SENSOR MANUAL Document No M (Revision L)

Size: px
Start display at page:

Download "SENSOR MANUAL Document No M (Revision L)"

Transcription

1 SENSOR MANUAL Document No M (Revision L) Sensidyne, LP th Circle N, Suite 0 St. Petersburg, Florida USA (800) (727) Fax: (727) web: info@sensidyne.com

2

3 PACKING LIST & NOTICES The items listed below are shipped with each SensAlert Sensor Sensor Assembly Packing Foam and Sensor Container ALWAYS check to make certain you have received all of the items listed above. If you have any questions or need assistance, contact your Sensidyne Representative, or call (800) or (727) PROPRIETARY NOTICE This manual was prepared by Sensidyne, LP exclusively for the owner of the SensAlert Sensor. The material within this manual is the proprietary information of Sensidyne, LP and is to be used only to understand, operate, and service the instrument. By receiving this document, the recipient agrees that neither this document nor the information disclosed within nor any part shall be reproduced or transferred, physically, electronically or in any form or used or disclosed to others for manufacturing or for any other purpose except as specifically authorized in writing by Sensidyne, LP. COPYRIGHT NOTICE , Sensidyne, LP. All rights reserved. Information contained in this document is protected by copyright. No part of this document may be photocopied, reproduced, or translated to another program or system without prior written authorization from Sensidyne, LP. TRADEMARK NOTICE Sensidyne, the Sensidyne logo, SensAlert, and the SensAlert logo are registered trademarks of Sensidyne, LP. The trademarks and service marks of Sensidyne, LP are protected through use and registration in the United States. DISCLAIMER SENSIDYNE, LP ASSUMES NO RESPONSIBILITY WHATSOEVER, TO ANY PARTY WHOSOEVER, FOR ANY PROPERTY DAMAGE, PERSONAL INJURY, OR DEATH RECEIVED BY OR RESULTING FROM, IN WHOLE, OR IN PART, THE IMPROPER USE, INSTALLATION, OR STORAGE OF THIS PRODUCT BY THE USER, PERSON, FIRM, ENTITY, CORPORATION OR PARTY NOT ADHERING TO THE INSTRUCTIONS AND WARNINGS OR NOT ADHERING TO ALL FEDERAL, STATE, AND LOCAL ENVIRONMENTAL AND OCCUPATIONAL HEALTH AND SAFETY LAWS AND REGULATIONS. THE SELLER SHALL NOT BE LIABLE FOR DIRECT, INDIRECT, CONSEQUENTIAL, INCIDENTAL OR OTHER DAMAGES RESULTING FROM THE SALE AND USE OF ANY GOODS AND SELLER S LIABILITY HEREUNDER SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF ANY GOODS FOUND DEFECTIVE. THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR USE OR FOR A PARTICULAR PURPOSE WHICH ARE EXPRESSLY DISCLAIMED.

4 TABLE OF CONTENTS PREFACE Packing List & Notices...3 WARNINGS... 7 SECTION ONE: INTRODUCTION 1.1 Overview SensAlert Intelligent Sensors SensAlert Disposable Sensors Sensor Storage SensAlert Refillable Sensors... 9 SECTION TWO: CALIBRATION EQUIPMENT & SERVICES 2.1 Introduction Accessories & Calibration Gases...12 Calibration Accessories...13 Toxic, Oxygen, & Gases...13 Combustible Gases...14 Zero Gases Calibration Services...15 SensAlert Sensor Calibration-Exchange Program...15 Field Service On-Site Maintenance & Calibration Program...15 SECTION THREE: SENSALERT SENSOR LIST SensAlert Sensors SECTION FOUR: REFILLABLE SENSOR MAINTENANCE 4.1 Overview General Inspection Electrolyte & Membrane Cap Replacement

5 TABLE OF CONTENTS APPENDICES Appendix A: Sensor Specifications Using The Sensor Tables...23 Sensor (Maximum Range) Ammonia (50 ) [Low ]...24 Ammonia (0 ) [Low ]... Ammonia (300 ) [Low ]...26 Ammonia (500 ) [Low ]...27 Arsine (1.00 )...28 (2.00 %vol) [Infrared]...29 (0 )...30 (500 )...31 (00 )...32 Low Interference (0 )...33 Low Interference (500 )...34 (5.00 )...35 (.0 )...36 (.0 ) [Refillable]...37 (0 )...38 (1.00 )...39 (5.00 )...40 Combustible Gas (0 %) [Infrared for Acetylene]...41 Combustible Gas (0 %) [Catalytic Bead]...42 Combustible Gas (0 %) [Catalytic Bead for ]...43 Combustible Gas (0 %) [Infrared for Methane Gas]...44 Combustible Gas (0 %) [Infrared for non-methane Hydrocarbons] Diborane (1.00 )...46 Ethylene Oxide (.0 ) [shipped with PNº Battery Board] Fluorine (.0 )...48 Fluorine (.0 )...49 Germane (2.00 )...50 Hydrazine (1.00 )...51 (00 )...52 (0 %)...53 Chloride (.0 )...54 Chloride (.0 ) [Refillable]...55 Chloride (20.0 )...56 Chloride (0 ) [shipped with PNº Battery Board] Chloride (0 ) [Refillable]...58 (sensor list continues on following page) 5

6 TABLE OF CONTENTS APPENDICES Appendix A: Sensor Specifications [continued] Sensor (Maximum Range) Cyanide (20.0 )...59 Cyanide (0 )...60 Fluoride (.0 )...61 Fluoride (.0 ) [Refillable]...62 Fluoride (20.0 ) [Refillable]...63 Sulfide (50 )...64 Sulfide (0 )...65 Nitric Oxide (0 ) [shipped with PNº Battery Board]...66 Nitrogen (.0 )...67 Oxygen (.0 %vol) [Process]...68 Oxygen (.0 %vol) [Ambient Air]...69 Oxygen (.0 %vol) [Process]...70 Ozone (2.00 )...71 Phosgene (1.00 )...72 Phosphine (1.00 )...73 Silane (.0 )...74 Styrene (200 )...75 Sulfur (.0 )...76 Sulfur (.0 ) [Refillable]...77 Sulfur (20.0 )...78 Sulfur (20.0 ) [Filtered]...79 Sulfur (0 ) [Filtered]...80 Sulfur (0 ) [Refillable]...81 Appendix B: Methane & Propane K-factors for Catalytic Sensors Appendix C: Returned Material Authorization...86 FIGURES 1.1 SensAlert Sensor Styles Biased Sensor With Battery Board Refillable Sensor Maintenance...20 TABLES 1.1 Replacement/Service Parts: SensAlert Refillable Sensors

7 WARNINGS READ AND UNDERSTAND ALL WARNINGS BEFORE USE Read and understand ALL warnings before using this product. Failure to read, understand, and comply with ALL warnings could result in property damage, severe personal injury, or death. Read and understand ALL applicable federal, state, and local environmental health and safety laws and regulations, including OSHA. Ensure complete compliance with ALL applicable laws and regulations before and during use of this product. UNDER NO CIRCUMSTANCES should this product be used except by qualified, trained, technically competent personnel and not until the warnings, Operation & Service Manual, labels, and other literature accompanying this product have been read and understood. Each user MUST READ AND UNDERSTAND the pertinent Material Safety Data Sheets (MSDS) and the Operation & Service Manual before operating this product or using its accessories, in order to ensure proper and safe use and installation of this product, and to ensure familiarity with the proper treatment and safety procedures in the event of an accident. Failure to read and understand the MSDS s and the Operation & Service Manual may result in preventable severe personal injury or death. Additional MSDS s and manuals are available from Sensidyne. Call (800) , or (727) ALWAYS wash your hands thoroughly after handling, calibrating, or servicing this product, and before eating, drinking, smoking or applying cosmetics. If chemicals are ingested accidently, consult Material Safety Data Sheets and a physician IMMEDIATELY. Call Poison Control Center IMMEDIATELY. Consult the emergency section of your telephone directory for the telephone number of the Poison Control Center in your area. ALWAYS wear eye protection (such as safety goggles), face shield, chemical resistant gloves and chemical resistant clothing when handling chemicals, or calibration sources. DO NOT get chemicals, gases, fumes, or vapors in your eyes or on your skin, as they may cause severe burns to skin and eyes. If chemicals, gases, fumes, or vapors get in your eyes or on your skin, wash the affected area with copious amounts of water, consult MSDS s and call a physician immediately. ALWAYS avoid any contact of acids with your skin or eyes. Seek immediate medical attention for any contact with acids and consult ALL pertinent MSDS s. ALWAYS calibrate in a well ventilated area by opening ALL doors and windows. Prevent the buildup of ANY calibration sources or vapors. Avoid breathing ANY calibration fumes or vapors as they are hazardous to your health. ALWAYS dispose of chemicals and calibration sources in compliance with ALL applicable safety laws, regulations, and guidelines for proper disposal. Failure to do so may result in environmental and property damage, personal injury or death. ALWAYS close ALL containers of chemicals used with this product after use. 7

8 WARNINGS ALWAYS ensure that any compressed calibration substance sources are empty prior to disposal. ALWAYS use clean, dry, inert materials to contain and transfer substances used for calibration. DO NOT remove, cover, or alter any label or tag on this product, its accessories, or related products. DO NOT operate this product should it malfunction or require repair. Operation of a malfunctioning product, or a product requiring repair may result in serious injury or death. DO NOT attempt to repair or modify instrument, except as specified in the Operation & Service Manual. If repair is needed, contact the Sensidyne Service Dept. to arrange for a Returned Material Authorization (RMA). Use ONLY genuine SENSIDYNE replacement parts when performing any maintenance procedures described in this manual. Failure to do so may seriously impair instrument performance. Repair or alteration of the product beyond the scope of these maintenance instructions, or by anyone other than an authorized SENSIDYNE service person, could cause the product to fail to perform as designed and persons who rely on this product for their safety could sustain severe personal injury or death. DO NOT restrict the access of ambient air to the sensor. Note: Prolonged exposure to excessively high concentrations of toxic gas may cause the sensor to produce erroneous readings. All Sensidyne gas detection equipment is functionally tested and calibrated with known zero and span gases before shipment to the customer. To ensure proper operation calibrate at regular intervals. Failure to calibrate in accordance with the instructions in this manual and at regular intervals may result in the product malfunctioning. Verify concentration of calibration span gas before making calibration adjustments. Concentration can be altered by: Deterioration of the concentration of compressed calibration substance sources during storage. Interaction of the calibration substance with materials used to contain and transfer the substance, as for example, absorption onto and permeation through certain plastics. Interaction of the calibration substance with materials with ambient contaminants, as for example, absorption into water. Sudden changes in pressure may cause temporary fluctuations in the sensor reading. 8

9 SECTION ONE INTRODUCTION 1.1 OVERVIEW This manual provides specific information concerning calibration, specifications, and interferents of the Sensidyne SensAlert Sensor. Each sensor is capable of detecting the presence of potentially hazardous concentrations of toxic, combustible gas, or Oxygen. The sensor assembly is mounted to the transmitter by inserting the three guide pins on the sensor assembly into the 3 matching holes on the sensor interface assembly. The sensor assembly is then pushed upward to engage the sensor interface cable connector. SensAlert Sensors may be interchanged without any special tools or adjustments. When a new sensor is installed in the unit, the transmitter or monitor automatically recognizes the new gas type, range, and measuring units and adjusts the electronics accordingly. The sensor information is also sent to a SensAlert controller (if one is connected), where the controller makes the necessary electronic adjustments for the new sensor. NOTE Catalytic Combustible and Infrared sensors require a 3-wire transmitter for operation and will not work with 2-wire transmitters. Sensor specifications can be found in Appendix A: Specifications. Product specifications for transmitters, monitors, and controllers can be found in the respective operation and services manuals. 1.2 SENSALERT INTELLIGENT SENSORS SensAlert Disposable Sensors SensAlert Universal Transmitters and the SensAlarm Universal Gas Monitor can use either toxic, combustible gas, or Oxygen sensors (see Figure 1.1). Toxic and Oxygen sensors are electrochemical sensors designed to respond to the presence of a specific gas. Gas molecules permeate the sensor membrane and initiate an electrochemical reaction in the sensor. Combustible sensors are catalytic oxidation type or IR used to detect most combustible gases in the 0-0 % range. The sensor assembly consists of either a sealed or refillable sensor attached to a circuit board. Non-volatile memory on the circuit board contains essential sensor identification information that is communicated to the monitor during sensor installation. A female connector mounted on top of the circuit board mates with a male connector mounted inside the sensor interface assembly. Note Some biased sensors are shipped with attached battery boards to maintain biasing for up to 1 year (see Figure 1.2). Battery boards should remain attached to the assembly until just prior to sensor installation (maximum 15 seconds) Sensor Storage Sensors should be stored in their original containers at 3 20 C (37 68 F) Cooler temperatures are preferred. If not stored in their original sealed containers, sensor should be stored at %RH (sensors in sealed containers will reach their own equilibrium moisture level). Sensors should be allowed to re-equilibrate at their use temperature prior to testing or calibration (30 minutes to 1 hour is usually sufficient for cold stored sensors). Biased sensors must be stored on their battery bias boards or they could need hours to re-equilibrate once they are plugged in for use. Long term storage of oxygen sensors is not recommended since these sensors can be consumed during storage SensAlert Refillable Sensors In addition to the above items, SensAlert refillable sensors are mounted on their sensor board via a coaxial (BNC) connector. The sensors may be removed from their PCBs for inspection and servicing. A table of refillable replacement/service parts is shown below. Note Sensor membrane caps are custom fit to each refillable sensor to maximize sensor performance. Always insure that the number on the membrane cap matches the cap code number on the side of the sensor, otherwise performance may be degraded, or cap may be damaged. Number (*) Membrane Cap Assembly (* specify Cap No. when ordering) Silicone Grease Torque Tool Electrolyte, 30 ml bottle Fluoride Electrolyte, 30 ml bottle Chloride Electrolyte, 30 ml bottle Sulfur Electrolyte, 30 ml bottle Table 1.1 Replacement/Service Parts: SensAlert Refillable Sensors 9 Part Description

10 TYPICAL SENSOR STYLES Sensor Interface Assembly Sensor Interface O-Ring Sensor Assembly Sensor Retainer Rainshield O-Ring Rainshield (other types available) SEN HYDROGEN CH Part No S/N Figure 1.1 SensAlert Sensor Styles

11 BATTERY BOARD (used on Ethylene Oxide, Chloride [0-0], and Nitric Oxide sensors only) GUIDE PINS SENSOR ASSEMBLY Figure 1.2 Biased Sensor With Battery Board 11

12 SECTION TWO CALIBRATION EQUIPMENT & SERVICES 2.1 INTRODUCTION All Sensidyne gas detection equipment is functionally tested and calibrated with known zero and span gases before shipment to the customer. However, it is recommended that qualified personnel verify operation after initial installation using known zero and span gases. Repeat verification at 30 days and 60 days after initial installation, with deviations in zero and span recorded. The calibration or functional check interval can then be adjusted to suit that application s conditions. For further information, please consult the ISA Recommended Practices for gas detectors. 2.2 ACCESSORIES & CALIBRATION GASES Standard calibration accessories for SensAlert Transmitters and SensAlarm Monitors include gas bottles, regulators, carrying cases, and calibration cups. All calibration gases are shipped in either disposable aluminum or steel cylinders. Aluminum cylinders contain 58 liters of the calibration gas, while steel cylinders contain 3 liters. Each calibration gas cylinder is shipped with an MSDS sheet (a NIST traceable calibration certificate also is available upon request.) Contact Sensidyne for information on ordering specialty accessories or gases. 12

13 CALIBRATION EQUIPMENT Calibration Accessories Description Part Number Regulator (1 LPM), for use with all gases PVC Carrying Case (holds two gas cylinders, plus regulator, tubing & fitting) Humidification Kit (Recommended for setting acid gas zero) Calibration Cup (for disposable sensors) for all SensAlert Transmitters & SensAlarm Gas Monitor Calibration Cup (for refillable sensors) for all SensAlert Transmitters & SensAlarm Gas Monitor Teflon Tubing (3 feet) Toxic, Oxygen, & Gases Description & Calibration Method [if Span Gas not available] (size, shelf life) Part Number Ammonia [NH 3 ], in Nitrogen (58L, 8 mos.) Ammonia [NH 3 ], 50 in Nitrogen (58L, 8 mos.) Arsine [AsH 3 ], (Zero & Span gases) See Sensor Exchange Programs Some customers use Phosphine as a functional check gas. Bromine, [Br 2 ], (Diffusion Tube) See Sensor Exchange Programs [CO 2 ], 5000 in Air (3L, 12 mos.) [CO 2 ], 1.75 %vol in Nitrogen (3L, 12 mos.) [CO], 50 in Nitrogen (3L, 12 mos.) [CO], 0 in Nitrogen (3L, 12 mos.) [CO], 0 in Nitrogen (3L, 12 mos.) [Cl 2 ], 2 in Nitrogen (58L, 8 mos.) [Cl 2 ], 5 in Nitrogen (58L, 8 mos.) [Cl2], in Nitrogen (58L, 8 mos.) [Cl 2 ], 50 in Nitrogen (58L, 8 mos.) [ClO 2 ], (Zero gas & calibrated gas generator) See Sensor Exchange Programs Some customers use as a functional check gas. Diborane [B 2 H 6 ] (Zero & Span gases) See Sensor Exchange Programs Ethylene Oxide [C 2 H 4 O], 5 in Nitrogen (58L, 12 mos.) Fluorine, F 2 (Permeation Tube) See Sensor Exchange Programs Germane, GeH 4, (Zero & Span gases, Permeation Tube) See Sensor Exchange Programs [H 2 ] () Electrochemical, 500 in Air (3L,12 mos.) Chloride 3 [HCl], 5 in Nitrogen (58L, 6 mos.) Chloride 3 [HCl], HCl in Nitrogen (58L, 6 mos.) Chloride 3 [HCl], 50 HCl in Nitrogen (58L, 6 mos.) Cyanide 3 [HCN], in Nitrogen (58L, 12 mos.) Fluoride, HF (Permeation Tube) See Sensor Exchange Programs 1 Recommended practice for determining calibration frequency. Sensidyne equipment is tested and calibrated prior to shipment. After installation and stabilization of the gas monitor, qualified personnel should verify calibration by applying zero and span gases. This procedure should be performed at commissioning, then repeated 30 and 60 days afterwards, with deviations in zero and span recorded. The calibration or functional check interval can then be adjusted to suit that application s conditions. For further information, please consult the ISA Recommended Practices for gas detectors. 2 The use of functional check or surrogate gases do not represent actual calibration of the sensor with the target gas, and should not be employed as such, unless the user has established a known, repeatable correlation with the target gas. 3 The humidification kit is recommended for setting the sensor zero, ONLY for acid gases. 13

14 CALIBRATION EQUIPMENT Toxic, Oxygen, & Gases (continued) Description & Calibration Method [if Span Gas not available] (size, shelf life) Part Number Sulfide 2 [H 2 S], 5 in Nitrogen (58L, 12 mos.) Sulfide 2 [H 2 S], in Nitrogen (58L, 12 mos.) Sulfide 2 [H 2 S], in Nitrogen (58L, 12 mos.) Sulfide 2 [H 2 S], 50 in Nitrogen (58L, 12 mos.) Nitric Oxide [NO], 30 in Nitrogen (58L, 6 mos.) Nitrogen 3 [NO 2 ], 5 in Nitrogen (3L, 6 mos.) Nitrogen 3 [NO 2 ], in Nitrogen (3L, 6 mos.) Oxygen [O 2 ], %vol in Nitrogen (3L, 12 mos.) Oxygen [O 2 ], 18.5 %vol in Nitrogen (3L, 12 mos.) Oxygen [O 2 ], 20.9 %vol in Nitrogen (3L, 12 mos.) Ozone [O 3 ], (Zero gas & calibrated gas generator) See Sensor Exchange Programs Phosgene [COCl 2 ] (Permeation Tube) See Sensor Exchange Programs Phosphine [PH 3 ], 0.5 in Nitrogen (58L, 6 mos.) Silane [SiH 4 ], 5 in Nitrogen (58L, 6 mos.) Sulfur [SO 2 ], 5 in Nitrogen (58L, 12 mos.) Sulfur [SO 2 ], in Nitrogen (58L, 12 mos.) Sulfur [SO 2 ], 50 in Nitrogen (58L, 12 mos.) Combustible Gases (Select appropriate span gas using K-factor, see Appendix B) Description Part Number [H 2 ], 2 %vol / 50 % in Air (3L, 12 mos.) Methane [CH 4 ], 1.5 %vol / 30 % in Air (3L, 12mos.) Methane [CH 4 ], 2.5 %vol / 50 % in Air (3L, 12 mos.) Propane [C 3 H 8 ], 0.66 %vol / 30% in Air (3L, 12 mos.) Propane [C 3 H 8 ], 0.95 %vol / 50 % in Air (3L, 12 mos.) Butane [C 4 H ], 0.56%vol [30 % in Air (3L, 12 mos.) Zero Gases Description Part Number Zero Gas, 20.9% Oxygen in Nitrogen (3L, 12 mos.) Note: The Oxygen zero is inherent and a zero gas is not required. 3 The humidification kit is recommended for setting the sensor zero, ONLY for acid gases. 1 Recommended practice for determining calibration frequency. Sensidyne equipment is tested and calibrated prior to shipment. After installation and stabilization of the gas monitor, qualified personnel should verify calibration by applying zero and span gases. This procedure should be performed at commissioning, then repeated 30 and 60 days afterwards, with deviations in zero and span recorded. The calibration or functional check interval can then be adjusted to suit that application s conditions. For further information, please consult the ISA Recommended Practices for gas detectors. 2 The use of functional check or surrogate gases do not represent actual calibration of the sensor with the target gas, and should not be employed as such, unless the user has established a known, repeatable correlation with the target gas. 3 The humidification kit is recommended for setting the sensor zero, ONLY for acid gases. 14

15 2.3 CALIBRATION SERVICE SensAlert Sensor Calibration-Exchange Program Standard Sensor Calibration and Exchange Customer must purchase required number of extra sensors, which are warranted for one year. Sensidyne stores and ships calibrated sensors at user selected intervals. This service includes sensor refrigeration, biased sensor handling, and other sensor specific storage measures. Freight both ways is prepaid and UPS call tag is provided. Sensors are replaced as required and billed to the customer. Customer s sensors are labeled with service and calibration dates plus user name. Calibrated sensors are shipped with a calibration certificate. The standard warranty does not cover abuse, immersion, excessive exposure to gas, or other abnormal application conditions requiring premature replacement. Description Part Number Monthly Calibration and Exchange for Toxic Sensors Bimonthly Calibration and Exchange for Toxic Sensors Quarterly Calibration and Exchange for Toxic Sensors Monthly Calibration and Exchange for Combustible Sensors Bimonthly Calibration and Exchange for Combustible Sensors Quarterly Calibration and Exchange for Combustible Sensors Field Service On-Site Maintenance & Calibration Program Start-up and Initial Calibration With Informal Training of Owner s Personnel Consult factory for daily rates. Travel costs and living expenses will be charged above the daily rate. There is NO charge for domestic travel time. Purchase order required in advance. Note: There is a 4 hour minimum charge for added time after first day. Waiting time is charged at the same rate. Parts (except warrantee items) and calibration gases are billed at prevailing prices. Routine Calibration Service Rates Consult factory for daily rates. Travel costs and living expenses will be charged above the daily rate. There is NO charge for domestic travel time. Purchase order required in advance. Note: There is a 4 hour minimum charge for added time after first day. Waiting time is charged at the same rate. Parts (except warranted items) and calibration gases are billed at prevailing prices. Maintenance Contract at User-Selected Intervals Maintenance Contract: To be negotiated as part of agreement. There is NO charge for domestic travel time. Purchase order required in advance. Note: If time exceeds that agreed to in contract, at no fault of Sensidyne, daily rates will then apply. Parts (except warranty items) and calibration gases are billed at prevailing prices. 15

16 SECTION THREE SENSALERT SENSOR LIST Description (Maximum Range) Part Number Ammonia (50 ) [Low ] D-3X Ammonia (0 ) [Low ] D-1 Ammonia (300 ) [Low ] D-2X Ammonia (500 ) [Low ] D-1 Arsine (1.00 ) D-1 (2.00 %vol) [Infrared] * D-1 (0 ) D-1 (500 ) D-2 (00 ) D-3X [Low Interference] (0 ) D-1 Low Interference (500 ) D-2 (5.00 ) D-2 (.0 ) D-1 (.0 ) [Refillable] D-1 (0 ) D-3 (1.00 ) D-1 (5.00 ) D-2 Combustible Gas (0 %) [Infrared for Acetylene] * D-1 Combustible Gas (0 %) [Catalytic Bead] D-1 Combustible Gas (0 %) [Catalytic Bead for ] D-1 Combustible Gas (0 %) [Infrared for Methane Gas] * D-1 Combustible Gas (0 %) [Infrared for non-methane Hydrocarbons] * D-2 Diborane (1.00 ) D-1 Ethylene Oxide (.0 ) [shipped with PNº Battery Board] D-1 Fluorine (.0 ) D-1 Fluorine (.0 ) D-2 Germane (2.00 ) D-1 Hydrazine (1.00 ) D-1 (00 ) D-1 (0 %) D-1 (sensor list continues on following page) 16 Cannot be used with the XP Gas Transmitter * Can only be used with Infrared Transmitters (PNº x or PNº x)

17 SECTION THREE SENSALERT SENSOR LIST Description (Maximum Range) Part Number Chloride (.0 ) D-3X Chloride (.0 ) [Refillable] D-1X Chloride (20.0 ) D-1 Chloride (0 ) [shipped with PNº Battery Board] D-2 Chloride (0 ) [Refillable] D-2 Cyanide (20.0 ) D-1 Cyanide (0 ) D-2X Fluoride (.0 ) D-1 Fluoride (.0 ) [Refillable] D-1 Fluoride (20.0 ) [Refillable] D-2X Sulfide (50 ) D-1 Sulfide (0 ) D-2 Nitric Oxide (0 ) [shipped with PNº Battery Board] D-1 Nitrogen (.0 ) D-1 Oxygen (.0 %vol) [Process] D-2X Oxygen (.0 %vol) [Ambient Air] D-1 Oxygen (.0 %vol) [Process] D-3X Ozone (2.00 ) D-1 Phosgene (1.00 ) D-1 Phosphine (1.00 ) D-1 Silane (.0 ) D-1 Styrene (200 ) D-1X Sulfur (.0 ) D-3X Sulfur (.0 ) [Refillable] D-1X Sulfur (20.0 ) D-1 Sulfur (20.0 ) [Filtered] D-1 Sulfur (0 ) [Filtered] D-2X Sulfur (0 ) [Refillable] D-2X Cannot be used with the XP Gas Transmitter * Can only be used with Infrared Transmitters (PNº x or PNº x) 17

18 THIS PAGE INTENTIONALLY LEFT BLANK 18

19 SECTION FOUR REFILLABLE SENSOR MAINTENANCE 4.1 OVERVIEW Periodic preventive maintenance of SensAlert refillable sensors is required to ensure proper operation of the system. The following sections describe certain simple maintenance procedures that may be performed in the field. Field maintenance should be limited to : General inspection Electrolyte replacement Membrane cap replacement General inspection, electrolyte replacement, and membrane cap replacement should be performed at regular three (3) month intervals. These maintenance procedures should also be performed following prolonged exposure to high concentrations of the target gas. The sensor assembly should be replaced if it is damaged when it no longer performs properly, or as scheduled. NOTE If the sensor assembly is damaged or defective, DO NOT attempt to repair it. Return the entire sensor to Sensidyne for service. NOTE The three (3) month maintenance schedule is based on normal operating conditions (see Appendix A: Specifications). Excessively severe or dirty environments may require these maintenance procedures to be performed more frequently. 4.2 GENERAL INSPECTION This inspection procedure should be performed every three (3) months (normal operating conditions). Refer to Figure ) Perform a visual inspection of the area surrounding the transmitter to ensure air flow to the sensor is unrestricted. 2) Remove the sensor retainer from the transmitter 3) Remove the sensor assembly from the sensor interface assembly by gently pulling down. 4) Remove the sensor electrode from the PCB by holding the edges of the PCB and pulling the sensor off the coaxial (BNC) connector. 5) Position the sensor vertically with the membrane cap facing down. Hold the sensor up to a bright light at eye level. The electrolyte should be visible, with its level just below the vent hole. If the electrolyte is not visible, or the level is well below the vent hole, the electrolyte should be replaced. (Go to Section 4.3.) 6) Refer to Figure 4.1. With the membrane cap still facing down, inspect the cap for leakage. NOTE Leaks - regardless of their source - tend to migrate toward the face of the membrane cap. If a leak is visible, determine its source by first blotting off excess electrolyte, then placing the sensor (cap-side down) on a piece of paper. 19

20 TOP VIEW (B) X CAP CODE NUMBER RESERVOIR MEMBRANE CAP CENTER POST SIDE VIEW (A) MEMBRANE CAP SEN HYDROGEN CH Part No S/N VENT HOLE CAP CODE NUMBER (X) TORQUE TOOL BNC CONNECTOR Figure 4.1 Refillable Sensor Maintenance 20

21 7) Remove the sensor and examine the surface of the paper: If the wet area forms a ring, electrolyte is leaking from the edges of the cap, indicating that it may not be securely sealed. a) Remove the cap and apply silicone grease to the threads. (Do not get grease on the inside surface of the cap, on the face of the cap, or on the membrane itself. If you do, discard the cap.) b) Replace the cap and tighten it using the Sensidyne Torque Tool. Re-inspect for leaks. If leaks are still visible, replace the membrane. If the wet area forms a dot, the leak is coming from the membrane itself. a) Remove the membrane cap and closely inspect the membrane for tears or punctures. If tears or punctures are visible, replace the membrane cap. If the membrane appears undamaged, apply silicone grease to the threads of the cap. (Do not get grease on the inside surface of the cap, on the face of the cap, or on the membrane itself. If you do, discard the cap.) 8) Position the sensor with the membrane cap facing up. 9) Inspect the membrane cap to ensure that the holes exposing the membrane are not clogged. Gently blow away any dirt that may be clogging the holes. Do not touch the membrane. ) Inspect the holes again. If the dirt is still clogging the holes, the membrane cap must be replaced. (Proceed to Section 4.4.) 11) Once all applicable maintenance procedures have been performed, mount the sensor by presssing it into the coaxial (BNC) connector. Mount the sensor in the transmitter by inserting the three (3) guide pins into the sensor interface assembly. Rotate the sensor assembly until the guide pins are properly seated in the counter bore holes. Push the sensor assembly upward to engage the electrical connection. 12) Make certain the sensor retainer is properly installed. 13) Allow the sensor to stabilize for one (1) hour before attempting calibration. WARNING Vapors from silicone grease will quickly poison a catalytic bead combustible sensor. Do not use silicone grease where its vapors can migrate to a catalytic bead combustible sensor. b) Replace the cap and tighten it using the Sensidyne Torque Tool. Re-inspect for leaks. If leaks are still visible, replace the membrane. (Proceed to Section 4.4.) 21

22 4.3 ELECTROLYTE & MEMBRANE CAP REPLACEMENT Sensidyne recommends replacing the membrane cap whenever the electrolyte is replaced. Electrolyte replacement should be done every three months. The electrolyte replacement schedule is based on normal conditions (see Appendix A: Specifications). Temperatures and/or air speeds higher than specified, or humidity lower than specified will increase the frequency of electrolyte replacement. 8) Remove the new membrane cap from its packaging. Ensure that the cap code number matches the code number in parenthesis on the outer wall of the sensor (shown as X in Figure 4.1). CAUTION The sensor operation may be degraded by using a membrane cap bearing the wrong code number. CAUTION The code number on the replacement membrane cap must match the code number in parenthesis on the outer wall of the sensor. A mismatched membrane cap can seriously impair the performance of the sensor. Electrolyte replacement should be done before replacing the membrane cap to ensure the membrane is not contaminated with old electrolyte. Replace the membrane cap and electrolyte as follows (refer to Figure 4.1): 1) Remove the sensor retainer. 2) Remove the sensor assembly from the sensor interface assembly by gently pulling down. Remove the sensor from its PCB by pulling it off the coaxial (BNC) connector. 3) Position the sensor vertically with the membrane cap facing up. 4) Carefully unscrew the membrane cap and lift it off the sensor. The membrane will be retained within the cap. Do not touch the surface of the membrane or any internal part of the sensor. 5) Safely dispose of the spent electrolyte. Shake the sensor to remove all of the electrolyte. 6) Re-orient the sensor vertically with the sensor reservoir surface facing upward. Fill the reservoir with distilled or de-ionized water. Discard the water and shake out any excess droplets. 7) Slowly pour 4-5 ml of fresh electrolyte into the sensor. This amount is sufficient to fill the sensor to approximately cm ( in) below the threads of the reservoir. A low profile drop of electrolyte should adhere to the surface of the center post. 9) Ensure that the threads of the membrane cap are freshly prepared with silicone grease. Applying a sufficient amount of silicone grease to fill the threads of the cap. (Do not get grease on the inside surface of the cap, on the face of the cap, or on the membrane itself. If you do, discard the cap.) ) Replace and tighten the cap using the Sensidyne Torque Tool until it is secure. Wipe away any excess silicone grease that escapes from around the edges of the membrane cap. NOTE Do not use any tools other than the Sensidyne Torque Tool to tighten the membrane cap. The sensor membrane cap may be damaged by over-tightening. 11) Position the sensor with the membrane cap facing down. Lightly tap the sensor 3-4 times to moisten the inside of the new membrane. 12) Once all applicable maintenance procedures have been performed, mount the sensor by presssing it onto the coaxial (BNC) connector. Mount the sensor in the transmitter by inserting the three (3) guide pins into the sensor interface assembly. Rotate the sensor assembly until the guide pins are properly seated in the counter bore holes. Push the sensor assembly upward to engage the electrical connection. 13) Make certain the sensor retainer is properly installed. 14) Allow the sensor to stabilize for one (1) hour before attempting calibration. 22

23 APPENDIX A USING THE SENSOR INTERFERENT TABLES Although highly selective for a particular gas, a sensor is not always completely specific. Certain sensor types will also respond to another gas, particularly if the other gas has similar physical/chemical properties. This other gas is called an interferent. Every effort is made to identify and minimize these interferences. NOTE The Tables are intended for guidance only. For the most accurate measurements the monitor should be calibrated using the target gas under investigation. Example shown below: Chloride The (Threshold Limit Value) for each interferent is included in this column. The is listed in (parts per million), %v (% by volume) or in some cases as an (ACGIH, 1999). A C before a indicates a ceiling limit. The exposure concentration level for the interferent is listed here. A single number (or range of numbers) represents the amount of interferent needed to register a sensor response as if it were the target gas. If the number is paired with in the Column any gas amount below that number will have no effect on the sensor %v Ethylene 3.1 %v 0 %v Cyanide C %v 0 no data Sulfide %v Nitric Oxide 0 Nitrogen Sulfur The (Lower Explosive Limit) for each interferent is included in this column. The is listed in %v (% by volume). Three asterisks () mean the interferent is not combustible in air under normal conditions. The response of the sensor to the interferent is listed here. If the response is listed as then the interferent has no effect on the sensor below the amount listed in the column. If the response is listed as + 1 the effect on the interferent is positive (additive). This means that the sensor will act as if there were 1 more of the target gas then is actually the case. If the response is listed as 1 the effect of the interferent is negative (subtractive). This means that the sensor will act as if there were 1 less of the target gas then is actually the case. The unit of measure can be either, %vol, or %. INTERFERENT NOTE If an interferent is present and there is no target gas, certain transmitters will not display the interferent response until the EFFECT of the interferent reaches ± 4. This is due to a display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations AS WHOLE NUMBERS ONLY (no decimals) [see Sensor Table on page 3 for list of applicable sensors]. 23

24 APPENDIX A AMMONIA (Low ) (0-50 ) Part No D-3X Minimum Indicated Concentration... 2 Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 20 seconds, successive exposures T 90 : < 60 seconds, successive exposures Recovery Time (Fall)... T : < 120 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow rate LPM Oxygen Requirement... 1% by volume, minimum AMMONIA Alcohols Hydrocarbons ol range 1 %vol Chloride C 5 5 Cyanide C Sulfide + 1 * (*) Long term exposure may damage sensor. () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 24

25 AMMONIA (Low ) (0-0 ) Part No D-1 APPENDIX A Minimum Indicated Concentration... 4 Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 20 seconds, successive exposures T 90 : < 60 seconds, successive exposures Recovery Time (Fall)... T : < 120 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow rate LPM Oxygen Requirement... 1% by volume, minimum AMMONIA Alcohols Hydrocarbons ol range 1 %vol Chloride C 5 5 Cyanide C Sulfide + 1 * (*) Long term exposure may damage sensor. () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals).

26 APPENDIX A AMMONIA (Low ) (0-300 ) Part No D-2X Minimum Indicated Concentration... Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 20 seconds, successive exposures T 90 : < 60 seconds, successive exposures Recovery Time (Fall)... T : < 120 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow rate LPM Oxygen Requirement... 1% by volume, minimum AMMONIA Alcohols Hydrocarbons ol range 1 %vol Chloride C 5 5 Cyanide C Sulfide + 1 * (*) Long term exposure may damage sensor. () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 26

27 AMMONIA (Low ) (0-500 ) Part No D-1 APPENDIX A Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 20 seconds, successive exposures T 90 : < 90 seconds, successive exposures Recovery Time (Fall)... T : < 120 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow rate LPM Oxygen Requirement... 1% by volume, minimum AMMONIA Alcohols Hydrocarbons ol range 1 %vol Chloride C 5 5 Cyanide C Sulfide + 1 * (*) Long term exposure may damage sensor. () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 27

28 APPENDIX A ARSINE ( ) Part No D-1 Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 30 seconds, successive exposures Recovery Time (Fall)... T : < 60 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow rate LPM Oxygen Requirement... 1% by volume, minimum ARSINE Acetylene 2.5 -range No Data-Pos(+) Diborane Germane Hydrocarbons ol range Phosphine Silane Sulfur No Data () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 28

29 CARBON DIOXIDE (INFRARED) [use only with IR Transmitters] ( %vol) Part No D-1 APPENDIX A Minimum Indicated Concentration %vol Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Zero Drift... < 0.01 %vol change per month (typical) Span Drift... < 0.05 %vol change per month (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 30 seconds, successive exposures Temperature Range to 55 C (-4 to 131 F)** Recommended Calibration Flow rate LPM ** Above 51 C transmitter LCD begins to fade. 29

30 APPENDIX A CARBON MONOXIDE (0-0 ) Part No D-1 Minimum Indicated Concentration... 4 Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 30 seconds, successive exposures Recovery Time (Fall)... T : < 45 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow rate LPM Oxygen Requirement... 1% by volume, minimum CARBON MONOXIDE Ethylene Chloride C 5 5 Cyanide C Sulfide Nitric Oxide Nitrogen Sulfur () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 30

31 CARBON MONOXIDE (0-500 ) Part No D-2 APPENDIX A Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 30 seconds, successive exposures Recovery Time (Fall)... T : < 45 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow rate LPM Oxygen Requirement... 1% by volume, minimum CARBON MONOXIDE Ethylene Chloride C 5 5 Cyanide C Sulfide Nitric Oxide Nitrogen Sulfur () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 31

32 APPENDIX A CARBON MONOXIDE (0-00 ) Part No D-3X Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 30 seconds, successive exposures Recovery Time (Fall)... T : < 45 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow rate LPM Oxygen Requirement... 1% by volume, minimum CARBON MONOXIDE Ethylene Chloride C 5 5 Cyanide C Sulfide Nitric Oxide Nitrogen Sulfur () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 32

33 CARBON MONOXIDE (Low Interference) (0-0 ) Part No D-1 APPENDIX A Minimum Indicated Concentration... 4 Repeatability... ± 2% of reading Accuracy *... ± % of full scale Span Drift... < 2% change per month (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 30 seconds, successive exposures Recovery Time (Fall)... T : < 45 seconds Temperature Range to 45 C (-4 to 113 F) Recommended Calibration Flow rate LPM Oxygen Requirement... 1% by volume, minimum HYDROGEN CROSS-INTERFERENCE with Temperature (0 H2 Challenge) HYDROGEN CROSS-INTERFERENCE with Concentration (@ C) Temperature Sensor Output Concentration Sensor Output 23 C C ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 33

34 APPENDIX A CARBON MONOXIDE (Low Interference) (0-500 ) Part No D-2 Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± % of full scale Span Drift... < 2% change per month (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 30 seconds, successive exposures Recovery Time (Fall)... T : < 45 seconds Temperature Range to 45 C (-4 to 113 F) Recommended Calibration Flow rate LPM Oxygen Requirement... 1% by volume, minimum HYDROGEN CROSS-INTERFERENCE with Temperature (0 H2 Challenge) HYDROGEN CROSS-INTERFERENCE with Concentration (@ C) Temperature Sensor Output Concentration Sensor Output 23 C C ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 34

35 CHLORINE ( ) Part No D-2 APPENDIX A Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 30 seconds, successive exposures Recovery Time (Fall)... T : < 60 seconds Temperature Range to 50 C (-31 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum CHLORINE Bromine Fluorine Hydrocarbons ol range,000 Chloride C Cyanide C Sulfide Nitrogen Sulfur 20 * (*) Long term exposure may damage sensor. () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 35

36 APPENDIX A CHLORINE (0-.0 ) Part No D-1 Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 30 seconds, successive exposures Recovery Time (Fall)... T : < 60 seconds Temperature Range to 50 C (-31 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum CHLORINE Bromine Fluorine Hydrocarbons ol range,000 Chloride C Cyanide C Sulfide Nitrogen Sulfur 20 * (*) Long term exposure may damage sensor. () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 36

37 CHLORINE (Refillable) (0-.0 ) Part No D-1 APPENDIX A Minimum Indicated Concentration Repeatability... ± % of reading Accuracy *... ± % of reading Span Drift... < % change per month (typical) Time (Rise)... T 50 : < 30 seconds, successive exposures T 90 : < 60 seconds, successive exposures Recovery Time (Fall)... T : < 180 seconds Temperature Range (continuous) to 45 C (23 F to 113 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... Position Requirement... ± 30 from vertical CHLORINE Bromine Fluorine Hydrocarbons ol range Chloride C Cyanide C Sulfide Nitrogen Sulfur No Data () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 37

38 APPENDIX A CHLORINE (0-0 ) Part No D-3 Minimum Indicated Concentration... 4 Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 18 seconds, (typical) T 90 : < 60 seconds, successive exposures Recovery Time (Fall)... T : < 60 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum CHLORINE Bromine Ethylene Chloride C 5 5 Cyanide C Sulfide Nitric Oxide Nitrogen Ozone Sulfur () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 38

39 CHLORINE DIOXIDE ( ) Part No D-1 APPENDIX A Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 20 seconds, successive exposures T 90 : < 120 seconds, successive exposures Recovery Time (Fall)... T : < 60 seconds Temperature Range (continuous) to 50 C (-4 F to 122 F) Humidity Range (continuous) %RH, non-condensing Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum CHLORINE DIOXIDE Hydrocarbons ol range Cyanide C Sulfide Ozone 20 * * At exposures < 15 hrs, longer/higher exposures can break through the filter, causing negative readings as well as filter blockage to ClO 2. () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 39

40 APPENDIX A CHLORINE DIOXIDE ( ) Part No D-2 Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 20 seconds, successive exposures T 90 : < 120 seconds, successive exposures Recovery Time (Fall)... T : < 60 seconds Temperature Range (continuous) to 50 C (-4 F to 122 F) Humidity Range (continuous) %RH, non-condensing Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum CHLORINE DIOXIDE Hydrocarbons ol range Cyanide C Sulfide Ozone 20 * * At exposures < 15 hrs, longer/higher exposures can break through the filter, causing negative readings as well as filter blockage to ClO 2. () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 40

41 ACETYLENE (INFRARED) [use only with IR Transmitters] (0-0 %) Part No D-1 APPENDIX A Minimum Indicated Concentration... 4 % Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Zero Drift... 5% change per year (typical) Span Drift... % change per year (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 30 seconds, successive exposures Temperature Range to 55 C (-4 to 131 F)** Humidity Range (intermittent) %RH, non-condensing Recommended Calibration Flow rate LPM ** Above 51 C transmitter LCD begins to fade. CLASS I, DIVISION 2, GROUPS A, B, C, & D 41

42 APPENDIX A COMBUSTIBLE GAS (CATALYTIC BEAD) (0-0 %) Part No D-1 Minimum Indicated Concentration... 4 % Repeatability... ± 2% of reading Accuracy *... ± 3% 0-50 %, ± 5% 50-0 % Span Drift... < % change per year (typical) Time (Rise)... T 50 : < seconds, successive exposures T 90 : < 30 seconds, successive exposures Recovery Time (Fall)... T : < 30 seconds Temperature Range (continuous) to 55 C (-4 F to 131 F) Humidity Range (continuous) %RH, non-condensing Recommended Calibration Flow Rate LPM Oxygen Requirement... %vol, minimum Recovery times may increase to 5 minutes in explosion-proof versions due to gas trapped in the void space inside the sensor cover. NOTE For Methane and Propane k-factors see Appendix B 42

43 COMBUSTIBLE GAS (CATALYTIC BEAD for ) (0-0 %) Part No D-1 APPENDIX A Minimum Indicated Concentration... 4 % Repeatability... ± 2% of reading Accuracy *... ± 3% 0-50 %, ± 5% 50-0 % Span Drift... < % change per year (typical) Time (Rise)... T 50 : < seconds, successive exposures T 90 : < 30 seconds, successive exposures Recovery Time (Fall)... T : < 30 seconds Temperature Range (continuous) to 55 C (-4 F to 131 F) Humidity Range (continuous) %RH, non-condensing Recommended Calibration Flow Rate LPM Oxygen Requirement... %vol, minimum Recovery times may increase to 5 minutes in explosion-proof versions due to gas trapped in the void space inside the sensor cover. NOTE The specific catalytic bead combustible sensor will also sense gases detectable by the catalytic bead combustible sensor (PNº 1110-D-1), however the interferences are not linear. 43

44 APPENDIX A COMBUSTIBLE GAS (INFRARED for Methane Gas) [use only with IR Transmitters] (0-0 %) Part No D-1 Minimum Indicated Concentration... 4 % Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Zero Drift... 5% change per year (typical) Span Drift... % change per year (typical) Time (Rise)... T 50 : < 6 seconds, (typical) T 90 : < 20 seconds, successive exposures Temperature Range to 55 C (-4 to 131 F)** Recommended Calibration Flow rate LPM ** Above 51 C transmitter LCD begins to fade. NOTE The Methane IR sensor will also sense gases detectable by the non-methane IR sensor (PNº 1115-D-2), however the interferences are not linear. 44

45 COMBUSTIBLE GAS (INFRARED for non-methane Hydrocarbons) [use only with IR Transmitters] (0-0 %) Part No D-2 APPENDIX A Minimum Indicated Concentration... 4 % Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Zero Drift... 5% change per year (typical) Span Drift... % change per year (typical) Time (Rise)... T 50 : < 6 seconds, (typical) T 90 : < 20 seconds, successive exposures Temperature Range to 55 C (-4 to 131 F)** Recommended Calibration Flow rate LPM ** Above 51 C transmitter LCD begins to fade. Combustible Gas/Vapor Propane Equivalents Propane k-factor Acetone 2. 4 Butadiene 2. 5 Butane 1. 2 Cyclopentane 0. 8 Dimethyl Ether 1. 0 Ethyl Acetate 1. 4 Ethylene 2. 4 Ethylene Oxide 2. 2 Heptane (estimated data) 2. 0 Hexane 1. 6 Methanol 0. 6 Pentane (use for Gasoline) 1. 2 Propane 1. 0 Propylene 1. 8 Tetrahydrofuran 1. 1 o-xylene

46 APPENDIX A DIBORANE ( ) Part No D-1 Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 45 seconds, (typical) T 90 : < seconds, successive exposures Recovery Time (Fall)... T : < 30 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum DIBORANE Acetylene 2.5 range No Data-Pos(+) Arsine Hydrocarbons Germane ol range Phosphine Silane Sulfur No Data () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 46

47 ETHYLENE OXIDE (0-.0 ) Part No D-1 APPENDIX A Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 45 seconds, (typical) T 90 : < 140 seconds, successive exposures Recovery Time (Fall)... T : < 0 seconds Temperature Range ** to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum ETHYLENE OXIDE Ethanol Ethylene Sulfide Methanol Nitric Oxide Nitrogen Oxide Ozone Sulfur Toluene () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 47

48 APPENDIX A FLUORINE (0-.0 ) Part No D-1 Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 30 seconds, successive exposures Recovery Time (Fall)... T : < 60 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum FLUORINE Bromine Hydrocarbons 3.1 %-range 00 Chloride C Cyanide C Sulfide Nitric Oxide Nitrogen Sulfur 33-1 * () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). (*) Continuous exposure will poison cell. ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 48

49 FLUORINE (0-.0 ) Part No D-2 APPENDIX A Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 30 seconds, successive exposures Recovery Time (Fall)... T : < 60 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum FLUORINE Bromine Hydrocarbons 3.1 %-range 00 Chloride C Cyanide C Sulfide Nitric Oxide Nitrogen Sulfur 33-1 * () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). (*) Continuous exposure will poison cell. ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 49

50 APPENDIX A GERMANE ( ) Part No D-1 Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 30 seconds, successive exposures Recovery Time (Fall)... T : < 180 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum GERMANE Acetylene 2.5 range No Data-Pos(+) Arsine Diborane Hydrocarbons ol range 3000 Phosphine Silane Sulfur () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 50

51 HYDRAZINE ( ) Part No D-1 APPENDIX A Minimum Indicated Concentration Repeatability... ± 5% of reading Accuracy *... ± % of full scale Span Drift... < 5% change per month Time (Rise)... T 50 : < 30 seconds, (typical) T 90 : < 120 seconds, successive exposures Recovery Time (Fall)... T : < 120 seconds [estimated] Temperature Range ** to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum HYDRAZINE Arsine Hydrocarbons Chloride ol Range C Cyanide C Sulfide Nitrogen Phosphine () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 51

52 APPENDIX A HYDROGEN (0-00 ) Part No D-1 Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 30 seconds, successive exposures Recovery Time (Fall)... T : < 60 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum HYDROGEN Ethylene Chloride C 5 5 Cyanide C Sulfide Nitric Oxide Nitrogen Sulfur () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 52

53 HYDROGEN (0-0 %, Electrochemical) Part No D-1 APPENDIX A Minimum Indicated Concentration... 4 % Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 30 seconds, (typical) T 90 : < 120 seconds, successive exposures Recovery Time (Fall)... T : < 120 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum HYDROGEN Ethylene Chloride C 5 5 Cyanide C Sulfide Methane %vol Nitric Oxide Nitrogen () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 53

54 APPENDIX A HYDROGEN CHLORIDE (0-.0 ) Part No D-3X Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < 3% change per month (typical) Time (Rise)... T 50 : < 30 seconds, successive exposures T 90 : < 70 seconds, successive exposures Recovery Time (Fall)... T : < 90 seconds Temperature Range (continuous) to 50 C (-4 F to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum HYDROGEN CHLORIDE Hydrocarbons Cyanide C 4.7 % Range 1 %vol Sulfide H ydrogen Bromide Sulfur () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 54

55 HYDROGEN CHLORIDE (Refillable) (0-.0 ) Part No D-1X APPENDIX A Minimum Indicated Concentration Repeatability... ± % of reading Accuracy *... ± % of full scale Span Drift... < 3% change per month (typical) Time (Rise)... T 50 : < 40 seconds, successive exposures T 90 : < 180 seconds, successive exposures Recovery Time (Fall)... T : < 180 seconds Temperature Range (continuous) to 40 C (23 F to 4 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... Position Requirement... ± 30 from vertical HYDROGEN CHLORIDE Hydrocarbons Cyanide C 4.7 % Range 5 %vol Sulfide Fluoride Sulfur 200 C () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 55

56 APPENDIX A HYDROGEN CHLORIDE ( ) Part No D-1 Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < 3% change per month (typical) Time (Rise)... T 50 : < 30 seconds, successive exposures T 90 : < 70 seconds, successive exposures Recovery Time (Fall)... T : < 90 seconds Temperature Range (continuous) to 50 C (-4 F to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum HYDROGEN CHLORIDE Hydrocarbons Cyanide C 4.7 % Range 1 %vol Sulfide H ydrogen Bromide Sulfur () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 56

57 HYDROGEN CHLORIDE (0-0 ) Part No D-2 APPENDIX A Minimum Indicated Concentration... 4 Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 36 seconds, (typical) T 90 : < 120 seconds, successive exposures Recovery Time (Fall)... T : < 90 seconds Temperature Range (continuous) to 50 C (-4 F to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum HYDROGEN CHLORIDE Ethylene Cyanide C Sulfide Nitric Oxide Nitrogen Sulfur () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 57

58 APPENDIX A HYDROGEN CHLORIDE (Refillable) (0-0 ) Part No D-2 Minimum Indicated Concentration... 4 Repeatability... ± % of reading Accuracy *... ± % of full scale Span Drift... < 3% change per month (typical) Time (Rise)... T 50 : < 40 seconds, successive exposures T 90 : < 180 seconds, successive exposures Recovery Time (Fall)... T : < 180 seconds Temperature Range (continuous) to 40 C (23 F to 4 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... Position Requirement... ± 30 from vertical HYDROGEN CHLORIDE Hydrocarbons Cyanide C 4.7 % Range 5 %vol Sulfide Fluoride Sulfur 200 C () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 58

59 HYDROGEN CYANIDE ( ) Part No D-1 APPENDIX A Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < 5% change per month (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 50 seconds, successive exposures Recovery Time (Fall)... T : < 0 seconds Temperature Range to 50 C (-4 to 122 F) Humidity Range (continuous) %RH, non-condensing Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum HYDROGEN CYANIDE Hydrocarbons Nitric Nitrogen Oxide Chloride Sulfide 0 % range 1% V 5 no data 20 * *Note: 40 reading after filter breaktrough. () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 59

60 APPENDIX A HYDROGEN CYANIDE (0-0 ) Part No D-2X Minimum Indicated Concentration... 4 Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 45 seconds, (typical) T 90 : < 150 seconds, successive exposures Recovery Time (Fall)... T : < 0 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum HYDROGEN CYANIDE Ethylene Chloride C no data Sulfide Nitric Oxide Nitrogen Sulfur 15 (see Note) Note: Due to the extremely high cross sensitivity (> 350%), this sensor is not suitable for use in atmospheres containing H 2 S. () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 60

61 HYDROGEN FLUORIDE (0-.0 ) Part No D-1 APPENDIX A Minimum Indicated Concentration Repeatability... ± 5% of reading Accuracy *... ± % of full scale Span Drift... < % change per month (typical) Time (Rise)... T 50 : < 40 seconds T 90 : < 90 seconds Recovery Time (Fall)... T : < 70 seconds Temperature Range (continuous) to 50 C (-4 F to 122 F) (Note: High humidity causes HF absorbtion) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum Arsine Boron Trifluoride Trifluoride Germane Hydrochloric Acid Sulfide Nitrogen Silicon Fluoride Sulfur HYDROGEN FLUORIDE C * C no data C 5 1 %vol mg/ m (*) may damage sensor. () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 61

62 APPENDIX A HYDROGEN FLUORIDE ( ) Part No D-2X Minimum Indicated Concentration Repeatability... ± 5% of reading Accuracy *... ± % of full scale Span Drift... < % change per month (typical) Time (Rise)... T 50 : < 40 seconds T 90 : < 90 seconds Recovery Time (Fall)... T : < 70 seconds Temperature Range (continuous) to 50 C (-4 F to 122 F) (Note: High humidity causes HF absorbtion) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum Arsine Boron Trifluoride Trifluoride Germane Hydrochloric Acid Sulfide Nitrogen Silicon Fluoride Sulfur HYDROGEN FLUORIDE C * C no data C 5 1 %vol mg/ m (*) may damage sensor. () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 62

63 HYDROGEN FLUORIDE (Refillable) (0-.0 ) Part No D-1 APPENDIX A Minimum Indicated Concentration Repeatability... ± % of reading Accuracy *... ± % of reading Span Drift... < 3% change per month (typical) Time (Rise)... T 50 : < 60 seconds, successive exposures T 90 : < 300 seconds, successive exposures Recovery Time (Fall)... T : < 300 seconds Temperature Range (continuous) to 40 C (23 F to 4 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... Position Requirement... ± 30 from vertical HYDROGEN FLUORIDE Hydrocarbons Cyanide C 4.7 % Range 5 %vol Sulfide Chloride Sulfur 200 C () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). (#) will damage sensor. ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. * Level varies with moisture content. 63

64 APPENDIX A HYDROGEN SULFIDE (0-50 ) Part No D-1 Minimum Indicated Concentration... 2 Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 30 seconds, successive exposures Recovery Time (Fall)... T : < 60 seconds Temperature Range to 50 C (-31 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum HYDROGEN SULFIDE Ammonia Ethylene Chloride C 5 5 Cyanide C Nitric Oxide Nitrogen Sulfur () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 64

65 HYDROGEN SULFIDE (0-0 ) Part No D-2 APPENDIX A Minimum Indicated Concentration... 4 Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 30 seconds, successive exposures Recovery Time (Fall)... T : < 60 seconds Temperature Range to 50 C (-31 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum HYDROGEN SULFIDE Ammonia Ethylene Chloride C 5 5 Cyanide C Nitric Oxide Nitrogen Sulfur () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 65

66 APPENDIX A NITRIC OXIDE (0-0 ) Part No D-1 Minimum Indicated Concentration... 4 Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 5 seconds, (typical) T 90 : < 15 seconds, successive exposures Recovery Time (Fall)... T : < 30 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum NITRIC OXIDE Ethylene Chloride C Cyanide C Sulfide Nitrogen Sulfur () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 4. This is due to display blanking that occurs between -3 and + 3 on transmitters that display gas concentrations as whole numbers (no decimals). 66

67 NITROGEN DIOXIDE (0-.0 ) Part No D-1 APPENDIX A Minimum Indicated Concentration Repeatability... ± 5% of reading Accuracy *... ± % of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 15 seconds, (typical) T 90 : < 40 seconds, successive exposures Recovery Time (Fall)... T : < 45 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum NITROGEN DIOXIDE Ethylene Chloride C 5 5 Cyanide C Sulfide Nitric Oxide Sulfur () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 67

68 APPENDIX A OXYGEN (Process) (0-.0 %vol) Part No D-2X Min. Indicated Concentration Change %vol Span Drift... < % change per year (typical) Time (Rise)... T 90 : < seconds (from 0 to 20.9 %vol O 2 ) Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM OXYGEN [%vol] 5000 %vol + 1 %vol %vol < 1 %vol 0.5 > 1 %vol + 1 %vol Methane 50 %vol < 1 %vol 5.0 < 0 %vol Nitrogen Ozone 3 2 %vol + 1 %vol 0.05 > 1 %vol + 1 %vol () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 0.4. This is due to display blanking that occurs between %vol and %vol on transmitters that display gas concentrations with one digit after the decimal. 68

69 OXYGEN (Ambient Air) (0-.0 %vol) Part No D-1 APPENDIX A Min. Indicated Concentration Change %vol Span Drift... < % change per year (typical) Time (Rise)... T 90 : < seconds (from 0 to 20.9 %vol O 2 ) Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM OXYGEN [%vol] 5000 %vol + 1 %vol %vol < 1 %vol 0.5 > 1 %vol + 1 %vol Methane 50 %vol < 1 %vol 5.0 < 0 %vol Nitrogen Ozone 3 2 %vol + 1 %vol 0.05 > 1 %vol + 1 %vol () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 0.4. This is due to display blanking that occurs between %vol and %vol on transmitters that display gas concentrations with one digit after the decimal. 69

70 APPENDIX A OXYGEN (Process) (0-.0 %vol) Part No D-3X Min. Indicated Concentration Change %vol Span Drift... < % change per year (typical) Time (Rise)... T 90 : < seconds (from 0 to 20.9 %vol O 2 ) Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM OXYGEN [%vol] 5000 %vol + 1 %vol %vol < 1 %vol 0.5 > 1 %vol + 1 %vol Methane 50 %vol < 1 %vol 5.0 < 0 %vol Nitrogen Ozone 3 2 %vol + 1 %vol 0.05 > 1 %vol + 1 %vol () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± 0.4. This is due to display blanking that occurs between %vol and %vol on transmitters that display gas concentrations with one digit after the decimal. 70

71 OZONE ( ) Part No D-1 APPENDIX A Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 45 seconds, (typical) T 90 : < 150 seconds, successive exposures Recovery Time (Fall)... T : < 120 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum OZONE Ethylene Chloride C 5 5 Cyanide C Sulfide Nitric Oxide Nitrogen Sulfur () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 71

72 APPENDIX A PHOSGENE ( ) Part No D-1 Minimum Indicated Concentration Repeatability... ± 5% of reading Accuracy *... ± % of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 30 seconds, successive exposures T 90 : < 120 seconds, successive exposures Recovery Time (Fall)... T : consult factory Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum Arsine PHOSGENE 0.05 ca. + 1 Alcohols Hydrocarbons range Chloride Cyanide Sulfide Nitrogen Phosphine Sulfur 1 C 5 5 * C * 1 * a sphyxiant * Gases can poison sensor after filter breakthrough. () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 72

73 PHOSPHINE ( ) Part No D-1 APPENDIX A Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 30 seconds, successive exposures Recovery Time (Fall)... T : < 180 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum PHOSPHINE Acetylene 2.5 range No Data-Pos(+) Arsine Diborane Germane Hydrocarbons ol range 3000 Silane Sulfur No Data () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 73

74 APPENDIX A SILANE (0.0 ) Part No D-1 Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 30 seconds, (typical) T 90 : < 90 seconds, successive exposures Recovery Time (Fall)... T : < 90 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum SILANE Arsine Diborane Germane Hydrocarbons ol range 3000 Phosphine Sulfur No Data () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 74

75 STYRENE (0-200 ) Part No D-1X APPENDIX A Minimum Indicated Concentration... 7 Repeatability... ± 3% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < seconds, (typical) T 90 : < 75 seconds, successive exposures Recovery Time (Fall)... T : < 120 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum Acetylene Ammonia STYRENE yl Sulfide Ethylene Chloride Cyanide Sulfide Nitric Oxide Nitrogen Phosphine Sulfur Toluene C C () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 75

76 APPENDIX A SULFUR DIOXIDE (0.0 ) Part No D-3X Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 5 seconds, (typical) T 90 : < 15 seconds, successive exposures Recovery Time (Fall)... T : < 30 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum SULFUR DIOXIDE Ethylene Chloride C Cyanide C Sulfide Nitric Oxide Nitrogen () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 76

77 SULFUR DIOXIDE (Refillable) (0-.0 ) Part No D-1X APPENDIX A Minimum Indicated Concentration Repeatability... ± % of reading Accuracy *... ± % of reading Span Drift... < 3% change per month (typical) Time (Rise)... T 50 : < 40 seconds, successive exposures T 90 : < 240 seconds, successive exposures Recovery Time (Fall)... T : < 240 seconds Temperature Range (continuous) to 40 C (23 F to 4 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... Position Requirement... ± 30 from vertical SULFUR DIOXIDE Hydrocarbons Cyanide C 4.7 % Range 5 %vol Sulfide Chloride Fluoride 200 C C () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 77

78 APPENDIX A SULFUR DIOXIDE ( ) Part No D-1 Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 5 seconds, (typical) T 90 : < 15 seconds, successive exposures Recovery Time (Fall)... T : < 30 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum SULFUR DIOXIDE Ethylene Chloride C Cyanide C Sulfide Nitric Oxide Nitrogen () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 78

79 SULFUR DIOXIDE ( ) Part No D-1 APPENDIX A [Filtered] Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 5 seconds, (typical) T 90 : < 20 seconds, successive exposures Recovery Time (Fall)... T : < 30 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum SULFUR DIOXIDE (Filtered) Ethylene Chloride C 5 5 Cyanide C Sulfide Nitric Oxide Nitrogen () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 79

80 APPENDIX A SULFUR DIOXIDE (0 0 ) Part No D-2X [Filtered] Minimum Indicated Concentration Repeatability... ± 2% of reading Accuracy *... ± 2% of full scale Span Drift... < % change per year (typical) Time (Rise)... T 50 : < 5 seconds, (typical) T 90 : < 20 seconds, successive exposures Recovery Time (Fall)... T : < 30 seconds Temperature Range to 50 C (-4 to 122 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... 1% by volume, minimum SULFUR DIOXIDE (Filtered) Ethylene Chloride C 5 5 Cyanide C Sulfide Nitric Oxide Nitrogen () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 80

81 SULFUR DIOXIDE (Refillable) (0-0 ) Part No D-2X APPENDIX A Minimum Indicated Concentration... 4 Repeatability... ± % of reading Accuracy *... ± % of reading Span Drift... < 3% change per month (typical) Time (Rise)... T 50 : < 40 seconds, successive exposures T 90 : < 240 seconds, successive exposures Recovery Time (Fall)... T : < 240 seconds Temperature Range (continuous) to 40 C (23 F to 4 F) Recommended Calibration Flow Rate LPM Oxygen Requirement... Position Requirement... ± 30 from vertical SULFUR DIOXIDE Hydrocarbons Cyanide C 4.7 % Range 5 %vol Sulfide Chloride Fluoride 200 C C () means the substance is not combustible in air under normal conditions. C Denotes a ceiling (in column). ± This is due to display blanking that occurs between and on transmitters that display gas concentrations with two digits after the decimal. 81

82 APPENDIX B METHANE & PROPANE K-FACTORS FOR CATALYTIC SENSORS Catalytic Bead Combustible Sensor The table below shows the variation in response for the Combustible Gas Sensor as K-factors (multipliers). These figures are experimentally derived and are expressed relative to Methane Propane. Using the K-factor: Multiply the Methane or Propane % challenge concentration by the respective K-factor to obtain the span value. Note that SensAlert Sensors will not operate above 0 % span. Combustible Gas/Vapor Methane K-factor Propane K-factor Methane Acetaldehyde Acetic acid Acetic anhydride Acetone Acetonitrile Acetylene Ammonia Aniline Benzene ,3-Butadiene n-butane iso-butane Butene cis-butene trans-butene n-butyl alcohol n-butyric acid disulphide monoxide yl sulphide Chlorobenzene Cyanogen Cyclohexane Cyclopropane (table continued on next page) IMPORTANT Information about substances not listed in this table was unavailable at the time of this printing. Listed K-factors are not warranted, but can be used to estimate the % concentration of substances from direct readings of Methane or Propane. K-factors can be used to obtain approximate direct readings of % concentration of substances by calibration with Methane or Propane. Error varies from one sensor to another and with the age of the sensor assembly. The typical K-factor conversion error can be as high as 20 30%. In order to achieve the most accurate detection of a substance, calibration should be performed using a know % concentration of the substance. 82

83 Catalytic Bead Combustible Sensor APPENDIX B METHANE & PROPANE K-FACTORS FOR CATALYTIC SENSORS The table below shows the variation in response for the Combustible Gas Sensor as K-factors (multipliers). These figures are experimentally derived and are expressed relative to Methane Propane. Using the K-factor: Multiply the Methane or Propane % challenge concentration by the respective K-factor to obtain the span value. Note that SensAlert Sensors will not operate above 0 % span. Combustible Gas/Vapor Methane K-factor Propane K-factor Methane n-decane Diethyl ether Diiso-propyl ether Dimethyl ether Dimethyl sulphide Dimethylbutane Dimethylhydrazine Dimethylpentane ,4 Dioxane Ethane Ethyl acetate Ethyl alcohol Ethyl bromide Ethyl chloride Ethyl formate Ethyl mercaptan Ethyl methyl ether Ethylamine Ethylbenzene Ethylene Ethylene dichloride Ethylene oxide Ethylpentane Gasoline n-heptane (table continued on next page) IMPORTANT Information about substances not listed in this table was unavailable at the time of this printing. Listed K-factors are not warranted, but can be used to estimate the % concentration of substances from direct readings of Methane or Propane. K-factors can be used to obtain approximate direct readings of % concentration of substances by calibration with Methane or Propane. Error varies from one sensor to another and with the age of the sensor assembly. The typical K-factor conversion error can be as high as 20 30%. In order to achieve the most accurate detection of a substance, calibration should be performed using a know % concentration of the substance. 83

84 APPENDIX B METHANE & PROPANE K-FACTORS FOR CATALYTIC SENSORS Catalytic Bead Combustible Sensor The table below shows the variation in response for the Combustible Gas Sensor as K-factors (multipliers). These figures are experimentally derived and are expressed relative to Methane Propane. Using the K-factor: Multiply the Methane or Propane % challenge concentration by the respective K-factor to obtain the span value. Note that SensAlert Sensors will not operate above 0 % span. Combustible Gas/Vapor Methane K-factor Propane K-factor Methane ,4-Hexadiene n-hexane Hydrazine cyanide sufide iso-butyl alcohol iso-propyl alcohol Isobutylene Methyl acetate Methyl alcohol Methyl bromide Methyl chloride Methyl ethyl ketone Methyl formate Methyl mercaptan Methyl propionate Methyl propyl ketone Methylamine Methylcyclohexane Methylene chloride Methylhexane Methylhydrazine Methylpentane (table continued on next page) IMPORTANT Information about substances not listed in this table was unavailable at the time of this printing. Listed K-factors are not warranted, but can be used to estimate the % concentration of substances from direct readings of Methane or Propane. K-factors can be used to obtain approximate direct readings of % concentration of substances by calibration with Methane or Propane. Error varies from one sensor to another and with the age of the sensor assembly. The typical K-factor conversion error can be as high as 20 30%. In order to achieve the most accurate detection of a substance, calibration should be performed using a know % concentration of the substance. 84

85 APPENDIX B METHANE & PROPANE K-FACTORS FOR CATALYTIC SENSORS Catalytic Bead Combustible Sensor The table below shows the variation in response for the Combustible Gas Sensor as K-factors (multipliers). These figures are experimentally derived and are expressed relative to Methane Propane. Using the K-factor: Multiply the Methane or Propane % challenge concentration by the respective K-factor to obtain the span value. Note that SensAlert Sensors will not operate above 0 % span. Combustible Gas/Vapor Methane K-factor Propane K-factor Methane Nitromethane n-nonane n-octane n-pentane iso-pentane neo-pentane Pentene Propane Propene n-propyl alcohol n-propyl chloride n-propylamine ,2-Propylene oxide Propyne tert-butyl alcohol Tetrahydrofuran Toluene Triethylamine Trimethylamine Trimethylbutane Vinyl chloride o-xylene m-xylene p-xylene IMPORTANT Information about substances not listed in this table was unavailable at the time of this printing. Listed K-factors are not warranted, but can be used to estimate the % concentration of substances from direct readings of Methane or Propane. K-factors can be used to obtain approximate direct readings of % concentration of substances by calibration with Methane or Propane. Error varies from one sensor to another and with the age of the sensor assembly. The typical K-factor conversion error can be as high as 20 30%. In order to achieve the most accurate detection of a substance, calibration should be performed using a know % concentration of the substance. 85

86 APPENDIX C RETURNED MATERIAL AUTHORIZATION Sensidyne maintains an instrument service facility at the factory to provide its customers with both warranty and non-warranty repair. Sensidyne assumes no liability for service performed by personnel other than Sensidyne personnel. To facilitate the repair process, please contact the Sensidyne Service Department in advance for assistance with a problem which cannot be remedied and/or requires the return of the product to the factory. All returned products require a Returned Material Authorization (RMA) number. Sensidyne Service Department personnel may be reached at: Sensidyne th CIRCLE N, SUITE 0 ST. PETERSBURG, FL USA [Service Fax] All non-warranty repair orders will have a minimum fee of $50 whether the repair is authorized or not. This fee includes handling, administration and technical expenses for inspecting the instrument and providing an estimate. However, the estimate fee is waived if the repair is authorized. If you wish to set a limit to the authorized repair cost, state a not to exceed figure on your purchase order. Please indicate if a price quotation is required before authorization of the repair cost, understanding that this invokes extra cost and handling delay. Sensidyne s repair policy is to perform all needed repairs to restore the instrument to its full operating condition. Repairs are handled on a first in - first out basis. Your order may be expedited if you authorize an expediting fee. This will place your order next in line behind orders currently in process. Pack the instrument and its accessories (preferably in their original packing) and enclose your return address, purchase order, shipping and billing information, RMA number, a description of the problem encountered with your instrument and any special instructions. All prices are subject to change without notice. If this is the first time you are dealing directly with the factory, you will be asked to prepay or to authorize a COD shipment. Send the instrument, prepaid, to: SENSIDYNE th CIRCLE N, SUITE 0 ST. PETERSBURG, FL USA ATTENTION: Service Department RMA #: SERVICE OPTIONS The Sensidyne Service Department offers a variety of service options which will minimize costly interruptions and maintenance costs. These options include initial training, on-site technical assistance, and full factory repairs. Sensidyne has developed several programs which offer options best suited to your applications and needs. For further information, contact the Sensidyne Service Department at the following numbers: [Service Fax]. 86

87 THIS PAGE INTENTIONALLY LEFT BLANK

88 Sensidyne, LP th Circle N, Suite 0 St. Petersburg, Florida USA (800) (727) Fax: (727) web: info@sensidyne.com

UNIVERSAL TUBE HOLDER SYSTEM

UNIVERSAL TUBE HOLDER SYSTEM UNIVERSAL TUBE for Constant Low Flow & Constant-Pressure (Multi-Flow) Operation INSTRUCTION MANUAL 16333 Bay Vista Dr. Clearwater, Florida 33760 (800) 451-9444 (727) 530-3602 (727) 539-0550 [FAX] Revision

More information

Operation and Maintenance Manual

Operation and Maintenance Manual Operation and Maintenance Manual GDS 49 Remote 4 20mA Sensor Transmitter for Toxic Gases GDS Corp. 1245 Butler Road League City, TX 77573 409 927 2980 409 927 4180 (Fax) www.gdscorp.com CAUTION: FOR SAFETY

More information

O3 3E 1 F Gas Sensor Module

O3 3E 1 F Gas Sensor Module Product Information Pack O3 3E 1 F Gas Sensor Module (Ozone) CONTENTS Product Data Sheet Product Specification 2 Poisoning and Cross Sensitivities 3 Operating Instructions Introduction 4 Electrostatic

More information

CSA Sample Draw Aspirator Adapter Operator s Manual

CSA Sample Draw Aspirator Adapter Operator s Manual 30-0951-CSA Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0367 Revision: 0 Released: 4/30/15 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating

More information

Air Check Lite Toxic Gas Monitor

Air Check Lite Toxic Gas Monitor Air Check Lite Toxic Gas Monitor Instruction Manual PureAire Monitoring Systems, Inc. 557 Capital Drive Lake Zurich, Illinois 60047 Phone: 847-726-6000 Fax: 847-726-6051 Toll-Free: 888-788-8050 www.pureairemonitoring.com

More information

MODEL CALIBRATION GAS DELIVERY SYSTEM

MODEL CALIBRATION GAS DELIVERY SYSTEM MODEL 1200-26 CALIBRATION GAS DELIVERY SYSTEM Sierra Monitor Corporation 1991 Tarob Court, Milpitas, CA 95035 (408) 262-6611 MODEL 1200-26 CALIBRATION GAS DELIVERY SYSTEM APPLICABILITY & EFFECTIVITY This

More information

RK-IR Sample Draw Aspirator Adapter Operator s Manual

RK-IR Sample Draw Aspirator Adapter Operator s Manual 30-0951RK-IR Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0018RK Revision: A Released: 6/2/10 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment

More information

Calibration Requirements for Direct Reading Confined Space Gas Detectors

Calibration Requirements for Direct Reading Confined Space Gas Detectors : Calibration Requirements for Direct Reading Confined Space Gas Detectors However, the definition of bump test has always been a little slippery. Some manufacturers differentiate between a bump test that

More information

Verification Of Calibration for Direct-Reading Portable Gas Monitors

Verification Of Calibration for Direct-Reading Portable Gas Monitors U. S. Department of Labor Occupational Safety and Health Administration Directorate of Science, Technology and Medicine Office of Science and Technology Assessment Verification Of Calibration for Direct-Reading

More information

GasLab SAN-102 Micro Oxygen Monitor User Manual

GasLab SAN-102 Micro Oxygen Monitor User Manual GasLab SAN-102 Micro Oxygen Monitor User Manual The GasLab SAN-102 is a wearable, personal safety meter designed to monitor ambient oxygen levels in real time. It is designed to protect workers in confined

More information

GasLab SAN-102 Micro Oxygen Monitor User Manual

GasLab SAN-102 Micro Oxygen Monitor User Manual GasLab SAN-102 Micro Oxygen Monitor User Manual The GasLab SAN-102 is a wearable, personal safety meter designed to monitor ambient oxygen levels in real time. It is designed to protect workers in confined

More information

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL -

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL - Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL - Table of Contents 1. Introduction 3 2. Instrument Setup 4 2.1 Instrument Checklist 4 2.2 Optional Accessories 4 2.3 Gas Concentration Ranges Available

More information

Fire Escape Hood OPERATION AND INSTRUCTIONS

Fire Escape Hood OPERATION AND INSTRUCTIONS TM Fire Escape Hood OPERATION AND INSTRUCTIONS THIS MANUAL MUST BE CAREFULLY READ AND FOLLOWED BY ALL PERSONS WHO HAVE, OR WILL HAVE, THE RESPONSIBILITY FOR USING THE S-CAP FIRE ESCAPE HOOD. THESE ESCAPE

More information

Calibration and Bump Test Requirements for Direct Reading Portable Gas Monitors

Calibration and Bump Test Requirements for Direct Reading Portable Gas Monitors : Calibration and Bump Test Requirements for Direct Reading Portable Gas Monitors Manufacturers and regulatory agencies agree the safest and most conservative approach is to perform a functional test by

More information

This written Compressed Gas Plan is kept at the corporate office and in the written Safety & Health Program.

This written Compressed Gas Plan is kept at the corporate office and in the written Safety & Health Program. Safety & Health Program 10627 Midwest Industrial Boulevard, St. Louis, MO Phone: 314-785-6425 Fax: 314-785-6426 Compressed Gas Plan Purpose It is the policy of EMA to permit only trained and authorized

More information

Model ASSEMBLY and OPERATING INSTRUCTIONS

Model ASSEMBLY and OPERATING INSTRUCTIONS QUICK CHANGE AIR BRUSH KIT Model 93506 ASSEMBLY and OPERATING INSTRUCTIONS Due to continuing improvements, actual product may differ slightly from the product described herein. 3491 Mission Oaks Blvd.,

More information

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc. Advanced Calibration Designs, Inc. Calibration Gas Instrument INSTRUCTION MANUAL Release I www.goacd.com Instruction Manual Gas Generator Release I TABLE OF CONTENTS I. General Description Page 2 II. Start-Up

More information

ACRYLIC FLOW METERS 0-30, 0-70 & LPM

ACRYLIC FLOW METERS 0-30, 0-70 & LPM ACRYLIC FLOW METERS 0-30, 0-70 & 0-120 LPM ASSEMBLY INSTRUCTIONS & INSTRUCTIONS FOR USE R219P86 (0-120LPM) R219P87 (0-70LPM) R219P88 (0-30 LPM) R138P11 Rev. D TABLE OF CONTENTS: 1.0 PRODUCT OVERVIEW...1

More information

Phosgene Sensor. Sensoric COCl2 3E 1

Phosgene Sensor. Sensoric COCl2 3E 1 Phosgene Sensor 1 FEATURES Amperometric 3 electrode sensor cell Fixed organic electrolyte High reliability TYPICAL APPLICATIONS Ambient monitoring of TLV levels Chemical Industry, Homeland Security PART

More information

RGC-IR Remote Gas Calibrator for IR400

RGC-IR Remote Gas Calibrator for IR400 Remote Gas Calibrator for IR400 The information and technical data disclosed in this document may be used and disseminated only for the purposes and to the extent specifically authorized in writing by

More information

Sterile Visual Flow Indicator

Sterile Visual Flow Indicator Sterile Visual Flow Indicator Sanitary Process Connection Installation / Operation / Maintenance Manual P.O. Box 1116 Twinsburg, OH 44087 Phone: 330/405-3040 Fax: 330/405-3070 E-mail: view@ljstar.com Web

More information

200 A, 15 and 25 kv class loadbreak bushing insert installation instructions

200 A, 15 and 25 kv class loadbreak bushing insert installation instructions Loadbreak Apparatus Connectors Effective October 2013 Supersedes March 2011 200 A, 15 and 25 kv class loadbreak bushing insert installation instructions S500-12-1 DISCLAIMER OF WARRANTIES AND LIMITATION

More information

Department of Physics Standard Operating Procedures (SOP) Approved by: Physics Safety Committee

Department of Physics Standard Operating Procedures (SOP) Approved by: Physics Safety Committee Page: Page 1 of 5 1 Introduction Compressed gas cylinders can present a variety of hazards due to their pressure and/or contents. One of the contributing factor to these hazards is gas leak due to malfunction

More information

Portable Gas Monitor GX User Maintenance Manual (H4-0050)

Portable Gas Monitor GX User Maintenance Manual (H4-0050) H4E-0050 Portable Gas Monitor GX-8000 User Maintenance Manual (H4-0050) Need of Maintenance and Servicing This gas monitor must be maintained in a normal state at all times to prevent accidents due to

More information

! Warning, refer to accompanying documents.

! Warning, refer to accompanying documents. About this Manual To the best of our knowledge and at the time written, the information contained in this document is technically correct and the procedures accurate and adequate to operate this instrument

More information

Operation Manual. O2 Quickstick. Oxygen Analyzer 08.17

Operation Manual. O2 Quickstick. Oxygen Analyzer 08.17 Operation Manual O2 Quickstick Oxygen Analyzer 08.17 If you have any questions on this equipment please contact Technical Support at: Nuvair 1600 Beacon Place Oxnard, CA 93033 Phone: 805-815-4044 FAX:

More information

Model B Wire Wet Oxidant Gas Transmitter. 6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park

Model B Wire Wet Oxidant Gas Transmitter. 6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park Model B12-69 2-Wire Wet Oxidant Gas Transmitter Home Office European Office Analytical Technology, Inc. ATI (UK) Limited 6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park Collegeville, PA 19426 Delph

More information

PERSONAL AIR BREATHING UNIT

PERSONAL AIR BREATHING UNIT PERSONAL AIR BREATHING UNIT Operation & Maintenance Manual MARTECH SERVICES C O M P A N Y 1-800-831-1525 Table of Contents INTRODUCTION: Page 2 COMPONENTS DRAWING: Page 3 START UP: Page 4 OPERATION: Page

More information

Instruction manual. for. high purity gas pigtails 200 bar / 300 bar

Instruction manual. for. high purity gas pigtails 200 bar / 300 bar Instruction manual for high purity gas pigtails 200 bar / 300 bar Contents Contents... 2 1. Preface... 3 1.1 Overview... 3 1.2 General... 3 1.3 Intended use... 4 1.4 Personnel requirements... 4 2. For

More information

PRECISION GAS MIXTURES

PRECISION GAS MIXTURES PRECISION GAS MIXTURES 2. Company Profile 3. Reactive Gases 4. Regulators 5. Cylinders 6. Gas Mixtures 7. Accessories PRECISION GAS MIXTURES GASCO Affiliates, LLC is a leading manufacturer of Precision

More information

GAS DETECTION: CALIBRATION GAS

GAS DETECTION: CALIBRATION GAS ENVIRONMENTAL SURVEY/PURGING MONITORS GAS DETECTION: CALIBRATION GAS Gas Mixtures Non-Reactive Single & Two-Gas Mixtures 98 Reactive Single & Two-Gas Mixtures 98 Non-Reactive Multi Mixtures 99 Reactive

More information

Gas-Detection Instruments

Gas-Detection Instruments ALTAIR 5X Multigas Detector Accessories SPARE PARTS/ REPLACEMENTS CALIBRATION 10114835 Battery pack assembly, rechargeable 10114837 Battery pack, alkaline (includes belt clip) 10114839 Battery pack, rechargeable,

More information

RK LEL Sample Draw Aspirator Adapter Operator s Manual

RK LEL Sample Draw Aspirator Adapter Operator s Manual 30-0951RK LEL Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0017RK Revision: A Released: 6/2/10 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment

More information

SSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR

SSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR S S F U SSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR MACHINERY DIVISION OWNER S MANUAL UNIT INSTRUCTIONS Please follow all SSFU Safety Instructions. Contact your Duro Dyne Tech Service if you have

More information

Operation Manual. O2 Quickstick. Oxygen Analyzer

Operation Manual. O2 Quickstick. Oxygen Analyzer Operation Manual O2 Quickstick Oxygen Analyzer 01.11 www.nuvair.com If you have any questions on this equipment please contact Technical Support at: Nuvair 2949 West 5 th St. Oxnard, CA 93030 Phone: 805-815-4044

More information

INSTRUCTION MANUAL GAS TRAK

INSTRUCTION MANUAL GAS TRAK INSTRUCTION MANUAL GAS TRAK 166 Keystone Drive Montgomeryville, PA 18936 Telephone: 215-641-2700 Fax: 215-641-2714 INT-0234 rev A1 TABLE OF CONTENTS SERVICE 3 LIMITED WARRANTY 3-4 USER RESPONSIBILITY 4-5

More information

CALIBRATION KIT. Instruction Manual

CALIBRATION KIT. Instruction Manual CALIBRATION KIT Instruction Manual 2 CALIBRATION KIT CALIBRATION KIT 3 Table of Contents General description....4 Getting Started....4 GG-NH3....5 GG-NH3-2%....6 GG-CO....7 GG-CO2...8 GG-CL2...9 GG-H2-EC....

More information

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor M A R T E C H S E R V I C E S C O M P A N Y P.O. Box 7079 OFFICE: (507)

More information

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES THIS BOOKLET CONTAINS PROPRIETARY INFORMATION OF ADVANCED SPECIALTY GAS EQUIPMENT CORP. AND IS PROVIDED TO THE PURCHASER SOLELY FOR USE

More information

AND OPERATION INSTRUCTIONS

AND OPERATION INSTRUCTIONS ADI 2111-D Certified ISO 9001 MEDICAL FLOWMETER (Back Pressure Compensated Thorpe Tube) INSTALLATION AND OPERATION INSTRUCTIONS Before Installing or Operating, Read and Comply with These Instructions Controls

More information

GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL

GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL Operating and Installation Manual Warning Use this product only in the manner described in this manual. If the equipment is used in a

More information

Carnegie Mellon University. Toxic and Highly Toxic Gas Handling Program

Carnegie Mellon University. Toxic and Highly Toxic Gas Handling Program Carnegie Mellon University Toxic and Highly Toxic Gas Handling Program Toxic and Highly Toxic Gas Handling Program Scope and Background The Carnegie Mellon University Toxic and Highly Toxic Gas Handling

More information

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved

More information

TactAir PCP Fill Kit/Adapter. Operation Manual

TactAir PCP Fill Kit/Adapter. Operation Manual TactAir PCP Fill Kit/Adapter Operation Manual INTRODUCTION Thank you for choosing a TactAir PCP fill accessory! Always exercise caution when operating this item. It will allow you to fill your PCP airguns

More information

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante810 Benchtop Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante810 benchtop dissolved oxygen

More information

INSTRUCTIONS PARTS LIST

INSTRUCTIONS PARTS LIST 3066V-E INSTRUCTIONS PARTS LIST INSTRUCTIONS! This manual contains important warnings and information. READ AND KEEP FOR REFERENCE. AQ Poppet Color Valve (3066-41-00A) 304 Stainless Steel Body Autoquip,

More information

DFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER)

DFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER) DFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER) 0-3, 0-15, 0-30 & 0-70 LPM ASSEMBLY INSTRUCTIONS & INSTRUCTIONS FOR USE R219P87-400 (0-70 LPM) R219P88-400 (0-30 LPM) R219P79-400 (0-15 LPM) R219P99-400

More information

Operation and Maintenance Manual

Operation and Maintenance Manual Operation and Maintenance Manual GDS-50 Remote Infrared Sensor Transmitter for Combustibles & CO 2 AUTHORIZED DISTRIBUTOR: GasDetectorsUSA.com - Houston, Texas USA sales@gasdetectorsusa.com - 832-615-3588

More information

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP LIMITED WARRANTY Getz Equipment Innovators warrants its products, and component parts of any product manufactured by

More information

SDS -Series 4C-D201/-D202 Supplemental Drying System User s Manual

SDS -Series 4C-D201/-D202 Supplemental Drying System User s Manual SDS -Series 4C-D201/-D202 Supplemental Drying System User s Manual 8 Executive Drive P.O. Box 2105 Toms River, NJ 08754 (732) 244-0010 (800) 337-3762 fax (732) 244-8140 www.permapure.com info@permapure.com

More information

Handi+N2 OPERATING MANUAL & INSTRUCTIONS FOR USE. R218M06 Industrial. R218M06 Rev. E

Handi+N2 OPERATING MANUAL & INSTRUCTIONS FOR USE. R218M06 Industrial. R218M06 Rev. E E N2 Handi+N2 OPERATING MANUAL & INSTRUCTIONS FOR USE R218M06 Industrial R218M06 Rev. E Maxtec TEL (801) 748. 5355 2305 S 1070 W FAX (801) 270. 5590 Salt Lake City, Utah 84119 www.maxtec.com USA CLASSIFICATION

More information

E X R A D I N Spherical Ion Chambers A3. A4 REF DOC #

E X R A D I N Spherical Ion Chambers A3. A4 REF DOC # E X R A D I N Spherical Ion Chambers A3. A4 REF 92717. 92715 R DOC #80360-00 E X R A D I N Spherical Ion Chambers A3. A4 REF 92717. 92715 STANDARD IMAGING INC. 7601 Murphy Drive Middleton, WI 53562 TEL

More information

MODEL GT820 OXYGEN SENSOR

MODEL GT820 OXYGEN SENSOR INSTRUCTION MANUAL MODEL GT820 OXYGEN SENSOR 70046 The information and technical data disclosed by this document may be used and disseminated only for the purposes and to the extent specifically authorized

More information

Applications Note: Use of "pentane equivalent" calibration gas mixtures

Applications Note: Use of pentane equivalent calibration gas mixtures Introduction The gas that is used to verify accuracy is every bit as important as the detector itself when it comes to worker safety. Choosing (and using) the right mixture is critical to the success of

More information

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Bante820 Portable Dissolved Oxygen Meter Instruction Manual Bante820 Portable Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante820 Portable Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante820 portable dissolved oxygen

More information

Advanced Calibration Designs, Inc. goacd.com. Certified Calibration Gas Solutions

Advanced Calibration Designs, Inc. goacd.com. Certified Calibration Gas Solutions Advanced Calibration Designs, Inc. goacd.com Certified Calibration Gas Solutions Table of Contents We Provide The Ultimate Calibration Gas Solution... 4 How to Buy... 4 CAL2000 Instrument & Accessories...5

More information

INSTRUCTIONS FOR MODEL SG6500 GAS HEATERS

INSTRUCTIONS FOR MODEL SG6500 GAS HEATERS INSTRUCTIONS FOR MODEL SG6500 GAS HEATERS THIS BOOKLET CONTAINS PROPRIETARY INFORMATION OF ADVANCED SPECIALTY GAS EQUIPMENT CORP. AND IS PROVIDED TO THE PURCHASER SOLELY FOR USE IN CONJUNCTION WITH MODEL

More information

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. U S E R M A N U A L 1600 SERIES OXYGEN REGULATOR 168715G (Shown) SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090

More information

SPECIFICATION. date of manufacture. Intermittent: -20 o C to +50 o C. Pressure Range:

SPECIFICATION. date of manufacture. Intermittent: -20 o C to +50 o C. Pressure Range: The ECO-Sure X is a long life, high quality, cost effective 2-electrode electromechanical cell designed for the detection of Carbon Monoxide CO in a range of applications but particularly for domestic/residential

More information

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician. USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician. RECEIVING / INSPECTION Remove the Precision Medical,

More information

Instruction Manual LIMITED 1 YEAR WARRANTY. Hydraulic Punch Driver Read this material before using this product.

Instruction Manual LIMITED 1 YEAR WARRANTY. Hydraulic Punch Driver Read this material before using this product. Instruction Manual Hydraulic Punch Driver 902-483 LIMITED 1 YEAR WARRANTY We make every effort to assure that its products meet high quality and durability standards, and warrant to the original purchaser

More information

Installation, Operating & Maintenance Instructions. Variable leak valve with manual actuator. Series 590 DN 16 mm (I. D. ⅝") E

Installation, Operating & Maintenance Instructions. Variable leak valve with manual actuator. Series 590 DN 16 mm (I. D. ⅝) E Installation, Operating & Maintenance Instructions Variable leak valve with manual actuator DN 16 mm (I. D. ⅝") This manual is valid for the following product ordering numbers: 59024-. E01 -... Sample

More information

Operating Instructions

Operating Instructions Operating Instructions Sensors Type 705 H 2 S Sensor and Transmitter for Electrochemical Cells Ul Certified HELPING TO MAKE A SAFER WORLD Ensure that you read and understand these operating instructions

More information

INSTALLATION MANUAL Matheson Tri-Gas Cabinet Enclosures

INSTALLATION MANUAL Matheson Tri-Gas Cabinet Enclosures INSTALLATION MANUAL Matheson Tri-Gas Cabinet Enclosures MINT-0289-XX TABLE OF CONTENTS Limited Warranty... 3 User Responsibility... 3-4 General Service... 4 Safety Precautions.... 5 Physical Dimensions..

More information

Instrument Operating Manual

Instrument Operating Manual Instrument Operating Manual INDEX NO. SUBJECT PAGE 1 INFORMATION 3 2 SAFETY 4 3 INTENDED USE 4 4 COMMISSIONING & OPERATION 5 5 MAINTENANCE 6 2 INFORMATION The pressure gauges described in the operating

More information

Operating Manual. Models AQT-02 & AQT-COO2 Manual No. MON004 (Rev 2 April 2000)

Operating Manual. Models AQT-02 & AQT-COO2 Manual No. MON004 (Rev 2 April 2000) Models AQT-02 & AQT-COO2 Manual No. MON004 (Rev 2 April 2000) Operating Manual AIR SYSTEMS INTERNATIONAL, INC. 829 Juniper Crescent, Chesapeake, Va., 23320 Telephone (757) 424-3967 Toll Free 1-800-866-8100

More information

Operating Instruction Manual for Liquid Nitrogen Containers. CRY/CRYC Series. Hagavish st. Israel Tel: , Fax:

Operating Instruction Manual for Liquid Nitrogen Containers. CRY/CRYC Series. Hagavish st. Israel Tel: , Fax: Operating Instruction Manual for Liquid Nitrogen Containers CRY/CRYC Series Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC.12.14 TO USER Thank You! You have just purchased

More information

Instrumental Technique GAS CYLINDER

Instrumental Technique GAS CYLINDER Instrumental Technique GAS CYLINDER Amitava Srimany 05-10-2013 Gas cylinder is a pressure vessel used to store gas above atmospheric pressure. Used in: Diving Mining operations Oil refineries Petrochemical

More information

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS Using state-of-the-art measuring and microprocessor technologies, Advanced has assembled a series of systems which can measure mass

More information

CLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1

CLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1 ANA 74300 Oxygen Regulator with Yoke Connector User Manual Manual No. ANA 74300 001 Issue 1 Safety Thank you for purchasing this Clements Oxygen Regulator. For your safety it is imperative that this unit

More information

Handi+ OPERATING MANUAL & INSTRUCTIONS FOR USE. R218P15 Industrial. R218M15 Rev. D

Handi+ OPERATING MANUAL & INSTRUCTIONS FOR USE. R218P15 Industrial. R218M15 Rev. D Handi+ OPERATING MANUAL & INSTRUCTIONS FOR USE R218P15 Industrial R218M15 Rev. D Maxtec TEL (800) 748.5355 2305 S 1070 W FAX (801) 270.5590 Salt Lake City, Utah 84119 www.maxtec.com USA CLASSIFICATION

More information

GG-VL2-NH3 AMMONIA VENT LINE SENSOR. Installation and Operation Manual

GG-VL2-NH3 AMMONIA VENT LINE SENSOR. Installation and Operation Manual GG-VL2-NH3 AMMONIA VENT LINE SENSOR Installation and Operation Manual 2 GG-VL2-NH3 Warning Use this product only in the manner described in this manual. If the equipment is used in a manner not specified

More information

SAFETY DATA SHEET Isopropyl Alcohol 70% Product Number: 32

SAFETY DATA SHEET Isopropyl Alcohol 70% Product Number: 32 SAFETY DATA SHEET Emergency phone: 800 424 9300 (Chemtrec) NFPA Rating: Health 2, Flammability 3, Reactivity 2 Special 0 HMIS Rating: Health 2, Flammability 3, Physical 2, Reactivity, 0 1. Product Identity

More information

Technical Manual. Sensepoint XCD Gas Detector

Technical Manual. Sensepoint XCD Gas Detector Technical Manual Sensepoint XCD Gas Detector 19 Cross Interference and Cross Calibration 19.1 Cross Interference Table for Toxic and Oxygen This table shows the relative cross sensitivity of the Sensepoint

More information

GETZ EQUIPMENT INNOVATORS PART No. 4G59678 CARBON DIOXIDE MINI PUMP FOR EXTINGUISHER SERVICE (Revised )

GETZ EQUIPMENT INNOVATORS PART No. 4G59678 CARBON DIOXIDE MINI PUMP FOR EXTINGUISHER SERVICE (Revised ) GETZ EQUIPMENT INNOVATORS PART No. 4G59678 CARBON DIOXIDE MINI PUMP FOR EXTINGUISHER SERVICE (Revised 11-24-10) 2320 Lakecrest Drive, Pekin IL 61554 PH. (888) 747-4389 Fax (309) 495-0625 Website: www.getzequipment.com

More information

92831 TEL: (714) FAX:

92831 TEL: (714) FAX: Document N0. 1800-03 Copyright 2010 Terra Universal Inc. All rights reserved. Revised Sept. 2010 Terra Universal, Inc. TerraUniversal.com 800 S. Raymond Ave. Fullerton, CA 92831 TEL: (714) 578-6000 FAX:

More information

USER S INFORMATION MANUAL

USER S INFORMATION MANUAL USER S INFORMATION MANUAL UPFLOW, DOWNFLOW, UPFLOW/HORIZONTAL & HORIZONTAL ONLY INDUCED DRAFT GAS FURNACES Recognize this symbol as an indication of Important Safety Information If the information in this

More information

MATERIAL SAFETY DATA SHEET

MATERIAL SAFETY DATA SHEET C.S.S Date Revised: 07-14-2004 SECTION 1 MATERIAL AND MANUFACTURER IDENTIFICATION Product Identifier: Product Use: CSS Color stabilizer and souring agent. MANUFACTURER CTI (Chemical Technologies International

More information

SPECIFICATION. date of manufacture. Intermittent: -20 o C to +50 o C. Pressure Range:

SPECIFICATION. date of manufacture. Intermittent: -20 o C to +50 o C. Pressure Range: The ECO-Sure (2e) is a high quality, cost effective 2-electrode electromechanical cell designed for the detection of Carbon Monoxide CO in a range of applications but particularly for domestic/residential

More information

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator. PRS(TC)4,8 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to

More information

Compressed Gases Safety Program

Compressed Gases Safety Program Compressed Gases Safety Program Summary: Compressed gas cylinders can present a variety of hazards due to their pressure and /or contents. This program covers requirements which must be followed for the

More information

Material Safety Data Sheet

Material Safety Data Sheet 1 Section 1 1145 Sutton Street Howell, MI 48843 Telephone Numbers: (517) 295-4196 Product Information () Transportation Emergency Chemtrec: (800) 424-9300 MATERIAL IDENTIFICATION PRODUCT NAME: LCA PEI

More information

Installation, Operation & Maintenance Manual

Installation, Operation & Maintenance Manual Original Instructions Installation, Operation & Maintenance Manual Sentry Cobra FV Sampler Low Emission Samplers S-MS-IOM-00543-0 08-18 Table of Contents Safety Information... 3 General Safety Precautions...

More information

Operating Manual. Check and Go Disposable Medical Depot, Inc Port Washington N.Y

Operating Manual. Check and Go Disposable Medical Depot, Inc Port Washington N.Y Check and Go Disposable Operating Manual 2005 Medical Depot, Inc Port Washington N.Y. 11050 visit us at www. drivemedical.com Check and Go Disposable Part# 18580 2005 Medical Depot, Inc Port Washington

More information

Unisense In Situ UniAmp Connector System

Unisense In Situ UniAmp Connector System Unisense In Situ UniAmp Connector System Overview The Unisense In Situ UniAmp Connector system is designed for easy replacement of sensors. Unisense In Situ UniAmp Connector system is a dedicated system

More information

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual OxyCheq Expedition-X Oxygen Analyzer Operator s Manual OxyCheq 3528 Russell Road Marianna, Florida 32446 USA Tel: 850.482.0385 Email: oxy@oxycheq.com Web: http://oxycheq.com CONTENTS 1.0 Introduction..

More information

INSTRUCTIONS FOR SINGLE STAGE LECTURE BOTTLE REGULATORS

INSTRUCTIONS FOR SINGLE STAGE LECTURE BOTTLE REGULATORS INSTRUCTIONS FOR SINGLE STAGE LECTURE BOTTLE REGULATORS THIS BOOKLET CONTAINS PROPRIETARY INFORMATION OF ADVANCED SPECIALTY GAS EQUIPMENT CORP. AND IS PROVIDED TO THE PURCHASER SOLELY FOR USE IN CONJUNCTION

More information

GILMONT ACCUCAL FLOWMETERS

GILMONT ACCUCAL FLOWMETERS OPERATING MANUAL GILMONT ACCUCAL FLOWMETERS 28W092 Commercial Ave. Barrington, IL U.S.A. 60010-2392 (847) 381-4888 (847) 381-7053 (Fax) 800-962-7142 www.barnant.com e-mail: barnant@barnant.com A-1299-0766

More information

High Pressure Inlet Kits for CONCOA BlendMites and BlendMasters

High Pressure Inlet Kits for CONCOA BlendMites and BlendMasters ADI 3210-A ADI 3210-A Certified ISO 9001 Certified ISO 9001 High Pressure Inlet Kits for CONCOA BlendMites and BlendMasters Warning: An appropriately sized pressure relief device downstream of the regulator

More information

Pressure Relief Valve Instruction Manual

Pressure Relief Valve Instruction Manual CVR3-M0_062017 Pressure Relief Valve Instruction Manual MODEL: CVR3 SFA Companies 10939 N. Pomona Ave. Kansas City, MO 64153 Tel: 888-332-6419 * Fax: 816-448-2142 E-mail: sales@bvahydraulics.com Website:

More information

OP08 - Electrochemical 3-Electrode Un-biased Toxic Sensors CONTENTS. Introduction 2. Operating Principles 2. Recommended Circuitry 4.

OP08 - Electrochemical 3-Electrode Un-biased Toxic Sensors CONTENTS. Introduction 2. Operating Principles 2. Recommended Circuitry 4. OP08 - Electrochemical 3-Electrode CONTENTS Introduction 2 2 Recommended Circuitry 4 Start-up Time 5 Calibration 5 Selectivity and Filters 6 Safety Monitoring 6 Continuous Exposure to Target Gas 7 Effects

More information

Yoke Block Instruction Manual

Yoke Block Instruction Manual Yoke Block Instruction Manual ! WARNING IMPORTANT: READ MANUAL COMPLETELY BEFORE OPERATING THIS DEVICE This manual contains instructions on periodically required checks to be performed by the user. These

More information

Respiratory Protection Self-Inspection Checklist

Respiratory Protection Self-Inspection Checklist Respiratory Protection Self-Inspection Checklist Guidelines: This checklist applies to the use of either atmosphere-supplying or air-purifying respirators being worn, voluntarily or otherwise, for comfort

More information

Atmospheric Hazards. Hazardous Atmosphere 3/29/2017. Oxygen deficiency and enrichment Flammable gases and vapors Toxic contaminants

Atmospheric Hazards. Hazardous Atmosphere 3/29/2017. Oxygen deficiency and enrichment Flammable gases and vapors Toxic contaminants 2017 Wisconsin Safety Council Safety & Health Conference, Presentation by Tim Gackle, OSHA Training Institute April 11, 2017 Atmospheric Hazards Oxygen deficiency and enrichment Flammable gases and vapors

More information

Yoke Block Instruction Manual

Yoke Block Instruction Manual Yoke Block Instruction Manual ! WARNING IMPORTANT: READ MANUAL COMPLETELY BEFORE OPERATING THIS DEVICE This manual contains instructions on periodically required checks to be performed by the user. These

More information

MATERIAL SAFETY DATA SHEET

MATERIAL SAFETY DATA SHEET Natural Fiber Cleaner Liquid Date Revised: 07-09-2004 SECTION 1 MATERIAL AND MANUFACTURER IDENTIFICATION Product Identifier: Natural Fiber cleaner liquid MANUFACTURER CTI (Chemical Technologies International

More information

ROPV R40 E Series User Manual

ROPV R40 E Series User Manual HARBIN ROPV INDUSTRY DEVELOPMENT CENTER ROPV R40 E Series User Manual For Use with the Following ROPV Pressure Vessel Models: R40 300E R40 450E Headquarters Tel:(+86)451-82267301 Fax:(+86)451-82267303

More information

User s Guide Temperature Sensor Converter TSC-599

User s Guide Temperature Sensor Converter TSC-599 User s Guide Temperature Sensor Converter TSC-599 ILX Lightwave Corporation 31950 Frontage Road Bozeman, MT, U.S.A. 59715 U.S. & Canada: 1-800-459-9459 International Inquiries: 406-556-2481 Fax 406-586-9405

More information

INSTRUCTIONS FOR USE

INSTRUCTIONS FOR USE Rebar Chain Assembly INSTRUCTIONS FOR USE 7260XX Rebar Chain Assembly Complies with the current ANSI Z359.1-2007 and all applicable OSHA regulations and requirements. Reliance Industries P.O. Box 2046

More information

SEC Signature DIR Process Monitor

SEC Signature DIR Process Monitor SEC Signature DIR Process Monitor Instruction and Operation Manual Sensor Electronics Corporation 12730 Creek View Avenue Savage, Minnesota 55378 USA (952) 938-9486 Fax (952) 938-9617 email sales@sensorelectronics.com

More information