ZIRCONIA OXYGEN DETECTOR SERVICE MANUAL
|
|
- Jewel Stone
- 5 years ago
- Views:
Transcription
1 Service Manual ZIRCONIA OXYGEN DETECTOR SERVICE MANUAL TYPE: ZFK 3, 4, 7 TN5A08a-E
2 CONTENTS. PREFACE.... MEASUREMENT PRINCIPLE ADJUSTMENT Zero (Air) Adjustment Span Adjustment TROUBLE CHECK SHEET REPAIR...5 APPENDIX Wiring Diagram...6 Output characteristic table
3 . PREFACE This service manual explains repair and adjustment procedure of the Zirconia oxygen detector (ZFK 3, 4, 7) as a single unit. This manual is intended for use with the instruction manual (INZ-TN4ZFK3).. MEASUREMENT PRINCIPLE When yttria-stabilized zirconia ceramic is heated to hightemperature, only the oxygen ion becomes a movable solid electrolyte. In a zirconia element, the both sides of which are fitted with a platinum electrode, the electrode reactions in the following expressions occur for the oxygen concentrations P and P (P > P ). Electrode of P side O + 4e 0 (negative electrode) Electrode of P side 0 O + 4e (positive electrode) Filter P Measured gas Heater Electrode Insulating material Casing Zirconia element Fig. Detector structure Air P Output That is, the oxygen ion moves from the P side, where oxygen concentration is higher, to P side, where it is lower. This energy for moving the ion generates the electric force (E) indicated in the following expression. RT P E= n (O) (V) 4F P (O) P (O) =50.74 log (mv) (at 800 C) P (O ) E : Electric force P (O ) : Comparative (Air) oxygen concentration P (O ) : Oxygen concentration in the measured gas R : Gas constant (J mol - K - ) T : Absolute temperature F : Faraday constant (c mol - ) Coefficient is the value when the zircornia element is 750 C. (The sensor unit becomes 750 C with 800 C temperature control.) Therefore, unknown oxygen concentration P (O ) can be calculated from the electric force (E). --
4 3. ADJUSTMENT 3. Zero (Air) Adjustment ) Remove the 3 screws fixing the case cover. ) Feed the zero gas (Dry Air) through the sample gas inlet (on the lower side of the main unit) at the specified flow rate (0.5L /min). 3) After the voltage indication between the external terminals block (4) and (5) (the terminal block on the front panel 7P) is stabilized (feed zero gas for to 3 minutes), adjust the indication to 0 ± mv by VRZ on the printed board; ZFK 3, 4. For ZFK7, make sure that the indication is within ± 3mV. (No adjustment) 3. Span Adjustment ) Feed the span gas ( to % O /N ) at the specified flow rate. ) After the indication is stabilized, adjust it by VRS; ZFK 3, 4 Find the adjustment value using the following expression. E (Output:V) = 0.38 log /Px Px:Calibration gas concentration Example) In case of.% O /N E=0.38 log /. = v Make an adjustment with the VRs control so that V is obtained. Span VR VRS Zero VR VRZ 8 C4 0 R4 3 6 R C CN3 JAPAN VRS R7 R3 CN R R8 CN C3 R5 VRZ ZD ZD ZD4 ZD3 C TK4B569R SSR Fig. Printed board mounting diagram; ZFK 3, 4 For ZFK7, make sure that the indication is within 50 to 70mV when the gas with concentration of to % O /N is flowed. (Refer to Appendix Table.) (No adjustment) -3-
5 4. TROUBLE CHECK SHEET Phenomena Cause Checking methods (normal value) Protective measures Slow response Sensor element Change over between zero and span Replace deterioration gas and check if 60 seconds or longer Sample gas is not supplied or output is abnormal. Sensor outlet is clogged with crystallized sulfate mist. Output is unstable ZFK 3, 4 Disconnection of connector or poor contact Output is not deflected or remains the same. Output swings over. is needed for 90% response. Visually check for clogging of the outlet of the gas introduction case. Check if the connector is disconnected or move the connector part to check for the indication change. Remove the clogging. Replace with a gas introduction case (provided with gas outlet pot). Repair Disturbance noise Check if output is hunting. Eliminate noise. Sensor element Flow zero/span gas continuously for Replace the sensor deterioration 0 minutes or so, and observe the element. Fuse is blown out or power is not supplied. ZFK 3, 4 Disconnection of connector or poor contact Sensor is defective. Solid relay is defective. Printed board is defective. ZFK 3, 4 Constant voltage power supply is defective. Temperature controller is defective. Measuring range is exceeded. indication status. Check the fuse and the supply voltage specification. Check if the connector is disconnected or move the connector part to check for the indication change. Change over between zero and span gas and check for the indication change. Remove one end of the sensor terminal block (3) (4), and check for continuity. The range should be within to 3 Ω. When 4V DC is applied to the outer end (7) (8) (-), check if AC power between () () on the sensor element is turned on or off. (Check with a tester.) Change over between zero and span gas and check if the voltage value between the sensor outer end (5) (6) (+) indicates as shown below. Air: ± 3mV or lower to % O /N : 50 to 70mV. Check if the CN of the constant voltage power supply indicates ± 0.V. Check if the temperature indication is constant in a range of 800 ± 5 C. Flow measured gas after zero/span calibration. Replace the fuse or check the power supply. Repair Replace the sensor element. Replace the printed board or the solid relay. Replace the printed board or the solid relay or sensor element. Replace the constant voltage power supply. Replace the temperature controller. Review the specifications. -4-
6 5. REPAIR () Replacement of the sensor element ) Change the measuring gas line to the Air atmosphere and then turn off the power switch. When the device is operating, install an oxygen meter by bypass piping. Leave the device for about 30 minutes till it cools down. ) Remove the case cover (a). 3) Remove the 6 wires connected to the sensor terminal block (b) and the 3 M4 screws fixing the sensor (c). 4) Pull out the sensor (d) from the sensor case (e), and assemble the new sensor in reverse order of disassembly. After wiring is checked, perform energization and running to calibrate zero/span. When replacing the sensor element, also replace the packing with a new one. () Replacement of the sensor case ), ) Follow the procedure in Item (). 3) Remove the 3 M4 screws fixing the sensor (c). 4) Remove the screws fixing the sensor case from the bottom of the main unit, and remove the sensor case from the top. 5) Assemble a new sensor in reverse order of disassembly. (3) Replacement of the printed board (ZFK 3, 4) ), ) Follow the procedure in Item (). 3) Remove the 8 wires connected to the terminal block of the printed board and 3 screws fixing the printed board. 4) Assemble a new printed board in reverse order of disassembly. Perform energization and running to calibrate zero/span. (4) Replacement of the temperature controller ), ) Follow the procedure in Item (). 3) Remove the screws from the top of the temperature controller and the socket from the backside, and pull out the controller forward. 4) Assemble a new temperature controller in reverse order of disassembly. (5) Replacement of the constant voltage power supply (ZFK 3, 4) ), ) Follow the procedure in Item (). 3) Remove the 5 wires and the side fixing screws connected to the upper terminal block of the constant voltage power supply (g), and remove the constant voltage power supply. 4) Assemble a new constant voltage power supply in reverse order of disassembly. (a) Case cover (e) Sensor case (c) Screw (d) Sensor (b) Wiring (h) Printed board: ZFK 3, 4 Solid relay; ZFK 7. (f) Temperature controller (g) Constant voltage power supply: ZFK 3, 4 Not mounted on ZFK 7. Fig.3 Internal diagram of the oxygen analyzer -5-
7 R T ZFK3,4 APPENDIX Wiring Diagram SW Power supply F. 3A 85 to 65V AC Temperature controller 0 PXR to 65V AC 5 4 Approx. 8mV/800 C Sensor mv 85 to 65V AC Constant voltage power supply AC AC V 0 DC POWER SUPPLY IN DCV OUT SSR Heat R OUT IN ZFK CN CN3 Amp. P.C.B O Sensor 00 to 5V AC or 00 to 0V AC R T E Main power Temperature supply alarm Output 0 0 C Output Temperature alarm When the setting value (800 C) is exceeded by ±0 C R T ZFK7 SW *00 to 40V AC for CE compatible types Power supply F. 3A 85 to 65V AC Temperature controller 0 PXR SSR O Sensor 00 to 5V AC or 00 to 0V AC Heat R OUT ZFK 0 0 C Temperature alarm C Output -6-
8 APPENDIX TABLE ZFK 3, 4 Zirconia oxygen Analyzer Output characteristic table (ZFK 3, 4 output) %O OUTPUT (mv) %O OUTPUT (mv) E (OUTPUT : V) = 0.38 log /O % APPENDIX TABLE ZFK 3, 4, 7 Zirconia oxygen Analyzer Detector Output characteristic (internal sensor) %O OUTPUT (mv) %O OUTPUT (mv) E (OUTPUT : mv) = log /O % -7-
9
INFRARED GAS ANALYZER
Service Manual INFRARED GAS ANALYZER TYPE: ZRJ-5 INZ-TN513433a-E PREFACE This service manual describes the infrared gas analyzer (Type: ZRJ-5). This service manual is intended for use with the instruction
More informationZIRCONIA OXYGEN ANALYZER CONVERTER
Instruction Manual ZIRCONIA OXYGEN ANALYZER CONVERTER TYPE: ZRY OXYGEN ANALYZER Vol% O2 PREFACE We are grateful for your purchase of Fuji Zirconia Oxygen Analyzer Converter (ZRY). First read this instruction
More informationZIRCONIA OXYGEN ANALYZER CONVERTER
Instruction Manual ZIRCONIA OXYGEN ANALYZER CONVERTER (SINGLE-CHANNEL TYPE) TYPE: ZRM Fuji Electric Co.,Ltd. INZ-TN1ZRMa-E PREFACE We are grateful for your purchase of Fuji Electric s Zirconia Oxygen Analyzer
More informationColumbus Instruments
0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011
More informationSpecifications of Oxygen Sensor Module V-8344A
Specifications of Oxygen Sensor Module V-8344A 1. General: This document describes about the specification of Ceramic Oxygen Sensor module, FUJIKURA's FCX-MWL-F-(AC). Table shown below is revision records
More informationStand-Alone Bubble Detection System
Instruction Sheet P/N Stand-Alone Bubble Detection System 1. Introduction The Bubble Detection system is designed to detect air-bubble induced gaps in a bead of material as it is being dispensed. When
More informationOWNER S MANUAL NMB SERIES
OWNER S MANUAL NMB SERIES NMB Manual V2.0.doc 1. PARTS IDENTIFICATION LIST 1. 5. 2. 4. 3. 6(a). 6(b). Figure 1.1 1. Powerpack with timer option (x1) 2. Cell (x1) 3. Cell Housing (x1) 4. O-ring (x1) 5.
More informationINFRARED GAS ANALYZER
Service Manual INFRARED GAS ANALYZER TYPE: ZPA, ZPB, ZPG INZ-TN5A1191-E PREFACE This service manual describes the infrared gas analyzer (Type: ZPA, ZPB, ZPG). This service manual is intended for use with
More informationTransmitter CS 21 Operation Manual
Transmitter CS 21 Operation Manual 1194 Oak Valley Drive, Suite 20, Ann Arbor, MI 48108 800-959-0573 734-769-1888 Content Page For your Safety 3 General Description 3 Detection Principle 4 Operation 4
More informationIR200 Infrared Gas Analyzer
User's Manual IR200 Infrared Gas Analyzer 3rd Edition PREFACE We are grateful for your purchase of Yokogawa s Infrared Gas Analyzer (Model : IR200). First read this instruction manual carefully until an
More informationModel ZX8D Zirconia Oxygen Analyzer
User s Manual Model ZX8D Zirconia Oxygen Analyzer IM 11G00E01-01EN IM 11G00E01-01EN 6th Edition u Introduction Thank you for purchasing the ZX8D Zirconia Oxygen Analyzer. Please read the following respective
More informationZirconia Oxygen Sensors. Model 1230 Series
Zirconia Oxygen Sensors Model 1230 Series March 2003 USING THIS MANUAL CAUTION 1 The probe or sensor heater is supplied with mains voltage. This supply has electrical shock danger to maintenance personnel.
More informationTransmitter CS 21 Operation Manual
Transmitter CS 21 Operation Manual Content Page For your Safety 3 General Description 3 Detection Principle 4 Operational Notes 4 Design 4 Mounting Position of CS21 5 Mounting 6 Installation of Electrical
More informationMEGAS 2.0. Gas analysis with direct C-Level Calculation and Bus-Connection. Data sheet
Data sheet MEGAS 2.0 Gas analysis with direct C-Level Calculation and Bus-Connection Mesa Industrie-Elektronik GmbH Neckarstraße 19, D-45768 Marl info@mesa-gmbh.de +49 (0) 2365-97 45 1-0 +49 (0) 2365-97
More informationUSER MANUAL. JET TST and JET TST S.
USER MANUAL JET TST and JET TST S www.sisteven.com Types of Jet fans: JET TST UNI: Impulse axial fan single direction JET TST S UNI: Impulse axial fan single direction (short case) JET TST REV: Impulse
More informationACV-10 Automatic Control Valve
ACV-10 Automatic Control Valve Installation, Operation & Maintenance General: The Archer Instruments ACV-10 is a precision automatic feed rate control valve for use in vacuum systems feeding Chlorine,
More informationBI-680 Online Dissolved Oxygen Controller Instruction Manual
BI-680 Online Dissolved Oxygen Controller Instruction Manual BANTE INSTRUMENTS CO., LTD BI-680 Online Dissolved Oxygen Controller 1 Introduction Thank you for selecting the BI-680 online dissolved oxygen
More informationInstruction and Maintenance Manual
Instruction and Maintenance Manual GRYF OXY Z 02/100/2 E Contact GRYF HB, spol. s r.o. Cechova 314 Havlickuv Brod 580 01 tel.: +420 569 426 627 fax: +420 569 426 627 Czech Republic www.gryf.eu Technical
More informationMODEL 400 CLD INSTRUCTION MANUAL
MODEL 400 CLD INSTRUCTION MANUAL This manual describes installation, calibration and operation of oxides of nitrogen gas analyzer. To assure correct operation and accurate results, it is recommended that
More informationInstruction and Maintenance Manual
Instruction and Maintenance Manual GRYF OXY ZM 02/100/2 E Contact GRYF HB, spol. s r.o. Cechova 314 Havlickuv Brod 580 01 tel.: +420 569 426 627 fax: +420 569 426 627 www.gryf.eu ver.: 11.2.2015 OEM module
More information97C COMPRESSOR KIT 12V PART NO C COMPRESSOR KIT 24V PART NO C COMPRESSOR KIT PART NO
97C COMPRESSOR KIT 12V PART NO. 00097 97C COMPRESSOR KIT 24V PART NO. 02497 98C COMPRESSOR KIT PART NO. 00098 97C 98C IMPORTANT: It is essential that you and any other operator of this product read and
More informationIN-SITU ZIRCONIA OXYGEN ANALYZER < ZIRCOMAT-C >
IN-SITU ZIRCONIA OXYGEN ANALYZER < ZIRCOMAT-C > DATA SHEET ZFK, ZRY This oxygen analyzer is used to continuously measure oxygen concentration in combustion exhaust gas of industrial boilers or furnaces,
More informationCCT-7320/ROC-2313 Reverse Osmosis Controller
CCT-7320/ROC-2313 Reverse Osmosis Controller 1 General The instrument is a combined control instrument of a reverse osmosis controller and an on-line conductivity instrument. It can perform the operation
More informationModel 130M Pneumatic Controller
Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate
More informationMODEL GT820 OXYGEN SENSOR
INSTRUCTION MANUAL MODEL GT820 OXYGEN SENSOR 70046 The information and technical data disclosed by this document may be used and disseminated only for the purposes and to the extent specifically authorized
More informationAlphasense 4-20 ma transmitters offer convenience and easy maintenance for toxic sensors:
ALPHASENSE USER MANUAL Page 1 4-20mA Transmitter for Toxic Sensors UMTOX-1 Issue 4 1 INTRODUCTION The Transmitter PCB includes circuitry for a three electrode toxic sensor to convert the µa output signal
More informationCONDUCTIVITY POINT/S LEVEL SWITCH JAYCEEAQUA 4000 SERIES
I N S T R U C T I O N - M A N U A L CONDUCTIVITY POINT/S LEVEL SWITCH JAYCEEAQUA 4000 SERIES FUNCTION: Conductivity level limit switches are based on the principle of measuring level through the electrical
More informationZIRCONIA OXYGEN ANALYZER CONVERTER
Instruction Manual ZIRCONIA OXYGEN ANALYZER CONVERTER Type: ZKM INZ-TN1ZKMf-E PREFACE We are grateful for your purchase of Fuji Direct Insertion Type Zirconia Oxygen Analyzer Converter (ZKM). First read
More informationZIRON Oxygen Measuring Technique. Oxygen Measuring Device ECO1000. Manual
ZIRON Oxygen Measuring Technique Oxygen Measuring Device ECO1000 Manual Sales and Service Center Shanghai ZIRON Gasanalyzer Co., Ltd 404/6, No.10, Lanxi Road, Shanghai P.R. China Tel. 86-21-6254 6905,
More informationC&G Level 3 Award in the Periodic Inspection, Testing and Certification of Electrical Installations
C&G 2395-01 Level 3 Award in the Periodic Inspection, Testing and Certification of Electrical Installations Test instruments and recommended testing sequence 1 Outcomes of this Session describe the different
More informationManual Oxygen Analyzer A4009T
Manual Oxygen Analyzer A4009T Revision 01-EN DIVIZE B.V.- info@divize.com November 25, 2014 Contents 1 Measuring principle 3 2 Safety 4 3 Installation 5 3.1 Mounting................................................
More informationOxyfire-LT Modular Oxygen Sensor Installation / Operation Manual
Oxyfire-LT Modular Oxygen Sensor Installation / Operation Manual United Process Controls, Inc. Revision 3 January 27, 2015 Copyright United Process Controls, Inc. No part of this document may be stored
More informationDF-550E PROCESS ANALYSERS APPLICATIONS FEATURES
PROCESS ANALYSERS DF-550E The DF-550E Nano Trace oxygen analyser is a Coulometric sensor based analyser designed to measure oxygen as a contaminant at ultra trace levels in ultra high purity electronic
More informationInert Air (N2) Systems Manual
INSTRUCTION MANUAL Inert Air (N2) Systems Manual N2-MANUAL 2.10 READ AND UNDERSTAND THIS MANUAL PRIOR TO OPERATING OR SERVICING THIS PRODUCT. GENERAL INFORMATION Positive pressure nitrogen gas pressurizing
More informationINSTRUCTION MANUAL. FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model Version Perma Pure LLC Tel:
PERMA PURE INSTRUCTION MANUAL FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model 3300 Version 4.06 Perma Pure LLC Tel: 732-244-0010 P.O. Box 2105, 8 Executive Drive Tel: 800-337-3762 (toll free US)
More informationUser's Manual. OX100 General-Purpose Oxygen Analyzer. IM 11M10A01-01E 2nd Edition IM 11M10A01-01E
User's Manual OX100 General-Purpose Oxygen Analyzer 2nd Edition Safety Precautions Safety Precautions This Analyzer is designed for instrument panel or rack mounting. To ensure safe operation of this instrument,
More informationGas Sampling Fitting with MK2.1 and L-Probe
measuring controlling automation measuring controlling automation measuring controlling automation Page i von 8 Operation Manual Gas Sampling Fitting with MK2.1 and L-Probe Page ii von 8 Manufacturer MESA
More informationSensoric 4-20 ma Transmitter Board Operation Manual
Sensoric 4-20 ma Transmitter Board Operation Manual 1 Content Features Operation Manual Technical Data Mechanical Dimensions Remarks & Contact information 2 Features Soldered sensor cell (non replaceable)
More informationVentam 85 Installation & Commissioning Instructions
Ventam Systems Ltd Unit D4 Seedbed Business Centre Vanguard Way Shoeburyness Essex SS3 9QY Phone 01702 382 307 Fax 01702 382 340 Ventam 85 Installation & Commissioning Instructions 1 General The Ventam
More informationOEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE
OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE 1 These instructions generally describe the installation, operation, and maintenance of subject equipment. The manufacturer reserves the right
More informationDPC-30 DPC-100. Reference Manual
DPC-30 DPC-100 Reference Manual 1. Introduction 1.1 Description The Martel DPC Digital Pneumatic Calibrator improves upon traditional dial gauge pneumatic calibrators. The Martel DPC improves accuracy,
More informationInstallation, operating and maintenance Instructions for Seemag bypass level indicator
Issue: S Date: 05-09-14 Type G35 General information The Seetru bypass magnetic level indicator, abbreviate SEEMAG, serves to show the filling level of fluids in tanks, basins, tubes etc. The Seemag operates
More informationHigh-performance submersible pressure transmitter For level measurement Model LH-10
Electronic pressure measurement High-performance submersible pressure transmitter For level measurement Model LH-10 WIKA data sheet PE 81.09 Applications Level measurement in rivers and lakes Deep well
More informationHigh-performance submersible pressure transmitter For level measurement Model LH-10
Electronic pressure measurement High-performance submersible pressure transmitter For level measurement Model LH-10 WIKA data sheet PE 81.09 Applications Level measurement in rivers and lakes Deep well
More information100C Air Compressor Kit
10010 100C Air Compressor (standard mounting bracket, CE Spec) 10014 100C Air Compressor (no leader hose or check valve, CE Spec) 10016 100C Air Compressor (with Omega Bracket, CE Spec) IMPORTANT: It is
More information200 PSI COMPRESSORS - MODEL NUMBERS
200 PSI COMPRESSORS - MODEL NUMBERS 380C AIR COMPRESSOR KIT PART NO. 38033 480C AIR COMPRESSOR KIT PART NO. 48043 380C 480C IMPORTANT: It is essential that you and any other operator of this product read
More informationIMPORTANT SAFETY INSTRUCTIONS
IMPORTANT SAFETY INSTRUCTIONS CAUTION - To reduce risk of electrical shock: - Do not disassemble. Do not attempt repairs or modifications. Refer to qualified service agencies for all service and repairs.
More informationBest Practice Guide, Servomex 2700
For full installations details refer to the. Best Practice Guide, Servomex 2700 Mounting: General Guidelines: Servomex 2700 Control Units and air supplies (utilities units) should, ideally, be mounted
More information420C AIR COMPRESSOR KIT PART NO C AIR COMPRESSOR KIT PART NO
420C AIR COMPRESSOR KIT PART NO. 42042 460C AIR COMPRESSOR KIT PART NO. 46043 420C 460C IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of this
More informationSTEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE
STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE GENERAL INFORMATION This family of dissolved oxygen electrodes is of the steam sterilizable polarographic type, designed to be interchangeable with NBS
More informationRC 195 Receiver-Controller
Document No. 129-082 RC 195 Receiver-Controller Product Description The POWERS RC 195 Receiver-Controller is a pneumatic instrument that receives one, two or three pneumatic inputs. It produces a pneumatic
More informationGround Fault Circuit Interrupter(GFCI) Policy
NewStar Netronics, LLC American Products for the World Ground Fault Circuit Interrupter(GFCI) Policy 3926 East 3 rd Street Tulsa, OK 74112 Phone (918)894-5006 Fax (918)836-9909 tech@newstarnetronics.com
More informationUser's Manual. OX100 General-Purpose Oxygen Analyzer. IM 11M10A01-01E 2nd Edition IM 11M10A01-01E
User's Manual OX100 General-Purpose Oxygen Analyzer 2nd Edition Safety Precautions Safety Precautions This Analyzer is designed for instrument panel or rack mounting. To ensure safe operation of this instrument,
More informationGASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL
GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL Operating and Installation Manual Warning Use this product only in the manner described in this manual. If the equipment is used in a
More informationINFRARED GAS ANALYZER
Instruction Manual INFRARED GAS ANALYZER TYPE: ZSU-5 INZ-TN5ZSUa-E PREFACE We are grateful for your purchase of Fuji Electric s Infrared Gas Analyzer, TYPE: ZSU. First read this instruction manual carefully
More informationO 2 SENSORS Zirconium Dioxide (ZrO 2 ) Oxygen Sensor Operating Principle Guide
O 2 SENSORS Zirconium Dioxide (ZrO 2 ) Oxygen Sensor Operating Principle Guide This document describes the physics and concepts behind CO2Meter s range of dynamic and highly accurate oxygen sensors. When
More informationInoxair. Repair Manual. Content. Edition: 12/2010. Subject to technical changes for reasons of the continous development.
Inoxair Repair Manual Edition: 12/2010 Content 1) Serial number 2) Technical data 3) Functional principle 4) Problem analysis 5) Removal 6) Repairs 7) Circuit diagram and cable-connection-plan 8) Pneumatic
More informationRHPS Series RD(H)F40 User Manual. Read the complete manual before installing and using the regulator.
RHPS Series RD(H)F40 User Manual Read the complete manual before installing and using the regulator. 2 WARNING Before removing a regulator from the system for service, you must depressurize system purge
More information400H HARDMOUNT AIR COMPRESSOR KIT PART NO H HARDMOUNT AIR COMPRESSOR KIT PART NO
400H HARDMOUNT AIR COMPRESSOR KIT PART NO. 40042 450H HARDMOUNT AIR COMPRESSOR KIT PART NO. 45042 400H 450H IMPORTANT: It is essential that you and any other operator of this product read and understand
More informationProcedure Troubleshooting the Interconnect Cables
Procedure 6.1 - Troubleshooting the Interconnect Cables Troubleshooting the upper interconnect cable 1. Remove the display housing per Procedure 7.13. Disconnect the upper interconnect cable from the upper
More informationAMS 2710 PCB pressure sensor module with V output
FEATURES Universal pressure sensor module with 0.. 10 V voltage output Fully calibrated and temperature compensated sensor module Variants for (bidirectional) differential, gage, absolute and barometric
More informationO3 3E 1 F Gas Sensor Module
Product Information Pack O3 3E 1 F Gas Sensor Module (Ozone) CONTENTS Product Data Sheet Product Specification 2 Poisoning and Cross Sensitivities 3 Operating Instructions Introduction 4 Electrostatic
More informationOverview. Front Panel: Keypad and Display
Overview The GA-200B is an analyzer that integrates a gas sampling system with sensors to measure and display the concentrations of oxygen and carbon dioxide in a sample as the percentage of a gas in the
More informationOXYGEN MONITORING SYSTEM OMS
User Manual OXYGEN MONITORING SYSTEM OMS 420 Read and observe before commissioning! Ex Situ oxygen analyzer based on ZrO 2 sensor ENGLISH Caution! Please inspect in the presence of the transporting agent
More informationIR400 Infrared Gas Analyzer
User's Manual IR400 Infrared Gas Analyzer Yokogawa Electric Corporation 3rd Edition PREFACE We are grateful for your purchase of Yokogawa s Infrared Gas Analyzer, Model:IR400. First read this instruction
More informationBante820 Portable Dissolved Oxygen Meter Instruction Manual
Bante820 Portable Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante820 Portable Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante820 portable dissolved oxygen
More informationFlamco Flamco-Fill STA Flamco-Fill STM
STA STM Typ STA 5936 Typ STM 5937 GB Installation and operating instructions 1999, 2 Contents GB UK Limited P.O. Box 9, Washway Lane, St. Helens, Merseyside WA10 6FE United Kingdom Telephone: 01744 744744
More informationABB MEASUREMENT & ANALYTICS DATA SHEET. AK100 Series ATEX compliant gas analyzer system for hydrogen-cooled alternators
ABB MEASUREMENT & ANALYTICS DATA SHEET AK100 Series ATEX compliant gas analyzer system for hydrogen-cooled alternators AK100 S E RIE S ATEX COMP LI AN T GAS ANALY ZE R SYSTE M DS/AK100-E N RE V. O Measurement
More informationThe Ins and Outs of I/P Transducers
The Ins and Outs of I/P Transducers By Mark B. Levine, ControlAir Inc. General description I/P transducers are versatile instruments that use an electrical control signal to proportionally regulate gas
More information444C DUAL PERFORMANCE VALUE PACK
(Chrome) PART NO. 44432 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of this manual before installing and using this product. SAVE THIS MANUAL
More informationOxygen Flue Gas Analyzer BA 2000
Gas Analysis Oxygen Flue Gas Analyzer BA 2000 Some combustion processes, e.g. process heaters, steam boilers or heating furnaces the air required to achieve optimal system efficiency can easily fluctuate.
More informationCrest Heating Boilers FBN(L) Series &
Replacement Parts List PARTS & SERVICE DEPARTMENT Nashville, Tennessee 877-554-5544 Fax: 615-882-2918 parts_team@lochinvar.com www.lochinvar.com Crest Heating Boilers FBN(L) 751-2001 Series 100-101 & 110-111
More information250C-IG COMPRESSOR KIT 12V PART NO C-IG COMPRESSOR KIT 24V PART NO
250C-IG COMPRESSOR KIT 12V PART NO. 25050 250C-IG COMPRESSOR KIT 24V PART NO. 25058 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of this manual
More informationDF-310E PROCESS ANALYSERS APPLICATIONS FEATURES
PROCESS ANALYSERS DF-310E The DF-310E is a Coulometric sensor based oxygen analyzer designed to measure trace and percent level oxygen in pure and multi-gas backgrounds for process and quality control
More information250C-IG COMPRESSOR KIT 12V PART NO C-IG COMPRESSOR KIT 24V PART NO
250C-IG COMPRESSOR KIT 12V PART NO. 25050 250C-IG COMPRESSOR KIT 24V PART NO. 25058 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of this manual
More informationAUTOMATIC GAS MANIFOLDS
AUTOMATIC GAS MANIFOLDS Phoenix Pipeline Products Limited. Unit 8, McKenzie Industrial Park, Tel No.: 44 (0) 161 428 7200 Bird Hall Lane, Fax No.: 44 (0) 161 428 7010 Stockport, Email: info@p3-phoenix.com
More informationAK100 Series ATEX Compliant Gas Analyzer System for Hydrogen-cooled Alternators
Data sheet DS/AK EN Rev. J AK Series ATEX Compliant Gas Analyzer System for Hydrogen-cooled Alternators Superior technology and quality from the world leader in hydrogen measurement ATEX compliant intrinsically
More informationMass Flow Meter (MFM) for gases
Mass Flow Meter (MFM) for gases Type can be combined with Inline MFM for nominal flow rates from 25 l N /min to 1,500 l N /min; 1/4 to 3/4 High accuracy Fast settling time Fieldbus option Special version
More information42045 Heavy Duty ADA Base Model Kit: 85/105 PSI (ADA Compressor Only) Heavy Duty ADA Base Model Kit: 110/145 PSI (ADA Compressor Only)
42045 Heavy Duty ADA Base Model Kit: 85/105 PSI (ADA Compressor Only) 42047 Heavy Duty ADA Base Model Kit: 110/145 PSI (ADA Compressor Only) 45052 Constant Duty ADA Base Model Kit: 85/105 PSI (ADA Compressor
More informationThis test shall be carried out on all vehicles equipped with open type traction batteries.
5.4. Determination of hydrogen emissions page 1 RESS-6-15 5.4.1. This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4.2. The test shall be conducted following
More informationDigital Flow Switch. Series PFM7
Compatible with rgon (r), Carbon Dioxide (CO) and the Mixed Gas (r+co) Digital Flow witch eries PFM7 Flow adjustment valve Fluid r, CO r + CO RoH -color display ee abnormal values at a glance. llows flow
More informationRD(H)20/25 Pressure-Reducing Regulator User Manual
RD(H)20/25 Pressure-Reducing Regulator User Manual Read the complete manual before installing and using the regulator. 2 Safe Product Selection When selecting a product, the total system design must be
More informationOXYGEN MINERALE ABSOLUTE RANGE OWNER S MANUAL OXY MINERALE 600/1000 SERIES
PURAPOOL OXYGEN MINERALE ABSOLUTE RANGE OWNER S MANUAL OXY MINERALE 600/1000 SERIES 1. PARTS IDENTIFICATION LIST 1. 5. 2. 4. 3. 6. 6. Figure 1.1 1. OXYGEN Minerale Power pack with timer option (x1) 2.
More information2/ Port G1/4
CATALOGUE > Release 8.8 > Series ER100 digital electro-pneumatic regulators Series ER100 digital electro-pneumatic regulators Port G1/4 Compact design Digital display Analog and digital input Programmable
More informationCommercial Softeners. Installation, Operation & Maintenance Guide. Manual 002.1
Commercial Softeners Installation, Operation & Maintenance Guide Manual 002.1 Contents Page 1. Unpacking Instructions 3 2. Installation 3 Pre-installation checks Fitting the bottom distribution system
More informationINSTRUCTION MANUAL. FLOAT TYPE LEVEL TRANSMITTER HT-100R Series HITROL CO., LTD.
HITROL CO., LTD. HEAD OFFICE.FACTORY.R&D INSTITUDE HITROL CO.,LTD 141, Palhakgol-gil, Jori-eup Paju-si, Gyeonggi-do, Korea TEL. : (00)-82-31-950-9700 FAX. : (00)-82-31-943-5600 www.hitrol.com INSTRUCTION
More informationBante810 Benchtop Dissolved Oxygen Meter Instruction Manual
Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante810 Benchtop Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante810 benchtop dissolved oxygen
More information400C & 450C DUAL PERFORMANCE VALUE PACKS
(Chrome) PART NO. 40013 (Silver) PART NO. 45012 (Chrome) PART NO. 45013 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of this manual before
More informationTWC Services, Inc. Structured On-The-Job Training for JT- 1, 2, 3
TWC Services, Inc. Structured On-The-Job Training for JT- 1, 2, 3 The following tasks parallel the Jr. Tech (JT-1, 2, 3) description and development plan and use the Refrigeration and Air conditioning
More informationNIV EAU MATIC ST User Manual Made in Canada
NIV EAU MATIC ST User Manual Made in Canada www.niveaumatic.com info@niveaumatic.com -2- Table of contents Introduction Section 1 Page 4 Installation Section 2 Pages 5 to 10 2.1 Control box 2.2 Sensor
More informationEPP4 Pressure Regulator Basic G 1/4"
EPP4 Pressure Regulator Basic G 1/4" Flow Curve 1/4" Dimensions Outlet pressure - bar 9 8 7 6 5 4 3 2 1 0 0 20 40 60 70 80 100 120 Flow rate - Nm 3 /h The male connector adopted on the EPP4 is a standard
More informationEngineered Solutions for Gas Detection and Analysis. High Purity Instruments MODEL 3-LPM-N1-SS
Engineered Solutions for Gas Detection and Analysis High Purity Instruments MODEL 3-LPM-N1-SS Remote Oxygen Sensor Module and Positive Pressure Sampling System for Model 3100 PERCENT to PPM RANGE Oxygen
More informationSeries 6517 Katharometers. The ideal equipment for Process Monitoring and Control. Robust no moving parts. Designed for continuous industrial use
Data sheet Katharometers The ideal equipment for Process Monitoring and Control Designed for continuous industrial use Long working life Suitable for flammable gases CENELEC certified (ATEX) to EExia IIC
More informationReverse Osmosis System Installation Guide and Operation Manual.
Reverse Osmosis System Installation Guide and Operation Manual. Table of Contents 1 Introduction...2 2 Installation...3 2.1 Feed water connection...3 2.2 Permeate and concentrate plumbing...3 2.3 Electrical
More informationAHE58/59 AC Servo System
AHE58/59 AC Servo System HMI-12 User Manual Safely INstruction Please read this manual carefully, also with related manual for the machine head before use. For perfect operation and safety, installing
More informationTSE Blood Pressure Monitor Invasive series
TSE Blood Pressure Monitor Invasive 209100series For Invasive Measurements in Laboratory Animals 2 Specifications subject to change without notice USA / Canada / Mexico Phone 1-989-698-3067 Fax 1-989-698-3068
More informationIntelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter
Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter Overview C % ppm 4~20mA Analog Output ppb Power Supply 100~240 VAC mg/l Dimensions 96 x 96 x 132mm RS-485 Digital Output (for DC-5310-RS only)
More informationTECHNICAL INFORMATION
TECHNICAL INFORMATION Models No. TD0101, TD0101F Description Impact Driver L PRODUCT P 1/ 14 CONCEPT AND MAIN APPLICATIONS Models TD0101 and TD0101F are cost-competitive 100N.m-class impact driver developed
More informationInstallation, use and maintenance instructions. Gas burner (5)
Installation, use and maintenance instructions Gas burner MODEL GAS 4 TYPE 516 T80 291 (5) TECHNICAL FEATURES Thermal output 180-470 kw 154.800-404.200 kcal/h Fuel Natural gas Pci 8-10 kwh/m 3 = 7000-8600
More informationNavigator 600 Silica analyzers
ABB MEASUREMENT & ANALYTICS INFORMATION INF09/020 REV. B Navigator 600 Silica analyzers Troubleshooting Measurement made easy Navigator 600 silica analyzers 1 Introduction This publication details troubleshooting
More information