Electron Microscopy Centre

Size: px
Start display at page:

Download "Electron Microscopy Centre"

Transcription

1 Electron Microscopy Centre Title: STANDARD PROCEDURES TO USE THE CRITICAL POINT DRYER Equipment: BIO-RAD E3000 Critical Point Dryer Revision: 1.0 Effective Date: 01/10/2006 Author: X. Yang 1

2 Warning! Before you attempt to operate this equipment for the first time, please make sure you are aware of the precautions that you must take to ensure your own safety. Warning! This apparatus is a PRESSURE VESSEL and operates at pressures in the range of 800 to 2000 psi (54 to 136 bar) WHICH IS POTENTIALLY DANGEROUS IF HANDLED CARELESSLY. Warning! Operation of the instrument involves Carbon Dioxide (CO 2 ), which is deemed to be lethal if sufficient amount of gas has been inhaled. Please refer to MSDS information of CO 2 previous to the operation. 2

3 Standard Procedures to Use the Critical Point Dryer Understand the CPD Valves (Figure 1): 1. Inlet Valve 2. Vent Valve 3. Drain Valve Checkouts before a Drying run: 1. Check the cold/hot water line is connected to the apparatus with no visible leaks. 2. Check the CO 2 high pressure cylinder is connected through a gas line to the inlet valve of the CPD. 3. Make sure the door of the lab is wide opened and the exhaust fans are turned on. 4. Make sure the tissues required to be dried have been washed, fixed, and dehydrated in either ethanol or acetone to 100% according to proper drying protocol previous to CPD. 5. Make sure the CO 2 inlet valve on the CPD apparatus is closed. Procedures for critical point drying of tissues: 1. Ensure the apparatus is correctly installed and run cold water to cool the chamber to 20 C or lower. 2. Place tissue in metal baskets, or plastic containers, taking care to prevent the tissue samples from drying. The tissue should be in the solvent all the time. 3. Fill the transfer boat with enough of the dehydrating agent to almost cover the samples and then place the baskets/containers in the boat. Hints: in order to prevent the tissue from air drying during the transferring, you could also fill the liquid transfer boat with dehydrating agent, put the metal baskets of plastic container in the boat then transfer the tissue to the baskets. The dehydrating agent (100% alcohol or acetone) are highly evaporative so that you may have to keep filling the boat with dehydrating agent during the transferring if you have more than one tissue required to be dried in a run. 4. Open the back door of the CPD apparatus, load the transfer boat into the chamber and then screw the door back on. Make sure the door is tight. 3

4 5. Turn on the CO 2 cylinder valve and open the CO 2 inlet valve on the CPD apparatus. At the same time, slightly open the vent valve to avoid back-pressure and allow a quick fill. Note: Filling with liquid CO 2 gas should be rapid. The picture on the right demonstrates how to proper handle the three valves. Gloves should be worn all the time. Please refer to a separate manual for detailed high pressure gas cylinder operation information. Note: The valves seal by contact between a ground steel cone and a brass knife edge; this metal to metal contact should not be over tightened as this will impair the efficiency of the valve. 6. Leave the inlet valve fully open with the vent valve slightly open to maintain liquid level, and then open the drain valve at the bottom to remove most of the dehydrating fluid. Allow the chamber to flush for about 30 sec, and then close both the vent and drain valves. Leave the inlet valve open to maintain liquid CO 2 level. 7. Flush the apparatus as outlined above every half-hour for 30 sec or so, to allow specimens to infiltrate with CO 2 and remove dehydrating fluid. Large specimens will require about hr. total time, while small specimens will be done in about 1.5 hr. Remember to leave the room door wide open to dissipate the CO 2 gas that is being vented. 8. After flushing the bulk of the substitution liquid, close the inlet valve and lower the liquid level to just below the top of the boat by venting off excess gas. Note: It is very important to have to liquid level just above the top of the boat because it provides the best view of the liquid level and the disappearance of the liquid/gas boundary line would be a good indicator of CO 2 liquid to gas transformation. Note: For the safety issue, do not observe the liquid level by direct watching the viewing window, watch the reflection of the chamber in the mirror provided. 9. Run hot water through the water jacket and monitor the temperature and pressure, with all valves closed. 4

5 10. When the temperature reaches about 36 to 38 C, and the pressure rises to1200 lbs/sq. in. or slightly above, the liquid/gas boundary line will disappear and the specimens are above the critical point. 11. Once the critical point has been reached, vent the gas off slowly (over a 5 min. period) to avoid condensation. The temperature should remain around 40 C during this time. Note: Condensation can be caused by rapid venting or by the presence of acetone or amyl acetate. 12. Shut off the water, open the door, remove the boat and sniff it to check for residual dehydrating solvent. 13. After taking out the specimens, replace the boat and containers in the drawer, seal the CPD apparatus again, and turn off the CO 2 cylinder. 14. Record the usage of Critical Point Dryer on the log book. Reference: 1. Instruction Manual for E3000 Series Critical Point Drying Apparatus. BIO-RAD. 5

SPI-DRY TM Critical Point Dryer Operation Manual. SPI Supplies Part #13200JE-AB

SPI-DRY TM Critical Point Dryer Operation Manual. SPI Supplies Part #13200JE-AB SPI-DRY TM Critical Point Dryer Operation Manual SPI Supplies Part #3200JE-AB Rev. 2. October 204 Structure Probe, Inc. / SPI Supplies Street address for UPS, FedEx, DHL, other couriers, trucks 206 Garfield

More information

Electron Microscopy Sciences

Electron Microscopy Sciences Electron Microscopy Sciences INSTRUCTIONAL MANUAL CAT 62057-10 Frigimat Cub Dry Ice Maker sgkcck@aolcom PO Box 550, 1560 Industry Road, Hatfield, PA 19440 Safety Symbols Used in this Instruction Manual

More information

School of Chemistry SOP For Operation Of Glove Boxes

School of Chemistry SOP For Operation Of Glove Boxes School SOP for Operation of Glove Boxes: The following SOP provides guidelines on how to adequately and safely operating a standard laboratory glove box and its associated devices. General Information:

More information

Working correctly with gas

Working correctly with gas Working correctly with gas When working with a reactive gas in a cylinder, there is an inherent risk. Special care has to be taken to allow safe handling. This example is with HCl. 1) Fixing the gas bottle.

More information

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax: 2 Table Of Contents 1. Introduction 3 2. About this Manual 3 3. Contacting YZ Systems 3 4. Vessel Components 4 5. Specifications 5 6. Application 6 7. Theory of Operation 7 8. DuraSite Installation & Use

More information

HIGH RATE BASKET STRAINER

HIGH RATE BASKET STRAINER Installation and Operation Instruction Manual INSMAN-105 High Rate Basket Strainer Phone: 951.656.6716 Toll-Free: 800.854.4788 www.yardneyfilters.com Yardney Water Management Systems, Inc. 6666 Box Springs

More information

OXY FUEL SYSTEM SET UP & SHUT DOWN PROCEDURE CHECKLIST ACETYLENE FUEL GAS

OXY FUEL SYSTEM SET UP & SHUT DOWN PROCEDURE CHECKLIST ACETYLENE FUEL GAS OXY FUEL SYSTEM SET UP & SHUT DOWN PROCEDURE CHECKLIST ACETYLENE FUEL GAS Items needed for demonstrations: Fire extinguisher PPE (Personal Protective Equipment): o Lab coats/welding jackets o Leather gloves

More information

School of Plant Biology - Liquid Nitrogen Induction

School of Plant Biology - Liquid Nitrogen Induction School of Plant Biology - Liquid Nitrogen Induction Liquid Nitrogen (LN 2 ) Usage Guidelines If you wish to use LN 2, there are a few rules that have been put in place to ensure your safety and to ensure

More information

Cryogenics is the science of ultra low temperatures. Low temperatures are achieved by the liquefaction of gases.

Cryogenics is the science of ultra low temperatures. Low temperatures are achieved by the liquefaction of gases. CRYOGENICS SAFETY Introduction Cryogenics is the science of ultra low temperatures. Low temperatures are achieved by the liquefaction of gases. The gases which are most widely used in industry and research

More information

The Discussion of this exercise covers the following points:

The Discussion of this exercise covers the following points: Exercise 5-3 Wet Reference Leg EXERCISE OBJECTIVE Learn to measure the level in a vessel using a wet reference leg. DISCUSSION OUTLINE The Discussion of this exercise covers the following points: Measuring

More information

Liquid Nitrogen for Duct Cleaning

Liquid Nitrogen for Duct Cleaning Liquid Nitrogen for Duct Cleaning Copyright 2017 NADCA, All Rights Reserved No part of this publication may be reproduced or distributed by any means, electronic or mechanical, including photocopy, recording,

More information

INSTALLATION & MAINTENANCE INSTRUCTION

INSTALLATION & MAINTENANCE INSTRUCTION ARCHON Industries, Inc Liquid Level Gauges Models: BT-LLG ND-LLG INSTALLATION & MAINTENANCE INSTRUCTION Instruction No.: 1014.2 Revision Issued: 3/01/03 Approved: Engineering Manager Warning ONLY QUALIFIED

More information

GREIG FILTERS, INC. OPERATION MANUAL

GREIG FILTERS, INC. OPERATION MANUAL GREIG FILTERS, INC. OPERATION MANUAL MODEL DCFH 3P 15/3 100 S4 DOE/222 DUAL CARTRIDGE FILTER HOUSING TABLE OF CONTENTS I. INTRODUCTION II. CARTRIDGE FILTER HOUSING UNIT A. OPERATION B. INSTALLATION C.

More information

Eaton Filtration, LLC

Eaton Filtration, LLC Eaton Filtration, LLC 900 Fairmount Avenue, Elizabeth, NJ 07207 Phone: 908-787-1000 Fax: 908-351-7893 E-Mail: filtration@eaton.com Web: www.filtration.eaton.com Installation, Operation & Service Manual

More information

Eaton Filtration, LLC

Eaton Filtration, LLC Eaton Filtration, LLC 900 Fairmount Avenue, Elizabeth, NJ 07207 Phone: 908-787-1000 Fax: 908-351-7893 E-Mail: filtration@eaton.com Web: www.filtration.eaton.com Installation, Operation & Service Manual

More information

Thermocirc Installation and Maintenance Instructions

Thermocirc Installation and Maintenance Instructions 1570072/2 IM-P157-36 ST Issue 2 Thermocirc Installation and Maintenance Instructions 1 General safety information 2 General product information 3 Installation 4 Commissioning 5 Operation 6 Maintenance

More information

S & M Water Consultants

S & M Water Consultants S & M Water Consultants Test Procedures In This Manual TEST Reduced Pressure Assembly One Hose Method PAGE 1 Double Check Assembly One Hose Method 5 Pressure Vacuum Breaker One Hose Method 7 Spill Proof

More information

K32230 DEW-POINT APPARATUS

K32230 DEW-POINT APPARATUS Installation, Operating and Maintenance Instructions K32230 DEW-POINT APPARATUS APPLICATIONS This manual has been prepared to assist in the preparation, use and care of the equipment. It is 1595 Sycamore

More information

Safe Work Instructions For Oxygen Acetylene Equipment

Safe Work Instructions For Oxygen Acetylene Equipment Safe Work Instructions For Oxygen Acetylene Equipment Policy Sponsor: Assistant Vice President Facilities Management Responsible Unit: Trade Services Approval Date: September 2010 Revisions: January 2018

More information

Instruction Manual VCG-6

Instruction Manual VCG-6 Specifications Amps / Volts requirements 400 ma @ 24 VDC Propane pressure (VCG-6LP) 11 WC / 2.8 kpa Natural Gas pressure (VCG-6NG) 4.5 WC / 1.15 kpa Cu ft per hour / CO2 3-6 SCFH BTU Rating (Variable)

More information

Operating Instructions

Operating Instructions Operating Instructions Before operating the thin film evaporator, please be aware of all safety concerns associated with this experiment: Burn hazard from the column and steam lines, Chemical hazards associated

More information

Installation, Operation, and Maintenance Manual

Installation, Operation, and Maintenance Manual Installation, Operation, and Maintenance Manual Welker Dehydration Assembly Model DA-1 Drawing No.: AD079CO, AD285CR Manual No.: IOM-018 The information in this manual has been carefully checked for accuracy

More information

Related Products: Auto Drain Valve. Model/Specifications. Model

Related Products: Auto Drain Valve. Model/Specifications. Model Related Products: Auto Drain Valve /600 Drainage is automatically discharged in a reliable manner, without requiring human operators. Highly resistant to dust and corrosion, operates reliably, and a bowl

More information

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane Gas requirements Gases for packed columns The carrier gas you use depends upon the type of detector and the performance requirements. Table 520-1 lists gas recommendations for packed column use. In general,

More information

ARC Unofficial Manual to the Solvent Dispensing System. Designed by J. C. Meyer

ARC Unofficial Manual to the Solvent Dispensing System. Designed by J. C. Meyer RC Unofficial Manual to the Solvent Dispensing System Designed by J. C. Meyer 1 Table of Contents: I. Introduction...3 II. rgon Supply...3 Tanks Regulators Over Pressurization III. Solvent Kegs...4 Solvent

More information

1. IDENTIFICATION OF THE SUBSTANCE / PREPARATION AND COMPANY PICRIC ACID (SATURATED AQUEOUS)

1. IDENTIFICATION OF THE SUBSTANCE / PREPARATION AND COMPANY PICRIC ACID (SATURATED AQUEOUS) Safety Data Sheet According to regulation (EC) # 1907/2006 1. IDENTIFICATION OF THE SUBSTANCE / PREPARATION AND COMPANY PRODUCT NAME PRODUCT CODE PICRIC ACID (SATURATED AQUEOUS) SUPPLIER Clin-Tech Limited

More information

Liquid Gas Cylinder (LGC)

Liquid Gas Cylinder (LGC) Liquid Gas Cylinder (LGC) www.soxal.com SOXAL Liquid Gas Cylinder (LGC) A Liquid Gas Cylinder (LGC) is a portable super-insulated vacuum vessel holding liquid nitrogen, oxygen, argon and carbon dioxide.

More information

INFLATOR TDUX-IT-16 AND GAS CYLINDER E

INFLATOR TDUX-IT-16 AND GAS CYLINDER E INFLATOR TDUX-IT-16 AND GAS CYLINDER E7512-0160 Operating manual Inflator TDUX-IT-16 together with the special gas cylinder E7512-0160 is specifically designed for inflating TDUX duct seals. The inflator

More information

TITAN FLOW CONTROL, INC.

TITAN FLOW CONTROL, INC. PREFACE: This manual contains information concerning the installation, operation, and maintenance of Titan Flow Control (Titan FCI) Simplex Basket Strainers. To ensure efficient and safe operation of Titan

More information

Hi-Force Limited Prospect Way Daventry Northants NN11 8PL United Kingdom Tel: +44(0) : Fax: +44(0) : Website:

Hi-Force Limited Prospect Way Daventry Northants NN11 8PL United Kingdom Tel: +44(0) : Fax: +44(0) : Website: 1.0 Inspection of the product upon receipt: On receipt of the product, visually inspect the item for any evidence of shipping damage. Please note shipping damage is not covered by warranty. If shipping

More information

Assembly Drawing: W-311B-A01, or as applicable Parts List: W-311B-A01-1, or as applicable Special Tools: , , &

Assembly Drawing: W-311B-A01, or as applicable Parts List: W-311B-A01-1, or as applicable Special Tools: , , & REDQ Regulators Model 411B Barstock Design Powreactor Dome Regulator OPERATION AND MAINTENANCE Contents Scope..............................1 Installation..........................1 General Description....................1

More information

CASH VALVE TYPE KP BACK PRESSURE VALVES

CASH VALVE TYPE KP BACK PRESSURE VALVES A high capacity pilot operated back pressure valve that offers accurate control and dependable protection against overpressure conditions FEATURES Automatically maintains maximum pressure in a vessel or

More information

PRESSURISED PAINT CONTAINER

PRESSURISED PAINT CONTAINER PRESSURISED PAINT CONTAINER MODEL NO: CPP2B PART NO: 3082115 OPERATION & MAINTENANCE INSTRUCTIONS GC0913 INTRODUCTION Thank you for purchasing this CLARKE Pressurised Paint Container. Before attempting

More information

Installation, Operation, and Maintenance Manual

Installation, Operation, and Maintenance Manual Installation, Operation, and Maintenance Manual Welker The information in this manual has been carefully checked for accuracy and is intended to be used as a guide for the installation, operation, and

More information

ROTATING DISK VALVES INSTALLATION AND MAINTENANCE 1. SCOPE 3 2. INFORMATION ON USAGE 3 3. VALVE TYPES 3 4. OPERATORS 5 5. VALVE CONSTRUCTION 6

ROTATING DISK VALVES INSTALLATION AND MAINTENANCE 1. SCOPE 3 2. INFORMATION ON USAGE 3 3. VALVE TYPES 3 4. OPERATORS 5 5. VALVE CONSTRUCTION 6 Sub Section INDEX Page Number 1. SCOPE 3 2. INFORMATION ON USAGE 3 3. VALVE TYPES 3 4. OPERATORS 5 5. VALVE CONSTRUCTION 6 6. INSTALLATION AND OPERATION 6 7. MAINTENANCE 8 8. REPAIR 9 9. ASSEMBLY 10 10.

More information

Propane Conversion Kit Instruction

Propane Conversion Kit Instruction Propane Conversion Kit Instruction Condensing gas boiler Required Input Rates GB142-24 84,800 btu/hr GB142-30 106,000 btu/hr GB142-45 160,900 btu/hr GB142-60 214,800 btu/hr This kit and instructions are

More information

Cash Valve TYPE KP PILOT OPERATED BACK PRESSURE VALVE. ISSUED - DECEMBER 2000 CAVMC-0518-US-0208 ISO 9001 Certified

Cash Valve TYPE KP PILOT OPERATED BACK PRESSURE VALVE. ISSUED - DECEMBER 2000 CAVMC-0518-US-0208 ISO 9001 Certified Cash Valve PILOT OPERATED BACK PRESSURE VALVE ISSUED - DECEMBER 2000 CAVMC-0518-US-0208 ISO 01 Certified KP HIGH CAPACITY PILOT OPERATED BACK PRESSURE VALVE DESCRIPTION The Cash Valve Type KP is a pilot

More information

CSA Sample Draw Aspirator Adapter Operator s Manual

CSA Sample Draw Aspirator Adapter Operator s Manual 30-0951-CSA Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0367 Revision: 0 Released: 4/30/15 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating

More information

Title/Description: Fuel Cell Fill Procedure

Title/Description: Fuel Cell Fill Procedure Fuel Cell Procedure Document No.: Author: Kurt Stiffel Title/Description: Approved by: Revision: 1-00 Effective Date: 1.0 Reference Documents All released OOI documents in Alfresco 3161-20000 Fuel Cell

More information

RARS5000 AIR BODY SAW OWNER S OPERATING MANUAL

RARS5000 AIR BODY SAW OWNER S OPERATING MANUAL RARS5000 AIR BODY SAW OWNER S OPERATING MANUAL DESCRIPTION 1. No mar 2. No mar tip 3. Housing grip 4. Trigger 5. Air inlet 6. Air inlet plug 7. Plastic board Important! It is essential that you read the

More information

Binks SV100 Conventional Suction Feed Spray Gun

Binks SV100 Conventional Suction Feed Spray Gun Binks SV100 Conventional Suction Feed Spray Gun The following instructions provide the necessary information for the proper maintenance of the Binks SV100 suction feed spray gun. Please read and understand

More information

Mass Spec will not pump down

Mass Spec will not pump down Mass Spec will not pump down Applies to 5973A/N MSD There are several situations that could cause the 5973A/N to have problems pumping down. The following scenarios should help in isolating problems related

More information

Propane Conversion Kit Instruction

Propane Conversion Kit Instruction Propane Conversion Kit Instruction Condensing gas boiler Required Input Rates GB42-24 84,800 btu/hr Tab. GB42-30 06,000 btu/hr GB42-45 49,000 btu/hr GB42-60 24,800 btu/hr This kit and instructions are

More information

Approval Block. Prepared by: Signature Date. Evan Parnell 08 NOV Reviewed by: Signature Date. Approved by: Signature Date

Approval Block. Prepared by: Signature Date. Evan Parnell 08 NOV Reviewed by: Signature Date. Approved by: Signature Date ATS-SOI-3490 Page: 1 of 6 Approval Block Prepared by: Signature Date Evan Parnell 08 NOV 2013 Reviewed by: Signature Date Brian Flynn 08 NOV 2013 Approved by: Signature Date Kristal Jewell 08 NOV 2013

More information

MATERIAL SAFETY DATA SHEET

MATERIAL SAFETY DATA SHEET 1/5 Label 2.2: Non flammable, non toxic gas. NFPA RATING 1 IDENTIFICATION OF THE SUBSTANCE / PREPARATION AND OF THE COMPANY / UNDERTAKING Trade Name : Chemical Formula : He Chemical Family : (Rare Gas)

More information

FILTER REGULATORS MODEL NO: CAT155 & CAT156 FITTING & MAINTENANCE INSTRUCTIONS PART NO: & ORIGINAL INSTRUCTIONS

FILTER REGULATORS MODEL NO: CAT155 & CAT156 FITTING & MAINTENANCE INSTRUCTIONS PART NO: & ORIGINAL INSTRUCTIONS FILTER REGULATORS MODEL NO: CAT155 & CAT156 PART NO: 3120169 & 3120170 FITTING & MAINTENANCE INSTRUCTIONS ORIGINAL INSTRUCTIONS GC0117 INTRODUCTION Thank you for purchasing this CLARKE Filter/Regulator.

More information

CA44, CA44S, CA46 and CA46S Air and Gas Traps

CA44, CA44S, CA46 and CA46S Air and Gas Traps 1484450/3 IM-P148-37 ST Issue 3 CA44, CA44S, CA46 and CA46S Air and Gas Traps Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4. Commissioning

More information

Chapter 13 Selection and Safe usage of cleaning solvents

Chapter 13 Selection and Safe usage of cleaning solvents Selection and Safe usage of cleaning solvents NO MORE ODS Application of R-141b Class II ozone-depleting substance (ODS) Characteristics of R-141b Non-flammable High solvency power Easy to store and handle

More information

POP Safety Valve. POP Safety Valve INTRODUCTION DEFINITIONS

POP Safety Valve. POP Safety Valve INTRODUCTION DEFINITIONS POP Safety Valve POP Safety Valve INTRODUCTION The effects of exceeding safe pressure levels in an unprotected pressure vessel or system, can have catastrophic effects on both plant and personnel. Safety

More information

MATERIAL SAFETY DATA SHEET

MATERIAL SAFETY DATA SHEET Sasol Solvents Trading as Sasol Chemicals North America LLC 900 Threadneedle Houston, Texas 77079 MATERIAL SAFETY DATA SHEET EMERGENCY NUMBERS: CHEMTREC (800) 424-9300 MSDS or Product Information: 281

More information

THERMAL EVAPORATION UNIT (for Al evaporation)

THERMAL EVAPORATION UNIT (for Al evaporation) THERMAL EVAPORATION UNIT (for Al evaporation) System Owner: NeerajPanwar 9619507210 panwar.iitr@gmail.com System Operator: 1. Bhimraj Sable 9930189878 bhimrajsable71@gmail.com Authorized User: 1. Ramesh

More information

Spot Weld Drill Spot Weld Removal System Instruction Manual

Spot Weld Drill Spot Weld Removal System Instruction Manual 85-1001 Spot Weld Drill Spot Weld Removal System Instruction Manual Version 1.1 TABLE OF CONTENTS COMPONENT CHECKLIST 2 GENERAL SAFETY - READ BEFORE USING 3 INSTALL AND STORAGE MAINTENANCE INSTRUCTION

More information

Procedure for Operating Columbia (Nevis) LHe Electron Bubble Chamber Cryostat. Version 5.1

Procedure for Operating Columbia (Nevis) LHe Electron Bubble Chamber Cryostat. Version 5.1 Project: Novel Electron Bubble Particle Detector Procedure for Operating Columbia (Nevis) LHe Electron Bubble Chamber Cryostat Version 5.1 (Using side filling line) Hand Processed Changes HPC No. Date

More information

Dri-Line Mk3 Monnier Compressed Air Drain Trap

Dri-Line Mk3 Monnier Compressed Air Drain Trap 5044050/2 IM-P504-24 CH Issue 2 Dri-Line Mk3 Monnier Compressed Air Drain Trap Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation and Operation

More information

Exhibit 4. Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N)

Exhibit 4. Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N) 1 APPLICABILITY Exhibit 4 Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N) Definitions common to all certification and test procedures

More information

Series Environmental Chambers

Series Environmental Chambers 3119-600 Series Environmental Chambers Challenges in Non-Ambient Testing Testing at non-ambient temperatures adds another layer of challenges to your testing laboratory. Ensuring you get accurate and stable

More information

1. Study the performance of a binary distillation column operated in batch mode.

1. Study the performance of a binary distillation column operated in batch mode. Goals for batch distillation using the East distillation column: 1. Study the performance of a binary distillation column operated in batch mode. 2. Determine the overall and local efficiency of the column

More information

Pneumatic Service Unit

Pneumatic Service Unit Pneumatic Service Unit Type 9 2406 0600 9 2406 0630 9 2406 0640 9 2406 0650 Illustration can differ from the original Installation and Operation Manual Compiled: 09.09.16 924060600_0630_0640_0650_en_Version_02

More information

Jerry Martin s Co 2 Snow Apparatus

Jerry Martin s Co 2 Snow Apparatus Introduction: Jerry Martin fashioned this Co 2 snow optics cleaning device similar to the ones that he built at McDonald Observatory. We made changes to fit our particular needs and I believe improved

More information

Auto EVAP Quick Start Guide

Auto EVAP Quick Start Guide Auto EVAP Quick Start Guide Pg. 13 Note: Never directly hit or drop the leak detector on the CO2 sensor tip, this will cause the sensor to become damaged! Do not get Leak Seeker Solution directly on the

More information

SSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR

SSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR S S F U SSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR MACHINERY DIVISION OWNER S MANUAL UNIT INSTRUCTIONS Please follow all SSFU Safety Instructions. Contact your Duro Dyne Tech Service if you have

More information

Gas Countertop Griddle

Gas Countertop Griddle Gas Countertop Griddle This manual contains important information regarding your purchased equipment. Please read the manual thoroughly prior to equipment set-up, operation and maintenance. Failure to

More information

Chapter 7. Coal Cargoes. Properties and Characteristics 7.1

Chapter 7. Coal Cargoes. Properties and Characteristics 7.1 Chapter 7 Coal Cargoes This chapter, and all of the chapters in Part 2, should be read in conjunction with the booklet Carrying Solid Bulk Cargoes Safely (Reference 16), published by Lloyd s Register/UK

More information

Gilson Aqua-Check Moisture Tester MA-26X

Gilson Aqua-Check Moisture Tester MA-26X OPERATING MANUAL Gilson Aqua-Check Moisture Tester MA-26X Rev: 01/2019 PHONE: 800-444-1508 P.O. Box 200, Lewis Center, Ohio 43035-0200 740-548-7298 E-mail: customerservice@gilsonco.com Product Web Page:

More information

Visit our Web site: http: //

Visit our Web site: http: // CONTENTS: Medium Duty Victor Style Welding And Cutting Outfit MANUFACTURED BY AN ISO 9001 REGISTERED FIRM Visit our Web site: http: // www.northerntool.com TABLE OF CONTENTS Page INTRODUCTION... 1 SECTION

More information

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation CHE 4115 Chemical Processes Laboratory 2 Experiment 1 Batch Distillation BACKGROUND Distillation is one of the most commonly used unit operations in chemical engineering. In general, a distillation operation

More information

ATD LB PRESSURE BLASTER INSTRUCTION MANUAL

ATD LB PRESSURE BLASTER INSTRUCTION MANUAL ATD-8402 90LB PRESSURE BLASTER INSTRUCTION MANUAL SAVE THESE INSTRUCTIONS SAFETY INSTRUCTIONS FOR SANDBLASTER 1. Before opening the tank release the air pressure on the sand tank. To do this, turn off

More information

OFI Testing Equipment - Aging Cell Instructions Page 1 of 4. Instructions OFITE AGING CELLS

OFI Testing Equipment - Aging Cell Instructions Page 1 of 4. Instructions OFITE AGING CELLS OFI Testing Equipment - Aging Cell Instructions Page 1 of 4 Instructions OFITE AGING CELLS The OFITE Aging Cell is a patented pressure vessel that enables samples to be subjected to temperatures higher

More information

FloWash. Industrial Filter Cart

FloWash. Industrial Filter Cart OPERATION AND MAINTENANCE MANUAL FloWash Industrial Filter Cart Read all instructions before installation or operation of equipment. Failure to comply with these instructions could result in bodily injury

More information

CAST IRON VALVES FLOAT VALVE

CAST IRON VALVES FLOAT VALVE CAST IRON VALVES FLOAT VALVE NETAFIM CAST IRON VALVES FLOAT VALVE INLINE 25-400MM ANGLE 50-100MM SPECIFICATIONS Manufactured to ISO 9001 Available Valve sizes INLINE Flanged 50mm to 600mm & Threaded 19mm

More information

OXYGEN ACETYLENE WELDING & CUTTING EQUIPMENT

OXYGEN ACETYLENE WELDING & CUTTING EQUIPMENT OXYGEN ACETYLENE WELDING & CUTTING EQUIPMENT Introduction: The oxy-acetylene torch can be used for various tasks. It can be used for cutting steel, welding steel, brazing and for heating metal. The oxy-acetylene

More information

University of Saskatchewan Department of Mechanical Engineering Standard Operating Procedure # Mat0002

University of Saskatchewan Department of Mechanical Engineering Standard Operating Procedure # Mat0002 Procedure Title: General Chemical Usage Guidelines Minimum Review Requirements: Annually Creation Date: November, 2011 Date of Next Review: November, 2012 Creator of Procedure: Authorized by: Title Page

More information

Intro: by firechim on October 3, 2013

Intro:   by firechim on October 3, 2013 by firechim on October 3, 2013 Intro: Step 1: Guide for installing chimney liners with a tee connection This guide is to give you step by step instructions to install a chimney liner that utilizes a tee

More information

OPERATING INSTRUCTIONS AND PARTS LIST HAND OPERATED LPG PISTON PUMP WITH SCALE 301-DH-SP

OPERATING INSTRUCTIONS AND PARTS LIST HAND OPERATED LPG PISTON PUMP WITH SCALE 301-DH-SP OPERATING INSTRUCTIONS AND PARTS LIST HAND OPERATED LPG PISTON PUMP WITH SCALE 301-DH-SP www.adceng.co.za P O BOX 5413, JOHANNESBURG 2000 Phone: 011 493-8209 Fax: 011 493-1459 Email: adceng@mweb.co.za

More information

Hot Compressed Air Lance Operation Manual. LAB Manufacturing 9483 Reading Road Cincinnati, OH (800)

Hot Compressed Air Lance Operation Manual.  LAB Manufacturing 9483 Reading Road Cincinnati, OH (800) Hot Compressed Air Lance Operation Manual www.heatlances.com LAB Manufacturing 9483 Reading Road Cincinnati, OH 45215 (800) 776-5830 PARTS CHECK LIST NOTE: Check all fittings & connections prior to operation.

More information

CA10S, CA14 and CA14S Air and Gas Traps

CA10S, CA14 and CA14S Air and Gas Traps 1448150/9 IM-P148-13 ST Issue 9 CA10S, CA14 and CA14S Air and Gas Traps Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4. Commissioning 5.

More information

User Information Sheet 013

User Information Sheet 013 User Information Sheet 013 Formerly LPGA Technical Memorandum NO.81 Issue 2 September 2007 ACME threaded couplings 1 INTRODUCTION ACME threaded couplings/end fittings are widely used for LPG transfer operations

More information

1. Cryogen Tanks. N2 Fill Tutorial for Vanderbilt Magnets. Contents: CSB NMR Facility

1. Cryogen Tanks. N2 Fill Tutorial for Vanderbilt Magnets. Contents: CSB NMR Facility CSB NMR Facility NMR-Home N2 Fill Tutorial for Vanderbilt Magnets This page gives an overview of the cryogen filling procedures for our instruments at Vanderbilt. User Info Access General Information:

More information

Instruction Manual. B&O WITH SLOPE BACK TENDER Live Steam

Instruction Manual. B&O WITH SLOPE BACK TENDER Live Steam Instruction Manual B&O 0-4-0 WITH SLOPE BACK TENDER Live Steam AML is proud to offer the famous B&O 0-4-0 of the Baltimore Waterfront. Four switchers were built in 1912 by Baldwin for the B&O. Initially

More information

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300 Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300 The Thermo Scientific icap 6300 Inductively Coupled Plasma Optical Emission

More information

Cryogenic Materials Operating Procedure and Safety Requirements

Cryogenic Materials Operating Procedure and Safety Requirements VERSION 2.0 JANUARY 2, 2018 Cryogenic Materials Operating Procedure and Safety Requirements UNC CHARLOTTE 9201 UNIVERSITY CITY BLVD., CHARLOTTE, NC 28223 1 I. Overview of Liquid Nitrogen and other Cryogenic

More information

Tips for Proper Pipette handling and Maintenance. D. Muruganand, Ph.D ILQA - Bangalore, Feb 2014

Tips for Proper Pipette handling and Maintenance. D. Muruganand, Ph.D ILQA - Bangalore, Feb 2014 Tips for Proper Pipette handling and Maintenance D. Muruganand, Ph.D ILQA - Bangalore, Feb 2014 Liquid handling tools Manual pipettes Electronic pipette and dispenser Automated systems Principles of Pipetting

More information

Glove Box Installation Manual

Glove Box Installation Manual Glove Box Installation Manual 1998 by M. Braun Company File: GB-UNI-INS.DOC! Edition 08-00 by M. Boutin! Subject to be changed without notice Glovebox Installation Your Glove box has been fully assembled,

More information

Sea Rebel Kayak Owner s Manual

Sea Rebel Kayak Owner s Manual Sea Rebel Kayak Owner s Manual www.ravesports.com Table of Contents Product instructions and warnings...2 Inflating instructions...3 Proper use.. 4 Deflating and storage.........6 Cleaning and troubleshooting...8

More information

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER FILTER DRYER

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER FILTER DRYER INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER FILTER DRYER MODEL F-4 DRAWING NUMBERS AD042C[ ] AD054C[ ] MANUAL NUMBER IOM-046 REVISION Rev. F, 4/23/2018 TABLE OF CONTENTS SAFETY 3 1. PRODUCT

More information

Installation, Operation and Maintenance Instructions for Mild Steel Buffer Vessel

Installation, Operation and Maintenance Instructions for Mild Steel Buffer Vessel OM006 Installation, Operation and Maintenance Instructions for Mild Steel Buffer Vessel The operating and maintenance instructions contained within this package are for standard mild steel buffer vessels

More information

GEN-1e CO2 Generator OVERVIEW INSTALLATION PROPANE GENERATORS ONLY NATURAL GAS GENERATORS ONLY

GEN-1e CO2 Generator OVERVIEW INSTALLATION PROPANE GENERATORS ONLY NATURAL GAS GENERATORS ONLY GEN-1e CO2 Generator Custom Automated Products offers economical and safe carbon dioxide generators. They produce CO2 by burning either propane or natural gas. We have designed our CO2 generator to allow

More information

Class 2 Gases Learning Tool

Class 2 Gases Learning Tool Class 2 Gases Learning Tool This work has been produced by DGL (Aust) Pty Ltd This Learner s Guide is about the skills and knowledge required to store and warehouse Class 2 Gases at DGL (Aust) facilities.

More information

CS150 CAP STAPLER OWNER S MANUAL

CS150 CAP STAPLER OWNER S MANUAL Operation Revised 6/2013 www.stingerworld.com CS150 CAP STAPLER OWNER S MANUAL! Maintenance Safety Warranty PLEASE READ! This manual contains important information about product safety. WELCOME TO STINGER

More information

Manual. Bypass Feeder. Installation Maintenance Repair Manual. Chemical Addition and Filtering

Manual. Bypass Feeder. Installation Maintenance Repair Manual. Chemical Addition and Filtering Manual Bypass Feeder Chemical Addition and Filtering Installation Maintenance Repair Manual Advantage Controls P.O. Box 1472 Muskogee, OK 74402 Phone: 918-686-6211 Fax: 918-686-6212 www.advantagecontrols.com

More information

[USING THE VITROBOT April 9, Using the Vitrobot

[USING THE VITROBOT April 9, Using the Vitrobot Using the Vitrobot The Vitrobot Mk 3 (aka Mark III, Mark 3, etc.) can control both the local temperature and the local humidity during the process of plunge-freezing a grid for cryoem. It does require

More information

CITY AND COUNTY OF DENVER CR&CF RISK UNIT Compressed Gas Safety Standard

CITY AND COUNTY OF DENVER CR&CF RISK UNIT Compressed Gas Safety Standard CITY AND COUNTY OF DENVER CR&CF RISK UNIT 65.5.11 Compressed Gas Safety Standard 1.0 Introduction 2.0 Scope This standard has been developed to protect all City and County of Denver employees and contractors

More information

Anti-flood device Model B-1

Anti-flood device Model B-1 December 4, 2009 Dry Systems 123a 1. DESCRIPTION The Anti-flood Device is required when Viking accelerators are installed on dry systems according to Viking Model E-1 Accelerator Trim Charts. In the SET

More information

HAYWARD FLOW CONTROL Series PBV Back Pressure Valve and Series RPV Pressure Relief Valve INSTALLATION, OPERATION, AND MAINTENANCE INSTRUCTIONS

HAYWARD FLOW CONTROL Series PBV Back Pressure Valve and Series RPV Pressure Relief Valve INSTALLATION, OPERATION, AND MAINTENANCE INSTRUCTIONS HAYWARD FLOW CONTROL Series PBV Back Pressure Valve and Series RPV Pressure Relief Valve INSTALLATION, OPERATION, AND MAINTENANCE INSTRUCTIONS Page 1 of 20 Page 2 of 20 TABLE OF CONTENTS Safety Warnings

More information

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST OIL-LESS TYPE IMPORTANT: PLEASE READ CAREFULLY BEFORE STARTING OPERATIONS. THE CONTENTS ARE FOR GENERAL INFORMATION OF ALL THE SIMILAR MODELS. Record

More information

OWNER S TECHNICAL MANUAL

OWNER S TECHNICAL MANUAL EL SERIES OWNER S TECHNICAL MANUAL DP7002 1 Air Operated Diaphragm Pump Description The DP7002 1 air operated diaphragm pump is the ideal device for the pumping, transfer and dispensing of chemical liquids,

More information

Spray Gun 134-A. 1 Liter High- Pressure Gel Coat. Usage Instructions & Parts List

Spray Gun 134-A. 1 Liter High- Pressure Gel Coat. Usage Instructions & Parts List 1 Liter High- Pressure Gel Coat Spray Gun 134-A Usage Instructions & Parts List www.fibreglast.com 1.800.821.3283 385 Carr Drive Brookville, OH 45309 Operating Instructions & Suggestions This is a high

More information

BS ml Bottle Sampling Unit

BS ml Bottle Sampling Unit BS500 500ml Bottle Sampling Unit User Guide www.mpfiltri.co.uk 200.307-EN Covers Model Numbers BS500 SAFETY WARNING Hydraulic systems contain dangerous fluids at high pressures and temperatures. Installation,

More information

SCA Series Inverted Bucket Steam Traps

SCA Series Inverted Bucket Steam Traps 0770050/5 IM-P077-06 ST Issue 5 SCA Series Inverted Bucket Steam Traps Installation and Maintenance Instructions 1. General safety information 2. General product information 3. Installation 4. Commissioning

More information