DART-Bruker Vapur Interface Manual SI-410-A-R Rev 1-3

Size: px
Start display at page:

Download "DART-Bruker Vapur Interface Manual SI-410-A-R Rev 1-3"

Transcription

1 DART-Bruker Vapur Interface Manual SI-410-A-R Rev 1-3 For All Bruker Mass Spectrometer Models (Except Bruker Esquire) Rev 1 Document # IonSense Inc. 999 Broadway Suite 404 Saugus, MA 01906

2 Copyright by IonSense Inc. All rights reserved. The information in this document has been carefully checked and is believed to be reliable. However, no responsibility is assumed for inaccuracies. Statements in the document not intended to create any warranty, expressed or implied. Specification and performance characteristics of the hardware and software described in the manual may be changed at any time without notice. IonSense Inc. reserves the right to make changes in any product herein in order to improve reliability, design, or function. IonSense does not assume any liability arising out of application or use of any product or circuit described nor does it cover any license under its patent rights or the rights of others. The apparatus and application of the apparatus described in this document is protected by US Patent Number 6,949,741 and used under license; additional patents pending. All trademarks are properties of their respective owners. 2

3 This manual details the steps necessary to install a DART source on all Bruker atmospheric pressure inlet mass spectrometers (except Bruker Esquire). This document uses a DART SVP 100 or 200 series source; however, the instructions should also provide general guidance for installation of a DART 100 or 100S. Table of Contents Installation of the Bruker Vapur Interface... 4 Attaching the DART to the Vapur Interface... 6 Addressing Software Vacuum Reading Errors... 7 Recommended Operating Parameters

4 Installation of the Bruker Vapur Interface For TOF instruments it is suggested to calibrate the instrument before switching to the DART ion source. Vapur Interface Installation 1. First, place the instrument in standby mode and remove all gas and electrical connections from the existing Bruker ESI (or other) source. 2. Remove the Bruker ion source and carefully place it to the side or in storage. Remove the Bruker end cap and atmospheric pressure inlet (API) inlet splash guard from the atmospheric pressure inlet area. The end cap and splash guard are pictured in Figure 1. a. CAUTION: These pieces will be extremely hot. Use gloves or wait for capillary and source temperatures to cool before removing. Figure 1: Bruker end cap and splash guard removed. A Figure 2: Bruker API inlet with end cap and splash guard removed. B Figure 3: A) IonSense Vapur Interface hinges slide down into the slots on the Bruker instrument in the same way that the Bruker source housing fits. B) The liquid drainage barbed fitting under the API inlet is sealed with the fitting provided by IonSense and the liquid drainage tube is slid over the fitting in order to seal the tube. 4

5 Figure 4: Vapur Interface in place on Bruker maxis 4G in the open position. Note: The latching bracket on the Bruker instrument will need to be adjusted for the DART Vapur Interface. To adjust the length of the latch, loosen the thumb nut and then rotate the entire latch to lengthen the latch. Tighten the thumb nut to lock the latch length. Inserting the Ceramic Ion Transfer Tube into the Vapur Interface Remove the Swagelok nut from the front of the Vapur Interface and insert the ceramic ion transfer tube into the interface. Secure the ceramic tube with one of the provided graphite Vespel ferrules making sure to set a 2 mm gap inside the interface. Note: To easily set the 2 mm gap, push the ceramic tube against the Bruker capillary, secure the Swagelok fitting with ferrule and then mark the tube with a pencil. Pull the tube out as shown in Figure 5 below so that a 2 mm gap is created inside the interface. Figure 5: Ceramic ion transfer tube set correctly in the Vapur Interface. Figure 6: Take care to attach the black rubber pumping line from diaphragm pump over the barbed pumping port fitting on the underside of the Vapur Interface. 5

6 Attaching the DART to the Vapur Interface Once the Vapur Interface has been successfully installed, you are ready to attach the DART source. Refer to the DART SVP Hardware and Network Installation Manual (pages 6-8) for more specifics on interfacing the DART SVP source to the Vapur Interface. IonSense website: ( 6

7 Addressing Software Vacuum Reading Errors Because the mass spectrometer is not designed to accept the large flows of helium gas required to run the DART, you may receive errors informing you that the Bruker vacuum is too high. This is not indicative of system damage and will not damage your instrument. It simply means that the vacuum gauge, which is calibrated for nitrogen gas is giving a falsely high readback. The error message can be addressed by changing the upper vacuum limit for the fore vacuum gauge. To do this, perform the following procedure: Go to the Service tab in the Bruker software: Select the MPC 555 Board from the list on the left side. _ Then enter the command VACU:SRPS=10 in the "Command" line at the bottom of page. Note: 10 mbar is the recommended value for the fore vacuum upper limit value. The Bruker default factory setting is 5 mbar on the maxis. Click "Send" button on the page to submit the command. Other Useful Commands "VACU:HELP?" shows a list of all commands. Enter "VACU:SRPS?" to see what the current pressure threshold is set too. It should be set to 10. 7

8 Recommended Operating Parameters Ion Source: Turbo Spray Bruker source settings in the software: End Plate Offset: 0 V Capillary Voltage has no effect on signal intensity set to 0 V Dry Gas: 0 L/min o The signal intensity was approximately 10 x lower when the dry gas was on and set at 4 L/min than when the Dry Gas was set to 0 L/min The Dry Gas Temperature has almost no effect on the DART signal. 8

9 Revision History REV DCR # Description of Change Effective Date 1 43 Initial Release 11/11/2013 9

ID-Cube. Installation and Operations Manual For ID-Cube Source and Power Supply. Revision 2 Document #:

ID-Cube. Installation and Operations Manual For ID-Cube Source and Power Supply. Revision 2 Document #: ID-Cube Installation and Operations Manual For ID-Cube Source and Power Supply Revision 2 Document #: 7.1.002 IonSense Inc. 999 Broadway Suite 404 Saugus, MA 01906 Copyright 2014 by IonSense Inc. All rights

More information

Mass Spec will not Autotune

Mass Spec will not Autotune Mass Spec will not Autotune Applies to 5973A/N MSD What could be the problem? There could be several things that would cause your Mass Spec not to Autotune. The most common, easily corrected Autotune problems

More information

Agilent Auxiliary Gas Module AGM 2. User s Guide

Agilent Auxiliary Gas Module AGM 2. User s Guide Agilent Auxiliary Gas Module AGM 2 User s Guide Notices Agilent Technologies, Inc. 1992, 1996, 1999, 2010 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

The Open-Split Capillary Interface. Part No

The Open-Split Capillary Interface. Part No The Open-Split Capillary Interface Part No. 113532 2 INTRODUCTION With the advent of fused silica capillary columns and their high degree of flexibility, it is obviously attractive to be able to take the

More information

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve USER'S GUIDE Nanoflow Probe 3-4mm Red Stripe ptfe Sleeve To Injector ZU1XC Fused Silica 20µm 90µm TTF115 ptfe Sleeve Fused Silica 25µm 375µm Micromass UK Limited Floats Road Wythenshawe M23 9LZ Tel: +44

More information

Booster Pump PB4-60 Replacement Kits

Booster Pump PB4-60 Replacement Kits Booster Pump PB4-60 Replacement Kits FOR YOUR SAFETY - This product must be installed and serviced by a contractor who is licensed and qualified in pool equipment by the jurisdiction in which the product

More information

Agilent ICP-MS Interactive Troubleshooting tool for Plasma ignition problem

Agilent ICP-MS Interactive Troubleshooting tool for Plasma ignition problem Agilent ICP-MS Interactive Troubleshooting tool for Plasma ignition problem Rev. B 2017, January, 20 Next > Preconditions 7900, 7800, 7700, 8800, or 8900 ICP-MS is used. Sample introduction type is PeriPump.

More information

Techcon Systems TS1254 Pressure Pot

Techcon Systems TS1254 Pressure Pot Techcon Systems TS1254 Pressure Pot User Guide Copyright OK International, Inc. TABLE OF CONTENT 1. SAFETY. 3 1.1 Intended Use 3 1.2 Safety Precaution 3 2. FEATURES. 5 3. SPECIFICATIONS 4 4. INSTALLATION

More information

Agilent 1290 Infinity Pump Head Maintenance

Agilent 1290 Infinity Pump Head Maintenance Agilent 1290 Infinity Pump Head Maintenance Technical Note Agilent Technologies Notices Agilent Technologies, Inc. 2012-2015, 2016 No part of this manual may be reproduced in any form or by any means (including

More information

Mass Spec will not pump down

Mass Spec will not pump down Mass Spec will not pump down Applies to 5973A/N MSD There are several situations that could cause the 5973A/N to have problems pumping down. The following scenarios should help in isolating problems related

More information

Step By Step Instructions for VT Experiments on the Bruker 300 MHz Spectrometer

Step By Step Instructions for VT Experiments on the Bruker 300 MHz Spectrometer Step By Step Instructions for VT Experiments on the Bruker 300 MHz Spectrometer (1) Use the drop down menu to create a new experiment/dataset for acquiring the room temperature 1D spectrum. The name of

More information

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet: OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle

More information

Variable Temperature Operation

Variable Temperature Operation Variable Temperature Operation AVIII 400 MHz Spectrometer November 11, 2012 The CBC NMR Laboratory University of Delaware AVIII 400 User Guide: Variable Temperature Operation 1 INTRODUCTION... 5 1.1 SAFETY...

More information

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300 Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300 The Thermo Scientific icap 6300 Inductively Coupled Plasma Optical Emission

More information

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 QUICK START GUIDE P/N 6006154, REVISION C MAY 2013 Model 8120 Mask Integrity Tester is patented under U.S. Patent No. 8,312,761. Additional patents are pending.

More information

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures

More information

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...

More information

Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity

Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity Rev. B 2017,Janyary,20 Next > Preconditions 7900, 7800, 7700, or 8800 ICP-MS is used. Sample introduction type is PeriPump. If you are

More information

3M Liqui-Cel EXF-14x28 and 14x40 Series Membrane Contactor

3M Liqui-Cel EXF-14x28 and 14x40 Series Membrane Contactor Membrane Contactors 3M Liqui-Cel EXF-14x28 and 14x40 Series Membrane Contactor Assembly and Disassembly Instructions 3M.com/Liqui-Cel TABLE OF CONTENTS I. Safety Information 3 II. Required Materials 4

More information

Operator Quick Guide ORBISPHERE 3654

Operator Quick Guide ORBISPHERE 3654 Operator Quick Guide ORBISPHERE 3654 Revision H - 14/03/2008 Operating Information About this Guide The information in this guide has been carefully checked and is believed to be accurate. However, Hach

More information

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS Fabrication Laboratory Revision: A Rev Date: 09/29/03 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Amray 1830 SEM. All users

More information

RC 195 Receiver-Controller

RC 195 Receiver-Controller Document No. 129-082 RC 195 Receiver-Controller Product Description The POWERS RC 195 Receiver-Controller is a pneumatic instrument that receives one, two or three pneumatic inputs. It produces a pneumatic

More information

Safety, Operating and Maintenance Manual. TRAIN Rat DS9-TRAIN

Safety, Operating and Maintenance Manual. TRAIN Rat DS9-TRAIN Safety, Operating and Maintenance Manual TRAIN Rat DS9-TRAIN Industry s first air powered entry tool! Quickly, quietly and reliably opens passenger train doors! Introducing the TRAIN Rat. This innovative

More information

COOPER POWER SERIES. D-73P3 bypass switches installation instructions. Switches MN008005EN. Effective April 2016 Supersedes S June 2013

COOPER POWER SERIES. D-73P3 bypass switches installation instructions. Switches MN008005EN. Effective April 2016 Supersedes S June 2013 Switches MN008005EN Effective April 2016 Supersedes S328-120-1 June 2013 D-73P3 bypass switches installation instructions COOPER POWER SERIES DISCLAIMER OF WARRANTIES AND LIMITATION OF LIABILITY The information,

More information

Addendum to the Waters Micromass Quattro Premier XE Mass Spectrometer Operator s Guide

Addendum to the Waters Micromass Quattro Premier XE Mass Spectrometer Operator s Guide Addendum to the Waters Micromass Quattro Premier XE Mass Spectrometer Operator s Guide 715001216/Revision A Copyright Waters Corporation 2006. All rights reserved. Copyright notice 2006 WATERS CORPORATION.

More information

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents Title: CHA E-Beam Evaporator Page 1 of 7 Table of Contents Purpose/Scope... 2 2.0 Reference Documents... 2 3.0 Equipment/Supplies/Material... 2 4.0 Safety... 2 5.0 Set Up Procedures... 2 5.1 PC Logon and

More information

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii Operator s Manual Table of Contents Operator Safety... 1 Introduction... 2 Unpacking and Setup... 3 Unpacking... 3 Setup... 4 FLYER Overview... 5 FLYER Diagram... 5 Capabilities... 5 Control Panel... 6

More information

ECHO MANUAL WARNING. L B A ltim e te rs. ECHO is a trademark of LB Altimeters, Denmark

ECHO MANUAL WARNING. L B A ltim e te rs. ECHO is a trademark of LB Altimeters, Denmark ECHO MANUAL L B A ltim e te rs ECHO is a trademark of LB Altimeters, Denmark LB Altimeters operates a policy of continuous development Therefore, we reserve the right to make changes and improvements to

More information

Standard Operating Procedure Porosity Analyzer - The Micromeritics Tri-Star II (surface area) with Micrometrics VacPrep 061 (Sample Degas System)

Standard Operating Procedure Porosity Analyzer - The Micromeritics Tri-Star II (surface area) with Micrometrics VacPrep 061 (Sample Degas System) Standard Operating Procedure Porosity Analyzer - The Micromeritics Tri-Star II (surface area) with Micrometrics VacPrep 061 (Sample Degas System) The Micromeritics Tri-Star II Surface Area and Porosity

More information

Pressure Injection Cell Operator s Manual

Pressure Injection Cell Operator s Manual Pressure Injection Cell Operator s Manual Operator s Manual Pressure Injection Cell CONGRATULATIONS! Congratulations on your purchase of a Next Advance Pressure Injection Cell. Please read this operator

More information

Best Practice for Identifying Leaks in GC and GC/MS Systems

Best Practice for Identifying Leaks in GC and GC/MS Systems Best Practice for Identifying Leaks in GC and GC/MS Systems Technical Overview Maintaining a leak-free GC or GC/MS system is critical for obtaining optimal system performance with reliable, reproducible,

More information

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings SomnoSuite FAQ V1.3 January 2015 Setup 1. How do I connect the SomnoSuite to my oxygen source? 2. Is there a way to speed up the downward movement of the pusher block when setting the empty position? 3.

More information

Agilent Flowmeter ADM1000

Agilent Flowmeter ADM1000 Agilent Flowmeter ADM1000 Operating Instructions Notices Agilent Technologies, Inc. 2003 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval

More information

Techcon Systems TS1258 Pressure Pot

Techcon Systems TS1258 Pressure Pot Techcon Systems TS1258 Pressure Pot User Guide TABLE OF CONTENT 1. SAFETY. 3 1.1 Intended Use 3 1.2 Safety Precaution 3 2. CERTIFICATE OF COMFORMANCE 4 3. FEATURES. 5 4. SPECIFICATIONS 5 5. INSTALLATION

More information

[needs to be discussed with faculty if any new material to be done]

[needs to be discussed with faculty if any new material to be done] Tool Name: EVO 18 Version 1 Authorized Users: Gayatri, Vishnu kant Materials Used: Sample holder has facility to mount 9 stubs: Si, Glass, Materials allowed : [needs to be discussed with faculty if any

More information

Digital Vacuum Regulator

Digital Vacuum Regulator Temperature Control for Research and Industry Digital Vacuum Regulator User s Manual Model DVR-380 INDEX SECTION PAGE SAFETY NOTICES................................................. 3 1. QUICK OPERATING

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual ARC12M Sputter Copyright 11.2015 by Hong Kong University of Science & Technology. All rights reserved. Page 1 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness

More information

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT ATTENTION SYRINGE REPLACEMENT Please read through the document completely before starting any repairs. Refer to the proper section in the service manual for complete removal and replacement procedures.

More information

COMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS

COMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS COMPACT TRIPLE WALL OUTLET STATION MODEL 6255-1 INSTALLATION AND OPERATING INSTRUCTIONS To assure safe operation and conformation to local fire codes, all Porter Outlet Stations are designed to be used

More information

SA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Knob Adjustment)

SA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Knob Adjustment) 3810050/5 IM-P381-01 CH Issue 5 SA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Knob Adjustment) Installation and Maintenance Instructions 1. Safety information 2. Use 3.

More information

User Manual 1 P a g e Rev. V1.6-EN 11/08/2014

User Manual 1 P a g e Rev. V1.6-EN 11/08/2014 User Manual 1 P a g e Rev. V1.6-EN 11/08/2014 Copyright Disclaimer Trademarks and patents Intended use Contact info 2011 Inflotrolix, Inc. This document may not be copied in whole or in part or otherwise

More information

Model 6202 Reduced Oxygen Breathing Device 2

Model 6202 Reduced Oxygen Breathing Device 2 Model 6202 Reduced Oxygen Breathing Device 2 Simulator Accessory Package Operator s Guide Revision 2 26 FEB 2013 Environics Inc. 69 Industrial Park Road East Tolland, CT 06084-2805 USA Phone (860) 872-1111

More information

Model 130M Pneumatic Controller

Model 130M Pneumatic Controller Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate

More information

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr A DIVISION OF THE FREDERICKS COMPANY 2400 PHILMONT AVE. HUNTINGDONVALLEY, PA 19006 PARTS LIST 1 3 4 2 # QTY ITEM DESCRIPTION

More information

Connecting a SSL/PTV Backflush System

Connecting a SSL/PTV Backflush System This section provides instructions for connecting your SSLBKF/PTVBKF injector modules with pre-column and analytical capillary column into the oven. Refer to the chapters SSLBKF Injector Module and PTVBKF

More information

M3-LED. Operator s Manual. Operator s Manual MADE IN THE USA USA

M3-LED. Operator s Manual. Operator s Manual MADE IN THE USA USA M3-LED Operator s Manual Operator s Manual MADE IN THE USA USA Section TABLE OF CONTENTS Page Warnings and Cautions... 1 M3-LED Parts Diagram... 4 Mounting... 6 Dismounting... 8 Switch Operation... 10

More information

MB-SP Series Solvent Purification System

MB-SP Series Solvent Purification System Operation Manual MB-SP Series Solvent Purification System Table of contents 1. System overview 2 2. Column activation.. 5 3. System purging.. 6 4. Glove box integration.... 7 5. Solvent reservoir installation....

More information

Steel Spring Driven Reel Instruction Manual

Steel Spring Driven Reel Instruction Manual Steel Spring Driven Reel Instruction Manual WARNING: Read carefully and understand all INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety precautions may result in

More information

Equipment Operating Procedure Glove Box

Equipment Operating Procedure Glove Box Equipment Operating Procedure Glove Box Page 1 0.0 Changing the Compressed Gas Cylinder 1. Complete Compressed Gas Cylinder training from EHS website before manually exchanging gas cylinders. In order

More information

airplaine kite Model assembly & Operating Instructions Important: If damage is caused due to a crash; your warranty is void.

airplaine kite Model assembly & Operating Instructions Important: If damage is caused due to a crash; your warranty is void. airplaine kite Model 95081 assembly & Operating Instructions Important: If damage is caused due to a crash; your warranty is void. Visit our website at: http://www.harborfreight.com Read this material

More information

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA Page 1 of 9 SSI Solaris 150 RTA The Solaris 150 RTA is a rapid thermal annealing system capable of handling sample sizes up to 100mm (4 diameter) or smaller. The system can anneal in N 2 and Forming gas

More information

6900 Maintenance Instruction System Flush

6900 Maintenance Instruction System Flush Equipment Required FA74005 Damper Drain Tube FA16005 Cover Removal Tool FA900005 Beaker 0.25 Litre FA900003 Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves Latex

More information

Glove Box Installation Manual

Glove Box Installation Manual Glove Box Installation Manual 1998 by M. Braun Company File: GB-UNI-INS.DOC! Edition 08-00 by M. Boutin! Subject to be changed without notice Glovebox Installation Your Glove box has been fully assembled,

More information

Latvin Luxury Shower Panel. Telephone Product Specification. ~ Minimum Working Pressure 1.0 bar ~ Maximum Working Pressure 3.

Latvin Luxury Shower Panel. Telephone Product Specification. ~ Minimum Working Pressure 1.0 bar ~ Maximum Working Pressure 3. Product Specification ~ Minimum Working Pressure 1.0 bar ~ Maximum Working Pressure 3.0 bar Latvin Luxury Shower Panel ~ Fixing Centres 150mm +/- 10mm ~ Outlet size 1/2" Bottom Outlet Always maintain a

More information

WSN: XXXXXX 101 ROME COURT FORT COLLINS, CO, PHONE #: (970) FAX #: (970) TO-Clean Manual

WSN: XXXXXX 101 ROME COURT FORT COLLINS, CO, PHONE #: (970) FAX #: (970) TO-Clean Manual WSN: XXXXXX 101 ROME COURT FORT COLLINS, CO, 80524 PHONE #: (970) 221-9179 FAX #: (970) 221-9364 TO-Clean Manual Copyright 2008 by Wasson ECE Instrumentation, Inc. (SJC 02/05/2008) All rights reserved

More information

Section 3- Repair and Calibration Procedures

Section 3- Repair and Calibration Procedures Section 3- Repair and Calibration Procedures 1. Refrigerant Removal Procedures 2. Scale/PC Board Recalibration Procedure 3. PC Board Configuration Procedure of AR400 4. Scale Tare Procedure 5. Gauge Calibration

More information

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008 UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method

More information

GE Oil & Gas Model. Masoneilan* Air Lock-up Valve Instruction Manual. GE Data Classification : Public

GE Oil & Gas Model. Masoneilan* Air Lock-up Valve Instruction Manual. GE Data Classification : Public GE Oil & Gas 77-6 Model Masoneilan* Air Lock-up Valve Instruction Manual GE Data Classification : Public THESE INSTRUCTIONS PROVIDE THE CUSTOMER/OPERATOR WITH IMPORTANT PROJECT-SPECIFIC REFERENCE INFORMATION

More information

COOPER POWER SERIES. 600 A, 15, 25, and 35 kv Class Cleer Grounding Elbow Installation Instructions. Loadbreak Apparatus Connectors MN650056EN

COOPER POWER SERIES. 600 A, 15, 25, and 35 kv Class Cleer Grounding Elbow Installation Instructions. Loadbreak Apparatus Connectors MN650056EN Loadbreak Apparatus Connectors MN650056EN Effective July 2018 Supersedes May 2017 COOPER POWER SERIES 600 A, 15, 25, and 35 kv Class Cleer Grounding Elbow Installation Instructions DISCLAIMER OF WARRANTIES

More information

FX134, FX1234, FX3030 Problem and Solutions

FX134, FX1234, FX3030 Problem and Solutions FX134, FX1234, FX3030 Problem and Solutions Unit Will Not Power Up Turn the power switch on. The unit LCD will display revision program and filter life within 3 to 5 seconds after turning on unit. If this

More information

AA100 Acoustic Actuator

AA100 Acoustic Actuator AA100 Acoustic Actuator Operating Manual REV. A, September 2001 COPYRIGHT 2001. ALL RIGHTS RESERVED Return Procedure It is necessary to obtain from ORE Offshore a Returned Material Evaluation (RMA) number

More information

3M Liqui-Cel EXF-10x28 Series Membrane Contactor

3M Liqui-Cel EXF-10x28 Series Membrane Contactor Membrane Contactors 3M Liqui-Cel EXF-10x28 Series Membrane Contactor Assembly and Disassembly Instructions 3M.com/Liqui-Cel TABLE OF CONTENTS I. Safety and Warning 3 II. Assembly Parts 4 III. Part Orientation

More information

Login to ilab Kiosk. Revised 05/22/2018. Load your sample:

Login to ilab Kiosk. Revised 05/22/2018. Load your sample: Login to ilab Kiosk Load your sample: 1. Check: The analysis chamber pressure is

More information

Operator s/parts Manual

Operator s/parts Manual Operator s/parts Manual 24 & 30 Folding No-Till Drill Harrow Manufacturing, Inc. P.O. Box 5060 Salina, Kansas 67402-5060! Read the Operator s manual entirely. When you see this symbol, the subsequent instructions

More information

BGA244 Binary Gas Analyzer

BGA244 Binary Gas Analyzer Quick Start Guide Revision 1.0 Certification Warranty Service certifies that this product met its published specification at the time of shipment. This product is warranted against defects in materials

More information

Fisher Lever-Lock Handlever Actuators for POSI-SEAL A81 Rotary Valves

Fisher Lever-Lock Handlever Actuators for POSI-SEAL A81 Rotary Valves Instruction Manual D103319X012 Lever-Lock Handlever Actuator Fisher Lever-Lock Handlever Actuators for POSI-SEAL A81 Rotary Valves Contents Introduction................................. 1 Scope of Manual.........................

More information

U.S. Patent No. 7,922,246. Patents Pending

U.S. Patent No. 7,922,246. Patents Pending U.S. Patent No. 7,922,246 Patents Pending 2 Table of Contents Page General Information... 3 Warnings and Cautions... 4 Tools... 6 SmartDock Parts... 6 Initial Set-Up and Adjustment... 7 Select Valve Retaining

More information

Installation, Operation, and Maintenance Manual

Installation, Operation, and Maintenance Manual Installation, Operation, and Maintenance Manual Welker Instrument Supply Pressure System Model WIC The information in this manual has been carefully checked for accuracy and is intended to be used as a

More information

N2 Blanketing Valve DST100 / DST200 TYPE INSTRUCTION MANUAL CONTENTS K.S.P.C. General Description Operation. Installation Maintenance

N2 Blanketing Valve DST100 / DST200 TYPE INSTRUCTION MANUAL CONTENTS K.S.P.C. General Description Operation. Installation Maintenance DST100 / DST200 TYPE N2 Blanketing Valve INSTRUCTION MANUAL CONTENTS General Description Operation Installation Maintenance K.S.P.C 488-1 Wolha-ro, Tongjin-eup, Gimpo-si, Gyeonggi-Do, Korea Tel : +82-31-998-3825~7

More information

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION BEST KNOWN METHODS Transpector XPR3 Gas Analysis System DESCRIPTION The Transpector XPR3 is a third-generation, quadrupole-based residual gas analyzer that operates at PVD process pressures and is the

More information

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual Installation, Operation, and Maintenance Manual Welker Sampler Model GSS-1 The information in this manual has been carefully checked for accuracy and is intended to be used as a guide to operations. Correct

More information

Velocity USER GUIDE CPU WATER BLOCK. 1st Revision, September 6th 2018

Velocity USER GUIDE CPU WATER BLOCK. 1st Revision, September 6th 2018 Velocity CPU WATER BLOCK 1st Revision, September 6th 2018 USER GUIDE This product is intended for installation only by expert users. Please consult with a qualified technician for installation. Improper

More information

DEH Instructions. Three-Phase, Platform-Mounted Distribution Transformers

DEH Instructions. Three-Phase, Platform-Mounted Distribution Transformers g DEH 40050 Instructions Three-Phase, Platform-Mounted Distribution Transformers Introduction The equipment covered by these instructions should be operated and serviced only by competent technicians familiar

More information

FX IMPACT OWNER S MANUAL

FX IMPACT OWNER S MANUAL made in sweden FX IMPACT OWNER S MANUAL Warranty Information All FX Airguns carry a One Year Warranty against faulty workmanship and defective materials. If it becomes necessary, first contact the dealer

More information

Met One E-BAM Particulate Monitor

Met One E-BAM Particulate Monitor STANDARD OPERATING PROCEDURES Met One E-BAM Particulate Monitor AMBIENT AIR MONITORING PROGRAM for the 130 LIBERTY STREET DECONSTRUCTION PROJECT LOWER MANHATTAN DEVELOPMENT CORPORATION 1 Liberty Plaza

More information

Installation Instructions

Installation Instructions Installation Instructions COLONY SOFT 7.0 Centerset Lavatory Faucet 7.0 with Speed Connect Drain Congratulations on purchasing your American Standard faucet with the Speed Connect drain, a features found

More information

Digital Vacuum Regulator

Digital Vacuum Regulator Temperature Control for Research and Industry Digital Vacuum Regulator User s Manual Model DVR-280 INDEX SECTION PAGE 1. QUICK OPERATING INSTRUCTIONS............................. 3 KEM-NET DATA LOGGING

More information

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions Compact Triple Cabinet Outlet Station Model 6258-1 Installation and Operating Instructions The Porter Compact Triple Outlet Station (6258-1) provides a quick, safe, and reliable method of connection to

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310 www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Varian 310 310 LC/MS Manual System INCLUDES: 310-MS LC/MS triple quadrople

More information

Agilent G3180B Two-Way Splitter Kit

Agilent G3180B Two-Way Splitter Kit Agilent G3180B Two-Way Splitter Kit With Makeup Gas Installation and Operation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2006 No part of this manual may be reproduced in any form or

More information

foldable/adjustable sawhorse

foldable/adjustable sawhorse foldable/adjustable sawhorse Model 96506 Set up And Operating Instructions Diagrams within this manual may not be drawn proportionally. Due to continuing improvements, actual product may differ slightly

More information

Procedure for Operating Columbia (Nevis) LHe Electron Bubble Chamber Cryostat. Version 5.1

Procedure for Operating Columbia (Nevis) LHe Electron Bubble Chamber Cryostat. Version 5.1 Project: Novel Electron Bubble Particle Detector Procedure for Operating Columbia (Nevis) LHe Electron Bubble Chamber Cryostat Version 5.1 (Using side filling line) Hand Processed Changes HPC No. Date

More information

Halsey Taylor Owners Manual

Halsey Taylor Owners Manual Halsey Taylor Owners Manual Refrigerated Marblyte Fountains with Back Panel HRFG - ER HRFG - SR HRFG - SER Installer To assure you install this model easily and correctly, PLEASE READ THESE SIMPLE INSTRUCTIONS

More information

P9000 DESCRIPTION. For parts requiring RoHS compliance, please contact factory. P February /6

P9000 DESCRIPTION. For parts requiring RoHS compliance, please contact factory. P February /6 P9000 High accuracy through digital compensation High thermal stability Rugged stainless steel construction Ideal for test stands High burst pressure limit DESCRIPTION The P9000 Series is a range of advanced,

More information

User Manual. Quantos Automated Dosing Liquid Module

User Manual. Quantos Automated Dosing Liquid Module User Manual Liquid Module 1 Safety Information 1.1 Definition of warnings and symbols Signal Words WARNING for a hazardous situation with medium risk, possibly resulting in severe injuries or death if

More information

Instruction Manual Model Borehole Pressure Cell

Instruction Manual Model Borehole Pressure Cell Instruction Manual Model 3200 Borehole Pressure Cell No part of this instruction manual may be reproduced, by any means, without the written consent of Geokon, Inc. The information contained herein is

More information

Universal Gas Analyzers Lite

Universal Gas Analyzers Lite Operation Manual and Programming Reference Universal Gas Analyzers Lite UGALT100, UGALT200, UGALT300 Revision 1.1 May, 2014 Certification certifies that this product met its published specifications at

More information

Notice. VeraCode Vortex Incubator User Guide

Notice. VeraCode Vortex Incubator User Guide Notice This publication and its contents are proprietary to Illumina, Inc., and are intended solely for the contractual use of its customers and for no other purpose than to operate the system described

More information

CLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1

CLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1 ANA 74300 Oxygen Regulator with Yoke Connector User Manual Manual No. ANA 74300 001 Issue 1 Safety Thank you for purchasing this Clements Oxygen Regulator. For your safety it is imperative that this unit

More information

Product Information News September 26, 2003

Product Information News September 26, 2003 Product Information News September 26, 2003 LEVER HEIGHT INSPECTION FOR FIREHAWK MMR MSA is announcing a revision to the instructions for the Firehawk MMR Second Stage Regulator as they relate to the air

More information

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot. Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 TestEquipmentDepot.com P5513 Pneumatic Comparison Test Pump Users Manual PN 3963372 November 2010 2010 Fluke Corporation. All

More information

200 A, 15 and 25 kv class loadbreak bushing insert installation instructions

200 A, 15 and 25 kv class loadbreak bushing insert installation instructions Loadbreak Apparatus Connectors Effective October 2013 Supersedes March 2011 200 A, 15 and 25 kv class loadbreak bushing insert installation instructions S500-12-1 DISCLAIMER OF WARRANTIES AND LIMITATION

More information

Universal Anchoring Adapter For Mercedes

Universal Anchoring Adapter For Mercedes Universal Anchoring Adapter For Mercedes Anchoring Pin & Tube Set (For Most Mercedes Models) Users Manual 006 Chief Automotive Technologies, Inc. Chief s Limited One-Year Warranty & Liability CHIEF'S

More information

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist Thank you for purchasing an Agilent system. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is

More information

3M Liqui-Cel EXF-10x28 Series Membrane Contactor with ANSI or JIS Connections with Integrated End Caps

3M Liqui-Cel EXF-10x28 Series Membrane Contactor with ANSI or JIS Connections with Integrated End Caps Membrane Contactors 3M Liqui-Cel EXF-10x28 Series Membrane Contactor with ANSI or JIS Connections with Integrated End Caps Assembly and Disassembly Instructions 3M.com/Liqui-Cel TABLE OF CONTENTS I. Safety

More information

VMC - Hydraulic Counterbalance - Pressure Switch - Replacement

VMC - Hydraulic Counterbalance - Pressure Switch - Replacement Haas Technical Documentation VMC - Hydraulic Counterbalance - Pressure Switch - Replacement AD0331 Rev A Applies to machines built from: May, 1995 to May, 2012 Scan code to get the latest version of this

More information

INSTRUCTION MANUAL. January 23, 2003, Revision 0

INSTRUCTION MANUAL. January 23, 2003, Revision 0 INSTRUCTION MANUAL Model 810A In-Vitro Test Apparatus for 310B Muscle Lever January 23, 2003, Revision 0 Copyright 2003 Aurora Scientific Inc. Aurora Scientific Inc. 360 Industrial Parkway S., Unit 4 Aurora,

More information

Masoneilan* 78 Model Air Filter Regulator

Masoneilan* 78 Model Air Filter Regulator GE Oil & Gas Masoneilan* 78 Model Air Filter Regulator Instruction Manual [Rev. D] GE Data Classification : Public THESE INSTRUCTIONS PROVIDE THE CUSTOMER/OPERATOR WITH IMPORTANT PROJECT-SPECIFIC REFERENCE

More information

UNITY 2. Uni-DAAMS Manual. Version 1.3 (Changes to Section 1.1) March 2010

UNITY 2. Uni-DAAMS Manual. Version 1.3 (Changes to Section 1.1) March 2010 UNITY 2 Uni-DAAMS Manual Version 1.3 (Changes to Section 1.1) March 2010 1. Introduction...2 1.1. Installing the DAAMS tube ovens...2 1.2. Swapping between tube types...5 1.3. Changing tube oven o-rings...6

More information

4400 TwinHybrid Gas Seal

4400 TwinHybrid Gas Seal 4400 TwinHybrid Gas Seal EQUIPMENT PREPARATION MECHANICAL SEAL INSTALLATION INSTRUCTIONS 1 2 200.005" 0,13 mm 125 µ" 3,2 µm R a 3 4 32 µ" 0,8 µm R a 1000 ±.002" 0,05mm CAUTIONS These instructions are general

More information