1 The MC commissioning of CyberKnife with MLC (Tips and Tricks) Alain Guemnie Tafo, PhD UPMC Pinnacle, Harrisburg, PA Institute Gustave Roussy, Villejuif, France
2 Disclaimer & Disclosure The views expressed in this presentation are those of the presenters and do not necessarily reflect the views or policies of Accuray Incorporated or its subsidiaries. No official endorsement by Accuray Incorporated or any of its subsidiaries of any vendor, products or services contained in this presentation is intended or should be inferred. An honorarium is provided by Accuray for this presentation
3 Guidelines MC Data/Commissioning MC source model MC Validation MC clinical implementation
4 MC Data/Commissioning In air Output Factor: same as for Ray Tracing Open field dose profile: Profile acquired with no collimator fixed to the Linac head 25mm depth) Central PDD for the 60mm collimator Material: Diode PTW Accuray s excel spread sheet is useful to compare with composite data OF correction factor in literature (Francescon et al 2017 Phys. Med. Biol )
5 MC source model Adjustment of the Energy factor: allow to adjust the photon energy spectrum to match the PDDs (choice of the photon mean energy and w/o beam hardening). Adjustment of the Gaussian spread: Allow to adjust the source size to match the penumbras (choice of the sigma of a Gaussian) Adjustment of Leaves related parameters
6 TPR Model (ECF factor) TPR 6MeV, 1mm, 7.6mm TPR 6.4MeV, 1.8mm, 7.6mm TPR 7MeV, 4mm, 7.6mm TPR 6MeV, 1mm, 84.6mm TPR 6.4MeV, 1.8mm, 84.6mm TPR 7MeV, 4mm, 84.6mm
7 OCR Model (Source size) TPR 6MeV, 1mm, 7.6mm, d=100mm TPR 6.4MeV, 1.8mm, 7.6mm, d=100mm TPR 7MeV, 4mm, 7.6mm, d=100mm TPR 6MeV, 1mm, 84.6mm, d=100mm TPR 6.4MeV, 1.8mm, 84.6mm, d=100mm TPR 7MeV, 4mm, 84.6mm, d=100mm
8 OCR Model (Source size) TPR 6.4MeV, 1.8mm, 7.6mm, d=300mm TPR 6.4MeV, 1.8mm, 7.6mm, d=300mm TPR 6.4MeV, 1.8mm, 7.6mm, d=300mm Proposed Source size and E spectrum are a good starting point No multiple scatter source in the model so the shoulder is not well-matched for large field size and deep depth Using a Diamond detector is recommended for tails measurement accuracy Underestimation of the shoulder for large field size and depth Overestimation of the tail using a Diode detector (correction possible using correction factors, Francescon 2017)
9 MC Validation Beam Configuration Single Beam in a homogeneous phantom Single Beam in a heterogeneous phantom Multiple Beam in a homogeneous phantom Multiple beam in a heterogeneous phantom Multiple Beam in an anthropomorphic phantom
10 Single Beam in a Homogeneous Phantom Field size ranging from 23.1x23.1cm to 100x100.1 cm 800mm SAD PTW (0.125cc) Chamber positioned in the central axis of the beam at 1.5, 55, 10 and 20 cm depth.
11 Single Beam in a Heterogeneous Phantom Water-Lung-Water Configuration (EPE) WATER 8 cm Water-Lung-Water-Lung- Water Configuration -Low (WLWLW-Low) WATER LUNG WATER LUNG 5 cm 8.5 cm 4 cm 5 cm LUNG WATER 2 cm Ionization Chamber Water-Lung-Water-Lung- Water Configuration - High (WLWLW-High) WATER LUNG WATER LUNG 2 cm Ionization Chamber 13.5 cm 8 cm 5 cm 8.5 cm 4 cm 5 cm Field size ranging from 23.1x23.1cm to 100x100.1 cm 800 mm SAD 2 setups slab phantom: Water Lung Water and Water-Lung-Water-Lung-Water PTW (0.125cc) Chamber positioned in the central axis of the beam at 23.5cm depth in the WLW configuration and 15.5cm and 24.5cm for the WLWLW-High and WLWLW-Low configurations WATER 2 cm 8 cm WATER 5 cm Ionization Chamber
12 Results Monte Carlo calculation is in good agreement with measurement for all configurations (<2%) FSPBLS calculation showed a good agreement?
13 Multiple Beam in a Homogeneous Phantom PTW Ionization chamber 5cm depth Field sizes from 23x23.1cm to 100x100.1cm Sequential optimization with PTV size matching the field size Monte Carlo results show a larger discrepancy as compared to FSPB or FSPB- LS but stays within 2%.
14 Multiple beam in a Heterogeneous Phantom Setup 1: Ionization Chamber in a water equivalent plug located in the Mediastinum region Setup 2: Ionization Chamber in a water equivalent plug located in the Lung region Phantom CIRS 002LFC Ionization chamber PTW (0.007cc) 2 setups: Mediastinum and Lung (with tissue like material chamber plug) for Isocentric and Sequential beam optimization Measurement/Calculation agreed within <2% FSPBLS was in good agreement with measurement
15 Multiple Beam in an Anthropomorphic Phantom Anthropomorphic phantom ATOM 701 (CIRS) TLD measurement from EQUAL ESTRO (European Accredited Laboratory) 2 different plans with 30mm and 50mm cone Uniform dose distribution
16 TLD Results MC calculation are in a good agreement with the measurement (within 4% in the non gradient region) FSPBLS dose calculation is in better agreement with measurement as compared to FSPB
17 MLC Clinical gain 7 Lung cases Plans were calculated on fix MC and have been recalculated using the MLC collimator Time reduction is observed (20%) but also highest conformality for the studied cases (5%)
18 MC Clinical gain 4 Liver cases Plans were calculated on fix MC and have been recalculated using the MLC collimator Treatment time reduction observed (25%) but also highest conformality
19 Summary MLC allows to quicken the treatment time for the patient MLC allow to treat larger target MC for MLC is a great addition for extracranial treatments Commissioning and modeling are pretty straight forward Modeling showed a lack of scatter visible on the profile shoulders for large fields and large depth Validation showed good agreements for 1D measurement in homogeneous, heterogeneous and anthropomorphic configurations FSPB-LS seems to be a more accurate starting point for optimization prior to MC calculation Lung and Liver treatment with MLC showed a more conformal and faster treatment time
20 THANK YOU!
Relative Dosimetry. Photons
Relative Dosimetry Photons What you need to measure! Required Data (Photon) Central Axis Percent Depth Dose Tissue Maximum Ratio Scatter Maximum Ratio Output Factors S c & S cp! S p Beam profiles Wedge
More informationMonitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription
Monitor Unit Calculations Part 1 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Return to our first patient 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt
More informationCHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS
47 CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 4.1 INTRODUCTION Advanced treatment techniques use optimized radiation beam intensities to conform dose distribution to the
More informationHigh Speed Direct SAD Radiosurgery Beam Scanner
Open Access Original Article DOI: 10.7759/cureus.20 High Speed Direct SAD Radiosurgery Beam Scanner Walter Nikesch 1, James M. Hevezi 2, Irene Monterroso 3, Daniel Navarro 4, James G. Schwade 5 1. CyberKnife
More informationCommissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale
Commissioning an IMRT System for MLC Delivery Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Taking the broad view of commissioning Commissioning elements Validating the dosimetry system Commissioning the
More informationField size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator
Journal of BUON 8: 55-59, 2003 2003 Zerbinis Medical Publications. Printed in Greece ORIGINAL ARTICLE Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator
More informationMonitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting
Monitor Unit Calculations Part 2 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Calculation of machine setting reference dose machine setting =, reference dose
More informationValidation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.
International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2017, 6, 57-72 http://www.scirp.org/journal/ijmpcero ISSN Online: 2168-5444 ISSN Print: 2168-5436 Validation of Treatment
More informationMonitor Unit Verification for Small Fields
Monitor Unit Verification for Small Fields Patrick Higgins, Ph.D University of Minnesota Department of Radiation Oncology October 10, 2013 The Issues: How do we verify the monitor units calculated by the
More informationCommissioning and periodic tests of the Esteya electronic brachytherapy system
Review paper Educational Activity Commissioning and periodic tests of the Esteya electronic brachytherapy system Cristian Candela-Juan, PhD 1, Yury Niatsetski, MSc 2, Zoubir Ouhib, PhD 3, Facundo Ballester,
More informationA beam-matching concept for medical linear accelerators
Acta Oncologica, 2009; 48: 192200 ORIGINAL ARTICLE A beam-matching concept for medical linear accelerators DAVID SJÖSTRÖM, ULF BJELKENGREN, WIVIANN OTTOSSON & CLAUS F. BEHRENS Copenhagen University Hospital,
More informationCommissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN
22 Turkish Journal of Cancer Volume 37, No.1, 2007 Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN SATISH PELAGADE 1, KALPANA
More informationTG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:
Practical Implementation of TG-61: II. Guidelines for clinical implementation of TG-61 J.P. Seuntjens Medical Physics Unit McGill University, Montreal General Hospital Montréal, Canada jseuntjens@medphys.mcgill.ca
More informationDosimetric Calculations. Lonny Trestrail
Dosimetric Calculations Lonny Trestrail 20 October 2008 Objectives Dose Distribution Measurements PDD, OCR TAR, SAR, TPR, TMR, SPR, SMR Arc or Rotational Therapy Isodose Curves Point Dose Calculations
More informationIntroduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components
Introduction The CyberKnife SRS/SBRT System James M. Hevezi,, Ph.D., FACR CyberKnife Unit San Antonio, Texas CyberKnife Components Treatment Delivery (G3 Platform San Antonio) Robot (modified German auto
More informationOutline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface
Chapter 11 reatment Planning Single Beams Radiation Dosimetry I Outline Basic terminology Curved contour surface correction (bolus, compensators, wedges) Oblique beam incidence Correction for tissue inhomogeneities
More informationCommissioning of Elekta 6MV FFF Versa HD and Pinnacle
Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Poster No.: R-0044 Congress: Type: Authors: Keywords: DOI: 2014 CSM Scientific Exhibit L. Bendall, I. Patel, N. McGrath, C. Rowbottom; MANCHESTER/
More informationSnapShot IMRT with compensators and FFF beams
SnapShot IMRT with compensators and FFF beams Vladimir Feygelman, PhD (1) Moffitt Cancer Center, Tampa, FL, USA Disclosure VF has a sponsored research agreement with.decimal. A short history of radiotherapy
More informationTHE development of more advanced techniques in radiotherapy,
1 A Geant4-based simulation of an accelerator s head used for Intensity Modulated Radiation Therapy F. Foppiano, B. Mascialino, M.G. Pia, M. Piergentili. Abstract We present a Geant4-based simulation,
More informationReview of fundamental photon dosimetry quantities
Review of fundamental photon dosimetry quantities Narayan Sahoo Main sources of the materials included in this lecture notes are: (1) Radiation Oncology Physics: A Handbook for Teachers and Students Edited
More informationBasics of Proton Therapy. Proton Treatment Planning and Beam Optimization
Basics of Proton Therapy Proton Treatment Planning and Beam Optimization SAM Educational Session, WE D BRB 2 Mark Pankuch, PhD Northwestern Medicine Chicago Proton Center Today s objectives Review the
More informationThe Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.
FRANZCR Examination Part I Radiation Oncology Radiotherapeutic Physics Candidate No.: The Royal Australian and New Zealand College of Radiologists FRANZCR Examination Part I Radiation Oncology Radiotherapeutic
More informationTG219: IT'S USE, STRENGTHS AND WEAKNESSES
1 CONFLICT OF INTEREST TG219: IT'S USE, STRENGTHS AND WEAKNESSES SOTIRI STATHAKIS, PHD, DABR None ROSTER STATEMENT OF THE PROBLEM Timothy C. Zhu, University of Pennsylvania, Philadelphia, PA (Co- Chair)
More informationTG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009
TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 DISCLAIMER: This publication and associated spreadsheets and digital files are based on sources and
More informationConstancy checks of well-type ionization chambers with external-beam radiation units
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Constancy checks of well-type ionization chambers with external-beam radiation units Sara L. Hackett, 1a Benjamin Davis, 2 Andrew
More informationExradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers
EXRADIN Ion Chambers Exradin Ion Chambers Exradin (EXacting RADiation INstrumentation) Ion Chambers have been built for over 33 years, are recognized by top research institutes and standards laboratories,
More informationCLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions
CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions Rebecca M. Howell, Ph.D., D.A.B.R, The University of Texas at M.D. Anderson Cancer Center Overview 2 Commissioning
More informationJOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 Comparison of monitor unit calculations performed with a 3D computerized planning system and independent hand calculations: Results
More information13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.
13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13.1.1 TRAINING GOAL 13.1.1.1 Among the responsabilities of a medical
More informationPractical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11
MV Photon Dosimetry in the Clinic This document discusses both the scientific methods and the system of work required to calibrate a field instrument expanding upon the practical details of both.. Introduction
More information1. Question Answer cgy / MU cgy / MU 2. Question Answer
GS020113: Introduction to Medical Physics III: Therapy s to home work problem set assigned on 3/22/11 1. Question A patient is set up at 100 cm SSD on a 6 MVX machine. The dose rate at 10 cm in phantom
More informationGEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY
The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World Chattanooga, Tennessee, April 17-21, 2005, on CD-ROM, American Nuclear Society, LaGrange Park, IL (2005) GEANT4 SIMULATION OF
More informationInvestigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy
University of South Florida Scholar Commons Graduate Theses and Dissertations Graduate School 7-14-2010 Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy
More informationTG-51: Experience from 150 institutions, common errors, and helpful hints
TG-51: Experience from 150 institutions, common errors, and helpful hints Ramesh C. Tailor, Ph. D., William F. Hanson, Ph.D., and Geoffrey S. Ibbott, Ph.D. Department of Radiation Physics, University of
More informationExperimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media
Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media N Tyagi 1, B H Curran, P L Roberson, J M Moran, E Acosta and B A Fraass Department
More informationClinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas
Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas Current Implementation Status Current Implementation Status 1600 1400 TOTAL 1494 of 1623 ACTIVE INSTITUTIONS
More informationCOMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy
COMPASS 3D IMRT Verification in Patient Anatomy AK IMRT Bamberg April 2010 Dr. Lutz Müller Patient-specific Verification? CT Treatment unit Gantry =0 Treatmt unit Patient couch X-ray beam Patient couch
More informationOnly 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.
Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas 2009 AAPM Summer School Educational Objectives Improve your understanding of how to implement TG-51
More informationImplementation of respiratory-gated VMAT on a Versa HD linear accelerator
Received: 21 November 2016 Revised: 7 April 2017 Accepted: 5 June 2017 DOI: 10.1002/acm2.12160 RADIATION ONCOLOGY PHYSICS Implementation of respiratory-gated VMAT on a Versa HD linear accelerator Jeffrey
More informationZoubir Ouhib Lynn Cancer Institute
Zoubir Ouhib Lynn Cancer Institute April 2015 Speaker for ELEKTA Brachytherapy using miniature X-ray sources Energy
More informationProton Therapy QA Tools
Proton Therapy QA Tools A complete range of tools for absolute dosimetry, machine QA and patient plan verification Introduction During the past years more and more heavy particle therapy centers, especially
More informationIntroduction. Materials and Methods
STANDARD WEDGE AND TRAY TRANSMISSION VALUES FOR VARIAN, SEIMENS, ELEKTA/PHILIPS ACCELERATORS; A QUALITY ASSURANCE TOOL David S. Followill, Nadia Hernandez, and William F. Hanson Introduction The Radiological
More informationCommissioning of a new total body irradiation protocol
University of Wollongong Research Online University of Wollongong Thesis Collection University of Wollongong Thesis Collections 2012 Commissioning of a new total body irradiation protocol Zoe Baldwin University
More informationBeam modeling and VMAT performance with the Agility 160-leaf multileaf collimator
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 2, 2013 Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator James L. Bedford, a Michael D. R. Thomas and Gregory
More informationDynamic wedges dosimetry and quality control
Original Paper Received: 2005.06.07 Accepted: 2006.02.28 Published: 2006.04.28 Authors Contribution: A Study Design B Data Collection C Statistical Analysis D Data Interpretation E Manuscript Preparation
More informationClinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry
Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Tom Jordan Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE
More informationA simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation
Radiotherapy and Oncology 71 (2004) 235 241 Technical Note A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation Nadine Linthout*,
More informationOriginal Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams
Iranian Journal of Medical Physics Vol. 10, No. 3, Summer 2013, 147-155 Received: April 22, 2013; Accepted: July 21, 2013 Original Article Effective Point of Measurement in Cylindrical Ion Chamber for
More informationExposure System Selection
Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others
More informationClinical Implementation of Volumetric Modulated Arc Therapy
Clinical Implementation of Volumetric Modulated Arc Therapy UT M.D. Anderson Cancer Center Ramaswamy Sadagopan, Rebecca M. Howell, Weiliang Du and Peter Balter Definition 2 Intensity Modulated Arc therapy
More informationThe Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007
The Effect of Immobilisation Devices on Radiotherapy Dose Distributions A thesis submitted in fulfilment of the requirements for the Degree of Master of Applied Science (Medical and Health Physics) By
More informationIranian Journal of Medical Physics
Iranian Journal of Medical Physics ijmp.mums.ac.ir Verification of Monitor Unit Calculations for Eclipse Treatment Planning System by in-house Developed Spreadsheet Athiyaman Mayivaganan 1*, Hemalatha
More informationEvaluation of six TPS algorithms in computing entrance and exit doses
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Evaluation of six TPS algorithms in computing entrance and exit doses Yun I. Tan, 1,2a Mohamed Metwaly, 3 Martin Glegg, 3 Shaun P.
More informationRadiotherapy physics & Equipments
Radiotherapy physics & Equipments RAD 481 Lecture s Title: Beam Modification devices in Radiotherapy Dr. Mohammed Emam Vision :IMC aspires to be a leader in applied medical sciences, health care education
More informationMath Review. Overview
Drawn by Steve Yan, CMD Math Review Shirley Ann Pinegar-Johnston MS RT(R)(T)CMD Overview Drawn by Steve Yan, CMD Divergence & Intensity DIVERGENCE X-rays travel in Straight but Divergent lines INTENSITY
More informationANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL
ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL FOR THE DEGREE MASTER OF ARTS BY CHARLES TRAVIS WEBB
More informationOUTPUT at dmax in tissue 8x8 20x20 At 100cm SSD At 100cm SAD
RAPHEX 2001 T1. All of the following are true regarding Percentage Depth Dose (PDD) except: A. Increases with increasing energy. B. Depends on field size. C. Is the dose at depth expressed as a % of the
More informationIMRT in clinical practice at the UMC-Utrecht. Utrecht
3 IMRT in clinical practice at the UMC-Utrecht Utrecht Clinical use of IMRT Improvement of local control without increased incidence of normal tissue complications + Dose escalation in the prostate Step-and
More informationAbstract: Introduction:
Are photon and electron beam calibrations more consistent with TG-51 than with TG-21? Ramesh C. Tailor, William F. Hanson, Nathan Wells, and Geoffrey S. Ibbott. U.T. MD Anderson Cancer Center, Houston
More informationScientific Journal of Nuclear Medicine & Radiation Therapy
Scientific Journal of Nuclear Medicine & Radiation Therapy Research Article Radiotherapy Dose Calculation Program, for Monitoring and Commissioning Treatment Planning Systems - Ioannis Vamvakas*, Stella
More informationTEST REPORT BEAMSCAN WITH HALCYON
TEST REPORT BEAMSCAN WITH HALCYON UZ Leuven, November 20 th, 2017 + = Ideal BEAMSCAN tank dimensions to fit perfectly into the Halcyon bore Fast tank positioning using the Halcyon lasers, no electronics
More informationCommissioning and quality assurance for the treatment delivery components of the AccuBoost system
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 2, 2015 Commissioning and quality assurance for the treatment delivery components of the AccuBoost system Ileana Iftimia, 1,2a Mike Talmadge,
More informationThe dose distribution of medium energy electron boosts to the laryngectomy stoma
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 1, WINTER 2001 The dose distribution of medium energy electron boosts to the laryngectomy stoma Ellen D. Yorke* Memorial Sloan-Kettering Cancer
More informationDisclosures. What Is So Hard? End of ITV: Gating is the Best ITV Killer 8/3/2016. Daniel A. Low, Ph.D. UCLA. Varian Grant Siemens Grant Accuray Grant
Real Timefullness 8/3/2016 End of ITV: Gating is the Best ITV Killer Daniel A. Low, Ph.D. UCLA Varian Grant Siemens Grant Accuray Grant Disclosures What Is So Hard? Light Ultrasound MRI X Ray Planar X
More informationAbsorption measurements for a carbon fiber couch top and its modelling in a treatment planning system
Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system G. Kunz, F. Hasenbalg, P. Pemler 1 1 Klinik für Radio-Onkologie und Nuklearmedizin, Stadtspital Triemli
More informationUser Manual. RW3 Slab Phantom T29672 and T
User Manual RW3 Slab Phantom Contents Operating Manual Technical Manual Service Manual D188.131.00/07 en 2015-07 Hn General Information General Information The product bears the CE-mark "CE" in accordance
More informationRADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER
RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER Proven Rad-Check technology specifically designed to provide you with the ultimate in versatility and cost-effective operation. Fast and easy
More informationINTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS
A9744593 INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSLS Ladislav Czap, Georg Matscheko and Pedro Andreo osimetry and Medical Radiation Physics Section, ivision
More informationTechnical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry
Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry Bryan R. Muir a) and Malcolm R. McEwen Measurement Science and Standards, National Research Council of
More informationAAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams
AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams Peter R. Almond Brown Cancer Center, Louisville, Kentucky 40202 Peter J. Biggs Department of Radiation Oncology,
More informationVarian HDR surface applicators commissioning and clinical implementation
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Varian HDR surface applicators commissioning and clinical implementation Ileana Iftimia, 1,2,a Andrea B. McKee, 1,2 and Per H. Halvorsen
More informationHAEMODYNAMIC MONITORING SETUP IN THE CATHETERISATION LABORATORY MAHAT SAPION
HAEMODYNAMIC MONITORING SETUP IN THE CATHETERISATION LABORATORY MAHAT SAPION Haemodynamic Monitoring Functions of monitoring to detect that there is a problem blood pressure compensatory sympathetic stimulation
More information3D Treatment Planning and verification with hand calculations
3D Treatment Planning and verification with hand calculations Randy Holt, PhD, CEO Pacific Crest Medical Physics, Inc Chico, CA Dosimetric Calculations Calibration and Measurement Conditions Patient Conditions
More informationHow to Design Medical Accelerator Systems for Reliability: IBA PT System
How to Design Medical Accelerator Systems for Reliability: IBA PT System Yves Jongen Founder & Chief Research Officer Ion Beam Applications s.a. Belgium 1 Outline A short introduction of a proton therapy
More informationComparison of ionization chambers of various volumes for IMRT absolute dose verification
Comparison of ionization chambers of various volumes for IMRT absolute dose verification Leonid B. Leybovich, a) Anil Sethi, and Nesrin Dogan Department of Radiation Oncology, Loyola University Medical
More informationUSER MANUAL PROGRAM VERSION 1.0.1
R WWW..COM REF 92730-92731 USER MANUAL PROGRAM VERSION 1.0.1 STANDARD IMAGING, INC. 7601 Murphy Drive Middleton, WI 53562-2532 Dec / 2004 2004 Standard Imaging, Inc. DOC #80464-01 TEL 800.261.4446 TEL
More informationSurface buildup dose dependence on photon field delivery technique for IMRT
Surface buildup dose dependence on photon field delivery technique for IMRT Shigeru Yokoyama, Peter L. Roberson, Dale W. Litzenberg, Jean M. Moran, and Benedick A. Fraass Department of Radiation Oncology,
More informationReal-time Analysis of Industrial Gases with Online Near- Infrared Spectroscopy
Real-time Analysis of Industrial Gases with Online Near- Infrared Branch 5 Petrochemical, Biofuels Keywords Process, NIR, Gases, Gas Cell, Ethylene, Carbon Dioxide, Propane, Butane, Acetylene Summary Near-infrared
More informationA dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions
A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions Wendel Dean Renner a) Math Resolutions, LLC, 5975 Gales Lane, Columbia,
More informationValidation of SEC/GPC Software
Validation of SEC/GPC Software M. Gray, P. KiIz*, G. Reinhold* PSS-USA, 13531Cedar Creek Lane, Silver Spring, MD 20904 *PSS Polymer Standards Service GmbH, POB 3368, D-55023 Mainz Poster presented at the
More informationElectronic Oncology Systems the Management and the Safety Measures. Hansen Chen, Director, TDSI
Electronic Oncology Systems the Management and the Safety Measures Hansen Chen, Director, TDSI Radiation Oncology Department People Place Equipments Procedures Patients Hansen Chen 2 Radiation Oncology
More informationReview of backscatter measurement in kilovoltage radiotherapy using novel detectors and reduction from lack of underlying scattering material
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 6, 2013 Review of backscatter measurement in kilovoltage radiotherapy using novel detectors and reduction from lack of underlying scattering
More informationCOMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC
COMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC by TAN ZI XIANG Thesis submitted in fulfillment of the requirements for the degree of Master of Science (Medical Physics) February 2015 ACKNOWLEDGEMENT
More informationIPEM Working Party: D IThwaites(Chair), A R DuSautoy, T Jordan, MRMcEwen, A Nisbet, A E Nahum and W G Pitchford
INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 48 (2003) 2929 2970 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(03)65222-X The IPEM code of practice for electron dosimetry for radiotherapy beams of
More information7.65 ±0.05 mm of lead. The maximum energy of the bremsstrahlung photons was calibrated using the threshold for the D(-y,n)H reaction at
Am 7Roentgenolia6:ia6oia65, 1976 CENTRAL AXIS DEPTH DOSE FOR A 2.5 MV VAN DE GRAAFF GENERATOR DAVID W. ANDERSON, DAVID E. RAESIDE, REBA I. ADAMS, AND MYRON R. GOEDE 2 ABSTRACT: Central axis percentage
More informationVariSource High Dose Rate Afterloader Procedures. For Performing Breast Brachytherapy with. The SAVI TM Applicator
VariSource High Dose Rate Afterloader Procedures For Performing Breast Brachytherapy with The SAVI TM Applicator Contents I. CT Evaluation (~1 week prior to insertion) II. SAVI Applicator Implantation
More informationPlanning Assignment (3 field rectum)
Megan Sullivan Clinical Lab Rectum April 25, 2016 Planning Assignment (3 field rectum) Use a CT dataset of the pelvis. Create a CTV by contouring the rectum (start at the anus and stop at the turn where
More informationCHAPTER-2. Wedge filters are the most important beam modifiers used in
CHAPTER-2 2. WEDGE FLTERS 2.1. NTRODUCTON Wedge filters are the most important beam modifiers used in radiotherapy which are placed in the path of photon beams to modify their isodose diistributions. The
More informationNumerical Tools for LCLS-II Vacuum Systems
Numerical Tools for LCLS-II Vacuum Systems G. Lanza, D. Gill, SLAC National Accelerator Laboratory AVS, November 2 nd 2017 Doc. N. SLAC-PUB-17166 Work described in this presentation is supported by the
More informationA clinical comparison and analysis between conventional MLC based and solid compensator based IMRT treatment techniques
The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2009 A clinical comparison and analysis between conventional MLC based and solid compensator based IMRT treatment
More informationClinical Commissioning of a Pencil Beam Scanning Treatment Planning System for Proton Therapy
Clinical Commissioning of a Pencil Beam Scanning Treatment Planning System for Proton Therapy Jatinder Saini, MS 1 ; Ning Cao, PhD 2 ; Stephen R. Bowen, PhD 2,3 ; Miguel Herrera, BS 1 ; Daniel Nicewonger,
More informationIAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy
IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy Guidelines on standardized procedures at Secondary Standards Dosimetry Laboratories (SSDLs) and hospitals March 2002 The
More informationVariable Depth Bragg Peak Test Method
Variable Depth Bragg Peak Test Method C. Foster (FCS), P. O Neill, B. Reddell (NASA/JSC), K. Nguyen, B. Jones (Jacobs Technology), N. Roche, S. Buchner (NRL) Acknowledgement: A. Rusek, M. Sivertz, and
More informationQuality control of breast ultrasound: How to use a breast ultrasound quality assurance phantom?
Quality control of breast ultrasound: How to use a breast ultrasound quality assurance phantom? Poster No.: C-2282 Congress: ECR 2013 Type: Educational Exhibit Authors: S. J. Oh, B. K. Seo, S. E. Song,
More informationGas mixtures. Individual solutions specifically for your application
Gas mixtures Individual solutions specifically for your application For many applications besides pure gases also defined gas mixtures are required. For many routine applications, from operating gases
More informationMIT Analytical X-Ray Safety Policies and Procedures
5th Edition 2014 MIT Analytical X-Ray Safety Issued by: MIT Radiation Protection Committee Contents 1.0 Program Overview...1 2.0 Purpose...1 3.0 Scope and Application...2 4.0 Radiation Protection Program
More informationTreatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation
Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation By Ruth Ann M. Good, CMD. R.T. (R)(T) Virginia Commonwealth University, Richmond,
More informationJefferson Lab Bubble Chamber Experiment Update and Future Plans. C(a,g) 16 O. Claudio Ugalde
Jefferson Lab Bubble Chamber Experiment Update and Future Plans 12 C(a,g) 16 O Claudio Ugalde Collaboration Whitney Armstrong Melina Avila Kevin Bailey Tom O Connor Ernst Rehm Seamus Riordan Brad DiGiovine
More informationRadiation Producing Machines Radiation Safety Training for X-ray Diffraction Units
INTRODUCTION Radiation Producing Machines Radiation Safety Training for X-ray Diffraction Units RADIATION SAFETY GUIDE IMAGING AND MICROSCOPY FACILITY University of California, Merced Training Document
More informationTechnical protocol of APMP key comparison for measurement of air kerma for 60 Co (APMP.RI(I)-K1.1)
Technical protocol of APMP key comparison for measurement of air kerma for 60 Co (APMP.RI(I)-K1.1) 1. Introduction A regional APMP.RI(I)-K1 comparison of the standards for air kerma for 60 Co occurred
More information