1 J Neurosurg Pediatrics 12: , 2013 AANS, 2013 Programming jammed Codman Hakim programmable valves: study of an explanted valve and successful programming in a patient Report of 2 cases Sui-To Wong, M.B.B.S., Eleanor Wen, M.B.Ch.B., and Dawson Fong, M.B.B.S. Department of Neurosurgery, Tuen Mun Hospital, Tuen Mun, Hong Kong Malfunction of a Codman Hakim programmable valve due to jamming of its programmable component may necessitate shunt revision. The authors report a method for programming jammed Codman Hakim programmable valves by using a Strata II magnet and additional neodymium magnets. The programming method was derived after studying a jammed valve in the laboratory that was explanted from an 10-year-old boy with a history of fourth ventricle ependymoma. Programming the explanted valve with a Codman programmer failed, but rotating a Strata II magnet above the valve resulted in rotation of the spiral cam in the valve. It was found that the Strata II magnet could be used to program the jammed valve by rotating the magnet 90 or multiples of 90 above the valve. The strength of the magnetic field of the Strata II magnet was able to be increased by putting neodymium magnets on it. The programming method was then successfully used in a patient with a jammed Codman Hakim programmable valve. After successful programming using this method, clinical and radiological follow-up of the patient was advised. ( Key Words Codman Hakim programmable valve malfunction programming failure salvage method hydrocephalus Malfunction of a Codman Hakim programmable valve due to failure of its programmable component usually requires shunt revision; the reported rate of valve explantation for this reason is as high as 9%. 1,3 8,10,11,14 16 The causes of failure of the programmable component include loosening or even dislodgement of the spiral cam (ratchet wheel, stator) and the radiopaque marker, valve damage by direct impact, and extensive deposits within the valve. 7,8,12,13 The dislodgement of the spiral cam has been addressed with design changes in the valve implemented in However, the programmable mechanism of a Codman Hakim programmable valve always involves rotation of the spiral cam by externally applied magnetic force; 2 deposits within a valve always pose the risk of hindering the rotation of the spiral cam, which may cause programming failure, or specifically, a jammed valve. Trou bleshooting methods for a jammed Codman Hakim programmable valve include reverse programming (placing the transmitter upside down against the flow direction) and exposure to the magnetic field of an MRI magnet. 8 In this article, we first report the study of a Codman Hakim programmable valve that was explanted from a patient because of programming failure; we then describe the application of the information obtained in this study to a patient with a jammed Codman Hakim programmable valve. Case 1: Study of a Jammed Codman Hakim Programmable Valve History and Examination. A 10-year-old boy experienced a jammed Codman Hakim programmable valve 2 years after its insertion. This patient had had a WHO Grade II fourth-ventricle ependymoma that was gross-totally excised in June 2010; he also underwent insertion of a ventriculoperitoneal shunt with a Codman Hakim programmable valve because of hydrocephalus. Three months after the shunt insertion, adjuvant radiation therapy was delivered to the tumor bed with a 1-cm margin and a total dose of 46.7 Gy. The boy recovered well and had a Karnofsky Performance Scale score of 80. In the 2 years after the operations, the boy had 6 sets of MR images of the brain and spine obtained using either 1.5-T or 3-T MRI machines; the shunt was reprogrammed and its pressure setting was confirmed with fluoroscopy following each MRI examination. 160 J Neurosurg: Pediatrics / Volume 12 / August 2013

2 Programming jammed Codman Hakim programmable valves The shunt pressure setting was increased stepwise from 60 mm H 2 O to 140 mm H 2 O. In May 2012, after an MRI examination, the pressure setting of the valve decreased to 30 mm H 2 O, and could not be programmed back to the setting before the MRI examination (140 mm H 2 O). Reverse programming with the Codman programmer and exposing the patient to a T2- weighted MRI examination did not change the pressure setting of the valve, 8 and the valve was explanted. Parameters of the CSF samples obtained at the initial shunt insertion and at explantation were both unremarkable (Table 1); swabs of the valve and CSF within the valve at explantation yielded no microbial growth. Investigation of the Explanted Jammed Valve. The following assessment methods were used in the study: 1) visual inspection with an operative microscope was performed to detect structural changes in the valve; 2) fluoroscopy was used to read the pressure setting of the valve, 2 and the angles of a particular pressure setting were measured on the fluoroscopic images with reference to a predefined axis (Fig. 1A); and 3) the lactated Ringer solution infusion technique was used to determine the working pressure of the explanted valve, which was defined as the infusion pressure resulting in a fluid flow compatible with the definition of working pressure. 2 The effects of the following methods on the pressure setting of the explanted valve were assessed: 1) programming with a Codman programmer; 2) rotating a Strata II magnet (Medtronic) above the valve; and 3) exposing the valve to a 3-T MRI magnet. For a Strata II magnet, there are both a magnetic north pole and south pole on its undersurface, and the strength of its magnetic field can be increased by adding magnets (N48 neodymium magnets of inches; Apex Magnet) to its upper surface (Fig. 2). The setup of the Strata II magnet for programming the explanted valve is illustrated in Fig. 1. The explanted valve was placed below the Strata locator tool with the cam of the valve at the center; the Strata II magnet was then placed in the locator tool and turned either clockwise or counterclockwise. Programming with the Codman programmer was conducted by both the usual method and the reverse method. 8 The pressure settings of the explanted valve after each manipulation were recorded using fluoroscopy, and the reference angle measured. When programming using the Strata II magnet, a coupling index was calculated, which was defined as follows: Coupling Index = Change in the cam s angle after rotating the Strata II magnet by X / X A coupling index of 1.0 indicates that the spiral cam rotates together with the Strata magnet completely. A coupling index of 0.0 indicates that the spiral cam does not rotate at all despite rotating the Strata magnet. A smaller coupling index indicates a greater rotational force is required or a greater adhesive force must be overcome when other parameters remain constant. Investigation of a Normal Codman Hakim Programmable Valve. To characterize the effect of a Strata II magnet on a normal Codman Hakim programmable valve, a sample valve was studied using the setup of the experiment as in J Neurosurg: Pediatrics / Volume 12 / August 2013 TABLE 1: Cerebrospinal fluid parameters in Case 1* CSF Parameters Shunt Insertion Shunt Removal protein (g/l) glucose (mmol/l) WBCs (cells/mm 3 ) 4 1 RBCs (cells/mm 3 ) 12 2 smear, culture negative negative * RBCs = red blood cells; WBCs = white blood cells. the study of the explanted valve (Fig. 1). The sample valve was set at several specified pressure levels (30, 60, 90, 120, 150, 180, and 200 mm H 2 O) with a Codman programmer. A Strata II magnet was then rotated above it 90, 180, 270, or 360 in either a clockwise or counterclockwise direction. The coupling index was then calculated. Findings in the Explanted Jammed Valve. Examination under a microscope revealed that the explanted valve was intact, but there were whitish deposits coating the spiral cam, spring, and chassis. The valve pressure setting at explantation was confirmed to be 30 mm H 2 O using fluoroscopy; the corresponding working pressure of the measured valve was 35 mm H 2 O. The pressure settings of the explanted valve after 25 consecutive steps of programming with a Codman programmer or Strata II magnet, or exposure to a 3-T MRI magnet, were recorded. In a Codman Hakim programmable valve, the difference in angle between 2 adjacent pressure settings (a difference of 10 mm H 2 O) was 18 for pressure settings between 30 mm H 2 O and 200 mm H 2 O. When programming with the Codman programmer using the usual method or the reverse method, the pressure setting of the explanted valve did not change to any significant degree; the change in the angle of the spiral cam was at most 16. With the Strata II magnet rotating clockwise, the explanted valve s pressure setting could be increased from 30 mm H 2 O to 180 mm H 2 O; the coupling index was when the valve pressure setting was between 30 mm H 2 O and 110 mm H 2 O, but much smaller ( 0.08) when the pressure setting was greater than 110 mm H 2 O. When the Strata II magnet was rotated counterclockwise, the pressure setting could be decreased from 180 mm H 2 O to 30 mm H 2 O, but the spiral cam could not be rotated further counterclockwise due to the spiral feature of the cam 2 (the coupling index was between 0.10 and 0.21). The explanted valve was brought inside a 3-T MRI suite, placed on the MRI table, and then brought out of the suite; these maneuvers caused only a small change in the pressure setting, from 140 mm H 2 O to 130 mm H 2 O. The working pressure of the explanted valve was measured at 2 additional pressure levels. With pressure settings at 130 mm H 2 O and 180 mm H 2 O, the measured working pressures were 120 mm H 2 O and 165 mm H 2 O, respectively. Findings in the Normal Valve. In the normal valve, when the Strata II magnet was rotated clockwise, the coupling index ranged between 0.17 and 0.71 (mean 0.29), 161

3 S. T. Wong, E. Wen, and D. Fong Fig. 1. Diagrams and photographs showing the setup for programming a Codman Hakim programmable valve with a Strata II magnet. A: Illustration of the cam s angle in relation to the chassis. B and E: Positioning a Strata locator over a Codman Hakim programmable valve. C and F: After putting a Strata II magnet in the locator (B), the starting position of the Strata II magnet is shown. The arrowhead is its magnetic north pole; the Strata II magnet is turned either clockwise or counterclockwise (curved arrows) when programming the valve. D F: Sketch diagrams showing the use of a Strata II magnet on a patient. except for 5 readings when the starting positions of the cam were 60 mm H 2 O or 90 mm H 2 O. When the Strata II magnet was rotated counterclockwise, the coupling index ranged between 0.12 and 0.54 (mean 0.24), except when the starting position of the cam was 30 mm H 2 O, which was expected because of the shape of the spiral cam. 2 After programming the sample valve with the Strata II magnet with and without additional magnets on its upper surface (Fig. 2), the valve could still be successfully programmed with the Codman programmer to all pressure settings between 30 mm H 2 O and 200 mm H 2 O. with rotation of 90, 180, and 360 did not change the valve s pressure setting. With additional neodymium magnets (Fig. 2), rotating 360 clockwise led to an increase of 10 mm H 2 O in the valve s pressure setting. To reach the targeted pressure setting of 180 mm H 2 O, a second 360 rotation was performed and the target pressure setting was reached (Fig. 3); the coupling index was At the time of the 4-month follow-up, the patient was doing well clinically, and serial imaging studies showed increased size of the lateral ventricles compared with the slitlike baseline. Case 2: Programming a Jammed Codman Hakim Programmable Valve History and Examination. This 18-month-old boy with a jammed Codman Hakim programmable valve had undergone shunt insertion for hydrocephalus due to poststreptococcal meningitis at 3 months of age. Initially the shunt pressure was set at 100 mm H 2 O. His head circumference percentile progressively decreased and serial MRI showed small ventricle size; therefore the valve pressure setting was increased stepwise from 100 mm H 2 O to 160 mm H 2 O in an 8-month period. He had 4 sets of MR images obtained during that period, using either 1.5-T or 3-T MRI machines. At 9 months after shunting, after an MRI examination, the pressure setting was jammed at 160 mm H 2 O, despite attempts at increasing it to 180 mm H 2 O. The pressure setting stayed at 160 mm H 2 O despite the reverse programming method and 2 more sets of follow-up MRI examinations. Programming the Jammed Valve. The shunt was not revised because it was structurally intact on radiography, and was functioning clinically. The method we used for Case 2 was based on the study of Case 1 s explanted valve. However, using the Strata II magnet alone, programming Fig. 2. Photographs of the undersurface (A and C) and upper surface (B and D) of a Strata II magnet. The magnetic poles on the undersurface of the Strata II magnet (A) were demonstrated using iron fillings. The Strata II magnet is also pictured with additional magnets to increase the strength of its magnetic field (C and D). The additional magnets used in this study were N48 neodymium magnets inches in size. 162 J Neurosurg: Pediatrics / Volume 12 / August 2013

4 Programming jammed Codman Hakim programmable valves Fig. 3. Fluoroscopic pictures of the valve used in Case 2 before programming with the Strata II magnet (left), and after programming with the Strata II magnet (right) with additional neodymium magnets (Fig. 2). Discussion Jammed Codman Hakim programmable valves may necessitate shunt revision, which can potentially be avoided in some cases if a method for programming jammed valves is available. In the explanted valve we found whitish deposits, as has been reported in the literature. 6,8 However, we could not identify what had caused their formation; the CSF parameters at the time of shunt insertion were normal and there was no shunt infection. In patients with jammed valves, it is likely that deposits inside the valve chamber lead to increased adhesive force between the movable components of the valve, which results in an increase in the force required to rotate the cam. The small coupling index in the explanted valve and in Case 2 s valve supports this hypothesis. Apart from the strength of rotational force required, the coupling index is also affected by other factors, such as the strength of the magnetic fields of the spiral cam and of the Strata magnet, the distance between the valve and the Strata magnet, the thickness of the scalp, and the alignment of the center of the spiral cam with the center of the Strata magnet. Another factor for consideration when managing a jammed Codman Hakim programmable valve is the function of the valve. There is usually a concern that a jammed valve s pressure setting might not reflect its true working pressure. In our explanted valve, we found that there was still satisfactory correlation between the pressure setting and the measured working pressure of the jammed valve, but that was only a single example. Therefore, when a jammed valve is not revised with or without programming using the method illustrated in this report, the patient needs more vigilant clinical and radiological follow-up. Concerning the programming method described here, the Strata II magnet was chosen because it is widely available and the strength of its magnetic field is standardized. We aligned the magnetic north-pole-to-south-pole direction of the Strata magnet with the CSF flow direction of the valve as the starting position of the rotation of the Strata magnet for ease of demonstrating the method (Fig. 1). Actually, the starting position of the magnetic poles of the Strata II magnet is not important, because the cam consists of a circle of micromagnets with their south poles on the upper surface of the cam (Fig. 4). However, the direction of rotation of the Strata magnet (clockwise or counterclockwise) is more important; clockwise is the direction for increasing the pressure setting from 30 mm H 2 O to 200 mm H 2 O, and counterclockwise is the direction for decreasing the pressure setting. And due to specific features of the spiral cam, rotating clockwise beyond 200 mm H 2 O, the pressure falls to 30 mm H 2 O, but when rotating counterclockwise beyond 30 mm H 2 O, movement will be blocked and will not increase to 200 mm H 2 O. 2 Despite the potential use of this programming method, we believe the manufacturer should investigate the use of a standardized but higher power magnet for salvage of jammed Codman Hakim programmable valves. Presently, we use a protocol (Fig. 5) for management of jammed Codman Hakim programmable valves. In successful programming attempts, the patient is monitored clinically and with imaging studies for features of overdrainage and underdrainage; if the shunt system is suitable, measuring the intracranial pressure should be considered. 9 Whether failed Fig. 4. Lateral view of the cam on a suture needle showing the north pole of the cam (A), undersurface of the cam showing the small magnets (the north poles; B), and the upper surface of the cam (the south poles; C). J Neurosurg: Pediatrics / Volume 12 / August

5 S. T. Wong, E. Wen, and D. Fong Fig. 5. Flow chart showing the protocol for management of jammed Codman Hakim programmable shunt (CHPS) valves at our hospital. Asterisk indicates certain patients might be observed without immediate shunt revision. attempts should be followed immediately by shunt revision or a period of observation will depend on the individual clinical situation. In summary, we report a method for programming jammed Codman Hakim programmable valves by using a Strata II magnet and, when necessary, additional neodymium magnets. Disclosure The authors report no conflict of interest concerning the materials or methods used in this study or the findings specified in this paper. Author contributions to the study and manuscript preparation include the following. Conception and design: Wong. Acquisition of data: Wong, Wen. Analysis and interpretation of data: Wong. Drafting the article: Wong. Critically revising the article: all authors. Reviewed submitted version of manuscript: all authors. Approved the final version of the manuscript on behalf of all authors: Wong. Study supervision: Fong. Acknowledgment The authors would like to thank Professor Dachling Pang (Oakland, CA) for performing the tumor excision surgery in the patient in Case 1. References 1. Black PM, Hakim R, Bailey NO: The use of the Codman-Medos Programmable Hakim valve in the management of patients with hydrocephalus: illustrative cases. Neurosurgery 34: , Codman: Procedure Guide. Codman Hakim Programmable Valve System for Hydrocephalus. Raynham, MA: Cod man & Shurtleff Inc., 2006 ( com/binary/org/codman/pdf/val chpv-procguide.pdf) [Accessed April 24, 2013] 3. Katano H, Karasawa K, Sugiyama N, Yamashita N, Ohkura A, Kamiya K: Clinical evaluation of shunt implantations using Sophy programmable pressure valves: comparison with Codman- Hakim programmable valves. J Clin Neurosci 10: , Kay AD, Fisher AJ, O Kane C, Richards HK, Pickard JD: A clinical audit of the Hakim programmable valve in patients with complex hydrocephalus. Br J Neurosurg 14: , J Neurosurg: Pediatrics / Volume 12 / August 2013

6 Programming jammed Codman Hakim programmable valves 5. Kestle JRW, Walker ML: A multicenter prospective cohort study of the Strata valve for the management of hydrocephalus in pediatric patients. J Neurosurg 102 (2 Suppl): , Mangano FT, Menendez JA, Habrock T, Narayan P, Leonard JR, Park TS, et al: Early programmable valve malfunctions in pediatric hydrocephalus. J Neurosurg (6 Suppl) 103: , Mauer UM, Kunz U: More malfunctioning Medos Hakim programmable valves: cause for concern? Clinical article. J Neurosurg 115: , Mauer UM, Schuler J, Kunz U: The Hakim programmable valve: reasons for reprogramming failures. J Neurosurg 107: , Pang D, Grabb PA: Accurate placement of coronal ventricular catheter using stereotactic coordinate-guided free-hand passage. Technical note. J Neurosurg 80: , Pollack IF, Albright AL, Adelson PD: A randomized, controlled study of a programmable shunt valve versus a conventional valve for patients with hydrocephalus. Neurosurgery 45: , Rohde V, Mayfrank L, Ramakers VT, Gilsbach JM: Four-year experience with the routine use of the programmable Hakim valve in the management of children with hydrocephalus. Acta Neurochir (Wien) 140: , Sato K, Shimizu S, Utsuki S, Suzuki S, Oka H, Fujii K: Disparity between adjusted and actual opening cerebrospinal fluid pressure in a patient with the Codman Hakim programmable valve: occult form of shunt failure due to head banging. Case report. J Neurosurg (5 Suppl) 105: , Sgouros S, Dipple SJ: An investigation of structural degradation of cerebrospinal fluid shunt valves performed using scanning electron microscopy and energy-dispersive x-ray microanalysis. J Neurosurg 100: , Takahashi Y: Withdrawal of shunt systems clinical use of the programmable shunt system and its effect on hydrocephalus in children. Childs Nerv Syst 17: , Zemack G, Bellner J, Siesjö P, Strömblad LG, Romner B: Clinical experience with the use of a shunt with an adjustable valve in children with hydrocephalus. J Neurosurg 98: , Zemack G, Romner B: Seven years of clinical experience with the programmable Codman Hakim valve: a retrospective study of 583 patients. J Neurosurg 92: , 2000 Manuscript submitted September 10, Accepted April 22, Please include this information when citing this paper: published online May 24, 2013; DOI: / PEDS Address correspondence to: Sui-To Wong, M.B.B.S., Department of Neurosurgery, Tuen Mun Hospital, Tuen Mun, Hong Kong. wongsuito@hotmail.com. J Neurosurg: Pediatrics / Volume 12 / August

POLARIS. The first MRI-stable adjustable valve NEW MR SAFE

POLARIS. The first MRI-stable adjustable valve NEW MR SAFE POLARIS The first MRI-stable adjustable valve TESTED MR SAFE NEW Polaris Valve The adjustable Polaris valve is a major breakthrough for the safety of patients thanks to the patented self-locking system

More information

Urgent Medical Device Recall Medtronic StrataMR Adjustable Valves & Shunts

Urgent Medical Device Recall Medtronic StrataMR Adjustable Valves & Shunts Urgent Medical Device Recall Medtronic StrataMR Adjustable Valves & Shunts 125 Cremona Drive Goleta, CA 93117 Catalog Number 42955, STRATAMR VALVE, SMALL E11656, E12556, E12753, E13482, E13483, E13484,

More information

MAGNETIC FIELD INFLUENCES AND YOUR STRATA VALVE. Strata II Valves

MAGNETIC FIELD INFLUENCES AND YOUR STRATA VALVE. Strata II Valves MAGNETIC FIELD INFLUENCES AND YOUR STRATA VALVE Strata II Valves Magnetic field influences on your Strata valve You have a Medtronic adjustable valve to treat your hydrocephalus or other CSF disorder.

More information

progav 2.0 In touch with you

progav 2.0 In touch with you progav 2.0 In touch with you We understand the gravity of the situation. Gravitational valves by Miethke Aesculap Neurosurgery progav 2.0 In touch with you MIETHKE gravitational valves provide a reliable

More information

WE UNDERSTAND. NEUROSURGERY. progav 2.0 IN TOUCH WITH YOU

WE UNDERSTAND. NEUROSURGERY. progav 2.0 IN TOUCH WITH YOU WE UNDERSTAND. NEUROSURGERY progav 2.0 IN TOUCH WITH YOU 2 ADJUSTABLE VALVES HAVE EXISTED FOR DECADES. THE ADJUSTMENT AND TESTING OF THE VALVE SETTING IS STILL A CHALLENGE FOR THOSE INVOLVED (1). Valve

More information

Product Design Specifications

Product Design Specifications Product Design Specifications Hydrocephalus Shunt Valve February 2, 2017 Team: Emma Alley (Leader), Andrew Miller (Communicator/BPAG), Karl Fetsch (BWIG), Catharine Flynn (BSAC) Advisor: Beth Meyerand

More information

March 17, 2017 Chiari II decompression in patients with myelomeningocele in the National Spina Bifida Patient Registry (NSBPR) Irene Kim, MD

March 17, 2017 Chiari II decompression in patients with myelomeningocele in the National Spina Bifida Patient Registry (NSBPR) Irene Kim, MD March 17, 2017 Chiari II decompression in patients with myelomeningocele in the National Spina Bifida Patient Registry (NSBPR) Irene Kim, MD Fellow, Pediatric Neurosurgery Children s of Alabama / University

More information

Challenges in Measuring CSF Flow with MRI

Challenges in Measuring CSF Flow with MRI Challenges in Measuring CSF Flow with MRI Rafeeque A. Bhadelia, MD HARVARD MEDICAL SCHOOL Background MR imaging of CSF flow is difficult to use in clinical practice This is because: Qualitative visual

More information

ASAHI Neurovascular Guide Wire

ASAHI Neurovascular Guide Wire ASAHI Neurovascular Guide Wire AMK-DT244 Ver.1.00 / 11TS073 Legal manufacturer Consult instructions for use Do not use if package is damaged RX ONLY Caution: Federal (USA) Law restricts this device to

More information

Model 130M Pneumatic Controller

Model 130M Pneumatic Controller Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate

More information

prosa Adjustable MIETHKE Shunt Systems Your Simple Choice for Precise Patient Care Aesculap Neurosurgery MIETHKE

prosa Adjustable MIETHKE Shunt Systems Your Simple Choice for Precise Patient Care Aesculap Neurosurgery MIETHKE prosa Adjustable MIETHKE Shunt Systems Your Simple Choice for Precise Patient Care MIETHKE Aesculap Neurosurgery MIETHKE prosa Enhanced Patient Care Active people spend about 2/3 of the day walking, standing

More information

Strata Valve ADJUSTMENT REFERENCE GUIDE

Strata Valve ADJUSTMENT REFERENCE GUIDE Strata Valve ADJUSTMENT REFERENCE GUIDE Strata Valve Adjustment Tools Preoperative Procedures READING THE PERFORMANCE LEVEL 1 2 1. Position the Locator Tool above the valve while it is still in the sterile

More information

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13.1.1 TRAINING GOAL 13.1.1.1 Among the responsabilities of a medical

More information

Spasticity / Dystonia

Spasticity / Dystonia Spasticity / Dystonia How can a pump help my movement disorder? This therapy, called intrathecal infusion, is a treatment for controlling some movement disorders. It uses a surgically implanted pump and

More information

AQUARIUS Continuous Renal Replacement Therapy Competency

AQUARIUS Continuous Renal Replacement Therapy Competency Preparation Explain the procedure to the patient and significant others Prepare the patient. Prepare the environment. Ensure a safe environment for the patient and staff Gather equipment required to Setup

More information

*

* Email: * cheekong@krdl.org.sg 336 X. Ma et al. procedures such as stent placement. Commercially available devices are based on the original mechanical design of the Gruentzig s device. Physicians have

More information

Hemoglobin Reference Failure Troubleshooting

Hemoglobin Reference Failure Troubleshooting Hemoglobin Reference Failure Troubleshooting The HEMOGLOBIN REFERENCE FAILURE error ( HGB REF FAILURE ) is generated by the Vet ABC as a quality control check for hemoglobin measurement. Recall that the

More information

EP A1 (19) (11) EP A1 (12) EUROPEAN PATENT APPLICATION. (43) Date of publication: Bulletin 2010/47

EP A1 (19) (11) EP A1 (12) EUROPEAN PATENT APPLICATION. (43) Date of publication: Bulletin 2010/47 (19) (12) EUROPEAN PATENT APPLICATION (11) EP 2 23 32 A1 (43) Date of publication: 24.11. Bulletin /47 (21) Application number: 0916089.6 (1) Int Cl.: A61M 27/00 (06.01) F16K 1/02 (06.01) F16K 1/06 (06.01)

More information

Intracranial Pressure Monitor. Department of Biomedical Engineering University of Wisconsin-Madison BME 301. Final Report

Intracranial Pressure Monitor. Department of Biomedical Engineering University of Wisconsin-Madison BME 301. Final Report 1 Intracranial Pressure Monitor Department of Biomedical Engineering University of Wisconsin-Madison BME 301 Final Report Erin Main: Communicator Jess Hause: Team Leader Peter Strohm: BWIG Lacey Halfen:

More information

MODEL 200 KNIFE GATE VALVES INSTALLATION & MAINTENANCE MANUAL

MODEL 200 KNIFE GATE VALVES INSTALLATION & MAINTENANCE MANUAL MODEL 200 KNIFE GATE VALVES INSTALLATION & MAINTENANCE MANUAL Index 1. List of components / General arrangement 2. Description 3. Handling 4. Installation 5. Actuators / Operation 6. Maintenance a. Changing

More information

Narkomed MRI-2. A perfect fit for the MRI environment. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

Narkomed MRI-2. A perfect fit for the MRI environment. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care Narkomed MRI-2 A perfect fit for the MRI environment Emergency Care Perioperative Care Critical Care Perinatal Care Home Care Because you care The complete anesthesia solution for the MRI environment A

More information

THIS PAGE INTENTIONALLY BLANK

THIS PAGE INTENTIONALLY BLANK THIS PAGE INTENTIONALLY BLANK Page 1 STEREOTACTIC MINIFRAME AND ACCESSORIES INSTRUCTIONS FOR USE CAUTION Federal (U.S.A.) law restricts this device to sale by or on the order of a physician. Carefully

More information

EASYPUMP II THE FLEXIBLE SOLUTION FOR SHORT AND LONG TERM INFUSION THERAPIES

EASYPUMP II THE FLEXIBLE SOLUTION FOR SHORT AND LONG TERM INFUSION THERAPIES THE FLEXIBLE SOLUTION FOR SHORT AND LONG TERM INFUSION THERAPIES CONTENT 4 PATIENT INFORMATION 5 Patient data 6 Treatment monitoring 7 Further notes 8 PRODUCT INFORMATION 9 Easypump II 9 Working principle

More information

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay Summary This assay measures Human Granulocyte Colony Stimulating Factor (G-CSF) in a 96-well

More information

Review of fundamental photon dosimetry quantities

Review of fundamental photon dosimetry quantities Review of fundamental photon dosimetry quantities Narayan Sahoo Main sources of the materials included in this lecture notes are: (1) Radiation Oncology Physics: A Handbook for Teachers and Students Edited

More information

Rules for. Polyathlon. Version Released

Rules for. Polyathlon. Version Released Object of the game: What is a polyathlon? Rules for Polyathlon Version 151021 Released Polyathlon is a series of contests or events similar to the athletic decathlon where robots compete to determine the

More information

Analysis of Shear Lag in Steel Angle Connectors

Analysis of Shear Lag in Steel Angle Connectors University of New Hampshire University of New Hampshire Scholars' Repository Honors Theses and Capstones Student Scholarship Spring 2013 Analysis of Shear Lag in Steel Angle Connectors Benjamin Sawyer

More information

Procedia Engineering Procedia Engineering 2 (2010)

Procedia Engineering Procedia Engineering 2 (2010) Available online at www.sciencedirect.com Procedia Engineering Procedia Engineering 2 (2010) 002681 2686 (2009) 000 000 Procedia Engineering www.elsevier.com/locate/procedia 8 th Conference of the International

More information

VariSource High Dose Rate Afterloader Procedures. For Performing Breast Brachytherapy with. The SAVI TM Applicator

VariSource High Dose Rate Afterloader Procedures. For Performing Breast Brachytherapy with. The SAVI TM Applicator VariSource High Dose Rate Afterloader Procedures For Performing Breast Brachytherapy with The SAVI TM Applicator Contents I. CT Evaluation (~1 week prior to insertion) II. SAVI Applicator Implantation

More information

AQUARIUS Continuous Renal Replacement Therapy with Regional Citrate Anticoagulation (RCA) Competency

AQUARIUS Continuous Renal Replacement Therapy with Regional Citrate Anticoagulation (RCA) Competency Demonstrates an understanding of why RCA has been prescribed Explain the anticoagulation protocol Demonstrate an understanding of potential complications associated with CRRT. Explain potential indications

More information

WP2 Fire test for toxicity of fire effluents

WP2 Fire test for toxicity of fire effluents Pagina 3 di 89 TRA SFEU VTT 22.6.2009 v.2 WP2 Fire test for toxicity of fire effluents Task 2.1.2 Development of small-scale test method for fire effluents Step 1: Use of modeling Plans according to DoW:

More information

EXERCISE 4: MICROCOLLECTION AND HEMATOCRITS. Hematocrit Skills: 5 points Capillary Puncture Skills: 26 points

EXERCISE 4: MICROCOLLECTION AND HEMATOCRITS. Hematocrit Skills: 5 points Capillary Puncture Skills: 26 points EXERCISE 4: MICROCOLLECTION AND HEMATOCRITS Hematocrit Skills: 5 points Capillary Puncture Skills: 26 points Objectives: 1. List three types of patients who may be candidates for capillary puncture and

More information

Navigation Using Compass

Navigation Using Compass Navigation Using Compass Updated January 2015 Albi Sole 0 Outdoor Council of Canada / Conseil canadien de plein air Published by The Outdoor Council of Canada / Conseil canadien de plein air KNA-101, 2500

More information

Operation of the mask aligner MJB-55

Operation of the mask aligner MJB-55 John Paul Adrian Glaubitz Operation of the mask aligner MJB-55 Department of Physics Faculty of Mathematics and Natural Sciences University of Oslo 1 Introduction The mask aligner is an essential tool

More information

Nexus Detachable Coils

Nexus Detachable Coils Micro Therapeutics, Inc. Irvine, California 92618 USA Phone: +1 949-837-3700 Phone: +1 949-837-2044 www.1mti.com Authorized Representative in EU: MediMark Europe BP 2332 38033 Grenoble Cedex 2 France Nexus

More information

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Cover Page for Lab Report Group Portion. Head Losses in Pipes Cover Page for Lab Report Group Portion Head Losses in Pipes Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 February 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section

More information

Lifecycle Performance of Escape Systems

Lifecycle Performance of Escape Systems Lifecycle Performance of Escape Systems A look at laboratory vs field conditioning of aramid fiber based escape systems. By James Hunter, Cedric Smith, Ole Kils and Tyler Mayer for ITRS 2018 1.1 Introduction

More information

INTERVENTIONAL CARDIOLOGY

INTERVENTIONAL CARDIOLOGY Y-Connector Newtech Y connector it is easy, fast and accurate single handed intra vascular device insertion. Perfect insertion device respecting the integrity of the coated surface with Combination of

More information

User s Guide. High Pressure NMR Cell. With High Pressure Tether. Bruker & Varian Cells TABLE OF CONTENTS:

User s Guide. High Pressure NMR Cell. With High Pressure Tether. Bruker & Varian Cells TABLE OF CONTENTS: TABLE OF CONTENTS: User s Guide High Pressure NMR Cell With High Pressure Tether Bruker & Varian Cells Version 1.06 Copyright 2011 Daedalus Innovations LLC Aston, Pennsylvania United States GUIDELINES:...

More information

PRODUCT CATALOGUE OF

PRODUCT CATALOGUE OF PRODUCT CATALOGUE OF INTERVENTIONAL CARDIOLOGY AND GENERAL MEDICAL DISPOSABLES Table of Contents INFLATION DEVICE... 4 GUIDE WIRE (PTFE)... 5 MANIFOLD... 6 PTCA Y Connector... 7 INTRODUCER SHEATH (Femoral/

More information

STANDARD OPERATING PROCEDURE ADVANCED INTRAVENOUS TRAINING ARM (S )

STANDARD OPERATING PROCEDURE ADVANCED INTRAVENOUS TRAINING ARM (S ) Page 1 of 27 1. Scope This Standard Operating Procedure (SOP) applies to the staff and students using the Advanced Intravenous Training Arm in the Pharmacy Practice Resource Unit (PPRU) at the Pharmacy

More information

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system G. Kunz, F. Hasenbalg, P. Pemler 1 1 Klinik für Radio-Onkologie und Nuklearmedizin, Stadtspital Triemli

More information

THE DANGERS OF USING A THREE-WAY VALVE AS A STOPCOCK IN A TRANSDUCER SYSTEM FOR MEASURING VENOUS BLOOD PRESSURE

THE DANGERS OF USING A THREE-WAY VALVE AS A STOPCOCK IN A TRANSDUCER SYSTEM FOR MEASURING VENOUS BLOOD PRESSURE THE DANGERS OF USING A THREE-WAY VALVE AS A STOPCOCK IN A TRANSDUCER SYSTEM FOR MEASURING VENOUS BLOOD PRESSURE J. W. SIVIITH~ PH.D., P.ENr ~ THE SYSTEM shown in Figure 1 has been widely used for the measuring

More information

CORESTA RECOMMENDED METHOD N 6

CORESTA RECOMMENDED METHOD N 6 CORESTA RECOMMENDED METHOD N 6 DETERMINATION OF VENTILATION DEFINITIONS AND MEASUREMENT PRINCIPLES (2015 Revision September 2016) 1. SCOPE This CORESTA Recommended Method specifies a method for the determination

More information

2007 ASME/IEEE Joint Rail Conference & Internal Combustion March Engine 13-16, 2007, Spring Pueblo, Technical Colorado, Conference USA

2007 ASME/IEEE Joint Rail Conference & Internal Combustion March Engine 13-16, 2007, Spring Pueblo, Technical Colorado, Conference USA Proceedings of JRCICE2007 2007 ASME/IEEE Joint Rail Conference & Internal Combustion Engine Spring Proceedings Technical of JRCICE2007 Conference 2007 ASME/IEEE Joint Rail Conference & Internal Combustion

More information

Wind Flow Validation Summary

Wind Flow Validation Summary IBHS Research Center Validation of Wind Capabilities The Insurance Institute for Business & Home Safety (IBHS) Research Center full-scale test facility provides opportunities to simulate natural wind conditions

More information

Ch.6. Dr.Rajaa Pressure

Ch.6. Dr.Rajaa Pressure Ch.6. Dr.Rajaa Pressure أ.م.د. رجاء سهيل جنم Pressure: It is the force per unit area in a gas or liquid. (For solid the term pressure is replaced by "stress"). Units: In metric system pressure is measured

More information

FRA Spacer Instruments. For anterior lumbar interbody fusion (ALIF) using straight anterior or anterolateral approach.

FRA Spacer Instruments. For anterior lumbar interbody fusion (ALIF) using straight anterior or anterolateral approach. FRA Spacer Instruments. For anterior lumbar interbody fusion (ALIF) using straight anterior or anterolateral approach. Technique Guide Instruments and implants approved by the AO Foundation Table of Contents

More information

Making Life Easier Easypump II

Making Life Easier Easypump II Making Life Easier Easypump II The elastomeric pump system for short- and long-term infusion therapy Elastomeric Infusion Systems Technical Guide Easypump II Disposable Elastomeric Infusion Pump System

More information

Read each slide, some slides have information to record on your organizer. Some slides have numbers that go with the question or red and underlined

Read each slide, some slides have information to record on your organizer. Some slides have numbers that go with the question or red and underlined Read each slide, some slides have information to record on your organizer. Some slides have numbers that go with the question or red and underlined to use for answering the questions. Essential Question:

More information

Homeostasis and Negative Feedback Concepts and Breathing Experiments 1

Homeostasis and Negative Feedback Concepts and Breathing Experiments 1 Homeostasis and Negative Feedback Concepts and Breathing Experiments 1 I. Homeostasis and Negative Feedback Homeostasis refers to the maintenance of relatively constant internal conditions. For example,

More information

Advanced Neuromonitoring Solutions ICP. Oxygen Partial Pressure, Intracranial Pressure and. Temperature. Drainage. Measurement O 2.

Advanced Neuromonitoring Solutions ICP. Oxygen Partial Pressure, Intracranial Pressure and. Temperature. Drainage. Measurement O 2. Advanced Neuromonitoring Solutions ICP Oxygen Partial Pressure, Intracranial Pressure and Temperature Measurement Drainage Temperature Table of contents Page Microchip catheters with maximum precision

More information

SEP-1 Additional Notes for Abstraction for Version 5.0b

SEP-1 Additional Notes for Abstraction for Version 5.0b SEP-1 Additional Notes for Abstraction for Version 5.0b Data Element Administrative Contraindication to Care Blood Culture Collection Date Blood Culture Collection Time Broad Spectrum or Other Antibiotic

More information

IMPROVING QUALITY THROUGH INNOVATION: A FULLY AUTOMATED SOLUBLE SALT METER

IMPROVING QUALITY THROUGH INNOVATION: A FULLY AUTOMATED SOLUBLE SALT METER ARP INSTRUMENTS, INC. SSM MODEL # RPCT-07-001 IMPROVING QUALITY THROUGH INNOVATION: A FULLY AUTOMATED SOLUBLE SALT METER For more information contact: ARP Instruments, Inc. www.arpinstruments.biz arp.instruments@gmail.com

More information

[Type here] MANAGEMENT OF X-RAY GENERATING EQUIPMENT

[Type here] MANAGEMENT OF X-RAY GENERATING EQUIPMENT MANAGEMENT OF X-RAY GENERATING EQUIPMENT HEALTH AND SAFETY UNIT AUTUMN 2016 1 CONTENTS INTRODUCTION Page 3 1. ACQUISITION OF X-RAY GENERATING EQUIPMENT Page 3 2. X-RAY EQUIPMENT FOR USE WITH HUMAN SUBJECTS

More information

Advanced Neuromonitoring Solutions ICP. Oxygen Partial Pressure, Intracranial Pressure and Temperature Measurement. Drainage

Advanced Neuromonitoring Solutions ICP. Oxygen Partial Pressure, Intracranial Pressure and Temperature Measurement. Drainage Advanced Neuromonitoring Solutions Oxygen Partial Pressure, Intracranial Pressure and Temperature Measurement Drainage O2 800-961-2122 Temperature Table of contents Page Microchip catheters with maximum

More information

Needle valve. Contents. User s Manual. (1) Be sure to read the following warranty clauses of our product 1. (2) General operating instructions 2

Needle valve. Contents. User s Manual. (1) Be sure to read the following warranty clauses of our product 1. (2) General operating instructions 2 Serial No. H-V024-E-7 Needle valve User s Manual Contents (1) Be sure to read the following warranty clauses of our product 1 (2) General operating instructions 2 (3) General instructions for transportation,

More information

Exposure System Selection

Exposure System Selection Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others

More information

This article appeared in a journal published by Elsevier. The attached copy is furnished to the author for internal non-commercial research and

This article appeared in a journal published by Elsevier. The attached copy is furnished to the author for internal non-commercial research and This article appeared in a journal published by Elsevier. The attached copy is furnished to the author for internal non-commercial research and education use, including for instruction at the authors institution

More information

CS CANNULATION. Step-by-Step Guide. This guide contains recommendations as provided by Dr. Seth Worley, MD POWERED BY

CS CANNULATION. Step-by-Step Guide. This guide contains recommendations as provided by Dr. Seth Worley, MD POWERED BY This guide contains recommendations as provided by Dr. Seth Worley, MD POWERED BY 1 Advance dilator and CSG over 0.035 wire until dilator tip reaches the SVC. 2 Holding dilator stationary, advance CSG

More information

Calibration Procedure

Calibration Procedure Management Procedure 2565 Revision: A Date Issued: October 30, 2017 Date Revised: Calibration Procedure DeFelsko Corporation PosiTector RTR-3D Replica Tape Reader Probe Table of Contents 1 Introduction

More information

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone: MJB4 Mask Aligner Operating Procedure Effective Date: 07/12/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Karl Suss MJB4 Mask Aligner, Optical Lithography,

More information

RESPIRATORY HEALTH PRODUCTS

RESPIRATORY HEALTH PRODUCTS RESPIRATORY HEALTH PRODUCTS CONTENTS ADULT ENDOTRACHEAL TUBES HALYARD* MICROCUFF* ENDOTRACHEAL TUBES Oral/Nasal Magill page 4 HALYARD* MICROCUFF* SUBGLOTTIC SECRETION DRAINAGE ENDOTRACHEAL TUBE Subglottic

More information

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator Journal of BUON 8: 55-59, 2003 2003 Zerbinis Medical Publications. Printed in Greece ORIGINAL ARTICLE Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

More information

CHEMICAL ENGINEERING LABORATORY CHEG 239W. Control of a Steam-Heated Mixing Tank with a Pneumatic Process Controller

CHEMICAL ENGINEERING LABORATORY CHEG 239W. Control of a Steam-Heated Mixing Tank with a Pneumatic Process Controller CHEMICAL ENGINEERING LABORATORY CHEG 239W Control of a Steam-Heated Mixing Tank with a Pneumatic Process Controller Objective The experiment involves tuning a commercial process controller for temperature

More information

Policy for the commissioning of arthroscopic shoulder decompression surgery for the management of Pure Subacromial Shoulder Impingement

Policy for the commissioning of arthroscopic shoulder decompression surgery for the management of Pure Subacromial Shoulder Impingement Policy for the commissioning of arthroscopic shoulder decompression surgery for the management of Pure Subacromial Shoulder Impingement Version of: December 2018 Version Number: V 0.1 Changes Made: Policy

More information

የIትዮዽያ የደረጃዎች ኤጀንሲ ETHIOPIAN STANDARDS AGENCY ES DRAFT UPLOAD REQUEST FORM FOR PUBLIC COMMENTS

የIትዮዽያ የደረጃዎች ኤጀንሲ ETHIOPIAN STANDARDS AGENCY ES DRAFT UPLOAD REQUEST FORM FOR PUBLIC COMMENTS OF/ESA/SDD/04 EHS This form is used to maintain uniformity and consistency in uploading the list of draft documents for the development of Ethiopian standards for public comment. 1. The draft documents

More information

Worksheets Based on.

Worksheets Based on. Worksheets Based on https://itunes.apple.com/book/id14280245 Copyright April 2019 Rhonda Ryman-Kane and Robyn Hughes Ryman 1 Acknowledgements The authors wish to gratefully acknowledge the proofreading

More information

NEANN LONG SPINE BOARD

NEANN LONG SPINE BOARD NEANN LONG SPINE BOARD INFORMATION BOOKLET 2 The following booklet is an information guide for the NEANN LSB (Long Spine Board). This Board is the development of user feedback over a 10 year period following

More information

Physics and Quality Assurance

Physics and Quality Assurance Disclosure Physics and Quality Assurance Consultancy agreements with Nucletron, Hologic Mahta M. McKee, MS, DABR Department of Radiation Oncology Wake Forest University Health Sciences Agenda Current Treatment

More information

Item 404 Driving Piling

Item 404 Driving Piling Item Driving Piling 1. DESCRIPTION Drive piling. 2. EQUIPMENT 2.1. Driving Equipment. Use power hammers for driving piling with specified bearing resistance. Use power hammers that comply with Table 1.

More information

SECTION BUTTERFLY VALVES

SECTION BUTTERFLY VALVES SECTION 15112 BUTTERFLY VALVES PART 1 GENERAL 1.01 SUMMARY A. All butterfly valves shall be of the tight closing, rubber seated type and fully comply with the latest revision of AWWA Standard C504, Class

More information

Analysis of fatigue failure in D-shaped karabiners

Analysis of fatigue failure in D-shaped karabiners Analysis of fatigue failure in D-shaped karabiners K.B. Blair, D.R. Custer, J.M. Graham, and M.H. Okal Center for Sports Innovation, Massachusetts Institute of Technology, USA Abstract In order to determine

More information

FieldStrength. Using MotionTrak for motion correction in body imaging. Application tips

FieldStrength. Using MotionTrak for motion correction in body imaging. Application tips FieldStrength Publication for the Philips MRI Community Issue 35 September / October 2008 Using MotionTrak for motion correction in body imaging Application tips This article is part of Field Strength

More information

RESPIRATORY MOTION IN RADIATION THERAPY

RESPIRATORY MOTION IN RADIATION THERAPY RESPIRATORY MOTION IN RADIATION THERAPY Significant effect on targets in the chest and upper abdominal cavities Can cause difficulty in reproducing radiation planning protocols in radiation treatment sessions

More information

Cell Injury Controller II (CIC II)

Cell Injury Controller II (CIC II) Cell Injury Controller II (CIC II) CIC II Operation Manual Contents Introduction 2 Quick-Start 2 Theory of Operation 3 Injury Measurement Method 3 Components 4 External Requirements 4 Front Panel 5 Rear

More information

Aesculap ProSpace Titanium / PEEK

Aesculap ProSpace Titanium / PEEK Aesculap ProSpace Titanium / PEEK Posterior Lumbar Interbody Fusion System Aesculap Spine ProSpace Implant Design Content A Philosophy 3 B Implant Material 4 C Implant Features 6 D Surgical Technique D1

More information

Evaluation of estimated CEM43 values attained for realistic use cases under (local) SAR constraints imposed by IEC

Evaluation of estimated CEM43 values attained for realistic use cases under (local) SAR constraints imposed by IEC Evaluation of estimated CEM43 values attained for realistic use cases under (local) SAR constraints imposed by IEC60601-2-33 Johan van den Brink Principal Scientist MRI 28 May 2015 1 Outline Overview of

More information

G622. APPLIED SCIENCE Monitoring the Activity of the Human Body ADVANCED SUBSIDIARY GCE. Thursday 27 May 2010 Afternoon. Duration: 1 hour 30 minutes

G622. APPLIED SCIENCE Monitoring the Activity of the Human Body ADVANCED SUBSIDIARY GCE. Thursday 27 May 2010 Afternoon. Duration: 1 hour 30 minutes ADVANCED SUBSIDIARY GCE APPLIED SCIENCE Monitoring the Activity of the Human Body G622 *OCE/17533* Candidates answer on the Question Paper OCR Supplied Materials: None Other Materials Required: Electronic

More information

Recommendations for Standards. Monitoring. Cardiopulmonary Bypass

Recommendations for Standards. Monitoring. Cardiopulmonary Bypass Recommendations for Standards of Monitoring during Cardiopulmonary Bypass Membership of the working party (2006) Ian Curle Fiona Gibson Jonathan Hyde Alex Shipolini David Smith J P van Besouw Kate Wark

More information

Case Iatrogenic venous air embolism

Case Iatrogenic venous air embolism Case 10857 Iatrogenic venous air embolism La Pietra P, Sommario M, Marchini N, Venturi P Ospedale Bentivoglio, Asl Bologna, Dipartimento servizi; Via Marconi 40121 Bentivoglio, Italy; Email:plapietra@tin.it

More information

extricare NPWT Foam Kit Instruction for Use

extricare NPWT Foam Kit Instruction for Use extricare NPWT Foam Kit Instruction for Use www.devonmedicalinc.com 1.866.446.0092 Manufactured For: Devon Medical, Inc. 1100 1st Avenue, Suite 202 King of Prussia PA 19406, USA www.devonmedicalinc.com

More information

APPENDIX B TESTING PROTOCOLS. Method A: Straight Test Method B: Angular Deflection Test. Method C: Shear Load Test

APPENDIX B TESTING PROTOCOLS. Method A: Straight Test Method B: Angular Deflection Test. Method C: Shear Load Test APPENDIX B TESTING PROTOCOLS Method A: Straight Test Method B: Angular Deflection Test Method C: Shear Load Test B.1 METHOD A. STRAIGHT PIPE JOINT TEST In order to simulate the straight pipe-joint condition

More information

Using Experimental Procedure to Improve the Efficiency of the Two Stand Reversing Cold Mill

Using Experimental Procedure to Improve the Efficiency of the Two Stand Reversing Cold Mill Using Experimental Procedure to Improve the Efficiency of the Two Stand Reversing Cold Mill Mohammad Heydari Vini 1* 1 Instructor, Department of Engineering, Islamic Azad University, Mobarakeh Branch,

More information

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics. FRANZCR Examination Part I Radiation Oncology Radiotherapeutic Physics Candidate No.: The Royal Australian and New Zealand College of Radiologists FRANZCR Examination Part I Radiation Oncology Radiotherapeutic

More information

(Lab Interface BLM) Acceleration

(Lab Interface BLM) Acceleration Purpose In this activity, you will study the concepts of acceleration and velocity. To carry out this investigation, you will use a motion sensor and a cart on a track (or a ball on a track, if a cart

More information

Happy New Year! from. Jorgensen Laboratories, Inc JorVet s Annual Suture Sale. January 2019 Volume 23 Issue 1

Happy New Year! from. Jorgensen Laboratories, Inc JorVet s Annual Suture Sale. January 2019 Volume 23 Issue 1 Jorgensen Laboratories January 2019 Volume 23 Issue 1 Page 1 January Specials - Annual Suture Sale 2018 JorVet Catalog Page 2 February Specials - Cal-Gel / Keto-Gel - Pet Piller - Dental Machines Happy

More information

Evita 2 dura. First-class ventilation in your patient s best interests. Because you care. Emergency Care OR/Anesthesia Critical Care Home Care

Evita 2 dura. First-class ventilation in your patient s best interests. Because you care. Emergency Care OR/Anesthesia Critical Care Home Care M E D I C A L Evita 2 dura First-class ventilation in your patient s best interests Emergency Care OR/Anesthesia Critical Care Home Care Because you care A workhorse in everyday clinical routine For more

More information

Distributed Control Systems

Distributed Control Systems Unit 41: Unit code Distributed Control Systems M/615/1509 Unit level 5 Credit value 15 Introduction With increased complexity and greater emphasis on cost control and environmental issues, the efficient

More information

MRI Safety Training for First Responders

MRI Safety Training for First Responders MRI Safety Training for First Responders By Sandip Basak, MD University Radiology Group 2018 Objectives Explain what an MRI is and how it works Describe the hazards of MRI Explain the safety zones of an

More information

Employer s Unit of Competence Magnetic particle testing of materials, products and plant

Employer s Unit of Competence Magnetic particle testing of materials, products and plant Employer s Unit of Competence Magnetic particle testing of materials, products and plant Image - if cover page required Supported by lead employer Overview This unit identifies the competencies required

More information

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization Basics of Proton Therapy Proton Treatment Planning and Beam Optimization SAM Educational Session, WE D BRB 2 Mark Pankuch, PhD Northwestern Medicine Chicago Proton Center Today s objectives Review the

More information

RESPIRATORY HEALTH PRODUCTS

RESPIRATORY HEALTH PRODUCTS RESPIRATORY HEALTH PRODUCTS CONTENTS ADULT CLOSED SUCTION SYSTEMS HALYARD* CLOSED SUCTION SYSTEMS Double Swivel Elbows page 54 WET PAK* Double Swivel Elbows page 54 T-Pieces page 65 WET PAK* T-Pieces page

More information

User s manual for the SC32 and SC96 form-fitting recording chamber and microdrive systems

User s manual for the SC32 and SC96 form-fitting recording chamber and microdrive systems User s manual for the SC32 and SC96 form-fitting recording chamber and microdrive systems 1. Description of the chamber system ----------------------------------------- 2-3 2. Description of the microdrive

More information

Electrical Safety in Hospitals

Electrical Safety in Hospitals Electrical Safety in Hospitals Safety in Clinical Environment Electrical hazards Electrical shocks (micro and macro) due to equipment failure, failure of power delivery systems, ground failures, burns,

More information

ADVANCED SUBSIDIARY GCE G622 APPLIED SCIENCE

ADVANCED SUBSIDIARY GCE G622 APPLIED SCIENCE ADVANCED SUBSIDIARY GCE APPLIED SCIENCE Unit 3: Monitoring the activity of the human body TUESDAY 5 JUNE 2007 G622 Afternoon *CUP/T21601* Additional materials: Electronic calculator Ruler (cm/mm) Time:

More information

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface Chapter 11 reatment Planning Single Beams Radiation Dosimetry I Outline Basic terminology Curved contour surface correction (bolus, compensators, wedges) Oblique beam incidence Correction for tissue inhomogeneities

More information

Injector system for implantation of 1stQ foldable intraocular lenses

Injector system for implantation of 1stQ foldable intraocular lenses Instruction For Use 1st Inject Instrument Injector system for implantation of 1stQ foldable intraocular lenses The IFU is available electronically on our website: www.1stq.eu Content: The package contains

More information

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS 1. INTRODUCTION An anaesthesia delivery system includes any machine, equipment or apparatus which supplies gases, vapours, local anaesthesia and/or

More information

Intracranial Pressure Sensor Testing Device

Intracranial Pressure Sensor Testing Device Intracranial Pressure Sensor Testing Device Detection of Shunt Malfunction in Hydrocephalus Patients Final Design Report Team Members: Brad Lindevig- BSAC Dan Miller- Team Leader Jamon Opgenorth- Communicator

More information