FPS System. MICtreat TM. instruction manual p14 series. Complete Treatment System for Fire Protection Systems. US Patent #6,221,263 B1.
|
|
- Hollie Roberts
- 5 years ago
- Views:
Transcription
1 MICtreat TM FPS System Complete Treatment System for Fire Protection Systems US Patent #6,221,263 B1 instruction manual p14 series BTIproductsLP
2 Specialists in Microbial Detection & Control BTIproductsLP BEFORE YOU USE MICtreat TM FPS CHEMICAL, READ THE FOLLOWING CAREFULLY NOTE: 1 MICtreat TM FPS Chemical is sold by BTI Products, LP for use in treatment of fire protection sprinkler systems (FPS) to prevent occurrence or recurrence of microbiologically influenced corrosion (MIC). 2 MICtreat TM FPS Chemical should only be used in FPS that have been properly tested and inspected to assess the extent and severity of deposition and corrosion. 3 It is the user s responsibility to discuss cleaning and treatment plans, including proper disposal of MICtreat TM FPS Chemical, with their local authorities, water supplier, and all other appropriate authorities to gain their permission prior to using or disposing of MICtreat TM FPS Chemical or water treated with MICtreat TM FPS Chemical. 4 An assessment report and cleaning and treatment plan should have been developed by the client and provided to BTI Products for review. BTI Products' recommendation for type and amounts of cleaning and treatment agents and protocols should be reviewed prior to implementing any cleaning or treatment using MICtreat TM FPS Chemical. 5 The FPS must have been repaired and cleaned to provide an FPS that is free of mechanical defects, dirt, oil, debris, corrosion, and other deposits before treatment of the FPS is implemented. 6 Appropriate monitoring of treated FPS is required to ensure that treatment is successful. Use test materials and follow directions provided in the MICtreat TM FPS Monitor Kit. Report any failures of MICtreat TM FPS Chemical to meet specifications to BTI Products immediately ( ). 7 Any use of MICtreat TM FPS Chemical for any purpose other than treatment of FPS to prevent MIC, or which is not in compliance with the foregoing conditions, is done without the permission of BTI Products, and BTI Products shall not be held liable for such misuse of MICtreat TM FPS Chemical.
3 MICtreat TM FPS Delivery System Instructions MICtreat TM FPS DELIVERY SYSTEM: P14 Series (US Patent #6,221,263 B1) BTIproductsLP ASSEMBLY & INSTALLATION INSTRUCTIONS 1.0 Before Assembling Your MICtreat TM FPS Delivery System NOTE: Installation and use of the MICtreat TM FPS Delivery System will require you to provide the following: 1) flow switch for activation of MICtreat TM FPS Delivery System pump, 2) fitting for attachment of injection check valve, 3) power receptacle in which pump and flow switch wiring will be joined and connected to power supply, 4) independent 115-volt, 15-amp power supply, and 5) manual activation switch for pump. 1.1 Read all instructions and manufacturer s manuals. Should you have any questions about any aspect of installation or operation of the MICtreat TM FPS Delivery System, call BTI Products at before proceeding. 1.2 Refer to LMI pump manufacturer s manual for additional information. Sections from the LMI pump manual are referenced in italics throughout these instructions. Refer to those sections for greater detail. 1.3 Install a flow switch at a point just upstream of point where MICtreat TM FPS Chemical is to be injected this is typically just downstream of the valve allowing water into the fire protection system (FPS). Use a flow switch that is activated at the lowest possible flow rate and set at zero retard. 1.4 Flow switch should lead to power receptacle into which MICtreat TM FPS Delivery System pump will be plugged. 1.5 Install a shut-off valve assembly to accommodate injection check valve on discharge-side tubing assembly leading from pump, as shown in Figure 1 (for vertical pipe installation) and Figure 2 (for horizontal pipe installation) (pages 3 and 4 of these instructions). Install this fitting at the point where chemical will be injected into the FPS. The fitting must allow discharge into the top or side of the pipe and allow discharge-side assembly injection check valve to be vertical (Section 4.7). 1.6 Close shut-off valve on shut-off valve assembly. 2.0 Assembling Your MICtreat TM FPS Delivery System 2.1 Remove all materials which comprise the MICtreat TM FPS Delivery System from boxes. 2.2 Check the List of Contents on page 2 of these instructions to make sure all necessary materials are enclosed. Pictures of pump parts can be found in Section 2.0 of the LMI Pump Instruction Manual. 2.3 Remove red plug from top fitting of pump. 2.4 Top fitting should be wet. If it is not, place a few drops of water in fitting to lubricate. Failure to lubricate this fitting may result in improper pump functioning. 2.5 NOTE: All fittings should be made hand-tight. DO NOT overtighten. DO NOT use pipe dope or tape on threads, except for connection of discharge-side assembly injection check valve to shut-off valve assembly. Carefully tool-tighten discharge-side assembly fittings. 2.6 Mount three-way function valve on top of pump head (Section 4.5). 2.7 To connect suction-side tubing assembly: Put entire suction-side assembly into tank by passing through large fill hole on top of tank. Pass small end of suction-side assembly back through small tank hole on top of tank, and attach to bottom of pump head by screwing coupling nut onto pump head connection (Sections 4.3, 4.6). 2.8 Mount pump to molded recess in tank using tank mounting kit (Section 4.2.3). 2.9 To connect return line tubing: Make sure return line tubing is securely attached to side of three-way valve (Section 4.5). Place opposite end into tank through small tank hole To connect discharge-side tubing assembly: Attach discharge-side assembly to top of three-way function valve by screwing coupling nut onto function valve connection (Sections 4.3, 4.7). Install opposite end in shut-off valve assembly, as shown in Figures 1 and 2 (pages 3 and 4 of these instructions) Read all precautions regarding MICtreat TM FPS Chemical Using appropriate precautions in accordance with MICtreat TM FPS Chemical MSDS (i.e., protection for eyes, skin, clothes, and surroundings), add five to ten gallons of MICtreat TM FPS Chemical to tank Clean and dispose of empty MICtreat TM FPS Chemical containers and any materials which may have come into contact with chemical in the appropriate manner, in accordance with MICtreat TM FPS Chemical MSDS provided. 3.0 Priming & Operating Your MICtreat TM FPS Delivery System 3.1 Connect power cord from pump to power receptacle so that pump can be switched on manually or through activation of flow switch. 3.2 Turn black by-pass knob on three-way function valve on pump counterclockwise until knob stays open. 3.3 Set pump speed knob (top knob) on front of pump to 100%. 3.4 Verify that pump stroke knob (bottom knob) on front 1
4 MICtreat TM FPS Delivery System Instructions 3.0 Priming & Operating Your MICtreat TM FPS Delivery System (Cont d) of pump is set to 100%. 3.5 Manually turn power to pump on. 3.6 Watch return line tubing until there are no bubbles coming from pump and fluid is flowing from return line tubing into tank. The pump is now primed (Section 6.2). 3.7 Turn power to pump off. 3.8 Close by-pass knob on three-way function valve. 3.9 Open shut-off valve on shut-off valve assembly Turn pump power on using manual switch Watch discharge-side tubing to make sure chemical is flowing through tubing and into shut-off valve assembly Inspect for leaks at all fittings. If leaks occur, handtighten fittings until leaks stop Leave manual switch on. Using city pressure and a method best-suited to your configuration, adjust fill rate such that water flow is between 5 and 25 gallons per minute (gpm). Set speed knob on MICtreat TM FPS Delivery System pump to appropriate setting, based on your water flow rate. Refer to Figure 3 (page 5 of these instructions) Once FPS is filled, use MICtreat TM FPS Monitor Kit to test water from far end of FPS (inspectors test valve or equivalent) to make sure desired concentration of 400 parts per million (ppm) MICtreat TM FPS Chemical is achieved Turn manual switch to off position. Leave MICtreat TM FPS Delivery System so that it is activated by flow switch Monitor FPS according to directions in MICtreat TM FPS Monitor Kit, provided with MICtreat TM FPS Delivery System. Note: Consult pump manufacturer s booklets, included in the MICtreat TM FPS Delivery System Instruction Manual, for further details. SPARE PARTS LIST It is recommended that the MICtreat TM FPS Delivery System pump be refurbished every one to two years, depending upon usage. The following pump parts should be replaced: All four check valves Diaphragm Shaft seal Please contact BTI Products at for pricing and availability. LIST OF CONTENTS 1. 1 Electronic Metering Pump (#P141352SI) 2. 1, 10-Gallon Polyethylene Tank Assembly (#27421) 3. 1 Tank Mounting Kit (#10461) 4. 1 Discharge-Side Tubing Assembly 5. 1 Suction-Side Tubing Assembly 6. 1 Return Line Tubing 7. 5 Gallons MICtreat TM FPS Chemical 8. 1 MICtreat TM FPS Monitor Kit 9. 1 Instruction Manual (includes): 1 MICtreat TM FPS Delivery System Assembly/ Installation Instructions Pump and Tank Manufacturer s Manuals MICtreat TM FPS Chemical Technical Information and Registrations MICtreat TM FPS Monitor Kit Instructions Material Safety Data Sheets (MSDS) WARRANTY BTI Products, LP s products are warranted by BTI Products, LP to perform as described in the technical literature supplied with each product, provided the products are used, stored, and maintained in accordance with the directions provided. They must also be used before the expiration date, if any. Adequate quality control must be done by the user of the products. BTI Products, LP disclaims any implied warranty of merchantability or fitness of its products for any other purpose than described in its technical literature, and in no event shall BTI Products, LP be held liable for any consequential damages arising out of the aforesaid express warranty. Warranty on electronics is two years. It is recommended that the MICtreat TM FPS Delivery System pump be refurbished every one to two years, depending upon frequency of use. No express of implied warranty is made on non-mechanical parts. Should you have questions about this product or any of the products and services we provide, please call or write: BTI Products, LP 652 Silver Hills Road Bayfield, CO products@bti-labs.com We welcome all comments and inquiries. Need Help? Call Rev 10/25/06 2
5 MICtreatTM FPS DELIVERY SYSTEM (US Patent #6,221,263 B1) Injection Tap Schematic: Vertical Pipe Discharge-side assembly injection check valve (included with MICtreatTM FPS Delivery System) ¼ 900 bend ¼ close nipple ¼ to ½ bushing ½ check valve (250 psi maximum rating) BTIproductsLP ½ close nipple ½ gate valve ¼ to ½ bushing ¼ close nipple ¼ NPT tap into pipe *All NPT Information presented is for example purposes only. Not to be taken as specifications. 3 MICtreatTM FPS Delivery System Instructions Figure 1
6 MICtreatTM FPS DELIVERY SYSTEM (US Patent #6,221,263 B1) Injection Tap Schematic: Horizontal Pipe Discharge-side assembly injection check valve (included with MICtreatTM FPS Delivery System) ¼ 900 bend ¼ close nipple ¼ to ½ bushing ½ check valve (250 psi maximum rating) ½ close nipple BTIproductsLP ½ gate valve ¼ to ½ bushing ¼ close nipple ¼ 900 bend ¼ x 4 nipple (can be any length) ¼ NPT tap into pipe *All NPT Information presented is for example purposes only. Not to be taken as specifications. 4 MICtreatTM FPS Delivery System Instructions Figure 2
7 MICtreat TM FPS Delivery System Instructions Figure 3 MICtreat TM FPS DELIVERY SYSTEM (US Patent #6,221,263 B1) P-14 Series Flow Rate Calibration BTIproductsLP Pump Speed Setting Gallons of Water Flow Which Can be Treated to 400 ppm MICtreat TM FPS Chemical at This Speed Setting Gallons of MICtreat TM FPS Chemical Delivered per Minute Note: All measurements done at 100% stroke setting. 5
8 MICtreat TM FPS Delivery System Instructions
9 MICtreat TM FPS Delivery System Instructions
10 MICtreat TM FPS Delivery System Instructions
11 MICtreat TM FPS Delivery System Instructions
12 MICtreat TM FPS Delivery System Instructions
13 Manufacturer s Manuals Instruction Manual Electronic Metering Pumps! CAUTION Carefully read and understand all precautions before installing or servicing any metering pump.! CAUTION For file reference, please record the following data: Model No: Serial No: Installation Date: Installation Location: When ordering replacement parts for your LMI Metering Pump or Accessory, please include complete Model Number and Serial Number of your unit. 201 Ivyland Road Ivyland, PA TEL: (215) FAX: (215) Replaces same of Rev. N 8/ O 11/06 1
14 Manufacturer s Manuals Contents 1.0 Introduction Spare Parts Unpacking Pre-Installation Instructions Installation Pump Location and Installation Pump Mounting Tubing Connections Multi-Function Valves Multi-Function Valve Installation Foot Valve/Suction Tubing Installation Injection Check Valve Installation Liquid End Parts List Start-Up and Adjustment Output Adjustment Controls Start-Up/Priming for Pump Supplied with Multi-Function Valve Start-Up/Priming without Accessory Valve Output Adjustment Total Pump Output Methods of External Triggering or Pacing A7, B7,C7 and P7 Pumps Calibration Pressure Control Calibration Procedure - On-Site Volumetric Calibration in External Mode Spare Parts Replacement/Routine Maintenance Depressurizing the Discharge Line (For Pumps Equipped with a 3-FV or a 4-FV Only) Liquifram (Diaphragm) Replacement Cartridge Valves, Seal Rings/Value Balls and Injection Check Valve Spring Replacement Checking Pump for Proper Zero Position (Stroke Knob) Type I: Push-on Knob Type II: Collet Knob Troubleshooting EPU Resistance Chart
15 Manufacturer s Manuals 1.0 Introduction LMI is the world s most versatile manufacturer of economical and efficient metering pumps. This manual addresses the installation, maintenance and troubleshooting procedures for manually and externally controlled pumps. LMI has a worldwide network of stocking representatives and authorized repair centers to give you prompt and efficient service. Please review this manual carefully. Pay particular attention to warnings and precautions. Always follow good safety procedures, including the use of proper clothing, eye and face protection. This manual is for Series A, B, C, E, J5, and P pumps. 1.1 Spare Parts LMI recommends replacing the elastomeric components of the pump on an annual basis. RPM Pro Pacs and spare part kits are available from your local LMI Master Stocking Distributor. 4 4
16 Manufacturer s Manuals Example: Your pump consists of two main components: 1. The Drive Assembly; and 2. The Liquid Handling Assembly. A Drive S I Liquid Handling Assembly + 5 5
17 Manufacturer s Manuals 2.0 Unpacking Check List Your carton will contain many or all of the following items. Please notify the carrier immediately if there are any signs of damage to the pump or its parts. Please refer to the enclosed Instruction Supplement for an illustration and electrical diagram of your complete pump. Metering Pump Foot Valve Tubing Depending on the model, your carton may contain 0, 1, 2 or 3 rolls of tubing. Your carton may contain a roll of clear vinyl tubing; this is for connection to the SUCTION SIDE OF THE PUMP HEAD ONLY. Ceramic Foot Valve Weight 6 6
18 Manufacturer s Manuals Instruction Supplement LMI MILTONROY Injection Check Valve Instruction Supplement * MULTI-FUNCTION Valve and Tubing * Your carton may or may not contain a 3-FV, 4-FV, or bleed 4-FV accessory. 7 7
19 Manufacturer s Manuals 3.0 Pre-Installation Instructions The following precautions should be taken when working with LMI metering pumps. Please read this section carefully prior to installation. Precautions! CAUTION Protective Clothing ALWAYS wear protective clothing, face shield, safety glasses and gloves when working on or near your metering pump. Additional precautions should be taken depending on the solution being pumped. Refer to MSDS precautions from your solution supplier.! CAUTION Water Pre-Prime All LMI pumps are pre-primed with water when shipped from the factory. If your solution is not compatible with water, disassemble the Pump Head Assembly. Thoroughly dry the pump head, valves, seal rings, balls and Liquifram (diaphragm). Reassemble head assembly tightening screws in a crisscross pattern. Refill the pump head with the solution to be pumped before priming the pump. (This will aid in priming.)! CAUTION Solution Compatibility Determine if the materials of construction included in the liquid handling portion of your pump are adequate for the solution (chemical) to be pumped. LMI pumps are tested by NSF for use on muriatic acid and sodium hypochlorite. Always refer to the solution supplier and the LMI Chemical Resistance Chart for compatibility of your specific LMI metering pump. Contact your local LMI distributor for further information. 8 8
20 Manufacturer s Manuals! CAUTION Tubing Connections Inlet and outlet tubing or pipe sizes must not be reduced. Make certain that all tubing is SECURELY ATTACHED to fittings prior to start-up (see Section 4.3, Tubing Connections). ALWAYS use LMI supplied tubing with your pump, as the tubing is specifically designed for use with the pump fittings. It is recommended that all tubing be shielded to prevent possible injury in case of rupture or accidental damage. If tubing is exposed to sunlight, black UV resistant tubing should be installed. Check tubing frequently for cracks and replace as necessary.! CAUTION Fittings And Machine Threads All fittings should be hand-tightened. An additional 1/8-1/4 turn after the fitting contacts the seal ring may be necessary to provide a leak-proof seal. Excessive overtightening or use of a pipe wrench can cause damage to the fittings, seals, or pump head. All LMI pumps have straight screw machine threads on the head and fittings and are sealed by the seal rings or O-rings. DO NOT use Teflon tape or pipe dope to seal threads. Teflon Tape may only be used on the 1/2" NPT thread side of the Injection Check Valve as well as stainless steel liquid end connections.! CAUTION Plumbing Always adhere to your local plumbing codes and requirements. Be sure installation does not constitute a cross connection. Check local plumbing codes for guidelines. LMI is not responsible for improper installations.! CAUTION Back Pressure/Anti-Syphon Valve If you are pumping downhill or into low or no system pressure, a back pressure/anti-syphon device such as LMI's Four Function Valve should be installed to prevent overpumping or syphoning. Contact your LMI distributor for furthur information. 9 9
21 Manufacturer s Manuals! WARNING Electrical Connections To reduce the risk of electrical shock, install only on a circuit protected by a ground-fault circuit-interrupter (GFCI). The metering pump must be plugged into a grounded outlet with ratings conforming to the data on the pump control panel. The pump must be connected to a good ground. DO NOT USE ADAPTERS! All wiring must conform to local electrical codes. To LMI Pump To LMI Pump INCORRECT CORRECT 4.0 Installation 4.1 Pump Location and Installation 10 Locate pump in an area convenient to solution tank and electrical supply. The pump should be accessible for routine maintenance, and should not be subjected to ambient temperatures above 122 F (50 C). If the pump will be exposed to direct sunlight, LMI black, UV resistant tubing should be installed. 4.2 Pump Mounting The pump can be mounted in one of two ways: A. FLOODED SUCTION (ideal installation); or B. SUCTION LIFT - when suction lift is less than 5 feet (1.5 m) for solutions having a specific gravity of water. For denser solutions, consult distributor. Your LMI metering pump must be mounted so that the suction and discharge valves are vertical. NEVER position pump head and fittings horizontally. 10
22 Manufacturer s Manuals Flooded Suction! CAUTION The pump is mounted at the base of the storage tank. This installation is the most trouble-free, and is recommended for very low outputs, solutions that gasify, and high-viscosity solutions. Since the suction tubing is filled with solution, priming is accomplished quickly and the chance of losing prime is reduced. When pumping downhill or into low or no pressure system, a back pressure/anti-syphon device should be installed to prevent overpumping or syphoning. Although popular for all solutions, LMI recommends flooded suction installations for all high-viscosity fluid applications. Tee Injection Check Valve LMI 35 Gallon Solution Tank INCORRECT Avoid this type of false flooded suction. CORRECT Vent Injection Check Valve Polymer (Polyelectrolyte) Drum LMI Pump Pump Stand Barrel Cradle CORRECT 11 11
23 Manufacturer s Manuals Suction Lift - Wall Bracket Mount The pump may be mounted using an LMI Wall Mount Bracket Assembly (part no ) directly above the solution tank. A pump mounted in this manner allows for easy changing of solution tanks or drums. Tee Injection Check Valve Pressure Line LMI Pump Front Mount LMI Pump Rear Mount Solution Tank 2.0 in. (50 mm) Space for Sediment Accumulation 12 12
24 Manufacturer s Manuals Suction Lift - Tank Mount The pump may be mounted on a molded tank provided there is a recess to keep the pump stationary. LMI 10-gallon tank (part no ), 35-gallon tank (part no ), and 50-gallon tank (part no ) have molded recesses for pump mounting. Pressure Line Tee Injection Check Valve LMI Pump LMI 50 Gallon Solution Tank LMI Pump Ceramic Weight LMI 10 Gallon Solution Tank Foot Valve 2.0 in. (50 mm) Space for Sediment Accumulation Foot Valve 2.0 in. (50 mm) Space for Sediment Accumulation 13 13
25 Manufacturer s Manuals Suction Lift - Shelf Mount The pump may be mounted on a shelf (customer supplied) maintaining a suction lift of less than 5 ft (1.5 m). An LMI mounting kit (part number 10461) is available for securing the pump to a shelf. Flow Flow Injection Check Valve Anti-Syphon Pressure Relief Valve (Optional Accessory) LMI Pump Solution Drum 2.0 in. (50 mm) Space for Sediment Accumulation 14 14
26 Manufacturer s Manuals 4.3 Tubing Connections! CAUTION! CAUTION A. Use only LMI tubing. B. DO NOT USE CLEAR VINYL TUBING ON THE DISCHARGE SIDE OF THE PUMP. The pressure created by the pump can rupture vinyl tubing. C. Before installation, all tubing must be cut with a clean square end. D. Valve and head connections from the factory are capped or plugged to retain pre-prime water. Remove and discard these caps or plugs before connecting tubing. DO NOT USE PLIERS OR PIPE WRENCH ON COUPLING NUTS OR FITTINGS. Tubing Connection 1/4" O.D. (.250") Tubing Tubing Connection 3/8" O.D. (.375") Tubing 1/4" OD Tubing 3/8" OD Tubing Ferrule 1" (25 mm) Fitting Coupling Nut Bottom Tubing in Fitting. Push and hold tubing down while tightening coupling nut. Force tubing completely over entire nozzle into groove. Push and hold tubing down while tightening coupling nut. Coupling Nut Clamp Ring Nozzle Fitting Groove Tubing Connection 1/2" O.D. (.5") Tubing Pipe Thread Connection 1/4" or 1/2" NPT 1/2" OD Tubing Coupling Nut Clamp Ring NPT Pipe Connection Nozzle Fitting Force tubing completely over entire nozzle. Push and hold tubing down while tightening coupling nut. Apply Teflon tape here. (Customer supplied pipe connections) 15 15
27 Manufacturer s Manuals 4.4 Multi-Function Valves Your pump may be equipped with one of the following multifunction valves: 3-FV, 4-FV, Bleed 4-FV, or standard discharge valve. If your pump is not equipped with a multi-function valve and you feel it is needed in your application, it can be purchased as an accessory. Contact your local LMI stocking distributor Three Function Valve (3-FV) 1. Pressure Relief If the discharge line is over pressurized, the valve opens sending solution back to the supply tank. 2. Line Depressurization Opening the relief knob provides line drain back to the supply tank. 3. Priming Aid Opening the relief knob assists in priming the pump by venting the discharge line to the atmosphere Four Function Valve (4-FV) 1. Pressure Relief If the discharge line is over pressurized, the valve opens sending solution back to the supply tank. 2. Line Depressurization Opening the relief knob provides line drain back to the supply tank. 3. Anti-Syphon Prevents syphoning when pumping solution downhill or into a vacuum. 4. Back Pressure Supplies approximately 25 psi back pressure to prevent overpumping when little or no system back pressure is present
28 Manufacturer s Manuals Bleed Four Function Valve (Bleed 4-FV) DO NOT submerge return line in solution 1. Line Depressurization Opening the relief port provides line drain back to the supply tank. 2. Anti-Syphon Prevents syphoning when pumping solution downhill or into a vacuum. 3. Back Pressure Supplies approximately 25 psi back pressure to prevent overpumping when little or no system back pressure is present. 4. Bleed Function Manually adjusted valve provides continuous bleed of entrapped vapors from Sodium Hypochlorite or Hydrogen Peroxide. Flow Flow LMI 4-FV (Accessory) Prevents syphoning when pumping into suction side of recirculating pump (Vacuum) Recirculating Pump LMI Pump Tank LMI 4-FV (Accessory) Prevents over pumping and syphoning when pumping downhill into low or no pressure Tank Injection Check Valve To Injection Point Typical 4-FV Installation 17 17
29 Manufacturer s Manuals 4.5 Multi-Function Valve Installation To install the multi-function valve, remove the yellow screw cap on the top of the pump head and screw in the valve so that it contacts the seal ring. An additional 1/8-1/4 turn may be necessary to prevent leakage. 1/4" O.D. tubing connects to the side of the valve and acts as a return line to the solution tank. To ensure priming, this tubing must NOT be submerged in the solution.! CAUTION This return line tubing must be secured to ensure pumped solution will safely return to supply tank. Discharge 1/4" Tubing Connector To Solution Tank or Drum 1 Inch (25 mm) Bottom Tubing in Fitting Push and hold while tightening connector. To Pump Head Multi-Function Valve Tubing Connection 4.6 Foot Valve/Suction Tubing Installation The Foot Valve acts as a check valve to keep the pump primed in suction lift applications. The foot valve is designed to be submersed in the solution tank or drum and must sit in a vertical position at the bottom. Position approximately 2 inches (50 mm) off the bottom if the tank or drum contains sediment
30 Manufacturer s Manuals Pump models equipped with high-viscosity liquid ends are not equipped with foot valves. Flooded suction is recommended. A 1/ 2" NPT connector is included for flooded suction installations. The ceramic weight, when installed, positions the foot valve in a vertical position. 1. Attach the foot valve to one end of the suction tubing (see Tubing Connections, Section 4.3). 2. Slide the ceramic weight over the tubing end until it contacts the top of the foot valve coupling nut. 3. Place foot valve and tubing into the solution tank. Check that the foot valve is vertical and approximately 2 inches (50 mm) from the bottom of the tank or drum (see illustration). Connect the other end of the tubing to the suction side of the pump head (bottom side) (see Tubing Connections, Section 4.3). LMI Pump LMI Pump Use Ceramic Weight Foot Valve Must Remain Vertical Foot Valve Tilted Sideways WILL NOT PRIME INCORRECT 2.0 in. (50 mm) for Sediment Accumulation CORRECT Proper Foot Valve Position 19 19
31 Manufacturer s Manuals 4.7 Injection Check Valve and Discharge Tubing Installation The Injection Check Valve prevents backflow from a treated line. Connect the Injection Check Valve to your DISCHARGE (outlet) line. Any size NPTF fitting or pipe tee with a reducing bushing to 1/2" NPTF will accept the injection check valve. Use Teflon tape or pipe dope to seal the pipe threads only. When installing the Injection Check Valve, be sure to position it so that the valve enters the bottom of your pipe in a vertical position. Variations left and right within 80 are acceptable (see illustration below). After cutting an appropriate length of tubing, connect tubing to the injection check valve then back to the discharge side of the pump head valve or discharge fitting (top side), making sure it does not crimp or come into contact with hot or sharp surfaces (see Tubing Connections, Section 4.3). Pipe Cross Section 1 Pipe Tee Reducing Bushing 1 to 1/2 NPT (Customer Supplied) Injection Check Valve CORRECT Use Teflon Tape here on Pipe threads only Do NOT use Teflon tape On machined Threads 40 Variation Acceptable Injection Check Valve Flow Flow Typical Injection Check Valve Installations 20 20
32 Manufacturer s Manuals 5.0 Liquid End Parts List 1 Flapper valve 2 Injection check valve body 3 Injection check valve spring 4 Check valve ball 5 Seal ring 6 Cartridge valve 7 Cartridge valve o-ring 8 Cartridge valve washer 9 Valve seat 10 Clamp ring 11 Ferrule 12 Clamp sleeve 13 Tubing adapter 14 Coupling nut 15 Discharge tubing 16 Valve housing 17 Multi-function valve 18 High-viscosity spring 19 Liquifram 20 Pump head 21 Pump head screw 22 Suction tubing 23 Foot valve seat 24 Foot valve screen 25 High-viscosity valve seat 26 H.V. tubing clamp 27 H.V. suction tubing 28 H.V. Tubing x 1 /2 NPT connector 29 Injection check valve assembly 30 Discharge valve assembly 31 Suction valve assembly 32 Pump head assembly 33 Foot valve assembly 34 Injection Seat PTFE 35 Ceramic Weight 36 Return Line 37 Cap ASM (Black Knob) 38 Cap ASM (Yellow Knob) 39 Multi-Function Valve Body 40 Nut Multi-Function Valve 41 Screw Multi-Function Valve 42 Return Line Coupling Nut 43 Adjustment Screw B/4-FV 44 Cap B/4-FV 45 Plug B/4-FV 46 Gasket B/4-FV 47 Small O-Ring B/4-FV 48 Large O-Ring B/4-FV NOTE: This illustration is a visual representation of all LE components. Liquid ends will 21 not include all parts shown. 21
33 Manufacturer s Manuals 6.0 Start-up and Adjustment a.) The pump is normally self-priming if suction lift is 5 ft (1.5m) or less and the steps below are followed. b.) Pumps are shipped from the factory with water in the pump head to aid in priming. 6.1 Output Adjustment Controls Manual series pump controls are not equipped with pressure control. 1. Pressure Control Adjustment (if equipped): Pressure control provides the adjustment of the pump s pressure capability and power consumption, reducing heat, pipe shock and pulsation while increasing pump life. See Section 7.0 after priming for proper adjustment settings. 2. Speed Adjustment (Upper Knob) (if equipped): Speed control provides adjustment of the percent of maximum strokes per minute. Turning this knob clockwise increases stroke frequency (speed). 3. Stroke Adjustment (Lower Knob): Stroke control provides adjustment of the percent maximum of solution discharged during each pump actuation. Turning this knob clockwise increases solution displacement. A7 and P7 Only: When operating the pump in external mode, the speed control knob should be turned fully counter-clockwise. A34 and A37 Only: Pump comes equipped with a range selector switch which provides high or low speed adjustment. The high setting provides speed adjustments between 8 and 100 strokes per minute. The low setting provides accurate speed adjustments between 1 and 12.5 strokes per minute for applications requiring infrequent stroking
34 Manufacturer s Manuals 6.2 Start-Up/Priming for Pump Supplied with Multi-Function Valve! CAUTION Read this entire section completely before proceeding. When all precautionary steps have been taken, the pump is mounted, and the tubing is securely attached, you may now start priming the pump. 1. Plug in or switch the pump on. 2. While the pump is running, set the speed knob at 80% and the stroke knob at 100%. If the pump is equipped with pressure control, turn fully clockwise. 3. 1/4 turn open the relief side (black knob) of the multifunction valve. 3A. (Bleed 4FV only) With screwdriver rotate bleed adjustment screw counter-clockwise. 2 full turns. When solution begins to flow through translucent bleed return tubing, the pump is primed. Stop pump. 4. The suction tubing should begin to fill with solution from the tank. 5. A small amount of solution will begin to discharge out the return line of the multi-function valve. Once this happens, 1/4 turn or release the knob and SHUT THE PUMP OFF. (If pump is not equipped with an on/off switch, disconnect the power cord.) 6. The pump is now primed. 6A. (Bleed 4FV only) a. Start pump and let pump inject solution into the discharge line. b. Close the bleed adjustment screw by rotating it clockwise with a screwdriver. c. Now adjust the pump stroke length and/or speed (frequency) to a range approximately 25% higher than you would normally want for the process. d. Slowly rotate bleed adjustment screw counter-clockwise. until just a small amount of solution begins to trickle 23 23
35 Manufacturer s Manuals down inside the bleed return tubing. A small amount of solution pumped back to the tank with each stroke of the pump will allow gas and air to escape without air or gas locking in the pump head. 7. Proceed to output adjustment, Section 6.4. If the pump does not self-prime, remove the multi-function valve on the discharge side of the pump head. Remove the check valve and pour water or solution into the port until the head is filled. Replace valve, then follow start up/priming steps. 6.3 Start-Up/Priming without Multi-Function Valve! CAUTION Read this entire section completely before proceeding. When all precautionary steps have been taken, the pump is mounted, and the tubing is securely attached, you may now prime the pump. 1. Plug in or switch on the pump. 2. While the pump is running, set the speed knob at 80% and the stroke knob at 100%. If the pump is equipped with pressure control, turn fully clockwise. 3. The suction tubing should begin to fill with solution from the tank. 4. Once the solution begins to exit the pump head on the discharge side, SHUT THE PUMP OFF. (If pump is not equipped with an on/off switch, disconnect the power cord). 5. The pump is now primed
36 Manufacturer s Manuals 6. Proceed to output adjustment, Section 6.4. If the pump does not self-prime, remove the fitting on the discharge side of the pump head. Remove the ball and pour water or solution into the port until the head is filled. Replace valve, then follow start up/priming steps. 6.4 Output Adjustment Once the pump has been primed, an appropriate output adjustment MUST be made. Pump output should be calculated and adjustments made accordingly. 6.5 Total Pump Output Calculate the total output of the pump as follows: PUMP OUTPUT = MAX PUMP OUTPUT x % SPEED x % STROKE Example: A SI Use MAX Output (from dataplate on bottom center of pump control panel) = 24 GPD (24 gallons per day). If the pump is set at 60% speed and 70% stroke length, the approximate pump output is: 24.0 x 0.60 x 0.70 = GPD (gallons per day). Divide by 24 (hours in one day) to calculate in gallons per hour. If pump is not equipped with speed adjustment, calculate by Max Pump Output x % Stroke only
37 I T L A N S T A T I I Manufacturer s Manuals 7.0 Methods of Externally Triggering P R C E S T N O S U R R O E L S E P E D OFF 120 V or V Hz ONA A NI ON ON F OUND AT INTERNAL EXTERNAL TO ADJUST 70 TURN WHILE 0 UNIT IS 80 RUNNING ACTON, MA U.S.A STROKE EMPTY DISCONNECT POWER CORD BEFORE REMOVING PANEL E X T E R N A L I N P U T Optional Low Level Sensor Part No NPN Transistor Base goes high to trigger pump 3. PNP Transistor Base goes low to trigger pump Opto Method of Triggering LMI Pump Through 4-Pin Connector 1. Switch Closure Switch closing triggers pump 4. Isolator + White Black + White Black + White Black PIN or Remote Start/Stop Stop (Open) Start (Closed) 26 Switch or transistors must be capable of switching 15V DC at 2 milliamperes. Minimum time in low impedance state (ON) is 50 milliseconds. Minimum time in high impedance state (OFF) is 100 milliseconds
38 Manufacturer s Manuals or Pacing A7, B7, C7 and P7 Pumps 120 V or V Hz? X DPC- 40 Higher Frequency Pulses Lower Frequency Pulses DPC milliamp DC MICROPACE A/D CONVERTER MP-100 MICROPACE 4 DIGIT DIVIDER MP-400 MICROPACE 4 DIGIT MULTIPLIER MP-500M 4-20 milliamp DC Higher Frequency Pulses Lower Frequency Pulses Pulse Transmitter Programmable Divider Programmable Divider RFP Flowmeter FC Flowmeter/Contactor 27 27
39 Manufacturer s Manuals 8.0 Calibration Once installation is complete and the approximate output has been determined, the pump should be calibrated to adjust speed and stroke for your actual desired output. (Calibration cylinders may be purchased from your local LMI distributor, ref. publication 1798.). 1. Be sure the pump is primed, and discharge tubing and Injection Check Valve are installed as they would be in normal service (i.e., including factors such as injection pressure, fluid viscosity, and suction lift). 2. Place the Foot Valve in a graduated container with a volume of 1000 ml or more. 3. Plug in and switch pump to Internal Mode. Pump until all the air is exhausted from the suction line and head. 4. Turn the pump off. Refill graduated container to a level starting point. If pump is equipped with pressure control, see Section 8.1 before proceeding. 5. Using a stopwatch or timer, turn the pump on for a measured amount of time (50 pump strokes minimum). The longer the time period, the more confident you can be of the results. Be sure to count the number of strokes during the calibration period when making comparisons. 6. Turn the pump off. Note the time elapsed in relation to volume displaced in the graduate. Now, calculate the output in the time unit you choose (minutes, hours, days, etc.). 7. If the output is too low or too great, adjust speed and or stroke, estimating required correction and repeat steps
40 50 Manufacturer s Manuals Tee Pressure Line Injection Check Valve LMI Pump Suction Lift Graduate 1000 ml Min. Foot Valve 8.1 Pressure Control Adjust Pressure Control: While unit is running, turn Pressure Control Potentiometer slowly counter-clockwise until unit just begins to stall. From this stall point, now turn Pressure Control Potentiometer clockwise halfway between the stall point and maximum setting. This is the optimum pressure control setting for your application. Increase setting if back pressure is increased. Adjusting pressure control decreases pressure rating of pump. Adjust pressure control to reduce heat, shock, and pulsations; and to prolong pump life
41 Manufacturer s Manuals 8.2 Calibration Procedure - On-Site Volumetric Calibration in External Mode 1. Since pump output is governed by an external device such as Flowmeter-Pulser, Liquitron Controller, or 4-20 ma DC signal from an instrument with an LMI Analog-to-Digital Converter, only the output per stroke may be calibrated. 2. With pump primed and discharge tubing connected to the injection point as it would be in normal service, place Foot Valve Assembly in a graduated container with a volume of 1000 ml or more. 3. Switch pump to Internal mode with Speed Knob set at 100 until air is exhausted from suction line and pump head. 4. Adjust Pressure Control (if desired) - See Section Switch pump OFF and note solution level in graduated container. Refill graduate to a starting point. 6. Switch pump ON and count the number of strokes for exactly one minute, then switch pump OFF. 7. Note volume pumped during the calibration period of one minute. Divide into this the number of strokes to determine the volume of solution pumped per stroke. Example: 500 ml in 100 strokes = 5.0 ml per stroke. Multiply this by your expected stroke rate per minute, per hour or per day and compare with desired output requirements. 8. Adjust Stroke Length Knob (lower knob) to your best estimate of required correction and repeat calibration procedure. 9.0 Spare Parts Replacement Routine Maintenance 9.1 Depressurizing the Discharge Line (For Pumps Equipped with a 3-FV or a 4-FV only) CAUTION 30! ALWAYS wear protective clothing, face shield, safety glasses and gloves when performing any maintenance or replacement on your pump. 30
42 Manufacturer s Manuals! CAUTION! CAUTION Read steps 1 and 2 below before proceeding. 1. Be sure the Injection Check Valve is properly installed and is operating. If a shut off valve has been installed downstream of the Injection Valve, it should be closed. Be sure your relief tubing is connected to your multi-function valve and runs back to your solution drum or tank. 2. 1/4 turn the black knob on the valve. The discharge line is now depressurized. Keep valve open until solution drains back down the discharge tubing into solution drum or tank. Then 1/4 turn knob to normal position. 9.2 Liquifram (Diaphragm) Replacement! CAUTION ALWAYS wear protective clothing, face shield, safety glasses and gloves when working near or performing any maintenance or replacement on your pump. See MSDS information from solution supplier for additional precautions. LMI metering pumps are designed for trouble-free operation, yet routine maintenance of elastomeric parts is essential for optimum performance. This involves replacing the Liquifram, cartridge valves or seal rings/valve balls, multi-function valve cap assemblies and the injection check valve spring. LMI recommends replacing these parts at least once a year; however, frequency will depend on your particular application. When replacing the Liquifram and the cartridge valves or seal rings/valve balls, the injection check valve spring should also be replaced (see next Section 9.3). A Spare Parts Kit (SP-#) or RPM Pro Pac kit containing these parts may be obtained from your local distributor. Replacing the Liquifram : 1. Carefully depressurize, drain, and disconnect the discharge line (see Section 8.1 in this manual). Place the Foot Valve into a container of water or other neutralizing solution. Turn the pump on to flush the head assembly. Once the pump head has been flushed, lift the Foot Valve out of the solution and continue to pump air into the pump head until the pump head is purged of water or neutralizing solution
43 Manufacturer s Manuals! CAUTION If the liquid cannot be pumped due to Liquifram rupture using protective clothing, gloves and face shield, carefully disconnect the suction and discharge tubing. Remove the four screws to the head and immerse the head in water or other neutralizing solution. 2. Start the pump. While running, set the stroke knob to zero and turn the pump off. See Section 10.0 for proper zero. 3. With the unit off, unscrew the Liquifram by carefully grasping the outer edge and turning it counter-clockwise. Discard old Liquifram. Remove the Liquifram disk if so equipped (located behind the Liquifram ) and check that the size code matches the size code on the replacement Liquifram (see illustration). 4. Reinstall the disk so the alignment pin on the disk (if present) seats in the recessed hole in the EPU. Be careful not to scratch the Teflon face of the new Liquifram. For Series B, C, & E For Series A, J, & P Spacer Size Code Marking Liquifram Size Code Marking Liquifram 6.0 Size Code Marking on Back NOTE: Size code markings for pumps supplied with 6.0 Liquiframs (diaphragms) should be referenced to the 6.0 Black Adapter, not the 3.0 Spacer Liquifram Disk Recessed Hole Spacer Start the pump and turn the stroke knob to the setting indicated on the following Stroke Setting Chart which matches the pump series number located on the pump dataplate. With the pump stroking (running), screw on the new Liquifram clockwise until the center begins to buckle inwards. Stop the pump. 32
44 Manufacturer s Manuals Liquifram Stroke Setting Chart Pump Series Stroke Knob Setting All A, B, J, P, Z Series C10, C11, C12, C70, C71, C72, C76, 90% C90, C91, C92, E70, E71, E72 All L Series 85% C78 50% C13, C14, C73, C74, C77, C93, C94, E73, E74 70% All M Series 100% * * Liquifram on M Series pumps only, must be bottomed completely (turned all the way in). Do Not Use Straight Edge. 6. Grasp the outer edge of the Liquifram and adjust by screwing it in or out so that the center of the Liquifram is flush with the outside of the spacer edge (see illustration below). 7. Once the Liquifram is properly positioned, remount the pump head to the spacer using the four (4) screws. Tighten in a criss-cross pattern. After one week of operation, recheck the screws and tighten if necessary
45 Manufacturer s Manuals 9.3 Cartridge Valves, Seal Rings/Valve Balls and Injection Check Valve Spring Replacement! CAUTION ALWAYS wear protective clothing, face shield, safety glasses and gloves when working on or performing any maintenance or replacement on your pump. See MSDS information from solution supplier for additional precautions.! CAUTION 1. Refer to the LMI Metering Pump Price List for the proper Spare Parts Kit or RPM Pro Pac kit number or contact your local LMI stocking distributor. 2. Carefully depressurize and disconnect the discharge line (see Section 9.1 in this manual). Place the Foot Valve into a container of water or other neutralizing solution. Turn the pump on to flush the head assembly. Once the pump has been flushed, lift the Foot Valve out and continue to pump to let air into the pump head until pump is purged of water or neutralizing solution. Once the pump has been flushed, lift the Foot Valve out and continue to let air into the pump head until pump is purged of water or neutralizing solution. If the liquid cannot be pumped due to Liquifram rupture, with protective clothing, gloves and face shield, carefully disconnect the tubing and four screws to remove the head. Immerse the head in water or other neutralizing solution. Spare part replacement kits include specific instructions for valve replacement. Please follow the instructions included with the replacement kit. IMPORTANT: Before disassembling the check valves, note the orientation of the valve. 3. Carefully disconnect one tubing connection and fitting at a time, then remove and replace the worn valve. If necessary, carefully loosen stuck valves by prying side to side using a small screwdriver through the center hole of the valve. 4. Install new check valves in each location
46 Manufacturer s Manuals! CAUTION IMPORTANT: Note correct orientation of each check valve. 5. Install the new spring in the Injection Check Valve. Depressurize and drain pipeline (or isolate I.C.V. point using valves) so that I.C.V. can safely be disassembled Checking Pump for Proper Zero Position (Stroke Knob) 1. With pump running, turn stroke knob counter-clockwise toward zero or end of black or red band on dial. 2. LISTEN to the clicking as the pump is running. The pump should operate quietly at the zero position (no clicking). 3. If the pump continues to click at zero or stops clicking before zero is reached, the pump zero must be reset (see Section 10.1 or 10.2) Type I - Push on Knob Re-Zeroing and Stroke Knob Disassembly and Assembly ` 1. Remove stroke knob from the pump by grasping the knob firmly and pulling it toward you. 2. Pry off the yellow cap. 3. Place the knob on a flat surface. 4. Using needle-nose pliers, squeeze the inner section together while lifting the outer section up. 5. Push the inner section back onto the D shaped stroke shaft. 6. With the pump running, zero the pump by turning the inner section of the knob counter-clockwise until the pump stops clicking
47 Manufacturer s Manuals 7. Position the outer section of the knob so that the pointer aligns with zero on the nameplate or end of the black or red band. 8. Push down on the outer section (a snap sound indicates parts are locked together). 9. Replace the yellow cap over the outer section of the knob, aligning the tabs on the cap with the slots inside the knob. Stroke Knob Assembly (Type I) 36 36
48 Manufacturer s Manuals 10.2 Type II Collet Knob Re-Zeroing and Stroke Knob Disassembly and Assembly 1. Remove Yellow Cap. 2. Hold knob with soft jaw pliers. 3. Disconnect knob by loosening 5/16" (8 mm) collet nut. There is no need to remove nut. 4. Remove knob by pulling towards you. 5. With pump running, zero the pump using a screw driver to turn the stroke shaft counter-clockwise until the pump just stops clicking. 6. Pump is now zeroed. 7. Position knob at zero, or the end of the low range band, and tighten 5/16" (8 mm) collet nut. 8. Replace yellow cap
49 Manufacturer s Manuals 11.0 Troubleshooting PROBLEM POSSIBLE CAUSE Pump Will Not Prime 1. Pump not turned on or plugged in. 2. Output dials not set properly. 3. Foot Valve not in vertical position on bottom of tank. 4. Pump suction lift too high. 5. Suction tubing is curved or coiled in tank. 6. Fittings are over tightened. 7. Air trap in suction valve tubing. 8. Too much pressure at discharge. (Pumps without multi-function valve.) Pump Loses Prime 1. Solution container ran dry. 2. Foot Valve is not in a vertical position on the bottom of the tank. 3. Pump suction lift is too high. 4. Suction tubing is curved or coiled in tank. 5. Fittings are over tightened. 6. Air trap in suction valve tubing. 7. Air leak on suction side
50 Manufacturer s Manuals SOLUTION 1. Turn on pump/plug in pump. 2. Always prime pump with speed at 80% and stroke at 100%. 3. Foot Valve must be vertical (see Foot Valve Installation, Section 4.6). 4. Maximum suction lift is 5 ft (1.5 m). Pumps with High Viscosity Liquid Handling Assemblies require flooded suction. 5. Suction tubing must be vertical. Use LMI tubing straightener supplied with pump (see Section 4.6). 6. Do not overtighten fittings. This causes seal rings to distort and not seat properly which causes pump to leak back or lose prime. 7. Suction tubing should be as vertical as possible. AVOID FALSE FLOODED SUCTION! (see Section 4.2.1). 8. Shut off valves in pressurized line. Disconnect tubing at injection check valve (see Priming Section 6.0). When pump is primed, reconnect discharge tubing. 1. Refill container with solution and reprime (see Section 6.0). 2. Foot Valve must be vertical (see Foot Valve Installation, Section 4.6). 3. Maximum suction lift is 5 ft (1.5 m). Pumps with High Viscosity Liquid Handling Assemblies require flooded suction. 4. Suction tubing must be vertical. Use LMI tubing straightener supplied with pump (see Section 4.6). 5. DO NOT OVERTIGHTEN FITTINGS. This causes seal rings to distort and not seat properly which caused pump to leak back or lose prime. 6. Suction tubing should be as vertical as possible. AVOID FALSE FLOODED SUCTION! (see Section 4.2.1). 7. Check for pinholes, cracks. Replace if necessary
51 Manufacturer s Manuals Troubleshooting (continued) PROBLEM POSSIBLE CAUSE Leakage at tubing 1. Worn tubing ends. 2. Loose or cracked fitting. 3. Worn seal rings. 4. Solution attacking Liquid Handling Assembly material. Low Output or Failure 1. Pump s maximum pressure rating to Pump Against is exceeded by injection pressure. Pressure 2. Worn Seal Rings. 3. Ruptured Liquifram. 4. Incorrect stroke length. 5. Tubing run on discharge may be too long. 6. Clogged Foot Valve strainer. Failure to Run 1. Pump not turned on or plugged in. 2. EPU failure. 3. Pulser failure. Excessive Pump Output 1. Syphoning. (Pumping downhill without a multi-function valve). 2. Little or no pressure at injection point. 3. Excessive strokes per minute
52 Manufacturer s Manuals SOLUTION 1. Cut about 1 in (25 mm) off tubing and then replace as before. 2. Replace fitting if cracked. Carefully hand tighten fittings. DO NOT USE PIPE WRENCH. Once fitting comes into contact with seal ring, tighten an additional 1/8 or 1/4 turn. 3. Replace balls and seal rings (see Section 8.3) Spare Parts (SP-#). 4. Consult your local distributor for alternate materials. 1. Injection pressure cannot exceed pump s maximum pressure. See pump data plate. 2. Worn seal rings or cartridge valves may need replacement (see Section 9.3). Spare Parts (SP- #), or RPM Pro Pac kit. 3. Replace Liquifram (see Section 9.2). 4. Check zero on pump/re-zero pump (see Section 10.0). 5. Longer tubing runs may create frictional losses sufficient to reduce pump s pressure rating. Consult factory for more information. 6. Remove Foot Valve strainer when pumping slurries or when solution particles cause strainer to clog. 1. Turn on or plug in pump. 2. Disassemble pump and measure the resistance of the EPU across the EPU wires. Resistance reading should be in accordance to the EPU Resistance Chart (see Section 12.0). Also, check EPU leads to ground. Consult supplier or factory. 3. The pulser should be replaced if EPU checks out OK. Consult supplier or factory. 1. Move injection point to a pressurized location or install an LMI 4-FV (see Section 4.4). 2. If pressure at injection point is less than 25 psi (1.7 Bar), an LMI 4-FV should be installed (see Section 4.4). 3. Replace pulser or resistor. Consult factory
53 Manufacturer s Manuals 12.0 EPU Resistance Chart Pump Series Voltage Coil Resistance 20 o C (68 o F)* A14, A15, A16, A34 A74, A75, A76 A94, A95, A96 J02, J03, J04, J05, J VAC J13, J15, J VAC PW4, PW5, PW6 P04, P05, P06 P08, P14, P15 P16, P18, P74 P75, P76, P78 (see Note 1) A17, A37, A77, A97, A18, A VAC P02,P VAC P12,P13 (see Note 2) A17, A37, A77, A97, A18, A VAC P02, P VAC P12, P13 J54D, J55D, J56D 12 VDC D10, D11, D12, D13, D VAC D70, D71, D72, D73, D VAC E70, E71, E72, E73, E VAC VAC B11, B12, B13, B VAC B71, B72, B73, B VAC C10, C11, C12, C13, C VAC C70, C71, C72, C73, C VAC C76, C77, C VAC VAC * Let pump cool down completely before checking resistance. EPU checked within 10 hours of operation can increase coil resistance reading as much as 20%. NOTES: 1. Pumps with serial numbers LOWER than: Pumps with serial numbers HIGHER than: Ivyland Road Ivyland, PA TEL: (215) FAX: (215) LMI Milton Roy - All Rights Reserved Printed 44 in USA Specifications subject to change without notice. Liquifram, Liquitron, Micropace, and Pro Pac are trademarks of Liquid Metronics Incorporated, Teflon is a registered trademark of E. I. du Pont de Nemours & Co., Inc. 42
54 Manufacturer s Manuals Instruction Supplement Series P1 Electronic Metering Pump Metering Pump Component Diagram 201 Ivyland Road Ivyland, PA USA TEL: (215) FAX: (215) Replaces same of Rev. B 2/ C 2/03
55 Manufacturer s Manuals Series P1 Drive Assembly Parts List KEY MODEL PART NO DESCRIPTION QTY NO 1 P CE EPU W/DISK P123, P125, P CE EPU W/DISK P CE EPU W/DISK P133, P135, P CE EPU W/DISK P CE EPU W/DISK P143, P145, P CE EPU W/DISK P CE EPU W/DISK P153, P155, P CE EPU W/DISK P CE EPU W/DISK P163, P165, P CE EPU W/DISK P121, P HOUSING ASSEMBLY W/PULSER, 115V 1 P123, P HOUSING ASSEMBLY W/PULSER, 230V DIN 1 P125, P HOUSING ASSEMBLY W/PULSER, 240V UK 1 P127, P HOUSING ASSEMBLY W/PULSER, 230V SWISS 1 P141, P151, P HOUSING ASSEMBLY W/PULSER, 115V 1 P143, P153, P HOUSING ASSEMBLY W/PULSER, 230V DIN 1 P145, P155, P HOUSING ASSEMBLY W/PULSER, 240V UK 1 P147, P157, P HOUSING ASSEMBLY W/PULSER, 230V SWISS 1 3 P12, P DISK, P13, P DISK, P DISK, P SEAL 1 5 P121, P CE EPU 1 P123, P125, P CE EPU 1 P133, P135, P CE EPU 1 P141, P CE EPU 1 P143, P145, P CE EPU 1 P153, P155, P CE EPU 1 P CE EPU 1 P163, P165, P CE EPU 1 6 P O-RING 1 7 P WIRE TERMINAL (FEMALE) 1 8 P WIRE TERMINAL (FLAG) 4 9 P RETAINING RING 1 10 P WASHER, NYLON 1 11 P WASHER, RUBBER 1 12 P DUAL PIN 1 13 P PULSER 1 Key Model PART Description Qty. No. No. 14 P SPEED SHAFT 1 15 P O-RING 1 16 P WIRE TERMINAL (MALE) 1 17 P SPRING 1 18 P121, P131, P141, P151, P MOV/CAP ASM, 115V 1 P123, P133, P143, P153, P MOV/CAP ASM, 230V 1 P125, P135, P145, P155, P MOV/CAP ASM, 230V 1 P127, P137, P147, P157, P MOV/CAP ASM, 230V 1 19 P121, P131, P141, P151, P HOUSING ASSEMBLY, 115V 1 P123, P133, P143, P153, P HOUSING ASSEMBLY, 230V DIN 1 P125, P135, P145, P155, P HOUSING ASSEMBLY, 240V UK 1 P127, P137, P147, P157, P HOUSING ASSEMBLY, 230V SWISS 1 20 P GASKET 1 21 P SPEED KNOB 1 22 P STROKE DIAL 1 P STROKE DIAL 1 P STROKE DIAL 1 23 P STROKE KNOB 1 24 P SCREW 4
56 Series P1 Drive Assembly Exploded View Diagram Manufacturer s Manuals
57 Manufacturer s Manuals Series P1 Control Panel Detail 2003 LMI Milton Roy, USA - All Rights Reserved Printed in USA Specifications subject to change without notice. Series P1 Wiring Diagram
58 Manufacturer s Manuals Configuration Data Model P SI Manual Control Speed (stroking frequency) fixed and stroke length manually adjustable. P GPH(0.79 l/h) psi (10.3 Bar) P GPH (1.6 l/h) psi (7.6 Bar) P GPH (2.2 l/h) psi (17.3 Bar) P GPH (3.8 l/h) psi (7.6 Bar) P GPH (7.6 l/h) psi (3.5 Bar) P GPH (12.1 l/h) psi (1.5 Bar) Dual Manual Control Speed (stroking frequency) and stroke length manually adjustable. P GPH (0.79 l/h) psi (10.3 Bar) P GPH (1.6 l/h) psi (7.6 Bar) P GPH (2.2 l/h) psi (17.3 Bar) P GPH (3.8 l/h) psi (7.6 Bar) P GPH (7.6 l/h) psi (3.5 Bar) P GPH (12.1 l/h) psi (1.5 Bar) Instrument Responsive / Manual Manual adjustment features of Series P1 plus switch conversion to external control for automatic systems. P GPH (2.2 l/h) psi (17.3 Bar) P GPH (3.8 l/h) psi (7.6 Bar) P GPH (7.6 l/h) psi (3.5 Bar) P GPH (12.1 l/h) psi (1.5 Bar) Voltage Code VAC US Plug VAC US Plug VAC DIN Plug VAC, UK Plug VAC, Aust./NZ Plug VAC, Swiss Plug Liquid End See next page for complete liquid end specifications and selection. Data Sheet Series P Electronic Metering Pumps Dimensions Specifications Strokes Per Stroke Length Minute (Adjustable) Average (Adjustable) Recommended Input Power Shipping Series Min Max Max Speed Weight P02, P03 fixed % P04, P05 P06, P08 fixed % P12, P % 22 watts 7.6 lbs (3.5 kg) P14, P74 P15, P % P16, P76 P18, P " 215 mm 4.1" 104 mm * Dimensions shown are maximums for largest available Liquid End. Dimensions will vary depending on Liquid End selected. 3.50" *89 mm 9.0" *229 mm 0.105" 2.7 mm 4.07" 103 mm 4.45" 113 mm 6.30" 160 mm 3.66" 93 mm 8 Post Office Square Acton, MA USA TOLL FREE: (800) TEL: (978) FAX: (978) Replaces same of Rev. D 8/ E 8/01
59 Manufacturer s Manuals Configuration Data & Materials of Construction Drive Liquid Size Materials of Construction Tubing & Connections Assembly End No. Code Head & Fittings Balls Liquifram Check Valve Accessory Discharge Suction 458SI 0.5 PVC / PVC Ceramic Fluorofilm PVDF / Polyprel 4FV PE.250" O.D. P02-450SI 0.5 Acrylic / PVC Ceramic Fluorofilm PVDF / Polyprel 4FV PE.250" O.D. P04-450FI 0.5 Acrylic / PVC PTFE Fluorofilm PVDF / Polyprel 4FV PE.250" O.D. P12-358SI 0.5 PVC / PVC Ceramic Fluorofilm PVDF / Polyprel 4FV PE.250" O.D. P14-353SI 0.5 PVDF / PVDF Ceramic Fluorofilm PVDF / PTFE 4FV PE.250" O.D. P74-352SI 0.5 PVDF / PVDF Ceramic Fluorofilm PVDF / Polyprel 4FV PE.250" O.D. 156HV 0.5 Acrylic / PP 316 S.S. Fluorofilm Hypalon PE.5" O.D. Vinyl.938" O.D. 155HV 0.5 Polypropylene 316 S.S. Fluorofilm PTFE PE.5" O.D. Vinyl.938" O.D S.S. 316 S.S. Fluorofilm 316 S.S. Pipe 1/4" NPT M P03-498SI 0.9 PVC / PVC Ceramic Fluorofilm PVDF / Polyprel 4FV PE.375" O.D. P05-490SI 0.9 Acrylic / PVC Ceramic Fluorofilm PVDF / Polyprel 4FV PE.375" O.D. P13-490FI 0.9 Acrylic / PVC PTFE Fluorofilm PVDF / Polyprel 4FV PE.375" O.D. P15-398SI 0.9 PVC / PVC Ceramic Fluorofilm PVDF / Polyprel 4FV PE.375" O.D. P75-393SI 0.9 PVDF / PVDF Ceramic Fluorofilm PVDF / PTFE 4FV PE.375" O.D. 392SI 0.9 PVDF / PVDF Ceramic Fluorofilm PVDF / Polyprel 4FV PE.375" O.D. 86HV 0.9 Acrylic / PP 316 S.S. Fluorofilm Hypalon PE.5" O.D. Vinyl.938" O.D. 85HV 0.9 Polypropylene 316 S.S. Fluorofilm PTFE PE.5" O.D. Vinyl.938" O.D S.S. 316 S.S. Fluorofilm 316 S.S. Pipe 1/4" NPT M P06-468SI 1.8 PVC / PVC Ceramic Fluorofilm PVDF / Polyprel 4FV PE.375" O.D. P08-460SI 1.8 Acrylic / PVC Ceramic Fluorofilm PVDF / Polyprel 4FV PE.375" O.D. P16-460FI 1.8 Acrylic / PVC PTFE Fluorofilm PVDF / Polyprel 4FV PE.375" O.D. P18-368SI 1.8 PVC / PVC Ceramic Fluorofilm PVDF / Polyprel 4FV PE.375" O.D. P76-363SI 1.8 PVDF / PVDF Ceramic Fluorofilm PVDF / PTFE 4FV PE.375" O.D. P78-362SI 1.8 PVDF / PVDF Ceramic Fluorofilm PVDF / Polyprel 4FV PE.375" O.D. 76HV 1.8 Acrylic / PP 316 S.S. Fluorofilm Hypalon PE.5" O.D. Vinyl.938" O.D. 75HV 1.8 Polypropylene 316 S.S. Fluorofilm PTFE PE.5" O.D. Vinyl.938" O.D S.S. 316 S.S. Fluorofilm 316 S.S. Pipe 1/4" NPT M See front page for voltage code specifications. To specify ¼" NPT male, change 'I' to 'P'. To specify black, UV resistant tubing, change 'I' to 'U'. To specify Bleed 4FV, change 'S' to 'B'. To specify 3FV, change 'S' to't'. 3FV indicates that the pump is equipped with an LMI Three Function Valve (pressure relief, priming aid, line drain). 4FV indicates that the pump is equipped with an LMI Four Function Valve. This diaphragm type, anti-syhon/pressure relief valve isinstalled on the pump head. It provides anti-syphon protection and aids in priming, even under pressure. Fluorofilm is a copolymer of PTFE and PFA. Polyprel is an elastomeric PTFE copolymer. Polyprel is a registered trademark of Liquid Metronics Incorporated. Flourofilm and Liquifram are trademarks of Liquid Metronics Incorporated. Hypalon is a registered trademark of E. I. du Pont de Nemours & Co., Inc. Output Information Gallons per Hour Liters per Hour ml/cc per Minute ml/cc per Stroke Maximum Injection Series Min Max Min Max Min Max Min Max Pressure P psi (10.3 Bar) P psi (7.6 Bar) P psi (17.3 Bar) P psi (7.6 Bar) P psi (3.5 Bar) P psi (1.5 Bar) P psi (10.3 Bar) P psi (7.6 Bar) P14, P psi (17.3 Bar) P15, P psi (7.6 Bar) P16, P psi (7.6 Bar) P18, P psi (1.5 Bar) 2001 LMI Milton Roy - All Rights Reserved Printed in USA Specifications subject to change without notice.
60 Manufacturer s Manuals Information Sheet Liquid End Installation Instructions When pumping solutions, make certain that all tubing is securely attached to the fittings. It is recommended that tubing or pipe lines be shielded to prevent possible injury in case of rupture or accidental damage. Always wear protective clothing and face shield when working on or near your metering pump. Spare Parts: LMI recommends replacing the elastomeric components of the pump on an annual basis. RPM Pro Pacs and spare part kits are available from your local LMI Master Stocking Distributor. Tubing Connections Your LMI pump will be supplied with one of the following tubing connections. Before installation, all tubing must be cut with a clean square end. Valve and head connections from the factory may be capped. Remove and discard these caps before connecting the tubing. Ferrule Fitting DO NOT USE CLEAR VINYL TUBING ON THE DISCHARGE SIDE OF THE PUMP. (Pressure can rupture tubing.) DO NOT USE PLIERS OR PIPE WRENCH ON COUPLING NUTS OR FITTINGS. Tubing Connection 1/4" O.D. (.250" Tubing) 1" (25mm) 1/4" OD Tubing Tubing Connection 1/2" O.D. (.5" Tubing) Coupling Nut Bottom Tubing in Fitting. Push and hold tubing down while tightening coupling nut. 1/2" OD Tubing Coupling Nut Clamp Ring Tubing Connection 3/8" O.D. (.375" Tubing) Force tubing completely over entire nozzle into groove. Push and hold tubing down while tightening coupling nut. Pipe Thread Connection 1/4" or 1/2" NPT NPT Pipe Connection 3/8" OD Tubing Coupling Nut Clamp Ring Nozzle Fitting Groove Injection Check Valve and Discharge Tubing Installation The Injection Check Valve prevents backflow from a treated line. Connect the Injection Check Valve to your DISCHARGE (outlet) line. An NPTF fitting or pipe tee with a reducing bushing to 1/2" NPTF will accept the injection check valve. Use Teflon tape or pipe dope to seal the pipe threads only. When installing the Injection Check Valve, be sure to position it so that the valve enters the bottom of your pipe in a vertical position. Variations left and right within 40 are acceptable (see illustration below). After cutting an appropriate length of discharge tubing, connect tubing to the injection check valve then back to the discharge side of the pump head valve or discharge fitting. Make sure it does not crimp or come into contact with hot or sharp surfaces (see Tubing Connections). 1" Pipe Tee Reducing Bushing 1" to 1/2" NPT (Customer Supplied) Injection Check Valve Flow CORRECT Use Teflon tape here on pipe threads only CORRECT Do NOT use Teflon tape on machined threads Pipe Cross Section Flow Injection Check Valve 40 Variation Acceptable Force tubing completely over entire nozzle. Push and hold tubing down while tightening coupling nut. Apply Teflon tape. (Customer supplied pipe connections) Typical Injection Check Valve Installations 8 Post Office Square Acton, MA USA TEL: (978) FAX: (978) Rev. A 5/99
61 Manufacturer s Manuals Foot Valve/Suction Tubing Installation The Foot Valve acts as a check valve to keep the pump primed in suction lift applications. The foot valve is designed to be submersed in the solution tank or drum and must sit in a vertical position at the bottom. Position approximately 2 inches (50 mm) off the bottom if the tank or drum contains sediment. The ceramic weight, when installed, positions the foot valve in a vertical position. NOTE: Pump models equipped with high-viscosity liquid ends are not equipped with foot valves. Flooded suction is recommended. A 1/2" NPT connector is included for flooded suction installations. 1. Cut a piece of suction tubing to a length so that the foot valve hangs just above the bottom of the solution container. Maximum recommended vertical suction lift is 5 feet (1.5 m). 2. Attach the foot valve to one end of the suction tubing. 3. Slide the ceramic weight over the tubing end until it contacts the top of the foot valve coupling nut. 4. Place foot valve and tubing into the solution tank. Check that the foot valve is vertical and approximately 2 inches (50 mm) from the bottom of the tank or drum (see illustration). Connect the other end of the tubing to the suction side (bottom side) of the pump head (see Tubing Connections). Multi-Function Valve Installation To install the multi-function valve, remove the yellow screw cap on the top of the pump head and screw in the valve so that it contacts the seal ring. An additional 1/8 1/4 turn may be necessary to prevent leakage. 1/4" O.D. tubing connects to the side of the valve and acts as a return line to the solution tank. To ensure priming, this tubing must NOT be submerged in the solution. This return line tubing must be secured to ensure pumped solution will safely return to supply tank. Connector To Solution Tank or Drum 1/4" Tubing 1 Inch (25 mm) Bottom Tubing in Fitting Push and hold while tightening connector. Discharge To Pump Head LMI Pump LMI Pump Multi-Function Valve Tubing Connection Depressurizing Discharge Line Solution Tank Foot Valve Tilted Sideways WILL NOT PRIME INCORRECT 2.0 in. (50 mm) for Sediment Accumulation CORRECT Foot Valve Tubing Connection Use Ceramic Weight Foot Valve Must Remain Vertical (pumps equipped with multi-function valves) It is possible to depressurize the discharge line and pump head without removal of tubing or loosening of fittings. 1. Be sure injection check valve is properly installed and is operating. If a gate valve or globe valve has been installed downstream of injection check valve, it should be closed. Be certain relief tubing from the multi-function valve is connected and run to solution reservoir. 2. Turn Pressure Relief knob 1/4 turn. 3. Open the anti-syphon valve, if equipped 4. The discharge line is now depressurized. 5. If injection check valve is of higher elevation than pump head, disconnecting tubing at the injection check valve end will allow air to enter and cause solution to drain back to tank.
62 Manufacturer s Manuals 5.0 Liquid End Parts List 1 Flapper valve 2 Injection check valve body 3 Injection check valve spring 4 Check valve ball 5 Seal ring 6 Cartridge valve 7 Cartridge valve o-ring 8 Cartridge valve washer 9 Valve seat 10 Clamp ring 11 Ferrule 12 Clamp sleeve 13 Tubing adapter 14 Coupling nut 15 Discharge tubing 16 Valve housing 17 Multi-function valve 18 High-viscosity spring 19 Liquifram 20 Pump head 21 Pump head screw 22 Suction tubing 23 Foot valve seat 24 Foot valve screen 25 High-viscosity valve seat 26 H.V. tubing clamp 27 H.V. suction tubing 28 H.V. Tubing x 1/2 NPT connector 29 Injection check valve assembly 30 Discharge valve assembly 31 Suction valve assembly 32 Pump head assembly 33 Foot valve assembly 34 Injection Seat PTFE 35 Ceramic Weight 36 Return Line 37 Cap ASM (Black Knob) 38 Cap ASM (Yellow Knob) 39 Multi-Function Valve Body 40 Nut Multi-Function Valve 41 Screw Multi-Function Valve 42 Return Line Coupling Nut 43 Adjustment Screw B/4-FV 44 Cap B/4-FV 45 Plug B/4-FV 46 Gasket B/4-FV 47 Small O-Ring B/4-FV 48 Large O-Ring B/4-FV FV FV B/4FV /2'' NPT OR OR OR OR OR OR OR OR NOT INCLUDED IN OR 5 4 OR OR OR OR OR OR OR OR OR OR OR NOTE: This illustration is a visual representation of all LE components. Liquid ends will not include all parts shown.
63 Manufacturer s Manuals Start-Up/Priming for Pump Supplied with Multi-Function Valve Read this entire section completely before proceeding. When all precautionary steps have been taken, the pump is mounted, and the tubing is securely attached, you may now start priming the pump. 1. Plug in or switch the pump on. 2. While the pump is running, set the speed knob at 80% and the stroke knob at 100%. If the pump is equipped with pressure control, turn fully clockwise. 3. 1/4 turn open the relief side (black knob) of the multifunction valve. 3A. (Bleed 4FV only) With screwdriver rotate bleed adjustment screw counter-clockwise 2 full turns. When solution begins to flow through translucent bleed return tubing, the pump is primed. Stop pump. 4. The suction tubing should begin to fill with solution from the tank. 5. A small amount of solution will begin to discharge out the return line of the multi-function valve. Once this happens, 1/4 turn or release the knob and SHUT THE PUMP OFF. (If pump is not equipped with an on/off switch, disconnect the power cord.) 6. The pump is now primed. 6A. (Bleed 4FV only) a. Start pump and let pump inject solution into the discharge line. b. Close the bleed adjustment screw by rotating it clockwise with a screwdriver. c. Now adjust the pump stroke length and/or speed (frequency) to a range approximately 25% higher than you would normally want for the process. d. Slowly rotate bleed adjustment screw counterclockwise until just a small amount of solution begins to trickle down inside the bleed return tubing. A small amount of solution pumped back to the tank with each stroke of the pump will allow gas and air to escape without air or gas locking in the pump head. NOTE: If the pump does not self-prime, remove the multifunction valve on the discharge side of the pump head. Remove the check valve and pour water or solution into the port until the head is filled. Replace valve, then follow start-up/priming steps. Anti-Syphon Knob Start-Up/Priming without Multi- Function Valve Read this entire section completely before proceeding. When all precautionary steps have been taken, the pump is mounted, and the tubing is securely attached, you may now prime the pump. 1. Plug in or switch on the pump. 2. While the pump is running, set the speed knob at 80% and the stroke knob at 100%. If the pump is equipped with pressure control, turn fully clockwise. 3. The suction tubing should begin to fill with solution from the tank. 4. Once the solution begins to exit the pump head on the discharge side, SHUT THE PUMP OFF. (If pump is not equipped with an on/off switch, disconnect the power cord). 5. The pump is now primed. Rotate clockwise to close Bleed Adjustment Screw Space open to drain Bleed 4-FV Negative pressure at discharge closes diaphragm and prevents syphoning. Pump discharge stroke lifts diaphragm to allow for fluid flow. Bleed (Relief) Port NOTE: If the pump does not self-prime, remove the fitting on the discharge side of the pump head. Remove the ball and pour water or solution into the port until the head is filled. Replace valve, then follow start-up/priming steps LMI Milton Roy - All Rights Reserved Printed in USA Specifications subject to change without notice.
64 Manufacturer s Manuals GENERAL Chemical metering pumps shall be positive displacement, Liquifram type pumps that are UL and CUL approved. Output volume shall be adjustable while pumps are in operation from zero to maximum capacity of: P02, P GPH (0.79 liters per hour) P03, P GPH (1.6 liters per hour) P04, P14, P GPH (2.2 liters per hour) P05, P15, P GPH (3.8 liters per hour) P06, P16, P GPH (7.6 liters per hour) P08, P18, P GPH (12.1 liters per hour) Chemical metering pumps shall be capable, without a hydraulically backed diaphragm, of injecting solutions against pressures up to: P02, P12-150psig (10.3 Bar) P03, P13-110psig (7.6 Bar) P04, P14, P74-250psig (17.3 Bar) P05, P15, P75-110psig (7.6 Bar) P06, P16, P76-50psig (3.5 Bar) P08, P18, P78-22psig (1.5 Bar) SERIES P0 The stroke frequency will be fixed at the following rates. P02, P03-60 strokes per minute P04, P05, P06, P strokes per minute Stroke length shall be adjustable by readily accessible dial knob located opposite the liquid handling end. SERIES P1 Adjustment shall be by means of readily accessible dial knobs, one for changing stroke length and the other for changing stroke frequency. Both knobs are to be located opposite the liquid handling end. Specification Sheet Series P SERIES P7 Control of Series P7 metering pumps shall be selectable between internal and external pulsing by means of a switch integral with frequency control knob. "External" position to be below lowest internally paced frequency setting. Stroke length shall be adjustable by means of readily accessible dial knob. When in external pulsed mode, Series P7 units shall accept signals without the use of electrical timer or internal timer. DRIVE The pump drive shall be totally enclosed with no exposed moving parts. Solid state electronic pulser shall be fully encapsulated and supplied with quick connect terminals at least 3/16" (4.75 mm) wide. Electronics shall be housed in chemical resistant enclosure at the rear of the pump for maximum protection against chemical spillage. Electrical power consumption shall not exceed 22 watts per hour under full speed and maximum pressure conditions. Pump weight shall not exceed 14 lbs (6.5 kg). AUTOMATIC PRESSURE RELIEF To eliminate need for pressure relief valve, Liquifram TM shall automatically stop pulsating when discharge pressure exceeds pump pressure rating by not more than 35%. MATERIAL Chemical metering pump housing shall be of chemically resistant glass fiber reinforced thermoplastic. All exposed fasteners shall be stainless steel. Chemical metering pump valves shall be ball type, with ceramic balls 1. Valve seat and seal ring shall be renewable by replacing the combination seat-seal ring 2 or cartridge valve assembly. Pump head shall be of transparent acrylic 3 material capable of resisting the pumped chemical. Fittings and connections at pump head shall be PVC 4. CHECK VALVES AND TUBING A total of 16 ft (4.8 m) of polyethylene tubing 5 shall be provided per pump complete with compression connections. A foot valve with integral one piece strainer shall be provided for the suction line, and an injection check/back pressure valve with ½" NPT male connection for the injection point. The injection check valve shall incorporate a dilating orifice which prohibits scale formation and accumulation of crystalline deposits. Notes: 1. Type 316 stainless steel or PTFE may be specified. 2. Hypalon, PTFE or Polyprel may be specified. 3. PVDF, PVC, Polypropylene, or Type 316 stainless steel may be specified.4. PVDF, Polypropylene, or Type 316 stainless steel may be specified ft. (1.8 m) of vinyl suction tubing may be specified in place of polyethylene for the suction side only. 1/4" pipe thread may be specified. 8 Post Office Square Acton, MA USA TEL: (978) FAX: (978) Replaces same of Rev. B 8/ C 6/01
65 Manufacturer s Manuals 2001 Liquid Metronics Incorporated - All Rights Reserved Printed in USA Specifications subject to change without notice. Polyprel is a registered trademark of Liquid Metronics Incorporated. Liquifram, Liquitron, Micropace are trademarks of Liquid Metronics Incorporated. Hypalon and Viton are registered trademarks of E. I. du Pont de Nemours & Co., Inc.
66 Manufacturer s Manuals
67 MICtreat TM FPS Chemical Info/Registrations MICtreat TM FPS Chemical Industrial Antimicrobial for Fire Protection Systems *Rapid kill of microbes, including MIC-related bacteria & SRB *Reduces free oxygen in FPS, controlling all corrosion *Non-flammable *Compatible with FPS system components & waters *Economical *Unusually good (low) toxicity profile *Good environmental profile/readily biodegradable *Easy to apply & test *Simple & inexpensive to deactivate *Registered with the Environmental Protection Agency & States Sold exclusivey by: BTIproductsLP
68 MICtreat TM FPS Chemical Info/Registrations MICtreat TM FPS Chemical Industrial Antimicrobial for Fire Protection Systems BTI Products, LP has been providing industrial systems with testing, consulting, and treatment solutions to microbiologically influenced corrosion (MIC) and other microbial and corrosion problems for over 25 years. The company approaches diagnosis and treatment of such problems based on scientific analysis of data from actual facilities. To help their clients properly diagnose and treat microbial and corrosion problems, BTI Products provides on-site test kits, patented treatment devices, and treatment chemicals for a variety of industries. In order to better serve the specific needs of the fire protection industry, BTI Products has developed a line of testing and treatment products specifically for FPS: MICkit FPS and MIPkit TM FPS test kits, MICtreat TM FPS Chemical, and the patented MICtreat TM FPS System chemical delivery devices. MICtreat TM FPS Chemical is a new generation antimicrobial agent offering outstanding performance and a good environmental profile. Rapid control of MIC-related bacteria: -Aerobic & anaerobic bacteria (AERO & ANA) -Slime-formers (LNB) & iron-depositing bacteria (IRB) -Acid-producing bacteria (APB) -Sulfate-reducing bacteria (SRB) Reduces free oxygen System compatibility; Non-flammable Cost of treatment is $ per 2,500 gallons of water Compatible with many water treatment chemicals (e.g., surfactants) Biodegradable; Simple & inexpensive to deactivate in waters removed from FPS Good environmental profile Unusually good (low) toxicity profile Simple & quantitative field test for active ingredient Simple test for treatment effectiveness using MICkit technology Registered with EPA for use in FPS Manufactured to ISO 9002 Controls bacteria, fouling, corrosion, & hydrogen sulfide production Reduces an important nutrient for MIC-related bacteria & a reactant in many forms of severe corrosion, rusting, & severe pitting Safe to use in FPS Economical treatment; only new water needs additional treatment Treatment can be done right after cleaning Reduced risk to environment & added safety to maintenance personnel Low environmental impact Ease of handling & added safety to maintenance personnel Easy, on-site monitoring & control of treatment levels Accurate & easy on-site monitoring of treatment success EPA-registered specifically for FPS Quality assurance Toll Free: products@bti-labs.com Website:
69 MICtreatTM FPS Chemical Industrial Antimicrobial for Fire Protection Systems MECHANISM TEST RESULTS.&$)"/*4. THPS, the active ingredient in MICtreatTM FPS Chemical, disrupts bacterial cells and inhibits enzymes and energy production in microbes, including those causing MIC. THPS also reacts with oxygen (forming a non-toxic oxide), thus reducing an important nutrient for MIC-related bacteria and a very important reactant in many severe corrosion reactions. Results of treating steel FPS pipes with MICtreatTM FPS Chemical are shown in the following pictures and table. MICtreatTM FPS Chemical Info/Registrations Untreated Steel FPS Pipe New Steel FPS Pipe Treated with MICtreatTM FPS Chemical.*$53&"5UN '14 $)&.*$"- $IBMMFOHF 5FTU 3FTVMUT SAMPLE PIPE CLEANED TREATMENT BEFORE TEST None COLOR OF DISSOLVED RESIDUAL RESIDUAL WATER/PIPE OXYGEN CHLORINE THPS BACTERIA BACTERIA BACTERIA BACTERIA (PPM) (PPM) (PPM) (PER ML) (PER ML) (PER ML) (PER ML) >100,000 >100,000 >100,000 >100,000 Clear LOW NUTRIENT IRON-RELATED ACID-PRODUCING SULFATE-REDUCING TEST WATER N/A PIPE 1 No None Black/Black >100, >100,000 10,000 PIPE 2 No 50 ppm Chlorine Black/Black >100,000 0 >100, PIPE 3 Yes 50 ppm Chlorine Black/Black >100,000 0 PIPE 4 No 450 ppm MICtreat Clear/Clean PIPE 5 Yes 50 ppm MICtreat Clear/Clean These results demonstate that: 1. MICtreatTM FPS Chemical kills all types of MIC-related bacteria, even in some uncleaned pipes. 2. MICtreatTM FPS Chemical can be used at low concentrations to kill all types of MIC-related bacteria in cleaned pipes. 3. MICtreatTM FPS Chemical consumes oxygen, thus preventing microbial growth, tubercule formation, MIC, and other forms of corrosion. 4. MICtreatTM FPS Chemical residuals are maintained in FPS pipes, thus maintaining control over long periods of time. 5. MICtreatTM FPS Chemical aids in removing oil and dirt from FPS pipes allowing access of biocide to microbes. Toll Free: products@bti-labs.com Website:
70 MICtreat TM FPS Chemical Info/Registrations MICtreat TM FPS Chemical Industrial Antimicrobial for Fire Protection Systems MICtreat TM FPS Chemical contains the active ingredient tetrakis(hydroxymethyl) phosphonium sulfate (THPS), shown in the chemical formula below: CH 2 OH MICtreat TM FPS Chemical kills MICrelated bacteria & consumes oxygen, thus preventing microbial growth, tuberule formation, MIC, & other forms of corrosion. HOH 2 C P + CH 2 OH SO 4 2- CH 2 OH 2 Chemical Concentrate Use Concentrations Active Content 750,000 ppm 100 to 400 ppm Appearance Clear, colorless Clear to pale yellow Specific gravity (25 0 C) Typically 1.38 Typically 1.0 ph Typically 4.0 Depends on ph of water to which it s added; Has minor effect on ph MICtreat TM FPS Chemical starts out clear and ends up clear. Shown here: water from FPS pipe treated with MICtreat TM FPS Chemical. Flash point Solubility Stability Compatibility Non-flammable Miscible in all proportions with fresh water & sea water Long-term stability under normal handling & storage conditions Stable in FPS long enough to kill MIC-related microbes & reduce oxygen levels Compatible with many water treatment chemicals. Avoid: strong bases, strong acids, strong oxidizing agents, strong reducing agents Toll Free: products@bti-labs.com Website:
71 MICtreat TM FPS Chemical Info/Registrations MICtreat TM FPS Chemical Industrial Antimicrobial for Fire Protection Systems MICtreat TM FPS Chemical is compatible with all types systems, waters, & other treatment chemicals MICtreat TM FPS Chemical is compatible with FPS system components and is non-flammable, making it an excellent choice for use in FPS. MICtreat TM FPS Chemical is miscible in all proportions with every type of water and is compatible with a wide range of water treatment additives, allowing for greater treatment flexibility. A key advantage of treating with MICtreat TM FPS Chemical is its good environmental profile. It is readily biodegradable and can be simply and inexpensively deactivated under controlled conditions, resulting in reduced risk to the environment. Detailed information can be found in the Material Safety Data Sheet, available upon request. MICtreat TM FPS Chemical offers outstanding performance & good environmental & toxicity profiles The active ingredient in MICtreat TM FPS Chemical has been extensively tested and has a very good (low) toxicity profile. The added benefit of easy and inexpensive deactivation results in safer handling for maintenance personnel. For more detailed information, please refer to the Material Safety Data Sheet, available upon request. MICtreat TM FPS Chemical has a good environmental profile, which makes it an excellent treatment choice. Toll Free: products@bti-labs.com Website:
72 MICtreat TM FPS Chemical Info/Registrations Initial treatment of all parts of FPS should be done using water containing 400 parts per million (ppm) MICtreat TM FPS Chemical to ensure adequate initial treatment. All waters subsequently added to the FPS should also be treated with MICtreat TM FPS Chemical. BTI Products patented MICtreat TM FPS System automatically delivers the proper amount of MICtreat TM FPS Chemical to all waters entering the FPS. Cost of treatment with MICtreat TM FPS Chemical is $ per 2,500 gallons of water.* Concentrations of the active ingredient in MICtreat TM FPS Chemical is accurately and simply determined by site personnel in minutes using a self-contained field test kit. Because MICtreat TM FPS Chemical is a registered biocide, you should check with authorities having jurisdiction about the use and disposal of chemicals in your facility before ordering or using MICtreat TM FPS Chemical. Please contact BTI Products if you require assistance. The following support documents are available upon request: Material Safety Data Sheets (MSDS) Technical publications *Price subject to change without notice. Every effort has been made to ensure that the information provided herein is reliable, however, BTI Products cannot assume liability for any loss, injury, or damage, consequential or otherwise, which may result from its use. Toll Free: products@bti-labs.com Website: Rev 12/07/07
73 MICtreat TM FPS Chemical Info/Registrations
74 MICtreat TM FPS Chemical Info/Registrations
75 MICtreat TM FPS Chemical Info/Registrations
76 MICtreat TM FPS Chemical Info/Registrations
77 MICtreat TM FPS Chemical Info/Registrations
78 MICtreat TM FPS Chemical Info/Registrations
79 MICtreat TM FPS Monitor Kit Instructions MICtreat TM FPS MONITOR For Monitoring Treatment of Fire Protection Systems Using MICtreat TM FPS Chemical BTIproductsLP INSTRUCTIONS Read all instructions before sample collection Before collecting samples, review contents of this kit (page 3). This kit will allow you to test twenty liquid samples. If you require technical assistance, please contact BTI Products at (800) Collect samples from remote inspector s test valves or remote branch lines. Test 7, 30, and 60 days after initial treatment. Test thereafter at every inspectors test and flow test (minimum of every 6 months). Let water flow long enough to ensure water from FPS, not water/debris in drainpipe, is being collected. Depending upon FPS setup, you may need to flow water for several minutes before collecting water to test. Do not contaminate samples by touching with hands or tools. Process all samples immediately after collection. For technical assistance, to request MSDS, or to place an order: Call Toll Free: Or products@bti-labs.com A. DISSOLVED OXYGEN TEST 1. Place stopper with hose, small end first, up into pipe from which you will collect water. Stopper should seat so that half the stopper is inside pipe and other half is exposed. 2. Run water through stopper hose until water coming out of hose is a stream without air bubbles. 3. Place end of stopper hose into sampling tube such that hose touches bottom of tube. 4. Flow water through hose and into bottom of sampling tube until tube is overflowing and there are no air bubbles present in water in tube. 5. While still flowing water into sampling tube, place dissolved oxygen ampoule, tapered end first, into sampling tube until ampoule is 3 / 4 submerged in water. 6. Gently press end of ampoule against side of sampling tube to snap tip. Let ampoule fill with water. 7. Remove ampoule from sampling tube and invert several times, allowing bubble to travel from end to end to mix contents. Avoid injury from touching broken tip! 8. Wait 2 minutes for color development. 9. Compare ampoule color with attached Dissolved Oxygen Color Chart (page 4). Chart should be illuminated from above by a bright, white light. Be sure to place ampoule on both sides of color chart before concluding it gives the best match. Placing ampoule between and parallel to color bars aids in readings. 10. Record dissolved oxygen level in attached Data Sheet (page 6). 11. Discard sample, and remove stopper with hose from pipe. 1
80 MICtreat TM FPS Monitor Kit Instructions B. RESIDUAL MICtreat TM FPS CHEMICAL TEST 1. Rinse sampling tube with fresh water sample. 2. Fill to top. 3. Follow THPS test kit (in blue plastic box) instructions. 4. Record THPS levels in attached Data Sheet (page 6). 5. Discard sample. C. TOTAL IRON TEST 1. Rinse sampling tube with fresh water sample. 2. Fill to top. 3. Follow instructions on total iron (Fe 2+ /Fe 3+ ) bottle label. 4. Record total iron levels in attached Data Sheet (page 6). 5. Discard sample. D. MICROBIOLOGICAL TEST 1. Using a marking pen, label media tray and media bottles as shown in Figure 1. Figure 1 2. Spray or wipe inside and outside of pipe from which water will be collected with 70% isopropyl alcohol (available from grocery or drugstore). Wait one minute. 3. Rinse sampling tube once with 70% isopropyl alcohol. 4. Rinse sampling tube several times with fresh water sample. 5. Fill sampling tube with water sample. 6. Remove and discard wrappers from a sterile 1 ml syringe and an 18g needle. Without touching tip of syringe or opening of needle, place needle onto syringe. Tighten needle onto syringe by pushing in and turning needle shield clockwise. 7. Remove needle shield. Place syringe/needle into water in sampling tube. 8. Withdraw 1.0 ml of water by gently pulling up on syringe plunger until water reaches 1.0 ml mark. 9. Flip plastic cap off first purple-capped bottle (labeled #1). If you touch rubber stopper, wipe stopper with alcohol wipe. 10. Insert syringe needle through rubber stopper of first bottle. Inject water into bottle by depressing plunger. 11. Keep needle in bottle. Mix solution by drawing up 1.0 ml of media-sample mixture and then expelling it back into bottle. Repeat several times. 12. Using the same syringe and needle, now withdraw 0.1 ml (one-tenth!) of solution from purple bottle #1 and inject into purple bottle #2. 13.Repeat steps 2 through 12 for water samples #2 through 20 using a new syringe and needle for each water sample. 14. Store all bottles of media in kit box at room temperature. 15. After 2, 5, and 15 days, compare your test bottles to Positive Reactions Sheet (page 5) to determine number of bottles, if any, that are positive for bacteria. 16.Record LNB results in attached Data Sheet (page 6). Microbiological test results: A positive reaction in bottle #1 indicates low numbers of low nutrient bacteria (1 to 10 per ml of water sample). Positive reactions in bottles #1 and #2 indicate high levels of low nutrient bacteria (greater than or equal to 100 per ml of water sample). 2
81 MICtreat MICtreat TM TM FPS FPS Monitor Monitor Kit Kit Instructions Instructions E. ACCEPTABLE TEST RESULTS Following are acceptable results for FPS properly chemically cleaned and then treated using MICtreat TM FPS Chemical: Dissolved oxygen levels should be less than 1 ppm. Note: Dissolved oxygen levels may not drop immediately. MICtreat TM FPS Chemical residuals, measured as THPS, should be at least 100 ppm. Total iron levels should be less than 2 ppm. Bacteria levels should be zero. If any water sample has results outside the acceptable ranges, repeat the test. If the second test shows results outside the acceptable range, take appropriate action to remedy the problem. Contact BTI Products at (800) for assistance. Usage & Storage: Test kit must be used by expiration date printed on kit box label. Store all test materials in a cool, dry, dark place. Do not eat or drink any of the contents of this kit. Keep out of reach of children. Material Safety Data Sheets available upon request. Disposal of Test Materials: Properly dispose of all syringes, needles, ampoules, media bottles, and test materials. Needles must be destroyed before disposal by cutting or bending back the needle.syringes must be destroyed by breaking or shattering the barrel. Federal and local laws apply. Used media bottles must be properly disposed of according to local regulations. Alternatively, bottles may be returned to BTI Products, LP for proper disposal for a fee of $15.00 per kit. List of Kit Contents: Bottles BTI-LNB Medium (Purple Flip-Off Cap) 2. 1, 50 ml Sampling Tube 3. 1 Stopper with Plastic Hose 4. 20, 1 ml Syringes 5. 20, 18g Needles Alcohol Prep Pads Dissolved Oxygen Ampoules 8. 1 THPS Test Kit (Blue Box) 9. 1 Bottle Total Iron Test Strips WARRANTY BTI Products, LP s products are warranted by BTI Products, LP to perform as described in the technical literature supplied with each product, provided the products are used, stored, and maintained in accordance with the directions provided. They must also be used before the expiration date. Adequate quality control must be done by the user of the products. BTI Products, LP disclaims any implied warranty of merchantability or fitness of its products for any other purpose than described in its technical literature, and in no event shall BTI Products, LP be held liable for any consequential damages arising out of the aforesaid express warranty. Should you have questions about this product or any of the products and services we provide, please call or write: BTI Products, LP 652 Silver Hills Road Bayfield, CO products@bti-labs.com We welcome all comments and inquiries. Rev 07/18/07 Need Help? Call
82 MICtreat TM FPS Monitor Kit Instructions Dissolved Oxygen Color Chart 4
83 MICtreat TM FPS Monitor Kit Instructions MICtreat TM FPS Monitor Kit - Positive Reactions Sheet BTI-LNB Uninoculated (Negative) 2. Positive Cloudy 3. Positive Cloudy with slime BTIproductsLP 5
84 MICtreatTM FPS Monitor Kit Instructions 6
85 MICtreatTM FPS Monitor Kit Instructions 7
DataSheet. Series AA Electronic Metering Pumps. Configuration Data. Dimensions. Specifications. Model AA HI. ISO 9001 Certified
DataSheet Series AA Electronic Metering Pumps Configuration Data Model AA 9 4 1-358HI Control Code 1 --- Manual Control: Speed (stroking frequency) and stroke length manually adjustable. 7 --- Instrument
More informationLRS(H)4 Pressure-Reducing Regulator User Manual
LRS(H)4 Pressure-Reducing Regulator User Manual Read the complete manual before installing and using the regulator. 2 Safe Product Selection When selecting a product, the total system design must be considered
More informationInstallation, Operation and Maintenance Manual for Back Pressure Regulator
Installation, Operation and Maintenance Manual for Back Pressure Regulator Model 8860 2009 Groth Corporation IOM-8860 Rev. B 12541 Ref. ID: 95565 Page 2 of 13 Table of Contents I. INTRODUCTION 3 II. DESIGN
More informationHAYWARD FLOW CONTROL Series PBV Back Pressure Valve and Series RPV Pressure Relief Valve INSTALLATION, OPERATION, AND MAINTENANCE INSTRUCTIONS
HAYWARD FLOW CONTROL Series PBV Back Pressure Valve and Series RPV Pressure Relief Valve INSTALLATION, OPERATION, AND MAINTENANCE INSTRUCTIONS Page 1 of 20 Page 2 of 20 TABLE OF CONTENTS Safety Warnings
More informationTOP VALVE. Pat. #5,857,486 & 5,944,050. High Temperature: max. 300 F (149 C) Back Pressure And Pressure Relief Valves. Instruction Manual
TOP VALVE Pat. #5,857,486 & 5,944,050 High Temperature: max. 300 F (149 C) Back Pressure And Pressure Relief Valves Instruction Manual PLEASE NOTE: This instruction manual provides information and instructions
More informationTOP VALVE. Pat. #5,857,486 & 5,944,050. Mid-Range Pressure PSIG Back Pressure and Pressure Relief Valves. Instruction Manual
TOP VALVE Pat. #5,857,486 & 5,944,050 Mid-Range Pressure 50 232 PSIG Back Pressure and Pressure Relief Valves Instruction Manual Please Note: This instruction manual provides detailed information and instructions
More informationRD(H)20/25 Pressure-Reducing Regulator User Manual
RD(H)20/25 Pressure-Reducing Regulator User Manual Read the complete manual before installing and using the regulator. 2 Safe Product Selection When selecting a product, the total system design must be
More informationPressure Dump Valve Service Kit for Series 2300 Units
Instruction Sheet Pressure Dump Valve Service Kit for Series 00 Units. Overview The Nordson pressure dump valve is used to relieve hydraulic pressure instantly in Series 00 applicator tanks when the unit
More informationFreedom8 ShoeBox Compressor Manual
Freedom8 ShoeBox Compressor Manual Warning!! This product is not a toy! Use or misuse can cause severe injury or death! Use only with adult supervision. This unit is only to be used with tanks, hoses and
More informationModel VR6 System. Installation, Operation & Maintenance
Model VR6 System Installation, Operation & Maintenance General: All Archer Instruments chlorination systems are carefully designed and tested for years of safe, accurate field service. All Archer Instruments
More informationPressure Dump Valve Service Kit for Series 3000 Units
Instruction Sheet Pressure Dump Valve Service Kit for Series 000 Units. Overview The Nordson pressure dump valve is used to relieve hydraulic pressure instantly in Series 00, 400, 500, and 700 applicator
More informationLRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator.
LRS(H)4 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to the
More informationTECHNICAL DATA. Q = C v P S
Page 1 of 13 1. DESCRIPTION The Viking 6 Model G-6000 Dry Valve Riser Assembly consists of a small profile, light weight, pilot operated valve that is used to separate the water supply from the dry sprinkler
More informationModel 130M Pneumatic Controller
Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate
More informationINDUSTRIAL VALVES MODELS: C62-A; C62-D. INSTRUCTION MANUAL Installation Operation Parts Service DIAPHRAGM BYPASS PRESSURE REGULATING VALVES
INSTRUCTION MANUAL Installation Operation Parts Service IMPORTANT Record your Regulator model number and serial number here for easy reference: Model No. Serial No. Date of Purchase When ordering parts
More informationBlue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION
Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION Install your Blue River Technologies Port-A-Poly mixing system in a clean dry area.
More informationFloat Operated Level Controllers
CONTENTS Float Operated Level Controllers IM0015 Nov. 2014 PAGE Introduction 1 Scope 1 Description 1 Specification 1 Control Installation 2 INTRODUCTION Side Mount Back Mount Prior to installing, the instructions
More information200 PSI FAST-FILL AIR SOURCE KIT
200 PSI FAST-FILL AIR SOURCE KIT 55% Duty Compressor on 2.0 Gallon Air Tank PART NO. 20007 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of
More informationTITAN FLOW CONTROL, INC.
PREFACE: This manual contains information concerning the installation, operation, and maintenance of Titan Flow Control (Titan FCI) Simplex Basket Strainers. To ensure efficient and safe operation of Titan
More informationModel PSI Compressor with 3-Gallon Air Tank 12VDC
Model 6350 150 PSI Compressor with 3-Gallon Air Tank 12VDC IMPORTANT: It is essential that you and any other operator of this product read and understandd the contents of this manual before installing
More informationPRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.
PRS(TC)4,8 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to
More information100C Air Compressor Kit
10010 100C Air Compressor (standard mounting bracket, CE Spec) 10014 100C Air Compressor (no leader hose or check valve, CE Spec) 10016 100C Air Compressor (with Omega Bracket, CE Spec) IMPORTANT: It is
More informationSPECIFICATIONS Type: Twin stack, single phase Tank: 4 gallon Air Output: PSI; PSI Max PSI: 125 PSI HP: 1.
2 GALLON TWIN STACK AIR COMPRESSOR Model: 9526 DO NOT RETURN TO STORE. Please CALL 800-348-5004 for parts and service. CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or
More information200 PSI HIGH-FLOW AIR SOURCE KIT
200 PSI HIGH-FLOW AIR SOURCE KIT 50% Duty Compressor on 2.0 Gallon Air Tank PART NO. 20008 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of
More informationINSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER RELIEF VALVE
INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER RELIEF VALVE MODELS RV-1 RV-2 RV-2CP RV-3 DRAWING NUMBERS AD017A[ ] AD018A[ ] AD020A[ ] AD282BO MANUAL NUMBER IOM-033 REVISION Rev. E, 3/28/2016 TABLE
More informationCompact Triple Cabinet Outlet Station Model Installation and Operating Instructions
Compact Triple Cabinet Outlet Station Model 6258-1 Installation and Operating Instructions The Porter Compact Triple Outlet Station (6258-1) provides a quick, safe, and reliable method of connection to
More informationPRO SINK -R 1 and 2. Dilution system CONTENTS
PRO SINK -R 1 and 2 Dilution system CONTENTS 1.0 Product description 2 2.0 Warnings.. 3 3.0 Installation procedure 4 4.0 Plumbing connections 4 5.0 Technical features.. 5 6.0 Maintenance... 6 7.0 Troubleshooting......6
More informationInstruction Manual - Diaframless TM Ejector Chlorine, Sulfur Dioxide and Ammonia
Instruction Manual - Diaframless TM Ejector Chlorine, Sulfur Dioxide and Ammonia - 1-122.6010.6 These instructions describe the installation, operation and maintenance of the subject equipment. Failure
More information444C DUAL PERFORMANCE VALUE PACK
(Chrome) PART NO. 44432 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of this manual before installing and using this product. SAVE THIS MANUAL
More information200 PSI COMPRESSORS - MODEL NUMBERS
200 PSI COMPRESSORS - MODEL NUMBERS 380C AIR COMPRESSOR KIT PART NO. 38033 480C AIR COMPRESSOR KIT PART NO. 48043 380C 480C IMPORTANT: It is essential that you and any other operator of this product read
More informationWelker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual
Installation, Operation, and Maintenance Manual Welker Sampler Model GSS-1 The information in this manual has been carefully checked for accuracy and is intended to be used as a guide to operations. Correct
More information400C & 450C DUAL PERFORMANCE VALUE PACKS
(Chrome) PART NO. 40013 (Silver) PART NO. 45012 (Chrome) PART NO. 45013 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of this manual before
More information97C COMPRESSOR KIT 12V PART NO C COMPRESSOR KIT 24V PART NO C COMPRESSOR KIT PART NO
97C COMPRESSOR KIT 12V PART NO. 00097 97C COMPRESSOR KIT 24V PART NO. 02497 98C COMPRESSOR KIT PART NO. 00098 97C 98C IMPORTANT: It is essential that you and any other operator of this product read and
More informationMODEL NUMBER: PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank
IMPORTANT SAFETY INSTRUCTIONS CAUTION - To reduce risk of electrical shock or Electrocution: MODEL NUMBER: 20008 200 PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank IMPORTANT: It is
More informationRS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.
RS(H)10,15 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to
More informationOPERATION MANUAL NTF-60 Plus
OPERATION MANUAL NTF-60 Plus Nitrogen Tire Filling Valve Stem Caps (Qty=200) Order P/N 436075 RTI Technologies, Inc 10 Innovation Drive York, PA 17402 800-468-2321 www.rtitech.com 035-81264-00 (Rev A)
More informationMODEL NUMBER: M20005 AIR SOURCE KIT. 30% Duty Compressor on. 2.0 Gallon Air Tank SAVE THIS MANUAL FOR FUTURE REFERENCE
MODEL NUMBER: M20005 AIR SOURCE KIT 30% Duty Compressor on 2.0 Gallon Air Tank SAVE THIS MANUAL FOR FUTURE REFERENCE USER MANUAL IMPORTANT SAFETY INSTRUCTIONS CAUTION - To reduce risk of electrical shock
More informationROYTRONIC EXCEL TM Series AD Electronic Metering Pumps
0.20 5.1 ROYTRONIC EXCEL TM AD Electronic Metering Pumps LMI ROYTRONIC EXCEL AD pumps achieve a new level of innovation in solenoid-driven metering pump technology. Offering a robust design, micropressor
More informationBooster Pump PB4-60 Replacement Kits
Booster Pump PB4-60 Replacement Kits FOR YOUR SAFETY - This product must be installed and serviced by a contractor who is licensed and qualified in pool equipment by the jurisdiction in which the product
More informationTECHNICAL DATA. Q= Cv S
Page 1 of 13 1. DESCRIPTION The Viking 4 inch Model G-4000 Dry Valve Riser Assembly consists of a small profile, light weight, pilot operated valve that is used to separate the water supply from the dry
More informationBS Series Basket Strainer
BS Series Basket Strainer Operating, Installation, & Maintenance Manual Corrosion Resistant Fluid and Air Handling Systems. Dated 04-26-12 PRESSURE DROP SIMTECH strainers are engineered to offer the lowest
More information! Warning, refer to accompanying documents.
About this Manual To the best of our knowledge and at the time written, the information contained in this document is technically correct and the procedures accurate and adequate to operate this instrument
More informationOPERATING AND MAINTENANCE MANUAL
Series 4300 Engineered Performance TABLE OF CONTENTS 0 INTRODUCTION 1 1 Scope 1 2 Description 1 3 Specifications 1 0 INSTALLATION 1 1 Mounting 1 2 Piping 1 1 Connecting Process Pressure 2 2 Vent Connections
More informationInterface Devices, Inc. Hydraulic Mini Mule
Interface Devices, Inc. Hydraulic Mini Mule INSTALLATION, OPERATION & MAINTENANCE MANUAL IMPORTANT! FILE THIS MANUAL IN A SAFE PLACE FOR FUTURE SERVICE & PARTS NEEDS ALWAYS REFERENCE THE SERIAL NUMBER
More informationInstallation of Your SprayMaster System
Installation of Your SprayMaster System 1. At the installation site, remove all equipment from the corrugated box and the polyethylene drum and replace the drum lid. Check the picture to identify each
More informationULTRA-LIGHT DUTY ONBOARD AIR SYSTEM
ULTRA-LIGHT DUTY ONBOARD AIR SYSTEM PART NO. 10000 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of this manual before installing and using
More informationNeedle valve. Contents. User s Manual. (1) Be sure to read the following warranty clauses of our product 1. (2) General operating instructions 2
Serial No. H-V024-E-7 Needle valve User s Manual Contents (1) Be sure to read the following warranty clauses of our product 1 (2) General operating instructions 2 (3) General instructions for transportation,
More informationOPERATION MANUAL NTF-15
OPERATION MANUAL NTF-15 Nitrogen Tire Filling Valve Stem Caps (Qty=200) Order P/N 436075 RTI Technologies, Inc 10 Innovation Drive York, PA 17402 800-468-2321 www.rtitech.com 035-81235-00 (Rev B) TABLE
More informationTECHNICAL DATA 3 MODEL G-3000 DRY VALVE RISER ASSEMBLY
Page 1 of 13 1. DESCRIPTION The Viking 3 Model G-3000 Dry Valve Riser Assembly is equipped with a small profile, light weight, pilot operated valve that is used to separate the water supply from the dry
More information5 Gallon Pressure Pot with HVLP Spray Gun and Hose
California Air Tools 5 Gallon Pressure Pot with HVLP Spray Gun and Hose Model No. 365 Technical Data Type of feed.pressure Maximum pressure in the tank... 0,413Mpa (60PSI) Working pressure in the tank.0,
More informationWHEATLEY WHEATLEY SERIES 500 SWING CHECK VALVE. Installation, Operation and Maintenance Manual
WHEATLEY SERIES 500 SWING CHECK VALVE STANDARD INTEGRAL SEAT & OPTIONAL REMOVABLE SEAT 2" FP - 6" FP 150# - 1500# 8" FP - 12" FP 150# - 900# API 6D and B16.34 2" FP - 4" FP 5000# DRILLING PRODUCTION VALVE
More information420C AIR COMPRESSOR KIT PART NO C AIR COMPRESSOR KIT PART NO
420C AIR COMPRESSOR KIT PART NO. 42042 460C AIR COMPRESSOR KIT PART NO. 46043 420C 460C IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of this
More informationCompact Triple Cabinet Outlet Station Model B Installation and Operating Instructions
PORTER Parker Hannifin Corporation Porter Instrument Division 245 Township Line Rd. P.O. Box 907 Hatfield, PA 19440-0907 USA (215) 723-4000 / fax (215) 723-5106 Compact Triple Cabinet Outlet Station Model
More informationType S301 & S302 Gas Regulators INTRODUCTION INSTALLATION. Scope of Manual. Description. Specifications. Type S301 and S302. Instruction Manual
Fisher Controls Instruction Manual Type S301 & S302 Gas Regulators October 1981 Form 5180 WARNING Fisher regulators must be installed, operated, and maintained in accordance with federal, state, and local
More informationWHEATLEY Series 500 Swing Check Valve
Document Number: TC003001-13 Revision: 02 WHEATLEY Series 500 Swing Check Valve Installation, Operation, and Maintenance Manual TABLE OF CONTENTS BILL OF MATERIALS...3 SCOPE...5 INSTALLATION AND OPERATION
More informationIMPORTANT SAFETY INSTRUCTIONS
IMPORTANT SAFETY INSTRUCTIONS CAUTION - To reduce risk of electrical shock: - Do not disassemble. Do not attempt repairs or modifications. Refer to qualified service agencies for all service and repairs.
More informationTHE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301
THE HF-300 SERIES Operating and Service Manual Series includes all variants of HF-300/301 Issue A July 2015 1 TABLE OF CONTENTS 1. Description... 3 2. Installation... 3 3. Operation... 4 3.1. Spring Loaded...
More informationAquor House Hydrant V2
Aquor House Hydrant V2 IN-WALL OUTDOOR FAUCET SYSTEM FLUSH-MOUNT QUICK-CONNECT SELF-DRAINING ANTI-SIPHON NON-FREEZE VBH-SERIES AQUOR HOUSE HYDRANT V2: WALL HYDRANT / FREEZELESS SILLCOCK Freezeless sillcock
More information250C-IG COMPRESSOR KIT 12V PART NO C-IG COMPRESSOR KIT 24V PART NO
250C-IG COMPRESSOR KIT 12V PART NO. 25050 250C-IG COMPRESSOR KIT 24V PART NO. 25058 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of this manual
More information! WARNING. Model PFC-1-G (direct acting) PFC-1-GR (reverse acting) Modulating Pneumatic Liquid Level Controls INSTRUCTION MANUAL MM-110B
INSTRUCTION MANUAL MM-110B Model PFC-1-G (direct acting) PFC-1-GR (reverse acting) Modulating Pneumatic Liquid Level Controls APPLICATIONS: Use with other pneumatic devices, for liquid level sensing in
More informationGILMONT ACCUCAL FLOWMETERS
OPERATING MANUAL GILMONT ACCUCAL FLOWMETERS 28W092 Commercial Ave. Barrington, IL U.S.A. 60010-2392 (847) 381-4888 (847) 381-7053 (Fax) 800-962-7142 www.barnant.com e-mail: barnant@barnant.com A-1299-0766
More informationAnti-flood device Model B-1
December 4, 2009 Dry Systems 123a 1. DESCRIPTION The Anti-flood Device is required when Viking accelerators are installed on dry systems according to Viking Model E-1 Accelerator Trim Charts. In the SET
More informationInstallation Troubleshooting Maintenance Instructions Installation / Start-up
Model ZW207 Installation Troubleshooting Maintenance Instructions Installation / Start-up NOTE: Flushing of all pipe lines is to be performed to remove all debris prior to installing valve. 1. For making
More informationAnderson Greenwood Series 93 Positive Pressure POSRV Installation and Maintenance Instructions
Before installation these instructions must be fully read and understood Installation and maintenance instructions for Series 93 Positive Pressure Pilot Operated Safety Relief Valves (POSRV). The intent
More informationInstallation, Operation, and Maintenance Manual
Installation, Operation, and Maintenance Manual Welker Probe Instrument Regulator Model The information in this manual has been carefully checked for accuracy and is intended to be used as a guide for
More informationInstruction Manual O-Ring Ejector Chlorine & Sulfur Dioxide to 500 PPD (10 kg/h)
Instruction Manual O-Ring Ejector Chlorine & Sulfur Dioxide to 500 PPD (10 kg/h) 100 PPD (2 kg/h) Maximum 250 PPD (5 kg/h) Maximum 500 PPD (10 kg/h) Maximum - 1-122.6006.9 These instructions describe the
More informationSummary of Accumulators These operating instructions apply to the accumulators listed in the table shown below.
Page 1 of 15 Summary of Accumulators These operating instructions apply to the accumulators listed in the table shown below. Model Number A25-XXXX-3.6K Series A40-XXXX-3K Series A40X-XXXX-3.6K Series A60-XXXX-3K
More informationContents. Stainless Steel Side Block. 1.1 Separating the Side Block. Stainless Steel Side Block Reassembly of. Assembly from the Helmet Shell
Separating the Side Block Assembly from the Helmet Shell Contents SSB-1 SSB-3 SSB-5 SSB-5 SSB-7 1.1 Separating the Side Block Assembly from the Helmet Shell 1.2 Side Block Assembly Replacement 1.3 Defogger
More informationModels: C62/63/64-A/D
Installation & Service C62-991-24A3 Models: C62/63/64-A/D Diaphragm Bypass Pressure Regulating Valves IMPORTANT Record your pump model number and serial number here for easy reference: Model No. Serial
More informationTypes 749B and R130 Changeover Manifolds
Instruction Manual MCK-1179 Types 749B and R130 June 2012 Types 749B and R130 Changeover Manifolds TYPE HSRL-749B TYPE 64SR/122 TYPE R130/21 TYPE 749B/21 Figure 1. Changeover Manifolds and Regulator Assemblies
More informationOperation Manual Piston Sensed Gas Pressure Regulators
687 Technology Way Napa, CA 94558 Phone: (707) 259-0102 FAX: (707) 259-0117 www.aptech-online.com Operation Manual Piston Sensed Gas Pressure Regulators (Models KT9, KT10, Welded KT10, KT12) Table of Contents:
More informationRG1200 Service and Repair Manual
Dive Rite RG 1200 Regulator Service and Repair Manual Page 1 Text and Photography by Pete Nawrocky Copyright ( ) 1999-2000, Lamartek, Inc., dba Dive Rite RG1200 Service and Repair Manual First Stage.........................................
More informationTECHNICAL DATA. Q = C v P S
January 6, 2012 Preaction 348a 1. Description Viking supervised Surefire Preaction Systems utilize the Viking G-6000P Valve. The small profile, lightweight, pilot operated Viking G-6000P Valve comes complete
More informationDS05C,D,G Dial Set Pressure Regulating Valves
DS05C,D,G Dial Set Pressure Regulating Valves APPLICATION The Honeywell DS05C,D,G Dial Set Pressure Regulating Valve is a high quality pressure regulating valve that maintains a constant outlet pressure
More informationInstallation and Use Manual
Installation and Use Manual EMASMB & LMASMB Surface Mount Bottle Filling Stations Model EMASMB IMPORTANT THIS IS AN INDOOR APPLICATION ONLY! ALL SERVICE TO BE PERFORMED BY AN AUTHORIZED SERVICE PERSONNEL.
More informationChemical Injection Technologies Installation/Service Bulletin
Bulletin 4002 Chemical Injection Technologies Installation/Service Bulletin SUPERIOR Series CL-16/26/56 Automatic Switchover Gas Chlorinator - Installation & Operation IMPORTANT!! READ THESE PRECAUTIONS
More informationGETZ EQUIPMENT INNOVATORS RECOVERY SYSTEM CLEAN AGENT FE-36 / HALOTRON 1 PART NO. 4G59751 OPERATIONS MANUAL
GETZ EQUIPMENT INNOVATORS CLEAN AGENT FE-36 / HALOTRON 1 PART NO. 4G59751 OPERATIONS MANUAL GETZ EQUIPMENT INNOVATORS PEKIN, ILLINOIS, U.S.A. PHONE: (888) 747-4389 FAX: (309) 495-0625 Limited Warranty
More information42045 Heavy Duty ADA Base Model Kit: 85/105 PSI (ADA Compressor Only) Heavy Duty ADA Base Model Kit: 110/145 PSI (ADA Compressor Only)
42045 Heavy Duty ADA Base Model Kit: 85/105 PSI (ADA Compressor Only) 42047 Heavy Duty ADA Base Model Kit: 110/145 PSI (ADA Compressor Only) 45052 Constant Duty ADA Base Model Kit: 85/105 PSI (ADA Compressor
More information250C-IG COMPRESSOR KIT 12V PART NO C-IG COMPRESSOR KIT 24V PART NO
250C-IG COMPRESSOR KIT 12V PART NO. 25050 250C-IG COMPRESSOR KIT 24V PART NO. 25058 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of this manual
More informationTECHNICAL DATA Q = C. v P S. 2 Model G-2000 Dry valve. Page 1 of 13
Page 1 of 13 1. Description The Viking 2 Model G-2000 Dry Valve Riser Assembly consists of a small profile, light weight, pilot operated valve that is used to separate the water supply from the dry sprinkler
More informationSSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR
S S F U SSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR MACHINERY DIVISION OWNER S MANUAL UNIT INSTRUCTIONS Please follow all SSFU Safety Instructions. Contact your Duro Dyne Tech Service if you have
More informationR E D I C O N T R O L S
R E D I C O N T R O L S Operation & Maintenance Manual Portable Service Purger for Low Pressure Chillers Model: PSP-LP-1B For Refrigerants R-11, R-113, R-114 & R-123 & Other Similar Refrigerants File Literature
More informationDPC-30 DPC-100. Reference Manual
DPC-30 DPC-100 Reference Manual 1. Introduction 1.1 Description The Martel DPC Digital Pneumatic Calibrator improves upon traditional dial gauge pneumatic calibrators. The Martel DPC improves accuracy,
More informationTECHNICAL DATA. Q = C v P S
Preaction 346a 1. Description The 6 Model G-6000P Electric Release Preaction System Riser Assembly can be used as a Single Interlock Preaction System with Electric Release, or as a Double Interlock Preaction
More informationTechcon Systems TS1254 Pressure Pot
Techcon Systems TS1254 Pressure Pot User Guide Copyright OK International, Inc. TABLE OF CONTENT 1. SAFETY. 3 1.1 Intended Use 3 1.2 Safety Precaution 3 2. FEATURES. 5 3. SPECIFICATIONS 4 4. INSTALLATION
More informationTHE BP-301 SERIES. Operating and Service Manual. Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv)
THE BP-301 SERIES Operating and Service Manual Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv) Issue B October 2015 1 TABLE OF CONTENTS 1. Description... 3 2. Installation... 3 3. Operation...
More informationPANACEA P200 STAINLESS STEEL USER MANUAL
Basic Features Stainless Steel body with 1.75 outside diameter. Will fit into wells down to 2 (schedule 80). Has an internal chamber volume of 200mL. Maximum depth below ground surface of 3000ft. Other
More informationTECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1
Dry 131h 1. DESCRIPTION The Viking Model G-1 Maintenance Air Compressor is an electric motor-driven, aircooled, single-stage, oil-less compressor. The unit is equipped with a check valve and provides a
More informationCSA Sample Draw Aspirator Adapter Operator s Manual
30-0951-CSA Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0367 Revision: 0 Released: 4/30/15 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating
More informationYoke Block Instruction Manual
Yoke Block Instruction Manual ! WARNING IMPORTANT: READ MANUAL COMPLETELY BEFORE OPERATING THIS DEVICE This manual contains instructions on periodically required checks to be performed by the user. These
More informationInstallation, Operation, and Maintenance Manual
Installation, Operation, and Maintenance Manual Welker The information in this manual has been carefully checked for accuracy and is intended to be used as a guide for the installation, operation, and
More informationINSTALLATION INSTRUCTIONS
INSTALLATION INSTRUCTIONS HIGH PRESSURE PUMP To minimize vibration, it is best to build brackets on the motor itself, similar to alternator brackets. Use cardboard to construct a pattern first before making
More informationP5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.
Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 TestEquipmentDepot.com P5513 Pneumatic Comparison Test Pump Users Manual PN 3963372 November 2010 2010 Fluke Corporation. All
More informationVessel Installation and Operating Manual
For In-ground and Aboveground Pools Vessel Installation and Operating Manual Important Safety Information Please read this manual prior to installation. Nature 2 Express is designed to sanitize in-ground
More informationG92 Series BASOTROL Automatic Pilot Gas Valve
Installation Instructions Issue Date September 17, 2008 G92 Series BASOTROL Automatic Pilot Gas Valve Installation IMPORTANT: Only qualified personnel should install or service BASO Gas Products. These
More informationTECHNICAL DATA. Q = C v P S
January 6, 2012 Preaction 333a 1. Description Viking supervised Surefire Preaction Systems Utilize the Viking G-3000P Valve. The small profile, lightweight, pilot-operated Viking G-3000P Valve comes complete
More informationInstruction Manual Updated 7/26/2011 Ver. 2.2
4-Unit Model MB HTHP Filter Press #171-50-4: 115-Volt #171-51-4: 230-Volt Instruction Manual Updated 7/26/2011 Ver. 2.2 OFI Testing Equipment, Inc. 11302 Steeplecrest Dr. Houston, Texas 77065 U.S.A. Tele:
More informationPROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road
PROPORTIONING VALVE Model 150 INSTRUCTION MANUAL March 2017 IMS Company 10373 Stafford Road Telephone: (440) 543-1615 Fax: (440) 543-1069 Email: sales@imscompany.com 1 Introduction IMS Company reserves
More informationPENBERTHY SUBMERSIBLE AUTOMATIC SUMP DRAINER INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS
Before installation, these instructions must be read carefully and understood. PRODUCT WARRANTY Emerson warrants its Penberthy products as designed and manufactured to be free of defects in the material
More informationTechcon Systems TS1258 Pressure Pot
Techcon Systems TS1258 Pressure Pot User Guide TABLE OF CONTENT 1. SAFETY. 3 1.1 Intended Use 3 1.2 Safety Precaution 3 2. CERTIFICATE OF COMFORMANCE 4 3. FEATURES. 5 4. SPECIFICATIONS 5 5. INSTALLATION
More information