Bubble Viscometers A5-A1 / A-T / U-Z6 / Z7-Z10
|
|
- Clementine Griffin
- 6 years ago
- Views:
Transcription
1 Measure what you see. Bubble Viscometers A5-A1 / A-T / U-Z6 / Z7-Z10 Manual A member of Additives & Instruments
2 Bubble Viscometers A5-A1 / A-T / U-Z6 / Z7-Z10 Manual English PV-0500 Series PV-0600 Series BYK - Gardner USA 9104 Guilford Road Columbia, MD USA Phone Fax BYK-Gardner GmbH Lausitzer Str. 8 D Geretsried Germany Tel gardner ( ) Fax
3 DIRECTIONS FOR USING THE GARDNER BUBBLE VISCOMETER A-5 to A-1 CAT. NO. PV-0500 This instrument is based on the principle that the viscosity is directly proportional to the bubble speed. The viscosity of a sample of varnish or other transparent liquid is therefore determined by reference to the standard tubes in which an air bubble rises with the same speed as it does in a tube of the sample being tested. For this purpose two carefully calibrated empty tubes are furnished with each set. The bubble size of the sample is first adjusted as nearly as possible to that in the standard tubes. The tubes are then brought to a temperature of 25ºC (77ºF). This is essential due to the difference in the temperature coefficient of viscosity of mineral oils and varnishes. It is suggested that a water bath be used, which quickly brings the tubes to the proper temperature and insures greater accuracy. The bubble should be started from the flat end of the tube held in a true vertical position. The bottom meniscus of the bubble should be used for comparing the samples. It is only necessary that they start together. Owing to unavoidable differences in bubble sizes and in the shapes of the ends of the tubes, it may be necessary to raise one of the tubes so that the points of comparison will be on the same level. The observation is made after the bubbles are completely formes and moving upwards along a true vertical line. This takes place about 1½ to 2 cm from the bottom of the tube. The standard empty grade A tubes have been carefully selected to and I.D. of ± mm. and should be used for exact determinations. However, this close tolerance is relatively expensive, and for the sake of economy, tubes 2
4 having the same approximate diameter with a tolerance of ± mm. Grade B are used by many manufacturer for general factory control tests. This viscometer meets the requirements of ASTM D 154, D 555, D 1131, D 1545, D 1725, FTMS 14la-Method 4271, and AOCS Method Ka6. Approximate Viscosity in Stokes at 25º C (77º F) A A A A A
5 DIRECTIONS FOR USING THE GARDNER BUBBLE VISCOMETER A to T CAT. NO. PV-0510 This instrument is based on the principle that the viscosity is directly proportional to the bubble speed. The viscosity of a sample of varnish or other transparent liquid is therefore determined by reference to the standard tubes in which an air bubble rises with the same speed as it does in a tube of the sample being tested. For this purpose two carefully calibrated empty tubes are furnished with each set. The bubble size of the sample is first adjusted as nearly as possible to that in the standard tubes. The tubes are then brought to a temperature of 25ºC (77ºF). This is essential due to the difference in the temperature coefficient of viscosity of mineral oils and varnishes. It is suggested that a water bath be used, which quickly brings the tubes to the proper temperature and insures greater accuracy. The bubble should be started from the flat end of the tube held in a true vertical position. The bottom meniscus of the bubble should be used for comparing the samples. It is only necessary that they start together. Owing to unavoidable differences in bubble sizes and in the shapes of the ends of the tubes, it may be necessary to raise one of the tubes so that the points of comparison will be on the same level. The observation is made after the bubbles are completely formes and moving upwards along a true vertical line. This takes place about 1½ to 2 cm from the bottom of the tube. The standard empty grade A tubes have been carefully selected to and I.D. of ± mm. and should be used for exact determinations. 4
6 However, this close tolerance is relatively expensive, and for the sake of economy, tubes having the same approximate diameter with a tolerance of ± mm. Grade B are used by many manufacturer for general factory control tests. This viscometer meets the requirements of ASTM D 154, D 555, D 1131, D 1545, D 1725, FTMS 14la-Method 4271, and AOCS Method Ka6. Approximate Viscosity in Stokes at 25º C (77º F) A 0.54 B 0.69 C 0.93 D 1.03 E 1.23 F 1.52 G 1.60 H 2.11 I 2.24 J 2.68 K 2.88 L 3.02 M 3.35 N 3.45 O 3.78 P 4.09 Q 4.42 R 4.67 S 5.18 T
7 DIRECTIONS FOR USING THE GARDNER BUBBLE VISCOMETER U to Z-6 CAT. NO. PV-0540 This instrument is based on the principle that the viscosity is directly proportional to the bubble speed. The viscosity of a sample of varnish or other transparent liquid is therefore determined by reference to the standard tubes in which an air bubble rises with the same speed as it does in a tube of the sample being tested. For this purpose two carefully calibrated empty tubes are furnished with each set. The bubble size of the sample is first adjusted as nearly as possible to that in the standard tubes. The tubes are then brought to a temperature of 25ºC (77ºF). This is essential due to the difference in the temperature coefficient of viscosity of mineral oils and varnishes. It is suggested that a water bath be used, which quickly brings the tubes to the proper temperature and insures greater accuracy. The bubble should be started from the flat end of the tube held in a true vertical position. The bottom meniscus of the bubble should be used for comparing the samples. It is only necessary that they start together. Owing to unavoidable differences in bubble sizes and in the shapes of the ends of the tubes, it may be necessary to raise one of the tubes so that the points of comparison will be on the same level. The observation is made after the bubbles are completely formes and moving upwards along a true vertical line. This takes place about 1½ to 2 cm from the bottom of the tube. The standard empty grade A tubes have been carefully selected to and I.D. of ±
8 mm. and should be used for exact determinations. However, this close tolerance is relatively expensive, and for the sake of economy, tubes having the same approximate diameter with a tolerance of ± mm. Grade B are used by many manufacturer for general factory control tests. This viscometer meets the requirements of ASTM D 154, D 555, D 1131, D 1545, D 1725, FTMS 14la-Method 4271, and AOCS Method Ka6. Approximate Viscosity in Stokes at 25º C (77º F) U 6.66 V 8.89 W 10.7 X 12.0 Y 17.4 Z 22.9 Z Z Z Z Z Z
9 DIRECTIONS FOR USING THE GARDNER BUBBLE VISCOMETER Z-7 to Z-10 CAT. NO. PV-0560 This instrument is based on the principle that the viscosity is directly proportional to the bubble speed. The viscosity of a sample of varnish or other transparent liquid is therefore determined by reference to the standard tubes in which an air bubble rises with the same speed as it does in a tube of the sample being tested. For this purpose two carefully calibrated empty tubes are furnished with each set. The bubble size of the sample is first adjusted as nearly as possible to that in the standard tubes. The tubes are then brought to a temperature of 25ºC (77ºF). This is essential due to the difference in the temperature coefficient of viscosity of mineral oils and varnishes. It is suggested that a water bath be used, which quickly brings the tubes to the proper temperature and insures greater accuracy. The bubble should be started from the flat end of the tube held in a true vertical position. The bottom meniscus of the bubble should be used for comparing the samples. It is only necessary that they start together. Owing to unavoidable differences in bubble sizes and in the shapes of the ends of the tubes, it may be necessary to raise one of the tubes so that the points of comparison will be on the same level. The observation is made after the bubbles are completely formes and moving upwards along a true vertical line. This takes place about 1½ to 2 cm from the bottom of the tube. The standard empty grade A tubes have been carefully selected to and I.D. of ± mm. and should be used for exact determinations. However, this close tolerance is relatively expensive, and for the sake of economy, tubes having the same approximate diameter with a 8
10 tolerance of ± mm. Grade B are used by many manufacturer for general factory control tests. This viscometer meets the requirements of ASTM D 154, D 555, D 1131, D 1545, D 1725, FTMS 14la-Method 4271, and AOCS Method Ka6. Approximate Viscosity in Stokes at 25º C (77º F) Z Z Z Z
11
12 EDFIS 0408 Europe
Instructions Manual. Variable area flowmeter 60 M-1. Series. instrumentation for fluids. Technical Data. C-MI-60M1 Rev.: 0 English Version
instrumentation for fluids Variable area flowmeter Series 60 M-1 Instructions Manual Technical Data Accuracy: ± 2,5% of full scale value. Based on VDE/VDI 3513 Scales: In l/h, l/min, m³/h, kg/h, %, mm.,
More informationRotational Viscometers RP1 RP2
INTENDED USE RAYPA viscometers, models RP1 and RP2, are rotational viscometers for the quick determination of viscosity as specified in ISO 2555 and other ASTM norms. Instrument offers a precise digital
More informationNCERT. To study the variation in volume (V) with pressure (P) for a sample of air at constant temperature by plotting graphs between P and V, and
EXPERIMENT AIM To study the variation in volume (V) with pressure (P) for a sample of air at constant temperature by plotting graphs between P and V, and between P and V. APPARATUS AND MATERIAL REQUIRED
More informationbyko-visc Advanced Rotational Viscometer
Measure what you see. byko-visc Advanced Rotational Viscometer Manual A member of Additives & Instruments byko-visc Advanced Rotational Viscometer Manual 199 024 263 E 1608 BYK-Gardner GmbH Lausitzer
More informationApplication of Flex-Z release agent with a pneumatic gun
Application of Flex-Z release agent with a pneumatic gun Manual for spray-gun Granudan ApS Knud Bro Alle 4 DK-3650 Stenloese Denmark www.granudan.com www.granudan.cn Application of Release agent (Flex-Z)
More informationACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION
Application Note Doc. G8108-001 Rev. A - 23-Jul-02 ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION For more information regarding Digiquartz products contact: Paroscientific,
More informationOEM Liquid-Flowmeter
OEM Liquid-Flowmeter (With State-of-the-Art Technology) Rev.: 1.2 Seite 1 / 6 OEM-Liquid-Flowmeter Functional Principle The OEM Liquid-Flowmeters are working with ultrasound according to the Time-of-Flight
More informationPolarimeter tubes and accessories
Polarimeter tubes and accessories Opto-electronic measuring device since 1864 Polarimeter tubes of SCHMIDT+HAENSCH are designed for different applications. All tubes meet the ICUMSArecommendations, class
More informationTi-Pure TITANIUM DIOXIDE DETERMINATION OF SLURRY RHEOLOGY METHOD: T WP
Ti-Pure TITANIUM DIOXIDE DETERMINATION OF SLURRY RHEOLOGY METHOD: T4400.308.01.WP 1 TITANIUM DIOXIDE (Ti-Pure ) Rheology of Titanium Dioxide Slurries I. Principle Rheology is a measurement of the fluid
More informationThe water supply for a hydroelectric plant is a reservoir with a large surface area. An outlet pipe takes the water to a turbine.
Fluids 1a. [1 mark] The water supply for a hydroelectric plant is a reservoir with a large surface area. An outlet pipe takes the water to a turbine. State the difference in terms of the velocity of the
More informationTex-414-A, Air Content of Freshly Mixed Concrete by the Volumetric Method
by the Volumetric Method Contents: Section 1 Overview...2 Section 2 Apparatus...3 Section 3 Sampling Requirements...5 Section 4 Procedures...6 Section 5 Calculation...9 Section 6 Archived Versions...10
More informationAUTO RANGE Display of maximum measurable limits. Displayed Info Viscosity (cp or mpa s) % Torque Spindle Speed
Fungilab Quality + Low Budget = Viscolead Series Rotational Viscometers Viscolead One Original Fungilab Viscometer. Essential viscosity measurement. 2 years warranty Standard Delivery Viscometer head with
More informationMeasurements. Metric System
Measurements Measurements are basic to any scientific pursuit. A measurement has both a magnitude (numeric value) and a unit. Metric units are used in the sciences. Metric System In science, the metric
More informationVITLAB Volumetric instruments
VITLAB Volumetric instruments Standard Operating Procedure (SOP) VITLAB GmbH Linus-Pauling-Str.1 63762 Grossostheim Germany Telefon: +49 6026 97799-0 Fax: +49 6026 97799-30 E-mail: info@vitlab.com Internet:
More informationStandard Test Method for Viscosity of Paints and Related Materials by ISO Flow Cups 1
Designation: D 5125 97 (Reapproved 2005) Standard Test Method for Viscosity of Paints and Related Materials by ISO Flow Cups 1 This standard is issued under the fixed designation D 5125; the number immediately
More informationTransferpette -8/-12
Transferpette -8/-12 Testing Instructions (SOP) May 2009 1. Introduction The standard DIN EN ISO 8655 describes both the design and the testing of the piston operated pipette Transferpette 8/ 12. The following
More informationQuestions. theonlinephysicstutor.com. facebook.com/theonlinephysicstutor. Name: Edexcel Drag Viscosity. Questions. Date: Time: Total marks available:
Name: Edexcel Drag Viscosity Questions Date: Time: Total marks available: Total marks achieved: Questions Q1. A small helium balloon is released into the air. The balloon initially accelerates upwards.
More informationElcometer Viscosity Cup. Operating Instructions
Elcometer 2350 2351 2352 2353 2354 English Viscosity Cup Operating Instructions English is a registered trademark of Elcometer Limited. All other trademarks acknowledged. Copyright Elcometer Limited. 2009/2010.
More informationA Reliable and Tracer Gas Independent Leak Detector for Food Packages
19 th World Conference on Non-Destructive Testing 2016 A Reliable and Tracer Gas Independent Leak Detector for Food Packages Silvio DECKER 1 1 INFICON GmbH, Köln, Germany Contact e-mail: Silvio.Decker@inficon.com-
More informationL I Q UID C O N T R O L S. Mechanical Temperature Volume Compensators Models D5120, D5121, & D5122. Installation, Operation, and Parts M500-30:V3_0119
Mechanical Temperature Volume Compensators Models D5120, D5121, & D5122 Installation, Operation, and Parts L I Q UID C O N T R O L S An IDEX Energy & Fuels Business Mechanical Temperature Volume Compensators
More informationCH2250: Techniques in Laboratory Chemistry. Outline Measuring Mass Measuring Volume Significant figures. Mass Measurement
CH2250: Techniques in Laboratory Chemistry Outline Measuring Mass Measuring Volume Significant figures Mass Measurement Mass Measurement Measure mass not weight Mass is measured with a balance (a scale
More informationVITLAB micropipette. Standard Operating Procedure (SOP)
VITLAB micropipette Standard Operating Procedure (SOP) VITLAB GmbH Linus-Pauling-Str.1 63762 Grossostheim Germany tel: +49 6026 97799-0 fax: +49 6026 97799-30 info@vitlab.com www.vitlab.com Table of contents
More informationP3000 Series. Pneumatic Deadweight Testers Model P3000. Technical Data. Features
P3000 Series Pneumatic Deadweight Testers P3000 Technical Data Features Pressure ranges from Vacuum through 2 000 psi (140 bar) Accuracy better than 0.015 % of reading. (Increased accuracy option of 0.008
More informationWIKA INSTRUMENT CORPORATION
WIKA INSTRUMENT CORPORATION Instruction Manual DIFFERENTIAL PRESSURE GAUGE Series 1500 Series 1000 Series 300 WIKA Instrument Corporation 1000 Wiegand Boulevard Lawrenceville, GA 30043 1-888-945-2872 http://www.wika.com
More informationInstruction Manual. Pipe Friction Training Panel
Instruction Manual HL 102 Pipe Friction Training Panel 100 90 80 70 60 50 40 30 20 10 HL 102 Instruction Manual This manual must be kept by the unit. Before operating the unit: - Read this manual. - All
More information3 1 PRESSURE. This is illustrated in Fig. 3 3.
P = 3 psi 66 FLUID MECHANICS 150 pounds A feet = 50 in P = 6 psi P = s W 150 lbf n = = 50 in = 3 psi A feet FIGURE 3 1 The normal stress (or pressure ) on the feet of a chubby person is much greater than
More informationOPERATION AND MAINTENANCE MANUAL
OPERATION AND MAINTENANCE MANUAL Skum Bladder Tank Type Vertical. Table of Contents Page General Description 2 1. Installation 1.1 Safety 2 1.2 Transporting 2 1.3 Inspection 2 1.4 Filling Instruction 3-7
More informationGenerating Calibration Gas Standards
Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and
More informationLow Pressure Valve. Model Overview. Typical applications. Key features and benefits.
Low Pressure Valve Model 4047 Overview AMOT Model 4047 is used as a 2-way or 3-way high capacity, low pressure valve for settings from 0.007 bar (0.1) psi falling to 0.77 bar (11.2 psi) rising over several
More informationLaboratory #2 Pipetting Technique and Micropipette Calibration Skills=40 pts
Laboratory #2 Pipetting Technique and Micropipette Calibration Skills=40 pts Objectives: Upon completion of this unit, the student should be able to: 1. List and describe 3 categories of pipets. 2. List
More informationCommonwealth of Pennsylvania PA Test Method No. 742 Department of Transportation October Pages LABORATORY TESTING SECTION. Method of Test for
Commonwealth of Pennsylvania PA Test Method No. 742 Department of Transportation 14 Pages LABORATORY TESTING SECTION Method of Test for BITUMEN CONTENT OF BITUMINOUS CONCRETE MIXTURES (Pennsylvania Pycnometer
More informationROTARY VISCOMETER. Rotary Viscometers MYR V0/V1/V2 acc. to ISO 2555/ASTM (Brookfield method) SERIE VR 3000
ROTARY VISCOMETER Rotary Viscometers MYR V0/V1/V2 acc. to ISO 2555/ASTM (Brookfield method) SERIE VR 3000 VISCOMETERS VR 3000 The MYR viscometers, models V0, V1 and V2, are rotational viscometers for the
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 CORPORATE CONSULTING, SERVICE, & INSTRUMENTS, INC. University Park, 221 Beaver Street Akron, OH 44304 Michael Kent Warner Phone: 330 376 3600 CALIBRATION Valid
More informationAMS 6916 Board mount pressure sensor with ratiometric analog output
FEATURES Piezoresistive pressure sensor with amplified analog output Calibrated and temperature compensated Ratiometric voltage output, 0.5 4.5 V Digital signal conditioning, 12 bit output resolution Differential,
More informationAlpha Series. Main Features > Data displayed Selected speed r.p.m Selected spindle SP
Viscosímetros rotacionales Alpha series viscometers allow fast and accurate viscosity readings. They are low budget and easy to use. Alpha Series > Data displayed Selected speed r.p.m Selected spindle
More informationRotary Viscometers Rotary viscosimeters ST-2020
Rotary s Rotary viscosimeters ST-2020 INTRODUCTION Rheology is the study of the effects experimented in a substance when a mechanical force is applied on a it (flow and deformation) under different external
More informationResearch Grade Blood Pressure Transducer User's Manual
Research Grade Blood Pressure Transducer User's Manual Research Grade Blood Pressure Transducer, 110 VAC/60 Hz MA1 60-3002 Research Grade Blood Pressure Transducer, 220 VAC/50 Hz MA1 60-3003 WEEE/RoHS
More informationState of Nevada Department of Transportation Materials Division METHOD OF TEST FOR SAND EQUIVALENT
State of Nevada Department of Transportation Materials Division METHOD OF TEST FOR SAND EQUIVALENT SCOPE This test method is intended to determine the proportion of detrimental fines (dust or clay size
More informationTransferpette. Testing Instructions (SOP) 1. Introduction. October 1998
Transferpette Testing Instructions (SOP) October 1998 1. Introduction The standards ISO DIS 8655 and DIN 12650 describe both the design and the testing of the piston operated pipette Transferpette. The
More informationmax 80 cm / 2 6 Trittfrequenzmesser Capteur de Cadence Trapfrequentie Sensor Sensor de Cadencia Sensore di Cadenza Sensor de Cadência Kadencesensor
1 max 80 cm / 2 6 2a 2b 3 4 5 max 4 mm / 0 16 ENGLISH The Polar Cadence Sensor is designed to measure cadence when cycling. No other use is intended or implied. It is compatible with Polar Cycling Computers
More informationOPERATING INSTRUCTIONS
OPERATING INSTRUCTIONS Air Entrainment Meter ELE International Chartmoor Road, Chartwell Business Park Leighton Buzzard, Bedfordshire, LU7 4WG England phone: +44 (0) 1525 249200 fax: +44 (0) 1525 249249
More informationAdam Equipment. DENSITY DETERMINATION KIT for AFP, AAA and AAA/LE SERIES BALANCES OPERATOR S MANUAL
Adam Equipment DENSITY DETERMINATION KIT for AFP, AAA and AAA/LE SERIES BALANCES OPERATOR S MANUAL Adam Equipment Company 2002 ADAM EQUIPMENT CO. LTD. pn. 4144 Rev. A, October 2002 Adam Equipment Company
More informationTransferpette. Testing Instructions (SOP) 1. Introduction. May 2009
Transferpette Testing Instructions (SOP) May 2009 1. Introduction The standards DIN EN ISO 8655 describe both the design and the testing of the piston operated pipette Transferpette. The following Testing
More informationGas Sampling Fitting with MK2.1 and L-Probe
measuring controlling automation measuring controlling automation measuring controlling automation Page i von 8 Operation Manual Gas Sampling Fitting with MK2.1 and L-Probe Page ii von 8 Manufacturer MESA
More informationCoriolisMaster mass flowmeter Diagnostics, verification and proof test
ABB MEASUREMENT & ANALYTICS WHITE PAPER CoriolisMaster mass flowmeter Diagnostics, verification and proof test Automatic in situ verification of accuracy for CoriolisMaster mass flowmeters. By Frank Frenzel
More informationQuantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs
Quantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs Gravimetric Sample Preparation The Alternative to Volumetric Flasks From Volumetric to Gravimetric
More informationAging Cell, OFITE Style. Included Items
DRILLING FLUIDS EQUIPMENT For over 30 years OFI Testing Equipment (OFITE) has provided instruments and reagents for testing drilling fluids, well cements, completion fluids, and wastewater. In addition
More informationInstruction Manual Contact Pressure Vacuum Gauge
MS10 Instruction Manual Contact Pressure Vacuum Gauge Table of Contents 1. Safety Instructions 2. Intended Applications 3. Product Description and Functions 4. Installation 5. Commissioning 6. Maintenance
More informationBlue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION
Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION Install your Blue River Technologies Port-A-Poly mixing system in a clean dry area.
More informationThird measurement MEASUREMENT OF PRESSURE
1. Pressure gauges using liquids Third measurement MEASUREMENT OF PRESSURE U tube manometers are the simplest instruments to measure pressure with. In Fig.22 there can be seen three kinds of U tube manometers
More informationBSS 2 Bottle Sampling System. Instruction manual Version 1.8
BSS 2 Bottle Sampling System Instruction manual Version 1.8 Serial-No.:... Version-/Rev.-Date : 12.06.2012 Contents 1. General... 3 2. Safety Instructions... 4 3. Extent of Delivery... 5 4. Switch Panel...
More informationBT 800-OX Operating Manual
BT 800-OX Operating Manual DRUCK & TEMPERATUR Leitenberger GmbH Bahnhofstr. 33 72138 Kirchentellinsfurt GERMANY Tel.: +49-7121-90920-0 Fax: +49-7121-90920-99 www.druck-temperatur.de www.lr-cal.com INSTRUCTION
More informationGA / 3.02 TL 4 TL 6. Cat. No Calibrated leaks GA / 3.02 TL 4 TL 6. Cat. No
O P E R A T I N G I N S T R U C T I O N S GA 10.18 / 3.02 TL 4 TL 6 Cat. No. 155 65 155 66 Calibrated leaks O P E R A T I N G I N S T R U C T I O N S GA 10.18 / 3.02 TL 4 TL 6 Cat. No. 155 65 155 66 Calibrated
More informationand its weight (in newtons) when located on a planet with an acceleration of gravity equal to 4.0 ft/s 2.
1.26. A certain object weighs 300 N at the earth's surface. Determine the mass of the object (in kilograms) and its weight (in newtons) when located on a planet with an acceleration of gravity equal to
More informationDensity2Go. Hydrometers Pycnometers Digital Hydrometers. 3 Ways to Measure Density Know-How, Hints, and More
Density2Go Hydrometers Pycnometers Digital Hydrometers Know-How, Hints, and More Content 1. Introduction 4 2. Hydrometer 4 2.1 How to choose a hydrometer 5 2.2 How to use your hydrometer 5 2.3 Calibration
More informationMounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator
Self-operated Pressure Regulators Type 2357-31 Pressure Build-up Regulator with safety function and integrated excess pressure valve Type 2357-31 with non-return unit at port C Ports A and B with soldering
More informationGMC-I Service GmbH Calibration Center
2016 GMC-I Service GmbH Calibration Center DAkkS, ISO & Factorycalibration www.gmci-service.com From spare parts, reparation, rental service, maintenance of industrial health and safety standards, calibration
More informationNotes on equipment protection to DIN EN and NEMA
Technical information Notes on equipment protection to DIN EN 60529 and NEMA WIKA data sheet IN 00.18 General information This technical information describes the measures to prevent both the formation
More informationCOP700 / COP1400 Operating manual for comparison test pump COP700 / COP1400
COP700 / COP1400 Operating manual for comparison test pump COP700 / COP1400 Table of Contents 1 introduction...1 1.1 general product description... 1 1.2 schematics...1 1.3 specifications... 2 2 instrument
More informationGuide for the Sampling of Oil from Trench COTA Bushings for DGA
Instruction Guide for the Sampling of Oil from Trench COTA Bushings for DGA Forward This instruction is protected by copyright. Imparting to any third party and reproduction of any kind without written
More informationPMI Pulse Decay Permeameter for Shale Rock Characterization Yang Yu, Scientist Porous Materials Inc., 20 Dutch Mill Road, Ithaca NY 14850
PMI Pulse Decay Permeameter for Shale Rock Characterization Yang Yu, Scientist Porous Materials Inc., 20 Dutch Mill Road, Ithaca NY 14850 This document describes the application of Pulse Decay Permeameter
More informationFrom and
From http://www.school-for-champions.com/science/fluidpressure.htm and http://www.school-forchampions.com/science/fluidfloating.htm by Ron Kurtus, School for Champions Pressure in Fluids by Ron Kurtus
More informationSelf-operated Pressure Regulators Type 2405 Pressure Reducing Valve
Self-operated Pressure Regulators Type 2405 Pressure Reducing Valve ANSI version Application Pressure reducing valve for set points from 0.075 to 150 psi (5 mbar to 10 bar) Valve size NPS ½ to 2 1) (DN
More informationUSER S MANUAL. Model TES-SM-30V
USER S MANUAL Model TES-SM-30V SETTLEMENT MONITORING SYSTEM USING Vibrating wire Settlement Monitoring Sensors Doc. # WI 6002.94 Rev. 1 Printed 24/11/2009 1. Introduction The TRITECH Model TES-SM-30V settlement
More informationMECHANICAL BULKMETERS BM Series
PRO MEA 05 PRODUCTS - Measurement MECHANICAL BULKMETERS BM Series Positive displacement liquid measuring instruments Unequalled performance in measurement Free from installation effects Flow Rates of 25
More informationApplication Note Temperature dependence of viscosity of liquid crystal
Application Note dependence of viscosity of liquid crystal Industry : Organic Chemical Instrument : Viscometer Measurement method : Electro Magnetically Spinning Method s : 1. Scope Liquid crystal is a
More informationPrecision level sensing with low-pressure module MS
The task on hand Level sensing as it is understood in modern process automation is much more than simply "tank half full" or "tank a quarter full". Using suitable sensors, levels, inlets and outlets can
More informationFLOW RATES IN A SHORT LENGTH WET (BUBBLY) SEAL AND TEST RIG RE-DESIGN
TRC 32513/1519NS Year II MAKING A WET (FOAMY) SEAL AND ESTIMATING ITS DYNAMIC FORCE COEFFICIENTS FLOW RATES IN A SHORT LENGTH WET (BUBBLY) SEAL AND TEST RIG RE-DESIGN Qing Liu Research Assistant, Ph.D.
More informationAdam Equipment. DENSITY DETERMINATION KIT for AAA, ADA and ADP SERIES BALANCES OPERATOR S MANUAL
Adam Equipment DENSITY DETERMINATION KIT for AAA, ADA and ADP SERIES BALANCES OPERATOR S MANUAL ADAM EQUIPMENT CO. LTD. pn. 3478 Rev. E1, September 2000 Adam Equipment Company 2000 Adam Equipment Company
More informationSupratec Gesellschaft für Umweltund Verfahrenstechnik mbh OXYFLEX - M T / M T M E M B R A N E D I S C D I F F U S ER AERATION TECHNOLOGY
OXYFLEX - M T 2 3 5 / M T 3 0 0 M E M B R A N E D I S C D I F F U S ER UND GESELLSCHAFT FÜR UMWELT- VERFAHRENSTECHNIK MBH AERATION TECHNOLOGY von-drais-straße 7 D-55469 Simmern / Hunsrück Tel.: +49 6761
More informationBRML INM. National Metrology Institute. COOMET Comparison COOMET.L-S20 676/MD/15. Calibration of gauge blocks by comparison. Technical protocol
BRML INM National Metrology Institute COOMET Comparison COOMET.L-S20 676/MD/5 Calibration of gauge blocks by comparison Technical protocol Contents. INTRODUCTION... 3 2. ORGANISATION... 3 2.. Participants...
More informationSensor for Air Bubble Detection at Liquid Filled Tubes. SONOCHECK Type ABD06.xx. Operating Manual
Sensor for Air Bubble Detection at Liquid Filled Tubes SONOCHECK Type ABD06.xx Operating Manual Manufacturer: Model: Type: SONOTEC Ultraschallsensorik Halle GmbH Air Bubble Detector ABD06.xx SONOTEC Ultraschallsensorik
More informationPneumatic Hand Pumps with no Gauge Connection (scissor handle) AP0V (300 psi / 20 bar)
Pneumatic Hand Pumps with no Gauge Connection (scissor handle) AP0V (300 psi / 20 bar) The quickest and most economical way to perform pneumatic field pressure calibrations on pressure transmitters, switches
More informationDeadweight Tester / Pressure Balance High pressure up to 5,000 bar / 70,000 psi CPB5000-HP. Applications. Special features. Description. Rel.
High up to 5,000 bar / 70,000 psi CPB5000-HP Rel. 20150206 Applications Primary standard for defining the scale in a range up to 5,000 bar Reference instrument for testing, adjusting and calibrating measuring
More informationProduct Standard and Inspection Criteria for ilmasil PS Clear Fused Quartz Tubes
Product Standard and Inspection Criteria for Page 1 of 5 Version S-101-13 Index: C (June 2015) Application: This standard is applicable to direct drawn tubes produced with ilmasil PS base material 1 Optical
More informationBeamex PG PRESSURE GENERATORS. Beamex PGM PGV PGC PGHH PGPH PGL
Beamex PG PRESSURE GENERATORS 97 Beamex PGM PGV PGC PGHH PGPH PGL Beamex PG pressure generators are portable hand-operated pressure generators and ultimate field calibration pumps 98 Calibration pumps
More informationT 2523 EN Type 2406 Excess Pressure Valve Self-operated Pressure Regulators ANSI version
T 2523 EN Type 2406 Excess Pressure Valve Self-operated Pressure Regulators ANSI version Application Excess pressure valve for set points from 0.075 to 150 psi (5 mbar to 10 bar) Valve size NPS ½ to 2
More informationSeries 44 Self-operated Pressure Regulators. Type 44-1 B Pressure Reducing Valve. Type 44-6 B Excess Pressure Valve. ANSI version
Series 44 Self-operated Pressure Regulators Type 44-1 B Pressure Reducing Valve Type 44-6 B Excess Pressure Valve ANSI version Application Set points from 3 to 290 psi (0.2 to 20 bar) with valves ½ NPT
More informationMeasuring Mass and Volume
Measuring Mass and Volume Experiment 2 Expt 2 Measurement.wpd INTENT The purpose of this experiment is to introduce some fundamental aspects of the measurement making process as well as to introduce some
More informationNOVALYNX CORPORATION MODEL RAIN GAUGE CALIBRATOR INSTRUCTION MANUAL
NOVALYNX CORPORATION MODEL 260-2595 RAIN GAUGE CALIBRATOR INSTRUCTION MANUAL REVISION DATE: August 2018 Receiving and Unpacking Carefully unpack all components and compare to the packing list. Notify NovaLynx
More informationCover Page for Lab Report Group Portion. Drag on Spheres
Cover Page for Lab Report Group Portion Drag on Spheres Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 29 September 2017 Name 1: Name 2: Name 3: [Name 4: ] Date: Section number:
More informationAgilent Flowmeter ADM1000
Agilent Flowmeter ADM1000 Operating Instructions Notices Agilent Technologies, Inc. 2003 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval
More informationCOP700 / COP1400 Operating manual for comparison test pump COP700 / COP1400
COP700 / COP1400 Operating manual for comparison test pump COP700 / COP1400 Table of Contents 1 introduction...1 1.1 general product description...1 1.2 schematics...1 1.3 specifications...2 2 instrument
More informationPaper 2.2. Operation of Ultrasonic Flow Meters at Conditions Different Than Their Calibration
Paper 2.2 Operation of Ultrasonic Flow Meters at Conditions Different Than Their Calibration Mr William Freund, Daniel Measurement and Control Mr Klaus Zanker, Daniel Measurement and Control Mr Dale Goodson,
More informationBest Practice for Calibrating LTH Conductivity Instruments
Application Note Best Practice for Calibrating LTH Conductivity Instruments As accurate process measurement becomes an everyday requirement it is vital to be able to calibrate conductivity instruments
More informationLevel MEASUREMENT 1/2016
Level MEASUREMENT 1/2016 AGENDA 2 A. Introduction B. Float method C. Displacer method D. Hydrostatic pressure method E. Capacitance method G. Ultrasonic method H. Radar method I. Laser method J. Level
More informationUSER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve
USER'S GUIDE Nanoflow Probe 3-4mm Red Stripe ptfe Sleeve To Injector ZU1XC Fused Silica 20µm 90µm TTF115 ptfe Sleeve Fused Silica 25µm 375µm Micromass UK Limited Floats Road Wythenshawe M23 9LZ Tel: +44
More informationMODEL WEIGH MODULE
MODEL 65082 WEIGH MODULE INSTALLATION & OPERATING MANUAL P.O. Box 775 - Farmington, NH 03835 Tel: 603-755-3885 email: cands_nh@msn.com www.candscontrols.com Model 65023 Cantilever Beam Transducer Nickel-Plated
More informationPARTS & SERVICE. regulator (new) for NGW/ GBW applications. Rev. 02/2018 Id.No. DOK50026
PARTS & SERVICE Heating systems Retrofit instruction gas pressure regulator (new) for NGW/ GBW applications Rev. 02/2018 Id.No. DOK50026 Regulations / safety hints natural gas vehicles On behalf of the
More informationOperating instructions for variable area flow meter with needle valve. Model: UVR/UTR
Operating instructions for variable area flow meter with needle valve Model: UVR/UTR 1. Contents 1. Contents... 2 2. Note... 3 3. Instrument inspection... 3 4. Regulation use... 3 5. Operating principle...
More informationMembrane covered optical CO 2 sensor
Membrane covered optical CO 2 sensor - Manual - 30 mg/l CO 2, 4-20 ma output Exclusively Distributed & Supported By: www.edaphic.com.au info@edaphic.com.au Ph: 1300 430 928 Warning! Do never try to unscrew
More informationCoriolis Mass Flow Meter
Coriolis Mass Flow Meter TMFW Series Mini Type Coriolis Mass Flow Meter Xi an Tosilon Automation Co., Ltd No.299, Daqing Rd, Lianhu District, Xi'an Shaanxi, China Tel: +86-29-8823 8550 info@tosilon.com;
More informationAutomatic level / Optical level / Dumpy Level Instruction Manual www.level1lasers.com.au Contents Introduction 2 Diagram 3 Operation Instruction 4 Calibration Self-Check 5 Technical Specifications 6 Maintenance
More information1. The principle of fluid pressure that is used in hydraulic brakes or lifts is that:
University Physics (Prof. David Flory) Chapt_15 Thursday, November 15, 2007 Page 1 Name: Date: 1. The principle of fluid pressure that is used in hydraulic brakes or lifts is that: A) pressure is the same
More informationSERIES 500 VARIABLE RANGE PNEUMATIC DIFFERENTIAL PRESSURE TRANSMITTER
Man500e 09/2006 Installation Operation and Maintenance Instructions SERIES 500 VARIABLE RANGE PNEUMATIC DIFFERENTIAL PRESSURE TRANSMITTER INDEX 1. INSTALLATION 2. COMPRESSED AIR SUPPLY 3. FLOW MEASURE
More informationPermeability. Darcy's Law
Permeability Permeability is a property of the porous medium that measures the capacity and ability of the formation to transmit fluids. The rock permeability, k, is a very important rock property because
More informationThermcraft Tube Furnace General Use
Page 1 of 9 Thermcraft Tube Furnace General Use The Thermcraft furnace is a general use, 3 zone tube furnace capable of reaching temperatures of 1200 C. The large 6 diameter quartz tube is capable of holding
More informationNoth American 7347 High Pressure Gas Regulators
Combustion Noth American 7347 High Pressure Gas s Bulletin 7347 7347 s reduce high gas supply pressures to practical use levels. Since capacities will vary with the pressure drop across the regulator (see
More informationMEDAP-MOBILE GAS SUPPLY
MEDAP-MOBILE GAS SUPPLY This document is intended to provide information to an international audience outside of the US. MOBILE GAS SUPPLY 3 MOBILE GAS SUPPLY FOR TRANSPORTATION AND WARD PRESSURE REGULATORS,
More informationLevel switches Series LD Vibrating fork level switch for liquids and solids
Level switches Series LD Vibrating fork level switch for liquids and solids Robust and compact Suitable for liquids (model LD61) and solids (model LD60) No moving parts, low maintenance Corrosion resistant
More information