Relative Dosimetry. Photons
|
|
- Ross Stevenson
- 5 years ago
- Views:
Transcription
1 Relative Dosimetry Photons
2 What you need to measure! Required Data (Photon) Central Axis Percent Depth Dose Tissue Maximum Ratio Scatter Maximum Ratio Output Factors S c & S cp! S p Beam profiles Wedge Factors, Percent Depth Dose Dynamic Wedge Factors Block Tray Factors Block Factor MLC Transmission Factor Beam Profiles Off-Axis factors
3 PDD Source PDD pri = 100 D pri Q pri D P f = SSD How to Position IC? When properly placed at the surface, the chamber, viewed end-on from beneath the water, will appear to be a complete circle. P Q A z max FS. 4cm 2-40cm 2!"5cm,35-40cm depth z
4 PDD PDD pri = 100 D pri Q pri D P Pitfalls: D max Small filed size measurements d max shift Voltage bias effect Dose D s D s D s 0 d max Depth
5 PDD of MV Photon Beams PDD Co SSD = 100 cm cm Depth in water (cm)
6 PDD Build-up to dmax Higher energy beams have more penetrating photons, and create more energetic electrons. Thus, the maximum dose occurs at a deeper depth for high energy beams. Typical dmax Beam Energy Ortho 0 60 Co 0.5 4MV 1.0 6MV MV MV MV MV 3.5 dmax (cm)
7 Buildup Region Interaction of incident photons with phantom yield secondary electrons. Secondary electrons deposit energy downstream. The electron fluence and, thus, absorbed dose increases until reach a maximum (approximate range of electrons) Same time -- photon fluence decreases with depth at constant rate. Therefore, fewer photons to eject electrons as increase depth. The combination of all these factors yields the dose buildup region
8 PDD increase beam energy, increase depth of d max Higher energies more skin sparing Lower energies less skin sparing
9 PDD Field Size Dependence 15 MV Photon Beam 4x4 16x16
10 Percent Depth Dose Decreases with depth for all energies beyond build-up region. Increases with energy beyond build-up region. If inverse square and scattering are ignored, it follows exponential attenuation. Increase with Field Size.
11 Percent Depth Dose AAPM TG Report # 106 recommendations: Use 4-5 mm diameter ion chamber for depth beyond 1cm. Use parallel plate or extrapolation chamber to measure data near the surface. Diodes and diamond detectors are appropriate as long as data measured with these detectors is crossreferenced to data measured with an ion chamber. Prone to radiation damage and non-linear response.
12 Is depth ionization data depth dose? Yes! If: TCPE exists at the point of measurement. The energy spectrum of incident photons does not change with the depth. Fluence across the detector remains the same. These conditions are met at depths beyond the range of contaminant charged particles
13 Is depth ionization data depth dose? However at shallow depth, The contaminants and secondary electrons have energy spectra that change rapidly with depth. Results in a variation of ~10% in restricted mass stopping power ratio data for water and air. Translates into a spatial uncertainty of less than 1.5 mm in dose in the build up region
14 TMR TMR = D Q Source D Q max (b) Tissue Maximum Ratio Ratio of the dose at point Q in the phantom for various depths to the dose at point Q when the d Q =d max SAD Q A ref A Q z ref
15 TMR of MV Photon Beams Plot of TMR for 10-MV x-ray as a function of depth for a selection of field sizes.
16 PDD vs. TMR PDD TMR 1 SSD (Source to surface distance) is fixed, normally 100 cm. 1 SSD (Source to surface distance) varies. 2 Ion chamber moves in depth. Therefore, SPD (Source to point distance) varies. 2 Ion chamber does not move in depth. Therefore, SPD (Source to point distance) is fixed (i.e. 100 cm). 3 Measurements at different points are divided by the measurement at depth dmax. 3 Measurement at the same point in space (but different depth in water) is then divided by dose at that point in dmax depth.
17 Beam Profiles 6 MV Photon Beam, Depth of 5.0 cm, Field size of 4x4, 10.4x10.4, and 21x21 cm 2. The flatness of photon beams is extremely sensitive to change in energy of the incident beam. A small change in the penetrative quality of a photon beam results in very large change in beam flatness.
18 Beam Profiles Affected by the radially symmetric conical high Z- material flattening filter, which Flattens the beam by differentially absorbing more photons in the center and less in the periphery unwanted consequence of flattening the beam is the differential change in beam quality at off-axis points. hardens the beam Cross beam flatness is defined as: F = 100! D max " D min D max + D min One flattening filter for each clinical photon beam results in a compromise of beam flatness characteristics of small and large fields. Flattening filters are designed to give a gradually increasing radial intensity. This is referred to as horns on a cross-beam profile
19 Beam Profiles Note: Cross beam profiles may not be radially symmetric due to non circular focal spot. Therefore, cross-beam data is characterized by a set of two orthogonal dose profiles measured perpendicular to the beam s central axis at a given depth in a phantom
20 Beam Profiles 6 MV Photon Beam, Field Size of 10.4x10.4 cm2, Depths of 1.5, 5.0, 10.0, 15.0 and 25 cm
21 Isodose Distributions (20 X 20 Cm 2 ) 6 MV 18 MV Note contaminant electrons contribute to dose outside the field at shallow depths. The magnitude and extent of dose outside the geometric edge of a field at shallow depths increases with beam energy.
22 Cross Beam Measurements!"#$%&'#()#'%*#&+*,'#-.# /*'*,'-0#)(1*2## # 3"#45,-00*,'#6*&)70*6*5'# -.#8*57690�*:(-5## Diode CC04 CC13 Diameter Penumbra 20%~80% 0.8x0.8 mm 2 4 mm 6 mm 4.0 mm 6.1 mm 7.2 mm
23 Output factors A linear accelerator is calibrated by measuring ionization per monitor unit at the reference depth. This measurement is converted to dose and divided by appropriate PDD to determine the dose at d max for a standard field size and SSD (i.e, 10cm 2 and 100cm respectively). As the field size changes the radiation output changes. This change is quantified by measuring output at d max on CAX for each FS and dividing by output at d max on CAX for 10x10 cm 2 field)
24 Output factors OF(r) = S P (r)! S C (r) " S P (r) = S CP(r) S C (r) OF(r) or S cp : output factor at d max for field size rxr. S c (r): collimator scatter correction factor measured in air for field size rxr. S p (r): phantom scatter correction for filed size rxr. S cp should be measured for square field sizes from 2x2 to 40x40 cm 2
25 Measuring Sc & Sc,p
26 Measuring S c Mini-phantom Water-equivalent materials. 4g/cm2 diameter and 10g/cm2 depth to maintain lateral CPE and eliminate contaminant electron. For small segment fields(c<4cm),high Z material (Brass etc.) should be used. Corrections for energy absorption coefficients and energy spectra change are needed. TG 74 recommendations
27 How to get from PDD to TMR! TMR(d, r d ) = PDD(d,r, f ) $! # &i# " 100 % " f: SSD f + d f + d max $ & % 2! i S p(r d max ) $ # " S p (r d ) & % r: field size at the surface r d : field size at depth d; r d =r(f+d)/f r dmax : field size at d max ; r d max=r(f+d max )/f S p : Phantom Scatter correction factor
28 Scatter-Maximum ratios! SMR(d, r d ) = TMR(d, r d )i S (r ) $ p d # " S p (0) & ' TMR(d,0) % SMR is the ratio of scattered dose at a point to the effective primary dose at the same point at d max. TMR can be divided into two parts Primary component, TMR (d,0) Scatter component, SMR S p (0): phantom scatter correction for fs=0 (extrapolated)
29 Wedges Physical wedges are made of lead, brass or steel. When placed in a radiation beam, they cause a progressive decrease in the intensity across the beam and a tilt of isodose curves under normal beam incidence. Dynamic wedges provide the wedge effect on isodose curves through a closing motion of a collimator jaw during irradiation. The wedge angle is defined as the angle through which an isodose curve at a given depth in water (usually 10 cm) is tilted at the central beam axis under the condition of normal beam incidence.
30 Wedges Wedge/Open Comparison FS = 10 x 10 cm2 15 MV (W/O) 6 MV (W/O)
31 Wedge factors Measurment condition similar to that of S cp. Measurements for different wedged field d max divided by measurement for 10x10 nonewedged field. The IC should be placed on the CAX with its axis parallel to a constant thickness of the wedge. Two sets of reading is required with wedgeposition rotated 180 o between them. Don t forget wedge PDD and the d max shift.
32 Other factors Block Tray Factor MLC Transmission Factor Off-Axis factor, OAF
33 OAF The off-axis ratio (OAR) is usually defined as the ratio of dose at an offaxis point to the dose on the central beam axis at the same depth in a phantom. Relative dose Depth (cm) Distance from central axis axis (cm)
34 Approaches to Dose Computation Algorithms Data measured in water and in air Parameterize water data Correction based methods Model based methods Reconstitute water data Calculate inhomogeneity corrections to water data Calculate dose directly based on beam and phantom configurations
35 Correction vs. Model Based Methods Correction based methods Measured data used as basis for Dose Computation Require measurements with buildup cap in air or in a mini-phantom Require lots of data. Generating functions used to reduce size of data set for convenient clinical use (i.e. less storage space). Model based methods Measured data used to setup description of treatment beam. Require a parameter to estimate size of photon source at target. Require more time for tuning of model parameters. Patient dose distribution obtained by first computing Dose in water from generating function, then correcting for tissue heterogeneity, patient contour, and beam modifiers. computing beam and beam transport (i.e. beam interactions in treatment head and in patient) directly.
Review of fundamental photon dosimetry quantities
Review of fundamental photon dosimetry quantities Narayan Sahoo Main sources of the materials included in this lecture notes are: (1) Radiation Oncology Physics: A Handbook for Teachers and Students Edited
More informationDosimetric Calculations. Lonny Trestrail
Dosimetric Calculations Lonny Trestrail 20 October 2008 Objectives Dose Distribution Measurements PDD, OCR TAR, SAR, TPR, TMR, SPR, SMR Arc or Rotational Therapy Isodose Curves Point Dose Calculations
More informationOutline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface
Chapter 11 reatment Planning Single Beams Radiation Dosimetry I Outline Basic terminology Curved contour surface correction (bolus, compensators, wedges) Oblique beam incidence Correction for tissue inhomogeneities
More informationField size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator
Journal of BUON 8: 55-59, 2003 2003 Zerbinis Medical Publications. Printed in Greece ORIGINAL ARTICLE Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator
More information1. Question Answer cgy / MU cgy / MU 2. Question Answer
GS020113: Introduction to Medical Physics III: Therapy s to home work problem set assigned on 3/22/11 1. Question A patient is set up at 100 cm SSD on a 6 MVX machine. The dose rate at 10 cm in phantom
More informationMonitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription
Monitor Unit Calculations Part 1 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Return to our first patient 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt
More information13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.
13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13.1.1 TRAINING GOAL 13.1.1.1 Among the responsabilities of a medical
More informationMonitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting
Monitor Unit Calculations Part 2 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Calculation of machine setting reference dose machine setting =, reference dose
More informationMath Review. Overview
Drawn by Steve Yan, CMD Math Review Shirley Ann Pinegar-Johnston MS RT(R)(T)CMD Overview Drawn by Steve Yan, CMD Divergence & Intensity DIVERGENCE X-rays travel in Straight but Divergent lines INTENSITY
More informationThe Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.
FRANZCR Examination Part I Radiation Oncology Radiotherapeutic Physics Candidate No.: The Royal Australian and New Zealand College of Radiologists FRANZCR Examination Part I Radiation Oncology Radiotherapeutic
More informationCommissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale
Commissioning an IMRT System for MLC Delivery Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Taking the broad view of commissioning Commissioning elements Validating the dosimetry system Commissioning the
More informationMonitor Unit Verification for Small Fields
Monitor Unit Verification for Small Fields Patrick Higgins, Ph.D University of Minnesota Department of Radiation Oncology October 10, 2013 The Issues: How do we verify the monitor units calculated by the
More informationANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL
ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL FOR THE DEGREE MASTER OF ARTS BY CHARLES TRAVIS WEBB
More informationCommissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN
22 Turkish Journal of Cancer Volume 37, No.1, 2007 Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN SATISH PELAGADE 1, KALPANA
More informationRadiotherapy physics & Equipments
Radiotherapy physics & Equipments RAD 481 Lecture s Title: Beam Modification devices in Radiotherapy Dr. Mohammed Emam Vision :IMC aspires to be a leader in applied medical sciences, health care education
More informationTHE development of more advanced techniques in radiotherapy,
1 A Geant4-based simulation of an accelerator s head used for Intensity Modulated Radiation Therapy F. Foppiano, B. Mascialino, M.G. Pia, M. Piergentili. Abstract We present a Geant4-based simulation,
More informationSnapShot IMRT with compensators and FFF beams
SnapShot IMRT with compensators and FFF beams Vladimir Feygelman, PhD (1) Moffitt Cancer Center, Tampa, FL, USA Disclosure VF has a sponsored research agreement with.decimal. A short history of radiotherapy
More informationCHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS
47 CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 4.1 INTRODUCTION Advanced treatment techniques use optimized radiation beam intensities to conform dose distribution to the
More informationClinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry
Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Tom Jordan Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE
More informationOnly 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.
Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas 2009 AAPM Summer School Educational Objectives Improve your understanding of how to implement TG-51
More informationThe MC commissioning of CyberKnife with MLC (Tips and Tricks)
The MC commissioning of CyberKnife with MLC (Tips and Tricks) Alain Guemnie Tafo, PhD UPMC Pinnacle, Harrisburg, PA Institute Gustave Roussy, Villejuif, France Disclaimer & Disclosure The views expressed
More informationHigh Speed Direct SAD Radiosurgery Beam Scanner
Open Access Original Article DOI: 10.7759/cureus.20 High Speed Direct SAD Radiosurgery Beam Scanner Walter Nikesch 1, James M. Hevezi 2, Irene Monterroso 3, Daniel Navarro 4, James G. Schwade 5 1. CyberKnife
More informationClinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas
Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas Current Implementation Status Current Implementation Status 1600 1400 TOTAL 1494 of 1623 ACTIVE INSTITUTIONS
More informationZoubir Ouhib Lynn Cancer Institute
Zoubir Ouhib Lynn Cancer Institute April 2015 Speaker for ELEKTA Brachytherapy using miniature X-ray sources Energy
More informationCommissioning and periodic tests of the Esteya electronic brachytherapy system
Review paper Educational Activity Commissioning and periodic tests of the Esteya electronic brachytherapy system Cristian Candela-Juan, PhD 1, Yury Niatsetski, MSc 2, Zoubir Ouhib, PhD 3, Facundo Ballester,
More informationTG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:
Practical Implementation of TG-61: II. Guidelines for clinical implementation of TG-61 J.P. Seuntjens Medical Physics Unit McGill University, Montreal General Hospital Montréal, Canada jseuntjens@medphys.mcgill.ca
More informationCHAPTER-2. Wedge filters are the most important beam modifiers used in
CHAPTER-2 2. WEDGE FLTERS 2.1. NTRODUCTON Wedge filters are the most important beam modifiers used in radiotherapy which are placed in the path of photon beams to modify their isodose diistributions. The
More informationInvestigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy
University of South Florida Scholar Commons Graduate Theses and Dissertations Graduate School 7-14-2010 Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy
More informationExradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers
EXRADIN Ion Chambers Exradin Ion Chambers Exradin (EXacting RADiation INstrumentation) Ion Chambers have been built for over 33 years, are recognized by top research institutes and standards laboratories,
More informationAbsorption measurements for a carbon fiber couch top and its modelling in a treatment planning system
Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system G. Kunz, F. Hasenbalg, P. Pemler 1 1 Klinik für Radio-Onkologie und Nuklearmedizin, Stadtspital Triemli
More informationJOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 Comparison of monitor unit calculations performed with a 3D computerized planning system and independent hand calculations: Results
More informationBasics of Proton Therapy. Proton Treatment Planning and Beam Optimization
Basics of Proton Therapy Proton Treatment Planning and Beam Optimization SAM Educational Session, WE D BRB 2 Mark Pankuch, PhD Northwestern Medicine Chicago Proton Center Today s objectives Review the
More informationConstancy checks of well-type ionization chambers with external-beam radiation units
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Constancy checks of well-type ionization chambers with external-beam radiation units Sara L. Hackett, 1a Benjamin Davis, 2 Andrew
More information3D Treatment Planning and verification with hand calculations
3D Treatment Planning and verification with hand calculations Randy Holt, PhD, CEO Pacific Crest Medical Physics, Inc Chico, CA Dosimetric Calculations Calibration and Measurement Conditions Patient Conditions
More informationA beam-matching concept for medical linear accelerators
Acta Oncologica, 2009; 48: 192200 ORIGINAL ARTICLE A beam-matching concept for medical linear accelerators DAVID SJÖSTRÖM, ULF BJELKENGREN, WIVIANN OTTOSSON & CLAUS F. BEHRENS Copenhagen University Hospital,
More informationAAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams
AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams Peter R. Almond Brown Cancer Center, Louisville, Kentucky 40202 Peter J. Biggs Department of Radiation Oncology,
More informationTEST REPORT BEAMSCAN WITH HALCYON
TEST REPORT BEAMSCAN WITH HALCYON UZ Leuven, November 20 th, 2017 + = Ideal BEAMSCAN tank dimensions to fit perfectly into the Halcyon bore Fast tank positioning using the Halcyon lasers, no electronics
More informationWEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS
Acta Radiologica Oncology 23 (1984) Fusc. 6 FROM THE RADIATION THERAPY DEPARTMENT, ANTON1 VAN LEEUWENHOEK HOSPITAL, THE NETHERLANDS CANCER INSTITUTE, AMSTERDAM, THE PHILIPS MEDICAL SYSTEMS DIVISION, BEST,
More informationSurface buildup dose dependence on photon field delivery technique for IMRT
Surface buildup dose dependence on photon field delivery technique for IMRT Shigeru Yokoyama, Peter L. Roberson, Dale W. Litzenberg, Jean M. Moran, and Benedick A. Fraass Department of Radiation Oncology,
More informationBadan Jadrowych Nuclear Research Institute. Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P Differential Recombination Chamber
Badan Jadrowych Nuclear Research Institute Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P00100517 Differential Recombination Chamber by M. Zel'chinskij K. Zharnovetskij Warsaw, June 1966 Translated
More informationGEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY
The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World Chattanooga, Tennessee, April 17-21, 2005, on CD-ROM, American Nuclear Society, LaGrange Park, IL (2005) GEANT4 SIMULATION OF
More informationTG-51: Experience from 150 institutions, common errors, and helpful hints
TG-51: Experience from 150 institutions, common errors, and helpful hints Ramesh C. Tailor, Ph. D., William F. Hanson, Ph.D., and Geoffrey S. Ibbott, Ph.D. Department of Radiation Physics, University of
More informationIntroduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components
Introduction The CyberKnife SRS/SBRT System James M. Hevezi,, Ph.D., FACR CyberKnife Unit San Antonio, Texas CyberKnife Components Treatment Delivery (G3 Platform San Antonio) Robot (modified German auto
More informationPractical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11
MV Photon Dosimetry in the Clinic This document discusses both the scientific methods and the system of work required to calibrate a field instrument expanding upon the practical details of both.. Introduction
More informationInvestigation of Suction Process of Scroll Compressors
Purdue University Purdue e-pubs International Compressor Engineering Conference School of Mechanical Engineering 2006 Investigation of Suction Process of Scroll Compressors Michael M. Cui Trane Jack Sauls
More informationValidation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.
International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2017, 6, 57-72 http://www.scirp.org/journal/ijmpcero ISSN Online: 2168-5444 ISSN Print: 2168-5436 Validation of Treatment
More informationPlanning Assignment (3 field rectum)
Megan Sullivan Clinical Lab Rectum April 25, 2016 Planning Assignment (3 field rectum) Use a CT dataset of the pelvis. Create a CTV by contouring the rectum (start at the anus and stop at the turn where
More informationThe dose distribution of medium energy electron boosts to the laryngectomy stoma
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 1, WINTER 2001 The dose distribution of medium energy electron boosts to the laryngectomy stoma Ellen D. Yorke* Memorial Sloan-Kettering Cancer
More informationAbstract: Introduction:
Are photon and electron beam calibrations more consistent with TG-51 than with TG-21? Ramesh C. Tailor, William F. Hanson, Nathan Wells, and Geoffrey S. Ibbott. U.T. MD Anderson Cancer Center, Houston
More informationOUTPUT at dmax in tissue 8x8 20x20 At 100cm SSD At 100cm SAD
RAPHEX 2001 T1. All of the following are true regarding Percentage Depth Dose (PDD) except: A. Increases with increasing energy. B. Depends on field size. C. Is the dose at depth expressed as a % of the
More informationOriginal Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams
Iranian Journal of Medical Physics Vol. 10, No. 3, Summer 2013, 147-155 Received: April 22, 2013; Accepted: July 21, 2013 Original Article Effective Point of Measurement in Cylindrical Ion Chamber for
More informationTG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009
TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 DISCLAIMER: This publication and associated spreadsheets and digital files are based on sources and
More informationScientific Journal of Nuclear Medicine & Radiation Therapy
Scientific Journal of Nuclear Medicine & Radiation Therapy Research Article Radiotherapy Dose Calculation Program, for Monitoring and Commissioning Treatment Planning Systems - Ioannis Vamvakas*, Stella
More informationCommissioning of Elekta 6MV FFF Versa HD and Pinnacle
Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Poster No.: R-0044 Congress: Type: Authors: Keywords: DOI: 2014 CSM Scientific Exhibit L. Bendall, I. Patel, N. McGrath, C. Rowbottom; MANCHESTER/
More informationDynamic wedges dosimetry and quality control
Original Paper Received: 2005.06.07 Accepted: 2006.02.28 Published: 2006.04.28 Authors Contribution: A Study Design B Data Collection C Statistical Analysis D Data Interpretation E Manuscript Preparation
More informationTG219: IT'S USE, STRENGTHS AND WEAKNESSES
1 CONFLICT OF INTEREST TG219: IT'S USE, STRENGTHS AND WEAKNESSES SOTIRI STATHAKIS, PHD, DABR None ROSTER STATEMENT OF THE PROBLEM Timothy C. Zhu, University of Pennsylvania, Philadelphia, PA (Co- Chair)
More information7.65 ±0.05 mm of lead. The maximum energy of the bremsstrahlung photons was calibrated using the threshold for the D(-y,n)H reaction at
Am 7Roentgenolia6:ia6oia65, 1976 CENTRAL AXIS DEPTH DOSE FOR A 2.5 MV VAN DE GRAAFF GENERATOR DAVID W. ANDERSON, DAVID E. RAESIDE, REBA I. ADAMS, AND MYRON R. GOEDE 2 ABSTRACT: Central axis percentage
More informationProton Therapy QA Tools
Proton Therapy QA Tools A complete range of tools for absolute dosimetry, machine QA and patient plan verification Introduction During the past years more and more heavy particle therapy centers, especially
More informationIMRT in clinical practice at the UMC-Utrecht. Utrecht
3 IMRT in clinical practice at the UMC-Utrecht Utrecht Clinical use of IMRT Improvement of local control without increased incidence of normal tissue complications + Dose escalation in the prostate Step-and
More informationUnderwater measurement of photosynthetically active radiation
Underwater measurement of photosynthetically active radiation Mark Blonquist, Damon Nitzel, and Bruce Bugbee Apogee Instruments, Inc., Logan, Utah Introduction Quantum sensors (photosynthetically active
More informationGuideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation
Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation This intercomparison will be carried out under the lead of KRISS with the cooperation of ARPANSA. This comparison has been approved
More information1. BEAM MODIFIERS IN RADIOTHERAPY
CHAPTER-1 1. BEAM MODIFIERS IN RADIOTHERAPY INTRODUCTION In radiotherapy, tumours under varying body contour are very often encountered. In such cases, due to the irregular body surface, the dose distribution
More informationRadiation Producing Machines Radiation Safety Training for X-ray Diffraction Units
INTRODUCTION Radiation Producing Machines Radiation Safety Training for X-ray Diffraction Units RADIATION SAFETY GUIDE IMAGING AND MICROSCOPY FACILITY University of California, Merced Training Document
More informationCOMPUTATIONAL FLOW MODEL OF WESTFALL'S LEADING TAB FLOW CONDITIONER AGM-09-R-08 Rev. B. By Kimbal A. Hall, PE
COMPUTATIONAL FLOW MODEL OF WESTFALL'S LEADING TAB FLOW CONDITIONER AGM-09-R-08 Rev. B By Kimbal A. Hall, PE Submitted to: WESTFALL MANUFACTURING COMPANY September 2009 ALDEN RESEARCH LABORATORY, INC.
More informationHMS/SHMS Drift Chambers Calibration. Carlos Yero June 26, 2017
HMS/SHMS Drift Chambers Calibration Carlos Yero June 26, 2017 SHMS Detector Stack Drift Chambers are tracking detectors Calibration is necessary for high precision particle track reconstruction Calibration
More informationCOMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC
COMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC by TAN ZI XIANG Thesis submitted in fulfillment of the requirements for the degree of Master of Science (Medical Physics) February 2015 ACKNOWLEDGEMENT
More informationRADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER
RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER Proven Rad-Check technology specifically designed to provide you with the ultimate in versatility and cost-effective operation. Fast and easy
More informationA simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation
Radiotherapy and Oncology 71 (2004) 235 241 Technical Note A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation Nadine Linthout*,
More informationIntroduction. Materials and Methods
STANDARD WEDGE AND TRAY TRANSMISSION VALUES FOR VARIAN, SEIMENS, ELEKTA/PHILIPS ACCELERATORS; A QUALITY ASSURANCE TOOL David S. Followill, Nadia Hernandez, and William F. Hanson Introduction The Radiological
More informationPerturbation Correction of a Cylindrical Thimble-type Chamber. in a Graphite Phantom for 60Co Gamma Rays*
Rapport BIPM-86/l4 Perturbation Correction of a Cylindrical Thimble-type Chamber in a Graphite Phantom for 60Co Gamma Rays* by C.E. de Almeida**, A.-M. Perroche-Roux and M. Boutillon Bureau International
More informationTreatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation
Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation By Ruth Ann M. Good, CMD. R.T. (R)(T) Virginia Commonwealth University, Richmond,
More informationINTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS
A9744593 INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSLS Ladislav Czap, Georg Matscheko and Pedro Andreo osimetry and Medical Radiation Physics Section, ivision
More informationWIND-INDUCED LOADS OVER DOUBLE CANTILEVER BRIDGES UNDER CONSTRUCTION
WIND-INDUCED LOADS OVER DOUBLE CANTILEVER BRIDGES UNDER CONSTRUCTION S. Pindado, J. Meseguer, J. M. Perales, A. Sanz-Andres and A. Martinez Key words: Wind loads, bridge construction, yawing moment. Abstract.
More informationA dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions
A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions Wendel Dean Renner a) Math Resolutions, LLC, 5975 Gales Lane, Columbia,
More informationExposure System Selection
Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others
More informationChapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens
Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens The characteristic ionizing radiation response is defined by the relationship between the stimulus
More informationLarry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin
Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin NCCAAPM Meeting Oct 11,2013 Larry DeWerd has a partial interest in Standard Imaging Talks from people who have been
More informationIPEM Working Party: D IThwaites(Chair), A R DuSautoy, T Jordan, MRMcEwen, A Nisbet, A E Nahum and W G Pitchford
INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 48 (2003) 2929 2970 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(03)65222-X The IPEM code of practice for electron dosimetry for radiotherapy beams of
More informationCharacteristics and Production of Gas Sealed, Radiation Hard, Small Ionization Chambers
Characteristics and Production of Gas Sealed, Radiation Hard, Small Ionization Chambers Mayda Velasco 1 Gianni Tassotto 3 on behalf of the SIC team P. Ball 2, A.Dabrowsky 1, G. Graham 2, C. Kendziora 3,
More informationClinical Implementation of Volumetric Modulated Arc Therapy
Clinical Implementation of Volumetric Modulated Arc Therapy UT M.D. Anderson Cancer Center Ramaswamy Sadagopan, Rebecca M. Howell, Weiliang Du and Peter Balter Definition 2 Intensity Modulated Arc therapy
More informationCommissioning of a new total body irradiation protocol
University of Wollongong Research Online University of Wollongong Thesis Collection University of Wollongong Thesis Collections 2012 Commissioning of a new total body irradiation protocol Zoe Baldwin University
More informationUser Manual. RW3 Slab Phantom T29672 and T
User Manual RW3 Slab Phantom Contents Operating Manual Technical Manual Service Manual D188.131.00/07 en 2015-07 Hn General Information General Information The product bears the CE-mark "CE" in accordance
More informationMicro Motion Pressure Drop Testing
12/2018 Micro Motion Pressure Drop Testing www.emerson.com/micromotion Introduction Micro Motion has traditionally taken a very conservative approach to pressure drop, with single pressure measurements
More informationExperimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media
Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media N Tyagi 1, B H Curran, P L Roberson, J M Moran, E Acosta and B A Fraass Department
More informationThe Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007
The Effect of Immobilisation Devices on Radiotherapy Dose Distributions A thesis submitted in fulfilment of the requirements for the Degree of Master of Applied Science (Medical and Health Physics) By
More informationVacuum Simulations of the KATRIN Experiment
Vacuum Simulations of the KATRIN Experiment Marcel Krause (KATRIN collaboration) 31.03.2014 KIT University of the State of Baden-Wuerttemberg and National Research Center of the Helmholtz Association www.kit.edu
More informationThe Gas Attenuator of FLASH
INTRODUCTION The Gas Attenuator of FLASH K. Tiedtke, N. von Bargen, M. Hesse, U. Jastrow, U. Hahn The experimental hall of the FLASH user facility is located approximately 30m behind the last dipole magnet
More informationTrial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water
Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water M R McEwen, A J Williams and A R DuSautoy Centre for Ionising Radiation Metrology National Physical
More informationCLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions
CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions Rebecca M. Howell, Ph.D., D.A.B.R, The University of Texas at M.D. Anderson Cancer Center Overview 2 Commissioning
More informationEFFECTS OF SIDEWALL OPENINGS ON THE WIND LOADS ON PIPE-FRAMED GREENHOUSES
The Seventh Asia-Pacific Conference on Wind Engineering, November 8-12, 29, Taipei, Taiwan EFFECTS OF SIDEWALL OPENINGS ON THE WIND LOADS ON PIPE-FRAMED GREENHOUSES Yasushi Uematsu 1, Koichi Nakahara 2,
More informationQuestions OSCILLATIONS AND WAVES
Questions 1 (IB) a) A pendulum consists of a bob suspended by a light inextensible string from a rigid support. The pendulum bob is moved to one side and then released. The sketch graph shows how the displacement
More informationWave phenomena in a ripple tank
Wave phenomena in a ripple tank LEP Related topics Generation of surface waves, propagation of surface waves, reflection of waves, refraction of waves, Doppler Effect. Principle Water waves are generated
More informationSPIRIT III Radiometer Saturation Effect
Utah State University DigitalCommons@USU Space Dynamics Lab Publications Space Dynamics Lab 1-1-1997 SPIRIT III Radiometer Saturation Effect Joseph J. Tansock Follow this and additional works at: https://digitalcommons.usu.edu/sdl_pubs
More informationPaper 2.2. Operation of Ultrasonic Flow Meters at Conditions Different Than Their Calibration
Paper 2.2 Operation of Ultrasonic Flow Meters at Conditions Different Than Their Calibration Mr William Freund, Daniel Measurement and Control Mr Klaus Zanker, Daniel Measurement and Control Mr Dale Goodson,
More informationOn the Use of Superheated Bubble Detectors on Space Missions
On the Use of Superheated Bubble Detectors on Space Missions Eric Benton 1, Rachid Machrafi 2, Leena Tomi 3, Hisashi Kitamura 4 and Satoshi Kodaira 4 1 Oklahoma State University, Dept. of Physics, Stillwater,
More informationIAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy
IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy Guidelines on standardized procedures at Secondary Standards Dosimetry Laboratories (SSDLs) and hospitals March 2002 The
More informationTotal Ionization Dose (TID) Test Results of the RH117H Positive Adjustable Low Dose Rate (LDR) LDR = 10 mrads(si)/s
Total Ionization Dose (TID) Test Results of the RH117H Positive Adjustable Regulator @ Low Dose Rate (LDR) LDR = 10 mrads(si)/s 18 December 2014 Duc Nguyen, Sana Rezgui Acknowledgements The authors would
More informationComparison of ionization chambers of various volumes for IMRT absolute dose verification
Comparison of ionization chambers of various volumes for IMRT absolute dose verification Leonid B. Leybovich, a) Anil Sethi, and Nesrin Dogan Department of Radiation Oncology, Loyola University Medical
More informationThe Discussion of this exercise covers the following points:
Exercise 3-2 Orifice Plates EXERCISE OBJECTIVE In this exercise, you will study how differential pressure flowmeters operate. You will describe the relationship between the flow rate and the pressure drop
More informationIranian Journal of Medical Physics
Iranian Journal of Medical Physics ijmp.mums.ac.ir Verification of Monitor Unit Calculations for Eclipse Treatment Planning System by in-house Developed Spreadsheet Athiyaman Mayivaganan 1*, Hemalatha
More information