on start-up. 5. In the event of central Air line failure the Inbuilt compressor should activate automatically. 6. It should have in-built programmable nebulizer (vol. & pressure compensated) system. 7. It should be easy to use having a color in-built touch screen at least 12 inch or more in size with screen lock, intuitive menu structure,, Mode preset capability, Pressure bar graph/ breath indicator and prioritized alarms 8. Should have the following Ventilation Modes: a) Volume Controlled ventilation (Assisted / Control) VCV b) Pressure Controlled ventilation (Assisted / Control) PCV c) Synchronized intermittent mandatory Ventilation V-SIMV AND P-SIMV d) Pressure support ventilation (Spont, CPAP, PEEP) PSV e) Non invasive ventilation VCV, PCV, SIMV, PSV f) Volume assured pressure support/ Similar VAPS g) Mandatory rate ventilation or similar MRV h) Airway pressure release ventilation APRV/BI-PHASIC VENTILATION i) Pressure regulated volume control PRVC j) Intermittent positive release ventilation or similar IMPRV k) Continuous positive airway pressure CPAP l) Automatic weaning Tool
3 9. Should have the following Ventilation Settings and Ranges a) Tidal Volume 20 ml to 2000 ml or more b) I:E ratio 1:4 to 4:1 c) Inspiratory Peak Flow 200 LPM or more (Compensated) d) Maximum Inspiratory Peak Flow 200 l/min or more (depending on gas supply pressure) e) Respiratory Rate BPM f) Pressure support 0-40 cm H2O g) Inspiratory Pressure: 0-80 cm H2O h) FiO2 21% to 100% i) Ti sec or more j) PEEP/CPAP 0-50 cm H2O k) Inspiratory pause, Expiratory Pause, sustained exhalation, Programmable/ adjustable SIGH l) Inspiratory Trigger (pressure mbar/ flow trigger L/ min or more) 10. Should monitor the following parameters: Respiratory Phase & Type, Respiratory Rate, Exhaled Tidal Volume, Exhaled Min. Volume Total, % leak, Leak min. vol., Spont. Min. volume, I : E Ratio, Peak Inspiratory Pressure, Average Pressure, Plateau Pressure, End Expiratory Pressure, Auto/Intrinsic PEEP, % Oxygen Delivered, Frequency/ Tidal Volume (RSBI), main stream etco2(end tidal Co2). External etco2 monitor is not acceptable. 11. Should measure following Respiratory Mechanics/ Maneuvers: Static Compliance and Resistance, Low Inflection point (LIP) and upper inflection point (UIP), Maximum Inspiratory Pressure, Work of Breathing, Lung recruitment maneuver & monitoring 12. Should have a Graphics Mode with simultaneous display of all curves and loops a) Scalars Flow vs. Time Pressure vs. Time Adjustable Time Scale.
4 b) Loops Flow / Volume Pressure / Volume Pressure/flow c) Facility for Freeze Screen Individual Analysis of Each Curve Loop Save and Overlay Function Individual Analysis of Each Loop d) Calculated Values Inspiratory pause, Expiratory Pause 13. Scope of supply Ventilator 1 No Patient Tubing (adult) Fisher and Paykel or Equivalent: 2 Nos/unit Patient Tubing (paed) Fisher and Paykel or Equivalent: 2 Nos/unit Nebuliser Kit 10 Nos/ Ventilator NIV Mask with harness (Reusable) 5 No for adult and Pediatric each/ Ventilator Bacteriological filters 10 Nos/ Ventilator Humidifier-Automatic servo controlled (MR850 Fisher and Paykel ) with all accessories: 1 No/ Ventilator Probe for Et CO2 measurment 1 no. with 2 reusable adaptor SpO 2 sensor: 1 Test lung9 Adult and Paeds) :1 no. each Air & Oxygen hose : each 1 no. Support arm to hold pt. circuit - 1 no. Auto-clavable exhalation system with flow sensor- 2 no., with 5 yrs. Warranty. In case of Sensor going faulty, the firm should replace it free of cost basis during warranty as well as during CMC. 14. Should have audio-visual alarms along with appropriate message for : Inspiratory pressure, FiO2 (High/Low), Resp. Rate, Tidal volume, minute ventilation, PEEP, O2/Air fail, circuit leak-disconnection, Apnea 15. Alarm History of 1000 alarms
5 16. Should have Automatic Tube Compensation 17. Displayed Trends Values for hours atleast for above parameters. 18. The company should provide Five years comprehensive warranty (including labour and spares) and five years CMC, rates to be quoted separately. Prices of essential/spare itemsaccessories also to be quoted separately. 19. The unit shall be capable of operating continuously in ambient temperature of deg C and relative humidity of 15-90% 20. The unit shall be capable of being stored continuously in ambient temperature of 0-50 C and relative humidity of 15-90% 21. Power input to be VAC, Current: 2 Amps, 50Hz 22. Should be US FDA approved product 23. Company should have service centre in Delhi/ NCR and should provide service 24x Demonstration is a must 25. Certified for meeting IEC Medical electrical equipment - Part 1-4: General requirements for safety - Collateral Standard: Programmable electrical medical systems 26. The service provider should have the necessary equipments recommended by the manufacturer to carry out preventive maintenance test as per guidelines provided in the service/maintenance manual. 27. Must submit user list and performance report within last 5 years from major Central Govt./State Govt./reputed private hospitals 28. List of important spare parts and accessories with their part number and costing. 29. Compliance Report to be submitted in a tabulated and point wise manner clearly mentioning the page/para number of original catalogue/data sheet. Any point, if not substantiated with authenticated catalogue/manual, will not be considered.
6 Schedule No. 35 Tender Code No /2016-LH(S) REVISED SPECIFICATION FOR ICU BED (HIGH END) 1. Bed should have Dual Pedestal Design. 2. Should have built in Bed extender of 200mn for long patients. 3. Tuck away side rails with embedded controls. 4. Electrically operated controls should have lock out facility for electric controls at foot end 5. Backrest & leg section emergency release (CPR) 6. Should have Trendelenbug & covered with bed cover made of AB for the frequent disinfections and washing 7. Synchronized command for bed sprung bas with inclination - backrest 75 degree and above & Femur rest degree 8. Should have electric backrest and pelvic sections with simultaneous movement 9. Should have complying to international standards with acimulated power 1,5 degree 70 VA, electrical current protection class 1, shock protection level type B. 10. Release lever for the side rails with four sectors 11. Protection against water invasion 12. Should have following parameters: a. Length: > 210 cm b. Width: > 85 cm c. Height adjustment: cm d. Back section 0-90 e. Leg Section: 0-25 f. Trendlenberg: g. Reverse Trendlenberg: Should have collapsible /tuck away side rails 14. Should have double X-ray back rest with Cassette tray 15. Should have facility to fix IV rod at all the four corners and place for fixing accessories 16. Should have castors with central braking system and steering facility 17. Should be radiolucent from Neck though Pelvic region 18. Should have bumpers at all 4 corners 19. Should be provided with:
7 a. Bed ends fixed to the frame b. Side rails c. IV rods d. Mattress e. X-ray cassette tray, urine bottle holder, drainage bottle holder 20. Load bearing capacity for the bed should be at least 220kg or more, should have 90 degree upright backrest for chest Imaging. 21. Electrical safety conforms to standards for electrical safety IEC-60601/IS Should have Dual Castors of cm 23. Electric shock protection facility 24. Should be USFDA approved. 25. Zone controls in Bed to track movement of patient on Bed (Optional) 26. The company should provide Five years comprehensive warranty (including labour and spares) and five years CMC, rates to be quoted separately. Prices of essential/spare items-accessories also to be quoted separately. 27. Power input to be VAC, Current: 2 Amps, 50Hz 28. Company should have service centre in Delhi/ NCR and should provide service 24x Demonstration is a must 30. Must submit user list and performance report within last 5 years from major Central Govt./State Govt./reputed private hospitals
8 Schedule No. 36 Tender Code No /2016-LH(S) Revised Specification for Portable Ultrasound System Fully digital, compact portable Colour Ultrasound machine is required with the following Technical features. 1. The unit should be compact, lightweight (Upto 5 Kg: Max) and portable. 2. It should be suitable for vascular access and Nerve block applications (both Superficial and Deep) in both adults and pediatric patients. 3. The unit must have real time compound imaging for improved contrast resolution and eliminating ultrasound artifacts to achieve optimum image quality. 4. The unit must have automatic gain adjustment for imaging. 5. Scanning depth must be available up to 30 cm. 6. System should support transducer technologies like convex, linear, phased array, etc. 7. System must have frequency range from 2 15 MHz (+/- 1 MHz). 8. Imaging modes of Real time 2D, Colour Doppler, Pulsed wave Doppler, Continuous wave Doppler (in all cardiac probes), Power (energy) Doppler must be available. 9. System should have enhanced needle visualization. 10. System must have fast start up to scanning in less than 30 seconds as essential in critical and emergency situation in ICU, emergency, OT. 11. Unit must be sturdy, resistant to breakage & damage on fall/ hit against the wall or hard surface. Certification must be attached. 12. Cine memory on all modes. 13. DICOM ready system with facility to print, store, ready to connect to PACS. 14. Flat LCD/TFT monitor of at least 10 inches with flicker free image. 15. Alphanumeric soft keys keyboard with easy access scans controls, facility to sanitize the system keyboard to avoid cross contamination. 16. Onboard storage of at least images, USB port for connectivity to computer. 17. Should be able to operate both on AC and battery. Battery pack should be self-recharging and should last at least for 90 minwhen fully charged. 18. Transducers to be supplied as standard. a. High Frequency Linear transducer 4-15 ±1 MHz with less then 40 mm size for nerve blocks, vascular access Imaging. Higher frequency will be preferred. b. Convex transducer 2-5 ±1 MHz for deep nerve access e.g. sciatic nerve and abdominal applications.
9 c. High Frequency Linear transducer 6-12 ±1 MHz for nerve block, vascular access, Vascular Imaging, with a small foot print of 25 mm for paediatric applications d. 2-5 MHz Board Band phased array Probe for FAST, cardiac application e. Needle guide must be supplied for linear & convex probe. 19. System should be supplied complete with the following a) B/W Thermal Printer. b) Trolley with 3 universal active port connection and Carry bag of the same company. 20. Power input to be VAC, Current: 2 Amps, 50Hz 21. The unit shall be capable of operating continuously in ambient temperature of deg C and relative humidity of 15-90% 22. The unit shall be capable of being stored continuously in ambient temperature of 0-50 C and relative humidity of 15-90% 23. The equipment should be USFDA and/or European CE approved from notified body. 24. Company should have service centre in Delhi/ NCR and should provide service 24x Demonstration is a must 26. Warranty: The system, transducers and all accessories should be covered with comprehensive warranty for 5 years& 5 yrs comprehensive CMC on expiry of warranty with rates to be quoted separately. 27. Certified for meeting IEC Medical electrical equipment - Part 1-4: General requirements for safety - Collateral Standard: Programmable electrical medical systems 28. The service provider should have the necessary equipments recommended by the manufacturer to carry out preventive maintenance test as per guidelines provided in the service/maintenance manual. 29. Must submit user list and performance report within last 5 years from major Central Govt./State Govt./reputed private hospitals 30. Compliance Report to be submitted in a tabulated and point wise manner clearly mentioning the page/para number of original catalogue/data sheet. Any point,if not substantiated with authenticated catalogue/manual, will not be considered. 31. Hands on Training of the equipment is a must.
10 Schedule No. 37 Tender Code No /2016-LH(S) Revised Specifications for Defibrillator with Monitor and Pacer with AED 1. Defibrillator should be Bi- Phasic, light weight and latest model 2. Should monitor and display vital parameters: 5 lead ECG, NIBP, HR, SPO2 and EtCO2. 3. Should print the ECG on thermal recorders. 4. Should work on Manual and Automated external defibrillation (AED) mode. Should have external pacing facility for adults and pediatric patients 5. Should be capable of doing synchronised & asynchronised cardioversion 6. Can be operated from mains as well as battery 7. Should have defibrillator testing facility 8. Should be a Low Energy Biphasic defibrillator monitor with Recorder, with a maximum energy of 200 Joules or more. 9. Should monitor ECG through paddles, pads and monitoring electrodes and Defibrillate through pads and paddles. Should have Automatic Lead switching to see patient ECG through paddles or leads 10. Should measure and compensate for chest impedance for a range of 25 to 1500hms 11. Should have a built in 50mm strip printer/ thermal recorder 12. Should have charging time of less than 5 seconds for maximum energy. Charging indicator should be there. 13. Should have Display- TFT coloured LED at least 8 diagonal for viewing messages and ECG waveform of 5 seconds 14. Should have facility to monitor & display 5 lead ECG-2 channel with ST segment analysis, Pulse oximetry (Masimo technology) with waveform & Non Invasive BP (Oscillometric technology) and side stream capnometry.
11 15. should have pacing facility with demand and fixed mode with rate and current adjustments 16. Should have external paddles with paddles contact indicator for good paddle contact. Both Adult and pediatric paddles should be available. 17. Should have event summary facility for recording and printing at least 250 events and 50 waveforms. 18. Should have a battery capable of usage for at least 120 minutes and/or 100 discharges. 19. Should be capable of printing Reports on Event summary, configuration, self test, battery capacity etc 20. Should have facility for self test/check before usage and set up function 21. Should be capable of delivering energy in increments of 1-2 joules up to 10J and increments 5-10 J till 50 and up to a maximum of 50J thereafter. 22. Should have user friendly 1,2,3 color coded operation. Should have audible and audiovisual alarm with voice prompt 23. List of accessories: Defibrillator with both adult and Pediatric paddles -01 Patient cable -02 ECG Rolls -50 Disposable pads-10 for adult and 5 for pediatric NIBP Cuff Adult - 02 of each size: Small, Medium, Large and Extra Large NIBP Cuff Paediatrics- 02 NIBP Cuff Infants- 02 SPO2 Finger Probe-Adult -05 SPO2 Ear Probe Spo2 finger probe paediatric 02
12 Complete set of ECG Leads- 02 Side stream EtCO2 module with all accessories. 50 sampling tubes and 10 water traps 24. Company shall provide at least 5 yrs warranty & 5 yrs comprehensive CMC on expiry of warranty with rates to be quoted separately. 25. Demonstration of the equipment is essential. 26. Should be US FDA and European CE approved from notified agency. 27. The unit shall be capable of operating continuously in ambient temperature of 0 50 C and relative humidity of 15-90% 28. The unit shall be capable of being stored continuously in ambient temperature of C and relative humidity of 15-90% 29. Shall meet IEC :2001(Or Equivalent BIS) General Requirements of Safety for Electromagnetic Compatibility. 30. Electrical safety conforms to standards for electrical safety IEC General Requirements and IEC Safety of Electrocardiograms (OR EQUIVALENT BIS Standard) 31. Drop Test-Withstands 0.75 meters or above drop to any edge, corner or surface. 32. Should conform to international test protocols on exposure to shock forces and to vibration forces. The standard should be documented. 33. Should meet IEC 529 Level-2 (IP2X) for enclosure protection solid foreign object ingress. 34. Should meet IEC 529 Level 3 (IP3X) (spraying water) for enclosure protection, water ingress. 35. Should have local service facility.the service provider should have the necessary equipments recommended by the manufacturer to carry out preventive maintenance test as per guidelines provided in the service/maintenance manual
13 36. User Manual and Service manual in English 37. List of important spare parts and accessories with their part number and costing 38. Certificate of calibration and inspection from factory. 39. Log book with instruction for daily, weekly, monthly and quarterly maintenance checklist. 40. The job description of the hospital technician and company service engineer should be clearly spelt out 41. Compliance Report to be submitted in a tabulated and point wise manner clearly mentioning the page/para number of original catalogue/data sheet. Any point,if not substantiated with authenticated catalogue/manual, will not be considered. 42. Must submit user list and performance report within last 5 years from major Central Govt./State Govt./reputed private hospitals
14 Schedule No. 38 Tender Code No /2016-LH(S) Revised Specification for FLUID WARMING DELIVERY SYSTEM Description:- System that warms the infusion and irrigating solutions & guarantee the supply of this precisely heated infusion solution to the patient. System: 1. An ideal Modular type, Microprocessor controlled unit for warming & infusion of intravenous & irrigating solutions. 2. Temperature range: 37 0 C-42 0 C, with Accuracy of +/_ C of desired temp. Point. 3. Temperature (in Fahrenheit or Celsius) should be set & monitored through digital LED Display. 4. The desired temperature should be achieved in less than 5-10 sec. 5. Desirable feature of delivery of high volumes and flows while constantly maintaining the infusion temperature. 6. Should have safety feature of cut-off protection with alarms preferably audio-visual, if temp. exceeds C of set temperature. 7. Should operate on V AC/50 Hz. 8. The system should be provided with 100 disposable tubing kits with each machine 9. Power input to be V AC, Current: 2 Amps, 50Hz 10. The unit shall be capable of operating continuously in ambient temperature of deg C and relative humidity of 15-90% 11. The unit shall be capable of being stored continuously in ambient temperature of 0-50 deg C and relative humidity of 15-90% 12. The equipment should be USFDA and/or European CE approved from notified agency. 13. Company should have service centre in Delhi/ NCR and should provide service 24x Demonstration is a must 15. Company shall provide at least 5 yrs warranty & 5 yrs comprehensive CMC on expiry of warranty with rates to be quoted separately.
15 16. Certified for meeting IEC Medical electrical equipment - Part 1-4: General requirements for safety - Collateral Standard: Programmable electrical medical systems 17. The service provider should have the necessary equipments recommended by the manufacturer to carry out preventive maintenance test as per guidelines provided in the service/maintenance manual. 18. Must submit user list and performance report within last 5 years from major Central Govt./State Govt./reputed private hospitals 19. List of important spare parts and accessories with their part number and costing. 20. Compliance Report to be submitted in a tabulated and point wise manner clearly mentioning the page/para number of original catalogue/data sheet. Any point,if not substantiated with authenticated catalogue/manual, will not be considered.
16 Schedule No. 40 Tender Code No /2016-LH(S) Revised Specifications for Cerebral Oximeter 1. A portable tissue oximeter that uses spatially resolved spectroscopy (Near Infra Red Spectroscopy: NIRS Technology) to determine tissue oxygen saturations in a non-invasive fashion. 2. Monitors must be 4 channels (2+2) capable of monitoring two tissue sites concentrations simultaneously. 3. Can measure regional oxygen saturations in the range of rso Values must be updated at regular intervals typically every 5 seconds or less. 4. Must comply with biomedical safety standards of the USFDA or European CE. 5. Should be applicable to adult, pediatric and neonatal patients in any clinical setting where the brain and body are at risk of reduced-flow or no-flow ischemic states. 6. Accuracy for reading must be within ± 5 rso2 unit of actual. Repeatability must be ± 1-2 rso2 units. 7. It should give % Arterial and % Venous values 8. It should have analytical tool 9. It should be supplied with 75 adult sensors and 25 pediatric sensors, the cost of which to be quoted separately. 10. A trending memory of 24 hours must be available. 11. Battery backup at least 12VDC; approximately 20 minutes 12. Should have USB Port USB 2.0 Flash Memory 13. Should be provided with appropriate accessories and sensors.
17 14. Equipment must work on V 50 Hz AC 15. Company shall provide at least 5 yrs warranty & 5 yrs comprehensive CMC on expiry of warranty with rates to be quoted separately. 16. Demonstration of the equipment is essential. 17. Should be US FDA/ European CE approved from notified agency. 18. The unit shall be capable of operating continuously in ambient temperature of 0 50 C and relative humidity of 15-90% 19. The unit shall be capable of being stored continuously in ambient temperature of C and relative humidity of 15-90% 20. Shall meet IEC :2001(Or Equivalent BIS) General Requirements of Safety for Electromagnetic Compatibility. 21. Electrical safety conforms to standards for electrical safety IEC General Requirements and IEC Safety of Electrocardiograms (OR EQUIVALENT BIS Standard) 22. Should conform to international test protocols on exposure to shock forces and to vibration forces. The standard should be documented. 23. Should meet IEC 529 Level-2 (IP2X) for enclosure protection solid foreign object ingress. 24. Should meet IEC 529 Level 3 (IP3X)(spraying water) for enclosure protection, water ingress. 25. Should have local service facility.the service provider should have the necessary equipments recommended by the manufacturer to carry out preventive maintenance test as per guidelines provided in the service/maintenance manual. 26. User Manual and Service manual in English 27. List of important spare parts and accessories with their part number and costing 28. Certificate of calibration and inspection from factory. 29. Log book with instruction for daily, weekly, monthly and quarterly maintenance checklist.
18 30. The job description of the hospital technician and company service engineer should be clearly spelt out 31. Compliance Report to be submitted in a tabulated and point wise manner clearly mentioning the page/para number of original catalogue/data sheet.any point,if not substantiated with authenticated catalogue/manual, will not be considered. 32. Must submit user list and performance report within last 5 years from major Central Govt./State Govt./reputed private hospitals
19 Schedule No. 41 Tender Code No /2016-LH(S) Revised TECHNICAL SPECIFICATION FOR Anesthesia Workstation 1. Anesthesia workstation with integrated ventilator, and monitoring facility. 2. S.S top, shelves and drawers to be provided. 3. Front wheels having foot activated brakes. 4. Microprocessor controlled electronic fresh gas delivery gas system with electronic blender for oxygen, nitrous oxide and air. 5. Digital and virtual flow meter for oxygen, nitrous oxide and air with TFT display screen of minimum size 12 inches. 6. Single yokes for oxygen, nitrous oxide and compressed air, to have non interchangeable gas supply inlet- for oxygen, N2O and air with DISS system for central pipeline with separate pipeline adaptors for the same. 7. Electronic display for cylinder and pipeline pressure of oxygen, N2O and compressed air. 8. Safety features: a) Automatic cut off of nitrous oxide in the absence of oxygen supply. b) Power failure alarms c) Oxygen failure alarm both visual and audible d) Inbuilt hypoxic guard to ensure minimum 25% oxygen at all oxygen and nitrous oxide mixture. e) Facility for switching to manual in event of electronic failure in the anaesthesia machine. f) Separate FGF outlet for attachment of anesthesia circuit for manual / spontaneous ventilation. g) Illumination at the working surface area. 9. Emergency oxygen flow of at least 35 lit/min. with non-lockable push button to be provided. 10. Twin vaporizer manifold with inter-locking facility. 11. Supplied with halothane, isoflurane, desflurane and sevofluorane vaporizers should be temperature, pressure and flow compensated with key filling arrangement and should be quick mountable and allow operation of only one vaporizer at a time. 12. Vaporizer, which allows easy exchange between agents.
20 13. Agent capacity should be minimum 225ml. of free volatile anesthetic agent. 14. Should be fitted with double/single chamber type of CO2 adsorber system. Provision to bypass absorber system through a switch and ability to ventilate with bag; with condensate drain and provision for scavenging of waste anesthesia gases. 15. Bag to ventilator switch should be Bi-stable. 16. Should have anesthesia ventilator. 17. Electrically operated pneumatically or electrically driven anesthesia ventilator, single bellow for adult and pediatric patient, integrated with anesthesia system a) Controls for tidal volume, respiratory rate and inspiratory flow with facility of monitoring and display of minute volume, I:E ratio, Expiratory TV, Inspiratory TV etc b) Monitoring of tidal volume, minute volume and FIO2%. Tidal vol. range ml. electronic peep range 0-20 mbar, respiratory wave from display and able to adjust I:E ratio, plateau time and wave form type. c) Back lit LCD display with visual messages. d) Alarms for high and low Fio2, disconnection, high and low system/airway pressure e) Facility of adding inspiratory pause of 25% of inspiratory time. f) Internal battery backup of 20 min or more g) Ventilator should be mounted on anesthesia machine h) Ventilators mode- spontaneous, manual cycling, pressure support, volume control, pressure, SIMV. 18. INTEGRATED MONITOR. a) 8 channels monitor not less than 15 inches colored TFT touch screen. User adjustable color of parameters and parameter display. b) Two channel ECG display ECG 3 lead system and ECG 5 lead system with ST segment and arrhythmia analysis. Capable to monitor 12 lead ECG. c) Spo2 measurement masimo technology, adult, pediatric and neonatal probes with ear probe. d) Built-in NIBP with adult, pediatric and neonatal cuffs with interval adjustment- stat, and an interval of 1- minute up to 10 min. in one minute interval. e) Should be able to monitor 2 IBP invasive pressure Two temperature probes: skin and rectal End tidal CO 2 side stream technology
21 Anaesthesia agent monitoring and display on monitor or anaesthesia machine Display of oxygen, nitrous oxide and CO2 and anaesthetic agent in inspired and expired gases on monitor or anaesthesia machine Integrated Neuromuscular monitoring Entropy or BIS Audio visual alarms for all parameters with provision for alarm volume and colour adjustment 19. Inbuilt thermal printer with data transfer facility. 20. Trends of 24 hours of real time trends and patient information at the bedside as standard. Both numeric and graphic display of all parameters. 21. Should have monitoring of 5 anaesthetic agents, N2O, CO2, both inspiratory and expiratory valves. 22. Both pressure and capnography curves with MAC value. Should provide ventilation parameters and measurement should be integrated. 23. Provision of up-gradation with compatible, central monitoring station using LAN card. 24. Each machine should be provided with the following: a) Adult NIBP cuff: 2 each of medium, large and extra large size b) Pediatric and Neonatal NIBP Cuff: 2 each c) EtCO 2 complete gas sampling kit with condenser: 100 d) Pressure transducer cables: 2 e) Pressure transducer domes: 100 f) Pressure transducer kit for mounting of domes: 4 g) Temperature Probes- 2 nasopharyngeal, 2 rectal and 2 mid-esophageal each for adults, pediatric and neonates h) Sensors/ Electrodes for Entropy or BIS: 100 for pediatric and 100 for adults i) Reusable and autoclavable silicone ventilator circuits for adult and pediatric cases: 2 each j) Thermal Printer Paper: 50 rolls 25. Supplier must guarantee and ensure a minimum of 10 years of supply of spares and consumables specific to the workstation as whole. 26. The unit shall be capable of operating continuously in ambient temperature of 0-40deg C and relative humidity of 15-90%
22 27. The unit shall be capable of being stored continuously in ambient temperature of -20 to 60 deg C and relative humidity of 15-90% 28. Power input to be VAC, 50Hz fitted with Indian plug 29. Should be US FDA and/or European CE from notified agency approved product. 30. Shall meet the safety requirements as per IEC :1994 Medical electrical equipment Part 2: Particular requirements for the safety of electrocardiographic monitoring equipment. 31. Shall meet IEC :2001(Or Equivalent BIS) General Requirements of Safety for Electromagnetic Compatibility. or should comply with 89/366/EEC; EMC-directive 32. The service provider should have the necessary equipments recommended by the manufacturer to carry out preventive maintenance test as per guidelines provided in the service/maintenance manual. 33. User Manual in English and Service manual in English 34. Must submit user list and performance report within last 5 years from major Central Govt./State Govt./reputed private hospitals 35. List of important spare parts and accessories with their part number and costing. 36. Compliance Report to be submitted in a tabulated and point wise manner clearly mentioning the page/para number of original catalogue/data sheet. Any point, if not substantiated with authenticated catalogue/manual, will not be considered. 37. List of Equipments available for providing calibration and routine Preventive Maintenance Support as per manufacturer documentation in service/technical manual. 38. Log book with instruction for daily, weekly, monthly and quarterly maintenance checklist. The job description of the hospital technician and company service engineer should be clearly spelt out 39. Company should have service centre in Delhi/ NCR and should provide service 24x Demonstration is a must 41. Hands- on training of medical and paramedical staff after the equipments are installed. Company shall provide at least 5 yrs warranty & 5 yrs comprehensive CMC on expiry of warranty with rates to be quoted separately. The payments of CMC will be made on yearly basis after successful completion of contract. The rates quoted for CMC will be included for ranking/ comparison purpose.
23 Schedule No. 42 Tender Code No /2016-LH(S) Revised Specifications for Multiparameter Monitors for ICU/ Recovery 1. It should provide complete monitoring solution to meet the requirement of wide spectrum monitoring needs of post operative & critically ill patient 2. ICU monitors are placed at the bedside and with central station. 3. Capability of storage of patient data and printing of patient reports. 4. Minimum 17 inches multicoloured LCD display screen operation through both touch screen with or without knob. 5. Should have facility for enlarge numeric display format and multiple layout of screen. 6. Separate CPU/Module trolly. 7. Six digital and waveforms/traces display 8. Combination of single, dual and multiparameter modules. 9. Parameter modules freely exchangeable between all the monitors. 10. Multichannel (upto 12 leads) ST segment analysis. 11. Facility to monitor and display - ECG, Respiration, NIBP, SpO2 (Masimo technology), CO2 with side stream capnography, Temp, Cardiac output, NMT, BIS, EEG & dual IBP. 12. Minimal invasive/ invasive continuous cardiac output. 13. Should be able to measure anesthetic agents (N20 and inhalational agents) including their inspiratory and expiratory values. 14. Should display perfusion index (PI%) from SpO2 as an indication of pulse strength at the sensor site. 15. Automatic arrhythmia detection & alarm for standard and lethal arrhythmia. 16. NMT Module (Integrated/ Stand alone): For measurement and display of TOF count, TOF %, ST, DBS, Tetanic and Trend for continuous usage. Automatic measurement facility in selected time interval. Automatic selection of supramaximal current. Include standard accessories 17. EEG Module with all accessories. 18. Should have a central station for bedside monitors with independent controlls. 22" and above multi colour LCD Monitor, complete with Ethernet LAN cabling, alarm management, 72 hours trending, bed to bed viewing of waveforms and remote alarm management like silencing of alarms etc.
24 19. Should provide hemodynamic, oxygenation, Ventilation calculation package. 20. Should have drug calculation package. 21. Trend of at least 48 hours nos. event recall/snapshot facility both manually and automatically triggered by alarm. 23. Automatic Zoom in Facility in the monitor display. 24. It should have monitor to overview facility and data transfer over the network. 25. Web browsing facility to review each networked monitors data through hospital LAN via office PC in Hospital LAN Network and/or through dial up facility from remote location(optional) 26. Battery / UPS backup of upto 30 min. when fully charged. 27. Facility for HIS Communications with Information Management Systems Include Laser Printer in the central monitor and dual channel strip chart recorder in each monitor. 28. Accessories per monitor should include: a) ECG/Resp: 5 Lead ECG Cable with clip- 2 sets and 10 Lead ECG Cable with clip- 2 set. b) NIBP: Adult cuff (Large)- 5, Adult Cuff (X-tra Large)- 5, paediatric cuffs (small)- 5, Pediatric Cuff (Large)- 5, neonatal cuffs: 5, NIBP tubings: 2 c) SpO2: Adult SpO2 sensor with cable- 2, Paediatric SpO2 sensors- 2, Neonatal SpO2 sensors: 2. d) IBP: Include four nos. of reusable pressure transducer cable with 1 bracket, holder and 100 nos of disposable domes kit per monitor. e) Temperature: Nasopharyngeal Temperature Probe: 2 Adult and 2 Pediatric, Rectal temperature probe-2 Adult and 2 Pediatric, skin temperature probe- 2. f) Side stream EtCO2 module with all accessories. 50 sampling tubes and 10 water traps g) EEG Modules- with all accessories. Should display at least two channels. h) BIS Module: Adult Sensors-100 numbers, Pediatric Sensors- 50 numbers. Spectral analysis modules by compressed spectral array. 29. Necessary cabling for networking the monitors on turnkey basis. 30. Necessary mounting solution for monitors
25 31. The unit shall be capable of operating continuously in ambient temperature of 10-40deg C and relative humidity of 15-90% 32. The unit shall be capable of being stored continuously in ambient temperature of -20 to 60 deg C and relative humidity of 15-90% 33. Shall meet IEC :2001(Or Equivalent BIS) General Requirements of Safety for Electromagnetic Compatibility. or should comply with 89/366/EEC; EMC-directive. 34. Power input to be VAC, 50Hz fitted with Indian plug 35. Should be US FDA and European CE approved from notified body. 36. Shall meet the safety requirements as per IEC :1994 Medical electrical equipment Part 2: Particular requirements for the safety of electrocardiographic monitoring equipment. 37. The service provider should have the necessary equipments recommended by the manufacturer to carry out preventive maintenance test as per guidelines provided in the service/maintenance manual. 38. Back to back warranty to be taken by the supplier from the principal to supply spares for a minimum period 10 years. 39. User Manual in English and Service manual in English 40. Must submit user list and performance report within last 5 years from major Central Govt./State Govt./reputed private hospitals 41. List of important spare parts and accessories with their part number and costing. 42. Compliance Report to be submitted in a tabulated and point wise manner clearly mentioning the page/para number of original catalogue/data sheet. Any point,if not substantiated with authenticated catalogue/manual, will not be considered. 43. List of Equipments available for providing calibration and routine Preventive Maintenance Support. as per manufacturer documentation in service/technical manual. 44. Log book with instruction for daily, weekly, monthly and quarterly maintenance checklist. The job description of the hospital technician and company service engineer should be clearly spelt out 45. Company should have service centre in Delhi/ NCR and should provide service 24x Demonstration is a must 47. Company shall provide at least 5 yrs warranty & 5 yrs comprehensive CMC on expiry of warranty with rates to be quoted separately.
Technical Data and Specifications
Technical Data and Specifications INTENDED USE Ventilator designed to provide Invasive and Non-invasive ventilation for the critical care management of adult, pediatric and neonate-infant (including premature)
More informationSATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products
Saturn Evo SATURN EVO COLOR A N A E S T H E S I A S Y S T E M i n t e l l i g e n t s i m p l i c i t y Manufacturer of high quality anaesthesia products SATURN EVO COLOR PATIENT SAFETY AND USER PATIENT
More informationA9600 Anesthesia Workstation. Your healthcare, we care.
A9600 Anesthesia Workstation Your healthcare, we care. A9600 is an advanced yet easy to use anesthesia workstation that provides accurate, pneumatically driven and electronically controlled ventilation.
More informationPanther 5 Acute Care Ventilator
1 HIGHLIGHTS High performance and advanced features target the ICU environment Utilizes an internal blower with a specially designed flow control valve removing the need for using compressed air without
More informationINTENSIVE CARE VENTILATORS
INTENSIVE CARE VENTILATORS iternis ADV INTENSIVE CARE VENTILATION FOR ALL PATIENT CATEGORIES Design Features HEYER iternis ADV s 12-inch touch-screen display combines comprehensive, easy-to-follow user
More informationMilestones of Aeonmed Ventilators
Aeonmed Development History Milestones of Aeonmed s 2003 More than 1000 Aeonmed ventilators contributed to save lives during SARS 2008 Aeonmed nominated emergency ventilator supplier for 2008 Beijing Olympic
More informationThe anaesthesia workstation focused on you
by The anaesthesia workstation focused on you With the most advanced anaesthesia Accuracy and performance Equipped with the most advanced ventilation modes. Electronic rotameters with precise mass-flow
More informationSafe Accurate Reliable. CWM-301 Anesthesia System
Safe Accurate Reliable CWM-301 Anesthesia System CWM-301 Anesthesia System Extraordinary performance points 8.4 inch colour LCD screen Compact Anesthesia ventilation system with the standard of ISO 9001:13485
More informationAPUS Anesthesia Machines
APUS Anesthesia Machines APUS x1 Anesthesia Machine Breathing circuit applicable for complete range Autoclavable metal and plastic materials. Suitable for adult, pediatric and neonatal patients. Featured
More informationState-of-the-art anesthesia technology
State-of-the-art anesthesia technology anesthesia system Get closer to the MRI scanner Superb certified performance in close proximity to strong magnetic fields. Our innovative machine provides benefits
More informationThe centerpiece of a comprehensive anaesthesia workstation
The centerpiece of a comprehensive anaesthesia workstation IntelliSave AX700 anaesthesia machine A slim design with rich functionality The IntelliSave AX700 is a comprehensive anaesthesia machine from
More informationServo UNIVERSAL ADVANCED VENTILATION PLATFORM ALL PATIENT CATEGORIES
DATA ADVANCED VENTILATION PLATFORM ALL PATIENT CATEGORIES CRITICAL CARE HIGHLIGHTS! For all patient categories, from neonate and pediatric to adult! A platform for the future choose the level of system
More informationThe anaesthesia workstation focused on you
by The anaesthesia workstation focused on you The anaesthesia Accuracy and performance Equipped with the most advanced ventilation modes. Electronic rotameters with precise mass-flow sensors, allowing
More informationAnesthesia Workstation
Your healthcare, we care. A9800 Anesthesia Workstation Technical Parameters A9800 Ventilation Modes Alarms Volume Control Pressure Control SIMV with Pressure Support (SIMV + PS) - with Volume or Pressure-type
More informationSTANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA
STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA SOP#: 1157.1 Date Issued: 05/14 Date Revised: 5/15 Page 1 of 6 TITLE: SCOPE: RESPONSIBILITY: PURPOSE: SurgiVet
More informationDATA SHEET. VENTILATION SERVO-s
DATA SHEET VENTILATION The Gold Standard Critical Care 3 PROVEN SERVO SIMPLICITY, RELIABILITY AND SENSITIVITY MAQUET THE GOLD STANDARD Leading the way: MAQUET is a premier international provider of medical
More informationNo./DMER/Tender Notice/Equip.Pur/14-15/Corri./7M Date :- 20/09/2014
No./DMER/Tender Notice/Equip.Pur/14-15/Corri./7M Date :- 20/09/2014 E-Tender Notice CORRIGENDUM REGARDING SPECIFICATIONS FOR 1. EQ-C-66 Bubble C PAP 2. EQ-C-68 Neonatal High frequency ventilator 3. EQ-C-70
More informationYour patients, our primary focus. Siesta i Whispa anesthesia machine
Your patients, our primary focus Siesta i Whispa anesthesia machine Functional, and flexible At Dameca we are not just committed to meeting the healthcare needs of today. We also think about tomorrow.
More informationAespire 7100 Essential performance Compact design
GE Healthcare Aespire 7100 Essential performance Compact design Features Enhanced monitor integration capabilities with our Datex-Ohmeda Anesthesia Monitor and Compact Anesthesia monitor Lightweight and
More informationAnnexure-2 1. The technical Specification of Schedule 1 item no. 19, Medical mobile x-ray machine should be read as follows:- The mobile x ray
Annexure-2 1. The technical Specification of Schedule 1 item no. 19, Medical mobile x-ray machine should be read as follows:- The mobile x ray equipment would be required to perform x ray studies in emergency
More informationVENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV
VENTIlogic LS VENTIlogic plus 100 % Mobility and Reliability in IV and NIV VENTIlogic LS VENTIlogic plus Your requirements for reliability and mobility are our benchmark. VENTIlogic LS and VENTIlogic plus
More informationAirox Supportair Ventilator
Airox Supportair Ventilator A versatile hospital ventilator with a titration platform that enables a smooth transition to the patient s home care ventilator Titration Platform The Supportair ventilator
More informationFabius Tiro TM. Designed for your environment. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care
Fabius Tiro TM Designed for your environment Emergency Care Perioperative Care Critical Care Perinatal Care Home Care Traditional hospital ORs have their needs... With more surgical procedures being performed
More informationV8600 Ventilator. Integrated Invasive & Noninvasive Ventilation
V8600 Ventilator Integrated Invasive & Noninvasive Ventilation 0123 V8600 Ventilator Invasive Ventilation With state-of-the-art turbine technology, V8600 helps achieve sequential ventilation in various
More informationVENTILATION SERVO-s EASY AND RELIABLE PATIENT CARE
VENTILATION SERVO-s EASY AND RELIABLE PATIENT CARE Critical Care SERVO-s 3 SERVO-s SIMPLY MAKES SENSE MAQUET THE GOLD STANDARD Leading the way: MAQUET is a premier international provider of medical products
More informationPERSEO EVOLUTION. workstation for gaseous anaesthesia. for adults, children and newborns. Code: OM 403 / 3 gas module 3 rotameter
ESSE 3 Via Garibaldi 30 14022Castelnuovo D.B. (AT) tel +39 011 99 27 706 fax +39 011 99 27 506 e-mail esse3@chierinet.it web : www.4ci.com PERSEO EVOLUTION workstation for gaseous anaesthesia for adults,
More informationV8800 Ventilator System. Your healthcare, we care.
V8800 Ventilator System Your healthcare, we care. V8800 Ventilator System V8800 is a critical care ventilator for infants, children and adults. It has comprehensive functionality, a user-friendly design,
More informationNarkomed MRI-2. A perfect fit for the MRI environment. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care
Narkomed MRI-2 A perfect fit for the MRI environment Emergency Care Perioperative Care Critical Care Perinatal Care Home Care Because you care The complete anesthesia solution for the MRI environment A
More informationSLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters
SLE4000 Infant Ventilator with touch-screen operation When the smallest thing matters SLE4000 - The Total Solution for Conventional Infant Ventilation SLE is a world leader in the design and manufacture
More informationAPUS Anesthesia Machines
APUS Anesthesia Machines Adult and pediatrics applicable breathing circuit Autoclavable metal and plastic materials. Featured with stable pressuretightness. Visual observation of bellows and valves. Breathing
More informationMka- jke euksgj yksfg;k vlirky] ubz fnyyh
nwjhkk"k : 23365525 Mka- jk-e-yks-v- 27¼v½ Tele No. : 23365525 D.R.M.L.H. - 27 (A) Hkkjr ljdkj@government OF INDIA Mka- jke euksgj yksfg;k vlirky] ubz fnyyh DR. RAM MANOHAR LOHIA HOSPITAL, NEW DELHI (PROCUREMENT
More informationOperation and Maintenance of the EPV200 Portable Ventilator
Operation and Maintenance of the EPV200 Portable Ventilator 1 Applications of the EPV200 The EPV200 Portable Ventilator is a gas powered electronically controlled mechanical ventilator, designed to provide
More informationYour patients, our primary focus
Your patients, our primary focus Siesta i Whispa anesthesia machine Functional, fail-safe and flexible to your needs Take a good look, and you ll be surprised by the many forward-thinking features of the
More informationAmended Technical Specifications BLOOD GAS ANALYZER
BLOOD GAS ANALYZER 1. It should measure Blood Gas (full parameters) in its addition to measure Electrolytes like Na+, K+, Cl-, ph, po2,pco2. and Haematocrit. 2. Calculated parameters: TCO2, HCO3, Base
More informationSLE5000 Infant Ventilator with HFO
SLE5000 Infant Ventilator with HFO When the smallest thing matters SLE5000 - The Total Solution for Infant Ventilation Shown on optional roll stand. Ventilator may be mounted either way round on stand.
More informationDameca MRI 508. State-of-the-art MRI technology
Dameca MRI 508 State-of-the-art MRI technology The closer, the safer the better Hospitals have long awaited an anaesthesia machine that can be used close to their MRI scanner. The Dameca MRI 508 is it
More informationGE Healthcare. Centiva/5 Critical Care Ventilator. Meet a new level of expectations
GE Healthcare Centiva/5 Critical Care Ventilator Meet a new level of expectations Intergrating performance and value Marry form and function In an environment where time and space are tight, the design
More informationVENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV
VENTIlogic LS VENTIlogic plus 100 % Mobility and Reliability in IV and NIV VENTIlogic VENTIlog LS Your Requirements for Reliability and Mobility are Our Benchmark. VENTIlogic LS and VENTIlogic plus are
More informationFlight Medical presents the F60
Flight Medical presents the F60 Reliable Ventilation Across the Spectrum of Care Adult & Pediatric Pressure/Volume Control Basic/Advanced Modes Invasive/NIV High Pressure/Low Flow O2 Up to 12 hours batteries
More informationfabian HFO: Modular, Powerful, Versatile.
fabian HFO: Modular, Powerful, Versatile. Introducing the flagship of the ACUTRONIC FABIAN FAMILY. The FABIAN HFO, a ventilator developed specifically for the NICU and PICU needs, packed with many great
More informationOxylog 3000 plus Emergency & Transport Ventilation
Oxylog 3000 plus Emergency & Transport Ventilation Offering high ventilation performance with features such as AutoFlow, integrated capnography and non-invasive Ventilation, the compact and robust Oxylog
More informationSERVO-s v8.0 Data sheet
Data sheet This document is intended to provide information to an international audience outside of the US. 2 Critical Care Key features For adult and pediatric patients A ventilator system designed to
More informationAn intuitive and innovative answer to your anaesthesia challenges
An intuitive and innovative answer to your anaesthesia challenges Dameca AX500 anaesthesia machine Meeting your needs today and tomorrow As an expert in anaesthesia care, you are looking to continuously
More informationDepartment of Paediatrics
KASTURBA HEALTH SOCIETY'S MAHATMA GANDHI INSTITUTE OF MEDICAL SCIENCES, P.O. SEVAGRAM, WARDHA (DISTT.) MAHARASHTRA STATE - 442 102 Specifications for Equipment Department of Paediatrics 1. A) Ventilator
More informationAUTOVENT 4000 VENTILATOR
OVERVIEW AUTOVENT 4000 Only properly trained and approved Escambia County Bureau of Public Safety Paramedics are to use the AutoVent 4000 ventilator manufactured by LSP to transport patients already on
More informationNo./DMER/Tender Notice/Equip.Pur/16-17/Corri.(10)/7M Date :- 03/09/2016
No./DMER/Tender Notice/Equip.Pur/16-17/Corri.(10)/7M Date :- 03/09/2016 E-Tender Notice CORRIGENDUM REGARDING SPECIFICATIONS FOR 1. EQ-C-256/16-17 SyringePump/Infusion Pump 2. EQ-C-257/16-17 Video Gastroscope
More informationYour customized anesthesia workstation DRÄGER FABIUS GS PREMIUM. Dräger. Technology for Life
D-9355-2009 Your customized anesthesia workstation DRÄGER FABIUS GS PREMIUM Dräger. Technology for Life 02 DRÄGER FABIUS GS PREMIUM One of the most challenging places D-9354-2009 on earth The modern hospital
More information2015 Hamilton Medical AG. All rights reserved. Printed in Switzerland.
HAMILTON-T1 Quick Guide This Quick Guide is intended as a useful reference for ventilation of adult and pediatric patients. It does not replace the clinical judgment of a physician or the content of the
More informationfabian Therapy evolution Our 2-in-1 device
fabian Therapy evolution Our 2-in-1 device Ventilation Beyond Limits When It Comes to Life, There Is No Compromise Distributing worldwide ACUTRONIC is a privately held company that develops, manufactures,
More informationfabian HFO Our 4-in-1 device
fabian HFO Our 4-in-1 device Ventilation Beyond Limits When It Comes to Life, There Is No Compromise Distributing worldwide ACUTRONIC is a privately held company that develops, manufactures, and distributes
More informationFabius Tiro TM. Designed for your environment. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care
Fabius Tiro TM Designed for your environment Emergency Care Perioperative Care Critical Care Perinatal Care Home Care Because you care Traditional hospital ORs have their needs... With more surgical procedures
More informationGE Healthcare. Aestiva/5. More than superior ventilation
GE Healthcare Aestiva/5 More than superior ventilation A balance of superior ventilation and award-winning design SmartVent the shape of superior ventilation At the heart of the Datex-Ohmeda Aestiva /5
More informationAnesthesia Solutions. Prima SP and Prima SP2 Anesthesia Systems. The Future of Anesthesia. Partnership for Life
Anesthesia Solutions The Future of Anesthesia Prima SP and Prima SP2 Anesthesia Systems The Penlon Anesthesia System Range provides the user with their choice of advanced, easy to use anesthesia systems
More informationIEC C LASSIFICATIONS:
Enrich life Introducing Astral Introducing the Astral 100 and Astral 150 life support ventilators. From initial setup to everyday use, Astral offers greater freedom, confident care and designed efficiency
More informationMEDIANA HEARTON AED A15. Mediana Automated External Defibrillator A15
AED A5 Mediana Automated External Defibrillator A5 AED A5 COVER Protected icon sheet, shock button, adult/paediatric mode switch. 7 BATTERY Disposable LiMnO2 Non-Rechargeable. 2yrs shelf life, 5yrs standby
More informationSiesta i Whispa Integrated technology
Siesta i Whispa Integrated technology Functional, fail-safe and flexible to your needs Take a good look, and you ll be surprised by the many forward-thinking features of the Siesta i Whispa anaesthesia
More informationSLE5000 Neonatal Ventilator with High Frequency Oscillation
SLE5000 Neonatal Ventilator with High Frequency Oscillation The SLE5000 is superb: the best High Frequency Ventilator I ve used When the smallest thing matters SLE5000 - The Total Solution for Infant Ventilation
More informationINTENDED USE CLASSIFICATION PHYSICAL CHARACTERISTICS
INTENDED USE The GraphNet ts ventilator was designed to be used with all types of patients: adult, pediatric, and neonatal infants (optionally with the NEO-INF module) requiring invasive and noninvasive
More informationAUGUST ANESTHESIA MACHINES MEDICAL PRODUCTS ANESTHESIA MACHINES. Quality System Standard applied: ISO 9001, ISO 13485
ANESTHESIA MACHINES AUGUST - 2008 MEDICAL PRODUCTS ANESTHESIA MACHINES Quality System Standard applied: ISO 9001, ISO 13485 CE Certification 93/42/EEC. Arsiconsult 38, Av. Guy de Maupassant 78400 Chatou
More information6 th Accredited Advanced Mechanical Ventilation Course for Anesthesiologists. Course Test Results for the accreditation of the acquired knowledge
6 th Accredited Advanced Mechanical Ventilation Course for Anesthesiologists Course Test Results for the accreditation of the acquired knowledge Q. Concerning the mechanics of the newborn s respiratory
More informationHallowell EMC Anesthesia WorkStation Operating Manual
Hallowell EMC Anesthesia WorkStation Operating Manual PN 000a2770 Anesthesia WorkStation A compact (9" W x 9" D x 15" H), convenient system for use on animals under 7 kg (15 lb.) Hallowell Engineering
More informationEmergency Transport and Ventilation
Emergency Transport and Ventilation When you get a call and the patient is not breathing, are you and your equipment ready? Can you make the difference? Pneupac portable gas powered ventilators (PGPVs)
More informationVT900A + VAPOR. Technical data. One Solution. Auto-detection to Ensure Patient Safety. Key benefits and features
Technical data VT900A + VAPOR The Fluke Biomedical VT900A + VAPOR is designed to efficiently and reliably perform a full anesthesia machine PM, from ventilators to vaporizers. One Solution VT900A + VAPOR
More informationOperation and Maintenance of the MCV100 and MCV100-B Portable Ventilator
Operation and Maintenance of the MCV100 and MCV100-B Portable Ventilator 1 Applications of the MCV100(B) The MCV100 and MCV100-B Mass Casualty Ventilators are electronically controlled portable ventilators,
More informationPERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:
PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: 1. **Identify and name the filters on the 7200ae. 2. **Explain how each filter is sterilized. 3. **Trace the gas flow through the ventilator
More informationCarestation 650. Physical Specifications Product Description Carestation 620 A1 ANESTESI CARESTATION 650
Carestation 650 The Carestation 620 is a compact, versatile and easy to use anesthesia system designed to help clinicians deliver reliable anesthesia care to solve today s toughest challenges. Key Features
More informationHospital and Transport for Controlled Breathing
Hospital and Transport for led Breathing When transporting a critically ill patient you need a ventilator that can go anywhere in any situation. Smiths Medical Pneupac small portable gas powered ventilators
More informationEVE. The mobile Ventilation Concept from Emergency to Intensive Care. easy - light - smart. Ventilation
Ventilation EVE easy - light - smart The mobile Ventilation Concept from Emergency to Intensive Care + Ventilation at the push of a button + Cutting edge turbine technology + Critical care during transport
More informationAnnex (A) List and technical specifications of items
Annex (A) List and technical specifications of items Item LOT# A) ICU Bed LOT A B) Monitoring system LOT A C) Defibrillator LOT A D) Shower Trolley LOT A E) Crash Cart LOT A F) Portable Ventilator LOT
More informationThe Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)
The Crossvent 2i+ 1. How is this ventilator classified 2. Ventilator Concept (brief theory of operation and features) -Your Two Choices with this Ventilator 3. An overview of the device (an in-service)
More informationNEONATAL/INFANT VENTILATOR
n a!! i le ion b a a i l ve r s v w a lone o N d-a n sta FEATURES: MDES F VENTILATIN PRESSURE LIMITED/ CNTINUUS FLW FLW TRIGGERED VLUME VENTILATIN NENATAL/INFANT VENTILATR The Crossvent 2i+ ventilator
More informationFabius plus. The plus that makes all the difference. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care
Fabius plus The plus that makes all the difference Emergency Care Perioperative Care Critical Care Perinatal Care Home Care More features, more options add up to more advantages Economical To meet today
More informationVENTILATION SERVO-i INSPIRATION IN EVERY BREATH
VENTILATION SERVO-i INSPIRATION IN EVERY BREATH The Gold Standard Critical Care SERVO-i 3 PROVIDING THE HIGHEST LEVEL OF CLINICAL CARE MAQUET THE GOLD STANDARD Leading the way: MAQUET is a premier international
More informationS/5 Aespire. Quick Reference Guide
S/5 Aespire Quick Reference Guide Anesthesia system controls 1 Turn on the system 2 7100 controls and monitored data 3 Menu map 4 Set the ventilation mode 5 Set ventilator controls 6 Vaporizer controls
More informationO-Two Self-Study Guide. e600 Transport Ventilator Ventilation Modes
O-Two Self-Study Guide e600 Transport Ventilator Ventilation Modes e600 VENTILATION MODES The e600 ventilator has 13 ventilation modes: Manual Ventilation, Controlled Mandatory Ventilation (CMV), Assist
More informationWarranty The device shall have a 6-year warranty at minimum
Bid Specifications Defibrillator The AED must have a high-resolution liquid crystal display with capacitive touch panel. The AED must have an ON/OFF button. The AED must have a SHOCK button that illuminates
More informationRecommendations on Checking Anaesthesia Delivery Systems
Page 1 of 11 Recommendations on Checking Anaesthesia Delivery Version Effective Date 1 Oct 1992 (reviewed Feb 07, Feb 02) 2 2004 3 Nov 2011 4 Dec 2016 Document No. HKCA T1 v4 Prepared by College Guidelines
More informationekom Compressor for Medical Air Sonata with Ekom
Compressor for Medical Air The compressor A40 is an ideal accessory for the intensive ventilators elisa and sonata. The A40 can be used either as the main or as a back-up gas supply. In the event of a
More informationFabius plus. The plus that makes all the difference. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care
Fabius plus The plus that makes all the difference Emergency Care Perioperative Care Critical Care Perinatal Care Home Care Because you care More features, more options add up to more advantages Economical
More informationWorking version subject to Departmental approval. Drafted by Dr. Olympio and reviewed by clinical members of The Capital Equipment Committee
Wake Forest University Baptist Medical Center Department of Anesthesiology TECHNICIAN ABBREVIATED CHECKLIST OF RESPONSIBILITIES For Anesthesia Apparatus Checkout Specific to the Draeger APOLLO anesthesia
More informationHONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS
RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS 1. INTRODUCTION An anaesthesia delivery system includes any machine, equipment or apparatus which supplies gases, vapours, local anaesthesia and/or
More informationIMPACT. Uni-Vent. Eagle. Model 754 Portable Ventilator
IMPACT Uni-Vent Eagle Model 754 Portable Ventilator Uni-Vent Eagle Model 754 Portable Ventilator OPERATOR TRAINING PROGRAM FOR: Critical Care Transport Aero medicine Emergency Services Uni-Vent Eagle Model
More informationAutomatic Transport Ventilator
Automatic Transport Ventilator David M. Landsberg, MD, FACP, FCCP, EMT-P Luke J. Gasowski, RRT, NPS, ACCS, CCP-C, FP-C Christopher J. Fullagar, MD, FACEP, EMT-P Stan Goettel, MS, EMT-P Author credits /
More informationNew A7 Anesthesia Workstation
New A7 Anesthesia Workstation Visible Precision Pls refer to user manual for manufacturer information BeneVision Central Monitoring System-210X285X12PX20150329 Since 2006 Mindray has successfully installed
More informationIntroducing Astral. Greater freedom. Confident care. Designed efficiency
Enrich life Introducing Astral From initial setup to everyday use, the Astral 100 and Astral 150 life support ventilators offer greater freedom, confident care and designed efficiency to enrich life for
More informationSupporting your patients every breath
Supporting your patients every breath ResMed.com/Astral Stay informed of patient changes throughout the therapy journey Improve patient care through connectivity. Astral s connectivity to AirView gives
More informationVT650. Gas Flow Analyzer. Technical data. Key benefits and features:
VT650 Gas Flow Analyzer Technical data Accurately test gas flow equipment including ventilators with the all-in-one, portable Fluke Biomedical VT650 Gas Flow Analyzer. Accurate The VT650 Gas Flow Analyzer
More informationGE Healthcare. Avance Carestation. Upgrade promotions guide
GE Healthcare Avance Carestation Upgrade promotions guide Contents Display upgrade package 1 for Avance 3 Display upgrade package 2 for Avance 3 Ventilation modes upgrade (PCV-VG, PSVPro, SIMV, Cardiac
More informationpatient education program 4156 South 52 nd Street, Omaha, NE ChildrensOmaha.org/HomeHealthcare
patient education program 4156 South 52 nd Street, Omaha, NE 68117 800-747-7334 402-734-6741 ChildrensOmaha.org/HomeHealthcare We know children CHILDREN S HOME HEALTHCARE Trilogy Ventilator Instructions
More informationBunnell LifePulse HFV Quick Reference Guide # Bunnell Incorporated
Bunnell Incorporated n www.bunl.com n 800-800-4358 (HFJV) n info@bunl.com 436 Lawndale Drive n Salt Lake City, Utah 84115 n intl 801-467-0800 n f 801-467-0867 Bunnell LifePulse HFV Quick Reference Guide
More informationNOTE: If not used, provider must document reason(s) for deferring mechanical ventilation in a patient with an advanced airway
APPENDIX: TITLE: Mechanical Ventilator Use REVISED: November 1, 2017 I. Introduction: Mechanical Ventilation is the use of an automated device to deliver positive pressure ventilation to a patient. Proper
More informationWILAmed INTENSA. Intensive Care Ventilator System
EN WILAmed INTENSA Intensive Care Ventilator System 0197 WILAmed INTENSA Ventilator System The WILAmed INTENSA is a full invasive and non-invasive life-support ventilator, solely designed for the ventilation
More informationFabius MRI. From specialists for specialists. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care
Fabius MRI From specialists for specialists Emergency Care Perioperative Care Critical Care Perinatal Care Home Care Because you care Your team focuses on performance and safety. So does our breakthrough
More informationHow Ventilators Work. Chapter 3
How Ventilators Work Chapter 3 To care for a ventilator patient, you need to know: The various functions of the ventilator used How the ventilator interacts with the patient How changes in lung condition
More informationVT900. Gas Flow Analyzer. Technical data. Key benefits and features:
VT900 Gas Flow Analyzer Technical data The Fluke Biomedical VT900 is designed to accurately and reliably test all types of medical gas flow equipment ventilators, insufflators, oxygen meters especially
More informatione700 Automatic Transport Ventilator User Manual e700 User Manual (15PL Rev.14 28_8_2017)
e700 Automatic Transport Ventilator User Manual Table of Contents Chapter 1 Safety... 1 Chapter 2 Intended Use.. 4 Chapter 3 Overview.... 5 3.1 Control and Display Layout...... 5 3.2 Function Keys......
More informationUpdate to RS-232 commands. Changing patient from NIV to INVASIVE Vent Type. SNDF command
Changing patient from NIV to INVASIVE Vent Type Table 9 shows automatic settings changes that occur when changing the same patient from NIV to INVASIVE Vent Type. Table 9: Automatic settings changes NIV
More informationMechanical Ventilation. Mechanical Ventilation is a Drug!!! is a drug. MV: Indications for use. MV as a Drug: Outline. MV: Indications for use
Mechanical Ventilation is a Drug!!! Mechanical Ventilation is a drug I am an employee of Philips Healthcare Hospital Respiratory Care Group and they help me pay for my kids education Jim Laging, RRT, RCP
More informationSERVO-air ventilator. Datasheet for system version 2.1
SERV-air ventilator Datasheet for system version 2.1 This document is intended to provide information to an international audience outside of the US. Contents Technical specifications... 3 General... 3
More informationSpecification of ICU Ventilator (Respiratory Ventilator
Specification of ICU Ventilator (Respiratory Ventilator 1. Should have facility for Invasive and Non-Invasive ventilation. 2. Microprocessor Control suitable for Pediatric and adult ventilation. 3. Electromagnetic
More information