1 COMBUSTION ANALYSERS SERVOTOUGH Fluegas(2700) 2730C Utilities Unit User and Installation Manual Part Number: C Revision: 1 Language: UK English

2 This page intentionally blank

3 CONTENTS LIST CONTENTS LIST LIST OF FIGURES INTRODUCTION WARNINGS, CAUTIONS and NOTES Scope of this Manual General Safety Information Unpacking DESCRIPTION AND OPERATION Overview Air Set (Single and dual sensor) Air Set and Autocalibration (Single sensor) Air Set and Autocalibration (Dual sensor) Air Set with Blowback and Autocalibration (Single sensor) Air Set with Blowback and Autocalibration (Dual sensor) Air Set and Manual Calibration (Single and dual sensor) INSTALLATION Mechanical Installation Electrical Installation TECHNICAL SPECIFICATION Air Set (Single and dual sensor) Air Set and Manual Calibration (Single and dual sensor) Air Set with Autocalibration (Single sensor) Air Set with Autocalibration (Dual sensor) Air Set with Blowback and Autocalibration (Single sensor) Air Set with Blowback and Autocalibration (Dual sensor) 5. MAINTENANCE AND SPARES Spare parts for Single and dual sensor air set ( ) Spare parts for Single sensor air set with auto calibration ( ) Spare parts for dual sensor air set with auto calibration ( C) Spare parts for dual sensor air set with auto calibration ( C) Spare parts for Single Sensor air set with auto calibration and blowback ( ) Spare parts for dual sensor air set with autocalibration and blowback ( C) Spare parts for single and dual sensor air set with manual calibration ( ) Routine maintenance User and Installation Manual 1

4 LIST OF FIGURES Figure 1 Single and dual sensor air set flow diagram Figure 2 Single sensor air set with auto calibration flow diagram Figure 3 Dual sensor air set with auto calibration flow diagram Figure 4 Dual sensor air set with blow back and auto calibration flow diagram Figure 5 Dual sensor air set with blow back and auto calibration flow diagram Figure 6 Single and dual sensor air set with manual calibration flow diagram Figure 7 Single sensor air set installation drawing Figure 8 Air set with either auto calibration or blowback installation drawing Figure 9 Gland plate details, auto calibration and blowback versions Figure 10 Manual calibration utility unit version installation drawing Figure 11 Interconnecting wiring schematic User and Installation Manual

5 Ref No. Revision History Sheet C 2730 Operators Manual Page(s) Affected Summary of Change Date/ by 02730/001C/0 First issue G.Botting /001C/1 Filter regulator part number changed except on assembly M. Hubball User and Installation Manual 3

6 This page is left intentionally blank User and Installation Manual

7 1. INTRODUCTION 1.1 WARNINGS, CAUTIONS and NOTES This manual includes WARNINGS, CAUTIONS and NOTES which provide important information relating to the following: WARNINGS Hazards which could result in personal injury or death. CAUTIONS Hazards which could result in damage to equipment or property. NOTES Alerts the user to pertinent facts and conditions. 1.2 Scope of this Manual This instruction manual describes installation and operation of the 2730 Utilities Units that are available for use with the 2700C combustion gas analyser. For technical specifications, see section 4 of this manual. For the addresses for technical assistance and spares, see the rear of this manual. About this Manual Ref.02730/001C/1 Order as Part C CE Marking Where applicable the 2730 utilities units carry the CE mark which indicates conformity with the European Community Directives on CE Marking (93/68/EEC), Electromagnetic Compatibility (EMC 89/336/EEC) and Low Voltage Directive (LVD 73/23/EEC) User and Installation Manual 5

8 1.3 General Safety Information Specific standards of safety are not included in this manual, since requirements of relevant authorities vary widely. The safety precautions described in this manual ensure that the existing safety features are not impaired during normal operation and use. WARNINGS The electrical power used in this equipment is at a voltage high enough to endanger life. It is essential that only suitably trained and competent personnel are allowed to access hazardous live parts by removing or opening covers of the equipment. The installer must be satisfied that the 2730 installation conforms to the relevant safety requirements, National Electrical Code and any other local regulations, and that the installation is safe for any extremes of conditions which may be experienced in the operating environment of the equipment. This appliance must be connected to a protective earth. The electrical installation must include a means of isolating electrical power by a switch or circuit breaker external to the analyser and within easy reach of the operator. It must be marked as the disconnecting device for the equipment. The 2700 sensor head aspirator supply is suitable for use with compressed air only. Flammable gases must not be used for the aspirator supply. The 2730 Utilities unit is not suitable for use in hazardous areas / locations without the use of appropriate safety purge equipment. The 2730 Utilities unit is only to be used with the gasses specified in the 2700 Quickstart Manual, 2700 Installation Manual and/or this manual User and Installation Manual

9 WARNING - GAS HAZARDS If a gas leak occurs within the automatic calibration utilities units, the enclosure may become filled with one or both of the following gaseous hazards. a) 0.3% oxygen in nitrogen which is an asphyxiant. b) Up to 1% carbon monoxide in air which is toxic. Appropriate care must be taken when opening the utilities enclosure for any reason. Automatic calibration units have a breathing device fitted to the enclosure. Ensure that this device is not obstructed or blocked. Manual and Automatic calibration units MUST be located in a freely ventilated area. 1.4 Unpacking WARNING The 2730 Utilities Units supplied with Autocalibration weigh <16Kg (<35lb). Care must be taken when handling. Remove the 2730 components from their packing and inspect for obvious external damage. If damage has occurred, inform Servomex or its agent immediately. Retain all packing and shipping information for future use User and Installation Manual 7

10 This page is left intentionally blank User and Installation Manual

11 2. DESCRIPTION AND OPERATION 2.1 Overview The 2700 Combustion Gas Analyser measures the concentrations of oxygen and combustibles present in the sample, this provides the ability to maximise combustion efficiency and reduce emissions. The 2730 Utilities Units provide a convenient package to supply the utility gases required by the 2700 sensor head. The Utilities Unit may provide a number of different gas supplies to the sensor head. For optimum performance, only one analyser should be used in conjunction with an air supply or utilities unit. Aspirator air supply This provides the instrument grade compressed air supply required by the 2700 combustion gas analyser sample aspirator. The aspirator is used to extract the gas sample from the flue into the analyser sensor head. The compressed air supply should be clean, dry, and free from oil and dirt. The aspirator pressure range required is 3 to 5 psig. The flow rate is typically <1.5 l/min under typical flue pressure conditions. Auxiliary air supply The combustibles only and dual sensor versions of the 2700 require auxiliary air to supplement the level of oxygen present in the combustibles sensor. This is derived from the aspirator air supply within the sensor head to give a flow of between 80 to 120 ml/min. Calibration gas supply As good practice, Servomex recommend that the 2700 is checked/validated with check gas every month and then calibrated as required. This period can be extended as a result of operating experience. Calibration can be carried out either automatically or manually. The utility units provide the analyser with a controlled constant flow of the relevant calibration gases. Blowback air supply The blow back utilities units provide a 1 barg pressure compressed air supply to the sensor head calibration port. In extreme conditions blowback may be used to extend the service interval for the external sample probe filter by blowing off any dust coating. The blow back facility is not intended as a replacement for a suitably filtered sampling system User and Installation Manual 9

12 WARNING For the utilities units supplied with either manual or automatic calibration. Do not unscrew the needle valve controlling the calibration gas flow rate by more than 4 turns. Failure to do this may result in the valve assembly becoming detached from the valve body exposing the user to potentially toxic or asphyxiant gases. CAUTIONS All Utilities units are specified to -10 C (+14 F), however, if operated below +2 C (+35.6 F) the instrument grade compressed air and calibration gases used must be free from oil, water, dirt and have a dew point 10 C (18 F) below the ambient temperature, to prevent freezing within the components. Maintenance of the Utilities Units should not be performed at temperatures below 0 C (+32 F). The 2730 utilities unit is designed for use only with the 2700C Series combustion gas analyser and should not be used with any other products or systems. NOTES The (Hoffman/Schroff) enclosure that is supplied with the automatic calibration utilities units requires a special key (which is provided) to gain access for installation and service. All gas pressures stated in this manual are gauge pressures unless otherwise specified. 1 barg = 100 kpag 1 barg = psig User and Installation Manual

13 2.2 Air Set (Single and dual sensor) Refer to the flow diagram in figure 1. The single or dual sensor air set (part number ) consists of a filtered pressure regulator supplied complete with gauge, auto drain, inlet and outlet fittings, wall mounting bracket, screws, washers and nuts. The inlet pressure to the air set should be in the range 1 to 5 barg. The outlet pressure from the regulator should be set to the required aspirator inlet pressure as stated in the sensor head terminal box (typically 3.5 psig at a flow rate of <1.5 litres/min). The inlet and outlet fittings are suitable for use with 1/4" OD plastic or metal tubing. Figure 1. Single and dual sensor air set flow diagram Air Inlet 1 to 5 barg Aspirator Air Outlet (3.5 psig <1.5 l/min) typical Auto Drain 2730 User and Installation Manual 11

14 2.3 Air Set and Autocalibration (Single sensor) Refer to the flow diagram figure 2. The oxygen only air set with autocalibration (part number ) consists of a wall mounted enclosure inside which are fitted the following components: A filtered pressure regulator supplied complete with gauge and auto drain fitting. An unfiltered pressure regulator complete with gauge. A multi voltage power supply. Two solenoid valves. A ml/min flow meter and fine control valve. The inlet pressure to the unit should be in the range 1 to 5 barg. The typical flow rate required is <2.5 litres/min. Set the outlet pressure from the filter regulator to 1 barg. The outlet from the aspirator air precision pressure regulator is set to the required aspirator inlet pressure as stated in the sensor head terminal box (typically 3.5 psig at a flow rate of <1.5 litres/min). The pressure of the calibration gases to the autocalibration unit inlet port should be set to 1 barg (flow rate of <1 litre/min). The autocalibration facility consists of two solenoid valves to select the calibration gas required and a flow meter and needle valve to regulate and measure the gas flow rate. The outlet flow rate from the autocalibration unit to the calibration port of the 2700 should be set to 600 ml/min. The inlet and outlet bulkhead fittings are suitable for use with 1/4" NPT or BSPT male fittings and tubing. At the 2700 control unit, to which the utilities unit is attached, configure the relay outputs from the analyser using the ASSIGN RELAYS menu option as described in the 2700 Quickstart Manual. The relays should be set as follows. RELAY 1 GAS 1 RELAY 2 GAS User and Installation Manual

15 Figure 2. Single sensor air set with auto calibration flow diagram. Air Inlet 1 to 5 barg <2.5 l/min Set to 1 barg Air Inlet 1 to 5 barg <2.5 l/min Aspirator Air Outlet (3.5 psig <1.5 l/min typical) SV1 Auto Drain 0.3% Oxygen in Nitrogen -OR- SV2 Calibration Gas Inlet 1 barg <1 l/min 500ppm Carbon Monoxide in Air Calibration Gas Outlet 600 ml/min 2730 User and Installation Manual 13

16 2.4 Air Set and Autocalibration (Dual sensor) Refer to the flow diagram in figure 3. The dual sensor air set with autocalibration (part number C) consists of a wall mounted enclosure inside which are fitted the following components: A filtered pressure regulator supplied complete with gauge and auto drain fitting. A precision pressure regulator complete with gauge. A multi voltage power supply. Three solenoid valves. A ml/min flow meter and fine control valve. The inlet pressure to the unit should be in the range 1 to 5 barg. The typical flow rate required is <2.5 litres/min. Set the outlet pressure from the filter regulator to 1 barg. The outlet from the aspirator air precision pressure regulator is set to the required aspirator inlet pressure as stated in the sensor head terminal box (typically 3.5 psig at a flow rate of <1.5 litres/min). The pressure of the calibration gases to the autocalibration unit inlet ports should be set to 1 barg (flow rate of <1 litre/min). The autocalibration facility consists of three solenoid valves to select the calibration gas required and a flow meter and needle valve to regulate and measure the gas flow rate. The outlet flow rate from the autocalibration unit to the calibration port of the 2700 should be set to 600 ml/min. The inlet and outlet bulkhead fittings are suitable for use with 1/4" NPT or BSPT male fittings and tubing At the 2700 control unit, to which the utilities unit is attached, configure the relay outputs from the analyser using the ASSIGN RELAYS menu option as described in the 2700 Quickstart Manual. The relays should be set as follows. RELAY 1 GAS 1 RELAY 2 GAS 2 RELAY 3 GAS User and Installation Manual

17 Figure 3. Dual sensor air set with auto calibration flow diagram. Air Inlet 1 to 5 barg <2.5 l/min Set to 1 barg Aspirator Air Outlet (3.5 psig <1.5 l/min typical) SV1 Auto Drain SV2 0.3% Oxygen in Nitrogen Calibration Gas Inlet 1 barg <1 l/min 500ppm Carbon Monoxide in Air SV3 Calibration Gas Outlet 600 ml/min 2730 User and Installation Manual 15

18 2.5 Air Set with Blowback and Autocalibration (Single sensor) Refer to the flow diagram in figure 4. The single sensor air set with blowback and autocalibration (part number ) consists of a wall mounted enclosure inside which are fitted the following components: A filtered pressure regulator supplied complete with gauge and auto drain fitting. An unfiltered pressure regulator complete with gauge. A multi voltage power supply. Three solenoid valves. A ml/min flow meter and fine control valve. The inlet pressure to the unit should be in the range 1 to 5 barg. The typical flow rate required is <10 litres/min. Set the outlet pressure from the filter regulator to 1 barg. The outlet from the aspirator air precision pressure regulator is set to the required aspirator inlet pressure as stated in the sensor head terminal box (typically 3.5 psig at a flow rate of <1.5 litres/min). The autocalibration facility consists of two solenoid valves to select the calibration gas required and a flow meter and needle valve to regulate and measure the gas flow rate. The pressure of the calibration gases to the autocalibration unit inlet port should be set to 1 barg (flow rate of <1 litre/min). The outlet flow rate from the autocalibration unit to the calibration port of the 2700 should be set to 600 ml/min. The Blowback facility consists of a solenoid valve and bulkhead fitting. The Outlet pressure from the Blowback port is 1 barg. The inlet and outlet bulkhead fittings are suitable for use with 1/4" NPT or BSPT male fittings and tubing. At the 2700 control unit, to which the utilities unit is attached, configure the relay outputs from the analyser using the ASSIGN RELAYS menu option as described in the 2700 Quickstart Manual. The relays should be set as follows. RELAY 1 GAS 1 RELAY 2 GAS 2 RELAY 4 BLOWBACK AIR User and Installation Manual

19 Figure 4. Dual sensor air set with blow back and auto calibration flow diagram. Air Inlet 1 to 5 barg <10 l/min Set to 1 barg Aspirator Air Outlet (3.5 psig <1.5 l/min typical) SV4 Blowback Air Outlet SV1 Calibration Gas Inlet 1 barg <1 l/min Auto Drain 0.3% Oxygen in Nitrogen -OR- 500ppm Carbon Monoxide in Air SV2 Calibration Gas Outlet 600 ml/min 2730 User and Installation Manual 17

20 2.6 Air Set with Blowback and Autocalibration (Dual sensor) Refer to the flow diagram in figure 5. The dual sensor air set with blowback and autocalibration (part number C) consists of a wall mounted enclosure inside which are fitted the following components: A filtered pressure regulator supplied complete with gauge and auto drain fitting. A precision pressure regulator complete with gauge. A multi voltage power supply. Four solenoid valves. A ml/min flow meter and fine control valve. The inlet pressure to the unit is 1 to 5 barg. The typical flow rate required is<10 litres/min. Set the outlet pressure from the filter regulator to 1 barg. The outlet from the aspirator air precision pressure regulator is set to the required aspirator inlet pressure as stated in the sensor head terminal box (typically 3.5 psig at a flow rate of <1.5 litres/min). The pressure of the calibration gases to the autocalibration unit inlet ports should be set to 1 barg (flow rate of 1 litre/min). The autocalibration facility consists of three solenoid valves to select the calibration gas required and a flow meter and needle valve to regulate and measure the gas flow rate. The outlet flow rate from the autocalibration unit to the calibration port of the 2700 should be set to 600 ml/min. The blowback facility consists of a solenoid valve and bulkhead fitting. The outlet pressure from the blowback port is 1 barg. The inlet and outlet bulkhead fittings are suitable for use with 1/4" NPT or BSPT male fittings and tubing. At the 2700 control unit, to which the utilities unit is attached, configure the relay outputs from the analyser using the ASSIGN RELAYS menu option as described in the 2700 Quickstart Manual. The relays should be set as follows. RELAY 1 GAS 1 RELAY 2 GAS 2 RELAY 3 GAS 3 RELAY 4 BLOWBACK AIR User and Installation Manual

21 Figure 5. Dual sensor air set with blow back and auto calibration flow diagram. Air Inlet 1 to 5 barg <10 l/min Set to 1 barg Aspirator Air Outlet (3.5 psig <1.5 l/min typical) SV4 Blowback Air Outlet Auto Drain 0.3% Oxygen in Nitrogen SV1 SV2 Calibration Gas Inlet 1 barg <1 l/min 500ppm Carbon Monoxide in Air SV3 Calibration Gas Outlet 600 ml/min 2730 User and Installation Manual 19

22 2.7 Air Set and Manual Calibration (Single and dual sensor) Refer to the flow diagram in figure 6. The single and dual sensor air set with manual calibration (part number ) consists of a wall mounted panel upon which are fitted the following components: A filtered pressure regulator supplied complete with gauge and auto drain fitting. A precision pressure regulator complete with gauge. A ml/min flow meter and fine control valve. The inlet pressure to the unit should be in the range 1 to 5 barg. The typical flow rate required is <2.5 litres/min. Set the outlet pressure from the filter regulator to 1 barg. The outlet from the aspirator air precision pressure regulator is set to the required aspirator inlet pressure as stated in the sensor head terminal box (typically 3.5 psig at a flow rate of <1.5 litres/min). The manual calibration facility consists of a flowmeter and fine flow control valve. The pressure of the calibration gases to the manual calibration unit should be set to 1 barg (flow rate of <1 litre/min). The outlet flow rate from the manual calibration unit to the calibration port of the 2700 should be set to 600 ml/min. The inlet and outlet bulkhead fittings are suitable for use with 1/4" NPT or BSPT male fittings and tubing. Figure 6. Single and dual sensor air set with manual calibration flow diagram. Air Inlet 1 to 5 barg <2.5 l/min Set to 1 barg Aspirator Air Outlet (3.5 psig <1.5 l/min typical) Auto Drain Quick Connect Plug Quick Connect Plug Calibration Gas Outlet 600 ml/min User and Installation Manual

23 3. INSTALLATION 3.1 Mechanical Installation Select a location which allows convenient access for installation and maintenance and will minimise ambient temperature fluctuations, vibration and be freely ventilated. The utilities units should be mounted onto a rigid vertical surface capable of supporting the weight of the unit. The aspirator air set and manual calibration versions should be mounted with the filter regulator drain pointing vertically down. The autocalibration versions of the utilities units should be mounted with the electrical entries on the under side. Ensure that the operating ambient temperature is in the range of -10 C to +50 C and that the Utilities Unit is protected from radiant heat sources or from direct sunlight (in ambient temperatures over 30 C). This can be achieved by installing a Heat Shield between the Utilities Unit and the source; a thin metal sheet is usually adequate. All pressure regulators, flow meters and gauges exhibit variations of output with changes in ambient temperature (their temperature coefficient). Although Servomex has chosen products that have good rejection to ambient temperature variation, utility units should be installed in locations with relatively stable ambient conditions. Ideally temperature swings from one extreme to another should be avoided. Instrument air and other gases supplied to the utilities unit must have dew points of at least 10 C less than the minimum anticipated ambient temperature. Table 1 lists the figure numbers that contain the installation drawings for the specific utilities units. Table 1 Utility Units Installation Drawings Part No. Description Fig. No Air set (single/dual sensor) Air set with autocalibration (single sensor) 8, C Air set with autocalibration (dual sensor) 8, Air set with blowback and autocalibration (single sensor) 8, C Air set with blowback and autocalibration (dual sensor) 8, Air and manual calibration (single and dual sensor) 10 NOTE The liquid that may collect in the filter regulator bowl will be automatically discharged through the drain port. This liquid must be disposed of in an appropriate and safe manner User and Installation Manual 21

24 Figure 7. Single sensor air set installation drawing. f 8 32mm 7 99mm 93mm 8mm 9mm Ø5.5mm FLOW 124mm 165mm Kit includes two M5 x 12mm mounting bolts, M5 washers and M5 full hex nuts Key to figure 7. 1 Two off 5.5mm diameter mounting holes for 5mm mounting bolts. 2 Filter drain port (threaded 1/8" NPT male). 3 Filter bowl. 4 Outlet pressure gauge, 0-20 psig (part number ). 5 Compressed air inlet (1/4" OD compression fitting). 6 Aspirator air outlet (1/4" OD compression fitting). 7 Mounting bracket. 8 Pressure adjustment control User and Installation Manual

25 Figure 8. Air set with either auto calibration or blowback installation drawing 227mm 220mm 1 400mm 362mm TB3 500mm 540mm 578mm 593mm TB4 7 SV4 SV1 SV2 SV3 TB2 TB1 8 Ø10mm Key to figure 8. 1 Hoffman hinged enclosure. 2 Four off 10mm diameter mounting holes. 4 Filter regulator assembly (part number ). 5 Filter regulator output pressure gauge, 0-20 psig (part number ). 6 Aspirator air pressure regulator (see spares list for part numbers). 7 Blowback solenoid valve (if required, part number ). 8 Calibration gas solenoid valves (2 off for single sensor or 3 off for dual sensor, part number ). 9 Gland plate assembly (see figure 14). 10 Enclosure breather fitting (part number ). 11 Power supply PCB (part number ) and transparent mains protection cover. 12 Calibration gas flow meter and needle valve (part number ). 13 Aspirator air output pressure gauge, 0-20 psig (part number ). 15 Adaptor, 1/4" BSPT (M) to 1/4" NPT (F) ** User and Installation Manual 23

26 Figure 9. Gland plate details, auto calibration and blowback versions ASP.AIR OUTLET DRAIN AIR INLET MAX. 5 BAR CAL GAS/ BLOWBACK CAL GAS (O2 LOW) MAX. 1 BAR CAL GAS (CO HIGH) SIGNAL MAX. 1 BAR MAINS Key to figure 9 1 Aspirator air outlet bulkhead fitting **. 2 Calibration and blowback (if fitted) outlet bulkhead fitting **. 3 Unused 4 Enclosure breather fitting (part number ). 5 Main power inlet M20 cable gland. 6 Signal connection inlet M20 cable gland. 7 Combustibles sensor high calibration gas inlet bulkhead fitting **. 8 Oxygen sensor low calibration gas inlet bulkhead fitting **. 9 Compressed air inlet bulkhead fitting **. 10 Filter regulator automatic drain outlet **. NOTE ** Gas connections are 1/4" NPT(F) as supplied, 1/4" BSPT (F) is obtained by removal of adaptor item 15 of figure User and Installation Manual

27 Figure 10. Manual calibration utility unit version installation drawing mm 450mm 124mm 2 390mm 360mm mm mm 265mm 5 41mm 6 Ø8.5mm Key to figure 10 1 Mounting bracket. 2 Filter regulator assembly (part number ). 3 Four off 8.5mm diameter mounting holes. 4 Compressed air inlet bulkhead fitting **. 5 6mm OD Quick Connect fitting and tubing (part number ). 6 Quick connect socket. 7 Filter regulator output pressure gauge, 0-20 psig (part number ). 8 Filter regulator automatic drain outlet. 11 Calibration gas flow meter and needle valve (part number ). 13 Calibration gas outlet bulkhead fitting **. 14 Aspirator air outlet bulkhead fitting **. 15 Aspirator air outlet pressure gauge, 0-20 psig (part number ). 16 Aspirator air pressure regulator (part number ). 17 Adaptor, 1/4" BSPT (M) to 1/4" NPT (F) **. NOTE ** Gas connections are 1/4" NPT(F) as supplied, 1/4" BSPT (F) is obtained by removal of adaptor item Unit comes with two 1/4" OD compression Quick connect plugs, to fit into external calibration gasses (part number ) User and Installation Manual 25

28 3.2 Electrical Installation This section defines the electrical interconnection details for the automatic calibration versions of the utilities units. The part numbers for the utility units requiring an electrical power supply are , C, , and C. WARNING Refer to the warnings and conditions outlined in the General Safety Information in section 1.3 of this manual. CAUTIONS The utilities unit supply voltage is field configurable for either V or V (nominal). Ensure that the rated voltage and frequency is suitable for the installation. Ensure that the link on the PCB has been set to the correct supply voltage. The supply voltage is set by linking together the terminals at terminal block TB2. Ensure that the link setting is correct for the rated supply voltage. The sensor head is intended to be powered from a supply which has the Neutral conductor referenced to ground. Where this is unavailable, it is recommended that a suitable isolating transformer is installed and the secondary winding is referenced to ground. Alternatively, the electrical installation must include a fuse rated F 250 ma HBC in each supply conductor. The disconnect device (see also section 1.3) must disconnect each supply conductor. The fuses must be marked as being associated with this equipment. Ensure that the installation conforms to the relevant national and local safety requirements. NOTES The conduit or screen (if any) associated with the power supply should be grounded at the cable gland or conduit entry to the enclosure in accordance with local wiring regulations. Fuse F1 located on the PCB is rated F250mA HBC for V and V operation. Solenoid valve 4 only must be used for the Blowback facility as it has a larger internal diameter than Solenoid valves 1, 2 & User and Installation Manual

29 Refer to figure 11 for an interconnection wiring schematic. The 2730 utilities units are rated in accordance with IEC 644 for INSTALLATION CATEGORY II which is characterised as being local level (not distribution level), appliances and portable equipment with over voltage impulse withstand up to 2500V. Electrical terminations are suitable for 20 AWG (0.5 mm 2 ) to 14 AWG (2.5 mm 2 ) solid conductors or 16 AWG (1.5 mm 2 ) stranded conductors. Gland entry sizes and positions are shown in the appropriate installation drawing. Un-used entries should be fitted with appropriate plugs. Ensure that good EMC practices are observed when making connections. Refer also to the 2700 Installation manual. Where a screened cable is specified for the relay connections, in order to avoid ground loops, the screen should be terminated to ground only at the Control unit. All electrical connections and access to the fuse, F1, are made to the PCB inside of the unit enclosure. To gain access use the special key provided and open the door. After electrical connections are complete ensure the cover is fitted and the door is locked closed. Figure 11. Interconnecting wiring schematic 2700 Control Unit 2730 Control Unit Vdc TB N/O N/C COM RELAY 1 TB4 1 2 Solenoid Valve N/O N/C COM RELAY Solenoid Valve N/O N/C COM RELAY Solenoid Valve N/O N/C COM RELAY Solenoid Valve 4 TB3 L Nl E Live Neutral Earth Mains Input Mains Input TB1 L Nl E 0Vdc Supply Vac Vac 100Vac AC Voltage Link Selector. Shown connected for Vac operation TB2 Control unit PCB / / Utilities Unit PCB User and Installation Manual 27

30 This page is left intentionally blank User and Installation Manual

31 4. TECHNICAL SPECIFICATION 4.1 Air Set (Single and dual sensor) Part numbers Dimensions (W x D x H): : 93mm x 99mm x 165mm (3.7" x 3.9" x 6.5") Weight : < 400 gm (< 1.0 lb) Temperature Operating : -10 C to +50 C (+14 F to +122 F) Storage : -20 C to +55 C (-4 F to +131 F) Compressed Air Supply Pressure : 1 to 5 barg (15 to 72 psig) Flowrate : <1.5 litres/min Instrument grade compressed air, free from oil, water and dirt. External connections are suitable for 1/4" OD plastic or metal tubing. Materials of Construction Filtered Pressure Regulator Filter bowl : Polycarbonate to BS 6005 Filter element : Polypropylene, (40 µm) Regulator body : Zinc alloy Adjusting knob : Acetal resin Elastomers : Synthetic rubber Gauge (0-20 psi / bar) Body : Plastic and steel Face : Plastic Movement : Copper/brass Fittings : Brass (Fibre sealing washer) Wall mounting bracket : Mild steel Wall mounting fixings M5x12mm Bolts : Stainless steel M5 full hex nut : Stainless steel M5 washer : Stainless steel 2730 User and Installation Manual 29

32 4.2 Air Set and Manual Calibration (Single and dual sensor) Part numbers Dimensions (W x D x H) : 390mm x 174mm x 265mm (15.4" x 6.8" x 10.4") Weight : < 4.5 kg (< 10 lbs) Temperature Operating : -10 C to +50 C (+14 F to +122 F) Storage : -20 C to +55 C (-4 F to +131 F) Compressed Air Supply Pressure : 1 to 5 barg (15 to 72 psig) Flow rate : <2.5 litres/min Instrument grade compressed air, free from oil, water and dirt. Calibration Gas Supply Pressure : 1 barg (15 psig) Flow rate : 1 litre/min Customer connections are brass /nickel plated brass and suitable for use with 1/4" NPT or BSPT male fittings. Materials of Construction Filtered Pressure Regulator Filter bowl : Polycarbonate to BS 6005 Filter element : Polypropylene, (40 µm) Regulator body : Zinc alloy Adjusting knob : Acetal resin Elastomers : Synthetic rubber Gauge (0-20 psi / bar) Body : Plastic and steel Face : Plastic Movement : Copper/brass Precision Pressure Regulator Regulator body : Zinc alloy Adjusting knob : Acetal resin Elastomers : Synthetic rubber Flowmeter & Valve O rings Calibration Quick fit connections Body Seals Wall mounting bracket Bulkhead /Adaptor fittings Fittings Tubing : Polycarbonate and steel : Neoprene and Buna N : Stainless steel and brass : Buna : Epoxy powder painted mild steel : Brass /nickel plated brass : Plastic & Steel with NBR packing & O rings : Polyurethane User and Installation Manual

33 4.3 Air Set with Autocalibration (Single sensor) Air Set with Autocalibration (Dual sensor) Air Set with Blowback and Autocalibration (Single sensor) Air Set with Blowback and Autocalibration (Dual sensor) Part numbers , C, & C Dimensions (W x D x H) : 400mm x 220mm x 500mm (16" x 9" x 20") Wall mounting brackets : add an extra 40mm top and bottom Weight : < 16 kg (< 35 lb) Ingress protection : IP65/NEMA 12 Power Supply : Vac, 50/60 Hz, <20VA (Field configurable) : Vac, 50/60 Hz, <20VA Installation Category : II Temperature Operating : -10 C to +50 C (+14 F to +122 F) Storage : -20 C to +55 C (-4 F to +131 F) Compressed Air Supply Pressure : 1 to 5 barg (15 to 72 psig) Flow rate : <2.5 litres/min without blowback : <10 litres/min with blowback Instrument grade compressed air, free from oil, water and dirt. Calibration Gas Supply Pressure : 1 barg (15 psig) Flow rate : 1 litre/min Customer connections are nickel plated brass and suitable for use with 1/4" NPT or BSPT male fittings. Materials of Construction Filtered Pressure Regulator Filter bowl : Polycarbonate to BS 6005 Filter element : Polypropylene, (40 µm) Regulator body : Zinc alloy Adjusting knob : Acetal resin Elastomers : Synthetic rubber Gauge (0-20 psi / bar) Body : Plastic and steel Face : Plastic Movement : Copper/brass Precision Pressure Regulator Regulator body : Zinc alloy Adjusting knob : Acetal resin Elastomers : Synthetic rubber Solenoid Valve Body : Stainless steel Seal : Viton Flowmeter & Valve : Polycarbonate and steel O rings : Neoprene and Buna N Enclosure : Powder coated steel (RAL7032) Wall mounting brackets : Zinc plated steel Bulkhead / Adaptor fittings : Nickel plated brass and plastic Internal Fittings : Plastic & Steel with NBR packing & O rings 2730 User and Installation Manual 31

34 Tubing Venting device : Polyurethane : Stainless steel The four automatic calibration utilities units , C, , & C comply with the CE Marking Directive 93/68/EEC User and Installation Manual

35 5. MAINTENANCE AND SPARES 5.1 Spare parts for Single and dual sensor air set ( ) Part No. Description Filter regulator <3.5 bar Pressure gauge 0-20 psig 5.2 Spare parts for Single sensor air set with auto calibration ( ) Part No. Description Filter regulator <7 bar Aspirator air pressure regulator Pressure gauge 0-20 psig Calibration gas solenoid valve Power supply PCB Enclosure breather fitting Dwyer flow meter l/min 5.3 Spare parts for dual sensor air set with auto calibration ( C) Part No. Description Filter regulator <7 bar Pressure gauge 0-20 psig Precision pressure controller Calibration gas solenoid valve Power supply PCB Enclosure breather fitting Dwyer flow meter l/min 2730 User and Installation Manual 33

36 5.4 Spare parts for Single Sensor air set with auto calibration and blowback ( ) Part No. Description Filter regulator <7 bar Aspirator air pressure regulator Pressure gauge 0-20 psig Calibration gas solenoid valve Blowback gas solenoid valve Power supply PCB Enclosure breather fitting Dwyer flow meter l/min 5.5 Spare parts for dual sensor air set with autocalibration and blowback ( C) Part No. Description Filter regulator <7 bar Pressure gauge 0-20 psig Precision pressure controller Calibration gas solenoid valve Blowback gas solenoid valve Power supply PCB Enclosure breather fitting Dwyer flow meter l/min User and Installation Manual

37 5.6 Spare parts for single and dual sensor air set with manual calibration ( ) Part No. Description Filter regulator <7 bar Pressure gauge 0-20 psig Precision pressure controller Dwyer flow meter l/min Quick connect plug (1/4" OD compression) Quick connect plug (6mm OD compression) 5.7 Routine maintenance CAUTIONS All Utilities units are specified to -10 C (+14 F), however, if operated below +2 C (+35.6 F) the instrument grade compressed air and calibration gases used must be free from oil, water, dirt and have a dewpoint 10 C (18 F) below the ambient temperature, to prevent freezing within the components. Maintenance of the Utilities Units should not be performed at temperatures below 0 C (+32 F). On initial installation check the filter regulator bowl monthly for signs of contamination or liquid from the compressed air inlet stream. If required remove and clean the bowl and filter. If no contamination is evident then this time interval can be extended depending on experience. Every 3 months visually inspect the internal and external pipe work and fittings for signs of leaks or failure. Every 6 months check and if necessary adjust the aspirator air pressure and calibration gas flow rates. Autocalibration units in enclosures - ensure that the breather is not obstructed User and Installation Manual 35

38 NOTES User and Installation Manual

Best Practice Guide, Servomex 2700

Best Practice Guide, Servomex 2700 For full installations details refer to the. Best Practice Guide, Servomex 2700 Mounting: General Guidelines: Servomex 2700 Control Units and air supplies (utilities units) should, ideally, be mounted

More information

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES PROCESS ANALYSERS DF-550E The DF-550E Nano Trace oxygen analyser is a Coulometric sensor based analyser designed to measure oxygen as a contaminant at ultra trace levels in ultra high purity electronic

More information

Pressure Automated Calibration Equipment

Pressure Automated Calibration Equipment GE Measurement & control Pressure Automated Calibration Equipment Safety Instructions and User Guide - K0447 PACE5000 PACE6000 K0447 Issue No. 9 1 10 1 PACE5000 1 2 3 4 5 PACE6000 2 6 7 8 3 4 5 6 7 8 9

More information

SERVICE MANUAL MODEL BA600AMST-S1 BREATHING AIR PANEL

SERVICE MANUAL MODEL BA600AMST-S1 BREATHING AIR PANEL SERVICE MANUAL MODEL BA600AMST-S1 BREATHING AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the service manual enclosed with this device. 2/05 GENERAL

More information

Electropneumatic Transducer (I/P) Installation, Operation and

Electropneumatic Transducer (I/P) Installation, Operation and Nitra NCP1 Series Electropneumatic Transducer (I/P) Installation, Operation and Maintenance Instructions Ordering Information Contents Part Number I/P Transducers Output Range Input psig bar Section Description

More information

3-Way Decoking Valve Option - Installation and Operation Instructions Accessory G Background

3-Way Decoking Valve Option - Installation and Operation Instructions Accessory G Background 3-Way Decoking Valve Option - Installation and Operation Instructions Accessory G6600-60058 Background Accumulation of contaminants, such as column bleed, impurities in detector gases and even carbon from

More information

Model PDT Dewpoint Transmitter

Model PDT Dewpoint Transmitter Model PDT Dewpoint Transmitter Instruction Manual Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Tel: +44 1274 733100 Fax: +44 1274 733200 email: mail@amsytems.co.uk web: www.amsystems.co.uk

More information

Lesman Level Instrumentation LSB100/300 Series Basic Bubbler System

Lesman Level Instrumentation LSB100/300 Series Basic Bubbler System Lesman Level Instrumentation LSB100/300 Series Basic Bubbler System PART 1: GENERAL 1.1 SCOPE A. This section describes the specifications of a bubbler system. B. The following information shall be described

More information

AutoChanger Installation & User Guide Issue 2

AutoChanger Installation & User Guide Issue 2 1 INDEX Page 1. Introduction... 3 2. AutoChanger Components Guide 4 3. Installation. 5-11 a) Installation Guidelines.. 5 b) Installation Retrofit... 6-11 c) Installation New... 11 4. AutoChanger User Guide

More information

GM Series Dual-Block Multi-Function Gas Control Valves

GM Series Dual-Block Multi-Function Gas Control Valves Installation Sheets Manual 121 Gas Combustion Combination Controls and Systems Section G Technical Bulletin GM Issue Date 0297 GM Series Dual-Block Multi-Function Gas Control Valves Figure 1: GM Series

More information

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL www.modsafe.com SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the service manual enclosed with this device.

More information

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS Annex 3, page 2 ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS The text of existing chapter 15 is replaced by the following: "1 Application This

More information

Autocalibration Systems For In Situ Oxygen Analyzers

Autocalibration Systems For In Situ Oxygen Analyzers Product Data Sheet PDS 106-340AC SPS 4000, IMPS 4000, and MPS 3000 January, 1999 Autocalibration Systems For In Situ Oxygen Analyzers Cost-effective autocalibration systems for installations ranging from

More information

IX1 Spirax-Monnier International Compressed Air Filter Installation and Maintenance Instructions

IX1 Spirax-Monnier International Compressed Air Filter Installation and Maintenance Instructions 0575050/2 IM-P057-04 CH Issue 2 IX1 Spirax-Monnier International Compressed Air Filter Installation and Maintenance Instructions 1 Safety information 2 General product information 3 Installation and commissioning

More information

MF2M Monnier Miniature Compressed Air Filter

MF2M Monnier Miniature Compressed Air Filter 5040050 / 2 IM-P504-02 CH Issue 2 MF2M Monnier Miniature Compressed Air Filter Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation and commissioning

More information

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES PROCESS ANALYSERS DF-310E The DF-310E is a Coulometric sensor based oxygen analyzer designed to measure trace and percent level oxygen in pure and multi-gas backgrounds for process and quality control

More information

Infrared Linescanners

Infrared Linescanners Thermal Imaging and Temperature Profiles for Continuous Process Monitoring and Quality Control HD Infrared Linescanners PROCESS & ANALYTICAL INSTRUMENTS Mountings and Accessories LSPHD Mountings and Accessories

More information

SERVICE MANUAL MODEL BA020AMST BREATHING AIR PANEL

SERVICE MANUAL MODEL BA020AMST BREATHING AIR PANEL SERVICE MANUAL MODEL BA020AMST BREATHING AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the service manual enclosed with this device. 5/08 2 GENERAL

More information

SERVICE MANUAL MODEL BA200AA-S1 BREATHING AIR PANEL

SERVICE MANUAL MODEL BA200AA-S1 BREATHING AIR PANEL SERVICE MANUAL MODEL BA200AA-S1 BREATHING AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the service manual enclosed with this device. 1/00 2 GENERAL

More information

Customer Responsibilities. Important Customer Information. Cary 4000/5000/6000i UV-Vis spectrophotometer Site Preparation Checklist

Customer Responsibilities. Important Customer Information. Cary 4000/5000/6000i UV-Vis spectrophotometer Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

GAS FUEL VALVE FORM AGV5 OM 8-03

GAS FUEL VALVE FORM AGV5 OM 8-03 ALTRONIC AGV5 OPERATING MANUAL GAS FUEL VALVE FORM AGV5 OM 8-03 WARNING: DEVIATION FROM THESE INSTALLATION INSTRUCTIONS MAY LEAD TO IMPROPER ENGINE OPERATION WHICH COULD CAUSE PERSONAL INJURY TO OPERATORS

More information

Installation, operating and maintenance Instructions for Seemag bypass level indicator

Installation, operating and maintenance Instructions for Seemag bypass level indicator Issue: S Date: 05-09-14 Type G35 General information The Seetru bypass magnetic level indicator, abbreviate SEEMAG, serves to show the filling level of fluids in tanks, basins, tubes etc. The Seemag operates

More information

SERVICE MANUAL MODEL BA200AMST-S1 BREATHING AIR PANEL

SERVICE MANUAL MODEL BA200AMST-S1 BREATHING AIR PANEL 1 SERVICE MANUAL MODEL BA200AMST-S1 BREATHING AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the service manual enclosed with this device. 2 GENERAL

More information

Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes

Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes Instruction Bulletin Addendum IB PB1000 Part no. Serial no. Order no. Oxymitter 4000 Applicability: Used with pressure-balancing

More information

TECHNICAL DATA. Trimpac 251a. Spetember 16, 2013

TECHNICAL DATA. Trimpac 251a. Spetember 16, 2013 Spetember 16, 2013 Trimpac 251a 1. DEsCrIpTION DESCRIPTION TRIMPAC Model B-6 and B-6B is a factory assembled trim package for a double interlocked preaction system with an electric/pneu-lectric release

More information

Operating Instructions

Operating Instructions Operating Instructions ProMinent CO2 Feeder CO2 - ProMinent P/N: 7746944 (Revision C), Date 09-14-12 Please completely read through these operating instructions first! Do not discard! The warranty shall

More information

Type 4708 Supply Pressure Regulator

Type 4708 Supply Pressure Regulator Type 478 Pressure Regulator Application pressure regulators used to provide pneumatic measuring and control equipment with a constant air supply Set point ranges. to. bar ( to 4 psi). to bar (8 to 9 psi)

More information

Type 4708 Supply Pressure Regulator

Type 4708 Supply Pressure Regulator Type 478 Pressure Regulator Application pressure regulator used to provide pneumatic measuring and control equipment with a constant air supply Set point ranges. to.6 bar ( to 4 psi) or. to 6 bar (8 to

More information

TECHNICAL DATA. Trimpac 244a. September 16, 2013

TECHNICAL DATA. Trimpac 244a. September 16, 2013 September 16, 2013 Trimpac 244a 1. DEsCrIpTION DESCRIPTION TRIMPAC Model B-1 and B-1B is a factory-assembled trim package with an electric release module in a metal enclosure. The standard trim normally

More information

TECHNICAL DATA. TRIMPAC Model B-5 & B-5B

TECHNICAL DATA. TRIMPAC Model B-5 & B-5B September 16, 2013 Trimpac 250a 1. DESCRIPTION DEsCRIPTIoN is a factory assembled trim package for a ed an electric/pneumatic release module in a metal enclosure. The standard trim normally required on

More information

Columbus Instruments

Columbus Instruments 0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011

More information

System Pressure Manager Standard & System Pressure Manager Plus

System Pressure Manager Standard & System Pressure Manager Plus System Pressure Manager Standard & System Pressure Manager Plus Installation, Commissioning & Servicing Instructions Note: THESE INSTRUCTIONS MUST BE READ AND UNDERSTOOD BEFORE INSTALLING, COMMISSIONING,

More information

Workhorse 8 Oxygen System Model No 5665, 5673, 5679

Workhorse 8 Oxygen System Model No 5665, 5673, 5679 11436 Sorrento Valley Road, San Diego CA 92121 USA 858-558-0202 Fax 858-558-1915 Workhorse 8 Oxygen System Model No 5665, 5673, 5679 Part Number 1815 Revision B SeQual Technologies San Diego, CA USA Cautions,

More information

Nitrogen Tyre Inflator Owner s Manual

Nitrogen Tyre Inflator Owner s Manual Nitrogen Tyre Inflator Owner s Manual Model:PN-8890 Please read this manual carefully before attempting assembly or use. 1 Suit for: Motorcycle, Car, Light Truck, Bus Manual Model Nitrogen Tyre Inflation

More information

SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL

SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL Product Manual MST Portable Air Filtration System SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the

More information

Model Series 62 Constant Differential Relay

Model Series 62 Constant Differential Relay Siemens Industry, Inc. INSTALLATION AND SERVICE INSTRUCTION INTRODUCTION Model Series 62 Constant Differential Relay Rev 11 March 2011 Supersedes Rev 10 The Constant Differential Relay maintains a constant

More information

Technical Information

Technical Information Technical Information Installation, Operation, Installation, Operation and Maintenance and Maintenance Manual Manual Balston Model 75700-K728 Nitrogen Generator Figure 1-75700-K728 Overall Dimensions These

More information

PROCESS ANALYSERS. SERVOPRO PureGas. Gas Purifier. Operator Manual. Part Number: A Revision: 0 Language: UK English

PROCESS ANALYSERS. SERVOPRO PureGas. Gas Purifier. Operator Manual. Part Number: A Revision: 0 Language: UK English PROCESS ANALYSERS SERVOPRO PureGas Gas Purifier Operator Manual Part Number: 02005001A Revision: 0 Language: UK English This page is intentionally blank LIST OF CONTENTS 1 DESCRIPTION AND DEFINITIONS...

More information

Dri-Line Mk2 Spirax-Monnier Compressed Air Drain Trap

Dri-Line Mk2 Spirax-Monnier Compressed Air Drain Trap 0509950/2 IM-P050-21 CH Issue 2 Dri-Line Mk2 Spirax-Monnier Compressed Air Drain Trap Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation and

More information

Installing a Probe. Typical Installation

Installing a Probe. Typical Installation for hazardous locations. Prior to any installation in a Classified Hazardous Location, verify installation methods by the Control Drawing referenced on the product s name tag 2. The Feed Through Assemblies

More information

Standard Specifications

Standard Specifications NARM, 2 Ways of Connection Small Size Pressure Gauge ø5mm Reverse flow function available on the standard model Space Saving -B2 NARM-6A-NG Standard Specifications Fluid Proof pressure psig (MPa) Max.

More information

PSG 901. Pressure Switch for Gas. elektrogas.com EE141-07/12

PSG 901. Pressure Switch for Gas. elektrogas.com EE141-07/12 PSG 901 Pressure Switch for Gas elektrogas.com EE141-07/12 PSG 901 Pressure Switch for Gas (EN1854) Contents Description... 2 Features... 2 Functioning and application... 3 Technical specifications...

More information

DK46 - DK800 Supplementary instructions

DK46 - DK800 Supplementary instructions DK46 - DK800 Supplementary instructions Variable area flowmeter Device category II2G with electrical internals Additional Ex manual KROHNE CONTENTS DK46 - DK800 1 Safety instructions 3 1.1 General... 3

More information

Aztec ADS430 EZCLEAN Compressed air supply unit

Aztec ADS430 EZCLEAN Compressed air supply unit Operating instruction OI/ADS430/EZCLN EN Rev. A Aztec ADS430 EZCLEAN Compressed air supply unit Measurement made easy Introduction The Aztec ADS430 EZCLEAN compressed air supply unit is a rugged, reliable

More information

RCD - Ultra-Rugged Differential Pressure Flowmeter

RCD - Ultra-Rugged Differential Pressure Flowmeter 0.15-1 to 85-600 GPM Liquids 2.5-25 to 300-1700 SCFM Gases 1/2 to 3 Line Sizes Bronze or 316-Ti Stainless Steel Mechanical Indicator, Digital Indicator, Transmitters and Switches Available Custom Calibrations

More information

Installation, Operation, and Maintenance Manual Parker Balston Model N2-04 Nitrogen Generator

Installation, Operation, and Maintenance Manual Parker Balston Model N2-04 Nitrogen Generator Installation, Operation, and Maintenance Manual Parker Balston Figure 1 - N2-04 Nitrogen Generator These instructions must be thoroughly read and understood before installing and operating this product.

More information

Series SFD. Clean Air Filter. Variations. Made to Order. Type. Cartridge (replaceable element) Disposable (non-replaceable element) Up to 100

Series SFD. Clean Air Filter. Variations. Made to Order. Type. Cartridge (replaceable element) Disposable (non-replaceable element) Up to 100 Clean Air Filter Variations SFD SFD SFD SFD Made to Order Type Disposable (non-replaceable element) Cartridge (replaceable element) Flow rate L/min (ANR) (at inlet pressure.7 MPa) Up to 6 Up to 8 Up to

More information

TECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1

TECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1 Dry 131h 1. DESCRIPTION The Viking Model G-1 Maintenance Air Compressor is an electric motor-driven, aircooled, single-stage, oil-less compressor. The unit is equipped with a check valve and provides a

More information

PROCESS ANALYSERS. SERVOTOUGH OxyExact (2223) Oxygen Transmitter. Installation Manual. Revision:

PROCESS ANALYSERS. SERVOTOUGH OxyExact (2223) Oxygen Transmitter. Installation Manual. Revision: PROCESS ANALYSERS SERVOTOUGH OxyExact (2223) Oxygen Transmitter Installation Manual Part Number: Revision: Language: 02223/005A 10 UK English This page intentionally blank TABLE OF CONTENTS SERVOTOUGH

More information

Information Sheet. About this information sheet. Floor and wall loadings. Attachment of process module to floor. PlasmaPro Estrelas100

Information Sheet. About this information sheet. Floor and wall loadings. Attachment of process module to floor. PlasmaPro Estrelas100 Information Sheet PlasmaPro Estrelas100 About this information sheet This data sheet provides basic information about the installation of a PlasmaPro Estrelas100 tool. Some of the information in this sheet

More information

Workhorse-15 Oxygen Generator

Workhorse-15 Oxygen Generator 11436 Sorrento Valley Road, San Diego CA 92121 USA 858-558-0202 Fax 858-558-1915 Workhorse-15 Oxygen Generator Model No. 5719, 5727, 5735 for Domestic and International Units Part number 1432 Revision

More information

Pneumatic spool valve islands

Pneumatic spool valve islands Pneumatic spool valve islands P7.GB.R MEGA SPOOL VALVE ISLANDS MEGA pneumatic spool valve islands offer maximum performance along with flexibility, easy installation and simple operation. Due to their

More information

PRO-50 Instrument Supply Regulator

PRO-50 Instrument Supply Regulator Features CRN Approved The PRO-50 Regulator has been granted a Canadian Registration Number. Sour Service Capability Available in NACE configurations that comply with NACE MR0175/MR0103. Environmental limits

More information

Design Features. General Description

Design Features. General Description Model thermal Mass Flow Controllers are designed to indicate and control set flow rates of gases. The combines the characteristics and accuracy of conventional mass flow devices into a unique compact design

More information

NGP-250/500 Nitrogen Generator Quick Start Guide

NGP-250/500 Nitrogen Generator Quick Start Guide NGP-250/500 Nitrogen Generator Quick Start Guide Version: A July 2013 Potter Electric Signal Company, LLC 5757 Phantom Dr., Suite 125 P. O. Box 42037 Hazelwood, MO 63042 Phone: (314) 595-6900 Document

More information

Direct Operated Precision Regulator Modular Style. Standard Specifications

Direct Operated Precision Regulator Modular Style. Standard Specifications 1 Direct Operated Precision Regulator Modular Style Series P3 Courtesy of Steven Engineering, Inc.-23 Ryan Way, South San Francisco, CA 948-637-Main Office: (65) 588-92-Outside Local Area: (8) 258-92-www.stevenengineering.com

More information

5 Channel Calibrator

5 Channel Calibrator 212 N. Woodwork Lane Palatine, IL 60067 800-223-3977 5 Channel Calibrator Part # 0724 User Manual Questions? Contact us at 800-223-3977 or online at http://www.cleanair.com/equipment/express/main.html

More information

ECS Protector Nitrogen Generator PGEN-30 (PGEN-30E)

ECS Protector Nitrogen Generator PGEN-30 (PGEN-30E) ECS Protector PGEN-30 (PGEN-30E) Specifications For use under U.S. Patents 8,720,591, 9,144,700 and 9,186,533 Dimensions (cabinet): 24.5 (W) x 52.5 (H) x 8.5 (D) (622mm(W) x 1,334mm(H) x 216mm(D)) Dimensions

More information

Installation, Operation & Maintenance Manual

Installation, Operation & Maintenance Manual Original Instructions Installation, Operation & Maintenance Manual Sentry RPG Hazardous Controller Sampler Controllers S-AS-IOM-00528-0 12-17 Table of Contents Safety Information... 4 General Safety Precautions...

More information

Model PSI Compressor with 3-Gallon Air Tank 12VDC

Model PSI Compressor with 3-Gallon Air Tank 12VDC Model 6350 150 PSI Compressor with 3-Gallon Air Tank 12VDC IMPORTANT: It is essential that you and any other operator of this product read and understandd the contents of this manual before installing

More information

Oxford Instruments Plasma Technology. PlasmaPro NGP80. Installation Data Sheet. Original Instructions.

Oxford Instruments Plasma Technology. PlasmaPro NGP80. Installation Data Sheet. Original Instructions. Oxford Instruments Plasma Technology www.oxford-instruments.com About this data sheet This data sheet provides basic information about the installation of a tool. Floor and wall loadings Table 1 lists

More information

xendos 1800 Oxygen Analyser The xendos 1800 is a stable, accurate and highly specific oxygen analyser for safe areas.

xendos 1800 Oxygen Analyser The xendos 1800 is a stable, accurate and highly specific oxygen analyser for safe areas. xendos 1800 Oxygen Analyser The xendos 1800 is a stable, accurate and highly specific oxygen analyser for safe areas. Designed for safe area oxygen analysis Low maintenance and re-calibration requirements

More information

Electro-Pneumatic Converter YT-940 SERIES

Electro-Pneumatic Converter YT-940 SERIES Electro-Pneumatic Converter YT-940 SERIES PRODUCT MANUAL VERSION 1.00 Contents 1. Introduction 3 1.1 General information for the users. 3 1.2 Manufacturer Warranty 3 1.3 Explosion Proof Warning. 4 2. Product

More information

Pressure Dump Valve Service Kit for Series 3000 Units

Pressure Dump Valve Service Kit for Series 3000 Units Instruction Sheet Pressure Dump Valve Service Kit for Series 000 Units. Overview The Nordson pressure dump valve is used to relieve hydraulic pressure instantly in Series 00, 400, 500, and 700 applicator

More information

VACUUM REGULATORS CONTENTS

VACUUM REGULATORS CONTENTS CAD drawing data catalog is available. ACCESSORIES GENERAL CATALOG AIR TREATMENT, AUXILIARY, VACUUM, AND FLUORORESIN PRODUCTS CONTENTS Small Regulators Features 759 Specifications, Order Codes, Flow Rate

More information

UL Listed GM-20/25 and GM-40/45 Series Dual-block Multi-function Gas Control Valves

UL Listed GM-20/25 and GM-40/45 Series Dual-block Multi-function Gas Control Valves Installation Sheets Manual 121 Gas Combustion Combination Controls and Systems Section G Technical Bulletin GM-20/25 GM-40/45 Issue Date 1298 UL Listed GM-20/25 and GM-40/45 Series Dual-block Multi-function

More information

IP2, IP2A and IP2D. Spirax-Monnier International Compressed Air Filter/ Regulators

IP2, IP2A and IP2D. Spirax-Monnier International Compressed Air Filter/ Regulators Page 1 of 6 TI-P510-01 H Issue 5 IP2, IP2 and IP2D ISO 9001 Spirax-Monnier International ompressed ir Filter/ Regulators ert. No. LRQ 0963008 Description The IP range of Spirax-Monnier international compressed

More information

IECEx Certificate of Conformity Annexe

IECEx Certificate of Conformity Annexe for Certificate No.: IECEx TSA 09.004X Issue No.: 1 Equipment description pertaining to Issue 0 of this Certificate: A range of electrical power distribution, control, protection and monitoring equipment

More information

RPS900W Redundant Power Supply. Installation Guide.

RPS900W Redundant Power Supply. Installation Guide. RPS900W Redundant Power Supply Installation Guide www.edge-core.com Installation Guide RPS900W Redundant Power Supply Single DC Output Port with Dual Output Voltages RPS900W E10013-CS-R01 1500000081A

More information

Type Pressure Regulator. for increased air capacity. Fig. 1: Type Pressure Regulator. Mounting and Operating Instructions EB EN

Type Pressure Regulator. for increased air capacity. Fig. 1: Type Pressure Regulator. Mounting and Operating Instructions EB EN Type 4708-45 Pressure Regulator for increased air capacity Fig. 1: Type 4708-45 Pressure Regulator Mounting and Operating Instructions EB 8546-1 EN Edition March 2010 Definition of the signal words used

More information

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight Technical Specifications Operating Instructions Maintenance Information Troubleshooting Guide Parts Diagrams AC1810 / AC1810-A THE EVOLUTION OF PERFECTION CAUTION: Before attempting to use or service this

More information

Dual Solenoid Gas Valve Installation

Dual Solenoid Gas Valve Installation Installation IMPORTANT: These instructions are intended as a guide for qualified personnel installing or servicing FLYNN Gas Products. Carefully follow all instructions in this bulletin and all instructions

More information

CSA Sample Draw Aspirator Adapter Operator s Manual

CSA Sample Draw Aspirator Adapter Operator s Manual 30-0951-CSA Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0367 Revision: 0 Released: 4/30/15 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating

More information

AUTOMATIC GAS MANIFOLDS

AUTOMATIC GAS MANIFOLDS AUTOMATIC GAS MANIFOLDS Phoenix Pipeline Products Limited. Unit 8, McKenzie Industrial Park, Tel No.: 44 (0) 161 428 7200 Bird Hall Lane, Fax No.: 44 (0) 161 428 7010 Stockport, Email: info@p3-phoenix.com

More information

GFM. Typical Aluminum GFM Mass Flow Meter NIST MASS FLOW METERS. Principles of Operation. Design Features

GFM. Typical Aluminum GFM Mass Flow Meter NIST MASS FLOW METERS. Principles of Operation. Design Features Design Features Rigid metallic construction. Maximum pressure of 1000 psig (70 bars). Leak integrity 1 x 10-7 of helium. NIST traceable certification. Built-in tiltable LCD readout. 0-5 Vdc and 4-20 ma

More information

Plasma. Argon Production Helium Liquefaction Truck Loading Pure Gas Bottling

Plasma. Argon Production Helium Liquefaction Truck Loading Pure Gas Bottling SERVOPRO Plasma The SERVOPRO Plasma is a Plasma Emission Detector based analyser designed to continuously measure Nitrogen in pure Argon for process and quality control in Argon production within Cryogenic

More information

OPERATING INSTRUCTIONS MANUAL FOR QACV PNEUMATIC DOSING PUMP

OPERATING INSTRUCTIONS MANUAL FOR QACV PNEUMATIC DOSING PUMP This operating instructions contains safety information that if ignored can endanger life or result in serious injury. They are indicated by this icon. Use of this pump with radioactive chemicals is forbidden!

More information

B84G - Excelon Plus Modular System General purpose filter/regulator

B84G - Excelon Plus Modular System General purpose filter/regulator > Port size: 3/8" to 3/4" (PTF/ISO G) > Excelon Plus design allows in-line installation or modular installation with other Excelon Plus products > Push to lock adjusting knob with built in tamper resistant

More information

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features ANALOG MASS FLOW CONTROLLERS Model AF mass fl ow controllers are designed to indicate fl ow rates and control set fl ow rates of gases. Each of these units incorporates an advanced straight tube sensor

More information

Precision and Instrument Regulators 1/4" port size

Precision and Instrument Regulators 1/4 port size Section 8 Precision and Instrument Regulators /" port size Contents -8 Precision Air Pressure Regulator /" ports............ale-8-2 R38 Instrument Regulator Aluminum Model /" ports.........ale-8- -8 R38

More information

B74G - EXCELON Modular System General purpose filter/regulator

B74G - EXCELON Modular System General purpose filter/regulator > Port size: /8... / (ISO G/PTF) > Excelon design allows in-line installation or modular installation with other Excelon products > Push to lock adjusting knob with optional tamper resistant accessory

More information

INSTRUCTION MANUAL FOR MODEL 7360V VERTICAL CURING CHAMBER Revision E May

INSTRUCTION MANUAL FOR MODEL 7360V VERTICAL CURING CHAMBER Revision E May INSTRUCTION MANUAL FOR MODEL 7360V VERTICAL CURING CHAMBER Revision E May 2015 98-0520 S/N 2001 N. Indianwood Ave. Tulsa, Oklahoma 74012 U.S.A. TEL: (918) 250-7200 FAX: (918) 459-0165 E-mail: chandler.sales@ametek.com

More information

Digital Melting Point Apparatus

Digital Melting Point Apparatus Digital Melting Point Apparatus Heating Plateau Ramping Start/Stop Plateau set Ramp stop Hold User Guide Version 1.1 Heating Viewing tube Sample Chamber IEC power inlet socket Power on/off Temperature

More information

36E DSI, HSI & Proven Pilot Two-Stage Combination Gas Valve INSTALLATION INSTRUCTIONS

36E DSI, HSI & Proven Pilot Two-Stage Combination Gas Valve INSTALLATION INSTRUCTIONS INLET PRESS TAP WTE-RODGERS 36E96-314 DSI, HSI & Proven Pilot Two-Stage Combination Gas Valve INSTALLATION INSTRUCTIONS Operator: Save these instructions for future use! FAILURE TO READ AND FOLLOW ALL

More information

Dri-Line Mk3 Monnier Compressed Air Drain Trap

Dri-Line Mk3 Monnier Compressed Air Drain Trap 5044050/2 IM-P504-24 CH Issue 2 Dri-Line Mk3 Monnier Compressed Air Drain Trap Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation and Operation

More information

RK-IR Sample Draw Aspirator Adapter Operator s Manual

RK-IR Sample Draw Aspirator Adapter Operator s Manual 30-0951RK-IR Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0018RK Revision: A Released: 6/2/10 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment

More information

OPERATION MANUAL NTF-15

OPERATION MANUAL NTF-15 OPERATION MANUAL NTF-15 Nitrogen Tire Filling Valve Stem Caps (Qty=200) Order P/N 436075 RTI Technologies, Inc 10 Innovation Drive York, PA 17402 800-468-2321 www.rtitech.com 035-81235-00 (Rev B) TABLE

More information

Series K Air service units G 1/8, G 1/4

Series K Air service units G 1/8, G 1/4 Air service units G /8, G /4 Series FK Filter, G /8 G /4.02 Series RK Pressure regulator, G /8 G /4.04 Series FRK Filter/regulator, G /8 G /4.06 Series OK Lubricator, G /8 G /4.08 Series FROK Two piece

More information

ML3M Monnier Miniature Compressed Air Lubricator

ML3M Monnier Miniature Compressed Air Lubricator 5041050/3 IM-P504-04 CH Issue 3 ML3M Monnier Miniature Compressed Air Lubricator Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation and commissioning

More information

Pressure Equalising Valves Model , for Differential Pressure Gauges

Pressure Equalising Valves Model , for Differential Pressure Gauges Accessories Pressure Equalising Valves Model 910.25, for Differential Pressure Gauges WIKA Data Sheet AC 09.11 Applications T Isolating, pressure equalising as well as purging and air bleeding valves to

More information

ABB MEASUREMENT & ANALYTICS DATA SHEET. Endura AZ40 Oxygen and carbon monoxide equivalent (COe) analyzer

ABB MEASUREMENT & ANALYTICS DATA SHEET. Endura AZ40 Oxygen and carbon monoxide equivalent (COe) analyzer ABB MEASUREMENT & ANALYTICS DATA SHEET Endura AZ40 Oxygen and carbon monoxide equivalent (COe) analyzer 2 E N D U R A AZ 40 OX YG EN AN D CARBON MONOXIDE E QU IVALE NT (COE ) ANALY ZE R DS/AZ40-E N RE

More information

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model 50 Systems Outfitted with ABM-725 Monitor M A R T E C H S E R V I C E S C O M P A N Y OFFICE: (507) 843-4700 P.O. BOX 7079 Toll

More information

INSTALLATION & MAINTENANCE INSTRUCTIONS

INSTALLATION & MAINTENANCE INSTRUCTIONS DESCRIPTION / IDENTIFICATION The QBX series valve uses Proportion-Air closed loop technology for Pressure control. It gives an output pressure proportional to an electrical command signal input. The QB1X

More information

TECHNICAL DATA. Trimpac 257a. September 16, 2013

TECHNICAL DATA. Trimpac 257a. September 16, 2013 September 16, 2013 Trimpac 257a 1. DESCrIpTION DESCRIPTION The Viking Double SUREFIRE SIngle Interlock Preaction TRIMPAC Model D-2 and D-2B used with either a Model E or F Deluge Valve (A.1), a Viking

More information

Features. Main Voltage: Vdc. Input Control Signal: 0-5 Vdc or 4-20 madc (Sourcing) Monitor Output Voltage: 0-5 Vdc or 4-20 madc (Sourcing)

Features. Main Voltage: Vdc. Input Control Signal: 0-5 Vdc or 4-20 madc (Sourcing) Monitor Output Voltage: 0-5 Vdc or 4-20 madc (Sourcing) 1 X-Flow Mass Flow Controller Flow Range from 0.8 ml/min to 20 l/min Target Markets Laboratory and Process Instrumentation Pharma / Bio-Pharma Equipment Air Quality Monitoring Systems Furnace and Coatings

More information

PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING

PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING Please Note: Ranges above 500mbar are designed and manufactured in accordance with sound engineering

More information

Vertex VP Operation and Maintenance

Vertex VP Operation and Maintenance AIRBENCH LTD Vertex VP Operation and Maintenance This Edition: 2 October 2018 AirBench Ltd, 14 Grange Farm Road, Colchester CO2 8JW www.airbench.com +44 (0) 1206 791191 Copyright AirBench Ltd 2016 Page

More information

D10S/D20S Wall/Post Mount Inflator Quick Start Manual

D10S/D20S Wall/Post Mount Inflator Quick Start Manual PART NUMBER SERIAL NUMBER D10S/D20S Wall/Post Mount Inflator Quick Start Manual Please read and save these instructions. Read carefully before attempting to assemble, install, operate or maintain the product

More information

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual Installation, Operation, and Maintenance Manual Welker Sampler Model GSS-1 The information in this manual has been carefully checked for accuracy and is intended to be used as a guide to operations. Correct

More information

B84G - Excelon Plus Modular System General purpose filter/regulator

B84G - Excelon Plus Modular System General purpose filter/regulator > Port size: 3/8... 3/4 (ISO G/PTF) > Excelon Plus design allows in-line installation or modular installation with other Excelon Plus products > Push to lock adjusting knob with built in tamper resistant

More information