prosa Tools Instructions for Use CHRISTOPH MIETHKE GMBH & CO. KG USA

Size: px
Start display at page:

Download "prosa Tools Instructions for Use CHRISTOPH MIETHKE GMBH & CO. KG USA"

Transcription

1 prosa Tools Instructions for Use CHRISTPH MIETHKE GMBH & C. KG 0297

2 prosa Tools CNTENT Remark 4 Indication 5 prosa VERIFICATIN TL 5 prosa MASTERDISC 5 prosa VERIFICATIN CMPASS 6 prosa ADJUSTMENT TL 6 prosa ADJUSTMENT DISC 7 Technical description 8 Physics background 9 Function of the prosa 10 Selecting the appropriate prosa 12 Adjusting the prosa 14 Reading the pressure setting from an X-ray image 17 prosa CHECK-MATE 18 Cleaning and disinfecting the prosa CHECK-MATE 18 Recommendation for sterilisation 20 Recommendation for sterilisation of the prosa CHECK-MATE 20 Cleaning recommendation for prosa TLS which are not sterilisable 20 Medical products consultant 21 General information 21 3

3 REMARK Note: In order to locate, to verifiy and to adjust either the adjustable DP-unit or the adjustable gravitational unit, different Verification and Adjustment tools are used: Tools used for the adjustable DP-unit: progav VERIFICATIN TL progav Masterdisc progav Compass progav ADJUSTMENT TL progav ADJUSTMENT DISC progav CHECK-MATE Tools used for the adjustable gravitational unit: prosa VERIFICATIN TL prosa Masterdisc prosa Compass prosa ADJUSTMENT TL prosa ADJUSTMENT DISC prosa CHECK-MATE Please verify specifically before using any tool for verifying or adjusting the opening pressure: for the adjustable DP-unit, use only progav Tools and for the adjustable gravitational unit, use only prosa Tools CAUTIN Federal law restricts this device to sale by or on order of a physician INDICATIN The Miethke prosa Adjustable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum. PRSA VERIFICATIN TL The prosa VERIFICATIN TL is used for reading the valve opening pressure setting. 2 The marking (3) on the prosa VERIFICATIN TL must be in line with the proximal (ventricular) catheter. The tool contains two magnets. As soon as the button on the TL is pushed (2) the magnets in the tool align with the magnets in the valve. The opening pressure is shown on the scale (1). PRSA MASTERDISC The prosa VERIFICATIN TL can be easily checked by using the prosa MASTERDISC before measuring the opening pressure of the valve. n the disc the positions 0, 10, 20, 30 and 40 cmh 2 are indicated. If the prosa VERIFICATIN TL is placed on the disc the opening pressure shown by the TL should be aligned to the value of the prosa MASTERDISC. Example: The VERIFICATIN TL is put on the MASTERDISC so that the marking on the TL is in line with the value 10 cmh 2 on the MASTERDISC. The prosa VERIFICATIN TL should indicate the value of 10 cmh 2. The mode of operation of progav und prosa TLS are identical. Hence, the instructions on the following pages are classified for both progav and prosa TLS. Note: These TLS contain material which has been determined to be MR Unsafe. D NT use in or around strong magnetic fields such as MR imaging equipment. 1 3 Fig. 1: prosa VERIFICATIN TL Fig. 2: prosa MASTERDISC 4 5

4 prosa VERIFICATIN CMPASS Alongside the VERIFICATIN TL there is an additional device for measuring the adjusted opening pressure. The prosa VERIFICATIN CMPASS can be used to locate the valve when palpation is not possible. The VERIFICA- TIN CMPASS is set to the skin above the implanted valve. The opening pressure corresponds to the value indicated towards the direction of the ventricular catheter. Caution: Airbubbles inside the compass do not afffect its funcionality. prosa Adjustment tool The prosa ADJUSTMENT TL is used for adjusting the valve opening pressure. First the intended pressure setting is selected at the knurled dial (1), the opening pressure is shown on a scale (2). Then the prosa ADJUSTMENT TL is placed centrally on the valve. By pushing the button (3), the adjustment tip (4) appears, the brake is decoupled, the rotor turns and the adjusted pressure is set. The marking (5) on the prosa ADJUSTMENT TL has to point towards the proximal catheter (leading to the ventricle). prosa Adjustment disc The prosa ADJUSTMENT DISC offers another option to adjust the pressure setting. The prosa ADJUSTMENT DISC is placed centrally over the valve. The desired pressure setting should be aligned with the proximal catheter (leading to the ventricle). By pressing down the prosa ADJUSTMENT DISC on the valve, the brake is decoupled and the opening pressure of the prosa is changed. Ensure that the pressure setting is changed no more than 16 cmh 2. (see Adjusting the prosa ) 3 Caution: Due to magnets inside the prosa TLS, do not use the prosa TLS nearby pacemakers. Further more do not use the prosa TLS nearby MRI scanner, since there is a danger of damaging the MRI-scanner Fig. 3: prosa VERIFICATIN CMPASS Fig. 4: prosa ADJUSTMENT TL Fig. 5: prosa ADJUSTMENT DISC 6 7

5 prosa CHRISTPH MIETHKE prosa CHRISTPH MIETHKE TECHNICAL DESCRIPTIN The prosa is the first hydrocephalus valve that enables the adjustment of the opening pressure in the upright position, as well as for all angles of elevation between the upright and the supine positions. This opening pressure depends on the body position of the patient. The prosa combines a differential pressure unit and an adjustable gravitational unit, (Fig. 6). The prosa comprises a robust titanium casing (1) whose proximal end contains a ball-cone valve. The sapphire ball (2) ensures the precise closure of the valve and a bow spring (4) determines the opening pressure of the ball-in-cone valve. The tantalum weight (3) is connected to the spring and keeps the sapphire ball in its position. In this way, optimal CSF drainage is ensured for each individual patient in any body position. The tension of the spring, and thus the valve opening pressure, can be adjusted by turning the rotor (5), with the valve implanted under the patient s skin. The cerebrospinal fluid flows in through the inlet port (6) and flows out by way of the outlet port (7). Intraventricular pressure pening pressure in horizontal positi- (DP-unit only) pening pressure in vertical position (DP-unit + gravitational unit) Pressure in the abdominal cavity Hydrostatic pressure IVP P Vli on P Vst P B P Hyd horizontal: IVP = P Vli + P B vertical: IVP = P Hyd - P Vst - P B P Vli P Hyd IVP P Vst IVP P B P B 1 titanium casing 2 sapphire ball 3 tantalum weight 4 bow spring 5 rotor 6 inlet connector 7 outlet connector 8 spiral spring 9 ligature 10 catheter adjustable gravitational unit 4 5 differential pressure unit Fig. 7: Calculating the intraventricular pressure PHYSICS BACKGRUND The intraventricular pressure is positive in a healthy human in the horizontal position. To adjust this pressure through shunt drainage, the appropriate pressure range must be chosen, taking into account the abdominal cavity pressure. The resulting IVP is the sum of the shunt opening pressure and the abdominal cavity pressure (Fig. 7). The ventricular pressure in a healthy human when moving to the vertical position becomes slightly negative. To maintain this pressure by means of shunt drainage, the shunt opening pressure has to be significantly higher so that the shunt can compensate for the change in pressure differential. This is equal to the hydrostatic pressure minus the sum of the abdominal cavity pressure and the slightly negative intraventricular pressure. Conventional shunts open immediately as soon as the patient stands up, which can lead to critical overdrainage. 1 7 Fig. 6: Schematic cross section of the prosa 8 9

6 FUNCTIN F THE prosa Horizontal body position With the prosa in the horizontal position (Fig. 8), the weight (2) does not affect the sapphire ball. Hence, in this configuration the prosa is open, and the force brought to bear by the weight does not counteract the fluid pressure. The ventricular pressure is now controlled solely by the differential-pressure valve installed. Vertical body position In the vertical position, the DP-unit and the gravitational unit of the prosa work together. The total opening pressure is the sum of the opening pressures of the DP-unit and the adjustable gravitational unit. If the IVP (intraventricular pressure) of the patient and the hydrostatic pressure exceed this opening pressure, the closing ball of the prosa is pushed away from the cone seal and a gap opens for fluid drainage (Fig. 9). The valve system prevents increases in intraventricular pressure above the physiological range. To ensure optimal CSF drainage for each individual patient in any body position, the opening pressure of the adjustable gravitational unit can be adjusted from 0 to 40 cmh 2. The adjustment can be achieved non-invasively using the prosa TLSET, preventing the need for further surgery. 2 1 a) b) a) b) Fig. 8: DP-unit a) closed and b) open The operating principle of the DP-unit is illustrated in Fig. 8. a) and b) Fig. 8 a shows the DP-unit in the horizontal position. The ball-cone valve is closed and drainage is prevented. If the patient s IVP increases and continues to rise, the spring pressure of the ball-cone unit will be overcome. The sealing ball will move away from the cone and a gap will open for fluid drainage. Fig. 9: gravitational unit in vertical body position a) closed b) open 10 11

7 SELECTING THE APPRPRIATE prosa The prosa is a position-dependent valve, that means the opening pressure changes depending on the body position of the patient. Hence, one opening pressure for the horizontal position and one for the vertical position is set to fit the prosa individually to the patient. (see recommendation for pressure levels: Horizontal position In the horizontal position the gravitational unit doesn t have any resistance. nly the differential pressure unit determines the opening pressure of the whole shunt - ystem. As standard configuration we recommend a differential pressure unit with opening pressure of 5 cmh 2. Vertical position In the vertical position the opening pressure of the complete shunt system is the sum of the opening pressure of the differential unit and the opening pressure of the gravitational unit. The opening pressure of the gravitational unit should be chosen depending on height, weight and age of the patient. opening pressure differential pressure unit opening pressure gravitational unit opening pressure differential pressure unit recommendation opening pressure gravitational unit standard 5 cmh 2 children up to 5 years 20 cmh 2 opening pressure of the whole shunt system defensive* = 10 cmh cmh 2 opening pressure of the whole shunt system = + opening pressure of the differential unit adults up to 60 years 25 cmh 2 special** 15 cmh 2 adults over 60 years 20 cmh 2 *e.g. patients with extremely wide ventricles and highly elevated ICP or aqueductal stenosis **e.g. patients mit pseudotumor cerebri The recommendations are based on common patient treatments, but can vary depending on the individual patient s condition

8 ADJUSTING THE prosa Please verfify specifically before using any TL for verifying or adjusting the opening pressure: for the adjustable DP-unit, use only progav Tools 2. Measuring the opening pressure The prosa VERIFICATIN TL is positioned centrally over the valve. (Fig. 11). The notch on the TL must point towards the proximal catheter (leading to the ventricle). 3. Adjusting the opening pressure 3a. Adjustment with the prosa ADJUSTMENT TL The prosa ADJUSTMENT TL is set to the required opening pressure by turning the knurled dial of the unit (Fig. 13). and for the adjustable gravitational-unit, use only prosa Tools Abb. 14b: Adjusting the pressure setting Each prosa is calibrated under strict quality control procedures. The adjustable gravitational unit is preset to 20 cmh 2, but must be checked before implantation. The adjustment procedure is outlined in the steps below: 1. Locating the prosa The adjustable gravitational unit is located under the skin (Fig. 10). Fig. 11: Place centrally over the valve The arrow on the bottom side of the prosa VERIFICATIN TL indicates the direction of CSF-flow. The button is pushed and the pressure setting is read (Fig. 12). a) b) Fig. 13: Setting the prosa ADJUSTMENT TL Caution: Please ensure that the opening pressure is changed by no more than 16 cmh 2 per each adjustment. (see Nr. 4) The prosa ADJUSTMENT TL is positioned centrally on the valve. The notch on the TL and the scale (a) must point towards the proximal catheter (b) (leading to the ventricle), see Fig. 14a. b) a) Caution: Ensure that the TL remains close to the valve during the adjustment procedure. When treating patients who have a low tolerance to pain, local anaesthesia (e.g. applied through a plaster) should be considered, in cases where contraindication can be excluded. If the site of implantation is poorly selected or if the skin over the valve is too thick, an adjustment of the prosa can be difficult or sometimes impossible. The valve then behaves like a gravitational valve with one opening pressure for a given position. Patients should discuss the risks involved with their surgeon. progav CHRISTPH MIETHKE Caution: Improper use of the prosa ADJUST- MENT TL may result in the damage of the prosa valve. Such damage may lead to loss of shunt integrity, and necessitate premature surgical revision of the shunt system. Fig. 12: Measuring the pressure setting a) correctly, b) incorrectly Caution: Failing to centre the prosa VERI- FICATIN TL over the valve can lead to erroneous readings! Fig. 14a: Positioning the prosa ADJUSTMENT TL As soon as the prosa ADJUSTMENT TL has been positioned centrally over the valve, the button is pushed and the adjustment tip appears to apply pressure to the valve. This triggers the mechanical decoupling of the rotor and the valve setting is adjusted to the required opening pressure setting (Fig. 14b). When adjusting the prosa preoperatively through the packaging, moderate force with the prosa ADJUSTMENT TL is sufficient. D NT USE THE BUTTN! Strong pressure can cause damage to the housing, which might affect valve function. Caution: Due to postoperative swelling of the skin the adjustment of the valve setting may be difficult within the first few days. Fig. 10: Locating the valve Caution: X-ray confirmation may still be necessary for patients with scalp thicknesses greater than or equal to 5 mm thick

9 3b. Adjustment with the prosa ADJUST- MENT DISC Center the prosa ADJUSTMENT DISC over the gravitational unit of the prosa and align the desired pressure setting (b) on top of the disc in direction of the ventricular catheter (c), see Fig. 15a. For changing the opening pressure, press down the prosa adjustment disc and release (Fig. 15b). Do not press and turn. Finally, remove the prosa ADJUSTMENT DISC and confirm the setting with the prosa VERIFICATIN TL. 4. Verifying after adjustment After adjusting the valve by using the Adjustment Tool, it can be verified using the Verification Tool, as described in step 2. If the pressure measured now differs from the intended pressure level, the adjustment procedure has to be repeated from step 3. After adjusting the valve using the prosa ADJUSTMENT TL, it can be verified using the prosa VERIFICATIN TL and may be confirmed by radiograph (X-ray). During the adjustment, the opening pressure should not be changed more than 16 cmh 2 at a time. The following table shows the quantitative information regarding the overall agreement rates between the X-ray and the respective verification tool for the prosa valve. The maximum deviation from the actual valve readings used were 0, 1, 2, >2 cmh 2. For example at 2 cmh % of the measurements made by the prosa VERIFICATIN TL with the prosa valve deviate not more than ± 2 cmh 2 from the actual value of the valve. The agreement rates are based on non-clinical testing with readings of 15 valves by 15 clinical users under simulated use conditions, yielding a total of 225 readings with each valve and verification tool or compass. 40 cut-out rotor tip Fig. 17: Schematic X-ray image not adjustable range 10 0 c b Example: pening pressure is to be changed from 6 to 36 cmh 2. With only one adjustment step the rotor would turn in a counter clockwise direction (shortest path) and would stop at the position 0 cmh 2 (Fig. 16a). prosa Valve = 0 cmh 2 1 cmh 2 2 cmh 2 > 2 cmh 2 prosa VERIFICATIN TL 64.4 % 89.8 % 99.6 % 0.4 % cut-out rotor tip Abb. 15a: Center the prosa Adjustment disc on the valve range not adjustable range not adjustable 0 prosa Valve prosa VERIFICATIN CMPASS 64.9 % 96.0 % 99.6 % 0.4% Fig. 18: X-ray image position 31 cmh 2 b READING THE PRESSURE SETTING FRM AN X-RAY IMAGE c Abb. 15b: Press down slightly the prosa ADJUSTMENT DISC and release a) b) Fig. 16: Rotor rotation during adjustment The correct adjustment is performed in 2 steps: Adjustment from 6 to 22 cmh 2, and then from 22 to 36 cmh 2. The rotor turns now in a clockwise direction (Fig. 16b). Caution: If the pressure configuration of the valve cannot be determined with complete certainty by the prosa VERIFICATIN TL, the use of imaging techniques is recommended (excluding MRI: danger of artefacts). MRI examinations must be performed at field strengths no greater than 3.0 tesla. The pressure setting of the prosa should be measured with the prosa Verification Tool, but if there is any discrepancy between the desired adjustment setting and the setting that is read by the prosa Verification Tool, then a radiographic confirmation (X-ray) can be performed to confirm the actual valve setting. The magnets of the rotor are visible in the X-ray image as white dots (Fig. 17). The orientation of the rotor tip indicates the pressure setting. The rotor tip can take any position outside the region indicated shown below (Fig. 17). Thus, the opening pressure of the adjustable gravitational unit can be adjusted from 0 to 40 cmh 2, in increments of 1 cmh 2. To further assist correct identification of the adjusted opening pressure, a cut-out in the casing ring is visible under X-ray imaging (Fig. 18)

10 prosa CHECK-MATE The prosa CHECK-MATE is delivered sterile and is intended to be re-sterilised. It is possible to change and to verify an applied pressure setting on the valve directly in the R. To verify the actual pressure setting the prosa CHECK-MATE has to be put centrally over the valve. The prosa CHECK-MATE will immediately start to move. If it remains stable, the pressure setting can be read in allignment to the inlet connector. To adjust a new pressure setting, the prosa CHECK-MATE has to be placed centrally over the valve. The new pressure setting has to point towards the proximal catheter (leading to the ventricle). By pressing down slightly the prosa CHECK-MATE, the brake of the valve is decoupled, the rotor turns and the opening pressure of the prosa is changed. Please be aware that the steps for changing the pressure setting should not be more than 16 cmh 2 per step (see chapter Adjusting the prosa ). Fig. 19: prosa CHECK-MATE 0297 CLEANING AND DISINFECTING THE prosa CHECK-MATE Avoid damage to the product due to inappropriate cleaning/disinfecting agents and/or exessive temperatures! Use cleaning and disinfecting agents approved for surgical steels according to the manufacturer s instructions. bserve specifications regarding concentration, temperature and exposure time. Do not exceed the maximum allowable cleaning temperature of 55 C. Carry out ultrasound cleaning: as an effective mechanical supplement to manual cleaning/disinfecting. as a precleaning procedure for products with encrusted residues, in preparation for mechanical cleaning/disinfecting. as an integrated mechanical support measure for mechanical cleaning/disin fecting. for additional cleaning for products with residues left after mechanical cleaning/ disinfecting. Clean and disinfect TLs mechanically, provided they can be securely fixed in machines or storage devices in such a way that they will be thoroughly cleaned. Manual cleaning/disinfecting Check visible surfaces for residues after manual cleaning/disinfectin Repeat the cleaning process if neccessary. Mechanical cleaning/disinfecting Place the product on a tray that is suitable for cleaning (avoid rinsing blind spots). Mechanical cleaning/disinfecting with manual pre-cleaning Manual pre-cleaning with ultrasound Stage Step T ( C/ F) I Desinfecting ultrasound cleaning t (min) Stage I Clean the product in an ultrasound cleaning bath (frequency 35 khz). Make certain that all accessible surfaces are moistened. Conc. (%) Water quality Chemical RT (cold) 15 2 D-W BBraun Stabimed; aldehyde phenol and QAV free; ph = 9 II Intermediate rinse RT (cold) 1 - D-W D-W: Drinking water, RT: Room temperature Mechanical alkaline cleaning and thermal disinfecting Stage II Rinse the product completely (all accessible surfaces) under running water. Machine type: Single-chamber cleaning/disinfecting machine without ultrasound Place the product on a tray that is suitable for cleaning ( avoid rinsing blind spots) Stage Step T ( C/ F) t (min) Water quality I Prerinse <25/77 3 D-W - Chemical II Cleaning 55/ FD-W -Concentrate, alkaline: ph = 10.9 <5% anionic tensides -1% solution: ph = 10.5 III Neutralization 20/68 2 FD-W - Concentrate, acid: ph = 2,6 Basis: Citric acid - 1% solution: ph = 3,0 IV Intermediate rinse 70/158 1 FD-W - V Thermal disinfecting 94/ FD-W - VI Drying 90/ D-W: Drinking water, FD-W: Fully desalinated water (demineralized) Control, care and inspection Allow the product to cool down to room temperature. Inspect the product after each cleaning and disinfecting cycle to be sure it is clean, functioning properly, not damaged, has intact insulation and does not have any loose, bent, broken, cracked, worn, or fractured components. Set aside the product if it is damaged. We recommend putting the TL in a double sterile bag after cleaning. So the next use in the theatre (sterile area) will be prepared as its best

11 RECMMENDATIN FR STERILISATIN Except the prosa CHECK-MATE, prosa TLS can not be sterilised. RECMMENDATIN FR STERILISATIN F THE PRSA CHECK-MATE The prosa CHECK-MATE should be sterilised by steam sterlization (fractionated vacuum process) at 132 C and 4 minutes cycle time. Sterilization for prosa- CHECK-MATE in the Validated Cycle: Sterilizer Type: Minimum Temperature: Full Cycle Time: Minimum Dry Time: Sample Configuration: Prevacuum 132 degrees Celsius 4 minutes 20 minutes Sterilization Container (Aesculap Steril- Container) CLEANING RECMMENDATIN FR prosa TLS WHICH ARE NT STERILISABLE Attention: prosa TLS are made from thermal unstable components which are affectable by heat or humidity or chemical agressive substances. Do not steep the prosa TL in liquids and keep the inside of the TLs dry! Remove surface pollution of the prosa TLS after the use immediately with alcohol based cleaners (more than 75% alc.) by a wiping procedure. The time of impact should be more than 60 sec. and should be depending on the level of pollution. For final cleaning use a dry wipe. The following cleaning methods are not allowed for the cleaning of the prosa TLs (except prosa CHECK-MATE): Irradiation, Ultrasonic, Sterilization, Machine preparation, Inserting into liquids. MEDICAL PRDUCTS CNSULTANT In compliance with the requirements of the European law MDD 93/42/EEC, Christoph Miethke GmbH&Co. KG names medical pro duct consultants as the individuals to be addressed with all queries concerning the products: Dipl.-Ing. Christoph Miethke Dipl.-Ing. Roland Schulz Christoph Miethke GmbH & Co. KG Ulanenweg 2 D Potsdam Phone: +49(0) or Phone: +49(0) Fax: +49(0) info@miethke.com Please address any enquiries to: AESCULAP AG Am Aesculap Platz D Tuttlingen Phone: +49 (0) Fax: +49 (0) information@aesculap.de Service address in the US AESCULAP Inc. Attn. AESCULAP Technical Services 615 Lambert Pointe Road Hazelwood, M, GENERAL INFRMATIN Manufacturer Product name Intended use Christoph Miethke GmbH & Co. KG TLS for prosa Treatment of Hydrocephalus Store in a clean, dry place CFU should be determined and lower than 10CFU per item AESCULAP Repair Hotline Phone: +1 (800) Fax: +1 (314) Distributor in the US/ Contact in Canada AESCULAP Inc Corporate Parkway Center Valley, PA Phone:

12

13 0297 CE-Kennzeichnung gemäß Richtlinie 93/42/EWG CE marking according to directive 93/42/EEC Label CE conforme à la directive 93/42/CEE Identificatión CE en conformidad con la directriz 93/42/CEE Marchio CE conforme alla direttiva 93/42/CEE Technische Änderungen vorbehalten Technical alterations reserved Sous réserve de modifications techniques Sujeto a modificationes técnicas Con riserva di modifiche tecniche Manufacturer acc. MDD 93/42/EEC: CHRISTPH MIETHKE GMBH & C. KG Christoph Miethke GmbH & Co. KG Ulanenweg Potsdam Germany Phone +49 (0) Fax +49 (0) Distributed by: Aesculap AG Am Aesculap-Platz Tuttlingen Germany Phone +49 (0) Fax +49 (0) Aesculap - a B. Braun company TA GBA_73_05_0217_

prosa CHRISTOPH MIETHKE GMBH & CO. KG Instructions for Use USA

prosa CHRISTOPH MIETHKE GMBH & CO. KG Instructions for Use USA Instructions for Use CHRISTPH MIETHKE GMBH & C. KG 0297 INSTRUCTINS FR USE CNTENT INDICATIN FR USE 4 TECHNICAL DESCRIPTIN 4 PHYSICS BACKGRUND 5 FUNCTIN F THE 6 SELECTING THE APPRPRIATE PENING PRESSURE

More information

progav 2.0 CHRISTOPH MIETHKE GMBH & CO. KG Instructions for Use USA

progav 2.0 CHRISTOPH MIETHKE GMBH & CO. KG Instructions for Use USA progav 2.0 Instructions for Use CHRISTPH MIETHKE GMBH & C. KG 0297 INSTRUCTINS FR USE CNTENT INDICATIN 4 TECHNICAL DESCRIPTIN 4 FUNCTIN F THE VALVE PHYSICS BACKGRUND 6 SELECTING THE APPRPRIATE SHUNT 7

More information

progav 2.0 CHRISTOPH MIETHKE GMBH & CO. KG Instructions for Use USA

progav 2.0 CHRISTOPH MIETHKE GMBH & CO. KG Instructions for Use USA progav 2.0 Instructions for Use CHRISTOPH MIETHKE GMBH & CO. KG 0297 INSTRUCTIONS FOR USE CONTENT INDICATION 4 TECHNICAL DESCRIPTION 4 FUNCTION OF THE VALVE 5 PHYSICS BACKGROUND 6 SELECTING THE APPROPRIATE

More information

progav 2.0 In touch with you

progav 2.0 In touch with you progav 2.0 In touch with you We understand the gravity of the situation. Gravitational valves by Miethke Aesculap Neurosurgery progav 2.0 In touch with you MIETHKE gravitational valves provide a reliable

More information

WE UNDERSTAND. NEUROSURGERY. progav 2.0 IN TOUCH WITH YOU

WE UNDERSTAND. NEUROSURGERY. progav 2.0 IN TOUCH WITH YOU WE UNDERSTAND. NEUROSURGERY progav 2.0 IN TOUCH WITH YOU 2 ADJUSTABLE VALVES HAVE EXISTED FOR DECADES. THE ADJUSTMENT AND TESTING OF THE VALVE SETTING IS STILL A CHALLENGE FOR THOSE INVOLVED (1). Valve

More information

prosa Adjustable MIETHKE Shunt Systems Your Simple Choice for Precise Patient Care Aesculap Neurosurgery MIETHKE

prosa Adjustable MIETHKE Shunt Systems Your Simple Choice for Precise Patient Care Aesculap Neurosurgery MIETHKE prosa Adjustable MIETHKE Shunt Systems Your Simple Choice for Precise Patient Care MIETHKE Aesculap Neurosurgery MIETHKE prosa Enhanced Patient Care Active people spend about 2/3 of the day walking, standing

More information

Prestige Endoscopic Graspers Instructions for Use

Prestige Endoscopic Graspers Instructions for Use Prestige Endoscopic Graspers Instructions for Use DESCRIPTION The Prestige Endoscopic Graspers are designed to grasp soft tissue during endoscopic procedures. The graspers are designed to be used through

More information

Instructions for use Compressed air injector with water trap

Instructions for use Compressed air injector with water trap Instructions for use Compressed air injector with water trap Table of Contents Introduction 1. Description of Functions 2. Putting into Operation 3. Instructions for Cleaning and Care 3.1 Disassembling

More information

MULTICHANNEL. Labnet International, Inc.

MULTICHANNEL. Labnet International, Inc. Labnet International, Inc. A B A1 A2 C F D H E 4A 4B 4C 4D G CONTENTS 1 - INTRODUCTION 2 - VOLUME SETTING 3 - METHOD OF PIPETTING 4 - RECOMMENDATIONS - RECALIBRATION 6 - CLEANING AND STERYLIZATION 7 -

More information

MAGNETIC FIELD INFLUENCES AND YOUR STRATA VALVE. Strata II Valves

MAGNETIC FIELD INFLUENCES AND YOUR STRATA VALVE. Strata II Valves MAGNETIC FIELD INFLUENCES AND YOUR STRATA VALVE Strata II Valves Magnetic field influences on your Strata valve You have a Medtronic adjustable valve to treat your hydrocephalus or other CSF disorder.

More information

AUTOVENT 4000 VENTILATOR

AUTOVENT 4000 VENTILATOR OVERVIEW AUTOVENT 4000 Only properly trained and approved Escambia County Bureau of Public Safety Paramedics are to use the AutoVent 4000 ventilator manufactured by LSP to transport patients already on

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.12 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient,

More information

Type C-CUT. Quickstart. English Deutsch Français

Type C-CUT. Quickstart. English Deutsch Français Capillary modules for micro and ultrafiltration Kapillarmodule für die Mikro- und Ultrafiltration Modules capillaires pour micro- et ultrafiltration Quickstart English Deutsch Français We reserve the right

More information

Instruction Manual. ErgoLAP BIPOLAR

Instruction Manual. ErgoLAP BIPOLAR Instruction Manual ErgoLAP BIPOLAR EN Liability and guarantee This product complies with Directive 93/42 EEC concerning medical devices. Liability or guarantee is excluded, when the Instruction Manual

More information

Strata Valve ADJUSTMENT REFERENCE GUIDE

Strata Valve ADJUSTMENT REFERENCE GUIDE Strata Valve ADJUSTMENT REFERENCE GUIDE Strata Valve Adjustment Tools Preoperative Procedures READING THE PERFORMANCE LEVEL 1 2 1. Position the Locator Tool above the valve while it is still in the sterile

More information

Instructions for use Evacuation system compressed air injector SL 20 / SL 20 Max SL 30 / SL 30 Max

Instructions for use Evacuation system compressed air injector SL 20 / SL 20 Max SL 30 / SL 30 Max Instructions for use Evacuation system compressed air injector SL 20 / SL 20 Max SL 30 / SL 30 Max Table of Contents Introduction 1. Description of Functions 2. Putting into Operation 3. Instructions for

More information

Urgent Medical Device Recall Medtronic StrataMR Adjustable Valves & Shunts

Urgent Medical Device Recall Medtronic StrataMR Adjustable Valves & Shunts Urgent Medical Device Recall Medtronic StrataMR Adjustable Valves & Shunts 125 Cremona Drive Goleta, CA 93117 Catalog Number 42955, STRATAMR VALVE, SMALL E11656, E12556, E12753, E13482, E13483, E13484,

More information

U S E R M A N U A L AIR FLOWMETER. MODELS: 1MFA2001 (shown) 1MFA9001 CAUTION. ISO Certified

U S E R M A N U A L AIR FLOWMETER. MODELS: 1MFA2001 (shown) 1MFA9001 CAUTION. ISO Certified U S E R M A N U A L AIR FLOWMETER MODELS: 1MFA2001 (shown) 1MFA9001 ISO 13485 Certified Authorized Your EU local Representative: distributor: EMERGO EUROPE, INC. Molenstraat 15 2513866 BH -624 The -3952

More information

BlenderBuddy 2. Operating Manual & Instructions For Use ENGLISH R233M01 REV. A

BlenderBuddy 2. Operating Manual & Instructions For Use ENGLISH R233M01 REV. A BlenderBuddy 2 Operating Manual & Instructions For Use ENGLISH R233M01 REV. A Maxtec 2305 South 1070 West Salt Lake City, Utah 84119 USA TEL (800) 748.5355 FAX (801) 973.6090 www.maxtec.com E-mail: sales@maxtec.com

More information

Type 2000 INOX. Quickstart. English Deutsch Français. 2/2-way angle seat valve 2/2-Wege Schrägsitzventil Vanne à siège incliné 2/2 voies

Type 2000 INOX. Quickstart. English Deutsch Français. 2/2-way angle seat valve 2/2-Wege Schrägsitzventil Vanne à siège incliné 2/2 voies Type 2000 INOX 2/2-way angle seat valve 2/2-Wege Schrägsitzventil Vanne à siège incliné 2/2 voies Quickstart English Deutsch Français Contents 1 QUICKSTART... 2 2 CONTACT ADDRESS... 2 3 INTENDED USE...

More information

Dialflow Regulator. Instructions for Use

Dialflow Regulator. Instructions for Use Dialflow Regulator Instructions for Use 702-0030.11 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

VariSource High Dose Rate Afterloader Procedures. For Performing Breast Brachytherapy with. The SAVI TM Applicator

VariSource High Dose Rate Afterloader Procedures. For Performing Breast Brachytherapy with. The SAVI TM Applicator VariSource High Dose Rate Afterloader Procedures For Performing Breast Brachytherapy with The SAVI TM Applicator Contents I. CT Evaluation (~1 week prior to insertion) II. SAVI Applicator Implantation

More information

Injector system for implantation of 1stQ foldable intraocular lenses

Injector system for implantation of 1stQ foldable intraocular lenses Instruction For Use 1st Inject Instrument Injector system for implantation of 1stQ foldable intraocular lenses The IFU is available electronically on our website: www.1stq.eu Content: The package contains

More information

POLARIS. The first MRI-stable adjustable valve NEW MR SAFE

POLARIS. The first MRI-stable adjustable valve NEW MR SAFE POLARIS The first MRI-stable adjustable valve TESTED MR SAFE NEW Polaris Valve The adjustable Polaris valve is a major breakthrough for the safety of patients thanks to the patented self-locking system

More information

A N F G J I C D D L A B C D E F

A N F G J I C D D L A B C D E F A1 A2 B A N C D E F G J I C D L A B C D E F Model GM2 GM10 GM20 GM100 GM200 GM1000 GM5000 GM10000 Volume range [μl] 0.1-2 0.5-10 2-20 10-100 20-200 100-1000 1000-5000 1000-10000 CONTENTS 1 - INTRODUCTlON

More information

Contacts. Quick Start Guide

Contacts. Quick Start Guide Contacts Clinical Support Specialist: Phone: Cell Phone: Email: Fresenius Renal Technologies A division of Fresenius Medical Care North America 920 Winter Street Waltham, MA 02451 Technical Service Customer

More information

Installation, Operation & Maintenance Manual for Flo-Max Coupler Bracket Model FM150

Installation, Operation & Maintenance Manual for Flo-Max Coupler Bracket Model FM150 Installation, Operation & Maintenance Manual for Flo-Max Coupler Bracket Model FM150 January 2013 Form FVC 084 - Rev 02 IMPORTANT: KEEP THIS DOCUMENT WITH THE PRODUCT UNTIL IT REACHES THE END USER. 1.

More information

User manual. MNPG 102 Rev. 04 Edition 04/02/16. Pressotherapy model. Power-Q1000 Premium

User manual. MNPG 102 Rev. 04 Edition 04/02/16. Pressotherapy model. Power-Q1000 Premium User manual MNPG 102 Rev. 04 Edition 04/02/16 Pressotherapy model Power-Q1000 Premium I.A.C.E.R. Srl Via S. Pertini 24/A 30030 Martellago (VE) ITALY Tel. 041.5401356 Fax 041.5402684 e-mail: iacer@iacer.it

More information

ACV-10 Automatic Control Valve

ACV-10 Automatic Control Valve ACV-10 Automatic Control Valve Installation, Operation & Maintenance General: The Archer Instruments ACV-10 is a precision automatic feed rate control valve for use in vacuum systems feeding Chlorine,

More information

AIR/OXYGEN BLENDER INSTRUCTION MANUAL

AIR/OXYGEN BLENDER INSTRUCTION MANUAL BIO-MED DEVICES AIR/OXYGEN BLENDER INSTRUCTION MANUAL CATALOG #2120 REV 042203 BIO-MED DEVICES INC. 1445 BOSTON POST ROAD, GUILFORD, CT 06437 (203) 458-0202 FAX (203) 458-0440 www.biomeddevices.com TABLE

More information

Operation Manual. Hand tool for AMP DUOPLUG Customer Manual PN: en (Translation of the original German version)

Operation Manual. Hand tool for AMP DUOPLUG Customer Manual PN: en (Translation of the original German version) Operation Manual Hand tool for AMP DUOPLUG 2.5 549033-1 Customer Manual Nr.: 411-18031-1 _Rev. C Customer Manual PN: Language: en (Translation of the original German version) The data specified above only

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.9 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

Use this key to determine which sections of this Product Manual apply to your job.

Use this key to determine which sections of this Product Manual apply to your job. R710 Product Manual Contents Warnings and Important Information 3 Recommended Use 4 Product Information 5 Technical Data 5 Before & During Every Transfer 6 Inspection 6 SoloLift Scale (optional) 6 SoloVest

More information

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. U S E R M A N U A L 1600 SERIES OXYGEN REGULATOR 168715G (Shown) SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090

More information

Micro Dial-Flowmeter. Instructions for Use

Micro Dial-Flowmeter. Instructions for Use Micro Dial-Flowmeter Instructions for Use 702-0082.8 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

SAPAG. Safety valves, type 5700 Storage, Use, Operation and Maintenance Instructions. IMPORTANT NOTICE

SAPAG. Safety valves, type 5700 Storage, Use, Operation and Maintenance Instructions.  IMPORTANT NOTICE SAPAG IMPORTANT NOTICE Contents Important notice 1 0 Valve identification 2 1 Storage 2 2 Installation 2 3 Operation 2 4 Maintenance 3 4.1 Dismantling 3 4.2 Inspection 3 4.3 Repair 3 4.4 Assembly 4 4.5

More information

ACRYLIC FLOW METERS 0-30, 0-70 & LPM

ACRYLIC FLOW METERS 0-30, 0-70 & LPM ACRYLIC FLOW METERS 0-30, 0-70 & 0-120 LPM ASSEMBLY INSTRUCTIONS & INSTRUCTIONS FOR USE R219P86 (0-120LPM) R219P87 (0-70LPM) R219P88 (0-30 LPM) R138P11 Rev. D TABLE OF CONTENTS: 1.0 PRODUCT OVERVIEW...1

More information

Operating Instructions VETTER Industrial Lifting Bags 8 bar - Type IKV

Operating Instructions VETTER Industrial Lifting Bags 8 bar - Type IKV Industrial Pneumatics. Operating Instructions VETTER Industrial Lifting Bags 8 bar - Type IKV Article No. 87015301 Vetter GmbH I 03/12 I Changes and errors excepted. Vetter Industrial Lifting Bags 8 bar

More information

User Manual PTW-DensiX. D / Di/Zi

User Manual PTW-DensiX. D / Di/Zi User Manual PTW-DensiX D148.131.0/3 2005-05 Di/Zi General Information General Information The product bears the CE-mark "CE-0124" in accordance with the Council Directive 93/42/EEC about Medical Devices

More information

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician. U S E R M A N U A L Model: 19MFA1001 Series SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090 Northampton,

More information

MANUAL. Sesame. Thermoplastic Tank Technologies

MANUAL. Sesame. Thermoplastic Tank Technologies MANUAL Sesame Thermoplastic Tank Technologies INSTALLATION AND USER GUIDE CONTENT 1. GENERAL 3 2. IMPORTANT 3 3. INSTALLATION EXPANSION VESSEL 4 4. USE EXPANSION VESSEL 5 5. AIR CELL REPLACEMENT 5 5.1

More information

X-Tech CLV User Manual. English

X-Tech CLV User Manual. English X-Tech CLV User Manual English Sommary 03. Use 04. Maintenance 05. Caution 06. Technical specifications 08. Cold water use (

More information

RB70 Automatic Diluent Valve Maintenance Manual. Version 1.1 November 2006 Written by Tino de Rijk. Page 1 of 23

RB70 Automatic Diluent Valve Maintenance Manual. Version 1.1 November 2006 Written by Tino de Rijk. Page 1 of 23 RB70 Automatic Diluent Valve Maintenance Manual Version 1.1 November 2006 Written by Tino de Rijk Page 1 of 23 Table of Contents 1. Introduction... 3 2. ADV diagram and parts list (Pre June 2006)... 4

More information

CJMJ.DPNQBDU. Instruction for use. Neonatal phototherapy device. BILI-COMPACT WY 1816/ Year of manufacture D2017_00

CJMJ.DPNQBDU. Instruction for use. Neonatal phototherapy device. BILI-COMPACT WY 1816/ Year of manufacture D2017_00 Instruction for use Neonatal phototherapy device CJMJ.DPNQBDU D2017_00 GmbH > BILI-COMPACT WY 1816/ Year of manufacture Herrenhöhe 4 D-51515 Kürten/Germany 02207-9698-0 02207-4750 E-mail: info@weyermed.com

More information

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician. USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician. RECEIVING / INSPECTION Remove the Precision Medical,

More information

Shukla Medical. TITLE: Manual Surgical Orthopedic Instruments Recommendations for Care, Cleaning, Maintenance and Sterilization

Shukla Medical. TITLE: Manual Surgical Orthopedic Instruments Recommendations for Care, Cleaning, Maintenance and Sterilization Page 1 of 5 1. PROCESSING INSTRUCTIONS These processing instructions are intended to assist the hospital and central supply management in developing procedures for safe and effective reprocessing of both,

More information

STAG. Balancing valves Balancing valve with grooved ends DN

STAG. Balancing valves Balancing valve with grooved ends DN STAG Balancing valves Balancing valve with grooved ends DN 65-300 IMI TA / Balancing valves / STAG STAG A grooved end, ductile iron balancing valve that delivers accurate hydronic performance in an impressive

More information

Instruction Manual Contact Pressure Vacuum Gauge

Instruction Manual Contact Pressure Vacuum Gauge MS10 Instruction Manual Contact Pressure Vacuum Gauge Table of Contents 1. Safety Instructions 2. Intended Applications 3. Product Description and Functions 4. Installation 5. Commissioning 6. Maintenance

More information

User Instruction Manual

User Instruction Manual User Instruction Manual Dual Sliding Beam Anchor Model # GF-DSBA-BA01202 Model # GF-DSBA-BA01802 Model # GF-DSBA-BA02402 WWW.GFORCE-SAFETY.COM Tacklestore Ltd Unit R1D Rockingham Gate Cabot Park, Poplar

More information

CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR RESUSCITATOR

CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR RESUSCITATOR CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR OXYLATOR FR-300 RESUSCITATOR Symbols used in the Operating Manual for the Oxylator FR-300 and/or on the Oxylator FR-300 unit: WARNING! Do not use grease or

More information

EMERGENCY WALL MOUNTED HAND OPERATED DELUGE SHOWER

EMERGENCY WALL MOUNTED HAND OPERATED DELUGE SHOWER EMERGENCY WALL MOUNTED HAND OPERATED DELUGE SHOWER Installation, Operating & Maintenance Instructions EW1050 I00232_Mar 2016 NOTE: THIS DOCUMENT IS TO BE LEFT ONSITE WITH FACILITY MANAGER AFTER INSTALLATION

More information

MONT BLANC 10400X XL XLED 10401X XL XLED 10403X XL XLED

MONT BLANC 10400X XL XLED 10401X XL XLED 10403X XL XLED MONT BLANC CONTR-ANGL D IMPLANTOLOGI 20:1 IMPLANTOLOGY CONTRA-ANGL 20:1 CONTRAANGULO PARA IMPLANTOLOGIA 20:1 CONTRANGOLO D IMPLANTOLOGIA 20:1 IMPLANTOLOGI WINKLSTÜCK 20:1 CONTRA-ÂNGULO PARA IMPLANTOLOGIA

More information

130400A 1/2 NPT Female A 3/4 NPT Female A 1 NPT Female A 1 1/4 NPT Female A 1 1/2 NPT Female A 2 NPT Female

130400A 1/2 NPT Female A 3/4 NPT Female A 1 NPT Female A 1 1/4 NPT Female A 1 1/2 NPT Female A 2 NPT Female 88.0A www.caleffi.com Venturi Style Balancing Valve Copyright 0 Caleffi 0 Series Function Product Range Balancing valves are hydraulic devices that can precisely control the flow rate of the fluid that

More information

Installation, Operating and Maintenance Instructions. V914 Cast Iron Flanged Swing Check Valve V914

Installation, Operating and Maintenance Instructions. V914 Cast Iron Flanged Swing Check Valve V914 Installation, Operating and Maintenance Instructions V914 Cast Iron Flanged Swing Check Valve V914 THE PRESSURE EQUIPMENT DIRECTIVE 97/23/EC and CE MARKING The Pressure Equipment Regulations 1999 (SI 1999/2001)

More information

MANUAL SEALLESS STEEL STRAPPING TOOL MODEL A

MANUAL SEALLESS STEEL STRAPPING TOOL MODEL A OPERATION MANUAL / SPARE PARTS LIST MANUAL SEALLESS STEEL STRAPPING TOOL MODEL A337.0001 13.1912.01 13191201.en/MAS/ 10.02 INDEX PAGE 1 SAFETY INSTRUCTIONS 2 2 WARRANTY CONDITIONS AND LIABILITY 4 3 APPROPRIATE

More information

Operation & Maintenance instructions for the Söll Vi-Go type fall protection system with guided type fall arrester

Operation & Maintenance instructions for the Söll Vi-Go type fall protection system with guided type fall arrester Operation & Maintenance instructions for the Söll Vi-Go type fall protection system with guided type fall arrester Söll Vi-Go is personal protective equipment against falls from a height: Guided type fall

More information

CALIBRATION AND ADJUSTMENT

CALIBRATION AND ADJUSTMENT CALIBRATION AND ADJUSTMENT All SMART pipettes are been quality tested according to ISO8655/DIN 12650. The quality control according to ISO 8655/DIN 12650 involves gravimetric testing of each pipette with

More information

Operating and maintenance manual Filter and reducing station Series / 1.0

Operating and maintenance manual Filter and reducing station Series / 1.0 Operating and maintenance manual Filter and reducing station Series 961 04.2017 / 1.0 Original instructions ARCA Regler GmbH. All rights reserved. Cover picture background: Freepik.com ARCA Regler GmbH

More information

Helium-Oxygen Blender

Helium-Oxygen Blender Helium-Oxygen Blender Service Manual Model No. PM5400 Series PM5500 Series (shown) SAVE THESE INSTRUCTIONS 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA 18067 USA Fax: (+001) 610-262-6080 www.precisionmedical.com

More information

ATS430 turbidity sensor Retractable insertion assembly

ATS430 turbidity sensor Retractable insertion assembly A MEASUREMENT & ANALYTICS INSTRUCTION ATS430 turbidity sensor Retractable insertion assembly Measurement made easy 1 Introduction This publication details installation procedures for the retractable insertion

More information

INSTRUCTIONS FOR USE (US)

INSTRUCTIONS FOR USE (US) INSTRUCTIONS FOR USE (US) RaplixaSpray TM, Raplixa TM Delivery Device For the application of Raplixa powder to stop bleeding. Raplixa is a hemostatic fibrin sealant product. It is a powder formulation

More information

Type Operating Instructions. 2/2-way solenoid valve 2/2-Wege Magnetventil Electrovanne 2/2 voies.

Type Operating Instructions. 2/2-way solenoid valve 2/2-Wege Magnetventil Electrovanne 2/2 voies. 2/2-way solenoid valve 2/2-Wege Magnetventil Electrovanne 2/2 voies We reserve the right to make technical changes without notice. Technische Änderungen vorbehalten. Sous réserve de modifications techniques.

More information

Speed control assembly model A-1

Speed control assembly model A-1 Pressure Regulation 533a 1. DESCRIPTION The Viking Speed Control Assembly provides adjustment of the opening speed of Viking Deluge Valves, and adjustment of both the opening and closing speed of Viking

More information

Gebrauchsanweisung Mode d' emploi Instrucciones para el uso Istruzioni per l'uso Инструкции по зксплуатации. ri-fox N

Gebrauchsanweisung Mode d' emploi Instrucciones para el uso Istruzioni per l'uso Инструкции по зксплуатации. ri-fox N Rudolf Riester GmbH Postfach 35 DE-72417 Jungingen Germany Tel.: +49 (0)74 77-92 70-0 Fax: +49 (0)74 77-92 70-70 info@riester.de www.riester.de Operator s manual Gebrauchsanweisung Mode d' emploi Instrucciones

More information

AMBLER SURGICAL INSTRUCTIONS FOR USE - OPHTHALMIC DUET-STYLE SCISSORS

AMBLER SURGICAL INSTRUCTIONS FOR USE - OPHTHALMIC DUET-STYLE SCISSORS AMBLER SURGICAL INSTRUCTIONS FOR USE - OPHTHALMIC DUET-STYLE SCISSORS Ambler Item # 6112E Duet-style MICS Packer/Chang IOL cutter, 1 3/8", 19 gauge, straight shaft, for use with Ambler #6100T INDICATIONS

More information

USER MANUAL. AIR-OXYGEN BLENDER (DISS and NIST Connections) Model No. PM5200 Series PM5300 Series (shown) SAVE THESE INSTRUCTIONS

USER MANUAL. AIR-OXYGEN BLENDER (DISS and NIST Connections) Model No. PM5200 Series PM5300 Series (shown) SAVE THESE INSTRUCTIONS USER MANUAL (DISS and NIST Connections) Model No. PM5200 Series PM5300 Series (shown) SAVE THESE INSTRUCTIONS CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician.

More information

XpressFill XF2500 Counter Pressure Filler. Operating Instructions

XpressFill XF2500 Counter Pressure Filler. Operating Instructions XpressFill XF2500 Counter Pressure Filler Operating Instructions ii XPRESSFILL Systems LLC lling machine. ank you for choosing our handcrafted bottle filler as the technology to bottle your passion. We

More information

TJF-Q180V Cleaning and Disinfection Checklist

TJF-Q180V Cleaning and Disinfection Checklist TJF-Q180V Cleaning and Disinfection Checklist TJF-Q180V Cleaning and Disinfection Checklist This checklist is used to evaluate and confirm if cleaning and disinfection of the TJF-Q180V has been performed

More information

OPERATION MANUAL Please read this Operation Manual carefully before use, and file for future reference.

OPERATION MANUAL Please read this Operation Manual carefully before use, and file for future reference. English Lubrication Free Air Turbine Handpiece with Water Spray OPERATION MANUAL Please read this Operation Manual carefully before use, and file for future reference. OM-T0286E 001 Thank you for purchasing

More information

Instructions for Use (IFU) Electrical Leg- & Arm lift, model , ver. 2012

Instructions for Use (IFU) Electrical Leg- & Arm lift, model , ver. 2012 Instructions for Use (IFU) Electrical Leg- & Arm lift, model 2009-01, ver. 2012 1 Opitek International Always be careful and pay attention when using the equipment to avoid over strain of limbs. To avoid

More information

UsER manual for Watersens ph -REDOX

UsER manual for Watersens ph -REDOX UsER manual for Watersens -REDOX Cl 8 1 2 6 3 3 7 7 4 4 4 4 Parts List 1 Redox Probe 1 x 2 PH Probe 1 x 5 Tube Weight 2 x 6 Connection Valve 1 x chlorine 3 Chlorine and Pumps 2 x 7 Dosing Valve 2 x 5 5

More information

J138RSH and J150RSH 3/8" J138RSH 1/2" J150RSH. 3/8" Drive Sealed Head Ratchet 1/2" Drive Sealed Head Ratchet. Operation and Maintenance Manual WARNING

J138RSH and J150RSH 3/8 J138RSH 1/2 J150RSH. 3/8 Drive Sealed Head Ratchet 1/2 Drive Sealed Head Ratchet. Operation and Maintenance Manual WARNING J138RSH and J150RSH Operation and Maintenance Manual 3/8" J138RSH 1/2" J150RSH WARNING To reduce the risk of injury, read and understand these safety warnings and instructions before using the tool. Keep

More information

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR: PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: 1. **Identify and name the filters on the 7200ae. 2. **Explain how each filter is sterilized. 3. **Trace the gas flow through the ventilator

More information

COMBINATION AIR RELEASE DEGASSING (CARD) VALVES INSTALLATION AND MAINTENANCE MANUAL

COMBINATION AIR RELEASE DEGASSING (CARD) VALVES INSTALLATION AND MAINTENANCE MANUAL COMBINATION AIR RELEASE DEGASSING (CARD) VALVES INSTALLATION AND MAINTENANCE MANUAL SPECIFICATIONS: The CARD series air valves are available in 3 pipe sizes, 1, 2 and 4 NPT or socket. Maximum inlet pressure

More information

Type /2-Way Globe Valve 3/2-Wege-Geradsitzventil Vanne à siège droit 3/2 voies Operating Instructions Bedienungsanleitung Manuel d utilisation

Type /2-Way Globe Valve 3/2-Wege-Geradsitzventil Vanne à siège droit 3/2 voies Operating Instructions Bedienungsanleitung Manuel d utilisation 3/2-Way Globe Valve 3/2-Wege-Geradsitzventil Vanne à siège droit 3/2 voies Operating Instructions Bedienungsanleitung Manuel d utilisation We reserve the right to make technical changes without notice.

More information

20 Ton SD Shop Press Operating Instructions

20 Ton SD Shop Press Operating Instructions 20 Ton SD Shop Press Operating Instructions MODEL NO. 850SD Hazard Symbols Used in the Manuals This manual includes the hazard symbols defined below when the operations or maintenance job involves a potential

More information

75 Ton SD Shop Press Operating Instructions

75 Ton SD Shop Press Operating Instructions 75 Ton SD Shop Press Operating Instructions MODEL NO. 856SD Hazard Symbols Used in the Manuals This manual includes the hazard symbols defined below when the operations or maintenance job involves a potential

More information

PremAire INSTRUCTIONS FOR VORTEX TUBE MODE OF OPERATION

PremAire INSTRUCTIONS FOR VORTEX TUBE MODE OF OPERATION DUAL SUPPLY REGULATOR ESCAPE CYLINDER VORTEX MAIN INLET PremAire INSTRUCTIONS FOR VORTEX TUBE MODE OF OPERATION WARNING THIS MANUAL MUST BE READ CAREFULLY BY ALL PERSONS WHO HAVE OR WILL HAVE THE RESPONSIBILITY

More information

Pressure Equipment Directive (PED) 97/23/EC Page 033 of 124

Pressure Equipment Directive (PED) 97/23/EC Page 033 of 124 Pressure Equipment Directive (PED) 97/23/EC Page 033 of 124 13.7 Pressure Equipment Directive (PED) 97/23/EC 1 The Pressure Equipment Directive (PED) 97/23/EC applies to the design, manufacturing and conformity

More information

Operation and Maintenance Manual for GENTEC Model 881VR Continuous/Intermittent Digital Suction Regulators

Operation and Maintenance Manual for GENTEC Model 881VR Continuous/Intermittent Digital Suction Regulators Operation and Maintenance Manual for GENTEC Model 881VR Continuous/Intermittent Digital Suction Regulators Genstar Technologies Co., Inc. 4525 Edison Avenue Chino, CA 91710 USA TEL 909-0-272 FAX 909-0-485

More information

Valves. Diaphragm shut-off valves, kidney-type

Valves. Diaphragm shut-off valves, kidney-type Valves Diaphragm shut-off valves, kidney-type Application When leakage of even the smallest quantities must be avoided with certainty For maximum requirements to operating safety For maximum requirements

More information

Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST Industrial Drive, Orlinda, TN 37141

Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST Industrial Drive, Orlinda, TN 37141 0 Operation Manual Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST 615-654-4441 sales@specialtyh2o.com 615-654-4449 fax TABLE OF CONTENTS Section 1 GENERAL 1.1 Warnings and

More information

F-330 Instruction manual. Magic Glove Bio Skin Smoother High Frequency Ultrasound Galvanic

F-330 Instruction manual. Magic Glove Bio Skin Smoother High Frequency Ultrasound Galvanic F-330 Instruction manual Magic Glove Bio Skin Smoother High Frequency Ultrasound Galvanic Warning Never under any circumstances attempt to open or inspect the internal components or accessories of the

More information

Service and Repair Manual

Service and Repair Manual II stage R2 Ice/ Special, II stage R 1 Pro DOWNSTREAM 2 nd STAGE REGULATOR Service and Repair Manual Introduction Safety Precautions...4 General Procedures, Maintenance Schedules...5 Initial Inspection

More information

User s Manual. USB Golf Simulator. Contents

User s Manual. USB Golf Simulator. Contents USB Golf Simulator User s Manual Contents safety information...1 identifying the components...2 assembly in 3 easy steps...3 plug & play...4 playing tips...5 care...6 troubleshooting...7 notes...8 warranty...9

More information

ROTATING DISK VALVES INSTALLATION AND MAINTENANCE 1. SCOPE 3 2. INFORMATION ON USAGE 3 3. VALVE TYPES 3 4. OPERATORS 5 5. VALVE CONSTRUCTION 6

ROTATING DISK VALVES INSTALLATION AND MAINTENANCE 1. SCOPE 3 2. INFORMATION ON USAGE 3 3. VALVE TYPES 3 4. OPERATORS 5 5. VALVE CONSTRUCTION 6 Sub Section INDEX Page Number 1. SCOPE 3 2. INFORMATION ON USAGE 3 3. VALVE TYPES 3 4. OPERATORS 5 5. VALVE CONSTRUCTION 6 6. INSTALLATION AND OPERATION 6 7. MAINTENANCE 8 8. REPAIR 9 9. ASSEMBLY 10 10.

More information

VWR UHP j :01 Page 1. ENGLISH 1 22 DEUTSCH FRANÇAIS ESPAπOL ITALIANO

VWR UHP j :01 Page 1. ENGLISH 1 22 DEUTSCH FRANÇAIS ESPAπOL ITALIANO VWR UHP 01.09.2010 5j 240.9001.0.09 1-09-10 16:01 Page 1 1 22 DEUTSCH 23 44 FRANÇAIS 45 66 ESPAπOL 67 88 ITALIANO 89 110 VWR UHP 01.09.2010 5j 240.9001.0.09 1-09-10 16:01 Page 2 Pipettors are equipped

More information

Laboratory Mortar Mixer (Testing)

Laboratory Mortar Mixer (Testing) TomTom-Tools GmbH Zelgli 20 8905 Arni info@tomtom-tools.com Switzerland www.tomtom-tools.com User Manual Version February 22, 2015 Laboratory Mortar Mixer (Testing) 1 Introduction The Laboratory Mortar

More information

FLOWMETER MODELS: 1MFA, 4MFA, 6MFA and 8MFA Series SAVE THESE INSTRUCTIONS

FLOWMETER MODELS: 1MFA, 4MFA, 6MFA and 8MFA Series SAVE THESE INSTRUCTIONS USER MANUAL FLOWMETER MODELS: 1MFA, 4MFA, 6MFA and 8MFA Series 1MFA3001 (Shown) 8MFA1001 (Shown) SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

More information

Installation instructions for the Yepp Maxi Easyfit

Installation instructions for the Yepp Maxi Easyfit Installation instructions for the Yepp Maxi Easyfit The Yepp Maxi Easyfit is suitable for: - fitting to an Easyfit luggage carrier* (fig 1A) or - fitting in combination with the Yepp Maxi Easyfit carrier

More information

Animal Bed Mouse Cryo

Animal Bed Mouse Cryo Animal Bed Mouse Cryo User Manual Version 003 Innovation with Integrity NMR Copyright by Bruker Corporation All rights reserved. No part of this publication may be reproduced, stored in a retrieval system,

More information

Varipette Operating manual

Varipette Operating manual Operating manual Copyright 2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior permission of the copyright owner. Eppendorf and the Eppendorf logo are registered

More information

M3-LED. Operator s Manual. Operator s Manual MADE IN THE USA USA

M3-LED. Operator s Manual. Operator s Manual MADE IN THE USA USA M3-LED Operator s Manual Operator s Manual MADE IN THE USA USA Section TABLE OF CONTENTS Page Warnings and Cautions... 1 M3-LED Parts Diagram... 4 Mounting... 6 Dismounting... 8 Switch Operation... 10

More information

SEQUENTIAL SHOWER VALVE INSTRUCTION MANUAL PLEASE LEAVE THIS MANUAL WITH THE END USER

SEQUENTIAL SHOWER VALVE INSTRUCTION MANUAL PLEASE LEAVE THIS MANUAL WITH THE END USER SEQUENTIAL SHOWER VALVE INSTRUCTION MANUAL PLEASE LEAVE THIS MANUAL WITH THE END USER CONTENTS 1. INTRODUCTION & SAFETY 1 2. DIMENSIONS 2 3. TECHNICAL DATA 3 4. OPERATION 4 5. COMPONENTS 5 6. SITE INSTALLATION

More information

USER MANUAL. JET TST and JET TST S.

USER MANUAL. JET TST and JET TST S. USER MANUAL JET TST and JET TST S www.sisteven.com Types of Jet fans: JET TST UNI: Impulse axial fan single direction JET TST S UNI: Impulse axial fan single direction (short case) JET TST REV: Impulse

More information

EasyFit Lite Nasal Mask, Silicone. Nasal Mask. EasyFit Lite Nasal Mask. Device Description and Instructions for Use. Caution:

EasyFit Lite Nasal Mask, Silicone. Nasal Mask. EasyFit Lite Nasal Mask. Device Description and Instructions for Use. Caution: EasyFit Lite Nasal Mask EasyFit Lite Nasal Mask, Silicone Nasal Mask Device Description and Instructions for Use Caution: Federal Law (U.S.A.) restricts this device to sale by or on the order of a physician.

More information

STEYR evo 10 E STEYR evo 10 E Compact

STEYR evo 10 E STEYR evo 10 E Compact TM STEYR evo 10 E STEYR evo 10 E Compact MATCH Bedienungsanleitung Operator's manual Mode d'emploi Instrucciones de funcionamiento STEYR evo 10 E STANDARD STEYR evo 10 E COMPACT ATTENTION This operator

More information

Instruction Manual for Configura Cushionair Portable Pump

Instruction Manual for Configura Cushionair Portable Pump Instruction Manual for Configura Cushionair Portable Pump Fitted with battery powered pump, suitable for Configura Portable chairs V E R S I O N O N E M A Y 2 0 1 6 Contents Introduction 3 Set up of Cushionair

More information

Product Design Specifications

Product Design Specifications Product Design Specifications Hydrocephalus Shunt Valve February 2, 2017 Team: Emma Alley (Leader), Andrew Miller (Communicator/BPAG), Karl Fetsch (BWIG), Catharine Flynn (BSAC) Advisor: Beth Meyerand

More information

Gas Pressure Regulator HON 300

Gas Pressure Regulator HON 300 Product information serving the gas industry worldwide Applications, characteristics, technical data Applications direct acting gas pressure regulator, for systems in accordance with DVGW working instruction

More information