1 Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin NCCAAPM Meeting Oct 11,2013

2 Larry DeWerd has a partial interest in Standard Imaging

3 Talks from people who have been in the field for awhile tend to be historical or philosophical My attempt at philosophical and hopefully informational material.

4

5 I don t have enough time Why worry it will all work out. I don t have to think about it it is late. Didn t the manufacturer tell me it was okay.

6 Some Manufacturers harp on the time saved, etc but their instrument is not as precise and can result in errors. For example, some digital sampling electrometers (depending on their sampling algorithm) may miss some signal from pulses of the accelerator. The physicist should ask.

7 Technicians use numbers, Medical physicists use understanding. Do it right good enough is not good enough

8 Transfer of Knowledge is Important. ADCLs transfer latest knowledge Â\y \ {täx áxxç yâüà{xü à{tç Éà{xÜá? à á uç áàtçw Çz âñéç à{x á{éâäwxüá Éy z tçàáê f Ü \áttv axãàéç? DIJI f Ü \áttv

9

10 Just doing a measurement and applying ADCL calibration does not mean that it is right You need to understand the equipment and see through the measurement. Numbers no matter how precise cannot by themselves imply anything. Precision and accuracy of dose measurements and reporting of the measurement details should be sufficient to allow the work to be interpreted and repeated and to allow valid comparisons to be made, both in the same laboratory and by other laboratories.

11 Ion chambers require calibration for the most precise measurements ADCLs willing to discuss measurements and methodology ADCL discuss the operation of instrumentation.

12 1 0.9 primary standard measurement frequency of result clinical measurement conventional true value a.u.

13 NBS petitioned AAPM to create Regional Calibration Laboratories in In 1983 called ADCLs. NBS acknowledges traceability to primary standards (Proficiency tests) Agreement for Proficiency tests and round robins < 0.5% UWADCL founded 1981 by LAD

14 International & National Standards Primary Calibration Secondary Standards ADCL Hospital or Clinical Standards Secondary Calibration Tertiary Calibration Operational Standards

15 AAPM in conjunction with NIST set up secondary laboratories for medical ionizing radiation calibrations - For therapy applications the precision is within 0.5%. Traceability of radiation dose is dependent on standards that are the same throughout the world at primary labs.

16 All calibrations (especially for medical applications) should have a traceability train to a standard This provides confidence in the dose given because it was measured and can be traced back to a national standard at NIST However, the SSDL network is not the equivalent to the ADCL network

17 The ADCLs have proven track records of providing precise calibrations of equipment for Therapy and Diagnostic applications Safety and treatment delivery or imaging is improved because of ADCL calibration and research.

18 N u m b e r Mean of ADCLs/ADCL

19 These tests have been in place over 20 years NIST and ADCLs agree within 0.5% for Cobalt-60 beams NIST and ADCLs agree less than 2.0 % for x-ray beams between 20 kvp and 250 kvp

20 Maintenance of accuracy and precision is very important for Medical Applications Knowledge of characteristics of chambers is very important Medical Applications rely heavily upon precise traceable calibrations.

21 Calibration of chambers should be done across the energy range of use. Bracket your point of interest. Generally there is a reference energy in use that is the minimum energy point used.

22 Because of variation in response with energy for chambers a range of calibration is necessary. The calibration should be done under correct criteria e.g. TG 61 There is no protocol using an absorbed dose to water calibration for x-ray beams. An SSDL offers absorbed dose in water but no protocol.

23 Within + 1% for Mammography range: 24 kvp to 35 kvp, 0.15 to 1.0 mm Al Within + 2 % or 2.5% for entire diagnostic range, up to 150 kvp For superficial or orthovoltage (radiobiology) 1 % to 2 % is desirable

24 There can be electron contamination and full buildup is needed for calibration of low energy (< 100kV) clinical beams This requires an addition to the window to be put over a plane-parallel chamber The window of the chamber (for example 2.5 mg / cm 2) should be subtracted from the values in the following table for the thickness of the foil for full buildup.

25 Tube potential (kv) Total wall thickness (mg cm -2 )

26 Physicist should understand uncertainties as delivered by the ADCL For example, for cobalt calibrations

27 The expanded uncertainty (at k=2) is expected to include the conventional true value. The chance of the population mean (the true value) being within the measured value has a 95% certainty. Or we can be sure that our measurement indicates the correct value within 95%

28 ADCL Standard Chamber Parameter Type A Type B Charge Timing Air Density Ionic Recombination 0 0 Distance from the source Beam Uniformity Air attenuation Quadratic Sum Type A and B Quadratic Sum NIST Source Calibration u NIST-ADCL Chamber Calibration or at k=2 level, 1.26 %

29 Clinic Chamber Parameter Type A Type B Charge Timing Air Density Ionic Recombination Distance from the source Beam Uniformity Quadratic Sum Type A and B Quadratic Sum ADCL Electrometer Readout ADCL Transfer Calibration u ADCL-Clinic Chamber Calibration or at k=2 level, 1.36 %

30 Standards changing Absorbed dose to water done with precision and uniformly. Calibration of chambers and sources essential for radiation therapy: Majority traceable through Cobalt and Cesium. Unfortunately, some manufacturers improvise dosimetry to market their product

31 Manufacturers need to pass some hoops to market product However, 510k from FDA does not mean that they have a calibration regimen for their machine. Also some manufacturers change without letting physicists know and assume everything still applies.

32

33 What happens when manufacturers improvise for Standards? Errors! NIST, ADCLs, AAPM should insist on a standard There is a need to have traceability delivered by the Accredited Dosimetry Calibration Laboratories Hospital Physicists generally have requests Users must insist on traceable standards from NIST through the ADCLs

34 Sr-90 applicators introduced in Plane or concave. 28 year half life Typical 8mm diameter with a beta shield Extremely high dose rate Calibrated by each mfr wrt Bragg-Gray theory Bizarre units like reps, beta-roentgens Hospitals asked UW ADCL accuracy of dose.

35 Wing-like growth in conjunctiva Requires surgery to remove Surgery fails 90% of time One dose with Sr-90 prevents recurrence in roughly 90% of patients treated

36

37 One Chicago hospital had an Amersham Model SIA-20 applicator calibrated at NBS: calibrations differed by 38%. Each lab stood by their calibration! NIST had to review their calibrations

38 Sr-90 sources had only been intercompared IN AIR at distances of 20 to 30cm Desired dose rate is IN CONTACT with Sr-90 on surface of silver matrix Chris Soares developed an extrapolation chamber for calibration.

39 NIST traceable determination of the absorbed dose to water rate in the central 4 mm of the applicator including color enhanced contour plots and two dimensional dose profiles. Routine ADCL calibrations from UW ADCL using radiochromic film began in NIST quotes uncertainty of 7 % Calibration disagreement (old versus new) average -27 %.

40 Example of a color enhanced contour plot of a uniform Sr-90 ophthalmic applicator. The dose weighted isocenter is equidistant from the hash marks. The outer circle represents the source physical diameter

41 2-Dimensional dose profiles for a uniform source. Horizontal Profile Vertical Profile Dose Rate (Gy/sec) Dose Rate (Gy/sec) Distance (mm) Distance (mm)

42 Example of a color enhanced contour plot of a skewed-sr-90 ophthalmic applicator. Notice the offset or shift of the dose weighted isocenter from the physical source center, and the nonuniform dose distribution

43 Horizontal Profile Vertical Profile Distance (mm) Distance (mm)

44 The average difference between the old calibration and the present calibration is -27%. The distribution of source activity can be very unique for the individual source.

45 NIST standard Pd-103 has very short half-life (17.0 d) so NIST traceable 109 Cd source (half-life of 463.2d) was used as a reference source for 12 years, then replaced in 1997 However, self-shielding of the source encapsulation was different between these two isotopes This resulted in a sudden 9% shift in calibration by letter to users in 1997

46

47 There can be significant errors If traceable to NIST, even if wrong, everyone is consistent if traceable standard- at least traceable through ADCL NIST, ADCLs and CIRMS need to insist upon standards. Manufacturers try hard but they need to rely on NIST and ADCLs for standards

48 Âg{Éáx ã{é wé ÇÉà ÜxÅxÅuxÜ à{x Ñtáà tüx véçwxåçxw àé ÜxÑxtà àaê George Santayana, Harvard Professor and poet

49 Thanks are due to Students and staff of the UW ADCL All those who send us calibration instruments that support the research program of the UW Medical Radiation Research Center.

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13.1.1 TRAINING GOAL 13.1.1.1 Among the responsabilities of a medical

More information

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61: Practical Implementation of TG-61: II. Guidelines for clinical implementation of TG-61 J.P. Seuntjens Medical Physics Unit McGill University, Montreal General Hospital Montréal, Canada jseuntjens@medphys.mcgill.ca

More information

Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens

Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens The characteristic ionizing radiation response is defined by the relationship between the stimulus

More information

Commissioning and periodic tests of the Esteya electronic brachytherapy system

Commissioning and periodic tests of the Esteya electronic brachytherapy system Review paper Educational Activity Commissioning and periodic tests of the Esteya electronic brachytherapy system Cristian Candela-Juan, PhD 1, Yury Niatsetski, MSc 2, Zoubir Ouhib, PhD 3, Facundo Ballester,

More information

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 DISCLAIMER: This publication and associated spreadsheets and digital files are based on sources and

More information

Relative Dosimetry. Photons

Relative Dosimetry. Photons Relative Dosimetry Photons What you need to measure! Required Data (Photon) Central Axis Percent Depth Dose Tissue Maximum Ratio Scatter Maximum Ratio Output Factors S c & S cp! S p Beam profiles Wedge

More information

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 47 CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 4.1 INTRODUCTION Advanced treatment techniques use optimized radiation beam intensities to conform dose distribution to the

More information

Constancy checks of well-type ionization chambers with external-beam radiation units

Constancy checks of well-type ionization chambers with external-beam radiation units JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Constancy checks of well-type ionization chambers with external-beam radiation units Sara L. Hackett, 1a Benjamin Davis, 2 Andrew

More information

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams Peter R. Almond Brown Cancer Center, Louisville, Kentucky 40202 Peter J. Biggs Department of Radiation Oncology,

More information

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Commissioning an IMRT System for MLC Delivery Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Taking the broad view of commissioning Commissioning elements Validating the dosimetry system Commissioning the

More information

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers EXRADIN Ion Chambers Exradin Ion Chambers Exradin (EXacting RADiation INstrumentation) Ion Chambers have been built for over 33 years, are recognized by top research institutes and standards laboratories,

More information

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08. Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas 2009 AAPM Summer School Educational Objectives Improve your understanding of how to implement TG-51

More information

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription Monitor Unit Calculations Part 1 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Return to our first patient 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt

More information

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas Current Implementation Status Current Implementation Status 1600 1400 TOTAL 1494 of 1623 ACTIVE INSTITUTIONS

More information

Zoubir Ouhib Lynn Cancer Institute

Zoubir Ouhib Lynn Cancer Institute Zoubir Ouhib Lynn Cancer Institute April 2015 Speaker for ELEKTA Brachytherapy using miniature X-ray sources Energy

More information

TG-51: Experience from 150 institutions, common errors, and helpful hints

TG-51: Experience from 150 institutions, common errors, and helpful hints TG-51: Experience from 150 institutions, common errors, and helpful hints Ramesh C. Tailor, Ph. D., William F. Hanson, Ph.D., and Geoffrey S. Ibbott, Ph.D. Department of Radiation Physics, University of

More information

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Tom Jordan Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE

More information

Victoreen B. Operators Manual. Image Intensifier Ion Chamber

Victoreen B. Operators Manual. Image Intensifier Ion Chamber Victoreen 6000-530B Image Intensifier Ion Chamber Operators Manual March 2005 Manual No 6000-530B-1 Rev. 4 2004, 2005 Fluke Corporation, All rights reserved. All product names are trademarks of their respective

More information

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER Proven Rad-Check technology specifically designed to provide you with the ultimate in versatility and cost-effective operation. Fast and easy

More information

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation This intercomparison will be carried out under the lead of KRISS with the cooperation of ARPANSA. This comparison has been approved

More information

IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy

IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy Guidelines on standardized procedures at Secondary Standards Dosimetry Laboratories (SSDLs) and hospitals March 2002 The

More information

Monitor Unit Verification for Small Fields

Monitor Unit Verification for Small Fields Monitor Unit Verification for Small Fields Patrick Higgins, Ph.D University of Minnesota Department of Radiation Oncology October 10, 2013 The Issues: How do we verify the monitor units calculated by the

More information

Abstract: Introduction:

Abstract: Introduction: Are photon and electron beam calibrations more consistent with TG-51 than with TG-21? Ramesh C. Tailor, William F. Hanson, Nathan Wells, and Geoffrey S. Ibbott. U.T. MD Anderson Cancer Center, Houston

More information

E X R A D I N. Parallel Plate Ion Chambers User Manual DOC #

E X R A D I N. Parallel Plate Ion Chambers User Manual DOC # E X R A D I N Parallel Plate Ion Chambers User Manual R DOC #80332-02 E X R A D I N Parallel Plate Ion Chambers User Manual STANDARD IMAGING INC. 7601 Murphy Drive Middleton, WI 53562 TEL 800.261.4446

More information

Proton Therapy QA Tools

Proton Therapy QA Tools Proton Therapy QA Tools A complete range of tools for absolute dosimetry, machine QA and patient plan verification Introduction During the past years more and more heavy particle therapy centers, especially

More information

Review of fundamental photon dosimetry quantities

Review of fundamental photon dosimetry quantities Review of fundamental photon dosimetry quantities Narayan Sahoo Main sources of the materials included in this lecture notes are: (1) Radiation Oncology Physics: A Handbook for Teachers and Students Edited

More information

INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS

INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS A9744593 INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSLS Ladislav Czap, Georg Matscheko and Pedro Andreo osimetry and Medical Radiation Physics Section, ivision

More information

Maintenance and general care of secondary standard and field instruments

Maintenance and general care of secondary standard and field instruments Maintenance and general care of secondary standard and field instruments This is without doubt the most important lecture that you will be given during this course. That rather rash statement is based

More information

THE development of more advanced techniques in radiotherapy,

THE development of more advanced techniques in radiotherapy, 1 A Geant4-based simulation of an accelerator s head used for Intensity Modulated Radiation Therapy F. Foppiano, B. Mascialino, M.G. Pia, M. Piergentili. Abstract We present a Geant4-based simulation,

More information

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL FOR THE DEGREE MASTER OF ARTS BY CHARLES TRAVIS WEBB

More information

Technical protocol of APMP key comparison for measurement of air kerma for 60 Co (APMP.RI(I)-K1.1)

Technical protocol of APMP key comparison for measurement of air kerma for 60 Co (APMP.RI(I)-K1.1) Technical protocol of APMP key comparison for measurement of air kerma for 60 Co (APMP.RI(I)-K1.1) 1. Introduction A regional APMP.RI(I)-K1 comparison of the standards for air kerma for 60 Co occurred

More information

ANSI N323A and ANSI N323B

ANSI N323A and ANSI N323B ANSI N323A and ANSI N323B Test and Calibration of Portable Survey Instruments Presented at AMUG Conference Las Vegas, NV April 23-27, 2012 This presentation describes the additions, deletions, and modifications

More information

Comparison of ionization chambers of various volumes for IMRT absolute dose verification

Comparison of ionization chambers of various volumes for IMRT absolute dose verification Comparison of ionization chambers of various volumes for IMRT absolute dose verification Leonid B. Leybovich, a) Anil Sethi, and Nesrin Dogan Department of Radiation Oncology, Loyola University Medical

More information

In air calibration of an HDR 192 Ir brachytherapy source using therapy ion chambers

In air calibration of an HDR 192 Ir brachytherapy source using therapy ion chambers Original Article Free full text available from www.cancerjournal.net In air calibration of an HDR 192 Ir brachytherapy source using therapy ion chambers ABSTRACT The Gammamed Plus 192 Ir high dose rate

More information

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics. FRANZCR Examination Part I Radiation Oncology Radiotherapeutic Physics Candidate No.: The Royal Australian and New Zealand College of Radiologists FRANZCR Examination Part I Radiation Oncology Radiotherapeutic

More information

IMRT in clinical practice at the UMC-Utrecht. Utrecht

IMRT in clinical practice at the UMC-Utrecht. Utrecht 3 IMRT in clinical practice at the UMC-Utrecht Utrecht Clinical use of IMRT Improvement of local control without increased incidence of normal tissue complications + Dose escalation in the prostate Step-and

More information

SnapShot IMRT with compensators and FFF beams

SnapShot IMRT with compensators and FFF beams SnapShot IMRT with compensators and FFF beams Vladimir Feygelman, PhD (1) Moffitt Cancer Center, Tampa, FL, USA Disclosure VF has a sponsored research agreement with.decimal. A short history of radiotherapy

More information

IAEA-TECDOC-1455 Implementation of the International Code of Practice on Dosimetry in Radiotherapy (TRS 398): Review of testing results

IAEA-TECDOC-1455 Implementation of the International Code of Practice on Dosimetry in Radiotherapy (TRS 398): Review of testing results IAEA-TECDOC-1455 Implementation of the International Code of Practice on Dosimetry in Radiotherapy (TRS 398): Revie of testing results Final report of the Coordinated Research Projects on Implementation

More information

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization Basics of Proton Therapy Proton Treatment Planning and Beam Optimization SAM Educational Session, WE D BRB 2 Mark Pankuch, PhD Northwestern Medicine Chicago Proton Center Today s objectives Review the

More information

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities Specific Accreditation Criteria Calibration ISO IEC 17025 Annex Mass and related quantities January 2018 Copyright National Association of Testing Authorities, Australia 2014 This publication is protected

More information

General Accreditation Guidance. User checks and maintenance of laboratory balances

General Accreditation Guidance. User checks and maintenance of laboratory balances General Accreditation Guidance User checks and maintenance of laboratory balances January 2018 Copyright National Association of Testing Authorities, Australia 2010 All intellectual property rights in

More information

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components Introduction The CyberKnife SRS/SBRT System James M. Hevezi,, Ph.D., FACR CyberKnife Unit San Antonio, Texas CyberKnife Components Treatment Delivery (G3 Platform San Antonio) Robot (modified German auto

More information

Dosimetric Calculations. Lonny Trestrail

Dosimetric Calculations. Lonny Trestrail Dosimetric Calculations Lonny Trestrail 20 October 2008 Objectives Dose Distribution Measurements PDD, OCR TAR, SAR, TPR, TMR, SPR, SMR Arc or Rotational Therapy Isodose Curves Point Dose Calculations

More information

Exposure System Selection

Exposure System Selection Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others

More information

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World Chattanooga, Tennessee, April 17-21, 2005, on CD-ROM, American Nuclear Society, LaGrange Park, IL (2005) GEANT4 SIMULATION OF

More information

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11 MV Photon Dosimetry in the Clinic This document discusses both the scientific methods and the system of work required to calibrate a field instrument expanding upon the practical details of both.. Introduction

More information

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions Rebecca M. Howell, Ph.D., D.A.B.R, The University of Texas at M.D. Anderson Cancer Center Overview 2 Commissioning

More information

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system G. Kunz, F. Hasenbalg, P. Pemler 1 1 Klinik für Radio-Onkologie und Nuklearmedizin, Stadtspital Triemli

More information

A beam-matching concept for medical linear accelerators

A beam-matching concept for medical linear accelerators Acta Oncologica, 2009; 48: 192200 ORIGINAL ARTICLE A beam-matching concept for medical linear accelerators DAVID SJÖSTRÖM, ULF BJELKENGREN, WIVIANN OTTOSSON & CLAUS F. BEHRENS Copenhagen University Hospital,

More information

Operational experiences with the EuroLoop Liquid Hydrocarbon Flow Facility

Operational experiences with the EuroLoop Liquid Hydrocarbon Flow Facility Operational experiences with the EuroLoop Liquid Hydrocarbon Flow Facility Jos G.M. van der Grinten, NMi EuroLoop Bart van der Stap, Flowways Dick van Driel, Krohne Abstract The operational experiences

More information

OBJECTIVE 6: FIELD RADIOLOGICAL MONITORING - AMBIENT RADIATION MONITORING

OBJECTIVE 6: FIELD RADIOLOGICAL MONITORING - AMBIENT RADIATION MONITORING OBJECTIVE 6: FIELD RADIOLOGICAL MONITORING - AMBIENT RADIATION MONITORING OBJECTIVE Demonstrate the appropriate use of equipment and procedures for determining field radiation measurements. INTENT This

More information

Requirements and Methods for Calibration of Equipment in Clinical Laboratories. NK Chan (HKRCBTS) 24 August 2007

Requirements and Methods for Calibration of Equipment in Clinical Laboratories. NK Chan (HKRCBTS) 24 August 2007 Requirements and Methods for Calibration of Equipment in Clinical Laboratories NK Chan (HKRCBTS) 24 August 2007 1 Contents Basic Requirements of Equipment Calibration Calibration of Electronic Balance

More information

Commissioning and quality assurance for the treatment delivery components of the AccuBoost system

Commissioning and quality assurance for the treatment delivery components of the AccuBoost system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 2, 2015 Commissioning and quality assurance for the treatment delivery components of the AccuBoost system Ileana Iftimia, 1,2a Mike Talmadge,

More information

Math Review. Overview

Math Review. Overview Drawn by Steve Yan, CMD Math Review Shirley Ann Pinegar-Johnston MS RT(R)(T)CMD Overview Drawn by Steve Yan, CMD Divergence & Intensity DIVERGENCE X-rays travel in Straight but Divergent lines INTENSITY

More information

Investigating the effects of interchanging components used to perform ripple assessments on calibrated vector network analysers

Investigating the effects of interchanging components used to perform ripple assessments on calibrated vector network analysers Abstract Investigating the effects of interchanging components used to perform ripple assessments on calibrated vector network analysers Andrew G Morgan and Nick M Ridler Centre for Electromagnetic and

More information

A3 / A4 / A5 / A6 / A8 Spherical Ionization Chambers REF / / / / User Manual

A3 / A4 / A5 / A6 / A8 Spherical Ionization Chambers REF / / / / User Manual A3 / A4 / A5 / A6 / A8 Spherical Ionization Chambers REF 92717 / 92715 / 92724 / 92716 / 92737 User Manual WWW..COM A3 / A4 / A5 / A6 / A8 Spherical Ionization Chambers REF 92717 / 92715 / 92724 / 92716

More information

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Poster No.: R-0044 Congress: Type: Authors: Keywords: DOI: 2014 CSM Scientific Exhibit L. Bendall, I. Patel, N. McGrath, C. Rowbottom; MANCHESTER/

More information

Victoreen and

Victoreen and Victoreen 06-526-5240 and 06-526-5242 RAD-CHECK MICRO-R Operators Manual February 2005 Manual No. 06-526-5240-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names

More information

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN 22 Turkish Journal of Cancer Volume 37, No.1, 2007 Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN SATISH PELAGADE 1, KALPANA

More information

BLOCKAGE LOCATION THE PULSE METHOD

BLOCKAGE LOCATION THE PULSE METHOD BLOCKAGE LOCATION THE PULSE METHOD Presented by John Pitchford Pitchford In-Line Author James Pitchford ABSTRACT Pipeline blockages can result from a number of different mechanisms: wax or solid hydrates

More information

Radiation Producing Machines Radiation Safety Training for X-ray Diffraction Units

Radiation Producing Machines Radiation Safety Training for X-ray Diffraction Units INTRODUCTION Radiation Producing Machines Radiation Safety Training for X-ray Diffraction Units RADIATION SAFETY GUIDE IMAGING AND MICROSCOPY FACILITY University of California, Merced Training Document

More information

Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water

Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water M R McEwen, A J Williams and A R DuSautoy Centre for Ionising Radiation Metrology National Physical

More information

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface Chapter 11 reatment Planning Single Beams Radiation Dosimetry I Outline Basic terminology Curved contour surface correction (bolus, compensators, wedges) Oblique beam incidence Correction for tissue inhomogeneities

More information

How to Design Medical Accelerator Systems for Reliability: IBA PT System

How to Design Medical Accelerator Systems for Reliability: IBA PT System How to Design Medical Accelerator Systems for Reliability: IBA PT System Yves Jongen Founder & Chief Research Officer Ion Beam Applications s.a. Belgium 1 Outline A short introduction of a proton therapy

More information

ON THE RELATIVE ADEQUACY OF SOME AMERICAN CHAMBERS FOR MEASUREMENTS IN THE GRENZ RAY RANGE*

ON THE RELATIVE ADEQUACY OF SOME AMERICAN CHAMBERS FOR MEASUREMENTS IN THE GRENZ RAY RANGE* ON THE RELATIVE ADEQUACY OF SOME AMERICAN CHAMBERS FOR MEASUREMENTS IN THE GRENZ RAY RANGE* VICTOR H. WITTEN, M.D., EDGAR N. GRISEWOOD, M.A. AND ELEANOR OSHRY, B.S.t The use of grenz radiation has been

More information

Improve Process Reliability

Improve Process Reliability Convenient Conductivity Calibration to Meet Pharmacopeia Regulations Pharmacopeias require that conductivity measurement systems be periodically calibrated to ensure their precision. The range of accessories

More information

Thimble Ionization Chambers

Thimble Ionization Chambers Thimble Ionization Chambers User Manual WWW..COM Thimble Ionization Chambers STANDARD IMAGING INC. 3120 Deming Way Middleton, WI 53562-1461 Dec / 2010 2010 Standard Imaging Inc. DOC #80322-16 TEL 800.261.4446

More information

Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System

Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System Amit Sawant 1, Sonja Dieterich 1, Michelle Svatos 2 and Paul Keall 1 1 Stanford University, Stanford, CA 2 Varian Medical Systems, Palo

More information

Review of backscatter measurement in kilovoltage radiotherapy using novel detectors and reduction from lack of underlying scattering material

Review of backscatter measurement in kilovoltage radiotherapy using novel detectors and reduction from lack of underlying scattering material JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 6, 2013 Review of backscatter measurement in kilovoltage radiotherapy using novel detectors and reduction from lack of underlying scattering

More information

Increased Project Bankability : Thailand's First Ground-Based LiDAR Wind Measurement Campaign

Increased Project Bankability : Thailand's First Ground-Based LiDAR Wind Measurement Campaign Increased Project Bankability : Thailand's First Ground-Based LiDAR Wind Measurement Campaign Authors: Velmurugan. k, Durga Bhavani, Ram kumar. B, Karim Fahssis As wind turbines size continue to grow with

More information

MET-101 Introduction to Metrology

MET-101 Introduction to Metrology Pasig city Tel: 654-13-42 MET-101 Introduction to Metrology This 2-day course offers a basic introduction into Metrology, meant for people who are working in calibration and testing labs, maintenance services

More information

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator Journal of BUON 8: 55-59, 2003 2003 Zerbinis Medical Publications. Printed in Greece ORIGINAL ARTICLE Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

More information

Electronic Oncology Systems the Management and the Safety Measures. Hansen Chen, Director, TDSI

Electronic Oncology Systems the Management and the Safety Measures. Hansen Chen, Director, TDSI Electronic Oncology Systems the Management and the Safety Measures Hansen Chen, Director, TDSI Radiation Oncology Department People Place Equipments Procedures Patients Hansen Chen 2 Radiation Oncology

More information

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Cover Page for Lab Report Group Portion. Head Losses in Pipes Cover Page for Lab Report Group Portion Head Losses in Pipes Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 February 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section

More information

Badan Jadrowych Nuclear Research Institute. Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P Differential Recombination Chamber

Badan Jadrowych Nuclear Research Institute. Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P Differential Recombination Chamber Badan Jadrowych Nuclear Research Institute Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P00100517 Differential Recombination Chamber by M. Zel'chinskij K. Zharnovetskij Warsaw, June 1966 Translated

More information

Technical Procedure for General Laboratory Equipment

Technical Procedure for General Laboratory Equipment Technical Procedure for General Laboratory Equipment 1.0 Purpose - This procedure specifies the required elements for the use of general laboratory equipment. 2.0 Scope - This procedure applies to all

More information

20. Technical Equipment. Archived. This section applies to all equipment used in forensic examinations, such as

20. Technical Equipment. Archived. This section applies to all equipment used in forensic examinations, such as Instruments and Equipment Introduction 5.5. Authorization to use technical equipment 5.5.3 Use of instruments and equipment Instruction and maintenance manuals 5.4. 5.5.3 Use of nonlaboratory equipment

More information

Ti-Pure TITANIUM DIOXIDE DETERMINATION OF SLURRY RHEOLOGY METHOD: T WP

Ti-Pure TITANIUM DIOXIDE DETERMINATION OF SLURRY RHEOLOGY METHOD: T WP Ti-Pure TITANIUM DIOXIDE DETERMINATION OF SLURRY RHEOLOGY METHOD: T4400.308.01.WP 1 TITANIUM DIOXIDE (Ti-Pure ) Rheology of Titanium Dioxide Slurries I. Principle Rheology is a measurement of the fluid

More information

ATION TITLE. Survey QC, Decision Making, and a Modest Proposal for Error Models. Marc Willerth, MagVAR

ATION TITLE. Survey QC, Decision Making, and a Modest Proposal for Error Models. Marc Willerth, MagVAR 1 a Modest Proposal for Error Models ATION TITLE Marc Willerth, MagVAR Speaker Information Marc Willerth VP of Survey Technologies April 11, 2018 MagVAR 2 Speaker Bio Marc Willerth Magnetic Variation Services,

More information

MIT Analytical X-Ray Safety Policies and Procedures

MIT Analytical X-Ray Safety Policies and Procedures 5th Edition 2014 MIT Analytical X-Ray Safety Issued by: MIT Radiation Protection Committee Contents 1.0 Program Overview...1 2.0 Purpose...1 3.0 Scope and Application...2 4.0 Radiation Protection Program

More information

Cover Page for Lab Report Group Portion. Pump Performance

Cover Page for Lab Report Group Portion. Pump Performance Cover Page for Lab Report Group Portion Pump Performance Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 March 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section number:

More information

Temperature, salinity, density, and the oceanic pressure field

Temperature, salinity, density, and the oceanic pressure field Chapter 2 Temperature, salinity, density, and the oceanic pressure field The ratios of the many components which make up the salt in the ocean are remarkably constant, and salinity, the total salt content

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 Comparison of monitor unit calculations performed with a 3D computerized planning system and independent hand calculations: Results

More information

CERTIFICATE OF ACCREDITATION ISO/IEC 17025:2005 ANSI/NCSL Z (R2002)

CERTIFICATE OF ACCREDITATION ISO/IEC 17025:2005 ANSI/NCSL Z (R2002) CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Antibus Scales & Systems, Inc. 1919 Research

More information

Impact of imperfect sealing on the flow measurement of natural gas by orifice plates

Impact of imperfect sealing on the flow measurement of natural gas by orifice plates Impact of imperfect sealing on the flow measurement of natural gas by orifice plates Rubens Silva Telles 1, Kazuto Kawakita 2 1 IPT Instituto de Pesquisas Tecnológicas, São Paulo, Brazil, rtelles@ipt.br

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. PRECISION MEASURING CORP Groesbeck Hwy. Fraser, MI Robert Meyer Phone:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. PRECISION MEASURING CORP Groesbeck Hwy. Fraser, MI Robert Meyer Phone: SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 PRECISION MEASURING CORP. 33250 Groesbeck Hwy. Fraser, MI 48026 Robert Meyer Phone: 586 293 4420 CALIBRATION Valid To: September 30, 2019 Certificate Number:

More information

Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry

Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry Bryan R. Muir a) and Malcolm R. McEwen Measurement Science and Standards, National Research Council of

More information

Testing of Active Flow Control actuators at harsh environment

Testing of Active Flow Control actuators at harsh environment Testing of Active Flow Control actuators at harsh environment Ionut BRINZA (INCAS) 6 th CEAS CONFERENCE 16-20 October 2017 Bucharest ROMANIA Introduction Results of the harsh environment tests Conclusions

More information

HUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0

HUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0 High Accuracy Humidity Calibrator The role of humidity calibrations that are accurate, reproducible, and documentable is becoming more and more important. ISO quality guidelines and regulations according

More information

Vaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated.

Vaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated. Asset #: 123 Calibration - Certificate No: 2083.01 Customer: Condition: Action Taken: Sample Inc. 123 Sample Rd. Sample, MA 01234 The instrument was operational upon receipt. The instrument was adjusted

More information

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010 Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600 November 2010 Table of Contents ``Abstract...3 ``Introduction...4 ``Methods & Results...6 ``Data Analysis...9 ``Conclusion...12

More information

EXPERIMENTAL RESULTS OF GUIDED WAVE TRAVEL TIME TOMOGRAPHY

EXPERIMENTAL RESULTS OF GUIDED WAVE TRAVEL TIME TOMOGRAPHY 18 th World Conference on Non destructive Testing, 16-20 April 2012, Durban, South Africa EXPERIMENTAL RESULTS OF GUIDED WAVE TRAVEL TIME TOMOGRAPHY Arno VOLKER 1 and Hendrik VOS 1 TNO, Stieltjesweg 1,

More information

LAKE WASHINGTON SOCKEYE SALMON STUDIES. Richard E. Thorne and James J. Dawson

LAKE WASHINGTON SOCKEYE SALMON STUDIES. Richard E. Thorne and James J. Dawson FRI-UW-7613 October 1976 LAKE WASHINGTON SOCKEYE SALMON STUDIES 1975 1976 by Richard E. Thorne and James J. Dawson Final Report Service Contract No. 648 Washington State Department of Fisheries For the

More information

Application Note AN-107

Application Note AN-107 SPEC Sensor TM Characterization & Calibration Considerations Scope This document is provided to describe the considerations needed to characterize, calibrate, verify and validate the measurement performance

More information

Picus NxT Electronic Pipettes Ergonomic, Efficient, & Safe

Picus NxT Electronic Pipettes Ergonomic, Efficient, & Safe Picus NxT Electronic Pipettes Ergonomic, Efficient, & Safe Picus NxT Electronic Pipettes Excel in Ergonomics, Efficiency, and Safety Are you looking for the following qualities in a pipette and are not

More information

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2017, 6, 57-72 http://www.scirp.org/journal/ijmpcero ISSN Online: 2168-5444 ISSN Print: 2168-5436 Validation of Treatment

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results. Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions

More information

Technical Report. 5th Round Robin Test for Multi-Capillary Ventilation Calibration Standards (2016/2017)

Technical Report. 5th Round Robin Test for Multi-Capillary Ventilation Calibration Standards (2016/2017) Physical Test Methods Sub-Group Technical Report 5th Round Robin Test for Multi-Capillary Ventilation Calibration Standards (2016/2017) June 2018 Author: James Vincent Cerulean, Milton Keynes, United Kingdom

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION doi:10.1038/nature16448 Supplementary Table 1. Summary of the stone artefact distributions, T2 and T4. T2 Age and context T4 Type

More information

Gun Vacuum Performance Measurements February 2003

Gun Vacuum Performance Measurements February 2003 JLAB-TN-03-013 Gun Vacuum Performance Measurements February 2003 Marcy Stutzman Abstract. The vacuum in the polarized photoguns at Jefferson Lab is crucial to long lifetime operation of the polarized source.

More information