Fiche de vie glider potame (Slocum #070)

Size: px
Start display at page:

Download "Fiche de vie glider potame (Slocum #070)"

Transcription

1 Fiche de vie glider potame (Slocum #070) Date Lieu Objet Document de référence Observations 12/01/07 LPO Brest Shipped by WRC via Nantes Airshipped from Boston / Paris / Nantes and then truck Nantes to Brest. Nantes is the closest airport with cargo facilities. Due to the box sizes, it can only go in a cargo. Open the box and visually inspect the glider and the box content. Nothing to note /01/07 LPO, Brest Power on the glider for the first time Compte rendu de préparation et déploiement d un planeur Spray et du Slocum profond du LPO en Méditerranée Occidentale dans le cadre de MERSEA WP3, T.Terre, P.Lherminier, February 2007 Activate the SIM Iridium card. Change the autoexec with glider name and phone number to call. Check functions : pump, GPS, Iridium calls. Check dockserver 28/01/07 LPO, Brest Ship to Mallorca, Imedea for Med Door to door shipping by special truck deployment 30/01/07 IMEDEA, Calanova Prepare and ballast Check CTD + oxygen sensor 31/01/07 IMEDEA, Calanova Check ballast for potame Verification of pool water characterisitics showed a significance diffrence with the previously assumed numbers. 01/02/07 Off Soller, Mallorca First try to pdeploy the glider but cancelled Deepy alone was deployed due to bad weather 02/02/07 Off Soller, Mallorca Deploy the glider but unsuccesfull. Decide to get it back to Brest for pressure tank tests. During the first try of the mission at 30 m, the glider never came back to the surface due to a bad functionning of the vehicle pressure sensor. Vaseline was introduced in the capillar for corrosion problem but it prevents the pressure to be correctly transmitted to the sensor. Back to shore with the glider. 04/02/07 IMEDEA, Calanova Prepare for shipping back to Brest 18/02/07 LPO, Brest Glider back to lab Door to door truck shipping with the same carrier 02/03/07 LPO, Brest Open and separate the 3 sections This is to change the vehicule pressure sensor capillar tube 1/5

2 05/03/07 LPO, Brest Dismount tube and check the pressure sensor 12/03/07 LPO, Brest Replace the capillar tube and check the overall pressure sensor circuit. Close the glider and put a vacuum at 7.4 inhg 13-16/03/07 LPO, Brest Check the vacuum is not decreasing. record Vacuum and internal temperature. Open the glider again to prepare the pressure tests by setting a cable for communication thru the pressure tank 19-20/03/07 LPO, Brest C. Jones visit to LPO Pressure tank tests while running the pump at 250 db and 1000 db Pool trials 12-20/04/07 OOB, Banyuls Preparation and deployment with N/O Nereis 17/04-18/06 Western Mediterranean Contribution to the convection monitoring with NOC and LOCEAN teams Report on Clayton Jones visit to Ifremer/LPO, T.Terre, march 2007 Compte-rendu de récupération d un planeur Spray et de préparation et déploiement du Slocum profond du LPO en Méditerranée Occidentale dans le cadre de Mersea WP3, T.Terre, mai 2007 Report on operations with Slocum Potane (unit #70) in Western Mediterranean Sea. T.Terre, July 2007 Vaseline clearly seen at the entrance of the pressure sensor. Check pressure variation with vacuum pump and nitrogen injection (adtest) Same vacuum pump and nitrogen use Vehicle pressure sensor OK Pump ran OK in presure tank Pool tests showed offset between vehicle pressure sensor and SBE CTD pressure sensor Pump was set OUT OF SERVICE due to an OUT OF DEADBAND error (later corrected by soft) Shipping door to door with Ifremer van Start in mission without DO sensor running. Set up Later by changing sbdlist.dat Diificulties to connect with Iridium : got an abort MS_ABORT_DEVICE_ERROR T18:43:57 18/05 Leak detect Abort; Recovery and check; Loose screws shortening the leak detect sensor! Redeploy on 19/04 Start in mission without DO sensor running. Set up later by changing sbdlist.dat Difficulties to connect with Iridium : got an abort 06/05 : MS_ABORT_DEVICE_ERROR T18:43:57 20/05 : Deep engine out of service OOD:oil_volume:OUT OF DEADBAND: M_su: C_su: change f_de_oil_vol_deadz_width from 40 to something bigger 10/06 : abort due to a bad syntwx in a mafile. 2/5

3 18/06/07 OOB, Banyuls Recovery Compte rendu de récupération d un planeur Slocum profond du LPO en Méditerranée Occidentale dans le cadre de MERSEA WP3, T.Terre, juin /09/07 LPO, Brest Pressure tank tests and batteries discharge Report on Iridium, pump test and batteries discharge after Medterranean Sea dployment During all the deployment low success in Iridium comms ~20 % low H-moment as first cause 2 nd cause could be turned air bladder? Questions raised : 1. Test of Iridum link 2. purge of deep engine return to WRC? 3. air bladder attitude, does oxygen sensor have an effect? 4. Cycles the pump at 1000 db in pressure tank to empty the lithium batteries 5. Change pressure hull from aluminium to composite? Run the pump for 140 cycles at 100bar with the batteries used during the experiment. Discharge the batteries on resistive loads 20/10-4/11 2 nd EGO Workshop and Glider School, Mallorca Potame send as instrument for the school Transport Brest-Mallorc-Brest by Ifremer van, driver TT /11/07 Shipping to USA Potame is sent back to WRC Air purge + switch to composite hull 26/11/07 Delivered to WRC Thanksgiving around! 21/12/07 Shipping WRC to LPO 15/01/08 Delivered to LPO List of upgrades : - composite hulls for FW/AFT sections - peek for CTD pressure sensor circuit - new CTD pressure transducer block - new vacuum plug - new vacuum tool - suppress pinger compense weight by lead bottle - revise altimeter (170 khz) block - change some components - new sofware version - full tests in Ashumet Pond 22-23/01/08 02/02/08 Shipping Brest-Toulon Compte-rendu de préparation et déploiement d un planeur Slocum profond du LPO en Méditerranée occidentale dans le cadre de MERSEA WP3., T.Terre, février 2008 Shipping Brest-Toulon on 22-23/01 by Ifremer van Banyuls on 26/01 3/5

4 23-25/01/08 Ifremer, Toulon Preparation for EGO deployment in gulf of Lion Training on air purge in the field 26/01/08 Shipping Toulon - Banyuls OOB, Banyuls End of preparation for deployment Shipping Toulon-Banyuls on 26/01 by Ifremer van 27/01/08-28/01/08 2 wires on connector 5 points i.e. cable for the optode are broken. No tool to redo the connector soldier the wires?! 29/01/08 Off shore Banyuls Deployment with Nereis II No problem except a false alarm with the altimeter. 1 CTD with 5 samples at 5 depths OOB, Banyuls Pilot the glider while PLH is away Dives to 300, 500, 800 and 1000 m. 30/01-01/02/08 29/01-07/04 Western Mediterranean Pilots PLH & TT Problems encountered - 04/02 abort device_error due to optode ; turn it off - 16/02 abort same_depth : landing on the bottom during the slope approach off Barcelona - 27/02 abort device_error : link to the pump but happens once and did nothing - 04/03 abort device_error : link to the pump send a command to reset a flag on Potame!put x_de_avg_oil_vol_err_on_descent /03 abort no_tickle : Potame suddenly too light cant dive; very bad waetaher, not able to connect for 4 days - 08/04 abort overdepth : still not able to dive change the pump volume to fix the case 27-28/03/08 Shipping Brest-Banyuls Compte-rendu de récupération de 2 planeurs Slocum profonds en Méditerranée occidentale dans le cadre de MERSEA WP3., T.Terre, février /03/08 OOB, Banuyls Recovery of Himilcon with Nereis II Recovery of Potame and Himilcon Problems with Himilcon - its 2 wings were broken - Heavy corrosion!! 4/5

5 07/04/08 OOB, Banyuls After a week, waiting for the good weather window, recovery of Potame with Nereis II 08/04/08 Shipping Banyuls Brest 18/04/08 LPO, Brest Pool test : deploy Potame together with an SBE25 to post calibrate the CTD. Oxygene and salinity samples. 10/11/08 Shipping to USA Need repair on : - Air purge - Optode connexion - Big connector Problem with Potame - Its 2 wings were broken I is the origin of the problem encountered 05-08/03/08 during the bad weather. - Corrosion on the nose Ifremer van Rapport de Stage J.Burdallet. Bend a pin on the big connector! 5/5

Deep-Arvor: a CTD & DO profiling float for Argo

Deep-Arvor: a CTD & DO profiling float for Argo Deep-Arvor: a CTD & DO profiling float for Argo Dr Emina Mamaca on behalf of S. LE RESTE, V. DUTREUIL, X. ANDRE, C. TRAUTMANN, T. BESCOND, AST-14, Wellington, New Zealand The first prototype of Deep-Arvor

More information

Copyright 2004 by the Thomas G. Faria Corporation, Uncasville CT No part of this publication may by reproduced in any form, in an electronic

Copyright 2004 by the Thomas G. Faria Corporation, Uncasville CT No part of this publication may by reproduced in any form, in an electronic Copyright 2004 by the Thomas G. Faria Corporation, Uncasville CT No part of this publication may by reproduced in any form, in an electronic retrieval system or otherwise, without the prior written permission

More information

Burner Management System DEMO Operating instructions

Burner Management System DEMO Operating instructions Burner Management System DEMO Operating instructions Burner Management System DEMO Operating Instructions Startup Summary - Normal startup is accomplished in four basic steps: 1. Leak Test a. Safety Valve

More information

New product release. Universal Rebreather Monitor (URBM) Three cell / Independent backup

New product release. Universal Rebreather Monitor (URBM) Three cell / Independent backup New product release Closed Circuit Research is pleased to announce the launch of our range of Universal Rebreather Monitors Universal Rebreather Monitor (URBM) Three cell / Independent backup Key features

More information

New product release. Universal Rebreather Monitor (URBM) Single O2 cell and Dual HP package

New product release. Universal Rebreather Monitor (URBM) Single O2 cell and Dual HP package New product release Closed Circuit Research is pleased to announce the launch of our range of Universal Rebreather Monitors Key features and benefits include: Universal Rebreather Monitor (URBM) Single

More information

FX134, FX1234, FX3030 Problem and Solutions

FX134, FX1234, FX3030 Problem and Solutions FX134, FX1234, FX3030 Problem and Solutions Unit Will Not Power Up Turn the power switch on. The unit LCD will display revision program and filter life within 3 to 5 seconds after turning on unit. If this

More information

Table of Contents Seabird 9Plus CTD Components and Overview... 2 CTD Deployment\Recovery Procedure... 7

Table of Contents Seabird 9Plus CTD Components and Overview... 2 CTD Deployment\Recovery Procedure... 7 Seabird CTD overview and Deployment Standard Operating Procedures (SOPs) Gordon Furey P&O Maritime Table of Contents Seabird 9Plus CTD Components and Overview... 2 CTD Deployment\Recovery Procedure...

More information

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet: OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle

More information

WEBB RESEARCH CORPORATION

WEBB RESEARCH CORPORATION Use spacebar or arrow keys to advance slide show WEBB RESEARCH CORPORATION Falmouth, MA APEX Profiler SLOCUM Glider Rev: 06/22/08 WEBB RESEARCH CORPORATION Specialists in neutrally buoyant drifters and

More information

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc. Advanced Calibration Designs, Inc. Calibration Gas Instrument INSTRUCTION MANUAL Release I www.goacd.com Instruction Manual Gas Generator Release I TABLE OF CONTENTS I. General Description Page 2 II. Start-Up

More information

Additions, Revisions, or Updates

Additions, Revisions, or Updates 1 3 01-12 SUBJECT DATE SPN 4334/FMI 2, 3, 4, 7 and SPN 4375/FMI 6 February 2012 Additions, Revisions, or Updates Publication Number / Title Platform Section Title Change SPN 4334/FMI 2 DDC-SVC-MAN-0084

More information

Water Samplers, Water Sampling with Internal Recording, Cabling, and Deployment

Water Samplers, Water Sampling with Internal Recording, Cabling, and Deployment Water Samplers, Water Sampling with Internal Recording, Cabling, and Deployment Water Samplers Care and feeding SeatermAF User interface for internally recording instruments firing water samplers with

More information

Troubleshooting Basics: Profiling CTDs

Troubleshooting Basics: Profiling CTDs Technical Note Troubleshooting Basics: Profiling CTDs Methods for Diagnosing Erratic and Unreasonable Data Greg Ikeda, October 2018 Introduction Any CTD or sensor may eventually exhibit data spikes and

More information

Additions, Revisions, or Updates

Additions, Revisions, or Updates 1 2 06-11 SUBJECT DATE SPN 4375/FMI 6 DEF Pump Supply Current High February 2011 Additions, Revisions, or Updates Publication Number / Title Platform Section Title Change DDC-SVC-MAN-0084 EPA10 DD SPN

More information

AcornVac Vacuum Plumbing Systems - Trouble Shooting Guide

AcornVac Vacuum Plumbing Systems - Trouble Shooting Guide AcornVac Vacuum Plumbing Systems - Trouble Shooting Guide 1. Accumulators Problem Accumulator is overflowing If Accumulator continues to overflow Correction 1. Push and hold the manual activation button

More information

Standard Operating Procedure #COE-SOP-0001 Chemical Fume Hood Operation

Standard Operating Procedure #COE-SOP-0001 Chemical Fume Hood Operation Standard Operating Procedure # Chemical Fume Hood Operation Facility: NMSU College of Engineering Laboratories Written by: Juanita Miller, Safety Specialist, miljgh@nmsu.edu (575)-646-1292 Scope: This

More information

LENNOX SLP98UHV DIAGNOSTIC CODES

LENNOX SLP98UHV DIAGNOSTIC CODES Code Status of Equipment Action required to clear and recover - Idle mode (Decimal blinks at 1 Hertz -- 0.5 second ON, 0.5 second OFF) A Cubic feet per minute (cfm) setting for indoor blower (1 second

More information

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards.

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards. 1. Scope Description of requirements and procedures for compact provers to be used as verification standards. 2. Reference documents Trade Metrology Act SANS1698 3. Policy A. BASIC REQUIREMENTS Compact

More information

Maintenance of the Electronic ( EPC ) Pressure Controllers on GCs

Maintenance of the Electronic ( EPC ) Pressure Controllers on GCs Gas flows in all GCs are controlled by Electronic Pressure Controllers ( EPC ). There may be up to 7 EPCs mounted in one GC depending on which combination of injectors and detectors are installed. Each

More information

PROVOR/ ARVOR ARGO profiling floats

PROVOR/ ARVOR ARGO profiling floats PROVOR/ ARVOR ARGO profiling floats Euroargo 24-25 june 2008 Patrice Brault Kannad A long partnership A family of floats 1st float MARVOR : multicycle Rafos float in 90-94 PROVOR for ARGO program in 99

More information

AFDXXX(X)AC Series Operators Manual Please read this manual thoroughly before attempting to operate your water maker.

AFDXXX(X)AC Series Operators Manual Please read this manual thoroughly before attempting to operate your water maker. AFDXXX(X)AC Series Operators Manual Please read this manual thoroughly before attempting to operate your water maker. E & O E Danger High Voltage AFDXXX(X) Series water makers operate on a 240vAC electricity

More information

OMNISPORT 2000 SWIMMING QUICK REFERENCE

OMNISPORT 2000 SWIMMING QUICK REFERENCE Before During After OMNISPORT 2000 SWIMMING QUICK REFERENCE 1. Confirm that the timer is reset. 2. Set the timer to the correct event and heat. The right LCD, line 2 displays E:(event) H:(heat) R:(round)

More information

Unisense In Situ UniAmp Connector System

Unisense In Situ UniAmp Connector System Unisense In Situ UniAmp Connector System Overview The Unisense In Situ UniAmp Connector system is designed for easy replacement of sensors. Unisense In Situ UniAmp Connector system is a dedicated system

More information

Periodical surveys of cargo installations on ships carrying liquefied gases in bulk

Periodical surveys of cargo installations on ships carrying liquefied gases in bulk (June 1999) (Rev.1 Mar 2006) (Rev.2 May 2007) (Rev.3 Mar 2010) (Corr.1 Feb 2011) (Rev.4 Oct 2013) Periodical surveys of cargo installations on ships carrying liquefied gases in bulk 1 General 1.1 Scope

More information

CO 2. (R744) Service Station Presentation. Contents. Goals Basic functions Specification Recovery R477 Vacuum Charge Safety Other features Prototype

CO 2. (R744) Service Station Presentation. Contents. Goals Basic functions Specification Recovery R477 Vacuum Charge Safety Other features Prototype (R744) Service Station Presentation Contents Goals Basic functions Specification Recovery R477 Vacuum Charge Safety Other features Prototype Page 1 (R744) Service Station Goals Usability Full automatic

More information

Q2: Which instruments are necessary for monitoring the groundwater level?

Q2: Which instruments are necessary for monitoring the groundwater level? Micro-Diver 1. Q: What do Divers actually do? 2. Q: Which instruments are necessary for monitoring the groundwater level? 3. Q: Can I place the Diver close to an electrical pump? 4. Q: How should I install

More information

Deep SOLO. Nathalie Zilberman, Dean Roemmich, and SIO float lab. 1. Deep SOLO float characteristics. 2. Deep SOLO float Deployment

Deep SOLO. Nathalie Zilberman, Dean Roemmich, and SIO float lab. 1. Deep SOLO float characteristics. 2. Deep SOLO float Deployment Deep SOLO Nathalie Zilberman, Dean Roemmich, and SIO float lab 1. Deep SOLO float characteristics 2. Deep SOLO float Deployment 3. Deep SOLO temperature and salinity Deep SOLO Float deployment, R/V Tangaroa

More information

Mass Spec will not Autotune

Mass Spec will not Autotune Mass Spec will not Autotune Applies to 5973A/N MSD What could be the problem? There could be several things that would cause your Mass Spec not to Autotune. The most common, easily corrected Autotune problems

More information

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel TANK MANAGER FOR TWO TANKS OPERATING MANUAL 10/31/11 C-More T6C L color touch panel 1 TABLE OF CONTENTS GENERAL...3 INSTALLATION...4 STONE TEST PROCEDURE...7 OPERATIONAL SUMMARY...7 AUTO CARBONATION...10

More information

ATON System Workshop

ATON System Workshop ATON System Workshop ATON System: Real-time current measurements from Coast Guard navigation buoys in ports, bays and the coastal ocean Aid-to-Navigation = ATON ATON System Workshop Outline: Overview of

More information

PRODUCT MANUAL. Diver Smart Interface Cable AS346

PRODUCT MANUAL. Diver Smart Interface Cable AS346 PRODUCT MANUAL Diver Smart Interface Cable AS346 Contact details: Van Essen Instruments B.V. Van Essen Instruments - Canada Delftechpark 20 630 Riverbend Drive, Suite 100 2628 XH Delft Kitchener, ON, The

More information

Vacuum Pumpdown and Venting Procedure, CRaTER Thermal Vacuum System. Dwg. No

Vacuum Pumpdown and Venting Procedure, CRaTER Thermal Vacuum System. Dwg. No Rev. ECO Description Checked Approval Date 01 32- Release M. Smith Vacuum Pumpdown and Venting Procedure, CRaTER Thermal Vacuum System Dwg. No. 32-06003.05 Revision 01 June 18, 2007 1 1. Introduction 1.1.

More information

The Military CYPRES Quick Guide For Operators

The Military CYPRES Quick Guide For Operators The Military CYPRES Quick Guide For Operators The military CYPRES is available in the following standard models: All models can be adapted with a one or two-pin cutter. 1000/35 A 1500/35 A 1000/35 indicates

More information

MAINTENANCE PROCEDURE FOR X 650

MAINTENANCE PROCEDURE FOR X 650 MAINTENANCE PROCEDURE FOR X 650 X 650 25. juli 2005-1/6 MAINTENANCE PROCEDURE FOR X 650 2 ND STAGE WARNING: This maintenance procedure is only for appointed Scubapro technicians that completed a course

More information

INTRODUCTION TO NETWORK WIND 3 MOUNTING THE UNIT 14 SELECTING THE DISPLAY MODE 5 ABBREVIATIONS AND DEFINITIONS 17

INTRODUCTION TO NETWORK WIND 3 MOUNTING THE UNIT 14 SELECTING THE DISPLAY MODE 5 ABBREVIATIONS AND DEFINITIONS 17 CONTENTS CONTENTS 1 INSTALLATION 14 GENERAL INTRODUCTION TO B&G NETWORK 2 SITING THE UNIT 14 INTRODUCTION TO NETWORK WIND 3 MOUNTING THE UNIT 14 EXAMPLE SYSTEMS USING NETWORK WIND 4 SPECIFICATION 16 SELECTING

More information

Instruction Manual VCG-6

Instruction Manual VCG-6 Specifications Amps / Volts requirements 400 ma @ 24 VDC Propane pressure (VCG-6LP) 11 WC / 2.8 kpa Natural Gas pressure (VCG-6NG) 4.5 WC / 1.15 kpa Cu ft per hour / CO2 3-6 SCFH BTU Rating (Variable)

More information

EZ Boom 2010 System for the EZ Guide 500 Lightbar Triangle Ag Services Users Guide

EZ Boom 2010 System for the EZ Guide 500 Lightbar Triangle Ag Services Users Guide EZ Boom 2010 System for the EZ Guide 500 Lightbar Triangle Ag Services Users Guide Parts of the Controller (For details on the parts of the EZ Boom controller refer to Appendix F) Status Indicator Rate

More information

SVP-B drifter built by METOCEAN Use and deployment instructions

SVP-B drifter built by METOCEAN Use and deployment instructions Centre de Meteorologie Marine Tel: +33 2 98 22 18 52 Meteo-France Fax: +33 2 98 22 18 49 13, rue du Chatellier E-mail: Pierre.Blouch@meteo.fr BP 90411 29604 BREST CEDEX France SVP-B drifter built by METOCEAN

More information

Agilent ICP-MS Interactive Troubleshooting tool for Plasma ignition problem

Agilent ICP-MS Interactive Troubleshooting tool for Plasma ignition problem Agilent ICP-MS Interactive Troubleshooting tool for Plasma ignition problem Rev. B 2017, January, 20 Next > Preconditions 7900, 7800, 7700, 8800, or 8900 ICP-MS is used. Sample introduction type is PeriPump.

More information

DEEPSEA. LED Micro SeaCam POWER & LIGHT. Operator s Manual. Original Instruction Ruffin Road San Diego, CA USA

DEEPSEA. LED Micro SeaCam POWER & LIGHT. Operator s Manual. Original Instruction Ruffin Road San Diego, CA USA DEEPSEA POWER & LIGHT LED Micro SeaCam T: 858-576-1261 F: 858-576-0219 4033 Ruffin Road San Diego, CA 92123-1817 USA www.deepsea.com sales@deepsea.com Original Instruction REV 02/26/18 LED Micro SeaCam

More information

INSTRUCTION MANUAL LOW PRESSURE CALIBRATION PUMP MODEL PGL

INSTRUCTION MANUAL LOW PRESSURE CALIBRATION PUMP MODEL PGL INSTRUCTION MANUAL LOW PRESSURE CALIBRATION PUMP MODEL PGL Dear user, We have made every effort to ensure the accuracy of the contents of this manual. Should any errors be detected, we would greatly appreciate

More information

living with the lab control of salinity 2012 David Hall

living with the lab control of salinity 2012 David Hall control of salinity 2012 David Hall General Idea The objective is to keep the salinity close to a setpoint which will be provided by your instructor The salinity sensor measures the analog voltage output

More information

Micrio Tester for the Raymarine ST50, ST60, and ST60+ Wind Instruments. Rev 1.1

Micrio Tester for the Raymarine ST50, ST60, and ST60+ Wind Instruments. Rev 1.1 Micrio Tester for the Raymarine ST50, ST60, and ST60+ Wind Instruments. Rev 1.1 The Micrio Wind Tester allows quick and easy testing of the Raymarine ST50/ST60/ST60+ masthead wind instrument. The Micrio

More information

Guidance on the maintenance and inspection of fixed CO 2 fire-extinguishing systems

Guidance on the maintenance and inspection of fixed CO 2 fire-extinguishing systems Translation. Only the Danish document has legal validity. Guidance no. 9514 of 9 July 2014 issued by the Danish Maritime Authority Guidance on the maintenance and inspection of fixed CO 2 fire-extinguishing

More information

RR10106 Cruise Plan ITOP Mooring Recovery and PhilSea10 Seaglider Operation (R/V Revelle November 6-26, 2010)

RR10106 Cruise Plan ITOP Mooring Recovery and PhilSea10 Seaglider Operation (R/V Revelle November 6-26, 2010) RR10106 Cruise Plan ITOP Mooring Recovery and PhilSea10 Seaglider Operation (R/V Revelle November 6-26, 2010) Chief Scientist: Ren-Chieh Lien, Applied Physics Lab, University of Washington Co-Chief Scientist:

More information

Laboratory Mortar Mixer (Testing)

Laboratory Mortar Mixer (Testing) TomTom-Tools GmbH Zelgli 20 8905 Arni info@tomtom-tools.com Switzerland www.tomtom-tools.com User Manual Version February 22, 2015 Laboratory Mortar Mixer (Testing) 1 Introduction The Laboratory Mortar

More information

Specialists in HTM Medical Gas Pipeline Equipment AVSU/Module Operating and Maintenance Instructions

Specialists in HTM Medical Gas Pipeline Equipment AVSU/Module Operating and Maintenance Instructions The Area Value Service Units and Module format (AVSU) provide a local gas isolation facility for use during normal installation and maintenance or in the event of an emergency. They are built in accordance

More information

STS ICP-RIE. Scott Munro (2-4826,

STS ICP-RIE. Scott Munro (2-4826, STS ICP-RIE LOCATION: Plasma Etch Area PRIMARY TRAINER: Scott Munro (2-4826, email@address.com) 1. OVERVIEW The STS ICP-RIE is available to users who require deep anisotropic silicon etching with near

More information

CORROSION AT THE AIR-WATER INTERFACE REMOVE THE GAS

CORROSION AT THE AIR-WATER INTERFACE REMOVE THE GAS CORROSION AT THE AIR-WATER INTERFACE There are three (3) physical attributes of oxygen that explain why almost all corrosion in wet pipe fire sprinkler systems occurs in close proximity to trapped air

More information

Besides all the normal preparation, this highlights the typical rms features

Besides all the normal preparation, this highlights the typical rms features How to use the scrubber monitoring system on the rms Version 0.1 22 02 2012 not final yet: updates will follow Latest modifications in yellow Unfinished parts in italic 1 preparation of the unit Besides

More information

Instructions for Use

Instructions for Use Select-380 T-Auto T-Auto Contents Page Instructions for Use Ref: 3.0 IFU 380 T-Auto Mar 18 2 Schematic layout of the doser 3 Quick-fit instructions 4 Description/Installation/Operation Pump tubes & Water

More information

TECHNICAL DATA. Q = C v P S

TECHNICAL DATA. Q = C v P S January 6, 2012 Preaction 348a 1. Description Viking supervised Surefire Preaction Systems utilize the Viking G-6000P Valve. The small profile, lightweight, pilot operated Viking G-6000P Valve comes complete

More information

WPB 5 / 7 / 10 INSTALLATION, OPERATION, & MAINTENANCE

WPB 5 / 7 / 10 INSTALLATION, OPERATION, & MAINTENANCE PACIFIC LIQUID & AIR SYSTEMS WPB 5 / 7 / 10 INSTALLATION, OPERATION, & MAINTENANCE IN THE BEGINNING Before installing or operating this system, familiarize yourself with these instructions. You should

More information

Government of Bermuda Department of Maritime Administration BERMUDA SHIPPING NOTICE

Government of Bermuda Department of Maritime Administration BERMUDA SHIPPING NOTICE Government of Bermuda Department of Maritime Administration 2015 027 BERMUDA SHIPPING NOTICE Maintenance and Inspection of Fixed Carbon Dioxide Fire Extinguishing Systems Ref: SOLAS 74/88 as amended, Regulation

More information

- Repeat calibration until desired results are reached otherwise set SW2 and SW1-8 open.

- Repeat calibration until desired results are reached otherwise set SW2 and SW1-8 open. Calibration Procedures on Cobe Centrysystem 3: I) Balance Chamber Calibration: - Verify that pressure regulators PRV and PRV2 are calibrated as well as pressure transducers PDI and PDO with switches SW2

More information

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System) Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System) Revision date October 2014 2 Emergency Plan for Nano Furnace Standard Operating Procedures for Emergencies Contact information Person

More information

Evaporative Emissions (EVAP) system, checking

Evaporative Emissions (EVAP) system, checking Page 1 of 12 24-136 Evaporative Emissions (EVAP) system, checking CAUTION! Whenever carrying out work on the fuel supply and/or fuel injection systems, read and carefully follow the Cautions, Warnings,

More information

User Manual. Quantos Automated Dosing Liquid Module

User Manual. Quantos Automated Dosing Liquid Module User Manual Liquid Module 1 Safety Information 1.1 Definition of warnings and symbols Signal Words WARNING for a hazardous situation with medium risk, possibly resulting in severe injuries or death if

More information

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products Saturn Evo SATURN EVO COLOR A N A E S T H E S I A S Y S T E M i n t e l l i g e n t s i m p l i c i t y Manufacturer of high quality anaesthesia products SATURN EVO COLOR PATIENT SAFETY AND USER PATIENT

More information

NGP-250/500 Nitrogen Generator Quick Start Guide

NGP-250/500 Nitrogen Generator Quick Start Guide NGP-250/500 Nitrogen Generator Quick Start Guide Version: A July 2013 Potter Electric Signal Company, LLC 5757 Phantom Dr., Suite 125 P. O. Box 42037 Hazelwood, MO 63042 Phone: (314) 595-6900 Document

More information

TECHNICAL DATA. Q = C v P S

TECHNICAL DATA. Q = C v P S Preaction 346a 1. Description The 6 Model G-6000P Electric Release Preaction System Riser Assembly can be used as a Single Interlock Preaction System with Electric Release, or as a Double Interlock Preaction

More information

MicroTim XB. User Manual. Precision Digital Barometric Altimeter / Barometer / VSI. Document Revision 1.0 Firmware Version 3.0

MicroTim XB. User Manual. Precision Digital Barometric Altimeter / Barometer / VSI. Document Revision 1.0 Firmware Version 3.0 MicroTim XB Precision Digital Barometric Altimeter / Barometer / VSI User Manual Document Revision 1.0 Firmware Version 3.0 Table of Contents Table of Contents...2 1 General Operation...5 1.1 Altitude

More information

Archimer

Archimer Please note that this is an author-produced PDF of an article accepted for publication following peer review. The definitive publisher-authenticated version is available on the publisher Web site Sea Technology

More information

DG100 Digital Depth Gauge and Timer. Training Module, Rev B

DG100 Digital Depth Gauge and Timer. Training Module, Rev B DG100 Digital Depth Gauge and Timer Training Module, Rev B DG100 Digital Depth Gauge Dive and Surface Timer Stopwatch Mode Designed for Military Divers Combat Swimmers DG100 Used as a Replacement for Analog

More information

Water Mist Systems Inspection, Testing, and Maintenance of Water Mist Systems

Water Mist Systems Inspection, Testing, and Maintenance of Water Mist Systems Water Mist Systems Inspection, Testing, and Maintenance of Water Mist Systems Name of Property: Address: Phone Number: Inspector: Contract No.: Date: This Report Covers: Monthly Quarterly Annual Other

More information

Fire Apparatus Operator: ARFF

Fire Apparatus Operator: ARFF Fire Apparatus Operator: ARFF Skill Sheet Package Based on NFPA 1002: Standard on Fire Apparatus Driver/ Operator Professional Qualifications, 2017 Edition. Alabama Fire College November 26, 2018 Alabama

More information

2017 Float and Sensor Workshop. Conveners: Brian King, Steve Riser, Susan Wijffels Local Host: Steve Riser, U. Washington

2017 Float and Sensor Workshop. Conveners: Brian King, Steve Riser, Susan Wijffels Local Host: Steve Riser, U. Washington 2017 Float and Sensor Workshop Conveners: Brian King, Steve Riser, Susan Wijffels Local Host: Steve Riser, U. Washington Objectives The goals of the workshop were to Increase the overall efficiency of

More information

frequently asked questions

frequently asked questions Hydra Pilot Fault Codes What do the fault codes for the Hydra Pilot mean? Fault Cause FAULT 100 FAULT 101 FAULT 102 FAULT 103 FAULT 104 FAULT 105 FAULT 106 FAULT 108 FAULT 109 FAULT 110 FAULT 111 FAULT

More information

Operating manual. TA 197 Oxi. Dissolved Oxygen Depth Armature

Operating manual. TA 197 Oxi. Dissolved Oxygen Depth Armature Operating manual Dissolved Oxygen Depth Armature ba25302e05 04/2009 Copyright Weilheim 2009, WTW GmbH Reprinting - even as excerpts - is only allowed with the explicit written authorization of WTW GmbH,

More information

OPERATING INSTRUCTIONS Sherlock Chamber Testing System

OPERATING INSTRUCTIONS Sherlock Chamber Testing System OPERATING INSTRUCTIONS Sherlock Chamber Testing System Arrangement Equipment Kit FL90 & FL100 Corbel Test Plate with Gasket Seal S8, S14, S15 & S16 Sump Test Plate with Gasket Seal S7 & S5 Chamber Test

More information

Fire Apparatus Operator Aircraft Rescue and Fire Fighting

Fire Apparatus Operator Aircraft Rescue and Fire Fighting Fire Apparatus Operator Aircraft Rescue and Fire Fighting Skill Sheet Package Based on NFPA 1002: Standard on Fire Apparatus Driver/ Operator Professional Qualifications, 2014 Edition. Alabama Fire College

More information

NOVALYNX CORPORATION MODEL WEATHER STATION INSTRUCTION MANUAL

NOVALYNX CORPORATION MODEL WEATHER STATION INSTRUCTION MANUAL NOVALYNX CORPORATION MODEL 100-1950 WEATHER STATION INSTRUCTION MANUAL Receiving and Unpacking Carefully unpack all components and compare to packing list. Notify NovaLynx Corporation immediately concerning

More information

DIVE PLAN Leg 1 Dive 11 Recover Wally 1

DIVE PLAN Leg 1 Dive 11 Recover Wally 1 DIVE PLAN Leg 1 Dive 11 Recover Wally 1 Page 1 of 6 Location: Barkley Canyon Date: 9 May, 2014 14:00 Constraints: Weather, Sufficient deck space ROV Dive # OE 114 Objectives Prepare Wally I cable for recovery

More information

Oxygen measurements: sensors accuracy and scientific needs

Oxygen measurements: sensors accuracy and scientific needs Oxygen measurements: sensors accuracy and scientific needs L. Coppola (CNRS-INSU), E.Diamond (CNRS-INSU), L.Delauney (IFREMER), D.Hydes (NOC), M.Haller (HZG), J.Karstensen (IFM-GEOMAR), D.Lefevre (CNRS-INSU),

More information

AMS 2710 PCB pressure sensor module with V output

AMS 2710 PCB pressure sensor module with V output FEATURES Universal pressure sensor module with 0.. 10 V voltage output Fully calibrated and temperature compensated sensor module Variants for (bidirectional) differential, gage, absolute and barometric

More information

Oxygen Meter User Manual

Oxygen Meter User Manual Oxygen Meter User Manual Monday, July 23, 2007 1. Outline...2 2. Program...3 2.1. Environment for program execution...3 2.2. Installation...3 2.3. Un installation...3 2.4. USB driver installation...3 2.5.

More information

MOYNO FLUID DETECTION CONTROL SERVICE MANUAL. Section: ACCESSORIES. Page: 1 of 8 Date: May 1, 1998 GENERAL

MOYNO FLUID DETECTION CONTROL SERVICE MANUAL. Section: ACCESSORIES. Page: 1 of 8 Date: May 1, 1998 GENERAL MOYNO Section: ACCESSORIES Page: 1 of 8 Date: May 1, 1998 SERVICE MANUAL FLUID DETECTION CONTROL GENERAL The Moyno Fluid Detection Control prevents unnecessary loss of production time or repair costs to

More information

3.0 Pressure Transmitter Selection

3.0 Pressure Transmitter Selection 3.0 Pressure Transmitter Selection Each Tronic Line pressure transmitter has different features to meet specific performance, environmental, and price requirements. It is not possible to describe every

More information

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS Annex 3, page 2 ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS The text of existing chapter 15 is replaced by the following: "1 Application This

More information

E-AIMS. R&D on float technology Synthesis D2.6.1

E-AIMS. R&D on float technology Synthesis D2.6.1 Research Project co-funded by the European Commission Research Directorate-General 7 th Framework Programme Project No. 284391 E-AIMS Euro-Argo Improvements for the GMES Marine Service R&D on float technology

More information

Quality control of data streams at ANFOG

Quality control of data streams at ANFOG Quality control of data streams at ANFOG Christine Hanson & Claire Gourcuff Australian National Facility for Ocean Gliders (ANFOG), School of Environmental Systems Engineering (SESE) and UWA Oceans Institute

More information

Chromat Calibration Updated October 27th, 2017

Chromat Calibration Updated October 27th, 2017 Chromat Calibration Updated October 27th, 2017 Calibrating the Chromatograph Performing the Chromat Calibration is highly recommended when starting a new well. The MLogger already has a default calibration

More information

TECHNICAL DATA. Q = C v P S

TECHNICAL DATA. Q = C v P S January 6, 2012 Preaction 333a 1. Description Viking supervised Surefire Preaction Systems Utilize the Viking G-3000P Valve. The small profile, lightweight, pilot-operated Viking G-3000P Valve comes complete

More information

H-DIL Hydrogen/Oxygen Sample-Draw Detector Operator s Manual

H-DIL Hydrogen/Oxygen Sample-Draw Detector Operator s Manual 35-3001-06H-DIL Hydrogen/Oxygen Sample-Draw Detector Operator s Manual Part Number: 71-0346 Revision: A Released: 8/7/17 www.rkiinstruments.com WARNING Read and understand this instruction manual before

More information

Annual Inspection and Calibration Checklist VAF Oilcon MK-6(M) MEPC.108(49) Page 1 of 5

Annual Inspection and Calibration Checklist VAF Oilcon MK-6(M) MEPC.108(49) Page 1 of 5 Page 1 of 5 Vessel Name: Serial number MCU: Serial Number EPU: Serial Number Skid: Mains Supply: V AC @ Hz Software version MCU: Software version EPU: K-factor Cell: K-factor Skid Flowmeter: P/L Maximum

More information

How well do CMIP5 climate models reproduce Southern Ocean bottom temperature? Model climatology

How well do CMIP5 climate models reproduce Southern Ocean bottom temperature? Model climatology How well do CMIP5 climate models reproduce Southern Ocean bottom temperature? Model climatology (colours) plus August (dashed) & February (solid) sea ice edge. Heuzé et al. (2013, GRL) Autosub autonomous

More information

Inert Air (N2) Systems Manual

Inert Air (N2) Systems Manual INSTRUCTION MANUAL Inert Air (N2) Systems Manual N2-MANUAL 2.10 READ AND UNDERSTAND THIS MANUAL PRIOR TO OPERATING OR SERVICING THIS PRODUCT. GENERAL INFORMATION Positive pressure nitrogen gas pressurizing

More information

RAM 4021 Operation Manual

RAM 4021 Operation Manual RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For your safety...3 Description...3 Set-up mode...4 Annunciator lights/alarms...4 Operation...5 Calibration...6

More information

Technical

Technical Control Settings SETTING Set point Range HOT Set point Differential ECO Limit Timings VALUE 90 F (32 C) to 160 F (71 C) 120 F (49 C) 15 F (9 C) 199 F (93 C) Notice the BLUE temperature adjustment knob.

More information

Low Flying Introduction

Low Flying Introduction Advanced Manoeuvres Low Flying Introduction Commonly, low flying refers to any flight at or below 500 feet agl that may be practised only in designated low flying zones. By maintaining good situational

More information

NOVALYNX CORPORATION MODEL BAROMETRIC PRESSURE SENSOR PCB MODEL BAROMETRIC PRESSURE SENSOR IN NEMA ENCLOSURE INSTRUCTION MANUAL

NOVALYNX CORPORATION MODEL BAROMETRIC PRESSURE SENSOR PCB MODEL BAROMETRIC PRESSURE SENSOR IN NEMA ENCLOSURE INSTRUCTION MANUAL NOVALYNX CORPORATION MODEL 230-600 BAROMETRIC PRESSURE SENSOR PCB MODEL 230-601 BAROMETRIC PRESSURE SENSOR IN NEMA ENCLOSURE INSTRUCTION MANUAL REVISION DATE: 11/15/2006 Receiving and Unpacking Carefully

More information

Simple, Robust, Reliable Measures ph Doses either acid (lower) or alkali (raise) Displays ph, Total Dose Count (DCT) and Dose Count per hour (DC/h)

Simple, Robust, Reliable Measures ph Doses either acid (lower) or alkali (raise) Displays ph, Total Dose Count (DCT) and Dose Count per hour (DC/h) Simple, Robust, Reliable Measures ph Doses either acid (lower) or alkali (raise) Displays ph, Total Dose Count (DCT) and Dose Count per hour (DC/h) Integrated peristaltic dosing pump When using automatic

More information

La Crosse Technology, Ltd. Page 1

La Crosse Technology, Ltd. Page 1 XG-55 FAQS The links below will work in most PDF viewers and link to the topic area by clicking the link. We recommend Adobe Reader version 10 or greater available at: http://get.adobe.com/reader CONTENTS

More information

R E D I C O N T R O L S

R E D I C O N T R O L S R E D I C O N T R O L S Operation & Maintenance Manual Portable Service Purger for Low Pressure Chillers Model: PSP-LP-1B For Refrigerants R-11, R-113, R-114 & R-123 & Other Similar Refrigerants File Literature

More information

Technical

Technical Control Settings SETTING Set point Range HOT Set point Differential ECO Limit Timings TECHNICAL SERVICE DEPARTMENT VALUE 90 F (32 C) to 160 F (71 C) 120 F (49 C) 15 F (9 C) 199 F (93 C) Notice the BLUE

More information

X-dock Software History

X-dock Software History X-dock Software History SW Version Bugfixes Features 01.01.01 01.02.2013 01.01.02 01.02.2013 01.02.05 04.03.2013 01.02.10 Start of history Pac 7000 support with FW 4.0 (older firmware locked) Improvements

More information

UsER manual for Watersens ph -REDOX

UsER manual for Watersens ph -REDOX UsER manual for Watersens -REDOX Cl 8 1 2 6 3 3 7 7 4 4 4 4 Parts List 1 Redox Probe 1 x 2 PH Probe 1 x 5 Tube Weight 2 x 6 Connection Valve 1 x chlorine 3 Chlorine and Pumps 2 x 7 Dosing Valve 2 x 5 5

More information

PRACTICAL EXPERIENCES IN THE PRODUCTION OF INSULATION FOAMS WITH HYDROCARBON BLOWING AGENT IN DEVELOPING AND DEVELOPED COUNTRIES

PRACTICAL EXPERIENCES IN THE PRODUCTION OF INSULATION FOAMS WITH HYDROCARBON BLOWING AGENT IN DEVELOPING AND DEVELOPED COUNTRIES PRACTICAL EXPERIENCES IN THE PRODUCTION OF INSULATION FOAMS WITH HYDROCARBON BLOWING AGENT IN DEVELOPING AND DEVELOPED COUNTRIES R I S T O O J A L A, U N I D O F O A M S E C T O R C O N S U LTA N T SAFETY

More information

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr A DIVISION OF THE FREDERICKS COMPANY 2400 PHILMONT AVE. HUNTINGDONVALLEY, PA 19006 PARTS LIST 1 3 4 2 # QTY ITEM DESCRIPTION

More information

FAULT CODE TROUBLESHOOTING INDEX

FAULT CODE TROUBLESHOOTING INDEX FAULT CODE TROUBLESHOOTING INDEX 1. Display indicates Change Filters 2. Display indicates Drip Tray Full Continuous Alarm will Sound 3. Display indicates Cold Fault 4. Display indicates Hot Fault 5. Display

More information