A beam-matching concept for medical linear accelerators

Size: px
Start display at page:

Download "A beam-matching concept for medical linear accelerators"

Transcription

1 Acta Oncologica, 2009; 48: ORIGINAL ARTICLE A beam-matching concept for medical linear accelerators DAVID SJÖSTRÖM, ULF BJELKENGREN, WIVIANN OTTOSSON & CLAUS F. BEHRENS Copenhagen University Hospital, Department of Oncology (R), Division of Radiophysics (51AA), Herlev, Denmark Abstract The flexibility in radiotherapy can be improved if a patient can be moved between any one of the department s medical linear accelerators without the need to change anything in the patient s treatment plan. For this to be possible, the dosimetric characteristics of the various accelerators must be the same, or nearly the same i.e. the accelerators must be beam-matched. During a period of nine months, eight Varian ix accelerators with 6 and 15 MV photon beams and 618 MeV electron beams (only four of the eight) were installed at our clinic. All accelerators fulfilled the vendor-defined fine beam-match criteria, and a more extensive set of measurements was carried out during commissioning. The measured absorbed dose data for each accelerator were compared with the first accelerator, chosen as reference, and the TPS calculations. Two of the eight accelerators showed a larger discrepancy for the 15 MV beam not revealed by the vendor-defined acceptance criteria, whereas the other six accelerators were satisfactorily matched. The beam-matching acceptance criteria defined by the vendor are not strict enough to guarantee optimal beam-match. Deviations related to dose calculations and to beam-matched accelerators may add up. The safest and most practical way to ensure that all accelerators are within clinical acceptable accuracy is to include TPS calculations in the evaluation. Further, comparisons between measurements and calculations should be done in absolute dose terms. The success of curative intended radiation therapy largely depends on the ability to deliver the prescribed dose to the patient within a narrow tolerance. A 5% difference in the delivered dose may result in changes in the order of 1020% in tumor control probability and 2030% in normal tissue complication probability [1]. Thus, there is a need for accurate treatment planning and delivery. This imposes narrow tolerance limits on all links in the treatment chain from planning to delivery, including imaging and structure delineation, absorbed dose measurements and calculations, treatment machine performance, and patient setup. Several publications present recommendations for such tolerance levels [e.g. 25]. Ahnesjö et al. [6] estimates that with the present techniques the uncertainty of the dose delivered to the patient is 3.9% (1 SD) excluding uncertainty related to dose calculation. Thus, if the uncertainty in dose calculation is between 1.0 and 5.0% (1 SD) the total uncertainty in the dose delivered to the patient will be between 4.0 and 6.3% (1 SD) [6]. Wambersie [7] and Mijnheer et al. [8] recommend a total uncertainty limit of 3.5% (1 SD) for the dose prescription point although 5% (1SD) is considered more appropriate for other points in the planning target volume [7]. Often cited is the ICRU 24 [2]95% accuracy requirement that can be interpreted as an action level [6] or as 1.5 SD [9,10]. These recommendations on uncertainty limits and estimations of what can be achieved demonstrate that there is not much room for inaccurate dose calculations. Efforts must be made to reduce the uncertainty of every link in the treatment chain from planning to delivery. Treatment flexibility and efficiency of a radiotherapy department are very important, and can be greatly improved if several linear accelerators (linacs) can be considered dosimetrically identical and patients can be treated using any linac without the need to adjust the treatment plans. In order to make this possible, the dosimetric characteristics of the various accelerators must be the same, or nearly the Correspondence: David Sjöström, Copenhagen University Hospital, Herlev, Department of Oncology (R), Division of Radiophysics (51AA), Herlev Ringvej 75, DK-2730 Herlev, Denmark. Tel: Fax: davsjo01@heh.regionh.dk (Received 20 May 2008; accepted 4 June 2008) ISSN X print/issn X online # 2009 Informa UK Ltd. (Informa Healthcare, Taylor & Francis AS) DOI: /

2 Beam matching concept 193 same i.e. the accelerators must be beam-matched. Beam-match results for Varian accelerators have previously been presented e.g. [11,12]. In addition to beam-matched accelerators only one set of data representing all the accelerators needs to be implemented in the treatment planning system (TPS). Due to the strict requirements on accuracy in radiotherapy, the following two questions arises: Is it possible to beam-match several medical accelerators to such a degree that they can be acceptably represented by only one set of data in the TPS? If so, how should this be undertaken? The aim of this study is to address these two questions using the commissioning data obtained from eight newly installed accelerators. Material and methods During a period of about nine months, eight Varian 2300 ix (Varian Medical Systems, Inc., Palo Alto, CA, USA) linacs were installed at our clinic. The linacs have two photon beams of nominal energies 6 and 15 MV and four of the eight linacs have electron beams with nominal energies 6, 9, 12, 15, and 18 MeV. The beam-matching procedure is described in detail in the vendor Customer Acceptance Procedure (CAP) manual [13] and the vendor accelerator specifications [14] and also by Hrbacek et al. [11]. The accelerators were purchased with the Table I. Measurements carried out for photons. vendor-defined fine beam-matching option [11,14]. The beam-matching criteria are based on depth ionization curves as well as inline and crossline profiles measured in the vendor-defined prescribed geometry. During the CAP of the accelerators, all the vendor-defined criteria were fulfilled. Even if the measurements were within the vendor-defined criteria adjustments were made to improve the match between each accelerator and the first accelerator, chosen as a reference during the CAP. Primarily adjustments were made to make both the photon and electron profiles as symmetrical as possible, but for some of the photon and electron beams it was necessary to tune the energy in order to improve the match of the depth dose curves. Following the CAP more extensive commissioning measurements were undertaken (Tables I, II) in view of the fact that; (i) the vendor-defined criteria does not consider full profiles or depth dose curves to evaluate the quality of beam-matching but instead only some points on the curves, (ii) all vendor-specified measurements for both electrons and photons are done with an ionization chamber and there is no or incorrect ionization chamber offset, (iii) the vendor-defined criteria are limited to only a few parameters for a limited set of beam geometries, (iv) the beammatching criteria do not include output factors and therefore a risk of good agreement solely due to Geometry Measurements Equipment FSD [cm] Depth [cm] Field size [cm 2 ] Absolute dosimetry Farmer chamber NE2571***, electrometer NE2570/1***, water fantom** TPR 2010 CC13 ionization chamber*, Blue phantom*, OmniPro Accept v.6.5a* Output factors (OF) open fields in water Waterproof farmer chamber 30006**, Unidos electrometer**, Blue phantom* , 8, 10, 15, 20, 30, 40 Depth dose open fields CC13 ion chamber*, Blue phantom* OmniPro Accept v.6.5a* 90 5, 8, 10, 15, 20, 30, 40 Inline and crossline Photon diode PFD-36*, Blue phantom*, and , 40 profiles open fields OmniPro Accept v.6.5a* 2D distribution open Seven29**, water-equivalent RW3 slab 100 5, 10, 15 5, 10, 15, 20 fields phantom**, Verisoft** 2D distribution EDW30 Seven29**, water-equivalent RW3 slab 100 5, 10, 15 5, 10, 15, 20 phantom**, Verisoft** 2D distribution EDW45 Seven29**, water-equivalent RW3 slab 100 5, 10, 15 5, 10, 15, 20 phantom**, Verisoft** 2D distribution EDW60 Seven29**, water-equivalent RW3 slab 100 5, 10, 15 5, 10, 15, 20 phantom**, Verisoft** Head scatter factors (HSc) Farmer chamber FC65-P* with brass buildup cap, Unidos electrometer**, in air measurements 98,55/98.17 for 6/15 MV **** Brass****: 0.25/0.63 for 6/15 MV 2-10(step 1), 13, 15-40(step 5) *IBA Dosimetry, AB, Uppsala, Sweden. **PTW, Freiburg, Germany. ***Nuclear Enterprises, Beenham, Reading, England. ****Effective point of measurement at the isocenter, brass buildup cap different for 6 and 15 MV.

3 194 D. Sjöström et al. Table II. Measurements carried out for electrons. Geometry Measurements Equipment FSD [cm] Depth [cm] Applicator [cm 2 ] Absolute dosimetry Roos ionization chamber TW34001**, 100 Zref 10 Unidos electrometer**, water fantom** Output factors Electron diode EFD-36*, Blue phantom* 100 Zref 6, 10, 15, 20, 25 Depth dose Electron diode EFD-36*, Blue phantom*, 100 6, 10, 15, 20, 25 OmniPro Accept v.6.5a* Inline and crossline profiles Electron diode EFD-36*, Blue phantom*, OmniPro Accept v.6.5a* 100 zref 6, 10, 15, 20, 25 *IBA Dosimetry, AB, Uppsala, Sweden. **PTW, Freiburg, Germany. normalization may exist, (v) only open static fields are included in the CAP. To ensure that the beam-matching was clinically acceptable for photons the commissioning measurements were compared with the TPS, Eclipse 8.0 (Varian Medical Systems), pencil beam algorithm (PBC) calculations. Furthermore, all data were compared in absolute terms, i.e. depth dose curves and profiles for both open and wedged beams were not relatively normalized to the same value at a specific point but to the output factors for each individual accelerator. The 2D array measurements were normalized by defining a calibration factor for each depth for the open 1010 cm 2 field size that gave the same value as the reference conditions i.e. measurements in water. In that way, any possible differences in attenuation between the solid water and the water were eliminated. Three of the linacs were prepared and used for IMRT treatments and as a further test of the beam-matching the same IMRT plans were delivered with these accelerators, measured with the 2D array and compared with TPS calculation. The input data for the TPS were obtained from the first accelerator, and in the TPS only one set of data is used to represent all accelerators. The commissioning of the TPS will not be described in detail in this paper. All electron beam treatment plans utilized at our clinic today are calculated manually using the output factors and depth dose curves and without sophisticated calculation algorithms such as TPS calculations. Hence, the commissioning measurements were not compared with TPS calculations but instead with the average of the measured values for all four accelerators. The reference depth (z ref ) used for absolute calibration and output factor measurements is defined in IAEA TRS 398 [15] and the same z ref was used for all linacs. The reference depth for 6, 9, 12, 15, and 18 MeV are 13.9, 21.2, 29.8, 37.4 and 45.0 mm, respectively. On the basis of the output factors at z ref and the depth dose curves the output at dose maximum (d max ) can be calculated. Other parameters such as the 85% depth (R 85 ) and the 50% depth (R 50 ) can also be extracted from the depth dose curves. Isodose plots for IMRT were taken directly from Verisoft version 3.1 (PTW, Freiburg, Germany) as screen dumps. All other figures were created in Matlab R2006a. TPS calculations were exported as DICOM and imported in Matlab using functionalities in the Matlab Image Processing Toolbox and Emiliano Spezis DICOMRT toolbox [16]. 2D array measurements were saved as bitmap files in Verisoft and depth dose and profile measurements were converted to and saved as ASCII files in OmniPro Accept (IBA Dosimetry, AB, Uppsala, Sweden) and imported into Matlab. Results Photons All results are output factor normalized using the calibration geometry for each specific beam energy. Hence, the normalization retains all absolute dose properties and avoids any ambiguities related to normalization. The calibration geometry used is a 1010 cm 2 field, FSD100 cm at depth 10 cm for both the 6 and 15 MV beam. The TPR 20,10 and the percentage depth dose at 10 cm depth (PDD 10 ) vary slightly between the accelerators for the calibration geometry. For both the 6 and 15 MV beam the mean values BPDD 10 and BTPR 20,10 of all eight accelerators are used for absolute dose calibration. The mean values were also used as reference for the output factor normalization. For the 6 MV beam the mean values and intervals ([min;max]) are: BPDD %, [66.2;66.4]% and BTPR , [0.666; 0.670]. For the 15 MV beam the corresponding numbers are BPDD %,

4 Figure 1. Head scatter factors (HSc) and output factors in water (OF) versus field size for 15 MV photon beams. [76.3;76.7]% and BTPR , [0.760; 0.763]. For accelerators no. 1 to 6, the measured output factors (OF) are within 0.2% of the mean Beam matching concept 195 ( 16 ). Accelerators 7 and 8 are the most recently installed and not yet taken into clinical use. Larger discrepancies for these two accelerators are found for the 15 MV beam ranging from 71.2% to 0.9% relative to 16 for field size 55 cm 2 and 3030 cm 2, respectively (Figure 1). The discrepancies in OF between these two accelerators and the rest reflect a difference in head scatter (Figure 1). Further investigations into the cause and possible solution to the problem remains and will be undertaken in the near future. Based on a preliminary analysis of the data for accelerator 7 and 8, however, we expect the results presented bellow to apply also to these two accelerators except where explicit noted. The relative depth dose data for the various accelerators are within 1%, when the measured output factors for the different accelerators are ignored. Larger systematic deviations are noticed in the comparison between the measured and calculated depth dose data, owing to the differences in calculated and measured OF (Figure 2). The largest Figure 2. The deviations between measured and TPS calculated depth dose data for each individual accelerator for the 55 and 4040 cm 2 field size for the 6 MV and 15 MV open beams. The solid (thin) lines represent accelerator 16 and the dotted (thick) lines represent accelerator 78.

5 196 D. Sjöström et al. Figure 3. Comparison between measured and calculated dose profiles at 10 cm depth for 6 and 15 MV open field. The lines correspond to the result of a g 2%,2mm evaluation between the measured and calculated dose profiles for each accelerator. The solid (thin) lines represent accelerator 16 and the dotted (thick) lines represent accelerator 78. deviation found is for 15 MV 55 cm 2 field size for accelerator 7 and 8, where the measured output factor are around 2% lower than the calculated (Figure 2). For field sizes larger than 1010 cm 2, the measured output factors increase for accelerators 7 and 8 relative to the other accelerators, resulting in slightly better agreement with the TPS calculations for these accelerators. Except for 15 MV 55 cm 2 field size, all points in the output factor normalized depth dose data beyond the depth of d max are within 92% compared with the TPS calculations. The percentage deviations are larger at shallower depths. However, all data points for all field sizes and depths for all accelerators for both the 6 and 15 MV beams pass a 2%/2 mm gamma evaluation criteria, i.e. g 2%,2mm 51. In most cases, the result of a gamma evaluation between measured and calculated profiles for open beams along the major axes (gun-target and AB: perpendicular to and crossing the central axis) pass the g 2%,2mm. However, close to the field edges for larger field sizes only some of the accelerators pass the criteria (Figure 3). The comparison of relative dose profile measurements from one accelerator with another, normalized to 100% at the central axis show that the profiles generally are within 1% except for larger field sizes close to the field edges were the largest deviation between two accelerators generally are around 1.5% for both the 6 and 15 MV beams. Comparing the measured and calculated 2D dose distribution for open beams the 6 MV does not fulfill the g 2%,2mm for all accelerators for 1515 cm 2 nor a g 3%,2mm for the 2020 cm 2 field size. Only data points that are off-axis fail. Along the major axes data agree well with the results of measurements in water. A comparison between the 2D measurements and calculated 2D dose distribution for the 15 MV open beams reveal a better result since all the accelerators pass g 3%,2mm and only a few points for two accelerators for the 20x20 cm 2 field size fail g 2%,2mm for all depths. The penumbra (20% 80%) for the eight accelerators were determined from the measured profiles for the 1010 cm 2 field size. For the 6 MV beams the values are mm for the upper jaws and mm for the lower jaws. For 15 MV the corresponding values are mm and mm. For wedged beams, the deviations among the accelerators are larger than for open beams. The measured data are generally within 23% for both the 6 MV and the 15 MV beams (Figure 4B, D). For the 608 enhanced dynamic wedge (EDW60), 6 MV beams, 1515 and 2020 cm 2 field sizes, most of the accelerators fail g 3%,3mm in both the low and the high dose region (Figure 4A) when compared to the TPS calculations. Agreement is better for smaller wedge angles and smaller field sizes. For 15 MV, all the accelerators pass g 3%,3mm for 1515 cm 2 field size or smaller but fail in the low dose region for the 2020 cm 2 field size (Figure 4C). The comparison of a measured and a calculated dynamic IMRT field for three accelerators is

6 Beam matching concept 197 Figure 4. Comparison between measured and calculated 2D absorbed dose distributions for the EDW60 beams at 10 cm depth for the 6 MV and 15 MV beams. Each square (I) in A and C corresponds to one of the 729 ionization chambers in the 2D array. The numbers within the squares correspond to the number of accelerators that fails the g 3%,3mm for that specific chamber. The dotted line indicates the field size. The percentage deviations between the measured and calculated dose distributions along the arrows in A and C, i.e. in the wedge direction at the center of the beam, are presented in B and D, respectively. presented in Figure 5. Both the profile (Figure 5B) and the 2D dose distribution (Figure 5A) demonstrate that measurements and the calculation agree well and that the accelerators are well matched. The g 3%,3mm fail at exactly the same point in the high dose region for the three accelerators. Two additional points fail for accelerator 1 as compared to accelerator 2 and 3 in the lower dose area (Figure 5A). However, the dose in this area corresponds to about 5% of the maximum and the measurements are therefore very sensitive. Electrons The largest deviation (M R85,M R50 ) is 0.7 mm for all electron energies and applicators when comparing the depth of the 85% (R 85 ) and the depth of the 50% dose (R 50 ) with the mean value (Table III). Also the output factors at d max are in good agreement. The largest deviation of OF between any two accelerators is about 1% and compared to the mean OF the largest deviation (M OF ) is 0.6% (Table III). In addition, the profiles agree within a clinically acceptable level.

7 198 D. Sjöström et al. Figure 5. Comparison between measured and calculated 2D dose distributions for an 15 MV IMRT field for three different accelerators measured at 5.5 cm depth and SSD94.5 cm. The squares (I) in A corresponds to ionization chambers in the 2D array that fails g 3%,3mm. Solid and dashed lines in A corresponds to the measured and calculated data, respectively. The calculated and the measured dose profiles along arrows in A, perpendicular to the MLC, are presented in B. Better agreement is generally found for smaller applicators (Figure 6). The largest deviation between profiles for any of two accelerators is less than 2%, and the agreement is generally within 1% for applicators smaller than 2020 cm 2. Discussion We conclude that all photon and electron beams, except the 15 MV photon beam from accelerators no. 7 and 8, can be represented by one set of data. For 15 MV it might be necessary to implement an additional set of data in the TPS to represent the two deviant accelerators. To ensure a satisfactory beammatch the dose prediction (e.g. TPS calculations) must be adopted in the beam-matching procedure. Further, comparisons between measurements and calculations should be done in absolute dose terms. The beam-matching performed during the CAP far from ensures that the linacs are matched to a clinically acceptable degree. The discrepancy in measured OF for accelerators no. 7 and 8 compared Table III. The maximum deviation (MR50,, and MOF) between the individual values for R50, R85 and output factors (OF) for all accelerators compared to the mean value of the four accelerators (BR50, BR85, and ). Applicator size [cmcm] MOF [%] MR50 MOF [%] MR50 MOF [%] MOF [%] MR50 MR50 MOF [%] MR50 E [MeV]

8 Figure MeV electron beam inline profiles from four linacs. Profiles are at z ref for the 66, 1010, 1515, 2020, and 2525 cm 2 applicators. The profiles were first normalized to 100% at the central axis (CAX) then multiplied by the output factors. The curves are separated by shifting the 66, 1010, 1515, 2020, and 2525 cm 2 applicator and the difference at the CAX for one applicator corresponds to the difference in output factors between the different linacs. to the other accelerators is one example proving this. These two accelerators passed all tests in the CAP and showed results similar to all other accelerators. This illustrates that even if all relative comparisons, e.g. during the acceptance procedure, indicate a high-quality beam-match the result might turn out to be unacceptable when compared in absolute dose terms. Since the discrepancies in OF between accelerator no.7 and 8 and the other six are evident only for 15 MV our hypothesis is that this is due to a difference in the flattening filter. A systematic deviation between TPS calculations and measurements is caused by a difference in the measured and calculated OF (Figure 2). However, this was expected. The OF was measured for FSD 90 cm, at 10 cm depth. If isocentric OF is used as input to Eclipse the deviation between the calculated and measured dose for the calibration geometry (FSD 100, 10 cm depth) will be around 0.5%. According to personal communication with the vendor, zero deviation for the calibration geometry can only be achieved if Eclipse OF are defined at d max for FSD 100 cm. Several possibilities exist to overcome the discrepancy. The OF can be found at d max by direct measurements or calculations. The OF can be manually adjusted in the TPS until the match is improved. Alternatively, the calibration of the accelerators can be changed to minimize the difference. Since we will implement the Eclipse AAA calculation algorithm in the near future we expect to do neither. Larger differences among the linacs for the measured profiles were observed for EDW60 beams Beam matching concept 199 than for open beams. The EDW delivery is more complicated than open beam delivery and the EDW beam profiles are more likely to be influenced by variations in beam spectrum and jaw calibration than open beams. Thus, slight difference in jaw calibration may cause some of the differences seen between the linacs for EDW delivery. However, any difference in jaw calibration between the accelerators used in this study reflects what one can expect in clinical practice since all jaws were well calibrated. All electron beams utilized at our clinic today are calculated manually. In practice, the profiles are not used directly for selecting an applicator or insert although the necessary margins needed to achieve full coverage laterally are considered. The measurements for electrons reveals deviations smaller than 1% in OF and 1 mm in depth compared to the mean values of all accelerators. If more sophisticated treatment planning calculations are used to predict the dose, e.g. the Eclipse Electron Monte Carlo calculation algorithm, the matching of electron profiles with TPS calculations might be a matter of concern. One approach to ensure a satisfactory beammatch could be to use one set of beam data from one linear accelerator as a reference. The other linac(s) should then be matched to this data set as closely as possible and within certain tolerance levels. However, it is equally important that the TPS calculated dose closely resembles the delivered dose from all accelerators. When one accelerator is used as a reference there is a risk that another accelerator will be considered well matched while it does not fulfill the requirements in terms of accuracy when compared with the TPS. For instance, if the TPS predicts a slightly higher dose than delivered by the reference linac and the other linac delivers a dose slightly lower than the reference linac, the two uncertainties add up. Therefore we advocate for the approach to include the TPS calculations in the beam-matching procedure as presented in this work. In future QA procedures, we recommend using the TPS as reference. If the QA procedures reveal deviations that are too large to be ignored the procedure is to adjust the accelerator to fit the TPS instead of vice versa. Obviously, also comparisons between the different accelerators can be included in the QA procedures but either much narrower tolerance levels must be applied or a direct comparison with the TPS is needed in order to ensure that the deviations between calculated and delivered dose stay within acceptable limits. Declaration of interest: The authors report no conflicts of interest. The authors alone are responsible for the content and writing of the paper.

9 200 D. Sjöström et al. References [1] Chetty IJ, Curran B, Cygler JE, DeMarco JJ, Ezzell G, Faddegon BA, et al. Report of the AAPM Task Group No. 105: Issues associated with clinical implementation of Monte Carlo-based photon and electron external beam treatment planning. Med Phys 2007;34: [2] ICRU. Report 24: Determination of absorbed dose in a patient irradiated by beams of X- or gamma rays in radiotherapy procedures. International Commission on Radiation Units and Measurements. Washington, D.C., USA; [3] ICRU. Report 42: Use of computers in external beam radiotherapy procedures with high-energy photons or electrons. International Commission on Radiation Units and Measurements. Bethesda, MD, USA; [4] Mijnheer B, Olszewska A, Fiorino C, Hartmann G, Knöös T, Rosenwald JC, et al. Quality assurance of treatment planning systems. Practical Examples for Non-IMRT Photon Beams. Estro: Booklet 7, ISBN ; [5] Venselaar J, Welleweerd H, Mijnheer B. Tolerances for the accuracy of photon beam dose calculations of treatment planning systems. Radiother Oncol 2001;60: [6] Ahnesjö A, Aspradakis MM. Dose calculations for external photon beams in radiotherapy. Phys Med Biol 1999;44:R99 R155. [7] Wambersie A. What accuracy is required and can be achieved in radiation therapy (Review of Radiobiological and Clinical Data). Radiochimica Acta 2001;89: [8] Mijnheer BJ, Battermann JJ, Wambersie A. What degree of accuracy is required and can be achieved in photon and neutron therapy? Radiother Oncol 1987;8: [9] Goitein M. Nonstandard deviations. Med Phys 1983;10: [10] Goitein M. Calculation of the uncertainty in the dose delivered during radiation therapy. Med Phys 1985;12: [11] Hrbacek J, Depuydt T, Nulens A, Swinnen A, Heuvela FVD. Quantitative evaluation of a beam-matching procedure using one-dimensional gamma analysis. Med Phys 2007;34: [12] Watts RJ. Comparative measurements on a series of accelerators by the same vendor. Med Phys 1999;26: [13] Varian. High energy clinac: Clinac ix & Triology. Customer Acceptance Procedure (CAP). Revision E. 660 N.McCarthy Blvd.,Milpitas, CA 95035: Varian Medical Systems; [14] Varian. Specifications. Clinac ix Accelerator. Innovative platform for today, solid foundation for the future. Oncology Systems, 3100 HansenWay, Palo Alto, CA, : Varian Medical Systems; [15] IAEA. TRS-398. Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absorbed dose to water. IAEA; 2000 (2004: V.11b). [16] Spezi E, Lewis DG, Smith CW. A DICOM-RT-based toolbox for the evaluation and verification of radiotherapy plans. Phys Med Biol 2002;47:

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN 22 Turkish Journal of Cancer Volume 37, No.1, 2007 Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN SATISH PELAGADE 1, KALPANA

More information

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 47 CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 4.1 INTRODUCTION Advanced treatment techniques use optimized radiation beam intensities to conform dose distribution to the

More information

Relative Dosimetry. Photons

Relative Dosimetry. Photons Relative Dosimetry Photons What you need to measure! Required Data (Photon) Central Axis Percent Depth Dose Tissue Maximum Ratio Scatter Maximum Ratio Output Factors S c & S cp! S p Beam profiles Wedge

More information

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator Journal of BUON 8: 55-59, 2003 2003 Zerbinis Medical Publications. Printed in Greece ORIGINAL ARTICLE Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

More information

Constancy checks of well-type ionization chambers with external-beam radiation units

Constancy checks of well-type ionization chambers with external-beam radiation units JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Constancy checks of well-type ionization chambers with external-beam radiation units Sara L. Hackett, 1a Benjamin Davis, 2 Andrew

More information

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13.1.1 TRAINING GOAL 13.1.1.1 Among the responsabilities of a medical

More information

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Commissioning an IMRT System for MLC Delivery Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Taking the broad view of commissioning Commissioning elements Validating the dosimetry system Commissioning the

More information

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription Monitor Unit Calculations Part 1 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Return to our first patient 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt

More information

Review of fundamental photon dosimetry quantities

Review of fundamental photon dosimetry quantities Review of fundamental photon dosimetry quantities Narayan Sahoo Main sources of the materials included in this lecture notes are: (1) Radiation Oncology Physics: A Handbook for Teachers and Students Edited

More information

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2017, 6, 57-72 http://www.scirp.org/journal/ijmpcero ISSN Online: 2168-5444 ISSN Print: 2168-5436 Validation of Treatment

More information

THE development of more advanced techniques in radiotherapy,

THE development of more advanced techniques in radiotherapy, 1 A Geant4-based simulation of an accelerator s head used for Intensity Modulated Radiation Therapy F. Foppiano, B. Mascialino, M.G. Pia, M. Piergentili. Abstract We present a Geant4-based simulation,

More information

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 DISCLAIMER: This publication and associated spreadsheets and digital files are based on sources and

More information

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics. FRANZCR Examination Part I Radiation Oncology Radiotherapeutic Physics Candidate No.: The Royal Australian and New Zealand College of Radiologists FRANZCR Examination Part I Radiation Oncology Radiotherapeutic

More information

Commissioning and periodic tests of the Esteya electronic brachytherapy system

Commissioning and periodic tests of the Esteya electronic brachytherapy system Review paper Educational Activity Commissioning and periodic tests of the Esteya electronic brachytherapy system Cristian Candela-Juan, PhD 1, Yury Niatsetski, MSc 2, Zoubir Ouhib, PhD 3, Facundo Ballester,

More information

Comparison of ionization chambers of various volumes for IMRT absolute dose verification

Comparison of ionization chambers of various volumes for IMRT absolute dose verification Comparison of ionization chambers of various volumes for IMRT absolute dose verification Leonid B. Leybovich, a) Anil Sethi, and Nesrin Dogan Department of Radiation Oncology, Loyola University Medical

More information

The MC commissioning of CyberKnife with MLC (Tips and Tricks)

The MC commissioning of CyberKnife with MLC (Tips and Tricks) The MC commissioning of CyberKnife with MLC (Tips and Tricks) Alain Guemnie Tafo, PhD UPMC Pinnacle, Harrisburg, PA Institute Gustave Roussy, Villejuif, France Disclaimer & Disclosure The views expressed

More information

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Poster No.: R-0044 Congress: Type: Authors: Keywords: DOI: 2014 CSM Scientific Exhibit L. Bendall, I. Patel, N. McGrath, C. Rowbottom; MANCHESTER/

More information

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL FOR THE DEGREE MASTER OF ARTS BY CHARLES TRAVIS WEBB

More information

Clinical Implementation of Volumetric Modulated Arc Therapy

Clinical Implementation of Volumetric Modulated Arc Therapy Clinical Implementation of Volumetric Modulated Arc Therapy UT M.D. Anderson Cancer Center Ramaswamy Sadagopan, Rebecca M. Howell, Weiliang Du and Peter Balter Definition 2 Intensity Modulated Arc therapy

More information

Monitor Unit Verification for Small Fields

Monitor Unit Verification for Small Fields Monitor Unit Verification for Small Fields Patrick Higgins, Ph.D University of Minnesota Department of Radiation Oncology October 10, 2013 The Issues: How do we verify the monitor units calculated by the

More information

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface Chapter 11 reatment Planning Single Beams Radiation Dosimetry I Outline Basic terminology Curved contour surface correction (bolus, compensators, wedges) Oblique beam incidence Correction for tissue inhomogeneities

More information

Dosimetric Calculations. Lonny Trestrail

Dosimetric Calculations. Lonny Trestrail Dosimetric Calculations Lonny Trestrail 20 October 2008 Objectives Dose Distribution Measurements PDD, OCR TAR, SAR, TPR, TMR, SPR, SMR Arc or Rotational Therapy Isodose Curves Point Dose Calculations

More information

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions Rebecca M. Howell, Ph.D., D.A.B.R, The University of Texas at M.D. Anderson Cancer Center Overview 2 Commissioning

More information

SnapShot IMRT with compensators and FFF beams

SnapShot IMRT with compensators and FFF beams SnapShot IMRT with compensators and FFF beams Vladimir Feygelman, PhD (1) Moffitt Cancer Center, Tampa, FL, USA Disclosure VF has a sponsored research agreement with.decimal. A short history of radiotherapy

More information

Evaluation of six TPS algorithms in computing entrance and exit doses

Evaluation of six TPS algorithms in computing entrance and exit doses JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Evaluation of six TPS algorithms in computing entrance and exit doses Yun I. Tan, 1,2a Mohamed Metwaly, 3 Martin Glegg, 3 Shaun P.

More information

Dynamic wedges dosimetry and quality control

Dynamic wedges dosimetry and quality control Original Paper Received: 2005.06.07 Accepted: 2006.02.28 Published: 2006.04.28 Authors Contribution: A Study Design B Data Collection C Statistical Analysis D Data Interpretation E Manuscript Preparation

More information

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Tom Jordan Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE

More information

Scientific Journal of Nuclear Medicine & Radiation Therapy

Scientific Journal of Nuclear Medicine & Radiation Therapy Scientific Journal of Nuclear Medicine & Radiation Therapy Research Article Radiotherapy Dose Calculation Program, for Monitoring and Commissioning Treatment Planning Systems - Ioannis Vamvakas*, Stella

More information

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World Chattanooga, Tennessee, April 17-21, 2005, on CD-ROM, American Nuclear Society, LaGrange Park, IL (2005) GEANT4 SIMULATION OF

More information

Proton Therapy QA Tools

Proton Therapy QA Tools Proton Therapy QA Tools A complete range of tools for absolute dosimetry, machine QA and patient plan verification Introduction During the past years more and more heavy particle therapy centers, especially

More information

High Speed Direct SAD Radiosurgery Beam Scanner

High Speed Direct SAD Radiosurgery Beam Scanner Open Access Original Article DOI: 10.7759/cureus.20 High Speed Direct SAD Radiosurgery Beam Scanner Walter Nikesch 1, James M. Hevezi 2, Irene Monterroso 3, Daniel Navarro 4, James G. Schwade 5 1. CyberKnife

More information

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers EXRADIN Ion Chambers Exradin Ion Chambers Exradin (EXacting RADiation INstrumentation) Ion Chambers have been built for over 33 years, are recognized by top research institutes and standards laboratories,

More information

Zoubir Ouhib Lynn Cancer Institute

Zoubir Ouhib Lynn Cancer Institute Zoubir Ouhib Lynn Cancer Institute April 2015 Speaker for ELEKTA Brachytherapy using miniature X-ray sources Energy

More information

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas Current Implementation Status Current Implementation Status 1600 1400 TOTAL 1494 of 1623 ACTIVE INSTITUTIONS

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 Comparison of monitor unit calculations performed with a 3D computerized planning system and independent hand calculations: Results

More information

Abstract: Introduction:

Abstract: Introduction: Are photon and electron beam calibrations more consistent with TG-51 than with TG-21? Ramesh C. Tailor, William F. Hanson, Nathan Wells, and Geoffrey S. Ibbott. U.T. MD Anderson Cancer Center, Houston

More information

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting Monitor Unit Calculations Part 2 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Calculation of machine setting reference dose machine setting =, reference dose

More information

Iranian Journal of Medical Physics

Iranian Journal of Medical Physics Iranian Journal of Medical Physics ijmp.mums.ac.ir Verification of Monitor Unit Calculations for Eclipse Treatment Planning System by in-house Developed Spreadsheet Athiyaman Mayivaganan 1*, Hemalatha

More information

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08. Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas 2009 AAPM Summer School Educational Objectives Improve your understanding of how to implement TG-51

More information

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11 MV Photon Dosimetry in the Clinic This document discusses both the scientific methods and the system of work required to calibrate a field instrument expanding upon the practical details of both.. Introduction

More information

A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions

A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions Wendel Dean Renner a) Math Resolutions, LLC, 5975 Gales Lane, Columbia,

More information

Introduction. Materials and Methods

Introduction. Materials and Methods STANDARD WEDGE AND TRAY TRANSMISSION VALUES FOR VARIAN, SEIMENS, ELEKTA/PHILIPS ACCELERATORS; A QUALITY ASSURANCE TOOL David S. Followill, Nadia Hernandez, and William F. Hanson Introduction The Radiological

More information

VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests

VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 3, 2016 VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests Michael P. Barnes, 1,2a Pejman Rowshanfarzad,

More information

Math Review. Overview

Math Review. Overview Drawn by Steve Yan, CMD Math Review Shirley Ann Pinegar-Johnston MS RT(R)(T)CMD Overview Drawn by Steve Yan, CMD Divergence & Intensity DIVERGENCE X-rays travel in Straight but Divergent lines INTENSITY

More information

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization Basics of Proton Therapy Proton Treatment Planning and Beam Optimization SAM Educational Session, WE D BRB 2 Mark Pankuch, PhD Northwestern Medicine Chicago Proton Center Today s objectives Review the

More information

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system G. Kunz, F. Hasenbalg, P. Pemler 1 1 Klinik für Radio-Onkologie und Nuklearmedizin, Stadtspital Triemli

More information

TG-51: Experience from 150 institutions, common errors, and helpful hints

TG-51: Experience from 150 institutions, common errors, and helpful hints TG-51: Experience from 150 institutions, common errors, and helpful hints Ramesh C. Tailor, Ph. D., William F. Hanson, Ph.D., and Geoffrey S. Ibbott, Ph.D. Department of Radiation Physics, University of

More information

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007 The Effect of Immobilisation Devices on Radiotherapy Dose Distributions A thesis submitted in fulfilment of the requirements for the Degree of Master of Applied Science (Medical and Health Physics) By

More information

Surface buildup dose dependence on photon field delivery technique for IMRT

Surface buildup dose dependence on photon field delivery technique for IMRT Surface buildup dose dependence on photon field delivery technique for IMRT Shigeru Yokoyama, Peter L. Roberson, Dale W. Litzenberg, Jean M. Moran, and Benedick A. Fraass Department of Radiation Oncology,

More information

TEST REPORT BEAMSCAN WITH HALCYON

TEST REPORT BEAMSCAN WITH HALCYON TEST REPORT BEAMSCAN WITH HALCYON UZ Leuven, November 20 th, 2017 + = Ideal BEAMSCAN tank dimensions to fit perfectly into the Halcyon bore Fast tank positioning using the Halcyon lasers, no electronics

More information

Original Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams

Original Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams Iranian Journal of Medical Physics Vol. 10, No. 3, Summer 2013, 147-155 Received: April 22, 2013; Accepted: July 21, 2013 Original Article Effective Point of Measurement in Cylindrical Ion Chamber for

More information

IMRT in clinical practice at the UMC-Utrecht. Utrecht

IMRT in clinical practice at the UMC-Utrecht. Utrecht 3 IMRT in clinical practice at the UMC-Utrecht Utrecht Clinical use of IMRT Improvement of local control without increased incidence of normal tissue complications + Dose escalation in the prostate Step-and

More information

Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy

Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy University of South Florida Scholar Commons Graduate Theses and Dissertations Graduate School 7-14-2010 Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy

More information

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams Peter R. Almond Brown Cancer Center, Louisville, Kentucky 40202 Peter J. Biggs Department of Radiation Oncology,

More information

Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media

Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media N Tyagi 1, B H Curran, P L Roberson, J M Moran, E Acosta and B A Fraass Department

More information

Implementation of respiratory-gated VMAT on a Versa HD linear accelerator

Implementation of respiratory-gated VMAT on a Versa HD linear accelerator Received: 21 November 2016 Revised: 7 April 2017 Accepted: 5 June 2017 DOI: 10.1002/acm2.12160 RADIATION ONCOLOGY PHYSICS Implementation of respiratory-gated VMAT on a Versa HD linear accelerator Jeffrey

More information

A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation

A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation Radiotherapy and Oncology 71 (2004) 235 241 Technical Note A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation Nadine Linthout*,

More information

1. Question Answer cgy / MU cgy / MU 2. Question Answer

1. Question Answer cgy / MU cgy / MU 2. Question Answer GS020113: Introduction to Medical Physics III: Therapy s to home work problem set assigned on 3/22/11 1. Question A patient is set up at 100 cm SSD on a 6 MVX machine. The dose rate at 10 cm in phantom

More information

INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS

INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS A9744593 INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSLS Ladislav Czap, Georg Matscheko and Pedro Andreo osimetry and Medical Radiation Physics Section, ivision

More information

TG219: IT'S USE, STRENGTHS AND WEAKNESSES

TG219: IT'S USE, STRENGTHS AND WEAKNESSES 1 CONFLICT OF INTEREST TG219: IT'S USE, STRENGTHS AND WEAKNESSES SOTIRI STATHAKIS, PHD, DABR None ROSTER STATEMENT OF THE PROBLEM Timothy C. Zhu, University of Pennsylvania, Philadelphia, PA (Co- Chair)

More information

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61: Practical Implementation of TG-61: II. Guidelines for clinical implementation of TG-61 J.P. Seuntjens Medical Physics Unit McGill University, Montreal General Hospital Montréal, Canada jseuntjens@medphys.mcgill.ca

More information

User Manual. RW3 Slab Phantom T29672 and T

User Manual. RW3 Slab Phantom T29672 and T User Manual RW3 Slab Phantom Contents Operating Manual Technical Manual Service Manual D188.131.00/07 en 2015-07 Hn General Information General Information The product bears the CE-mark "CE" in accordance

More information

COMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC

COMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC COMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC by TAN ZI XIANG Thesis submitted in fulfillment of the requirements for the degree of Master of Science (Medical Physics) February 2015 ACKNOWLEDGEMENT

More information

Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water

Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water M R McEwen, A J Williams and A R DuSautoy Centre for Ionising Radiation Metrology National Physical

More information

Commissioning of a new total body irradiation protocol

Commissioning of a new total body irradiation protocol University of Wollongong Research Online University of Wollongong Thesis Collection University of Wollongong Thesis Collections 2012 Commissioning of a new total body irradiation protocol Zoe Baldwin University

More information

Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator

Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 2, 2013 Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator James L. Bedford, a Michael D. R. Thomas and Gregory

More information

IPEM Working Party: D IThwaites(Chair), A R DuSautoy, T Jordan, MRMcEwen, A Nisbet, A E Nahum and W G Pitchford

IPEM Working Party: D IThwaites(Chair), A R DuSautoy, T Jordan, MRMcEwen, A Nisbet, A E Nahum and W G Pitchford INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 48 (2003) 2929 2970 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(03)65222-X The IPEM code of practice for electron dosimetry for radiotherapy beams of

More information

Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry

Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry Bryan R. Muir a) and Malcolm R. McEwen Measurement Science and Standards, National Research Council of

More information

Electronic Oncology Systems the Management and the Safety Measures. Hansen Chen, Director, TDSI

Electronic Oncology Systems the Management and the Safety Measures. Hansen Chen, Director, TDSI Electronic Oncology Systems the Management and the Safety Measures Hansen Chen, Director, TDSI Radiation Oncology Department People Place Equipments Procedures Patients Hansen Chen 2 Radiation Oncology

More information

23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN APRIL 2007

23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN APRIL 2007 23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN 16-20 APRIL 2007 AN INVESTIGATION INTO THE INTERRELATION BETWEEN THE INTERNAL AND EXTERNAL BALLISTICS OF FIRING A TP-T TANK AMMUNITION M. H.

More information

COMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy

COMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy COMPASS 3D IMRT Verification in Patient Anatomy AK IMRT Bamberg April 2010 Dr. Lutz Müller Patient-specific Verification? CT Treatment unit Gantry =0 Treatmt unit Patient couch X-ray beam Patient couch

More information

Planning Assignment (3 field rectum)

Planning Assignment (3 field rectum) Megan Sullivan Clinical Lab Rectum April 25, 2016 Planning Assignment (3 field rectum) Use a CT dataset of the pelvis. Create a CTV by contouring the rectum (start at the anus and stop at the turn where

More information

Commissioning and quality assurance for the treatment delivery components of the AccuBoost system

Commissioning and quality assurance for the treatment delivery components of the AccuBoost system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 2, 2015 Commissioning and quality assurance for the treatment delivery components of the AccuBoost system Ileana Iftimia, 1,2a Mike Talmadge,

More information

Badan Jadrowych Nuclear Research Institute. Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P Differential Recombination Chamber

Badan Jadrowych Nuclear Research Institute. Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P Differential Recombination Chamber Badan Jadrowych Nuclear Research Institute Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P00100517 Differential Recombination Chamber by M. Zel'chinskij K. Zharnovetskij Warsaw, June 1966 Translated

More information

Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin

Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin NCCAAPM Meeting Oct 11,2013 Larry DeWerd has a partial interest in Standard Imaging Talks from people who have been

More information

Varian HDR surface applicators commissioning and clinical implementation

Varian HDR surface applicators commissioning and clinical implementation JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Varian HDR surface applicators commissioning and clinical implementation Ileana Iftimia, 1,2,a Andrea B. McKee, 1,2 and Per H. Halvorsen

More information

Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens

Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens The characteristic ionizing radiation response is defined by the relationship between the stimulus

More information

Technical protocol of APMP key comparison for measurement of air kerma for 60 Co (APMP.RI(I)-K1.1)

Technical protocol of APMP key comparison for measurement of air kerma for 60 Co (APMP.RI(I)-K1.1) Technical protocol of APMP key comparison for measurement of air kerma for 60 Co (APMP.RI(I)-K1.1) 1. Introduction A regional APMP.RI(I)-K1 comparison of the standards for air kerma for 60 Co occurred

More information

Radiotherapy physics & Equipments

Radiotherapy physics & Equipments Radiotherapy physics & Equipments RAD 481 Lecture s Title: Beam Modification devices in Radiotherapy Dr. Mohammed Emam Vision :IMC aspires to be a leader in applied medical sciences, health care education

More information

WEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS

WEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS Acta Radiologica Oncology 23 (1984) Fusc. 6 FROM THE RADIATION THERAPY DEPARTMENT, ANTON1 VAN LEEUWENHOEK HOSPITAL, THE NETHERLANDS CANCER INSTITUTE, AMSTERDAM, THE PHILIPS MEDICAL SYSTEMS DIVISION, BEST,

More information

7.65 ±0.05 mm of lead. The maximum energy of the bremsstrahlung photons was calibrated using the threshold for the D(-y,n)H reaction at

7.65 ±0.05 mm of lead. The maximum energy of the bremsstrahlung photons was calibrated using the threshold for the D(-y,n)H reaction at Am 7Roentgenolia6:ia6oia65, 1976 CENTRAL AXIS DEPTH DOSE FOR A 2.5 MV VAN DE GRAAFF GENERATOR DAVID W. ANDERSON, DAVID E. RAESIDE, REBA I. ADAMS, AND MYRON R. GOEDE 2 ABSTRACT: Central axis percentage

More information

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience -

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience - Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience - Dr. Wolfgang Jockel, Martin Schneider, TÜV Rheinland Group, D-51105 Cologne / Germany 1. Introduction A new basis for the certification

More information

The dose distribution of medium energy electron boosts to the laryngectomy stoma

The dose distribution of medium energy electron boosts to the laryngectomy stoma JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 1, WINTER 2001 The dose distribution of medium energy electron boosts to the laryngectomy stoma Ellen D. Yorke* Memorial Sloan-Kettering Cancer

More information

CHAPTER-2. Wedge filters are the most important beam modifiers used in

CHAPTER-2. Wedge filters are the most important beam modifiers used in CHAPTER-2 2. WEDGE FLTERS 2.1. NTRODUCTON Wedge filters are the most important beam modifiers used in radiotherapy which are placed in the path of photon beams to modify their isodose diistributions. The

More information

How to Design Medical Accelerator Systems for Reliability: IBA PT System

How to Design Medical Accelerator Systems for Reliability: IBA PT System How to Design Medical Accelerator Systems for Reliability: IBA PT System Yves Jongen Founder & Chief Research Officer Ion Beam Applications s.a. Belgium 1 Outline A short introduction of a proton therapy

More information

Maintenance and general care of secondary standard and field instruments

Maintenance and general care of secondary standard and field instruments Maintenance and general care of secondary standard and field instruments This is without doubt the most important lecture that you will be given during this course. That rather rash statement is based

More information

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation This intercomparison will be carried out under the lead of KRISS with the cooperation of ARPANSA. This comparison has been approved

More information

Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System

Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System Amit Sawant 1, Sonja Dieterich 1, Michelle Svatos 2 and Paul Keall 1 1 Stanford University, Stanford, CA 2 Varian Medical Systems, Palo

More information

IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy

IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy Guidelines on standardized procedures at Secondary Standards Dosimetry Laboratories (SSDLs) and hospitals March 2002 The

More information

Quality control for Intensity-modulated radiation therapy

Quality control for Intensity-modulated radiation therapy Schweizerische Gesellschaft für Strahlenbiologie und Medizinische Physik Société Suisse de Radiobiologie et de Physique Médicale Società Svizzera di Radiobiologia e di Fisica Medica Swiss Society for Radiobiology

More information

Treatment plan complexity metrics for predicting IMRT pretreatment. assurance results. Scott Crowe, QUT

Treatment plan complexity metrics for predicting IMRT pretreatment. assurance results. Scott Crowe, QUT Treatment plan complexity metrics for predicting IMRT pretreatment quality assurance results Scott Crowe, QUT Pre-treatment QA Verification of fluence delivery to ensure that dose can be delivered accurately

More information

An investigation of kinematic and kinetic variables for the description of prosthetic gait using the ENOCH system

An investigation of kinematic and kinetic variables for the description of prosthetic gait using the ENOCH system An investigation of kinematic and kinetic variables for the description of prosthetic gait using the ENOCH system K. OBERG and H. LANSHAMMAR* Amputee Training and Research Unit, University Hospital, Fack,

More information

In air calibration of an HDR 192 Ir brachytherapy source using therapy ion chambers

In air calibration of an HDR 192 Ir brachytherapy source using therapy ion chambers Original Article Free full text available from www.cancerjournal.net In air calibration of an HDR 192 Ir brachytherapy source using therapy ion chambers ABSTRACT The Gammamed Plus 192 Ir high dose rate

More information

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components Introduction The CyberKnife SRS/SBRT System James M. Hevezi,, Ph.D., FACR CyberKnife Unit San Antonio, Texas CyberKnife Components Treatment Delivery (G3 Platform San Antonio) Robot (modified German auto

More information

IAEA-TECDOC-1455 Implementation of the International Code of Practice on Dosimetry in Radiotherapy (TRS 398): Review of testing results

IAEA-TECDOC-1455 Implementation of the International Code of Practice on Dosimetry in Radiotherapy (TRS 398): Review of testing results IAEA-TECDOC-1455 Implementation of the International Code of Practice on Dosimetry in Radiotherapy (TRS 398): Revie of testing results Final report of the Coordinated Research Projects on Implementation

More information

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER Proven Rad-Check technology specifically designed to provide you with the ultimate in versatility and cost-effective operation. Fast and easy

More information

Methods to model and predict the ViewRay treatment deliveries to aid patient scheduling and treatment planning

Methods to model and predict the ViewRay treatment deliveries to aid patient scheduling and treatment planning Washington University School of Medicine Digital Commons@Becker Open Access Publications 2016 Methods to model and predict the ViewRay treatment deliveries to aid patient scheduling and treatment planning

More information

Part 1: General principles

Part 1: General principles Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 19030-1 First edition 2016-11-15 Ships and marine technology Measurement of changes in hull and propeller performance Part 1: General principles

More information

Queue analysis for the toll station of the Öresund fixed link. Pontus Matstoms *

Queue analysis for the toll station of the Öresund fixed link. Pontus Matstoms * Queue analysis for the toll station of the Öresund fixed link Pontus Matstoms * Abstract A new simulation model for queue and capacity analysis of a toll station is presented. The model and its software

More information

Total Ionization Dose (TID) Test Results of the RH117H Positive Adjustable Low Dose Rate (LDR) LDR = 10 mrads(si)/s

Total Ionization Dose (TID) Test Results of the RH117H Positive Adjustable Low Dose Rate (LDR) LDR = 10 mrads(si)/s Total Ionization Dose (TID) Test Results of the RH117H Positive Adjustable Regulator @ Low Dose Rate (LDR) LDR = 10 mrads(si)/s 18 December 2014 Duc Nguyen, Sana Rezgui Acknowledgements The authors would

More information