Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale
|
|
- Cleopatra Newman
- 5 years ago
- Views:
Transcription
1 Commissioning an IMRT System for MLC Delivery Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale
2 Taking the broad view of commissioning
3 Commissioning elements Validating the dosimetry system Commissioning the delivery system Commissioning the planning/delivery systems for dosimetric accuracy Learning how to use the planning system for particular clinical problems Caveat: our experience is with Corvus + Varian
4 Step 0: the dosimetry system Decide on the system that will be used to measure IMRT dose distributions Determine the measurement uncertainty - Compare measurements of unmodulated fields with this system to ion chamber scans Needed to establish acceptance criteria
5 Delivery system tests Create test patterns independent of planning system - We use a text editor to create Varian files Need to check - Effect of rounded leaf ends (if applicable) - MLC positioning accuracy - MLC dynamic motion accuracy (if applicable) - MU linearity over range of interest...
6 Radiation field offset for rounded leaf ends For rounded MLC leaf ends, there is an offset between the light and radiation field edges: ~0.6 mm Important in IMRT Light Radiation
7 Important for IMRT because there are so many junctions within a field
8 Measuring the offset Irradiate a series of strips that create matchlines Use different values of offset Select offset value that gives best uniformity
9 Measuring the offset No offset 0.6 mm offset 1.0 mm offset Here, 0.6 appears best, i.e. subtract 0.6 mm from MLC settings What is really needed?? most uniform at junction? Integrated dose? Check planning system requirements.
10 Matchlines are very sensitive to positioning Offset 0.5 mm Offset 0.7 mm Offset 0.9 mm
11 MLC accuracy Tenths of millimeters are important for IMRT Use patterns of matchline strips for a sensitive test - Test at different gantry and collimator angles - Select subset for routine QA
12 DMLC tests For dynamic MLC motion, leaf speed and dose rate also need to be controlled Need to test for range of leaf speed and dose rates, as well as gantry and collimator angles Create patterns that move a 1 cm gap using all leaves at same or variable rates
13 Sweep all leaves at same rate Chamber reading at center should be proportional to MU Film should show uniformity
14 Sweep leaves at different rates e.g. Travel 5, 7, 9, 11, 13 cm in same MU Check relative dose
15 Special problem of matching accelerators In 99 Mayo Clinic Scottsdale had two matched Varian 2100C linacs, but IMRT doses differed by ~2.5% Needed to adjust MLC calibration on one machine by 0.13 mm (Leaf gap error parameter)
16 Delivery system QA Sweep a 5mm gap across a 10 cm span with Farmer chamber at center Take ratio to 10x10 open field 2.5% for 0.2 mm change Ratio to standard 5mm gap Gap error
17 Routine QA checks for positioning accuracy Matchline films 1 mm gap films Check at different gantry, collimator angles Decide QA frequency
18 Mayo monthly includes matchline films at 5 combinations of gantry, collimator
19 Check visually. Patterns include programmed deviations 0.5 mm closed 0.5 mm open orientation 0.5 mm open 0.5 mm closed
20 Mayo monthly includes one profile analyzed quantitatively
21 Mayo monthly includes ratio of swept 5 mm gap to 10x10 open Similar test used with daily QA device
22 Summary for the delivery system Commission the delivery system independently of the planning system Establish QA tests to monitor performance
23 on to the planning system
24 Planning system: Dosimetric accuracy For IMRT, the MLC leaves move through the area of interest Final distribution is created by summing many beamlets New things become important - Leakage through MLC leaves - Penumbra defined by MLC leaves - Small fields
25 MLC leakage - measure average value Leakage through leaf (~2%) Between neighboring leaves (~5%) Measure using a pattern that fully closes all leaves - careful not to be under carriage or jaw
26 Penumbra Measure with film, diode, or microchamber, conventional scanning chamber too wide Subtle effects make a difference in IMRT Beam model based on penumbra measured with chamber Beam model based on penumbra measured with film
27 Dosimetric validation (1) Start with the basics Define simple open fields irradiating a simple phantom - check calculated output, PDD, profiles against standard measurements - (just like commissioning any planning system) And shaped static fields also
28 6MV 21EX Profiles 10x Relative dose Meas D=5 Peregrine D=5 Corvus D= Off axis
29 6MV 21EX Profiles 10x30 alternating Relative dose Meas D=5 Peregrine D=5 Corvus D= Off axis
30 Dosimetric validation (2) Define simply-modulated fields with wide regions that can be measured with chamber and film (or equivalent) Check: - Leakage/transmission - Basic modulation - Penumbra - Out of field dose
31 Leak100 Leak0 patterns Corvus/Meas Peregrine/Meas Leak Leak Leak Leak Leak
32 Ridge and Trench patterns Useful for deciding minimum segment width 100% centers, 10% wings 10% centers, 100% wings Incline patterns 50% centers, 100% -10% wings
33 Ridge40 Ridge Dose (cgy) Corvus Peregrine Chamber Film Dose (cgy) Corvus Peregrine Chamber Film cm off axis cm off axis Trench40 Trench Dose (cgy) Corvus Peregrine Chamber Film Dose (cgy) Corvus Peregrine Chamber Film cm off axis cm off axis
34 5mm strips In c lin e Dose (cgy) Corvus Peregrine Chamber Film cm off axis
35 10mm strips Incline Dose (cgy) Corvus Peregrine Chamber Film cm off axis
36 Band100 Band0 patterns Bands Dose (cgy) Corvus Peregrine Chamber Film Film Repeat Chamber % cm off axis
37 Finally, highly modulated single fields Random array
38 What to do about differences? May need to adjust the beam model May need to live with it - That is, take known deficiencies into account when evaluating plans - Very important to know about it, especially for critical structures
39 And once you are over the simple stuff
40 Dosimetric validation (3) Define targets and structures in a simple phantom: mock clinical situations Create inverse plan Measure doses with chamber, film, Evaluate dose/mu in low gradient region Evaluate isodoses on many planes
41 Mock prostate
42
43 Colors: film, Gray: plan
44 Mock Head/Neck
45
46 Colors: film, Gray: plan
47 Mock spine
48 Critical structure plane
49 Target plane
50 Colors: film, Gray: plan
51 Another question What is the role of quantitative assessment of individual IMRT fields vs. composite doses?
52 Some problems show up more clearly on individual fields than composite distributions e.g. Dynamic delivery with Corvus/Varian Step/Shoot Dynamic
53 Dosimetric validation (4) Test calculations in the presence of inhomogeneities Measured/Corvus: 0.985
54 Dose/MU at 5 cm below 12 cm inhomogeneity Measured/Corvus Bone Air Lung Water 0.978
55 LUNG Meas d=1.25 Meas d=5.0 Meas d=10.0 Meas d=20.0 Corvus d=1.25 Corvus d=5.0 Corvus d=10.0 Corvus d=20.0 Peregrine d=1.25 Peregrine d=5.0 Peregrine d=10.0 Peregrine d=
56 Dosimetric evaluation: Acceptance criteria What kind of agreement to expect in high dose, low gradient regions? - ~2-4% is reasonable (chamber) What kind of agreement to expect for isodoses? - Much harder to specify - ~2-4 mm for 50% - 90% lines
57 How to be more quantitative? Acceptance criteria should be stated statistically, such as: within the region of interest, 95% of the calculation points should agree with the measurement to within 5%, which combines the acceptable degree of agreement with reality (3%) with the two-sigma uncertainty in the measurement technique (4%).
58 Questions What is the region of interest? What is the acceptable degree of agreement with reality? What is the uncertainty in the measurement technique? What percentage of the points should agree to the specified tolerance?
59 Questions Should the agreement be expressed as a percent of the local dose, prescription dose, or some other dose? For the points that do not agree to that tolerance, is there an outer limit of acceptability? Should distance to agreement be incorporated?
60 Dosimetric validation - Summary Know your measurement uncertainties Start with simple situations that can be more easily analyzed Design single beam tests that focus on specific parameters, e.g. - Transmission - Small or narrow fields - Penumbra and dose outside field Finish with multiple fields and mock clinical situations
61 Welcome to IMRT there is a way out
TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009
TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 DISCLAIMER: This publication and associated spreadsheets and digital files are based on sources and
More informationCLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions
CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions Rebecca M. Howell, Ph.D., D.A.B.R, The University of Texas at M.D. Anderson Cancer Center Overview 2 Commissioning
More informationCHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS
47 CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 4.1 INTRODUCTION Advanced treatment techniques use optimized radiation beam intensities to conform dose distribution to the
More informationCommissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN
22 Turkish Journal of Cancer Volume 37, No.1, 2007 Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN SATISH PELAGADE 1, KALPANA
More information13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.
13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13.1.1 TRAINING GOAL 13.1.1.1 Among the responsabilities of a medical
More informationSnapShot IMRT with compensators and FFF beams
SnapShot IMRT with compensators and FFF beams Vladimir Feygelman, PhD (1) Moffitt Cancer Center, Tampa, FL, USA Disclosure VF has a sponsored research agreement with.decimal. A short history of radiotherapy
More informationRelative Dosimetry. Photons
Relative Dosimetry Photons What you need to measure! Required Data (Photon) Central Axis Percent Depth Dose Tissue Maximum Ratio Scatter Maximum Ratio Output Factors S c & S cp! S p Beam profiles Wedge
More informationClinical Implementation of Volumetric Modulated Arc Therapy
Clinical Implementation of Volumetric Modulated Arc Therapy UT M.D. Anderson Cancer Center Ramaswamy Sadagopan, Rebecca M. Howell, Weiliang Du and Peter Balter Definition 2 Intensity Modulated Arc therapy
More informationMonitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription
Monitor Unit Calculations Part 1 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Return to our first patient 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt
More informationReview of fundamental photon dosimetry quantities
Review of fundamental photon dosimetry quantities Narayan Sahoo Main sources of the materials included in this lecture notes are: (1) Radiation Oncology Physics: A Handbook for Teachers and Students Edited
More informationMonitor Unit Verification for Small Fields
Monitor Unit Verification for Small Fields Patrick Higgins, Ph.D University of Minnesota Department of Radiation Oncology October 10, 2013 The Issues: How do we verify the monitor units calculated by the
More informationValidation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.
International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2017, 6, 57-72 http://www.scirp.org/journal/ijmpcero ISSN Online: 2168-5444 ISSN Print: 2168-5436 Validation of Treatment
More informationTHE development of more advanced techniques in radiotherapy,
1 A Geant4-based simulation of an accelerator s head used for Intensity Modulated Radiation Therapy F. Foppiano, B. Mascialino, M.G. Pia, M. Piergentili. Abstract We present a Geant4-based simulation,
More informationIMRT in clinical practice at the UMC-Utrecht. Utrecht
3 IMRT in clinical practice at the UMC-Utrecht Utrecht Clinical use of IMRT Improvement of local control without increased incidence of normal tissue complications + Dose escalation in the prostate Step-and
More informationCommissioning of Elekta 6MV FFF Versa HD and Pinnacle
Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Poster No.: R-0044 Congress: Type: Authors: Keywords: DOI: 2014 CSM Scientific Exhibit L. Bendall, I. Patel, N. McGrath, C. Rowbottom; MANCHESTER/
More informationANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL
ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL FOR THE DEGREE MASTER OF ARTS BY CHARLES TRAVIS WEBB
More informationConstancy checks of well-type ionization chambers with external-beam radiation units
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Constancy checks of well-type ionization chambers with external-beam radiation units Sara L. Hackett, 1a Benjamin Davis, 2 Andrew
More informationOutline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface
Chapter 11 reatment Planning Single Beams Radiation Dosimetry I Outline Basic terminology Curved contour surface correction (bolus, compensators, wedges) Oblique beam incidence Correction for tissue inhomogeneities
More informationComparison of ionization chambers of various volumes for IMRT absolute dose verification
Comparison of ionization chambers of various volumes for IMRT absolute dose verification Leonid B. Leybovich, a) Anil Sethi, and Nesrin Dogan Department of Radiation Oncology, Loyola University Medical
More informationGEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY
The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World Chattanooga, Tennessee, April 17-21, 2005, on CD-ROM, American Nuclear Society, LaGrange Park, IL (2005) GEANT4 SIMULATION OF
More informationMonitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting
Monitor Unit Calculations Part 2 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Calculation of machine setting reference dose machine setting =, reference dose
More informationMath Review. Overview
Drawn by Steve Yan, CMD Math Review Shirley Ann Pinegar-Johnston MS RT(R)(T)CMD Overview Drawn by Steve Yan, CMD Divergence & Intensity DIVERGENCE X-rays travel in Straight but Divergent lines INTENSITY
More informationThe MC commissioning of CyberKnife with MLC (Tips and Tricks)
The MC commissioning of CyberKnife with MLC (Tips and Tricks) Alain Guemnie Tafo, PhD UPMC Pinnacle, Harrisburg, PA Institute Gustave Roussy, Villejuif, France Disclaimer & Disclosure The views expressed
More informationDosimetric Calculations. Lonny Trestrail
Dosimetric Calculations Lonny Trestrail 20 October 2008 Objectives Dose Distribution Measurements PDD, OCR TAR, SAR, TPR, TMR, SPR, SMR Arc or Rotational Therapy Isodose Curves Point Dose Calculations
More informationIntroduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components
Introduction The CyberKnife SRS/SBRT System James M. Hevezi,, Ph.D., FACR CyberKnife Unit San Antonio, Texas CyberKnife Components Treatment Delivery (G3 Platform San Antonio) Robot (modified German auto
More informationTG219: IT'S USE, STRENGTHS AND WEAKNESSES
1 CONFLICT OF INTEREST TG219: IT'S USE, STRENGTHS AND WEAKNESSES SOTIRI STATHAKIS, PHD, DABR None ROSTER STATEMENT OF THE PROBLEM Timothy C. Zhu, University of Pennsylvania, Philadelphia, PA (Co- Chair)
More informationAbsorption measurements for a carbon fiber couch top and its modelling in a treatment planning system
Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system G. Kunz, F. Hasenbalg, P. Pemler 1 1 Klinik für Radio-Onkologie und Nuklearmedizin, Stadtspital Triemli
More informationTreatment plan complexity metrics for predicting IMRT pretreatment. assurance results. Scott Crowe, QUT
Treatment plan complexity metrics for predicting IMRT pretreatment quality assurance results Scott Crowe, QUT Pre-treatment QA Verification of fluence delivery to ensure that dose can be delivered accurately
More informationA beam-matching concept for medical linear accelerators
Acta Oncologica, 2009; 48: 192200 ORIGINAL ARTICLE A beam-matching concept for medical linear accelerators DAVID SJÖSTRÖM, ULF BJELKENGREN, WIVIANN OTTOSSON & CLAUS F. BEHRENS Copenhagen University Hospital,
More informationOnly 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.
Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas 2009 AAPM Summer School Educational Objectives Improve your understanding of how to implement TG-51
More informationA dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions
A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions Wendel Dean Renner a) Math Resolutions, LLC, 5975 Gales Lane, Columbia,
More informationField size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator
Journal of BUON 8: 55-59, 2003 2003 Zerbinis Medical Publications. Printed in Greece ORIGINAL ARTICLE Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator
More informationCOMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy
COMPASS 3D IMRT Verification in Patient Anatomy AK IMRT Bamberg April 2010 Dr. Lutz Müller Patient-specific Verification? CT Treatment unit Gantry =0 Treatmt unit Patient couch X-ray beam Patient couch
More informationClinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas
Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas Current Implementation Status Current Implementation Status 1600 1400 TOTAL 1494 of 1623 ACTIVE INSTITUTIONS
More informationDynamic wedges dosimetry and quality control
Original Paper Received: 2005.06.07 Accepted: 2006.02.28 Published: 2006.04.28 Authors Contribution: A Study Design B Data Collection C Statistical Analysis D Data Interpretation E Manuscript Preparation
More information1. Question Answer cgy / MU cgy / MU 2. Question Answer
GS020113: Introduction to Medical Physics III: Therapy s to home work problem set assigned on 3/22/11 1. Question A patient is set up at 100 cm SSD on a 6 MVX machine. The dose rate at 10 cm in phantom
More informationZoubir Ouhib Lynn Cancer Institute
Zoubir Ouhib Lynn Cancer Institute April 2015 Speaker for ELEKTA Brachytherapy using miniature X-ray sources Energy
More informationIntroduction. Materials and Methods
STANDARD WEDGE AND TRAY TRANSMISSION VALUES FOR VARIAN, SEIMENS, ELEKTA/PHILIPS ACCELERATORS; A QUALITY ASSURANCE TOOL David S. Followill, Nadia Hernandez, and William F. Hanson Introduction The Radiological
More informationExperimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media
Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media N Tyagi 1, B H Curran, P L Roberson, J M Moran, E Acosta and B A Fraass Department
More informationThe Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.
FRANZCR Examination Part I Radiation Oncology Radiotherapeutic Physics Candidate No.: The Royal Australian and New Zealand College of Radiologists FRANZCR Examination Part I Radiation Oncology Radiotherapeutic
More informationA simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation
Radiotherapy and Oncology 71 (2004) 235 241 Technical Note A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation Nadine Linthout*,
More informationClinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry
Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Tom Jordan Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE
More informationCommissioning and periodic tests of the Esteya electronic brachytherapy system
Review paper Educational Activity Commissioning and periodic tests of the Esteya electronic brachytherapy system Cristian Candela-Juan, PhD 1, Yury Niatsetski, MSc 2, Zoubir Ouhib, PhD 3, Facundo Ballester,
More informationScientific Journal of Nuclear Medicine & Radiation Therapy
Scientific Journal of Nuclear Medicine & Radiation Therapy Research Article Radiotherapy Dose Calculation Program, for Monitoring and Commissioning Treatment Planning Systems - Ioannis Vamvakas*, Stella
More informationInvestigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy
University of South Florida Scholar Commons Graduate Theses and Dissertations Graduate School 7-14-2010 Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy
More informationVMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 3, 2016 VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests Michael P. Barnes, 1,2a Pejman Rowshanfarzad,
More informationDisclosures. What Is So Hard? End of ITV: Gating is the Best ITV Killer 8/3/2016. Daniel A. Low, Ph.D. UCLA. Varian Grant Siemens Grant Accuray Grant
Real Timefullness 8/3/2016 End of ITV: Gating is the Best ITV Killer Daniel A. Low, Ph.D. UCLA Varian Grant Siemens Grant Accuray Grant Disclosures What Is So Hard? Light Ultrasound MRI X Ray Planar X
More informationLarry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin
Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin NCCAAPM Meeting Oct 11,2013 Larry DeWerd has a partial interest in Standard Imaging Talks from people who have been
More informationSurface buildup dose dependence on photon field delivery technique for IMRT
Surface buildup dose dependence on photon field delivery technique for IMRT Shigeru Yokoyama, Peter L. Roberson, Dale W. Litzenberg, Jean M. Moran, and Benedick A. Fraass Department of Radiation Oncology,
More informationBasics of Proton Therapy. Proton Treatment Planning and Beam Optimization
Basics of Proton Therapy Proton Treatment Planning and Beam Optimization SAM Educational Session, WE D BRB 2 Mark Pankuch, PhD Northwestern Medicine Chicago Proton Center Today s objectives Review the
More informationFailure Mode and Effect Analysis (FMEA) for a DMLC Tracking System
Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System Amit Sawant 1, Sonja Dieterich 1, Michelle Svatos 2 and Paul Keall 1 1 Stanford University, Stanford, CA 2 Varian Medical Systems, Palo
More informationElectronic Oncology Systems the Management and the Safety Measures. Hansen Chen, Director, TDSI
Electronic Oncology Systems the Management and the Safety Measures Hansen Chen, Director, TDSI Radiation Oncology Department People Place Equipments Procedures Patients Hansen Chen 2 Radiation Oncology
More informationQuality control for Intensity-modulated radiation therapy
Schweizerische Gesellschaft für Strahlenbiologie und Medizinische Physik Société Suisse de Radiobiologie et de Physique Médicale Società Svizzera di Radiobiologia e di Fisica Medica Swiss Society for Radiobiology
More informationBeam modeling and VMAT performance with the Agility 160-leaf multileaf collimator
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 2, 2013 Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator James L. Bedford, a Michael D. R. Thomas and Gregory
More informationImplementation of respiratory-gated VMAT on a Versa HD linear accelerator
Received: 21 November 2016 Revised: 7 April 2017 Accepted: 5 June 2017 DOI: 10.1002/acm2.12160 RADIATION ONCOLOGY PHYSICS Implementation of respiratory-gated VMAT on a Versa HD linear accelerator Jeffrey
More informationJOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 Comparison of monitor unit calculations performed with a 3D computerized planning system and independent hand calculations: Results
More informationPractical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11
MV Photon Dosimetry in the Clinic This document discusses both the scientific methods and the system of work required to calibrate a field instrument expanding upon the practical details of both.. Introduction
More informationA clinical comparison and analysis between conventional MLC based and solid compensator based IMRT treatment techniques
The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2009 A clinical comparison and analysis between conventional MLC based and solid compensator based IMRT treatment
More informationThe Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007
The Effect of Immobilisation Devices on Radiotherapy Dose Distributions A thesis submitted in fulfilment of the requirements for the Degree of Master of Applied Science (Medical and Health Physics) By
More informationTG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:
Practical Implementation of TG-61: II. Guidelines for clinical implementation of TG-61 J.P. Seuntjens Medical Physics Unit McGill University, Montreal General Hospital Montréal, Canada jseuntjens@medphys.mcgill.ca
More informationGuideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation
Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation This intercomparison will be carried out under the lead of KRISS with the cooperation of ARPANSA. This comparison has been approved
More informationTEST REPORT BEAMSCAN WITH HALCYON
TEST REPORT BEAMSCAN WITH HALCYON UZ Leuven, November 20 th, 2017 + = Ideal BEAMSCAN tank dimensions to fit perfectly into the Halcyon bore Fast tank positioning using the Halcyon lasers, no electronics
More informationHigh Speed Direct SAD Radiosurgery Beam Scanner
Open Access Original Article DOI: 10.7759/cureus.20 High Speed Direct SAD Radiosurgery Beam Scanner Walter Nikesch 1, James M. Hevezi 2, Irene Monterroso 3, Daniel Navarro 4, James G. Schwade 5 1. CyberKnife
More informationProton Therapy QA Tools
Proton Therapy QA Tools A complete range of tools for absolute dosimetry, machine QA and patient plan verification Introduction During the past years more and more heavy particle therapy centers, especially
More informationMethods to model and predict the ViewRay treatment deliveries to aid patient scheduling and treatment planning
Washington University School of Medicine Digital Commons@Becker Open Access Publications 2016 Methods to model and predict the ViewRay treatment deliveries to aid patient scheduling and treatment planning
More informationCOMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC
COMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC by TAN ZI XIANG Thesis submitted in fulfillment of the requirements for the degree of Master of Science (Medical Physics) February 2015 ACKNOWLEDGEMENT
More informationTG-51: Experience from 150 institutions, common errors, and helpful hints
TG-51: Experience from 150 institutions, common errors, and helpful hints Ramesh C. Tailor, Ph. D., William F. Hanson, Ph.D., and Geoffrey S. Ibbott, Ph.D. Department of Radiation Physics, University of
More informationRadiotherapy physics & Equipments
Radiotherapy physics & Equipments RAD 481 Lecture s Title: Beam Modification devices in Radiotherapy Dr. Mohammed Emam Vision :IMC aspires to be a leader in applied medical sciences, health care education
More informationExradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers
EXRADIN Ion Chambers Exradin Ion Chambers Exradin (EXacting RADiation INstrumentation) Ion Chambers have been built for over 33 years, are recognized by top research institutes and standards laboratories,
More informationATION TITLE. Survey QC, Decision Making, and a Modest Proposal for Error Models. Marc Willerth, MagVAR
1 a Modest Proposal for Error Models ATION TITLE Marc Willerth, MagVAR Speaker Information Marc Willerth VP of Survey Technologies April 11, 2018 MagVAR 2 Speaker Bio Marc Willerth Magnetic Variation Services,
More informationThe dose distribution of medium energy electron boosts to the laryngectomy stoma
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 1, WINTER 2001 The dose distribution of medium energy electron boosts to the laryngectomy stoma Ellen D. Yorke* Memorial Sloan-Kettering Cancer
More informationLab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow
Georgia Institute of Technology School of Earth and Atmospheric Sciences EAS 4641, Spring 2008 Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow Purpose of Lab 3: 1) To gain a
More information3D Treatment Planning and verification with hand calculations
3D Treatment Planning and verification with hand calculations Randy Holt, PhD, CEO Pacific Crest Medical Physics, Inc Chico, CA Dosimetric Calculations Calibration and Measurement Conditions Patient Conditions
More informationGerald D. Anderson. Education Technical Specialist
Gerald D. Anderson Education Technical Specialist The factors which influence selection of equipment for a liquid level control loop interact significantly. Analyses of these factors and their interactions
More informationPlanning Assignment (3 field rectum)
Megan Sullivan Clinical Lab Rectum April 25, 2016 Planning Assignment (3 field rectum) Use a CT dataset of the pelvis. Create a CTV by contouring the rectum (start at the anus and stop at the turn where
More informationCommissioning of a new total body irradiation protocol
University of Wollongong Research Online University of Wollongong Thesis Collection University of Wollongong Thesis Collections 2012 Commissioning of a new total body irradiation protocol Zoe Baldwin University
More informationStadium Project - DP-70 Atlanta, GA, 9/30/2014
Page 1 of 13 The enclosed report contains the data and analysis summary for the SoniCaliper shaft caliper, performed at Stadium Project (DP-70), Atlanta, GA on Tuesday, September 30, 2014 by Chris Kohlhof.
More informationBadan Jadrowych Nuclear Research Institute. Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P Differential Recombination Chamber
Badan Jadrowych Nuclear Research Institute Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P00100517 Differential Recombination Chamber by M. Zel'chinskij K. Zharnovetskij Warsaw, June 1966 Translated
More informationNew Highly Productive Phased Array Ultrasonic Testing Machine for Aluminium Plates for Aircraft Applications
19 th World Conference on Non-Destructive Testing 2016 New Highly Productive Phased Array Ultrasonic Testing Machine for Aluminium Plates for Aircraft Applications Christoph HENKEL 1, Markus SPERL 1, Walter
More informationOUTPUT at dmax in tissue 8x8 20x20 At 100cm SSD At 100cm SAD
RAPHEX 2001 T1. All of the following are true regarding Percentage Depth Dose (PDD) except: A. Increases with increasing energy. B. Depends on field size. C. Is the dose at depth expressed as a % of the
More informationEvaluation of six TPS algorithms in computing entrance and exit doses
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Evaluation of six TPS algorithms in computing entrance and exit doses Yun I. Tan, 1,2a Mohamed Metwaly, 3 Martin Glegg, 3 Shaun P.
More informationAnemometry. Anemometry. Wind Conventions and Characteristics. Anemometry. Wind Variability. Anemometry. Function of an anemometer:
Anemometry Anemometry Function of an anemometer: Measure some or all of the components of the wind vector In homogeneous terrain, vertical component is small express wind as -D horizontal vector For some
More information(Lab Interface BLM) Acceleration
Purpose In this activity, you will study the concepts of acceleration and velocity. To carry out this investigation, you will use a motion sensor and a cart on a track (or a ball on a track, if a cart
More informationAN31E Application Note
Balancing Theory Aim of balancing How an unbalance evolves An unbalance exists when the principle mass axis of a rotating body, the so-called axis of inertia, does not coincide with the rotational axis.
More informationCHAPTER-2. Wedge filters are the most important beam modifiers used in
CHAPTER-2 2. WEDGE FLTERS 2.1. NTRODUCTON Wedge filters are the most important beam modifiers used in radiotherapy which are placed in the path of photon beams to modify their isodose diistributions. The
More informationITTC Recommended Procedures and Guidelines
Page 1 of 6 Table of Contents 1. PURPOSE...2 2. PARAMETERS...2 2.1 General Considerations...2 3 DESCRIPTION OF PROCEDURE...2 3.1 Model Design and Construction...2 3.2 Measurements...3 3.5 Execution of
More informationintended velocity ( u k arm movements
Fig. A Complete Brain-Machine Interface B Human Subjects Closed-Loop Simulator ensemble action potentials (n k ) ensemble action potentials (n k ) primary motor cortex simulated primary motor cortex neuroprosthetic
More informationTutorial for the. Total Vertical Uncertainty Analysis Tool in NaviModel3
Tutorial for the Total Vertical Uncertainty Analysis Tool in NaviModel3 May, 2011 1. Introduction The Total Vertical Uncertainty Analysis Tool in NaviModel3 has been designed to facilitate a determination
More informationBreast Treatment System REFERENCE MANUAL
Breast Treatment System REFERENCE MANUAL This manual may not be reproduced in whole or in part, by mimeograph or any other means, without the written permission of Diacor, Inc. The MammoRx Breast Treatment
More informationCommercial Practice Test Method Internal Vapor Analysis of Hermetic Devices
Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Oneida Research Services, Inc. 8282 Halsey Road Whitesboro, NY 13492 Phone: (315) 736-5480 FAX: (315) 736-9321 1.0 Purpose The
More informationHardware Triaxial and Consolidation Testing Systems Pressure Measurement and Control
25 GDS Helpsheet Hardware Triaxial and Consolidation Testing Systems Pressure Measurement and Control World Leaders in Computer Controlled Testing Systems for Geotechnical Engineers and Geologists 1. Digital
More informationHarry M. Cullings Radiation Effects Research Foundation Hiroshima and Nagasaki, Japan. EURADOS Winter School Milan, Italy February 11,
Harry M. Cullings Radiation Effects Research Foundation Hiroshima and Nagasaki, Japan EURADOS Winter School Milan, Italy February 11, 2016 1 Selection of Life Span Study cohort Available data Calculation
More informationAAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams
AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams Peter R. Almond Brown Cancer Center, Louisville, Kentucky 40202 Peter J. Biggs Department of Radiation Oncology,
More informationCommissioning and quality assurance for the treatment delivery components of the AccuBoost system
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 2, 2015 Commissioning and quality assurance for the treatment delivery components of the AccuBoost system Ileana Iftimia, 1,2a Mike Talmadge,
More informationCycling Volume Estimation Methods for Safety Analysis
Cycling Volume Estimation Methods for Safety Analysis XI ICTCT extra Workshop in Vancouver, Canada Session: Methods and Simulation Date: March, 01 The Highway Safety Manual (HSM) documents many safety
More informationEvaluation of High Lift System with Oscillatory Blowing in 2.5D Configuration
Evaluation of High Lift System with Oscillatory Blowing in 2.5D Configuration Cǎtǎlin NAE, Mihai-Victor PRICOP Corresponding author INCAS - National Institute for Aerospace Research Elie Carafoli Bdul
More informationTreatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation
Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation By Ruth Ann M. Good, CMD. R.T. (R)(T) Virginia Commonwealth University, Richmond,
More informationQueue analysis for the toll station of the Öresund fixed link. Pontus Matstoms *
Queue analysis for the toll station of the Öresund fixed link Pontus Matstoms * Abstract A new simulation model for queue and capacity analysis of a toll station is presented. The model and its software
More informationClinical Commissioning of a Pencil Beam Scanning Treatment Planning System for Proton Therapy
Clinical Commissioning of a Pencil Beam Scanning Treatment Planning System for Proton Therapy Jatinder Saini, MS 1 ; Ning Cao, PhD 2 ; Stephen R. Bowen, PhD 2,3 ; Miguel Herrera, BS 1 ; Daniel Nicewonger,
More informationkvue IGRT Couch top One couch top, multiple treatment solutions For Varian, Elekta, Siemens, Proton and more The Best of Both Worlds
TM kvue IGRT Couch top One couch top, multiple treatment solutions For Varian, Elekta, Siemens, Proton and more kvue The Best of Both Worlds kvue IGRT COUCH TOP One couch top, multiple treatment solutions
More information