Hydro user manual. Malvern Instruments Limited

Size: px
Start display at page:

Download "Hydro user manual. Malvern Instruments Limited"

Transcription

1 Hydro 2000 µp Malvern Instruments Limited Printed in England MRK user manual Enigma Business Park Grovewood Road, Malvern Worcs, WR14 1XZ, U.K. Tel: +44 (0) Fax: +44 (0) I N S T R U M E N T S

2

3 Hydro 2000µP User Manual MAN0389 Issue 1.0 March 2007

4 Malvern Instruments Ltd Malvern Instruments makes every effort to ensure that this document is correct. However, due to Malvern Instruments policy of continual product development we are unable to guarantee the accuracy of this, or any other document after the date of publication. We therefore disclaim all liability for any changes, errors or omissions after the date of publication. No reproduction or transmission of any part of this publication is allowed without the express written permission of Malvern Instruments Ltd. Head office: Malvern Instruments Ltd. Enigma Business Park, Grovewood Road, Malvern, Worcestershire WR14 1XZ United Kingdom. Tel + [44] (0) Fax + [44] (0) Windows 2000 and XP are registered trademarks of the Microsoft Corporation. Tygon is a registered trademark of Norton Co. Perlast is a registered trademark of Precision Polymer Engineering. Printed in England

5 Table of Contents Part 1 - Operator s Guide Introduction to this manual Introduction Access to the dispersion unit Assumed information Where to get help Hardware features Introduction What the dispersion unit does How the dispersion unit is controlled How the dispersion unit works Hardware features Software features Introduction Making a measurement Manually controlling the dispersion unit Writing an SOP for the dispersion unit Making a measurement Introduction Measurement guidelines Instrument preparation Anaerobic fill Measuring new samples Changing cell holder position Introduction Hydro 2000µP Page i

6 Table of Contents Hydro 2000µP Part 2 - Appendices Specification Specification A-1 Chemical compatibility Introduction B-1 Regulatory statements CE Declaration of Conformity C-2 Index Page ii MAN 0389

7 Part 1 - Operator s Guide

8

9 1 Introduction to this manual Introduction This manual describes the operation of the Micro Precision Hydro 2000µP sample dispersion unit, one of the following: Instrument Hydro 2000µP (with ultrasonics) Hydro 2000µP (without ultrasonics) Model number AWA2003 AWA2004 This manual is a supplement to the following manuals: Mastersizer 2000 User Manual (termed the main User Manual ). Mastersizer 2000 Essentials Manual (termed the Essentials Manual ). This dispersion unit manual focuses on some specific issues of the dispersion unit that are not covered by the above manuals. It aims to: Describe what the dispersion unit is and explain in simple terms how it works. Identify the physical features of the dispersion unit. Describe the software and explain how to use the dispersion unit to make a measurement on the system. Warning! The dispersion unit or the samples to be measured may be hazardous if misused. Users must read the Health and Safety information in the Essentials Manual before operating the system. We recommend that users who have never operated a Malvern particle analyser read this manual fully before starting the first measurement. Those who are more familiar with particle size analysers may wish to jump straight to Chapter 4 of the main User Manual which gives details on making measure- Hydro 2000µP Page 1-1

10 1 Chapter 1 Introduction to this manual ments. However, the importance of sample preparation before measurement cannot be overstated so we recommend reading the chapter on sample preparation (Chapter 8 of the main User Manual) as a priority. Access to the dispersion unit Within this manual, reference is made to the various people who will have access to the dispersion unit. The following is a list of these people and their responsibilities: Malvern personnel Malvern personnel (service engineers, representatives, etc.) have full access to the dispersion unit and are authorised to perform all service procedures that may require the removal of the covers. Supervisor The supervisor is the person responsible for the management/safety of the dispersion unit and of its operation. The supervisor is responsible for the training of the operators. The supervisor can perform all user maintenance routines identified in Chapter 4 of the Essentials Manual. Warning! The supervisor/operator must never remove the covers of the instrument or dispersion unit. Removal of the covers by unauthorised personnel will invalidate the warranty of the dispersion unit. Operator An operator is a person trained in the use of the dispersion unit. The operator can perform all user maintenance routines identified in Chapter 4 of the Essentials Manual. Warning! Failure to follow this guideline could result in exposure to hazardous voltages. Page 1-2 MAN 0389

11 Introduction to this manual Chapter 1 Assumed information Naming convention Within this manual: The Mastersizer 2000 optical bench is referred to as the optical bench or the instrument. The Hydro 2000µP is referred to in full or as the dispersion unit. The combination of the optical bench, one or more dispersion units and the computer is referred to as the system. Menu commands Menu commands from the Malvern software are referred to in the form main menu-menu item. As an example, the command Configure-New SOP refers to selecting the New SOP item in the Configure menu. The same rules apply for sub-menus of sub-menus, so that Tools-Options-Instrument Port refers to the Instrument Port item in the Options sub-menu, which itself is a sub-menu of the Tools menu. Menu commands are always shown in bold text. Where to get help Full details on where and how to obtain help can be found in Chapter 1 of the Mastersizer 2000 User Manual. The Essentials Manual gives information on the following: Site requirements. Health and Safety. Maintenance. Installation (in case the system has to be moved after its initial installation by Malvern Instruments personnel). Hydro 2000µP Page 1-3

12 Chapter 1 Introduction to this manual Page 1-4 MAN 0389

13 2 Hardware features Introduction This chapter gives a simple overview of the dispersion unit. All key aspects of the dispersion unit s operation are explained in very simple terms. This chapter covers: What the dispersion unit does. How the dispersion unit is controlled. How the dispersion unit works. The hardware features in detail. Hydro 2000µP Page 2-1

14 2 Chapter 2 What the dispersion unit does Hardware features The sole purpose of any sample dispersion unit is to prepare the sample then deliver it to the measurement zone of the optical bench so that it can be measured. The Hydro 2000µP allows the Mastersizer to be used for particle-in-liquid particle sizing and has been specifically designed so that only small amounts of sample and suspending liquids (dispersants) are required. The small volume of the Hydro 2000µP is ideal when using solvent dispersants or when samples and dispersants are either expensive or hazardous. The materials used in the manufacture of the dispersion unit maximise the range of samples which can be measured. Other features of the dispersion unit include: Variable speed pump allows a wide range of particle sizes and densities to be suspended and circulated. Ultrasonic system (optional) allows particle agglomerates to be dispersed and assists in removing bubbles from the system. Warning! When using solvents, especially those with low ignition points, minimise the use of ultrasonics. There is a remote possibility that localised heating caused by the ultrasonic action may cause an explosive reaction. Check the Material Safety Data Sheets for all substances used. Temperature sensing an internal sensor monitors the temperature of the dispersant. The software can be configured so that measurements are only made between user specified temperature limits. This can be used to ensure that the instrument can only be used when it has reached the correct temperature. How the dispersion unit is controlled The dispersion unit is controlled using a single software dialogue with sliders for the pump speed and ultrasonic power (Chapter 3 has details). The dialogue also has pump pause, drain valve and Anaerobic fill buttons. When controlled through a Standard Operating Procedure (SOP), the software tells the user what value to set for each parameter. Page 2-2 MAN 0389

15 Hardware features Chapter 2 How the dispersion unit works The dispersion unit looks like this: The dispersion unit comprises two main units: The control unit which contains most of the control electronics. The cell unit which contains the cell, the sample well, pump, ultrasonic probe (optional) and the drain facilities. The dispersant is injected into the cell via the dispersant inlet port using a syringe. Once preliminary measurements ( backgrounds ) are made, the sample to be measured is added to the well. The pump agitates the dispersant to ensure that the sample does not settle out and continually circulates the sample through the cell and then back to the well. Within the cell, the dispersant and sample flow between two glass windows. The windows are precisely mounted within the measurement zone of the optical bench where a laser is passed through the sample/dispersant mixture. The particles within the sample scatter the laser light. This scattered laser light is measured by the optical bench and used to calculate the particle size distribution of the sample. ill 5473 Hydro 2000µP Page 2-3

16 Chapter 2 Hardware features The optional ultrasonic probe can be used to break up particle agglomerates and help to remove bubbles from the system. Once the sample has been measured, an internal drain valve is operated and the dispersant and sample are drained to an external beaker. The lid for the well should always be fitted during a measurement. The lid is locked in place by rotating the locking lever. All functions of the dispersion unit are controlled using the Malvern software. Hardware features This section describes the hardware in detail. Main system components The diagram below shows a typical installation for the Hydro 2000µP. Its main components are identified. These components are examined in more detail later in this section Power supply The external DC Power Supply Unit (PSU) supplies all power requirements to the dispersion unit. ill 5407 Page 2-4 MAN 0389

17 Hardware features Chapter 2 Warning! The PSU used on the Hydro 2000µP is not designed for operation in wet conditions. Do not place it in areas that are prone to spillages. Warning! Only use the power supply and cables supplied by Malvern Instruments. Control unit The control unit contains most of the control electronics for the dispersion unit. Power and communications cables are connected to the control unit. A control cable is connected from the front of the control unit to the front of the cell unit. Cell unit The cell unit contains the cell, well, pump and drain facilities. The cell unit is designed to circulate the sample and dispersant mixture and pass it through the analysing laser beam of the optical bench. Sample well The well in the cell unit is where sample to be measured is added. The sample well and connecting pipes have a capacity of less than 20ml. Dispersant input/syringe The dispersant for suspending the sample and flushing the system is injected into this port using a syringe. Always use a syringe with a maximum capacity of 20 to 25ml. Drain beaker Once the measurement is complete (or during a flush routine to clean the system) an internal drain valve can be operated. The contents of the cell will drain through the drain port into the drain beaker. If the end of the drain pipe is under the surface of the waste liquid, the back pressure created may prevent the cell from draining correctly. Typically ensure that the end of the tube is approximately 3cm below the rim of the beaker (slide the beaker holder until the correct length is achieved). Hydro 2000µP Page 2-5

18 Chapter 2 Hardware features ill 5643 Warning! If using hazardous dispersants or samples, do not allow the beaker to become over-full. If dispersants or samples are noxious, the instrument should ideally be used in a fume cupboard. Drain beaker holder The drain beaker may hold hazardous samples and dispersants. The drain beaker holder secures the drain beaker to prevent it from being accidentally knocked over when the cell unit is removed from the optical bench. We recommend that the beaker holder is always used. Cell unit holder The cell unit holder is a parking area for the cell when it is not in use. This is particularly useful when more than one dispersion unit is in use. It allows the cell to be removed and stored without the need to disconnect it from the dispersion unit. When the cell is parked for a short period make sure that the cell is full of dispersant. If it is to be parked for longer periods, the unit should be drained and the cell windows dried. The Essentials manual describes these procedures. Well lid The well lid is used to reduce the escape of hazardous fumes when measuring using hazardous samples and dispersants. It also reduces the risk of accidental splashing from the well. Always use the well lid when making measurements or moving the cell unit. Always lock the lid into place with the lid locking lever. Lid locking lever The lid locking lever is used to lock the well lid in place when measurements are taking place. Page 2-6 MAN 0389

19 Hardware features Chapter 2 Dispersant valve The dispersant valve is used to prevent sample or dispersant leaking when the dispersant input syringe is removed. Two types of valve are available: Aqueous valve this is disposable and can be used with non aggressive dispersants such as water. The valve is easily identifiable as it is made of transparent, purple plastic. Non-aqueous valve suitable for use with a large range of chemically aggressive dispersants. The non-aqueous valve is white and has a tap on its side. Replacement valves are available from a Malvern representative. Controller unit This section examines the controller unit in more detail. As stated above, the controller unit contains most of the control electronics for the dispersion unit. Rear panel The rear panel of the controller unit is shown below and its main features identified. DRAIN OPEN/CLOSED 4 CAUTION Flash tests can permanently damage this equipment WARNING: To prevent electric shock do not remove screws. No user serviceable parts inside. Refer servicing to Malvern trained service personnel. DC IN 24V 2.9A 1 ACCESSORY COMMS IN OUT 2 3 DC power input Power input from the external power supply. ill 5409 Hydro 2000µP Page 2-7

20 Chapter 2 Hardware features Warning! The PSU is not designed for operation in wet conditions. Do not place it in areas prone to spillages. Warning! Only use the PSU and cables supplied by Malvern Instruments. Accessory comms in connector The communications cable from the optical bench or from another dispersion unit connects here. Accessory comms out connector If more than one dispersion unit is connected to the system and this is the first dispersion unit in line, a communication cable will be connected from this connector to the Accessory comms - in connector on the second dispersion unit. Caution! The termination plug supplied with the optical bench must be fitted here if this is the only dispersion unit connected or it is the last dispersion unit in the line. Manual drain The manual drain button can be pressed to drain the well. Users may wish to do this if, for example, they need to drain the dispersion unit without having to start the control software. Pressing the button once (for about half a second) will open the drain valve. Pressing the button a second time will close the drain valve. If the drain valve is left open for more than 15 minutes it will automatically close. Page 2-8 MAN 0389

21 Hardware features Chapter 2 Front panel The front panel of the controller unit is shown below and its main features identified. 1 2 CONTROL ULTRASOUND 3 4 HYDRO Status indicator The status indicator illuminates in one of three colours when the dispersion unit is powered up: Green if the dispersion unit is functioning correctly and the cell unit has been loaded into the optical bench (i.e. the dispersion unit is active ). Amber if the dispersion unit is functioning correctly but the cell has not been loaded into the optical bench (i.e. the dispersion unit is at standby and is parked in the cell unit holder). Red if the software has detected a error. If an error message does not appear on the screen, selecting Configure-Accessories displays the accessory control dialogue which should show the type of error. Power on/off This is the main power switch for the dispersion unit. Control cable The control cable carries all power and communication signals between the control unit and the cell unit (with the exception of ultrasonics). ill 5410 Hydro 2000µP Page 2-9

22 Chapter 2 Hardware features Ultrasonic control cable (optional) If the ultrasonic option is fitted, this cable will carry power and control signals for the ultrasonic probe between the control unit and the cell unit. Warning! The Ultrasonic control cable carries high voltages. Under normal operating procedures this is perfectly safe. Inspect the cable at regular intervals for any sign of damage. Never operate the equipment if the cable is damaged. Cell unit The cell unit is shown below with its main features identified Well cone The well cone is where the sample is added to the cell unit. Typically, dispersant is added by filling the well to within 1 to 2mm of the top of the well. Perform an ill 5495 Page 2-10 MAN 0389

23 Hardware features Chapter 2 anaerobic fill (to remove bubbles) and then remove dispersant directly from the well using a pipette so that the well is about a quarter full. Overflow If the well is overfilled any excess is directed to the drain port via the overflow. Control cable The control cable carries all power and communication signals between the control unit and the cell unit (with the exception of ultrasonics). Ultrasonic control cable (optional) If the ultrasonic option is fitted this cable carries control signals for the ultrasonic transducer between the control unit and the cell unit. Dispersant inlet port The dispersant for suspending the sample and flushing the system is injected into this port using a syringe. To avoid introducing bubbles into the system, use a slow, steady pressure when injecting the dispersant; see Chapter 2 for details. Warning! Always consult the Materials Safety Data Sheet for any handling and safety information for all dispersants used. Only use syringes with a maximum capacity of 20-25ml. The dispersant for suspending the sample and flushing the system is injected into this port using a syringe. To avoid introducing bubbles into the system, use a slow, steady pressure when injecting the dispersant; see Chapter 2 for details. Fill the well until dispersant is visible within the well cone. Dispersant valve The dispersant valve must be used on the dispersant inlet to prevent sample/dispersant from leaking when the dispersant syringe is removed. Two types of valve are available, as described earlier. Replacement valves are available from a Malvern representative. Drain port When the drain valve is operated, the contents of the cell will flow out through this port. A tube will guide the contents to a waste beaker. Always ensure that the waste beaker has enough capacity before draining the cell. Well lid The well lid should always be in place when making measurements (except when adding sample). The lid reduces the amount of vapour released when using solvents as a dispersant. It also stops dispersant from welling up and stops air being sucked in. Hydro 2000µP Page 2-11

24 Chapter 2 Hardware features Additionally, when measurements are performed using solvents, users may occasionally find that the background levels are very high. This is due to temperature gradients within the solvent. The normal cure is to wait for several minutes for thermal equilibration, or even to warm the solvent beforehand. In many cases, a much simpler and quicker solution is to ensure that the lid is in place. Lid locking lever The lid locking lever is used to lock the well lid into place during measurements. Cell windows The cell windows allow the analyser beam of the optical bench to pass through the cell and hence the particle field. The cell windows can be removed to allow cleaning or replacement. Caution! The cell windows are delicate and should be treated with caution; refer to the Essentials Manual for details on cleaning the windows. Window ring lock screws The windows can be removed for cleaning. The window ring lock screws can be loosened to allow the window ring and window to be removed; refer to the Essentials Manual for details. Page 2-12 MAN 0389

25 3 Software features Introduction This chapter describes all the physical features of the dispersion unit. It covers: Making a measurement the basics of making a measurement using the dispersion unit. Manually controlling the dispersion unit. Writing a Standard Operating Procedure (SOP) for the dispersion unit. Software features the features of the Malvern software that are specific to the dispersion unit. Hydro 2000µP Page 3-1

26 3 Chapter 3 Making a measurement Software features The dispersion unit can either be controlled manually or automatically by running a Standard Operating Procedure (SOP). Making a measurement is fully documented in Chapter 4 of the main User manual. The system automatically detects which dispersion unit and cell is connected. If more than one dispersion unit is connected, the system detects all dispersion units connected, but only the dispersion unit that has its cell installed on the optical bench will be active. Manually controlling the dispersion unit To control the dispersion unit manually either select Configure-Accessory or, when making a manual measurement, click the Accessory button in the Measurement Display window. This accessory control dialogue appears: ill 7776 This dialogue is specific to the Hydro 2000µP. It has separate Control and Errors tabs. Use it to alter the pump speed and (optionally) the ultrasonic power by moving the relevant slider bar. Page 3-2 MAN 0389

27 Software features Chapter 3 It features the following: Pump Speed control The speed of the pump (which controls the speed at which the sample is circulated) is altered by moving the slider. Use the up and down arrows under the slider for fine adjustment of the pump speed. Ultrasound control (optional) Warning! When using solvents, especially those with low ignition points, minimise the use of ultrasonics. There is a remote possibility that localised heating caused by the ultrasonic action may cause an explosive reaction. Check the Material Safety Data Sheets for all substances used. The power of the ultrasonic probe is altered by moving the slider. Use the up and down arrows under the slider for fine adjustment of the ultrasonic power. Dispersant temperature The temperature of the dispersant within the cell is continually monitored and displayed here. Temperature monitoring can be used in conjunction with the Sampler Settings dialogue to limit the range of temperatures at which measurements can be made. Pause Pump button Pressing the Pause Pump button stops the pump motor. Pressing the button a second time restarts the pump. The indicator to the right of the button shows red if the pump is paused, and green when the pump is running. i Note If having difficulty with bubbles in the system, they can often be removed by turning the pump on and off in quick succession using Pause Pump. Drain Valve button Pressing the Drain Valve button will open the drain valve, allowing the contents of the cell to be emptied into the waste beaker. The indicator to the right of the button will show red if the drain is open and green when the drain is closed. Anaerobic fill button Due to the small volume of the Hydro 2000µP and associated problems of removing bubbles from the system, a special fill sequence called Anaerobic fill can be used. Chapter 4 gives details. Hydro 2000µP Page 3-3

28 Chapter 3 Software features Error indicators The following indicators on the Errors tab illuminate if an error condition has been detected. Connection box disconnected shows red if the control cable from the control unit to the cell unit is not connected. Dispersant Temperature Probe shows red if the temperature is outside the range 10 C - 35 C or there is an internal error with the temperature sensor. Contact a Malvern representative if this error condition persists. Drain Valve illuminates if the drain valve has failed to drive (i.e. it has stuck open or closed). This may be caused by leaving a sample in the cell for long periods - this may have set and bonded the drain valve. Never leave the sample in the cell for long periods - always flush the cell with clean dispersant. Contact a Malvern representative if this error indicator illuminates. Pump illuminates if there is a problem with the pump controller electronics Pump over temperature illuminates if the pump is over its operating temperature range. The pump will stop until its temperature returns to the operating range. The pump will typically overheat if the sample dispersant is too viscous or the sample material is too large, causing jamming. Pump speed control illuminates if the difference between the speed demanded and the actual speed exceeds 500rpm for over 10 seconds. Ultrasound illuminates if there is an error with the ultrasound control electronics. Ultrasound displacement illuminates if the current supplied to drive the ultrasound probe is too high. Ultrasound timeout illuminates if the ultrasound is left on for over 15 minutes. Status indicators These indicators give the current status of the pump and ultrasonic probe. If this indicator shows green then the pump or ultrasonic probe are currently in use. If the Ultrasonics indicator is greyed out then the unit does not have the ultrasonic option installed. Page 3-4 MAN 0389

29 Software features Chapter 3 Writing an SOP for the dispersion unit A Standard Operating Procedure (SOP) can be configured to control all settings for the dispersion unit automatically. When an SOP is run the software (depending on how the SOP is set up) automatically runs a measurement, giving instructions to the operator such as when to add dispersant, sample, etc. The SOP even reports if the wrong cell unit is fitted. The SOP wizard is described in full in the main User Manual. Here we just describe those dialogues which are specific to the Hydro 2000µP. These are the Sampler Selection and Sampler Settings dialogues. Sampler Selection dialogue The Sampler Selection dialogue is shown below: 1 2 It features: ill 7774 Sample handling unit list Select the dispersion unit to be used from the drop down list. All dispersion units available will be listed, even those that are not attached to the system. This allows an SOP to be written remotely from the system. All systems the SOP is distributed to must have the dispersion unit connected. The system allows up to two of each dispersion unit to be connected to the optical bench. In this situation, the dispersion units will be identified as unit A and B - in Hydro 2000µP Page 3-5

30 Chapter 3 Software features the case of this dispersion unit there would be an entry for Hydro 2000µP (A) and Hydro 2000µP (B). If using just one dispersion unit, select the A option. Sample handling unit options This dispersion unit is supplied in two versions, with ultrasonics and without ultrasonics. Use this pull down list to select the model in use. If the user tries to run an SOP that has been defined for an dispersion unit with ultrasonics on an dispersion unit without that option, a message is displayed, telling them to connect the correct dispersion unit. Sampler Settings dialogue The Sampler Settings dialogue is shown below: Its features are: ill 7775 Pump speed The pump speed is altered by moving the slider or entering the required speed in the RPM box to the right of the slider. Selecting the Advanced button displays the Vari-flow options. Some materials such as metal flakes have been found to form structures when they are circulated through the cell at normal pump speeds. These structures give rise to a secondary diffraction pattern that is incorrectly interpreted as a small peak at 1000 microns. Vari-flow allows these materials to be dispersed in the well in the normal way and then the pump is briefly turned off or reduced before measurement of the sample. When the pump is turned off, the structures in the sample are randomised and the Page 3-6 MAN 0389

31 Software features Chapter 3 measurement can be made without the occurrence of the secondary diffraction pattern. When the flake material is suspended in a volatile solvent, such as white spirit, variflow is even more useful since it assists the rapid thermal stabilisation of the optics. When the sample well is filled with fresh solvent, very high backgrounds can often be observed. These arise from thermal gradients in the solvent causing steering of the laser beam onto the inner rings of the internal detector. As the temperature equilibrates, the background will fall to normal levels (80-100) and the measurement can proceed. Pausing the pump after the solvent has had a chance to thoroughly disperse through the system will accelerate the fall of the background to normal levels. Ultrasonics Warning! When using solvents, especially those with low ignition points, minimise the use of ultrasonics. There is a remote possibility that localised heating caused by the ultrasonic action may cause an explosive reaction. Check the Material Safety Data Sheets for all substances used. The ultrasonics radio buttons enable the ultrasonic transducer. The options are: None no ultrasound. Continuous (from sample addition) ultrasound is activated after sample is added and will run continuously. Pre-measurement ultrasound is activated for a set time prior to measurement. For pre-measurement mode, type in the required duration in seconds in the Duration box. Continuous (from start) ultrasound is active throughout the measurement. Ultrasound will start after the electrical background is complete. The stabilising period will add a delay between the electrical background and the optical alignment. During this delay ultrasound will be active to allow bubbles to be driven from the dispersant. Either click and drag the slider bar to set the level, or type the exact level required in the box to the right of the slider bar. Determine the optimum ultrasound level and duration using manual control, ensuring that the lowest setting giving satisfactory dispersion is obtained. Record the settings and transfer these to the SOP in the Sampler Settings dialogue using the slider controls. Hydro 2000µP Page 3-7

32 Chapter 3 Software features Pre-Measurement Delay A pre-measurement delay can be added before the measurement to allow thermal equilibration to occur after ultrasound has been applied. If this is not done the analysis may show the presence of large particles which are not in the sample. This is caused by the refractive index differences observed in the sample cell when the dispersant is not equilibrated. Choose a delay sufficiently long enough for adequate thermal equilibration to occur. To set the delay, highlight either the minutes or seconds in the entry box, then use the selection arrows to alter the time; alternatively the time can be typed in. A delay of between 0 and minutes can be chosen. Temperature Monitoring It is possible to monitor the temperature of the sample and dispersant and to set the software to only make measurements between specified limits. Temperature monitoring is enabled by selecting the Allow measurements only. check box. The options are: Lower/upper limit this section specifies the temperature range within which the sample/dispersant must be for the measurement to start. This limit must be in the range of 10 to 35 C. Stabilising temperature the time for which the temperature must be stable before the measurement proceeds. Time-out period the measurement will be abandoned if the temperature is not stabilised within this time period. Page 3-8 MAN 0389

33 4 Making a measurement Introduction This chapter describes how to make a measurement using the Hydro 2000µP. It covers: Measurement guidelines. Instrument preparation. Anaerobic fill. Measuring new samples. Hydro 2000µP Page 4-1

34 4 Chapter 4 Measurement guidelines Making a measurement The Hydro 2000µP uses very low volumes (approximately 20ml) of dispersant and sample. When using such small quantities, take the following precautions when making a measurement: Minimise the amount of sample lost during sample preparation, or through sample adhering to the sides of the well. Cleanliness of the system is critical: very small amounts of cross-contamination can bias the results. Manage the pump speed to prevent the introduction of bubbles, while also ensuring that the sample is kept in suspension. This is true for all wet sample dispersion units but particularly important with the small dispersant volumes of the Hydro 2000µP. The basic procedure Add dispersant to the well by injecting the requested dispersant through the inlet port. Inject the dispersant slowly until the level of the dispersant is within 1 to 2mm of the internal rim of the well. Because the stirring and the circulation of the sample is carried out in the chamber below the well cone, keep the level of dispersant in the well low. If sample is seen adhering to the sides of the well cone, use a pipette to withdraw some of the dispersant from the well to rinse the sides of the well. Once the sample has been added, it is important to replace the well lid and lock it in place using the locking lever. Minimising sample loss Sample loss by adhesion To prevent sample adhering to the side of the sample well during sample addition, it is typical to have the dispersant at a level a quarter of the way up the well. This allows the sample to be taken into the well cavity by the dispersant. A typical procedure is: 1. Fill the well to 1-2mm below the top of the well. 2. Run an anaerobic fill sequence (press the Anaerobic fill button on the accessory control dialogue). 3. Once complete, use a pipette to remove dispersant so that the well is about a quarter full. Page 4-2 MAN 0389

35 Making a measurement Chapter 4 Sample loss during sample preparation Choose a sample preparation method that ensures that all particles are transferred to the well. A recommended method is to prepare a slurry in a small container and add this to the well using a pipette. Add as a maximum 2ml of sample volume, but 1ml is preferred. Material will usually be seen collecting in the bottom of the container. To take a representative sample of the material: 1. Make the slurry more dilute than is normal for larger volume sample dispersion units. This will ensure that adding slurry does not lead to an excessive concentration. 2. Tilt the container at an angle. 3. Agitate the dispersant by continually filling and discharging the pipette to ensure the material is suspended in the dispersant (rather than sinking to the bottom). 4. Once dispersed, finally fill the pipette with as much of the dispersant as possible and transfer it to the well drop by drop, pausing between drops to allow the sample to fully disperse i.e. monitor the obscuration value until it stabilises. 5. When adding the sample, ensure that it is added directly to the dispersant rather than trickling it down the sides of the well walls. This prevents sample loss through adhesion. Cleanliness Cleanliness is essential; since only 10-50mg of sample is added, it only takes a small amount of cross-contamination from a previous measurement to bias results significantly. Consider the following points when making measurements: When cleaning, rinse the system three or four times. It doesn t use much dispersant, and gives better performance for the next sample. The first rinse of a clean is sometimes best done by injecting 5-10ml bursts of dispersant into the well while the pump is running and the drain valve open. This tends to rinse larger particles through the system. When changing dispersants, blow each dispersant out of the fill tube with a syringe full of air (with the drain open) to reduce cross contamination. When changing over from an oily organic solvent (vegetable oil, etc.) to water, use a biological drain cleaner (containing food enzymes) to clean out the last traces of oil. Hydro 2000µP Page 4-3

36 Chapter 4 Making a measurement Bubbles and pump speeds Control the pump speed to prevent the introduction of bubbles while ensuring that the sample is kept in suspension. This is true for all wet dispersion units but particularly important with the small dispersant volumes of the Hydro 2000µP. The Mastersizer sees bubbles as particles and measures these. Bubbles will vary in size but are typically in the region of 100 microns. In many cases bubbles can be clearly seen as a separate peak when the measurement data is analysed. Consider the following points when making measurements: When filling the system with dispersant, turn off the pump to prevent it introducing bubbles. Fill the system with dispersant slowly to avoid trapping bubbles. When the system is first filled, bring the pump speed up slowly to clear any remaining bubbles, without breaking them into smaller ones. Ideally use the Anaerobic fill sequence (described later) to remove air and bubbles from the system. Take care with the pump speed. The system is designed, and balanced, for water. Other dispersants may tend to well up in the well or pump chamber as the pump speed increases. Full speed (5000 rpm) is too much for most dispersants, so do not use 100% pump by default. See the table in the Anaerobic fill section later for examples of settings. Choose a pump speed high enough to keep the particles in suspension. Using too high a pump speed will tend to draw air into the system. Always fit the well lid during measurements. Finally, on the last wash before filling the system to make a measurement, we recommend that the pump is set to off whilst draining. This is because the action of running the pump whilst draining can cause the formation of small air bubbles in drops of dispersant that may remain in the system after draining. If the system is then filled immediately in order to perform a measurement, these air bubbles are circulated and sized in addition to the required sample. Page 4-4 MAN 0389

37 Making a measurement Chapter 4 Instrument preparation Before starting a measurement, ensure that the correct cell unit is installed in the cell area and then power up the system. The stages are outlined below. Powering up Power up the system by: Switching on the dispersion unit(s). Switching on the optical bench. Switching on the computer and starting the software by double-clicking on the Mastersizer 2000 icon. Filling the cell with clean dispersant Before a measurement fill the cell with clean, degassed dispersant (when the measurement is run the software asks if this has been done). Take care not to let air (either dissolved gasses or bubbles) into the system. The following procedure is recommended to fill the cell unit. Degassing the dispersant Gases dissolved in the dispersant can form bubbles in the cell. To prevent this from happening, we recommend that the dispersant is degassed prior to adding it to the cell unit. The following procedure is suggested if water is used as a dispersant. Warning! This procedure may not be suitable for all dispersants, especially dispersants with low boiling/ignition points. To degas water: 1. Choose a vessel that has a lid and fill the vessel with water. Do not fill the vessel completely at this point. 2. Loosely fit the lid, but do not tighten it. This ensures that pressure does not build up in the vessel, while preventing dust getting in. 3. Place the open vessel in an ultrasonic bath that can be heated. 4. Set the ultrasonic power to 60W and the water bath to 70 C. 5. Degas for 20 minutes. Hydro 2000µP Page 4-5

38 Chapter 4 Making a measurement Vessel Heater Ultrasonic bath ill Once the water has been degassed, a number of bubbles may have formed on the inside surface of the vessel. To remove these, fit the lid to the vessel and very carefully turn the vessel on its side (see the diagram below). Slowly rotate the bottle through one complete revolution - the bubbles will be removed as they meet the air/fluid interface. Bubbles Filling the syringe Take care to avoid air entering the syringe when filling it with dispersant. ill 5639 Warning! Avoid using a syringe with a capacity of more than 20ml. The additional capacity may cause the cell to overflow, causing hazardous conditions within the cell unit. Page 4-6 MAN 0389

39 Making a measurement Chapter 4 To fill the syringe: 1. Insert the syringe into the vessel of dispersant. 2. Draw back the syringe very slowly. Pulling back too quickly will cause boiling on the surface of the syringe plunger (see the diagram below) or air may be dragged back through the syringe seal. Syringe Syringe plunger Boiling Syringe body ill Hold the syringe with the tip pointing upwards, and push the plunger in to displace any air (as shown below). It may be necessary to tap the body of the syringe to displace any air bubbles. ill 5641 i Note Avoid having to refill the syringe half way through filling the cell, as disconnecting and reconnecting the syringe may introduce air into the cell. When filling the standard 20ml syringe, fully draw back the plunger to take up enough dispersant to fill the cell (its approximate capacity is 18ml). Hydro 2000µP Page 4-7

40 Chapter 4 Making a measurement Filling the cell Finally, fill the cell with the clean dispersant. To fill the cell: 1. Connect the syringe to the valve attached to the front of the cell unit. 2. Hold the syringe with the plunger facing upwards (as shown below) so that any bubbles in the syringe rise to the top and are not injected into the cell. Pipe to cell unit Valve 3. Filling the cell now depends on which of the two types of dispersant valve being used: ill 5642 For the non-aqueous valve (suitable for use with a large range of chemically aggressive dispersants) open the tap on the side of the valve. Slowly press the plunger of the syringe (typically over about a 20 second period). Fill the cell until the sample well is almost full (to within 1 or 2mm of the internal well rim. Once full, close the tap on the side of the valve. Note that there is no need to remove the syringe from the valve. The aqueous valve (used with non aggressive dispersants such as water) does not have a tap to control the flow through the valve but acts as a nonreturn valve. To fill the cell, slowly press the plunger of the syringe (typically over about 20 seconds). Fill the cell until the sample well is almost full (to within 1 or 2mm of the internal well rim). i Note For the valve to function correctly as a non-return valve, the syringe must be removed from the valve once the cell is full. Page 4-8 MAN 0389

41 Making a measurement Chapter 4 Anaerobic fill Due to the small volume of the Hydro 2000µP and associated problems of removing bubbles from the system, a special fill sequence called Anaerobic fill can be used. During an anaerobic fill the pump speed is increased in steps to a target speed. A short pause is added between steps to allow bubbles to disperse. To use anaerobic fill click the Anaerobic fill button on the accessory control dialogue s Control tab. The Anaerobic fill dialogue will appear. This contains a field for setting the final pump speed. This is the speed the pump will run at after the fill process has finished. During a manual measurement the last value used will be recalled. During an SOP this will be the pump speed specified in the SOP. For best results set the pump speed field 500rpm above the final speed required. Once the Anaerobic fill sequence is complete the final pump speed can be reset. Example values are given below. Material Pump speed (SOP) Anaerobic fill Glass Beads in: Water Oils Alcohols 2500rpm 3000rpm 2500rpm 3000rpm 3500rpm 3000rpm Latex in water 2000rpm 2500rpm The dialogue prompts the user to fill the well. Once it s full press the Go button. As bubbles escape the level of the dispersant may drop. Once complete, ensure that the well is approximately a quarter full, using a pipette to remove excess dispersant if necessary. Hydro 2000µP Page 4-9

42 Chapter 4 Making a measurement Measuring new samples The above procedures are ideal if the SOP is set up or all the parameters for a manual measurement have been specified. Users measuring a new material must either create or modify an SOP or set up the options for a new manual measurement. This section outlines the setup for measuring a new sample. Measuring new samples - manual measurement The basic procedure is the same as outlined above. When the user selects the Options button in the measurement display, the Measurement Options dialogue appears. This dialogue controls all aspects of the measurement. Use the Materials tab to select a different material and dispersant or define the optical properties of the new material. This tab also allows users to specify options regarding the result calculation. i Note If a material is created or modified the computer calculates a new scattering model. This usually takes between one and 10 minutes, depending on the power of the computer. However, if the ratio of the refractive index of the sample to the dispersant is in the range 0.99 to 1.01, then this time may substantially increase. Under these conditions we recommend that the model is generated overnight. Use the Measurement tab to set up measurement times, set measurement alarms, etc. Press the Document button in the Measurement display to open the Document dialogue. Add relevant details that will allow a user to reproduce the measurement at a later date. All the options available are basically the same when setting up an SOP. For more details of these options see Chapter 4 of the main User Manual. Measuring new samples - SOP measurements Creating an SOP is easy. Select Configure-New SOP; the software will run through an SOP Wizard. This wizard asks a series of questions, such as: what are the optical properties of the sample? what is it suspended in? how fast should the pump be? how much ultrasonic dispersion is required? do any additives need to be added to the sample to aid dispersion? Page 4-10 MAN 0389

43 Making a measurement Chapter 4 how will the result be calculated and displayed? These are all the questions that need to be answered in order to define a measurement. Anyone can create an SOP but typically it is the responsibility of the system administrator. An SOP can be created by a specialist and then distributed to all Mastersizer 2000 users in an organisation. Remember that no instrument needs to be connected to the software to write an SOP. The Malvern software can be installed on a remote computer and SOPs created and edited before they are tried out on an instrument. (The Essentials Manual has details on installing the software on another computer.) A completely new SOP can be created by selecting Configure-New SOP and following the SOP Wizard. An existing SOP can be easily edited either to correct it or to create a new SOP based on it. If several slightly varying SOPs are being saved, it is useful to include the differentiating features in the SOP name e.g. glopalene ultrasound 60, glopalene ultrasound 40. To edit an SOP select Configure-Existing SOP. A tabbed dialogue will appear. Each tab corresponds to a screen from the SOP Wizard. Select a tab and change any of the entries as required. Click OK to save the changes. To create a new SOP based on an existing one, first modify the existing SOP using Configure-Existing SOP. Once all the changes have been made, the Save-As dialogue will appear. Save the SOP under a different name. Hydro 2000µP Page 4-11

44 Chapter 4 Making a measurement Page 4-12 MAN 0389

45 5 Changing cell holder position Introduction The Essentials Manual covers most maintenance procedures. This chapter describes one non-essential procedure, how to move the cell holder from its position on the right of the optical bench, should this be necessary. The Hydro 2000µP is typically configured to be positioned on the right of the optical bench. For users who wish to locate the dispersion unit on the left of the optical bench we recommend that the cell holder be swapped to the other side of the control unit. 1 2 ill 5433 Hydro 2000µP Page 5-1

46 5 Chapter 5 Changing cell holder position To swap over the cell holder: 1. Disconnect all connectors from the control unit. 2. Carefully tip the control unit and the cell holder onto its back to expose the feet and connection plate (as shown on the illustration below). 3. Using a 4mm Allen key, remove the eight screws securing the connection plate. 4. Swap the cell holder to the other side of the control unit. 5. Flip the connection plate over and screw it back into position. 6. Unscrew the four feet and move them to the alternative locations (now located at the outside edge of the assembly) ill 5434 Page 5-2 MAN 0389

47 Part 2 - Appendices

48

49 A Specification Specification Dispersion type Capacity Dispersion mechanisms Modes of operation Weight Cell Control box/cell holder Dimensions Cell Control box/cell holder Particle size range Power requirement Wet. Approx. 20ml AWA Continuously variable and independent pump and ultrasound. AWA Continuously variable pump. Automatic via SOPs. Manual via computer on-screen operating dialogues. 4.25kg (AWA2003), 4kg (AWA2004) 8kg (AWA2003), 7.75kg (AWA2004) Width 108mm Height: 242mm (excluding motor and handle) Height: 335mm (including motor and handle) Depth: 252mm Width: 284mm Height: 338mm Depth: 252mm µm, depending on particle shape and density V / V AC, 48W Only use the PSU and cables provided by Malvern. Using any other PSU may compromise safety and will void any warranty. Hydro 2000µP Page A-1

50 A Appendix A Pump Speed Accuracy Max dispersant viscosity Safety: Ultrasound Frequency Power Safety Specification 0 and rpm rpm, +/-5% of F.S Poise (0.054 Pa.s) Over current and over temperature protected. 48 khz (nominal) 0-20W (dependent upon dispersant) Over current protected. Page A-2 MAN 0389

51 B Chemical compatibility Introduction The Hydro 2000µP is manufactured from materials that give the widest protection from chemical attack. However, it is important to check that any sample or dispersant used is chemically compatible with the materials it will come into contact with in the dispersion unit. This appendix lists all materials that come into contact with the sample and dispersant during normal operation. Component Cell, pump cavity, drain, well, drain tube Internal pipes Impeller, input syringe tube, non-aqueous dispersant inlet valve Drain sump Windows Internal seals Drain and overflow pipes Non-return valve, well lid Beaker Materials 316 stainless steel FEP PTFE Polypropylene Glass Perlast (G70G) Tygon (MH2075) PEEK Pyrex Hydro 2000µP Page B-1

52 B Appendix B Chemical compatibility Page B-2 MAN 0389

53 C Regulatory statements This appendix presents regulatory information. Hydro 2000µP Page C-1

54 C Appendix C CE Declaration of Conformity Regulatory statements The CE badge on this product, shown below, signifies conformance to European Commission Directives. ill 8005 Page C-2 MAN 0389

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02 ! Read this first Zetasizer nano series Self installation and Quick start guide I N S T R U M E N T S MRK825-02 Zetasizer Nano series Self installation and Quick start guide MAN0383 Issue 1.1 July 2007

More information

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI NanoSight NS300 Operation instructions Injection/flushing brief overview: 1. Do not exceed flow of 1 ml per 20 seconds. 2. Inject two 1 ml syringes with nano-pure or DI water. 3. If the water does not

More information

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the

More information

6900 Maintenance Instruction System Flush

6900 Maintenance Instruction System Flush Equipment Required FA74005 Damper Drain Tube FA16005 Cover Removal Tool FA900005 Beaker 0.25 Litre FA900003 Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves Latex

More information

ORB-400 BUBBLE/HAZE MACHINE. Item ref: UK User Manual

ORB-400 BUBBLE/HAZE MACHINE. Item ref: UK User Manual ORB-400 BUBBLE/HAZE MACHINE Item ref: 160.462UK User Manual Caution: Please read this manual carefully before operating Damage caused by misuse is not covered by the warranty Introduction Thank you for

More information

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1 BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1 The front door of the main body of the instrument may be open when you arrive. Take the sample chamber and slide it into position with the

More information

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet: OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle

More information

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS 2000 Dionex Corporation Document No. 031277 Revision 03 April 2000 2000 Dionex Corporation All rights reserved worldwide. Printed in the United States of

More information

Using the Akta Prime plus October 22, 2012

Using the Akta Prime plus October 22, 2012 Some starting precautions: 1. Vacuum filter all buffers. Removes any large particles/debris that may clog your column De-gases the buffers 2. Clarify lysates first by centrifugation and then filtration

More information

SOP for use of the SX19 Stopped-Flow Instrument

SOP for use of the SX19 Stopped-Flow Instrument SOP for use of the SX19 Stopped-Flow Instrument The main components of the stopped flow system are: Always book the system before use. Starting up and preparing the SX19 for use: 1. Turn on the compressed

More information

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008 UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method

More information

6800 Maintenance Instruction System Flush Procedure

6800 Maintenance Instruction System Flush Procedure Equipment Required FA74005 FA65318 FA900005 FA900003 Damper Drain Tube 6800 Cover Removal Tool Beaker 0.25 Litre Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves

More information

BD Cytopeia Fluidic Kit User s Guide

BD Cytopeia Fluidic Kit User s Guide BD Cytopeia Fluidic Kit User s Guide For Research Use Only 23-17618-00 9/2015 Becton, Dickinson and Company BD Biosciences 2350 Qume Drive San Jose, CA 95131 USA BD Biosciences European Customer Support

More information

BGA244 Binary Gas Analyzer

BGA244 Binary Gas Analyzer Quick Start Guide Revision 1.0 Certification Warranty Service certifies that this product met its published specification at the time of shipment. This product is warranted against defects in materials

More information

Armfield Distillation Column Operation Guidelines

Armfield Distillation Column Operation Guidelines Armfield Distillation Column Operation Guidelines 11-2016 R.Cox Safety SAFETY GLASSES ARE REQUIRED WHEN OPERATING THE DISTILLATION COLUMN Wear gloves when mixing alcohol feedstock The column will become

More information

Microfluidic Demonstration

Microfluidic Demonstration Microfluidic Demonstration Co-flow of liquids across a microfluidic chip Create a microfluidic controller to flow liquids across a microfluidic chip. 1. Assemble fluidic circuits and attach to breadboard.

More information

Calcium Reactor. CalcFeeder PRO. User Manual 1.1a. Model. CalcFeeder AC1 PRO CalcFeeder AC2 PRO CalcFeeder AC3 PRO CalcFeeder AC4 PRO

Calcium Reactor. CalcFeeder PRO. User Manual 1.1a. Model. CalcFeeder AC1 PRO CalcFeeder AC2 PRO CalcFeeder AC3 PRO CalcFeeder AC4 PRO Calcium Reactor CalcFeeder PRO User Manual 1.1a Model CalcFeeder AC1 PRO CalcFeeder AC2 PRO CalcFeeder AC3 PRO CalcFeeder AC4 PRO Page 1 Dear Customer, Thank you for purchasing an Pacific Sun CalcFeeder

More information

SECTION 2 SMART PAYOUT MANUAL SET FIELD SERVICE MANUAL

SECTION 2 SMART PAYOUT MANUAL SET FIELD SERVICE MANUAL SECTION 2 SMART PAYOUT MANUAL SET FIELD SERVICE MANUAL Innovative Technology assume no responsibility for errors, omissions, or damages resulting from the use of information contained within this manual.

More information

VLI Visible Level Interface Operation & Maintenance Manual

VLI Visible Level Interface Operation & Maintenance Manual VLI Visible Level Interface Operation & Maintenance Manual All rights reserved. Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted

More information

ACV-10 Automatic Control Valve

ACV-10 Automatic Control Valve ACV-10 Automatic Control Valve Installation, Operation & Maintenance General: The Archer Instruments ACV-10 is a precision automatic feed rate control valve for use in vacuum systems feeding Chlorine,

More information

The Univentor 1250 Anaesthesia Unit

The Univentor 1250 Anaesthesia Unit THE UNIVENTOR 1200/1250 ANAESTHESIA UNIT The Univentor 1250 Anaesthesia Unit TABLE OF CONTENTS EDITION 1 Section 1 - WARRANTY & SERVICE 1.1. WARRANTY 2 1.2. DAMAGED SHIPMENTS 2 1.3. SERVICE 2 Section 2

More information

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev -

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev - 1 W. W. Hansen Experimental Physics Laboratory STANFORD UNIVERSITY STANFORD, CALIFORNIA 94305-4085 Gravity Probe B Relativity Mission OPERATION OF LIQUID PARTICLE COUNTER GP-B P0045 Rev - May 23, 1997

More information

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings SomnoSuite FAQ V1.3 January 2015 Setup 1. How do I connect the SomnoSuite to my oxygen source? 2. Is there a way to speed up the downward movement of the pusher block when setting the empty position? 3.

More information

1)! DO NOT PROCEED BEYOND THIS MARK

1)! DO NOT PROCEED BEYOND THIS MARK Operating Instructions for X-ray Photoelectron Spectrometer: Physical Electronics Model 555 XPS/AES (John H. Thomas, III, Ph.D., Electron Spectroscopy) Sample Insertion: figure 1. Sample insertion rod

More information

REACTOR 40 MECHANICAL Configuration Guide

REACTOR 40 MECHANICAL Configuration Guide REACTOR 40 MECHANICAL Configuration Guide Important Safety Information WARNING See the Important Safety and Product Information guide in the product box for product warnings and other important information.

More information

Pipettor. User Manual

Pipettor. User Manual Pipettor User Manual Product Code Description Single Chanel Pipettor Variable 550.002.055 Volume 0.5 to 10ul 550.002.060 2 to 20ul 550.002.065 10 to 100ul 550.002.070 20 to 200ul 550.002.075 100 to 100ul

More information

Lipodrop 2.0 device coating

Lipodrop 2.0 device coating Lipodrop 2.0 device coating Compiled by K.V., K.Ž, J.R AIM: To selectively coat channels in order to make them hydrophobic (pre junction channels) or hydrophilic (post junction channels). REAGENTS USED:

More information

FRDS GEN II SIMULATOR WORKBOOK

FRDS GEN II SIMULATOR WORKBOOK FRDS GEN II SIMULATOR WORKBOOK Trotter Control Inc 2015 Document# Revision Revised 9001-0038 FRDS GEN II Simulator Workbook E 02/15/2015 by DC FRDS GEN II Simulator Workbook This workbook is a follow-on

More information

256 Pneumatic Pressure Indicator

256 Pneumatic Pressure Indicator 256 Pneumatic Pressure Indicator 51425699 Copyright 2002 Slope Indicator Company. All Rights Reserved. This equipment should be installed, maintained, and operated by technically qualified personnel. Any

More information

Vapor Sorption Analyzer

Vapor Sorption Analyzer Vapor Sorption Analyzer Quick Start Guide 1. Getting Started Place the AquaLab VSA on a flat surface. Use the bubble level(figure 1: Adjustment and bubble level) and use the adjustable feet to level the

More information

SOP: Buck Scientific BLC-20P HPLC Operation

SOP: Buck Scientific BLC-20P HPLC Operation Page 1 of 11 Approvals Preparer: John Buford Reviewer: Tim Kull Reviewer: Dr. Margaret Bryans Date: 16OCT13 Date: 30OCT13 Date: 31OCT13 1. Purpose 1.1. Basic operation of the Buck Scientific BLC-20P isocratic

More information

Product Manual B-Safety ClassicLine & PremiumLine Emergency Eyewash and Eye/Face Wash Equipment

Product Manual B-Safety ClassicLine & PremiumLine Emergency Eyewash and Eye/Face Wash Equipment Product Manual B-Safety ClassicLine & PremiumLine Emergency Eyewash and Eye/Face Wash Equipment 1. Application Emergency eyewash and eye/face wash equipment are prescribed first aid installations for workplaces

More information

Budget Range Operators Handbook

Budget Range Operators Handbook Budget Range Operators Handbook BAMBI AIR COMPRESSORS LTD 152 Thimble Mill Lane Heartlands Birmingham B7 5HT United Kingdom Tel: 0121 322 2299 Fax: 0121 322 2297 Email: sales@bambi-air.co.uk www.bambi-air.co.uk

More information

User Manual for the Mars Calibration Bench

User Manual for the Mars Calibration Bench User Manual for the Mars Calibration Bench Fall 2013 Table of Contents Table of Contents Table of Contents... iii Introduction... v Chapter 1: The Mars Calibration Bench... 1 What Is the Mars Calibration

More information

User Manual. Quantos Automated Dosing Liquid Module

User Manual. Quantos Automated Dosing Liquid Module User Manual Liquid Module 1 Safety Information 1.1 Definition of warnings and symbols Signal Words WARNING for a hazardous situation with medium risk, possibly resulting in severe injuries or death if

More information

High Pressure Chem-SCAN Operating Manual

High Pressure Chem-SCAN Operating Manual GAS INLET VALVES REACTIVE GAS PRESSURE RELIEF VALVE INERT GAS VENT VENT VALVE REACTOR INLET VALVES PRESSURE TRANSDUCERS REACTORS STIRRER & THERMOWELL HEATING JACKET STIRRER MOTORS High Pressure Chem-SCAN

More information

AKTA pure 25 New Owner s Intro

AKTA pure 25 New Owner s Intro AKTA pure 25 New Owner s Intro The exercise below will give a quick demonstration of how easy and intuitive the AKTA pure 25 will be for you in demonstrating downstream processing to your students. Steps

More information

Aerobic Respiration. Evaluation copy

Aerobic Respiration. Evaluation copy Aerobic Respiration Computer 17 Aerobic cellular respiration is the process of converting the chemical energy of organic molecules into a form immediately usable by organisms. Glucose may be oxidized completely

More information

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents AQUARIUS Operating Instructions for the Type 70 Electrolytic Gas Soldering / Welding Unit Table of Contents 1. Important General Remarks 2 2. Safety Instructions 2 3. Description of the Soldering Unit

More information

Instructions for Installation, Use and Maintenance FSI Safety Tank Shower Systems

Instructions for Installation, Use and Maintenance FSI Safety Tank Shower Systems Instructions for Installation, Use and Maintenance FSI Safety Tank Shower Systems per DIN 12 899 T1-3 and ANSI Z 358-2004 Table of Contents for Instructions for Use Section 1 Manufacturer, Safety Shower

More information

INSTRUCTION MANUAL FOR MODEL 7360V VERTICAL CURING CHAMBER Revision E May

INSTRUCTION MANUAL FOR MODEL 7360V VERTICAL CURING CHAMBER Revision E May INSTRUCTION MANUAL FOR MODEL 7360V VERTICAL CURING CHAMBER Revision E May 2015 98-0520 S/N 2001 N. Indianwood Ave. Tulsa, Oklahoma 74012 U.S.A. TEL: (918) 250-7200 FAX: (918) 459-0165 E-mail: chandler.sales@ametek.com

More information

VERTICAL AIR COMPRESSORS

VERTICAL AIR COMPRESSORS VERTICAL AIR COMPRESSORS MODEL NO: VE11C150, VE15C150, VE18C150 PART NO: 2226005, 2226000, 2226015 OPERATION & MAINTENANCE INSTRUCTIONS LS0615 INTRODUCTION Thank you for purchasing this CLARKE Vertical

More information

Instruction Manual MediPrep 120

Instruction Manual MediPrep 120 Instruction Manual MediPrep 120 ERWEKA GmbH * Ottostr. 20 22 * 63150 Heusenstamm * Germany * Tel: +49 (0) 6104 6903-0 Fax: +49 (0) 6104 6903-40 * E-mail: info@erweka.com * www.erweka.com Directors: Werner

More information

[USING THE VITROBOT April 9, Using the Vitrobot

[USING THE VITROBOT April 9, Using the Vitrobot Using the Vitrobot The Vitrobot Mk 3 (aka Mark III, Mark 3, etc.) can control both the local temperature and the local humidity during the process of plunge-freezing a grid for cryoem. It does require

More information

UsER manual for Watersens ph -REDOX

UsER manual for Watersens ph -REDOX UsER manual for Watersens -REDOX Cl 8 1 2 6 3 3 7 7 4 4 4 4 Parts List 1 Redox Probe 1 x 2 PH Probe 1 x 5 Tube Weight 2 x 6 Connection Valve 1 x chlorine 3 Chlorine and Pumps 2 x 7 Dosing Valve 2 x 5 5

More information

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 QUICK START GUIDE P/N 6006154, REVISION C MAY 2013 Model 8120 Mask Integrity Tester is patented under U.S. Patent No. 8,312,761. Additional patents are pending.

More information

RayPette Plus Autoclavable Pipette. User Manual

RayPette Plus Autoclavable Pipette. User Manual RayPette Plus Autoclavable Pipette User Manual CONTENTS 1. YOUR NEW PIPETTE... 1 1.1. Adjustable volume pipettes... 1 1.3 Fully autoclavable... 2 2. UNPACKING... 3 3. INSTALLING THE PIPETTE HOLDER... 4

More information

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road PROPORTIONING VALVE Model 150 INSTRUCTION MANUAL March 2017 IMS Company 10373 Stafford Road Telephone: (440) 543-1615 Fax: (440) 543-1069 Email: sales@imscompany.com 1 Introduction IMS Company reserves

More information

The pipettes cover a volume range from 0.1µl to 10ml.

The pipettes cover a volume range from 0.1µl to 10ml. CONTENTS 1. YOUR NEW PIPETTE 1 1.1. Adjustable volume pipettes 1 1.2. Fixed volume pipettes 2 1.3 Fully autoclavable 3 2. UNPACKING 3 3. INSTALLING THE PIPETTE HOLDER 4 4. PIPETTE COMPONENTS 5 5. PIPETTE

More information

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure PPMS Service Note 1099-412 In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure Introduction: Loss of electricity to the PPMS EverCool II should not cause damage to

More information

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...

More information

RM-80 respiration monitor

RM-80 respiration monitor RM-80 respiration monitor User Manual September 18, 2015 0025-003M 950 North Hague Avenue Columbus, Ohio 43204-2121 USA Sales: sales@colinst.com Service: service@colinst.com Phone: (614) 276-0861 Fax:

More information

Instructions for Use

Instructions for Use Select-380 T-Auto T-Auto Contents Page Instructions for Use Ref: 3.0 IFU 380 T-Auto Mar 18 2 Schematic layout of the doser 3 Quick-fit instructions 4 Description/Installation/Operation Pump tubes & Water

More information

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA Page 1 of 9 SSI Solaris 150 RTA The Solaris 150 RTA is a rapid thermal annealing system capable of handling sample sizes up to 100mm (4 diameter) or smaller. The system can anneal in N 2 and Forming gas

More information

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text. Hach LDO Sensor Instruction Sheet 00745589 Safety Precautions Please read this entire instruction sheet before operating this sensor. Pay particular attention to all danger and caution statements. Failure

More information

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer July 2011 OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer WARNING! Before operating this product, read and understand this Operator s Manual. Become familiar with the potential hazards of this unit. Contact

More information

VERTICAL AIR COMPRESSORS

VERTICAL AIR COMPRESSORS VERTICAL AIR COMPRESSORS MODEL NO: VE15C150, VE18C150, VE25C150 PART NO: 2226010, 2226020, 2226025 OPERATION & MAINTENANCE INSTRUCTIONS LS0715 INTRODUCTION Thank you for purchasing this CLARKE Vertical

More information

Boyle s Law: Pressure-Volume Relationship in Gases

Boyle s Law: Pressure-Volume Relationship in Gases Boyle s Law: Pressure-Volume Relationship in Gases The primary objective of this experiment is to determine the relationship between the pressure and volume of a confined gas. The gas we will use is air,

More information

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures

More information

MP15 Jockey Pump Controller

MP15 Jockey Pump Controller Setup and Operating Instructions MP15 Jockey Pump Controller This manual provides general information, installation, operation, maintenance, and system setup information for Metron Model MP15 Jockey Pump

More information

THE UNIVENTOR 400 ANAESTHESIA UNIT

THE UNIVENTOR 400 ANAESTHESIA UNIT THE UNIVENTOR 400 ANAESTHESIA UNIT TABLE OF CONTENTS EDITION 3 Section 1 - WARRANTY & SERVICE 1.1. WARRANTY 2 1.2. DAMAGED SHIPMENTS 2 1.3. SERVICE 2 Section 2 - INTRODUCTION 2.1. INTRODUCTION 3 Section

More information

STARLOG. Capacitive Water Depth Probe

STARLOG. Capacitive Water Depth Probe STARLOG Capacitive Water Depth Probe Model 6521 User Manual Supplement 6219 Revision D July 10. 1998 Copyright Notice Copyright Unidata Australia 1998. All rights reserved. No part of this publication

More information

NanoSight LM14. Operating Manual P551J NanoSight Ltd. NanoSight Ltd., Minton Park, Amesbury, Wiltshire SP4 7RT, UK

NanoSight LM14. Operating Manual P551J NanoSight Ltd. NanoSight Ltd., Minton Park, Amesbury, Wiltshire SP4 7RT, UK NanoSight LM14 Operating Manual P551J 2012 NanoSight Ltd. NanoSight Ltd., Minton Park, Amesbury, Wiltshire SP4 7RT, UK 1 Contents IMPORTANT NOTES... 3 USAGE... 3 SAFETY... 3 SERVICING... 4 MAINTENANCE...

More information

Navigator 600 Silica analyzers

Navigator 600 Silica analyzers ABB MEASUREMENT & ANALYTICS INFORMATION INF09/020 REV. B Navigator 600 Silica analyzers Troubleshooting Measurement made easy Navigator 600 silica analyzers 1 Introduction This publication details troubleshooting

More information

OC Panel High Limit Aquastat Kit, Manual Reset p/n

OC Panel High Limit Aquastat Kit, Manual Reset p/n OC Panel High Limit Aquastat Kit, Manual Reset p/n 233202 Instruction Sheet APPLICATION The OC (Option Control) Panel High Limit Aquastat Kit provides electronic temperature sensing in a UL limit-rated

More information

GV Standard X-Vent. Setup, Commissioning & Installation Guide

GV Standard X-Vent. Setup, Commissioning & Installation Guide GV Standard X-Vent Setup, Commissioning & Installation Guide Technical experts in the design, manufacture and supply of precision engineered, architectural rooflights for residential and commercial buildings.

More information

XpressFill XF4500 Counter Pressure Filler. Operating Instructions

XpressFill XF4500 Counter Pressure Filler. Operating Instructions XpressFill XF4500 Counter Pressure Filler Operating Instructions ii XPRESSFILL Systems LLC lling machine. ank you for choosing our handcrafted bottle filler as the technology to bottle your passion. We

More information

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel TANK MANAGER FOR TWO TANKS OPERATING MANUAL 10/31/11 C-More T6C L color touch panel 1 TABLE OF CONTENTS GENERAL...3 INSTALLATION...4 STONE TEST PROCEDURE...7 OPERATIONAL SUMMARY...7 AUTO CARBONATION...10

More information

product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels

product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels 12.09 product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels Introduction: This manual covers the installation and operation of Humboldt FlexPanels for Triaxial and Permeability

More information

PH01 Perfusion Cannula Manual

PH01 Perfusion Cannula Manual PH01 Perfusion Cannula Manual Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted without the express written permission of Multi

More information

Using the UltraRAE. Firmware 2.35

Using the UltraRAE. Firmware 2.35 Using the UltraRAE Firmware 2.35 Training Agenda UltraRAE features Setting up the UltraRAE Turning on the UltraRAE Idle Operation RAE-Sep Tubes Prepping for a measurement Taking a measurement Alarm modes

More information

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021-PR Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-PR For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4

More information

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany Manual Degasser DDG-75 April 2007 1 Manual Online Degasser DDG-75 DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE-68766 Hockenheim Germany Tel. +49-6205-9450-0 Fax. +49-6205-9450-33 email info@duratec.com

More information

Agilent 1220 Infinity II LC Mobile Upgrade Kit

Agilent 1220 Infinity II LC Mobile Upgrade Kit Agilent 1220 Infinity II LC Mobile Upgrade Kit Note 1220 Infinity II LC Mobile Upgrade Kit This note describes the procedures to install an Mobile Upgrade kit to an existing Agilent 1220 Infinity II LC.

More information

RH800 & 2000 BASIC OPERATIONS

RH800 & 2000 BASIC OPERATIONS RH800 & 2000 BASIC OPERATIONS 2 General information This manual contains technical information regarding Bayer SeedGrowth Equipment. Please read and understand these instructions completely before proceeding

More information

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by: 1 AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by: Materials/Reagents/Equipment Vendor 1. 0.2 um filtered Water + 0.05% sodium azide 2. 0.2 um filtered

More information

RAM 4021 Operation Manual

RAM 4021 Operation Manual RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For your safety...3 Description...3 Set-up mode...4 Annunciator lights/alarms...4 Operation...5 Calibration...6

More information

Installation, operating and maintenance Instructions for Seemag bypass level indicator

Installation, operating and maintenance Instructions for Seemag bypass level indicator Issue: S Date: 05-09-14 Type G35 General information The Seetru bypass magnetic level indicator, abbreviate SEEMAG, serves to show the filling level of fluids in tanks, basins, tubes etc. The Seemag operates

More information

MB-SP Series Solvent Purification System

MB-SP Series Solvent Purification System Operation Manual MB-SP Series Solvent Purification System Table of contents 1. System overview 2 2. Column activation.. 5 3. System purging.. 6 4. Glove box integration.... 7 5. Solvent reservoir installation....

More information

Laboratory Mortar Mixer (Testing)

Laboratory Mortar Mixer (Testing) TomTom-Tools GmbH Zelgli 20 8905 Arni info@tomtom-tools.com Switzerland www.tomtom-tools.com User Manual Version February 22, 2015 Laboratory Mortar Mixer (Testing) 1 Introduction The Laboratory Mortar

More information

XpressFill XF4500C Counter Pressure Can Filler. Operating Instructions

XpressFill XF4500C Counter Pressure Can Filler. Operating Instructions XpressFill XF4500C Counter Pressure Can Filler Operating Instructions ii XPRESSFILL Systems LLC lling machine. ank you for choosing our handcrafted bottle filler as the technology to bottle your passion.

More information

RB70 Automatic Diluent Valve Maintenance Manual. Version 1.1 November 2006 Written by Tino de Rijk. Page 1 of 23

RB70 Automatic Diluent Valve Maintenance Manual. Version 1.1 November 2006 Written by Tino de Rijk. Page 1 of 23 RB70 Automatic Diluent Valve Maintenance Manual Version 1.1 November 2006 Written by Tino de Rijk Page 1 of 23 Table of Contents 1. Introduction... 3 2. ADV diagram and parts list (Pre June 2006)... 4

More information

200 PSI FAST-FILL AIR SOURCE KIT

200 PSI FAST-FILL AIR SOURCE KIT 200 PSI FAST-FILL AIR SOURCE KIT 55% Duty Compressor on 2.0 Gallon Air Tank PART NO. 20007 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

12V MINI AIR COMPRESSOR

12V MINI AIR COMPRESSOR 12V MINI AIR COMPRESSOR STOCK No.65958 PART No.DA12/250A INSTRUCTIONS IMPORTANT: PLEASE READ THESE INSTRUCTIONS CAREFULLY TO ENSURE THE SAFE AND EFFECTIVE USE OF THIS TOOL. 08/2001 GENERAL INFORMATION

More information

Installation of Your SprayMaster System

Installation of Your SprayMaster System Installation of Your SprayMaster System 1. At the installation site, remove all equipment from the corrugated box and the polyethylene drum and replace the drum lid. Check the picture to identify each

More information

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL -

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL - Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL - Table of Contents 1. Introduction 3 2. Instrument Setup 4 2.1 Instrument Checklist 4 2.2 Optional Accessories 4 2.3 Gas Concentration Ranges Available

More information

BS ml Bottle Sampling Unit

BS ml Bottle Sampling Unit BS500 500ml Bottle Sampling Unit User Guide www.mpfiltri.co.uk 200.307-EN Covers Model Numbers BS500 SAFETY WARNING Hydraulic systems contain dangerous fluids at high pressures and temperatures. Installation,

More information

THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY

THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY Page: 1 of 9 1. Risk Assessment: This risk assessment has been prepared using the National Safety Council of Australia Risk Score Calculator

More information

TQC Bresle Chloride test 2008

TQC Bresle Chloride test 2008 TQC Bresle Chloride test 2008 SP7310 MANUAL See also : ISO8502-6 ISO8502-9 1 DETERMINATION OF SOLUBLE CONTAMINANTS ON SURFACES IN CONFORMITY WITH ISO 1.1 PREPARATION With this system very low concentrations

More information

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone: MJB4 Mask Aligner Operating Procedure Effective Date: 07/12/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Karl Suss MJB4 Mask Aligner, Optical Lithography,

More information

- Repeat calibration until desired results are reached otherwise set SW2 and SW1-8 open.

- Repeat calibration until desired results are reached otherwise set SW2 and SW1-8 open. Calibration Procedures on Cobe Centrysystem 3: I) Balance Chamber Calibration: - Verify that pressure regulators PRV and PRV2 are calibrated as well as pressure transducers PDI and PDO with switches SW2

More information

Assembly drawing and the names of features

Assembly drawing and the names of features Assembly drawing and the names of features M-2T model main body Power switch Timer Amplitude control switch STOP key START key TIMER Timer operation indicator Regulator output indicator Reset indicator

More information

Operating instructions Capacitive sensors KG AC (M18) / / 2010

Operating instructions Capacitive sensors KG AC (M18) / / 2010 Operating instructions Capacitive sensors KG AC (M18) UK 704923 / 00 09 / 2010 Contents 1 Preliminary note 3 2 Safety instructions 3 3 Functions and features 3 3.1 Application examples 4 4 Installation

More information

User Manual. Heads-Up Display (HUD) DiveCAN. Mechanical Button Version

User Manual. Heads-Up Display (HUD) DiveCAN. Mechanical Button Version User Manual Heads-Up Display (HUD) Mechanical Button Version DiveCAN Table of Contents 1. Introduction...4 1.1 Features...4 2. Physical Description...5 3. Reading the PPO2...6 3.1 Modified Smither s Code...7

More information

444C DUAL PERFORMANCE VALUE PACK

444C DUAL PERFORMANCE VALUE PACK (Chrome) PART NO. 44432 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of this manual before installing and using this product. SAVE THIS MANUAL

More information

Reflow Oven HHL3000 INSTRUCTION MANUAL POHUA - jedyny autoryzowany przedstawiciel w Polsce

Reflow Oven HHL3000 INSTRUCTION MANUAL POHUA - jedyny autoryzowany przedstawiciel w Polsce POHUA - jedyny autoryzowany przedstawiciel w Polsce www.pohua.pl AOYUE TONGYI ELECTRONIC EQUIPMENT FACTORY Jishui Industrial Zone, Nantou, Zhongshan City, Guangdong Province, P. R. China www.aoyue.com

More information

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved

More information

air/hydraulic nut riveter heavy-duty vacuum system

air/hydraulic nut riveter heavy-duty vacuum system instructions for air/hydraulic nut riveter heavy-duty vacuum system model no: SA317 Thank you for purchasing a Sealey product. Manufactured to a high standard, this product will, if used according to these

More information

MODEL NUMBER: PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank

MODEL NUMBER: PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank IMPORTANT SAFETY INSTRUCTIONS CAUTION - To reduce risk of electrical shock or Electrocution: MODEL NUMBER: 20008 200 PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank IMPORTANT: It is

More information