Biometry tips. Marilyn Watkins-Ramdin, COMT ROUB

Size: px
Start display at page:

Download "Biometry tips. Marilyn Watkins-Ramdin, COMT ROUB"

Transcription

1 Biometry tips Marilyn Watkins-Ramdin, COMT ROUB

2 Types of Biometry Ultrasound (A-Scan) Contact Immersion Laser Interferometry (IOL Master)

3 Patient History Use the oldest possible MRx to include for calculations since myopic shifts up to 10 diopters can be caused by the changes in the lens. Necessary for Holladay 2 formula Ask about prior refractive surgeries: LASIK, PRK these can change K readings and require special calculations

4 Patient History Remember Myopic eyes are usually longer /- 1.23mm Hyperopic eyes are usually shorter /- 0.6m Emmentropic eyes are usually 23.5mm long

5 Patient History con t Retinal surgery: Scleral buckles can elongate the eye mm Pseudophakic patients find out what IOL material was used: PMMA, Silicone or Acrylic Aphakic: IOL setting must be adjusted PK or corneal opacity Slit lamp exam: Ocular surface disease, Eg. Severe dry eyes

6 Patient History con t Ocular history: glaucoma, prior surgery, eye meds Medical history: prostate; leads to floppy iris syndrome if patient on Flomax Social history: job type, social activities, drug use

7 Keratometry Manual Automatic: Autorefractor / IOL Master Pick a method and stick with it Use two methods for K readings to verify Eyes should be within 1D of each other NB: Myopic eyes are long and have flat OA's Hyperopic eyes are short with steep K's Average eye is 23.5mm Average K reading D

8 Keratometry P=(n-1)/r P= power n=standardized keratometric index of refraction usually The index of refraction refers to theoretical single refractive lens with a stable ratio of anterior to posterior corneal curvature

9 Keratometry con t

10 Keratometry con t Manual keratometry considered the Gold standard seldom used due to better automated keratometers Always focus your eye piece first as in lensometry since uncorrected refractive errors of the examiner will result in erroneous measurements

11 Keratometry con t Keep both eyes open when measuring Focus the horizontal meridian first then measure Then refocus mires to measure the vertical axis Much needed in high myopes NB 1D error K s =1D post-op refractive error

12 Keratometry con t CL s wearers: Soft CL wearers should take lenses out for about one week prior to measurements GP and HL should be out until K s are stable NB. Prior to K reading no procedures ointments or drops are to be used that will distort mires 1 drop of anaesthics GTTs can be used to smooth mires Calibrate weekly

13 Topography Topography is used to detect any deformities of the corneal surface e.g. Keratoconus, PK, RK The cornea is provides 70% of the eye s refractive power Keratometry measures the central 3mm of cornea VS Topography measuring entire anterior corneal surface More accurate for corneas << 40D or >> 46D & irregular Best when used in conjunction with Keratometry. An average taken between when K s are disparate much needed when measuring for toric IOL s

14 A-Scans Immersion - You will need scleral shell : Hansen or Prager BSS & anesthetic gtts Position patient as for contact method Have the machine in a position where you can see and easily access screen as for contact Can be done manually or automatically Instill gtts

15 A-Scans Contact vs Immersion Contact Patient must be comfortable and informed of what you are going to do Patient can be sitting or reclining Instill anesthetic gtts Place probe directly on central cornea Align probe along visual axis come off cornea before making any major movements other than tilting

16 A-Scans Contact vs Immersion Immersion Insert shell with probe attached for Prager or in the case of Hansen shell just insert into eye Immerse probe in BSS Align probe along visual axis Lateral movements are used to obtain measurements

17 A-Scans Contd. For either method take 10 readings per eye Immersion is more accurate but difficult at first to learn Messy at first Takes longer at first Contact Corneal compression Corneal abrasion Fluid bridge Dry Eye

18 Sources of Error You may have some situations that require contact so perfect the skill Causes of errors Short measurements Corneal compression Velocity too low Improper gate placement Gain too high Misalignment of sound beams

19 Errors Long measurements Fluid bridge Velocity too high Improper gate placement Gain too low Misalignment of sound beam Improper techniques Poor patient fixation: Strabismus, NLP in one eye Internal vs External fixation Pathology: disciform scar or saphyloma

20 Macula Localizing Technique Find optic nerve Tilt probe to direct sound until scleral spoke shows

21 Laser Interferometry IOL master uses light to measure axial length Advantages Non-contact Very accurate Disadvantages Cannot penetrate dense cataracts Difficulty measuring Pseudophakic eyes due to high degree of reflection

22 Axial Length Measurements Good fixation Good corneal tear film Not a fool proof method Nuclear cataracts may require off-axis measuring technique Dense opacity of posterior capsule may produce error message Ametropia of more than 6D may require patient to wear glasses for best results

23 IOL Master Useful for measuring eye with posterior staphyloma (protrusion of sclera causing an irregular shaped posterior eye) this is because the axial length measurements is required along the axis of patient fixation NB: when using an A-Scan proper alignment is at best estimated with a staphyloma

24 Corneal Curvature Measurements K Reading's Susceptible to the same error factors as the keratometer Poor tear film Corneal irregularly Image obscured by eyelids or eyelashes Improper focusing

25 Anterior Chamber Depth Many of the IOL calculation formulas use ACD as part of calculation IOL master uses a projected slit beam to measure ACD Excessive accommodation by patient can cause divinations, if so measure after dilation ACD will NOT be accurate on a Pseudophakic eye

26 IOL Calculations & Formulas 4 primary pieces of information needed for IOL calculations: A-Scan axial length measurements Average K readings A-constant for IOL to be implanted Desired post-op refraction Formulas: Haigis, HofferQ, Holladay, SRK II, SRK/T Can be linked to Holladay II program and other network system Know which formulas are most appropriate for the measurements you acquired.

27 IOL Calculations & Formulas Both eyes must be measured to identify any potential sources of error IOL power should be within 1D between both eyes Theoretical formulas 1 st Gen Fyodorov, Binkhorst, Holladay Shammas used A-scan ultrasound and K readings to devise these formulas Are all based on the same equation P=n/L-ACD-n x K/n-K x ACD Regression formulas SRK P=A-2.5L-0.9K Gave better results with normal eyes than the theoretical formulas

28 Formulas 2 nd Gen formulas were used to improve the accuracy of IOL calculations for short & long eyes Increase in PIOL vs Iris-plane IOLs a factor SRK altered by combining a linear regression analysis for short and long eyes this became SRKII 3 rd Gen formulas arose after doctors express discomfort with a formula based on an artificial linear model. The authors of SRK devised then SRK/T an new theoretical formula. SRK/T is similar to the Holladay & relies on the ACD constant rather than surgeon factor

29 Formulas 4 th & 5 th Gen formulas are: Holladay,Haigis, Olsen & Barrett Universal II These can use as many as seven variables (keratometry, axial length, anterior chamber depth, lens thickness, horizontal white-to-white measurement, age and pre-op refraction in the case of the Holladay II

30 Formulas To get the most from many of the latestgeneration formulas, you also need to have an optical biometer such as the IOLMaster 700 or the Lenstar. This is because two of the newer formulas, the Holladay II and the Olsen, require a measurement of the lens thickness that these instruments provide

31 Clean & Calibrate IOL Master Clean Chin rest and head rest with Alcohol wipes between patients Once per week calibrate with test eye if available or staff member with stable Rx Ultrasound Machine Soak shells in Hydrogen Peroxide prior and after each patient but rinse with water prior to inserting into the eye Probe tip can be soaked in peroxide or wipe with alcohol Calibrate using a staff member

32 Summary For successful biometry Examiner must have knowledge of proper techniques High quality consistent scans IOL calculations Biometry standards Have a 2nd examiner re-measure patient anytime you are unsure Triple check your work Never stop learning!

33 NB: take home points For successful post-op outcomes take into consideration : The surgeon s expectations The patient s expectations The increasing use of Premium IOLs The world of biometry is heading to optical vs ultrasound with the increasing demands for optimum visual outcomes

34 References/ Resources A-Scan -Axial Eye Length Measurements A Handbook For IOL Calculations- Sandra Frazier Bryne Grove Park Publishers

35 References/ Resources /article/power-calculation-how-to-upyour-game Ultrasound of the Eye & Orbit 2nd Sandra Frazier Bryne Ronald L. Green Mosby Inc ISBN

36 THANK YOU!

Ophthalmology Clinical Pearls A Quick Guide To Crystalens

Ophthalmology Clinical Pearls A Quick Guide To Crystalens Ophthalmology Clinical Pearls A Quick Guide To Crystalens Crystalens Accommodating Posterior Chamber Intraocular Lens BRIEF STATEMENT Rx only. Indications for Use: The Crystalens is intended for primary

More information

LENSTAR LS 900 Setting the standard in optical biometry

LENSTAR LS 900 Setting the standard in optical biometry LENSTAR LS 900 Setting the standard in optical biometry 02 03 LENSTAR BIOMETER LENSTAR Better clinical outcomes The LENSTAR LS 900 provides multiple highly-accurate laser optic measurements of nine sections

More information

Scleral Fitting Guide

Scleral Fitting Guide Scleral Fitting Guide Scleral Lens Design...3 Fitting Philosophy...4 Select a Diagnostic Lens...5 Initial Lens Evaluation...6 Assess Scleral Landing Zone (SLZ), Sagittal Depth, Limbal Clearance, Lens Power...7-10

More information

Getting Started GUIDE LEAVE A LEGACY OF EXCELLENT OUTCOMES FOR PATIENTS WITH ASTIGMATISM. TECNIS TORIC 1-PIECE IOL pg 1

Getting Started GUIDE LEAVE A LEGACY OF EXCELLENT OUTCOMES FOR PATIENTS WITH ASTIGMATISM. TECNIS TORIC 1-PIECE IOL pg 1 Getting Started GUIDE LEAVE A LEGACY OF EXCELLENT OUTCOMES FOR PATIENTS WITH ASTIGMATISM. TECNIS TORIC 1-PIECE IOL pg 1 GETTING STARTED WITH THE TECNIS TORIC IOL The IOL you leave behind is the quality

More information

Sofmed breathables XW (comfilcon A) Sofmed breathables XW premium (comfilcon A) Sofmed breathables XW toric (comfilcon A)

Sofmed breathables XW (comfilcon A) Sofmed breathables XW premium (comfilcon A) Sofmed breathables XW toric (comfilcon A) Sofmed breathables XW (comfilcon A) Sofmed breathables XW premium (comfilcon A) Sofmed breathables XW toric (comfilcon A) Sofmed breathables XW multifocal (comfilcon A) SOFT (HYDROPHILIC) CONTACT LENSES

More information

Appendix C: Skill Stations Examination - Marking Rubric

Appendix C: Skill Stations Examination - Marking Rubric Appendix C: Skill Stations Examination - Marking Rubric This document provides the assessment sheets which are used by the examiners for the Skills Station examination component of the OCANZ Competency

More information

TONOMETRY AND PACHYMETRY

TONOMETRY AND PACHYMETRY TONOMETRY AND PACHYMETRY Caroline Pate, OD Associate Professor UAB School of Optometry Birmingham, AL Intraocular Pressure (IOP) The fluid pressure inside the eye The pressure created by the continual

More information

LENS EXCHANGE WITH IMPLANTATION OF A BITORIC INTRAOCULAR LENS IN EYES WITH SIGNIFICANT ASTIGMATISM AFTER PREVIOUS CORNEAL SURGERIES

LENS EXCHANGE WITH IMPLANTATION OF A BITORIC INTRAOCULAR LENS IN EYES WITH SIGNIFICANT ASTIGMATISM AFTER PREVIOUS CORNEAL SURGERIES LENS EXCHANGE WITH IMPLANTATION OF A BITORIC INTRAOCULAR LENS IN EYES WITH SIGNIFICANT ASTIGMATISM AFTER PREVIOUS CORNEAL SURGERIES Dimitrii Dementiev, MD, Anna Shipunova, MD, PhD Irina Ivanchikova, MD,

More information

Acknowledgments. Paragon Vision Sciences Adoption of CRT certification slide for this generic talk. Material Considerations

Acknowledgments. Paragon Vision Sciences Adoption of CRT certification slide for this generic talk. Material Considerations Corneal Reshaping Jeffrey J. Walline, OD PhD The Ohio State University College of Optometry Acknowledgments Paragon Vision Sciences Adoption of CRT certification slide for this generic talk Corneal Refractive

More information

LASIK Wavefront Guided idesign LASIK with IntraLase

LASIK Wavefront Guided idesign LASIK with IntraLase LASIK Wavefront Guided idesign LASIK with IntraLase What is LASIK? LASIK (Laser Assisted In-Situ Keratomileusis), commonly referred to as laser eye surgery or laser vision correction is a safe High Definition

More information

Pre- and postoperative refraction after cataract extraction with implantation of standard power IOL

Pre- and postoperative refraction after cataract extraction with implantation of standard power IOL British Journal of Ophthalmology, 1988, 72, 231-235 Pre- and postoperative refraction after cataract extraction with implantation of standard power IOL THOMAS OLSEN From the Department of Ophthalmology,

More information

Orthokeratology. What is Orthokeratology? How does it work?

Orthokeratology. What is Orthokeratology? How does it work? What is? The goal of, OrthoK or corneal refractive therapy is to use a rigid gas permeable contact lens to reshape the cornea and provides correction for refractive errors. is a non-surgical procedure

More information

Appendix C: Skill Stations Examination - Marking Rubric

Appendix C: Skill Stations Examination - Marking Rubric Appendix C: Skill Stations Examination - Marking Rubric This document provides the assessment sheets which are used by the examiners for the Skills Station examination component of the OCANZ Competency

More information

POLYMACON HYDROPHILIC CONTACT LENSES PRACTITIONER FITTING GUIDE

POLYMACON HYDROPHILIC CONTACT LENSES PRACTITIONER FITTING GUIDE POLYMACON HYDROPHILIC CONTACT LENSES PRACTITIONER FITTING GUIDE Part Number: PFG01005 Page 1 of 10 Table of Contents INTRODUCTION:... 3 MATERIAL:... 3 DESIGN:... 4 AVAILABLE LENS PARAMETERS... 4 INDICATIONS:...

More information

Dr. Thomas R. Reim President, DreimLens, Inc.

Dr. Thomas R. Reim President, DreimLens, Inc. DreamLens Fitting Guide Dr. Thomas R. Reim President, DreimLens, Inc. The introduction of the DreamLens in 1997 has revolutionized orthokeratology and now made it a viable procedure for the general population.

More information

An introduction to modern scleral lenses

An introduction to modern scleral lenses An introduction to modern scleral lenses The subject of scleral lenses seems to divide eye care practitioners more than just about any other type of contact lens. Views range considerably-some believing

More information

How images are formed in the eye

How images are formed in the eye Refractive Surgery How images are formed in the eye Rays of light enter the eye (Figure 1) through its front surface, which is called the cornea (A), go through the hole in the iris, which is called the

More information

EVERY TREATMENT IS A TRUE ORIGINAL

EVERY TREATMENT IS A TRUE ORIGINAL EVERY TREATMENT IS A TRUE ORIGINAL : The STAR S4 IR Excimer Laser System and the idesign Refractive Studio is indicated for wavefront-guided laser assisted in situ keratomileusis () to achieve monovision

More information

The BE Retainer System Certification Course. Supplement

The BE Retainer System Certification Course. Supplement The System Certification Course Supplement The BE Retainer System Certification Course The System for Vision Shaping Treatment Certification Course Welcome to the System Certification Course. This handout

More information

V E J A R A N O P R O T O C O L FOR T H E M A N A G M E N T O F Q U E R A T O C O N O SURGICAL T I PS

V E J A R A N O P R O T O C O L FOR T H E M A N A G M E N T O F Q U E R A T O C O N O SURGICAL T I PS V E J A R A N O P R O T O C O L FOR T H E M A N A G M E N T O F Q U E R A T O C O N O SURGICAL T I PS SURGICAL TIPS, 5 STEPS TO FOLLOW: To achieve excellent results in the management of comprehensive treatment

More information

Pardon the Objection (PTO) Contact Lens Clinical Case Debate (2 hours)

Pardon the Objection (PTO) Contact Lens Clinical Case Debate (2 hours) Pardon the Objection (PTO) Contact Lens Clinical Case Debate (2 hours) Mile Brujic, O.D. Melissa Barnett, O.D. Glenda Secor, O.D. Jeff Sonsino, O.D. Summary The contact lens options that we have continue

More information

FITTING AND TROUBLESHOOTING SCLERAL LENSES

FITTING AND TROUBLESHOOTING SCLERAL LENSES FITTING AND TROUBLESHOOTING SCLERAL LENSES ANDREW J. BIONDO, OD, FSLS SCLERAL LENS ADVANTAGES Comfort Vision Centration Stability Health Moisture Easy to fit 1 WHO S A GOOD CANDIDATE Any irregular cornea

More information

Fitting Guide. Minutes-to-Fit 2.0. by SynergEyes Moving Vision Forward

Fitting Guide. Minutes-to-Fit 2.0. by SynergEyes Moving Vision Forward Fitting Guide Minutes-to-Fit 2.0 by SynergEyes Moving Vision Forward Minutes-to-Fit 2.0 Everyday success made easy! Thank you for making Duette HD, by SynergEyes, a part of your contact lens practice.

More information

Basic Concept & Techniques of Phaco Prechop

Basic Concept & Techniques of Phaco Prechop Basic Concept & Techniques of Phaco Prechop Mitsui Memorial Hospital Takayuki Akahoshi, MD The author has no financial interest in the products shown in this presentation Divide & Conquer U/S Energy Loss

More information

Financial Disclosures Toric IOL Selec-on and Posi-oning

Financial Disclosures Toric IOL Selec-on and Posi-oning Financial Disclosures Toric IOL Selec-on and Posi-oning Abbo$ Medical Op/cs Inc. L Guided by Intraopera-ve Aberrometry Allergan, Inc. - C, L Kathryn M. Hatch, M.D. Assistant Clinical Professor of Ophthalmology,

More information

NAVAL SUBMARINE MEDICAL RESEARCH LABORATORY

NAVAL SUBMARINE MEDICAL RESEARCH LABORATORY NAVAL SUBMARINE MEDICAL RESEARCH LABORATORY SUBMARINE BASE, GROTON, CONN. REPORT NO. 1102 EFFECTS OF HYPERBARIC CONDITIONS ON CORNEAL PHYSIOLOGY WITH HYDROGEL CONTACT LENSES by Joseph F. Molinari and James

More information

NormalEyes Guidelines For Successful Fitting RIGID GAS PERMEABLE SCLERAL CONTACT LENSES. Manufactured in Paragon HDS 1OO (paflufocon D)

NormalEyes Guidelines For Successful Fitting RIGID GAS PERMEABLE SCLERAL CONTACT LENSES. Manufactured in Paragon HDS 1OO (paflufocon D) NormalEyes 15.5 Guidelines For Successful Fitting RIGID GAS PERMEABLE SCLERAL CONTACT LENSES Manufactured in Paragon HDS 1OO (paflufocon D) FOR DAILY WEAR PARAGON VISION SCIENCES Just watch us 2013 Paragon

More information

Alberta Police Recruit Selection Standards

Alberta Police Recruit Selection Standards Alberta Police Recruit Selection Standards SUPPLEMENTAL VISION REPORT FORMS For Police Constable Applicants Alberta Solicitor General and Public Security Police Recruit Selection Standards Corneal Refractive

More information

THE TOP 100 KEYWORDS FOR AN

THE TOP 100 KEYWORDS FOR AN THE TOP 100 KEYWORDS FOR AN OPHTHALMOLOGY PRACTICE Introduction to Search Engine Optimization for Ophthalmology Search Engine Optimization, also known as SEO, is an online marketing technique that aims

More information

Injector system for implantation of 1stQ foldable intraocular lenses

Injector system for implantation of 1stQ foldable intraocular lenses Instruction For Use 1st Inject Instrument Injector system for implantation of 1stQ foldable intraocular lenses The IFU is available electronically on our website: www.1stq.eu Content: The package contains

More information

International Council of Ophthalmology s Ophthalmology Surgical Competency Assessment Rubric (ICO-OSCAR) ICO-OSCAR Instructor Directions

International Council of Ophthalmology s Ophthalmology Surgical Competency Assessment Rubric (ICO-OSCAR) ICO-OSCAR Instructor Directions International Council of Ophthalmology s Ophthalmology Surgical Competency Assessment Rubric (ICO-OSCAR) The International Council of Ophthalmology s Ophthalmology Surgical Competency Assessment Rubrics

More information

Learn Connect Succeed. JCAHPO Regional Meetings 2017

Learn Connect Succeed. JCAHPO Regional Meetings 2017 Learn Connect Succeed JCAHPO Regional Meetings 2017 Refractometry: Knowledge and Skills for Today s Technicians Matthew L. Parker, PhD, LSSMBB, COMT, CST, OSC U.S. Army (Retired) I have absolutely no financial

More information

DISCLOSURES FITTING AND TROUBLESHOOTING SCLERAL LENSES WHO S A GOOD CANDIDATE SCLERAL LENS ADVANTAGES ANDREW J. BIONDO, OD, FSLS.

DISCLOSURES FITTING AND TROUBLESHOOTING SCLERAL LENSES WHO S A GOOD CANDIDATE SCLERAL LENS ADVANTAGES ANDREW J. BIONDO, OD, FSLS. FITTING AND TROUBLESHOOTING SCLERAL LENSES ANDREW J. BIONDO, OD, FSLS DISCLOSURES Disclosures: Paid consultant and lecturer for X Cel Specialty Contacts Paid consultant for SynergEyes Remote Education

More information

In November 1998, the Staar AA4203 toric intraocular. Early rotational stability of the longer Staar toric intraocular lens. Fifty consecutive cases

In November 1998, the Staar AA4203 toric intraocular. Early rotational stability of the longer Staar toric intraocular lens. Fifty consecutive cases Early rotational stability of the longer Staar toric intraocular lens Fifty consecutive cases David F. Chang, MD Purpose: To determine whether rotational stability is improved with the longer Staar toric

More information

Procedures for ETDRS 4M Refraction and Vision Testing. Katherine M. Burke, BS, CO, COMT CertifEYED Associates, LLC New York, New York

Procedures for ETDRS 4M Refraction and Vision Testing. Katherine M. Burke, BS, CO, COMT CertifEYED Associates, LLC New York, New York Procedures for ETDRS 4M Refraction and Vision Testing Katherine M. Burke, BS, CO, COMT Associates, LLC New York, New York K. Burke, FEB2017 Hello and thank you for joining this important meeting Consistency

More information

REFRACTIVE ERRORS in. CORRECTION of OPTICAL DIVERS USING OPTICAL MASKS

REFRACTIVE ERRORS in. CORRECTION of OPTICAL DIVERS USING OPTICAL MASKS Features CORRECTION of OPTICAL REFRACTIVE ERRORS in DIVERS USING OPTICAL MASKS photo by Marco A. Donnini part one INTRODUCTION Good vision underwater is an important safety factor in diving. More than

More information

New Highly Productive Phased Array Ultrasonic Testing Machine for Aluminium Plates for Aircraft Applications

New Highly Productive Phased Array Ultrasonic Testing Machine for Aluminium Plates for Aircraft Applications 19 th World Conference on Non-Destructive Testing 2016 New Highly Productive Phased Array Ultrasonic Testing Machine for Aluminium Plates for Aircraft Applications Christoph HENKEL 1, Markus SPERL 1, Walter

More information

PACKAGE INSERT RIGID GAS PERMEABLE CONTACT LENSES FOR ORTHOKERATOLOGY. (hexafocon B)

PACKAGE INSERT RIGID GAS PERMEABLE CONTACT LENSES FOR ORTHOKERATOLOGY. (hexafocon B) PACKAGE INSERT BOSTON XO 2 (hexafocon B) RIGID GAS PERMEABLE CONTACT LENSES FOR ORTHOKERATOLOGY Daily Wear IMPORTANT: Please read carefully and keep this information for future use. This package insert

More information

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13.1.1 TRAINING GOAL 13.1.1.1 Among the responsabilities of a medical

More information

Phacoemulsification and foldable posterior chamber intraocular lenses in aircrew: Clinical behaviour and aeromedical implications

Phacoemulsification and foldable posterior chamber intraocular lenses in aircrew: Clinical behaviour and aeromedical implications Original Article Phacoemulsification and foldable posterior chamber intraocular lenses in aircrew: Clinical behaviour and aeromedical implications Sqn Ldr NK Tripathy *, Wg Cdr GKG Prasad VSM (Retd) +

More information

Automatic level / Optical level / Dumpy Level Instruction Manual www.level1lasers.com.au Contents Introduction 2 Diagram 3 Operation Instruction 4 Calibration Self-Check 5 Technical Specifications 6 Maintenance

More information

Training Guide STAR S4 IR Excimer Laser

Training Guide STAR S4 IR Excimer Laser Training Guide STAR S4 IR Excimer Laser 1 P age TABLE OF CONTENTS Safety....3 Hardware.....5 System Start-up and Calibration......9 Software... 16 Environmental Conditions...24 System Shutdown.. 24 Maintenance......24

More information

Management of corneal wounds: some practical tips. Lee SH, MD, FRCS (C ), FAAO

Management of corneal wounds: some practical tips. Lee SH, MD, FRCS (C ), FAAO Clinical practice Management of corneal wounds: some practical tips Lee SH, MD, FRCS (C ), FAAO Abstract Corneal trauma is one of the leading causes of ocular morbidity. This paper aims at providing some

More information

Understanding How the Appearance of Optical Fiber Splices Relates to Splice Quality

Understanding How the Appearance of Optical Fiber Splices Relates to Splice Quality Understanding How the Appearance of Optical Fiber Splices Relates to Splice Quality Douglas Duke & David Mansperger Fusion Splicing Systems, AFL, Duncan, SC Doug.Duke@AFLglobal.com David.Mansperger@AFLglobal.com

More information

Practical method to calculate post-lasik corneal power: the Actual K a+p method

Practical method to calculate post-lasik corneal power: the Actual K a+p method 陨灶贼允韵责澡贼澡葬造皂燥造熏灾燥造援 3 熏晕燥援 4 袁 Dec.18, 圆园 10 www. IJO. cn 窑 Clinical Research 窑 Practical method to calculate post-lasik corneal power: the Actual K a+p method Foundation item: Shanghai Leading Academic

More information

Operating manual Analogue refractometer

Operating manual Analogue refractometer KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen, Germany E-mail: info@kern-sohn.com Operating manual Analogue refractometer KERN ORA 1 SA Seawater salinity ORA 2 SA NaCL% ORA 3 SA NACL% Brix% Tel: +49-[0]7433-9933-0

More information

Spherical Soft Contact Lens Fitting. J. Perrigin OPTO 6374 Spring 2017

Spherical Soft Contact Lens Fitting. J. Perrigin OPTO 6374 Spring 2017 Spherical Soft Contact Lens Fitting J. Perrigin OPTO 6374 Spring 2017 Reading Assignment Bennett & Henry: Chapter 11 pages 270-286 Tables 11.1 &11.2 Some Advantages of SCLs versus GPs Initial comfort,

More information

INTRASTROMAL CORRECTION OF PRESBYOPIA AND CLINICAL APPLICATION WITH TECHNOLAS FEMTOSECOND LASER

INTRASTROMAL CORRECTION OF PRESBYOPIA AND CLINICAL APPLICATION WITH TECHNOLAS FEMTOSECOND LASER INTRASTROMAL CORRECTION OF PRESBYOPIA AND CLINICAL APPLICATION WITH TECHNOLAS FEMTOSECOND LASER Nuernberg, June 19th 2009 LUIS ANTONIO RUIZ Bogota-Colombia The INTRACOR Procedure Eye Fixation with Minimal

More information

Andre Litwin BM FRCOphth Consultant Ophthalmologist and Oculoplastic Surgeon

Andre Litwin BM FRCOphth Consultant Ophthalmologist and Oculoplastic Surgeon Andre Litwin BM FRCOphth Consultant Ophthalmologist and Oculoplastic Surgeon Queen Victoria Hospital NHS Foundation Trust, The McIndoe Surgical Centre, Brighton Nuffield Hospital, Holtye Road, East Grinstead.

More information

Reusable Ophthalmic Instruments

Reusable Ophthalmic Instruments Reusable Ophthalmic Instruments Quality Reusable Instruments from Surgitrac l High quality range of precision reusable ophthalmic instruments l Made from Japanese medical grade Stainless Steel l Rigerous

More information

Sterile, Single-use, Instruments

Sterile, Single-use, Instruments Sterile, Single-use, Instruments Surgitrac instruments, single-use ophthalmic and refractive instruments. Single-use instruments look and feel like a reusable, but offer so much more: Safer: sterile, ready-to-use,

More information

TEST FOR STABILOMETER VALUE OF BITUMINOUS MIXTURES

TEST FOR STABILOMETER VALUE OF BITUMINOUS MIXTURES Test Procedure for TEST FOR STABILOMETER VALUE OF BITUMINOUS MIXTURES TxDOT Designation: Tex-208-F Effective Date: February 2005 1. SCOPE 1.1 Use this test method to determine the Hveem stability value

More information

Calculating Total Uncertainty of Temperature Calibration with a Dry Block. Calibration White Paper. Beamex.

Calculating Total Uncertainty of Temperature Calibration with a Dry Block. Calibration White Paper. Beamex. Beamex Calibration White Paper info@beamex.com Calculating Total Uncertainty of Temperature Calibration with a Dry Block World-class calibration solutions Calculating Total Uncertainty of Temperature Calibration

More information

Ocular hazards of playing squash

Ocular hazards of playing squash Brit. J. Ophthal. (I973) 57, 434 Ocular hazards of playing squash rackets D. V. INGRAM AND I. LEWKONIA Moorfields Eye Hospital, High Holborn, London, W.C.i Squash rackets is a game played at a fast pace,

More information

Ophthalmic Surgical Instruments

Ophthalmic Surgical Instruments Ophthalmic Surgical Instruments Welcome to Optico. Optico is a small, entrepreneurial company with our manufacturing facility in Hertfordshire, England. Since our formation we have seen strong growth and

More information

International Council of Ophthalmology s Ophthalmology Surgical Competency Assessment Rubric (ICO-OSCAR)

International Council of Ophthalmology s Ophthalmology Surgical Competency Assessment Rubric (ICO-OSCAR) International Council of Ophthalmology s Ophthalmology Surgical Competency Assessment Rubric (ICO-OSCAR) The International Council of Ophthalmology s Ophthalmology Surgical Competency Assessment Rubrics

More information

Triple EK/CE/IOL has a cataract needs both DMEK and CEIOL

Triple EK/CE/IOL has a cataract needs both DMEK and CEIOL 1 DMEK COURSE Advanced Surgical Techniques and Surgical Pearls ASCRS 2017 Los Angeles, CA Instructor: Thomas John, MD Faculty: Francis Price, MD,, 9002 N Meridian St, # 100, Indianapolis, Indiana 46260-5349

More information

Level MEASUREMENT 1/2016

Level MEASUREMENT 1/2016 Level MEASUREMENT 1/2016 AGENDA 2 A. Introduction B. Float method C. Displacer method D. Hydrostatic pressure method E. Capacitance method G. Ultrasonic method H. Radar method I. Laser method J. Level

More information

CLINICAL SCIENCES. Michael Goggin, FRCSI(Ophth), MS; Sacha Moore, FRCOphth; Adrian Esterman, AStat, PhD

CLINICAL SCIENCES. Michael Goggin, FRCSI(Ophth), MS; Sacha Moore, FRCOphth; Adrian Esterman, AStat, PhD CLINICAL SCIENCES Outcome of Toric Intraocular Lens Implantation After Adjusting for Anterior Chamber Depth and Intraocular Lens Sphere Equivalent Power Effects Michael Goggin, FRCSI(Ophth), MS; Sacha

More information

OPESAVER Super Irrigation System

OPESAVER Super Irrigation System OPESAVER Super Irrigation System Makoto Kishimoto, MD Dokkyo University School of Medicine Kishimoto Eye Clinic Moriyama City Shiga, Japan n ABSTRACT OPESERVER reduces the surge phenomenon during ultrasonic

More information

485 Annubar Primary Flow Element Installation Effects

485 Annubar Primary Flow Element Installation Effects ROSEMOUNT 485 ANNUBAR 485 Annubar Primary Flow Element Installation Effects CONTENTS Mounting hole diameter Alignment error Piping Geometry Induced Flow Disturbances Pipe reducers and expansions Control

More information

PROFESSIONAL FITTING AND INFORMATION GUIDE. Paragon CRT Manufactured in Paragon HDS (paflufocon B)

PROFESSIONAL FITTING AND INFORMATION GUIDE. Paragon CRT Manufactured in Paragon HDS (paflufocon B) PROFESSIONAL FITTING AND INFORMATION GUIDE Paragon CRT Manufactured in Paragon HDS (paflufocon B) or Paragon CRT 100 Manufactured in Paragon HDS 1OO (paflufocon D) RIGID GAS PERMEABLE CONTACT LENSES FOR

More information

Operating manual Analogue Brix Refractometer

Operating manual Analogue Brix Refractometer KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen, Germany E-mail: info@kern-sohn.com Operating manual Analogue Brix Refractometer Tel: +49-[0]7433-9933-0 Fax: +49 (0)7433-9933-149 Internet: www.kern-sohn.com

More information

RECOMMENDER REQUIRED PRE-AUTH GROUP A PRE-AUTH GROUP B

RECOMMENDER REQUIRED PRE-AUTH GROUP A PRE-AUTH GROUP B DESCRIPTION CONTACT LENS FITTING PROCEDURES INCLUDING INITIAL ISSUE - SPHERICAL 602560 01-Jan-14 O, OP $700.00/2 CY N SEE NOTE 5 CONTACT LENS FITTING PROCEDURES INCLUDING INITIAL ISSUE - TORIC 602156 01-Jan-14

More information

Figure 1 Schematic of opposing air bearing concept

Figure 1 Schematic of opposing air bearing concept Theoretical Analysis of Opposing Air Bearing Concept This concept utilizes air bearings to constrain five degrees of freedom of the optic as shown in the figure below. Three pairs of inherently compensated

More information

FL M FLUME FOR HYDRAULIC BENCH

FL M FLUME FOR HYDRAULIC BENCH This equipment is designed to study the behavior of fluids in open channels by conducting a wide range of practices and experiences. HIGHLIGHTS Possibility of negative and positive channel slope. Various

More information

UsER manual for Watersens ph -REDOX

UsER manual for Watersens ph -REDOX UsER manual for Watersens -REDOX Cl 8 1 2 6 3 3 7 7 4 4 4 4 Parts List 1 Redox Probe 1 x 2 PH Probe 1 x 5 Tube Weight 2 x 6 Connection Valve 1 x chlorine 3 Chlorine and Pumps 2 x 7 Dosing Valve 2 x 5 5

More information

UNITED STATES MARINE CORPS WEAPONS TRAINING BATTALION MARINE CORPS COMBAT DEVELOPMENT COMMAND QUANTICO, VIRGINIA

UNITED STATES MARINE CORPS WEAPONS TRAINING BATTALION MARINE CORPS COMBAT DEVELOPMENT COMMAND QUANTICO, VIRGINIA UNITED STATES MARINE CORPS WEAPONS TRAINING BATTALION MARINE CORPS COMBAT DEVELOPMENT COMMAND QUANTICO, VIRGINIA 22134-5040 DETAILED INSTRUCTOR GUIDE LESSON TITLE FUNDAMENTALS OF RIFLE MARKSMANSHIP COURSE

More information

James T. Schwiegerling CURRICULUM VITAE

James T. Schwiegerling CURRICULUM VITAE James T. Schwiegerling CURRICULUM VITAE Chronology of Education B.S. Optics, University of Rochester, 1990. M.S. Optics, University of Rochester, 1991. Ph.D. Optical Sciences, University of Arizona, 1995.

More information

ASICO Innovative I/A Tip System (Patented) Maximize your I/A efficiency for Today s MICS and FLACS procedure

ASICO Innovative I/A Tip System (Patented) Maximize your I/A efficiency for Today s MICS and FLACS procedure Call us today for special pricing on all I/A tip models Surgical Notes ASICO Innovative I/A Tip System (Patented) Jan 2019 Maximize your I/A efficiency for Today s MICS and FLACS procedure 0.3mm port Call

More information

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1 BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1 The front door of the main body of the instrument may be open when you arrive. Take the sample chamber and slide it into position with the

More information

Inspection of CANDU Reactor Pressure Tubes Using Ultrasonics

Inspection of CANDU Reactor Pressure Tubes Using Ultrasonics 17th World Conference on Nondestructive Testing, 25-28 Oct 2008, Shanghai, China Inspection of CANDU Reactor Pressure Tubes Using Ultrasonics Michael TRELINSKI Inspection & Maintenance Services Ontario

More information

LaserSnake Underwater Laser Cutting Phase 1 TWI

LaserSnake Underwater Laser Cutting Phase 1 TWI LaserSnake2 110128 Underwater Laser Cutting Phase 1 TWI-002 6-13 Contents 1 Introduction 1 2 Objectives 1 3 Underwater Cutting Nozzle Design and Operation 1 3.1 Design 1 3.2 Dry-zone testing 3 4 Equipment

More information

Ophthalmic Surgical Instruments

Ophthalmic Surgical Instruments Ophthalmic Surgical Instruments Welcome to Optico. Optico is a small, entrepreneurial company with our manufacturing facility in Hertfordshire, England. Since our formation we have seen strong growth

More information

Aerodrome Design Manual

Aerodrome Design Manual Doc 9157 AN/901 Aerodrome Design Manual Part 4 Visual Aids Approved by the Secretary General and published under his authority Fourth Edition 2004 International Civil Aviation Organization 8-12 Aerodrome

More information

Outline Chapter 7 Waves

Outline Chapter 7 Waves Outline Chapter 7 Waves 7-1. Water Waves 7-2. Transverse and Longitudinal Waves 7-3. Describing Waves 7-4. Standing Waves 7-5. Sound 7-6. Doppler Effect 7-7. Musical Sounds 7-8. Electromagnetic Waves 7-9.

More information

BIORAPTOR Knotless Suture Anchor Arthroscopic Shoulder Instability Repair. Shoulder Series Technique Guide by: Thomas Youm, MD

BIORAPTOR Knotless Suture Anchor Arthroscopic Shoulder Instability Repair. Shoulder Series Technique Guide by: Thomas Youm, MD BIORAPTOR Knotless Suture Anchor Arthroscopic Shoulder Instability Repair Shoulder Series Technique Guide by: Thomas Youm, MD As described by: Thomas Youm, MD Clinical Assistant Professor NYU Hospital

More information

Single-use Ophthalmic Instruments

Single-use Ophthalmic Instruments Single-use Ophthalmic Instruments Forceps Scissors Surgitrac instruments, single-use Ophthalmic instruments. Single-use instruments look and feel like a reusable, but offer so much more: Safer: sterile,

More information

CATALOG CATALOG. Pioneer in Ophthalmic Products. BACK FLUSH INSTRUMENTS VITREORETINAL INSTRUMENTS

CATALOG CATALOG. Pioneer in Ophthalmic Products.    BACK FLUSH INSTRUMENTS VITREORETINAL INSTRUMENTS Pioneer in Ophthalmic Products VITREORETINAL INSTRUMENTS 23G 25G 23G TROCAR SYSTEM BACK FLUSH INSTRUMENTS CATALOG CATALOG AN ISO 9001 & ISO 3485 CERTIFIED COMPANY VITREORETINAL INSTRUMENTS www.ovationint.com

More information

25+ instruments answer the most common barrier to 25G instrument flexibility.

25+ instruments answer the most common barrier to 25G instrument flexibility. Fine Membranes Heavy Membranes ILM Forceps End-grasping design ideal for maculorhexis or other fine membrane grasping Tip style improves visualization Multi-purpose grasping applications GRIESHABER MAXGRIP

More information

Single-use Ophthalmic Instruments

Single-use Ophthalmic Instruments Single-use Ophthalmic Instruments Surgitrac instruments, single-use Ophthalmic instruments. Single-use instruments look and feel like a reusable, but offer so much more: Safer: sterile, ready-to-use, single-use

More information

Keeler Digital Applanation Tonometer (D-KAT)

Keeler Digital Applanation Tonometer (D-KAT) Keeler Digital Applanation Tonometer (D-KAT) Instructions for use Home Back Next Contents 1. Introduction 1.1 Brief description of the instrument 1.2 Intended use / purpose of instrument 1.3 How the intraocular

More information

Instrumentation & Data Acquisition Systems

Instrumentation & Data Acquisition Systems Instrumentation & Data Acquisition Systems Section 3 -Level Robert W. Harrison, PE Bob@TheHarrisonHouse.com Made in USA 1 Level Section Question Which level measuring technology is the best solution when

More information

Waves. Kevin Small or

Waves. Kevin Small   or Waves Opening note: X-rays can penetrate your body. Sound waves can make thinks vibrate; water waves can knock you over in the sea. Infrared waves can warm you up and slinky waves are fun to play with.

More information

AE3610 Experiments in Fluid and Solid Mechanics OPTICAL/PRESSURE MEASUREMENTS IN SUPERSONIC FLOW

AE3610 Experiments in Fluid and Solid Mechanics OPTICAL/PRESSURE MEASUREMENTS IN SUPERSONIC FLOW AE3610 Experiments in Fluid and Solid Mechanics OPTICAL/PRESSURE MEASUREMENTS IN SUPERSONIC FLOW Objective This lab covers the use of taps, probes and transducers for measurement of static and stagnation

More information

AllPro Foot. Posterior Mount Product Manual

AllPro Foot. Posterior Mount Product Manual AllPro Foot Posterior Mount Product Manual Instructions The AllPro PM foot system has been designed and manufactured for specific patient weights. Failure to follow the weight guidelines and/or overload

More information

PAPER 2 THEORY QUESTIONS

PAPER 2 THEORY QUESTIONS PAPER 2 THEORY QUESTIONS 1 (a) Water waves are transverse waves. Sound is a longitudinal wave. (i) Describe the difference between transverse waves and longitudinal waves. In your account, draw a diagram

More information

Insert to. Sponsored by Alcon Laboratories, Inc. February 2014 EXPLORING THE CATARACT REFRACTIVE SUITE FEATURING. The CENTURION Vision System

Insert to. Sponsored by Alcon Laboratories, Inc. February 2014 EXPLORING THE CATARACT REFRACTIVE SUITE FEATURING. The CENTURION Vision System Insert to Sponsored by Alcon Laboratories, Inc. February 2014 EXPLORING THE CATARACT REFRACTIVE SUITE FEATURING The CENTURION Vision System The Centurion Vision System The CENTURION Vision System: The

More information

LESSON 9 THE SWING PLANE AND THE BODY (Part 1)

LESSON 9 THE SWING PLANE AND THE BODY (Part 1) LESSON 9 THE SWING PLANE AND THE BODY (Part 1) A BLOW IN PLANE FACE ON. In our very first lesson we understood the principle of striking a fence post with a hammer; we could make a glancing blow, but the

More information

CONTROL VALVE TESTING

CONTROL VALVE TESTING The optimal functioning of the Control valve not only exists of sufficient body & seat tightness, but more important, the total "performance" of the valve and its controls! For an accurate and reliable

More information

Oxygen tension under a contact lens. Kenneth A. Poise and Marianne Decker

Oxygen tension under a contact lens. Kenneth A. Poise and Marianne Decker Oxygen tension under a contact lens Kenneth A. Poise and Marianne Decker Cornea! thickness changes were measured on human subjects who wore gel lenses that varied in center thickness. Using these measurements

More information

Post-Standardisation

Post-Standardisation Version General Certificate of Education (A-level) Applied January 011 Applied Science SC08 (Specification 8771/877/8776/8777/8779) Unit 8: Medical Physics Post-Standardisation Mark Scheme Mark schemes

More information

Aeris Activity. Product Manual

Aeris Activity. Product Manual Aeris Activity Product Manual Instructions The Aeris Activity Foot System has been designed and manufactured for specific patient weights. Failure to follow the weight guidelines and/or overload conditions

More information

Measurement of court speed and bounce. Rod Cross, Physics Department, Sydney University 2nd July 2006

Measurement of court speed and bounce. Rod Cross, Physics Department, Sydney University 2nd July 2006 1 Measurement of court speed and bounce Rod Cross, Physics Department, Sydney University 2nd July 2006 The speed of a surface refers to the change in ball speed in a direction parallel to the surface.

More information

Underwater measurement of photosynthetically active radiation

Underwater measurement of photosynthetically active radiation Underwater measurement of photosynthetically active radiation Mark Blonquist, Damon Nitzel, and Bruce Bugbee Apogee Instruments, Inc., Logan, Utah Introduction Quantum sensors (photosynthetically active

More information

SMILE for low Myopia

SMILE for low Myopia SMILE for low Myopia Dr. Rainer Wiltfang North American Refractive Laser Symposium November 11-13, 2016 Smile for low Myopia Scientific literature clinical outcomes Smile vs. Lasik clinical outcomes Smile

More information

OP CHECKLIST FOR 1D CONSOLIDATION LABORATORY TEST

OP CHECKLIST FOR 1D CONSOLIDATION LABORATORY TEST Page 1 of 5 WORK INSTRUCTIONS FOR ENGINEERS NHB Compiled by : LSS Checked by : GSS Approved by : OP-3-31. CHECKLIST FOR 1D CONSOLIDATION LABORATORY TEST Page 2 of 5 31.0 CHECKLIST ITEMS *(refer to respective

More information

CHAPTER 7 : SMOKE METERS AND THEIR INSTALLATIONS

CHAPTER 7 : SMOKE METERS AND THEIR INSTALLATIONS CHAPTER 7 : SMOKE METERS AND THEIR INSTALLATIONS 1 Scope : This Chapter covers the requirements of smoke meters and their installation on engines for full load and free acceleration tests, mentioned in

More information

Rotator Cuff Solution. Combined Strengths. Precise Control. TWINFIX Ultra FOOTPRINT Ultra Suture Anchors

Rotator Cuff Solution. Combined Strengths. Precise Control. TWINFIX Ultra FOOTPRINT Ultra Suture Anchors Rotator Cuff Solution R Combined Strengths. Precise Control. TWINFIX Ultra FOOTPRINT Ultra Suture Anchors Trusted materials Absorbable Our absorbable HA suture anchor combines poly-l-lactic acid (PLLA)

More information