Using the Agilent 6890N GC/MS for Sampling Volatile Organic Compounds. Bob Castro, B.S. Chemical Eng., M.S. Environmental Eng.

Size: px
Start display at page:

Download "Using the Agilent 6890N GC/MS for Sampling Volatile Organic Compounds. Bob Castro, B.S. Chemical Eng., M.S. Environmental Eng."

Transcription

1 Using the Agilent 6890N GC/MS for Sampling Volatile Organic Compounds by Bob Castro, B.S. Chemical Eng., M.S. Environmental Eng. Edited by Don Marek, Lab Manager 1

2 Table of Contents GC/MS Concepts... 3 Figure 1-1: Stainless Steel Silonite Coated Canister... 4 Figure 1-2: GC/MS Conceptual Diagram...5 Figure 1-3: GC-MS Schematic Diagram... 6 GC/MS Standard Operations Procedure... 8 Figure Tuning with the PTFBA spectra... 8 Figure 2-2. Peak widths Figure 2-3 Relationship between peak width and AMU gain and AMU offset Figure 2-4: Sample Log Table Figure 2-5. Start sequence options...13 BFB Tuning Validation...13 Microscale Purge and Trap/ Pre-concentration...14 Figure: Schematic Diagram of the Pre-concentrator Quantification Standard Operation Procedures Figure 3-1: Butane Calibration Curve...19 Using the Chemstation Data Analysis Software to Qauntitate the Data...20 Figure 3-3: Manual Integration of the Isoprene Peak Figure 3-4: Peak Overlap...22 Standard Operation Procedure for Canister Cleaning...23 Standard Operation Procedure for Standards Preparation Purging the Regulator Standard Preparation by Static Dilution Internal Standard Preparation Standard Operation Procedure For VOC Sample Collection Figure: Schematic Diagram of Sample Collection System...28 References

3 GC/MS Concepts Introduction The Analytical tools of Gas Chromatography with Mass Spectrometry Detection (GC/MS) will be used to identify the Volatile Organic Carbons (VOC). The USEPA Compendium method TO- 15 will be followed in the sample collection and analysis of the ambient air samples. Compendium Method TO 15 The analytical strategy of Compendium Method TO -15 involves using a high resolution GC coupled to a mass spectrometer. Mass spectra for the individual peaks in the total ion chromatogram are examined with respect to fragmentation pattern of ions corresponding to various VOCs including the intensity of primary and secondary ions. The fragmentation pattern is compared with stored spectra taken under similar condition in order to identify the component. For any given compound, the intensity of the primary fragment is compared with the system response to the primary fragment for known amounts of the compound. This establishes the compound concentration that exists in the sample. Canisters The canisters used to collect the ambient air samples are made of stainless steel and coated with silonite. The canisters are 6 Liters in volume and are evacuated to a vacuum of 50 militorr. A picture of the canister is shown in Figure 1-1. below. When the valve to the can is opened, the air passes through a flow restrictor which is an orifice that restricts the flow rate into the can. A diaphragm below the flow restrictor, regulates the flow into the can so that the flow of air at a fixed rate. 3

4 Figure 1-1: Stainless Steel Silonite Coated Canister Preconcentration The air sample from the canister is directed to a pre-concentrator which traps the VOCs and remove much of the excess air in the sample. Additionally the pre-concentrator has a purge and trap to remove water vapor from the sample. The sample is desorbed off the trap and carried off with Helium into the GC column for separation. GC/MS GC/MS is the combination of two powerful analytical techniques. Gas chromatography is the physical separation of two or more compounds based on their volatility from the liquid phase to the gas phase. The gas chromatograph employs a carrier gas (mobile phase) to move a sample in the vapor phase through a column coated with a stationary phase where separation takes place. A detector converts the column eluent to an electrical signal that is measured and recorded. The out put of the GC is a plot of detector signal abundance versus time. The abundance remains at a low baseline level except when a separated sample component elutes from the column and produces a peak in the chromatogram plots. Chromatographic peaks can be identified from their 4

5 corresponding retention times measured from the time of sample injection to the time of the peak maximum. The retention time of any component peak is unaffected by the presence of other sample components. The height of the area of a peak may be used to measure the concentration of a component in the sample mixture. Mass Analysis A mass spectrometer is one kind of GC detector. As the separated sample component molecules elute from the column to the inside of the MS, they are bombarded with energy. This is to remove an electron from them and converts them into unipositive ions. Some chemical bonds may also break up in this process and the resulting fragments may rearrange or break up again to form more stable fragments. A mass spectrum is a recording of the masses of each of the ionized fragments, representing a unique fingerprint of a molecule that can be used in identification. How GC/MS works: Figure 1-2: GC/MS Conceptual Diagram As shown in Figure 1-2 above, the sample is separated into its components by the gas chromatograph. The components are then ionized and identified by their characteristic spectra produced by the mass spectrometer. What are Mass Spectra? ( Molecular Fingerprints ) A mass spectrum is a plot showing the mass/charge ratio (in a.m.u) versus abundance data for ions from the sample molecule and its fragments. The charge of ions formed in the GC/MC is +1 5

6 and the mass / charge ratio of any fragment is therefore normally equal to the mass for that fragment. The largest peak in the spectrum is called the BASE PEAK. Certain fragments are more prone to form from the parent molecule than others, due to the presence of functional groups in the molecule and their interconnection. The masses of these fragments are used to deduce the structure of the parent compound. The ionized parent molecule, when seen as part of the mass spectrum, is referred to as MOLECULAR ION. ABCD - e - ABCD +. ABCD +. A + + BCD. A. + BCD + BC + + D. ABCD. + D. + ABC + A + BC + The mass spectra of certain compounds exhibit cluster of mass peaks. These clusters represent naturally occurring impurities or isotopes that are present for carbon, nitrogen, sulfur, chlorine, bromine, and a few other elements. The relative percentage of these cluster ions provides more clues useful in identifying a parent molecule from its molecular fingerprints. GS/MS Flow Chart Figure 1-3: GC-MS Schematic Diagram GC Column: The column separates the mixture into its components. 6

7 Interface: The interface between the GC and MS must transfer the sample / carrier gas matrix from positive pressure inside the MS. It must reliably transmit the entire sample without compromising the performance of the GC or the MS GC Oven: The temperature of the GC oven can be programmed to optimize separation and resolution of the components of a given sample. Precise temperature control of the oven allows the chromatography to be reproducible. Ion Source: In the electron impact (EI) ion source, a filament wire emits electrons that strike the sample molecule as they elute from the column. Ions are produced from the electron/molecule collisions and a series of lenses propels and focuses them toward the quadrupole. Quadrapole: The quadrapole consists of four conductive rods, separates ionized fragments according to their mass/charge ratio. Voltage on the rods can be set to allow ions of a particular mass to pass through and ions of the wrong mass will be pumped away by the vacuum system. Vacuum System: The MS operates under a vacuum of about 10-5 torr. The reduced pressure increases the distance between the molecules, minimizing the number of intermolecular collisions inside the ion source. Detector: The detector counts the ions that pass through the quadrapole. This signal is small and must be amplified. An electron multiplier detector measures the abundance of each component ion, and the presence of an ion is ultimately recorded as a mass peak in a mass spectrum. Data System: The data system is responsible for the total control of the GC/MS system. This includes GC temperatures, tuning the MS system, controlling the voltages on the quadruple during the data acquisition, detecting the abundance of each ion, and processing the acquired data. 7

8 GC/MS Standard Operations Procedure Before samples can be analyzed using the GC/MS, the mass spectrometer must be tuned. The MS is tuned by adjusting several parameters while the Per-fluoro tributylamine (PFTBA)is injected to the MS chamber Below is the procedure for tuning the GC/MS 1) Click on the Instrument #1 icon to start the Chemstation software. 2) From the View menu select Manual Tune. 3) When prompted be sure changes are saved. Switch View Now?, select Yes 4) In the Manual Tune view select the AdjParam menu to adjust the tune parameters. 5) From The AdjParam menu select Edit MS Params menu 6) From the edit MS Params menu select Scan and you should see a similar PTFBA spectra a s shown in Figure 2-1. Wait 30 seconds for the PFTBA levels to stabilize before changing any tune parameters. Figure Tuning with the PTFBA spectra The scan from amu should have less than 200 peaks. If there are more than 200 peaks then some sort of contamination is entering the MS chamber and further trouble shooting should be done. 8

9 The mass spectrum of PFTBA has a base peak of 69 amu. The relative abundance of the mass peak of 219 to the base peak in the PTFBA spectra should be between 40 and 85%. The relative abundance of the mass peak of 502 to the base peak in the PTFBA spectra should be between 2.5 and 5%. In the mass column in Figure 2-1, if the masses of 69, 219, and 502 are off by more than 0.2amu then the mass axis should be adjusted. This is done by selecting Mass Axis from the calibrate menu. By doing this the mass axis is automatically adjusted by the software. The Rel Abund column shows the relative abundance of the 69, 219, and 502 masses. To adjust the relative abundance, the EN lens and the EntLens offset parameters should be adjusted. The Entrance lens parameters are inversely proportional to the relative abundance, which means that if you wish to increase the relative abundance you should lower the entrance lens setting and vise versa. The abundance of mass 69 should be between 200,000 and 400,000. To adjust the abundance the EM volts parameter should be adjusted. The EM volts are directly proportional to the abundance. An increase in the EM volts parameter will increase the abundance. Click on the Prof button to get a profile of the 69, 219, and 502 peaks. The profiles are shown in Figure 2-2. The peak with at 50% of the peak height (Pw50) should be between

10 Figure 2-2. Peak widths To adjust the peak widths the amu gain and the amu off set parameters should be adjusted. The relationship between the peak widths and the amu gain and the AMU off set are shown in Figure 2-3. In order to increase the peak width the amu gain and the amu off set parameters should be decreased, and vice versa to decrease the peak width. Figure 2-3 Relationship between peak width and AMU gain and AMU offset 7) After changing the tune parameters print the parameter settings by selecting print from the file menu. 8) Click on OK to leave the edit Parameters view. 9) Save the tune values by selecting save tune values from the file menu. Save the file as ATUNE.U. 10) Select Top from the view menu, to switch to the top view. When Prompted be sure changes are saved. Switch View Now?, select Yes Note, if changes have not been saved go ahead and do so now other wise the changes will be lost once the view has been switched. After the mass spectrometer has been tuned then a sequence table will be prepared to run the samples using the AAVOC1 method. The instrument parameters for the GC/MS-FID system are shown in the table below: 10

11 Columns Flow Rate Injector Temperature MS transfer interface FID settings MS settings Temperature Program DB-1. 60m long, 0.32mm OD, 1 micron coating Gas Pro, 30m long, 0.32mmOD He; 1.3 ml/min Mode: Splitless 150 C 280 C Temperature: 250 C Hydrogen flow: 40 ml/min Air flow: 450 ml/min Make up gas flow: 45 ml/min Scan range: atomic mass units Scan Rate: 4.3 scans per second Solvent delay: 4.17 min MS Source: 230 C MS Quad: 150 C Initial oven temperature:20 C Initial hold time:2.7 First ramp:2.76 C/min Next temp: 90 C Hold: 0 min Second ramp: C/min Next temp: 130 C Hold: 0 min Third ramp: 6.18 C/min Next temp: 220 C Hold: 18 min Total Run Time: ) From the top view select edit sample log table from the sequence table. A sample sequence table is shown in Figure 2-4: 11

12 Figure 2-4: Sample Log Table The vial number will always be 1 The data file name is limited to eight characters The method for GC/MS analysis that the class will use is AAVOC1 The sample name can be more descriptive and can be longer than eight characters. More information about the sample can be entered in the Miscellaneous Information Box. Normally such information and the sample volume and internal standard volume injected and perhaps the canister size are entered here. The Expected Barcode Box is left blank. At the end of each sequence, there should be a data file called end 2 which is ran with method end2. The end2 method turns off the liquid nitrogen at the end of the sequence.. 12) Click on OK to close the sample log table. 13) Select Run from the Sequence menu to run the sequence. Figure 2-5 shows the start sequence options 14) Select Full Method, select Inject anyway on Barcode Mismatch. Do not select Overwrite existing data files. 15) Entering a sequence comment is optional. 16) Enter your name under the operator name. 17) Enter C:\msdchem\1\DATA\Measurements Class\ for the data file directory 18) Click on the Run Sequence button to run the sequence. 12

13 Figure 2-5. Start sequence options BFB Tuning Validation For quality control and quality assurance, to verify that the MS has been properly tuned, BFB tuning validation must be performed. A sample volume of 20 ml of 103 ppb Bromofluorobenzene (BFB) is used for this validation process. After the BFB sample run is complete, the BFB data file is analyzed using the Chemstation Data Analysis software. 1) Click on the Instrument #1 Data analysis icon. 2) Open the BFB data file from the data file directory chosen, by selecting Load Data File from the file menu. 3) Load method voc0316.m, by selecting load method from the file menu 4) Zoom in on the BFB peak which has a retention time of 33.3 minutes. Zooming is performed by dragging a square around the peak, while pressing the left mouse button. 5) Obtain an average spectrum of the BFB peak by dragging a square inside the peak while pressing the right mouse button. 6) Select Evaluate BFB from the Tuner menu 7) Evaluate the BFB to the screen 8) If any of the ions in the BFB tuning evaluation fail, then trouble shooting must be done to determine why it failed. 13

14 Microscale Purge and Trap/ Pre-concentration. In order to avoid interferences of the MS results, the air, carbon dioxide and water must be removed from the sample. These components are removed with the Entech 7100a preconcentrator. In Module 1 (M1) of the pre-concentrator the VOC s are trapped onto glass beads while all gases more volatile than methane, which include carbon dioxide and air, are allowed to pass through. Module 2 (M2) of the pre-concentrator contains a sorbent material called Tenax which absorbs the water from the sample. In Module 3 (M3), the focusing module, the VOC s in the sample are liquefied and then quickly vaporized, and injected into the GC transfer line. A schematic diagram of the pre-concentrator is shown in the figure below. The settings for the pre-concentrator are shown in the tables below: Figure: Schematic Diagram of the Pre-concentrator. 14

15 M1 M1 bulk head M2 M2 bulk head M3 Event ( C) ( C) ( C) ( C) ( C) Concentration Preheat No -- Desorption Bake out Module Event Temperatures Zone ( C) Rotary Valve 100 GC Transfer Line 100 Manifold Transfer Line 100 Auto sampler Rotary Valve 100 Sample ambient Zone Temperatures Flows and Volumes Pre-flush Flow rate Volume Medium (sec) (cc/min) (cc) Internal Standard Analytical Standard Sample Final sweep/purge flush M1-M2 Transfer Flows and Volumes Event (Min) Transfer to Focuser 2.5 Injection 2 Bakeout (M1 and M2) 5 Focuser Bake out 3 Wait Time 64 Event durations 1) Connect the canisters to the sample ports. Make the connections finger tight first, then tighten with a wrench. 2) Click on the NT 7000.exe icon to open the Entec Pre-concentrator Software. 3) Enter the sample names in the same order as the sample log table for the GC/MS 4) The internal standard volume for all samples, will be set at 20 ml 5) After entering the sample name click on the sample port that the can is connected to. 6) Enter the sample volume of 400 ml 7) Enter C:\Smart\tceq4.mpt as the method 8) Click on the add icon to add the sample to the sequence table 15

16 9) After all the samples have been added to the sequence table, click on the leak icon to leak check the connections to the canisters. It is important that there are no leaks otherwise air from the lab can dilute the sample drawn into the pre-concentrator. Make sure that all the canister valves are closed before leak checking. 10) Click Go to start the leak check. The leak check fails if the initial pressure is not the same as the final pressure. If a sample port fails, undo the connection, reconnect, retighten, and leak check again. 11) After all sample ports pass the leak check, click on the green Go Icon. Make sure that all canister valves are open before pressing Go. 16

17 Quantification Standard Operation Procedures Calibration curves have been developed to quantitate 59 different VOC components. A list of the components is given below in table 3-1. The first five components ethane through propylene, are measured using a flame ionization detector (FID). All the other components are measured using the mass spectrometer as the detector. The components are listed in the order of their retention time. The target ion used to quantitate the component and the qualifying ions used to identify the component are also listed in Table 3-1. An asterisk * is placed before compounds which are internal standards. In each sample run the same amount of internal standard is added. For each sample run, 5.15 ppb internal standard is added. Each VOC component is measured relative to the internal standard component listed above it in table 3-1. Table 3-1: VOC Component List Retention Time Target Qualifying Qualifying Qualifying Compound Name (min) Ion Ion (1) Ion (2) Ion (3) ethane 5.41 FID ethylene 6.09 FID Acetylene 10.2 FID propane FID propylene FID *bromochloro methane isobutane butene butane trans-2-butene cis-2-butene isopentane pentene n-pentane isoprene trans-2-pentene cis-2-pentene ,2-dimethylbutane ,3-dimethylbutane methylpentane methylpentane hexene n-hexane ,4-dimethylpentane methylcyclopentane *1,4 difluoro benzene benzene cyclohexane methylhexane ,3-dimethylpentane methylhexane

18 2,2,4-trimethylpentane n-heptane methylcyclohexane ,3,4-trimethylpentane methylheptane toluene methylheptane n-octane *chlrorbenzene-d ethyl benzene para-xylene meta-xylene styrene ortho-xylene n-nonane isopropylbenzene propylbenzene m-ethyl toluene p-ethyl toluene ,3,5-trimethylbenzene o-ethyl toluene ,2,4-trimethylbenzene decane ,3 trimethylbenzene m-diethyl benzene p-diethyl benzene Undecane Dodecane

19 Figure 3-1: Butane Calibration Curve The Calibration curve for butane is shown in Figure 3-1. The response ratio on the Y-axis is calculated by the following equation. Response Ratio = A c /A i Where A c is the peak response area of the component, and A i is the peak response area of the internal standard. The internal standard for the component butane, is, bromochloro methane. The amount ratio on the X-axis is calculated by the following equation. Amont ratio = C c /C i Where C c is the concentration of the component and C i is the concentration of the internal standard. C i is always the same, it is 5.15 ppb. The calibration curve is a 6 point curve made by measuring the VOC Standard concentrations of 24.3, 9.72, 2.43, 0.972, 0.413, and ppb. 19

20 Using the Chemstation Data Analysis Software to Qauntitate the Data 1) Click on the Instrument #1 Data Analysis icon to start the software. 2) We will be using a method called VOC0316W.M to quantitate the data. Load the method by selecting load method from the File menu. Select VOC0316W.M from the methods folder. 3) Load your data file by selecting Load Data File from the File menu. 4) We will initially quantitiate the data by generating a report. Select Generate a Report from the Quant menu. The report generated shows the estimated concentrations in the sample. The estimated concentrations are estimated based on the peak areas measured in the chromatogram based on an algorithm in the software. Unfortunately this algorithm is not perfect and occasionally makes mistakes, so the peak areas need to be measured or integrated manually. 5) You can integrate the peaks manually by selecting Qedit Quant Result from the Quant menu. For the most part peaks the Chemstation algorithm for integrating peaks works fine for smooth peaks. If the peak is jagged as shown in Figure 3-2. below, the peak will need to be manually integrated. The peak should be located within the retention time window. The peak should be located on the retention time marker. The retention time marker is the cyan dashed vertical line. The area of integration is marked by the dark blue line. Since the isoprene peak is jagged the algorithm incorrectly started measuring the peak late and ended early. This can be corrected by the following steps: Figure 3-2: Qedit, Peak Integration of Isoprene 20

21 6) Zoom into the peak by dragging a square over the peak while holding down the left mouse button. Double clicking the left mouse button zooms out. 7) Manually integrate the peak by clicking the right mouse button on the beginning of the peak and dragging to the end point of the peak. Note, if the end of the peak peak tails like the last peak on the right in Figure 3-2, the end point of the peak is where the peak signal begins to become horizontal. After this step is completed the dark blue peak integration area line should look as it is in Figure 3-3 below. Note, that the response has increased but the estimated concentration is still 1.0 ppb. This indicates that the integrated peak area response is below the lowest level in the calibration curve. Figure 3-3: Manual Integration of the Isoprene Peak 21

22 Figure 3-4: Peak Overlap When peaks over lap as shown in Figure 3-4 the end point of the peak is where the next peak starts to rise. In Figure 3-4, the 2,3 dimethyl butane peak which has a retention time of 9.8 minutes is overlapping with the 2-methylpentane peak which has a retention time of 9.9 minutes. 8) After each peak has been examined or manually integrated click on the Exit Icon in the Quick Edit activated window. When prompted select Yes to save your changes made to the quantitation results. 9) Select View Quant Results from the Quant menu. This generates a text file with the quantitation results. Save the text file. 10) The last and final step is to get the text data into a more usable format. Cut the header information from the text and save the resultant text file under a different name. The resultant text file can be imported into an excel spreadsheet using fixed width separation to put the text into columns. 22

23 Standard Operation Procedure for Canister Cleaning After the sample has been analyzed by the GC/MS the 6L canisters are cleaned for reuse. The canisters are cleaned by flushing the canisters with humidified zero grade air, evacuating the canister to 0.2 psia with a rough pump, and further evacuating the canister to 100 militorr. The flushing and evacuations are repeated for four cycles. In the final evacuation the canisters are evacuated to 50 militorr. During the flushing/evacuation cycles heating bands heat the canisters to vaporize any condensation along the canister walls. 1) Make sure the cylinder pressure of the zero grade air is above 500 psig. If the Pressure is below 500 psig, replace the zero grade air cylinder with a new one. Make sure the line pressure is at 50 psig. 2) Make sure that the water level of canister cleaner is between 80-20%. If the water level is below 20% then HPLC grade water needs to be added. Before removing the humidifying vessel from the canister cleaner be sure to close the valve to the zero grade air cylinder. Fill the humidifying vessel with HPLC grade water till the level is 80% full. Re connect the humidifying vessel to the can cleaner. Open the valve to the zero grade air cylinder. Check for an air leak on the nut of the humidifying vessel. Spray the nut with a 50/50 mixture of methanol and water. If bubbles form at the nut. Then tighten the nut till the leak stops. 3) Wrap the heating bands around the can. Turn on the power strip which the heating bands are plugged into. 4) Connect can to Canister Cleaner. Tighten the nuts by hand first. Then tighten them with a 9/16 wrench. Make sure the valves to the canisters are closed. 5) Open the software to control the canister cleaner by clicking on the canister cleaner icon. 6) Click on Open and select canclean.m30 to load the set points for the method. 7) Click on Run to get to the run control screen. 8) Before the can cleaning process is started, the connection of the canisters to the canister cleaner must be leak checked. Click on the rough pump button wait till the pressure of the system is approximately 0.1 psia 9) Click on the all off button. The system should be able to maintain a vacuum for 1 minute. If the pressure increases by 0.1 psi during this time then there is a leak in the system. If there is a leak in the system, trouble shooting will need to be done to identify the leak If the pressure of the system does not increase during the 1 minute, then the leak check has passed. 10) Open the valves to the cans. 11) Click on the Go Icon to start the can cleaning process. 12) After the cans have gone through 4 cleaning cycles and final evacuation. Close the valves to the can. Hit the all off button. Turn off the power to the heating bans. 23

24 Standard Operation Procedure for Standards Preparation Purging the Regulator There are three standards, which are the Internal Standard, the 4-Bromofluro-Benzene (BFB) and the Ozone Precursor Mix. Each standard cylinder has a regulator. When a new standard cylinder is received the regulator must be flushed with the new standard to remove pockets of lab air in the regulator. 1. The cylinder valve on the cylinder should be open the regulator valve should be closed. The needle valve to the cylinder should be closed. 2. Connect the cylinder to the canister cleaner. 3. Perform a manual leak check of the system. Click on the rough pump button wait till the pressure of the system is approximately 0.1 psia 4. Click on the all off button. The system should be able to maintain a vacuum for 1 minute. If the pressure increases by 0.1 psi during this time then there is a leak in the system. If there is a leak in the system, trouble shooting will need to be done to identify the leak If the pressure of the system does not increase during the 1 minute, then the leak check has passed. 5. Close the regulator valve on the regulator. 6. Open the needle valve to the cylinder. 7. Click on the rough pump button and evacuate the system to about 0.1 psia. 8. Click on the all off button 9. Open the regulator valve on the regulator slowly till the gauge on the regulator reads about -5 psig. 10. Close the needle valve to the cylinder. 11. Click in the rough pump button and evacuate the system to about 0.1 psia. 12. Click on the all off button 13. Open the needle valve to the cylinder to allow some gas to flow through. The gauge on the regulator should increase to -5 psig. 14. Repeat steps 9-12 six times to flush the regulator. 24

25 Standard Preparation by Static Dilution A VWR Traceable Pressure Meter will be used in preparing a static dilution. This pressure meter should be re-calibrated every year. The pressure meter reads gauge pressure so add barometric pressure in the room (usually 14.7psi) to the meter reading to convert the pressure into absolute terms. A diagram of the connections of the pressure meter, canister and the cylinder should be connected to the canister cleaner is shown in the figure below. 1) The cylinder valve on the cylinder should be open the regulator valve should be closed. The needle valve to the cylinder should be closed. 2) Connect the cylinder, a clean evacuated canister, and the pressure meter to the canister cleaner. Note: it important that the canister is at room temperature. If the canister has just been cleaned allow the canister 1 hour to cool to room temperature before using in the preparation of a static dilution. 3) Perform a manual leak check of the system. Click on the rough pump button wait till the pressure of the system is approximately 0.1 psia. (Use reading from canister cleaner system) 4) Click on the all off button. The system should be able to maintain a vacuum for 1 minute. If the pressure increases by 0.1 psi during this time then there is a leak in the system. If there is a leak in the system, trouble shooting will need to be done to identify the leak If the pressure of the system does not increase during the 1 minute, then the leak check has passed. 5) Click on the High Vacuum button. Wait till the system is evacuated to 25 militorr. 6) Click on the all off button 25

26 7) Open the needle valve to the cylinder 8) Slowly open the regulator valve till the pressure on the VWR Traceable pressure meter is about psi 9) Open the valve to the clean evacuated cylinder. 10) Close the needle valve to the cylinder when the pressure on the VWR Traceable pressure meter is about psi. 11) Record the pressure reading on the VWR Traceable pressure meter. This is the initial pressure reading. 12) Click on the diluent button till the pressure on the VWR Traceable pressure meter is about ) Click on the all off button. (Tip: Click on the all off button when the VWR Traceable pressure meter is at 20 psi. The pressure will increase a few psi after the all off button has been clicked.) 14) Record the pressure reading on the VWR Traceable pressure meter. This is the final pressure reading 15) Allow the system 15 minutes to equilibrate before closing the valve to the canister. 16) Close the valve to the canister. 17) Convert the pressure readings from the VWR Traceable pressure meter to absolute pressure. 18) The concentration in the Canister can be calculated by the following equation. Initial Pr essure Re ading( psia) Concentrat ion = * OriginalCylinderConcentration Final Pr essure Re ading( psia) 26

27 Internal Standard Preparation The Internal standard does not need any dilution, however the standard must transferred from the cylinder into a six liter canister. A diagram of how the cylinder, canister and canister cleaner should be connected is shown in the figure below: 1) The cylinder valve on the cylinder should be open the regulator valve should be closed. The needle valve to the cylinder should be closed. 2) Connect the cylinder, and the canister to the tee as show in figure 1. Connect the tee to the canister cleaner system. 3) Perform a manual leak check of the system. Click on the rough pump button wait till the pressure of the system is approximately 0.1 psia. 4) Click on the all off button. The system should be able to maintain a vacuum for 1 minute. If the pressure increases by 0.1 psi during this time then there is a leak in the system. If there is a leak in the system, trouble shooting will need to be done to identify the leak If the pressure of the system does not increase during the 1 minute, then the leak check has passed. 5) Click on the High Vacuum button. Wait till the system is evacuated to 25 militorr. 6) Click on the all off button 7) Close the needle valve to the canister cleaner. 8) Open the needle valve to the cylinder. 9) Open the regulator valve slowly till the line pressure is 35 psig 10) Open the valve to the evacuated canister. Wait till the line pressure returns to 35 psig. 11) Close the valve to the canister. 12) Close the needle to the cylinder. 13) Close the regulator valve. 27

28 Standard Operation Procedure For VOC Sample Collection The VOC sample collection system is designed to collect four three hour samples. The four samples consist of a morning, noon, afternoon, and night time sample. The sampling times are 5:00 8:00, 11:00 14:00, 17: :00 and 23:00 2:00. A schematic diagram of the VOC sample collection system is shown in the figure below. The housing for the sample collection system is ventilated so that the temperature inside the housing remains cool. A small pump pumps air from outside the housing into the sample system tubing. The canisters are evacuated to 50 militorr. When the solenoid valve is opened the air passes through a flow restrictor which is an orifice that restricts the flow rate into the can. The flow controller which is a diaphragm below the flow restrictor, regulates the flow into the can so that the flow of air is at a constant rate. Without the flow controller the flow rate into the can would decrease as the pressure in the can approaches ambient pressure and majority of the volume collected in the can would be over representative of the 1 st hour of collection. The canister timers control the opening and closing of the solenoid valve. The main timer is used to set the day of which the samples are to be collected. Figure: Schematic Diagram of Sample Collection System. Below are the steps for the operation of the VOC sample collection system: 1. Press the On/Off button of the main timer to turn on power to the canister timers. 2. Turn on the canister timer knob (counter clockwise one click) and record the pressure gauge reading on the back of the tag of the canister. Mark as Pf for final pressure in the can 28

29 3. Turn off the canister timer knob (counter clockwise one click). and record the pressure gauge reading on the back of the tag of the canister. Mark as Pg for gauge pressure when the solenoid valve is closed. 4. If 0 < (Pf Pg ) <5 then the timer setting is good. If (Pf Pg ) = 0 then decrease the sample time on the canister timer by 15minutes. Note: each notch on the timer is about 15 minutes. If (Pf-Pg) >5 then increase the sample time on the canister timer by 15 minutes. Repeat steps 3-6 for each canister. 5. Press the On/Off button of the main timer to turn off power to the canister timers 6. Make sure all canister timers are set at 10pm. 7. Close the canister valves on all the canisters. 8. Disconnect the cans from the sampling let the samplers rest on top of the can before moving the cans out of the housing. This is to prevent the sample system tubing from snapping off. Remove the cans from the housing 9. Now that the housing is empty it is a good time to stick your head in and change the time on the main timer. a) Check the time on the main timer and maker sure that it is displaying the current time and day. b) press the Program button to see program 1 on screen make sure that the program on time for program 1 is 10pm. Change the Day to today s date by pressing the day button. c) Press the on/off button to see the off time for program 1. Make sure the that the program on time for program 1 is 10pm. Change the day to tomorrow s date by pressing the day button. d) Press the time select button when finished changing program 1. e) Make sure that the main timer is in the auto mode. If the words Automode are not shown on the display then press the mode button. 10. Place the new cans in the housing and put samplers on top of the canisters before connecting them (again to avoid the sample line tubing from snapping off). 11. Connect the cans to the sampling system. Make sure the connections are tight. It is recommend tightening the nut with your fingers first to avoid striping the threads of the nut. Then tighten with the wrench. 12. Put the labels on the cans. 13. Open the valves on the cans. 29

30 References Compendium of Methods for the Determination of Toxic Organic Compounds in Ambient Air Second Edition Compendium Method TO-15 Determination Of Volatile Organic Compounds (VOCs) In Air Collected In Specially-Prepared Canisters And Analyzed By Gas Chromatography/Mass Spectrometry (GC/MS) Center for Environmental Research Information Office of Research and Development, U.S. Environmental Protection Agency, Cincinnati, OH 45268, January Methods of Air Sampling and Analysis, Third Edition, James P. Lodge Jr., c1989, Catalog# TD, 890, M488,

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008 UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method

More information

GCMSD-Headspace Analysis SOP

GCMSD-Headspace Analysis SOP Before you start GCMSD-Headspace Analysis SOP Method and Sequences names are restricted to 50 characters (MSD program will crash otherwise) The GC oven and Injection Ports need to be cooled to 50 oc to

More information

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Quantitative Analysis of Hydrocarbons by Gas Chromatography Quantitative Analysis of Hydrocarbons by Gas Chromatography Introduction Gas-liquid chromatography (GLC) accomplishes a separation by partitioning solutes between a mobile gas phase and a stationary liquid

More information

Experiment GC : Analysis of BTEX by GC-FID

Experiment GC : Analysis of BTEX by GC-FID 46 Experiment GC : Analysis of BTEX by GC-FID Learning Goals: Familiarity with gas chromatography Gain experience in temperature programming and method development Correctly use an internal standard, and

More information

FROG-4000 TM Chemical Analysis System

FROG-4000 TM Chemical Analysis System FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies, Inc. 2015 Vol.5 This page is intentionally left blank. FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies,

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 14 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCE AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Mass Spec will not Autotune

Mass Spec will not Autotune Mass Spec will not Autotune Applies to 5973A/N MSD What could be the problem? There could be several things that would cause your Mass Spec not to Autotune. The most common, easily corrected Autotune problems

More information

5890II GC Standard Operating Procedure 9/2/2005

5890II GC Standard Operating Procedure 9/2/2005 5890II GC Standard Operating Procedure 9/2/2005 This procedure is for the analysis of natural gas using the HP 5890II GC with the Restek ShinCarbon column. Detector Type: TCD Column Type: Restek Micropacked,

More information

FROG-4000 TM Chemical Analysis System

FROG-4000 TM Chemical Analysis System FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies, Inc. 2014 Vol.4 This page is intentionally left blank. FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies,

More information

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note Gas Chromatography Author Matthew S. Klee Agilent Technologies 2850 Centerville Road Wilmington, DE 19808-1610 USA Abstract

More information

Multiple Gas#5 GC configuration Jan 2016

Multiple Gas#5 GC configuration Jan 2016 History: Unfortunately there is no single column that can separate: Hydrogen Oxygen Nitrogen Methane CO CO2 Ethane Water Propane Butane Pentane Over the years SRI Instruments has devised several solutions

More information

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion: Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: The RSK-175 standard operating procedure was developed in order to determine the amount of dissolved gas in water. Due to the expansion

More information

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY Nengbing Xu, Hongmei Ying and Libo Zhu Ningbo Environmental Monitoring

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 9 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 3.1 Canister Receipt 3.2 Canister Storage 4.0 INTERFERENCE AND POTENTIAL PROBLEMS

More information

FROG-5000TM Chemical Analysis System User s Manual

FROG-5000TM Chemical Analysis System User s Manual FROG-5000TM Chemical Analysis System User s Manual Defiant Technologies, Inc. 2018 Rev.3 How to assemble a ferrule. This problem has arisen so often and caused so many problems, that we have added it as

More information

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor Anwar Sutan, Metco Services Ltd. Charles Johnson, Metco Services Ltd. Jason Laidlaw,

More information

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES 1. Applicability and Principle 1.1 Applicability. This method applies to the analysis of carbon dioxide

More information

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Jamie Brown R&D Scientist Supelco 595 North Harrison RD. Bellefonte, PA 16823 National

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results. Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions

More information

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere! The portable Gas Chromatograph is now more versatile than ever. DPS Instruments is pleased to present the newest Portable Companion 2 Gas Chromatograph with room for 2 Detectors. The Companion 2 GC was

More information

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler Optimizing Vial Pressurization Parameters for the Analysis of Residual Solvents Using the 7697A Headspace Sampler Application Note Pharmaceuticals Author Roger L Firor Agilent Technologies, Inc.

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs. Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography

More information

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE. Gases&Technology FEATURE Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. B Y M A T T H E W M O N A G L E Abstract Bulk gases are often delivered to

More information

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION BEST KNOWN METHODS Transpector XPR3 Gas Analysis System DESCRIPTION The Transpector XPR3 is a third-generation, quadrupole-based residual gas analyzer that operates at PVD process pressures and is the

More information

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly

More information

Rapid and Reliable Detection of Dissolved Gases in Water

Rapid and Reliable Detection of Dissolved Gases in Water Rapid and Reliable Detection of Dissolved Gases in Water Andrea Caruso and Massimo Santoro Thermo Fisher Scientific, Milan, Italy Application Note 005 Key Words Chromeleon CDS, Environmental, Fracking,

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 14 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 5.1 Subatmospheric

More information

Best Practice for Identifying Leaks in GC and GC/MS Systems

Best Practice for Identifying Leaks in GC and GC/MS Systems Best Practice for Identifying Leaks in GC and GC/MS Systems Technical Overview Maintaining a leak-free GC or GC/MS system is critical for obtaining optimal system performance with reliable, reproducible,

More information

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection This user guide does not have the intention to introduce you to any of the theory and/or principles behind

More information

Operation of the Perkin Elmer TGA-GC/MS

Operation of the Perkin Elmer TGA-GC/MS Operation of the Perkin Elmer TGA-GC/MS Summary of the TGA-GC/MS: The TGA-GC/MS allows a user to decompose a sample by heating, measure its loss of mass, and simultaneously analyze the chemical composition

More information

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Cover Page for Lab Report Group Portion. Head Losses in Pipes Cover Page for Lab Report Group Portion Head Losses in Pipes Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 February 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section

More information

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation User s Booklet for the Wyatt minidawn Light Scattering Instrumentation The Wyatt minidawn Light Scattering instrument is able to measure the weight average molecular weight of a synthetic polymer or a

More information

System Overview. TCD Detector temperature setpoint Regulator

System Overview. TCD Detector temperature setpoint Regulator System Overview POPULAR CONFIGURATION GCs Your educational TCD GC is configured on the compact 310 chassis. It is equipped with a TCD Detector, a temperature programmable Column Oven, a 3 Silica Gel packed

More information

3 Flow and Pressure Control

3 Flow and Pressure Control 3 Hydrogen shutdown Column shutdown Turning gas flows on and off EPC-controlled streams NonEPC-controlled streams Electronic Pneumatic Control (EPC) Interpreting flow and pressure readings Configuration

More information

Carrier Gases in Capillary GC

Carrier Gases in Capillary GC Carrier Gases in Capillary GC LC Columns and Consumables Mark Sinnott Application Engineer January 15, 2009 CARRIER GAS Mobile Phase Carries the solutes down the column Selection and velocity influences

More information

no peaks missing peaks too small peaks increasing retention decreasing retention

no peaks missing peaks too small peaks increasing retention decreasing retention GC Troubleshooting We can help you. 1. Observation: no peaks no peaks missing peaks too small peaks increasing retention decreasing retention declining baseline rising baseline bleeding plateaus interfering

More information

Chromat Calibration Updated October 27th, 2017

Chromat Calibration Updated October 27th, 2017 Chromat Calibration Updated October 27th, 2017 Calibrating the Chromatograph Performing the Chromat Calibration is highly recommended when starting a new well. The MLogger already has a default calibration

More information

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS) Toxicology Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose - This procedure specifies the required elements for the calibration and use of the Agilent Gas Chromatograph interfaced to the Agilent

More information

Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS)

Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS) Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS) 1.0 Purpose - This procedure specifies the required elements for the calibration and use of the Agilent 6890 GC interfaced

More information

Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions

Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions Introduction: There is a bit of explanation of the health effects of benzene in the Introduction section of the first part

More information

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer application Note Gas Chromatography/ Mass Spectrometry Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer Overview A mass spectrometer

More information

Retention Time Locking: Concepts and Applications. Application

Retention Time Locking: Concepts and Applications. Application Retention Time Locking: Concepts and Applications Application Gas Chromatography December 1997 Authors Vince Giarrocco Bruce Quimby Matthew Klee Agilent Technologies, Inc. 2850 Centerville Road Wilmington,

More information

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Oneida Research Services, Inc. 8282 Halsey Road Whitesboro, NY 13492 Phone: (315) 736-5480 FAX: (315) 736-9321 1.0 Purpose The

More information

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more! After years of development and testing, DPS Instruments is pleased to present the newest, most expandable and versatile Gas Chromatography Systems in history. The DPS 600 Series GC systems are the world

More information

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager Ghost Peaks: How to Fix a Haunting Problem Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager Outline of Topics What is a ghost peak? How to identify ghost peaks and their source How to correct

More information

SOP: Buck Scientific BLC-20P HPLC Operation

SOP: Buck Scientific BLC-20P HPLC Operation Page 1 of 11 Approvals Preparer: John Buford Reviewer: Tim Kull Reviewer: Dr. Margaret Bryans Date: 16OCT13 Date: 30OCT13 Date: 31OCT13 1. Purpose 1.1. Basic operation of the Buck Scientific BLC-20P isocratic

More information

Gas Diluter Portable Gas Dilution System User s Manual

Gas Diluter Portable Gas Dilution System User s Manual Gas Diluter Portable Gas Dilution System User s Manual Defiant Technologies, Inc. 2015 Vol.1 This page is intentionally left blank. Gas Diluter Portable Gas Dilution System User s Manual Defiant Technologies,

More information

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD) Procedure No: SOP-026 Revision No: 1.0 January 24, 2011 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to meet

More information

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (16) 17 R QUALIFICATION OF EQUIPMENT ANNEX 2: QUALIFICATION OF GC EQUIPMENT Full document title and reference Document type Legislative

More information

Agilent 490 Micro GC Natural Gas Analyzer

Agilent 490 Micro GC Natural Gas Analyzer Agilent 490 Micro GC Natural Gas Analyzer User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by any means (including electronic

More information

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12 LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/ This experiment will introduce you to the kinetic properties of low-pressure gases. You will make observations on the rates with which selected

More information

This GC has two columns (I think new columns can be ordered from Restek)

This GC has two columns (I think new columns can be ordered from Restek) SRI 8610C GC Ann s Notes 11-17-2014 Introduction This is the GC as it came from the manufacturer This GC has two columns (I think new columns can be ordered from Restek) 1. Hayesep D: Column 1 a. Separating

More information

FLUORESCENCE DETERMINATION OF OXYGEN

FLUORESCENCE DETERMINATION OF OXYGEN FLUORESCENCE DETERMINATION OF OXYGEN Objectives This experiment will familiarize the students with the principles of fluorescence quenching as well as the Ocean Optics FOXY Fiber Optic Sensor System. Spectra

More information

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane Gas requirements Gases for packed columns The carrier gas you use depends upon the type of detector and the performance requirements. Table 520-1 lists gas recommendations for packed column use. In general,

More information

AKTA pure 25 New Owner s Intro

AKTA pure 25 New Owner s Intro AKTA pure 25 New Owner s Intro The exercise below will give a quick demonstration of how easy and intuitive the AKTA pure 25 will be for you in demonstrating downstream processing to your students. Steps

More information

The instrument should have a mass range from 2 to 2000 m/z or better Interface

The instrument should have a mass range from 2 to 2000 m/z or better Interface Revised Specifications for Liquid Chromatography Mass Spectrometry (LC-MS/MS ) Triple Quadrupole Specification for Ultra-Fast Liquid Chromatography system Solvent Delivery 1) A quaternary Pump system for

More information

1)! DO NOT PROCEED BEYOND THIS MARK

1)! DO NOT PROCEED BEYOND THIS MARK Operating Instructions for X-ray Photoelectron Spectrometer: Physical Electronics Model 555 XPS/AES (John H. Thomas, III, Ph.D., Electron Spectroscopy) Sample Insertion: figure 1. Sample insertion rod

More information

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013 The SRI ASTM 3612/TOGA/ Dissolved Gas GC configuration ( TOGA GC ) permits measurement of gasses dissolved in water, oil or other liquids, and is based on the requirements of ASTM method 3612, option C,

More information

6890 GC Site Preparation

6890 GC Site Preparation 6890 GC Site Preparation (a16011) This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this material. The information

More information

Calibration of a GC955

Calibration of a GC955 Calibration of a GC955 1 Contents of the power point instructions Standard calibration Automatic validation methods and calibration of multi component mixes Autolinearisation: test of the linearity of

More information

Automated Determination of Dissolved Gases in Water Anne Jurek

Automated Determination of Dissolved Gases in Water Anne Jurek Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: Tapping the natural gas reservoirs throughout the United States has long been a viable solution for energy independence; however

More information

Figure 1: Plug in ion inlet of mass spec

Figure 1: Plug in ion inlet of mass spec Starting procedure: by Vicky - is the vacuum pump running (should run all time, otherwise it ll need 4-5 hours to have a adequate vacuum for the MS) otherwise something has gone wrong- F.T - Remove plug

More information

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator Designed, built, and supported by InstruQuest Inc. Universal pycnometer, no

More information

Generating Calibration Gas Standards

Generating Calibration Gas Standards Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and

More information

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1 Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost HERMAN HOLM SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 4C CONFERENCE APRIL 2018 SAN ANTONIO, TX SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC

More information

Carrier Gases in Capillary GC

Carrier Gases in Capillary GC Carrier Gases in Capillary GC GC Columns and Consumables Abby Folk Application Scientist January 15, 2008 CARRIER GAS Mobile Phase Carries the solutes down the column Selection and velocity influences

More information

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions Varian 3300 Gas Chromatograph Abbreviated Operating Instructions 1. Introduction and Startup 2 2. Basic Configurations for the GC 3 3. Sample Preparation 3 4. Building a Method 3 5. Split/Splitless Injection

More information

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek Abstract: Due to the expansion of natural gas drilling through horizontal fracturing, there has been

More information

Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity

Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity Rev. B 2017,Janyary,20 Next > Preconditions 7900, 7800, 7700, or 8800 ICP-MS is used. Sample introduction type is PeriPump. If you are

More information

Markes Thermal Desorption Systems Site Preparation Page 1 of 8

Markes Thermal Desorption Systems Site Preparation Page 1 of 8 Thermal Desorption Systems Site Preparation Purpose of site preparation Your site must meet this specification to assure a successful and timely installation of your Agilent thermal desorption equipment.

More information

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation CHE 4115 Chemical Processes Laboratory 2 Experiment 1 Batch Distillation BACKGROUND Distillation is one of the most commonly used unit operations in chemical engineering. In general, a distillation operation

More information

May 1, By John Heim,Doug Staples

May 1, By John Heim,Doug Staples 1 of 7 7/11/2014 3:39 PM May 1, 2010 By John Heim,Doug Staples Anabolic steroid screening analysis in urine is complex and labor intensive requiring sensitive instrumentation and optimized chromatographic

More information

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator Designed, built, and supported by InstruQuest Inc. Temperature controlled, multi-technique volumetric

More information

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic. 458 LCGC NORTH AMERICA VOLUME 24 NUMBER 5 MAY 2006 www.chromatographyonline.com LC TROUBLESHOOTING Dwell Volume Revisited Gradient transfer does not have to be problematic. John W. Dolan LC Troubleshooting

More information

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial Supporting Information for Micro-Collection of Gases in a Capillary Tube: Preservation of Spatial and Temporal Resolution AUTHOR INFORMATION Kristin A. Herrmann Favela, 1 * Pieter Tans, 2 Thomas Jaeckle,

More information

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems Inorganic Analysis Chromatography Molecular Spectroscopy Thermal/Elemental Analysis Informatics Gas Chromatography MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass

More information

ISO 7941 First edition

ISO 7941 First edition INTERNATIONAL STANDARD ISO 7941 First edition 1988-08-01 INTERNATIONAL ORGANIZATION FOR STANDARDIZATION ORGANISATION INTERNATIONALE DE NORMALISATION MEXaYHAPOAHAR OPrAHM3A4MR fl0 CTAH~APTM3A~MM Commercial

More information

Manual for continuous distillation

Manual for continuous distillation Manual for continuous distillation 1. Week 1: Objectives: Run the column at total reflux. When steady state is reached, take the sample from the top and bottom of the column in order to determine the overall

More information

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER Michael T Tang, Ph.D. Grace Feng Greg Merideth Rui Huang Matheson Gas Applied Lab

More information

B. Assemble the Method-25 sample train with a heated probe/filter.

B. Assemble the Method-25 sample train with a heated probe/filter. 6/30/2015 VOC Reporting, Inc. Method 25 Heated Controller Operating Instructions. pg. 1 of 5 INSTRUCTIONS FOR EPA METHOD-25 FIELD SAMPLE COLLECTION USING A VRi HEATED PROBE AND FILTER ASSEMBLY Introduction:

More information

Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems

Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems The bucket brigade air sampler. For the square Tedlar bags with the push/pull valves: Put the valve of a tedlar

More information

X-Flow Mass Flow Controller 授权代理 : 北京品超思瑞科技学院公司

X-Flow Mass Flow Controller 授权代理 : 北京品超思瑞科技学院公司 X-Flow Mass Flow Controller 2 Highlights Parker Hannifin Precision Fluidics Division is excited to introduce X-Flow, a new easy to use general purpose mass flow controller for your instrument, lab, or

More information

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures

More information

EXPERIMENT XI. Careful!! Improper handling of the vacuum line may result in the release of SO 2 which is an irritating and suffocating gas.

EXPERIMENT XI. Careful!! Improper handling of the vacuum line may result in the release of SO 2 which is an irritating and suffocating gas. Chem 366-3 Page XI - 1 EXPERIMENT XI INFRARED SPECTRUM OF SO2 (S&G, 5th ed. Expt 36, 6th ed. Expt. 35) 1. Pre-Lab preparation. The description of this experiment has disappeared from the more recent editions

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310 www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Varian 310 310 LC/MS Manual System INCLUDES: 310-MS LC/MS triple quadrople

More information

CHM Gas Chromatography: Our Instrument Charles Taylor (r10)

CHM Gas Chromatography: Our Instrument Charles Taylor (r10) 1/8 Introduction You should already be familiar with the basic concepts of gas chromatography. The purpose of this note pack is to familiarize you with the specific instrument that we will use in our labs.

More information

A New GC/MS System Designed for Helium Carrier Gas Conservation

A New GC/MS System Designed for Helium Carrier Gas Conservation A New GC/MS System Designed for Helium Carrier Gas Conservation Approaches for the Smarter Use of Helium Dale R. Walker GCMS Applications Specialist August, 7, 2013 1 Helium Facts How it's produced Where

More information

Total Pressure Measurement Using Residual Gas Analyzers

Total Pressure Measurement Using Residual Gas Analyzers Total Pressure Measurement Using Residual Gas Analyzers M. Maskell, Old Dominion University; G. Myneni, P. Adderley, Jefferson Lab; G. Brucker, Stanford Research Systems; C. Day, Forschungszentrum, Karlsruhe

More information

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction A Column-Flow Independent Configuration for QuickSwap Application Authors Matthew S. Klee and Bruce Quimby Agilent Technologies, Inc. 80 Centerville Road Wilmington, DE 08 USA Abstract As can be seen from

More information

Zebron Gas Management

Zebron Gas Management Zebron www.phenomenex.com/gasmanagement What are the key benefits of Zebron filters and traps? The high adsorption capacity of the Zebron Filters removes contaminants from mobile phase gases, reducing

More information

Agilent G3185B QuickSwap Accessory

Agilent G3185B QuickSwap Accessory Agilent G3185B QuickSwap Accessory Reference Manual Agilent Technologies Notices Agilent Technologies, Inc. 2005 No part of this manual may be reproduced in any form or by any means (including electronic

More information

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler. Exercise 5-2 Bubblers EXERCISE OBJECTIVE Learn to measure the level in a vessel using a bubbler. DISCUSSION OUTLINE The Discussion of this exercise covers the following points: Bubblers How to measure

More information

During today's webinar we're going to look at the following topics:

During today's webinar we're going to look at the following topics: Alternative Carrier Gas for GC and GC-MS Presentation Transcript Hello and welcome to the Peak scientific webinar which today will concern alternative carrier gas for GC and GC-MS. My name is Ed Connor,

More information

Agilent G3180B Two-Way Splitter Kit

Agilent G3180B Two-Way Splitter Kit Agilent G3180B Two-Way Splitter Kit With Makeup Gas Installation and Operation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2006 No part of this manual may be reproduced in any form or

More information

LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary

LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary ADH 1/7/014 LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary This experiment will introduce you to the kinetic properties of low-pressure gases. You will make observations on the

More information

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the 1151 METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS 1.0 Scope and Application. 1.1 Analytes. Analyte Volatile Organic Compounds (VOC) CAS No. No CAS number assigned 1.2 Scope. This method

More information

Gas Leak Measurement of TRT Barrel Module

Gas Leak Measurement of TRT Barrel Module DUKHEP01-07-02 Gas Leak Measurement of TRT Barrel Module Shannon Watson, Seog Oh Department of Physics Duke University Durham, NC, 27708, USA 1. Introduction One of the requirements for the TRT module

More information

Boyle s Law VC 09. Experiment 9: Gas Laws. Abstract

Boyle s Law VC 09. Experiment 9: Gas Laws. Abstract Experiment 9: Gas Laws VC 09 Abstract In this laboratory activity, you will experimentally confirm Boyle s Law, determine absolute zero from Gay-Lussac's Law, and determine the molecular weight of acetone,

More information

Impact of Air Leaks on the Productivity of GC and GC/MS Systems

Impact of Air Leaks on the Productivity of GC and GC/MS Systems Impact of Air Leaks on the Productivity of GC and GC/MS Systems Application Note Environmental Author Ken Lynam Agilent Technologies, Inc. Abstract Agilent 7890A GC dual-channel FID, Agilent 7890B GC,

More information

Armfield Distillation Column Operation Guidelines

Armfield Distillation Column Operation Guidelines Armfield Distillation Column Operation Guidelines 11-2016 R.Cox Safety SAFETY GLASSES ARE REQUIRED WHEN OPERATING THE DISTILLATION COLUMN Wear gloves when mixing alcohol feedstock The column will become

More information