7.65 ±0.05 mm of lead. The maximum energy of the bremsstrahlung photons was calibrated using the threshold for the D(-y,n)H reaction at
|
|
- Lorraine Daniel
- 6 years ago
- Views:
Transcription
1 Am 7Roentgenolia6:ia6oia65, 1976 CENTRAL AXIS DEPTH DOSE FOR A 2.5 MV VAN DE GRAAFF GENERATOR DAVID W. ANDERSON, DAVID E. RAESIDE, REBA I. ADAMS, AND MYRON R. GOEDE 2 ABSTRACT: Central axis percentage depth dose values and isodose curves for the bremsstrahlung beam from a 2.5 MV Van de Graaff generator were measured with a water phantom at 100 cm targettosurface distance. Tissueair ratios were calculated from the central axis depth dose data. Use of the 2.5 MV percentage depth dose values are necessary for treatment planning since they are substantially larger than the values given in compilations for 2.0 MV beams. INTRODUCTION Increased output and increased percentage depth dose are two advantages to operation of a Van de Graaff bremsstrahlung generator at 2.5 MV rather than 2.0 MV. The generator used for radiation therapy at the University of Oklahoma Health Sciences Center was recently modified for 2.5 MV operation with the result that the dose rate was increased by a factor of I.8. Subsequent to the modification, measurements required for obtaining central axis percentage depth dose tables and isodose curves for a variety of fields were made. The complete tables and several of the isodose curves are presented here since they may prove useful in other institutions anticipating similar modifications. EXPERIMENTAL APPARATUS AND TECHNIQUE The generator used as the radiation source was an electrostatic electron accelerator (Van de Graaff) manufactured by High Voltage Engineering Corporation ( Burlington, Massachusetts) and operated at 2.5 MV. A tungsten transmission target 2.3 mm thick and mounted on a 1.8 mm copper backing in the end ofthe accelerator tube was used to produce the radiation beam. No field flattener was used, but additional filtration due to cooling water jacket, an exposure rate monitor, and a light field mirror was sufficient to produce a beam with a halfvalue thickness of 7.65 ±0.05 mm of lead. The maximum energy of the bremsstrahlung photons was calibrated using the threshold for the D(y,n)H reaction at MeV [I]. A sample of 25 g of heavy water was irradiated at various indicated terminal voltages with a silver sheet wrapped around the teflon sample holder. The f3 activity induced in the silver was counted after the irradiation as the mdicator of reaction yield. A linear plot of yield versus indicated voltage was easily extended to zero yield to verify threshold terminal voltage to within ± io kv. The focal spot size for the accelerator beam was measured using a device made of laminated strips of lead alternating with paper spacers in a holder i cm long with a film packet placed at the end. The first exposure was followed by a second with the lamination direction rotated through 90#{176}. The effective focal spot size was 2.5X3.0 mm on the film. Dose profile measurements for the bremsstrahlung beam were made at ioo cm targettosurface distance for a I0XI0 cm field and a 25 X 25 cm field. In separate measurements, lithium borate thermoluminescent dosimeters and later a low sensitivity nonscreen radiographic film were sandwiched between pieces of plexiglass 0.5 cm thick. Film density measurements were corrected using the characteristic curve for the film for 60Co radiation. The 1 Department of Radiological Sciences, University of Oklahoma Health Sciences Center, Oklahoma City, Oklahoma Present address: St. Francis Hospital, Tulsa, Oklahoma
2 CENTRAL AXIS DEPTH DOSE 1261 measurements both showed a narrow penumbra region at the field edges with the dose falling from 90% to 10% of the central axis value within I cm of lateral distance. A tail on the dose profile extended laterally from the field edge with relative dose less than 2% of the central axis value at 4 cm lateral distance. Since the transmission of the 7.6 cm lead collimator blocks was less than 1%, the tail must be partially attributed to scatter from the collimator block faces. Since no flattening filter was used, the dose profile was noticeably forward peaked. At central axis and laterally to within I cm of the field edge, the dose was well described by the form D(x) = D(o) exp (ax2). The variable x is defined here as the radial distance in centimeters from the central axis at the usual treatment distance. The coefficient a is the peaking parameter. A least square fit of the 2.5 MV data to this function gave the value avdg (6. ± o.8)io. This is substantially larger than the peaking parameter for our 60Co teletherapy machine obtained in a similar manner at 8o cm without trimmers. For that machine, ac0 = (.o ± o.6)io. The effects of the forward peaking on the isodose curves are discussed later. Determination of the depth necessary for electron buildup was accomplished using thermoluminescent dosimeters made of lithium borate imbedded in teflon. The dosimeters were io.o mm in diameter and only 0.1 mm thick. They were placed in a phantom consisting of plexiglass sheets of I.0 mm thickness. The arrangement for irradiation was such that no two dosimeters overlapped. Irradiation at 2.5 MV and 100 cm distance with a lox 10 cm field was chosen. The dosimeter response showed a maximum at a depth of 3.8 ±0.1 mm. Under the assumption that the depth of the dose maximum is proportional to electron range in the medium, a value of ±0. mm was calculated for the buildup depth in water [2]. Dose versus depth data were taken with a computer controlled system (Artonix, Inc., St. Louis, Missouri). A pair of ionization chambers 0.5 cm in diameter and i. cm in length with tissueequivalent walls were used. The chambers used meet the recommended standards of the ICRU [s]. Mounting rods were made of teflon. One chamber served as the beam intensity monitor and was located at the periphery of the field. The other chamber served as the underwater probe and was moved about in a plexiglass water tank with a surface area of 52 X 53 cm and a depth of 50 cm. This volume was sufficient to allow ample peripheral water even for our largest field (20 X 20 cm) and deepest measurement ( 28 cm). The water temperature was 21#{176} C (densi ty = g/cm3). Two shielded preamplifiers were used to drive the chamber signals to the electrometer located outside of the radiation room. Data points were taken as the ratio of the response of the underwater probe to that of the monitor probe to minimize the effects of fluctuating exposure rates. With the automated scan several minutes were required to acquire a full set ofdata for a given field. The true probe depths (controlled remotely) were found to be within. mm of the desired depths in all situations. The targettosurface distance for the irradiations was determined to be 100.0±0.5 cm. The effects of scattered radiation on the preamplifiers and cables were tested by placing the movable probe near the bottom of the tank and outside the field limits for a I 5 X I 5 cm field. The response was small but measurable with the beam on. A similar reading taken with a thick cylindrical lead cap over the chamber was reduced substantially. This indicated that the preamplifier and cable response due to scattered radiation was very small. The magnitude of the preamplifier and cable response during operation is estimated to be less than i% of central axis values at com
3 1262 ANDERSON ET AL. I I I I F I I I I 1041 U) 0 0 I 0 z U ,( The ion chamber response ratios ob 3( 2( I.l l t_ I IZ DEPTH FIG. I.Central axis percentage depth doses calculated from functional fit. Radius of data points indicates standard deviation. parable depths for moderate field size and depth. As an overall check on the experimental technique, several runs were made at 2.0 MV. The central axis percentage depth dose data obtained were similar to values in standard compilations [4]. The magnitude of the deviations between our raw values and the accepted values was i..% of the local dose for depths between i and 20 cm for a ioxio cm field. The data runs at 2.5 MV were then made. A total of 40 separate and complete isodose curves were obtained with i,440 individual measurements for central axis response ratios. The standard deviations for the central axis values for a single field and depth were less than i% on the average. (cm) tamed at the central axis were averaged at each depth for each field size investigated and normalized to ioo.o% at the depth of 0.45 cm to produce the percentage depth dose values P(d). To minimize subjective errors, a regression analysis technique was used to generate smooth curves from the data. When such an analysis is performed with a computer, a quantitative measure of the precision of the fit can easily be calculated as part of the program. A similar measure of the precision of the fit has rarely been given when curves are drawn by hand through raw data points. In this case, the values were fit by cornputer to a smooth curve with the form /d+ P(d) = ioo() exp [y(d a\ RESULTS where d is the depth in water, m is the depth of the maximum dose, and a, f3, and
4 CENTRAL AXIS DEPTH DOSE I 263 Fic. 2.Isodose curves for I 2 X I 2 cm and 6 X 6 cm fields. Solid lines indicate shapes measured for 2. 5 MV beam. Dashed curves indicate shapes for beam from 60Co machine at 8o cm without trimmers. Details in edgeoffield effects masked on these curves because ionization chamber used in measurement had 0.5 cm diameter. 7 are constants determined by the cornputer analysis for each field. Scattered radiation in the field at central axis is known to be a function of the areaperimeter ratio [57] defined as (zl/p)d= field area at deph d/field perimeter at depth d. Because of this, the values of a, fi, and y were fit by computer to a power series in.1/p so that all field sizes could be represented. Using the results of this procedure, the final percentage depth dose values, defined as F[d, (il/p)m], are calculable for all depths and fields. Values of this percentage depth dose are tabulated in table for common square fields. It should be noted that the depth dose values for the oxo cm field in table are the result of extrapolation to A/P=o of measurements for 5everal small fields. To check the analysis technique, the calculated values taken from the table were compared with the data for each field. The precision of the fit was indicated by the average magnitude of the difference between these quantities. For our data the average deviation was only o.8% of the local values, and the largest deviation was I.% of the local value. Thus the fitting procedure was adequate for the data obtained. Figure I shows two of the continuous central axis percentage depth dose curves plotted from the calculated ppints in table I and from the function. The experimental data points are also given on the figure for comparison. The functional form and fitting technique is shown to be adequate. Table 2 contains the values of the central axis tissueair ratios for the 2.5 MV beam. These numbers were calculated from the smooth curves for percentage depth dose. For this calculation, the following expres
5 1264 ANDERSON ET AL. TABLE I PERCENTAGE DEPTH DOSES AT 2.5 MV FIELD (CM) AND BACKSC AFTER FACTOR oxo 4X4 Xg 6X6 8X8 ioxio 12X12 Xi aoxao DEPTH (CM) I.025 i.o i o ! o o o o I #{231}o o o I o i8.o i i io NOTE.IOO cm. targettosurface distance. sion was used [8]: T[d, (A/P)d] = (TSD + d B[(A/P)] P[d, (A/P)m] \TSD+m/ ( 100 In this equation the tissueair ratio (T) is a function of depth d and the area perimeter ratio at that depth, (il/p)d. The symbol TSD denotes the targettosurface distance (Ioo cm), and m is the depth of the maximum dose. Note that the relationship between the areaperimeter ratios is /100 + d (A/P)d (A/P)m ( \I00 + m The backscatter factors B[(il/P)m] were measured with the chamber described previously. The results in tables I and 2 can be cornpared with the published results for 2.0 MV beams [4] and 60Co beams at 100 cm. Central axis percentage depth dose values at 2.5 MV are generally from 1% to #{231}% larger than those for 2.0 MV beams for most fields at substantial depths. However, the 2.5 MV percentage depth dose values generally are not as large as values for 60Co beams at 100 cm distances. Figure 2 shows isodose curves obtained for a 6 x 6 cm field and a I 2 X I 2 cm field. As expected, the 60Co curves show considerably more penumbra than the 2.5 MV curves. However, the 2.5 MV isodose lines show more forward peaking than the 60Co lines. As depth increases, the effects of penumbra and forward peaking become more difficult to distinguish, and compa
6 CENTRAL AXIS DEPTH DOSE I 265 TABLE 2 TIssuEAIR RATIOS AT 2.5 MV FIELD (CM) DEPTH(CM) oxo 4X4 6X6 8X8 I0XI0 I2XII I5X15 20X #{149} o o i6.o 27.0 x8.o o.866 o.8o o o.858 o.8o o.6o ! rable isodose lines from the two sources have similar curvature. ACKNOWLEDGMENT Our thanks to Dr. Joseph C. Giarratano for suggesting the use of heavy water in the machine energy calibration. REFERENCES I. Mattauch L, Thiele J, Wapstra A. Mass excess values. Nuci Phys 67 : 140, Pages L, Bertel E, Joifre H, Sklavenitis L: Energy loss, range, and bremsstrahlung yield for io kev to 100 MeV electrons. Atomic Data :69125, ! 0.46! o.85i o.8io ! o.6o ! o x.o6 i.oo o.88o ! International Commission on Radiation Units and Measurements: Report 23, Measurement of absorbed dose in a phantom irradiated by a single beam of x or gamma rays, 1973, p 3 4. Cohen M, Jones DEA, Green D: Central axis depth dose data for use in radiotherapy. Br 7 Radiol I I, suppl : 475 I, I Sterling TD, Perry H, Katz L: Automation of radiation treatment planning. Br 7 Radiol 37: , Wrede DE : Practical aspects of area/perimeter. Bul/Am Assoc Phys Med 6 737k, I Wrede DE: Central axis tissueair ratios as a function of area/perimeter at depth. Phys Med Biol 17:548554, Johns HE, Cunningham JR: The Physics of Radiology. Springfield, Thomas, 1969
7 This article has been cited by: 1. A. Bridier, H. Beauvais, A. Dutreix On the use of a quality index to specify high energy photon beams. Radiotherapy and Oncology 5:1, [CrossRef]
Relative Dosimetry. Photons
Relative Dosimetry Photons What you need to measure! Required Data (Photon) Central Axis Percent Depth Dose Tissue Maximum Ratio Scatter Maximum Ratio Output Factors S c & S cp! S p Beam profiles Wedge
More informationField size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator
Journal of BUON 8: 55-59, 2003 2003 Zerbinis Medical Publications. Printed in Greece ORIGINAL ARTICLE Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator
More informationTHE development of more advanced techniques in radiotherapy,
1 A Geant4-based simulation of an accelerator s head used for Intensity Modulated Radiation Therapy F. Foppiano, B. Mascialino, M.G. Pia, M. Piergentili. Abstract We present a Geant4-based simulation,
More information13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.
13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13.1.1 TRAINING GOAL 13.1.1.1 Among the responsabilities of a medical
More informationConstancy checks of well-type ionization chambers with external-beam radiation units
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Constancy checks of well-type ionization chambers with external-beam radiation units Sara L. Hackett, 1a Benjamin Davis, 2 Andrew
More informationCHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS
47 CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 4.1 INTRODUCTION Advanced treatment techniques use optimized radiation beam intensities to conform dose distribution to the
More informationReview of fundamental photon dosimetry quantities
Review of fundamental photon dosimetry quantities Narayan Sahoo Main sources of the materials included in this lecture notes are: (1) Radiation Oncology Physics: A Handbook for Teachers and Students Edited
More informationCommissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN
22 Turkish Journal of Cancer Volume 37, No.1, 2007 Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN SATISH PELAGADE 1, KALPANA
More informationHigh Speed Direct SAD Radiosurgery Beam Scanner
Open Access Original Article DOI: 10.7759/cureus.20 High Speed Direct SAD Radiosurgery Beam Scanner Walter Nikesch 1, James M. Hevezi 2, Irene Monterroso 3, Daniel Navarro 4, James G. Schwade 5 1. CyberKnife
More informationThe Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.
FRANZCR Examination Part I Radiation Oncology Radiotherapeutic Physics Candidate No.: The Royal Australian and New Zealand College of Radiologists FRANZCR Examination Part I Radiation Oncology Radiotherapeutic
More informationZoubir Ouhib Lynn Cancer Institute
Zoubir Ouhib Lynn Cancer Institute April 2015 Speaker for ELEKTA Brachytherapy using miniature X-ray sources Energy
More informationMath Review. Overview
Drawn by Steve Yan, CMD Math Review Shirley Ann Pinegar-Johnston MS RT(R)(T)CMD Overview Drawn by Steve Yan, CMD Divergence & Intensity DIVERGENCE X-rays travel in Straight but Divergent lines INTENSITY
More informationRadiotherapy physics & Equipments
Radiotherapy physics & Equipments RAD 481 Lecture s Title: Beam Modification devices in Radiotherapy Dr. Mohammed Emam Vision :IMC aspires to be a leader in applied medical sciences, health care education
More informationGEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY
The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World Chattanooga, Tennessee, April 17-21, 2005, on CD-ROM, American Nuclear Society, LaGrange Park, IL (2005) GEANT4 SIMULATION OF
More informationExradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers
EXRADIN Ion Chambers Exradin Ion Chambers Exradin (EXacting RADiation INstrumentation) Ion Chambers have been built for over 33 years, are recognized by top research institutes and standards laboratories,
More informationClinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas
Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas Current Implementation Status Current Implementation Status 1600 1400 TOTAL 1494 of 1623 ACTIVE INSTITUTIONS
More informationCommissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale
Commissioning an IMRT System for MLC Delivery Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Taking the broad view of commissioning Commissioning elements Validating the dosimetry system Commissioning the
More informationOnly 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.
Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas 2009 AAPM Summer School Educational Objectives Improve your understanding of how to implement TG-51
More informationOutline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface
Chapter 11 reatment Planning Single Beams Radiation Dosimetry I Outline Basic terminology Curved contour surface correction (bolus, compensators, wedges) Oblique beam incidence Correction for tissue inhomogeneities
More informationAbsorption measurements for a carbon fiber couch top and its modelling in a treatment planning system
Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system G. Kunz, F. Hasenbalg, P. Pemler 1 1 Klinik für Radio-Onkologie und Nuklearmedizin, Stadtspital Triemli
More informationMonitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription
Monitor Unit Calculations Part 1 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Return to our first patient 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt
More informationOriginal Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams
Iranian Journal of Medical Physics Vol. 10, No. 3, Summer 2013, 147-155 Received: April 22, 2013; Accepted: July 21, 2013 Original Article Effective Point of Measurement in Cylindrical Ion Chamber for
More informationWEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS
Acta Radiologica Oncology 23 (1984) Fusc. 6 FROM THE RADIATION THERAPY DEPARTMENT, ANTON1 VAN LEEUWENHOEK HOSPITAL, THE NETHERLANDS CANCER INSTITUTE, AMSTERDAM, THE PHILIPS MEDICAL SYSTEMS DIVISION, BEST,
More informationProton Therapy QA Tools
Proton Therapy QA Tools A complete range of tools for absolute dosimetry, machine QA and patient plan verification Introduction During the past years more and more heavy particle therapy centers, especially
More informationANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL
ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL FOR THE DEGREE MASTER OF ARTS BY CHARLES TRAVIS WEBB
More informationDosimetric Calculations. Lonny Trestrail
Dosimetric Calculations Lonny Trestrail 20 October 2008 Objectives Dose Distribution Measurements PDD, OCR TAR, SAR, TPR, TMR, SPR, SMR Arc or Rotational Therapy Isodose Curves Point Dose Calculations
More informationPractical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11
MV Photon Dosimetry in the Clinic This document discusses both the scientific methods and the system of work required to calibrate a field instrument expanding upon the practical details of both.. Introduction
More informationChapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens
Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens The characteristic ionizing radiation response is defined by the relationship between the stimulus
More informationMonitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting
Monitor Unit Calculations Part 2 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Calculation of machine setting reference dose machine setting =, reference dose
More informationINTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS
A9744593 INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSLS Ladislav Czap, Georg Matscheko and Pedro Andreo osimetry and Medical Radiation Physics Section, ivision
More informationGuideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation
Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation This intercomparison will be carried out under the lead of KRISS with the cooperation of ARPANSA. This comparison has been approved
More informationClinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry
Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Tom Jordan Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE
More informationRadiation Producing Machines Radiation Safety Training for X-ray Diffraction Units
INTRODUCTION Radiation Producing Machines Radiation Safety Training for X-ray Diffraction Units RADIATION SAFETY GUIDE IMAGING AND MICROSCOPY FACILITY University of California, Merced Training Document
More informationAbstract: Introduction:
Are photon and electron beam calibrations more consistent with TG-51 than with TG-21? Ramesh C. Tailor, William F. Hanson, Nathan Wells, and Geoffrey S. Ibbott. U.T. MD Anderson Cancer Center, Houston
More informationPerturbation Correction of a Cylindrical Thimble-type Chamber. in a Graphite Phantom for 60Co Gamma Rays*
Rapport BIPM-86/l4 Perturbation Correction of a Cylindrical Thimble-type Chamber in a Graphite Phantom for 60Co Gamma Rays* by C.E. de Almeida**, A.-M. Perroche-Roux and M. Boutillon Bureau International
More informationSnapShot IMRT with compensators and FFF beams
SnapShot IMRT with compensators and FFF beams Vladimir Feygelman, PhD (1) Moffitt Cancer Center, Tampa, FL, USA Disclosure VF has a sponsored research agreement with.decimal. A short history of radiotherapy
More informationAAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams
AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams Peter R. Almond Brown Cancer Center, Louisville, Kentucky 40202 Peter J. Biggs Department of Radiation Oncology,
More informationComparison of ionization chambers of various volumes for IMRT absolute dose verification
Comparison of ionization chambers of various volumes for IMRT absolute dose verification Leonid B. Leybovich, a) Anil Sethi, and Nesrin Dogan Department of Radiation Oncology, Loyola University Medical
More informationCommissioning and periodic tests of the Esteya electronic brachytherapy system
Review paper Educational Activity Commissioning and periodic tests of the Esteya electronic brachytherapy system Cristian Candela-Juan, PhD 1, Yury Niatsetski, MSc 2, Zoubir Ouhib, PhD 3, Facundo Ballester,
More informationTG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009
TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 DISCLAIMER: This publication and associated spreadsheets and digital files are based on sources and
More informationTG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:
Practical Implementation of TG-61: II. Guidelines for clinical implementation of TG-61 J.P. Seuntjens Medical Physics Unit McGill University, Montreal General Hospital Montréal, Canada jseuntjens@medphys.mcgill.ca
More informationA beam-matching concept for medical linear accelerators
Acta Oncologica, 2009; 48: 192200 ORIGINAL ARTICLE A beam-matching concept for medical linear accelerators DAVID SJÖSTRÖM, ULF BJELKENGREN, WIVIANN OTTOSSON & CLAUS F. BEHRENS Copenhagen University Hospital,
More informationUnderwater measurement of photosynthetically active radiation
Underwater measurement of photosynthetically active radiation Mark Blonquist, Damon Nitzel, and Bruce Bugbee Apogee Instruments, Inc., Logan, Utah Introduction Quantum sensors (photosynthetically active
More informationVictoreen B. Operators Manual. Image Intensifier Ion Chamber
Victoreen 6000-530B Image Intensifier Ion Chamber Operators Manual March 2005 Manual No 6000-530B-1 Rev. 4 2004, 2005 Fluke Corporation, All rights reserved. All product names are trademarks of their respective
More informationBadan Jadrowych Nuclear Research Institute. Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P Differential Recombination Chamber
Badan Jadrowych Nuclear Research Institute Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P00100517 Differential Recombination Chamber by M. Zel'chinskij K. Zharnovetskij Warsaw, June 1966 Translated
More information1. Question Answer cgy / MU cgy / MU 2. Question Answer
GS020113: Introduction to Medical Physics III: Therapy s to home work problem set assigned on 3/22/11 1. Question A patient is set up at 100 cm SSD on a 6 MVX machine. The dose rate at 10 cm in phantom
More informationEquation 1: F spring = kx. Where F is the force of the spring, k is the spring constant and x is the displacement of the spring. Equation 2: F = mg
1 Introduction Relationship between Spring Constant and Length of Bungee Cord In this experiment, we aimed to model the behavior of the bungee cord that will be used in the Bungee Challenge. Specifically,
More informationCommissioning of Elekta 6MV FFF Versa HD and Pinnacle
Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Poster No.: R-0044 Congress: Type: Authors: Keywords: DOI: 2014 CSM Scientific Exhibit L. Bendall, I. Patel, N. McGrath, C. Rowbottom; MANCHESTER/
More informationExperiment. THE RELATIONSHIP BETWEEN VOLUME AND TEMPERATURE, i.e.,charles Law. By Dale A. Hammond, PhD, Brigham Young University Hawaii
Experiment THE RELATIONSHIP BETWEEN VOLUME AND TEMPERATURE, i.e.,charles Law By Dale A. Hammond, PhD, Brigham Young University Hawaii The objectives of this experiment are to... LEARNING OBJECTIVES introduce
More informationJefferson Lab Bubble Chamber Experiment Update and Future Plans. C(a,g) 16 O. Claudio Ugalde
Jefferson Lab Bubble Chamber Experiment Update and Future Plans 12 C(a,g) 16 O Claudio Ugalde Collaboration Whitney Armstrong Melina Avila Kevin Bailey Tom O Connor Ernst Rehm Seamus Riordan Brad DiGiovine
More informationRADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER
RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER Proven Rad-Check technology specifically designed to provide you with the ultimate in versatility and cost-effective operation. Fast and easy
More informationTrial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water
Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water M R McEwen, A J Williams and A R DuSautoy Centre for Ionising Radiation Metrology National Physical
More informationAC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW
AC 2010-2145: MEASUREMENT OF HYDROGEN IN HELIUM FLOW Randy Buchanan, University of Southern Mississippi Christopher Winstead, University of Southern Mississippi Anton Netchaev, University of Southern Mississippi
More informationINSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE
TELEDYNE HASTINGS TECHNICAL PAPERS INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE Proceedings of FEDSM 98 1998 ASME Fluids Engineering
More informationON THE RELATIVE ADEQUACY OF SOME AMERICAN CHAMBERS FOR MEASUREMENTS IN THE GRENZ RAY RANGE*
ON THE RELATIVE ADEQUACY OF SOME AMERICAN CHAMBERS FOR MEASUREMENTS IN THE GRENZ RAY RANGE* VICTOR H. WITTEN, M.D., EDGAR N. GRISEWOOD, M.A. AND ELEANOR OSHRY, B.S.t The use of grenz radiation has been
More informationCHAPTER-2. Wedge filters are the most important beam modifiers used in
CHAPTER-2 2. WEDGE FLTERS 2.1. NTRODUCTON Wedge filters are the most important beam modifiers used in radiotherapy which are placed in the path of photon beams to modify their isodose diistributions. The
More informationMonitor Unit Verification for Small Fields
Monitor Unit Verification for Small Fields Patrick Higgins, Ph.D University of Minnesota Department of Radiation Oncology October 10, 2013 The Issues: How do we verify the monitor units calculated by the
More informationTotal Ionization Dose (TID) Test Results of the RH117H Positive Adjustable Low Dose Rate (LDR) LDR = 10 mrads(si)/s
Total Ionization Dose (TID) Test Results of the RH117H Positive Adjustable Regulator @ Low Dose Rate (LDR) LDR = 10 mrads(si)/s 18 December 2014 Duc Nguyen, Sana Rezgui Acknowledgements The authors would
More informationA dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions
A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions Wendel Dean Renner a) Math Resolutions, LLC, 5975 Gales Lane, Columbia,
More informationSurface buildup dose dependence on photon field delivery technique for IMRT
Surface buildup dose dependence on photon field delivery technique for IMRT Shigeru Yokoyama, Peter L. Roberson, Dale W. Litzenberg, Jean M. Moran, and Benedick A. Fraass Department of Radiation Oncology,
More informationBasics of Proton Therapy. Proton Treatment Planning and Beam Optimization
Basics of Proton Therapy Proton Treatment Planning and Beam Optimization SAM Educational Session, WE D BRB 2 Mark Pankuch, PhD Northwestern Medicine Chicago Proton Center Today s objectives Review the
More informationThe Gas Attenuator of FLASH
INTRODUCTION The Gas Attenuator of FLASH K. Tiedtke, N. von Bargen, M. Hesse, U. Jastrow, U. Hahn The experimental hall of the FLASH user facility is located approximately 30m behind the last dipole magnet
More informationCalibration Procedure
Management Procedure 2565 Revision: A Date Issued: October 30, 2017 Date Revised: Calibration Procedure DeFelsko Corporation PosiTector RTR-3D Replica Tape Reader Probe Table of Contents 1 Introduction
More informationTG-51: Experience from 150 institutions, common errors, and helpful hints
TG-51: Experience from 150 institutions, common errors, and helpful hints Ramesh C. Tailor, Ph. D., William F. Hanson, Ph.D., and Geoffrey S. Ibbott, Ph.D. Department of Radiation Physics, University of
More informationTotal Ionization Dose (TID) Test Results of the RH1021BMH-10 Precision 10V Low Dose Rate (LDR) LDR = 10 mrads(si)/s
Total Ionization Dose (TID) Test Results of the RH1021BMH-10 Precision 10V Reference @ Low Dose Rate (LDR) LDR = 10 mrads(si)/s 22 September 2014 Duc Nguyen, Sana Rezgui Acknowledgements The authors would
More informationarxiv: v1 [physics.ins-det] 16 Aug 2014
Preprint typeset in JINST style - HYPER VERSION RPC Gap Production and Performance for CMS RE4 Upgrade arxiv:1408.3720v1 [physics.ins-det] 16 Aug 2014 Sung Keun Park, Min Ho Kang and Kyong Sei Lee Korea
More informationClinical Implementation of Volumetric Modulated Arc Therapy
Clinical Implementation of Volumetric Modulated Arc Therapy UT M.D. Anderson Cancer Center Ramaswamy Sadagopan, Rebecca M. Howell, Weiliang Du and Peter Balter Definition 2 Intensity Modulated Arc therapy
More informationUser Manual. RW3 Slab Phantom T29672 and T
User Manual RW3 Slab Phantom Contents Operating Manual Technical Manual Service Manual D188.131.00/07 en 2015-07 Hn General Information General Information The product bears the CE-mark "CE" in accordance
More informationIN-PLACE DENSITY OF BITUMINOUS MIXES USING THE NUCLEAR MOISTURE-DENSITY GAUGE FOP FOR WAQTC TM 8
OF BITUMINOUS MIXES USING THE NUCLEAR MOISTURE-DENSITY GAUGE FOP FOR WAQTC TM 8 Scope This test method describes a test procedure for determining the density of bituminous mixes by means of a nuclear gauge
More informationAnemometry. Anemometry. Wind Conventions and Characteristics. Anemometry. Wind Variability. Anemometry. Function of an anemometer:
Anemometry Anemometry Function of an anemometer: Measure some or all of the components of the wind vector In homogeneous terrain, vertical component is small express wind as -D horizontal vector For some
More informationJOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 Comparison of monitor unit calculations performed with a 3D computerized planning system and independent hand calculations: Results
More informationTeacher s Assistant Laboratory Guide
Teacher s Assistant Laboratory Guide Joseph B. Hakanson 1, Warner Meeks 2, and Dr. Joshua L. Rovey 3 Missouri University of Science and Technology, Rolla, Missouri, 65409 These instructions will discuss
More informationThe MC commissioning of CyberKnife with MLC (Tips and Tricks)
The MC commissioning of CyberKnife with MLC (Tips and Tricks) Alain Guemnie Tafo, PhD UPMC Pinnacle, Harrisburg, PA Institute Gustave Roussy, Villejuif, France Disclaimer & Disclosure The views expressed
More informationELEVATION CORRECTION FACTORS FOR RADON MONITORS*
ELEVATION CORRECTION FACTORS FOR E-PERM@ RADON MONITORS* P. Kotrappa and L. R. Stiefl Abstract-E-PERMm radon monitors are based on the principle of electret ion chambers and are usually calibrated in a
More informationResidual Stresses in Railway Axles
Residual Stresses in Railway Axles ŘEHA BOHUSLAV 1,a,*, VÁCLAVÍK JAROSLAV 2,b NÁVRAT TOMÁŠ 3,c and DAVID HALABUK 3,d 1 GHH-BONATRANS GROUP a.s., Revoluční 1234, 735 94 Bohumín, Czech Republic 2 Výzkumný
More informationDEPARTMENT OF THE NAVY DIVISION NEWPORT OFFICE OF COUNSEL PHONE: FAX: DSN:
IMAVSBA WARFARE CENTERS NEWPORT DEPARTMENT OF THE NAVY NAVAL UNDERSEA WARFARE CENTER DIVISION NEWPORT OFFICE OF COUNSEL PHONE: 401 832-3653 FAX: 401 832-4432 DSN: 432-3653 Attorney Docket No. 85031 Date:
More informationTotal Ionization Dose (TID) Test Results of the RH1028MW Ultralow Noise Precision High Speed Operational Low Dose Rate (LDR)
Total Ionization Dose (TID) Test Results of the RH1028MW Ultralow Noise Precision High Speed Operational Amplifiers @ Low Dose Rate (LDR) LDR = 10 mrads(si)/s 20 February 2015 Duc Nguyen, Sana Rezgui Acknowledgements
More informationDEVELOPMENT OF A ROBUST PUSH-IN PRESSUREMETER
DEVELOPMENT OF A ROBUST PUSH-IN PRESSUREMETER Roger Failmezger, P.E. In-Situ Soil Testing, L.C., Lancaster, Virginia, USA ABSTRACT A push-in pressuremeter was developed using slotted steel casing with
More informationMIL-STD-883G METHOD
STEADY-STATE LIFE 1. PURPOSE. The steady-state life test is performed for the purpose of demonstrating the quality or reliability of devices subjected to the specified conditions over an extended time
More informationLab 1. Adiabatic and reversible compression of a gas
Lab 1. Adiabatic and reversible compression of a gas Introduction The initial and final states of an adiabatic and reversible volume change of an ideal gas can be determined by the First Law of Thermodynamics
More informationFluid Flow. Link. Flow» P 1 P 2 Figure 1. Flow Model
Fluid Flow Equipment: Water reservoir, output tubes of various dimensions (length, diameter), beaker, electronic scale for each table. Computer and Logger Pro software. Lots of ice.temperature probe on
More informationHydrostatics Physics Lab XI
Hydrostatics Physics Lab XI Objective Students will discover the basic principles of buoyancy in a fluid. Students will also quantitatively demonstrate the variance of pressure with immersion depth in
More information6. EXPERIMENTAL METHOD. A primary result of the current research effort is the design of an experimental
6. EXPERIMENTAL METHOD 6.1 Introduction A primary result of the current research effort is the design of an experimental setup that can simulate the interaction of a windmill with a vortex wake and record
More informationLab 1c Isentropic Blow-down Process and Discharge Coefficient
058:080 Experimental Engineering Lab 1c Isentropic Blow-down Process and Discharge Coefficient OBJECTIVES - To study the transient discharge of a rigid pressurized tank; To determine the discharge coefficients
More informationExperiment 13: Make-Up Lab for 1408/1420
Experiment 13: Make-Up Lab for 1408/1420 This is only for those that have approval. Students without approval will not be allowed to perform the lab. The pre-lab must be turned in at the beginning of lab.
More informationIPEM Working Party: D IThwaites(Chair), A R DuSautoy, T Jordan, MRMcEwen, A Nisbet, A E Nahum and W G Pitchford
INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 48 (2003) 2929 2970 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(03)65222-X The IPEM code of practice for electron dosimetry for radiotherapy beams of
More informationAnalysis of Pressure Rise During Internal Arc Faults in Switchgear
Analysis of Pressure Rise During Internal Arc Faults in Switchgear ASANUMA, Gaku ONCHI, Toshiyuki TOYAMA, Kentaro ABSTRACT Switchgear include devices that play an important role in operations such as electric
More informationThe Discussion of this exercise covers the following points:
Exercise 3-2 Orifice Plates EXERCISE OBJECTIVE In this exercise, you will study how differential pressure flowmeters operate. You will describe the relationship between the flow rate and the pressure drop
More informationLarry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin
Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin NCCAAPM Meeting Oct 11,2013 Larry DeWerd has a partial interest in Standard Imaging Talks from people who have been
More informationInstruction Manual. Pipe Friction Training Panel
Instruction Manual HL 102 Pipe Friction Training Panel 100 90 80 70 60 50 40 30 20 10 HL 102 Instruction Manual This manual must be kept by the unit. Before operating the unit: - Read this manual. - All
More informationIn air calibration of an HDR 192 Ir brachytherapy source using therapy ion chambers
Original Article Free full text available from www.cancerjournal.net In air calibration of an HDR 192 Ir brachytherapy source using therapy ion chambers ABSTRACT The Gammamed Plus 192 Ir high dose rate
More informationKill-painting of hypoxic tumors in charged particle therapy
Supplementary Information: Kill-painting of hypoxic tumors in charged particle therapy Walter Tinganelli 1,2,, Marco Durante 1,3 *, Ryoichi Hirayama 2, Michael Krämer 1, Andreas Maier 1, Wilma Kraft-Weyrather
More informationInvestigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy
University of South Florida Scholar Commons Graduate Theses and Dissertations Graduate School 7-14-2010 Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy
More informationWP2 Fire test for toxicity of fire effluents
Pagina 3 di 89 TRA SFEU VTT 22.6.2009 v.2 WP2 Fire test for toxicity of fire effluents Task 2.1.2 Development of small-scale test method for fire effluents Step 1: Use of modeling Plans according to DoW:
More informationEEC 686/785 Modeling & Performance Evaluation of Computer Systems. Lecture 6. Wenbing Zhao. Department of Electrical and Computer Engineering
EEC 686/785 Modeling & Performance Evaluation of Computer Systems Lecture 6 Department of Electrical and Computer Engineering Cleveland State University wenbing@ieee.org Outline 2 Review of lecture 5 The
More informationSPIRIT III Radiometer Saturation Effect
Utah State University DigitalCommons@USU Space Dynamics Lab Publications Space Dynamics Lab 1-1-1997 SPIRIT III Radiometer Saturation Effect Joseph J. Tansock Follow this and additional works at: https://digitalcommons.usu.edu/sdl_pubs
More informationMIT Analytical X-Ray Safety Policies and Procedures
5th Edition 2014 MIT Analytical X-Ray Safety Issued by: MIT Radiation Protection Committee Contents 1.0 Program Overview...1 2.0 Purpose...1 3.0 Scope and Application...2 4.0 Radiation Protection Program
More informationCommissioning and quality assurance for the treatment delivery components of the AccuBoost system
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 2, 2015 Commissioning and quality assurance for the treatment delivery components of the AccuBoost system Ileana Iftimia, 1,2a Mike Talmadge,
More informationOutline. Terminology. EEC 686/785 Modeling & Performance Evaluation of Computer Systems. Lecture 6. Steps in Capacity Planning and Management
EEC 686/785 Modeling & Performance Evaluation of Computer Systems Lecture 6 Department of Electrical and Computer Engineering Cleveland State University wenbing@ieee.org Outline Review of lecture 5 The
More informationInvestigation of Thermal Effects of CO 2 on Earth-Atmosphere System
Investigation of Thermal Effects of CO 2 on Earth-Atmosphere System Design Team Michel Beguin, Tim Bevins Dan Jakiela, Frank Kuchinski Design Advisor Prof. Yiannis Levendis Abstract This Capstone Project
More information