13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

Size: px
Start display at page:

Download "13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS."

Transcription

1 13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS TRAINING GOAL Among the responsabilities of a medical physicist, the implementation of as quality assurance program (a, b, c) is one of the most important. Many checks are required to assure the safety of the patients and of the personnel in radiation therapy. The purpose of this task is to give knowledge on how to perform the main checks on a teletherapy unit. This submodule deals with visual checks made to verify that the collimators are in good order Minimum time to be spent on the task: 1 day Competencies Addressed Ability to check the position of the collimator isocenter, the jaws symmetry, the field size and collimator angles indicated by the readout system Equipment, Materials Needed for the Task Mechanical front pointer (optional); sheets of paper; ruler Depending on local circumstances hands-on training may not be possible Important Note The use of equipment must be supervised by a qualified medical physicist Procedure Collimator Isocenter Place a sheet of paper horizontally on the table at the standard SSD. Project the image of a rectangular field on the paper and draw its diagonals, then marks their intersection. Rotate the collimator to 90, 180, 270 and repeat the procedure. Measure the maximum distance between the centers of the fields.

2 Jaws With the same setup used in procedure set the gantry to 0 and mark the corners of a rectangular light field. Rotate the collimators through 180 and check the correspondence of the new light field with the previous one Light Field vs Indicators With the same set-up used in mark the corners of a rectangular field and the image of the cross hairs on the paper. Measure the sizes of the light field and compare them with the readout values. Draw the diagonals of the light field and compare their intersection with the image of the cross hairs Collimator Angle Indicators With the same setup used in procedure rotate the collimator through different angles and check the correspondence between the effective angles and the readout values Measurements Integral part of the procedure Calculations No calculation is required Observations, Interpretations and Conclusions Observe that in procedures and the check is made on the light field. In order to check the position of the mechanical collimator isocenter a mechanical front pointer is needed (see optional) Compare the results to tolerance levels defined by the locally adopted quality assurance protocol Optional (Relative to procedure ) For a check of the mechanical collimator isocenter a mechanical front pointer held to the jaws is necessary. If this device is available, the procedure can be performed in the same way, by marking on the paper the point indicated by the pointer tip.

3 (Relative to procedure ) If a mechanical front pointer is available, check that the position of its tip is congruent with the center of the light field References W.R.Hendee, G.S.Ibbot, "Radiation Therapy Physics", 2nd ed., Mosby, 1996 J.R. Williams, D.I. Twaites, "Radiotherapy Physics", Oxford Medical Publications, 1993 UNICEN Routine Checks and Dosimetry of Therapeutic X-ray Units, AAPM code of practice for radiotherapy accelerators: Report of AAPM Radiation Therapy Task Group No Verification Signature and date by the trainer: Name of the trainee: Comments: Date: Trainer's sign:

4 13.2 GANTRY ISOCENTER, GANTRY ANGLE INDICATOR, COUCH ISOCENTER,COUCH ANGLE, COUCH POSITION INDICATORS Training Goal This submodule deals with visual checks made to verify that the gantry and the couch are in good order Minimum time to be spent on the task: 1 day Competencies Addressed Ability to check the position of the gantry and the couch, to verify the readout relative to gantry and couch Equipment, Materials needed for the Task Mechanical front pointer; rod with fine point; ballast (weight of approximately 50 Kg); ruler Depending on local circumstances hands-on training may not be possible Important Note The use of equipment must be supervised by a qualified medical physicist Procedure Gantry Isocenter Mount the front pointer and fix the rod on the table so that its tip juts out and touches the tip of the front pointer. Then rotate the gantry from 0 to 360 with a 45 step. For each position measure the distance between the tips. Find the maximum value Gantry Angle Indicator Check that the gantry angle corresponds to the indicated value for at least four positions (0, 90, 180,270 ).

5 Couch Isocenter The general set-up is the same used for Measure the distance between the tips for different couch positions Couch Angle Check that the couch angle corresponds to the indicated value for at least four positions (0, 90, 180,270 ) Couch Position Indicators Put the ballast over the couch on the top. Move the couch along the three axes and check the correspondence between the real position and the redout indicators Measurements Integral part of the procedure Calculations No calculation is required Observations, Interpretations and Conclusions Observe that the position of the gantry and couch mechanical isocenter should be compared to the position of the radiation isocenter Compare the results to tolerance levels defined by the locally adopted quality assurance protocol Optional Nothing optional is suggested.

6 References W.R.Hendee, G.S.Ibbot, "Radiation Therapy Physics", 2nd ed., Mosby, 1996 J.R. Williams, D.I. Twaites, "Radiotherapy Physics", Oxford Medical Publications, 1993 UNICEN Routine Checks and Dosimetry of Therapeutic X-ray Units, AAPM code of practice for radiotherapy accelerators: Report of AAPM Radiation Therapy Task Group No Verification Signature and date by the trainer: Name of the trainee: Comments: Date: Trainer's sign:

7 13.3 LOCALIZING LASER, LIGHT VS RADIATION FIELD CONGRUENCE, COLLIMATOR ROTATION, GANTRY ROTATION, COUCH ROTATION Training Goal The main purpose of this submodule is to give knowledge about how to find the radiation field and the radiation isocenter. This step in quality assurance is fundamental to check finally the coincidence between radiation, light and mechanical isocenter and field Minimum time to be spent on the task: 1 day Competencies Addressed Learning how to check the laser localization system. Understanding how to find the radiation isocenter for the collimators, the gantry and the couch Equipment, Materials Needed for the Task Ready pack films ; water equivalent plastic sheets; photodensitometer; pin Depending on local circumstances hands-on training may not be possible Important Note The use of equipment must be supervised by a qualified medical physicist Procedure Localizing Laser Place a paper sheet horizontally on the table and set the gantry to 0. Open the jaws to the maximum field size and project the light field on the paper. Then compare the field cross hairs to laser cross hairs and measure the maximum shifting between them. The check should be done also in vertical planes Light vs Radiation Field Congruence The check should be done for at least three fields: 5 cm x 5 cm, 10 cm x 10 cm and the maximum field size.

8 Place the film over at least 5 cm of plastic sheets and perpendicular to the beam axis, at the standard SSD. Then, make holes in the film in order to mark the corners and the center of the light field. Mark the film outside the field in a known position, in order to recognize its orientation with respect to the collimators after the development. Cover the film by other plastic sheets to ensure electronic buildup. Make an exposure in order to have a linear response by the film. Develop the film, then scan it with the photodensitometer along the two collimators axes to find the limits of the radiation field, defined where the net optical density is 50% of the central axis value. Compare the radiation field size to the light field size (corresponding to the holes) Collimator Rotation Arrange the film and the plastic sheets as in , without piercing it. Close one couple of jaws and open the other to obtain a strip-like field. Expose the film for different collimator angles. Develop the film, then draw the field midline for the different angles with a sharp pen. Sign the intersections of the fields on the film and measure the maximum shift between them. Repeat the procedure for a new film by inverting the collimators Gantry Rotation Place the film between two plastic sheets in the plane defined by the beam axis when the gantry rotates. Arrange the collimators to obtain a strip-like field parallel to the rotation axis of the gantry. Expose the film for different gantry positions, then develop it. Draw the field midline for the different angles with a sharp pen. Sign the intersections of the fields on the film and measure the maximum shift between them Couch Rotation Arrange the film on the table at the standard SSD between two plastic sheets as indicated in Close one couple of jaws and open the other to obtain a strip-like field. Expose the film for different couch angles. Develop the film, then draw the field midline for the different angles with a sharp pen. Sign the intersections of the fields on the film and measure the maximum shift between them. Repeat the procedure for a new film by inverting the collimators Measurements Integral part of the procedure.

9 Calculations No calculation is required Observations, Interpretations and Conclusions Compare the results to tolerance levels defined by the locally adopted quality assurance protocol Optional Nothing optional is suggested References W.R.Hendee, G.S.Ibbot, "Radiation Therapy Physics", 2nd ed., Mosby, 1996 J.R. Williams, D.I. Twaites, "Radiotherapy Physics", Oxford Medical Publications, 1993 UNICEN Routine Checks and Dosimetry of Therapeutic X-ray Units, AAPM code of practice for radiotherapy accelerators: Report of AAPM Radiation Therapy Task Group No Verification Signature and date by the trainer: Name of the trainee: Comments: Date: Trainer's sign:

10 13.4 OUTPUT CONSTANCY AND LINEARITY Training goal This submodule deals with the verification that the timing and dosimetry properties of a teletherapy unit are in good order Related submodules: Calibration of a MVXRBeam Competencies Addressed Learning how to check the output constancy, the time constancy, the linearity Equipment, Materials Needed for the Task Water equivalent phantom; ionization chamber (e.g. 0.6 cc FARMER type ) with associated electrometer; termometer, barometer; stopwatch Depending on local circumstances hands-on training may not be possible Important Note The use of equipment must be supervised by a qualified medical physicist Procedure Output Constancy Check the output constancy by measuring the dose in reference condition. Repeat the measurement at least 10 times and calculate the standard deviation of the results. An index of output stability is the percent ratio of the standard deviation to the mean: check that it is smaller than 1% Linearity Measure the dose in reference condition for different MU pre-sets. Plot the resulted doses against the MU set-up and check the linearity of the curve Measurements Integral part of the procedure.

11 Calculations Integral part of procedure Observations, Interpretations and Conclusions Compare the results to tolerance levels defined by the locally adopted quality assurance protocol Optional Optionally, the output constancy should be checked also during the day, by repeating the measurement in different moments References W.R.Hendee, G.S.Ibbot, "Radiation Therapy Physics", 2nd ed., Mosby, 1996 J.R. Williams, D.I. Twaites, "Radiotherapy Physics", Oxford Medical Publications, 1993 UNICEN Routine Checks and Dosimetry of Therapeutic X-ray Units, AAPM code of practice for radiotherapy accelerators: Report of AAPM Radiation Therapy Task Group No Verification Signature and date by the trainer: Name of the trainee: Comments: Date: Trainer's sign:

12 13.5 PDD (OR TPR), BEAM FLATNESS AND SYMMETRY Training Goal The purpose of this task is to check the constance of PDD (or TPR) for a linear accelerator Minimum time to be spent on the task: 1 day Related submodules: Measurement of PDD, Measurement of TPR Competencies Addressed Equipment, Materials Needed for the Task Water scanning system; couple of small ionization chamber (0.1 cc volume) Depending on local circumstances hands-on training may not be possible Important Note The use of equipment must be supervised by a qualified medical physicist Procedure For this measurement see the related submodules (Measurement of PDD, Measurement of TPR) Measurements Integral part of the procedure Calculations Integral part of the procedure.

13 Observations, Interpretations and Conclusions Find D2010 or TPR2010 and check that they are congruent with the expected values Optional Nothing optional is suggested References W.R.Hendee, G.S.Ibbot, "Radiation Therapy Physics", 2nd ed., Mosby, 1996 J.R. Williams, D.I. Twaites, "Radiotherapy Physics", Oxford Medical Publications, 1993 UNICEN Routine Checks and Dosimetry of Therapeutic X-ray Units, AAPM code of practice for radiotherapy accelerators: Report of AAPM Radiation Therapy Task Group No Verification Signature and date by the trainer: Name of the trainee: Comments: Date: Trainer's sign:

14 13.6 BEAM FLATNESS AND SYMMETRY, TRANSMISSION FACTOR OF WEDGE FILTERS Training Goal The purpose of this task is to check the main charateristics of the beams: depth dose, flatness and symmetry (a, b). The beam profile of linacs can change because the flattening filter and the source are misaligned Minimum time to be spent on the task: 1 day Related submodules: Measuring MVXR Isodose Curves, Measurement of the Wedge Transmission Competencies Addressed Ability to check that the beam flatness and symmetry are within the tolerance limits. Ability to check the wedge transmission factors Equipment, Materials Needed for the Task Water scanning system; ionisation chambers; termometer, barometer; calculator. The materials are the same used for the measurement of isodose curves and wedge transmission factors (see the related submodules) Depending on local circumstances hands-on training may not be possible Important Note The use of equipment must be supervised by a qualified medical physicist Procedure Beam Flatness and Symmetry Measure the beam profiles at the buildup depth and at 10 cm depth along the X and Y directions as indicated in the related submodule. Calculate the beam flatness: R R max max R + R min min 100

15 where R max and R min are the maximum and minimum readings in the interval defined by the 80% of the central axis reading (reference region). Calculate the beam symmetry as the maximum percent difference between the doses at the left and right sides of the field. Optionally, repeat the assessment along the diagonals directions Transmission Factor of Wedge Filters Measure the transmission factors of all wedge filters for some fields as indicated in the related submodule. Compare the resulted values to those registered in the machine book Measurements Integral part of the procedure Calculations Integral part of the procedure Observations, Interpretations and Conclusions The manufacturers give tolerance limits both for the flatness and the symmetry: compare these limits to the the measured values Optional Nothing optional is suggested References W.R.Hendee, G.S.Ibbot, "Radiation Therapy Physics", 2nd ed., Mosby, 1996 J.R. Williams, D.I. Twaites, "Radiotherapy Physics", Oxford Medical Publications, 1993 UNICEN Routine Checks and Dosimetry of Therapeutic X-ray Units, AAPM code of practice for radiotherapy accelerators: Report of AAPM Radiation Therapy Task Group No. 45.

16 Verification Signature and date by the trainer: Name of the trainee: Comments: Date: Trainer's sign:

Constancy checks of well-type ionization chambers with external-beam radiation units

Constancy checks of well-type ionization chambers with external-beam radiation units JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Constancy checks of well-type ionization chambers with external-beam radiation units Sara L. Hackett, 1a Benjamin Davis, 2 Andrew

More information

Relative Dosimetry. Photons

Relative Dosimetry. Photons Relative Dosimetry Photons What you need to measure! Required Data (Photon) Central Axis Percent Depth Dose Tissue Maximum Ratio Scatter Maximum Ratio Output Factors S c & S cp! S p Beam profiles Wedge

More information

Dosimetric Calculations. Lonny Trestrail

Dosimetric Calculations. Lonny Trestrail Dosimetric Calculations Lonny Trestrail 20 October 2008 Objectives Dose Distribution Measurements PDD, OCR TAR, SAR, TPR, TMR, SPR, SMR Arc or Rotational Therapy Isodose Curves Point Dose Calculations

More information

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Commissioning an IMRT System for MLC Delivery Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Taking the broad view of commissioning Commissioning elements Validating the dosimetry system Commissioning the

More information

THE development of more advanced techniques in radiotherapy,

THE development of more advanced techniques in radiotherapy, 1 A Geant4-based simulation of an accelerator s head used for Intensity Modulated Radiation Therapy F. Foppiano, B. Mascialino, M.G. Pia, M. Piergentili. Abstract We present a Geant4-based simulation,

More information

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface Chapter 11 reatment Planning Single Beams Radiation Dosimetry I Outline Basic terminology Curved contour surface correction (bolus, compensators, wedges) Oblique beam incidence Correction for tissue inhomogeneities

More information

Review of fundamental photon dosimetry quantities

Review of fundamental photon dosimetry quantities Review of fundamental photon dosimetry quantities Narayan Sahoo Main sources of the materials included in this lecture notes are: (1) Radiation Oncology Physics: A Handbook for Teachers and Students Edited

More information

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator Journal of BUON 8: 55-59, 2003 2003 Zerbinis Medical Publications. Printed in Greece ORIGINAL ARTICLE Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

More information

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 DISCLAIMER: This publication and associated spreadsheets and digital files are based on sources and

More information

Math Review. Overview

Math Review. Overview Drawn by Steve Yan, CMD Math Review Shirley Ann Pinegar-Johnston MS RT(R)(T)CMD Overview Drawn by Steve Yan, CMD Divergence & Intensity DIVERGENCE X-rays travel in Straight but Divergent lines INTENSITY

More information

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN 22 Turkish Journal of Cancer Volume 37, No.1, 2007 Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN SATISH PELAGADE 1, KALPANA

More information

Zoubir Ouhib Lynn Cancer Institute

Zoubir Ouhib Lynn Cancer Institute Zoubir Ouhib Lynn Cancer Institute April 2015 Speaker for ELEKTA Brachytherapy using miniature X-ray sources Energy

More information

Commissioning and periodic tests of the Esteya electronic brachytherapy system

Commissioning and periodic tests of the Esteya electronic brachytherapy system Review paper Educational Activity Commissioning and periodic tests of the Esteya electronic brachytherapy system Cristian Candela-Juan, PhD 1, Yury Niatsetski, MSc 2, Zoubir Ouhib, PhD 3, Facundo Ballester,

More information

TEST REPORT BEAMSCAN WITH HALCYON

TEST REPORT BEAMSCAN WITH HALCYON TEST REPORT BEAMSCAN WITH HALCYON UZ Leuven, November 20 th, 2017 + = Ideal BEAMSCAN tank dimensions to fit perfectly into the Halcyon bore Fast tank positioning using the Halcyon lasers, no electronics

More information

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas Current Implementation Status Current Implementation Status 1600 1400 TOTAL 1494 of 1623 ACTIVE INSTITUTIONS

More information

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08. Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas 2009 AAPM Summer School Educational Objectives Improve your understanding of how to implement TG-51

More information

1. Question Answer cgy / MU cgy / MU 2. Question Answer

1. Question Answer cgy / MU cgy / MU 2. Question Answer GS020113: Introduction to Medical Physics III: Therapy s to home work problem set assigned on 3/22/11 1. Question A patient is set up at 100 cm SSD on a 6 MVX machine. The dose rate at 10 cm in phantom

More information

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics. FRANZCR Examination Part I Radiation Oncology Radiotherapeutic Physics Candidate No.: The Royal Australian and New Zealand College of Radiologists FRANZCR Examination Part I Radiation Oncology Radiotherapeutic

More information

Proton Therapy QA Tools

Proton Therapy QA Tools Proton Therapy QA Tools A complete range of tools for absolute dosimetry, machine QA and patient plan verification Introduction During the past years more and more heavy particle therapy centers, especially

More information

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World Chattanooga, Tennessee, April 17-21, 2005, on CD-ROM, American Nuclear Society, LaGrange Park, IL (2005) GEANT4 SIMULATION OF

More information

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11 MV Photon Dosimetry in the Clinic This document discusses both the scientific methods and the system of work required to calibrate a field instrument expanding upon the practical details of both.. Introduction

More information

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions Rebecca M. Howell, Ph.D., D.A.B.R, The University of Texas at M.D. Anderson Cancer Center Overview 2 Commissioning

More information

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription Monitor Unit Calculations Part 1 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Return to our first patient 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt

More information

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL FOR THE DEGREE MASTER OF ARTS BY CHARLES TRAVIS WEBB

More information

High Speed Direct SAD Radiosurgery Beam Scanner

High Speed Direct SAD Radiosurgery Beam Scanner Open Access Original Article DOI: 10.7759/cureus.20 High Speed Direct SAD Radiosurgery Beam Scanner Walter Nikesch 1, James M. Hevezi 2, Irene Monterroso 3, Daniel Navarro 4, James G. Schwade 5 1. CyberKnife

More information

Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin

Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin NCCAAPM Meeting Oct 11,2013 Larry DeWerd has a partial interest in Standard Imaging Talks from people who have been

More information

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting Monitor Unit Calculations Part 2 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Calculation of machine setting reference dose machine setting =, reference dose

More information

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 47 CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 4.1 INTRODUCTION Advanced treatment techniques use optimized radiation beam intensities to conform dose distribution to the

More information

Clinical Implementation of Volumetric Modulated Arc Therapy

Clinical Implementation of Volumetric Modulated Arc Therapy Clinical Implementation of Volumetric Modulated Arc Therapy UT M.D. Anderson Cancer Center Ramaswamy Sadagopan, Rebecca M. Howell, Weiliang Du and Peter Balter Definition 2 Intensity Modulated Arc therapy

More information

Monitor Unit Verification for Small Fields

Monitor Unit Verification for Small Fields Monitor Unit Verification for Small Fields Patrick Higgins, Ph.D University of Minnesota Department of Radiation Oncology October 10, 2013 The Issues: How do we verify the monitor units calculated by the

More information

A beam-matching concept for medical linear accelerators

A beam-matching concept for medical linear accelerators Acta Oncologica, 2009; 48: 192200 ORIGINAL ARTICLE A beam-matching concept for medical linear accelerators DAVID SJÖSTRÖM, ULF BJELKENGREN, WIVIANN OTTOSSON & CLAUS F. BEHRENS Copenhagen University Hospital,

More information

IMRT in clinical practice at the UMC-Utrecht. Utrecht

IMRT in clinical practice at the UMC-Utrecht. Utrecht 3 IMRT in clinical practice at the UMC-Utrecht Utrecht Clinical use of IMRT Improvement of local control without increased incidence of normal tissue complications + Dose escalation in the prostate Step-and

More information

Abstract: Introduction:

Abstract: Introduction: Are photon and electron beam calibrations more consistent with TG-51 than with TG-21? Ramesh C. Tailor, William F. Hanson, Nathan Wells, and Geoffrey S. Ibbott. U.T. MD Anderson Cancer Center, Houston

More information

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers EXRADIN Ion Chambers Exradin Ion Chambers Exradin (EXacting RADiation INstrumentation) Ion Chambers have been built for over 33 years, are recognized by top research institutes and standards laboratories,

More information

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components Introduction The CyberKnife SRS/SBRT System James M. Hevezi,, Ph.D., FACR CyberKnife Unit San Antonio, Texas CyberKnife Components Treatment Delivery (G3 Platform San Antonio) Robot (modified German auto

More information

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2017, 6, 57-72 http://www.scirp.org/journal/ijmpcero ISSN Online: 2168-5444 ISSN Print: 2168-5436 Validation of Treatment

More information

Comparison of ionization chambers of various volumes for IMRT absolute dose verification

Comparison of ionization chambers of various volumes for IMRT absolute dose verification Comparison of ionization chambers of various volumes for IMRT absolute dose verification Leonid B. Leybovich, a) Anil Sethi, and Nesrin Dogan Department of Radiation Oncology, Loyola University Medical

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 Comparison of monitor unit calculations performed with a 3D computerized planning system and independent hand calculations: Results

More information

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams Peter R. Almond Brown Cancer Center, Louisville, Kentucky 40202 Peter J. Biggs Department of Radiation Oncology,

More information

A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions

A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions Wendel Dean Renner a) Math Resolutions, LLC, 5975 Gales Lane, Columbia,

More information

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61: Practical Implementation of TG-61: II. Guidelines for clinical implementation of TG-61 J.P. Seuntjens Medical Physics Unit McGill University, Montreal General Hospital Montréal, Canada jseuntjens@medphys.mcgill.ca

More information

INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS

INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS A9744593 INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSLS Ladislav Czap, Georg Matscheko and Pedro Andreo osimetry and Medical Radiation Physics Section, ivision

More information

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Poster No.: R-0044 Congress: Type: Authors: Keywords: DOI: 2014 CSM Scientific Exhibit L. Bendall, I. Patel, N. McGrath, C. Rowbottom; MANCHESTER/

More information

Scientific Journal of Nuclear Medicine & Radiation Therapy

Scientific Journal of Nuclear Medicine & Radiation Therapy Scientific Journal of Nuclear Medicine & Radiation Therapy Research Article Radiotherapy Dose Calculation Program, for Monitoring and Commissioning Treatment Planning Systems - Ioannis Vamvakas*, Stella

More information

TG-51: Experience from 150 institutions, common errors, and helpful hints

TG-51: Experience from 150 institutions, common errors, and helpful hints TG-51: Experience from 150 institutions, common errors, and helpful hints Ramesh C. Tailor, Ph. D., William F. Hanson, Ph.D., and Geoffrey S. Ibbott, Ph.D. Department of Radiation Physics, University of

More information

SnapShot IMRT with compensators and FFF beams

SnapShot IMRT with compensators and FFF beams SnapShot IMRT with compensators and FFF beams Vladimir Feygelman, PhD (1) Moffitt Cancer Center, Tampa, FL, USA Disclosure VF has a sponsored research agreement with.decimal. A short history of radiotherapy

More information

Radiotherapy physics & Equipments

Radiotherapy physics & Equipments Radiotherapy physics & Equipments RAD 481 Lecture s Title: Beam Modification devices in Radiotherapy Dr. Mohammed Emam Vision :IMC aspires to be a leader in applied medical sciences, health care education

More information

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Tom Jordan Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE

More information

Implementation of respiratory-gated VMAT on a Versa HD linear accelerator

Implementation of respiratory-gated VMAT on a Versa HD linear accelerator Received: 21 November 2016 Revised: 7 April 2017 Accepted: 5 June 2017 DOI: 10.1002/acm2.12160 RADIATION ONCOLOGY PHYSICS Implementation of respiratory-gated VMAT on a Versa HD linear accelerator Jeffrey

More information

Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation

Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation By Ruth Ann M. Good, CMD. R.T. (R)(T) Virginia Commonwealth University, Richmond,

More information

The MC commissioning of CyberKnife with MLC (Tips and Tricks)

The MC commissioning of CyberKnife with MLC (Tips and Tricks) The MC commissioning of CyberKnife with MLC (Tips and Tricks) Alain Guemnie Tafo, PhD UPMC Pinnacle, Harrisburg, PA Institute Gustave Roussy, Villejuif, France Disclaimer & Disclosure The views expressed

More information

Breast Treatment System REFERENCE MANUAL

Breast Treatment System REFERENCE MANUAL Breast Treatment System REFERENCE MANUAL This manual may not be reproduced in whole or in part, by mimeograph or any other means, without the written permission of Diacor, Inc. The MammoRx Breast Treatment

More information

TG219: IT'S USE, STRENGTHS AND WEAKNESSES

TG219: IT'S USE, STRENGTHS AND WEAKNESSES 1 CONFLICT OF INTEREST TG219: IT'S USE, STRENGTHS AND WEAKNESSES SOTIRI STATHAKIS, PHD, DABR None ROSTER STATEMENT OF THE PROBLEM Timothy C. Zhu, University of Pennsylvania, Philadelphia, PA (Co- Chair)

More information

Viva TPS. TS11/15 Total Stations Check and Adjust Procedure. October Summary

Viva TPS. TS11/15 Total Stations Check and Adjust Procedure. October Summary Viva TPS October 2010 TS11/15 Total Stations Summary Leica builds total stations to the highest quality and calibrates each instrument before it leaves the Factory. After the instrument is shipped or used

More information

IN-PLACE DENSITY AND MOISTURE CONTENT OF SOIL AND SOIL- AGGREGATE BY NUCLEAR METHODS (SHALLOW DEPTH) FOP FOR AASHTO T 310

IN-PLACE DENSITY AND MOISTURE CONTENT OF SOIL AND SOIL- AGGREGATE BY NUCLEAR METHODS (SHALLOW DEPTH) FOP FOR AASHTO T 310 AND MOISTURE CONTENT OF SOIL AND SOIL- AGGREGATE BY NUCLEAR METHODS (SHALLOW DEPTH) FOP FOR AASHTO T 310 Scope This procedure covers the determination of density, moisture content, and relative compaction

More information

Original Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams

Original Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams Iranian Journal of Medical Physics Vol. 10, No. 3, Summer 2013, 147-155 Received: April 22, 2013; Accepted: July 21, 2013 Original Article Effective Point of Measurement in Cylindrical Ion Chamber for

More information

Electronic Oncology Systems the Management and the Safety Measures. Hansen Chen, Director, TDSI

Electronic Oncology Systems the Management and the Safety Measures. Hansen Chen, Director, TDSI Electronic Oncology Systems the Management and the Safety Measures Hansen Chen, Director, TDSI Radiation Oncology Department People Place Equipments Procedures Patients Hansen Chen 2 Radiation Oncology

More information

Gripping rotary modules

Gripping rotary modules Gripping rotary modules Gripping rotary modules GRIPPING ROTARY MODULES Series Size Page Gripping rotary modules RP 314 RP 1212 318 RP 1216 322 RP 1520 326 RP 2120 330 RP 2128 334 RC 338 RC 1212 342 RC

More information

IN-PLACE DENSITY OF BITUMINOUS MIXES USING THE NUCLEAR MOISTURE-DENSITY GAUGE FOP FOR WAQTC TM 8

IN-PLACE DENSITY OF BITUMINOUS MIXES USING THE NUCLEAR MOISTURE-DENSITY GAUGE FOP FOR WAQTC TM 8 OF BITUMINOUS MIXES USING THE NUCLEAR MOISTURE-DENSITY GAUGE FOP FOR WAQTC TM 8 Scope This test method describes a test procedure for determining the density of bituminous mixes by means of a nuclear gauge

More information

2 Available: 1390/08/02 Date of returning: 1390/08/17 1. A suction cup is used to support a plate of weight as shown in below Figure. For the conditio

2 Available: 1390/08/02 Date of returning: 1390/08/17 1. A suction cup is used to support a plate of weight as shown in below Figure. For the conditio 1. A suction cup is used to support a plate of weight as shown in below Figure. For the conditions shown, determine. 2. A tanker truck carries water, and the cross section of the truck s tank is shown

More information

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007 The Effect of Immobilisation Devices on Radiotherapy Dose Distributions A thesis submitted in fulfilment of the requirements for the Degree of Master of Applied Science (Medical and Health Physics) By

More information

Maintenance and general care of secondary standard and field instruments

Maintenance and general care of secondary standard and field instruments Maintenance and general care of secondary standard and field instruments This is without doubt the most important lecture that you will be given during this course. That rather rash statement is based

More information

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system G. Kunz, F. Hasenbalg, P. Pemler 1 1 Klinik für Radio-Onkologie und Nuklearmedizin, Stadtspital Triemli

More information

OUTPUT at dmax in tissue 8x8 20x20 At 100cm SSD At 100cm SAD

OUTPUT at dmax in tissue 8x8 20x20 At 100cm SSD At 100cm SAD RAPHEX 2001 T1. All of the following are true regarding Percentage Depth Dose (PDD) except: A. Increases with increasing energy. B. Depends on field size. C. Is the dose at depth expressed as a % of the

More information

Introduction. Materials and Methods

Introduction. Materials and Methods STANDARD WEDGE AND TRAY TRANSMISSION VALUES FOR VARIAN, SEIMENS, ELEKTA/PHILIPS ACCELERATORS; A QUALITY ASSURANCE TOOL David S. Followill, Nadia Hernandez, and William F. Hanson Introduction The Radiological

More information

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER Proven Rad-Check technology specifically designed to provide you with the ultimate in versatility and cost-effective operation. Fast and easy

More information

WEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS

WEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS Acta Radiologica Oncology 23 (1984) Fusc. 6 FROM THE RADIATION THERAPY DEPARTMENT, ANTON1 VAN LEEUWENHOEK HOSPITAL, THE NETHERLANDS CANCER INSTITUTE, AMSTERDAM, THE PHILIPS MEDICAL SYSTEMS DIVISION, BEST,

More information

IAEA-TECDOC-1455 Implementation of the International Code of Practice on Dosimetry in Radiotherapy (TRS 398): Review of testing results

IAEA-TECDOC-1455 Implementation of the International Code of Practice on Dosimetry in Radiotherapy (TRS 398): Review of testing results IAEA-TECDOC-1455 Implementation of the International Code of Practice on Dosimetry in Radiotherapy (TRS 398): Revie of testing results Final report of the Coordinated Research Projects on Implementation

More information

COMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC

COMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC COMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC by TAN ZI XIANG Thesis submitted in fulfillment of the requirements for the degree of Master of Science (Medical Physics) February 2015 ACKNOWLEDGEMENT

More information

7.65 ±0.05 mm of lead. The maximum energy of the bremsstrahlung photons was calibrated using the threshold for the D(-y,n)H reaction at

7.65 ±0.05 mm of lead. The maximum energy of the bremsstrahlung photons was calibrated using the threshold for the D(-y,n)H reaction at Am 7Roentgenolia6:ia6oia65, 1976 CENTRAL AXIS DEPTH DOSE FOR A 2.5 MV VAN DE GRAAFF GENERATOR DAVID W. ANDERSON, DAVID E. RAESIDE, REBA I. ADAMS, AND MYRON R. GOEDE 2 ABSTRACT: Central axis percentage

More information

Surface buildup dose dependence on photon field delivery technique for IMRT

Surface buildup dose dependence on photon field delivery technique for IMRT Surface buildup dose dependence on photon field delivery technique for IMRT Shigeru Yokoyama, Peter L. Roberson, Dale W. Litzenberg, Jean M. Moran, and Benedick A. Fraass Department of Radiation Oncology,

More information

CHAPTER-2. Wedge filters are the most important beam modifiers used in

CHAPTER-2. Wedge filters are the most important beam modifiers used in CHAPTER-2 2. WEDGE FLTERS 2.1. NTRODUCTON Wedge filters are the most important beam modifiers used in radiotherapy which are placed in the path of photon beams to modify their isodose diistributions. The

More information

Underwater measurement of photosynthetically active radiation

Underwater measurement of photosynthetically active radiation Underwater measurement of photosynthetically active radiation Mark Blonquist, Damon Nitzel, and Bruce Bugbee Apogee Instruments, Inc., Logan, Utah Introduction Quantum sensors (photosynthetically active

More information

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation This intercomparison will be carried out under the lead of KRISS with the cooperation of ARPANSA. This comparison has been approved

More information

AN31E Application Note

AN31E Application Note Balancing Theory Aim of balancing How an unbalance evolves An unbalance exists when the principle mass axis of a rotating body, the so-called axis of inertia, does not coincide with the rotational axis.

More information

COMMITMENT & SOLUTIONS THE EFFECTS OF PROPELLANT POSITION ON SMALL CALIBER BALLISTIC PERFORMANCE UNPARALLELED

COMMITMENT & SOLUTIONS THE EFFECTS OF PROPELLANT POSITION ON SMALL CALIBER BALLISTIC PERFORMANCE UNPARALLELED THE EFFECTS OF PROPELLANT POSITION ON SMALL CALIBER BALLISTIC PERFORMANCE Connie M. Lusto Physicist, U.S. Army ARDEC 04-MAY-2017 UNPARALLELED COMMITMENT & SOLUTIONS U.S. ARMY ARMAMENT RESEARCH, DEVELOPMENT

More information

Lab 7 Rotational Equilibrium - Torques

Lab 7 Rotational Equilibrium - Torques Lab 7 Rotational Equilibrium - Torques Objective: < To test the hypothesis that a body in rotational equilibrium is subject to a net zero torque and to determine the typical tension force that the biceps

More information

Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy

Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy University of South Florida Scholar Commons Graduate Theses and Dissertations Graduate School 7-14-2010 Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy

More information

Commissioning and quality assurance for the treatment delivery components of the AccuBoost system

Commissioning and quality assurance for the treatment delivery components of the AccuBoost system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 2, 2015 Commissioning and quality assurance for the treatment delivery components of the AccuBoost system Ileana Iftimia, 1,2a Mike Talmadge,

More information

VariSource High Dose Rate Afterloader Procedures. For Performing Breast Brachytherapy with. The SAVI TM Applicator

VariSource High Dose Rate Afterloader Procedures. For Performing Breast Brachytherapy with. The SAVI TM Applicator VariSource High Dose Rate Afterloader Procedures For Performing Breast Brachytherapy with The SAVI TM Applicator Contents I. CT Evaluation (~1 week prior to insertion) II. SAVI Applicator Implantation

More information

VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests

VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 3, 2016 VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests Michael P. Barnes, 1,2a Pejman Rowshanfarzad,

More information

3D Treatment Planning and verification with hand calculations

3D Treatment Planning and verification with hand calculations 3D Treatment Planning and verification with hand calculations Randy Holt, PhD, CEO Pacific Crest Medical Physics, Inc Chico, CA Dosimetric Calculations Calibration and Measurement Conditions Patient Conditions

More information

Anemometry. Anemometry. Wind Conventions and Characteristics. Anemometry. Wind Variability. Anemometry. Function of an anemometer:

Anemometry. Anemometry. Wind Conventions and Characteristics. Anemometry. Wind Variability. Anemometry. Function of an anemometer: Anemometry Anemometry Function of an anemometer: Measure some or all of the components of the wind vector In homogeneous terrain, vertical component is small express wind as -D horizontal vector For some

More information

Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens

Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens The characteristic ionizing radiation response is defined by the relationship between the stimulus

More information

New Highly Productive Phased Array Ultrasonic Testing Machine for Aluminium Plates for Aircraft Applications

New Highly Productive Phased Array Ultrasonic Testing Machine for Aluminium Plates for Aircraft Applications 19 th World Conference on Non-Destructive Testing 2016 New Highly Productive Phased Array Ultrasonic Testing Machine for Aluminium Plates for Aircraft Applications Christoph HENKEL 1, Markus SPERL 1, Walter

More information

Scleral Fitting Guide

Scleral Fitting Guide Scleral Fitting Guide Scleral Lens Design...3 Fitting Philosophy...4 Select a Diagnostic Lens...5 Initial Lens Evaluation...6 Assess Scleral Landing Zone (SLZ), Sagittal Depth, Limbal Clearance, Lens Power...7-10

More information

Commissioning of a new total body irradiation protocol

Commissioning of a new total body irradiation protocol University of Wollongong Research Online University of Wollongong Thesis Collection University of Wollongong Thesis Collections 2012 Commissioning of a new total body irradiation protocol Zoe Baldwin University

More information

Calibration Procedure

Calibration Procedure Management Procedure 2565 Revision: A Date Issued: October 30, 2017 Date Revised: Calibration Procedure DeFelsko Corporation PosiTector RTR-3D Replica Tape Reader Probe Table of Contents 1 Introduction

More information

User Manual. RW3 Slab Phantom T29672 and T

User Manual. RW3 Slab Phantom T29672 and T User Manual RW3 Slab Phantom Contents Operating Manual Technical Manual Service Manual D188.131.00/07 en 2015-07 Hn General Information General Information The product bears the CE-mark "CE" in accordance

More information

Measurement of court speed and bounce. Rod Cross, Physics Department, Sydney University 2nd July 2006

Measurement of court speed and bounce. Rod Cross, Physics Department, Sydney University 2nd July 2006 1 Measurement of court speed and bounce Rod Cross, Physics Department, Sydney University 2nd July 2006 The speed of a surface refers to the change in ball speed in a direction parallel to the surface.

More information

Victoreen B. Operators Manual. Image Intensifier Ion Chamber

Victoreen B. Operators Manual. Image Intensifier Ion Chamber Victoreen 6000-530B Image Intensifier Ion Chamber Operators Manual March 2005 Manual No 6000-530B-1 Rev. 4 2004, 2005 Fluke Corporation, All rights reserved. All product names are trademarks of their respective

More information

The dose distribution of medium energy electron boosts to the laryngectomy stoma

The dose distribution of medium energy electron boosts to the laryngectomy stoma JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 1, WINTER 2001 The dose distribution of medium energy electron boosts to the laryngectomy stoma Ellen D. Yorke* Memorial Sloan-Kettering Cancer

More information

Evaluation of six TPS algorithms in computing entrance and exit doses

Evaluation of six TPS algorithms in computing entrance and exit doses JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Evaluation of six TPS algorithms in computing entrance and exit doses Yun I. Tan, 1,2a Mohamed Metwaly, 3 Martin Glegg, 3 Shaun P.

More information

Characteristics and Production of Gas Sealed, Radiation Hard, Small Ionization Chambers

Characteristics and Production of Gas Sealed, Radiation Hard, Small Ionization Chambers Characteristics and Production of Gas Sealed, Radiation Hard, Small Ionization Chambers Mayda Velasco 1 Gianni Tassotto 3 on behalf of the SIC team P. Ball 2, A.Dabrowsky 1, G. Graham 2, C. Kendziora 3,

More information

Dynamic wedges dosimetry and quality control

Dynamic wedges dosimetry and quality control Original Paper Received: 2005.06.07 Accepted: 2006.02.28 Published: 2006.04.28 Authors Contribution: A Study Design B Data Collection C Statistical Analysis D Data Interpretation E Manuscript Preparation

More information

[Type here] MANAGEMENT OF X-RAY GENERATING EQUIPMENT

[Type here] MANAGEMENT OF X-RAY GENERATING EQUIPMENT MANAGEMENT OF X-RAY GENERATING EQUIPMENT HEALTH AND SAFETY UNIT AUTUMN 2016 1 CONTENTS INTRODUCTION Page 3 1. ACQUISITION OF X-RAY GENERATING EQUIPMENT Page 3 2. X-RAY EQUIPMENT FOR USE WITH HUMAN SUBJECTS

More information

Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water

Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water M R McEwen, A J Williams and A R DuSautoy Centre for Ionising Radiation Metrology National Physical

More information

How to Design Medical Accelerator Systems for Reliability: IBA PT System

How to Design Medical Accelerator Systems for Reliability: IBA PT System How to Design Medical Accelerator Systems for Reliability: IBA PT System Yves Jongen Founder & Chief Research Officer Ion Beam Applications s.a. Belgium 1 Outline A short introduction of a proton therapy

More information

OBJECTIVE 6: FIELD RADIOLOGICAL MONITORING - AMBIENT RADIATION MONITORING

OBJECTIVE 6: FIELD RADIOLOGICAL MONITORING - AMBIENT RADIATION MONITORING OBJECTIVE 6: FIELD RADIOLOGICAL MONITORING - AMBIENT RADIATION MONITORING OBJECTIVE Demonstrate the appropriate use of equipment and procedures for determining field radiation measurements. INTENT This

More information

COMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy

COMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy COMPASS 3D IMRT Verification in Patient Anatomy AK IMRT Bamberg April 2010 Dr. Lutz Müller Patient-specific Verification? CT Treatment unit Gantry =0 Treatmt unit Patient couch X-ray beam Patient couch

More information