Monitor Unit Verification for Small Fields
|
|
- Jennifer Briggs
- 5 years ago
- Views:
Transcription
1 Monitor Unit Verification for Small Fields Patrick Higgins, Ph.D University of Minnesota Department of Radiation Oncology October 10, 2013
2 The Issues: How do we verify the monitor units calculated by the planning system are accurate when we have multiple issues of small fields and heterogeneous media? Do we have to run IMRT-like phantom tests (which avoid the problem of heterogeneity) to verify MU s? Can we do some kind of hand calculation to verify MU s? If so, how do we approximate heterogeneity effects? What level of check accuracy is reasonable to avoid catastrophe?
3 What do people do? As best as we have been able to tell, they do either: 1. Simple hand calcs for confirmation. But MuCheck and Rad Calc use deff for heterogeneous calcs. This is maybe the simplest of 1-d calculations and loses accuracy as the depth (and effective depth) increases. or, 2. Measure dose using Mapcheck or its equivalent (with or without point measurement) as is done for IMRT qa. But now we replace a hard problem with a much easier test.
4 What do we do? Historically, at Minnesota we have relied on hand calculations as being the most accurate, least biased method for monitor unit definition. The planning system MU s were considered a secondary check. Prior to Pinnacle (i.e. superposition/convolution) heterogeneity corrections were rarely done, except for lung. Even then we tried to be sure the central axis minimally overlapped lung so that simple hand calculations could still be done. Only in the last 5 or 6 years have we become comfortable with the predictive quality of Pinnacle for heterogeneous media and only after we convinced ourselves that we could do confirmative hand calculations for 6 MV to within about ±5%. Another great impacting factor on us was the introduction of SBRT which forced us to find a way to check doses in the presence of significant heterogeneity.
5 Planning System Accuracy The best we can hope for beam models is probably 2-5% accuracy when we include heterogeneities (AAPM report 85). For systems like Pinnacle that accuracy is further confounded by the subjectivity of the modeling process. Beam models (and modelers!) are further stretched in the limits of small fields where both the input data is more prone to error and heterogeneity modeling is not included. For IMRT plans we are forced to transfer heterogeneous plans to homogeneous phantoms in which we can do direct measurements to confirm monitor units and bypass the heterogeneity issue. For static field SBRT cases some places use the IMRT QA approach. We felt that, at least for static fields, we should be able to find a way to verify Pinnacle MU s without resorting to special (albeit flawed) measurements.
6 Planning system verification vs. Hand calculation As part of an M.S. project (Rishik Saxena) we measured the validity and range of application of several 1-d heterogeneity corrections that could easily be incorporated into a hand calculation for planning system MU verification. We built composite lung-soft tissue and bone-soft tissue phantoms to measure doses and compared against Pinnacle, Eclipse pencil beam and 3 1-d methods over a range of field sizes and energies: 1. Effective depth (used by MuCheck and other vendor supplied programs). 2. Effective attenuation (originally used in diagnostic applications). 3. Ratio of TMR s. All 3 methods use effective depth in some form or other. For this presentation we are going to concentrate on effective attenuation.
7 Effective Depth This is the simplest possible estimate of the effect of electron density on depth for the purpose of calculating dose in homogeneous media of density different from water. It is so primitive that it is not discussed in textbooks as a correction model. deff is the effective pathlength or depth in a medium based on the weighted electron densities of the elements in the medium. This has been used to adjust measurement depths over a small range to make up for the difference in density in water-replacement plastics for calibration purposes. It has been used in place of the physical depth in some vendor-supplied monitor unit calculation programs.
8 Effective attenuation An effective attenuation coefficient is used to estimate the correction to TERMA due to differences in physical and effective depths: ICF = exp( µ ( d eff d)) (Inhomogeneity Correction Factor) µ is the effective attenuation coefficient of water for a particular beam quality. (For this we have just used the linear attenuation coefficient for the peak photon energy.) d is the physical depth d eff is the effective depth In this method the homogeneously calculated pdd or TMR is corrected by the ICF.
9 Ratio of tissue-air ratios (RTAR) This method was one of the most commonly used methods in older commercial treatment planning systems and as a method of spot checking dose. Other names have been given to this method. The 1-d version is given by ICF = TAR( d eff, W ) TAR( d, W ) One might expect the inclusion of field size effects would be an advantage over simple attenuation differences, but as you will see we found no particular improvement in accuracy.
10 Model Evaluation In order to evaluate dose calculation accuracy for our planning systems we built a lung/soft tissue phantom. Measurements were made 1 at a number of depths (EDR2 film) and in interfaces for a range of field sizes and energies. These measurements have been used for some time for baseline testing of our planning systems. Using this data we could also evaluate the viability of different correction methods. 1 Saxena R. and Higgins PD, Measurement and Evaluation of Inhomogeneity Corrections and Monitor Unit Verification for Treatment Planning. Medical Dosimetry, 35:19-27, 2010.
11 Lung Phantom for Testing Solid Water 3 cm Cork Solid Water Cork Solid Water 3.0 cm 5.0 cm 7.0 cm 7.5 cm 8.5 cm 10.5 cm 11.5 cm 13.5 cm 15.5 cm 17.5 cm 4 cm 4.5 cm 4 cm 5 cm
12 4x D-effective %-difference Depth (cm) Attenuation corr Eclipse Eq. TAR Pinnacle 6 MV Pdd calculation vs. 10x10 measurement %-difference D-effective Attenuation corr Eclipse Eq. TAR Pinnacle Depth (cm)
13 4x %-difference Attenuation corr Eclipse Eq. TAR Pinnacle 10 MV Depth (cm) 10x10 Pdd calculation vs. measurement %-difference Attenuation corr Eclipse Eq. TAR Pinnacle Depth (cm)
14 25.0 4x4 %-difference Depth (cm) x10 Attenuation corr Eclipse Eq. TAR Pinnacle 18 MV Pdd calculation vs. measurement 10.0 %-difference Attenuation corr Eclipse Eq. TAR Pinnacle Depth (cm)
15 How good can Pinnacle be? Pdd vs. Depth, Varian 25 MV, field size 2x Pdd (%) Measured Pinnacle Depth (cm)
16 Measurement Summary 1. All of the 1-d correction errors are in the direction of overestimating doses. 2. Pinnacle appears accurate to within 5% for field sizes as small as 2x2 for energies of 6-25 MV. 3. For large field sizes (>10x10) all of the simple 1-d correction methods work fairly well (within 5%). 4. For small fields (4x4) and low energy (6 MV) the attenuation correction appeared promising, at least in a simple phantom setting.
17 Clinical Small Field MU Evaluation Phantom tests confirmed the accuracy of Pinnacle. The next step was to apply the simple 1-d corrections for real test cases. For our first 26 SBRT lung cases, we compared Pinnacle MU settings with hand monitor unit calculations, based on the physical depth, but corrected using the effective attenuation ICF factor. 2 2 Higgins PD, Adolfson T, Cho LC and Saxena R. Monitor Unit Checking in Heterogeneous SBRT Treatment Planning; Medical Dosimetry, Vol. 36 (3): , 2011
18 SBRT #19 Comparison of Pinnacle with ICFcorrected Point calc. Field d deff FS Eff FS ICF Pinnacle MU Point MU Point/ICF Delta a % b % c % d % e % f % g % h % I % j % ICF = exp{ µ w ( d eff d )} µ water ( 6 x ) = / cm
19 MU comparison with Pinnacle vs. ICF ICF Pinnacle Calculation Comparison: Attenuation Correction Most outliers are correlated with very low lung density % -15.0% -10.0% -5.0% 0.0% 5.0% 10.0% 15.0% 20.0% % Diff calc vs Pinnacle MU n Ave ICF St Dev Max Min Ave St Dev Max Min 1.80% 4.10% 14.70% % Pass number % 15% % 10% % 5% %
20 MU comparison with Pinnacle vs. Field Size Pinnacle Calculation Comparison: Field Size Effective Field Size % -15.0% -10.0% -5.0% 0.0% 5.0% 10.0% 15.0% 20.0% % Diff calc vs Pinnacle MU
21 Ratio of TMR Comparison of MU s with Pinnacle vs. ICF Pinnacle Calculation Comparison: Ratio of TMR's ICF % -5.0% 0.0% 5.0% 10.0% 15.0% 20.0% 25.0% 30.0% 35.0% 40.0% % Diff calc vs Pinnacle MU
22 SBRT Test Summary 1. Check calculations using effective attenuation to approximate the effect of heterogeneities on lung plans appear to be accurate to ±5% for 6 MV and field sizes 3x3 and larger. 2. For those fields where the check is more than 10% different from Pinnacle, we review the planned fields for potential problems, increasing the field size, if needed, to obtain better agreement. Even though we have the highest confidence in Pinnacle, we may not know if the discrepancy is due to the lousy 1-d test, or to modeling error in small field extrapolation. So it is better to err on the cautious side.
23 SBRT Test Summary 3. We have found that conditions where the point calculation disagrees the most with Pinnacle are for the situation of very low density lung tissue (COPD) and in areas of bone-lung interfaces where Pinnacle is less accurate. This led us to do the following study of density effects, relative to this simple point calculation check.
24 Effects of Lung and Target Density The amount of heterogeneity in the scattering environment certainly contributes to the treatment planning problem. Low density lungs require larger PTV volumes or block margins to achieve the same coverage as higher density lungs. We set up a water phantom model in Pinnacle with a spherical lung insert and varied the size and density of a target as well as the lung density. We then exercised our simple correction by calculating MU s vs. Pinnacle MU s for a single field ranging in size to just matching the target to a few cm larger.
25 Effects of Lung and Target Density 6 MV Field Unit density 3.5 cm Lung 15 cm PTV
26 Effects of Lung and Target Density cm Target Density 1.0 MuCalc MU/Pinnacle MU Lung density 0.25 Lung density 0.2 Lung density 0.15 Lung density 0.1 Lung density 0.05 Lung density Field Size ICF-corrected MU vs. Pinnacle MU Single Small Field of Different Widths: 5-10% overestimates.
27 Effects of Lung and Target Density cm Target Density 0.6 MuCalc MU/Pinnacle MU Field Size Lung density 0.25 Lung density 0.2 Lung density 0.15 Lung density 0.1 Lung density 0.05 Lung density 0.0 ICF-Corrected MU vs. Pinnacle for Typical Target Density: ±5% Range.
28 Effects of Lung and Target Density cm Target Density 0.3 MuCalc MU/Pinnacle MU Field Size Lung density 0.25 Lung density 0.2 Lung density 0.15 Lung density 0.1 Lung density 0.05 Lung density 0.0 ICF-Corrected MU vs. Pinnacle for Typical Target Density: 10+% underestimate depending on field size.
29 Effects of Lung and Target Density cm Target Density 0.3 MuCalc MU/Pinnacle MU Field Size Lung density 0.25 Lung density 0.2 Lung density 0.15 Lung density 0.1 Lung density 0.05 Lung density 0.0 ICF-Corrected MU vs. Pinnacle for Typical Target Density: Match improves with target and field size.
30 Conclusions 1. For low energy (6 MV) a 1-d attenuation-based correction can be used to verify the planned MU s for a well-tested planning system, avoiding the need for phantom test measurements and giving us some level of confidence that there is no catastrophic problem in the plan. 2. Small targets in SBRT lung cases pose additional problems for planning systems depending on how diffuse the targets are and on the surrounding lung density. It is better to increase the field margin to satisfy a simple 1-d test than use very tight margins and prescribe to low isodose lines. 3. Even though it may be argued that hand calculations are much inferior to a full, well-tested beam model, experience has shown that the risk of not performing some independent test is not worth taking.
31 Thank you!
32
33 1.6 cm Target Density Ratio 0.8 cm Lat:to:Center Dose Lung density 0.25 Lung density 0.2 Lung density 0.15 Lung density 0.1 Lung density 0.05 Lung density 0.0 Dose homogeneity: Ratio of dose at lateral edges of targets to the dose at the center. Field Size Ratio 1.5 cm Lat:to:Center Dose 3 cm Target Density Lung density 0.25 Lung density 0.15 Lung density 0.1 Lung density 0.05 Lung density 0.0 Effect of field size and densities of target and lungs on dose coverage. Field Size
34 Estimates of Uncertainty (in terms of one standard deviation) in absolute dose in the patient for megavoltage photons (AAPM Report 85). Source of Uncertainties Uncertainty at Uncertainty in Present (%) Future (%) Dose at the calibration point in water Additional uncertainty for other points Beam Monitor stability Beam flatness Patient data Patient set up and organ motion Overall (excluding dose calculation) Dose calculation algorithm 1.0 / 2.0 / 3.0 / / 2.0 / 3.0 TOTAL 4.4 / 4.7 / 5.2 / / 3.2 / 3.9
35 Table 7. Categorization of different inhomogeneity correction algorithms according to the level of anatomy sampled (1D or 3D) and the inclusion or exclusion of electron transport (AAPM Report #85) TERMA Local energy deposition (No electron transport) DOSE Non-local energy deposition(electron transport) 1d Category Linear attenuation 1.2 Ratio of TAR (RTAR) (Equivalent path length, effective SSD, isodose shift) 1.3 Power law (Batho) 3d Category Equivalent TAR (ETAR) 2.2 Differential SAR (DSAR) 2.3 Delta volume (DVOL) 2.5 3D Beam Subtraction Method Category Convolution (pencil beam) 3.2 FFT techniques Category Superposition/Convolution 4.2 Monte Carlo 2.4 Differential TAR (dtar)
36 1-d Methods for Evaluating Monitor Units 6 MV 2300cd pdd pdd (uncorrected) pdd (deff) Measured ICF-corrected pdd Pinnacle depth (cm)
Dosimetric Calculations. Lonny Trestrail
Dosimetric Calculations Lonny Trestrail 20 October 2008 Objectives Dose Distribution Measurements PDD, OCR TAR, SAR, TPR, TMR, SPR, SMR Arc or Rotational Therapy Isodose Curves Point Dose Calculations
More informationRelative Dosimetry. Photons
Relative Dosimetry Photons What you need to measure! Required Data (Photon) Central Axis Percent Depth Dose Tissue Maximum Ratio Scatter Maximum Ratio Output Factors S c & S cp! S p Beam profiles Wedge
More informationCommissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale
Commissioning an IMRT System for MLC Delivery Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Taking the broad view of commissioning Commissioning elements Validating the dosimetry system Commissioning the
More informationOutline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface
Chapter 11 reatment Planning Single Beams Radiation Dosimetry I Outline Basic terminology Curved contour surface correction (bolus, compensators, wedges) Oblique beam incidence Correction for tissue inhomogeneities
More informationCHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS
47 CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 4.1 INTRODUCTION Advanced treatment techniques use optimized radiation beam intensities to conform dose distribution to the
More informationReview of fundamental photon dosimetry quantities
Review of fundamental photon dosimetry quantities Narayan Sahoo Main sources of the materials included in this lecture notes are: (1) Radiation Oncology Physics: A Handbook for Teachers and Students Edited
More informationThe Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.
FRANZCR Examination Part I Radiation Oncology Radiotherapeutic Physics Candidate No.: The Royal Australian and New Zealand College of Radiologists FRANZCR Examination Part I Radiation Oncology Radiotherapeutic
More informationTG219: IT'S USE, STRENGTHS AND WEAKNESSES
1 CONFLICT OF INTEREST TG219: IT'S USE, STRENGTHS AND WEAKNESSES SOTIRI STATHAKIS, PHD, DABR None ROSTER STATEMENT OF THE PROBLEM Timothy C. Zhu, University of Pennsylvania, Philadelphia, PA (Co- Chair)
More informationSnapShot IMRT with compensators and FFF beams
SnapShot IMRT with compensators and FFF beams Vladimir Feygelman, PhD (1) Moffitt Cancer Center, Tampa, FL, USA Disclosure VF has a sponsored research agreement with.decimal. A short history of radiotherapy
More informationClinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas
Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas Current Implementation Status Current Implementation Status 1600 1400 TOTAL 1494 of 1623 ACTIVE INSTITUTIONS
More informationOnly 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.
Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas 2009 AAPM Summer School Educational Objectives Improve your understanding of how to implement TG-51
More informationMonitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting
Monitor Unit Calculations Part 2 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Calculation of machine setting reference dose machine setting =, reference dose
More informationTG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009
TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 DISCLAIMER: This publication and associated spreadsheets and digital files are based on sources and
More informationThe MC commissioning of CyberKnife with MLC (Tips and Tricks)
The MC commissioning of CyberKnife with MLC (Tips and Tricks) Alain Guemnie Tafo, PhD UPMC Pinnacle, Harrisburg, PA Institute Gustave Roussy, Villejuif, France Disclaimer & Disclosure The views expressed
More informationCLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions
CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions Rebecca M. Howell, Ph.D., D.A.B.R, The University of Texas at M.D. Anderson Cancer Center Overview 2 Commissioning
More informationMath Review. Overview
Drawn by Steve Yan, CMD Math Review Shirley Ann Pinegar-Johnston MS RT(R)(T)CMD Overview Drawn by Steve Yan, CMD Divergence & Intensity DIVERGENCE X-rays travel in Straight but Divergent lines INTENSITY
More information1. Question Answer cgy / MU cgy / MU 2. Question Answer
GS020113: Introduction to Medical Physics III: Therapy s to home work problem set assigned on 3/22/11 1. Question A patient is set up at 100 cm SSD on a 6 MVX machine. The dose rate at 10 cm in phantom
More information13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.
13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13.1.1 TRAINING GOAL 13.1.1.1 Among the responsabilities of a medical
More informationZoubir Ouhib Lynn Cancer Institute
Zoubir Ouhib Lynn Cancer Institute April 2015 Speaker for ELEKTA Brachytherapy using miniature X-ray sources Energy
More informationMonitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription
Monitor Unit Calculations Part 1 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Return to our first patient 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt
More informationValidation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.
International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2017, 6, 57-72 http://www.scirp.org/journal/ijmpcero ISSN Online: 2168-5444 ISSN Print: 2168-5436 Validation of Treatment
More informationTG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:
Practical Implementation of TG-61: II. Guidelines for clinical implementation of TG-61 J.P. Seuntjens Medical Physics Unit McGill University, Montreal General Hospital Montréal, Canada jseuntjens@medphys.mcgill.ca
More informationJOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 Comparison of monitor unit calculations performed with a 3D computerized planning system and independent hand calculations: Results
More informationCommissioning of Elekta 6MV FFF Versa HD and Pinnacle
Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Poster No.: R-0044 Congress: Type: Authors: Keywords: DOI: 2014 CSM Scientific Exhibit L. Bendall, I. Patel, N. McGrath, C. Rowbottom; MANCHESTER/
More informationField size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator
Journal of BUON 8: 55-59, 2003 2003 Zerbinis Medical Publications. Printed in Greece ORIGINAL ARTICLE Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator
More informationAbsorption measurements for a carbon fiber couch top and its modelling in a treatment planning system
Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system G. Kunz, F. Hasenbalg, P. Pemler 1 1 Klinik für Radio-Onkologie und Nuklearmedizin, Stadtspital Triemli
More informationExperimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media
Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media N Tyagi 1, B H Curran, P L Roberson, J M Moran, E Acosta and B A Fraass Department
More informationTHE development of more advanced techniques in radiotherapy,
1 A Geant4-based simulation of an accelerator s head used for Intensity Modulated Radiation Therapy F. Foppiano, B. Mascialino, M.G. Pia, M. Piergentili. Abstract We present a Geant4-based simulation,
More informationClinical Implementation of Volumetric Modulated Arc Therapy
Clinical Implementation of Volumetric Modulated Arc Therapy UT M.D. Anderson Cancer Center Ramaswamy Sadagopan, Rebecca M. Howell, Weiliang Du and Peter Balter Definition 2 Intensity Modulated Arc therapy
More informationConstancy checks of well-type ionization chambers with external-beam radiation units
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Constancy checks of well-type ionization chambers with external-beam radiation units Sara L. Hackett, 1a Benjamin Davis, 2 Andrew
More informationA beam-matching concept for medical linear accelerators
Acta Oncologica, 2009; 48: 192200 ORIGINAL ARTICLE A beam-matching concept for medical linear accelerators DAVID SJÖSTRÖM, ULF BJELKENGREN, WIVIANN OTTOSSON & CLAUS F. BEHRENS Copenhagen University Hospital,
More informationCommissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN
22 Turkish Journal of Cancer Volume 37, No.1, 2007 Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN SATISH PELAGADE 1, KALPANA
More informationThe Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007
The Effect of Immobilisation Devices on Radiotherapy Dose Distributions A thesis submitted in fulfilment of the requirements for the Degree of Master of Applied Science (Medical and Health Physics) By
More informationPlanning Assignment (3 field rectum)
Megan Sullivan Clinical Lab Rectum April 25, 2016 Planning Assignment (3 field rectum) Use a CT dataset of the pelvis. Create a CTV by contouring the rectum (start at the anus and stop at the turn where
More informationBasics of Proton Therapy. Proton Treatment Planning and Beam Optimization
Basics of Proton Therapy Proton Treatment Planning and Beam Optimization SAM Educational Session, WE D BRB 2 Mark Pankuch, PhD Northwestern Medicine Chicago Proton Center Today s objectives Review the
More informationANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL
ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL FOR THE DEGREE MASTER OF ARTS BY CHARLES TRAVIS WEBB
More informationTG-51: Experience from 150 institutions, common errors, and helpful hints
TG-51: Experience from 150 institutions, common errors, and helpful hints Ramesh C. Tailor, Ph. D., William F. Hanson, Ph.D., and Geoffrey S. Ibbott, Ph.D. Department of Radiation Physics, University of
More informationThe dose distribution of medium energy electron boosts to the laryngectomy stoma
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 1, WINTER 2001 The dose distribution of medium energy electron boosts to the laryngectomy stoma Ellen D. Yorke* Memorial Sloan-Kettering Cancer
More informationAbstract: Introduction:
Are photon and electron beam calibrations more consistent with TG-51 than with TG-21? Ramesh C. Tailor, William F. Hanson, Nathan Wells, and Geoffrey S. Ibbott. U.T. MD Anderson Cancer Center, Houston
More informationIMRT in clinical practice at the UMC-Utrecht. Utrecht
3 IMRT in clinical practice at the UMC-Utrecht Utrecht Clinical use of IMRT Improvement of local control without increased incidence of normal tissue complications + Dose escalation in the prostate Step-and
More informationCommissioning of a new total body irradiation protocol
University of Wollongong Research Online University of Wollongong Thesis Collection University of Wollongong Thesis Collections 2012 Commissioning of a new total body irradiation protocol Zoe Baldwin University
More informationClinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry
Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Tom Jordan Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE
More informationScientific Journal of Nuclear Medicine & Radiation Therapy
Scientific Journal of Nuclear Medicine & Radiation Therapy Research Article Radiotherapy Dose Calculation Program, for Monitoring and Commissioning Treatment Planning Systems - Ioannis Vamvakas*, Stella
More informationIranian Journal of Medical Physics
Iranian Journal of Medical Physics ijmp.mums.ac.ir Verification of Monitor Unit Calculations for Eclipse Treatment Planning System by in-house Developed Spreadsheet Athiyaman Mayivaganan 1*, Hemalatha
More informationEvaluation of six TPS algorithms in computing entrance and exit doses
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Evaluation of six TPS algorithms in computing entrance and exit doses Yun I. Tan, 1,2a Mohamed Metwaly, 3 Martin Glegg, 3 Shaun P.
More informationCOMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy
COMPASS 3D IMRT Verification in Patient Anatomy AK IMRT Bamberg April 2010 Dr. Lutz Müller Patient-specific Verification? CT Treatment unit Gantry =0 Treatmt unit Patient couch X-ray beam Patient couch
More informationA dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions
A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions Wendel Dean Renner a) Math Resolutions, LLC, 5975 Gales Lane, Columbia,
More informationIntroduction. Materials and Methods
STANDARD WEDGE AND TRAY TRANSMISSION VALUES FOR VARIAN, SEIMENS, ELEKTA/PHILIPS ACCELERATORS; A QUALITY ASSURANCE TOOL David S. Followill, Nadia Hernandez, and William F. Hanson Introduction The Radiological
More informationHigh Speed Direct SAD Radiosurgery Beam Scanner
Open Access Original Article DOI: 10.7759/cureus.20 High Speed Direct SAD Radiosurgery Beam Scanner Walter Nikesch 1, James M. Hevezi 2, Irene Monterroso 3, Daniel Navarro 4, James G. Schwade 5 1. CyberKnife
More informationIntroduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components
Introduction The CyberKnife SRS/SBRT System James M. Hevezi,, Ph.D., FACR CyberKnife Unit San Antonio, Texas CyberKnife Components Treatment Delivery (G3 Platform San Antonio) Robot (modified German auto
More informationOriginal Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams
Iranian Journal of Medical Physics Vol. 10, No. 3, Summer 2013, 147-155 Received: April 22, 2013; Accepted: July 21, 2013 Original Article Effective Point of Measurement in Cylindrical Ion Chamber for
More informationGEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY
The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World Chattanooga, Tennessee, April 17-21, 2005, on CD-ROM, American Nuclear Society, LaGrange Park, IL (2005) GEANT4 SIMULATION OF
More informationComparison of ionization chambers of various volumes for IMRT absolute dose verification
Comparison of ionization chambers of various volumes for IMRT absolute dose verification Leonid B. Leybovich, a) Anil Sethi, and Nesrin Dogan Department of Radiation Oncology, Loyola University Medical
More informationCommissioning and periodic tests of the Esteya electronic brachytherapy system
Review paper Educational Activity Commissioning and periodic tests of the Esteya electronic brachytherapy system Cristian Candela-Juan, PhD 1, Yury Niatsetski, MSc 2, Zoubir Ouhib, PhD 3, Facundo Ballester,
More informationA simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation
Radiotherapy and Oncology 71 (2004) 235 241 Technical Note A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation Nadine Linthout*,
More informationDisclosures. What Is So Hard? End of ITV: Gating is the Best ITV Killer 8/3/2016. Daniel A. Low, Ph.D. UCLA. Varian Grant Siemens Grant Accuray Grant
Real Timefullness 8/3/2016 End of ITV: Gating is the Best ITV Killer Daniel A. Low, Ph.D. UCLA Varian Grant Siemens Grant Accuray Grant Disclosures What Is So Hard? Light Ultrasound MRI X Ray Planar X
More informationRadiotherapy physics & Equipments
Radiotherapy physics & Equipments RAD 481 Lecture s Title: Beam Modification devices in Radiotherapy Dr. Mohammed Emam Vision :IMC aspires to be a leader in applied medical sciences, health care education
More informationPractical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11
MV Photon Dosimetry in the Clinic This document discusses both the scientific methods and the system of work required to calibrate a field instrument expanding upon the practical details of both.. Introduction
More informationInvestigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy
University of South Florida Scholar Commons Graduate Theses and Dissertations Graduate School 7-14-2010 Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy
More informationTreatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation
Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation By Ruth Ann M. Good, CMD. R.T. (R)(T) Virginia Commonwealth University, Richmond,
More information3D Treatment Planning and verification with hand calculations
3D Treatment Planning and verification with hand calculations Randy Holt, PhD, CEO Pacific Crest Medical Physics, Inc Chico, CA Dosimetric Calculations Calibration and Measurement Conditions Patient Conditions
More informationOUTPUT at dmax in tissue 8x8 20x20 At 100cm SSD At 100cm SAD
RAPHEX 2001 T1. All of the following are true regarding Percentage Depth Dose (PDD) except: A. Increases with increasing energy. B. Depends on field size. C. Is the dose at depth expressed as a % of the
More informationExradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers
EXRADIN Ion Chambers Exradin Ion Chambers Exradin (EXacting RADiation INstrumentation) Ion Chambers have been built for over 33 years, are recognized by top research institutes and standards laboratories,
More informationWEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS
Acta Radiologica Oncology 23 (1984) Fusc. 6 FROM THE RADIATION THERAPY DEPARTMENT, ANTON1 VAN LEEUWENHOEK HOSPITAL, THE NETHERLANDS CANCER INSTITUTE, AMSTERDAM, THE PHILIPS MEDICAL SYSTEMS DIVISION, BEST,
More informationPaper 2.2. Operation of Ultrasonic Flow Meters at Conditions Different Than Their Calibration
Paper 2.2 Operation of Ultrasonic Flow Meters at Conditions Different Than Their Calibration Mr William Freund, Daniel Measurement and Control Mr Klaus Zanker, Daniel Measurement and Control Mr Dale Goodson,
More informationCHAPTER-2. Wedge filters are the most important beam modifiers used in
CHAPTER-2 2. WEDGE FLTERS 2.1. NTRODUCTON Wedge filters are the most important beam modifiers used in radiotherapy which are placed in the path of photon beams to modify their isodose diistributions. The
More informationINTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS
A9744593 INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSLS Ladislav Czap, Georg Matscheko and Pedro Andreo osimetry and Medical Radiation Physics Section, ivision
More informationUnderwater measurement of photosynthetically active radiation
Underwater measurement of photosynthetically active radiation Mark Blonquist, Damon Nitzel, and Bruce Bugbee Apogee Instruments, Inc., Logan, Utah Introduction Quantum sensors (photosynthetically active
More informationUSER MANUAL PROGRAM VERSION 1.0.1
R WWW..COM REF 92730-92731 USER MANUAL PROGRAM VERSION 1.0.1 STANDARD IMAGING, INC. 7601 Murphy Drive Middleton, WI 53562-2532 Dec / 2004 2004 Standard Imaging, Inc. DOC #80464-01 TEL 800.261.4446 TEL
More informationA clinical comparison and analysis between conventional MLC based and solid compensator based IMRT treatment techniques
The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2009 A clinical comparison and analysis between conventional MLC based and solid compensator based IMRT treatment
More informationLarry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin
Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin NCCAAPM Meeting Oct 11,2013 Larry DeWerd has a partial interest in Standard Imaging Talks from people who have been
More informationProton Therapy QA Tools
Proton Therapy QA Tools A complete range of tools for absolute dosimetry, machine QA and patient plan verification Introduction During the past years more and more heavy particle therapy centers, especially
More informationAAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams
AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams Peter R. Almond Brown Cancer Center, Louisville, Kentucky 40202 Peter J. Biggs Department of Radiation Oncology,
More informationCommissioning and quality assurance for the treatment delivery components of the AccuBoost system
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 2, 2015 Commissioning and quality assurance for the treatment delivery components of the AccuBoost system Ileana Iftimia, 1,2a Mike Talmadge,
More informationOBJECTIVE 6: FIELD RADIOLOGICAL MONITORING - AMBIENT RADIATION MONITORING
OBJECTIVE 6: FIELD RADIOLOGICAL MONITORING - AMBIENT RADIATION MONITORING OBJECTIVE Demonstrate the appropriate use of equipment and procedures for determining field radiation measurements. INTENT This
More informationTreatment plan complexity metrics for predicting IMRT pretreatment. assurance results. Scott Crowe, QUT
Treatment plan complexity metrics for predicting IMRT pretreatment quality assurance results Scott Crowe, QUT Pre-treatment QA Verification of fluence delivery to ensure that dose can be delivered accurately
More informationRon Gibson, Senior Engineer Gary McCargar, Senior Engineer ONEOK Partners
New Developments to Improve Natural Gas Custody Transfer Applications with Coriolis Meters Including Application of Multi-Point Piecewise Linear Interpolation (PWL) Marc Buttler, Midstream O&G Marketing
More informationImplementation of respiratory-gated VMAT on a Versa HD linear accelerator
Received: 21 November 2016 Revised: 7 April 2017 Accepted: 5 June 2017 DOI: 10.1002/acm2.12160 RADIATION ONCOLOGY PHYSICS Implementation of respiratory-gated VMAT on a Versa HD linear accelerator Jeffrey
More informationTutorial for the. Total Vertical Uncertainty Analysis Tool in NaviModel3
Tutorial for the Total Vertical Uncertainty Analysis Tool in NaviModel3 May, 2011 1. Introduction The Total Vertical Uncertainty Analysis Tool in NaviModel3 has been designed to facilitate a determination
More informationDynamic wedges dosimetry and quality control
Original Paper Received: 2005.06.07 Accepted: 2006.02.28 Published: 2006.04.28 Authors Contribution: A Study Design B Data Collection C Statistical Analysis D Data Interpretation E Manuscript Preparation
More informationA method for checking high dose rate treatment times for vaginal applicators
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 4, FALL 2001 A method for checking high dose rate treatment times for vaginal applicators Charles S. Mayo* Department of Radiation Oncology,
More informationTechnical Report. 5th Round Robin Test for Multi-Capillary Ventilation Calibration Standards (2016/2017)
Physical Test Methods Sub-Group Technical Report 5th Round Robin Test for Multi-Capillary Ventilation Calibration Standards (2016/2017) June 2018 Author: James Vincent Cerulean, Milton Keynes, United Kingdom
More informationDevelopment of Graphical User Interfaces (GUI) software and database for radiation therapy applications
The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2005 Development of Graphical User Interfaces (GUI) software and database for radiation therapy applications
More informationPredicting climate conditions for turbine performance
Predicting climate conditions for turbine performance Mark Žagar, Vinay Belathur Krishna, Alvaro Matesanz Gil Vestas Data Engineering & Analytics / Advanced Plant Modelling Resource assessment, power curve,
More informationGeneral Accreditation Guidance. User checks and maintenance of laboratory balances
General Accreditation Guidance User checks and maintenance of laboratory balances January 2018 Copyright National Association of Testing Authorities, Australia 2010 All intellectual property rights in
More informationCHAPTER 7 : SMOKE METERS AND THEIR INSTALLATIONS
CHAPTER 7 : SMOKE METERS AND THEIR INSTALLATIONS 1 Scope : This Chapter covers the requirements of smoke meters and their installation on engines for full load and free acceleration tests, mentioned in
More informationVarian HDR surface applicators commissioning and clinical implementation
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Varian HDR surface applicators commissioning and clinical implementation Ileana Iftimia, 1,2,a Andrea B. McKee, 1,2 and Per H. Halvorsen
More informationHardware Triaxial and Consolidation Testing Systems Pressure Measurement and Control
25 GDS Helpsheet Hardware Triaxial and Consolidation Testing Systems Pressure Measurement and Control World Leaders in Computer Controlled Testing Systems for Geotechnical Engineers and Geologists 1. Digital
More informationEvaluations on CO2 Rich Natural Gas. Jörg Wenzel - Sick
Evaluations on CO2 Rich Natural Gas Jörg Wenzel - Sick ULTRASONIC GAS FLOW METER - EVALUATION ON CO 2 -RICH NATURAL GAS Author: Co-Author: Jörg Wenzel SICK AG E. Riedel, S. Stoof, Dr. V. Herrmann SICK
More informationGait Analyser. Description of Walking Performance
Gait Analyser Description of Walking Performance This brochure will help you to understand clearly the parameters described in the report of the Gait Analyser, provide you with tips to implement the walking
More informationComputationally Efficient Determination of Long Term Extreme Out-of-Plane Loads for Offshore Turbines
Computationally Efficient Determination of Long Term Extreme Out-of-Plane Loads for Offshore Turbines Anand Natarajan Senior Scientist Wind Energy Department, Risø DTU Denmark Introduction IEC 61400-1
More informationTorrild - WindSIM Case study
Torrild - WindSIM Case study Note: This study differs from the other case studies in format, while here another model; WindSIM is tested as alternative to the WAsP model. Therefore this case should be
More informationUser Manual. RW3 Slab Phantom T29672 and T
User Manual RW3 Slab Phantom Contents Operating Manual Technical Manual Service Manual D188.131.00/07 en 2015-07 Hn General Information General Information The product bears the CE-mark "CE" in accordance
More informationExposure System Selection
Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others
More informationDatasheet: K-30 ASCII Sensor
Datasheet: K-30 ASCII Sensor The K30 ASCII sensor is a low cost, infrared and maintenance free transmitter module intended to be built into different host devices that require CO2 monitoring data. The
More informationExecutive Summary of Accuracy for WINDCUBE 200S
Executive Summary of Accuracy for WINDCUBE 200S The potential of offshore wind energy has gained significant interest due to consistent and strong winds, resulting in very high capacity factors compared
More informationVariSource High Dose Rate Afterloader Procedures. For Performing Breast Brachytherapy with. The SAVI TM Applicator
VariSource High Dose Rate Afterloader Procedures For Performing Breast Brachytherapy with The SAVI TM Applicator Contents I. CT Evaluation (~1 week prior to insertion) II. SAVI Applicator Implantation
More informationBadan Jadrowych Nuclear Research Institute. Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P Differential Recombination Chamber
Badan Jadrowych Nuclear Research Institute Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P00100517 Differential Recombination Chamber by M. Zel'chinskij K. Zharnovetskij Warsaw, June 1966 Translated
More informationRotor Average wind speed for power curve performance. Ioannis Antoniou (LAC), Jochen Cleve (LAC), Apostolos Piperas (LAC)
Rotor Average wind speed for power curve performance Ioannis Antoniou (LAC), Jochen Cleve (LAC), Apostolos Piperas (LAC) March 2, 23 Contents Rotor Average wind speed EU flat terrain wind profiles vs.
More informationGuideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation
Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation This intercomparison will be carried out under the lead of KRISS with the cooperation of ARPANSA. This comparison has been approved
More information