TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:

Size: px
Start display at page:

Download "TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:"

Transcription

1 Practical Implementation of TG-61: II. Guidelines for clinical implementation of TG-61 J.P. Seuntjens Medical Physics Unit McGill University, Montreal General Hospital Montréal, Canada TG-61 deals ith: Calibration of absolute output of orthovoltage x-ray units in terms of D Relative dosimetry of orthovoltage beams Clinical issues electron contamination dose in biological tissues This part of the refresher course: Aspects and pitfalls of absolute output calibration Peculiarities in relative dosimetry of kilovoltage beams Other clinical issues Absolute calibration: Summary of recommendations tube potential < 100 kv in-airmethodmandatory tube potential 100 kv in-phantommethodalloed choice for medium energies to be made based on the location of the point of interest (target volume) Chambers Chambers: lo energy tube potential < 70 kv: soft x-ray parallel plate chambers tube potential 70 kv: cylindrical chambers ith flat energy response medium energy cylindrical chambers ith flat energy response Phantoms no phantom for in-air method ater phantom for in-phantom reference dosimetry for relative dosimetry plastic phantoms for stability checks

2 I. Absolute calibration: a. In-air method Absolute Calibration: In-air method µ en D (0 cm) = MN K P stem,air B ρ air air M: corrected chamber reading N K : air-kerma calibration factor P stem,air : stem correction factor free-in-air µ en mass-energy absorption coefficient ρ ratio ater to air, free-in-air air air B : back-scatter factor In-air method: 2 conceptual steps In-air method: 2 conceptual steps Step 1: measure air-kerma in clinical beam get and interpolate N K from a standards dosimetry laboratory (ADCL s, NIST, NRCC) establish P stem,air K air (clinical beam) =MN K P stem,air Step 2: look-up conversion factor and backscatter factor µ en D K =K,air B = K air ρ air B air STEP 1: Getting a calibration factor from standards lab & Evaluating P stem,air Get N K from standards lab calibration in terms of air-kerma (for both in-air and in-phantom method): beam quality specification is in terms of both HVL and tube potential

3 NIST calibration qualities disseminated by ADCL s Beam Code First HVL Hom. Coeff. (mm Al) (mm Cu) (Al) L L M M M M M Energy dependence of chamber calibration factor?... calibration factors should not vary significantly beteen to calibration points so that the estimated uncertainty in the calibration factor for a clinical beam beteen the to calibration points is ithin 2%. Energy dependence: in practical terms chamber response should be knon approximately, e.g., from manufacturer <2% calibration for at least to radiation qualities that bracket the radiation quality used in the clinic N K (x-rays)/n K ( 60 Co) should be as expected

4 In-air method: evaluate P stem,air for cylindrical chambers, stem effect in-air is usually less than 1% so P stem,air <1% stem scatter for large body endindo or superficial therapy chambers can be appreciable and should be checked Grimbergen,1995 Measuring P stem,air comparison to chamber ith knon P stem,air M chamber reading, M ref reference chamber reading f c field size at calibration lab; f u field size in clinical beam Grimbergen,1995 Mf ( ) ref ( ) stem,air( ) c M f P f u u = Pstem,air( fc) Mf ( u) Mref ( fc) STEP 2: Data on mass-energy absorption coefficients & Backscatter factors ±1.5%

5 Backscatter factors B is a kerma based quantity: K,phantom B = K,free-air Based on Monte Carlo calculations TG61 adopted data sets from Grossendt (1990, 1993) Surface kerma is very difficult to measure ith chambers electron contamination chamber response issues Clinical issues of in-air method strain on accurate knoledge of energy dependence of chamber response clinical relevance of electron contamination at the surface validity of backscatter factors for small cones Overall uncertainty: In-air method In-air method: Uncertainty at the reference point Type of quantity or procedure Uncertainty N K from standards lab 0.7% Effect of beam-quality difference 2.0% Backscatter factor B 1.5% P stem,air 1.0% µ en 1.5% ρ air Free-air measurement in user s 1.5% beam Combined D,z=0 3.5%

6 Summary: In-phantom method I. Absolute Calibration: b. In-phantom method µ = MN K P sheath P en D (2 cm) Q,cham ρ air M: corrected in-phantom chamber reading N K : air-kerma calibration factor P sheath : sheath correction factor P Q,cham : chamber correction factor µ en mass-energy absorption coefficient ρ ratio ater to air, in-phantom air Background: In-phantom method Background: In-phantom method Step 1: measure air kerma in ater: get and interpolate N K from a standards laboratory (see: in-air method) perform in-phantom measurement M (2 cm) Step 2: lookup conversion factor, chamber correction factor, sheath correction factor K air ()= MN K D µ en K ( ) = K' air ( ) PQ, champsheath ρ air K air () Clinical issues of in-phantom method positioning uncertainties strain on accuracy of PDD if point of interest is more shallo than the calibration depth of 2 cm (i.e., errors get blon up )

7 In-phantom method: Uncertainty at the reference point Ma et al, 1998 In-phantom method Type of quantity or procedure Uncertainty N K from standards lab 0.7% Effect of beam-quality difference 2.0% Chamber correction factor P Q,cham 1.5% Sleeve correction factor P sheath 0.5% µ en 1.5% ρ air ater In-ater measurement in user s 2.0% beam Combined D,z=2cm 3.6% II. Relative Dosimetry Relative dosimetry - hat do e require hich detectors can one use to measure PDD and profiles? spatial resolution requirements energy dependence requirements µ D ( z) N ( ) P en = KMz Q, champsheath ( z) ρ air Ma and Seuntjens, 1997

8 Ma et al, 1998 Ma and Seuntjens, 1997 RK RK 110 NACP 100 NACP diode 90 diode Farmer Farmer 100 Spokas 80 Spokas TLD 70 TLD N N Li et al, depth (g/cm 2 ) Markus Capintec depth (g/cm 2 ) Li et al, 1999 Markus Capintec Dosimeters for relative dosimetry Phantoms Acceptable dosimeters At depth: cylindrical chambers ith favourable energy response Unacceptable dosimeters General: chambers ith bad in-phantom response characteristics At surface: plane parallel Chambers ith high-z electron chambers (tested: components in or near NACP, Markus) detection volume radiochromic film, diamond detectors, liquid ionization chambers Diodes, silver-based film, TLD? plastic phantom materials useful for QA of output NOT for absolute calibrations unless investigated

9 depth (g/cm 2 ) NACP/H2O Farmer/H2O NACP/Acrylic Farmer/MS11 NACP/MS11 NACP/Polyst. Li et al, 1999 Uncertainties in the dose at other points in ater Type of quantity or procedure Combined D,z=2cm (in-phantom) 3.6% Combined D,z=0cm (in-air) 3.5% Determination of dose at other points in ater Uncertainty 3.0% Overall 4.7% Surface dose and electron contamination Other clinical issues TG-61 is a kerma-based protocol, i.e, surface dose cannot be derived from an air-kerma protocol Surface dose, should be assessed using thin indo plane parallel chambers and dealt ith if clinically important Podgorsak et al, 1990 Podgorsak et al, 1990

10 Dose in biological tissues TG-61 provides a method to calculate dose at the surface of tissue phantoms med med B med µ en C = B ρ air Dose in biological tissues for soft tissues the ratio of the backscatter factors is ignored; for compact bone, a table is provided no recommendations on ho to calculate dose at depth in the patient Seuntjens and Ma, 1999 Seuntjens and Ma, 1999 Conclusions TG-61 is an air-kerma based protocol recommending in-air method (lo and medium energy if point of interest is at the surface) in-phantom method (medium energy if point of interest is at a depth in-phantom) We discussed: calibration issues relative dosimetry some clinical issues

Zoubir Ouhib Lynn Cancer Institute

Zoubir Ouhib Lynn Cancer Institute Zoubir Ouhib Lynn Cancer Institute April 2015 Speaker for ELEKTA Brachytherapy using miniature X-ray sources Energy

More information

Abstract: Introduction:

Abstract: Introduction: Are photon and electron beam calibrations more consistent with TG-51 than with TG-21? Ramesh C. Tailor, William F. Hanson, Nathan Wells, and Geoffrey S. Ibbott. U.T. MD Anderson Cancer Center, Houston

More information

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers EXRADIN Ion Chambers Exradin Ion Chambers Exradin (EXacting RADiation INstrumentation) Ion Chambers have been built for over 33 years, are recognized by top research institutes and standards laboratories,

More information

Relative Dosimetry. Photons

Relative Dosimetry. Photons Relative Dosimetry Photons What you need to measure! Required Data (Photon) Central Axis Percent Depth Dose Tissue Maximum Ratio Scatter Maximum Ratio Output Factors S c & S cp! S p Beam profiles Wedge

More information

Commissioning and periodic tests of the Esteya electronic brachytherapy system

Commissioning and periodic tests of the Esteya electronic brachytherapy system Review paper Educational Activity Commissioning and periodic tests of the Esteya electronic brachytherapy system Cristian Candela-Juan, PhD 1, Yury Niatsetski, MSc 2, Zoubir Ouhib, PhD 3, Facundo Ballester,

More information

IAEA-TECDOC-1455 Implementation of the International Code of Practice on Dosimetry in Radiotherapy (TRS 398): Review of testing results

IAEA-TECDOC-1455 Implementation of the International Code of Practice on Dosimetry in Radiotherapy (TRS 398): Review of testing results IAEA-TECDOC-1455 Implementation of the International Code of Practice on Dosimetry in Radiotherapy (TRS 398): Revie of testing results Final report of the Coordinated Research Projects on Implementation

More information

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08. Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas 2009 AAPM Summer School Educational Objectives Improve your understanding of how to implement TG-51

More information

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas Current Implementation Status Current Implementation Status 1600 1400 TOTAL 1494 of 1623 ACTIVE INSTITUTIONS

More information

Review of fundamental photon dosimetry quantities

Review of fundamental photon dosimetry quantities Review of fundamental photon dosimetry quantities Narayan Sahoo Main sources of the materials included in this lecture notes are: (1) Radiation Oncology Physics: A Handbook for Teachers and Students Edited

More information

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Tom Jordan Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE

More information

Monitor Unit Verification for Small Fields

Monitor Unit Verification for Small Fields Monitor Unit Verification for Small Fields Patrick Higgins, Ph.D University of Minnesota Department of Radiation Oncology October 10, 2013 The Issues: How do we verify the monitor units calculated by the

More information

The MC commissioning of CyberKnife with MLC (Tips and Tricks)

The MC commissioning of CyberKnife with MLC (Tips and Tricks) The MC commissioning of CyberKnife with MLC (Tips and Tricks) Alain Guemnie Tafo, PhD UPMC Pinnacle, Harrisburg, PA Institute Gustave Roussy, Villejuif, France Disclaimer & Disclosure The views expressed

More information

Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin

Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin NCCAAPM Meeting Oct 11,2013 Larry DeWerd has a partial interest in Standard Imaging Talks from people who have been

More information

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams Peter R. Almond Brown Cancer Center, Louisville, Kentucky 40202 Peter J. Biggs Department of Radiation Oncology,

More information

TG-51: Experience from 150 institutions, common errors, and helpful hints

TG-51: Experience from 150 institutions, common errors, and helpful hints TG-51: Experience from 150 institutions, common errors, and helpful hints Ramesh C. Tailor, Ph. D., William F. Hanson, Ph.D., and Geoffrey S. Ibbott, Ph.D. Department of Radiation Physics, University of

More information

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics. FRANZCR Examination Part I Radiation Oncology Radiotherapeutic Physics Candidate No.: The Royal Australian and New Zealand College of Radiologists FRANZCR Examination Part I Radiation Oncology Radiotherapeutic

More information

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13.1.1 TRAINING GOAL 13.1.1.1 Among the responsabilities of a medical

More information

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation This intercomparison will be carried out under the lead of KRISS with the cooperation of ARPANSA. This comparison has been approved

More information

Review of backscatter measurement in kilovoltage radiotherapy using novel detectors and reduction from lack of underlying scattering material

Review of backscatter measurement in kilovoltage radiotherapy using novel detectors and reduction from lack of underlying scattering material JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 6, 2013 Review of backscatter measurement in kilovoltage radiotherapy using novel detectors and reduction from lack of underlying scattering

More information

Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry

Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry Bryan R. Muir a) and Malcolm R. McEwen Measurement Science and Standards, National Research Council of

More information

INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS

INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS A9744593 INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSLS Ladislav Czap, Georg Matscheko and Pedro Andreo osimetry and Medical Radiation Physics Section, ivision

More information

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface Chapter 11 reatment Planning Single Beams Radiation Dosimetry I Outline Basic terminology Curved contour surface correction (bolus, compensators, wedges) Oblique beam incidence Correction for tissue inhomogeneities

More information

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting Monitor Unit Calculations Part 2 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Calculation of machine setting reference dose machine setting =, reference dose

More information

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2017, 6, 57-72 http://www.scirp.org/journal/ijmpcero ISSN Online: 2168-5444 ISSN Print: 2168-5436 Validation of Treatment

More information

Original Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams

Original Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams Iranian Journal of Medical Physics Vol. 10, No. 3, Summer 2013, 147-155 Received: April 22, 2013; Accepted: July 21, 2013 Original Article Effective Point of Measurement in Cylindrical Ion Chamber for

More information

Proton Therapy QA Tools

Proton Therapy QA Tools Proton Therapy QA Tools A complete range of tools for absolute dosimetry, machine QA and patient plan verification Introduction During the past years more and more heavy particle therapy centers, especially

More information

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 DISCLAIMER: This publication and associated spreadsheets and digital files are based on sources and

More information

Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens

Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens The characteristic ionizing radiation response is defined by the relationship between the stimulus

More information

Constancy checks of well-type ionization chambers with external-beam radiation units

Constancy checks of well-type ionization chambers with external-beam radiation units JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Constancy checks of well-type ionization chambers with external-beam radiation units Sara L. Hackett, 1a Benjamin Davis, 2 Andrew

More information

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Commissioning an IMRT System for MLC Delivery Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Taking the broad view of commissioning Commissioning elements Validating the dosimetry system Commissioning the

More information

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 47 CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 4.1 INTRODUCTION Advanced treatment techniques use optimized radiation beam intensities to conform dose distribution to the

More information

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11 MV Photon Dosimetry in the Clinic This document discusses both the scientific methods and the system of work required to calibrate a field instrument expanding upon the practical details of both.. Introduction

More information

THE development of more advanced techniques in radiotherapy,

THE development of more advanced techniques in radiotherapy, 1 A Geant4-based simulation of an accelerator s head used for Intensity Modulated Radiation Therapy F. Foppiano, B. Mascialino, M.G. Pia, M. Piergentili. Abstract We present a Geant4-based simulation,

More information

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL FOR THE DEGREE MASTER OF ARTS BY CHARLES TRAVIS WEBB

More information

Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water

Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water M R McEwen, A J Williams and A R DuSautoy Centre for Ionising Radiation Metrology National Physical

More information

In air calibration of an HDR 192 Ir brachytherapy source using therapy ion chambers

In air calibration of an HDR 192 Ir brachytherapy source using therapy ion chambers Original Article Free full text available from www.cancerjournal.net In air calibration of an HDR 192 Ir brachytherapy source using therapy ion chambers ABSTRACT The Gammamed Plus 192 Ir high dose rate

More information

Technical protocol of APMP key comparison for measurement of air kerma for 60 Co (APMP.RI(I)-K1.1)

Technical protocol of APMP key comparison for measurement of air kerma for 60 Co (APMP.RI(I)-K1.1) Technical protocol of APMP key comparison for measurement of air kerma for 60 Co (APMP.RI(I)-K1.1) 1. Introduction A regional APMP.RI(I)-K1 comparison of the standards for air kerma for 60 Co occurred

More information

High Speed Direct SAD Radiosurgery Beam Scanner

High Speed Direct SAD Radiosurgery Beam Scanner Open Access Original Article DOI: 10.7759/cureus.20 High Speed Direct SAD Radiosurgery Beam Scanner Walter Nikesch 1, James M. Hevezi 2, Irene Monterroso 3, Daniel Navarro 4, James G. Schwade 5 1. CyberKnife

More information

1. Question Answer cgy / MU cgy / MU 2. Question Answer

1. Question Answer cgy / MU cgy / MU 2. Question Answer GS020113: Introduction to Medical Physics III: Therapy s to home work problem set assigned on 3/22/11 1. Question A patient is set up at 100 cm SSD on a 6 MVX machine. The dose rate at 10 cm in phantom

More information

IPEM Working Party: D IThwaites(Chair), A R DuSautoy, T Jordan, MRMcEwen, A Nisbet, A E Nahum and W G Pitchford

IPEM Working Party: D IThwaites(Chair), A R DuSautoy, T Jordan, MRMcEwen, A Nisbet, A E Nahum and W G Pitchford INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 48 (2003) 2929 2970 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(03)65222-X The IPEM code of practice for electron dosimetry for radiotherapy beams of

More information

Dosimetric Calculations. Lonny Trestrail

Dosimetric Calculations. Lonny Trestrail Dosimetric Calculations Lonny Trestrail 20 October 2008 Objectives Dose Distribution Measurements PDD, OCR TAR, SAR, TPR, TMR, SPR, SMR Arc or Rotational Therapy Isodose Curves Point Dose Calculations

More information

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator Journal of BUON 8: 55-59, 2003 2003 Zerbinis Medical Publications. Printed in Greece ORIGINAL ARTICLE Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

More information

IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy

IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy Guidelines on standardized procedures at Secondary Standards Dosimetry Laboratories (SSDLs) and hospitals March 2002 The

More information

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World Chattanooga, Tennessee, April 17-21, 2005, on CD-ROM, American Nuclear Society, LaGrange Park, IL (2005) GEANT4 SIMULATION OF

More information

Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media

Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media N Tyagi 1, B H Curran, P L Roberson, J M Moran, E Acosta and B A Fraass Department

More information

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Poster No.: R-0044 Congress: Type: Authors: Keywords: DOI: 2014 CSM Scientific Exhibit L. Bendall, I. Patel, N. McGrath, C. Rowbottom; MANCHESTER/

More information

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription Monitor Unit Calculations Part 1 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Return to our first patient 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt

More information

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN 22 Turkish Journal of Cancer Volume 37, No.1, 2007 Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN SATISH PELAGADE 1, KALPANA

More information

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system G. Kunz, F. Hasenbalg, P. Pemler 1 1 Klinik für Radio-Onkologie und Nuklearmedizin, Stadtspital Triemli

More information

Maintenance and general care of secondary standard and field instruments

Maintenance and general care of secondary standard and field instruments Maintenance and general care of secondary standard and field instruments This is without doubt the most important lecture that you will be given during this course. That rather rash statement is based

More information

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER Proven Rad-Check technology specifically designed to provide you with the ultimate in versatility and cost-effective operation. Fast and easy

More information

arxiv: v1 [physics.ins-det] 16 Aug 2014

arxiv: v1 [physics.ins-det] 16 Aug 2014 Preprint typeset in JINST style - HYPER VERSION RPC Gap Production and Performance for CMS RE4 Upgrade arxiv:1408.3720v1 [physics.ins-det] 16 Aug 2014 Sung Keun Park, Min Ho Kang and Kyong Sei Lee Korea

More information

E X R A D I N. Parallel Plate Ion Chambers User Manual DOC #

E X R A D I N. Parallel Plate Ion Chambers User Manual DOC # E X R A D I N Parallel Plate Ion Chambers User Manual R DOC #80332-02 E X R A D I N Parallel Plate Ion Chambers User Manual STANDARD IMAGING INC. 7601 Murphy Drive Middleton, WI 53562 TEL 800.261.4446

More information

Exposure System Selection

Exposure System Selection Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others

More information

User Manual. RW3 Slab Phantom T29672 and T

User Manual. RW3 Slab Phantom T29672 and T User Manual RW3 Slab Phantom Contents Operating Manual Technical Manual Service Manual D188.131.00/07 en 2015-07 Hn General Information General Information The product bears the CE-mark "CE" in accordance

More information

Math Review. Overview

Math Review. Overview Drawn by Steve Yan, CMD Math Review Shirley Ann Pinegar-Johnston MS RT(R)(T)CMD Overview Drawn by Steve Yan, CMD Divergence & Intensity DIVERGENCE X-rays travel in Straight but Divergent lines INTENSITY

More information

Perturbation Correction of a Cylindrical Thimble-type Chamber. in a Graphite Phantom for 60Co Gamma Rays*

Perturbation Correction of a Cylindrical Thimble-type Chamber. in a Graphite Phantom for 60Co Gamma Rays* Rapport BIPM-86/l4 Perturbation Correction of a Cylindrical Thimble-type Chamber in a Graphite Phantom for 60Co Gamma Rays* by C.E. de Almeida**, A.-M. Perroche-Roux and M. Boutillon Bureau International

More information

Commissioning and quality assurance for the treatment delivery components of the AccuBoost system

Commissioning and quality assurance for the treatment delivery components of the AccuBoost system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 2, 2015 Commissioning and quality assurance for the treatment delivery components of the AccuBoost system Ileana Iftimia, 1,2a Mike Talmadge,

More information

A beam-matching concept for medical linear accelerators

A beam-matching concept for medical linear accelerators Acta Oncologica, 2009; 48: 192200 ORIGINAL ARTICLE A beam-matching concept for medical linear accelerators DAVID SJÖSTRÖM, ULF BJELKENGREN, WIVIANN OTTOSSON & CLAUS F. BEHRENS Copenhagen University Hospital,

More information

Comparison of ionization chambers of various volumes for IMRT absolute dose verification

Comparison of ionization chambers of various volumes for IMRT absolute dose verification Comparison of ionization chambers of various volumes for IMRT absolute dose verification Leonid B. Leybovich, a) Anil Sethi, and Nesrin Dogan Department of Radiation Oncology, Loyola University Medical

More information

Victoreen B. Operators Manual. Image Intensifier Ion Chamber

Victoreen B. Operators Manual. Image Intensifier Ion Chamber Victoreen 6000-530B Image Intensifier Ion Chamber Operators Manual March 2005 Manual No 6000-530B-1 Rev. 4 2004, 2005 Fluke Corporation, All rights reserved. All product names are trademarks of their respective

More information

Evaluation of six TPS algorithms in computing entrance and exit doses

Evaluation of six TPS algorithms in computing entrance and exit doses JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Evaluation of six TPS algorithms in computing entrance and exit doses Yun I. Tan, 1,2a Mohamed Metwaly, 3 Martin Glegg, 3 Shaun P.

More information

Varian HDR surface applicators commissioning and clinical implementation

Varian HDR surface applicators commissioning and clinical implementation JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Varian HDR surface applicators commissioning and clinical implementation Ileana Iftimia, 1,2,a Andrea B. McKee, 1,2 and Per H. Halvorsen

More information

General Accreditation Guidance. User checks and maintenance of laboratory balances

General Accreditation Guidance. User checks and maintenance of laboratory balances General Accreditation Guidance User checks and maintenance of laboratory balances January 2018 Copyright National Association of Testing Authorities, Australia 2010 All intellectual property rights in

More information

Radiation Protection Service RADIOISOTOPES : SPECIFIC PROCEDURES AND HAZARD ASSESSMENT

Radiation Protection Service RADIOISOTOPES : SPECIFIC PROCEDURES AND HAZARD ASSESSMENT Radiation Protection Service RADIOISOTOPES : AND Revised November 2008 1 TABLE OF CONTENTS Alpha Emitters... 3 H3, C14 & S35... 3 Ba-133... 4 Ca-45... 4 Cl-36... 5 Co-57... 6 Cr-51... 7 Cu-64... 7 I-125

More information

Density of Brass: Accuracy and Precision

Density of Brass: Accuracy and Precision Density of Brass: Accuracy and Precision Introduction Density is a measure of a substance s mass-to-volume ratio. For liquids and solids, density is usually expressed in units of g/ml or g/cm 3 ; these

More information

HEALTH AND SAFETY SERVICES

HEALTH AND SAFETY SERVICES . HEALTH AND SAFETY SERVICES WELLBEING, SAFETY AND HEALTH LOCAL RULES FOR THE USE OF X-RAY EQUIPMENT??FACULTY??? /??FACILITY?? DOCUMENT CONTROL These local rules should be reviewed annually. Date Signed

More information

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization Basics of Proton Therapy Proton Treatment Planning and Beam Optimization SAM Educational Session, WE D BRB 2 Mark Pankuch, PhD Northwestern Medicine Chicago Proton Center Today s objectives Review the

More information

HMS/SHMS Drift Chambers Calibration. Carlos Yero June 26, 2017

HMS/SHMS Drift Chambers Calibration. Carlos Yero June 26, 2017 HMS/SHMS Drift Chambers Calibration Carlos Yero June 26, 2017 SHMS Detector Stack Drift Chambers are tracking detectors Calibration is necessary for high precision particle track reconstruction Calibration

More information

ARCHIMEDES PRINCIPLE AND THE COMPUTATION OF BUOYANT FORCES. Alexis Rodriguez-Carlson

ARCHIMEDES PRINCIPLE AND THE COMPUTATION OF BUOYANT FORCES. Alexis Rodriguez-Carlson ARCHIMEDES PRINCIPLE AND THE COMPUTATION OF BUOYANT FORCES Alexis Rodriguez-Carlson September 20, 2006 Purpose: The purpose of this experiment is to show that the buoyant force acting on an object submerged

More information

Measurements. Metric System

Measurements. Metric System Measurements Measurements are basic to any scientific pursuit. A measurement has both a magnitude (numeric value) and a unit. Metric units are used in the sciences. Metric System In science, the metric

More information

Radiotherapy physics & Equipments

Radiotherapy physics & Equipments Radiotherapy physics & Equipments RAD 481 Lecture s Title: Beam Modification devices in Radiotherapy Dr. Mohammed Emam Vision :IMC aspires to be a leader in applied medical sciences, health care education

More information

Phys101 Lectures Fluids I. Key points: Pressure and Pascal s Principle Buoyancy and Archimedes Principle. Ref: 10-1,2,3,4,5,6,7.

Phys101 Lectures Fluids I. Key points: Pressure and Pascal s Principle Buoyancy and Archimedes Principle. Ref: 10-1,2,3,4,5,6,7. Phys101 Lectures 21-22 Fluids I Key points: Pressure and Pascal s Principle Buoyancy and Archimedes Principle Ref: 10-1,2,3,4,5,6,7. Page 1 10-1 Phases of Matter The three common phases of matter are solid,

More information

HASTAC High stability Altimeter SysTem for Air data Computers

HASTAC High stability Altimeter SysTem for Air data Computers HASTAC High stability Altimeter SysTem for Air data Computers André Larsen R&D Director Memscap Sensor Solution SIXTH FRAMEWORK PROGRAMME PRIORITY 4 Aeronautics and Space EC contract no. AST4-CT-2005-012334

More information

UNIVERSITY OF EDINBURGH GENERIC RISK ASSESSMENT - RADIATION RP/RA8 - WORK WITH X-RAY FLUORESCENCE ANALYSERS

UNIVERSITY OF EDINBURGH GENERIC RISK ASSESSMENT - RADIATION RP/RA8 - WORK WITH X-RAY FLUORESCENCE ANALYSERS UNIVERSITY OF EDINBURGH GENERIC RISK ASSESSMENT - RADIATION RP/RA8 - WORK WITH X-RAY FLUORESCENCE ANALYSERS Foreword This is a generic risk assessment for use by persons involved in or effected by the

More information

Underwater measurement of photosynthetically active radiation

Underwater measurement of photosynthetically active radiation Underwater measurement of photosynthetically active radiation Mark Blonquist, Damon Nitzel, and Bruce Bugbee Apogee Instruments, Inc., Logan, Utah Introduction Quantum sensors (photosynthetically active

More information

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components Introduction The CyberKnife SRS/SBRT System James M. Hevezi,, Ph.D., FACR CyberKnife Unit San Antonio, Texas CyberKnife Components Treatment Delivery (G3 Platform San Antonio) Robot (modified German auto

More information

Performance of the Liquid Hydrogen Target

Performance of the Liquid Hydrogen Target Performance of the Liquid Hydrogen Target Presented by Chad Gillis Indiana University at NPDGamma Collaboration Meeting July 2012, Oak Ridge, Tennessee Since upgrading the target for operation at the SNS,

More information

Victoreen and

Victoreen and Victoreen 06-526-5240 and 06-526-5242 RAD-CHECK MICRO-R Operators Manual February 2005 Manual No. 06-526-5240-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names

More information

Radiation Producing Machines Radiation Safety Training for X-ray Diffraction Units

Radiation Producing Machines Radiation Safety Training for X-ray Diffraction Units INTRODUCTION Radiation Producing Machines Radiation Safety Training for X-ray Diffraction Units RADIATION SAFETY GUIDE IMAGING AND MICROSCOPY FACILITY University of California, Merced Training Document

More information

Badan Jadrowych Nuclear Research Institute. Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P Differential Recombination Chamber

Badan Jadrowych Nuclear Research Institute. Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P Differential Recombination Chamber Badan Jadrowych Nuclear Research Institute Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P00100517 Differential Recombination Chamber by M. Zel'chinskij K. Zharnovetskij Warsaw, June 1966 Translated

More information

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007 The Effect of Immobilisation Devices on Radiotherapy Dose Distributions A thesis submitted in fulfilment of the requirements for the Degree of Master of Applied Science (Medical and Health Physics) By

More information

DENSITY CALIBRATIONS AT MIKES. Heikki Kajastie, Jorma Manninen and Kari Riski Centre for Metrology and Accreditation (MIKES) ABSTRACT

DENSITY CALIBRATIONS AT MIKES. Heikki Kajastie, Jorma Manninen and Kari Riski Centre for Metrology and Accreditation (MIKES) ABSTRACT DENSITY CALIBRATIONS AT MIKES Heikki Kajastie, Jorma Manninen and Kari Riski Centre for Metrology and Accreditation (MIKES) ABSTRACT The density measurement instruments at MIKES will be described. Also

More information

IN-PLACE DENSITY AND MOISTURE CONTENT OF SOIL AND SOIL- AGGREGATE BY NUCLEAR METHODS (SHALLOW DEPTH) FOP FOR AASHTO T 310

IN-PLACE DENSITY AND MOISTURE CONTENT OF SOIL AND SOIL- AGGREGATE BY NUCLEAR METHODS (SHALLOW DEPTH) FOP FOR AASHTO T 310 AND MOISTURE CONTENT OF SOIL AND SOIL- AGGREGATE BY NUCLEAR METHODS (SHALLOW DEPTH) FOP FOR AASHTO T 310 Scope This procedure covers the determination of density, moisture content, and relative compaction

More information

The dose distribution of medium energy electron boosts to the laryngectomy stoma

The dose distribution of medium energy electron boosts to the laryngectomy stoma JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 1, WINTER 2001 The dose distribution of medium energy electron boosts to the laryngectomy stoma Ellen D. Yorke* Memorial Sloan-Kettering Cancer

More information

WMO LABORATORY INTERCOMPARISON OF RAIN INTENSITY GAUGES

WMO LABORATORY INTERCOMPARISON OF RAIN INTENSITY GAUGES WMO LABORATORY INTERCOMPARISON OF RAIN INTENSITY GAUGES Christophe ALEXANDROPOULOS and Muriel LACOMBE Météo-France, Direction des Systèmes d Observation, BP 202-78195 Trappes France christophe.alexandropoulos@meteo.fr

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 Comparison of monitor unit calculations performed with a 3D computerized planning system and independent hand calculations: Results

More information

Validation of Custody Transfer Metering Skid at Site After Laboratory Proving

Validation of Custody Transfer Metering Skid at Site After Laboratory Proving flotek.g 2017- Innovative Solutions in Flow Measurement and Control - Oil, Water and Gas August 28-30, 2017, FCRI, Palakkad, Kerala, India Validation of Custody Transfer Metering Skid at Site After Laboratory

More information

How to Design Medical Accelerator Systems for Reliability: IBA PT System

How to Design Medical Accelerator Systems for Reliability: IBA PT System How to Design Medical Accelerator Systems for Reliability: IBA PT System Yves Jongen Founder & Chief Research Officer Ion Beam Applications s.a. Belgium 1 Outline A short introduction of a proton therapy

More information

Clinical Implementation of Volumetric Modulated Arc Therapy

Clinical Implementation of Volumetric Modulated Arc Therapy Clinical Implementation of Volumetric Modulated Arc Therapy UT M.D. Anderson Cancer Center Ramaswamy Sadagopan, Rebecca M. Howell, Weiliang Du and Peter Balter Definition 2 Intensity Modulated Arc therapy

More information

The use of the analytical balance, and the buret.

The use of the analytical balance, and the buret. 1211L Experiment 1. Density 2015 by H. Patterson Instructor Notes: Students make measurements individually then share data to make the graph. There are four volumetric measurements to be studied; 3.00

More information

Phys101 Lectures Fluids I. Key points: Pressure and Pascal s Principle Buoyancy and Archimedes Principle. Ref: 10-1,2,3,4,5,6,7.

Phys101 Lectures Fluids I. Key points: Pressure and Pascal s Principle Buoyancy and Archimedes Principle. Ref: 10-1,2,3,4,5,6,7. Phys101 Lectures 24-25 luids I Key points: Pressure and Pascal s Principle Buoyancy and Archimedes Principle Ref: 10-1,2,3,4,5,6,7. Page 1 10-1 Phases of Matter The three common phases of matter are solid,

More information

Health & Safety Policy and Procedures Manual SECTION 38 RADIATION SAFETY PROGRAM

Health & Safety Policy and Procedures Manual SECTION 38 RADIATION SAFETY PROGRAM SECTION 38 RADIATION SAFETY PROGRAM 1. GENERAL REQUIREMENTS 1. Maul Electric, Inc radiation safety program along with 29CFR 1910.1096 and 10CFR 20.1101 include provisions for project sites to keep radiation

More information

EXPERIMENT 5 FIELD DENSITY TEST: SAND REPLACEMENT METHOD (PREPARED BY : MUHAMMAD MUNSIF AHMAD)

EXPERIMENT 5 FIELD DENSITY TEST: SAND REPLACEMENT METHOD (PREPARED BY : MUHAMMAD MUNSIF AHMAD) 1.0 OBJECTIVE EXPERIMENT 5 FIELD DENSITY TEST: SAND REPLACEMENT METHOD (PREPARED BY : MUHAMMAD MUNSIF AHMAD) To determine the in-situ density of fine and medium rained soil. 2.0 INTRODUCTION In most specifications

More information

Vaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated.

Vaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated. Asset #: 123 Calibration - Certificate No: 2083.01 Customer: Condition: Action Taken: Sample Inc. 123 Sample Rd. Sample, MA 01234 The instrument was operational upon receipt. The instrument was adjusted

More information

deltaflowc deltaflowc Venturi or Probe

deltaflowc deltaflowc Venturi or Probe deltaflowc Mass Flowmeter for Gases - Multivariable with ultra fast dp, p and T-sensors - Compact, accurate and user-friendly - Ideal for OEMs deltaflowc Venturi or Probe Precise mass flow metering deltaflowc

More information

2014 Programme of Training Courses in Measurement & Calibration

2014 Programme of Training Courses in Measurement & Calibration 2014 Programme of Training Courses in Measurement & Calibration NSAI National Metrology Laboratory offers public courses to help individuals gain the skills that they require in the field of Measurement

More information

Pressure Metrology in Mexico

Pressure Metrology in Mexico Pressure Metrology in Mexico Jorge C. TORRES G., Benjamin SORIANO C. and Victor ARANDA C. Centro Nacional de Metrologia-CENAM km 4,5 Carretera a Los Cues, El Marques, Queretaro, MEXICO e-mail: jorge.torres@cenam.mx

More information

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards.

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards. 1. Scope Description of requirements and procedures for compact provers to be used as verification standards. 2. Reference documents Trade Metrology Act SANS1698 3. Policy A. BASIC REQUIREMENTS Compact

More information

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision Minnesota Rules, Chapter 4732 X-ray Revision DRAFT X-RAY SYSTEMS FOR SUSPECTED HAZARDS, 2.0 4732.#### X-RAY SYSTEMS FOR SUSPECTED HAZARDS. Summary of Changes MDH made a number of changes to the X-Ray Systems

More information