Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization

Size: px
Start display at page:

Download "Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization"

Transcription

1 Basics of Proton Therapy Proton Treatment Planning and Beam Optimization SAM Educational Session, WE D BRB 2 Mark Pankuch, PhD Northwestern Medicine Chicago Proton Center

2 Today s objectives Review the concepts of CTV / ITV / PTV when treating with protons Discuss the general planning parameters used in proton planning Present Aperture / Compensator, forward based, treatment planning methods Discuss the methods and clinical benefits of Intensity Modulated Proton Therapy (IMPT)

3 ICRU Definitions Patient CTV GTV ITV

4 ICRU Definitions Patient ITV PTV = ITV + SM

5 ICRU Definitions Patient ITV PTV = ITV + SM

6 ICRU Definitions Patient ITV PTV = ITV + SM

7 So. Distal margins of Protons fields need to consider potential path length differences (Smearing and Robustness) But are there other considerations for the distal edge margins?? What about Proton Range Uncertainties??

8 Where do range uncertainties come from and how big are they?? It depends on who you ask Paganetti, Phys. Med Biol (57), 2012

9 Moyers : Ion Stopping Powers and CT Numbers Moyers, Medical Dosimetry(35), 2010

10 Yang : Comprehensive analysis of proton range uncertainties related to patient stopping power ratio estimation using the stoichiometric calibration Yang, Phys. Med. Biol, (57) 2012

11 Paganetti : Range uncertainties in proton therapy and the role of Monte Carlo simulations Paganetti, Phys. Med Biol (57), 2012

12 Setup and Range Uncertainty with Protons : Field Specific Margins Perpendicular Expansion Avoid a geometric miss Physical Distance (cm) Parallel Expansion Avoid a range miss Radiobiological Depth (Water Equiv. Thickness)

13

14

15 CT is the Patient Map : Areas of Specific Concern for Protons Conversion from HU to RSP has inherent problems Noise Beam hardening Trying to make our CT scanner a spectrometer Two tissues can have same HU but different RSP Anything not natural can have large errors. Contrast Fillings Implants

16 Chestwall Expander

17 Breast Prosthesis Moyers et.al Med Dosim Spring;39(1):

18 Is there any hope for improvements? Dual Energy CT (kv / MVCT) Proton activation (PET/SPECT) Tomography Prompt Gamma verification Proton Radiography Proton Tomography

19 Treatment Planning Methods depend on Proton Delivery Methods Double Scatter (DS) Uniform Scanning (US) Aperture / Compensator Based Planning Pencil Beam Scanning (PBS) Intensity Modulated Proton Therapy (IMPT) Inversed planned proton spot intensity optimization 19

20 Aperture / Compensator based Planning Strategies Cover the target with appropriate margins Spare the critical structures Plan with fields that deliver the most robust plan

21 Tools to do our job Range : The depth of the Bragg peak (Distal 90%) Modulation : The spread of the Bragg peak Apertures : Shaping the beam perpendicular to the path Compensators : Distal Shaping

22 The Physics of Protons Spread Out Bragg Peak (SOBP) Relative Dose Healthy Tissue Tumor Depth in Tissue (cm) Healthy Tissue

23 Range and Modulation

24 Spreading the beam across the field 24

25 Patient Specific Devices Aperture

26 Aperture Design

27 27

28 Penumbra at Various Air Gaps 1 2 cm cm cm 20 2 cm cm cm Distance (cm)

29 Penumbra as Mid SOBP at various ranges 1 Penumbra at Various Ranges, mid SOPB (4cm) 0.8 Relative Intensity 0.6 Range 10cm Range 10cm Range 20cm 10cm Range 30cm 30m 20cm Distance (cm)

30 For treatments, the snout is extended to be close to the patient Reduced Air Gap = smaller penumbra 30

31 Compensators for Distal Shaping

32 Target Area Proton Beam Aperture Inhomogeneity (Air Pocket)

33 Compensator Target Area Aperture Inhomogeneity (Air Pocket)

34

35

36 Compensator Target Area Aperture Inhomogeneity (Air Pocket)

37 Compensator Target Area Aperture Inhomogeneity (Air Pocket)

38

39 Compensator Target Area Aperture Inhomogeneity (Air Pocket)

40 Smearing Sacrificing distal conformity to ensure you have enough range (and modulation) to cover the target in the case of anticipated misalignments Accounts for the fact that treatment path lengths may be different than planned path lengths due to set up errors and internal motion. Can easily be built into compensator design Is not directly mechanically accounted for in IMPT (No compensator)

41 Intensity Modulated Proton Therapy (IMPT) Layers of spot patterns delivered over the target volume Variable Intensity Control Dose uniformity Simultaneous Intergraded Boost Distal AND Proximal conformity The ability to perform Single Field or Multi Field Optimizations Movie clip from Varian ProBeam :

42 Spot Intensity for SFUD plan BEV of the spot intensity patterns

43 Spot Positions and Intensity Impossible to manually define spot positions and intensities and hope they relate to each other. Inverse planning is required Objective function is defined An iterative process is used to minimize the objective function 43

44 Use of a Compensator for distal shaping

45 Distal conformity using a Compensator No Compensator With Compensator

46 Advantage of PBS : The addition of Proximal Conformity PBS With Compensator

47 IMPT Optimization Methods Single Field Optimization (SFO) Uniform Dose is delivered to the entire target by each field individually Multi Field Optimization (MFO) Spot weights of all fields are optimized together. The spot weight of one field will rely on another field s dose to create an integrated uniform target dose Less sparing of critical structures Better for sparing critical structures Less sensitive to Set up/range errors More sensitive to Set up/range errors

48 Single Field Optimized : (SFO) OAR < 100% of Dose 100% of Dose

49 Multi Field Optimized: (MFO) OAR < 100% of Dose 100% of Dose

50 Single Field Optimized with a systematic range error OAR < 100% of Dose 100% of Dose

51 Multi Field Optimized with a systematic range error OAR < 100% of Dose 100% of Dose

52 A need to Quantify and account for the effects of: Non ideal set up Range uncertainty Intra fraction motion Respiratory motion Inter fraction motion Anatomical consistency

53 Quantify and account for the effects of: Two methods to do this: Prospectively : Robustness Optimization Retrospectively : Robustness Evaluation

54 PBS Robust Optimization Add penalties into the cost function for robustness Allow the planning system to score robustness on a spot to spot basis AND how one spot will effect the overall sensitivity to potential plan degradation. Spots with poor robustness (high sensitivity to plan degradation) will be penalized by iteratively decreasing, and potentially, eliminating their intensity

55 Robustness Evaluation Process of evaluating several potential scenarios to understand potential worse case results Translate and/or rotate individual fields and recalculate Mimic Set up errors Re assign systematically shifted HU conversion curves and recalculate Mimic HU conversion errors Move Target structures and recalculate Mimic Internal Motion and Anatomical Variances

56 Robust Evaluations

57 In Conclusion.. Because the protons stop, the standard planning methods used in photon planning may not be directly applicable. Treatment Planning with proton requires several addition consideration. The proper application of proton planning techniques can generate some very wonderful treatment plans.

58 Thanks You for listening!

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 47 CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 4.1 INTRODUCTION Advanced treatment techniques use optimized radiation beam intensities to conform dose distribution to the

More information

SnapShot IMRT with compensators and FFF beams

SnapShot IMRT with compensators and FFF beams SnapShot IMRT with compensators and FFF beams Vladimir Feygelman, PhD (1) Moffitt Cancer Center, Tampa, FL, USA Disclosure VF has a sponsored research agreement with.decimal. A short history of radiotherapy

More information

Disclosures. What Is So Hard? End of ITV: Gating is the Best ITV Killer 8/3/2016. Daniel A. Low, Ph.D. UCLA. Varian Grant Siemens Grant Accuray Grant

Disclosures. What Is So Hard? End of ITV: Gating is the Best ITV Killer 8/3/2016. Daniel A. Low, Ph.D. UCLA. Varian Grant Siemens Grant Accuray Grant Real Timefullness 8/3/2016 End of ITV: Gating is the Best ITV Killer Daniel A. Low, Ph.D. UCLA Varian Grant Siemens Grant Accuray Grant Disclosures What Is So Hard? Light Ultrasound MRI X Ray Planar X

More information

Relative Dosimetry. Photons

Relative Dosimetry. Photons Relative Dosimetry Photons What you need to measure! Required Data (Photon) Central Axis Percent Depth Dose Tissue Maximum Ratio Scatter Maximum Ratio Output Factors S c & S cp! S p Beam profiles Wedge

More information

Clinical Implementation of Volumetric Modulated Arc Therapy

Clinical Implementation of Volumetric Modulated Arc Therapy Clinical Implementation of Volumetric Modulated Arc Therapy UT M.D. Anderson Cancer Center Ramaswamy Sadagopan, Rebecca M. Howell, Weiliang Du and Peter Balter Definition 2 Intensity Modulated Arc therapy

More information

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers EXRADIN Ion Chambers Exradin Ion Chambers Exradin (EXacting RADiation INstrumentation) Ion Chambers have been built for over 33 years, are recognized by top research institutes and standards laboratories,

More information

How to Design Medical Accelerator Systems for Reliability: IBA PT System

How to Design Medical Accelerator Systems for Reliability: IBA PT System How to Design Medical Accelerator Systems for Reliability: IBA PT System Yves Jongen Founder & Chief Research Officer Ion Beam Applications s.a. Belgium 1 Outline A short introduction of a proton therapy

More information

The MC commissioning of CyberKnife with MLC (Tips and Tricks)

The MC commissioning of CyberKnife with MLC (Tips and Tricks) The MC commissioning of CyberKnife with MLC (Tips and Tricks) Alain Guemnie Tafo, PhD UPMC Pinnacle, Harrisburg, PA Institute Gustave Roussy, Villejuif, France Disclaimer & Disclosure The views expressed

More information

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions Rebecca M. Howell, Ph.D., D.A.B.R, The University of Texas at M.D. Anderson Cancer Center Overview 2 Commissioning

More information

Monitor Unit Verification for Small Fields

Monitor Unit Verification for Small Fields Monitor Unit Verification for Small Fields Patrick Higgins, Ph.D University of Minnesota Department of Radiation Oncology October 10, 2013 The Issues: How do we verify the monitor units calculated by the

More information

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Commissioning an IMRT System for MLC Delivery Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Taking the broad view of commissioning Commissioning elements Validating the dosimetry system Commissioning the

More information

VariSource High Dose Rate Afterloader Procedures. For Performing Breast Brachytherapy with. The SAVI TM Applicator

VariSource High Dose Rate Afterloader Procedures. For Performing Breast Brachytherapy with. The SAVI TM Applicator VariSource High Dose Rate Afterloader Procedures For Performing Breast Brachytherapy with The SAVI TM Applicator Contents I. CT Evaluation (~1 week prior to insertion) II. SAVI Applicator Implantation

More information

RESPIRATORY MOTION IN RADIATION THERAPY

RESPIRATORY MOTION IN RADIATION THERAPY RESPIRATORY MOTION IN RADIATION THERAPY Significant effect on targets in the chest and upper abdominal cavities Can cause difficulty in reproducing radiation planning protocols in radiation treatment sessions

More information

Review of fundamental photon dosimetry quantities

Review of fundamental photon dosimetry quantities Review of fundamental photon dosimetry quantities Narayan Sahoo Main sources of the materials included in this lecture notes are: (1) Radiation Oncology Physics: A Handbook for Teachers and Students Edited

More information

Proton Therapy QA Tools

Proton Therapy QA Tools Proton Therapy QA Tools A complete range of tools for absolute dosimetry, machine QA and patient plan verification Introduction During the past years more and more heavy particle therapy centers, especially

More information

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface Chapter 11 reatment Planning Single Beams Radiation Dosimetry I Outline Basic terminology Curved contour surface correction (bolus, compensators, wedges) Oblique beam incidence Correction for tissue inhomogeneities

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 Comparison of monitor unit calculations performed with a 3D computerized planning system and independent hand calculations: Results

More information

IMRT in clinical practice at the UMC-Utrecht. Utrecht

IMRT in clinical practice at the UMC-Utrecht. Utrecht 3 IMRT in clinical practice at the UMC-Utrecht Utrecht Clinical use of IMRT Improvement of local control without increased incidence of normal tissue complications + Dose escalation in the prostate Step-and

More information

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN 22 Turkish Journal of Cancer Volume 37, No.1, 2007 Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN SATISH PELAGADE 1, KALPANA

More information

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription Monitor Unit Calculations Part 1 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Return to our first patient 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt

More information

The dose distribution of medium energy electron boosts to the laryngectomy stoma

The dose distribution of medium energy electron boosts to the laryngectomy stoma JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 1, WINTER 2001 The dose distribution of medium energy electron boosts to the laryngectomy stoma Ellen D. Yorke* Memorial Sloan-Kettering Cancer

More information

4D Target definition for SBRT : how to deal with moving targets, a practical approach

4D Target definition for SBRT : how to deal with moving targets, a practical approach 4D Target definition for SBRT : how to deal with moving targets, a practical approach Dr. Alina Mihai, MD MSc Beacon Hospital, Dublin, Ireland From 3D to 4D Target Definitions (ICRU50/62) 3D CT 4D CT ICRU

More information

Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media

Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media N Tyagi 1, B H Curran, P L Roberson, J M Moran, E Acosta and B A Fraass Department

More information

Comparison of ionization chambers of various volumes for IMRT absolute dose verification

Comparison of ionization chambers of various volumes for IMRT absolute dose verification Comparison of ionization chambers of various volumes for IMRT absolute dose verification Leonid B. Leybovich, a) Anil Sethi, and Nesrin Dogan Department of Radiation Oncology, Loyola University Medical

More information

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 DISCLAIMER: This publication and associated spreadsheets and digital files are based on sources and

More information

Kill-painting of hypoxic tumors in charged particle therapy

Kill-painting of hypoxic tumors in charged particle therapy Supplementary Information: Kill-painting of hypoxic tumors in charged particle therapy Walter Tinganelli 1,2,, Marco Durante 1,3 *, Ryoichi Hirayama 2, Michael Krämer 1, Andreas Maier 1, Wilma Kraft-Weyrather

More information

Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System

Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System Amit Sawant 1, Sonja Dieterich 1, Michelle Svatos 2 and Paul Keall 1 1 Stanford University, Stanford, CA 2 Varian Medical Systems, Palo

More information

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas Current Implementation Status Current Implementation Status 1600 1400 TOTAL 1494 of 1623 ACTIVE INSTITUTIONS

More information

Planning Assignment (3 field rectum)

Planning Assignment (3 field rectum) Megan Sullivan Clinical Lab Rectum April 25, 2016 Planning Assignment (3 field rectum) Use a CT dataset of the pelvis. Create a CTV by contouring the rectum (start at the anus and stop at the turn where

More information

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator Journal of BUON 8: 55-59, 2003 2003 Zerbinis Medical Publications. Printed in Greece ORIGINAL ARTICLE Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

More information

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL FOR THE DEGREE MASTER OF ARTS BY CHARLES TRAVIS WEBB

More information

Zoubir Ouhib Lynn Cancer Institute

Zoubir Ouhib Lynn Cancer Institute Zoubir Ouhib Lynn Cancer Institute April 2015 Speaker for ELEKTA Brachytherapy using miniature X-ray sources Energy

More information

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08. Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas 2009 AAPM Summer School Educational Objectives Improve your understanding of how to implement TG-51

More information

High Speed Direct SAD Radiosurgery Beam Scanner

High Speed Direct SAD Radiosurgery Beam Scanner Open Access Original Article DOI: 10.7759/cureus.20 High Speed Direct SAD Radiosurgery Beam Scanner Walter Nikesch 1, James M. Hevezi 2, Irene Monterroso 3, Daniel Navarro 4, James G. Schwade 5 1. CyberKnife

More information

Implementation of respiratory-gated VMAT on a Versa HD linear accelerator

Implementation of respiratory-gated VMAT on a Versa HD linear accelerator Received: 21 November 2016 Revised: 7 April 2017 Accepted: 5 June 2017 DOI: 10.1002/acm2.12160 RADIATION ONCOLOGY PHYSICS Implementation of respiratory-gated VMAT on a Versa HD linear accelerator Jeffrey

More information

Exposure System Selection

Exposure System Selection Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others

More information

A gated deep inspiration breath-hold radiation therapy technique using a linear position transducer

A gated deep inspiration breath-hold radiation therapy technique using a linear position transducer JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 1, WINTER 2005 A gated deep inspiration breath-hold radiation therapy technique using a linear position transducer Svetlana I. Denissova, 1

More information

Dosimetric Calculations. Lonny Trestrail

Dosimetric Calculations. Lonny Trestrail Dosimetric Calculations Lonny Trestrail 20 October 2008 Objectives Dose Distribution Measurements PDD, OCR TAR, SAR, TPR, TMR, SPR, SMR Arc or Rotational Therapy Isodose Curves Point Dose Calculations

More information

A beam-matching concept for medical linear accelerators

A beam-matching concept for medical linear accelerators Acta Oncologica, 2009; 48: 192200 ORIGINAL ARTICLE A beam-matching concept for medical linear accelerators DAVID SJÖSTRÖM, ULF BJELKENGREN, WIVIANN OTTOSSON & CLAUS F. BEHRENS Copenhagen University Hospital,

More information

Physics and Quality Assurance

Physics and Quality Assurance Disclosure Physics and Quality Assurance Consultancy agreements with Nucletron, Hologic Mahta M. McKee, MS, DABR Department of Radiation Oncology Wake Forest University Health Sciences Agenda Current Treatment

More information

Abstract: Introduction:

Abstract: Introduction: Are photon and electron beam calibrations more consistent with TG-51 than with TG-21? Ramesh C. Tailor, William F. Hanson, Nathan Wells, and Geoffrey S. Ibbott. U.T. MD Anderson Cancer Center, Houston

More information

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007 The Effect of Immobilisation Devices on Radiotherapy Dose Distributions A thesis submitted in fulfilment of the requirements for the Degree of Master of Applied Science (Medical and Health Physics) By

More information

Commissioning and periodic tests of the Esteya electronic brachytherapy system

Commissioning and periodic tests of the Esteya electronic brachytherapy system Review paper Educational Activity Commissioning and periodic tests of the Esteya electronic brachytherapy system Cristian Candela-Juan, PhD 1, Yury Niatsetski, MSc 2, Zoubir Ouhib, PhD 3, Facundo Ballester,

More information

Radiotherapy Treatment Planning and Delivery

Radiotherapy Treatment Planning and Delivery Radiotherapy Treatment Planning and Delivery SCOPE 1: Study of Chemoradiotherapy in Oesophageal Cancer Plus or Minus Erbitux A randomised phase II/III multi-centre clinical trial of definitive chemo-radiation,

More information

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system G. Kunz, F. Hasenbalg, P. Pemler 1 1 Klinik für Radio-Onkologie und Nuklearmedizin, Stadtspital Triemli

More information

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting Monitor Unit Calculations Part 2 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Calculation of machine setting reference dose machine setting =, reference dose

More information

A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation

A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation Radiotherapy and Oncology 71 (2004) 235 241 Technical Note A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation Nadine Linthout*,

More information

Audiovisual (AV) biofeedback improves anatomical position management in breath-hold

Audiovisual (AV) biofeedback improves anatomical position management in breath-hold Audiovisual (AV) biofeedback improves anatomical position management in breath-hold (b) Taeho Kim, Sean Pollock, Danny Lee and Paul Keall Radiation Physics Laboratory, Sydney Medical School,The University

More information

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator)

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator) VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator) VT PLUS HF provides a special mode for evaluating the performance of high frequency ventilators while connected

More information

Inspired by the challenge to create a better solution for the thoracic aorta, RelayPlus was specifically designed to:

Inspired by the challenge to create a better solution for the thoracic aorta, RelayPlus was specifically designed to: A H E A D O F T H E C U R V E C R A F T E D F O R P R E C I S I O N Inspired by the challenge to create a better solution for the thoracic aorta, RelayPlus was specifically designed to: Respect the thoracic

More information

kvue IGRT Couch top One couch top, multiple treatment solutions For Varian, Elekta, Siemens, Proton and more The Best of Both Worlds

kvue IGRT Couch top One couch top, multiple treatment solutions For Varian, Elekta, Siemens, Proton and more The Best of Both Worlds TM kvue IGRT Couch top One couch top, multiple treatment solutions For Varian, Elekta, Siemens, Proton and more kvue The Best of Both Worlds kvue IGRT COUCH TOP One couch top, multiple treatment solutions

More information

3D Treatment Planning and verification with hand calculations

3D Treatment Planning and verification with hand calculations 3D Treatment Planning and verification with hand calculations Randy Holt, PhD, CEO Pacific Crest Medical Physics, Inc Chico, CA Dosimetric Calculations Calibration and Measurement Conditions Patient Conditions

More information

THE development of more advanced techniques in radiotherapy,

THE development of more advanced techniques in radiotherapy, 1 A Geant4-based simulation of an accelerator s head used for Intensity Modulated Radiation Therapy F. Foppiano, B. Mascialino, M.G. Pia, M. Piergentili. Abstract We present a Geant4-based simulation,

More information

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics. FRANZCR Examination Part I Radiation Oncology Radiotherapeutic Physics Candidate No.: The Royal Australian and New Zealand College of Radiologists FRANZCR Examination Part I Radiation Oncology Radiotherapeutic

More information

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components Introduction The CyberKnife SRS/SBRT System James M. Hevezi,, Ph.D., FACR CyberKnife Unit San Antonio, Texas CyberKnife Components Treatment Delivery (G3 Platform San Antonio) Robot (modified German auto

More information

COMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy

COMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy COMPASS 3D IMRT Verification in Patient Anatomy AK IMRT Bamberg April 2010 Dr. Lutz Müller Patient-specific Verification? CT Treatment unit Gantry =0 Treatmt unit Patient couch X-ray beam Patient couch

More information

Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin

Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin NCCAAPM Meeting Oct 11,2013 Larry DeWerd has a partial interest in Standard Imaging Talks from people who have been

More information

Original Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams

Original Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams Iranian Journal of Medical Physics Vol. 10, No. 3, Summer 2013, 147-155 Received: April 22, 2013; Accepted: July 21, 2013 Original Article Effective Point of Measurement in Cylindrical Ion Chamber for

More information

Processing SPECT VQ Studies

Processing SPECT VQ Studies Processing SPECT VQ Studies Joseph O'Brien Clinical Scientist joseph.o'brien@nhs.net City Hospital, Birmingham How do we get from this to this? Perf Vent Topics QC Checks Processing Display Raw Data Verification

More information

Modeling of Estimated Respiratory Waveform

Modeling of Estimated Respiratory Waveform Modeling of Estimated Respiratory Waveform ALEKSEI E. ZHDANOV, LEONID G. DOROSINSKY Institute of Radioelectronics and Information Technology - RTF Ural Federal University named after the first President

More information

Methods to model and predict the ViewRay treatment deliveries to aid patient scheduling and treatment planning

Methods to model and predict the ViewRay treatment deliveries to aid patient scheduling and treatment planning Washington University School of Medicine Digital Commons@Becker Open Access Publications 2016 Methods to model and predict the ViewRay treatment deliveries to aid patient scheduling and treatment planning

More information

1. Question Answer cgy / MU cgy / MU 2. Question Answer

1. Question Answer cgy / MU cgy / MU 2. Question Answer GS020113: Introduction to Medical Physics III: Therapy s to home work problem set assigned on 3/22/11 1. Question A patient is set up at 100 cm SSD on a 6 MVX machine. The dose rate at 10 cm in phantom

More information

A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions

A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions Wendel Dean Renner a) Math Resolutions, LLC, 5975 Gales Lane, Columbia,

More information

Math Review. Overview

Math Review. Overview Drawn by Steve Yan, CMD Math Review Shirley Ann Pinegar-Johnston MS RT(R)(T)CMD Overview Drawn by Steve Yan, CMD Divergence & Intensity DIVERGENCE X-rays travel in Straight but Divergent lines INTENSITY

More information

TG-51: Experience from 150 institutions, common errors, and helpful hints

TG-51: Experience from 150 institutions, common errors, and helpful hints TG-51: Experience from 150 institutions, common errors, and helpful hints Ramesh C. Tailor, Ph. D., William F. Hanson, Ph.D., and Geoffrey S. Ibbott, Ph.D. Department of Radiation Physics, University of

More information

COMPUTATIONAL FLOW MODEL OF WESTFALL'S LEADING TAB FLOW CONDITIONER AGM-09-R-08 Rev. B. By Kimbal A. Hall, PE

COMPUTATIONAL FLOW MODEL OF WESTFALL'S LEADING TAB FLOW CONDITIONER AGM-09-R-08 Rev. B. By Kimbal A. Hall, PE COMPUTATIONAL FLOW MODEL OF WESTFALL'S LEADING TAB FLOW CONDITIONER AGM-09-R-08 Rev. B By Kimbal A. Hall, PE Submitted to: WESTFALL MANUFACTURING COMPANY September 2009 ALDEN RESEARCH LABORATORY, INC.

More information

3100A Competency Exam

3100A Competency Exam NAME DATE (Circle the appropriate answer) 3100A Competency Exam 1. Of the following, which best describes the mechanics of ventilation used by the 3100A? a. Active inspiration with passive exhalation b.

More information

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams Peter R. Almond Brown Cancer Center, Louisville, Kentucky 40202 Peter J. Biggs Department of Radiation Oncology,

More information

Study of the IBF of Double/Triple THGEM

Study of the IBF of Double/Triple THGEM Study of the IBF of Double/Triple THGEM Qian Liu, Carlos, Hongbang LIU, Zhengyang Heng NED 2012, Mianyang 1 Outline PMT/Gas PMT THGEM introduction IBF of DTHGEM study Triple THGEM study Combined analysis

More information

OptonPro. High-power laser up to 7000 mw. Laser therapy natural healing with the power of light

OptonPro. High-power laser up to 7000 mw. Laser therapy natural healing with the power of light GB High-power laser up to 7000 mw Laser therapy natural healing with the power of light Application of high-power light for targeted penetration of high amounts of energy deep into the tissue. Initiates

More information

Optimization in high dose rate vaginal cylinder for vaginal cuff irradiation

Optimization in high dose rate vaginal cylinder for vaginal cuff irradiation Rep Pract Oncol Radiother, 2008; 13(1): 35-48 Original Paper Received: 2007.07.09 Accepted: 2007.12.13 Published: 2008.02.29 Authors Contribution: A Study Design B Data Collection C Statistical Analysis

More information

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61: Practical Implementation of TG-61: II. Guidelines for clinical implementation of TG-61 J.P. Seuntjens Medical Physics Unit McGill University, Montreal General Hospital Montréal, Canada jseuntjens@medphys.mcgill.ca

More information

Seismic Survey Designs for Converted Waves

Seismic Survey Designs for Converted Waves Seismic Survey Designs for Converted Waves James A. Musser* GMG/AXIS Inc., Denver, CO 1720 Red Cloud Road, Longmont, CO, 80501, USA jmusser@gmg.com ABSTRACT Designing converted wave 3D surveys is considerably

More information

Commissioning and quality assurance for the treatment delivery components of the AccuBoost system

Commissioning and quality assurance for the treatment delivery components of the AccuBoost system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 2, 2015 Commissioning and quality assurance for the treatment delivery components of the AccuBoost system Ileana Iftimia, 1,2a Mike Talmadge,

More information

Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation

Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation By Ruth Ann M. Good, CMD. R.T. (R)(T) Virginia Commonwealth University, Richmond,

More information

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13.1.1 TRAINING GOAL 13.1.1.1 Among the responsabilities of a medical

More information

Clinical Commissioning of a Pencil Beam Scanning Treatment Planning System for Proton Therapy

Clinical Commissioning of a Pencil Beam Scanning Treatment Planning System for Proton Therapy Clinical Commissioning of a Pencil Beam Scanning Treatment Planning System for Proton Therapy Jatinder Saini, MS 1 ; Ning Cao, PhD 2 ; Stephen R. Bowen, PhD 2,3 ; Miguel Herrera, BS 1 ; Daniel Nicewonger,

More information

Constancy checks of well-type ionization chambers with external-beam radiation units

Constancy checks of well-type ionization chambers with external-beam radiation units JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Constancy checks of well-type ionization chambers with external-beam radiation units Sara L. Hackett, 1a Benjamin Davis, 2 Andrew

More information

Surface buildup dose dependence on photon field delivery technique for IMRT

Surface buildup dose dependence on photon field delivery technique for IMRT Surface buildup dose dependence on photon field delivery technique for IMRT Shigeru Yokoyama, Peter L. Roberson, Dale W. Litzenberg, Jean M. Moran, and Benedick A. Fraass Department of Radiation Oncology,

More information

A Review of the Bed Roughness Variable in MIKE 21 FLOW MODEL FM, Hydrodynamic (HD) and Sediment Transport (ST) modules

A Review of the Bed Roughness Variable in MIKE 21 FLOW MODEL FM, Hydrodynamic (HD) and Sediment Transport (ST) modules A Review of the Bed Roughness Variable in MIKE 1 FLOW MODEL FM, Hydrodynamic (HD) and Sediment Transport (ST) modules by David Lambkin, University of Southampton, UK 1 Bed roughness is considered a primary

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310 www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Varian 310 310 LC/MS Manual System INCLUDES: 310-MS LC/MS triple quadrople

More information

Resist round robin ELETTRA

Resist round robin ELETTRA Resist round robin outgassing @ ELETTRA Trieste - Italy M. Bertolo, P. Lacovig, L.Rumiz, D. Giuressi, R. Sergo, A. Stolfa Experimental setup (1) 100 mm linear translator Sample manipulator Mask manipulator

More information

INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS

INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS A9744593 INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSLS Ladislav Czap, Georg Matscheko and Pedro Andreo osimetry and Medical Radiation Physics Section, ivision

More information

Investigation of Suction Process of Scroll Compressors

Investigation of Suction Process of Scroll Compressors Purdue University Purdue e-pubs International Compressor Engineering Conference School of Mechanical Engineering 2006 Investigation of Suction Process of Scroll Compressors Michael M. Cui Trane Jack Sauls

More information

Review of backscatter measurement in kilovoltage radiotherapy using novel detectors and reduction from lack of underlying scattering material

Review of backscatter measurement in kilovoltage radiotherapy using novel detectors and reduction from lack of underlying scattering material JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 6, 2013 Review of backscatter measurement in kilovoltage radiotherapy using novel detectors and reduction from lack of underlying scattering

More information

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World Chattanooga, Tennessee, April 17-21, 2005, on CD-ROM, American Nuclear Society, LaGrange Park, IL (2005) GEANT4 SIMULATION OF

More information

Variable Depth Bragg Peak Test Method

Variable Depth Bragg Peak Test Method Variable Depth Bragg Peak Test Method C. Foster (FCS), P. O Neill, B. Reddell (NASA/JSC), K. Nguyen, B. Jones (Jacobs Technology), N. Roche, S. Buchner (NRL) Acknowledgement: A. Rusek, M. Sivertz, and

More information

Sontek RiverSurveyor Test Plan Prepared by David S. Mueller, OSW February 20, 2004

Sontek RiverSurveyor Test Plan Prepared by David S. Mueller, OSW February 20, 2004 Sontek RiverSurveyor Test Plan Prepared by David S. Mueller, OSW February 20, 2004 INTRODUCTION Sontek/YSI has introduced new firmware and software for their RiverSurveyor product line. Firmware changes

More information

ADDENDUM IMPORTANT DOCUMENT INVITATION TO BID ADDENDUM. OPENING DATE & TIME: January 3, 2018 at 3:00PM

ADDENDUM IMPORTANT DOCUMENT INVITATION TO BID ADDENDUM. OPENING DATE & TIME: January 3, 2018 at 3:00PM Procurement Services ADDENDUM IMPORTANT DOCUMENT INVITATION TO BID ADDENDUM ITB NUMBER: 1708MC OPENING DATE & TIME: January 3, 2018 at 3:00PM ITB TITLE: X-Ray Photoelectron Spectroscopy System ADDENDUM

More information

Evaluation of six TPS algorithms in computing entrance and exit doses

Evaluation of six TPS algorithms in computing entrance and exit doses JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Evaluation of six TPS algorithms in computing entrance and exit doses Yun I. Tan, 1,2a Mohamed Metwaly, 3 Martin Glegg, 3 Shaun P.

More information

Commissioning of a new total body irradiation protocol

Commissioning of a new total body irradiation protocol University of Wollongong Research Online University of Wollongong Thesis Collection University of Wollongong Thesis Collections 2012 Commissioning of a new total body irradiation protocol Zoe Baldwin University

More information

Rotation Centers of the Equine Digit and their Use in Quantifying Conformation

Rotation Centers of the Equine Digit and their Use in Quantifying Conformation Presented at the Roth Kinematics Symposium, Stanford University, June 21, 2003. Rotation Centers of the Equine Digit and their Use in Quantifying Conformation John J. Craig, Phd ; Monique F. Craig, BSCE;

More information

Radiotherapy physics & Equipments

Radiotherapy physics & Equipments Radiotherapy physics & Equipments RAD 481 Lecture s Title: Beam Modification devices in Radiotherapy Dr. Mohammed Emam Vision :IMC aspires to be a leader in applied medical sciences, health care education

More information

Mechanical power and opening pressure. Fellowship training program Intensive Care Radboudumc, Nijmegen

Mechanical power and opening pressure. Fellowship training program Intensive Care Radboudumc, Nijmegen Mechanical power and opening pressure Fellowship training program Intensive Care Radboudumc, Nijmegen Mechanical power Energy applied to the lung: Ptp * V (Joule) Power = Energy per minute (J/min) Power

More information

Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens

Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens The characteristic ionizing radiation response is defined by the relationship between the stimulus

More information

Iranian Journal of Medical Physics

Iranian Journal of Medical Physics Iranian Journal of Medical Physics ijmp.mums.ac.ir Verification of Monitor Unit Calculations for Eclipse Treatment Planning System by in-house Developed Spreadsheet Athiyaman Mayivaganan 1*, Hemalatha

More information

USER MANUAL PROGRAM VERSION 1.0.1

USER MANUAL PROGRAM VERSION 1.0.1 R WWW..COM REF 92730-92731 USER MANUAL PROGRAM VERSION 1.0.1 STANDARD IMAGING, INC. 7601 Murphy Drive Middleton, WI 53562-2532 Dec / 2004 2004 Standard Imaging, Inc. DOC #80464-01 TEL 800.261.4446 TEL

More information

Can audio coached 4D CT emulate free breathing during the treatment course?

Can audio coached 4D CT emulate free breathing during the treatment course? Acta Oncologica ISSN: -6X (Print) 65-6X (Online) Journal homepage: http://www.tandfonline.com/loi/ionc Can audio coached D CT emulate free breathing during the treatment course? Gitte F. Persson, Ditte

More information

3D Inversion in GM-SYS 3D Modelling

3D Inversion in GM-SYS 3D Modelling 3D Inversion in GM-SYS 3D Modelling GM-SYS 3D provides a wide range of inversion options including inversion for both layer structure and physical properties for gravity and magnetic data. There is an

More information

Pressure Sensitive Paint (PSP) / Temperature Sensitive Paint (TSP) Part 1

Pressure Sensitive Paint (PSP) / Temperature Sensitive Paint (TSP) Part 1 AerE 545 class notes #12 Pressure Sensitive Paint (PSP) / Temperature Sensitive Paint (TSP) Part 1 Hui Hu Department of Aerospace Engineering, Iowa State University Ames, Iowa 50011, U.S.A Introduction

More information