A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation

Size: px
Start display at page:

Download "A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation"

Transcription

1 Radiotherapy and Oncology 71 (2004) Technical Note A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation Nadine Linthout*, Dirk Verellen, Swana Van Acker, Guy Storme Department of Radiotherapy, Oncology Center, Academic Hospital Free University Brussels (AZ-VUB), Laarbeeklaan 101, B-1090 Brussels, Belgium Received 31 October 2002; received in revised form 4 November 2003; accepted 26 February Abstract A spreadsheet based program is presented to perform an independent Monitor Unit (MU) calculation verification for the Quality Assurance (QA) of Intensity Modulated Radiation Therapy (IMRT) using Dynamic MultiLeaf Collimation (DMLC). The computed dose value is compared to the planned dose by calculating the percent dose difference per Intensity Modulated Beam (IMB) and absolute dose difference per IMB. The proposed acceptability levels are ^5.0% or ^2.0 cgy for the percent dose difference per IMB and the absolute dose difference per IMB, respectively. For percent dose difference per treatment, an acceptability level of ^2.0% is proposed. The presented program is considered adequate for checking the treatment plans calculated for IMRT treatments using DMLC as a part of the QA procedure. q 2004 Elsevier Ireland Ltd. All rights reserved. Keywords: Monitor unit calculation; Dynamic multileaf collimation; Beam intensity modulation; Quality assurance 1. Introduction Intensity Modulated Radiation Therapy (IMRT) has proven advantages over Conformal Radiation Therapy (CRT) for several clinical cases [1,5,7,8,13,19,21]. The application of IMRT compared to CRT, could improve the target dose uniformity and reduce the dose delivered to the radiosensitive organs close to the target [4,20]. The different possible techniques that have been developed by now to achieve IMRT are grouped together and described by Webb [23]. One of those techniques has received recently increased attention in development and clinical implementation. The technique is Dynamic Multi- Leaf Collimation (DMLC) and uses a MultiLeaf Collimator (MLC) of which the leaves move continuously across the treatment field during irradiation to create the Intensity Modulated Beam (IMB). It is recognized that this technique requires more stringent Quality Assurance (QA) since it is a much more complicated form of treatment compared to conventional treatments. Not only the MLC but also the Treatment Planning System (TPS) must meet more stringent requirements for the delivery of IMRT treatments. If the calculated IMB is accurately converted by the leaf sequencing algorithm incorporated in the TPS into a * Corresponding author. collimation scheme deliverable by the MLC, this collimation scheme will result in the correct intensity modulated treatment field for both geometry and absolute dose. The absolute dose given by an IMB using DMLC is influenced by an important factor being the amount of Monitor Units (MU) that is used to deliver the IMB. The MU verification can, amongst others, be used to check the accuracy of a treatment planning system s dose calculation algorithm for a single point in a target volume. The most accurate yet cumbersome approach is absolute dose measurement. Because an independent MU calculation can in some cases be used as an alternative for these absolute dose measurements, it is considered an important asset in the QA procedure as it can reduce the workload. In CRT, the amount of MU that a treatment field requires can be estimated relatively easy, but in IMRT using DMLC it is very hard to intuitively estimate or evaluate the MUs needed to deliver the IMB. Some previous studies have performed MU calculation verification for DMLC IMRT [3,11] based on programs that require programming skills and specific software to be able to implement the described algorithm. Although meanwhile there is already software commercially available to perform an independent MU calculation verification in IMRT treatments, this work presents an /$ - see front matter q 2004 Elsevier Ireland Ltd. All rights reserved. doi: /j.radonc

2 236 N. Linthout et al. / Radiotherapy and Oncology 71 (2004) in-house developed program using basic formulae and spreadsheet based software to perform a fast and relatively accurate MU calculation verification for IMRT using DMLC. The presented approach to verify the MU calculation is simple and easy to implement in a department. MLC transmission and scattering are taken into account by the program as well as radiation field offset of the segments of the IMB. The original measured beam data are used together with a minimum of output parameters of the treatment plan to recalculate the dose in the normalization point with a calculation algorithm independent from the one used by the TPS. 2. Methods and materials The system that is used in our department to deliver IMRT treatments is the Novalis w system: a single energy (6 MV-photon) linear accelerator (linac) with an integrated mini-multileaf Collimator (mmlc), dedicated for stereotactic radiosurgery. The system offers next to conformal beam and dynamic field shaping rotation therapy [12] also IMRT capable of using Static MultiLeaf Collimation (SMLC) as well as DMLC. The mmlc M3 w is described in detail by Cosgrove et al. and Xia et al. [6,24]. The linac is capable of generating a variable dose rate up to 800 cgy per minute (calibrated to deliver 1 cgy/mu at depth of maximum dose in fixed SSD configuration). The latter aspect, combined with the zero delay time before the beam is on, makes the machine very suitable to perform IMRT. BrainSCAN w is the Treatment Planning System (TPS) that is combined with the Novalis w system. The software (BrainSCAN w v5.1: BrainLAB AG, Heimstetten, Germany) offers an inverse planning modality for IMRT [22]. In the inverse planning, four plans are optimized with different organ at risk priorities using a Dynamically Penalized Likelihood algorithm [15]. During the optimization the transmission calculation is handled by integrating a quick leaf sequencing from the start of the optimization. The leaf sequencing is based on the theory of Bortfeld et al. [2]. When the inverse planning is finished, the actual leaf sequencing is performed in three parts: the calculation of the leaf movements to deliver the fluence (i.e. the number of segments and the step size), the tongue-and-groove optimization and the transmission calculation. The produced fluence maps are used in the forward calculation of the dose distribution based on the Pencil Beam dose calculation algorithm developed by Mohan et al. [16 18]. One of the four calculated plans can be chosen or the calculation parameters and planning constraints can be changed to achieve the required dose distribution. When one of the four plans is selected for which the MU calculations will be verified, the reference point (normalization point) was defined in accordance with the International Commission on Radiation Units and Measurements (ICRU) 50 report [9] recommendations. In the cases that are verified in this study the normalization point was placed at the isocenter. For each single treatment beam, the relative and absolute dose delivered in the normalization point is calculated by the TPS using the corresponding amount of MU and the calculated segments of the IMB. To verify this MU calculation, a simple spreadsheetbased program is developed. A minimum number of parameters are used from the plan to recalculate the dose predicted by the TPS in the normalization point for each IMB, hereafter called D IMB (TPS). Each segment of a DMLC IMB, defined by the leaf-sequencing algorithm of the TPS, is considered as a uniform intensity beam to calculate the dose delivered by that IMB in the normalization point. This computed dose value will be referred to as D IMB (verif). The variation of the tissue contour of the patient within the treatment fields is neglected in the program and the patient is assumed to be homogeneous and water equivalent. The three parameters that need to be imported in the program from the treatment plan are the MUs for the IMB, the beam shape file of the IMB, and the depth to the normalization point. The MUs for each IMB are copied from the printout of the treatment parameters. The beam shape file includes the shape (determined by the positions of all mmlc leaves) and the relative weight of all segments of that IMB calculated by the TPS. The beam shape file can be exported from the TPS for each IMB used in the treatment plan. The depth to the normalization point for the IMB is defined manually on the TPS because it is not included in the output parameters of the treatment plan. The equivalent depth defined by the TPS is not used in the spreadsheet program because the patient is assumed to be water equivalent. The parameters used in the program that are not imported from the treatment plan are the Tissue Maximum Ratio (TMR) data, the total scatter factors (S c,p ) and the Off Axis Ratios (OAR). These parameters are taken from the original measured beam data. The equivalent square field that corresponds with each segment is calculated with the Sterling formula using the contour of each segment defined by the leaf settings of that segment to calculate the area and the perimeter of that segment. The equivalent square field size is used together with the depth to the normalization point to find the correct TMR value for each segment of the IMB. The corresponding S c,p is determined for the equivalent square field size formed by the leaves of the mmlc. Since all the IMBs are delivered with the jaws positioned at 4.90 cm, the interpolation for the S c,p is only performed between different fields sizes formed by the leaves and for a fixed field size of cm 2 formed by the jaws. To determine the OAR corresponding with every segment, the coordinates of the

3 N. Linthout et al. / Radiotherapy and Oncology 71 (2004) gravitation point of every segment are calculated using Eqs (1) and (2): x g ¼ y g ¼ xydx ydx 1 2 y 2 dx ydx Based on these coordinates, the radial distance between the gravitation point of the segment and the normalization point is determined. This diagonal radius and the depth to the normalization point allow one to define the OAR that corresponds with every segment based on the original measured beam data. The total amount of MUs of the IMB is distributed over the segments of the IMB. Therefore, the relative weight of every segment ðw i Þ is calculated by subtracting the indices from the successive segments in the exported beam shape file. There is no threshold value defined to eliminate segments with a low relative weight, every segment of the IMB is taken in consideration. The contribution of each segment (con i ) to the dose delivered in the normalization point is defined by determination of the degree to which the segment covers the normalization point. If the normalization point is located within and at least 0.50 cm from the borders of the segment, the contribution of the dose delivered by that segment to the normalization point dose is considered maximal (con i ¼ 1). For the segments where the normalization point is blocked by the mmlc leaves by at least 0.50 cm in both the inplane and crossplane direction, only the mmlc transmission dose to the normalization point dose is taken into account (con i ¼ 0.02). When the normalization point is positioned between the two previous margins (^0.50 cm), the contribution of that segment to the normalization point dose is calculated by linear interpolation depending on the position of the normalization point with respect to the mmlc leaves. This interpolation interval covers the penumbra as well as the shoulder and the heel of the field border. By multiplying the relative weight and the contribution with the total amount of MUs of the IMB, the MUs of the IMB are distributed over all segments: MU i ¼ MU IMB ðtpsþ w i con i ð3þ with MU IMB (TPS) representing the total amount of MUs calculated by the TPS for that IMB, MU i the MUs for segment i that contribute to the dose delivered in the normalization point, w i the relative weight of segment i and con i the contribution of the dose delivered by segment i to the normalization point dose. ð1þ ð2þ The dose that every segment delivers to the normalization point is calculated with the formula: D i ¼ MU i TMR i S c;pi OAR i ð4þ CF SAD where D i is the dose delivered by segment i; MU i, the MUs for segment i; TMR i, the tissue maximum ratio for segment i corresponding with the depth and the field size of segment i; S c,pi the total scatter factor for segment i corresponding with the field size of segment i; OAR i the off axis ratio for segment i based on the position of the gravitation point of segment i and CF SAD the conversion factor for isocentric treatments since the linac is calibrated to deliver 1 cgy/mu at depth of maximum dose with a fixed SSD geometry that is calculated with Eq. (5). CF SAD ¼ððSSD norm þ d norm Þ=SIDÞ 2 with SSD norm representing the source surface distance used for dose monitor calibration of the linac performed at a defined depth d norm and SID the source isocenter distance. The dose delivered by that IMB in the normalization point is calculated by summing the doses of all segments of the IMB together: D IMB ðverifþ ¼ X D i ð6þ i The calculated doses D IMB (verif) are compared with the predicted doses by the BrainSCAN software D IMB (TPS). The relative difference is reported as percent dose differences per IMB and is computed using: %diff ¼ ðd IMBðverifÞ 2 D IMB ðtpsþþ 100 ð7þ D IMB ðtpsþ The absolute difference between the two doses per IMB is also defined: diffðcgyþ ¼D IMB ðtpsþ 2 D IMB ðverifþ 3. Results In order to test the program, 166 IMBs have been generated by the BrainSCAN w TPS and have been verified with the spreadsheet program. The IMRT plans that are used for this study included 11 prostate cases, 9 brain and 6 head and neck cases and three imaginary targets defined in a homogeneous phantom to cover different degrees of treatment complexities. The average number of IMBs of the 29 verified treatment plans was 5.5 with a standard deviation (SD) of 1.1 (range 4 9). The delivered dose per treatment had a mean value of cgy with a SD of 73.8 cgy (range cgy). The average amount of ð5þ ð8þ

4 238 N. Linthout et al. / Radiotherapy and Oncology 71 (2004) Fig. 1. The frequency of the discrepancies between the planned doses and the dose values computed with the spreadsheet-based program expressed in percent dose differences per DMLC IMB. The curve is the normal distribution curve fitted to the resulting frequencies. MUs per treatment was 825 MU with a SD of 376 MU (range MU). In Fig. 1, the frequency is displayed of the percent dose differences per IMB calculated with Eq. (7). The mean percent difference per IMB was 21.1% with a SD of 6.5% (range þ 20.7%). The percent differences between the total dose calculated by the TPS in the normalization point per treatment and the total dose value computed by the spreadsheet program for the same treatment are also calculated. The mean percent dose difference per treatment was 20.6% with a SD of 2.9% (range 25.2 þ 5.6%). The frequency of the absolute dose differences per IMB calculated with Eq. (8) is shown in Fig. 2. The mean absolute dose difference was 20.2 cgy with a SD of 2.0 cgy (range þ 7.5 cgy). 4. Discussion and conclusion All used IMBs are typically composed of 29 segments, corresponding with the root-mean-square (RMS) error that is set to 1.0 in the treatment plan optimization parameters. All segments are imported at the same time in the program when the beam shape file is imported. Therefore, the number of segments of the IMB does not influence the workload of using the program. The spread of the percent dose differences per IMB was rather wide due to a not completely optimized program to calculate the dose values. Some of the imperfections of the program can be found in: (a) the linear approximation of the shape of the field border of the treatment beam, (b) the scatter in the patient that is not taken into account because a homogeneous patient is assumed, (c) no changes in the outer contour of the patient are taken into account, (d) borderline situations where the normalization point is blocked by the majority of the segments of the IMB, may not be fully covered. (a) The change of the penumbra with the field size and the position of the field border, which is less than 0.02 cm for the Novalis w system, is not taken into account with the linear approximation of the shape of the field border. The interpolation is always performed between a contribution of 1 and 0.02 for a leaf position ranging from 0.50 cm (open field) to cm (overtravel). The relative leaf position of ^ 0.50 cm with respect to the position of the normalization point is chosen as interpolation margin because the penumbra (lateral distance between the 80% and the 20% isodose line) of the largest possible treatment field of the Novalis w system ( cm 2 ) is 0.43 cm while the smallest treatment field ( cm 2 ) has a penumbra of 0.41 cm. The penumbra as well as the shoulder and the heel of the field border are covered in the linear interpolation range. The calculated dose to the normalization point can be underestimated due to the linear approximation of the field border compared to the planned dose if the normalization point is located between the 50 and 100% isodose line of the treatment field. Similar, if the normalization point is located between the 0.02 and the 50% isodose line of the treatment field the dose to the normalization point is overestimated by the program.

5 N. Linthout et al. / Radiotherapy and Oncology 71 (2004) Fig. 2. The frequency of the discrepancies between the planned doses and the dose values computed with the spreadsheet-based program expressed in absolute dose differences per DMLC IMB. The curve is the normal distribution curve fitted to the resulting frequencies. (b) Deviations between planned and computed dose values can be introduced due to the assumed homogeneous patient when large changes in electron density appeared in the area traversed by the IMB. While air cavities will cause the computed dose value to underestimate the actual or planned dose [14], bony structures will introduce the opposite effect. If a large negative difference is observed during the verification of the dose and a large air cavity is traversed by the treatment beam before the normalization point is reached, one can recalculate the dose value with the equivalent depth to evaluate the observed difference. Similar, a large positive difference could be observed when a lot of bony structures appear in the pathway of the treatment beam before the normalization point, if the equivalent depth is used for the calculation of the dose value instead of the manually measured depth. An observed difference for the verification of the calculated dose of an IMB can of course also be evaluated based on the absolute dose measurement of that IMB. (c) The changes in the outer contour of the patient within the geometrical limits of the IMB can cause the actual distance between the central entrance point of a segment and the normalization point to differ from the measured depth to the normalization point. The TMR, S c,p and OAR values for that segment will be altered by the difference in depth. (d) Borderline situations are found in IMBs where very little segments contribute to the dose in the normalization point. For instance, in the IMB for which the percent dose difference amounts to 224.8% the normalization point is covered partially by only 2 out of 29 segments. Large percent dose differences can be introduced by this fact in combination with the linear approximation of the penumbra shape. The spreadsheet program is not optimized to account for these borderline situations in order to keep the program simple because it is only used for double-checking the calculated plan. If the percent dose difference of an IMB is extremely high or low, a closer look has to be taken at the separate segments of the IMB to verify if their shape can cause the large discrepancy between the doses. If the latter is not the case, a deviation might be found in one of the other treatment parameters in either the value used by the TPS or the value imported in the spreadsheet program. Due to these limitations of the presented program, an accuracy of ^ 5.0% is found acceptable for the calculated percent dose difference per IMB. At the moment there is no general protocol established that defines the required accuracy for the 3-dimensional computer planning of IMRT treatments. This protocol would refer to the agreement between the calculated and the measured dose in a point under the same irradiation conditions. Therefore an accuracy of ^5.0% is proposed as accuracy level of the for the percent dose difference per IMB. Because the spread of the percent dose differences per IMB was considered wide, the percent dose difference per treatment was also calculated to allow an evaluation of the entire treatment plan. These results had a much smaller spread. In the ICRU 42 report [10] a relative accuracy of 2.0% is prescribed for the dose calculated by the TPS. In this report is not referred to IMRT treatments but the same accuracy level can be used as an acceptability level for the percent dose difference per treatment calculated with the presented program since it refers to a dose difference

6 240 N. Linthout et al. / Radiotherapy and Oncology 71 (2004) resulting from an entire treatment plan. This acceptability level of ^ 2.0% should provide a reasonable guarantee that the treatment preparation of the IMRT treatment is performed within the uncertainties of all involved parameters. Nevertheless, the verification of the MU calculation of the IMBs separately is necessary to give supplementary information about the treatment compared to the verification of the MU calculation for the entire treatment plan because the latter can obscure large deviations in the verification of the IMBs separately that need to be looked at in detail. During commissioning of IMRT with the Novalis w system, the agreement between the dose to a point calculated by the TPS and the absolute measured dose in that point was found within 2.0% in a waterphantom, as recommended by the ICRU42 report [10]. During the pretreatment verification of a patient treatment the delivered dose of the entire treatment plan is verified absolutely and the fluence patterns of all IMBs of the treatment plan are verified. The results of these verifications are beyond the scope of this study and not reported here. The agreement between D IMB (verif) and the actual dose delivered by a single IMB was not verified for all IMBs in this study, but for all IMBs the computed dose values and planned doses are compared absolutely. The comparison of the absolute doses has given another, sometimes more relevant, appreciation of the differences between the doses. Especially when the planned dose for a certain IMB is very small, e.g. 5.0 cgy, a deviation of 0.4 cgy (mean absolute dose difference) results in a percent dose difference of 8.0%, what exceeds the proposed accuracy level of ^ 5.0%. For the absolute dose verification an accuracy of ^ 2.0 cgy is proposed as acceptability level, that takes the uncertainties of the involved parameters into account. If the example of the IMB with a percent dose difference of % is reviewed, the absolute dose difference for this IMB amounts to 21.5 cgy what is within the accuracy of the absolute dose verification. When the percent dose difference per IMB exceeds the acceptability level but the absolute dose difference per IMB does not, the comparison of the absolute doses eliminates a further evaluation of the IMB segments. Therefore the absolute dose verification is considered a complementary tool in the MU calculation verification program. For IMBs, where neither of the acceptability criteria were fulfilled (what happened in about 10% of the verified IMBs) a further investigation of the dose delivered by that IMB was required. Performing an absolute dose measurement for these IMBs covered the latter. The discrepancies in the outcome of the program have proved that a MU calculation verification can certainly not be used as the only verification that is performed for IMRT treatments. A program like the presented one could be used to reduce the workload of IMRT treatment verification because it is not longer required to verify all IMBs by absolute dose measurement. The proposed accuracy levels for both percent dose and absolute dose differences per IMB are found stringent enough to indicate important errors in the treatment plan. For instance, if another beam data file is used for the calculations of the IMRT plan, all the IMBs used in that plan will result in a percent dose and an absolute dose difference per IMB that exceeds the proposed accuracy levels. Any possible uncertainty embedded by the recalculation of the measured TMR data to PDD data that the TPS uses for the treatment plan calculations is avoided by using the original TMR data for the calculation verification. Since all treatments performed with the Novalis w system are isocentric treatments, it is found reasonable to compute the dose values based on the originally measured TMR data. The BrainSCAN w software uses also a MU calibration factor, the Nominal Linac Output factor (NOF), that defines the dose delivered at calibration depth using the calibration setup for 100 MUs. This factor does not need to be included in the calculation program because the linac is calibrated to deliver 1 cgy/mu at a depth of maximum dose in a fixed SSD configuration, the original measured TMR beam data are normalized to 100% at depth of maximum dose and the conversion factor CF SAD is used to take the fact that the linac is not calibrated in SAD geometry into consideration. In this way, a minimum number of parameters are used from the MU calculation performed by the TPS making the presented program an independent MU calculation verification that can be implemented in the QA procedure for IMRT treatments using DMLC. The achieved accuracy of the presented program has shown that it is not always necessary to have the programming skills to write a computer program or to buy software that performs the MU calculation verification in the QA procedure. A more simple approach can, notwithstanding some loss of accuracy, be sufficient to result in an adequate QA for IMRT treatments. References [1] Benedict SH, Cardinale RM, Wu Q, Zwicker RD, Broaddus WC, Mohan R. Intensity-modulated stereotactic radiosurgery using dynamic micro-multileaf collimation. Int J Radiat Oncol Biol Phys 2001;50(3): [2] Bortfeld TR, Kahler DL, Waldron TJ, Boyer AL. X-ray field compensation with multileaf collimators. Int J Radiat Oncol Biol Phys 1994;28(3): [3] Boyer A, Xing L, Ma CM, et al. Theoretical considerations of monitor unit calculations for intensity modulated beam treatment planning. Med Phys 1999;26(2): [4] Brahme A. Optimization of stationary and moving beam radiation therapy techniques. Radiother Oncol 1988;12: [5] Chang SX, Deschesne KM, Cullip TL, Parker SA, Earnhart J. A comparison of different intensity modulation treatment techniques for tangential breast irradiation. Int J Radiat Oncol Biol Phys 1999;45(5): [6] Cosgrove VP, Jahn U, Phaender M, Bauer S, Budach V, Wurm RE. Commissioning of a micro multi-leaf collimator and planning system for stereotactic radiosurgery. Radiother Oncol 1999;50: [7] De Neve W, Claus F, Van Houtte P, Derycke S, De Wagter C.

7 N. Linthout et al. / Radiotherapy and Oncology 71 (2004) Intensity modulated radiotherapy with dynamic multileaf collimator. Technique and clinical experience. Cancer Radiother 1999;3(5): [8] Hunt MA, Zelefsky MJ, Wolden S, et al. Treatment planning and delivery of intensity-modulated radiation therapy for primary nasopharynx cancer. Int J Radiat Oncol Biol Phys 2001;49(3): [9] International Commission on Radiation Units and Measurements (ICRU), Prescribing, Recording and Reporting Photon Beam Therapy. ICRU Report 50, Baltimore, MD: ICRU; [10] International Commission on Radiation Units and Measurements (ICRU), Use of computers in external beam radiotherapy procedures with high-energy photons and electrons. ICRU Report 42, Baltimore, MD: ICRU; [11] Kung JH, Chen GTY, Kuchnir JK. A monitor unit verification calculation in intensity modulated radiotherapy as a dosimetry quality assurance. Med Phys 2000;27(10): [12] Leavitt DD, Gibbs FA, Heilbrun MP, Moeller JH, Takack GA. Dynamic field shaping to optimize stereotactic radiosurgery. Int J Radiat Oncol Biol Phys 1991;21: [13] Linthout N, Verellen D, De Coninck P, Bel A, Storme G. Threedimensional conformal radiation therapy: the tomotherapy approach. Cancer Radiother 2000;4(6): [14] Linthout N, Verellen D, Van Acker S, Voordeckers M, Bretz A, Storme G. Evaluation of dose calculation algorithms for dynamic arc treatments of head and neck treatments. Radiother Oncol 2002;64(1): [15] Llacer J. Inverse radiation treatment planning using the dynamically penalized likelihood method. Med Phys 1997;24(11): [16] Mohan R, Chui C. Use of Fast Fourier Transforms in calculating dose distributions for irregularly shaped fields for three-dimensional treatment planning. Med Phys 1987;14:70 7. [17] Mohan R, Chui C, Lidofsky L. Differential pencil beam dose computation model for photons. Med Phys 1986;13(1): [18] Mohan R, Chui C, Lidofsky L. Energy and angular distributions of photons from medical linear accelerators. Med Phys 1985;12: [19] Nutting CM, Bedford JL, Cosgrove VP, Tait DM, Dearnaley DP, Webb S. A comparison of conformal and intensity-modulated techniques for oesophageal radiotherapy. Radiother Oncol 2001;61: [20] Nutting CM, Dearnaley DP, Webb S. Intensity modulated radiotherapy: a clinical review. Br J Radiol 2000;54: [21] Nutting CM, Rowbottom CG, Cosgrove VP, et al. Optimization of radiotherapy for carcinoma of the parotid gland: a comparison of conventional, three-dimensional conformal, and intensity-modulated techniques. Radiother Oncol 2001;60: [22] Verellen D, Linthout N, Soete G, Van Acker S, De Roover P, Storme G. Considerations on treatment efficiency of different conformal radiation therapy techniques. Radiother Oncol 2001;63(1): [23] Webb S. Rotation IMRT: tomotherapy; IMRT using a multileaf collimator. In: Orton CG, Spaan JAE, Webster JG, editors. Intensitymodulated radiation therapy. Philadelphia, PA: Institute of Physics Publishing; p [24] Xia P, Geis P, Xing L, et al. Physical characteristics of a miniature multileaf collimator. Med Phys 1999;26(1):65 70.

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 47 CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 4.1 INTRODUCTION Advanced treatment techniques use optimized radiation beam intensities to conform dose distribution to the

More information

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN 22 Turkish Journal of Cancer Volume 37, No.1, 2007 Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN SATISH PELAGADE 1, KALPANA

More information

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Commissioning an IMRT System for MLC Delivery Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Taking the broad view of commissioning Commissioning elements Validating the dosimetry system Commissioning the

More information

Comparison of ionization chambers of various volumes for IMRT absolute dose verification

Comparison of ionization chambers of various volumes for IMRT absolute dose verification Comparison of ionization chambers of various volumes for IMRT absolute dose verification Leonid B. Leybovich, a) Anil Sethi, and Nesrin Dogan Department of Radiation Oncology, Loyola University Medical

More information

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription Monitor Unit Calculations Part 1 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Return to our first patient 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt

More information

Relative Dosimetry. Photons

Relative Dosimetry. Photons Relative Dosimetry Photons What you need to measure! Required Data (Photon) Central Axis Percent Depth Dose Tissue Maximum Ratio Scatter Maximum Ratio Output Factors S c & S cp! S p Beam profiles Wedge

More information

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator Journal of BUON 8: 55-59, 2003 2003 Zerbinis Medical Publications. Printed in Greece ORIGINAL ARTICLE Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

More information

Dosimetric Calculations. Lonny Trestrail

Dosimetric Calculations. Lonny Trestrail Dosimetric Calculations Lonny Trestrail 20 October 2008 Objectives Dose Distribution Measurements PDD, OCR TAR, SAR, TPR, TMR, SPR, SMR Arc or Rotational Therapy Isodose Curves Point Dose Calculations

More information

Clinical Implementation of Volumetric Modulated Arc Therapy

Clinical Implementation of Volumetric Modulated Arc Therapy Clinical Implementation of Volumetric Modulated Arc Therapy UT M.D. Anderson Cancer Center Ramaswamy Sadagopan, Rebecca M. Howell, Weiliang Du and Peter Balter Definition 2 Intensity Modulated Arc therapy

More information

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions Rebecca M. Howell, Ph.D., D.A.B.R, The University of Texas at M.D. Anderson Cancer Center Overview 2 Commissioning

More information

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 DISCLAIMER: This publication and associated spreadsheets and digital files are based on sources and

More information

Treatment plan complexity metrics for predicting IMRT pretreatment. assurance results. Scott Crowe, QUT

Treatment plan complexity metrics for predicting IMRT pretreatment. assurance results. Scott Crowe, QUT Treatment plan complexity metrics for predicting IMRT pretreatment quality assurance results Scott Crowe, QUT Pre-treatment QA Verification of fluence delivery to ensure that dose can be delivered accurately

More information

High Speed Direct SAD Radiosurgery Beam Scanner

High Speed Direct SAD Radiosurgery Beam Scanner Open Access Original Article DOI: 10.7759/cureus.20 High Speed Direct SAD Radiosurgery Beam Scanner Walter Nikesch 1, James M. Hevezi 2, Irene Monterroso 3, Daniel Navarro 4, James G. Schwade 5 1. CyberKnife

More information

SnapShot IMRT with compensators and FFF beams

SnapShot IMRT with compensators and FFF beams SnapShot IMRT with compensators and FFF beams Vladimir Feygelman, PhD (1) Moffitt Cancer Center, Tampa, FL, USA Disclosure VF has a sponsored research agreement with.decimal. A short history of radiotherapy

More information

Math Review. Overview

Math Review. Overview Drawn by Steve Yan, CMD Math Review Shirley Ann Pinegar-Johnston MS RT(R)(T)CMD Overview Drawn by Steve Yan, CMD Divergence & Intensity DIVERGENCE X-rays travel in Straight but Divergent lines INTENSITY

More information

THE development of more advanced techniques in radiotherapy,

THE development of more advanced techniques in radiotherapy, 1 A Geant4-based simulation of an accelerator s head used for Intensity Modulated Radiation Therapy F. Foppiano, B. Mascialino, M.G. Pia, M. Piergentili. Abstract We present a Geant4-based simulation,

More information

Review of fundamental photon dosimetry quantities

Review of fundamental photon dosimetry quantities Review of fundamental photon dosimetry quantities Narayan Sahoo Main sources of the materials included in this lecture notes are: (1) Radiation Oncology Physics: A Handbook for Teachers and Students Edited

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 Comparison of monitor unit calculations performed with a 3D computerized planning system and independent hand calculations: Results

More information

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface Chapter 11 reatment Planning Single Beams Radiation Dosimetry I Outline Basic terminology Curved contour surface correction (bolus, compensators, wedges) Oblique beam incidence Correction for tissue inhomogeneities

More information

Monitor Unit Verification for Small Fields

Monitor Unit Verification for Small Fields Monitor Unit Verification for Small Fields Patrick Higgins, Ph.D University of Minnesota Department of Radiation Oncology October 10, 2013 The Issues: How do we verify the monitor units calculated by the

More information

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting Monitor Unit Calculations Part 2 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Calculation of machine setting reference dose machine setting =, reference dose

More information

1. Question Answer cgy / MU cgy / MU 2. Question Answer

1. Question Answer cgy / MU cgy / MU 2. Question Answer GS020113: Introduction to Medical Physics III: Therapy s to home work problem set assigned on 3/22/11 1. Question A patient is set up at 100 cm SSD on a 6 MVX machine. The dose rate at 10 cm in phantom

More information

The MC commissioning of CyberKnife with MLC (Tips and Tricks)

The MC commissioning of CyberKnife with MLC (Tips and Tricks) The MC commissioning of CyberKnife with MLC (Tips and Tricks) Alain Guemnie Tafo, PhD UPMC Pinnacle, Harrisburg, PA Institute Gustave Roussy, Villejuif, France Disclaimer & Disclosure The views expressed

More information

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL FOR THE DEGREE MASTER OF ARTS BY CHARLES TRAVIS WEBB

More information

A beam-matching concept for medical linear accelerators

A beam-matching concept for medical linear accelerators Acta Oncologica, 2009; 48: 192200 ORIGINAL ARTICLE A beam-matching concept for medical linear accelerators DAVID SJÖSTRÖM, ULF BJELKENGREN, WIVIANN OTTOSSON & CLAUS F. BEHRENS Copenhagen University Hospital,

More information

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system G. Kunz, F. Hasenbalg, P. Pemler 1 1 Klinik für Radio-Onkologie und Nuklearmedizin, Stadtspital Triemli

More information

A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions

A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions Wendel Dean Renner a) Math Resolutions, LLC, 5975 Gales Lane, Columbia,

More information

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13.1.1 TRAINING GOAL 13.1.1.1 Among the responsabilities of a medical

More information

Surface buildup dose dependence on photon field delivery technique for IMRT

Surface buildup dose dependence on photon field delivery technique for IMRT Surface buildup dose dependence on photon field delivery technique for IMRT Shigeru Yokoyama, Peter L. Roberson, Dale W. Litzenberg, Jean M. Moran, and Benedick A. Fraass Department of Radiation Oncology,

More information

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization Basics of Proton Therapy Proton Treatment Planning and Beam Optimization SAM Educational Session, WE D BRB 2 Mark Pankuch, PhD Northwestern Medicine Chicago Proton Center Today s objectives Review the

More information

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Poster No.: R-0044 Congress: Type: Authors: Keywords: DOI: 2014 CSM Scientific Exhibit L. Bendall, I. Patel, N. McGrath, C. Rowbottom; MANCHESTER/

More information

IMRT in clinical practice at the UMC-Utrecht. Utrecht

IMRT in clinical practice at the UMC-Utrecht. Utrecht 3 IMRT in clinical practice at the UMC-Utrecht Utrecht Clinical use of IMRT Improvement of local control without increased incidence of normal tissue complications + Dose escalation in the prostate Step-and

More information

Constancy checks of well-type ionization chambers with external-beam radiation units

Constancy checks of well-type ionization chambers with external-beam radiation units JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Constancy checks of well-type ionization chambers with external-beam radiation units Sara L. Hackett, 1a Benjamin Davis, 2 Andrew

More information

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas Current Implementation Status Current Implementation Status 1600 1400 TOTAL 1494 of 1623 ACTIVE INSTITUTIONS

More information

Iranian Journal of Medical Physics

Iranian Journal of Medical Physics Iranian Journal of Medical Physics ijmp.mums.ac.ir Verification of Monitor Unit Calculations for Eclipse Treatment Planning System by in-house Developed Spreadsheet Athiyaman Mayivaganan 1*, Hemalatha

More information

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08. Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas 2009 AAPM Summer School Educational Objectives Improve your understanding of how to implement TG-51

More information

Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media

Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media N Tyagi 1, B H Curran, P L Roberson, J M Moran, E Acosta and B A Fraass Department

More information

Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator

Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 2, 2013 Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator James L. Bedford, a Michael D. R. Thomas and Gregory

More information

Implementation of respiratory-gated VMAT on a Versa HD linear accelerator

Implementation of respiratory-gated VMAT on a Versa HD linear accelerator Received: 21 November 2016 Revised: 7 April 2017 Accepted: 5 June 2017 DOI: 10.1002/acm2.12160 RADIATION ONCOLOGY PHYSICS Implementation of respiratory-gated VMAT on a Versa HD linear accelerator Jeffrey

More information

Zoubir Ouhib Lynn Cancer Institute

Zoubir Ouhib Lynn Cancer Institute Zoubir Ouhib Lynn Cancer Institute April 2015 Speaker for ELEKTA Brachytherapy using miniature X-ray sources Energy

More information

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World Chattanooga, Tennessee, April 17-21, 2005, on CD-ROM, American Nuclear Society, LaGrange Park, IL (2005) GEANT4 SIMULATION OF

More information

Methods to model and predict the ViewRay treatment deliveries to aid patient scheduling and treatment planning

Methods to model and predict the ViewRay treatment deliveries to aid patient scheduling and treatment planning Washington University School of Medicine Digital Commons@Becker Open Access Publications 2016 Methods to model and predict the ViewRay treatment deliveries to aid patient scheduling and treatment planning

More information

Scientific Journal of Nuclear Medicine & Radiation Therapy

Scientific Journal of Nuclear Medicine & Radiation Therapy Scientific Journal of Nuclear Medicine & Radiation Therapy Research Article Radiotherapy Dose Calculation Program, for Monitoring and Commissioning Treatment Planning Systems - Ioannis Vamvakas*, Stella

More information

Dynamic wedges dosimetry and quality control

Dynamic wedges dosimetry and quality control Original Paper Received: 2005.06.07 Accepted: 2006.02.28 Published: 2006.04.28 Authors Contribution: A Study Design B Data Collection C Statistical Analysis D Data Interpretation E Manuscript Preparation

More information

Electronic Oncology Systems the Management and the Safety Measures. Hansen Chen, Director, TDSI

Electronic Oncology Systems the Management and the Safety Measures. Hansen Chen, Director, TDSI Electronic Oncology Systems the Management and the Safety Measures Hansen Chen, Director, TDSI Radiation Oncology Department People Place Equipments Procedures Patients Hansen Chen 2 Radiation Oncology

More information

Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System

Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System Amit Sawant 1, Sonja Dieterich 1, Michelle Svatos 2 and Paul Keall 1 1 Stanford University, Stanford, CA 2 Varian Medical Systems, Palo

More information

A clinical comparison and analysis between conventional MLC based and solid compensator based IMRT treatment techniques

A clinical comparison and analysis between conventional MLC based and solid compensator based IMRT treatment techniques The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2009 A clinical comparison and analysis between conventional MLC based and solid compensator based IMRT treatment

More information

TG219: IT'S USE, STRENGTHS AND WEAKNESSES

TG219: IT'S USE, STRENGTHS AND WEAKNESSES 1 CONFLICT OF INTEREST TG219: IT'S USE, STRENGTHS AND WEAKNESSES SOTIRI STATHAKIS, PHD, DABR None ROSTER STATEMENT OF THE PROBLEM Timothy C. Zhu, University of Pennsylvania, Philadelphia, PA (Co- Chair)

More information

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Tom Jordan Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE

More information

Commissioning and periodic tests of the Esteya electronic brachytherapy system

Commissioning and periodic tests of the Esteya electronic brachytherapy system Review paper Educational Activity Commissioning and periodic tests of the Esteya electronic brachytherapy system Cristian Candela-Juan, PhD 1, Yury Niatsetski, MSc 2, Zoubir Ouhib, PhD 3, Facundo Ballester,

More information

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2017, 6, 57-72 http://www.scirp.org/journal/ijmpcero ISSN Online: 2168-5444 ISSN Print: 2168-5436 Validation of Treatment

More information

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics. FRANZCR Examination Part I Radiation Oncology Radiotherapeutic Physics Candidate No.: The Royal Australian and New Zealand College of Radiologists FRANZCR Examination Part I Radiation Oncology Radiotherapeutic

More information

Quality control for Intensity-modulated radiation therapy

Quality control for Intensity-modulated radiation therapy Schweizerische Gesellschaft für Strahlenbiologie und Medizinische Physik Société Suisse de Radiobiologie et de Physique Médicale Società Svizzera di Radiobiologia e di Fisica Medica Swiss Society for Radiobiology

More information

23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN APRIL 2007

23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN APRIL 2007 23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN 16-20 APRIL 2007 AN INVESTIGATION INTO THE INTERRELATION BETWEEN THE INTERNAL AND EXTERNAL BALLISTICS OF FIRING A TP-T TANK AMMUNITION M. H.

More information

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007 The Effect of Immobilisation Devices on Radiotherapy Dose Distributions A thesis submitted in fulfilment of the requirements for the Degree of Master of Applied Science (Medical and Health Physics) By

More information

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11 MV Photon Dosimetry in the Clinic This document discusses both the scientific methods and the system of work required to calibrate a field instrument expanding upon the practical details of both.. Introduction

More information

Radiotherapy physics & Equipments

Radiotherapy physics & Equipments Radiotherapy physics & Equipments RAD 481 Lecture s Title: Beam Modification devices in Radiotherapy Dr. Mohammed Emam Vision :IMC aspires to be a leader in applied medical sciences, health care education

More information

Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy

Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy University of South Florida Scholar Commons Graduate Theses and Dissertations Graduate School 7-14-2010 Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy

More information

COMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy

COMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy COMPASS 3D IMRT Verification in Patient Anatomy AK IMRT Bamberg April 2010 Dr. Lutz Müller Patient-specific Verification? CT Treatment unit Gantry =0 Treatmt unit Patient couch X-ray beam Patient couch

More information

Commissioning of a new total body irradiation protocol

Commissioning of a new total body irradiation protocol University of Wollongong Research Online University of Wollongong Thesis Collection University of Wollongong Thesis Collections 2012 Commissioning of a new total body irradiation protocol Zoe Baldwin University

More information

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams Peter R. Almond Brown Cancer Center, Louisville, Kentucky 40202 Peter J. Biggs Department of Radiation Oncology,

More information

Quality Management Determined from Risk Assessment

Quality Management Determined from Risk Assessment Quality Management Determined from Risk Assessment Bruce Thomadsen University of Wisconsin and The Center for the Assessment of Radiological Sciences Disclosure I am the President of the Center for the

More information

Kill-painting of hypoxic tumors in charged particle therapy

Kill-painting of hypoxic tumors in charged particle therapy Supplementary Information: Kill-painting of hypoxic tumors in charged particle therapy Walter Tinganelli 1,2,, Marco Durante 1,3 *, Ryoichi Hirayama 2, Michael Krämer 1, Andreas Maier 1, Wilma Kraft-Weyrather

More information

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components Introduction The CyberKnife SRS/SBRT System James M. Hevezi,, Ph.D., FACR CyberKnife Unit San Antonio, Texas CyberKnife Components Treatment Delivery (G3 Platform San Antonio) Robot (modified German auto

More information

VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests

VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 3, 2016 VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests Michael P. Barnes, 1,2a Pejman Rowshanfarzad,

More information

WEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS

WEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS Acta Radiologica Oncology 23 (1984) Fusc. 6 FROM THE RADIATION THERAPY DEPARTMENT, ANTON1 VAN LEEUWENHOEK HOSPITAL, THE NETHERLANDS CANCER INSTITUTE, AMSTERDAM, THE PHILIPS MEDICAL SYSTEMS DIVISION, BEST,

More information

Evaluation of six TPS algorithms in computing entrance and exit doses

Evaluation of six TPS algorithms in computing entrance and exit doses JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Evaluation of six TPS algorithms in computing entrance and exit doses Yun I. Tan, 1,2a Mohamed Metwaly, 3 Martin Glegg, 3 Shaun P.

More information

General Accreditation Guidance. User checks and maintenance of laboratory balances

General Accreditation Guidance. User checks and maintenance of laboratory balances General Accreditation Guidance User checks and maintenance of laboratory balances January 2018 Copyright National Association of Testing Authorities, Australia 2010 All intellectual property rights in

More information

Planning Assignment (3 field rectum)

Planning Assignment (3 field rectum) Megan Sullivan Clinical Lab Rectum April 25, 2016 Planning Assignment (3 field rectum) Use a CT dataset of the pelvis. Create a CTV by contouring the rectum (start at the anus and stop at the turn where

More information

Proton Therapy QA Tools

Proton Therapy QA Tools Proton Therapy QA Tools A complete range of tools for absolute dosimetry, machine QA and patient plan verification Introduction During the past years more and more heavy particle therapy centers, especially

More information

Original Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams

Original Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams Iranian Journal of Medical Physics Vol. 10, No. 3, Summer 2013, 147-155 Received: April 22, 2013; Accepted: July 21, 2013 Original Article Effective Point of Measurement in Cylindrical Ion Chamber for

More information

Optimization in high dose rate vaginal cylinder for vaginal cuff irradiation

Optimization in high dose rate vaginal cylinder for vaginal cuff irradiation Rep Pract Oncol Radiother, 2008; 13(1): 35-48 Original Paper Received: 2007.07.09 Accepted: 2007.12.13 Published: 2008.02.29 Authors Contribution: A Study Design B Data Collection C Statistical Analysis

More information

EXPERIMENTAL RESULTS OF GUIDED WAVE TRAVEL TIME TOMOGRAPHY

EXPERIMENTAL RESULTS OF GUIDED WAVE TRAVEL TIME TOMOGRAPHY 18 th World Conference on Non destructive Testing, 16-20 April 2012, Durban, South Africa EXPERIMENTAL RESULTS OF GUIDED WAVE TRAVEL TIME TOMOGRAPHY Arno VOLKER 1 and Hendrik VOS 1 TNO, Stieltjesweg 1,

More information

The Study of Half-tooth Master Gear Structure Optimization Yazhou Xie a, Zhiliang Qian b

The Study of Half-tooth Master Gear Structure Optimization Yazhou Xie a, Zhiliang Qian b X 3rd International Conference on Mechatronics, Robotics and Automation (ICMRA 2015) The Study of Half-tooth Master Gear Structure Optimization Yazhou Xie a, Zhiliang Qian b School of Mechanical and Electric

More information

Modeling of Hydraulic Hose Paths

Modeling of Hydraulic Hose Paths Mechanical Engineering Conference Presentations, Papers, and Proceedings Mechanical Engineering 9-2002 Modeling of Hydraulic Hose Paths Kurt A. Chipperfield Iowa State University Judy M. Vance Iowa State

More information

ESTIMATION OF THE DESIGN WIND SPEED BASED ON

ESTIMATION OF THE DESIGN WIND SPEED BASED ON The Seventh Asia-Pacific Conference on Wind Engineering, November 8-12, 2009, Taipei, Taiwan ESTIMATION OF THE DESIGN WIND SPEED BASED ON UNCERTAIN PARAMETERS OF THE WIND CLIMATE Michael Kasperski 1 1

More information

TG-51: Experience from 150 institutions, common errors, and helpful hints

TG-51: Experience from 150 institutions, common errors, and helpful hints TG-51: Experience from 150 institutions, common errors, and helpful hints Ramesh C. Tailor, Ph. D., William F. Hanson, Ph.D., and Geoffrey S. Ibbott, Ph.D. Department of Radiation Physics, University of

More information

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience -

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience - Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience - Dr. Wolfgang Jockel, Martin Schneider, TÜV Rheinland Group, D-51105 Cologne / Germany 1. Introduction A new basis for the certification

More information

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers EXRADIN Ion Chambers Exradin Ion Chambers Exradin (EXacting RADiation INstrumentation) Ion Chambers have been built for over 33 years, are recognized by top research institutes and standards laboratories,

More information

This test shall be carried out on all vehicles equipped with open type traction batteries.

This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4. Determination of hydrogen emissions page 1 RESS-6-15 5.4.1. This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4.2. The test shall be conducted following

More information

Safety Manual. Process pressure transmitter IPT-1* 4 20 ma/hart. Process pressure transmitter IPT-1*

Safety Manual. Process pressure transmitter IPT-1* 4 20 ma/hart. Process pressure transmitter IPT-1* Safety Manual Process pressure transmitter IPT-1* 4 20 ma/hart Process pressure transmitter IPT-1* Contents Contents 1 Functional safety 1.1 General information... 3 1.2 Planning... 4 1.3 Instrument parameter

More information

The dose distribution of medium energy electron boosts to the laryngectomy stoma

The dose distribution of medium energy electron boosts to the laryngectomy stoma JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 1, WINTER 2001 The dose distribution of medium energy electron boosts to the laryngectomy stoma Ellen D. Yorke* Memorial Sloan-Kettering Cancer

More information

Transactions on Ecology and the Environment vol 12, 1996 WIT Press, ISSN

Transactions on Ecology and the Environment vol 12, 1996 WIT Press,   ISSN Open boundary condition for unsteady open-channel flow K. Mizumura Civil Engineering Department, Kanazawa Institute of Technology, 7-1 Ogigaoka, Nonoichimachi, Ishikawa Pref. 921, Japan Abstract Initial

More information

Safety Manual VEGAVIB series 60

Safety Manual VEGAVIB series 60 Safety Manual VEGAVIB series 60 NAMUR Document ID: 32005 Contents Contents 1 Functional safety... 3 1.1 General information... 3 1.2 Planning... 4 1.3 Adjustment instructions... 6 1.4 Setup... 6 1.5 Reaction

More information

3D Treatment Planning and verification with hand calculations

3D Treatment Planning and verification with hand calculations 3D Treatment Planning and verification with hand calculations Randy Holt, PhD, CEO Pacific Crest Medical Physics, Inc Chico, CA Dosimetric Calculations Calibration and Measurement Conditions Patient Conditions

More information

How to Design Medical Accelerator Systems for Reliability: IBA PT System

How to Design Medical Accelerator Systems for Reliability: IBA PT System How to Design Medical Accelerator Systems for Reliability: IBA PT System Yves Jongen Founder & Chief Research Officer Ion Beam Applications s.a. Belgium 1 Outline A short introduction of a proton therapy

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 1940-1 Second edition 2003-08-15 Mechanical vibration Balance quality requirements for rotors in a constant (rigid) state Part 1: Specification and verification of balance tolerances

More information

THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST

THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST SYSTEM Yixiong Xu Mechatronic Engineering, Shanghai University of Engineering Science, Shanghai, China ABSTRACT The pressure signal

More information

PREDICTING THE ABILITY OF SURVIVAL AFTER DAMAGE IN TANKERS. José Poblet Martínez - SENER, (Spain) José Javier Díaz Yraola - SENER, (Spain)

PREDICTING THE ABILITY OF SURVIVAL AFTER DAMAGE IN TANKERS. José Poblet Martínez - SENER, (Spain) José Javier Díaz Yraola - SENER, (Spain) 8 th International Conference on 477 PREDICTING THE ABILITY OF SURVIVAL AFTER DAMAGE IN TANKERS Abstract José Poblet Martínez - SENER, (Spain) José Javier Díaz Yraola - SENER, (Spain) To meet the demand

More information

Abstract: Introduction:

Abstract: Introduction: Are photon and electron beam calibrations more consistent with TG-51 than with TG-21? Ramesh C. Tailor, William F. Hanson, Nathan Wells, and Geoffrey S. Ibbott. U.T. MD Anderson Cancer Center, Houston

More information

VariSource High Dose Rate Afterloader Procedures. For Performing Breast Brachytherapy with. The SAVI TM Applicator

VariSource High Dose Rate Afterloader Procedures. For Performing Breast Brachytherapy with. The SAVI TM Applicator VariSource High Dose Rate Afterloader Procedures For Performing Breast Brachytherapy with The SAVI TM Applicator Contents I. CT Evaluation (~1 week prior to insertion) II. SAVI Applicator Implantation

More information

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61: Practical Implementation of TG-61: II. Guidelines for clinical implementation of TG-61 J.P. Seuntjens Medical Physics Unit McGill University, Montreal General Hospital Montréal, Canada jseuntjens@medphys.mcgill.ca

More information

OUTPUT at dmax in tissue 8x8 20x20 At 100cm SSD At 100cm SAD

OUTPUT at dmax in tissue 8x8 20x20 At 100cm SSD At 100cm SAD RAPHEX 2001 T1. All of the following are true regarding Percentage Depth Dose (PDD) except: A. Increases with increasing energy. B. Depends on field size. C. Is the dose at depth expressed as a % of the

More information

Sontek RiverSurveyor Test Plan Prepared by David S. Mueller, OSW February 20, 2004

Sontek RiverSurveyor Test Plan Prepared by David S. Mueller, OSW February 20, 2004 Sontek RiverSurveyor Test Plan Prepared by David S. Mueller, OSW February 20, 2004 INTRODUCTION Sontek/YSI has introduced new firmware and software for their RiverSurveyor product line. Firmware changes

More information

A COMPARATIVE STUDY OF MIX FLOW PUMP IMPELLER CFD ANALYSIS AND EXPERIMENTAL DATA OF SUBMERSIBLE PUMP

A COMPARATIVE STUDY OF MIX FLOW PUMP IMPELLER CFD ANALYSIS AND EXPERIMENTAL DATA OF SUBMERSIBLE PUMP IMPACT: International Journal of Research in Engineering & Technology (IMPACT: IJRET) ISSN 2321-8843 Vol. 1, Issue 3, Aug 2013, 57-64 Impact Journals A COMPARATIVE STUDY OF MIX FLOW PUMP IMPELLER CFD ANALYSIS

More information

Proceedings of Meetings on Acoustics

Proceedings of Meetings on Acoustics Proceedings of Meetings on Acoustics Volume 9, 2010 http://acousticalsociety.org/ 159th Meeting Acoustical Society of America/NOISE-CON 2010 Baltimore, Maryland 19-23 April 2010 Session 1pBB: Biomedical

More information

A method for checking high dose rate treatment times for vaginal applicators

A method for checking high dose rate treatment times for vaginal applicators JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 4, FALL 2001 A method for checking high dose rate treatment times for vaginal applicators Charles S. Mayo* Department of Radiation Oncology,

More information

Windcube FCR measurements

Windcube FCR measurements Windcube FCR measurements Principles, performance and recommendations for use of the Flow Complexity Recognition (FCR) algorithm for the Windcube ground-based Lidar Summary: As with any remote sensor,

More information

Autodesk Moldflow Communicator Process settings

Autodesk Moldflow Communicator Process settings Autodesk Moldflow Communicator 212 Process settings Revision 1, 3 March 211. Contents Chapter 1 Process settings....................................... 1 Profiles.................................................

More information

Organisation Internationale de Métrologie Légale

Organisation Internationale de Métrologie Légale Organisation Internationale de Métrologie Légale OIML INTERNATIONAL RECOMMENDATION Pressure balances Manomètres à piston OIML R 110 Edition 1994 (E) CONTENTS Foreword... 3 1 Scope... 4 2 Terminology...

More information