Modular Analytical Panel System with Sample preparation for Steam and Power Applications
|
|
- Cassandra Stevens
- 5 years ago
- Views:
Transcription
1 Special Documentation Modular Analytical Panel System with Sample preparation for Steam and Power Applications Catalogue and Operating Instructions SD_Modular_Panels_Power_V2.0.doc/04_12
2 Table of Contents: 1 SAFETY INSTRUCTIONS DESIGNATED USE INSTALLATION, COMMISSIONING AND OPERATION OPERATIONAL SAFETY RETURN SAFETY ICONS OVERVIEW OF AVAILABLE PANELS GENERAL INFORMATION SAMPLE PREPARATION PANELS SL-100-L/H C-A /72 SL-200-L/H C- A / SINGLE LINE SAMPLE PANEL SL100L/H C-A / SINGLE LINE SAMPLE PANEL SL200L/H C-A / SAMPLE PREPARATION PANELS PE 3A C-A PE 2A C-A SAMPLE PREPARATION PANEL PE 3A (WATER) C-A SAMPLE PREPARATION PANEL PE 2A (STEAM) C-A SAMPLE PREPARATION PANELS KEP02 C-A KEP05-4 C-A KEP05-8 C-A SAMPLE PREPARATION PANEL KEP02 (WATER) C-A SAMPLE PREPARATION PANEL KEP05-4 (STEAM) C-A SAMPLE PREPARATION PANEL KEP05-8 (STEAM) C-A CONDUCTIVITY / DISSOLVED OXYGEN / PH PANEL C-A / CONDUCTIVITY PANEL C-A CONDUCTIVITY + DISSOLVED OXYGEN OR CONDUCTIVITY + PH C-A DISSOLVED OXYGEN + PH C-A CATIONIC AND DIFFERENTIAL CONDUCTIVITY PANEL C-A CATIONIC AND DIFFERENTIAL CONDUCTIVITY PANEL C-A / CA71 ANALYZER PANEL C-A SODIUM (NA) ANALYZER PANEL C-A PRELIMINARY DESIGN! ANALYZER FOR DEGASSED CATIONIC CONDUCTIVITY C-A ANALYZER FOR DEGASSED CATIONIC CONDUCTIVITY C-A ACCESSORIES FOR MOUNTING PLATES ALUMINIUM PROFILE FRAME C-A WEATHER PROTECTION ROOF C-A DRAIN GUTTER C-A IDENTIFICATION NAME PLATE SCOPE OF DELIVERY INSTALLATION INCOMING ACCEPTANCE, TRANSPORT, STORAGE...33 Endress+Hauser 2
3 5.2 INSTALLATION REQUIREMENTS CONNECTIONS ELECTRICAL CONNECTION MEDIUM CONNECTION CONNECTION BETWEEN TWO OR MORE PANELS OPERATION MAINTENANCE GENERAL ACCESSORIES TROUBLE SHOOTING TECHNICAL DATA PROCESS CONDITIONS...40 Endress+Hauser 3
4 1 Safety Instructions 1.1 Designated Use The Modular Analytical Mounting Plate System is a sampling system for measuring ph, ORP, conductivity, cationic conductivity or dissolved oxygen in water and steam circuits of power plants in NON-EX area. Depending on the used Sample Preparation Panel up to three panels can be easily connected to a full analytical system. See chapter 2.2 and 2.3 for further details. The mounting panels shall be wall mounted in dry rooms protected from frosty ambient conditions. They can also be installed in monitoring containers. Main area of application: Quality measurements in water and steam circuits of power plants, heating plants and incineration plants. Features and benefits: compact, vertical arrangement and easy connection of plates reliable operation as special sensor properties are taken into account in the layout. choice of different sample preparation panels for different demands max. pressure at sample inlet: up to 245 bar depending on the sample panel max. temperature at sample inlet: up to 550 C depending on the sample panel optional thermal shut off valve in sample line after cooler to protect operating personal and analyzers optional flow meter with switching contact or flow indicator for cooling water circuit adjustable flow rate for every measurement parameter optional flow meter with switching contact at every sample line integration of Third Party components in the system on request 1.2 Installation, Commissioning and Operation Please note the following items: Installation, commissioning, operation and maintenance of the measuring system must only be carried out by trained technical personnel. The technical personnel must be authorized for the specified activities by the system operator. Electrical connection must only be carried out by a certified electrician. Technical personnel must have read and understood these Operating Instructions and must adhere to them. Before commissioning the entire measuring point, check all the connections for correctness. Ensure that electrical cables and hose connections are not damaged. Before commissioning check all fittings and retighten them if necessary to avoid leakages. Do not operate damaged products and secure them against unintentional commissioning. Mark the damaged product as being defective. Endress+Hauser 4
5 Measuring point faults may only be rectified by authorized and specially trained personnel. If faults can not be rectified, the products must be taken out of service and secured against unintentional commissioning. Repairs are not described in these Operating Instructions and may only be carried out at the manufacturer or by the service organisation. 1.3 Operational Safety The Analytical Panel has been designed and tested according to the state of the art and left the factory in perfect functioning order. Relevant regulations and European standards have been met. As the user, you are responsible for complying with the following safety conditions: EMC Installation instructions Local prevailing standards and regulations. The instruments mounted on the panels have been tested for electromagnetic compatibility in industrial use according to applicable European standards. Protection against interference as specified above is valid only for an instrument connected according to the instructions in these Standard Operating Instructions. 1.4 Return If a mounting plate has to be repaired, please return it cleaned to the sales center responsible. Please use the original packaging, if possible. 1.5 Safety Icons # Warning! This symbol alerts you to hazards. They can cause serious damage to the instrument or to persons if ignored. " Caution! This symbol alerts you to possible faults which could arise from incorrect operation. They could cause damage to the instrument if ignored.! Note! This symbol indicates important items of information. Endress+Hauser 5
6 2 Overview of Available Panels 2.1 General Information Tubing: Stainless Steel tubing (316 or 316L): 25 mm or ½ inch for cooling water lines, 12 mm or ¼ inch sample in- and outlet lines, 6 mm sample lines to flow vessels, Swagelok fittings (316) and clamping ring connections Mounting plates: Stainless Steel (304) or partially powder coated aluminum Media: Boiler feed water, condensate, heating water, steam etc. Dimensions: Sample preparation plate Sentry SL100/200: H x W x D: 1000 x 305mm Sample preparation plate Thiedig PE02/PE03: H x W x D: 1000 x 450mm Analysis plates: H x W x D: 1000 x 300mm Sample preparation plate Woesthoff KEP02/KEP05: H x W x D: 1000 x 500mm Analysis plates: H x W x D: 1000 x 400mm Features: Sensor and transmitter selectable for best performance must be ordered separately with a suitable cable Depending on the type of Sample Preparation Panel up to three panels can easily be connected. Flow meters or flow guards with switching contact can be ordered as an option. This allows to adjust an appropriate flow on each panel.! Note! All standard components like sensors, assemblies (if mentioned), cables, transmitters etc. are not included in the scope of delivery of the mounting plate. They have to be ordered separately! Endress+Hauser 6
7 2.2 Sample Preparation Panels SL-100-L/H C-A /72 SL-200-L/H C-A /84! Note! The panel containing the cooler and the pressure reducing device must be used always as the first panel in a row. These panels are intended to be used for different pressure and temperature sample conditions, primary for water or condensate. Depending on the type, one (with SL100) respectively up to three (with SL200) analytical panels can be attached. Figure 1.: Example of a sample conditioning panel with different options (exemplary, may vary with the order), brand Sentry Endress+Hauser 7
8 2.2.1 Single Line Sample Panel SL100L/H C-A /72! Note! The panel containing the cooler and the pressure reducing device must be always used as the first panel in a row. The designated use is grab sample only, it can be used with one further attached analytical panel with low (L-version) and high (H-version) pressure and temperature sample inlet conditions after check up in particular case. optional optional optional optional Figure 2.: P&I scheme of the sample conditioning panel SL100L/H with different options (may vary with the order). Ordering Information SL100L SL100H Process conditions Sample Inlet P max. = 34 bar; T max. = 232 C P max. = bar; T max. = 538 C For further details and optional components please refer to the Single Line Sample Panel Manual SPD which is available in the TSP modification database. The process data of the cooling water as pressure, T in, T out/max and Flowrate must be provided. This defines the operational range and the type of the cooler. Endress+Hauser 8
9 2.2.2 Single Line Sample Panel SL200L/H C-A /84! Note! The panel containing the cooler and the pressure reducing device must be always used as the first panel in a row. It can be used with a max. of three further attached analytical panel with low (L-version) and high (Hversion) pressure and temperature sample inlet conditions. optional optional optional optional optional optional optional optional Figure 3.: P&I scheme of the sample conditioning panel SL200L/H with different options (may vary with the order). Ordering Information SL200L SL200H Process conditions Sample Inlet P max. = 34 bar; T max. = 232 C P max. = bar; T max. = 538 C For further details and optional components please refer to the Single Line Sample Panel Manual SPD which is available in the TSP modification database. The process data of the cooling water as pressure, T in, T out/max and Flowrate must be provided. This defines the operational range and the type of the cooler. Endress+Hauser 9
10 2.3 Sample preparation panels PE 3A C-A PE 2A C-A ! Note! The panel containing the cooler and the pressure reducing device must be always used as the first panel in a row. cooling water sight glass (option) Thermometer protection insulation temperature shut off valve (option) inlet blow down valve (option) back pressure regulator (option) flow guard (option) back plate coated aluminium connection to analysis panel (Swagelok compression fitting for 12mm tube) Figure 4.: Example of a high temperature sample preparation panel with steam cooler, brand Dr. Thiedig, with optional accessories Endress+Hauser 10
11 Features and benefits: high pressure inlet valve, welded at sample inlet and with protection cover needle valve for pressure reduction to ensure constant sample flow to analysis a backpressure regulator is available as option cooler material or for high temperature application aluminium back plate with powder coating Available documentation on request (with additional costs): order related drawings certificate of compliance with the order (EN ) pressure test and functional test for the high-pressure valves (EN ) material certificate for the valve body (EN ) pressure test for the cooler (EN ) material certificate for the coil (EN ) Advantages: High-Pressure Sample Cooler higher cooling capacity, reduced pressure drop in the sample line and cooling water consumption the capacity of the coolers is designed to 60 l/h in accordance with VGB 450 coolers are fabricated free of brass and are 100% austenitic steel counter flow principle for cooling water and sample cooling water flow of 1,6m³/h (steam cooler) and 1,1m³/h (water cooler) leads to sample outlet temperature of 5-10K above the cooling water inlet temperature cooler can be maintained and cleaned in a short time without complete dismounting in place coolers can be opened without de-installation cooling coils made of thermal and chemical resistant Austenite (Alloy 28) for temperatures > 400 C High-Pressure Valves robust needle valves for cold pressure reduction up to 320 bar with 200 C (PN400) robust inlet valves with differential thread for handling minimized wear and secure shut-off by non-turning spindle thread of the spindle is not in contact with the sample media identical internal diameters of inlet valve and inlet sample line, this reduces pressure gap and fouling stuffing boxes can be adjusted without opening the valves forged valve bodies and corrosion resistant spindles lead to extended life-time valve bodies made of heat resistant Austenite for temperature > 400 C Endress+Hauser 11
12 System and material of panels PE02 / PE03 equipment for inlet temperatures of up to 630 C all high pressure components are tested separately only designed for hot blow down before coolers to achieve the required capacities and acceleration standards for bending ranges according design temperature and pressure are recognized Made in Germany high pressure components seal metal to metal for stable sealing low pressure gap in the sampling line for optimal supply of the analyzers system design according to VGB 450L and ASTM designation design according ergonomic principles for improved safety of operation personal touch guard for lines and valve bodies upstream cooling welded inlet lines, valves and coolers welding testing according AD 2000 and EN / possible wear parts of the valves exchangeable without demounting Austenite (316 Ti) for the high pressure tubing and all high pressure components in contact to the sample Endress+Hauser 12
13 2.3.1 Sample preparation panel PE 3A (water) C-A cooling water inlet valve thermometer pressure reducing valve > needle valve for constant flow backplate aluminum; stainless steel as option Figure 5.: sample preparation panel with water cooler PE03 (without additional options) Ordering Information Optional cooling water sight glass, flow guard, inlet blow down valve, back pressure regulator, temperature shutoff 55 C, backplate stainless steel Process conditions Sample inlet P max. = 245 bar; T max. = 250 C cooling water inlet P max. = 16 bar; T max. = 100 C All high pressure tubes are welded, low pressure tubes are connected with Swagelok compression fittings. Welding certificates according to the European Pressure Equipment Directive 97/23/EG and DIN ISO plus inspection certificates EN can be ordered additionally. The process data of the cooling water as pressure, T in, T out/max and Flowrate must be provided. This defines the operational range. accessories please see chapter 3 Endress+Hauser 13
14 2.3.2 Sample preparation panel PE 2A (steam) C-A option: cooling water sight glass option: inlet blow down valve backplate coated aluminium; stainless steel as option option: temperature shutoff valve (55 C) option: backpressure regulator option: flow guard with limit switch Figure 6.: sample preparation panel with steam cooler PE02 (picture shows additional options) Ordering Information Optional cooling water sight glass, flow guard, inlet blow down, back pressure regulator, temperature shutoff 55 C, backplate stainless steel, high temperature version Process conditions Sample inlet P max. = 245 bar; T max. = 400 C / P max. = 245 bar; T max. = 550 C (high temperature version) cooling water inlet P max. = 16 bar; T max. = 100 C All high pressure tubes are welded, low pressure tubes are connected with Swagelok compression fittings. Welding certificates according to the European Pressure Equipment Directive 97/23/EG and DIN ISO plus inspection certificates EN can be ordered additionally. The process data of the cooling water as pressure, T in, T out/max and Flowrate must be provided. This defines the operational range. accessories please see chapter 3 Endress+Hauser 14
15 2.4 Sample preparation panels KEP02 C-A KEP05-4 C-A KEP05-8 C-A ! Note! The panel containing the cooler and the pressure reducing device must be always used as the first panel in a row. sample inlet, Option: 2nd and 3rd inlet valve Options: cooling water sight glass; sample filter; blow down valve back panel TRESPA; Option: stainless steel Options: flow meter; mechanical temperature shutoff valve; pressure gauge Option: pressure relief valve in cooling water inlet Option: backpressure regulator with manual grab sample Figure 7.: Example of a high temperature sample preparation panel with steam cooler, brand Woesthoff, with optional equipment Endress+Hauser 15
16 2.4.1 Sample preparation panel KEP02 (water) C-A Figure 8.: sample preparation panel with water cooler G02S (picture shows additional options) Ordering Information Optional e.g. backplate stainless steel, cooling water sight glass, flow guard, inlet blow down, back pressure regulator, temperature shutoff 50 C, second sample inlet valve (please see TSP data base for full options list) Process conditions Sample inlet P max. = 65 bar; T max. = 200 C cooling water inlet P max. = 16 bar; T max. = 100 C All tubes are connected with Swagelok compression fittings. Inspection certificates EN can be ordered additionally. The process data of the cooling water as pressure, T in, T out/max and Flowrate must be provided. This defines the operational range. accessories please see chapter 3 Endress+Hauser 16
17 2.4.2 Sample preparation panel KEP05-4 (steam) C-A Figure 9.: sample preparation panel with water cooler G05S-4 (picture shows additional options) Ordering Information Optional e.g. backplate stainless steel, cooling water sight glass, flow guard, inlet blow down, back pressure regulator, temperature shutoff 50 C, second sample inlet valve (please see TSP data base for full options list) Process conditions Sample inlet P max. = 160 bar; T max. = 400 C / P max. = 320 bar; T max. = 400 C (high pressure version) cooling water inlet P max. = 16 bar; T max. = 100 C All tubes are connected with Swagelok compression fittings. Inspection certificates EN can be ordered additionally. The process data of the cooling water as pressure, T in, T out/max and Flowrate must be provided. This defines the operational range. accessories please see chapter 3 Endress+Hauser 17
18 2.4.3 Sample preparation panel KEP05-8 (steam) C-A Figure 10.: sample preparation panel with water cooler G05S-8 (picture shows additional options) Ordering Information Optional e.g. backplate stainless steel, cooling water sight glass, flow guard, inlet blow down, back pressure regulator, temperature shutoff 50 C, second sample inlet valve (please see TSP data base for full options list) Process conditions Sample inlet P max. = 250 bar; T max. = 550 C / P max. = 250 bar; T max. = 650 C (high temperature version) cooling water inlet P max. = 16 bar; T max. = 100 C All tubes are connected with Swagelok compression fittings. Inspection certificates EN can be ordered additionally. The process data of the cooling water as pressure, T in, T out/max and Flowrate must be provided. This defines the operational range. accessories please see chapter 3 Endress+Hauser 18
19 2.5 Conductivity / Dissolved Oxygen / ph panel C-A / -71 Pmax=4bar Tmax=80 C Figure 11.: Conductivity/DO panel. Ordering Information Conductivity Dissolved Oxygen or ph Sensor CLS15D-#1A1 COS21D-C11## or CPS11D-7AS21 or CPS41D-7AB2B1 with CPY7-B0 (not included) Cable CYK10-A881 (1,5m) CYK10-A881 (1,5m) Assembly Included (SAP ) Included (SAP ) Transmitter CM442-########### Optional VA-Flow meter for sample See chapter 2.1 for further details. 0,1 0,6 L/min, switching contact (Non-Ex, 200 V, 1 A, 20 VA, IP65) accessories please see chapter 3 Endress+Hauser 19
20 2.6 Conductivity panel C-A Pmax=4bar Tmax=80 C Figure 12.: Conductivity panel. Ordering Information Conductivity Sensor CLS21D-C1E1 Cable CYK10-A881 (1,5m) Assembly Included (SAP ) Transmitter CM442-########### Optional VA-Flow meter for sample See chapter 2.1 for further details. 0,1 0,6 L/min, switching contact (Non-Ex, 200 V, 1 A, 20 VA, IP65) accessories please see chapter 3 Endress+Hauser 20
21 2.7 Conductivity + Dissolved Oxygen or Conductivity + ph C-A Pmax=4bar Tmax=80 C Figure 13.: Conductivity / DO or conductivity / ph panel. Ordering Information Conductivity Dissolved Oxygen or ph Sensor CLS15D-#1A1 COS21D-C11## or CPS11D-7AS21 or CPS41D-7AB2B1 with CPY7-B0 (not included) Cable CYK10-A881 (1,5m) CYK10-A881 (1,5m) Assembly Included (SAP ) Included (SAP ) Transmitter CM442-##M2####### Optional VA-Flow meter for sample 0,1 0,6 L/min, switching contact (Non-Ex, 200 V, 1 A, See chapter 2.1 for further details. 20 VA, IP65) accessories please see chapter 3 Endress+Hauser 21
22 2.8 Dissolved Oxygen + ph C-A Pmax=4bar Tmax=80 C Figure 14.: DO / ph panel Ordering Information Dissolved Oxygen ph Sensor COS21D-C11## CPS11D-7AS21 or CPS41D-7AB2B1 with CPY7-B0 (not included) Cable CYK10-A881 (1,5m) CYK10-A881 (1,5m) Assembly Included (SAP ) Included (SAP ) Transmitter CM442-##M2####### Optional VA-Flow meter for sample 0,1 0,6 L/min, switching contact (Non-Ex, 200 V, 1 A, See chapter 2.1 for further details. 20 VA, IP65) accessories please see chapter 3 Endress+Hauser 22
23 2.9 Cationic and Differential Conductivity panel C-A CM442 (not included) backplate Trespa; stainless steel as option CLS15D (not included) flow indicator as option cation filter with assembly for 2x CLS15D in price included Pmax=4bar Tmax=55 C free outlet Figure 15.: Cationic and Differential Conductivity panel. Ordering Information Sensor Cable Assembly Transmitter Option 2 x CLS15D-#1A1 for differential conductivity 2 x CYK10-A881 (1,5m) Included in cation filter (0.6 litre volume); resin included; flow indicator in cation filter included CM442-##M2####### backplate stainless steel, larger filter housing for 1.2L resin, flow indicator accessories please see chapter 3 Endress+Hauser 23
24 2.10 Cationic and Differential Conductivity panel C-A / 81 CM442 (not included) backplate coated aluminium; stainless steel as option CLS15D (not included) cation filter with flow indicator and assembly for 2x CLS15D in price included Pmax=3,2bar Tmax=60 C free outlet Figure 16.: Cationic and Differential Conductivity panel. Ordering Information Sensor 2 x CLS15D-#1A1 for differential conductivity 1 x CLS15D-#1A1 for cationic conductivity Cable 2 x CYK10-A881 (1,5m) 1 x CYK10-A881 (1,5m) Assembly Included in cation filter (1.5 litre volume); resin included; flow indicator in cation filter included Transmitter CM442-##M2####### CM442-##M1####### Option backplate stainless steel backplate stainless steel accessories please see chapter 3 Endress+Hauser 24
25 2.11 CA71 analyzer panel C-A Pmax inlet=10bar Tmax=45 C Figure 17.: CA71 analyzer panel (picture shows optional 2-channel version) Ordering Information Analyzer Optional VA-Flow meter for sample See chapter 2.1 for further details. 1-channel version: CA71##-#1#B3## 2-channel version: CA71##-#2#D3## 0,1 0,6 L/min, switching contact (Non Ex, 200 V, 1 A, 20 VA, IP65) accessories please see chapter 3 Endress+Hauser 25
26 2.12 Sodium (Na) analyzer panel C-A Preliminary design! Na-Analyzer Waltron 9032 Pmax=2bar Tmax=55 C Figure 18.: Design draft of Sodium analyzer panel. Ordering Information Analyzer Optional VA-Flow meter for sample See chapter 2.1 for further details. Sodium Analyzer Waltron 9032 (included) analysis range: 0,1ppb 10ppm 0,1 0,6 L/min, switching contact (Non Ex, 200 V, 1 A, 20 VA, IP65) accessories please see chapter 3 Endress+Hauser 26
27 2.13 Analyzer for degassed cationic conductivity C-A Figure 19.: Picture and function principle of DAC analyzer Ordering Information Analyzer Options C-A Degassed cationic conductivity DAC analysis range: 0-200mS/cm, ph VAC 50/60 Hz, 50VA assembling on a stainless steel panel 1000x800x3mm for rack mounting Endress+Hauser 27
28 2.14 Analyzer for degassed cationic conductivity C-A Figure 20.: Picture of DAC analyzer panel Ordering Information Analyzer Remark Options Degassed cationic conductivity DAC 230 VAC 50Hz; HxW=1000mm x 600mm works only in combination with pre-conditioning with cation filter (e.g. panel ) stainless steel back plate Endress+Hauser 28
29 3 Accessories for Mounting Plates 3.1 Aluminium Profile Frame C-A This frame can be used when the panels shall be wall mounted. Variable width Y from 300mm 1200mm Figure 21. : Aluminium back frame for one or up to four modular panels. Ordering Information Back Frame made of Aluminium Remarks Required width Y must be indicated with the order configure rack wise! Endress+Hauser 29
30 3.2 Weather Protection Roof C-A Variable width Y from 300mm 500mm Figure 22. : Weather protection roof for transmitter (for one or max. two panels in a row). Ordering Information Weather Protection Roof Remarks Required width Y must be indicated with the order Endress+Hauser 30
31 3.3 Drain Gutter C-A Variable width Y from 300mm 1200mm Figure 23. : Drain gutter for sample or cooling water outlet, drain on one side, for direct mounting on plate or offset mounting on rack (for one or up to four panels in a row). Figure 24. : Optional tray for sample container, which can be placed in the gutter. Ordering Information Drain Gutter Optional holder for sample container (stainless W x H x D = 150mm x 60mm x 92mm steel) Remarks Required width Y must be indicated with the order configure rack wise Endress+Hauser 31
32 4 Identification 4.1 Name Plate You can find the following name plate on each mounting plate: Figure 25.: Name plate. Endress+Hauser Conducta GmbH+Co. KG Dieselstrasse Gerlingen, Germany Modification-No. C-########### Serial Number: Year of Construction: 4.2 Scope of delivery Please see the modifications in the TSP-database for the scope of delivery. In general the scope of delivery for a typical mounting plate comprises the following components: A Stainless Steel mounting plate The complete SS-tubing (12 and 6 mm) and different connection parts Cable duct the mounting of all components like tubing, transmitter, flow assembly etc.! Note! All standard components like sensors, cables, transmitters etc. are not part of the delivery scope of the mounting plate. They have to be ordered separately! " Caution! The sensors are not installed in the flow assemblies for shipment. Endress+Hauser 32
33 5 Installation! Note! For installation, commissioning and operation of the SL100/200-L/H panels, the PE02A/PE03A panels and the panels with third party products, see the separate documentation which comes with these panels and can be found in the TSP modification database. 5.1 Incoming Acceptance, Transport, Storage Make sure the packaging is undamaged! Inform the supplier about damage to the packaging. Keep the damaged packaging until the matter has been settled. Make sure the contents are undamaged! Inform the supplier about damage to the delivery contents. Keep the damaged products until the matter has been settled. Check that the scope of delivery is complete and agrees with your order and the shipping documents. The packaging material used to store or to transport the product must provide shock protection and humidity protection. The original packaging offers the best protection. Also, keep to the approved ambient conditions (see "Technical Data Section"). If you have any questions, please contact your supplier or your sales centre responsible. 5.2 Installation Requirements " Attention! The mounting panels shall be wall mounted in dry rooms (Non-Ex-areas) protected from frosty ambient conditions. They can also be installed in monitoring containers or on free standing racks. Endress+Hauser 33
34 6 Connections! Note! For installation, commissioning and operation of the SL100/200-L/H panels, the PE02A/PE03A panels, the KEP02/KEP05 panels and the panels with third party products, see the separate documentation which comes with these panels and can be found in the TSP modification database. 6.1 Electrical Connection Please read the operating instructions of the mounted transmitter for connection the power supply and the sensors.! Note! To wire the switching contacts of the optional flow meters, you can loosen the fittings of the flow meter and rotate them 90 degrees. Then you can wire the angle plug of the flow meter. Rotate the flow meter back into the original position and retighten the fittings again. Please assure via a pressure test that the connection is leak-proof. Ensure that the protective earth is connected. " Caution! Before you switch on the power supply, ensure that all transmitters are correctly connected to the power supply. 6.2 Medium Connection " Caution! Before connecting the medium supply, check that all sensors are properly installed and connected to the transmitters. Check if cooling water and sample is available. If a sample preparation panel is used please read the operating instruction of this panel carefully. Always turn on the cooling water first then turn on the medium. 6.3 Connection between two or more panels All components to connect the sample conditioning panels to a measuring panel (e.g. C- A , C-A , C-A ) are in the scope of delivery. Connect the sample conditioning panel with fittings for 12 mm pipes to a measuring panel. The main sample line is blocked at the right end. To connect a further panel to the right, this blind plug must be removed. Endress+Hauser 34
35 7 Operation! Note! For installation, commissioning and operation of the SL100/200-L/H panels, the PE02A/PE03A panels, the KEP02/KEP05 panels and the panels with third party products, see the separate documentation which comes with these panels and can be found in the TSP modification database.! Note! The operation of the instruments is described in detail in the Operating Instructions of the instruments. Please see the appropriate Operating Instruction! Please ensure the max. allowed process pressure, process temperature, flow rate and ambient conditions comply with the requirements of the different instruments.! Note! Each mounting plate has an inlet needle valve which allows to separate each measurement line. To separate a measurement line, e.g. for maintenance, first close the valve of the incoming measurement line. The flow can be adjusted with the inlet needle valve. The optional flow meter allows measuring the flow and offering an electrical switch contact, which indicates a flow below the minimum. # Warning! For max. pressure and temperature conditions for the media at the sample inlet, refer to the technical data in chapter 11! The outlet of the measurement plates should have no counter pressure. Endress+Hauser 35
36 ! Note! For installation, commissioning and operation of the SL100/200-L/H panels, the PE02A/PE03A panels, the KEP02/KEP05 panels and the panels with third party products, see the separate documentation which comes with these panels and can be found in the TSP modification database. Endress+Hauser 36
37 8 Maintenance " Caution! Avoid that sand, fibres and other abrasive material can enter the system!! Note! Please follow the maintenance recommendations in the Operating Instructions of the sensors, cables, instruments, analyzers and assemblies. Endress+Hauser 37
38 9 General accessories! Note! For accessories please see the Operating Instructions of the standard products. For the mounting plates are several accessories available. Please see chapter 3. Endress+Hauser 38
39 10 Trouble shooting! Note! Please follow the trouble shooting recommendations in the Operating Instructions of the standard products (e.g. instruments). Endress+Hauser 39
40 11 Technical Data 11.1 Process Conditions! Note! This chapter describes only the technical data of the mounting panels. Concerning the technical data of the assemblies, sensors, cables, instruments and analyzers, please refer to the operating instructions of these products. Panel Sample Conditioning Panel Low Pressure Panel Max. pressure sample inlet 34 bar Pressure at the sample inlet before pressure reduction valve Max. sample inlet temperature 232 C Sample inlet temperature before cooler Panel Sample Conditioning Panel High Pressure Panel Max. pressure sample inlet bar Pressure at the sample inlet before pressure reduction valve Max. sample inlet temperature 538 C Sample inlet temperature before cooler Panel Sample Conditioning Panel for water / condensate High Pressure / Low Temp. Max. pressure sample inlet 245 bar Pressure at the sample inlet before pressure reduction valve Max. sample inlet temperature 250 C Sample inlet temperature before cooler Panel Sample Conditioning Panel for steam High Pressure / High Temp. Max. pressure sample inlet 245 bar Pressure at the sample inlet before pressure reduction valve Max. sample inlet temperature 400 C / 550 C Sample inlet temperature before cooler Panel Sample Conditioning Panel for water / condensate High Pressure / High Temp. Max. pressure sample inlet 65 bar Pressure at the sample inlet before pressure reduction valve Max. sample inlet temperature 200 C Sample inlet temperature before cooler Endress+Hauser 40
41 Panel Sample Conditioning Panel for water / condensate High Pressure / High Temp. Max. pressure sample inlet 160 / 320 bar Pressure at the sample inlet before pressure reduction valve Max. sample inlet temperature 400 C Sample inlet temperature before cooler Panel Sample Conditioning Panel for water / condensate High Pressure / High Temp. Max. pressure sample inlet 250 / 320 bar Pressure at the sample inlet before pressure reduction valve Max. sample inlet temperature 550 C / 650 C Sample inlet temperature before cooler Panel / Conductivity (CLS15D) or DO (COS21D) or ph (CPS11D, CPS41D) panel Max. pressure sample inlet 4 bar Pressure at the sample inlet Max. sample inlet temperature 80 C Sample inlet temperature Max. counter pressure outlet free outlet Panel Conductivity (CLS21D) panel Max. pressure sample inlet 4 bar Pressure at the sample inlet Max. sample inlet temperature 80 C Sample inlet temperature Max. counter pressure outlet free outlet Panel Conductivity (CLS15D) + DO (COS21D) or Conductivity (CLS15D) + ph (CPS11D, CPS41D) Max. pressure sample inlet 4 bar Pressure at the sample inlet Max. sample inlet temperature 80 C Sample inlet temperature Max. counter pressure outlet free outlet Panel DO (COS21D) + ph (CPS11D, CPS41D) Max. pressure sample inlet 4 bar Pressure at the sample inlet Max. sample inlet temperature 80 C Sample inlet temperature Max. counter pressure outlet free outlet Endress+Hauser 41
42 Panel Cationic / diff. conductivity panel (CLS15D) Max. pressure sample inlet 4 bar Pressure at the sample inlet Max. sample inlet temperature 70 C Sample inlet temperature Max. counter pressure outlet free outlet Panel / Cationic / diff. conductivity panel (CLS15D) Max. pressure sample inlet 3,2 bar Pressure at the sample inlet Max. sample inlet temperature 60 C Sample inlet temperature Max. counter pressure outlet free outlet Panel CA71## Analyzer panel Max. pressure sample inlet 10 bar Pressure at the sample inlet Max. sample inlet temperature 40 C Sample inlet temperature Max. counter pressure outlet free outlet Panel Sodium analyzer panel Max. pressure sample inlet 2 bar Pressure at the sample inlet Max. sample inlet temperature 55 C Sample inlet temperature Max. counter pressure outlet free outlet Panel C-A Degassed conductivity Max. pressure sample inlet on request Pressure at the sample inlet Max. sample inlet temperature on request Sample inlet temperature Max. counter pressure outlet free outlet Panel Degassed conductivity Max. pressure sample inlet on request Pressure at the sample inlet Max. sample inlet temperature on request Sample inlet temperature Max. counter pressure outlet free outlet General information Power Supply Depending on transmitter and analyzer type Ambient temperature range + 2 to + 40 C Humidity Max. 95 % relative humidity, non-condensing Endress+Hauser 42
43
44 Endress+Hauser 44
Technical Information Conducal CLY421
TI00496C/07/EN/01.12 71142150 Products Solutions Services Technical Information Conducal CLY421 Conductivity calibration set for ultrapure water applications Factory calibrated measurement box Certified
More informationOperating Instructions Flowfit CUA252
BA01281C/07/EN/02.15 71284726 Products Solutions Services Operating Instructions Flowfit CUA252 Flow assembly for CUS52D turbidity sensor Document information Warnings The structure, signal words and safety
More informationSiphon vessel / Priming aid
Manufacturer: sera GmbH sera-straße 1 34376 Immenhausen Germany Tel.: +49 5673 999-00 Fax: +49 5673 999-01 info@sera-web.com Keep the operating manual for future use! Record the exact type and serial number
More informationMeasurement accessories METPOINT OCV for the measurement in systems up to 40 bar
EN - english Instructions for installation and operation Measurement accessories METPOINT OCV for the measurement in systems up to 40 bar Dear customer, Thank you for deciding in favour of the METPOINT
More informationRS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.
RS(H)10,15 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to
More informationGM Series Dual-Block Multi-Function Gas Control Valves
Installation Sheets Manual 121 Gas Combustion Combination Controls and Systems Section G Technical Bulletin GM Issue Date 0297 GM Series Dual-Block Multi-Function Gas Control Valves Figure 1: GM Series
More information6301 TYPE CAST IRON SAFETY VALVES
Pressure (bar) 6301 TYPE CAST IRON SAFETY VALVES FEATURES The 6301 type safety valve is a device designed to protect installations against possible overpressure. It operates automatically and closes when
More informationInstruction Manual Differential Pressure Transmitter
DE61 Instruction Manual Differential Pressure Transmitter Table of Contents 1. Safety Instructions 2. Intended Applications 3. Product Description and Functions 4. Installation 5. Commissioning 6. Maintenance
More informationDK48 - DK800. Quick Start. Variable area flowmeter
Quick Start Variable area flowmeter General safety notes You can find the newest and/or additional information in the handbook, the datasheet, special manuals, certificates and the download center at www.krohne.com.
More informationOPERATING INSTRUCTIONS MANUAL FOR QACV PNEUMATIC DOSING PUMP
This operating instructions contains safety information that if ignored can endanger life or result in serious injury. They are indicated by this icon. Use of this pump with radioactive chemicals is forbidden!
More informationOUS31. Technical Information
Technical Information OUS Turbidity sensor Installation and immersion sensor for drinking water and industrial water according to the 90 scattered light method Application All phases of drinking water
More informationInstruction Manual Contact Pressure Vacuum Gauge
MS10 Instruction Manual Contact Pressure Vacuum Gauge Table of Contents 1. Safety Instructions 2. Intended Applications 3. Product Description and Functions 4. Installation 5. Commissioning 6. Maintenance
More informationOperating Instructions. Ball valve fitting according to ZB For pressure transmitter VEGABAR 82. Document ID: 50027
Operating Instructions Ball valve fitting according to ZB 2553 For pressure transmitter VEGABAR 82 Document ID: 50027 Contents Contents 1 About this document 1.1 Function... 3 1.2 Target group... 3 1.3
More informationLIQUID LEVEL GAUGE FOR WELDING ZGAU 716 O&M, 717 O&M, 718 O&M, 719 O&M
LIQUID LEVEL GAUGE FOR WELDING ZGAU 7 O&M, 77 O&M, 78 O&M, 79 O&M Edition: 07/0 Date: 0.07.0 GO WITH OUR FLOW SPIS TREŚCI. Introduction. Construction. Installation. Installation of liquid level gauge of
More informationMiniature flowmeters DK 46, DKR 46 DK 47, DK 48, DK 800
KROHNE 1/22 7.236.22. GR Miniature flowmeters DK 46, DKR 46 DK 47, DK 48, DK 8 Variable area flowmeters Vortex flowmeters Flow controllers Electromagnetic flowmeters Ultrasonic flowmeters Mass flowmeters
More information2-port and 3-port valves PN 16
OEM 2-port valves VVP459.10-0.63 to VVP459.25-6.3 3-port valves VXP459.10-0.63 to VXP459.25-6.3 3-port valves with bypass VMP459.10-0.63 to VMP459.20-4 2-port valves VVP459.25-6.3 to VVP459.40-25 3-port
More informationbetter measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.
Simply a question of better measurement SCHMIDT Flow Sensor SS 20.261 The cost-effective alternative in pressurised systems up to 10 bars. Compressed air technology Industrial processes A cost analysis
More informationLRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator.
LRS(H)4 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to the
More informationCustomer Responsibilities. Important Customer Information. Cary 4000/5000/6000i UV-Vis spectrophotometer Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationG type DUCTED EXHAUST SAFETY VALVE 2871 AND 288X SERIES. Model/Ref:
Model/Ref: 28713 www.lauridsenindustri.com CHARACTERISTICS The G type safety valves are dedicated to protect the equipment from potential overpressure. They are automatic and close when the pressure conditions
More informationPneumatic spool valve islands
Pneumatic spool valve islands P7.GB.R MEGA SPOOL VALVE ISLANDS MEGA pneumatic spool valve islands offer maximum performance along with flexibility, easy installation and simple operation. Due to their
More informationPressure Relief Valve DHV 718
Advantages frictionless components low maintenance low pressure increase up to fully opened valve constant low vibration controlling hermetically sealed by diaphragm for oscillating pumps for viscous media
More informationSafety. Operating instructions Solenoid valve for gas VG 6 VG 15/10 DANGER. Contents WARNING CAUTION. Changes to edition 09.14
15 Elster GmbH Edition 7.15 Translation from the German 519 D F NL I E DK S N P GR TR CZ PL RUS H www.docuthek.com Operating instructions Solenoid valve for gas VG VG 15/1 Contents Solenoid valve for gas
More information3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289
3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS Item #31289 The EASTWOOD 3 GALLON, OILLESS PANCAKE COMPRESSOR, with an Integral Air Regulator, efficiently supplies all compressed air requirements for
More informationPRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.
PRS(TC)4,8 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to
More informationMounting and Operating Instructions EB 3007 EN. Self-operated Pressure Regulators. Differential Pressure Regulators (opening) Type Type 42-25
Self-operated Pressure Regulators Differential Pressure Regulators (opening) Type 42-20 Type 42-25 Type 42-20 Differential Pressure Regulator Type 42-25 Differential Pressure Regulator Mounting and Operating
More informationOperating Instructions
Operating Instructions Light-metal Ex d enclosures / flameproof enclosure > 8265/0 Empty enclosure > 8265/4 Control panel, integrated in Ex e enclosure > 8265/5 Control panel Table of Contents 1 Table
More informationMounting and Operating Instructions EB EN. Type Supply Pressure Regulator. with increased air capacity
Type 4708-45 Supply Pressure Regulator with increased air capacity Translation of original instructions Mounting and Operating Instructions EB 8546-1 EN Edition March 2016 Note on these mounting and operating
More informationGS-20/25 and GS-40/45 Series Single-block Multi-function Gas Control Valves
Installation Sheets Manual 121 Gas Combustion Combination Controls and Systems Section G Technical Bulletin GS-20/25 GS-40/45 Issue Date 1298 GS-20/25 and GS-40/45 Series Single-block Multi-function Gas
More informationInstruction Manual Differential Pressure Transmitter
DE13 Instruction Manual Differential Pressure Transmitter Table of Contents 1. Safety Instructions 2. Intended Applications 3. Product Description and Functions 4. Installation 5. Commissioning 6. Maintenance
More informationSystem Pressure Manager Standard & System Pressure Manager Plus
System Pressure Manager Standard & System Pressure Manager Plus Installation, Commissioning & Servicing Instructions Note: THESE INSTRUCTIONS MUST BE READ AND UNDERSTOOD BEFORE INSTALLING, COMMISSIONING,
More informationOxymax COS61/COS61D. Technical Information
Technical Information Oxymax / Dissolved oxygen measurement Optical sensor acc. to the fluorescence quenching principle, with or without Memosens protocol Application The continuous measurement of the
More informationHorizontal Bladder Tanks
DATA SHEET Horizontal Bladder Tanks Features UL Listed and FM Approved for use with various ANSUL proportioners and foam concentrates 175 psi (12.1 bar) maximum allowable working pressure (design pressure)
More informationMounting and Operating Instructions EB 3007 EN. Self-operated Pressure Regulators. Type Type Differential Pressure Regulators (opening)
Self-operated Pressure Regulators Type 42-20 Type 42-25 Differential Pressure Regulators (opening) Translation of original instructions Type 42-20 Differential Pressure Regulator Type 42-25 Differential
More informationVariable Area Flow Meters
Variable Area Flow Meters Over sixty years of variable area meter expertise Brooks started designing and manufacturing variable area (VA) meters, also referred to as rotameters, in 1946. Now, with more
More informationModel MTB-ASME Vertical Bladder Tanks
DATA SHEET Model MTB-ASME Vertical Bladder Tanks Features n UL Listed for use with various proportioners and foam concentrates n 175 psi (12.1 bar) maximum allowable working pressure (design pressure)
More informationDK46 - DK800 Supplementary instructions
DK46 - DK800 Supplementary instructions Variable area flowmeter Device category II2G with electrical internals Additional Ex manual KROHNE CONTENTS DK46 - DK800 1 Safety instructions 3 1.1 General... 3
More informationPRESSURE REGULATORS AND FLOWMETERS, ALL MEDICAL GASES
PRESSURE REGULATORS ALL MEDICAL GASES In this section we propose our pressure regulators (FM series, MU and MERCURY ) and our Rs flowmeters for applications other than oxygen therapy (presented in the
More informationCAST IRON SAFETY VALVE TYPE 6301
CHARACTERISTICS The 6301 safety valve is dedicated to protect the equipment from potential overpressure. This is an automatic device that closes when the pressure conditions are back to normal. It is a
More informationOxford Instruments Plasma Technology. PlasmaPro NGP80. Installation Data Sheet. Original Instructions.
Oxford Instruments Plasma Technology www.oxford-instruments.com About this data sheet This data sheet provides basic information about the installation of a tool. Floor and wall loadings Table 1 lists
More informationModel MTB-ASME Vertical Bladder Tanks
DATA SHEET Model MTB-ASME Vertical Bladder Tanks Features n UL Listed for use with various proportioners and foam concentrates n 175 psi (12.1 bar) maximum allowable working pressure (design pressure)
More informationRelated Products: Auto Drain Valve. Model/Specifications. Model
Related Products: Auto Drain Valve /600 Drainage is automatically discharged in a reliable manner, without requiring human operators. Highly resistant to dust and corrosion, operates reliably, and a bowl
More informationSafety Please read and keep in a safe place Operating instructions Linear flow control VFC Linear flow control with actuator IFC
05 Elster GmbH Edition 0.5 Translation from the German 058 D F NL I E DK S N P GR TR CZ PL RUS H www.docuthek.com Operating instructions Linear flow control VFC Linear flow control with actuator IFC Contents
More informationPressure relief valve DHV 716 set range: 0,5-10,0 bar
Pressure relief valve DHV 7 set range: 0,5-0,0 bar Advantage pressure setting possible at any time, also during operation optimum monitoring valves high reproducibility of the set pressure high level of
More informationGENO -Flushing Compressor Product Data Sheet F 02
GENO -flushing compressor 1988 K Fig. 1: GENO -flushing compressor 1988 K Designated application is suitable for the following fields of application: Flushing of pipes with water/air mix prior to the start-up
More informationFlamco Airfix D-E. Installation and operating instructions. 1999, Flamco eps
Airfix D-E 6442.eps GB Installation and operating instructions 1999, Exclusion of liability All operation and maintenance related technical information, data and instructions contained in these installation
More informationSolenoid gas valve. The solenoid gas valve is also applicable as a main gas valve Pilot gas valve for forced draft burners
7 634 VGS1... with AGA67 connector VGS2... with AGA67 connector Solenoid gas valve VGS The solenoid gas valve is also applicable as a main gas valve Pilot gas valve for forced draft burners The VGS and
More informationPressure maintenance valve SDV-P / SDV-P-E
Translation of the original manual Pressure maintenance valve SDV-P / SDV-P-E Assembly and Operating Manual Superior Clamping and Gripping Imprint Imprint Copyright: This manual remains the copyrighted
More informationTHE COMPACT LINE. Mobile power packs in convincing BAUER quality l/min 225 / 330 bar TOUGH RELIABLE WELL ESTABLISHED
THE COMPACT LINE Mobile power packs in convincing BAUER quality 100 140 l/min 225 / 330 bar TOUGH RELIABLE WELL ESTABLISHED COMPACT LINE mobile breathing air compressors for high demands Due to its legendary
More informationContents. Operating instructions. Pressure switch, heavy-duty version Model PSM-520
Operating instructions Contents EN Pressure switch, heavy-duty version Model PSM-520 1. General information 2. Design and function 3. Safety 4. Transport, packaging and storage 5. Commissioning, operation
More informationHigh-performance submersible pressure transmitter For level measurement Model LH-10
Electronic pressure measurement High-performance submersible pressure transmitter For level measurement Model LH-10 WIKA data sheet PE 81.09 Applications Level measurement in rivers and lakes Deep well
More informationPressure Reducing Valve DMV 750 set range: bar
Pressure Reducing Valve DMV 750 set range:.0-6.0 bar Advantage pressure setting possible at any time, also during operation hermetically sealed by valve diaphragm high level of operating safety and long
More informationInstallation, operation & maintenance manual - original version
Installation, operation & maintenance manual - original version AVK gate valves for water and wastewater Series 01, 02, 06, 12, 15, 18, 20, 26, 32, 33, 36, 38, 50, 55 and 636 COPYRIGHT AVK GROUP A/S 2018
More informationVertical Bladder Tanks
DATA SHEET Vertical Bladder Tanks Features UL Listed and FM Approved for use with various ANSUL proportioners and foam concentrates 175 psi (12.1 bar) maximum allowable working pressure (design pressure)
More informationTwo-stage valve type NE
Two-stage valve type NE Product documentation Operating pressure pmax: Flow rate Qmax: 700 bar (High pressure) 80 bar (Low pressure) 25 lpm (High pressure) 180 lpm (Low pressure) D 7161 03-2017-1.0 by
More informationHydrostatic. Continuous level measurement. Overview Page 188 VEGAWELL 52 Page 190 VEGABAR 86, 87 Page 192 Accessory Page 196 AL-EN
Hydrostatic Continuous level measurement 8 Overview Page 188 VEGAWELL 52 Page 190 VEGABAR 86, 87 Page 192 Accessory Page 196 187 Hydrostatic Level and pressure measurement Measuring principle The pressure
More informationInstruction Manual Contact Pressure Gauge
MS11 Instruction Manual Contact Pressure Gauge Table of Contents 1. Safety Instructions 2. Intended Applications 3. Product Description and Functions 4. Installation 5. Commissioning 6. Maintenance 7.
More informationKMW3000C. User Manual. Water Chiller for Xcalibur and Gemini systems. Version 1.6, September 2010
KMW3000C Water Chiller for Xcalibur and Gemini systems User Manual Version 1.6, September 2010 Agilent Technologies XRD Products 10 Mead Road, Yarnton, Oxfordshire. OX5 1QU, UK Tel: +44 (0)1865 291600
More informationBrooks Flow Controllers for Gas and Liquid Service Model FC 8744, Series FC 8800 and 8900
Data Sheet Flow Controllers for Gas and Liquid Service Brooks Flow Controllers for Gas and Liquid Service Model FC 8744, Series FC 8800 and 8900 Model FC 8800 Flow Controller Model 1350/55 with FC 8900
More informationType 0404 / /2-way solenoid valve 2/2-Wege-Magnetventil Électrovanne 2/2 voies Operating Instructions Bedienungsanleitung Manuel d utilisation
Type 0404 / 5404 2/2-way solenoid valve 2/2-Wege-Magnetventil Électrovanne 2/2 voies Operating Instructions Bedienungsanleitung Manuel d utilisation 1 OPERATING INSTRUCTIONS The operating instructions
More informationOFFICINE OROBICHE S.p.A. 1/9
OFFICINE OROBICHE S.p.A. 1/9 INSTRUCTIONS MANUAL FOR LEVEL SWITCHES SERIES 7000 mod. 7250 AND 7400 Electric 1. INSTRUMENT DESCRIPTION Series 7000 level switches have been designed for external lateral
More informationType 1367 High-Pressure Instrument Supply System with Overpressure Protection
Instruction Manual D100343X012 Type 1367 November 2017 Type 1367 High-Pressure Instrument Supply System with Overpressure Protection TYPE 252 FILTER 2ND-STAGE TYPE 67CF FILTER-STYLE REGULATOR INLET TYPE
More informationNeotecha SAPRO - Aseptic sampling valve SV Installation, operation and maintenance instructions
Before installation these instructions must be fully read and understood 2 Safety Please read these instructions carefully. 2.1 General danger potential Not paying attention to this instruction Incautious
More informationValve Proving System VDK 200 A S06*
Valve Proving System VDK 200 A S06* Valve proving system with the following approvals. UL Recognized File # MH17004 CSA Certified File # 1637485 CSA Requirement No. 4-01 (USA) Technical Information Letter
More informationDual Solenoid Gas Valve Installation
Installation IMPORTANT: These instructions are intended as a guide for qualified personnel installing or servicing FLYNN Gas Products. Carefully follow all instructions in this bulletin and all instructions
More informationModel PDT Dewpoint Transmitter
Model PDT Dewpoint Transmitter Instruction Manual Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Tel: +44 1274 733100 Fax: +44 1274 733200 email: mail@amsytems.co.uk web: www.amsystems.co.uk
More informationPTF4 Pivotrol Pump (patented) version Dual Mechanism - Pressure Powered Pump
Local regulations may restrict the use of this product to below the conditions quoted. In the interests of development and improvement of the product, we reserve the right to change the specification without
More informationDifferential pressure controller (PN 16) AHP - return mounting, adjustable setting
Data sheet Differential pressure controller (PN 16) AHP - return mounting, adjustable setting Description DN 15-40 DN 50 DN 65-100 AHP is a self-acting differential pressure controller primarily for use
More informationVGG... VGF... VGH... Gas Valves. Building Technologies HVAC Products VGG... VGF... VGH...
7 66 VGG... VGF... VGH... Gas Valves VGG... VGF... VGH... Single valves of class A for installation in gas trains Safety shutoff valves conforming to EN 161 in connection with SKP... actuators Suitable
More informationModel MTB-ASME Horizontal Bladder Tanks
DATA SHEET Model MTB-ASME Horizontal Bladder Tanks Features n UL Listed and FM Approved for use with various proportioners and foam concentrates n 175 psi (12.1 bar) maximum allowable working pressure
More informationMobile High Pressure Compressor Unit for Compressing Air and Breathing Air. PE 250-TE with compressor control (optional equipment)
Mobile High Pressure Compressor Unit for Compressing Air and Breathing Air Types: PE 200-TE PE 250-TE PE 300-TE Production status: F02 PE 250-TE with compressor control (optional equipment) General Medium
More informationPressure Automated Calibration Equipment
GE Measurement & control Pressure Automated Calibration Equipment Safety Instructions and User Guide - K0447 PACE5000 PACE6000 K0447 Issue No. 9 1 10 1 PACE5000 1 2 3 4 5 PACE6000 2 6 7 8 3 4 5 6 7 8 9
More informationOxymax W COS41. Technical Information. Potentiostatic amperometric two-electrode sensor Long-term stable sensor for water and wastewater applications
Technical Information Oxymax W COS41 Potentiostatic amperometric two-electrode sensor Long-term stable sensor for water and wastewater applications Application The continuous measurement of the dissolved
More informationInstallation and Maintenance Instruction Manual
Installation and Maintenance Instruction Manual Bi-metal Thermometer Model A and E in a configuration - pursuant to ASME B40.200: ##=E#=### or ##=ERT#=### - pursuant to EN 13190: ###=A#=### or ##=ART#=###
More informationBY GREGGERSEN CATA LOG
MEDICAL DEVICES BY GREGGERSEN CATA LOG COMBINATION UNITS t h e r e i s n o s u b s t i t u t e CATALOG COMBINATION UNITS Hawk variants AND THEIR AREAS OF APPLICATION The HAWK PORTABLE EMERGENCY TREATMENT
More informationPROCESS ANALYSERS. SERVOPRO PureGas. Gas Purifier. Operator Manual. Part Number: A Revision: 0 Language: UK English
PROCESS ANALYSERS SERVOPRO PureGas Gas Purifier Operator Manual Part Number: 02005001A Revision: 0 Language: UK English This page is intentionally blank LIST OF CONTENTS 1 DESCRIPTION AND DEFINITIONS...
More informationBIMBAR INFLATABLE PACKERS AND ACCESSORIES
BIMBAR INFLATABLE PACKERS AND ACCESSORIES Geopro supplies a complete range of inflatable packers in nine different diameters from 28 up to 170mm. All our packers made of BIMBAR rubber technology are reinforced
More informationOverview of types. T5-R B2/R B2 en v Subject to changes 1 / 4. k vs (Sequence 2)
Technical data sheet R3015-..-..-B2 / R3020-..-..-B2 Characterised control valves, 6-way, with internal threads Two sequences (cooling/heating) With a rotary actuator 90 Water-side switching or modulating
More informationAirfix. Flamco. Flamco. Pressure expansion vessels for sanitary appliances and potable water systems AIRFIX A AIRFIX D AIRFIX D-E
AIRFIX EXPANSION VESSELS Flamco Airfix Pressure expansion vessels for sanitary appliances and potable water systems AIRFIX A AIRFIX AIRFIX -E Flamco Edition 2005 / EXP Flamco Flamco offers you a wide choice
More informationFPD20 Series. Gear wheel flow sensor
. FPD20 Series Gear wheel flow sensor Series FPD20-2 - M-5455/0315 Series FPD20 Table of contents page 0 About this operating manual... 4 1 Device description... 5 1.1 Intended use... 5 1.2 Exclusion of
More informationE 158 E 198 E 248. Tank top mounting Connection up to G11 / 4 Nominal flow rate up to 250 l/min e d
R e t u r n - S u c t i o n F i l t e rs E 158 E 198 E 248 Tank top mounting Connection up to G11 / 4 Nominal flow rate up to 250 l/min 20.90-5e 0.0-2d D e s c r i p t i o n Application For operation in
More informationDifferential Pressure Regulator Type Type 45-6 (0.1 to 1 bar, DN 15) Mounting and Operating Instructions EB 3226 EN
Differential Pressure Regulator Type 45-6 Type 45-6 (0.1 to 1 bar, DN 15) Mounting and Operating Instructions EB 3226 EN Edition March 2008 Contents Contents Page 1 Design and principle of operation...................
More informationSafety Please read and keep in a safe place Operating instructions Linear flow control VFC Linear flow control with actuator IFC
07 Elster GmbH Edition 06.7 Translation from the German 058 D GB F NL I E DK S N P GR TR CZ PL RUS H www.docuthek.com Operating instructions Linear flow control VFC Linear flow control with actuator IFC
More informationIndumax CLS54D. Operating Instructions
Operating Instructions Indumax CLS54D Inductive conductivity sensor with hygienic design for applications in the food, beverage and pharmaceutical industries and in biotechnology BA00508C/07/EN/01.12 71189125
More informationMounting and Operating Instructions EB 2172 EN. Series 43 Temperature Regulators. Type 43-5 and Type Type Type 43-6
Series 43 Temperature Regulators Type 43-5 and Type 43-7 Type 43-6 Type 43-6 Type 43-7 with flanged valve body, DN 32 to 50 Type 43-5 Type 43-7 with welding ends Mounting and Operating Instructions EB
More informationACCUMULATOR OPERATING & MAINTENANCE INSTRUCTIONS
ACCUMULATOR OPERATING & MAINTENANCE INSTRUCTIONS READ ALL INSTRUCTIONS PRIOR TO INSTALLATION AND OPERATION TO AVOID POSSIBLE INJURY Warning: Always consider any accumulator to contain pressure until proven
More informationMAINTENANCE AND OPERATING MANUAL
MAINTENANCE AND OPERATING MANUAL Cyclone condensate separator Type ASA EN EN Maintenance and operating manual Purpose and appropriate use Dear Customer, thank you for choosing our product. In order to
More informationCompact differential pressure switch Flameproof enclosure Ex d Models DE, DEC
Mechatronic pressure measurement Compact differential pressure switch Flameproof enclosure Ex d Models DE, DEC WIKA data sheet PV 35.41 Process Compact Series Applications Differential pressure monitoring
More informationInstallation and operating manual. Pneumatic control station LK product no: PCS 1-10
LK product no: PCS 1-10 Article no: 74503 Revision:8 Article no: 74503 Revision: 8 2 (23) Contents 1. General information... 5 2. Safety precautions... 5 2.1 Significance of symbols... 5 2.2 Explanatory
More informationGM-7000 Series CE Approved Gas Control Valve
Installation Instructions GM-7000 Issue Date November 9, 2012 GM-7000 Series CE Approved Gas Control Valve Installation IMPORTANT: These instructions are intended as a guide for qualified personnel installing
More informationMV-01. Features. 2/2-way Solenoid Valve for Fluids. Description: Application:
MV-01 2/2-way Solenoid Valve for Fluids Description: Features / Nominal diameters 1/4-2 / Pressure up to 10 bar / 24 V DC and all common AC variants / Forced-lifting The pilot-controlled full-way valve
More informationSafety. Operating instructions Solenoid valve for gas VG 10/15 VG 65 DANGER. Contents WARNING CAUTION. Changes to edition 11.14
2 Elster GmbH Edition. Translation from the German 22 D F NL I E DK S N P GR TR CZ PL RUS H www.docuthek.com Operating instructions Solenoid valve for gas G / G Contents Solenoid valve for gas G / G....
More informationerc-o 2 -tech Ò innovative gas systems Markwerbener Str Weißenfels Germany
Compart Umwelttechnik GmbH erc-o 2 -tech Ò innovative gas systems Markwerbener Str. 24 06667 Weißenfels Germany Phone: 0049 (0) 3443/2901-0 Fax: 0049 (0) 3443/2901-12 Mail: compart-weissenfels@t-online.de
More informationSafety. Operating instructions Magnetic relief valve VAN DANGER. Contents WARNING CAUTION. Changes to edition 11.11
0084 Edition 0. D F L I E DK S P GR TR CZ PL RUS H www.docuthek.com Operating instructions Magnetic relief valve Translation from the German 0 Elster GmbH Contents Magnetic relief valve... Contents...
More informationPremium Series. Economical Price & Reliable Performance. High Quality. Rust-Resistant. Long Life. Tamper-Resistant locking Screw.
Premium Series High Quality Forged Brass Body Rust-Resistant Nickel Plated Finish Long Life Corrosion Resistant Valve Seat Economical Price & Reliable Performance Tamper-Resistant locking Screw Reduces
More informationOverview of types. Technical data. Safety notes
Technical data sheet R2..-S.. Open-close ball valves, 2-way, with internal thread For open and closed cold and warm water systems For shut-off functions on the water side and 2-point controls in U and
More information4/2 way Pneumatic Solenoid Valve
4/2 way Pneumatic Solenoid Valve Compact design Push-over solenoid coil Exhaust air can be regulated Tube, threaded and sub-base connections Type combined with Seat valve version Type 2508 Type 2510/11
More informationFlow VA 520. incl. temperature measurement. Intelligent solutions for ac- for compressed air and gases R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25
VA 520 incl. temperature R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25 R 1 1/4 (DN 32) DN 40 R 2 (DN 50) Intelligent solutions for ac- for compressed air and gases work according to the approved calorimetric
More informationFlow VA 520. incl. temperature measurement. Intelligent solutions for ac- for compressed air and gases R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25
VA 520 incl. temperature R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25 R 1 1/4 (DN 32) DN 40 R 2 (DN 50) Intelligent solutions for ac- for compressed air and gases work according to the approved calorimetric
More informationAutomatic Balancing Valve Pressure difference controller with integrated flow limiter AB-PM DN
Data sheet Automatic Balancing Valve Pressure difference controller with integrated flow limiter AB-PM DN 40-100 Description AB-PM is ideal for staged installation and commissioning AB-PM is a combined
More information