Reference method for the enumeration of erythrocytes and leucocytes

Size: px
Start display at page:

Download "Reference method for the enumeration of erythrocytes and leucocytes"

Transcription

1 Clin. lab. Haemat , Reference method for the enumeration of erythrocytes and leucocytes INTERNATIONAL COUNCIL FOR STANDARDIZATION IN HAEMATOLOGY; PREPARED BY THE EXPERT PANEL ON CYTOMETRY Members: J.M. England (Chairman, UK), R.M. Rowan (Secretary, UK), 0. W. van Assendelft (USA), B.S. Bull (USA), W.H. Coulter (USA), K. Fujimoto (Japan), W. Groner (USA), A.R. Jones (USA), J.A. Koepke (USA), S.M. Lewis (UK), N. Tatsumi (Japan) & R.L. Verwilghen (Belgium). Consultant: C.E. McLaren (USA) Accepted for publication 14 January 1994 Introduction The International Council for Standardization in Haematology (ICSH) defines a Reference Method (1991) as a clearly and exactly described technique for a particular determination which, in the opinion of a defined authority, provides sufficiently accurate and precise laboratory data for it to be used to assess the validity of other laboratory methods for this determination. The accuracy of the reference method must be established by comparison with a definitive method where one exists. A reference method should be traceable to a primary metrological standard and the degree of inaccuracy and imprecision must be stated. Accurate measurement, traceable to primary metrological standards, can be obtained for Hb and PCV by means of specified reference procedures defined by the International Council for Standardization in Haematology (ICSH 1980; 1987). Accurate measurements for other parameters of the blood count have been less easy to achieve. The traditional methods for red and white cell counting have been based on direct haemocytometry of diluted blood in a counting chamber, however, the imprecision and inaccuracy of this technique renders it unsuitable as a reference method. ICSH has recommended that chamber counting be replaced by a single channel semi-automated electronic counter using the aperture-impedance principle (ICSH 1988) provided that a known volume of diluted blood is displaced through the orifice. However this procedure too has recognized sources of error amongst which are inaccurate displacement volume, recirculation of cells inside the orifice tube, cell adherence to the counting vial and flow pathway, inclusion of signals due to artefact and coincidence counting., Correspondence: Dr R.M. Rowan, Department of Haematology, Western Infirmary, Dumbarton Road, Glasgow, GI1 6NT, Scotland, UK. 131

2 132 ICSH Expert Panel on Cytometry Coincidence remains an important source of error unless correction is applied. Several formulae have been proposed for coincidence correction, and these have either been integrated into the instrument counting system or used to prepare correction tables. Most of these formulae assume that coincidence has a constant relationship at a particular orifice size and is unaffected by other factors such as cell volume or flexibility. Recently Lewis et al. (1989) described a method for accurate correction of red blood cell count for coincidence in aperture-impedance electronic blood cell counters which was based on extrapolation using counts obtained with several dilutions of the sample. This document extends the previous recommendations (ICSH 1988) to include more accurate coincidence corrections. BLOOD SAMPLES Fresh venous blood specimens should be obtained by syringe or evacuated container. Specimens should be rejected if there is visible haemolysis or if microclots are present. Specimens should be anticoagulated with K, EDTA (ICSH 1993) in a final concentration of pmol ( mg/ml) and placed in containers which allow sufficient air remaining to enable proper mixing. Specimens should be maintained at 20 2 C until processing. Ideally the period between venous sampling and processing should not exceed 4 h. Before testing the specimen should be well mixed by gently inverting the container. The number of inversions of a container, to achieve adequate homogeneity, will depend on the type and dimensions of the container. Standard evacuated containers, x 75 mm, containing 5 ml blood require an air bubble comprising at least 20% of the tube volume. Non-standard tubes, particularly narrower tubes require more complete inversions (NCCLS 1993). Vortex mixers should not be used. GLASSWARE Pipettes Positive displacement pipettes should be used, calibrated to an accuracy of +0.5%, validated by a process traceable to a primary metrological standard (Appendix). Volumetric flasks Grade A volumetric flasks made of borosilicate glass, should be used, each with a stated volume which has been tested by an appropriate national standards body, or which has been checked by the individual user by means of a gravimetric method using certified weights and corrections for buoyancy and temperature (Appendix). Counting vials The counting vial should have a minimum volume of 10ml. It should be of sufficient height that the entry orifice of the electronic cell counter is at approxi-

3 Reference method for enumeration of erythrocytes and leucocytes 133 mately half the depth of the fluid before counting commences and there should be more than 1 cm of fluid above that orifice after the completion of counting. Before use, the vial should be cleaned free of chemical contaminants and adventitious particles. It is necessary to confirm that cells do not adhere to the counting vial and cause a progressive fall in the count. This is ascertained by testing a number of vials from each batch. Counts are made at measured intervals after the dilutions have been placed in the counting vials (see below). The dilution should be mixed in the vial before counting. INSTRUMENT SPECIFICATIONS The design of the electronic cell counting instrument should be such that each cell is counted once and only once. The probability of re-circulation of particles through the sensing zone must be low and any impulses so created should be below the appropriate lower threshold. The detection electronics should have the ability to discriminate, with a low probability of error, by pulse amplitude between the cells of interest, other cells and electrical noise. Although in principle these criteria may be met with optical sensing counters, at present, only semi-automated single channel aperture-impedance particle counters are available. The instrument should have an orifice diameter pm and length % of the diameter. The volume displaced during the counting period must be known to within an accuracy of 1 % traceable to a national or international metrological standard. This displacement can be achieved by the motion of a mercury column or other suitable means. The differential displacement due to thermal effects should not be greater than 0.1% per "C. To avoid artefactual signals, ICSH recommends threshold settings which ensure that the signal to noise ratio is greater than 100: 1. REAGENTS Diluent The diluent must be a sterile, non-toxic, buffered salt solution suitable for the measurement system. Its tonicity should be stated and should fall in the range 280f 15 mosms. The constituents and their concentrations must be given on the label of the container. When counted in the vial it must contain less than 1 x lo5 particles per litre measurable at the threshold setting for the cells which are to be counted. The diluent must not cause cell lysis nor alter cell volume by > 2 fl over a 20 min period. Lytic agent When white blood cells are to be counted in a whole blood sample the red cells must first be lysed. The lytic agent must not affect the ability of the white cells to

4 134 ICSH Expert Panel on Cytometry be counted whilst the red cell stroma must be reduced to a residue which does not contribute to the white cell count. Flushing solution The flushing solution should consist of diluent degassed by vacuum or heating to 90 C. METHOD FOR RED CELL COUNT I. Flushing Immediately before carrying out each individual count, flush the instrument through with flushing solution. 2. Dilution Coincidence correction should be performed for every blood specimen which is to be counted. Make two primary dilutions (each of 0.1 ml blood plus 20 ml diluent). Take 0.02, 0.04 and 0.08ml of the primary dilutions and add these to 20ml aliquots of diluent to make two sets of four secondary dilutions, the most concentrated of these, i.e., ml, having a dilution factor of 1/ Each secondary dilution is then counted as follows: Secondary dilutions No. of counts Transfer to counting vial The diluted sample must be transferred to counting vials and gently mixed for about 30 s to get rid of bubbles. The count must be completed within 5 min of performing the dilution. 4. Counting procedure It is expected that thresholding will be verified as described below on a sample-bysample basis. Ideally 0.5ml should be drawn by displacement through the sampling aperture within s. The time required should be in conformity with that recommended by the manufacturer. This check is designed to recognize partial blockage of the aperture. The counts should be replicated as explained above and the replicate counts in each secondary dilution summated. 5. Validating thresholds Using a pulse height analyser the lower counting threshold is set in the valley between platelets and red cells. The count per channel measured at this position can be compared with the count per channel at a position corresponding to the red cell modal volume. Provided the count channel in the latter position is at least 50 times that of the former, the lower threshold is deemed to be satisfactory. The

5 Reference method for enumeration of erythrocytes and leucocytes 135 reference method is inapplicable if this criterion does not apply. The upper threshold is then set to the maximum possible value (it should be remembered that this will include leucocytes as well as red cells). 6. Correction for coincidence A correction for coincidence is made by using regression analysis with analysis of variance to check for non-linearity of the regression; the data points for the analysis are derived from the two sets of four secondary dilutions (Figure 1). The y values are summated counts at each level of secondary dilution and the x values are given in an arbitrary scale on which the most concentrated secondary dilution ( ml) is given the value of 1.0. Secondary dilution (ml) x value oooo If the anaiysis of variance fails to show evidence of non-linearity, the intercept of the regression line represents the coincidence corrected count (Lewis et al. 1989). This count gives the number of red cells aspirated through the orifice tube at the dilution level of the most concentrated secondary dilution ( ml). Divide this result by three because of the three replicate counts. Express the results as counts per ml of secondary dilution knowing the volume of secondary dilution aspirated through the orifice. Finally, multiply this count by to obtain the RBC. Summated count i Coincidence corrected count I I I I o Arbitrary scale Figure 1. Principle of coincidence correction using summation of replicated counts on two sets of four secondary dilutions. -X

6 136 ICSH Expert Panel on Cytometry METHOD FOR WHITE CELL COUNT The principles described above also apply for white cell counting. Make two dilutions, each of 0.08 ml blood plus 20 ml diluent. Prior to counting add the lytic agent. Count the white cells as soon as the red cell stroma has been reduced to a residue which does not contribute to the white cell count but before the ability to count the white cells has been affected. The same orifice can be used as for red cell counting. Validation of thresholds This follows the same principle as for RBC with the lower counting threshold placed between the noise from red cell stroma and leucocyte signals. Coincidence correction This is performed as for red cell counting save that four primary dilutions of whole blood in diluent are required: ml ml ml ml Analyse the data as explained above; the intercept of the regression line represents the coincidence corrected count for the most concentrated primary dilution ( ml). Correct for replication and for volume aspirated through the orifice as explained above. Also correct for dilution by the lytic agent. Finally, multiply this count by 251 to obtain the WBC. ANALYSIS OF ERROR The maximum permissible bias is 2.0% for red cell counts and 4.0% for white cell counts. Potential sources of error are listed below and apply to both red and white cell counting. Sampling error This error arises if the measured sample is not representative of the original whole blood specimen. Factors to be considered include poor mixing, haemorheological effects in extracting the sample and the inaccuracy in dilution. Transport error Errors due to transport occur if each and every cell is not counted in a known volume of diluted blood. Such errors are caused by sedimentation, cell loss and/or recirculation, as well as the inaccuracy of the displaced volume. Counting error Errors can arise if the final count is distorted because of improper discrimination, spurious counts or inaccurate correction for coincidence loss. The imprecision of

7 Reference method for enumeration of erythrocytes and leucocytes 137 the reference cell count should be no greater than 1%. Potential sources of error include extracting the sample, dilution, the imprecision of the displaced volume and the statistical limitation of the number of cells counted. References ICSH (1980) Recommendation for reference method for determination by centrifugation of packed cell volume of blood. J. Clzn. Pathol. 33, 1-2 ICSH (1987) Recommendation for reference method for haemoglobinometry in human blood (ICSH) Standard (1986) and specification for international haemiglobincyanide reference preparation (Third Edition). Clin. lab. Haemat. 9, ICSH (1988) The assignment of values to fresh blood used for calibrating automated blood cell counters. Clin. lab. Haemat. 10, ICSH (1991) Rules and Operating Procedures, p. 8, ICSH Secretariat, Leuven, Belgium. ICSH (1993) Recommendations of the International Council for Standardization in Haematology for Ethylenediaminetetraacetic Acid Anticoagulation of Blood for Blood Cell Counting and Sizing. Am. J. Clin. Pathol. 100, LEWIS S.M., ENGLAND J.M. & KUBOTA F. (1989) Coincidence correction in red blood cell counting. Phys. Med. Bid. 34, NCCLS (1993) Methods for the Erythrocyte Sedimentation Rate Test. Third edition. Candidate Standard for Approval. NCCLS Document H2-A3

8 138 ICSH Expert Panel on Cytometry Appendix 1. CALIBRATION OF VOLUMETRIC PIPETTES The pipette is filled to the calibration mark with distilled water, which is then transferred to a weighed beaker in accordancc with the normal usage of the pipette. Care must be taken to avoid evaporation. The beaker is reweighed. To achieve imprecision < 1% on a 10 ml pipette requires weighing which is reliable to 0.1 mg. Thus, the analytic balance should be stable for measurement at 0.01 mg, i.e. at the 1 pg level. If this is too stringent, then at least a 5 pg level is required. The ambient temperature is noted. The volume of the pipette (in ml) is calculated by dividing the weight of the water (in g) by one of the following factors depending on temperature: Temperature ("C) Factor The calibration must be performed in duplicate for each pipette. In calibrating a 20 pl pipette by weighing, imprecision should not exceed 1 %. 2. CALIBRATION OF VOLUMETRIC FLASKS The pre-weighed flask is filled to the calibration mark with distilled water and re-weighed. The ambient temperature is noted and the calculation carried out as in section 1.

ICSH recommendations for measurement of erythrocyte sedimentation rate

ICSH recommendations for measurement of erythrocyte sedimentation rate 1989 Clin Pathol 1993;46:198-203 ICSH recommendations for measurement of erythrocyte sedimentation rate Members of Expert Panel: B S Bull (USA), M Caswell (UK), E Ernst (Austria), J M Jou (Spain), A Kallner

More information

CORESTA RECOMMENDED METHOD N 6

CORESTA RECOMMENDED METHOD N 6 CORESTA RECOMMENDED METHOD N 6 DETERMINATION OF VENTILATION DEFINITIONS AND MEASUREMENT PRINCIPLES (2015 Revision September 2016) 1. SCOPE This CORESTA Recommended Method specifies a method for the determination

More information

Info Counting chamber (haemacytometer)

Info Counting chamber (haemacytometer) Info Counting chamber (haemacytometer) 1. What is a counting chamber and what is it used for? A counting chamber is a precision measuring instrument made of special optical glass. It is used to count cells

More information

CORESTA RECOMMENDED METHOD Nº 67

CORESTA RECOMMENDED METHOD Nº 67 CORESTA RECOMMENDED METHOD Nº 67 DETERMINATION OF WATER IN THE MAINSTREAM SMOKE OF CIGARS BY GAS CHROMATOGRAPHIC ANALYSIS (November 2005) 1. FIELD OF APPLICATION The method is applicable to the particulate

More information

Laboratory #2 Pipetting Technique and Micropipette Calibration Skills=40 pts

Laboratory #2 Pipetting Technique and Micropipette Calibration Skills=40 pts Laboratory #2 Pipetting Technique and Micropipette Calibration Skills=40 pts Objectives: Upon completion of this unit, the student should be able to: 1. List and describe 3 categories of pipets. 2. List

More information

VOLUMETRIC TECHNIQUES

VOLUMETRIC TECHNIQUES REVISED 10/14 CHEMISTRY 1101L VOLUMETRIC TECHNIQUES Volume measurements are important in many experimental procedures. Sometimes volume measurements must be exact; other times they can be approximate.

More information

DXH Troubleshooting FlowCharts

DXH Troubleshooting FlowCharts DXH Troubleshooting FlowCharts WARNINGS AND PRECAUTIONS Read all product manuals and consult with Beckman Coulter-trained personnel before attempting to operate instrument. Beckman Coulter, INC., urges

More information

Technical Procedure for General Laboratory Equipment

Technical Procedure for General Laboratory Equipment Technical Procedure for General Laboratory Equipment 1.0 Purpose - This procedure specifies the required elements for the use of general laboratory equipment. 2.0 Scope - This procedure applies to all

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 8655-4 First edition 2002-09-15 Piston-operated volumetric apparatus Part 4: Dilutors Appareils volumétriques à piston Partie 4: Diluteurs Reference number ISO 8655-4:2002(E)

More information

Quantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs

Quantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs Quantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs Gravimetric Sample Preparation The Alternative to Volumetric Flasks From Volumetric to Gravimetric

More information

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service? Pipette specifications according to ISO 8655 Chapter 5 5.1 The ISO 8655 standard gives the accuracy and precision limits as both absolute and relative values. Specifications will depend on the technique

More information

Tips for Proper Pipette handling and Maintenance. D. Muruganand, Ph.D ILQA - Bangalore, Feb 2014

Tips for Proper Pipette handling and Maintenance. D. Muruganand, Ph.D ILQA - Bangalore, Feb 2014 Tips for Proper Pipette handling and Maintenance D. Muruganand, Ph.D ILQA - Bangalore, Feb 2014 Liquid handling tools Manual pipettes Electronic pipette and dispenser Automated systems Principles of Pipetting

More information

CHM250 Calibration and Measurement Lab. Balance Calibration

CHM250 Calibration and Measurement Lab. Balance Calibration CHM250 Calibration and Measurement Lab Green Profile Balance Calibration Introduction: Balances that are properly operated, calibrated and maintained are crucial for laboratory operations. The accuracy

More information

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries. Informal Commentary Limit of Diethylene Glycol (DEG) and Ethylene Glycol (EG) in Sorbitol Solution, Sorbitol Sorbitan Solution and Noncrystallizing Sorbitol Solution December 2009 Monograph/Section(s):

More information

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay Summary This assay measures Human Granulocyte Colony Stimulating Factor (G-CSF) in a 96-well

More information

PERFORMANCE TEST INSTRUCTIONS FOR BIOHIT PROLINE PIPETTORS

PERFORMANCE TEST INSTRUCTIONS FOR BIOHIT PROLINE PIPETTORS PERFORMANCE TEST INSTRUCTIONS FOR BIOHIT PROLINE PIPETTORS VER3150999 1 Content: 1.0 Introduction...2 2.0 Existing standards....3 3.0 Biohit gravimetric test process, flow chart 4 4.0 The Calibration Service.

More information

<788> PARTICULATE MATTER IN INJECTIONS

<788> PARTICULATE MATTER IN INJECTIONS PARTICULATE MATTER IN INJECTIONS This general chapter is harmonized with the corresponding texts of the European Pharmacopoeia and/or the Japanese Pharmacopoeia. These pharmacopeias have undertaken

More information

Standard Operating Procedure. Air Displacement Pipette Calibration

Standard Operating Procedure. Air Displacement Pipette Calibration University of Colorado at Denver October 28, 2003 - Revision 1.00 Page 1 of 7 1 Background: Standard Operating Procedure An accurate pipette is one of the most important tools in performing accurate analytical

More information

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010 Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600 November 2010 Table of Contents ``Abstract...3 ``Introduction...4 ``Methods & Results...6 ``Data Analysis...9 ``Conclusion...12

More information

VIAFLO 96. Electronic 96 Channel Hand Held Pipettor

VIAFLO 96. Electronic 96 Channel Hand Held Pipettor VIAFLO 96 Electronic 96 Channel Hand Held Pipettor VIAFLO 96 Unique Multichannel microplate pipetting is a growing task because of the need to increase productivity and improve the quality of the data

More information

1. Scope: This document describes the procedure for verifying the performance of vaccine vial monitors.

1. Scope: This document describes the procedure for verifying the performance of vaccine vial monitors. PQS Independent type-testing protocol WHO/PQS/E06/IN05.VP.2 Original: English Distribution: General TITLE: Vaccine Vial Monitor Product verification protocol: E06/IN05.VP.2 Applies to specification ref(s):

More information

INTRODUCTION TO THE SPECTROPHOTOMETER AND PIPETTING SKILLS

INTRODUCTION TO THE SPECTROPHOTOMETER AND PIPETTING SKILLS INTRODUCTION TO THE SPECTROPHOTOMETER AND PIPETTING SKILLS Section A: Intro to the spectrophotometer A commonly used instrument in the analysis of cellular extracts is the Spectrophotometer. Today you

More information

Monosed ESR Non-Vacuum Tube USER S MANUAL

Monosed ESR Non-Vacuum Tube USER S MANUAL Monosed ESR Non-Vacuum Tube USER S MANUAL Manual code MAN-134 Revision 02 Revised date: November 2015 Instructions for use for in vitro diagnostic instruments for professional use. PRODUCT NAME PRODUCT

More information

biosensis Human IGF-II, Insulin-like growth factor II, Somatomedin-A ELISA Kit Protocol

biosensis Human IGF-II, Insulin-like growth factor II, Somatomedin-A ELISA Kit Protocol biosensis Human IGF-II, Insulin-like growth factor II, Somatomedin-A ELISA Kit Protocol Catalog No: BEK-2029-1P For quantitative detection of human IGF-II in cell culture supernatants, cell lysates, tissue

More information

Requirements and Methods for Calibration of Equipment in Clinical Laboratories. NK Chan (HKRCBTS) 24 August 2007

Requirements and Methods for Calibration of Equipment in Clinical Laboratories. NK Chan (HKRCBTS) 24 August 2007 Requirements and Methods for Calibration of Equipment in Clinical Laboratories NK Chan (HKRCBTS) 24 August 2007 1 Contents Basic Requirements of Equipment Calibration Calibration of Electronic Balance

More information

METHOD 204F--VOLATILE ORGANIC COMPOUNDS CONTENT IN LIQUID INPUT STREAM (DISTILLATION APPROACH) 1.1 Applicability. This procedure is applicable for

METHOD 204F--VOLATILE ORGANIC COMPOUNDS CONTENT IN LIQUID INPUT STREAM (DISTILLATION APPROACH) 1.1 Applicability. This procedure is applicable for METHOD 204F--VOLATILE ORGANIC COMPOUNDS CONTENT IN LIQUID INPUT STREAM (DISTILLATION APPROACH) 1. INTRODUCTION 1.1 Applicability. This procedure is applicable for determining the input of volatile organic

More information

Any laboratory is equipped with specific tools, equipment,

Any laboratory is equipped with specific tools, equipment, Use of Laboratory Equipment and Supplies 3 When you have completed this exercise, you will be able to: 1. Use a balance. 2. Use pipettes and graduated cylinders to measure the volume of liquids. 3. Use

More information

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (09) 64 R5 QUALIFICATION OF EQUIPMENT ANNEX 6: QUALIFICATION OF PISTON PIPETTES Full document title and reference Document type

More information

RayPette Plus Autoclavable Pipette. User Manual

RayPette Plus Autoclavable Pipette. User Manual RayPette Plus Autoclavable Pipette User Manual CONTENTS 1. YOUR NEW PIPETTE... 1 1.1. Adjustable volume pipettes... 1 1.3 Fully autoclavable... 2 2. UNPACKING... 3 3. INSTALLING THE PIPETTE HOLDER... 4

More information

Standard Operating Procedure (SOP) SOP-WP5-002/Version 1.0

Standard Operating Procedure (SOP) SOP-WP5-002/Version 1.0 Page 1 of 12 (SOP) SOP-WP5-002/Version 1.0 Title: Manual blood processing Validity: valid after: valid until: insert end date Replaced versions : insert previous version Modification note: insert modifications

More information

Contact Person(s) : Isabel M. Fisenne APPLICATION

Contact Person(s) : Isabel M. Fisenne APPLICATION Ra-02-RC RADIUM-226 - EMANATION PROCEDURE Contact Person(s) : Isabel M. Fisenne APPLICATION This procedure is specific and may be applied to almost any matrix which can be converted to a homogeneous solution.

More information

CH2250: Techniques in Laboratory Chemistry. Outline Measuring Mass Measuring Volume Significant figures. Mass Measurement

CH2250: Techniques in Laboratory Chemistry. Outline Measuring Mass Measuring Volume Significant figures. Mass Measurement CH2250: Techniques in Laboratory Chemistry Outline Measuring Mass Measuring Volume Significant figures Mass Measurement Mass Measurement Measure mass not weight Mass is measured with a balance (a scale

More information

Laboratory Activity Measurement and Density. Average deviation = Sum of absolute values of all deviations Number of trials

Laboratory Activity Measurement and Density. Average deviation = Sum of absolute values of all deviations Number of trials Laboratory Activity Measurement and Density Background: Measurements of mass and volume are very common in the chemistry laboratory. The analytical balance is used to measure mass, and the graduated cylinder,

More information

Calibration of Volumetric Glassware

Calibration of Volumetric Glassware Calibration of Volumetric Glassware Introduction This set of laboratory experiments is designed to introduce you to some of the apparatus and operations you will be using during the remainder of this course,

More information

Performance Verification of BioTek s Precision 2000 using Artel s Multichannel Volume Verification (MVV ) System

Performance Verification of BioTek s Precision 2000 using Artel s Multichannel Volume Verification (MVV ) System Performance Verification of BioTek s Precision 2000 using Artel s Multichannel Volume Verification (MVV ) System Authors: Paul Held 1, John Bradshaw 2, Jason Greene 1, Alex Rogers 2, and Tanya Knaide 2

More information

The pipettes cover a volume range from 0.1µl to 10ml.

The pipettes cover a volume range from 0.1µl to 10ml. CONTENTS 1. YOUR NEW PIPETTE 1 1.1. Adjustable volume pipettes 1 1.2. Fixed volume pipettes 2 1.3 Fully autoclavable 3 2. UNPACKING 3 3. INSTALLING THE PIPETTE HOLDER 4 4. PIPETTE COMPONENTS 5 5. PIPETTE

More information

The use of the analytical balance, and the buret.

The use of the analytical balance, and the buret. 1211L Experiment 1. Density 2015 by H. Patterson Instructor Notes: Students make measurements individually then share data to make the graph. There are four volumetric measurements to be studied; 3.00

More information

Pipettor. User Manual

Pipettor. User Manual Pipettor User Manual Product Code Description Single Chanel Pipettor Variable 550.002.055 Volume 0.5 to 10ul 550.002.060 2 to 20ul 550.002.065 10 to 100ul 550.002.070 20 to 200ul 550.002.075 100 to 100ul

More information

Pipetting and Determining Protein Concentration

Pipetting and Determining Protein Concentration Pipetting and Determining Protein Concentration Background Information: When performing experiments in Cell Biology, it is often necessary to use very small volumes of reagents sometimes because the reagents

More information

DISSOLUTION TEST FOR SOLID DOSAGE FORMS

DISSOLUTION TEST FOR SOLID DOSAGE FORMS Seite 1 von 15 01/2010:20903 corrected 6.8 2.9.3. DISSOLUTION TEST FOR SOLID DOSAGE FORMS This test is provided to determine compliance with the dissolution requirements for solid dosage forms administered

More information

II. MATERIALS REQUIRED FOR PERFORMANCE TESTING

II. MATERIALS REQUIRED FOR PERFORMANCE TESTING TSP-016 P/N L038-058 4-001 Rev. D February 8, 2000 Page 1 of 14 DISTRIBUTION: VistaLab Technologies Customers EQUIPMENT: MLA Pipettes SUBJECT: Pipette Performance Verification Protocol PURPOSE: To evaluate

More information

QAM-I-117 Volumetric Equipment Calibration Verification

QAM-I-117 Volumetric Equipment Calibration Verification 1. Applicability and Purpose This procedure applies to all adjustable and fixed volume pipetters and any labware used to deliver measured volumes of liquid by laboratory analysts at the Texas Institute

More information

A SIMPLE TITRATION METHOD FOR DETERMINING THE SPECIFIC GRAVITY ON ONE DROP OF URINE

A SIMPLE TITRATION METHOD FOR DETERMINING THE SPECIFIC GRAVITY ON ONE DROP OF URINE J. clin. Path. (1951), 4, 491. A SIMPLE TITRATION METHOD FOR DETERMINING THE SPECIFIC GRAVITY ON ONE DROP OF URINE BY From the Pathological Laboratory, the Peace Memorial Hospital, Watford (RECEIVED FOR

More information

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI NanoSight NS300 Operation instructions Injection/flushing brief overview: 1. Do not exceed flow of 1 ml per 20 seconds. 2. Inject two 1 ml syringes with nano-pure or DI water. 3. If the water does not

More information

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist ASTM WK 43267 Standard Test Method for Dissolved Gases Anne Jurek Applications Chemist Rationale Hydraulic Fracturing is becoming more and more common. Nearby water well can be affected during the fracking

More information

25/04/56. LTC Carmen A. Bell Chief, Laboratory Operations USAMC-AFRIMS

25/04/56. LTC Carmen A. Bell Chief, Laboratory Operations USAMC-AFRIMS Pipette Theory and Maintenance LTC Carmen A. Bell Chief, Laboratory Operations USAMC-AFRIMS carmen.bell@afrims.org +6602-696-2725 1 References Pharma Express: http://pharma.financialexpress.com/2008063

More information

Transferpette -8/-12

Transferpette -8/-12 Transferpette -8/-12 Testing Instructions (SOP) May 2009 1. Introduction The standard DIN EN ISO 8655 describes both the design and the testing of the piston operated pipette Transferpette 8/ 12. The following

More information

CHM 317H1S Winter Section P Procedures and Tables

CHM 317H1S Winter Section P Procedures and Tables CHM 317H1S Winter 2018 Section P Procedures and Tables Procedures Page 1 Standard Operating Procedures Throughout the laboratory portion of this course, you will be required to perform a number of operations

More information

Gas mixtures. Individual solutions specifically for your application

Gas mixtures. Individual solutions specifically for your application Gas mixtures Individual solutions specifically for your application For many applications besides pure gases also defined gas mixtures are required. For many routine applications, from operating gases

More information

Ensuring pipette & weighing device performance: Verification & Calibration

Ensuring pipette & weighing device performance: Verification & Calibration Ensuring pipette & weighing device performance: Verification & Calibration Wan Melissa Diyana bt Wan Normazlan Central Research Laboratories, Faculty of Medicine terms What is preventive maintenance? Regularly

More information

MULTICHANNEL. Labnet International, Inc.

MULTICHANNEL. Labnet International, Inc. Labnet International, Inc. A B A1 A2 C F D H E 4A 4B 4C 4D G CONTENTS 1 - INTRODUCTION 2 - VOLUME SETTING 3 - METHOD OF PIPETTING 4 - RECOMMENDATIONS - RECALIBRATION 6 - CLEANING AND STERYLIZATION 7 -

More information

PIPETMAN F ENGLISH. Other languages available on our website:

PIPETMAN F ENGLISH. Other languages available on our website: PIPETMAN F ENGLISH Other languages available on our website: www.gilson.com. PIPETMAN F CONTENTS page 1 - INTRODUCTION 2 2 - SPECIFICATIONS 3 3 - OPERATION 4 4 - USER ADJUSTMENT 5 5 - PRE-RINSING 6 6 -

More information

biosensis Rat Glial cell line-derived neurotrophic factor/gdnf total /ATF ELISA Kit Protocol

biosensis Rat Glial cell line-derived neurotrophic factor/gdnf total /ATF ELISA Kit Protocol biosensis Rat Glial cell line-derived neurotrophic factor/gdnf total /ATF ELISA Kit Protocol Catalog No: BEK-2020-1P For quantitative detection of rat GDNF in cell culture supernatants, cell lysates, tissue

More information

Rat myeloperoxidase-antineutrophil cytoplasmic antibody(mpo-anca) ELISA Kit

Rat myeloperoxidase-antineutrophil cytoplasmic antibody(mpo-anca) ELISA Kit Rat myeloperoxidase-antineutrophil cytoplasmic antibody(mpo-anca) ELISA Kit Catalog Number. CSB-E08675r For the quantitative determination of rat myeloperoxidase-antineutrophil cytoplasmic antibody(mpo-anca)

More information

Validation of Custody Transfer Metering Skid at Site After Laboratory Proving

Validation of Custody Transfer Metering Skid at Site After Laboratory Proving flotek.g 2017- Innovative Solutions in Flow Measurement and Control - Oil, Water and Gas August 28-30, 2017, FCRI, Palakkad, Kerala, India Validation of Custody Transfer Metering Skid at Site After Laboratory

More information

TECHNICAL NOTE. October 23, Procedures for Verifying Temperature Calibration

TECHNICAL NOTE. October 23, Procedures for Verifying Temperature Calibration TECHNICAL NOTE October 23, 2002 SUBJECT: Procedures for Verifying Temperature Calibration PROBLEM: DATATRACE users frequently need a procedure to verify that the Tracer calibration is in specification.

More information

(NIST Special Publication )

(NIST Special Publication ) (NIST Special Publication 260-126) THE NIST TRACEABLE REFERENCE MATERIAL PROGRAM FOR GAS STANDARDS by Franklin R. Guenther, William D. Dorko, Walter R. Miller and George C. Rhoderick Analytical Chemistry

More information

IMPROVING QUALITY THROUGH INNOVATION: A FULLY AUTOMATED SOLUBLE SALT METER

IMPROVING QUALITY THROUGH INNOVATION: A FULLY AUTOMATED SOLUBLE SALT METER ARP INSTRUMENTS, INC. SSM MODEL # RPCT-07-001 IMPROVING QUALITY THROUGH INNOVATION: A FULLY AUTOMATED SOLUBLE SALT METER For more information contact: ARP Instruments, Inc. www.arpinstruments.biz arp.instruments@gmail.com

More information

v /07/17 Page 1 of 18

v /07/17 Page 1 of 18 A. PURPOSE AND SCOPE Details the on-site measurement of free and total chlorine in water samples using the Palintest DPD,GP/HR and HACH DPD methods. B. CONTENTS A. PURPOSE AND SCOPE... 1 B. CONTENTS...

More information

VITLAB Volumetric instruments

VITLAB Volumetric instruments VITLAB Volumetric instruments Standard Operating Procedure (SOP) VITLAB GmbH Linus-Pauling-Str.1 63762 Grossostheim Germany Telefon: +49 6026 97799-0 Fax: +49 6026 97799-30 E-mail: info@vitlab.com Internet:

More information

Level MEASUREMENT 1/2016

Level MEASUREMENT 1/2016 Level MEASUREMENT 1/2016 AGENDA 2 A. Introduction B. Float method C. Displacer method D. Hydrostatic pressure method E. Capacitance method G. Ultrasonic method H. Radar method I. Laser method J. Level

More information

MVS Volume Verification Using Any Microtiter Plate or Small Volume Container, Such as a Tube or Vial

MVS Volume Verification Using Any Microtiter Plate or Small Volume Container, Such as a Tube or Vial MVS Volume Verification Using Any Microtiter Plate or Small Volume Container, Such as a Tube or Vial Keith J. Albert, Ph.D. ARTEL Abstract. This application note discusses methods for measuring target

More information

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS Frank M. Gavila F&J SPECIALTY PRODUCTS, INC. 404 Cypress Road Ocala, Florida, USA 3447 fandj@fjspecialty.com INTRODUCTION The calibration

More information

11/22/ (4) Harmonization: <846> SPECIFIC SURFACE AREA

11/22/ (4) Harmonization: <846> SPECIFIC SURFACE AREA BRIEFING 846 Specific Surface Area, USP 27 page 2385. The European Pharmacopoeia is the coordinating pharmacopeia for the international harmonization of the Specific Surface Area General Chapter, as part

More information

biosensis Mouse CXCL10/IP-10 ELISA Kit Protocol

biosensis Mouse CXCL10/IP-10 ELISA Kit Protocol biosensis Mouse CXCL10/IP-10 ELISA Kit Protocol Catalogue No: BEK-2124-2P TABLE OF CONTENTS I Materials provided...2 II Equipment required but not supplied...2 III Technical hints....2 IV Storage of kit

More information

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction Application Note Rapid performance verification of AZURA systems with refractive index detector Method Keywords ID HPLC Quality control, system verification, refractive index, AZURA Analytical HPLC Plus

More information

INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION.

INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION. Andrzej Hantz RADWAG BALANCES AND SCALES INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION. The article presents the influence of comparison process automatization

More information

Analysis of Deuterium Enrichment by Fourier Transform Infrared Spectrometry (FTIR): Practice. Christine Slater PhD

Analysis of Deuterium Enrichment by Fourier Transform Infrared Spectrometry (FTIR): Practice. Christine Slater PhD Analysis of Deuterium Enrichment by Fourier Transform Infrared Spectrometry (FTIR): Practice Christine Slater PhD Nutrition Specialist C.Slater@iaea.org International Atomic Energy Agency Preparation of

More information

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany Manual Degasser DDG-75 April 2007 1 Manual Online Degasser DDG-75 DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE-68766 Hockenheim Germany Tel. +49-6205-9450-0 Fax. +49-6205-9450-33 email info@duratec.com

More information

Operating Instructions METTLER TOLEDO Pipette Check Application for AX and MX/UMX Balances Version 1.xx

Operating Instructions METTLER TOLEDO Pipette Check Application for AX and MX/UMX Balances Version 1.xx Operating Instructions METTLER TOLEDO Pipette Check Application for AX and MX/UMX Balances Version 1.xx Contents 1 Introducing the Pipette Check application... 3 2 important notes... 3 3 Selecting the

More information

THERMODYNAMICS OF A GAS PHASE REACTION: DISSOCIATION OF N 2 O 4

THERMODYNAMICS OF A GAS PHASE REACTION: DISSOCIATION OF N 2 O 4 THERMODYNAMICS OF A GAS PHASE REACTION: DISSOCIATION OF N 2 O 4 OBJECTIVES 1. To measure the equilibrium constant, enthalpy, entropy, and Gibbs free energy change of the reaction N2O4(g) = 2 NO2(g). 2.

More information

THE USE OF Z FACTORS IN GRAVIMETRIC VOLUME MEASUREMENTS

THE USE OF Z FACTORS IN GRAVIMETRIC VOLUME MEASUREMENTS THE USE OF Z FACTORS IN GRAVIMETRIC VOLUME MEASUREMENTS Author: T. Mautjana NMISA Private Bag X 34 Lynwood Ridge; 0040 Email: tmautjana@nmisa.org Phone: 012 841 4798; Fax: 012 841 2131 Abstract Using gravimetric

More information

Eliminating Sources of Pipetting Error in the Forensic Laboratory David M. Epstein Director of Scientific Operations. Abstract.

Eliminating Sources of Pipetting Error in the Forensic Laboratory David M. Epstein Director of Scientific Operations. Abstract. Page 1 of 6 Table of Contents Back Issues Search Editors About FSC Submissions Forensic Science Communications October 2003 Volume 5 Number 4 Research and Technology Eliminating Sources of Pipetting Error

More information

download instant at Experiment 2 A Submarine Adventure: Density Saves the Day

download instant at  Experiment 2 A Submarine Adventure: Density Saves the Day Experiment 2 A Submarine Adventure: Density Saves the Day Instructor Notes and Lab Preparation: Chemicals and Equipment: various metal shapes of copper, nickel, lead, aluminum, brass, iron and magnesium

More information

How to Measure R7.1. Reference. I. Linear dimensions

How to Measure R7.1. Reference. I. Linear dimensions How to Measure Written by Connie Russell I. Linear dimensions Measuring linear dimensions (the distance between two points) is usually associated with using a ruler or a tape measure. For measuring objects

More information

Density Determination Kit Instruction Manual

Density Determination Kit Instruction Manual Visit www.balances.com to see the full line of Ohaus Balances Instruction Manual This Kit is compatible with OHAUS Adventurer, Adventurer Pro (except AV5x), Adventurer SL (except AS15x), Pioneer, Explorer

More information

STD-3-V1M4_1.7.1_AND_ /15/2015 page 1 of 6. TNI Standard. EL-V1M4 Sections and September 2015

STD-3-V1M4_1.7.1_AND_ /15/2015 page 1 of 6. TNI Standard. EL-V1M4 Sections and September 2015 page 1 of 6 TNI Standard EL-V1M4 Sections 1.7.1 and 1.7.2 September 2015 Description This TNI Standard has been taken through all of the voting stages and has received consensus approval by the TNI membership.

More information

711 DISSOLUTION. Portions of the present general chapter text that are national USP text, and therefore not part of the

711 DISSOLUTION. Portions of the present general chapter text that are national USP text, and therefore not part of the Change to read: 711 DISSOLUTION This general chapter is harmonized with the corresponding texts of the European Pharmacopoeia and/or the Japanese Pharmacopoeia. The texts of these pharmacopeias are therefore

More information

RayBio Human vwf ELISA Kit

RayBio Human vwf ELISA Kit RayBio Human vwf ELISA Kit Catalog #: ELH-vWF User Manual Last revised April 15, 2016 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite 100 Norcross, GA

More information

Any use of this material except in accordance with a written agreement with Western Power is prohibited.

Any use of this material except in accordance with a written agreement with Western Power is prohibited. September 2006 December 2005 Client Economic Regulation Authority Energy Safety Project Name Metering Code 2005 Document Number 3274025 Document Title Revision Status METERING MANAGEMENT PLAN Approved

More information

Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow

Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow Georgia Institute of Technology School of Earth and Atmospheric Sciences EAS 4641, Spring 2008 Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow Purpose of Lab 3: 1) To gain a

More information

CALIBRATION SYSTEM REQUIREMENTS. ESCC Basic Specification No

CALIBRATION SYSTEM REQUIREMENTS. ESCC Basic Specification No Page 1 of 8 CALIBRATION SYSTEM REQUIREMENTS ESCC Basic Specification Issue 4 February 2014 Document Custodian: European Space Agency see https://escies.org PAGE 2 LEGAL DISCLAIMER AND COPYRIGHT European

More information

Gas Laws. Introduction

Gas Laws. Introduction Gas Laws Introduction In 1662 Robert Boyle found that, at constant temperature, the pressure of a gas and its volume are inversely proportional such that P x V = constant. This relationship is known as

More information

Transfusion equipment for medical use. Part 4: Transfusion sets for single use, gravity feed

Transfusion equipment for medical use. Part 4: Transfusion sets for single use, gravity feed Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 1135-4 Sixth edition 2015-12-01 Transfusion equipment for medical use Part 4: Transfusion sets for single use, gravity feed Matériel de transfusion

More information

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev -

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev - 1 W. W. Hansen Experimental Physics Laboratory STANFORD UNIVERSITY STANFORD, CALIFORNIA 94305-4085 Gravity Probe B Relativity Mission OPERATION OF LIQUID PARTICLE COUNTER GP-B P0045 Rev - May 23, 1997

More information

Quick Reference Guide

Quick Reference Guide HematologyAnalyzer Quick Reference Guide For Veterinary Use Only Running a Blank Blanks must be run every 12 hours, after a Soak Cleaning has been performed, after a reagent pack change, and after the

More information

APPLICATION NOTE. GC Integrated Permeation Device

APPLICATION NOTE. GC Integrated Permeation Device GC Integrated Permeation Device GC-integrated Design PPB to PPM Level Calibration No Cylinders or Regulators Required Option for Dual Ovens Cost Effective, Safe, Clean, Flexible, NIST Traceable Keywords:

More information

biosensis Rat Fibronectin ELISA Kit Protocol

biosensis Rat Fibronectin ELISA Kit Protocol biosensis Rat Fibronectin ELISA Kit Protocol Catalog No: BEK-2017-2P For quantitative detection of rat Fibronectin in cell culture supernatants, serum, and citrate, heparin, or EDTA plasma samples only

More information

Department of Laboratories St. Louis, MO ISSUE DATE: June 2007 REVISION DATE: July 2010 REVIEWED DATE: May 2017 PRINCIPILE:

Department of Laboratories St. Louis, MO ISSUE DATE: June 2007 REVISION DATE: July 2010 REVIEWED DATE: May 2017 PRINCIPILE: Page 1 of 6 PROCEDURE: HEMOCUE HEMOGLOBIN 201+ ANALYZER ISSUE DATE: June 2007 REVISION DATE: July 2010 REVIEWED DATE: May 2017 PRINCIPILE: The HemoCue B-Hemoglobin 201+ system consists of disposable microcuvettes

More information

HemoCue Hb Procedure Template

HemoCue Hb Procedure Template HemoCue Hb 201 + Procedure Template PURPOSE The HemoCue Hb 201 + System is used for the quantitative determination of hemoglobin in blood using a specially designed analyzer, HemoCue Hb 201 +, and specially

More information

Improve pipetting results in pharmaceutical formulation by using MICROMAN E

Improve pipetting results in pharmaceutical formulation by using MICROMAN E Improve pipetting results in pharmaceutical formulation by using MICROMAN E Application Note LT805005 Nadège Belhadj, Gilson, France nbelhadj@gilson.com In pharmaceutical formulation, properties of used

More information

biosensis Human TNFα/Cachectin/TNFSF2 ELISA Kit Protocol

biosensis Human TNFα/Cachectin/TNFSF2 ELISA Kit Protocol biosensis Human TNFα/Cachectin/TNFSF2 ELISA Kit Protocol Catalog No: BEK-2100-1P For quantitative detection of human TNFα in cell culture supernatants, serum, and heparin, EDTA or citrate treated plasma

More information

SEMATECH Provisional Test Method for Pressure Cycle Testing Filter Cartridges Used in UPW Distribution Systems

SEMATECH Provisional Test Method for Pressure Cycle Testing Filter Cartridges Used in UPW Distribution Systems Provisional Test Method for Pressure Cycle Testing Filter Cartridges Used in UPW Distribution Systems Technology Transfer 92010941B-STD and the logo are registered service marks of, Inc. 1996, Inc. Provisional

More information

SEMLATA3-12 Measuring, weighing and preparing compounds and solutions for laboratory use

SEMLATA3-12 Measuring, weighing and preparing compounds and solutions for laboratory use Measuring, weighing and preparing compounds and solutions for Overview This unit covers the competences you need to measure, weigh and prepare compounds and solutions for laboratory investigations. Prior

More information

ab IgG1 Human ELISA Kit

ab IgG1 Human ELISA Kit ab100548 IgG1 Human ELISA Kit Instructions for Use For the quantitative measurement of Human IgG1 in serum and plasma This product is for research use only and is not intended for diagnostic use. Version

More information

Experiment #2. Density and Measurements

Experiment #2. Density and Measurements Experiment #2. Density and Measurements Goals 1. To measure and record length, volume and mass accurately with the correct number of significant figures 2. To use significant figures correctly in calculations.

More information

ISO 7218 INTERNATIONAL STANDARD

ISO 7218 INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 7218 Third edition 2007-08-15 AMENDMENT 1 2013-08-01 Corrected version 2014-04-15 Microbiology of food and animal feeding stuffs General requirements and guidance for microbiological

More information

biosensis Mouse Vascular endothelial growth factor A/VEGF-A/VEGF-164/VEGF-1/VEGF- 120/VEGF-2 ELISA Kit Protocol

biosensis Mouse Vascular endothelial growth factor A/VEGF-A/VEGF-164/VEGF-1/VEGF- 120/VEGF-2 ELISA Kit Protocol biosensis Mouse Vascular endothelial growth factor A/VEGF-A/VEGF-164/VEGF-1/VEGF- 120/VEGF-2 ELISA Kit Protocol Catalog No: BEK-2110-1P For quantitative detection of mouse VEGF-A (VEGF164&VEGF120) in mouse

More information

Physics 1021 Experiment 4. Buoyancy

Physics 1021 Experiment 4. Buoyancy 1 Physics 1021 Buoyancy 2 Buoyancy Apparatus and Setup Materials Force probe 1000 ml beaker Vernier Calipers Plastic cylinder String or paper clips Assorted bars and clamps Water Attach the force probe

More information

VITLAB micropipette. Standard Operating Procedure (SOP)

VITLAB micropipette. Standard Operating Procedure (SOP) VITLAB micropipette Standard Operating Procedure (SOP) VITLAB GmbH Linus-Pauling-Str.1 63762 Grossostheim Germany tel: +49 6026 97799-0 fax: +49 6026 97799-30 info@vitlab.com www.vitlab.com Table of contents

More information