Equipment Testing and Validation ASSESSMENT OF FRACTIONAL EXPIRED GASES AND AIR FLOW BY AN AMBULATORY METABOLIC ANALYZER

Size: px
Start display at page:

Download "Equipment Testing and Validation ASSESSMENT OF FRACTIONAL EXPIRED GASES AND AIR FLOW BY AN AMBULATORY METABOLIC ANALYZER"

Transcription

1 Ambulatory Metabolic Analysis 20 JEPonline Journal of Exercise Physiologyonline Official Journal of The American Society of Exercise Physiologists (ASEP) ISSN An International Electronic Journal Volume 2 Number 4 October 1999 Equipment Testing and Validation ASSESSMENT OF FRACTIONAL EXPIRED GASES AND AIR FLOW BY AN AMBULATORY METABOLIC ANALYZER PATRICK L. JACOBS, MARK S. NASH AND C. DAVID MINTZ Departments of Neurological Surgery and Orthopaedics & Rehabilitation, and The Miami Project University of Miami School of Medicine, Miami, Florida ABSTRACT ASSESSMENT OF FRACTIONAL EXPIRED GASES AND AIR FLOW BY AN AMBULATORY METABOLIC ANALYZER. PATRICK L. JACOBS, MARK S. NASH AND C. DAVID MINTZ. JEPONLINE, 1999, 2(4): Fractional expired oxygen (F E O 2 ) and carbon dioxide (F E CO 2 ), and air volume flow were compared using an ambulatory metabolic analyzer (KB1-C, Aerosport, Inc.) and a laboratory metabolic analyzer (Horizon MMC, Sensormedics, Inc.). Ten subjects with paraplegia exercised to exhaustion using their arms, and ten nondisabled subjects ran to exhaustion on a treadmill. Simultaneous F E O 2 and F E CO 2 measurements were taken during exercise by inserting the KB1-C pneumotachometer in series with the MMC. Measurement accuracy of air flow was examined by sending known air volumes through the KB1-C at three pneumotachometer flow settings (low = 3 to 30 L. min -1 ; medium = 3 to 120 L. min -1 ; and high = 3 to 200 L. min -1 ). Results showed a significant positive association for F E O 2 (R 2 = 0.94) with minimal but statistically significant absolute differences (SE E = 0.26%, p<0.001)). The R 2 for F E CO 2 was 0.89, although average KB1-C measurements were slightly lower (SE E = 0.28%, p<0.001). Differences from the reference standard for the low flow pneumotachometer were 12.3 and 15.5% at 3 L/min and 6 L/min, respectively, but averaged between 2.5 and 4.0% at 9-45 L/min. For the medium flow setting, average differences from the reference standard ranged from 0.5% at 72 L/min to 4.6% at 117 L/min. For the high flow setting, measurements varied dramatically from the reference standard at flow rates up to nine L/min, and by % at 27 to 54 L/min. Otherwise, differences of less than 2% were observed at rates of L/min. This study observed similar measurements for F E O 2 and F E CO 2 when measured by an ambulatory metabolic analyzer and a laboratory standard, and found that accurate measurement of air flow depended upon careful matching of the pneumotachometer setting with the expected air volume ranges. Key Words: Exercise, Metabolism, Ventilation, Fractional Gases, Portable INTRODUCTION Metabolic analysis using open-circuit spirometry is a widely used standard for assessing fitness, calculating work intensities, and estimating caloric expenditures. Unfortunately, the size, relative immobility, and need for residential/commercial (110 or 240 V) electrical current restricts the use of most metabolic analyzers to the controlled laboratory setting. Such restrictions also limit the ability of investigators to address study questions about physical activity when conducted under field conditions of ambient temperature and humidity. Widespread use and universal acceptance of portable

2 Ambulatory Metabolic Analysis 21 metabolic analyzers has been challenged by their size (1-3), lack of direct carbon dioxide (CO 2 ) gas measurement (4-7), and poor measurement validity 3,5,9,11). However, an ambulatory metabolic testing system claiming to solve these problems has recently been marketed (Aerosport KB1-C, Aerosport, Inc., Ann Arbor, MI). This system weighs about 1.0 kg, has dimensions that favor its portability, measures both oxygen (O 2 ) and CO 2 gases, and operates under rechargeable battery power. Unlike many laboratory analyzers, it uses a flat plate pressuredifferential pneumotachometer to measure expired gas volumes, which eliminates the need for bulky expired gas lines. In addition, the sampling of expired air from the pneumotachometer eliminates the needs for an expired gas-mixing chamber. Despite favorable physical characteristics and portability, a study comparing measurements obtained from the KB1-C unit with those of an industry standard has yet to be reported for both rest and exercise conditions. Various study methods have been used to make such comparisons for other portable metabolic analyzers. Some studies have compared VO 2 measurements at matched work intensities during separate exercise tests that use identical test protocols (1,4-7). Such analyses may be compromised by small differences in day-to-day exercise physiology and test-retest calibration errors. Other studies have simultaneously measured VO 2 during the same exercise test by placing the portable gas analyzer in series (2,3) with a laboratory standard. Unfortunately, the physics of air flow by which a pneumotachometer accurately measures ventilation volume may render such methods suspect. For example, accurate pneumotachometer measurement of minute ventilation (V E ) favors bidirectional air flow under low pressure conditions (11), while many open-circuit methods used for such comparisons employ unidirectional flow in a closed circuit of descending ventilation tube diameters and building pressure. The accuracy of in-series air flow comparisons may therefore be compromised, as the KB1-C computer program assumes the air pressure on the downstream side of its pneumotachometer plate (by calibration) to be ambient. The accurate determination of VO 2 and VCO 2 by open-circuit spirometry requires simultaneous measurement of fractional expired gasses (O 2 and CO 2 ) and minute ventilation. Several investigations have compared VO 2 alone, or VO 2 and V E together (2-6) during graded and steady state work. For those cases in which differences between portable and laboratory standards were observed, the source of this difference has never been established (e.g. measurement error for gas concentration versus volume). The isolation of such measurement error is important, as the KB1-C permits the selection of three fixed pneumotachometer settings. While each pneumotachometer may be considered accurate throughout a defined flow range, that specified range may be exceeded during exercise testing. Measurement of VO 2 alone, as reported by most studies, fails to identify the source of this potential error. Thus, the purpose of this study was twofold. First, the study compared simultaneous measurements of fractional expired oxygen (F E O 2 ) and carbon dioxide (F E CO 2 ) gas concentrations obtained during graded exercise on the KB1-C ambulatory metabolic analyzer with those of a laboratory standard. Second, we compared KB1-C pneumotachometer volume measurements obtained both within and outside its three rated ranges with those of a calibration standard. METHODS Instrumentation The KB1-C is a battery powered, microprocessorcontrolled metabolic analyzer designed for ambulatory and field-based studies. The unit computes VO 2, VCO 2, and respiratory exchange ratios using measurements derived from gas volume flow and fractional inspired and expired gas concentrations. Ventilation volumes are calculated using a flat-plate orifice within an open pneumotachometer. Air flow over the orifice produces a drop in pressure proportional to the square root of its flow. Pressure differentials across the plate are measured via two pneumotachometer

3 Ambulatory Metabolic Analysis 22 pressure lines using a bonded strain gauge positioned across the orifice plate. The pneumotachometer has three flow ranges (low, medium, and high). The low flow range (LFR) is recommended for flow rates between two and 30L/min, the medium flow range (MFR) from 10 to 120 L/min, and the high flow range (HFR) from 25 to 200 L/min. Fractional concentrations of expired O 2 and CO 2 are measured in gas samples drawn from the pneumotachometer. Oxygen concentration is assayed by a galvanic fuel cell contained in the unit, and computed using a line of identity drawn through null and the ambient air O 2 concentration (20.93%). Carbon dioxide concentration is measured using nondispersive infrared gas analysis, which uses ambient CO 2 as a reference. These O 2 and CO 2 values are then corrected for ambient temperature and pressure. The reference system used in this study was a Horizon System MMC (Sensormedics, Inc.). This system also measures O 2 and CO 2 concentrations using a galvanic fuel cell and nondispersive infrared gas analysis, respectively. Both analyzers were calibrated before each exercise test using certified gases of known concentration. Subjects Twenty male volunteers between the ages of years participated in this study. Ten of these subjects had neurologically complete paraplegia (ASIA A) between T 5 and T 12 for more than one year. The remaining ten subjects had no neurological deficits. All subjects were in good health - operationally defined as asymptomatic for acute treatable illnesses - and had no known musculoskeletal or cardiac conditions that would otherwise limit their exercise activities. Written consent to participate in this study was obtained according to guidelines of the University of Miami Medical Sciences Subcommittee for the Protection of Human Subjects. Exercise Testing All subjects underwent graded exercise testing to volitional fatigue. Criteria for exercise termination were consistent with those recommended by the American College of Sports Medicine (8). Subjects with paraplegia underwent continuous graded arm tests on an Upper Body Ergometer (Cybex, Inc., Owatonna, MN) starting at an initial workload of 400 kg/m/min and progressing by 100 kg/m/min every three minutes. Subjects without neurological deficits underwent graded continuous treadmill exercise using the protocol of Bruce (9). Gas and Volume Measurements Fractional expired gas percentages for O 2 and CO 2 during each exercise test were simultaneously measured using the KB1-C and the MMC. These systems were placed in series, with the KB1-C MFR pneumotachometer positioned immediately before the gas mixing chamber of the MMC. To ensure complete purging of calibration and ambient gases from the mixing chamber and gas sampling lines, resting gas collection was started four minutes before the start of exercise testing. Measurement of Gas Volumes The accuracy of gas flow measured by the three KB1-C pneumotachometers settings was assessed on another occasion by manually injecting ambient air through the pneumotachometers. A certified three liter syringe (Hans Rudolf Inc., Kansas City, MO) was used to pass known volumes of air through the KB1-C pneumotachometer within 20 second periods, the shortest time interval over which the system can be programmed to calculate and display information. The pneumotachometer was connected to the syringe by a six centimeter length of 3.5 centimeter diameter corrugated plastic tubing. Each of the three pneumotachometer range settings (LFR, MFR, and HFR) was tested at multiple flow rates across and beyond the ventilation ranges recommended by the manufacturer. The manufacturer recommended ranges for the LFR, MFR, and HFR are two to 30 L/min, 10 to 120 L/min, 25 to 200 L/min, respectively. The three pneumotachometer settings were evaluated at three, six, and nine L/min, and at increasing nine L/min intervals thereafter. The upper limits of the flow rate testing were 63, 126 and 216 L/min for the LFR, MFR, and HFR, respectively. Thus this protocol

4 Ambulatory Metabolic Analysis 23 allowed the examination of possible measurement inaccuracies for flow rates exceeding the recommendations for each setting. Ten ventilation tests were performed at each of these flow rates, making a grand total of 510 flow tests. Statistical Analysis Fractional expired gas concentrations were examined between KBC-1 and MMC by pooling the simultaneous measurements of the twenty subjects across workloads. Regression analyses were performed, and correlation coefficients calculated, which examined the associations between the simultaneously derived KB1-C and MMC measurements of F E O 2 and F E CO 2. Measurements derived from the two metabolic analyzers were compared using Student s paired t-tests. Standard errors of estimate (SE E ) were calculated using the formula: SE E = S y (1- r 2 ) ½, where r 2 is the reliability coefficient and S y is the standard deviation of the MMC values. Comparison of reference gas volumes with those obtained from the KB1-C was made using Student s paired t-tests. Means and standard deviations for KB1-C measurements were calculated for each flow rate. The measurements of air flow made by the KB1- C were compared with those of the standard across all tested flow rates for each of the three pneumotachometer settings (Student s paired t-tests and SE E ). Linear regression analysis was performed for each pneumotachometer range to examine the association between values obtained from the KB1-C and the reference system. Statistical significance for all tests was accepted at p<0.05. Based on 20 subjects, acceptable power of at least 0.80, and a standard deviation for ventilation of 0.25 L/min, the mean difference that could be detected was 0.20 L/min. Pooling the data by the ten trial repetitions and multiple pneumotachometer settings provided additional improvements to statistical power. RESULTS Comparisons of Reference and KB1-C Measurements for FEO 2 and F E CO 2 Figure 1 shows subpeak and peak F E O 2 values simultaneously measured by the MMC analyzer and the KB1-C. A significant positive association for this relationship was obtained (r 2 =0.94), although significant differences between the two means were also observed. These differences were explained by lower F E O 2 values measured by the KB1-C than the MMC (p<0.001). However, the magnitude of test differences was small, as the SE E for F E O 2 was only 0.26%. This was less than 1.7% of the mean reference values for F E O 2 and well within the sensitivity error for many metabolic analyzers (10). When examining the same relationships for F E CO 2, the coefficient of determination was 0.89, although the means measured by the KB1-C were lower than the MMC (p<0.001) (Figure 2). The SE E for F E CO 2 values was 0.28 %, which was 6.5% of the mean reference value. Figure 1 - The relationship between expired air O 2 concentration measured using the Sensormedics MMC and the Aerosport KB1-C during graded incremental exercise. Comparisons of Reference and KB1-C Measurements for Air Flow Measurements of air flow for all three pneumotachometer settings are shown in Table 1. These values represent the mean and standard deviation of ten tests at each selected flow volume. For the LFP, large differences from the reference

5 Ambulatory Metabolic Analysis 24 Table 1: Air volume measurements (mean±sd) and mean errors using three pneumotachometer settings of the Aerosport KB1-C ambulatory metabolic analyzer (ten tests each) in reference to volumes of air injected with a calibration syringe. High Flow Calibation Low Flow Medium Flow (L/min) mean ± SD mean error mean ± SD Mean error mean ± SD mean error ± 0.22 (12.3%) 1.89 ± 0.37 (37.0%) 0.15 ± 0.05 (95.0%) ± 0.08 (15.5%) 5.82 ± 0.15 (3.0%) 0.65 ± 0.29 (89.2%) ± 0.07 (4.0%) 8.84 ± 0.08 (1.8%) 2.87 ± 0.57 (68.1%) ± 0.07 (3.5%) ± 0.09 (2.9%) ± 0.44 (14.1%) ± 0.41 (2.8%) ± 0.55 (1.5%) ± 0.21 (8.8%) ± 0.27 (3.2%) ± 0.47 (2.6%) ± 0.23 (8.2%) ± 0.08 (2.5%) ± 0.78 (1.5%) ± 0.22 (7.5%) ± 3.71 (10.5%) ± 0.32 (0.9%) ± 0.41 (7.0%) ± 0.74 (19.3%) ± 0.27 (1.0%) ± 0.37 (5.6%) ± 0.21 (0.5%) ± 0.29 (3.9%) ± 0.24 (0.6%) ± 0.26 (3.2%) ± 0.60 (0.5%) ± 0.46 (2.2%) ± 0.41 (0.8%) ± 0.31 (0.5%) ± 1.10 (1.1%) ± 0.40 (1.2%) ± 1.51 (4.6%) ± 0.41 (1.2%) ± 1.49 (7.5%) ± 0.36 (1.6%) ± 0.60 (0.8%) ± 0.95 (1.5%) ± 0.34 (2.3%) ± 0.57 (0.8%) ± 0.56 (0.6%) ± 2.83 (1.3%) ± 1.68 (1.2%) ± 1.58 (1.8%) ± 1.58 (3.1%) ± 2.78 (3.3%) standard were observed at flow rates of 3 L/min and 6 L/min (12.3 and 15.5% different, respectively). Differences averaged between 2.5 and 4.0% for rates ranging from 9-45 L/min, and increased to errors of 10.5% and 19.3% at 54 and 63 L/min, respectively. The SE E of the LFP measurements across flow rates ranging from 3 to 45 L/min was 4% of the mean reference values. Figure 3a compares these measurements with a line of unity. The greatest source of variation within this flow

6 Ambulatory Metabolic Analysis 25 range occurred at the two lowest flow rates tested, 3 and 6 L/min. The SE E of the LFP flow rate measurements obtained across flow rates ranging from 9 to 45 L/min was 2.4% of the mean of the reference system values. The absolute differences in measurement across this range (9 to 45 L/min) ranged from 0.63 to 1.12 L/min. flow rate measurements differed dramatically from the reference system values at flow rates equal to, or less than, 9 L/min; and by about 7-8% at the 27 to 54 L/min rates. Otherwise, differences of less than 2% from the standard were observed at flow rates of L/min. At flow rates from 72 to 216 L/min, the SE E of the HFP measurements represented 1.5% of the mean reference values. Within this range, the absolute differences between the HFP and the reference standard ranged from 0.46 to 3.64 L/min. DISCUSSION Figure 2 - The relationship between expired air CO 2 concentration measured using the Sensormedics MMC and the Aerosport KB1-C during graded incremental exercise. Air flow volume measurements using the MFP are reported in Table 1. The largest deviations from the standard in this range (3 to 126 L/min) occurred at flow rates well below and above the recommended range for the MFP. Otherwise, average differences from the reference ranged from 0.5% at 72 L/min to 4.6% at 117 L/min. Figure 3b displays the relationship of these measurements when compared with a line of unity. The SE E for the MFP values at flow rates between 6 and 117 L/min was 2.9% of the mean reference values and the absolute differences ranged from 0.18 to 5.35 L/min. Mean values and standard deviations of the HFP flow volume assessments are shown in Table 1, while Figure 3c graphically presents the comparisons of the KB1-C measurements with the line of unity. The HFP The findings of this study investigating the accuracy of an ambulatory metabolic analyzer agree with those of previous reports showing reliable and valid measurements between selected portable systems and standard laboratory metabolic analysis models. This consistency appears to be dependent upon direct measurement of both O 2 and CO 2, as the accuracy of measurement for portable analyzers that algebraically computed VCO 2 from a measured VO 2 and an assumed RER of unity failed to be as accurate (4,5,7). Several studies have compared a portable analyzer having both O 2 and CO 2 sensors with laboratory standards. For the Teem 100 (11), VO 2 measured during graded work performed on two occasions showed no differences from VO 2 measured with a computer-based component system using a ventilation meter and gas analyzers during three steady-state exercise trials of walking, jogging, and running (1). Others have placed this portable analyzer in series with other laboratory analyzers. In one study, no significant differences of VO 2 were observed between the Teem 100 and a laboratory standard during incremental treadmill running and stepping (3). In another study where the analysis systems were placed in series, no differences were observed for VO 2 during incremental cycling (2). The present investigation was novel in several respects. This study was the first to assess the measurement accuracy of O 2 and CO 2 gas sensors and the expired air flow in an ambulatory metabolic analysis system. Previous ambulatory devices were

7 Ambulatory Metabolic Analysis 26 Figure 3 - The relationship of air flow volume measured with a calibration syringe and the Aerosport KB1-C using the low flow (A), the medium flow (B), and the high flow pneumotachometer (C), relative to the lines of identity. A C B limited by assessment of metabolic activity based on direct measurement of F E O 2 and the assumption of an RER of unity (4-7). As accurate computation of VO 2 and VCO 2 require measurement of fractional expired gases for O 2 and CO 2, the design of this study allowed us to detect whether errors would be introduced into this computation by inaccuracies in direct measurement of expired gases. The O 2 analyzer used in the KB1-C is a galvanic fuel cell like that used in the MMC. Measurement of F E O 2 showed a strong correlation between the analyzers, with an R 2 of Direct comparisons showed significantly lower values when measured by the KB1-C, although the magnitude of the difference was small and its clinical significance questionable. The R 2 for F E CO 2 was 0.89, with the KB1-C also producing slightly lower values of FECO 2 than the MMC. It is important to note that the statistical significance of the relatively small differences between the MMC and the KB1-C assessments of F E O 2 and F E CO 2 are a function of the large amount of statistical power available as a result of a large sample size. The mean differences between the two means of measurement were 0.26% and 0.28% for F E O 2 and F E CO 2, respectively, which are minimal effect sizes when one considers the F E O 2 ranged during the testing from 14 to 19% and the F E CO 2

8 Ambulatory Metabolic Analysis 27 ranged from 2.5 to 6.5%. It is also reasonable to expect that small differences in measurement might be observed between the two units, as they draw their gas sample by different methods. The KB1-C draws a micro sample of expired gas passing through the pneumotachometer based upon pressure differences sensed across its fixed plate. The sample for each expired breath is drawn into the gas sensors, analyzed, and then averaged over a programmed sampling period of 20 seconds to one minute. By contrast, expired gasses entering the Horizon system undergo mixing in a three liter chamber, in this case slightly downstream from the pneumotachometer, from which a single sample is drawn and then analyzed. Thus, differences in methods of analysis due to the time of sampling may explain the small differences in the F E O 2 and F E CO 2 measurements. Unlike previous reports, this study compared air flow rates measured by the three system pneumotachometers with a calibrated standard. Each of these pneumotachometers has a recommended flow range. When steady state measurements within the recommended flow ranges are taken, it is more likely that measurement of flow is accurate. However, when graded work is performed, expired air volumes may approach or exceed the end-ranges for which they are rated. The pneumotachometer setting cannot be adjusted once exercise testing has been initiated. As such, the choice of pneumotachometer for the KB1-C becomes an important component in accurate measurement of exercise metabolism. For this reason, the three pneumotachometer settings were examined beyond the recommended ranges. No other study has undertaken such an examination which delimits the air flow ranges in which this analyzer can be used. For the low flow pneumotachometer rated from 2-30 L/min, errors were small for flow rates between 9-45 L/min. However, pneumotachometer flow measurements deviated from the calibration syringe by 0.37 L/min (12.3%) at 3 L/min and 0.93 L/min (15.5 %) at 6 L/min. These results indicate that the low flow pneumotachometer setting would best be used for resting studies, resting values taken before low intensity work, or measurement of peak work in highly deconditioned persons. Caution should be exercised when interpreting data derived with the low flow pneumotachometer in conditioned persons at moderate or high levels of exertion. For the medium flow head, which is rated for flow between L/min, error was less than 3.0% percent when measuring flow rates 6 L/min to 108 L/min, and 4.6% at 117 L/min. Interestingly, the flow measurements had less error for the medium than low pneumotachometer setting at all flow rates between 6 and 63 L/min. For the high flow setting, errors at or less than 3.3% were observed for flow rates ranging from L/min. However, from 27 to 72 L/min, errors of 3.9 to 8.8% were observed. The design of this study did not allow comparison of VO 2 simultaneously derived with the KB1-C and a laboratory standard. Open bi-directional flow is required for accurate air volume assessment with the KB1-C pneumotachometers. Therefore, assessment of V E with the KB1-C system in series with a unidirectional system of descending pathway diameter is rendered suspect. It is acknowledged that the assessment of VO 2 as measure of energy uptake is a primary utilization of metabolic analyzers. Future studies are recommended to examine the accuracy of VO 2, VCO 2, and RER as assessed with this ambulatory metabolic analysis system. CONCLUSIONS This study found that fractional gas measurements were similar when assayed by an ambulatory metabolic analyzer and a laboratory standard. Accurate measurement of air flow depended upon selection of a pneumotachometer that could accommodate the peak ventilation input. The MFR satisfied this need best, and was consistently accurate at flow rates between L/min. Therefore, this pneumotachometer setting is

9 Ambulatory Metabolic Analysis 28 appropriate for most activity assessments, but would exclude accurate metabolic measurement during graded incremental testing of conditioned persons who have higher maximal ventilation values. While use of the HFP may satisfy this need at maximal work, its use will sacrifice accuracy of some data collected at submaximal work intensities. Otherwise, the physical characteristics of the unit make it practical for field testing of human performance. REFERENCES 1. Melanson EL, Freedson PS, Hendelman D, Debold E (1996) Reliability and validity of a portable metabolic measurement system. Can J Appl Physiol 21(2): Novitsky S, Segal KR, Chartr-Aryamontri B, Guvakov D, Katch VL (1995) Validity of a new portable indirect calorimeter: the Aerosport TEEM 100. Eur J Appl Physiol 70: Wideman L, Stoudemire NM, Pass KA, McGinnes CL, Gaesser GA, Weltman A (1996) Assessment of the Aerosport TEEM 100 portable metabolic measurement system. Med Sci Sports Exerc 28(4): American College of Sports Medicine (1995) ACSM s Resource Guidelines for Exercise Testing and Prescription. Baltimore, MD: Williams and Wilkins; Bruce RA, Kusumi F, Hosmer D (1973) Maximal oxygen intake and nomographic assessment of functional aerobic impairment in cardiovascular disease. Am Heart J 85: Versteeg PGA, Kippersluis GJ. (1989) Automated systems for measurement of oxygen uptake during exercise testing. Int J Sports Med 10: Aerosport Inc. (1995) KB1-C: Ambulatory Metabolic Measurement System (1995) Ann Arbor, MI:Aerosport Inc. Address Correspondence to: Patrick L. Jacobs, Ph.D., The Miami Project to Cure Paralysis, University of Miami School of Medicine, 1600 Northwest 10th Avenue, R-48, Miami, Florida 33136; Phone: ; Fax: PJACOBS@miamiproj.med.miami.edu 4. Crandall DG, Taylor SL, Raven PB (1994) Evaluation of the Cosmed K2 portable telemetric oxygen uptake analyzer. Med Sci Sports Exerc 26(1): Lothian F, Farrally MR, Mahoney C (1993) Validity and reliability of the Cosmed K2 to measure oxygen uptake. Can J Appl Physiol 18(2): Lucia A, Fleck SJ, Gotshall RW, Kearney JT (1993) Validity and reliability of the Cosmed K2 instrument. Int J Sports Med 14(7): Peel C, Utsey C (1993) Oxygen consumption using the K2 telemetry system and a metabolic cart. Med Sci Sports Exerc 25(3):

Validity of a new portable indirect calorimeter: the AeroSport TEEM 100

Validity of a new portable indirect calorimeter: the AeroSport TEEM 100 Eur J Appl Physiol (1995) 70:462-467 Springer-Verlag 1995 S. Novitsky K. R. Segal B. Chatr-Aryamontri D. Guvakov V. L. Katch Validity of a new portable indirect calorimeter: the AeroSport TEEM 100 Accepted:

More information

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance Introduction Expired Gas Analysis or indirect calorimetry, can be used to measure ventilation and the fractions of oxygen and carbon dioxide in expired air. From these measurements, the body's oxygen consumption

More information

JEPonline Journal of Exercise Physiologyonline

JEPonline Journal of Exercise Physiologyonline Walking Technique and Estimated VO 2 max Values 21 JEPonline Journal of Exercise Physiologyonline Official Journal of The American Society of Exercise Physiologists (ASEP) ISSN 1097-9751 An International

More information

Understanding the Issues Affecting the Accuracy of Measuring Oxygen Consumption Utilizing Metabolic Carts

Understanding the Issues Affecting the Accuracy of Measuring Oxygen Consumption Utilizing Metabolic Carts Understanding the Issues Affecting the Accuracy of Measuring Oxygen Consumption Utilizing Metabolic Carts Thank you for joining us. The webinar will begin shortly. Tweet #LifeScienceWebinar #ISCxAEITech

More information

Gas exchange measurement in module M- COVX

Gas exchange measurement in module M- COVX Gas exchange measurement in module M- COVX This document explains how the gas exchange measurement in the gas module M-COVX works. 1. Basic equations To obtain the oxygen consumption VO2 of a patient,

More information

IMPC phenotyping SOPs in JMC

IMPC phenotyping SOPs in JMC IMPC phenotyping SOPs in JMC Indirect Calorimetry IMPC_CAL_002 Purpose Indirect calorimetry provides detailed information on the energy metabolism of mutant mice. Energy expenditure is evaluated through

More information

iworx Sample Lab Experiment HE-5: Resting Metabolic Rate (RMR)

iworx Sample Lab Experiment HE-5: Resting Metabolic Rate (RMR) Experiment HE-5: Resting Metabolic Rate (RMR) Before Starting 1. Read the procedures for the experiment completely before beginning the experiment. Have a good understanding of how to perform the experiment

More information

Application Note 268 Metabolic Analysis Systems and Propane-based Verification

Application Note 268 Metabolic Analysis Systems and Propane-based Verification Application Note 268 Metabolic Analysis Systems and Propane-based Verification 8.10.11 This application note examines a conventional equipment setup for metabolic measurements. The setup is exposed to

More information

Specificity of training is perhaps the most significant

Specificity of training is perhaps the most significant Energy system contribution during 200- to 1500-m running in highly trained athletes MATT R. SPENCER and PAUL B. GASTIN Human Performance Laboratory, Department of Human Movement and Sport Sciences, University

More information

iworx Sample Lab Experiment HE-5: Resting Metabolic Rate (RMR)

iworx Sample Lab Experiment HE-5: Resting Metabolic Rate (RMR) Experiment HE-5: Resting Metabolic Rate (RMR) Before Starting 1. Read the procedures for the experiment completely before beginning the experiment. Have a good understanding of how to perform the experiment

More information

Application Note 183. Page 1 of Aero Camino, Goleta, CA Tel (805) Fax (805)

Application Note 183. Page 1 of Aero Camino, Goleta, CA Tel (805) Fax (805) APPLICATION NOTE 42 Aero Camino, Goleta, CA 93117 Tel (805) 685-0066 Fax (805) 685-0067 info@biopac.com www.biopac.com 7.10.18 Application Note 183 VO 2 and RER Measurement For simplified VO 2 measurement

More information

This test shall be carried out on all vehicles equipped with open type traction batteries.

This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4. Determination of hydrogen emissions page 1 RESS-6-15 5.4.1. This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4.2. The test shall be conducted following

More information

Columbus Instruments

Columbus Instruments 0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011

More information

Submaximal Treadmill Exercise Test to Predict VO 2 max in Fit Adults

Submaximal Treadmill Exercise Test to Predict VO 2 max in Fit Adults Measurement in Physical Education and Exercise Science ISSN: 1091-367X (Print) 1532-7841 (Online) Journal homepage: http://www.tandfonline.com/loi/hmpe20 Submaximal Treadmill Exercise Test to Predict VO

More information

Article published in: ACSM s Medicine & Science in Sports & Exercise Vol. 27, No. 4, April 1995

Article published in: ACSM s Medicine & Science in Sports & Exercise Vol. 27, No. 4, April 1995 Article published in: ACSM s Medicine & Science in Sports & Exercise Vol. 7, No. 4, April 1995 ENERGY EXPENDITURE DURING SUBMAXIMAL WALKING WITH EXERSTRIDERS Carol D. Rodgers, Jaci L. VanHeest, and Candice

More information

Evaluation of Three New Fit Test Protocols for Use with the TSI PortaCount

Evaluation of Three New Fit Test Protocols for Use with the TSI PortaCount Evaluation of Three New Fit Test Protocols for Use with the TSI PortaCount by Roy T. McKay, Ph.D. Jeff Bradley, RRT Occupational Pulmonary Services Department of Environmental Health University of Cincinnati

More information

Experiment HE-9: Resting, Active, and Exercising Metabolic Rates

Experiment HE-9: Resting, Active, and Exercising Metabolic Rates Experiment HE-9: Resting, Active, and Exercising Metabolic Rates Before Starting 1. Read the procedures for the experiment completely before beginning the experiment. Have a good understanding of how to

More information

Experiment HE-9: Resting, Active, and Exercising Metabolic Rates

Experiment HE-9: Resting, Active, and Exercising Metabolic Rates Experiment HE-9: Resting, Active, and Exercising Metabolic Rates Before Starting 1. Read the procedures for the experiment completely before beginning the experiment. Have a good understanding of how to

More information

12. Laboratory testing

12. Laboratory testing 12. Laboratory testing The performance lab of a Sports Medical Center offers various tests. In this paper we elaborate the testing of the aerobic system of a runner on a treadmill. To test the aerobic

More information

Assessment of an International Breaststroke Swimmer Using a Race Readiness Test

Assessment of an International Breaststroke Swimmer Using a Race Readiness Test International Journal of Sports Physiology and Performance, 2009, 4, 139-143 2009 Human Kinetics, Inc. Assessment of an International Breaststroke Swimmer Using a Race Readiness Test Kevin G. Thompson

More information

Content Display. - Introduction. Laboratory Unit : Lab 2 - Measurement of Oxygen Consumption. KINE xxxx Exercise Physiology

Content Display. - Introduction. Laboratory Unit : Lab 2 - Measurement of Oxygen Consumption. KINE xxxx Exercise Physiology Content Display Laboratory Unit : Lab 2 - Measurement of Oxygen KINE xxxx Exercise Physiology 6 Laboratory Unit 2 Lab 2 - Measurement of Oxygen 1 LAB2P1 - Introduction The goal of lab 2 is to facilitate

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results. Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions

More information

iworx Sample Lab Experiment HE-4: Respiratory Exchange Ratio (RER)

iworx Sample Lab Experiment HE-4: Respiratory Exchange Ratio (RER) Experiment HE-4: Respiratory Exchange Ratio (RER) Before Starting 1. Read the procedures for the experiment completely before beginning the experiment. Have a good understanding of how to perform the experiment

More information

The role of fitness testing in the evaluation of primary school running programmes

The role of fitness testing in the evaluation of primary school running programmes The role of fitness testing in the evaluation of primary school running programmes Making Strides in School Symposium National Centre for Sport and Exercise Medicine 7 th September 2016 Prof. Stuart Fairclough

More information

Using Hexoskin Wearable Technology to Obtain Body Metrics During Trail Hiking

Using Hexoskin Wearable Technology to Obtain Body Metrics During Trail Hiking Technical Note Using Hexoskin Wearable Technology to Obtain Body Metrics During Trail Hiking JEFF MONTES 1, TORI M. STONE 1, JACOB W. MANNING 2, DAMON MCCUNE 1, DEBRA K. TACAD 1, JOHN C. YOUNG 1, MARK

More information

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Background There are two main sources of energy available for animal metabolism: carbohydrates (CHO) and fats. These molecules are broken

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs. Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography

More information

MIL-STD-883G METHOD

MIL-STD-883G METHOD STEADY-STATE LIFE 1. PURPOSE. The steady-state life test is performed for the purpose of demonstrating the quality or reliability of devices subjected to the specified conditions over an extended time

More information

PICU Resident Self-Study Tutorial The Basic Physics of Oxygen Transport. I was told that there would be no math!

PICU Resident Self-Study Tutorial The Basic Physics of Oxygen Transport. I was told that there would be no math! Physiology of Oxygen Transport PICU Resident Self-Study Tutorial I was told that there would be no math! INTRODUCTION Christopher Carroll, MD Although cells rely on oxygen for aerobic metabolism and viability,

More information

POWER Quantifying Correction Curve Uncertainty Through Empirical Methods

POWER Quantifying Correction Curve Uncertainty Through Empirical Methods Proceedings of the ASME 2014 Power Conference POWER2014 July 28-31, 2014, Baltimore, Maryland, USA POWER2014-32187 Quantifying Correction Curve Uncertainty Through Empirical Methods ABSTRACT Christopher

More information

Anaerobic and aerobic contributions to 800 m and 8 km season bests

Anaerobic and aerobic contributions to 800 m and 8 km season bests Short Communication Anaerobic and aerobic contributions to 8 m and 8 km season bests Julia C. Blumkaitis, Christopher L. Sandefur, Christopher A. Fahs, Lindy M. Rossow Objective: Both anaerobic and aerobic

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) DOCUMENT NO: STP 2203 VERSION: 2008-07-01 Page 1 of 6 1.0 PURPOSE This test procedure is used by the Electrical Safety

More information

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS Frank M. Gavila F&J SPECIALTY PRODUCTS, INC. 404 Cypress Road Ocala, Florida, USA 3447 fandj@fjspecialty.com INTRODUCTION The calibration

More information

New Generation System M, leading the World in the Non-Invasive Measurement of Critical Real-Time Parameters.

New Generation System M, leading the World in the Non-Invasive Measurement of Critical Real-Time Parameters. New Generation System M, leading the World in the Non-Invasive Measurement of Critical Real-Time Parameters. System M Spectrum Medicals total commitment to continuous product improvement is demonstrated

More information

A Re-Examination of Running Energetics in Average and Elite Distance Runners

A Re-Examination of Running Energetics in Average and Elite Distance Runners University of Colorado, Boulder CU Scholar Integrative Physiology Graduate Theses & Dissertations Integrative Physiology Spring 1-1-2013 A Re-Examination of Running Energetics in Average and Elite Distance

More information

Instruction and Maintenance Manual

Instruction and Maintenance Manual Instruction and Maintenance Manual GRYF OXY ZM 02/100/2 E Contact GRYF HB, spol. s r.o. Cechova 314 Havlickuv Brod 580 01 tel.: +420 569 426 627 fax: +420 569 426 627 www.gryf.eu ver.: 11.2.2015 OEM module

More information

Recording established using the VO2 and RER graph template file (h19.gtl)

Recording established using the VO2 and RER graph template file (h19.gtl) Updated 10-14-13 BSL PRO Lesson H19: VO 2 and RER Measurement Recording established using the VO2 and RER graph template file (h19.gtl) Overview Oxygen Consumption (VO 2 ) and Respiratory Exchange Ratio

More information

Measuring Lung Capacity

Measuring Lung Capacity Name Class Date Chapter 37 Circulatory and Respiratory Systems Measuring Lung Capacity Introduction The amount of air that you move in and out of your lungs depends on how quickly you are breathing. The

More information

A CO 2 Waveform Simulator For Evaluation and Testing of Respiratory Gas Analyzers

A CO 2 Waveform Simulator For Evaluation and Testing of Respiratory Gas Analyzers 2011 ROCKY MOUNTAIN NASA SPACE GRANT CONSORTIUM 1 A CO 2 Waveform Simulator For Evaluation and Testing of Respiratory Gas Analyzers Christina Long, and Joseph Orr, Ph.D. Department of Bioengineering, University

More information

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Background There are two main sources of energy available for animal metabolism: carbohydrates (CHO) and fats. These molecules are broken

More information

EVALUATION OF A CANOPY SYSTEM AND A SIMPLE CALORIMETER FOR RESTING METABOLISM USING A RESPIRATORY SIMULATOR

EVALUATION OF A CANOPY SYSTEM AND A SIMPLE CALORIMETER FOR RESTING METABOLISM USING A RESPIRATORY SIMULATOR EVALUATION OF A CANOPY SYSTEM AND A SIMPLE CALORIMETER FOR RESTING METABOLISM USING A RESPIRATORY SIMULATOR N. Ichinoseki 1, 2, T. Yoshimura 2, M. Nambu 2, and T. Tamura 2 1 Japan Foundation for Aging

More information

SERIAL DETERMINATION OF CARDIAC OUTPUT DURING PROLONGED WORK BY A CO2-REBREATHING TECHNIQUE

SERIAL DETERMINATION OF CARDIAC OUTPUT DURING PROLONGED WORK BY A CO2-REBREATHING TECHNIQUE J. Human Ergol., 4: 35-41,1975 SERIAL DETERMINATION OF CARDIAC OUTPUT DURING PROLONGED WORK BY A CO2-REBREATHING TECHNIQUE Brian A. WILSON and R. T. HERMISTON* Department of Human Kinetics, University

More information

Outside Air Nonresidential HVAC Stakeholder Meeting #2 California Statewide Utility Codes and Standards Program

Outside Air Nonresidential HVAC Stakeholder Meeting #2 California Statewide Utility Codes and Standards Program 1 Outside Air Nonresidential HVAC Stakeholder Meeting #2 California Statewide Utility Codes and Standards Program Jim Meacham, CTG Energetics Agenda 2 In Situ Testing Results Reduced Ventilation after

More information

BREATH-BY-BREATH METHOD

BREATH-BY-BREATH METHOD BREATH-BY-BREATH METHOD COR-MAN-0000-005-IN / EN Issue A, Rev. 2 2013-07 INNOISION ApS Skovvænge DK-5620 Glamsbjerg Denmark Tel.: +45 65 95 91 00 Fax: +45 65 95 78 00 info@innovision.dk www.innovision.dk

More information

Gerald D. Anderson. Education Technical Specialist

Gerald D. Anderson. Education Technical Specialist Gerald D. Anderson Education Technical Specialist The factors which influence selection of equipment for a liquid level control loop interact significantly. Analyses of these factors and their interactions

More information

O 6 = 6 CO H 2 O H 2 = 16 CO 2. consumed during the oxidation of fatty acids, a ratio of 0.7.

O 6 = 6 CO H 2 O H 2 = 16 CO 2. consumed during the oxidation of fatty acids, a ratio of 0.7. Experiment HE-4: Respiratory Exchange Ratio (RER) Background The two sources of energy available for human metabolism are carbohydrates (CHO) and fats. These molecules are broken down, or catabolized,

More information

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300 Technical Note GA-300 Overview The GA-300 CO2 and O2 Gas Analyzer is easy to use, robust, and adaptable to human, animal, and plant applications. The GA-300 has two analog outputs to allow recording and

More information

CDI Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist

CDI Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist Original Article Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist David W. Fried, MS Ed, CCP; Joseph J. Leo, BS, CCP; Gabriel J. Mattioni, BS, CCP; Hasratt Mohamed, CCP; Raymond

More information

The estimation of energy expenditure (EE) is of interest

The estimation of energy expenditure (EE) is of interest Estimation of energy expenditure using CSA accelerometers at hip and wrist sites ANN M. SWARTZ, SCOTT J. STRATH, DAVID R. BASSETT, JR., WILLIAM L. O BRIEN, GEORGE A. KING, and BARBARA E. AINSWORTH Department

More information

INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION.

INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION. Andrzej Hantz RADWAG BALANCES AND SCALES INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION. The article presents the influence of comparison process automatization

More information

Journal of Exercise Physiologyonline (JEPonline)

Journal of Exercise Physiologyonline (JEPonline) Effect of Hiking Poles 20 Journal of Exercise Physiologyonline (JEPonline) Volume 11 Number 3 June 2008 Managing Editor Tommy Boone, Ph.D. Editor-in-Chief Jon Linderman, Ph.D. Revi ew Board Todd Astorino,

More information

Congress Science and Cycling 29 & 30 june 2016 Caen. Théo OUVRARD, Julien Pinot, Alain GROSLAMBERT, Fred GRAPPE

Congress Science and Cycling 29 & 30 june 2016 Caen. Théo OUVRARD, Julien Pinot, Alain GROSLAMBERT, Fred GRAPPE Congress Science and Cycling 29 & 30 june 2016 Caen Exposure Variation Analysis (EVA) method to monitor ability to optimally regulate exercise intensity of professional cyclists during time-trial competitions

More information

Exploring the relationship between Heart Rate (HR) and Ventilation Rate (R) in humans.

Exploring the relationship between Heart Rate (HR) and Ventilation Rate (R) in humans. Exploring the relationship between Heart Rate (HR) and Ventilation Rate (R) in humans. The Research Question In this investigation I will be considering the following general research question: Does increased

More information

Comparing Garmin Forerunner 405CX GGP and Nike + ipod to Accurately Measure Energy Expenditure, Distance, and Speed of Overground Running

Comparing Garmin Forerunner 405CX GGP and Nike + ipod to Accurately Measure Energy Expenditure, Distance, and Speed of Overground Running Cleveland State University EngagedScholarship@CSU ETD Archive 2010 Comparing Garmin Forerunner 405CX GGP and Nike + ipod to Accurately Measure Energy Expenditure, Distance, and Speed of Overground Running

More information

HHS Public Access Author manuscript Int J Cardiol. Author manuscript; available in PMC 2016 April 15.

HHS Public Access Author manuscript Int J Cardiol. Author manuscript; available in PMC 2016 April 15. FITBIT : AN ACCURATE AND RELIABLE DEVICE FOR WIRELESS PHYSICAL ACTIVITY TRACKING Keith M. Diaz 1, David J. Krupka 1, Melinda J Chang 1, James Peacock 1, Yao Ma 2, Jeff Goldsmith 2, Joseph E. Schwartz 1,

More information

CHAPTER 7 : SMOKE METERS AND THEIR INSTALLATIONS

CHAPTER 7 : SMOKE METERS AND THEIR INSTALLATIONS CHAPTER 7 : SMOKE METERS AND THEIR INSTALLATIONS 1 Scope : This Chapter covers the requirements of smoke meters and their installation on engines for full load and free acceleration tests, mentioned in

More information

that, as a means of progression, walking is suitable for lower speeds

that, as a means of progression, walking is suitable for lower speeds 2 6I2 744.22 ENERGY EXPENDITURE IN WALKING AND RUNNING. BY M. OGASAWARA. (From the Department of Industrial Physiology, London School of Hygiene and Tropical Medicine.) (Received February 28, 1934.) IT

More information

AC : ACCURATE CRYOCHAMBER FOR A SMALL LABORATORY WITH SMALL BUDGET

AC : ACCURATE CRYOCHAMBER FOR A SMALL LABORATORY WITH SMALL BUDGET AC 2007-1108: ACCURATE CRYOCHAMBER FOR A SMALL LABORATORY WITH SMALL BUDGET Matthew Braley, University of Idaho Paul Anderson, University of Idaho Tracey Windley, University of Idaho Kevin Buck, University

More information

Changes in a Top-Level Soccer Referee s Training, Match Activities, and Physiology Over an 8-Year Period: A Case Study

Changes in a Top-Level Soccer Referee s Training, Match Activities, and Physiology Over an 8-Year Period: A Case Study Case Study International Journal of Sports Physiology and Performance, 2011, 6, 281-286 2011 Human Kinetics, Inc. Changes in a Top-Level Soccer Referee s Training, Match Activities, and Physiology Over

More information

Delay time adjustments to minimize errors in breath-by-breath measurement of V O2 during exercise

Delay time adjustments to minimize errors in breath-by-breath measurement of V O2 during exercise Delay time adjustments to minimize errors in breath-by-breath measurement of V O2 during exercise DAVID N. PROCTOR AND KENNETH C. BECK Department of Anesthesia and Division of Pulmonary and Critical Care

More information

Journal of Exercise Physiologyonline

Journal of Exercise Physiologyonline 39 Journal of Exercise Physiologyonline June 2017 Volume 20 Number 3 Editor-in-Chief Official Research Journal of Tommy the American Boone, PhD, Society MBA of Review Exercise Board Physiologists Todd

More information

Review of gas frac9ons in atmospheric air 2/2/16. Metabolic Calculations of Indirect Calorimetry. Expired gas analysis.

Review of gas frac9ons in atmospheric air 2/2/16. Metabolic Calculations of Indirect Calorimetry. Expired gas analysis. Metabolic Calculations of Indirect Calorimetry Understanding atmospheric air and gas volumes Calcula9ng VO2 The Haldane Transforma9on Calcula9ng VCO2 Calcula9ng RER Calcula9ng caloric expenditure Expired

More information

Instruction and Maintenance Manual

Instruction and Maintenance Manual Instruction and Maintenance Manual GRYF OXY Z 02/100/2 E Contact GRYF HB, spol. s r.o. Cechova 314 Havlickuv Brod 580 01 tel.: +420 569 426 627 fax: +420 569 426 627 Czech Republic www.gryf.eu Technical

More information

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE TELEDYNE HASTINGS TECHNICAL PAPERS INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE Proceedings of FEDSM 98 1998 ASME Fluids Engineering

More information

Steeplechase Hurdle Economy, Mechanics, and Performance

Steeplechase Hurdle Economy, Mechanics, and Performance Brigham Young University BYU ScholarsArchive All Theses and Dissertations 2013-06-27 Steeplechase Hurdle Economy, Mechanics, and Performance Sarah Ingebretsen Brigham Young University - Provo Follow this

More information

CARDIOVIT AT-104 ergospirometry

CARDIOVIT AT-104 ergospirometry E D E R I T Z M A N U F A C T U R E D I N A N D S W I G N S E D L A N D CARDIOVIT AT-104 ergospirometry Cardiopulmonary exercise tests (CPET) with touch screen by SCHILLER it s the experience that counts.

More information

Validity and Reproducibility of the Garmin Vector Power Meter When Compared to the SRM Device

Validity and Reproducibility of the Garmin Vector Power Meter When Compared to the SRM Device Journal of Sports Science 5 (2017) 235-241 doi: 10.17265/2332-7839/2017.05.001 D DAVID PUBLISHING Validity and Reproducibility of the Garmin Vector Power Meter When Compared to the SRM Device Randolph

More information

Lung Volumes and Capacities

Lung Volumes and Capacities Lung Volumes and Capacities Normally the volume of air entering the lungs during a single inspiration is approximately equal to the volume leaving on the subsequent expiration and is called the tidal volume.

More information

Applied. Physiology and Occupational Physiology Springer-Verlag 1981

Applied. Physiology and Occupational Physiology Springer-Verlag 1981 Eur J Appl Physiol (1981) 47: 385-391 European Journal of Applied Physiology and Occupational Physiology Springer-Verlag 1981 V0 2 Peak During Free Swimming Using the Backward Extrapolation of the 0 2

More information

SEMEM3-78 Testing and calibrating instrumentation and control equipment and circuits

SEMEM3-78 Testing and calibrating instrumentation and control equipment and circuits Testing and calibrating instrumentation and control equipment and Overview This unit identifies the competences you need to carry out tests and calibration of instrumentation and control equipment and,

More information

RESPIRATORY REGULATION DURING EXERCISE

RESPIRATORY REGULATION DURING EXERCISE RESPIRATORY REGULATION DURING EXERCISE Respiration Respiration delivery of oxygen to and removal of carbon dioxide from the tissue External respiration ventilation and exchange of gases in the lung Internal

More information

COMPARISON OF RESULTS OF AEROBIC POWER VALUE DERIVED FROM DIFFERENT MAXIMUM OXYGEN CONSUMPTION TESTING METHODS

COMPARISON OF RESULTS OF AEROBIC POWER VALUE DERIVED FROM DIFFERENT MAXIMUM OXYGEN CONSUMPTION TESTING METHODS COMPARISON OF RESULTS OF AEROBIC POWER VALUE DERIVED FROM DIFFERENT MAXIMUM OXYGEN CONSUMPTION TESTING METHODS LI JIA A THESIS SUBMITTED IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE MASTER DEGREE

More information

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV 16-1 INTRODUCTION 16-1.1 Purpose. The purpose of this chapter is to familiarize divers with the techniques used to mix divers breathing gas. 16-1.2 Scope. This chapter

More information

Validity and Reliability of Predicting Maximum Oxygen Uptake via Field Tests in Children and Adolescents

Validity and Reliability of Predicting Maximum Oxygen Uptake via Field Tests in Children and Adolescents Pediatric Exercise Science, 1991.3, 250-255 Validity and Reliability of Predicting Maximum Oxygen Uptake via Field Tests in Children and Adolescents Michael J. Buono, Julia J. Roby, Frank G. Micale, James

More information

This article is downloaded from.

This article is downloaded from. This article is downloaded from http://researchoutput.csu.edu.au It is the paper published as: Author: R. Duffield, B. Dawson and C. Goodman Title: Energy system contribution to 100-m and 200-m track running

More information

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly

More information

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies IMPROVED CALIBRATION AND OPERATION OF A STABLE-GAS QUANTIFICATION MANIFOLD AS PART OF A RADIOXENON COLLECTION SYSTEM ABSTRACT Richard M. Williams, James C. Hayes, and Randy R. Kirkham Pacific Northwest

More information

University of Canberra. This thesis is available in print format from the University of Canberra Library.

University of Canberra. This thesis is available in print format from the University of Canberra Library. University of Canberra This thesis is available in print format from the University of Canberra Library. If you are the author of this thesis and wish to have the whole thesis loaded here, please contact

More information

A COMPARATIVE STUDY ON RESPIRATORY PARAMETERS BETWEEN SHORT DISTANCE AND LONG DISTANCE SWIMMERS

A COMPARATIVE STUDY ON RESPIRATORY PARAMETERS BETWEEN SHORT DISTANCE AND LONG DISTANCE SWIMMERS A COMPARATIVE STUDY ON RESPIRATORY PARAMETERS BETWEEN SHORT DISTANCE AND LONG DISTANCE SWIMMERS Baghel Priya PhD scholar, LNIPE, Gwalior, M.P., India ABSTRACT The purpose of the study was to find out the

More information

The Relationship between Speech and Peak flow values at various levels of work

The Relationship between Speech and Peak flow values at various levels of work A STUDY OF HOW SPEECH AFFECTS PEAK INSPIRATORY AIR FLOW VALUES, AT VARIOUS LEVELS OF WORK AND HOW THE TEST RESULTS SPREAD BETWEEN INDIVIDUALS IN A TEST GROUP. Submitted for publication at the 1997 ISRP

More information

ABSTRACT THE INFLUENCE OF BODY COMPOSITION ON CADENCE EFFICIENCY IN COMPETITIVE CYCLISTS. by Tate Bross Devlin

ABSTRACT THE INFLUENCE OF BODY COMPOSITION ON CADENCE EFFICIENCY IN COMPETITIVE CYCLISTS. by Tate Bross Devlin ABSTRACT THE INFLUENCE OF BODY COMPOSITION ON CADENCE EFFICIENCY IN COMPETITIVE CYCLISTS by Tate Bross Devlin The primary aim of this investigation was to determine the relationship between body composition

More information

Competitive Performance of Elite Olympic-Distance Triathletes: Reliability and Smallest Worthwhile Enhancement

Competitive Performance of Elite Olympic-Distance Triathletes: Reliability and Smallest Worthwhile Enhancement SPORTSCIENCE sportsci.org Original Research / Performance Competitive Performance of Elite Olympic-Distance Triathletes: Reliability and Smallest Worthwhile Enhancement Carl D Paton, Will G Hopkins Sportscience

More information

The validity of the Moxus Modular metabolic system during incremental exercise tests: impacts on detection of small changes in oxygen consumption

The validity of the Moxus Modular metabolic system during incremental exercise tests: impacts on detection of small changes in oxygen consumption Eur J Appl Physiol (2014) 114:941 950 DOI 10.1007/s00421-014-2825-x Original Article The validity of the Moxus Modular metabolic system during incremental exercise tests: impacts on detection of small

More information

Mass Flowmeters. Mass Flowmeters for Gases

Mass Flowmeters. Mass Flowmeters for Gases Mass Flowmeters Mass Flowmeters for Gases Fast, Accurate, Low Pressure Drop for critical measurement applications! Flow (Std l/min) Response to a step change in flowrate 300 200 100 4 Milliseconds 63%

More information

A Nomogram Of Performances In Endurance Running Based On Logarithmic Model Of Péronnet-Thibault

A Nomogram Of Performances In Endurance Running Based On Logarithmic Model Of Péronnet-Thibault American Journal of Engineering Research (AJER) e-issn: 2320-0847 p-issn : 2320-0936 Volume-6, Issue-9, pp-78-85 www.ajer.org Research Paper Open Access A Nomogram Of Performances In Endurance Running

More information

CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS

CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS Brit. J. Anaesth. (1956), 28, 196 CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS BY RUSSELL M. DAVIES, I. R. VERNER Queen Victoria Hospital, East Grinstead AND A. BRACKEN Research and Development Centre,

More information

23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN APRIL 2007

23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN APRIL 2007 23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN 16-20 APRIL 2007 AN INVESTIGATION INTO THE INTERRELATION BETWEEN THE INTERNAL AND EXTERNAL BALLISTICS OF FIRING A TP-T TANK AMMUNITION M. H.

More information

GE Healthcare. CARESCAPE R860 Nutritional Assessment Tools Appliguide

GE Healthcare. CARESCAPE R860 Nutritional Assessment Tools Appliguide GE Healthcare CARESCAPE R860 Nutritional Assessment Tools Appliguide 1 Contents Introduction... 3 Scope of this appliguide... 3 Structure of this appliguide... 3 Indirect Calorimetry... 4 Limitations of

More information

UNIVERSITY OF CINCINNATI

UNIVERSITY OF CINCINNATI UNIVERSITY OF CINCINNATI DATE: November 7, 2003 I, Christina Lynn Stewart, hereby submit this as part of the requirements for the degree of: Master of Science in: Nutrition It is entitled: Comparison of

More information

LEVEL OF VO2 MAX CAPACITY VOLLEYBALL PLAYERS

LEVEL OF VO2 MAX CAPACITY VOLLEYBALL PLAYERS LEVEL OF VO2 MAX CAPACITY VOLLEYBALL PLAYERS Wojciech BAJOREK ABCDEF, Joanna PIECH BCD Faculty of Physical Education, University of Rzeszow, Rzeszow, Poland Keywords: volleyball oxygen uptake beep test

More information

Gas Measurement Fundamentals Certification. Curriculum

Gas Measurement Fundamentals Certification. Curriculum Gas Certification Institute, LLC P.O. Box 131525 Houston, Texas 77219-1525 281-598-7200 Phone 281-598-7199 Fax contact@gascertification.com www.gascertification.com Curriculum Copyright 2007-2012 Gas Certification

More information

APPENDIX C. 40 CFR PART 60 App. B SPEC. 1-4, 6

APPENDIX C. 40 CFR PART 60 App. B SPEC. 1-4, 6 APPENDIX C 40 CFR PART 60 App. B SPEC. 1-4, 6 Part 60, Appendix B - Performance Specification 2 Specifications and Test Procedures for SO2 and NOx Continuous Emission Monitoring Systems in Stationary Sources

More information

A Description of Variability of Pacing in Marathon Distance Running

A Description of Variability of Pacing in Marathon Distance Running Original Research A Description of Variability of Pacing in Marathon Distance Running THOMAS A. HANEY JR. and JOHN A. MERCER Department of Kinesiology and Nutrition Sciences, University of Nevada, Las

More information

The Discussion of this exercise covers the following points:

The Discussion of this exercise covers the following points: Exercise 3-2 Orifice Plates EXERCISE OBJECTIVE In this exercise, you will study how differential pressure flowmeters operate. You will describe the relationship between the flow rate and the pressure drop

More information

THE PHYSICAL AND PHYSIOLOGICAL VALUE OF 5-A-SIDE SOCCER TRAINING TO

THE PHYSICAL AND PHYSIOLOGICAL VALUE OF 5-A-SIDE SOCCER TRAINING TO 1 THE PHYSICAL AND PHYSIOLOGICAL VALUE OF 5-A-SIDE SOCCER TRAINING TO 11-A-SIDE MATCH PLAY Jason D. Allen 1, Ronald Butterly 2, Michael A. Welsch 1, and Robert Wood 1 1 Department of Kinesiology, Louisiana

More information

Syllabus Number: 3.B.35 / BOD n. 188 ( ) CMAS CCR Diver diluent air/nitrox Training Program Minimum Training Program Content

Syllabus Number: 3.B.35 / BOD n. 188 ( ) CMAS CCR Diver diluent air/nitrox Training Program Minimum Training Program Content Syllabus Number: 3.B.35 / BOD n. 188 ( 09-06-2014 ) CMAS CCR Diver diluent air/nitrox Training Program Minimum Training Program Content 1. Required theoretical knowledge 1.1 Subject Area 1: Introduction

More information

Unpacking Interval. Lisa A. Workman

Unpacking Interval. Lisa A. Workman Unpacking Interval Training Lisa A. Workman Sunday, September 24, 2017 Overview History Exercise Physiology 101 The Three Energy Systems Training Continuum Types of Intervals Benefits History Joe Binks

More information

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE. Gases&Technology FEATURE Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. B Y M A T T H E W M O N A G L E Abstract Bulk gases are often delivered to

More information

Micro Channel Recuperator for a Reverse Brayton Cycle Cryocooler

Micro Channel Recuperator for a Reverse Brayton Cycle Cryocooler Micro Channel Recuperator for a Reverse Brayton Cycle Cryocooler C. Becnel, J. Lagrone, and K. Kelly Mezzo Technologies Baton Rouge, LA USA 70806 ABSTRACT The Missile Defense Agency has supported a research

More information

OXYGEN POWER By Jack Daniels, Jimmy Gilbert, 1979

OXYGEN POWER By Jack Daniels, Jimmy Gilbert, 1979 1 de 6 OXYGEN POWER By Jack Daniels, Jimmy Gilbert, 1979 Appendix A Running is primarily a conditioning sport. It is not a skill sport as is basketball or handball or, to a certain extent, swimming where

More information