1 Translation of Validation Summary Report DOMETIC MT8B Original report in German established by page 1
2 Table of Contents Description of the Validation Method...3 Test Readings MT8B at +10ºC Outside Temperature over 24 Hours...5 Performance Report...6 Chart...7 Test Readings MT8B at +32ºC Outside Temperature over 24 Hours...8 Performance Report...9 Chart...10 Test Readings MT8B "Maximum Cold Life" at +32ºC Outside Temperature...11 Performance Report...12 Chart...13 Test Readings MT8B "Maximum Cold Life" at +43ºC Outside Temperature...14 Performance Report...15 Chart...16 Validation Results...17 Validation Report MT8B...18 page 2
3 Validation of the Dometic MT8B Containment for the Transport of Blood Preparations 1 Introduction Object of the validation is the transport of erythrocyte concentrates (EC) in the Dometic MT8B transport container whilst maintaining the cold chain required for these preparations. The relevant German and European laws and directives require a storage temperature of 4ºC ± 2K or a transport temperature between 1ºC and 10ºC. The question of this validation is under what environmental conditions can erythrocyte concentrates be transported legally safely (and therefore without quality loss) over what period using the Dometic MT8B. 2 Basic Conditions 2.1 Description of System The transport containers of the Dometic Series MT8B are cooled passively by means of ice packs. The casings are manufactured by rotation process and receive an insulation of polyurethane foam. The complete functional unit always consists of an outer and an inner container. Summary of Individual Features Description MT8B Outer dimensions (HxWxD in mm) 435x590x290 Weight 7.0 kg Gross interior volume 18 litres Storage capacity (ECs at 270 ml each) 14 blood bags Number of ice packs (at 600 ml each) at 2ºC to +10ºC outside temperature None at +10ºC to +20ºC outside temperature 2 at outside temperature of above +20ºC 4 In order to compensate for statistical deviations and series variations, three different test patterns are used for this validation. 2.2 Ambient Temperatures Three ambient temperature ranges have been chosen for this validation: +10ºC, +32ºC, and +43ºC. As minimum or maximum limits, these temperature ranges cover more than 90% of the transport scenarios imaginable for Central Europe, including safety margins. 2.3 Charge For the 24 hours tests at +32ºC and +43ºC the typical lower charge limit of blood bags is chosen as charge. These low charge levels are more unstable, and the resulting test readings are very telling in respect to critical temperature ranges. With increasing charges the preparations' temperatures in the secondary container become increasingly stable. In fact this means a charge of 2 blood bags at 270 ml each. In order to determine the Maximum Cold Life at +32ºC and +43ºC, the maximum charge (i.e. 14 blood bags at 270 ml each) is used to determine the absolute performance limits. page 3
4 2.4 Time Frame The validation is carried out for a transport time of 24 hours. This period covers almost 100% of the genuine range of usage times for Central Europe. In addition, the time value "Maximum Cold Life" is determined for the ambient temperature ranges +32ºC and +43ºC. This value indicates the maximum amount of time the blood products can be held in the defined transport temperature ranges. 2.5 Check Points The recording of test readings is carried out at ten points in the interior container of the transport container: eight check points in the corners of the container, always at a distance of two centimeters of all three nearest side walls, one check point at the intersection of the diagonals, and another one inside a blood bag. If more than one blood bag is used, the bag containing the sensor is placed in the centre. Top View Front View The erythrocyte concentrate is substituted by a mixture consisting of 110 ml PAGGS-M and 160 ml HAES-Sterile 6% infusion solution. This mixture has very similar characteristics with respect to density and viscosity. Composelect by Fresenius HemoCare is used as blood bags. 2.6 Other The ice packs are frozen to a core temperature of between -18ºC and -22ºC by storing them for at least 24 hours at the corresponding temperature. Before the boxes are equipped with the cooling elements, the elements are allowed to defrost at room temperature for minutes. Before starting a series of readings, the open (both interior and outer container are open) transport containers are exposed to the ambient temperature for at least six hours. page 4
5 Test Readings MT8B At an outside temperature of +10ºC over 24 hours page 5
6 Ambient temperature: +10ºC Number of measuring runs: 3 Duration of validation: 24 hours Validation Summary Report Dometic Series MT Performance Report for Dometic MT8B at +10ºC over 24 h Ambient temperature Measuring run 1 Measuring run 2 Measuring run 3 Accumulated Absolute Average temperature 10.13ºC 10.16ºC 10.14ºC 10.14ºC 10.14ºC Minimum value 8.22ºC 8.34ºC 8.33ºC 8.68ºC 8.22ºC Maximum value 12.10ºC 12.10ºC 12.09ºC 11.77ºC 12.10ºC Test readings MT8B Charge: 2 ECs at 270 ml each; two cooling elements at 600 ml each Temp 30 (air at top) Average temperature 4.52ºC 4.45ºC 4.66ºC 4.53ºC 4.66ºC Minimum value 4.07ºC 3.97ºC 4.17ºC 4.07ºC 4.17ºC Maximum value 19.57ºC 19.50ºC 19.70ºC 18.98ºC 19.70ºC Temp 31 (air at top) Average temperature 4.44ºC 4.37ºC 4.57ºC 4.45ºC 4.57ºC Minimum value 3.99ºC 3.87ºC 4.07ºC 3.99ºC 4.07ºC Maximum value 19.56ºC 19.44ºC 19.66ºC 18.88ºC 19.66ºC Temp 32 (air at top) Average temperature 4.39ºC 4.32ºC 4.52ºC 4.40ºC 4.52ºC Minimum value 3.91ºC 3.80ºC 4.02ºC 3.92ºC 4.02ºC Maximum value 19.45ºC 19.34ºC 19.59ºC 18.84ºC 19.59ºC Temp 33 (air at top) Average temperature 4.50ºC 4.43ºC 4.64ºC 4.51ºC 4.64 ºC Minimum value 4.03ºC 3.92ºC 4.13ºC 4.04ºC 4.13 ºC Maximum value 19.59ºC 19.51ºC 19.67ºC 18.97ºC ºC Temp 34 (air at bottom) Average temperature 4.30ºC 4.23ºC 4.44ºC 4.32ºC 4.44ºC Minimum value 3.88ºC 3.77ºC 3.99ºC 3.88ºC 3.99ºC Maximum value 15.87ºC 15.72ºC 15.99ºC 15.17ºC 15.99ºC Temp 35 (air at bottom) Average temperature 4.43ºC 4.36ºC 4.57ºC 4.44ºC 4.57ºC Minimum value 3.98ºC 3.88ºC 4.09ºC 4.01ºC 4.09ºC Maximum value 16.00ºC 15.81ºC 16.14ºC 15.29ºC 16.14ºC Temp 36 (air at bottom) Average temperature 4.47ºC 4.40ºC 4.61ºC 4.48ºC 4.61ºC Minimum value 4.02ºC 3.91ºC 4.13ºC 4.04ºC 4.13ºC Maximum value 15.98ºC 15.94ºC 16.09ºC 15.31ºC 16.09ºC Temp 37 (air at bottom) Average temperature 4.39ºC 4.3 ºC 4.53ºC 4.40ºC 4.53ºC Minimum value 3.94ºC 3.83ºC 4.06ºC 3.95ºC 4.06ºC Maximum value 15.95ºC 15.86ºC 16.02ºC 15.28ºC 16.02ºC Temp 38 (air at centre) Average temperature 4.44ºC 4.37ºC 4.58ºC 4.45ºC 4.58ºC Minimum value 4.00ºC 3.89ºC 4.10ºC 4.01ºC 4.10ºC Maximum value 18.36ºC 18.25ºC 18.48ºC 17.69ºC 18.48ºC Temp 29 (EC) Initial temperature 3.73ºC 3.84ºC 3.85ºC 3.81ºC Average temperature 4.53ºC 4.47ºC 4.67ºC 4.56ºC 4.56ºC Minimum value 3.73ºC 3.84ºC 3.85ºC 3.81ºC 3.73ºC Maximum value 5.34ºC 5.35ºC 5.47ºC 5.38ºC 5.47ºC Final temperature 4.71ºC 4.76ºC 5.00ºC 4.82ºC Storage temperatures for ECs (2ºC - 6ºC) were maintained for: > 24:00:00 Transport temperatures for ECs (1ºC - 10ºC) were maintained for: > 24:00:00 page 6
7 Temperature chart for Dometic MT8B over 24 hours at an ambient temperature of 10 C (accumulated results) Ambient temperature MT8B Top Temp 30 MT8B Top Temp 31 MT8B Top Temp 32 MT8B Top Temp 33 MT8B Bottom Temp 34 MT8B Bottom Temp 35 MT8B Bottom Temp 36 MT8B Bottom Temp 37 MT8B Centre Temp 38 MT8B EC Temp 39 Time (hours) page 7
8 Test Readings MT8B At an outside temperature of +32ºC over 24 hours page 8
9 Ambient temperature: +32ºC Number of measuring runs: 3 Duration of validation: 24 hours Performance Report for Dometic MT8B at +32ºC over 24 h Ambient temperature Measuring run 1 Measuring run 2 Measuring run 3 Accumulated Absolute Average temperature 32.06ºC 32.07ºC 32.07ºC 32.07ºC 32.07ºC Minimum value 28.08ºC 27.50ºC 28.01ºC 27.87ºC 27.50ºC Maximum value 35.07ºC 35.07ºC 34.67ºC 34.65ºC 35.07ºC Test readings MT8B Charge: 2 ECs at 270 ml each; two cooling elements at 600 ml each Temp 30 (air at top) Average temperature 6.62ºC 6.71ºC 6.77ºC 6.69ºC 6.77ºC Minimum value 1.59ºC 1.69 ºC 1.75ºC 1.68ºC 1.75ºC Maximum value 15.49ºC ºC 15.63ºC 14.73ºC 15.63ºC Temp 31 (air at top) Average temperature 6.54ºC 6.62ºC 6.69ºC 6.61ºC 6.69ºC Minimum value 1.51ºC 1.58ºC 1.65ºC 1.60ºC 1.65ºC Maximum value 15.46ºC 15.50ºC 15.61ºC 14.62ºC 15.61ºC Temp 32 (air at top) Average temperature 6.49ºC 6.57ºC 6.63ºC 6.56ºC 6.63ºC Minimum value 1.44ºC 1.55ºC 1.60ºC 1.54ºC 1.60ºC Maximum value 15.38ºC 15.46ºC 15.51ºC 14.61ºC 15.51ºC Temp 33 (air at top) Average temperature 6.60ºC 6.68ºC 6.75ºC 6.67ºC 6.75ºC Minimum value 1.56ºC 1.64ºC 1.70ºC 1.67ºC 1.70ºC Maximum value 15.51ºC 15.57ºC 15.62ºC 14.69ºC 15.62ºC Temp 34 (air at bottom) Average temperature 5.38ºC 5.46ºC 5.53ºC 5.45ºC 5.53ºC Minimum value 1.18ºC 1.25ºC 1.31ºC 1.26ºC 1.31ºC Maximum value 16.24ºC 16.26ºC 16.30ºC 15.56ºC 16.30ºC Temp 35 (air at bottom) Average temperature 5.50ºC 5.59ºC 5.65ºC 5.57ºC 5.65ºC Minimum value 1.27ºC 1.35ºC 1.42ºC 1.38ºC 1.42ºC Maximum value 16.30ºC 16.39ºC 16.45ºC 15.66ºC 16.45ºC Temp 36 (air at bottom) Average temperature 5.54ºC 5.63ºC 5.69ºC 5.61ºC 5.69ºC Minimum value 1.31ºC 1.42ºC 1.48ºC 1.41ºC 1.48ºC Maximum value 16.34ºC 16.44ºC 16.51ºC 15.70ºC 16.51ºC Temp 37 (air at bottom) Average temperature 5.46ºC 5.55ºC 5.61ºC 5.53ºC 5.61ºC Minimum value 1.22ºC 1.32ºC 1.36ºC 1.31ºC 1.36ºC Maximum value 16.30ºC 16.40ºC 16.44ºC 15.62ºC 16.44ºC Temp 38 (air at centre) Average temperature 6.20ºC 6.28ºC 6.35ºC 6.27ºC 6.35ºC Minimum value 1.46ºC 1.51ºC 1.60ºC 1.53ºC 1.60ºC Maximum value 15.77ºC 15.85ºC 15.87ºC 14.97ºC 15.87ºC Temp 29 (EC) Initial temperature 4.22ºC 3.99ºC 4.48ºC 4.23ºC Average temperature 6.01ºC 6.08ºC 6.16ºC 6.08ºC 6.08ºC Minimum value 3.04ºC 3.04ºC 3.22ºC 3.10ºC 3.04ºC Maximum value 7.48ºC 7.57ºC 7.62ºC 7.55ºC 7.62ºC Final temperature 7.48ºC 7.57ºC 7.62ºC 7.55ºC Storage temperatures for ECs (2ºC - 6ºC) were maintained for: > 08:25:00 Transport temperatures for ECs (1ºC - 10ºC) were maintained for: > 24:00:00 page 9
10 Temperature chart for Dometic MT8B over 24 hours at an ambient temperature of 32 C (accumulated results) Ambient temperature MT8B Top Temp 30 MT8B Top Temp 31 MT8B Top Temp 32 MT8B Top Temp 33 MT8B Bottom Temp 34 MT8B Bottom Temp 35 MT8B Bottom Temp 36 MT8B Bottom Temp 37 MT8B Centre Temp 38 MT8B EC Temp 39 Time (hours) page 10
11 Test Readings "Maximum Cold Life" MT8B At an outside temperature of +32ºC page 11
12 Performance Report for Dometic MT8B "Maximum Cold Life" at +32ºC Note: All statistical values refer to the period of time until the 10ºC limit in the EC is reached (Temp 39) Ambient temperature: +32ºC Number of measuring runs: 3 Ambient temperature Measuring run 1 Measuring run 2 Measuring run 3 Accumulated Absolute Average temperature 32.57ºC 32.57ºC 32.57ºC 32.57ºC 32.57ºC Minimum value 31.53ºC 31.53ºC 31.56ºC 31.65ºC 31.53ºC Maximum value 34.08ºC 33.98ºC 34.07ºC 33.80ºC 34.08ºC Test readings MT8B Charge: 14 ECs at 270 ml each; 4 cooling elements at 600 ml each Temp 30 (air at top) Average temperature 5.32ºC 5.58ºC 5.36ºC 5.42ºC 5.42ºC Minimum value 2.42ºC 2.81ºC 2.48ºC 2.59ºC 2.42ºC Maximum value 11.57ºC 11.51ºC 11.57ºC 11.52ºC 11.57ºC Temp 31 (air at top) Average temperature 5.23ºC 5.49ºC 5.27ºC 5.34ºC 5.33ºC Minimum value 2.31ºC 2.71ºC 2.38ºC 2.47ºC 2.31ºC Maximum value 11.47ºC 11.42ºC 11.45ºC 11.44ºC 11.47ºC Temp 32 (air at top) Average temperature 5.18ºC 5.45ºC 5.22ºC 5.29ºC 5.28ºC Minimum value 2.27ºC 2.65ºC 2.33ºC 2.44ºC 2.27ºC Maximum value 11.44ºC 11.38ºC 11.44ºC 11.36ºC 11.44ºC Temp 33 (air at top) Average temperature 5.29ºC 5.56ºC 5.33ºC 5.40ºC 5.39ºC Minimum value 2.38ºC 2.76ºC 2.44ºC 2.56ºC 2.38ºC Maximum value 11.53ºC 11.49ºC 11.55ºC 11.49ºC 11.55ºC Temp 34 (air at bottom) Average temperature 5.03ºC 5.32ºC 5.08ºC 5.14ºC 5.14ºC Minimum value 2.67ºC 3.06ºC 2.73ºC 2.83ºC 2.67ºC Maximum value 9.95ºC 9.80ºC 9.89ºC 9.86ºC 9.95ºC Temp 35 (air at bottom) Average temperature 5.16ºC 5.45ºC 5.20ºC 5.27ºC 5.27ºC Minimum value 2.78ºC 3.16ºC 2.83ºC 2.95ºC 2.78ºC Maximum value 10.02ºC 9.90ºC 10.01ºC 10.00ºC 10.02ºC Temp 36 (air at bottom) Average temperature 5.20ºC 5.49ºC 5.24ºC 5.31ºC 5.31ºC Minimum value 2.82ºC 3.21ºC 2.88ºC 3.00ºC 2.82ºC Maximum value 10.13ºC 10.00ºC 10.02ºC 10.05ºC 10.31ºC Temp 37 (air at bottom) Average temperature 5.12ºC 5.41ºC 5.16ºC 5.23ºC 5.23ºC Minimum value 2.74ºC 3.12ºC 2.80ºC 2.91ºC 2.74ºC Maximum value 9.98ºC 9.89ºC 9.94ºC 9.92ºC 9.98ºC Temp 38 (air at centre) Average temperature 5.21ºC 5.48ºC 5.26ºC 5.32ºC 5.32ºC Minimum value 2.49ºC 2.87ºC 2.56ºC 2.65ºC 2.49ºC Maximum value 11.02ºC 10.94ºC 10.99ºC 10.94ºC 11.02ºC Temp 29 (EC) Initial temperature 3.91ºC 3.82ºC 4.36ºC 4.03ºC Average temperature 5.02ºC 5.29ºC 5.09ºC 5.14ºC 5.13ºC Minimum value 2.79ºC 3.17ºC 2.86ºC 2.94ºC 2.79ºC Maximum value 10.01ºC 10.00ºC 10.00ºC 10.00ºC 10.01ºC Storage temperatures for ECs (2ºC - 6ºC) were maintained for: 30:30 Transport temperatures for ECs (1ºC - 10ºC) were maintained for: 57:02 page 12
13 Validation for Dometic MT8B Transport Container Maximum Cold Life at +32 C (accumulated results) Time (hours) Ambient temperature MT8B Top Temp 30 MT8B Top Temp 31 MT8B Top Temp 32 MT8B Top Temp 33 MT8B Bottom Temp 34 MT8B Bottom Temp 35 MT8B Bottom Temp 36 MT8B Bottom Temp 37 MT8B Centre Temp 38 MT8B EC Temp 39 page 13
14 Test Readings "Maximum Cold Life" MT8B At an outside temperature of +43ºC page 14
15 Performance Report for Dometic MT8B "Maximum Cold Life" at +43ºC Note: All statistical values refer to the period of time until the 10ºC limit in the EC is reached (Temp 39) Ambient temperature: +43ºC Number of measuring runs: 3 Ambient temperature Measuring run 1 Measuring run 2 Measuring run 3 Accumulated Absolute Average temperature 43.40ºC 43.39ºC 43.39ºC 43.39ºC 43.39ºC Minimum value 36.21ºC 36.44ºC 36.25ºC 36.30ºC 36.21ºC Maximum value 44.28ºC 44.34ºC 44.29ºC 44.25ºC 44.34ºC Test readings MT8B Charge: 14 ECs at 270 ml each; 4 cooling elements at 600 ml each Temp 30 (air at top) Average temperature 5.32ºC 5.58ºC 5.36ºC 5.42ºC 5.42ºC Minimum value 2.42ºC 2.81ºC 2.48ºC 2.59ºC 2.42ºC Maximum value 11.57ºC 11.51ºC 11.57ºC 11.52ºC 11.57ºC Temp 31 (air at top) Average temperature 5.23ºC 5.49ºC 5.27ºC 5.34ºC 5.33ºC Minimum value 2.31ºC 2.71ºC 2.38ºC 2.47ºC 2.31ºC Maximum value 11.47ºC 11.42ºC 11.45ºC 11.44ºC 11.47ºC Temp 32 (air at top) Average temperature 5.18ºC 5.45ºC 5.22ºC 5.29ºC 5.28ºC Minimum value 2.27ºC 2.65ºC 2.33ºC 2.44ºC 2.27ºC Maximum value 11.44ºC 11.38ºC 11.44ºC 11.36ºC 11.44ºC Temp 33 (air at top) Average temperature 5.29ºC 5.56ºC 5.33ºC 5.40ºC 5.39ºC Minimum value 2.38ºC 2.76ºC 2.44ºC 2.56ºC 2.38ºC Maximum value 11.53ºC 11.49ºC 11.55ºC 11.49ºC 11.55ºC Temp 34 (air at bottom) Average temperature 5.03ºC 5.32ºC 5.08ºC 5.14ºC 5.14ºC Minimum value 2.67ºC 3.06ºC 2.73ºC 2.83ºC 2.67ºC Maximum value 9.95ºC 9.80ºC 9.89ºC 9.86ºC 9.95ºC Temp 35 (air at bottom) Average temperature 5.16ºC 5.45ºC 5.20ºC 5.27ºC 5.27ºC Minimum value 2.78ºC 3.16ºC 2.83ºC 2.95ºC 2.78ºC Maximum value 10.02ºC 9.90ºC 10.01ºC 10.00ºC 10.02ºC Temp 36 (air at bottom) Average temperature 5.20ºC 5.49ºC 5.24ºC 5.31ºC 5.31ºC Minimum value 2.82ºC 3.21ºC 2.88ºC 3.00ºC 2.82ºC Maximum value 10.13ºC ºC 10.02ºC 10.05ºC 10.31ºC Temp 37 (air at bottom) Average temperature 5.12ºC 5.41ºC 5.16ºC 5.23ºC 5.23ºC Minimum value 2.74ºC 3.12ºC 2.80ºC 2.91ºC 2.74ºC Maximum value 9.98ºC 9.89ºC 9.94ºC 9.92ºC 9.98ºC Temp 38 (air at centre) Average temperature 5.21ºC 5.48ºC 5.26ºC 5.32ºC 5.32ºC Minimum value 2.49ºC 2.87ºC 2.56ºC 2.65ºC 2.49ºC Maximum value 11.02ºC 10.94ºC 10.99ºC 10.94ºC 11.02ºC Temp 29 (EC) Initial temperature 3.91ºC 3.82ºC 4.36ºC 4.03ºC Average temperature 5.02ºC 5.29ºC 5.09ºC 5.14ºC 5.13ºC Minimum value 2.79ºC 3.17ºC 2.86ºC 2.94ºC 2.79ºC Maximum value 10.01ºC 10.00ºC 10.00ºC 10.00ºC 10.01ºC Storage temperatures for ECs (2ºC - 6ºC) were maintained for: 12:02 Transport temperatures for ECs (1ºC - 10ºC) were maintained for: 16:39 page 15
16 Validation for Dometic MT8B Transport Container Maximum Cold Life at +43 C (accumulated results) Time (hours) Ambient temperature MT8B Top Temp 30 MT8B Top Temp 31 MT8B Top Temp 32 MT8B Top Temp 33 MT8B Bottom Temp 34 MT8B Bottom Temp 35 MT8B Bottom Temp 36 MT8B Bottom Temp 37 MT8B Centre Temp 38 MT8B EC Temp 39 page 16
17 Validation Results page 17
18 Validation Report DOMETIC MT8B Object: Range of application: Results: Dometic MT8B Special transport container for the transport of blood preparations PROD-ID ; S/N , , Transport of erythrocyte concentrates under Central European average climatic conditions Temp. +10 C over 24 h; 2 blood bags at 270 ml each; 2 icepacks at 600 ml each Ambient temperature Average: C Absolute minimum: 8.68 C Absolute maximum: C Blood bag core temperature Initial temperature: 3.81 C Final temperature: 4.82 C Maximum temperature: 5.38 C Minimum temperature: 3.81 C Temp. +32 C over 24 h; 2 blood bags at 270 ml each; 4 icepacks at 600 ml each Ambient temperature Average: C Absolute minimum: C Absolute maximum: C Blood bag core temperature Initial temperature: 4.23 C Final temperature: 7.55 C Maximum temperature: 7.55 C Minimum temperature: 3.10 C Temp. +32 C "Max. Cold Life"; 14 blood bags at 270ml each; 4 icepacks at 600ml each Ambient temperature Average: C Absolute minimum: C Absolute maximum: C Blood bag core temperature Initial temperature: 4.01 C Minimum temperature: 4.01 C Time until 6 C: 30h:30min Time until 10 C: 57h:02min Temp. +43 C "Max. Cold Life"; 14 blood bags at 270ml each; 4 icepacks at 600ml each Ambient temperature Average: C Absolute minimum: C Absolute maximum: C Blood bag core temperature Initial temperature: 4.03 C Minimum temperature: 2.94 C Time until 6 C: 12h:02min Time until 10 C: 16h:39min page 18
19 Evaluation: The effect of the ambient temperature on the two erythrocyte concentrates stored in a transport container of the type Dometic MT8B over a period of 24 hours, is small. At an ambient temperature of +10 C, the preparations' core temperature within a MT8B charged with 2 blood bags at 270 ml each over more than 24 hours remains within the legally required storage temperature range. With the same charge and at an ambient temperature of +32 C, the storage temperature range of 2 C to 6 C is maintained for 8:25 hours and the transport temperature range is maintained for more than 24 hours. With a full charge of 14 blood bags at 270 ml each and at an ambient temperature of +32 C, the storage temperature range is maintained for over 30 hours and the transport temperature range for more than 57 hours, i.e. notably more than two days. Under extreme conditions, at an ambient temperature of +43 C and with a full charge, the storage temperatures are still maintained for more than 12 hours and the transport temperature range for over 16 hours. The passive design of the transport container makes the container entirely impervious to operating errors or power cuts under any environmental conditions if it is correctly configured. The casing of the transport container is extremely sturdy and almost entirely impervious against exterior forces, e.g. bumps and falls. Furthermore, it can be cleaned and disinfected easily and thoroughly. The interior does not feature any inaccessible corners and areas. Münster, 19th December 2003 page 19
CBMTG Laboratory Committee-Laboratory Guidance
Purpose: To confirm, by examination and objective evidence, that a selected cryopreservation bag is compatible with institutional policies and standard operating procedures and functions adequately so
More informationOutside Air Nonresidential HVAC Stakeholder Meeting #2 California Statewide Utility Codes and Standards Program
1 Outside Air Nonresidential HVAC Stakeholder Meeting #2 California Statewide Utility Codes and Standards Program Jim Meacham, CTG Energetics Agenda 2 In Situ Testing Results Reduced Ventilation after
More informationTR Test Report Complementary Testing RKN e1 Container (+5 C) Protection by Insulation Version: 01 Page: 2 (22)
RKN e1 Container (+5 C) Protection by Insulation Version: 01 Page: 2 (22) Document History Version Date of issue Revision(s) Issued by 01 16OCT2012 N/A (First version of document) David Graaf, Envirotainer
More informationSTAD-D. Balancing valves Balancing valve for domestic water systems, DN Balancing valves
STAD-D Balancing valves Balancing valve for domestic water systems, DN 15-50 Balancing valves IMI TA / Balancing valves / STAD-D STAD-D The STAD-D balancing valve delivers accurate hydronic performance
More informationSTA. Double regulating valves DN 15-50
STA Double regulating valves DN -0 IMI TA / Balancing valves / STA STA The STA double regulating valve delivers accurate hydronic performance in an impressive range of applications. Ideally suited for
More informationA Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers
A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers Slemeyer, A. University of Applied Sciences of Giessen, Dept. of Electrical Engineering 1 Wiesenstr. 14, D-35390
More informationCORESTA RECOMMENDED METHOD N 6
CORESTA RECOMMENDED METHOD N 6 DETERMINATION OF VENTILATION DEFINITIONS AND MEASUREMENT PRINCIPLES (2015 Revision September 2016) 1. SCOPE This CORESTA Recommended Method specifies a method for the determination
More informationSTA. Double regulating valves DN 15-50
STA Double regulating valves DN 15-50 IMI TA / Balancing valves / STA STA The STA double regulating valve delivers accurate hydronic performance in an impressive range of applications. Ideally suited for
More informationEPSILON. User Manual. EP-5KG (5000g x 1g) Copyright 2011 American Weigh Scales, Inc. All rights reserved. Rev. 1.0
User Manual EP-5KG (5000g x 1g) Copyright 2011 American Weigh Scales, Inc. All rights reserved. Rev. 1.0 Thank you for purchasing the American Weigh EPSILON digital kitchen scale. Please read all operating
More informationSTAG. Balancing valves Balancing valve with grooved ends DN
STAG Balancing valves Balancing valve with grooved ends DN 65-300 IMI TA / Balancing valves / STAG STAG A grooved end, ductile iron balancing valve that delivers accurate hydronic performance in an impressive
More informationSET OF EUROLEAGUE BACKSTOP UNITS (Reference PK120)
SET OF EUROLEAGUE BACKSTOP UNITS (Reference PK120) DESCRIPTION The EUROLEAGUE backstop unit is mainly designed for multi-sports pavilions and installations where the highest-level basketball competitions
More informationFOOD SAFETY RISK ASSESSMENT
FOOD SAFETY RISK ASSESSMENT FOR Eggbuckland Fish And Chips Ltd Membership Number 12078 Responsible Person - a wiseman This includes a prep kitchen Food Types Equipment Creation / Next Renewal Date Fish
More informationHI 2314 HI 2315 HI 23151
Instruction Manual HI 2314 HI 2315 HI 23151 Multi-Range Conductivity Meters for Laboratories www.hannainst.com Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction
More informationThis parts list is current as of the above revision date. See for the current revision.
Parts List This parts list is current as of the above revision date. See http://www.rheempartslists.net/92-42800-.pdf for the current revision. Heat Pumps - Commercial Table of Contents Model Explanation...................................................................................................................
More informationOxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument
OxyScan Graphic Operating Instructions UMS Micro-oxygen sensor 501 Microprocessor instrument Introduction Thank you for choosing the UMS Micro Oxygen Sensor 501 - a highly advanced product! Please read
More informationBERG BUZZY BERG BUZZY. Easy to ride, easy to master. age 2 5 BERGTOYS.COM. More stable, much faster and more fun than any tricycle!
BERG BUZZY Product brochure BERG BUZZY Product brochure Easy to ride, easy to master. Buzzy Yellow, Buzzy Bloom, Buzzy Nitro, Buzzy Racing Buzzy FIAT500 More stable, much faster and more fun than any tricycle!
More informationOperating manual Analogue Brix Refractometer
KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen, Germany E-mail: info@kern-sohn.com Operating manual Analogue Brix Refractometer Tel: +49-[0]7433-9933-0 Fax: +49 (0)7433-9933-149 Internet: www.kern-sohn.com
More informationP06 & P012 Pitot-Static Probes
P06 & P012 Pitot-Static Probes Operating Instructions: Measure in straight duct sections that are at least 1.5 diameters upstream and 8.5 diameters downstream of any duct disturbance such as elbows and
More informationP8 to P60 Pitot-Static Probes
P8 to P60 Pitot-Static Probes Operating Instructions: Install optional PX NPT fitting if required. See separate documentation for details. Measure in straight duct sections that are at least 1.5 diameters
More informationDrying ovens FD series Drying ovens with forced convection Leistungsmerkmale/Ausstattung:
Drying ovens www.matronic.se FD series Drying ovens with forced convection FD series units are primarily used in applications needing rapid drying and sterilization. Totally homogenous temperature distribution,
More informationSTAD. Balancing valves DN 15-50
STAD Balancing valves DN 15-50 IMI TA / Balancing valves / STAD STAD The STAD balancing valve delivers accurate hydronic performance in an impressive range of applications. Ideally suited for use on the
More informationRESISTORS, HEATERS, FLEXIBLE, SINGLE AND DOUBLE LAYER. ESCC Detail Specification No. 4009/002
Page 1 of 16 RESISTORS, HEATERS, FLEXIBLE, SINGLE AND DOUBLE LAYER ESCC Detail Specification Issue 8 February 2018 Document Custodian: European Space Agency see https://escies.org PAGE 2 LEGAL DISCLAIMER
More informationFlexAct BT (Bag Tester) Validation November, 2015
FlexAct BT (Bag Tester) Validation November, 2015 Agenda 1 Validation Approach and Method 2 Development Case Study 3 Materials Scope of the Validation Point-of-use leak test for Flexboy 2D bags from 50mL
More informationDigiquartz Water-Balanced Pressure Sensors for AUV, ROV, and other Moving Underwater Applications
Digiquartz Water-Balanced Pressure Sensors for AUV, ROV, and other Moving Underwater Applications Dr. Theo Schaad Principal Scientist Paroscientific, Inc. 2002 Paroscientific, Inc. Page 1 of 6 Digiquartz
More informationThe Sea Fish Industry Authority Seafish Technology
The Sea Fish Industry Authority Seafish Technology Trials to Compare the Thermal Performance of a New Design of Tri-pack Corrugated Plastic Nonreusable fish box with Expanded Polystyrene and Single Walled
More informationINSTRUCTION MANUAL FOR POLYKEG DISPOSABLE KEG
TABLE OF CONTENTS OF MANUAL Chapters 1.0 Product description 2.0 Safety rules for the utilisation of the keg 3.0 Instructions for manual filling 4.0 Instructions for automatic filling 5.0 Instructions
More informationFigure 1. Winning percentage when leading by indicated margin after each inning,
The 7 th Inning Is The Key By David W. Smith Presented June, 7 SABR47, New York, New York It is now nearly universal for teams with a 9 th inning lead of three runs or fewer (the definition of a save situation
More informationReport No. 371/ This report consists of 6 pages
Test Report This report consists of 6 pages Licence/Certificate No. KM 578199 Client Super Sealed Units Ltd 35 Lovett Way Woodside Industrial Estate Dunstable LU5 4TU Authority & date BSI Service Management
More informationCompressed gas. In high concentrations may cause asphyxiation
Broadman limited Weldgas: Pure Argon Product Name: Argon Compressed Chemical Formula: Ar Company identification: Broadman Ltd, Wiltshire Road, Hull. HU4 6PA Emergency phone No: 01482 506 050 Substance/Preparation:
More informationSCHOOL SPORT VICTORIA BASEBALL - SECONDARY
SCHOOL SPORT VICTORIA BASEBALL - SECONDARY Updated January 2017 These rules apply to State, Region and Division competitions. Any variations to these rules at Region and Division levels will be found on
More informationEC214 EC215 - EC215R Bench Conductivity Meters
Instruction Manual EC214 EC215 - EC215R Bench Conductivity Meters http://www.hannainst.com These Instruments are in Compliance with the CE Directives Dear Customer, Thank you for choosing a Hanna Instruments
More informationOperating Instructions METTLER TOLEDO Pipette Check Application for AX and MX/UMX Balances Version 1.xx
Operating Instructions METTLER TOLEDO Pipette Check Application for AX and MX/UMX Balances Version 1.xx Contents 1 Introducing the Pipette Check application... 3 2 important notes... 3 3 Selecting the
More informationGAS SYSTEM HYDRA GS. The gas installation system that sets new standards. Witzenmann GmbH
GAS SYSTEM The gas installation system that sets new standards. Witzenmann GmbH Östliche Karl-Friedrich-Str. 134 75175 Pforzheim, Germany Phone +49 7231 581-0 Fax +49 7231 581-820 wi@witzenmann.com www.witzenmann.com
More informationPQ Performance Qualification - Page 1/13
PQ Performance Qualification - Page 1/13 Customer: Location of installation: Item number: (manual) The PQ consists of inspections of the correct operation of the cabinet under predefined conditions and
More informationLevel switches Series LD Vibrating fork level switch for liquids and solids
Level switches Series LD Vibrating fork level switch for liquids and solids Robust and compact Suitable for liquids (model LD61) and solids (model LD60) No moving parts, low maintenance Corrosion resistant
More informationLevel switches Series LD Vibrating fork level switch for liquids and solids
Level switches Series LD Vibrating fork level switch for liquids and solids Robust and compact Suitable for liquids (model LD61) and solids (model LD60) No moving parts, low maintenance Corrosion resistant
More informationValidation of Measurements from a ZephIR Lidar
Validation of Measurements from a ZephIR Lidar Peter Argyle, Simon Watson CREST, Loughborough University, Loughborough, United Kingdom p.argyle@lboro.ac.uk INTRODUCTION Wind farm construction projects
More informationRoboCupJunior Soccer Rules for SuperTeams 2016
RoboCupJunior Soccer Rules for SuperTeams 2016 Soccer Technical Committee 2016: Paul Manns pfmanns@googlemail.com (Germany) Gio Spina profspina@gmail.com (Canada) Katsumi Masaoka k mac@sea.plala.or.jp
More informationRoboCupJunior Soccer Rules for Big Field SuperTeams 2014
RoboCupJunior Soccer Rules for Big Field SuperTeams 2014 RoboCupJunior Soccer Technical Committee 2014: Katsumi Masaoka (Japan), CHAIR, k-mac@sea.plala.or.jp Luis José López (Mexico), llopez@tecnovanguardia.com
More informationSite Summary. Wind Resource Summary. Wind Resource Assessment For King Cove Date Last Modified: 8/6/2013 By: Rich Stromberg & Holly Ganser
Site Summary Wind Resource Assessment For King Cove Date Last Modified: 8/6/2013 By: Rich Stromberg & Holly Ganser Station ID: 2857 Latitude: 55 7 45.8 N Longitude: 162 16 10.6 W Tower Type: 30 m NRG Tall
More informationApplication Notes. SLP85xD Load Cells
Application Notes Load Cells Table of Contents 1 Introduction 3 2 Description of the Filling Cycle 4 3 Filling Optimization 7 4 Filling Monitor 8 4.1 Weight-Based Filling Monitor... 8 4.2 Time-Based Filling
More informationIsofrigo. Cold rooms in Verona. Detail of the internal side
Cold rooms in Verona 126 Detail of the internal side EXAMPLES OF PROJECTS SPECIAL PRODUCTS 127 Legend pag. 18 Self-supporting metal faced panel, insulated with polyurethane, extremely versatile and easy
More information7000 Series OPTIMASS Mass Flowmeters. Titanium Hastelloy Stainless Steel
KROHNE 04/2003 7.02445.71.00 GR 7000 Series OPTIMASS Mass Flowmeters with single straight measuring tube Titanium Hastelloy Stainless Steel One tube no limits No limits with tube material No limits with
More informationAN LFPAK MOSFET thermal design guide - Part 2. Document information
Rev. 2 16 November 11 Application note Document information Info Keywords Abstract Content LFPAK, MOSFET, thermal analysis, design and performance, thermal considerations, thermal resistance, thermal vias,
More informationEN1279 part 3 assessment of insulating glass: IGU with Duralite sealant
TNO Quality Services BV TNO report De Rondom 1 P.O. Box 6235 5600 HE Eindhoven The Netherlands TQS-RAP-07-271 EN1279 part 3 assessment of insulating glass: IGU with Duralite sealant Date February 7, 2007
More informationph measuring system for hygienic process applications
ph measuring system for hygienic process applications Sterile design, CIP-compatible, in-line sterilizable Extremely robust and unbreakable construction Extremely long service life Long calibration intervals
More informationNEW: CO2 incubator with sterilizable sensor
CB series - NEW 160 C2 Incubators NEW: C2 incubator with sterilizable sensor The new CB series offers optimal growth conditions for cell cultures. The ANTI.PLENUM concept, the 180 C sterilization routine
More informationSTAD Series. Balancing and control valves. Technical Data Sheet. WattsIndustries.com
STAD Series Balancing and control valves Technical Data Sheet WattsIndustries.com Description STAD Series variable-orifice threaded valves are designed for flow control and monitoring in climate control
More informationThis test shall be carried out on all vehicles equipped with open type traction batteries.
5.4. Determination of hydrogen emissions page 1 RESS-6-15 5.4.1. This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4.2. The test shall be conducted following
More informationThe following drawings are concerned of this measure, which will be considered in the next revision of the documents as well as in LogiKal.
Based on an much easier assembly of the sealing frame insulating block on fixed glazings the sealing frame has been increased from 7 mm to 9 mm in the system FRAME + 75 WI. This measure has no impact on
More informationMass Split of Two-Phase-Flow
Mass Split of Two-Phase-Flow The split of two-phase-flow at horizontal side-t-junctions in unbalanced pipe systems for clean extinguishing agents Speaker: Gudrun Fay Minimax 2010 2013 Content 1. Applications
More informationStandard Operating Procedure (SOP) SOP-WP5-002/Version 1.0
Page 1 of 12 (SOP) SOP-WP5-002/Version 1.0 Title: Manual blood processing Validity: valid after: valid until: insert end date Replaced versions : insert previous version Modification note: insert modifications
More informationTechnical Data Sheet MF010-O-LC
Technical Data Sheet MF010-O-LC - 1 - 1. Properties The oxygen measuring system MF010-O-LC determines the oxygen content in gas mixtures up to a temperature of 250 C. It is particularly suitable for the
More informationThe next generation of Blood and Fluid Warming
The next generation of One of the contributing factors of unintended hypothermia is the intravenous (IV) delivery of cold blood and fluids, resulting in a reduction of the main body temperature by 0.25ºC
More informationMIL-STD-883G METHOD
STEADY-STATE LIFE 1. PURPOSE. The steady-state life test is performed for the purpose of demonstrating the quality or reliability of devices subjected to the specified conditions over an extended time
More informationInnovation Electrical Industry
Position Paper Potential for Optimising the Use of SF6 in Switchgear Automotive Industry 4.0 Energy Efficiency Digitalisation 4.0 Innovation Electrical Industry 16 November 2016 ZVEI German Electrical
More informationContinuous Gas Analysis In situ laser gas analyzers TÜV and MCERTS add-on for LDS 6 operating instructions Compact Operating Instructions
Continuous Gas Analysis In situ laser gas analyzers Compact Operating Instructions 1 LDS 6 TÜV/MCERTS certified analyzers for NH3/H2O, NH3 and H2O 1.1 Operation requirements Please observe the following:
More informationOperating manual Analogue refractometer
KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen, Germany E-mail: info@kern-sohn.com Operating manual Analogue refractometer KERN ORA 1 SA Seawater salinity ORA 2 SA NaCL% ORA 3 SA NACL% Brix% Tel: +49-[0]7433-9933-0
More informationVESSELS. DIAPHRAGM OR BLADDER TANKS For expansion and pressure boosting applications. Temperature range : -10 to C* Operating pressure :
operating limits Temperature range : -10 to + C* Operating pressure : to 20 bar* * according to models. VESSES IAPHRAGM OR BAER TANKS For expansion and pressure boosting applications AdVANTAGES APPICATIONS
More informationDTG - LCD Digital Temperature Gauge USER MANUAL
DTG - LCD USER MANUAL Page 1 of 7 USERS GUIDE page 1. Description 3 2. Function 3 3. Safety Instruction 3 3.1. Safety Conventions 3 3.2. Proper Use 3 4. Installation 4 4.1. Unpacking 4 4.2. Storage 4 4.3.
More informationDeadweight Tester / Pressure Balance High pressure up to 5,000 bar / 70,000 psi CPB5000-HP. Applications. Special features. Description. Rel.
High up to 5,000 bar / 70,000 psi CPB5000-HP Rel. 20150206 Applications Primary standard for defining the scale in a range up to 5,000 bar Reference instrument for testing, adjusting and calibrating measuring
More informationRoboCupJunior Soccer - SuperTeams Rules 2018
RoboCupJunior Soccer - SuperTeams Rules 2018 RoboCupJunior Soccer Technical Committee Marek Šuppa (Slovakia) CHAIR Felipe Nascimento (The Netherlands) James Riley (Australia) Javier Delgado (México) Michael
More informationK V A L I F I K. Our accreditation number: NAH /2015. Valid to:
KVALIFIK Ltd. has founded in 1991. In 1993 our calibration laboratory has get the accredited status and we have calibrated standards for the generally used physical parameters. Our standards are traceable
More informationOxford Instruments Plasma Technology. PlasmaPro NGP80. Installation Data Sheet. Original Instructions.
Oxford Instruments Plasma Technology www.oxford-instruments.com About this data sheet This data sheet provides basic information about the installation of a tool. Floor and wall loadings Table 1 lists
More informationPQ Performance Qualification - Page 1/13
PQ Performance Qualification - Page 1/13 Customer: Location of installation: Item number: (manual) The PQ consists of inspections of the correct operation of the cabinet under predefined conditions and
More informationMS-500. User Manual. MS-500 (500g x 0.1g) Copyright 2010 Fast Weigh Scales All rights reserved. Rev. 2.0
User Manual MS-500 (500g x 0.1g) Copyright 2010 Fast Weigh Scales All rights reserved. Rev. 2.0 Thank you for purchasing the Fast Weigh MS-500 digital pocket scale. Please read all operating instructions
More informationSpeed of opening, stability, descent rate
Speed of opening, stability, descent rate Steerable Yes Maximum weight in flight (1) [kg]: 125 Weight (2) [kg] 1430 volume packed [cm 3 ]: 4900 TR12F080100032 Test data (3) Test no. 1 Test no. 2 Place
More informationNEW: CO2 incubator with sterilizable sensor
CB series - NEW 160 C2 incubators NEW: C2 incubator with sterilizable sensor The new CB series offers optimal growth conditions for cell cultures. The ANTI.PLENUM concept, the 180 C sterilization routine
More informationPredicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010
Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600 November 2010 Table of Contents ``Abstract...3 ``Introduction...4 ``Methods & Results...6 ``Data Analysis...9 ``Conclusion...12
More informationExperiment 13: Make-Up Lab for 1408/1420
Experiment 13: Make-Up Lab for 1408/1420 This is only for those that have approval. Students without approval will not be allowed to perform the lab. The pre-lab must be turned in at the beginning of lab.
More informationCountry report. Swedish bathing water quality in Sweden. May Photo: Peter Kristensen/EEA
Country report Swedish bathing water quality in 2017 Sweden May 2018 Photo: Peter Kristensen/EEA BWD Report For the Bathing Season 2017 Sweden The report gives a general overview of information acquired
More informationMSDS (MATERIAL SAFETY DATA SHEET) north sea BUTANE. MSDS Nr: BOC(A) Page: 1/5
MSDS Nr: 302-25-2023BOC(A) Page: 1/5 1 IDENTIFICATION OF THE SUBSTANCE/PREPARATION AND OF THE COMPANY Product name Whip-it! North Sea Premium Butane Company identification see heading and/or footer Emergency
More information3-way control valves VK 3
Data sheet 3-way control valves VK 3 Description / Application VK 3 valves are a range of 3 port high specification flanged valves for chilled water, LPHW, MPHW, HPHW (low, medium or high pressure hot
More informationHTR Systems and Components
IAEA Course on HTR Technology Beijing, 22-26.October 2012 HTR Systems and Components Dr. Gerd Brinkmann Dieter Vanvor AREVA NP GMBH Henry-Dunant-Strasse 50 91058 Erlangen phone +49 9131 900 96840/95821
More informationConfidence Interval Notes Calculating Confidence Intervals
Confidence Interval Notes Calculating Confidence Intervals Calculating One-Population Mean Confidence Intervals for Quantitative Data It is always best to use a computer program to make these calculations,
More informationOperating instruction for the quick-change tap holders type:
type: HF 20 HF 30 Date of edition: 15.07.2008 Stage of alteration: 4 Please keep this for future use! Contents: 1 Application range, safety instructions and technical data... 3 1.1 Application range,
More informationMonitoring Gas Pressure in Vacuum Insulation Panels
Gas Pressure in VIP Monitoring Gas Pressure in Vacuum Insulation Panels Dr. Roland Caps, va-q-tec AG, Karl-Ferdinand-Braun-Str. 7, D-978 Würzburg, Germany, www.va-qtec.com, caps@va-q-tec.com, Hermann Beyrichen,
More informationE. Agu, M. Kasperski Ruhr-University Bochum Department of Civil and Environmental Engineering Sciences
EACWE 5 Florence, Italy 19 th 23 rd July 29 Flying Sphere image Museo Ideale L. Da Vinci Chasing gust fronts - wind measurements at the airport Munich, Germany E. Agu, M. Kasperski Ruhr-University Bochum
More informationSTAD. Balancing valves DN (MD60, MD61, MD62)
STAD Balancing valves DN 15-50 (MD60, MD61, MD62) IMI TA / Balancing valves / STAD STAD The STAD balancing valve delivers accurate hydronic performance in an impressive range of applications. Ideally suited
More informationPeak Pressure Indicator MSI-3
Peak Pressure Indicator MSI-3 Operating Instructions Cylinder Pressure Monitoring Digital Pressure Indicator DPI No installation Ready to use High quality sensor Large memory Easy handling Cost effective.com
More informationPROTOCOL FOR COMPRESSED AIR PROFICIENCY TESTING (CAPT) PROGRAM SAMPLE ANALYSIS
Contact Information: PROTOCOL FOR COMPRESSED AIR PROFICIENCY TESTING (CAPT) PROGRAM SAMPLE ANALYSIS No formal enrollment is required to participate in the CAPT program. To get added to the email notifications
More informationOperating Manual Body fat balance
KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen email: info@kern-sohn.com Phone: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operating Manual Body fat balance KERN MFB Version
More informationSLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters
SLE4000 Infant Ventilator with touch-screen operation When the smallest thing matters SLE4000 - The Total Solution for Conventional Infant Ventilation SLE is a world leader in the design and manufacture
More informationExtension and compensating cables, single-paired version - suitable for use in temperature measurement and manufacturing process control
PVC, silicone, FEP or glass fibre-insulated Extension and compensating cables, single-paired version - suitable for use in temperature measurement and manufacturing process control Info Available in many
More informationINSTITUTE AND FACULTY OF ACTUARIES. Curriculum 2019 AUDIT TRAIL
INSTITUTE AND FACULTY OF ACTUARIES Curriculum 2019 AUDIT TRAIL Subject CP2 Actuarial Modelling Paper One Institute and Faculty of Actuaries Triathlon model Objective Each year on the Island of IFoA a Minister
More informationOperation of Compressors Through Use of Load Stands
Purdue University Purdue e-pubs International Compressor Engineering Conference School of Mechanical Engineering 1972 Operation of Compressors Through Use of Load Stands W. R. Thornton Purdue University
More informationRapidmain Installation Guide
Rapidmain Installation Guide Aluminium Compressed Air Pipe Work Installation Guide Trafalgar Court, Waterloo Ind Estate, Widnes Cheshire. UK. Ref: Rapidmain installation guide 4b Index 1. Rapidmain fittings:
More informationCapacity and pressure similar to R-507. Flooded systems. Low temperature (-46.7 C). R22 replacement as long as the condenser has the right size.
TECHNICAL DATA SHEET R-428A () Features and uses of R-428A () R-428A (RS-45) is a non flammable azeotropic blend, with zero ODP, compatible with traditional mineral lubricants, alkyl benzene and also with
More informationSensing and Modeling of Terrain Features using Crawling Robots
Czech Technical University in Prague Sensing and Modeling of Terrain Features using Crawling Robots Jakub Mrva 1 Faculty of Electrical Engineering Agent Technology Center Computational Robotics Laboratory
More informationSTAD-C. Balancing valves DN 15-50
STAD-C Balancing valves 15-50 IMI TA / Balancing valves / STAD-C STAD-C The STAD-C balancing valve has been specially developed for use in indirect cooling systems but performs just as effectively in refrigerated
More informationCO2 incubator with additional process controls
CB series 150 C2 incubators C2 incubator with additional process controls The BINDER CB series C2 incubator is the premium class among the C2 incubators. It is suitable for all sensitive incubation applications
More informationFlamco Airfix D-E. Installation and operating instructions. 1999, Flamco eps
Airfix D-E 6442.eps GB Installation and operating instructions 1999, Exclusion of liability All operation and maintenance related technical information, data and instructions contained in these installation
More informationECL Comfort 110, application 131 (valid as of software version 2.00)
Operating Guide ECL Comfort 110, application 131 (valid as of software version 2.00) English version www.danfoss.com How to navigate? Adjust temperatures and values. Switch between menu lines. Select /
More informationHUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0
High Accuracy Humidity Calibrator The role of humidity calibrations that are accurate, reproducible, and documentable is becoming more and more important. ISO quality guidelines and regulations according
More informationApplication Note AN-107
SPEC Sensor TM Characterization & Calibration Considerations Scope This document is provided to describe the considerations needed to characterize, calibrate, verify and validate the measurement performance
More informationTechnical Information Conducal CLY421
TI00496C/07/EN/01.12 71142150 Products Solutions Services Technical Information Conducal CLY421 Conductivity calibration set for ultrapure water applications Factory calibrated measurement box Certified
More informationFigure 1 Schematic of opposing air bearing concept
Theoretical Analysis of Opposing Air Bearing Concept This concept utilizes air bearings to constrain five degrees of freedom of the optic as shown in the figure below. Three pairs of inherently compensated
More informationThe Agility Association of Canada
Check list to Prepare for Equipment Inspection 2018 Introduction: requires that all equipment used at sanctioned events be inspected and approved at least four months before a host s first trial, and regularly
More informationBaseCal 3, calculating the performance of Diaphragm Seals
BaseCal 3, calculating the performance of Diaphragm Seals BASECAL: THE DIAPHRAGM SEAL PERFORMANCE CALCULATION TOOL BaseCal is the web-based performance calculation tool for diaphragm seal applications,
More informationTESTING BENCHES for watermeters
TESTING BENCHES for watermeters +372 6 480 235 Haljas tee 25, 12012 TALLINN Fax +372 6 480 125 Testing benches are designed by Tepso Ltd. for using in calibration laboratories and also in watermeter production
More information