Michael H. Baumann, MD, FCCP; Praful B. Patel, MD; Chris W. Roney, MD; and Marcy F. Petrini, PhD, FCCP

Size: px
Start display at page:

Download "Michael H. Baumann, MD, FCCP; Praful B. Patel, MD; Chris W. Roney, MD; and Marcy F. Petrini, PhD, FCCP"

Transcription

1 Comparison of Function of Commercially Available Pleural Drainage Units and Catheters* Michael H. Baumann, MD, FCCP; Praful B. Patel, MD; Chris W. Roney, MD; and Marcy F. Petrini, PhD, FCCP Purpose: Flow rates and pressures generated by commercially available pleural drainage units (PDUs) and flow rates through available pleural drainage catheters (PDCs) are not known. This information may be important clinically depending on the volume of air leak associated with a bronchopleural fistula. Design: Eight PDUs were assessed for flow rates at various suction levels and for the percent accuracy of suction pressures generated at various settings. Eleven commonly used PDCs were assessed for flow rates at various suction control levels. All devices were donated by their manufacturer. Flow rates and pressures were measured by a RT 200 Calibration Analyzer (Timeter Instrument Corporation; St. Louis, MO) at body temperature, ambient pressure, saturated with water vapor. Five devices of each type were tested. Analysis of variance was performed with p < 0.05 being significant. Results: Multiple significant differences between PDUs were noted at a pressure of 20 cm H 2 O. The Argyle Sentinel Seal (Sherwood Medical; Tillamore, Ireland) had significantly lower flow rates (mean SD, L/min) compared with all other models. The Argyle Aqua-Seal (Sherwood Medical) had the highest PDU flow rate of devices tested ( L/min). The accuracy of PDUs at manufacturer-suggested settings varied from a mean percentage error of 0.0 to 15.5% from expected pressures; significant differences were noted in accuracy among multiple interdevice pressure comparisons. Similarly, multiple significant flow rate differences between PDCs were noted at 20 cm H 2 O. Lowest flow rates were noted with thoracentesis catheters (used as PDCs) containing side ports. Arrow drainage catheters (14F, pigtail and straight) [Arrow International; Reading, PA] both had significantly greater flow rates (both, L/min), compared with the 14F ( ) and 16F ( ) Cook devices (Cook; Bloomington, IN). Conclusions: These differences in flow rates for PDUs and PDCs may be clinically important, particularly in patients with large pneumothorax-related air leaks. Observed differences in PDU-generated pressures are likely not clinically important. (CHEST 2003; 123: ) Key words: bronchopleural fistula; catheter; chest tube; flow; pneumothorax; pleural drainage unit Abbreviations: ACCP American College of Chest Physicians; ANOVA analysis of variance; BTPS body temperature, ambient pressure, saturated with water vapor; PDC pleural drainage catheter; PDU pleural drainage unit After introduction of the three-bottle, water-seal system for pleural cavity drainage, many commercially available adaptations of this system have *From the Division of Pulmonary and Critical Care Medicine, University of Mississippi Medical Center, Jackson, MS. Presented in part at the American Thoracic Society Annual International Meeting, Toronto, ON, Canada (May 5 10, 2000), and American College of Chest Physicians Annual Meeting, San Francisco, CA (October 21 26, 2000). Manuscript received July 22, 2002; revision accepted December 3, Reproduction of this article is prohibited without written permission from the American College of Chest Physicians ( permissions@chestnet.org). Correspondence to: Michael H. Baumann, MD, FCCP, Professor of Medicine, Division of Pulmonary and Critical Care Medicine, University of Mississippi Medical Center, 2500 North State St, Jackson, MS ; mbaumann@medicine. umsmed.edu become available. These commercial pleural drainage units (PDUs) appear to dominate the marketplace to the extent that the original three-bottle system is a rarity in practice today. Portability, ease of use, and disposability are significant advantages of available PDUs; however, the question remains, how reliable are these devices? The only available assessments of commercial PDUs were published in the 1980s, 1,2 noting flow rates for these devices. The four assessed devices are no longer available. No assessment of PDU-generated pressure accuracy has been published. Similarly, no manufacturer-independent analysis of the myriad of commercially available pleural drainage catheters (PDCs) has been published Clinical Investigations

2 Given the recent American College of Chest Physicians (ACCP) recommendations for use of smallbore catheters in certain settings of spontaneous pneumothorax, 3 and their use in traumatic and iatrogenic pneumothoraces, knowledge of the potential air flow rate limitations of these small-bore catheters is key, especially in the setting of a bronchopleural fistula. 4 Limitation in the flow rates accommodated by the PDU or the PDC can lead to the development of a tension pneumothorax. PDU-delivered pressures may also affect device efficiency; hence, knowledge of the flow rate limitations of commercially available PDUs and PDCs and of the accuracy of PDUdelivered pressure would be valuable to the daily management of pneumothoraces. PDU Assessment Materials and Methods Eight PDU models were supplied free of charge by three leading manufacturers (Table 1). These models appear to be the most widely used units in the United States per information supplied by the respective manufacturers. All units are designed to represent the three chambers under water seal concept: a collection chamber, a water-seal chamber, and a suction control chamber. Suction control is achieved either by a mechanical device without need for water (termed dry suction control) or by using a water column. Of the eight units we tested, four units used dry suction control and four units used a water column. RT 200 Calibration Analyzer The RT 200 Calibration Analyzer (Timeter Instrument Corporation; St. Louis, MO) was used for measuring PDU flow rates and pressures. The RT 200 Calibration Analyzer (Fig 1) provides a digital readout of the measurements. A pneumotachometer measures gas flows with a resolution of 0.1 L/min in a range of 5 to 180 L/min with an accuracy of 1% or better of the readout. Pressures from 0 to 250 cm H 2 O are measured with a resolution of 0.1 cm H 2 O with an accuracy of 0.5%, according to the operation/service manual of the manufacturer. Our unit was last calibrated for accuracy by Onsite Calibration Service (Louisville, KY) 1 week prior to our data collection on the PDU. Calibration is suggested by the manufacturer to occur once yearly. This measurement device may be set to make all measurements at body temperature, ambient pressure, saturated with water vapor (BTPS). Measurement Methodology Flow Measurement: A bench set-up was created as in Figure 1. The PDUs were prepared following manufacturer instructions. The suction port of the PDU was connected to wall suction through a regulator. The patient connection tubing of the PDUs was connected to the flow outlet of RT-200 Calibration Analyzer. Extra connection tubing was required for this, and was minimized in length (measured at 15 cm) with a diameter (approximately 2.5 cm) significantly larger than that of the connection tubing supplied with the PDU to avoid additional resistance to flow. The flow inlet of RT-200 Calibration Analyzer was kept open to room air to simulate a continuous large bronchopleural fistula. Each PDU was tested for flow rates at suction control settings of 10 cm H 2 O, 20 cm H 2 O, and 40 cm H 2 O (if available), in that order. Five PDUs were tested for each given model. The PDU models were selected in random order. All five units were tested before proceeding to another model. After initiation of the suction, time was allowed to achieve a steady-state level (the digital readout on the RT-200 Calibrator Analyzer remained constant 0.01) prior to recording the data. All data were recorded at BTPS. Pressure Measurement: The PDUs were set up as above, except the patient connection tubing of each PDU was connected to the pressure measurement port on the RT-200 Calibrator Analyzer. Each PDU was tested for negative pressure generated at suction control settings as for flow measurements. After initiation of the suction, time was allowed to achieve a steadystate level (the digital readout on the RT-200 Calibrator Analyzer remained constant 0.01) prior to recording the data. All data were recorded at BTPS. The accuracy of the device (mean percentage of error) was used for all statistical comparisons of pressures generated by the PDU tested. Accuracy is the standard Table 1 PDU Models, Manufacturer, and Flow Rates* Suction Level, cm H 2 O PDUs Manufacturer/Model Atrium Medical Corporation (Hudson, NH) Atrium Ocean no 40 setting Atrium Atrium 3612 (pediatric device) Sherwood Medical (Tullamore, Ireland) Argyle Aqua-Seal no 40 setting Argyle Thora-Seal III no 40 setting Argyle Sentinel Seal no 40 setting Deknatel (Fall River, MA) Pleur-evac A-6000 (Code #A-6000) Pleur-evac SAHARA (Code #S11000FS) *Data are presented as L/min (mean SD). Water device: pressure regulation by water column. Dry device: pressure regulation not by water column. CHEST / 123 / 6/ JUNE,

3 Figure 1. The set-up to measure PDU flow. Shown (left to right), sequentially connected, are the RT-200 Calibration Analyzer, a PDU, and a wall suction regulator. biomedical instrumentation expression of how well the function of an instrument approaches the true or reference value, 5 in this case the expected delivered negative pressure. A final test of accuracy of pressure was performed after the above measurements were completed. In all other prior pressure accuracy tests outlined above, the directions of the manufacturer regarding the titration of external suction level (wall suction through a regulator) were closely followed. Many wall suction regulators offer the option of full vacuum. This essentially allows the full negative pressure of the suction source to be applied to the PDU. This appears in our experience to be a common practice utilized when setting up a PDU. To test the accuracy of the PDU in down-regulating the full force of the suction source to the pressure set on the PDU, each PDU was retested after all the measurements outlined above were obtained. For each of the pressure settings a PDU was capable of ( 10 cm H 2 O, 20 cm H 2 O, and 40 cm H 2 O, as available), the PDU was retested with the wall suction regulator set to full vacuum and the derived PDU pressure measured as outlined above. Statistical Analysis For both flow and pressure measurements, two-way analysis of variance (ANOVA) with repeated measures was used. Comparison of data at 40 cm H 2 O was made separately from data at 10 cm H 2 O and 20 cm H 2 O, because only four PDUs offered a 40 cm H 2 O suction control level. Post hoc multiple comparison analysis was performed, if appropriate, using a Tukey test. A probability value 0.05 was considered significant. Comparisons were done between all models at 10 cm H 2 O and 20 cm H 2 O, as not all PDUs offered a setting of 40 cm H 2 O. Pressure accuracy was defined as mean percentage error (deviation from the set value SD percentage). PDC Assessment Eleven different catheter types were obtained from three major manufacturers (Arrow International [Reading, PA]; Cook [Bloomington, IN]; and Argyle [Sherwood Medical]) for testing (Table 2). The catheters are listed in the order of ascending expected flow rates based on the bore and length of the catheter as described by the Fanning equation. Briefly, the larger the catheter bore and shorter its length the greater the expected flow accommodated (a complete explanation of the Fanning equation is provided in the Discussion ). Each company donated the needed test samples from everyday stock supplies. Five catheters of each type were tested. Flow Measurement: A bench set-up was created as in Figure 2. As with the PDUs, flow measurements were obtained using the RT 200 Calibration Analyzer. Measurements were made four months after the last calibration. The Pleur-evac A-6000 (Deknatel) PDU with 60 mm Hg of applied wall suction was used for testing of all catheter sets. The Pleur-evac A-6000 was chosen given its combination of suction accuracy and high flow rates, particularly at 20 cm H 2 O, accommodated in a dry unit. A water column unit was not chosen given the potential of variability in the water column height during testing of each PDC group. The testing apparatus consisted of a large carboy connected to the RT 200 Calibration Analyzer, which was then connected in serial to the PDU through which suction was applied. All connections were with rigid tubing with elastic foam sealant tape at all joints to prevent air leaks. The catheters themselves were placed in adjustable adapters supplied by Cook and sealed with silicon caulk. The catheter/adapter assembly was inserted into a large rubber cork, and the cork was inserted into the carboy for testing purposes. The RT 200 Calibration Analyzer was set at BTPS and zeroed prior to testing of each catheter. Three pressure settings ( 10 cm H 2 O, 20 cm H 2 O, and 40 cm H 2 O) were used during testing, and each catheter was occluded prior to each measurement while under suction to ensure the absence of leaks in the apparatus (0 L/min of flow). The catheters in a specific set were all tested at the same negative pressure to decrease the chances for variability. Negative pressures were increased sequentially, and testing of all catheters in each set was completed Clinical Investigations

4 Table 2 PDC Description and Flow Rates* Suction Level, cm H 2 O Model Manufacturer Size Description AK Arrow 8.0F by 16 cm Pneumothorax kit AK Arrow 8.0F by 12 cm Pleura-seal thoracentesis kit (three-way stopcock and selfsealing valve) Argyle Safety Argyle 8.0F by 10 cm Safety pneumothorax system Pneumothorax Argyle Safety Thoracentesis Argyle 8.0F by 8.5 cm Turkel safety thoracentesis system (catheter with sideport/three-way stopcock and self-sealing valve) C-TPTS FSNS Cook 8.5F by 7.5 cm Emergency pneumothorax set C-TPT-100 Cook 9.0F by 29 cm Pneumothorax set C-UPTP-1400-WAYNE Cook 14.0F by 29 cm Pneumothorax set AK Arrow 14.0F by 23 cm Percutaneous cavity drainage catheterization kit (curved catheter with sidearm port) AK Arrow 14.0F by 23 cm Percutaneous cavity drainage catheterization kit (straight catheter with sidearm port) C-TQTS-1600 Cook 16.0F by 41 cm Quick chest tube set C-TQTS-2400 Cook 24.0F by 41 cm Quick chest tube set *Data are presented as L/min (mean SD). Arranged in predicted increasing flow rates as would be predicted by catheter bore and length, those with the smallest bore size and, less importantly, longer length would be expected to have the lower flow rates. After all of the sample sets had been tested, the mean air flow for each product was calculated and used for comparison. Statistical Analysis Two-way ANOVA with repeated measures was used, with models of catheters as the random factor, and the three suction pressures as fixed factors. Post hoc multiple comparison analysis was performed using the Tukey test when differences were found by ANOVA. A probability value 0.05 was considered significant. PDU Assessment Results The PDU mean flow rates achieved at various water pressure suction levels appear in Table 1. Multiple statistically significant differences and areas of similarities are noted. There is nearly a fourfold variation from the PDU with highest flow (Argyle Aqua-Seal) and the PDU with the lowest flow (Argyle Sentinel Seal). The Atrium Ocean 2002, Argyle Thora-Seal III, and Argyle Aqua-Seal provide the highest flow rates, and the units are statistically similar at 10 cm H 2 O and 20 cm H 2 O of suction. The Argyle Sentinel Seal provides the lowest flow rates and was statistically different from all other PDUs at both 10 cm H 2 O and 20 cm H 2 O of pressure. The Pleur-evac A-6000, Pleur-evac SAHARA, Atrium 3600, and Atrium 3612 provide intermediate flow rates. Statistical assessment of those PDUs (n 4) producing 40 cm H 2 O of pressure demonstrates that each model is significantly different from each other (Table 1). The accuracy (mean percent error) of the PDUs tested using the directions of the manufacturers regarding external suction titration appears in Table 3. Multiple interdevice differences in accuracy are significantly different but likely not of clinical significance. Focusing on the PDU 20 cm H 2 O setting frequently incorporated in practice, three major findings are noted when following manufacturer instructions. The Atrium 3612 was the least accurate PDU, demonstrating a mean error of 15.5%, with this accuracy being different from all other PDU models tested at 20 cm H 2 O. The Sentinel Seal was the most accurate PDU, with a mean error of 0.0%, notably with manufacturer instructions necessitating wall suction being set at full vacuum. Accuracy of the other units was variable, but all devices had a mean error of 10%. The errors of all of the PDUs were in the positive direction (delivering less negative pressure than the PDU test value) except for the Thora-Seal III, which demonstrated a mean error of 0.8% in the direction of increased negative pressure from the test (PDU set) value. Regarding the question of the effects of fullvacuum vs manufacturer-directed external suction specifications at the PDU setting of 20 cm H 2 O CHEST / 123 / 6/ JUNE,

5 Figure 2. The set-up to measure PDC flow. Shown (left to right), sequentially connected, are the PDU (Pleur-evac A-6000), the RT-200 Calibration Analyzer, and a carboy with a PDC seated in its top opening. (within the same PDU model), several findings are noted (no data table provided). Most notable, full vacuum applied to any of these PDUs does not appear to create a clinically detrimental increase in negative pressure delivered to the patient. The Pleur-evac A-6000, Atrium Ocean 2002, and the Aqua-Seal produced similar accuracy (comparing same model PDUs to same model PDUs at the two different external suction settings). All other PDU models demonstrated statistically significant pressure accuracy error differences (but not likely clinically significant) between the two external suction settings (manufacturer-suggested wall suction setting PDUs Table 3 Accuracy of Pressures* Suction Level, cm H 2 O Atrium Ocean No 40 setting Atrium Atrium Aqua-Seal NT No 40 setting Thora-Seal III No 40 setting Sentinel Seal No 40 setting Pleur-evac A Pleur-evac SAHARA *Data are presented as mean percent error SD. Accuracy not tested (NT) at 10 cm H 2 O due to technical issues. vs full wall suction) in the negative direction (delivering more negative pressure than the PDU test value), with the Thora-Seal III being least accurate and producing a mean error of 27.0%. The next closest mean error is 8.7% by the Pleur-evac SAHARA. The Sentinel Seal was tested at all PDUset negative suction pressures utilizing external suction source at full vacuum, as necessitated by manufacturer directions, and delivered very accurate pressures (see above). Of the PDUs providing 40 cm H 2 O settings (Table 1), the Pleur-evac A-6000 and Pleur-evac SAHARA produced similar accuracy (comparing same-model PDU to samemodel PDU at the two different external suction settings). The Atrium 3600 and 3612 demonstrated significant pressure accuracy error differences between the two external suction settings in the negative direction (delivering more negative pressure than the PDU test value), with mean accuracy errors of 2.3% and 2.9%, respectively. PDC Assessment Flow rates achieved at various water pressure suction levels for the PDCs tested appear in Table 2. These data may be summarized in five major points: (1) the highest flow rate at all negative pressures is seen with the Cook 24F chest tube; (2) at all pressures, the 14F Arrow drainage catheters had 1882 Clinical Investigations

6 statistically greater flow rates than both the 16F and 14F Cook catheters (approximately 2 L/min and 4 L/min greater, respectively, at 20 cm H 2 O); (3) the majority of catheters (6 of 11 catheters) at all suction levels tested ( 10 cm H 2 O, 20 cm H 2 O, and 40 cm H 2 O) have flow rates 16 L/min; (4) at 20 cm H 2 O (an often-used level of suction pressure in clinical practice), only three catheters (Arrow 14F AK-01600, Arrow 14F AK-01601, and Cook 24F C-TQTS-2400) had flow rates 16 L/min; and (5) the lowest mean flow rates are seen in the two catheters with three-way stopcocks (Argyle Safety Thoracentesis and Arrow AK Pleural-seal Thoracentesis kit). Discussion Chest tube placement and subsequent connection to a PDU may be appropriate for many pneumothoraces. The settings appropriate for chest tube placement, the size of tube selected, including smallerbore tubes, and subsequent utilization of a PDU are well outlined for spontaneous pneumothoraces in the recent ACCP spontaneous pneumothorax consensus statement. 3 Consensus is less defined regarding chest tube utilization, tube size selection, and the role of PDUs in nonspontaneous pneumothoraces, but some evidence provides direction. 6 Key to the selection of a PDC and a PDU in the setting of both spontaneous and nonspontaneous pneumothoraces is the maximal air flow (the volume of air per unit of time) each device accommodates particularly if an air leak (bronchopleural fistula) persists. Gases leaked by a bronchopleural fistula are moist with turbulent flow characteristics. The Fanning equation determines the flow of moist gas with turbulent flow characteristics through a PDC (v 2 r 5 P/fl; v flow, r radius, l length, P pressure, f friction factor) Obviously, the critical factor in chest tube selection is the internal diameter (bore) of the tube and, less so, the tube length. Air flow may be compromised by the presence of viscous fluids such as blood. When considering air drainage alone, tube selection must account not only for removal of existing pleural air, but also the ongoing removal of additional air produced (bronchopleural fistula) and the rate of that air production. Patients with bronchopleural fistulae in the setting of chest trauma, thoracotomy, and ARDS may have air leaks ranging from 1 L/min up to 16 L/min. 2,8,11 Conceivably, patients with spontaneous pneumothoraces, especially those requiring mechanical ventilation, may also have large air leaks. Practical experience indicates that 16 L/min is not a fixed upper limit for air leak volume, with larger leaks possibly occurring in any patient receiving mechanical ventilation with a persistent bronchopleural fistula. The majority of the minute ventilation in an intubated patient receiving mechanical ventilation can exit a large bronchopleural fistula. This may be the case in a postpneumonectomy stump dehiscence. Appropriate PDC size selection is therefore key to preventing tension pneumothorax development. Because a PDU is often connected in series with a PDC, appropriate PDU selection is also key. However, when considering flow rates accommodated, the PDC is the most likely rate-limiting device when compared to the PDU (Tables 1, 2). No formula analogous to the Fanning equation exists to determine flow rates for commercially available PDUs. This is likely due to their relatively complex structure based on the three-bottle (compartment) system. This three-bottle system is commercially condensed into a convenient and easily mobile single module of variable design incorporating the three compartments. The three compartments, sequentially, include the following: (1) the collection bottle to trap liquid material and other debris from the patient s pleural space while allowing pleural air to pass through the next two compartments, (2) the water seal bottle to prevent air flow back to the patient s pleural space and allow detection of an air leak (bronchopleural fistula), and (3) the manometer bottle to regulate the amount of negative pressure transmitted back to the patient from the wall suction device (or equivalent suction source). The manometer bottle has an input and output tube and a central vent tube that may be raised or lowered. Adjusting the depth of the central vent tube in the chamber water determines the negative pressure transmitted from the wall suction back to the patient s pleural space. 9,12 The water chamber device has been replaced in some commercial PDUs with a spring-loaded or similar device to regulate the negative pressure transmitted to the patient. Given the complexity of PDUs and their design variability, it is understandable there is no formula to calculate flow. Meantime, using the Fanning equation to mathematically derive flow rates for a PDC is onerous and impractical. This problem is compounded by the absence of flow rate information in the packaging with either PDUs or PDCs. This absence of information leaves the clinician unable to compare flow characteristics of these devices and objectively choose the optimal device fitting the clinical situation. The absence of packaging information regarding the accuracy of pressures generated by PDUs further complicates the picture. These factors prompted the current study. CHEST / 123 / 6/ JUNE,

7 The multiple flow rate differences among the tested PDCs highlight the need for clinicians to pick the appropriate catheter for specific patient circumstances. This is particularly important in circumstances where high flow rates (persistent large bronchopleural fistulae) may be encountered. The popularity of smaller-bore PDCs, despite some offering potentially dangerous low flow rates, potentially worsens the situation. Contributing to this popularity is the intuitive notion that smaller PDCs, particularly those incorporating the Seldinger technique (guidewire placement), may be less painful to place than the traditional larger-bore tubes often placed by blunt dissection; however, a limited assessment of this issue indicates that smaller-bore catheters placed for simple aspiration of a pneumothorax appear less painful compared to traditional chest tube placement based on their average daily pain score, not from pain during tube placement. 13 The clinician should not hesitate to choose a traditional large-bore chest tube if the situation warrants; however, knowledge of the flow rates of the PDCs tested in this study may allow clinicians to safely select a smaller-bore chest tube. The maximal flow rate accommodated of the PDCs tested at 20 cm H 2 O, as expected, was from the largest-bore PDC tested, the 24F Cook C- TQTS-2400 (Table 2). This catheter had the highest flows at all pressures tested. Catheters 16F may be considered small bore and deliver significantly less flow than the only larger-bore PDC tested, the Cook C-TQTS Various clinical conditions outlined above may have bronchopleural fistula flow rates of up to 16 L/min. 2,8,11 The clinician, however, must be vigilant for air leaks 16 L/min, as noted earlier. The Arrow AK and AK are the only two 16F PDCs tested able to accommodate the arbitrary value of 16 L/min of flow at 20 cm H 2 O. Patients with spontaneous pneumothoraces who are not receiving mechanical ventilation are likely not at high risk of a large bronchopleural fistula and may be appropriate for a 16F PDC. The ACCP spontaneous pneumothorax guidelines suggest initial placement of a chest tube and hospital admission as preferred management of an unstable patient with a large ( 3 cm lung collapse) primary spontaneous pneumothorax, or any secondary spontaneous pneumothorax patient with a large pneumothorax or with clinical instability. Specifically, patients with a primary spontaneous pneumothorax (unlikely at risk for a large air leak) suitable for chest tube placement should have their lung re-expanded using a smallbore catheter ( 14F) or placement of a 16F to 22F chest tube. 3 When to place a chest tube and what size to select is less well defined for nonspontaneous pneumothoraces. A recent review suggests, based on available evidence, placement of larger-bore PDCs (28F to 36F) for noniatrogenic traumatic pneumothoraces, particularly in the patient receiving mechanical ventilation, to drain both an ongoing air leak and potential accompanying hemothorax. 6 Iatrogenic traumatic pneumothoraces may be treated with smaller-bore PDCs as long as patients are not receiving mechanical ventilation. 6 Perhaps in both spontaneous and nonspontaneous pneumothoraces the selection of a PDC with flow rate reserve would be wise to provide a safety net in case of unanticipated increased air flow rates. The premise that larger-bore PDCs should provide larger flow rates breaks down when assessing the Cook 14F (C-UPTP-1400-WAYNE) and 16F (C-TQTS-1600) catheters compared with the Arrow 14F catheters (AK and AK-01601) [Table 2]. Based on the Fanning equation, the clinically significant, lower flow rates of the Cook 14F and 16F catheters at 20 cm H 2 O (approximately 4 L/min and 2 L/min less, respectively) may be accounted for by their greater length (29 cm and 41 cm, respectively) compared to the Arrow 14F catheters (both 23 cm). Conceivably, these differences may also be accounted for by differences in the catheter bore size not reflected by the sizes reported by the manufacturers. Lastly, different PDC construction materials may be more or less collapsible under negative pressure. Ease of collapse would be particularly important in the patient where tissues surrounding the PDC could also drive collapse. PDC collapsibility and its affect on air flow were not assessed. Regardless, differences in air flow rates may not be recognized by a clinician selecting a catheter simply based on bore size. Lastly, using PDC equipment designed as thoracentesis equipment for pneumothorax management may not be a wise choice. Such issues are becoming more prevalent in a cost-containment environment when attempting to purchase one device to meet multiple clinical needs. Both the Argyle and Arrow thoracentesis PDCs have clinically significant less air flow rates compared to their comparable pneumothorax PDCs despite the same bore (each 8F) and the lowest flow rates of all PDCs tested at all tested pressure settings (Table 2). The Argyle thoracentesis PDC accommodates less than half the flow of the comparable Argyle pneumothorax PDC, and the Arrow thoracentesis PDC handles approximately 1.5 L/min less flow than the comparable Arrow pneumothorax PDC (all tests at 20 cm H 2 O). Interestingly, the lengths of both the Argyle and Arrow thoracentesis PDCs are shorter than their 1884 Clinical Investigations

8 comparable pneumothorax catheter of the same internal bore (Table 2); shorter length should enhance not reduce flow rates. The lower flow rates of the thoracentesis catheters may be accounted for by the equipment attached proximally to the thoracentesis catheters including a three-way stopcock and a self-sealing valve. Once a chest tube is placed, whether small or large bore, the tube may then be attached to a PDU. Similar to the selection of a PDC, the appropriate use of a PDU is well delineated in the ACCP spontaneous pneumothorax guidelines and less well defined for traumatic and iatrogenic pneumothoraces. Regardless, if a PDU is connected to a PDC, the same flow considerations exist. Appropriate size selection of a PDC to accommodate bronchopleural fistula air flow may be thwarted by selecting a PDU unable to handle the air flow with subsequent development of a tension pneumothorax. The mean flow rates handled by commercial PDU at 20 cm H 2 O vary considerably, from 10.8 to 42.1 L/min (Table 1). The Sentinel Seal PDU has the lowest flow rate at 10.8 L/min and is lower than any other PDU tested. This average flow is substantially less than may be encountered in various clinical situations, outlined above, and could lead to the development of a tension pneumothorax. Several PDUs deliver 16 L/min (discussed above) at 10 cm H 2 O (Table 1), but all PDUs except the Sentinel Seal deliver 16 L/min at 20 cm H 2 O. The accuracy of the pressures delivered by the PDU varies considerably (Table 3) with multiple interdevice differences; however, the clinical significance of these differences is questionable. For example, the Atrium 3612 was the least accurate PDU, with a mean error of 15.5% at 20 cm H 2 O. This means that the device actually delivers on average 16.9 cm H 2 O( 20 cm H 2 O less 20 cm H 2 O multiplied by 15.5%). This decrease in delivered negative pressure is likely of little clinical significance given the mean flow rate delivered by the device (20.3 L/min, Table 1) at the set PDU pressure of 20 cm H 2 O. Perhaps of more concern would be if the PDU delivered substantially greater negative pressure than indicated. Such elevated pressures theoretically could lead to damage to the underlying tissues including the lung and pericardium. The only device to deliver a negative pressure greater than the indicated PDU set pressure of 20 cm H 2 O (Table 3) is the Thora-Seal III, with a mean increase of only 0.8% (a mean increase in negative pressure of 0.16 cm H 2 O). Accuracy of the other units at the set pressure of 20 cm H 2 O was variable, with a mean error of 10% (other than the Atrium 3612, see above) in the direction of delivering less negative pressure than the PDU test values. Therefore, on average, the delivered negative pressure is reduced by 2cmH 2 O of pressure at the set value of 20 cm H 2 O (a delivered pressure 18 cm H 2 O). These pressure inaccuracies are likely of little clinical significance. Most concern for inappropriate negative pressure delivered to the pleural space might be expected when full-vacuum vs manufacturer-directed external suction specifications are used. Full vacuum does not appear to substantially adversely affect the delivered pressures when PDUs were tested at 20 cm H 2 O settings (no data table). The Pleur-evac A-6000, Atrium Ocean 2002, and the Aqua-Seal produced similar accuracy (comparing same-model PDU to same-model PDU at the two different external suction settings). The Sentinel Seal was tested at all set negative suction levels at full vacuum, based on manufacturer directions, and delivered very accurate pressures. All other PDU models demonstrated significant pressure accuracy error differences between the two external suction settings (manufacturersuggested wall suction setting vs full wall suction) in the negative direction (delivering more negative pressure than the PDU test value). The least accurate PDU at full wall suction was the Thora-Seal III, with a mean error of 27%. This translates to a mean delivered negative water pressure of 25.4 cm H 2 O. Similarly, the most inaccurate PDU with a 40 cm H 2 O setting when tested at full vacuum, the Atrium 3612, produced a mean error of 2.9% in the negative direction (a mean delivered negative pressure of 41.2 cm H 2 O). Such minimal increases in negative pressure are likely of little clinical consequence. No definitive comparison in the accuracy (Table 3) of the water column PDU vs dry PDU systems can be made (Table 1). Absolute mean accuracy (positive or negative error) of the water column units when tested at a set pressure of 20 cm H 2 O varied from a mean error of 0.0% (Sentinel Seal) to 4.6% (Atrium Ocean 2002). The dry unit absolute mean accuracy when tested at 20 cm H 2 O varied from 4.2% (Pleur-evac A-6000) to 15.5% error (Atrium 3612). Several problems exist with this study of PDUs and PDCs. First, although major US manufacturers were contacted, not all commercially available devices were tested. For example, the TRU-CLOSE Thoracic Vent Procedure Tray (Davis and Geck; Wayne, NJ), a PDC, was not tested due to the inability to adapt the device to our testing system. Larger-bore chest tubes, except the Cook C-TQTS- 2400, were not tested given that clinically significant potential flow restrictions were found to exist primarily with small-bore catheters. Products not available in the United States were not tested. Additionwww.chestjournal.org CHEST / 123 / 6/ JUNE,

9 ally, as product lines evolve, the devices reported in this study may no longer be available or be modified and carry the same or different name. Accordingly, the same name may not equate to the same flow and/or accuracy characteristics. This requires the clinician to always query the maximal flow rates that a PDU or PDC delivers and the accuracy of a PDU when incorporating new products in their practice. Lastly, our sham set-up was designed to assess optimal flows from each PDU and PDC, and the PDU accuracy of delivered pressures. This design, however, may not reflect clinical conditions, particularly for a PDC, wherein the pleural space and its non-air contents may affect the flow a particular device may be capable of handling. The presence of the lung or pleural debris such as blood may significantly decrease the air flows reported from our test design. In conclusion, multiple different flow rates are noted among different PDUs at the same negative water pressure settings. Although significant accuracy differences in delivered pressures of the PDU tested are seen, these may not be of clinical consequence. PDUs, in general, significantly protect the patient (pleural space) in the event of inadvertent use of full wall suction. Similarly, multiple significant differences in PDC flow rates exist, including between catheters of the same bore size. The significant differences in accommodated air flows may be particularly important in patients with large air leaks. Incorporation of such knowledge may preclude the development of a tension pneumothorax. References 1 Capps JS, Tyler ML, Rusch VW, et al. Potential of chest drainage units to evacuate broncho-pleural air leaks. Chest 1985; 88:57S 2 Rusch VW, Capps JS, Tyler ML, et al. The performance of four pleural drainage systems in an animal model of bronchopleural fistula. Chest 1988; 93: Baumann MH, Strange C, Heffner JE, et al. Management of spontaneous pneumothorax: an American College of Chest Physicians Delphi consensus statement. Chest 2001; 119: Baumann MH, Sahn SA. Medical management and therapy of bronchopleural fistulas in the mechanically ventilated patient. Chest 1990; 97: Latman NS, Lanier R. Expressions of accuracy in the evaluation of biomedical instrumentation. Biomed Instrum Technol 1998; 32: Baumann MH. Non-spontaneous pneumothorax. In: Light RW, Lee YCG, eds. Pleural disease: an international textbook. London, UK: Arnold Publishers, 2003 (in press) 7 Baumann MH, Strange C. Treatment of spontaneous pneumothorax: a more aggressive approach? Chest 1997; 112: Batchelder TL, Morris KA. Critical factors in determining adequate pleural drainage in both the operated and nonoperated chest. Am Surg 1962; 28: Miller KS, Sahn SA. Chest tubes: indications, technique, management and complications. Chest 1987; 91: Swenson EW, Birath G, Ahbeck A. Resistance to air flow in bronchospirometric catheters. J Thorac Surg 1957; 33: Powner DJ, Cline D, Rodman GH. Effect of chest-tube suction on gas flow through a bronchopleural fistula. Crit Care Med 1985; 13: Baumann MH. Chest tubes. In: Bouros D, ed. Pleural disease. New York, NY: Marcel Dekker, 2003 (in press) 13 Harvey J, Prescott RJ. Simple aspiration versus intercostal tube drainage for spontaneous pneumothorax in patients with normal lungs. BMJ 1994; 309: Clinical Investigations

Key words: biomedical engineering; flowmeters; laboratory; pleural drainage system; pneumothorax

Key words: biomedical engineering; flowmeters; laboratory; pleural drainage system; pneumothorax A Hydrodynamic Study of Pleural Drainage Systems* Some Practical Consequences Gerardo Manzanet, MD, PhD; Antonio Vela, MSc, PhD; Ricardo Corell, MD, PhD; Ramón Morón, MD; Rogelio Calderón, MD, PhD; and

More information

Prof AH Basson, Pr Eng March Departement Meganiese en Megatroniese Ingenieurswese. Department of Mechanical and Mechatronic Engineering

Prof AH Basson, Pr Eng March Departement Meganiese en Megatroniese Ingenieurswese. Department of Mechanical and Mechatronic Engineering An evaluation of the effectiveness of the Sinapi Chest Drain's flutter valve in comparison to a water seal as found in an UWD to resolve pneumo- or hemothorax Prof AH Basson, Pr Eng March 2010 Departement

More information

Chest Drains. All Covenant Health Intermediate Care Nursery staff. Needle Aspiration

Chest Drains. All Covenant Health Intermediate Care Nursery staff. Needle Aspiration Approved by: Chest Drains Gail Cameron Director, Maternal, Neonatal & Child Health Programs Neonatal Nursery Policy & Procedures Manual : April 2013 Next Review April 2016 Dr. Ensenat Medical Director,

More information

CORESTA RECOMMENDED METHOD N 6

CORESTA RECOMMENDED METHOD N 6 CORESTA RECOMMENDED METHOD N 6 DETERMINATION OF VENTILATION DEFINITIONS AND MEASUREMENT PRINCIPLES (2015 Revision September 2016) 1. SCOPE This CORESTA Recommended Method specifies a method for the determination

More information

Characterizers for control loops

Characterizers for control loops Characterizers for control loops By: F. G. Shinskey (May 1999) Introduction Commercial controllers such as the PID series (proportional, integral, derivative, and their combinations) are linear devices

More information

Guidelines for Care: Management of Patients with a Chest Drainage Unit

Guidelines for Care: Management of Patients with a Chest Drainage Unit GMHA NURSING SERVICES DEPARTMENT Guidelines for Care: Management of Patients with a Chest Drainage Unit Roseann Apuron, RNC OB, Staff Nurse Training Officer OBJECTIVES: 1. Identify the indications, contraindications,

More information

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Cover Page for Lab Report Group Portion. Head Losses in Pipes Cover Page for Lab Report Group Portion Head Losses in Pipes Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 February 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section

More information

Cover Page for Lab Report Group Portion. Pump Performance

Cover Page for Lab Report Group Portion. Pump Performance Cover Page for Lab Report Group Portion Pump Performance Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 March 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section number:

More information

Chest Drainage Systems for Cardiac & Thoracic Surgery. Including portable suction with Digital Air Leak Monitoring

Chest Drainage Systems for Cardiac & Thoracic Surgery. Including portable suction with Digital Air Leak Monitoring Chest Drainage Systems for Cardiac & Thoracic Surgery Including portable suction with Digital Air Leak Monitoring REDAX REDAX Simple Plus Unique, post-operative chest drainage for cardio-thoracic surgery.

More information

Experiment. THE RELATIONSHIP BETWEEN VOLUME AND TEMPERATURE, i.e.,charles Law. By Dale A. Hammond, PhD, Brigham Young University Hawaii

Experiment. THE RELATIONSHIP BETWEEN VOLUME AND TEMPERATURE, i.e.,charles Law. By Dale A. Hammond, PhD, Brigham Young University Hawaii Experiment THE RELATIONSHIP BETWEEN VOLUME AND TEMPERATURE, i.e.,charles Law By Dale A. Hammond, PhD, Brigham Young University Hawaii The objectives of this experiment are to... LEARNING OBJECTIVES introduce

More information

Single- or Two-Stage Compression

Single- or Two-Stage Compression The following article was published in ASHRAE Journal, August 2008. Copyright 2008 American Society of Heating, Refrigerating and Air- Conditioning Engineers, Inc. It is presented for educational purposes

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.9 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

Esophageal Simulator Joel Schmocker ~ Leader Luke Juckett ~ Communicator Ian Linsmeier ~ BSAC Tyler Klann ~ BWIG

Esophageal Simulator Joel Schmocker ~ Leader Luke Juckett ~ Communicator Ian Linsmeier ~ BSAC Tyler Klann ~ BWIG Esophageal Simulator Joel Schmocker ~ Leader Luke Juckett ~ Communicator Ian Linsmeier ~ BSAC Tyler Klann ~ BWIG Bonnie Reinke ~ Client Stephen Gorski ~ Client John Webster ~ Advisor Table of Contents

More information

Drilling Efficiency Utilizing Coriolis Flow Technology

Drilling Efficiency Utilizing Coriolis Flow Technology Session 12: Drilling Efficiency Utilizing Coriolis Flow Technology Clement Cabanayan Emerson Process Management Abstract Continuous, accurate and reliable measurement of drilling fluid volumes and densities

More information

Operation and Maintenance of the EPV200 Portable Ventilator

Operation and Maintenance of the EPV200 Portable Ventilator Operation and Maintenance of the EPV200 Portable Ventilator 1 Applications of the EPV200 The EPV200 Portable Ventilator is a gas powered electronically controlled mechanical ventilator, designed to provide

More information

Investigation of Thermal Effects of CO 2 on Earth-Atmosphere System

Investigation of Thermal Effects of CO 2 on Earth-Atmosphere System Investigation of Thermal Effects of CO 2 on Earth-Atmosphere System Design Team Michel Beguin, Tim Bevins Dan Jakiela, Frank Kuchinski Design Advisor Prof. Yiannis Levendis Abstract This Capstone Project

More information

Sizing Pulsation Dampeners Is Critical to Effectiveness

Sizing Pulsation Dampeners Is Critical to Effectiveness Sizing Pulsation Dampeners Is Critical to Effectiveness Pressure variation is an important consideration when determining the appropriate size pulsation dampener needed for an application. by David McComb,

More information

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE TELEDYNE HASTINGS TECHNICAL PAPERS INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE Proceedings of FEDSM 98 1998 ASME Fluids Engineering

More information

product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels

product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels 12.09 product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels Introduction: This manual covers the installation and operation of Humboldt FlexPanels for Triaxial and Permeability

More information

H16 Losses in Piping Systems

H16 Losses in Piping Systems H16 Losses in Piping Systems The equipment described in this manual is manufactured and distributed by TECQUIPMENT LIMITED Suppliers of technological laboratory equipment designed for teaching. BONSALL

More information

POWER Quantifying Correction Curve Uncertainty Through Empirical Methods

POWER Quantifying Correction Curve Uncertainty Through Empirical Methods Proceedings of the ASME 2014 Power Conference POWER2014 July 28-31, 2014, Baltimore, Maryland, USA POWER2014-32187 Quantifying Correction Curve Uncertainty Through Empirical Methods ABSTRACT Christopher

More information

The Discussion of this exercise covers the following points: Pumps Basic operation of a liquid pump Types of liquid pumps The centrifugal pump.

The Discussion of this exercise covers the following points: Pumps Basic operation of a liquid pump Types of liquid pumps The centrifugal pump. Exercise 2-3 Centrifugal Pumps EXERCISE OBJECTIVE In this exercise, you will become familiar with the operation of a centrifugal pump and read its performance chart. You will also observe the effect that

More information

Practical Guide. By Steven T. Taylor, P.E., Member ASHRAE

Practical Guide. By Steven T. Taylor, P.E., Member ASHRAE ractical Guide The following article was published in ASHRAE Journal, March 2003. Copyright 2003 American Society of Heating, Refrigerating and Air- Conditioning Engineers, Inc. It is presented for educational

More information

RSPT 1060 OBJECTIVES OBJECTIVES OBJECTIVES EQUATION OF MOTION. MODULE C Applied Physics Lesson #1 - Mechanics. Ventilation vs.

RSPT 1060 OBJECTIVES OBJECTIVES OBJECTIVES EQUATION OF MOTION. MODULE C Applied Physics Lesson #1 - Mechanics. Ventilation vs. RSPT 1060 MODULE C Applied Physics Lesson #1 - Mechanics OBJECTIVES At the end of this module, the student should be able to define the terms and abbreviations used in the module. draw & explain the equation

More information

3 1 PRESSURE. This is illustrated in Fig. 3 3.

3 1 PRESSURE. This is illustrated in Fig. 3 3. P = 3 psi 66 FLUID MECHANICS 150 pounds A feet = 50 in P = 6 psi P = s W 150 lbf n = = 50 in = 3 psi A feet FIGURE 3 1 The normal stress (or pressure ) on the feet of a chubby person is much greater than

More information

Medical Instruments in the Developing World

Medical Instruments in the Developing World 2.2 Ventilators 2.2.1 Clinical Use and Principles of Operation Many patients in an intensive care and the operating room require the mechanical ventilation of their lungs. All thoracic surgery patients,

More information

Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 6th Edition by Cairo

Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 6th Edition by Cairo Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 6th Edition by Cairo Link full download: http://testbankair.com/download/test-bank-for-pilbeams-mechanicalventilation-physiological-and-clinical-applications-6th-edition-by-cairo/

More information

VIAFLO 96. Electronic 96 Channel Hand Held Pipettor

VIAFLO 96. Electronic 96 Channel Hand Held Pipettor VIAFLO 96 Electronic 96 Channel Hand Held Pipettor VIAFLO 96 Unique Multichannel microplate pipetting is a growing task because of the need to increase productivity and improve the quality of the data

More information

Suctioning: Esophageal/Gastric

Suctioning: Esophageal/Gastric Approved by: Suctioning: Esophageal/Gastric Gail Cameron Senior Director Operations, Maternal, Neonatal & Child Health Programs Dr. Paul Byrne Medical Director, Neonatology Neonatal Policy & Procedures

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.12 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient,

More information

Test Method of Trap Performance for Induced Siphonage

Test Method of Trap Performance for Induced Siphonage VII.4 Test Method of Trap Performance for Induced Siphonage K. Sakaue (1), H. Kuriyama (2), H. Iizuka (3), M. Kamata (4) (1) sakaue@ isc.meiji.ac.jp School of Science and Technology, Meiji University,

More information

Cardiac Output Simulation for Specific Makes of Monitor. Each injection yields in a time-temperature curve whose area represents the cardiac output:

Cardiac Output Simulation for Specific Makes of Monitor. Each injection yields in a time-temperature curve whose area represents the cardiac output: Cardiac Output Simulation for Specific Makes of Monitor Theory The measurement of the volume of blood pumped by the heart is a valuable diagnostic tool in the management of patients undergoing major cardiovascular

More information

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW AC 2010-2145: MEASUREMENT OF HYDROGEN IN HELIUM FLOW Randy Buchanan, University of Southern Mississippi Christopher Winstead, University of Southern Mississippi Anton Netchaev, University of Southern Mississippi

More information

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS 1. INTRODUCTION An anaesthesia delivery system includes any machine, equipment or apparatus which supplies gases, vapours, local anaesthesia and/or

More information

Degassing ERC KK Saitama, Japan

Degassing ERC KK Saitama, Japan Degassing Degassing Solutions IDEX Health & Science degassers improve fluidic instrument precision and reliability by removing dissolved gases from fluids before they outgas and form problem-causing bubbles

More information

THE DANGERS OF USING A THREE-WAY VALVE AS A STOPCOCK IN A TRANSDUCER SYSTEM FOR MEASURING VENOUS BLOOD PRESSURE

THE DANGERS OF USING A THREE-WAY VALVE AS A STOPCOCK IN A TRANSDUCER SYSTEM FOR MEASURING VENOUS BLOOD PRESSURE THE DANGERS OF USING A THREE-WAY VALVE AS A STOPCOCK IN A TRANSDUCER SYSTEM FOR MEASURING VENOUS BLOOD PRESSURE J. W. SIVIITH~ PH.D., P.ENr ~ THE SYSTEM shown in Figure 1 has been widely used for the measuring

More information

Memorandum Background: Results and Discussion:

Memorandum Background: Results and Discussion: Memorandum To: Dr. Abercromby From: Jason Rapp and Matthew Stumbo Date: January 17, 2012 Subject: Vacuum Chamber Basics Technical Memo Background: We explored several aspects of vacuum chambers, in preparation

More information

Effect of Coiled Capillary Tube Pitch on Vapour Compression Refrigeration System Performance

Effect of Coiled Capillary Tube Pitch on Vapour Compression Refrigeration System Performance Effect of Coiled Capillary Tube Pitch on Vapour Compression Refrigeration System Performance Mutalubi Aremu Akintunde Federal University of Technology, Department of Mechanical Engineering Akure, Ondo

More information

Interface Pressure Mapping (IPM) Clinical Use of the Literature

Interface Pressure Mapping (IPM) Clinical Use of the Literature Interface Pressure Mapping (IPM) Clinical Use of the Literature Laura Titus OT Reg.(Ont.), PhD Student, Jan Miller Polgar PhD, OT Reg.(Ont.), FCAOT SJHC-Parkwood Seating Program London Ontario Faculty

More information

C O R P O R AT E B R O C H U R E

C O R P O R AT E B R O C H U R E CORPORATE BROCHURE OUR MISSION MORE THAN 40 YEARS OF EXPERIENCE EXPORT TO OVER 60 COUNTRIES MORE THAN 40 WORLDWIDE PATENTS O-Two Medical Technologies is dedicated to the provision of state-of-the-art

More information

Inspire rpap REVOLUTION FROM THE FIRST BREATH

Inspire rpap REVOLUTION FROM THE FIRST BREATH Inspire rpap TM REVOLUTION FROM THE FIRST BREATH The Inspire rpap The Inspire rpap is a revolutionary, non-invasive system for the initial stabilisation and resuscitation of infants. TM Its innovative,

More information

OPESAVER Super Irrigation System

OPESAVER Super Irrigation System OPESAVER Super Irrigation System Makoto Kishimoto, MD Dokkyo University School of Medicine Kishimoto Eye Clinic Moriyama City Shiga, Japan n ABSTRACT OPESERVER reduces the surge phenomenon during ultrasonic

More information

COURSE NUMBER: ME 321 Fluid Mechanics I Fluid statics. Course teacher Dr. M. Mahbubur Razzaque Professor Department of Mechanical Engineering BUET

COURSE NUMBER: ME 321 Fluid Mechanics I Fluid statics. Course teacher Dr. M. Mahbubur Razzaque Professor Department of Mechanical Engineering BUET COURSE NUMBER: ME 321 Fluid Mechanics I Fluid statics Course teacher Dr. M. Mahbubur Razzaque Professor Department of Mechanical Engineering BUET 1 Fluid statics Fluid statics is the study of fluids in

More information

Leadership in Fluid Management

Leadership in Fluid Management Leadership in Fluid Management Medical Vision leads the development in Fluid Management in arthroscopic surgery, to facilitate better visualization during surgery. Through our unique arthroscopy pump,

More information

Technical Data Sheet MF010-O-LC

Technical Data Sheet MF010-O-LC Technical Data Sheet MF010-O-LC - 1 - 1. Properties The oxygen measuring system MF010-O-LC determines the oxygen content in gas mixtures up to a temperature of 250 C. It is particularly suitable for the

More information

VENTILATORS PURPOSE OBJECTIVES

VENTILATORS PURPOSE OBJECTIVES VENTILATORS PURPOSE To familiarize and acquaint the transfer Paramedic with the skills and knowledge necessary to adequately maintain a ventilator in the interfacility transfer environment. COGNITIVE OBJECTIVES

More information

ELEVATION CORRECTION FACTORS FOR RADON MONITORS*

ELEVATION CORRECTION FACTORS FOR RADON MONITORS* ELEVATION CORRECTION FACTORS FOR E-PERM@ RADON MONITORS* P. Kotrappa and L. R. Stiefl Abstract-E-PERMm radon monitors are based on the principle of electret ion chambers and are usually calibrated in a

More information

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies IMPROVED CALIBRATION AND OPERATION OF A STABLE-GAS QUANTIFICATION MANIFOLD AS PART OF A RADIOXENON COLLECTION SYSTEM ABSTRACT Richard M. Williams, James C. Hayes, and Randy R. Kirkham Pacific Northwest

More information

INSTRUCTOR RESOURCES

INSTRUCTOR RESOURCES Gases: Dalton s Law INSTRUCTOR RESOURCES By Dale A. Hammond, PhD LEARNING OBJECTIVES introduce the concept of ideal gases. experimentally determine the relationship between pressure and amount of gas,

More information

Keeping anaesthetic pollution to minimum in theatre environment

Keeping anaesthetic pollution to minimum in theatre environment Vet Times The website for the veterinary profession https://www.vettimes.co.uk Keeping anaesthetic pollution to minimum in theatre environment Author : Gianluca Bini Categories : General, Vets Date : September

More information

Exercise 2-2. Second-Order Interacting Processes EXERCISE OBJECTIVE DISCUSSION OUTLINE. The actual setup DISCUSSION

Exercise 2-2. Second-Order Interacting Processes EXERCISE OBJECTIVE DISCUSSION OUTLINE. The actual setup DISCUSSION Exercise 2-2 Second-Order Interacting Processes EXERCISE OBJECTIVE Familiarize yourself with second-order interacting processes and experiment with the finer points of controller tuning to gain a deeper

More information

HAMILTON-C2 HAMILTON-C2. The universal ventilation solution

HAMILTON-C2 HAMILTON-C2. The universal ventilation solution HAMILTON-C2 HAMILTON-C2 The universal ventilation solution The universal ventilation solution HAMILTON-C2 - The compact ventilation solution The HAMILTON-C2 mechanical ventilator is a universal ventilation

More information

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay Summary This assay measures Human Granulocyte Colony Stimulating Factor (G-CSF) in a 96-well

More information

a personal guide to M A N A G I N G Chest Drainage

a personal guide to M A N A G I N G Chest Drainage a personal guide to M A N A G I N G Chest Drainage Table of Contents Your personal guide to Managing Chest Drainage is a quick and easy reference to help extend your understanding of chest tube drainage

More information

PURE SUBSTANCE. Nitrogen and gaseous air are pure substances.

PURE SUBSTANCE. Nitrogen and gaseous air are pure substances. CLASS Third Units PURE SUBSTANCE Pure substance: A substance that has a fixed chemical composition throughout. Air is a mixture of several gases, but it is considered to be a pure substance. Nitrogen and

More information

War Surgery Dr. Abdulwahid INTRODUCTION: AIRWAY, BREATHING

War Surgery Dr. Abdulwahid INTRODUCTION: AIRWAY, BREATHING War Surgery Dr. Abdulwahid INTRODUCTION: AIRWAY, BREATHING The aims of war surgery: Save life Avoid infectious complications Save limbs Minimize residual disability The outcome is influenced by: Type of

More information

Initiation and Management of Airway Pressure Release Ventilation (APRV)

Initiation and Management of Airway Pressure Release Ventilation (APRV) Initiation and Management of Airway Pressure Release Ventilation (APRV) Eric Kriner RRT Pulmonary Critical Care Clinical Specialist Pulmonary Services Department Medstar Washington Hospital Center Disclosures

More information

Unit 15 Manual Resuscitators

Unit 15 Manual Resuscitators 15-1 Unit 15 Manual Resuscitators GOAL On completion of this unit, the student should comprehend the proper operation of self-inflating resuscitation bags, flow-inflating resuscitation bags and gas-powered

More information

Outside Air Nonresidential HVAC Stakeholder Meeting #2 California Statewide Utility Codes and Standards Program

Outside Air Nonresidential HVAC Stakeholder Meeting #2 California Statewide Utility Codes and Standards Program 1 Outside Air Nonresidential HVAC Stakeholder Meeting #2 California Statewide Utility Codes and Standards Program Jim Meacham, CTG Energetics Agenda 2 In Situ Testing Results Reduced Ventilation after

More information

OIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply

OIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply OIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply 4.1.1 General The primary function of a system for handling fuel oil is to transfer oil from the storage tank to the oil burner at specified conditions

More information

Design of an Automated Perfusion Pump James J Higham

Design of an Automated Perfusion Pump James J Higham Design of an Automated Perfusion Pump James J Higham Abstract Department of Gastroenterology Repatriation General Hospital The design and construction of an automated perfusion pump commenced as part of

More information

23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN APRIL 2007

23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN APRIL 2007 23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN 16-20 APRIL 2007 AN INVESTIGATION INTO THE INTERRELATION BETWEEN THE INTERNAL AND EXTERNAL BALLISTICS OF FIRING A TP-T TANK AMMUNITION M. H.

More information

NOTE: If not used, provider must document reason(s) for deferring mechanical ventilation in a patient with an advanced airway

NOTE: If not used, provider must document reason(s) for deferring mechanical ventilation in a patient with an advanced airway APPENDIX: TITLE: Mechanical Ventilator Use REVISED: November 1, 2017 I. Introduction: Mechanical Ventilation is the use of an automated device to deliver positive pressure ventilation to a patient. Proper

More information

APPLYING VARIABLE SPEED PRESSURE LIMITING CONTROL DRIVER FIRE PUMPS. SEC Project No

APPLYING VARIABLE SPEED PRESSURE LIMITING CONTROL DRIVER FIRE PUMPS. SEC Project No APPLYING VARIABLE SPEED PRESSURE LIMITING CONTROL DRIVER FIRE PUMPS SEC Project No. 1803007-000 November 20, 2006 TABLE OF CONTENTS I. ABSTRACT...1 II. INTRODUCTION...1 III. HISTORY...2 IV. VARIABLE SPEED

More information

Methods of Rating Flow

Methods of Rating Flow Methods of Rating Flow What engineers need to know when using CV as a measurement of flow in pneumatic applications In the pneumatic industry, CV is one standard for expressing the flow capacity of devices

More information

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Lab 1c Isentropic Blow-down Process and Discharge Coefficient 058:080 Experimental Engineering Lab 1c Isentropic Blow-down Process and Discharge Coefficient OBJECTIVES - To study the transient discharge of a rigid pressurized tank; To determine the discharge coefficients

More information

Constant Pressure Inlet (CCN) Operator Manual

Constant Pressure Inlet (CCN) Operator Manual Constant Pressure Inlet (CCN) Operator Manual DOC-0125 Revision J 2545 Central Avenue Boulder, CO 80301-5727 USA C O P Y R I G H T 2 0 1 1 D R O P L E T M E A S U R E M E N T T E C H N O L O G I E S, I

More information

AUTOVENT 4000 VENTILATOR

AUTOVENT 4000 VENTILATOR OVERVIEW AUTOVENT 4000 Only properly trained and approved Escambia County Bureau of Public Safety Paramedics are to use the AutoVent 4000 ventilator manufactured by LSP to transport patients already on

More information

Experiment 8: Minor Losses

Experiment 8: Minor Losses Experiment 8: Minor Losses Purpose: To determine the loss factors for flow through a range of pipe fittings including bends, a contraction, an enlargement and a gate-valve. Introduction: Energy losses

More information

Solutions. Digital pressure management solutions. Copyright 2013 GaleMed Corporation

Solutions. Digital pressure management solutions. Copyright 2013 GaleMed Corporation Solutions Digital pressure management solutions Content 1. About GiO Solutions 2. What is Gio 3. Clinical Applications 4. Product Range 5. Features 6. Specifications 7. Accessories 8. Order Information

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results. Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions

More information

Alternate Suction to Reduce Prolonged Air Leak

Alternate Suction to Reduce Prolonged Air Leak Alternate Suction to Reduce Prolonged Air Leak Alessandro Brunelli Division of Thoracic Surgery Ancona, Italy AIR LEAKS Frequent Prolong the hospital stay Increase the hospital costs Increase the risk

More information

Chest Drain. Aims: To remove air I fluid from the pleural space. To prevent air I fluid going back to pleural space

Chest Drain. Aims: To remove air I fluid from the pleural space. To prevent air I fluid going back to pleural space Chest Drain Aims: To remove air fluid from the pleural space To prevent air fluid going back to pleural space To restore negative pressure in the pleural space so to re-expand the lungs . Mechanism of

More information

Major Design Features

Major Design Features Major Design Features 3Flex Features Design Benefits Discussion 316 SS fittings and pneumatically actuated, hard-seal valves Provides virtually leak-free gas management with the lowest outgassing rate

More information

Endotracheal Suctioning: In Line ETT

Endotracheal Suctioning: In Line ETT Approved by: Endotracheal Suctioning: In Line ETT Gail Cameron Senior Director Operations, Maternal, Neonatal & Child Health Programs Dr. Paul Byrne Medical Director, Neonatology Neonatal Policy & Procedures

More information

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE TITLE: ISSUED FOR: ENDO TRACHEAL SUCTIONING OF THE ADULT PATIENT (suc03) Nursing, Respiratory DATE: REVIEWED: PAGES: 11/82 12/18 1 of 5 RESPONSIBILITY: RN,

More information

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the 1151 METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS 1.0 Scope and Application. 1.1 Analytes. Analyte Volatile Organic Compounds (VOC) CAS No. No CAS number assigned 1.2 Scope. This method

More information

Leak Checking Large Vacuum Chambers

Leak Checking Large Vacuum Chambers Leak Checking Large Vacuum Chambers Technical Overview Vacuum Technologies Introduction Understanding the pump-down characteristics of a large vacuum vessel is critical for determining whether the vacuum

More information

TOPIC: KNOWLEDGE: K1.01 [2.5/2.7] An atmospheric pressure of 15.0 psia is equivalent to... A psig. B psig. C psig.

TOPIC: KNOWLEDGE: K1.01 [2.5/2.7] An atmospheric pressure of 15.0 psia is equivalent to... A psig. B psig. C psig. P73 An atmospheric pressure of 15.0 psia is equivalent to... A. 30.0 psig. B. 29.4 psig. C. 14.7 psig. D. 0.0 psig. ANSWER: D. P273 A pressure gauge on a condenser reads 27.0 inches Hg vacuum. What is

More information

Gas Laws: Boyle s and Amonton s Laws MCTC Chemistry v.9.17

Gas Laws: Boyle s and Amonton s Laws MCTC Chemistry v.9.17 Gas Laws: Boyle s and Amonton s Laws MCTC Chemistry v.9.17 Objective: The purpose of this experiment is confirm Boyle's and Amontons' Laws in the laboratory. Prelab Questions: Read through this lab handout

More information

Figure 1 Schematic of opposing air bearing concept

Figure 1 Schematic of opposing air bearing concept Theoretical Analysis of Opposing Air Bearing Concept This concept utilizes air bearings to constrain five degrees of freedom of the optic as shown in the figure below. Three pairs of inherently compensated

More information

General Accreditation Guidance. User checks and maintenance of laboratory balances

General Accreditation Guidance. User checks and maintenance of laboratory balances General Accreditation Guidance User checks and maintenance of laboratory balances January 2018 Copyright National Association of Testing Authorities, Australia 2010 All intellectual property rights in

More information

The NAMIC Dream Kit. Single Operator Control. Reduce Contrast Waste. Less Force to Inject. Compensator * Manifold. How the Compensator Manifold Works:

The NAMIC Dream Kit. Single Operator Control. Reduce Contrast Waste. Less Force to Inject. Compensator * Manifold. How the Compensator Manifold Works: Dream Kit The NAMIC Dream Kit Single Operator Control. Reduce Contrast Waste. Less Force to Inject. For more than 40 years, NAMIC * Fluid Management products have been providing cardiologists, technicians

More information

Micro Dial-Flowmeter. Instructions for Use

Micro Dial-Flowmeter. Instructions for Use Micro Dial-Flowmeter Instructions for Use 702-0082.8 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

Recommendations on Checking Anaesthesia Delivery Systems

Recommendations on Checking Anaesthesia Delivery Systems Page 1 of 11 Recommendations on Checking Anaesthesia Delivery Version Effective Date 1 Oct 1992 (reviewed Feb 07, Feb 02) 2 2004 3 Nov 2011 4 Dec 2016 Document No. HKCA T1 v4 Prepared by College Guidelines

More information

User Manual. Model 3930 Vacuum Assisted Venous Drainage Controller w/ WAGD Suction Flowmeter

User Manual. Model 3930 Vacuum Assisted Venous Drainage Controller w/ WAGD Suction Flowmeter User Manual Model 3930 Vacuum Assisted Venous Drainage Controller w/ WAGD Suction Flowmeter Boehringer Laboratories, LLC 300 Thoms Dr. Phoenixville, PA 19460 800-642-4945 3930.003 Rev D P/N 34031 1 of

More information

The Discussion of this exercise covers the following points:

The Discussion of this exercise covers the following points: Exercise 3-2 Orifice Plates EXERCISE OBJECTIVE In this exercise, you will study how differential pressure flowmeters operate. You will describe the relationship between the flow rate and the pressure drop

More information

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products Saturn Evo SATURN EVO COLOR A N A E S T H E S I A S Y S T E M i n t e l l i g e n t s i m p l i c i t y Manufacturer of high quality anaesthesia products SATURN EVO COLOR PATIENT SAFETY AND USER PATIENT

More information

DIAPHRAGM PUMPS WITH KNF STABILIZATION SYSTEM

DIAPHRAGM PUMPS WITH KNF STABILIZATION SYSTEM DIAPHRAGM PUMPS WITH KNF STABILIZATION SYSTEM DATA SHEET E 037 N 940.5 APE as AC version in IP 54 N 940.5 APE-W with multi-voltage power supply input (As an option adjustable with potentiometer or by external

More information

Astral in AirView: Improving patient care through connectivity. ResMed.com

Astral in AirView: Improving patient care through connectivity. ResMed.com Astral in AirView: Improving patient care through connectivity ResMed.com This guide will assist you with: Setting up the ResMed Connectivity Module for Astral 2 Troubleshooting the ResMed Connectivity

More information

Students measure the change in pressure by varying the volume of trapped air in a syringe while:

Students measure the change in pressure by varying the volume of trapped air in a syringe while: How Does a Trapped Gas Behave? Teacher Information Objective Students investigate the effect of changes in the volume of a confined gas on pressure at constant temperature. Using the pressure sensor, students

More information

How to specify a product. Process Sensors and Mechanical Instruments

How to specify a product. Process Sensors and Mechanical Instruments How to specify a product Process Sensors and Mechanical Instruments Keep the overview. Here is some guideline information on how to specify our products. Intended as supplementary help to specification

More information

TPM TIP. Oil Viscosity

TPM TIP. Oil Viscosity TPM TIP Oil Viscosity TIP #1 Classification according to viscosity is the most prevalent method of describing oils, and the most common classification systems are those of the SAE, AGMA, and ISO. Each

More information

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV 16-1 INTRODUCTION 16-1.1 Purpose. The purpose of this chapter is to familiarize divers with the techniques used to mix divers breathing gas. 16-1.2 Scope. This chapter

More information

2600T Series Pressure Transmitter Models 264DC Differential and 264HC Gage Level Transmitters. Kent-Taylor

2600T Series Pressure Transmitter Models 264DC Differential and 264HC Gage Level Transmitters. Kent-Taylor INDUSTRIAL INSTRUMENTS AND CONTROLS SPECIALIST Kent-Taylor 2600T Series Pressure Transmitter Models 264DC Differential and 264HC Gage Level Transmitters Features Include Base accuracy : ±0.075% Span limits

More information

Introductory Lab: Vacuum Methods

Introductory Lab: Vacuum Methods Introductory Lab: Vacuum Methods Experiments in Modern Physics (P451) In this lab you will become familiar with the various components of the lab vacuum system. There are many books on this topic one of

More information

Chem 110 General Principles of Chemistry

Chem 110 General Principles of Chemistry CHEM110 Worksheet - Gases Chem 110 General Principles of Chemistry Chapter 9 Gases (pages 337-373) In this chapter we - first contrast gases with liquids and solids and then discuss gas pressure. - review

More information

Improving distillation tower operation

Improving distillation tower operation Improving distillation tower operation Measuring differential pressure across long sections of distillation columns has always been challenging, but purpose-built sensor systems provide a solution Fast

More information

Module 6. Tightness Testing

Module 6. Tightness Testing Module 6 Tightness Testing IGE / UP / 1B New Tightness Testing Procedure Covers pipework up to 35mm and installation volumes of 0.035m3 Objectives By the end of Module 6, Tightness Testing and Direct Purging

More information

Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung. Susan Blonshine RRT, RPFT, FAARC, AE-C

Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung. Susan Blonshine RRT, RPFT, FAARC, AE-C Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung Susan Blonshine RRT, RPFT, FAARC, AE-C Joint ATS/ERS Taskforce Recent literature reviewed Surveyed current technical

More information