Gas Diluter Portable Gas Dilution System User s Manual

Size: px
Start display at page:

Download "Gas Diluter Portable Gas Dilution System User s Manual"

Transcription

1 Gas Diluter Portable Gas Dilution System User s Manual Defiant Technologies, Inc Vol.1

2 This page is intentionally left blank.

3 Gas Diluter Portable Gas Dilution System User s Manual Defiant Technologies, Inc Vol.1

4 Gas Diluter User s Manual This page is intentionally left blank. 4

5 TABLE OF CONTENTS Gas Diluter User s Manual Section 1: Getting to Know the Gas Diluter Manual Overview Operations Overview... 7 Section 2: Components Instrument Components... 8 Section 3: Instrument Setup Opening Exhaust Outlet Connecting to a Gas Cylinder Connecting to a FROG-4000 TM Selecting Valve Positions Powering On Section 4: Diluting a Gas Dilution Spreadsheet Diluting a Gas Section 5: Troubleshooting Rotameter Settings Are Not Stable

6 Gas Diluter User s Manual SECTION 1: GETTING TO KNOW THE GAS DILUTER 1.1 Manual Overview The Gas Diluter uses a practical approach to dilute gas mixtures and delivers metered amounts of VOCs to calibrate a portable GC. This manual includes the following: An overview of the Gas Diluter components and features. Detailed procedures for using the Gas Diluter. Troubleshooting information. 6

7 Gas Diluter User s Manual 1.2 Operations Overview The Gas Diluter mixes scrubbed ambient air with a calibration gas enabling the end user to perform a mutli-point calibration on a portable GC. The Gas Diluter is designed to be used with the FROG-4000 or other gas chromatographs. The basic steps for operating the system are: 1. Connecting a gas cylinder (containing a known concentration of calibration gas) to the Gas Diluter. 2. Connecting a FROG to the Gas Diluter. 3. Calculating the gas settings based on the desired concentrations for the calibration. 4. Diluting the gas. 7

8 SECTION 2: COMPENENTS Gas Diluter User s Manual 2.1. Instrument Components Components Reference for Gas Diluter A. Dilution Flow Rotameter B. Calibration Gas Mix Flow Rotameter A B 8

9 Gas Diluter User s Manual 2.1: Instrument Components Components Reference for Gas Diluter C. Pump Power Button D. Power Supply Port E. 1/4 inch Ambient Air Inlet F. 1/4 inch Nitrogen Inlet G. 2 Way Select Valve for Dilution Gas Allows selection of ambient air or nitrogen as the dilution gas H. 1/4 inch Gas Mix Inlet I. 2 Way Select Valve for Mode Allows for selection of Flush Mode or Dilution Flow J. 1/4 inch Exhaust K. Diluted Calibration Gas Outlet (1/8 inch outlet with a 1/8 to 1/16 adapter) E F G D C K J I H 9

10 SECTION 3: INSTRUMENT SETUP Gas Diluter User s Manual There are several procedures necessary to setup the Gas Diluter for proper use. The following procedures prepare the instrument for use: : Opening Exhaust Outlet : Connecting to a Gas Cylinder : Connecting to a FROG-4000 TM : Selecting Valve Positions : Powering On 10

11 Section 3: Instrument Setup Gas Diluter User s Manual 3.1: Opening Exhaust Outlet Remove cap from exhaust outlet O O Figure 3.1-1: Removing Exhaust Cap Figure 3.1-2: Exhaust Cap Removed 11

12 Gas Diluter User s Manual Section 3: Instrument Setup 3.2: Connecting to a Gas Cylinder Remove cap from gas mix inlet. Figure 3.2-1: Removing Gas Mix Inlet Cap N Q Figure 3.2-2: Gas Mix Inlet Cap Removed Obtain a gas cylinder containing known concentrations of analytes to the gas diluter. Figure 3.2-3: Pressurized Gas Cylinder 12

13 Gas Diluter User s Manual 3.2 Connecting to a Gas Cylinder Use proper fittings to connect the gas cylinder to the gas diluter via the Gas Mix Inlet. Figure 3.2-4: Connecting the gas cylinder via the Gas Mix Inlet Figure 3.2-5: Gas Diluter with gas cylinder connected 13

14 Gas Diluter User s Manual Section 3: Instrument Setup 3.3: Connecting to a FROG-4000 TM The Gas Diluter is designed to easily connect to the FROG-4000 TM. However, it is possible to use the Gas Diluter with other gas chromatography systems. The Instructions that follow are for connecting the Gas Diluter to a FROG-4000 TM. Obtain a FROG with an air sampler connected. For more detail about air sampling consult the FROG-4000 User s Manual. Connect the air sampler transfer line that is included with the FROG-4000 s air sampler kit to the Calibration Gas Outlet. Figure 3.3-1: Attaching transfer line to Calibration Gas Outlet. Figure 3.3-2: Transfer line attached to Calibration Gas Outlet. 14

15 Gas Diluter User s Manual 3.3. Connecting to a FROG-4000 Make sure the transfer line is also attached to the FROG s air sampler box. Figure 3.3-3: Transfer line connecting FROG-4000 and Gas Diluter. 15

16 Gas Diluter User s Manual Section 3: Instrument Setup 3.4: Selecting Valve Positions The 2 Way Select Valves direct the flow of gas in the Gas Diluter. Having these valves in the correct position is important for successful use of the Gas Diluter. I G Figure 3.4-1: 2 Way Select Valves G. 2 Way Select Valve for Dilution Gas: Allows selection of ambient air or nitrogen as the dilution gas If using air as dilution gas have valve pointed towards the A (Figure 3.4-2). Figure 3.4-2: Valve position for air as dilution gas If using nitrogen as dilution gas (attached to the N2 position) have valve pointed away from the A (Figure 3.4-3). Figure 3.4-3: Valve position for nitrogen as dilution gas 16

17 Gas Diluter User s Manual 3.4. Selecting Valve Positions I. 2 Way Select Valve for Mode: Allows for selection of Flush Mode or Dilution Flow If using gas diluter for dilution of a gas mix, valve should point toward the A (Figure 3.4-4). Figure 3.4-4: Valve position for gas dilution If flushing the lines of the Gas Diluter to clean it, the valve should point away from the A (Figure 3.4-5). Figure 3.4-5: Valve position for air as dilution gas 17

18 Gas Diluter User s Manual Section 3: Instrument Setup 3.5: Powering On Connect power supply to outlet and to Gas Diluter. To turn on Gas Diluter, press power button once (Figure 3.5-2). To turn off Gas Diluter, press power button once. To Outlet To Gas Diluter Figure 3.5-1: Power Supply for Gas Diluter. Figure 3.5-2: Plugged in power supply, power button. 18

19 Gas Diluter User s Manual 3.4 Powering On The Gas Diluter is now setup and ready to use. The figure below shows what the Gas Diluter should look like at this stage (Figure 3.5-3). Figure 3.5-3: Gas Diluter ready to use. 19

20 SECTION 4: DILUTING A GAS Gas Diluter User s Manual 4.1: Dilution Spreadsheet Included with the Gas Diluter is a spreadsheet containing calibration data. This calibration data is important for calculating the final concentration of the gas exiting the Gas Diluter. The first time the Gas Diluter is used the pressure at the location of use needs to be set. On the settings and concentrations spreadsheet enter the pressure at the location the Gas Diluter is being used in mmhg (Figure 4.1-1). Pressure at Location of Use (mmhg): 740 Figure 4.1-1: Pressure Input on Settings and Concentrations Spreadsheet To calculate the final concentration of the calibration gas, adjust the settings of the gas mix flow and the dilution flow in the settings and concentrations spreadsheet (Figure 4.1-2). In the example in Figure the bottle concentration (gas mix) is 1000ppb. To dilute the gas mix to 100ppb, the dilution gas setting is 15 and the gas mix setting is 100. This calculation is shown below. Note that the flow rates have already been calculated and are included in the spreadsheet. Also note that the bottle concentration should be changed to the concentration of the gas mix entering the gas diluter. 20

21 Dilution Gas Gas Mix Setting Flowrate (ml/min) Setting Flowrate (ml/min) Bottle Conc (ppbv) Final Conc (ppbv) Figure 4.1-2: Example Table of Settings and Concentrations. Gas Diluter User s Manual 21

22 Gas Diluter User s Manual 4.1 Dilution Spreashsheet For ease of use, the settings columns can be edited in the spreadsheet to produce different concentrations. For example, if 150 ppb is desired rather than 100ppb, the dilution setting can be adjusted to 10 and the gas mix setting can be adjusted to 110 (Figure 4.1-3). The Columns in Figure and Figure that are highlighted in blue (Dilution Gas Setting, Gas Mix Setting and Bottle Concentration) are the columns to edit to adjust the settings and to account for the bottle concentration. The Final Concentration column, highlighted in yellow will provide the final concentration based on the settings and the bottle concentration. 22

23 Dilution Gas Gas Mix Setting Flowrate (ml/min) Setting Flowrate (ml/min) Bottle Conc (ppbv) Final Conc (ppbv) Figure 4.1-3: Example Table of Settings and Concentrations. Gas Diluter User s Manual 23

24 4 Diluting a Gas Gas Diluter User s Manual 4.2: Diluting a Gas This section describes the procedure for using the gas diluter to dilute a gas mix to a desired concentration for calibration. 1. After determining which concentrations are desired use the spreadsheet to determine what gas mix and calibration flow settings correspond to the desired concentrations. 2. Once the settings are known, turn the knobs at the bottoms of both the gas mix flow rotameter and the dilution flow rotameter to reach the desired settings (Figure 4.2-1). Figure 4.2-1: Adjusting the Gas Mix Flow Rotameter. 24

25 Gas Diluter User s Manual 4.2 Diluting a Gas 3. It may take several minutes for the rotameter to stabilize at a setting. It is recommended that the user wait at least 90 seconds after obtaining a stable reading before collecting the calibration gas for samples. 4. After 90 seconds has elapsed since the rotameter settings have stabilized, the FROG-4000 can be used to take a sample. Please refer to the FROG-4000 s User s Manual for more detail on starting an air sample. 5. Repeat steps 2-4 until all samples have been taken. 25

26 Gas Diluter User s Manual SECTION 5: TROUBLE SHOOTING 5.1: Rotameter settings are not stable The rotameters are quite sensitive and take time to stabilize. However if after several minutes the setting is not stable it may be due to the pressure of the gas mix. It is recommended that the gas mix entering the gas diluter does not exceed 25psi. 26

27 Gas Diluter User s Manual Notes Scrubber 27

28 Notes Gas Diluter User s Manual 28

29 Notes Gas Diluter User s Manual 29

30 Notes Gas Diluter User s Manual 30

31

32

FROG-4000 TM Chemical Analysis System

FROG-4000 TM Chemical Analysis System FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies, Inc. 2015 Vol.5 This page is intentionally left blank. FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies,

More information

FROG-4000 TM Chemical Analysis System

FROG-4000 TM Chemical Analysis System FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies, Inc. 2014 Vol.4 This page is intentionally left blank. FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies,

More information

FROG-5000TM Chemical Analysis System User s Manual

FROG-5000TM Chemical Analysis System User s Manual FROG-5000TM Chemical Analysis System User s Manual Defiant Technologies, Inc. 2018 Rev.3 How to assemble a ferrule. This problem has arisen so often and caused so many problems, that we have added it as

More information

Calibration of a GC955

Calibration of a GC955 Calibration of a GC955 1 Contents of the power point instructions Standard calibration Automatic validation methods and calibration of multi component mixes Autolinearisation: test of the linearity of

More information

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008 UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method

More information

5000TOC Sensor Standard Operating Procedure for TOC Calibration Conductivity Calibration Temperature Calibration Flow Rate Calibration

5000TOC Sensor Standard Operating Procedure for TOC Calibration Conductivity Calibration Temperature Calibration Flow Rate Calibration Part No. 84468 5000TOC Sensor Standard Operating Procedure for TOC Calibration Conductivity Calibration Temperature Calibration Flow Rate Calibration This document contains proprietary information, which

More information

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Jamie Brown R&D Scientist Supelco 595 North Harrison RD. Bellefonte, PA 16823 National

More information

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator Procedure No: SOP-034 Revision No: 1.0 Revised Dec. 29, 2010 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE This procedure is intended to describe the operations of the Dasibi model 5008 calibrator. The Dasibi

More information

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...

More information

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly

More information

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler. Exercise 5-2 Bubblers EXERCISE OBJECTIVE Learn to measure the level in a vessel using a bubbler. DISCUSSION OUTLINE The Discussion of this exercise covers the following points: Bubblers How to measure

More information

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE. Gases&Technology FEATURE Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. B Y M A T T H E W M O N A G L E Abstract Bulk gases are often delivered to

More information

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions Varian 3300 Gas Chromatograph Abbreviated Operating Instructions 1. Introduction and Startup 2 2. Basic Configurations for the GC 3 3. Sample Preparation 3 4. Building a Method 3 5. Split/Splitless Injection

More information

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator Procedure No: SOP-029 Revision No: 1.1 (December 29, 2010) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting, air quality

More information

21700 Nordhoff St. Chatsworth, CA Phone

21700 Nordhoff St. Chatsworth, CA Phone 21700 Nordhoff St. Chatsworth, CA. 91311 Phone 818.882.2331 DILUTION SYSTEM DILUTION SYSTEM ~ 1 ~ 09 21 07 TABLE OF CONTENTS DEFINITIONS.. 3 INTRODUCTION.. 4 INSTALLATION & SETUP 5 CONVERSION TO PRESSURE

More information

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD) Procedure No: SOP-026 Revision No: 1.0 January 24, 2011 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to meet

More information

Generating Calibration Gas Standards

Generating Calibration Gas Standards Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and

More information

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300 Technical Note GA-300 Overview The GA-300 CO2 and O2 Gas Analyzer is easy to use, robust, and adaptable to human, animal, and plant applications. The GA-300 has two analog outputs to allow recording and

More information

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings SomnoSuite FAQ V1.3 January 2015 Setup 1. How do I connect the SomnoSuite to my oxygen source? 2. Is there a way to speed up the downward movement of the pusher block when setting the empty position? 3.

More information

Armfield Distillation Column Operation Guidelines

Armfield Distillation Column Operation Guidelines Armfield Distillation Column Operation Guidelines 11-2016 R.Cox Safety SAFETY GLASSES ARE REQUIRED WHEN OPERATING THE DISTILLATION COLUMN Wear gloves when mixing alcohol feedstock The column will become

More information

Exercise 2-2. Second-Order Interacting Processes EXERCISE OBJECTIVE DISCUSSION OUTLINE. The actual setup DISCUSSION

Exercise 2-2. Second-Order Interacting Processes EXERCISE OBJECTIVE DISCUSSION OUTLINE. The actual setup DISCUSSION Exercise 2-2 Second-Order Interacting Processes EXERCISE OBJECTIVE Familiarize yourself with second-order interacting processes and experiment with the finer points of controller tuning to gain a deeper

More information

Sampling Equipment Catalog. MicroPurge Low-Flow Pump Control

Sampling Equipment Catalog. MicroPurge Low-Flow Pump Control MicroPurge Low-Flow Pump Control Sampling Equipment Catalog 10 MicroPurge Controls The MicroPurge Controller (U.S. Patent Number 6,508,310) revolutionizes low-flow sampling with advanced logic control

More information

Hazards associated with the gas system and how to mitigate them

Hazards associated with the gas system and how to mitigate them Hazards associated with the gas system and how to mitigate them Do not enter the gas shack if the rotating red light at the entrance is on. This warns of an oxygen deficiency. In General the Gas Mixing

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 14 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 5.1 Subatmospheric

More information

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS 2000 Dionex Corporation Document No. 031277 Revision 03 April 2000 2000 Dionex Corporation All rights reserved worldwide. Printed in the United States of

More information

Agilent Inlet EPC Leak Test

Agilent Inlet EPC Leak Test Agilent Inlet EPC Leak Test Gas Chromatographs Instructions Agilent Technologies Notices Agilent Technologies, Inc. 2002 No part of this manual may be reproduced in any form or by any means (including

More information

Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272)

Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272) Manual Addendum Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272) P/N M07273 DATE 06/11/13 TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments 16830 Chestnut Street City of Industry,

More information

300 Held Drive Northampton Pa

300 Held Drive Northampton Pa 300 Held Drive Northampton Pa 18067 610.262.6090 June 10, 2015 Recommended testing for verifying device performance. Intervals of testing are to be set by the Healthcare Provider / Dealer. Table of Contents

More information

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2. STANDARD OPERATING PROCEDURES Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.5 ) Sampler (RAAS) AMBIENT AIR MONITORING PROGRAM for the 130 LIBERTY STREET DECONSTRUCTION

More information

SHIMADZU LC-10/20 PUMP

SHIMADZU LC-10/20 PUMP SHIMADZU LC-10/20 PUMP Clarity Control Module ENG Code/Rev.: M091/70C Date: 24.10.2017 Phone: +420 251 013 400 DataApex Ltd. Fax: +420 251 013 401 Petrzilkova 2583/13 clarity@dataapex.com 158 00 Prague

More information

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion: Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: The RSK-175 standard operating procedure was developed in order to determine the amount of dissolved gas in water. Due to the expansion

More information

HIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION

HIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION HIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION SCENTROID is the global leader in sensory technology providing intelligent environmental solutions across a wide range of industries

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 13 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION

Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION Install your Blue River Technologies Port-A-Poly mixing system in a clean dry area.

More information

RK-IR Sample Draw Aspirator Adapter Operator s Manual

RK-IR Sample Draw Aspirator Adapter Operator s Manual 30-0951RK-IR Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0018RK Revision: A Released: 6/2/10 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment

More information

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the 1151 METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS 1.0 Scope and Application. 1.1 Analytes. Analyte Volatile Organic Compounds (VOC) CAS No. No CAS number assigned 1.2 Scope. This method

More information

Exhibit 4. Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N)

Exhibit 4. Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N) 1 APPLICABILITY Exhibit 4 Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N) Definitions common to all certification and test procedures

More information

CircuFlow Quick Setup Guide

CircuFlow Quick Setup Guide CircuFlow 5200 Quick Setup Guide LB04.0001 Rev B 20140314 TABLE OF CONTENTS Setup Therapy Setup Gradient Mode Pressure Mode Lock Device Unlock Device Treatment 03 04 05 08 11 12 13 02 SETUP 01 01 Open

More information

NMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013

NMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013 NMISA NEW GAS FLOW CALIBRATION FACILITY Deona Jonker 7 October 2013 Introduction Flow the total volume of a fluid that flows past a fixed point in a given time Flow measurements used in industry as a means

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs. Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography

More information

Manual for continuous distillation

Manual for continuous distillation Manual for continuous distillation 1. Week 1: Objectives: Run the column at total reflux. When steady state is reached, take the sample from the top and bottom of the column in order to determine the overall

More information

Model of a Standard Operating Procedure (SOP) for SO 2 measurements with a UV fluorescence analyzer

Model of a Standard Operating Procedure (SOP) for SO 2 measurements with a UV fluorescence analyzer Model of a Standard Operating Procedure (SOP) for SO 2 measurements with a UV fluorescence analyzer How to use this specimen SOP This model SOP aims at giving support to network operators in setting-up

More information

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane Gas requirements Gases for packed columns The carrier gas you use depends upon the type of detector and the performance requirements. Table 520-1 lists gas recommendations for packed column use. In general,

More information

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY Nengbing Xu, Hongmei Ying and Libo Zhu Ningbo Environmental Monitoring

More information

Overview. Front Panel: Keypad and Display

Overview. Front Panel: Keypad and Display Overview The GA-200B is an analyzer that integrates a gas sampling system with sensors to measure and display the concentrations of oxygen and carbon dioxide in a sample as the percentage of a gas in the

More information

Chromat Calibration Updated October 27th, 2017

Chromat Calibration Updated October 27th, 2017 Chromat Calibration Updated October 27th, 2017 Calibrating the Chromatograph Performing the Chromat Calibration is highly recommended when starting a new well. The MLogger already has a default calibration

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results. Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions

More information

Using the UltraRAE. Firmware 2.35

Using the UltraRAE. Firmware 2.35 Using the UltraRAE Firmware 2.35 Training Agenda UltraRAE features Setting up the UltraRAE Turning on the UltraRAE Idle Operation RAE-Sep Tubes Prepping for a measurement Taking a measurement Alarm modes

More information

8.1 Calibration of gas sensors

8.1 Calibration of gas sensors 8.1 Calibration of gas sensors! CAUTION! Calibration gas handling and precautions Observe safety regulation and accident prevention regulations when using or handling test gases. Use test gases in well-ventilated

More information

The Discussion of this exercise covers the following points:

The Discussion of this exercise covers the following points: Exercise 5-3 Wet Reference Leg EXERCISE OBJECTIVE Learn to measure the level in a vessel using a wet reference leg. DISCUSSION OUTLINE The Discussion of this exercise covers the following points: Measuring

More information

Gas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation

Gas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation Gas Clean Filters Delivering Clean Gases for GC and GC/MS Operation Gas Clean Filters Fast, leak-free replacement without tools! Each Gas Clean Filter features a unique quick disconnect design which allows

More information

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by: 1 AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by: Materials/Reagents/Equipment Vendor 1. 0.2 um filtered Water + 0.05% sodium azide 2. 0.2 um filtered

More information

Gas Absorption Draft Standard Operating Procedure

Gas Absorption Draft Standard Operating Procedure Gas Absorption Draft Standard Operating Procedure R.C. 9-14 Scope Page 1 Equipment Overview Page 1 Safety Page 2 Operating Procedure Page 4 Equipment Description Page 16 Gas Absorption Experiment Draft

More information

6200E Installation Requirements

6200E Installation Requirements Requirements When purchasing a standalone 6200E unit with the calibrator, the analyzer will be supplied as three 19 rack mount units. Analyzer Converter Calibrator Included with the unit are two lengths

More information

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist Thank you for purchasing an Agilent system. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is

More information

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual OxyCheq Expedition-X Oxygen Analyzer Operator s Manual OxyCheq 3528 Russell Road Marianna, Florida 32446 USA Tel: 850.482.0385 Email: oxy@oxycheq.com Web: http://oxycheq.com CONTENTS 1.0 Introduction..

More information

Low-Flow Purging Procedure with MicroPurge basics Equipment. MP10 Bladder Pump Controller MP15 Control. MP30 Drawdown Meter.

Low-Flow Purging Procedure with MicroPurge basics Equipment. MP10 Bladder Pump Controller MP15 Control. MP30 Drawdown Meter. Low-Flow Purging Procedure with MicroPurge basics Equipment MP10 Bladder Pump Controller MP15 Control and Power Pack MP20 Flow Cell MP30 Drawdown Meter P.O. Box 326 Ann Arbor, MI 48106-326 USA 1-800-624-2026

More information

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Quantitative Analysis of Hydrocarbons by Gas Chromatography Quantitative Analysis of Hydrocarbons by Gas Chromatography Introduction Gas-liquid chromatography (GLC) accomplishes a separation by partitioning solutes between a mobile gas phase and a stationary liquid

More information

Operating Instructions

Operating Instructions Operating Instructions Preparation for Start-up 1. If the computer system is logged off, log into the computer system. Log into the operator account on the computer. Please ask your TA for username and

More information

User s Guide. CO2 Controller for miniature incubators

User s Guide. CO2 Controller for miniature incubators CO2 Control User s Guide CO2 Controller for miniature incubators Precise CO2 Control throughout the experiment Media ph control Compatible with any perfusion system Miniature incubators for any microscope

More information

Prep LC Training Manual S.V.

Prep LC Training Manual S.V. 1234 Hach Hall 515-294-5805 www.cif.iastate.edu Prep LC Training Manual 11-6-2013 S.V. Location: Contact: 1238 Hach Hall Kamel Harrata, 1236 Hach Hall and Steve Veysey, 1234 Hach Hall Safety: All researchers

More information

LARGE DRY CHEMICAL, TWIN-AGENT, AFFF, HALON 1211, AND AQUASONIC (NITROGEN-OPERATED SYSTEMS) Pressure Regulator Test Manual

LARGE DRY CHEMICAL, TWIN-AGENT, AFFF, HALON 1211, AND AQUASONIC (NITROGEN-OPERATED SYSTEMS) Pressure Regulator Test Manual LARGE DRY CHEMICAL, TWIN-AGENT, AFFF, HALON 1211, AND AQUASONIC (NITROGEN-OPERATED SYSTEMS) Pressure Regulator Test Manual 002694 IntroductIon 8-1-11 Page 1 ForEWord This manual is intended for use with

More information

Prep LC Training Manual

Prep LC Training Manual Prep LC Training Manual (11-6-2013) Location: 1238 Hach Hall Contacts: kamel Harrata and Steve Veysey, 1234 Hach Hall Safety: Introduction: The Agilent PrepStar SD-1 preparative HPLC is a student operated

More information

The Preparation of Low Concentration Hydrogen Sulfide Standards Gulf Coast Conference September 12, 2002 Galveston, TX Paper 050,12:45 pm Moody F

The Preparation of Low Concentration Hydrogen Sulfide Standards Gulf Coast Conference September 12, 2002 Galveston, TX Paper 050,12:45 pm Moody F The Preparation of Low Concentration Hydrogen Sulfide Standards Gulf Coast Conference September 12, 2002 Galveston, TX Paper 050,12:45 pm Moody F Robert Benesch, Malik Haouchine, and Tracey Jacksier, Ph.D.

More information

WATER CONTROL SYSTEM QUICK START

WATER CONTROL SYSTEM QUICK START SETTINGS MENU SYSTEM OPTIONS WATER LEVEL BAR GRAPH 3 DS SPRAY ON/OFF CS SPRAY ON/OFF BEACON WATER CONTROL SYSTEM QUICK START WWW.GFWORLDWIDE.COM +1 (208) 664-9291 SERVICE@GFWORLDWIDE.COM DISPLAY SCREEN

More information

AKTA pure 25 New Owner s Intro

AKTA pure 25 New Owner s Intro AKTA pure 25 New Owner s Intro The exercise below will give a quick demonstration of how easy and intuitive the AKTA pure 25 will be for you in demonstrating downstream processing to your students. Steps

More information

The instrument will process the samples one by one till all the samples have been cleaned up.

The instrument will process the samples one by one till all the samples have been cleaned up. Table of Contents SPE-01 Cleanup Station... 1 LC-04SP Valve System... 5 LC-04SP Standalone Valve... 10 LC-05 Auto Injector... 12 LC-06 Dosing Pump... 16 LC-08 Column Selector... 17 SPE-01 Cleanup Station

More information

LC, RSLC, Bio RS and Dual Gradient Systems

LC, RSLC, Bio RS and Dual Gradient Systems Thermo Scientific Dionex UltiMate 3000 LC, RSLC, Bio RS and Dual Gradient Systems P/N 4823.0205 Revision B December 2014 Automated Method Scouting Considerations All information in this manual is subject

More information

6000 SERIES OF PNEUMATIC DEADWEIGHT TESTERS

6000 SERIES OF PNEUMATIC DEADWEIGHT TESTERS 6000 SERIES OF PNEUMATIC DEADWEIGHT TESTERS USER MANUAL -1- SI-BARNET DEADWEIGHT TESTER 6000 SERIES Release: S31-0250 Revision: A Date: 05/08/03 MANUFACTURED BY: GE RUSKA RUSKA INSTRUMENT CORPORATION 10311

More information

Pressure Automated Calibration Equipment

Pressure Automated Calibration Equipment GE Measurement & control Pressure Automated Calibration Equipment Safety Instructions and User Guide - K0447 PACE5000 PACE6000 K0447 Issue No. 9 1 10 1 PACE5000 1 2 3 4 5 PACE6000 2 6 7 8 3 4 5 6 7 8 9

More information

Sales Guide. 700HPPK Pneumatic Test Pump Kit

Sales Guide. 700HPPK Pneumatic Test Pump Kit Sales Guide 700HPPK Pneumatic Test Pump Kit 1. Product Positioning The 700HPPK High Pressure Pneumatic Test Pump Kit is the pressure generation tool preferred by Instrument and Calibration Technicians

More information

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin Electronic Control Manual 216 Sensors and Stats Section S Technical Bulletin CDS-2000 Issue Date 0393 CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin Introduction 3 Pre-Verification

More information

Columbus Instruments

Columbus Instruments 0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011

More information

Installation Instructions

Installation Instructions Model: ACT-4500 ACT-4500-3 ACT-4500-6 SUN ELECTRIC CORPORATION Page: 1 of 5 Installation Instructions INSTALLATION MUST BE PERFORMED BY QUALIFIED SUN PERSONNEL ONLY INSTALLATION OVERVIEW: The Installation

More information

6890 GC Site Preparation

6890 GC Site Preparation 6890 GC Site Preparation (a16011) This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this material. The information

More information

Agilent G3180B Two-Way Splitter Kit

Agilent G3180B Two-Way Splitter Kit Agilent G3180B Two-Way Splitter Kit With Makeup Gas Installation and Operation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2006 No part of this manual may be reproduced in any form or

More information

Making OTR Measurements in Bottles using the. BottlePerm Fixture

Making OTR Measurements in Bottles using the. BottlePerm Fixture Application Note: Dated: June 24, 2013 Introduction Making OTR Measurements in Bottles using the BottlePerm Fixture Measurement of oxygen transmission rate (OTR) of bottles can be complicated and time

More information

RK LEL Sample Draw Aspirator Adapter Operator s Manual

RK LEL Sample Draw Aspirator Adapter Operator s Manual 30-0951RK LEL Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0017RK Revision: A Released: 6/2/10 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment

More information

911A Summa Canister Sampler

911A Summa Canister Sampler North Carolina Department of Environment & Natural Resources Division of Air Quality 911A Canister Sampler Page 1 of 7 Standard Operating Procedure for 911A Summa Canister Sampler 1.0 General Information

More information

ARC Unofficial Manual to the Solvent Dispensing System. Designed by J. C. Meyer

ARC Unofficial Manual to the Solvent Dispensing System. Designed by J. C. Meyer RC Unofficial Manual to the Solvent Dispensing System Designed by J. C. Meyer 1 Table of Contents: I. Introduction...3 II. rgon Supply...3 Tanks Regulators Over Pressurization III. Solvent Kegs...4 Solvent

More information

CSA Sample Draw Aspirator Adapter Operator s Manual

CSA Sample Draw Aspirator Adapter Operator s Manual 30-0951-CSA Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0367 Revision: 0 Released: 4/30/15 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating

More information

MV5: Automated concentration system

MV5: Automated concentration system MV5: Automated concentration system Abstract LabTech MV5 is the latest automatic nitrogen evaporator developed by our R&D Team able to process up to 54 samples simultaneously. This advanced instrument

More information

Test pumps. Pressure sources. Good reasons for proper and reliable calibration

Test pumps. Pressure sources. Good reasons for proper and reliable calibration Test pumps Pressure is one of the most commonly measured quantities in engineering, which is why exact and reliable pressure measurement is especially important. However, the characteristics of the even

More information

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation User s Booklet for the Wyatt minidawn Light Scattering Instrumentation The Wyatt minidawn Light Scattering instrument is able to measure the weight average molecular weight of a synthetic polymer or a

More information

Beamex PG PRESSURE GENERATORS. Beamex PGM PGV PGC PGHH PGPH PGL

Beamex PG PRESSURE GENERATORS. Beamex PGM PGV PGC PGHH PGPH PGL Beamex PG PRESSURE GENERATORS 97 Beamex PGM PGV PGC PGHH PGPH PGL Beamex PG pressure generators are portable hand-operated pressure generators and ultimate field calibration pumps 98 Calibration pumps

More information

Operation of the Perkin Elmer TGA-GC/MS

Operation of the Perkin Elmer TGA-GC/MS Operation of the Perkin Elmer TGA-GC/MS Summary of the TGA-GC/MS: The TGA-GC/MS allows a user to decompose a sample by heating, measure its loss of mass, and simultaneously analyze the chemical composition

More information

BIOGAS COMBUSTION PLANT MAINTENANCE MANUAL

BIOGAS COMBUSTION PLANT MAINTENANCE MANUAL BIOGAS COMBUSTION PLANT MAINTENANCE MANUAL (PLANT MAINTENANCE MANUAL) Puente Gallego Landfill, Rosario (Argentina) ARIA.BIZ S.A. Montevideo 589, 7 floor 1019 Capital federal Argentina Tel: +54 11 5171

More information

THE UNIVENTOR 400 ANAESTHESIA UNIT

THE UNIVENTOR 400 ANAESTHESIA UNIT THE UNIVENTOR 400 ANAESTHESIA UNIT TABLE OF CONTENTS EDITION 3 Section 1 - WARRANTY & SERVICE 1.1. WARRANTY 2 1.2. DAMAGED SHIPMENTS 2 1.3. SERVICE 2 Section 2 - INTRODUCTION 2.1. INTRODUCTION 3 Section

More information

490 Micro GC Site Preparation Checklist

490 Micro GC Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Glove Box Installation Manual

Glove Box Installation Manual Glove Box Installation Manual 1998 by M. Braun Company File: GB-UNI-INS.DOC! Edition 08-00 by M. Boutin! Subject to be changed without notice Glovebox Installation Your Glove box has been fully assembled,

More information

NITROGEN INFLATION SYSTEM NG-4, NG-6, NG-12, NG-18

NITROGEN INFLATION SYSTEM NG-4, NG-6, NG-12, NG-18 NITROGEN OPERATING INSTRUCTIONS NITROGEN INFLATION SYSTEM NG-4, NG-6, NG-12, NG-18 READ INSTRUCTIONS THOROUGHLY BEFORE OPERATING 3451 S. 40th Street Phoenix, AZ 85040 602.437.5020 800.223.4540 www.tsissg.com

More information

LNG Regulator Test Kit

LNG Regulator Test Kit OVERVIEW Chart provides a LNG Economizer/Regulator Test Kit to help save money and time by allowing a technician to more accurately and easily diagnose the operability of the pressure control regulator/economizer

More information

3 Flow and Pressure Control

3 Flow and Pressure Control 3 Hydrogen shutdown Column shutdown Turning gas flows on and off EPC-controlled streams NonEPC-controlled streams Electronic Pneumatic Control (EPC) Interpreting flow and pressure readings Configuration

More information

TEL/jlRE" Introduction. Display Features and Modes. Startup Procedure. Power-Up Procedure. Adjustment Modes

TEL/jlRE Introduction. Display Features and Modes. Startup Procedure. Power-Up Procedure. Adjustment Modes TEL/jlRE" Introduction The Telaire 7001 CO 2 /T emperature monitor (shown in Fi gt u e 1 below) is an easy to use hand-held instnunent, which provides stable and highly accurate readings due to Telaire

More information

ISOARC project. Samoylov water vapour isotopes analyser raw data

ISOARC project. Samoylov water vapour isotopes analyser raw data ISOARC project Samoylov water vapour isotopes analyser raw data Within the ISOARC project, the water vapour analyser on Samoylov is dedicated to the continuous measurements in ambient air of humidity level

More information

Adsorption Unit Operating Procedures: CO2 Adsorption on to Activated Carbon

Adsorption Unit Operating Procedures: CO2 Adsorption on to Activated Carbon Adsorption Unit Operating Procedures: CO2 Adsorption on to Activated Carbon Background Safety Concerns Description of the Unit Switching the Unit On Preparation of the Unit Configuration of the Unit for

More information

SOP: Buck Scientific BLC-20P HPLC Operation

SOP: Buck Scientific BLC-20P HPLC Operation Page 1 of 11 Approvals Preparer: John Buford Reviewer: Tim Kull Reviewer: Dr. Margaret Bryans Date: 16OCT13 Date: 30OCT13 Date: 31OCT13 1. Purpose 1.1. Basic operation of the Buck Scientific BLC-20P isocratic

More information

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE INSTALLATION, OPERATION & MAINTENANCE MANUAL INTERFACE DEVICES, INC. 230 Depot Road, Milford, CT 06460 Ph: (203) 878-4648, Fx: (203) 882-0885, E-mail: info@interfacedevices.com

More information

Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions

Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions Introduction: There is a bit of explanation of the health effects of benzene in the Introduction section of the first part

More information

MINIBRUTE ANNEAL TUBE STANDARD OPERATION PROCEDURE

MINIBRUTE ANNEAL TUBE STANDARD OPERATION PROCEDURE Arizona State University NanoFab MINIBRUTE ANNEAL TUBE STANDARD OPERATION PROCEDURE Rev A Title: MINIBRUTE ANNEAL TUBE STANDARD OPERATION PROCEDURE Table of Contents Issue: Rev A Contents Table of Contents...1

More information