DXH Troubleshooting FlowCharts

Size: px
Start display at page:

Download "DXH Troubleshooting FlowCharts"

Transcription

1 DXH Troubleshooting FlowCharts WARNINGS AND PRECAUTIONS Read all product manuals and consult with Beckman Coulter-trained personnel before attempting to operate instrument. Beckman Coulter, INC., urges its customers to comply with all national health and safety standards such as the use of barrier protection. This may include, but it is not limited to, protective eyewear, gloves and suitable laboratory attire when operating or maintaining this or any other automated laboratory equipment. INTENTION FOR USE This document is not intended to replace the information in your Instrument Instructions for Use Manual (IFU). Information in the Instructions for Use Manual supersedes information in any other manual.

2 Daily Checks Background fails Still fails which parameter? WBC RBC HGB PLT DIFF NRBC RETIC Zap Apertures x3 Clear appropriate Apertures (WBC or RBC) through Dx Tools>Maintenance tab Clean Apertures, Perform HGB Blank Verification, Likely causes: - Microbubbles in sweepflow lines - Electronic noise, or -Diluent contamination Prime Sweepflow Rinse Mix Chamber Go to Dx Tools>VCSn tab, select Fluidic>Rinse mix Chamber>Start Perform in order, as needed: - Flush Flow Cell (no cleaner) Both, (Repeat 5x) - Flush Flow Cell with cleaner (5 mins) - Clear Flow Cell Aperture 3x Clean Apertures, Shutdown for 30 mins, Still fails Run a cassette load of samples. Repeat Daily Checks Still fails When was the diluent last changed? Prime diluent Still fails Coulter Customer Support Center Run cassette of samples, Shutdown for 30 mins, Diff, NRBC still fails Retic still fails Prime the Retic reagents (Dx tools>common Services tab) 2

3 QC 6C, Retic-X Which results are affected? CBC Parameters? (WBC, HGB, RBC, PLT, DIFF, NRBC, RETIC Parameters (%) Refer to QC 1 Refer to QC 2

4 QC 1 CBC parameters (WBC, RBC, HGB, MCV, PLT, MPV) Out of Range? R Flags? Voteouts? Parameter affected in same way for each level (ie all running H or L; or showing a bias in same direction, but within range)? Reagent/s changed recently? All levels of QC affected and results confirmed by running new vials of each Are the other levels showing a bias on the same side of the mean? New lot? Try another box of reagent t Resolved Resolved Coulter Customer Support Center to advise of findings and request replacement Trending noted? Sudden shift? (except for MCV which is expected to trend up over time) - Has calibration factor been changed recently? - Has PMI been done recently? - Has FSE visited for an issue and made adjustments (changed hardware etc)? Check calibration and adjust as needed Pass Pass Pass Possible dilution issue?hardware - Clean outside of BSV - Clear appropriate Apertures (in Dx Tools>Maintenance) - Perform repeatability Fail Repeat repeatability Fail - Perform Clean Apertures (in Dx Tools>Maintenance) - Perform repeatability Fail Repeat repeatability Fail New vial gives same result? New lot of QC? Lab can set own target (as per IFU) as long as CVs are within performance specifications. Contact Beckman Coulter Customer Support Center to advise of findings Coulter Customer Support Center 4

5 QC 2 Out of Range? R Flags? Voteouts? DIFF, NRBC, RETIC Reagent/s changed recently? All levels affected? Try new vial of same level. Issue persists? New lot? Try another box of reagent Issue persists R flags? Results out of range or no results? Vial related. System is okay to run. Coulter Customer Support Center to advise of findings and request replacement Issue resolved Perform Rinse Mix Chamber Go to Dx Tools>VCSn tab, select Fluidic>Rinse mix Chamber>Start When complete, Finish Diagnostics Results in range? Are patient samples giving R flags? System is okay to run (Refer to 6C QC IFU> Expected Results section) New kit lot of QC? Coulter Customer Support Center to advise of findings Perform in order, as needed: - Flush Flow Cell (no cleaner) Go to Dx Tools>Maintenance tab, select Flush Flow Cell.>Choose Both >Start (Repeat 5x) - Flush Flow Cell with Cleaner for 5 mins - Clear Flow Cell Aperture When complete, Finish Diagnostics Issue persists 5

6 SAMPLES Which results are affected? All Parameters? (WBC, HGB, RBC, PLT, DIFF, NRBC, RETIC) CBC parameters only? (WBC, HB, RBC, PLT) AMTC module parameters only? (DIFF, NRBC, RETIC) Refer to Samples 1 Refer to Samples 2 Refer to Samples 3

7 SAMPLES 1 All Parameters (WBC, HGB, RBC, PLT, DIFF% #, NRBC% #, RETIC% #) Check the Sample Aspiration Probe - has it been recently replaced? Check the aspiration line connected at the top of the probe. Check the probe is seated & screwed in correctly. Ver 1.3 (vember 2016) Possible blocked Sample Probe OR Sample Aspiration Syringe /Motor issue Is blood being pulled through to the Blood Sample Valve? Is Sample Aspiration Syringe moving up & down freely? Signs of leaking around aspiration syringe? Replace Sample Aspiration Probe Confirm which results are affected (cannot be all) and refer to appropriate flowchart for further troubleshooting Check thumbscrew at bottom of syringe is finger tight. - If already tight, syringe needs replacing. Coulter Customer Support Center - If loose, tighten and check results again Could be faulty syringe motor or syringe itself. Coulter Customer Support Center Beckman Coulter Technical Training 7

8 SAMPLES 2 CBC parameters only (WBC, HGB, RBC, PLT) Is the BSV moving during the analysis cycle? Go to Dx Tools>CBC tab, select Mechanical>Cycle BSV. Does the BSV respond? Coulter Customer Support Center Voteouts? (Partial or Total) Which parameter/s? ALL Plt only - Clean outside of BSV -Perform Cycle BSV Still no CBC results? Is Sample Aspiration Syringe moving up & down freely? WBC or RBC only HGB only WBC & HB RBC & PLT Check PLT histogram spike at 2fL? Possible causes: - microbubbles in sweepflow line, prime sweepflow - electronic noise, check cover in place - possible diluent contamination, change diluent Voteouts? (Partial or Total) result Perform Zap Apertures x3 Voteouts persist Is HGB lamp on when analyser is powered on? Perform Clean Apertures and HGB Blank Verification Problem persists Clear appropriate Apertures (RBC or WBC) (Dx tools >Maintenance tab) 3x Clean Apertures (Dx tools >Maintenance tab) - Check when DxH Lyse was last changed and replace if coincides with onset of the problem. Contact Beckman Coulter if issue not resolved. - Possible dilution issue (if both parameters are affected in the same direction (ie H or L) - Check BSV and appropriate pump for movement - Clean outside of BSV - Perform Cycle BSV from Dx Tools> Maintenance - Unresolved? Coulter Customer Support Center Could be faulty syringe motor or syringe itself. Check thumbscrew at bottom of aspiration syringe is finger tight and look for signs of leaking - If loose, tighten and check results again - If already tight, syringe needs replacing. - Coulter Support Center Voteouts persist Voteouts persist Coulter Support Ver 1.3 (vember 2016) Beckman Coulter Technical Training 8

9 SAMPLES 3 Check SAM. Is sample being dispensed to the chambers? Is the aspiration Syringe moving up & down freely? ALL no results, low events, R flags AMTC parameters only* (DIFF, NRBC, RETIC) Perform in order, as needed: DIFF only NRBC only RETIC only Pathway or Reagent specific *te if you observe %, but not absolute values, then the issue is with the 1 o parameter used for calculating the abs value (eg WBC for DIFF, RBC for RETIC). DV or Flow Cell Issue Cycle DV Go to Dx Tools>CBC tab, select Mechanical>Cycle DV. Does the DV respond? Coulter Customer Support Center Still no results - Flush Flow Cell (no cleaner) Go to Dx Tools>Maintenance tab, select Flush Flow Cell.>Choose Both >Start (repeat 5x) - Flush Flow Cell (with Cleaner) for 5 mins (repeat if needed) - Clear Flow Cell Aperture Check counts (% and # affected?), flags (R?) and error messages Perform Rinse Mix Chamber Go to Dx Tools>VCSn tab, select Fluidic>Rinse mix Chamber>Start. If not resolved, proceed to next step. Low events, Low event rate, Required parameters for generating dataplots missing, R flags High event rate, R flags Unlock DV Go to Dx Tools>CBC tab, select Mechanical>Unlock DV. Does the DV respond? Possible blockage inside DV or input line of Flowcell Perform in order, as needed: - Flush Flow Cell (no cleaner) Go to Dx Tools>Maintenance tab, select Flush - Flow Cell.>Choose Both >Start (repeat 5x) - Repeat Flush Flow Cell with Cleaner for 5 mins (repeat if needed) - Clear Flow Cell Aperture Check reagents. - Near end of open expiry? - Nearly empty? - Changed recently? Change relevant reagent pack Ver 1.3 (vember 2016) Coulter Customer Support Center Issue persists Beckman Coulter Technical Training 9

Quick Reference Guide

Quick Reference Guide HematologyAnalyzer Quick Reference Guide For Veterinary Use Only Running a Blank Blanks must be run every 12 hours, after a Soak Cleaning has been performed, after a reagent pack change, and after the

More information

COULTER A C T diff Analyzer. Operator s Guide. diff. PN DB (March 2016) Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA U.S.A.

COULTER A C T diff Analyzer. Operator s Guide. diff. PN DB (March 2016) Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA U.S.A. TM COULTER A C T diff Analyzer Operator s Guide diff PN 4237416DB (March 2016) Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA 92821 U.S.A. WARNINGS AND PRECAUTIONS READ ALL PRODUCT MANUALS AND CONSULT

More information

COULTER LH 500 HEMATOLOGY ANALYZER

COULTER LH 500 HEMATOLOGY ANALYZER COULTER LH 500 HEMATOLOGY ANALYZER TRAINING MODULES This document is not intended to replace the information in your instrument Instructions for Use manual (IFU). Information in the Instructions for Use

More information

COULTER A C T diff 2 Analyzer. Operator s Guide. PN BA (August 2010) Beckman Coulter, Inc. 250 S. Kraemer Blvd.

COULTER A C T diff 2 Analyzer. Operator s Guide. PN BA (August 2010) Beckman Coulter, Inc. 250 S. Kraemer Blvd. COULTER A C T diff 2 Analyzer Operator s Guide (August 2010) Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA 92821 WARNINGS AND PRECAUTIONS READ ALL PRODUCT MANUALS AND CONSULT WITH BECKMAN COULTER-TRAINED

More information

COULTER LH 500 HEMATOLOGY ANALYZER

COULTER LH 500 HEMATOLOGY ANALYZER COULTER LH 500 HEMATOLOGY ANALYZER TRAINING MODULES WARNINGS AND PRECAUTIONS READ ALL PRODUCT MANUALS AND CONSULT WITH BECKMAN COULTER-TRAINED PERSONNEL BEFORE ATTEMPTING TO OPERATE INSTRUMENT. DO NOT

More information

COULTER LH 750 HEMATOLOGY ANALYZER COULTER LH 780 HEMATOLOGY ANALYZER COULTER LH SLIDEMAKER COULTER LH SLIDESTAINER Training Guide

COULTER LH 750 HEMATOLOGY ANALYZER COULTER LH 780 HEMATOLOGY ANALYZER COULTER LH SLIDEMAKER COULTER LH SLIDESTAINER Training Guide COULTER LH 750 HEMATOLOGY ANALYZER COULTER LH 780 HEMATOLOGY ANALYZER COULTER LH SLIDEMAKER COULTER LH SLIDESTAINER Training Guide This Training Guide belongs to: Coulter LH 700 Series Training Guide Page

More information

Automated Liquid Handling Station

Automated Liquid Handling Station Automated Liquid Handling Station Software User s Guide Manual Part Number 32-0442-048 Rev 0 2018 Teledyne Technologies Incorporated. All rights reserved. Printed in USA. Installation See the SimPrep Quick

More information

Hemoglobin Reference Failure Troubleshooting

Hemoglobin Reference Failure Troubleshooting Hemoglobin Reference Failure Troubleshooting The HEMOGLOBIN REFERENCE FAILURE error ( HGB REF FAILURE ) is generated by the Vet ABC as a quality control check for hemoglobin measurement. Recall that the

More information

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT ATTENTION SYRINGE REPLACEMENT Please read through the document completely before starting any repairs. Refer to the proper section in the service manual for complete removal and replacement procedures.

More information

HumaCount 30 TS and 60 TS

HumaCount 30 TS and 60 TS HumaCount 0 TS and 60 TS Service Manual 640/0 REVISION LIST OF THE MANUAL Rev. /DATE REVISION DESCRIPTION 0/00-09 First edition SYSTEM VERSION HumaCount 60 TS and HumaCount 0 TS with software version..

More information

Reference method for the enumeration of erythrocytes and leucocytes

Reference method for the enumeration of erythrocytes and leucocytes Clin. lab. Haemat. 1994 16, 131-138 Reference method for the enumeration of erythrocytes and leucocytes INTERNATIONAL COUNCIL FOR STANDARDIZATION IN HAEMATOLOGY; PREPARED BY THE EXPERT PANEL ON CYTOMETRY

More information

Observations Possible Causes Recommended Actions. Turquoise color

Observations Possible Causes Recommended Actions. Turquoise color ABI PRISM 3700 Instrument-Hardware-Array View Troubleshooting Module Please contact technical support if you have any questions regarding the document. Text in blue indicates a link to another section

More information

Navigator 600 Silica analyzers

Navigator 600 Silica analyzers ABB MEASUREMENT & ANALYTICS INFORMATION INF09/020 REV. B Navigator 600 Silica analyzers Troubleshooting Measurement made easy Navigator 600 silica analyzers 1 Introduction This publication details troubleshooting

More information

Proficiency Testing Corrective Action Checklist: PT PROVIDER: PT EVENT: TEST:

Proficiency Testing Corrective Action Checklist: PT PROVIDER: PT EVENT: TEST: Proficiency Testing Corrective Action Checklist: PT PROVIDER: PT EVENT: TEST: PT Process Package Received Who received it? Handling Upon arrival Was the kit cold? Was the kit damaged? Was the kit complete?

More information

Mass Spec will not pump down

Mass Spec will not pump down Mass Spec will not pump down Applies to 5973A/N MSD There are several situations that could cause the 5973A/N to have problems pumping down. The following scenarios should help in isolating problems related

More information

Specifications...Easy CLIA Classification: Sample Type: Blood gas analyzer operation has never been simpler Sample Size: Measured Parameters PCO2

Specifications...Easy CLIA Classification: Sample Type: Blood gas analyzer operation has never been simpler Sample Size: Measured Parameters PCO2 medicacorp.com easy inside and out With EasyBloodGas, Medica has redefined blood gas analyzer design The sophistication of traditional blood gas analyzers has been packaged in a new compact format with

More information

ABX Micros CRP 200. Technical manual. P/n: RAA028A. HORIBA ABX B.P MONTPELLIER Cedex 4 - FRANCE

ABX Micros CRP 200. Technical manual. P/n: RAA028A. HORIBA ABX B.P MONTPELLIER Cedex 4 - FRANCE ABX Micros CRP 200 Technical manual P/n: RAA028A HORIBA ABX B.P. 7290 34184 MONTPELLIER Cedex 4 - FRANCE Revisions User Manual Part number Technical Note Software revision Sections Date RAA028A RAN153A

More information

Hemoglobin g/dl Female Male 12 years and older 11.9 to to 17.5

Hemoglobin g/dl Female Male 12 years and older 11.9 to to 17.5 Reference Ranges: updated 6-21-18 Parameter Age WBC x 0 to 3 days 9.0 to 30.0 3 days to 2 weeks 5.0 to 21.0 2 weeks to 1 month 5.0 to 20.0 1 month to 2 years 6.0 to 17.5 2 years to 6 years 5.0 to 15.5

More information

Quality Control of Pulmonary Diagnostic Systems

Quality Control of Pulmonary Diagnostic Systems Quality Control of Pulmonary Diagnostic Systems James P. Sullivan, BA, RPFT Pulmonary Diagnostic Laboratories Memorial Sloan Kettering Cancer Center New York, New York QA, QC and QI: what s the difference?

More information

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS) Toxicology Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose - This procedure specifies the required elements for the calibration and use of the Agilent Gas Chromatograph interfaced to the Agilent

More information

AFDXXX(X)AC Series Operators Manual Please read this manual thoroughly before attempting to operate your water maker.

AFDXXX(X)AC Series Operators Manual Please read this manual thoroughly before attempting to operate your water maker. AFDXXX(X)AC Series Operators Manual Please read this manual thoroughly before attempting to operate your water maker. E & O E Danger High Voltage AFDXXX(X) Series water makers operate on a 240vAC electricity

More information

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS Fabrication Laboratory Revision: A Rev Date: 09/29/03 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Amray 1830 SEM. All users

More information

6800 Maintenance Instruction System Flush Procedure

6800 Maintenance Instruction System Flush Procedure Equipment Required FA74005 FA65318 FA900005 FA900003 Damper Drain Tube 6800 Cover Removal Tool Beaker 0.25 Litre Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves

More information

Department of Laboratories St. Louis, MO ISSUE DATE: June 2007 REVISION DATE: July 2010 REVIEWED DATE: May 2017 PRINCIPILE:

Department of Laboratories St. Louis, MO ISSUE DATE: June 2007 REVISION DATE: July 2010 REVIEWED DATE: May 2017 PRINCIPILE: Page 1 of 6 PROCEDURE: HEMOCUE HEMOGLOBIN 201+ ANALYZER ISSUE DATE: June 2007 REVISION DATE: July 2010 REVIEWED DATE: May 2017 PRINCIPILE: The HemoCue B-Hemoglobin 201+ system consists of disposable microcuvettes

More information

Overview. SEMEMI216 SQA Unit Code HF4G 04 Carrying out fault location on fluid power equipment and circuits

Overview. SEMEMI216 SQA Unit Code HF4G 04 Carrying out fault location on fluid power equipment and circuits Overview This standard identifies the competences you need to locate faults on fluid power equipment and circuits, on mobile or static plant, in accordance with approved procedures. You will be required

More information

MVS Volume Verification Using Any Microtiter Plate or Small Volume Container, Such as a Tube or Vial

MVS Volume Verification Using Any Microtiter Plate or Small Volume Container, Such as a Tube or Vial MVS Volume Verification Using Any Microtiter Plate or Small Volume Container, Such as a Tube or Vial Keith J. Albert, Ph.D. ARTEL Abstract. This application note discusses methods for measuring target

More information

OAI Model 200 Tabletop Mask Aligner Portland State University

OAI Model 200 Tabletop Mask Aligner Portland State University OAI Model 200 Tabletop Mask Aligner Portland State University WARNING: This machine exposes users to ultraviolet radiation. Do not touch the lens underneath the lamp hood as it may damage the machine and

More information

HemoCue Hb Procedure Template

HemoCue Hb Procedure Template HemoCue Hb 201 + Procedure Template PURPOSE The HemoCue Hb 201 + System is used for the quantitative determination of hemoglobin in blood using a specially designed analyzer, HemoCue Hb 201 +, and specially

More information

Rotary Abreast Automatic Teat Sprayer User Manual. Technical Manual Revision 0.01

Rotary Abreast Automatic Teat Sprayer User Manual. Technical Manual Revision 0.01 Rotary Abreast Automatic Teat Sprayer Technical Manual 006911 Revision 0.01 Contents Contents... 2 Important information - read this first... 3 Icons... 3 1. Introduction... 4 2. Intended Use... 4 3. General

More information

Balloon Gastrostomy (BG) Tube

Balloon Gastrostomy (BG) Tube Balloon Gastrostomy (BG) Tube Post insertion advice Please follow hospital/ Consultant advice for care of the tube and stoma site for the first 28 days (if the BG tube has been placed as a first gastrostomy

More information

Medonic M-series Art no May 2006

Medonic M-series Art no May 2006 Medonic M-series Contents PREFACE... 3 Introduction... 3 SECTION 1: SAFETY INSTRUCTIONS... 5 Section Overview... 5 1.1 Intended Use... 5 1.2 Safety Instruction... 6 1.3 Biohazards... 6 1.4 Emergency Procedure...

More information

SURVEYING THE HAZARDOUS MATERIALS INCIDENT

SURVEYING THE HAZARDOUS MATERIALS INCIDENT HAZARDOUS MATERIALS TECHNICIAN MANIPULATIVE SKILL OBJECTIVES SURVEYING THE HAZARDOUS MATERIALS INCIDENT 1. Given three hazardous materials, one of which is a solid, one a liquid and one a gas, and the

More information

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the

More information

Please refer F 90 of the present manual for the identification of the valves used in the detector.

Please refer F 90 of the present manual for the identification of the valves used in the detector. DN 25/DN 40 VAT valves assembly : Contact customer service. E 190 (MTR only) Background The detector is equipped with Bacosol valves and Minisol valves without valve block. Please refer F 90 of the present

More information

UsER manual for Watersens ph -REDOX

UsER manual for Watersens ph -REDOX UsER manual for Watersens -REDOX Cl 8 1 2 6 3 3 7 7 4 4 4 4 Parts List 1 Redox Probe 1 x 2 PH Probe 1 x 5 Tube Weight 2 x 6 Connection Valve 1 x chlorine 3 Chlorine and Pumps 2 x 7 Dosing Valve 2 x 5 5

More information

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet: OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle

More information

SEQUENTIAL SHOWER VALVE INSTRUCTION MANUAL PLEASE LEAVE THIS MANUAL WITH THE END USER

SEQUENTIAL SHOWER VALVE INSTRUCTION MANUAL PLEASE LEAVE THIS MANUAL WITH THE END USER SEQUENTIAL SHOWER VALVE INSTRUCTION MANUAL PLEASE LEAVE THIS MANUAL WITH THE END USER CONTENTS 1. INTRODUCTION & SAFETY 1 2. DIMENSIONS 2 3. TECHNICAL DATA 3 4. OPERATION 4 5. COMPONENTS 5 6. SITE INSTALLATION

More information

Gas Network Craftsperson

Gas Network Craftsperson Gas Network Craftsperson Unit EIAU06 Maintaining Instrumentation and Control Equipment Circuits This assessment specification has been developed as part of the network maintenance craftsperson standard

More information

Quality Management Determined from Risk Assessment

Quality Management Determined from Risk Assessment Quality Management Determined from Risk Assessment Bruce Thomadsen University of Wisconsin and The Center for the Assessment of Radiological Sciences Disclosure I am the President of the Center for the

More information

Mass Spec will not Autotune

Mass Spec will not Autotune Mass Spec will not Autotune Applies to 5973A/N MSD What could be the problem? There could be several things that would cause your Mass Spec not to Autotune. The most common, easily corrected Autotune problems

More information

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay Summary This assay measures Human Granulocyte Colony Stimulating Factor (G-CSF) in a 96-well

More information

Convergent Technologys product overview

Convergent Technologys product overview Convergent Technologys product overview Lab Diagnostics - Convergys 1100-1001 Convergys ISE Analyzer 1100-1003 Convergys ISE Analyzer 3 electrodes out of (K+, Na+, Li+, CL-, Ca++, ph)* 5 electrodes out

More information

BD Cytopeia Fluidic Kit User s Guide

BD Cytopeia Fluidic Kit User s Guide BD Cytopeia Fluidic Kit User s Guide For Research Use Only 23-17618-00 9/2015 Becton, Dickinson and Company BD Biosciences 2350 Qume Drive San Jose, CA 95131 USA BD Biosciences European Customer Support

More information

Calcium Reactor. CalcFeeder PRO. User Manual 1.1a. Model. CalcFeeder AC1 PRO CalcFeeder AC2 PRO CalcFeeder AC3 PRO CalcFeeder AC4 PRO

Calcium Reactor. CalcFeeder PRO. User Manual 1.1a. Model. CalcFeeder AC1 PRO CalcFeeder AC2 PRO CalcFeeder AC3 PRO CalcFeeder AC4 PRO Calcium Reactor CalcFeeder PRO User Manual 1.1a Model CalcFeeder AC1 PRO CalcFeeder AC2 PRO CalcFeeder AC3 PRO CalcFeeder AC4 PRO Page 1 Dear Customer, Thank you for purchasing an Pacific Sun CalcFeeder

More information

CDI Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist

CDI Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist Original Article Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist David W. Fried, MS Ed, CCP; Joseph J. Leo, BS, CCP; Gabriel J. Mattioni, BS, CCP; Hasratt Mohamed, CCP; Raymond

More information

Liquid Handling Systems

Liquid Handling Systems Liquid Handling Systems 1.0 Purpose - This procedure specifies the required elements for the maintenance and use of the Hamilton- Microlab 1000 Plus and the Microlab 500 Liquid Handling Systems. 2.0 Scope

More information

Multi-Probe Sampling System User Guide & Manual

Multi-Probe Sampling System User Guide & Manual Multi-Probe Sampling System User Guide & Manual Rev. 06/05/12 Part #22200012 Table of Contents CHAPTER 1: SYSTEM DESCRIPTION... 2 FUNCTION AND THEORY... 3 SYSTEM COMPONENTS... 4 CHAPTER 2: SYSTEM INSTALLATION...

More information

Guidelines. Milan Special School District. Automated External Defibrillator (AED) Program

Guidelines. Milan Special School District. Automated External Defibrillator (AED) Program Guidelines Milan Special School District Automated External Defibrillator (AED) Program Purpose: To provide management or administration of the MSSD school-based AED program. Sudden cardiac arrest (SCA)

More information

SEMPEO SQA Unit Code FP2P 04 Maintaining fluid power equipment

SEMPEO SQA Unit Code FP2P 04 Maintaining fluid power equipment Overview This standard covers a broad range of basic fluid power maintenance competences which will prepare you for entry into the engineering or manufacturing sectors, creating a progression between education

More information

6900 Maintenance Instruction System Flush

6900 Maintenance Instruction System Flush Equipment Required FA74005 Damper Drain Tube FA16005 Cover Removal Tool FA900005 Beaker 0.25 Litre FA900003 Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves Latex

More information

no peaks missing peaks too small peaks increasing retention decreasing retention

no peaks missing peaks too small peaks increasing retention decreasing retention GC Troubleshooting We can help you. 1. Observation: no peaks no peaks missing peaks too small peaks increasing retention decreasing retention declining baseline rising baseline bleeding plateaus interfering

More information

PositionMaster EDP300 Extended Diagnostics. Compact, well-proven, and flexible

PositionMaster EDP300 Extended Diagnostics. Compact, well-proven, and flexible Change from one to two columns Technical Information TI/EDP300_ED-EN Rev. A PositionMaster EDP300 Extended Diagnostics Compact, well-proven, and flexible The benefits of the extended diagnostics feature

More information

Technical Procedure for General Laboratory Equipment

Technical Procedure for General Laboratory Equipment Technical Procedure for General Laboratory Equipment 1.0 Purpose - This procedure specifies the required elements for the use of general laboratory equipment. 2.0 Scope - This procedure applies to all

More information

PH01 Perfusion Cannula Manual

PH01 Perfusion Cannula Manual PH01 Perfusion Cannula Manual Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted without the express written permission of Multi

More information

Mix 920 ml of 87% glycerol (on shelf) with 1080 ml of Isoton II (in cabinet), stir in a flask until homogenized (the glycerol will sink).

Mix 920 ml of 87% glycerol (on shelf) with 1080 ml of Isoton II (in cabinet), stir in a flask until homogenized (the glycerol will sink). Instructions for the care and use of the Multisizer 3 The Multisizer is intended for quantifying either cells or spheres. We are in possession of two apertures: one with a 100-µm diameter opening and the

More information

Parallel Reactor Standard Operating Procedure THIS SOP IS NOT MEANT TO REPLACE HANDS-ON TRAINING FROM QUALIFIED PERSONNEL!

Parallel Reactor Standard Operating Procedure THIS SOP IS NOT MEANT TO REPLACE HANDS-ON TRAINING FROM QUALIFIED PERSONNEL! Parallel Reactor Standard Operating Procedure THIS SOP IS NOT MEANT TO REPLACE HANDS-ON TRAINING FROM QUALIFIED PERSONNEL! This SOP describes the basic procedure for using the HEL Cat-24 parallel reactor,

More information

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings SomnoSuite FAQ V1.3 January 2015 Setup 1. How do I connect the SomnoSuite to my oxygen source? 2. Is there a way to speed up the downward movement of the pusher block when setting the empty position? 3.

More information

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager Ghost Peaks: How to Fix a Haunting Problem Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager Outline of Topics What is a ghost peak? How to identify ghost peaks and their source How to correct

More information

Exhibit 4. Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N)

Exhibit 4. Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N) 1 APPLICABILITY Exhibit 4 Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N) Definitions common to all certification and test procedures

More information

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI NanoSight NS300 Operation instructions Injection/flushing brief overview: 1. Do not exceed flow of 1 ml per 20 seconds. 2. Inject two 1 ml syringes with nano-pure or DI water. 3. If the water does not

More information

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax OPERATION MANUAL Zutron Medical, LLC 17501 W 98 th St #40-27 Lenexa, KS 66219 Phone 877-343-5873 Fax 913-967-5944 ZUT-Lab-004-30004 REV. 03312017 Table of Contents 2 Introduction 1. Intended Use 2. Labels,

More information

EAS Basic Soil Gas Sampler

EAS Basic Soil Gas Sampler EAS Basic Soil Gas Sampler Please Read The Following Important Information before Starting The EAS Basic Soil Gas Sampler has the traditional design where a pump or evacuated canister can be to purge the

More information

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere! The portable Gas Chromatograph is now more versatile than ever. DPS Instruments is pleased to present the newest Portable Companion 2 Gas Chromatograph with room for 2 Detectors. The Companion 2 GC was

More information

Using the Akta Prime plus October 22, 2012

Using the Akta Prime plus October 22, 2012 Some starting precautions: 1. Vacuum filter all buffers. Removes any large particles/debris that may clog your column De-gases the buffers 2. Clarify lysates first by centrifugation and then filtration

More information

Agilent Auxiliary Gas Module AGM 2. User s Guide

Agilent Auxiliary Gas Module AGM 2. User s Guide Agilent Auxiliary Gas Module AGM 2 User s Guide Notices Agilent Technologies, Inc. 1992, 1996, 1999, 2010 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

PP-1, PP-1W, PP-2, & PP-3

PP-1, PP-1W, PP-2, & PP-3 Welker Pitot Probe Model PP-1, PP-1W, PP-2, & PP-3 The information in this manual has been carefully checked for accuracy and is intended to be used as a guide for the installation, operation, and maintenance

More information

Cartridge Based & Point-Of-Care Blood Gas Analyser QAP Performances & Quality Management Systems.. Issues

Cartridge Based & Point-Of-Care Blood Gas Analyser QAP Performances & Quality Management Systems.. Issues Cartridge Based & Point-Of-Care Blood Gas Analyser QAP Performances & Quality Management Systems.. Issues AACB QC Workshop 4 Brisbane 16th Sept. 2016 John Calleja Objectives To discuss Pre-Analytical Issues

More information

CBMTG Laboratory Committee-Laboratory Guidance

CBMTG Laboratory Committee-Laboratory Guidance Purpose: To confirm, by examination and objective evidence, that a selected cryopreservation bag is compatible with institutional policies and standard operating procedures and functions adequately so

More information

Thermo Environmental 48C Carbon Monoxide QA Plan Section I Electronic Calibration Branch Responsibilities

Thermo Environmental 48C Carbon Monoxide QA Plan Section I Electronic Calibration Branch Responsibilities Page 1 of 12 Thermo Environmental 48C Carbon Monoxide QA Plan Section I Electronic Calibration Branch Responsibilities Page 2 of 12 Table of Contents 2.9.1 SAMPLE EQUIPMENT...3 2.9.1.1 Selection and Procurement...3

More information

The benefits of the extended diagnostics feature. Compact, well-proven, and flexible

The benefits of the extended diagnostics feature. Compact, well-proven, and flexible ABB MEASUREMENT & ANALYTICS TECHNICAL INFORMATION PositionMaster EDP300 Extended Diagnostics Compact, well-proven, and flexible The benefits of the extended diagnostics feature The PositionMaster EDP300

More information

UNIVERSITY OF ROCHESTER MEDICAL CENTER

UNIVERSITY OF ROCHESTER MEDICAL CENTER 1. Purpose/Scope: This procedure defines the operation, maintenance and calibration of the Beckman 2. General Policy: The HIC will adhere to the specific guidelines recommended by Beckman Coulter s instruction

More information

Glove Box Installation Manual

Glove Box Installation Manual Glove Box Installation Manual 1998 by M. Braun Company File: GB-UNI-INS.DOC! Edition 08-00 by M. Boutin! Subject to be changed without notice Glovebox Installation Your Glove box has been fully assembled,

More information

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more! After years of development and testing, DPS Instruments is pleased to present the newest, most expandable and versatile Gas Chromatography Systems in history. The DPS 600 Series GC systems are the world

More information

1. Introduction Why use the Flow-Through Cell? Component Identification... 1

1. Introduction Why use the Flow-Through Cell? Component Identification... 1 Contents 1. Introduction... 1 2. Why use the Flow-Through Cell?... 1 3. Component Identification... 1 3.1. Standard Components... 2 3.2. Optional Components... 2 4. Assembly and Probe Insertion... 3 4.1.

More information

Powers Controls RC 195 Multiple Input Receiver- Controller

Powers Controls RC 195 Multiple Input Receiver- Controller Powers Controls RC 195 Multiple Input Receiver- Controller Document No. 155-036P25 RC 195-1 Description Features Application The RC 195 Multiple lnput Receiver-Controller is a pneumatic instrument which

More information

Agilent 1290 Infinity Pump Head Maintenance

Agilent 1290 Infinity Pump Head Maintenance Agilent 1290 Infinity Pump Head Maintenance Technical Note Agilent Technologies Notices Agilent Technologies, Inc. 2012-2015, 2016 No part of this manual may be reproduced in any form or by any means (including

More information

SEMEM SQA Unit Code HWAW 04 Carrying out maintenance activities on fluid power equipment

SEMEM SQA Unit Code HWAW 04 Carrying out maintenance activities on fluid power equipment Carrying out maintenance activities on fluid power equipment Overview This standard identifies the competences you need to lead maintenance activities by carrying out corrective maintenance activities

More information

Headspace Troubleshooting Made Easy. Presented today by: Susan Scheib Sr. Technical Support Specialist Joe Barta FSE Chicago & Escalation Engineer

Headspace Troubleshooting Made Easy. Presented today by: Susan Scheib Sr. Technical Support Specialist Joe Barta FSE Chicago & Escalation Engineer Headspace Troubleshooting Made Easy Presented today by: Susan Scheib Sr. Technical Support Specialist Joe Barta FSE Chicago & Escalation Engineer Topics to cover 7000 Determining, setting and checking

More information

Data Sheet T 8389 EN. Series 3730 and 3731 Types , , , and. EXPERTplus Valve Diagnostic

Data Sheet T 8389 EN. Series 3730 and 3731 Types , , , and. EXPERTplus Valve Diagnostic Data Sheet T 8389 EN Series 3730 and 3731 Types 3730-2, 3730-3, 3730-4, 3730-5 and Type 3731-3 Electropneumatic Positioners EXPERTplus Valve Diagnostic Application Positioner firmware to detect potential

More information

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator Designed, built, and supported by InstruQuest Inc. Universal pycnometer, no

More information

PerkinElmer Aanalyst100

PerkinElmer Aanalyst100 Using of AAS Model PerkinElmer Aanalyst100 Mr.Pisipong Meunprasertdee Scientist Instrument Center Faculty of Science and Technology 1 Using of AAS Model PerkinElmer Aanalyst100 2 AAS Training Jan 5-6,

More information

Cover Page for Lab Report Group Portion. Pump Performance

Cover Page for Lab Report Group Portion. Pump Performance Cover Page for Lab Report Group Portion Pump Performance Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 March 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section number:

More information

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev -

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev - 1 W. W. Hansen Experimental Physics Laboratory STANFORD UNIVERSITY STANFORD, CALIFORNIA 94305-4085 Gravity Probe B Relativity Mission OPERATION OF LIQUID PARTICLE COUNTER GP-B P0045 Rev - May 23, 1997

More information

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 QUICK START GUIDE P/N 6006154, REVISION C MAY 2013 Model 8120 Mask Integrity Tester is patented under U.S. Patent No. 8,312,761. Additional patents are pending.

More information

LAMINATED SAND FILTERS Filtering Equipment. Manufactured in Spain for PQS Piscinas y Consumo SA

LAMINATED SAND FILTERS Filtering Equipment. Manufactured in Spain for PQS Piscinas y Consumo SA LAMINATED SAND FILTERS Filtering Equipment Manufactured in Spain for PQS Piscinas y Consumo SA LAMINATED SAND FILTERS Filtering Equipment Filtration is an impotant part of pool maintenance. However, a

More information

PROCESS CONTROL HANDBOOK Z-6

PROCESS CONTROL HANDBOOK Z-6 Reversal Bath This step prepares the film for the color-developer step. In the reversal bath, a chemical reversal agent is absorbed into the emulsion. No chemical reaction takes place until the film enters

More information

OPERATING INSTRUCTIONS Sherlock Chamber Testing System

OPERATING INSTRUCTIONS Sherlock Chamber Testing System OPERATING INSTRUCTIONS Sherlock Chamber Testing System Arrangement Equipment Kit FL90 & FL100 Corbel Test Plate with Gasket Seal S8, S14, S15 & S16 Sump Test Plate with Gasket Seal S7 & S5 Chamber Test

More information

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions Varian 3300 Gas Chromatograph Abbreviated Operating Instructions 1. Introduction and Startup 2 2. Basic Configurations for the GC 3 3. Sample Preparation 3 4. Building a Method 3 5. Split/Splitless Injection

More information

Rhino VE CE Bulk Unloader Operator s Card

Rhino VE CE Bulk Unloader Operator s Card Rhino VE CE Bulk Unloader Operator s Card Component Locator SeeFigure1andrefertoTable1. SHOWNINTHE OPEN POSITION 1 3 8 4 7 6 5 Figure 1 Component Locator (Some parts are not shown for clarity.) Item Description

More information

Implementing IEC Standards for Safety Instrumented Systems

Implementing IEC Standards for Safety Instrumented Systems Implementing IEC Standards for Safety Instrumented Systems ABHAY THODGE TUV Certificate: PFSE-06-607 INVENSYS OPERATIONS MANAGEMENT What is a Safety Instrumented System (SIS)? An SIS is designed to: respond

More information

ACQUITY UPLC Sample Manager - Fixed Loop Operator s Overview and Maintenance Information

ACQUITY UPLC Sample Manager - Fixed Loop Operator s Overview and Maintenance Information ACQUITY UPLC Sample Manager - Fixed Loop Operator s Overview and Maintenance Information Revision A Copyright Waters Corporation 2010 2011 All rights reserved Copyright notice 2010 2011 WATERS CORPORATION.

More information

Agilent ICP-MS Interactive Troubleshooting tool for Plasma ignition problem

Agilent ICP-MS Interactive Troubleshooting tool for Plasma ignition problem Agilent ICP-MS Interactive Troubleshooting tool for Plasma ignition problem Rev. B 2017, January, 20 Next > Preconditions 7900, 7800, 7700, 8800, or 8900 ICP-MS is used. Sample introduction type is PeriPump.

More information

LENNOX SLP98UHV DIAGNOSTIC CODES

LENNOX SLP98UHV DIAGNOSTIC CODES Code Status of Equipment Action required to clear and recover - Idle mode (Decimal blinks at 1 Hertz -- 0.5 second ON, 0.5 second OFF) A Cubic feet per minute (cfm) setting for indoor blower (1 second

More information

Twin & Triple Control Concealed Thermostatic Shower Valve

Twin & Triple Control Concealed Thermostatic Shower Valve Twin & Triple Control Concealed Thermostatic Shower Valve Installation & Operating Guide Please leave this installation & user guide with the end user CONTENTS: 1. Introduction & Safety 1 2. Dimensions

More information

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure PPMS Service Note 1099-412 In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure Introduction: Loss of electricity to the PPMS EverCool II should not cause damage to

More information

SEMPEO SQA Unit Code FP2M 04 Maintaining mechanical devices and equipment

SEMPEO SQA Unit Code FP2M 04 Maintaining mechanical devices and equipment Overview This standard covers a broad range of basic mechanical maintenance competences that will prepare you for entry into the engineering or manufacturing sectors, creating a progression between education

More information

Mounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator

Mounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator Self-operated Pressure Regulators Type 2357-31 Pressure Build-up Regulator with safety function and integrated excess pressure valve Type 2357-31 with non-return unit at port C Ports A and B with soldering

More information

CGMP KITS PROTOCOL ASSAY PRINCIPLE

CGMP KITS PROTOCOL ASSAY PRINCIPLE CGMP KITS PROTOCOL Part # 62GM2PEG & 62GM2PEH Test size#: 500 tests (62GM2PEG), 10,000 tests (62GM2PEH) - assay volume: 20 µl Revision: 02-Jan.2018 Store at: 2-8 C (62GM2PEG); 2-8 C (62GM2PEH) For research

More information

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection This user guide does not have the intention to introduce you to any of the theory and/or principles behind

More information

CONTENTS SPECIFICATIONS GENERAL INFORMATION RECOMMENDED USE OPERATING PRINCIPLE TIPS ON TAKING YOUR BLOOD PRESSURE 3-4 BATTERY INSTALLATION

CONTENTS SPECIFICATIONS GENERAL INFORMATION RECOMMENDED USE OPERATING PRINCIPLE TIPS ON TAKING YOUR BLOOD PRESSURE 3-4 BATTERY INSTALLATION IFU SBPMON107 CONTENTS SPECIFICATIONS GENERAL INFORMATION RECOMMENDED USE OPERATING PRINCIPLE TIPS ON TAKING YOUR BLOOD PRESSURE BATTERY INSTALLATION CORRECT POSITION FOR MEASUREMENT POSITIONING THE CUFF

More information