Bassett Medical Center Transfusion Services
|
|
- Victoria Jacobs
- 5 years ago
- Views:
Transcription
1 Bassett Medical Center Transfusion Services SIGNING OUT (ISSUE) and STORAGE OF RED CELL Products in the OR Signing out/issuing of red cell products for patient s in the Operating Room (OR) is performed by the Blood Bank (BB) Technologist at the time of surgery. It will be the responsibility of the OR Circulating RN to notify the Blood Bank by phone at least 15 minutes prior to the scheduled case using first and last name and MRN. This will allow time for packing, issuing and transport of red cell products to the OR by a validated OR assistant (ORA) or if staffing does not permit, the Blood Bank Technologist. It will be standard practice, unless OR staffing does not permit, for the ORA to be dispatched to the Blood Bank for up to four (4) units of crossmatch compatible red cell products, in the temperature regulated Hemo-Cool, Jr. II following the procedure below. All units stored in the Hemo-Cool, Jr. II will have a Hemo Temp II Indicator on the back of the blood bag to ensure that unit storage has been maintained at a temperature of 1 o C-6 o C upon return to the Blood Bank. For all procedures: Coolers will be issued to the OR for a maximum of 3 hours at which time the cooler must be returned to the Blood Bank to determine temperature of the cooler and the blood bag(s) inside. A Delivery Request Form must accompany the cooler if the cooler is to be reissued to the OR.. Blood Bank Technologist will be responsible for: 1. The HemoCool, Jr.ll coolers are routinely kept in a ready to use status in the Blood Bank. However, if not already plugged in, at least one hour prior to the scheduled surgery, the Hemo-Cool, Jr. II must be plugged in making certain the power adapter (black box) is placed into the metal vented protection bracket on the end of the cooler. If the thermal plate is not already inside the Hemo-Cool Jr., it can be found stored in the Helmer #5 refrigerator, on the bottom shelf. Remove the plate from the refrigerator and place in the Hemo- Cool Jr. with the vertical portion against the cold sink in the cooler. Note: The temperature switch must be in the low position to achieve a fast cool down rate. 2. Place the number of HemoTemp II Indicators, flower side down, that are needed into the HemoTemp Activator for a minimum of one minute up to a maximum of three days, being sure to include one indicator for each unit being stored and one for a mock product. Note: Prior to placing indicators in the heat block activator, place the current date on the back of the indicator. 3. After incubation, remove the indicators from the Activator. The irreversible flower will be a black color. Once the flower on the indicator turns a blue color (30-45 seconds), remove the back of the indicator strip. 4. Remove the units that are to be issued from Helmer #4 as well as the mock product from Helmer #2. Immediately press the indicator strip onto the bottom back of the red cell product and mock product. Note: The indicator strip must be directly on the product bag and not on the unit number stickers found on the back or on the front label of the unit bag. 5. To insure proper adhesion of the indicator strip once it has been applied, place the blood bag with the face of the strip down on a flat surface and rotate the bag. For proper activation and usage: Indicator must be heated and placed on the designated product that is at a temperature of 1 o C -4 o C. If the color is green or tan, the indicator has not been properly activated. This could be a result of the blood being warmer than 4 o C when the strip was applied. Remove the indicator and repeat steps Place red cell products and mock product back into Helmer #4. 1
2 7. Secure the Traceable Alarm Thermometer probe (bottle) in its designated location (on the bottom of the cooler). Note: The bottled glycol solution containing the green temperature-buffered sensor must be touching the bottom of the thermolyser plate inside the cooler. The display portion of the Traceable Alarm Thermometer is placed outside of the cooler) 8. A Patient Identification Card must be properly filled out indicating the patient s first and last name, MRN and unit numbers in the bottom section of the Patient Identification Card. Green card cardiac cases Orange card non-cardiac cases 9. Refer to the procedure on Signing out of Blood and Blood Components, Volume III being sure to include the patient identification card when checking patient information in the Laboratory Information System. 10. Transfusion slips are preprinted with a box containing information for recording HemoTemp ll information. Note: During Downtime procedures, transfusion slips are available above the sign-out station for manual entry of patient and unit information. Each slip used during downtime has a label attached to each ply containing the Hemotemp II information for documentation of temperature and blue flower being present at the time of transfusion. On the back side of this slip is information in the event of a transfusion reaction. 11. At the time of issue, for the mock product and each unit, document the temperature range color of the flower, and cooler identification (C, D, E, Y, or Z) on the Patient Identification Card (see example below). Indicate the color of the flower with the first letter of the color. For example, B for blue. Note: For product to be acceptable for issue, the HemoTemp Indicator should read 1 o C -3 o C or 4 o C -6 o C and the flower should be blue B. 12. Record the temperature of the cooler s digital display, the minimum temperature and the maximum temperature on the Patient Identification Card along with the date, time and initials of the technologist packing the cooler (Temp Upon Issue). Note: Temperature of the cooler must be in the range of 1 o C-6 o C on the digital, minimum and maximum temperature readings in order to use for storage. The acceptability of the blood product to issue is based upon the following criteria: Reversible Numbers Temperature Range Irreversible Flower Acceptability Visible 1 o -3 o C or 4 o -6 o C Blue OK Visible 7 o C-9 o C Blue Not OK Visible 1 o -3 o C or 4 o -6 o C or 7 o -9 o C Green or Tan Not OK Invisible No visible range Green or Tan Not OK Invisible No visible range Blue Not OK 13. Immediately place the blood bags inside the cooler chamber in a vertical position, facing toward the front of the cooler. Unit bags must be sitting on the cold plate in an upright position to maintain appropriate storage temperature. The maximum number of units to be transported/stored in the cooler is four (4) units. 14. The mock product is then placed vertically against the cold sink, perpendicular to the units. The mock product serves to moderate the unit temperature in the cooler. 15. Secure the cooler cover. 16. For all cases using the orange and green Patient Identification Card, note the time the cooler must be returned in the appropriate location on the card. Secure the card to the cooler. Note: Time of return is determined by adding three hours to the time of issue. 17. For cardiac case (green card) coolers only, the Time Tracker must be started and sent with the cooler. Quick Start of Time Tracker: Press the MENU key. Use the arrow keys to select AUTO. Press the ENTER key to select. Press START 2
3 Operating Room Circulating RN will be responsible for: 1. The circulating RN will phone the Blood Bank to indicate the OR is ready to receive the HemoCool Jr.. The RN will dispatch the validated ORA to the Blood Bank to pick up the cooler by signing a Delivery Request Form indicating the cooler is needed NOW. Note: If staffing does not permit an ORA to obtain the cooler, a Blood Bank tech will transport the HemoCool Jr. to the OR suite where it will be received by designated OR personnel at the entrance of the OR suite. 2. The ORA will be responsible for immediately returning to the designated OR room with the cooler and plugging the cooler into a designated wall outlet. Note: If the cooler is left unplugged or the cover is not securely fashioned, the HemoCool Jr. will maintain a digital temperature display of 1 o C -6 o C for a maximum of 30 minutes. 3. The circulating RN will sign the Blood Product Delivery Request form as the receiving RN and document time received. 4. On the line: Received in OR: Record the following: a. Temperature on the digital display of the thermometer as well as the minimum temperature and maximum temperature. Note: If the temperature of the digital display is greater than 6 o C or less than 1 o C, the Alarm Thermometer will alert the OR personnel with a visual and audible signal even if the temperature returns to a non-alarmable range (1 o C 6 o C). The circulating RN will notify the Blood Bank by phone and the ORA will be responsible for returning the cooler with the blood products back to the Blood Bank immediately. Additional product will be issued and transported to the OR by an ORA or Blood Bank tech following the above procedure. It will be the responsibility of the Blood Bank Technologist to determine acceptability of the red cell products returned. b. Date and time. c. The RN will compare the name and MRN on the Patient s Identification Card with the patient s wristband. The RN will initial and indicate by circling under the Comment section: Ok if Patient Identification Card matches the patient s wristband. Not Ok if Patient Identification Card does not match the patient s wristband. Notify the Blood Bank and return products immediately. The Patient Identification Card must remain with the cooler at all times to ensure proper identification of the cooler and content of red cell products inside. Anesthesia(CRNA or anesthesiologist/perfusion (perfusionist) is responsible for: 1. At the time of transfusion, the unit will be removed from the properly identified storage cooler. 2. If it has been determined that the product will not be transfused, the red cell product must be immediately returned to the cooler. 3. Refer to the Blood Administration policy for additional information in transfusing blood products. At the end of a case or at three hours from time of issue, the Circulating RN will be responsible for: 1. Coordinating and ensuring the Hemo Cool Jr storage container is returned to the Blood Bank. 2. If the cooler is still needed in the room after three hours, the circulating RN will sign the Blood Product Delivery Request form requesting another cooler with designated number of units up to four. Note: The Time Tracker sent with a cardiac cooler only, is set to alarm prior to the three hour time limit. This will alarm and stay lit at the following intervals: Green section - 2 hrs and 24 minutes Yellow section - 2 hrs and 33 minutes Red section 2 hrs 51 minutes at which time the alarm will say Time s Up 3
4 If it has been deemed necessary to have the cooler containing red cell products follow the patient to the ICU, the anesthesiologist/surgeon or designee will: 1. Contact the Blood Bank at 3720 to inform the Blood Bank. The Circulating RN will be responsible for: 1. Coordinating and ensuring the Hemo Cool Jr storage container is returned to the Blood Bank as well as the Time Tracker for the Cardiac Cases. 2. The Blood Product Delivery Request form is completed requesting a cooler with designated number of units up to four to be sent to ICU. The Blood Bank Technologist will be responsible for: 1. Determining product acceptability based on the criteria noted below. 2. For all cases, note the time of 0630 AM on the following day in the appropriate location on the card. Secure the card to the cooler. 3. Issue products. 4. Arrange to transport the cooler to the ICU, plugging the cooler into a designated wall outlet and communicating with the ICU RN that the cooler is in the room and needed for the patient 5. Documenting on the Patient Identification Card- on the line: Received in ICU: a. Temperature on the digital display of the thermometer as well as the minimum temperature and maximum temperature. Note: If the temperature of the digital display is greater than 6 o C or less than 1 o C, the Alarm Thermometer will alert the technologist. The tech ICU personnel with a visual and audible signal even if the temperature returns to a nonalarmable range (1 o C 6 o C). The ICU RN will notify the Blood Bank by phone and the Blood Bank will be responsible for removing the alarming cooler with a replacement cooler. b. Date and time. c. The technologist will compare the name and MRN on the card with the patient s wristband. The technologist will initial and indicate by circling under the Comment section: Ok if Patient Identification Card matches the patient s wristband. Not Ok if Patient Identification Card does not match the patient s wristband. Notify the Blood Bank and return products immediately. The Patient Identification Card must remain with the cooler at all times to ensure proper identification of the cooler and content of red cell products inside. ICU RN is responsible for: 1. At the time of transfusion, the unit will be removed from the properly identified storage cooler by the patient s RN. 2. The following information is immediately recorded on the attached transfusion slip. Temperature range of the HemoTemp Indicator strip. Blue color is noted on the transfusion slip as Yes or No. The acceptability of the blood product for transfusion is based upon the following criteria: Reversible Numbers Temperature Range Irreversible Flower Acceptability Visible 1 o -3 o C or 4 o -6 o C Blue (Yes) OK Visible 7 o C-9 o C Blue (Yes) Not OK Visible 1 o -3 o C or 4 o -6 o C or 7 o -9 o C Green or Tan (No) Not OK Invisible No visible range Green or Tan (No) Not OK Invisible No visible range Blue (Yes) Not OK Refer to the Blood Administration policy for additional information in transfusing blood products. 3. If the products are used, the Blood Bank can be notified by the ICU nurse and the HemoCool Jr will be returned to the Blood Bank to ensure the continued availability of products. 4. If the temperature of the digital display is greater than 6 o C or less than 1 o C, the Alarm Thermometer will alert the ICU nurse with a visual and audible signal even if the temperature returns to a non-alarm able range of (1 o C 6 o C). 4
5 The ICU nurse will notify the Blood Bank and the ICU PCA or Blood Bank Technologist will be responsible for returning the blood products to the Blood Bank immediately. Two (2)-four (4) additional units will be made available for the patient and transported to the ICU by the Blood Bank Technologist following the above procedure. Note: It will be the responsibility of the Blood Bank Technologist to determine acceptability of red cell products if the Alarm Thermometer alerts the ICU nurse of the HemoCool storage unit reaching an alarmable range of less than 1 0 C and greater than 6 0 C. 5. The HemoCool Jr. can remain in the patient s room until 0630 on the next day. Blood Bank Technologist: 1. If the cooler has followed the patient to the ICU, it will be the responsibility of the Blood Bank Technologist to pick up the cooler in the ICU at 06:30 AM unless other arrangements for pickup have been made. 2. At the time of pickup, the Blood Bank Technologist will record the temperature of the digital display, the minimum temperature and the maximum temperature, date, time and initials on the Patient Identification Card (Pickup in ICU). 3. Upon receipt in the Blood Bank, the temperature of the current digital display, the minimum temperature and the maximum temperature, date, time and initials upon receipt into the Blood Bank (Received in BB) will be recorded. 4. Visual inspection of the units will be performed. (Refer to the procedure Returning Components to the Blood Bank (Volume III Components) References: Acceptability of the red cells for reissue is based upon the maintained temperature of the cooler over 24 hours, the HemoTemp Indicator strip, and visual inspection criteria. Acceptability of Red Cells for Reissue (Note: Contingent on acceptable visual inspection of unit) Indicator Reading Digital, Minimum and Maximum Temperature Temp Flower <1 o C 1 o C - 6 o C >6 o C 1 o C - 3 o C (or) 4 o C - 6 o C Blue OK for Reissue OK for Reissue OK for Reissue 1 o C - 3 o C (or) 4 o C - 6 o C Green or Tan 7 o C - 9 o C Blue Green or Tan Not OK for Reissue Not OK for Reissue Not OK for Reissue No Visible Range Blue Record Yes or No in the Okay to Reissue? Section of the Patient Identification Card if the Alarm Thermometer is acceptable and the HemoTemp Indicator Strip is acceptable. 5. Follow-up, utilizing the RL6 Electronic Incident reporting, will occur if the cooler is returned after 30 minutes from the designated time noted on the cooler card. Technical Manual, American Association of Blood Banks, 17 th edition, 2011, ,
Laboratory Preparedness Checklists for Advanced Planning, Preparing to Close & Safe Re-Entry
Laboratory Preparedness Checklists for Advanced Planning, Preparing to Close & Safe Re-Entry Disruptions, utility outages, and severe weather events may require a laboratory closure. These events have
More informationEndo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax
OPERATION MANUAL Zutron Medical, LLC 17501 W 98 th St #40-27 Lenexa, KS 66219 Phone 877-343-5873 Fax 913-967-5944 ZUT-Lab-004-30004 REV. 03312017 Table of Contents 2 Introduction 1. Intended Use 2. Labels,
More informationENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL
1 of 7 The quality assurance plan as presented in the Quality Manual should assure that constant and consistent test conditions are met and verified and should be tailored to the laboratory s activities.
More informationUser Manual 1 P a g e Rev. V1.6-EN 11/08/2014
User Manual 1 P a g e Rev. V1.6-EN 11/08/2014 Copyright Disclaimer Trademarks and patents Intended use Contact info 2011 Inflotrolix, Inc. This document may not be copied in whole or in part or otherwise
More informationNatural Gas to L.P. Gas Conversion Kit
Natural Gas to L.P. Gas Conversion Kit For Bosch 80% AFUE Gas Furnace, BGS80 Model Installation Instructions 3124627 2 Natural Gas to L.P. Gas Conversion Kit Installation Instructions Data subject to change
More informationI. Subject: Medical Gas Systems Utility Failure Interim Procedure III. Purpose:
I. Subject: Medical Gas Systems Utility Failure Interim Procedure II. Purpose: To provide a systematic response to medical gas system failure and restoration of this critical utility. III. Procedure: Systems
More informationFace of the Fume Hood. Must work 6 Inches Back from Face of Hood. Horizontal Sash Configuration
S a s h This Fume Hood has a 6 Sash Configuration, some may have 8 Yellow EHS tag Digital Monitor Face of the Fume Hood Must work 6 nches Back from Face of Hood Horizontal Sash Configuration ertical Sash
More informationYellow EHS Tag. Face of the Fume Hood. Must work 6 Inches Back from Face of Hood
Sash Digital Monitor Face of the Fume Hood Yellow EHS Tag Must work 6 nches Back from Face of Hood Orange Failed Sticker The Maximum Height for Sash EHS Yellow Tag Do Not Obstruct Airflow PROPER USE OF
More informationAlfred ECMO Circuit Priming and Storage Guide
Alfred ECMO Circuit Priming and Storage Guide 1 March 2012 Check List Page 2. Assembly, Priming and Storage Page 3. Prepare ECMO trolley and Circuit immediately before use Page 6. Connection to Patient
More informationpatient education program 4156 South 52 nd Street, Omaha, NE ChildrensOmaha.org/HomeHealthcare
patient education program 4156 South 52 nd Street, Omaha, NE 68117 800-747-7334 402-734-6741 ChildrensOmaha.org/HomeHealthcare We know children CHILDREN S HOME HEALTHCARE Trilogy Ventilator Instructions
More informationPRE-TRANSFUSION GUIDELINES
PRE-TRANSFUSION GUIDELINES TACTICAL COMBAT CASUALTY CARE PATIENT STABILIZATION REQUIREMENTS HEMORRHAGE Bleeding is CONTROLLED by use of pressure dressing, tourniquet or fibrin bandage AIRWAY Secured and
More informationFace of the Fume Hood. Must work 6 Inches Back from Face of Hood
S a s h This Fume Hood has a 6 Sash Configuration, some may have 8 Yellow EHS tag Digital Monitor Face of the Fume Hood Must work 6 nches Back from Face of Hood EHS Yellow Tag Orange Failed Sticker Do
More informationUSER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website:
USER MANUAL Model: PM4300 Series SAVE THESE INSTRUCTIONS For the most current manual revision, please visit our Website: www.precisionmedical.com 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA
More informationUSER MANUAL SAVE THESE INSTRUCTIONS. 300 Held Drive Tel: (+001) Northampton, PA USA Fax: (+001) ISO Certified
USER MANUAL Model: PM4300 Series (PM4351 Shown) SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090 Northampton,
More informationMA PHIT- Food Certificate Program
Proper inspection equipment and the knowledge of how to use and care for the equipment is necessary for an inspector to be able to identify food safety hazards which could lead to foodborne illness. Day
More informationwith O 2 Controller Instruction Manual
14500 Coy Drive, Grass Lake, Michigan 49240 734-475-2200 E-mail: sales@coylab.com www.coylab.com Hypoxic Cabinets (In-Vivo / In-Vitro) with O 2 Controller Instruction Manual Index Page Warranty 2 Warnings
More informationNatural Gas to L.P. Gas Conversion Kit
Natural Gas to L.P. Gas Conversion Kit For Bosch 96% AFUE Gas Furnace, BGH96 Model Installation Instructions 3124627 2 Natural Gas to L.P. Gas Conversion Kit Installation Instructions Data subject to change
More informationPREPARED BY: Marshall K. Cheung, Ph.D., Laboratory Director. REVISED BY: Marshall K. Cheung, Ph.D., Laboratory Director
DOCUMENT TYPE: DOCUMENT CLASS: Standard Operating Procedure Physical Property Procedure TITLE: Conductivity, EPA 120.1 INSTRUMENTATON: HACH CO150 Conductivity Meter PREPARED BY: Marshall K. Cheung, Ph.D.,
More information2017 DirtFish SCCA RallyCross National Challenge and Championship
Below are the general Supplemental Regulations for all RallyCross National events. The organizers for each event will add more event specifics such as schedule, run/work order, etc. and it will be posted
More informationANNUAL SELF INSPECTION / SERVICING INSTRUCTION SHEET
ANNUAL SELF INSPECTION / SERVICING INSTRUCTION SHEET These instructions relate ONLY to the following models of inflatable PFD s: Watersnake Inflatable PFD & Watersnake Deluxe PFD (ADULT AND CHILDRENS SIZES)
More informationOXYGEN MINERALE ABSOLUTE RANGE OWNER S MANUAL OXY MINERALE 600/1000 SERIES
PURAPOOL OXYGEN MINERALE ABSOLUTE RANGE OWNER S MANUAL OXY MINERALE 600/1000 SERIES 1. PARTS IDENTIFICATION LIST 1. 5. 2. 4. 3. 6. 6. Figure 1.1 1. OXYGEN Minerale Power pack with timer option (x1) 2.
More informationU S E R M A N U A L. Model: PM4300 Series SAVE THESE INSTRUCTIONS
U S E R M A N U A L Model: PM4300 Series (PM4351 Shown) SAVE THESE INSTRUCTIONS CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090
More informationSTANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA
STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA SOP#: 1157.1 Date Issued: 05/14 Date Revised: 5/15 Page 1 of 6 TITLE: SCOPE: RESPONSIBILITY: PURPOSE: SurgiVet
More informationUNH Department of Forensic Sciences
UNH Department of Forensic Sciences LABORATORY SAFETY RULES, PRACTICES AND AGREEMENT (Professors, Graduate Students, Research Assistants, Teaching Assistants, Students Working on Research Projects and
More informationFAULT CODE TROUBLESHOOTING INDEX
FAULT CODE TROUBLESHOOTING INDEX 1. Display indicates Change Filters 2. Display indicates Drip Tray Full Continuous Alarm will Sound 3. Display indicates Cold Fault 4. Display indicates Hot Fault 5. Display
More informationJETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012
JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research and Teaching Facility TABLE OF CONTENTS 1. Introduction....2 1.1 Scope of Work.....2
More informationSERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL
www.modsafe.com SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the service manual enclosed with this device.
More information20. Technical Equipment. Archived. This section applies to all equipment used in forensic examinations, such as
Instruments and Equipment Introduction 5.5. Authorization to use technical equipment 5.5.3 Use of instruments and equipment Instruction and maintenance manuals 5.4. 5.5.3 Use of nonlaboratory equipment
More informationEnvironmental Health & Safety
Environmental Health & Safety Oklahoma State University AUTOMATED EXTERNAL DEFIBRILLATOR (AED) MANAGEMENT PROGRAM Environmental Health and Safety University Health Services Suite 002 Stillwater, OK 74078
More informationVet-Temp Ear Thermometer Frequently Asked Questions
1) What is the price of the Vet-Temp? Vet-Temp Ear Thermometer Frequently Asked Questions The Vet-Temp thermometer (VT-150) is $259.00 and a case of the disposable probe covers (DPC- 500) is $64.00. The
More informationINSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor
INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor M A R T E C H S E R V I C E S C O M P A N Y P.O. Box 7079 OFFICE: (507)
More information1.4 SWIMMER BRIEFING See also: RULES
1.4 SWIMMER BRIEFING See also: RULES Key Role Pre race instructions to the swimmers cover informational and safety points. Relevant Rules USMS Open Water Swimming Manual, section 304. The Race Director
More informationINSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y
INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model 50 Systems Outfitted with ABM-725 Monitor M A R T E C H S E R V I C E S C O M P A N Y OFFICE: (507) 843-4700 P.O. BOX 7079 Toll
More informationPERSONAL AIR BREATHING UNIT
PERSONAL AIR BREATHING UNIT Operation & Maintenance Manual MARTECH SERVICES C O M P A N Y 1-800-831-1525 Table of Contents INTRODUCTION: Page 2 COMPONENTS DRAWING: Page 3 START UP: Page 4 OPERATION: Page
More informationConcentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST Industrial Drive, Orlinda, TN 37141
0 Operation Manual Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST 615-654-4441 sales@specialtyh2o.com 615-654-4449 fax TABLE OF CONTENTS Section 1 GENERAL 1.1 Warnings and
More informationTYPICAL TOXIC GAS SYSTEM INSPECTION VIOLATIONS
TYPICAL TOXIC GAS SYSTEM INSPECTION VIOLATIONS The following is a list of typical violations often found by inspectors and a generic solution. You can use this list to improve the safety of your facility,
More informationOxygen Users Disaster Evacuation Planning Guide
March 2015 Edition Oxygen Users Disaster Evacuation Planning Guide Provided Compliments of: No Person Left Behind www.oxygen.nopersonleftbehind.org (239) 368-6846 Updated March 19, 2015 Page 1 of 13 Create
More informationBlocks must always run perpendicular (at 90 degrees) to the fence to ensure adequate stability.
This information paper is to provide minimum guidelines and information of the erection of temporary fencing for the purpose of barricading, delineation of a work area, and for protection of workers from
More informationCommon Discrepancies Uncovered on ADCI Diving Contractor & Training Program Audits
Common Discrepancies Uncovered on ADCI Diving Contractor & Training Program Audits Ways to Set Your Company or School Up For Success on ADCI Audits and Reduce Costs In most instances, all discrepancies
More informationOxygen Users Disaster Evacuation Planning Guide
May 2012 Edition Oxygen Users Disaster Evacuation Planning Guide Provided Compliments of: No Person Left Behind www.nopersonleftbehind.org (239) 368-6846 Updated June 1, 2012 Written by Linda R Carter
More information1020 Industrial Drive, Orlinda, TN fax
Operation Manual Tank Distribution System A-UPT Series 615-654-4441 sales@specialtyh2o.com 615-654-4449 fax TABLE OF CONTENTS Section 1 GENERAL 1.2 Warnings and Cautions... 1 1.2 Theory of Operation...
More informationOPERATOR S MANUAL Ar-Gone Weld Gas Analyzer
July 2011 OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer WARNING! Before operating this product, read and understand this Operator s Manual. Become familiar with the potential hazards of this unit. Contact
More informationBiology Unit 2, Structure of Life, Lab Activity 2-3
Biology Unit 2, Structure of Life, Lab Activity 2-3 Cellular respiration is the release of energy from organic compounds by metabolic chemical oxidation in the mitochondria within each cell. Cellular respiration
More informationQuick Reference Guide
HematologyAnalyzer Quick Reference Guide For Veterinary Use Only Running a Blank Blanks must be run every 12 hours, after a Soak Cleaning has been performed, after a reagent pack change, and after the
More informationRADIATION PROCEDURES MANUAL Procedure Cover Sheet
RADIATION PROCEDURES MANUAL Procedure Cover Sheet Procedure Title: Calibration of Eberline Portable Particulate Noble Gas Monitor Procedure Number: TSO-08-13-REV 1 Effective Date: July 1, 2008 Approved
More informationInstallation Instructions
Model: ACT-4100 ACT-4100-3 ACT-4100-6 SUN ELECTRIC CORPORATION Page: 1 of 5 Installation Instructions INSTALLATION MUST BE PERFORMED BY QUALIFIED SUN PERSONNEL ONLY INSTALLATION OVERVIEW: The Installation
More informationInstructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems
Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the
More informationInstallation Instructions
Model: ACT-4500 ACT-4500-3 ACT-4500-6 SUN ELECTRIC CORPORATION Page: 1 of 5 Installation Instructions INSTALLATION MUST BE PERFORMED BY QUALIFIED SUN PERSONNEL ONLY INSTALLATION OVERVIEW: The Installation
More informationScoreboard Operator s Instructions MPC Control
Scoreboard Operator s Instructions MPC Control Some features on the keyboard overlay may not be included on the particular model being operated. Since 1934 Retain this manual in your permanent files 1/21/2011
More informationRisk Assessment for: Roaches Pyrotec dyeing machines
School of Design Faculty of Performance, Visual Arts and Communications Risk Assessment for: Roaches Pyrotec dyeing machines Assessment No: 002 PROCESS / ACTIVITY LOCATION ASSESSED BY DATE(S) Small scale
More informationInstallation Instructions
SUN ELECTRIC CORPORATION Model: MRC-450 MRC-450-3 MRC-450-6 MRC-450-GR Page: 1 of 5 Installation Instructions INSTALLATION MUST BE PERFORMED BY QUALIFIED SUN PERSONNEL ONLY INSTALLATION OVERVIEW: The Installation
More informationINSTALLATION & OPERATION MANUAL. Dual Duct VAV Terminals. Redefine your comfort zone.
INSTALLATION & OPERATION MANUAL Dual Duct VAV Terminals IOM Dual Duct VAV Terminals IMPORTANT! READ BEFORE PROCEEDING! GENERAL SAFETY GUIDELINES This equipment is a relatively complicated apparatus. During
More informationBrevard County (Florida) Administrative Procedures
Page 1 of 8 Brevard County (Florida) Administrative Procedures 8452 - AUTOMATED EXTERNAL DEFIBRILLATOR Mission Statement The mission of the District is to plan and implement a sustainable Automated External
More informationBIOGAS COMBUSTION PLANT MAINTENANCE MANUAL
BIOGAS COMBUSTION PLANT MAINTENANCE MANUAL (PLANT MAINTENANCE MANUAL) Puente Gallego Landfill, Rosario (Argentina) ARIA.BIZ S.A. Montevideo 589, 7 floor 1019 Capital federal Argentina Tel: +54 11 5171
More informationInstruction Manual for Configura Cushionair Portable Pump
Instruction Manual for Configura Cushionair Portable Pump Fitted with battery powered pump, suitable for Configura Portable chairs V E R S I O N O N E M A Y 2 0 1 6 Contents Introduction 3 Set up of Cushionair
More informationInstalling ASI Umbilical Delivery Systems
Rev A 05 Nov 2015 Page 1 of 10 PURPOSE This guide provides an overview of tubing connections into an ASI wall-mounted Junction Box that has already been plumbed with a compressed air connection including
More informationMunicipal Boost w/pre-treatment Monitoring System K-A-MCB-PRE
Operation Manual Municipal Boost w/pre-treatment Monitoring System K-A-MCB-PRE 615-654-4441 sales@specialtyh2o.com 615-654-4449 fax TABLE OF CONTENTS Section 1 GENERAL 1.1 Warnings and Cautions... 1 1.2
More informationCircuFlow Quick Setup Guide
CircuFlow 5200 Quick Setup Guide LB04.0001 Rev B 20140314 TABLE OF CONTENTS Setup Therapy Setup Gradient Mode Pressure Mode Lock Device Unlock Device Treatment 03 04 05 08 11 12 13 02 SETUP 01 01 Open
More informationTRANSCUTANEOUS CO2 MONITORING (TCpCO2) PURPOSE
PURPOSE tcpco2 monitoring provides continuous non invasive monitoring of an infant s ventilatory status. The sensor consists of a Stow-Severinghaus-type electrode. The use of tcpco2 monitoring may facilitate
More informationMANITOBA RENAL PROGRAM
MANITOBA RENAL PROGRAM SUBJECT Fresenius AquaUNO Reverse Osmosis System; use of SECTION CODE 30.30.04 30.30 Water Treatment AUTHORIZATION Professional Advisory Committee, Manitoba Renal Program Nursing
More informationINSPIRE TM Inspired choice
Inspired choice Personalized perfusion because no two of your patients are the same Leveraging worldwide leadership Our vision becomes reality LivaNova is committed to providing solutions that advance
More informationSTANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.
STANDARD OPERATING PROCEDURES Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.5 ) Sampler (RAAS) AMBIENT AIR MONITORING PROGRAM for the 130 LIBERTY STREET DECONSTRUCTION
More informationGV Standard X-Vent. Setup, Commissioning & Installation Guide
GV Standard X-Vent Setup, Commissioning & Installation Guide Technical experts in the design, manufacture and supply of precision engineered, architectural rooflights for residential and commercial buildings.
More informationSERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL
Product Manual MST Portable Air Filtration System SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the
More informationMANITOBA RENAL PROGRAM
MANITOBA RENAL PROGRAM SUBJECT Fresenius 5008 Preparation for Hemodialysis using the ONLINEplus System SECTION CODE 30.10.02 30.10 Hemodialysis: Equipment and Procedures AUTHORIZATION Professional Advisory
More informationBARBECUE CONTEST 2018 GUIDELINES BARBECUE COMMITTEE 24 TH ANNUAL JUNE 2018 SATURDAY. Jenavee Barber PSSI Co-Chairman
PEACE 24 TH ANNUAL BARBECUE CONTEST JUNE 2018 16 SATURDAY BARBECUE COMMITTEE Mike Prpich Tyson Foods, Inc. Chairman Jenavee Barber PSSI Co-Chairman 2018 GUIDELINES CONTEST GUIDELINES 2018 2018 LOCATION
More informationTechnical Procedure for General Laboratory Equipment
Technical Procedure for General Laboratory Equipment 1.0 Purpose - This procedure specifies the required elements for the use of general laboratory equipment. 2.0 Scope - This procedure applies to all
More informationRelative Humidity Calibration Kit
Relative Humidity Calibration Kit For calibrating the following Kestrel Meters Model Numbers 3000, 3500, 4000, 4200, 4250, 4300, 4400, 4500, 4600 (Not to be used for calibrating Kestrel 5 Series units
More informationAny tool, at the ready. 3M DBI-SALA Fall Protection for Tools Pocket Reference Guide
Any tool, at the ready 3M DBI-SALA Fall Protection for Tools Pocket Reference Guide 3M DBI-SALA Fall Protection for Tools Product Catalog Download/Request a Printed Catalog Download the product catalog
More informationInstallation and Operation Manual
Installation and Operation Manual WIKA FLR-SBDF / BLR-SBDF Magnetic Level Transmitter (Please retain for future usage) Contact: Gayesco-WIKA USA, L.P. 229 Beltway Green Boulevard Pasadena, TX 77503 www.wika.com
More informationMisaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii
Operator s Manual Table of Contents Operator Safety... 1 Introduction... 2 Unpacking and Setup... 3 Unpacking... 3 Setup... 4 FLYER Overview... 5 FLYER Diagram... 5 Capabilities... 5 Control Panel... 6
More informationHow Ventilators Work. Chapter 3
How Ventilators Work Chapter 3 To care for a ventilator patient, you need to know: The various functions of the ventilator used How the ventilator interacts with the patient How changes in lung condition
More informationINSTRUCTION MANUAL. Includes: OPERATION SETUP USE & CARE SERVICE. SPORTS ATTACK LLC
Electronic Baseball Pitching Machine baseball INSTRUCTION MANUAL PART NO. B-EHABA-100 REV022111 INSTRUCTION MANUAL Includes: OPERATION SETUP USE & CARE SERVICE SPORTS ATTACK LLC 800-717-4251 www.sportsattack.com
More informationHZ-300 HAZER MACHINE USER MANUAL. Congratulations on the purchase of your new ANTARI HZ-300 Haze Machine.
HZ-300 HAZER MACHINE USER MANUAL Congratulations on the purchase of your new ANTARI HZ-300 Haze Machine. Read and save these instructions 2005 Antari Lighting and Effects Ltd 2 User Manual Antari HZ-300
More informationOperation Manual. H & R Healthcare 1750 Oak Street Lakewood, NJ Phone: Fax:
Operation Manual H & R Healthcare 1750 Oak Street Lakewood, NJ 08701 Phone: 800-801-5533 Fax: 732-367-6231 www.handrhealthcare.com Congratulations and thank you for purchasing this Relief Aire Alternating
More informationKonza District Pinewood Derby Track Parts And Assembly Instructions
Konza District Pinewood Derby Track Parts And Assembly Instructions - 1 - Pinewood Derby Track Parts List: Box 1 of 3 2 each - Race Track Flat Sections 1 each - Race Track Finish Line Section 1 each -
More informationSimrad GD10P Extended On-site Maintenance
Simrad Extended On-site Maintenance This procedure covers maintenance of gas detectors with the aid of a handheld terminal (re-zeroing tool). Please also refer to the standard operating manual (P3359E).
More informationCONTROL OF HAZARDOUS ENERGY (LOCKOUT/TAG OUT PROGRAM) Washington State University Tri-Cities (all departments)
CONTROL OF HAZARDOUS ENERGY (LOCKOUT/TAG OUT PROGRAM) Washington State University Tri-Cities (all departments) PURPOSE This program establishes the general and specific requirements for the control of
More informationInstruction and Practice Manual
Instruction and Practice Manual Axiom Sports Manufacturing A Division of Fitec International 3525 Ridge Meadow Parkway Memphis, TN 38175 (901) 366-9144 www.axiomsports.com 1 Thank you for purchasing the
More informationDigital Tourniquets. Product Details. For electricity VAC, 50/60 Hz
Digital Tourniquets For electricity 100-240 VAC, 50/60 Hz Product Details Digital Display Large color coded LED display is easy to read Precise monitoring shows actual cuff pressure Precision Pressure
More informationEngineering World Health Youth EWH Design Projects That Matter
Engineering World Health Youth EWH Design Projects That Matter Engineering World Health is an extraordinary fusion of engineers, scientists and physicians who donate their time and talents to positively
More informationINSTALLATION, OPERATION AND SERVICE MANUAL ABS AIR BAG LIFT
INSTALLATION, OPERATION AND SERVICE MANUAL ABS AIR BAG LIFT P.O. Box 1058 1058 West Industrial Avenue Guthrie, OK 73044-1058 405-282-5200 FAX: 405-282-8105 www.autoquip.com Item # 830ABS Version 1.0 07/2001
More informationLockout Tagout Policy
Office of Environmental Health & Safety www.moreheadstate.edu/ehs 606-783-2584 Lockout Tagout Policy PURPOSE To establish procedures for the de-energization and isolation of energy sources or the lockout
More informationOUTDOOR LED SCOREBOARD CARE & OPERATIONS MANUAL
OUTDOOR LED SCOREBOARD CARE & OPERATIONS MANUAL 98 0000 11 REV 11 19 08 DISCLAIMER This publication is not a service manual. There are no user serviceable components. If you require service, please contact
More informationWarner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax
DH-40i, Rev. 04.03.02 Micro-Incubation Platform For RC-40 Chambers Model DH-40i 1125 Dixwell Avenue, Hamden, CT 06514 (800) 599-4203 / (203) 776-0664 (203) 776-1278 - fax DH-40i, Rev. 04.03.02 Table of
More informationAQUARIUS Continuous Renal Replacement Therapy with Regional Citrate Anticoagulation (RCA) Competency
Demonstrates an understanding of why RCA has been prescribed Explain the anticoagulation protocol Demonstrate an understanding of potential complications associated with CRRT. Explain potential indications
More informationCover Page for Lab Report Group Portion. Head Losses in Pipes
Cover Page for Lab Report Group Portion Head Losses in Pipes Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 February 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section
More informationAutomated External Defibrillator Program. Developed by: The Office of Environmental Health and Safety
Automated External Defibrillator Program Developed by: The Office of Environmental Health and Safety Revised July 2018 I. Purpose and Scope This document serves to set forth responsibilities, inspections,
More informationRisk Assessment for: Fume cupboards
School of Design Faculty of Performance, Visual Arts and Communications Risk Assessment for: Fume cupboards Assessment No: 117 PROCESS / ACTIVITY LOCATION ASSESSED BY DATE(S) Carrying out activities that
More informationSIDEWINDER DRILLING RIG 122 CLEBURNE COUNTY, ARKANSAS
SIDEWINDER DRILLING RIG 122 CLEBURNE COUNTY, ARKANSAS March, 2015 Infrared Inspection Report Sidewinder Drilling Rig 122 Company Contact Jesse Russell Project: 07090315 2605 CHARTER OAK DRIVE, LITTLE ROCK,
More informationInstruction manual. for. high purity gas pigtails 200 bar / 300 bar
Instruction manual for high purity gas pigtails 200 bar / 300 bar Contents Contents... 2 1. Preface... 3 1.1 Overview... 3 1.2 General... 3 1.3 Intended use... 4 1.4 Personnel requirements... 4 2. For
More informationORB-400 BUBBLE/HAZE MACHINE. Item ref: UK User Manual
ORB-400 BUBBLE/HAZE MACHINE Item ref: 160.462UK User Manual Caution: Please read this manual carefully before operating Damage caused by misuse is not covered by the warranty Introduction Thank you for
More informationAutomated External Defibrillator Program. Developed by: The Department of Environmental Health and Safety
Automated External Defibrillator Program Developed by: The Department of Environmental Health and Safety Revised May 2017 I. Purpose and Scope This document serves to set forth responsibilities, inspections,
More informationOCCUPATIONAL SAFETY AND ENVIRONMENTAL HEALTH GUIDELINE
OCCUPATIONAL SAFETY AND ENVIRONMENTAL HEALTH GUIDELINE Date: 04-22-2009 Revision: 02 Page: 1 of 6 SUMMARY: SCOPE: REFERENCE REGULATIONS: Employees using laboratory fume hoods shall observe all applicable
More informationPressure Ulcers ecourse
Pressure Ulcers ecourse Module 5.2: Repositioning & Support Surfaces Handout 2015 Steppingstones Partnership, Inc. / Learning Nurse: All Rights Reserved See: Terms of Use at LearningNurse.org for acceptable
More informationOXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:
OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle
More informationUNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008
UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method
More informationAuto EVAP Quick Start Guide
Auto EVAP Quick Start Guide Pg. 13 Note: Never directly hit or drop the leak detector on the CO2 sensor tip, this will cause the sensor to become damaged! Do not get Leak Seeker Solution directly on the
More informationMedical oxygen. LIV IQ Digital therapy device.
Medical oxygen. LIV IQ Digital therapy device. BOC: Living healthcare 02 Cylinder instructions Cylinder instructions 03 Therapy device features. LIV IQ giving you safe guidance during patient treatment.
More informationHPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1
HPICAL-15000 Operation & Data Logging Procedures Click spacebar to advance through slides 1 WARNING Always wear proper safety equipment when using high pressure equipment. Do not exceed 125 psi air pressure.
More information