The effect of changes of lung characteristics on the functioning of automatic ventilators*

Size: px
Start display at page:

Download "The effect of changes of lung characteristics on the functioning of automatic ventilators*"

Transcription

1 VOL 17 NO 3 ANBSTHESA JULY 1962 The effect of changes of lung characteristics on the functioning of automatic ventilators* To study the effect of changes of lung characteristics on the functioning of automatic ventilators it is first necessary to understand the functioning of the various types of ventilator when connected to a normal lung. A set of concepts, with which it is possible to give a systematic account of the functioning of a ventilator, has already been expoundedl; but certain of these concepts are not very well understood. Therefore some of them will be considered here in rather more detail. GENERAL An automatic ventilator has to do four things: 1 nflate the patient s lungs 2 Change over from inspiration to expiration 3 Deflate the lungs - or permit passive expiration 4 Change over from expiration to inspiration t seems that the change-over, or cycling, processes are better understood than the processes of inflation or deflation. For instance, it is well-known that the change-over from inspiration to expiration can be either time-cycled, pressure-cycled or volume-cycled ; that is to say the process of inflation is stopped either after some preset time has elapsed, or as soon as some preset pressure is reached at the mouth (or at some other point), or when some preset volume has been delivered to the patient. n the same way the change-over from expiration to inspiration can be time-, pressure- or volume-cycled; in addition there is here the possibility of patient-cycling. What seems to be much less well understood is what goes on during the inspiratory phase. n some *Paper read at the Scientific Meeting of the Faculty of Anaesthetists of The Royal College of Surgeons of England, 12th May

2 ANBSTHESA 301 ventilators there may be equal confusion about what goes on during the expiratory phase, but in so many of them this is just a matter of allowing the patient s lungs to deflate passively to atmospheric pressure that it need not be considered further here. Therefore attention can be concentrated upon the inspiratory phase. During the inspiratory phase there are four factors to be considered : 1 Flowrate of gas into the lungs 2 Volume of gas in the lungs 3 Pressure in the alveoli 4 Pressure at the mouth For a complete understanding, the way in which these four factors vary from moment to moment throughout the inspiratory phase must be determined. These four factors are all inter-related in a rigorous cause-and-effect manner and the inter-relationships are determined entirely by the characteristics of the patient s lungs. Therefore only one of these four factors is free to be independently determined by the ventilator. n fact it is always either the flow into the patient s lungs or the pressure at the patient s mouth which is determined by the ventilator. Once one or other of these two factors has been determined by the ventilator the other three follow from the effect of the one factor on the characteristics of the patient s lungs. f it is the flowrate into the lungs which is determined by the ventilator then the ventilator may be called a flow generator; if it is the pressure applied to the mouth which is determined by the ventilator then it may be called a pressure generator. t seems to be possible to allot all ventilators to these two groups. FLOW GENERATORS: MECHANSMS The discussion will be restricted to the case of a constant-flow generator. A ventilator which operates as a constant-flow generator produces a flowrate into the patient s lungs which is constant throughout the inspiratory phase and furthermore, which is entirely determined by the ventilator and is uninfluenced by the patient s lung characteristics. There may be a knob on the ventilator to adjust the flowrate but once this is set, no matter what the lung characteristics, this constant flow will be driven into the patient for as long as the inspiratory phase lasts, that is until it is terminated by the cycling mechanism - unless some pressure-limit comes into operation earlier due, perhaps, to the opening of a safety valve. Once the ventilator has determined the flowrate into the lungs the other three factors listed above follow according to the lung characteristics.

3 302 ANESTHESA A number of mechanisms operate as constant-flow generators. n one the patient s lungs are connected more or less directly to an air compressor driven by a powerful constant-speed motor. (This system is found in the Aintree, Blease and Fazakerley* ventilators). This compressor goes on pumping out the same steady flow of gas throughout the inspiratory phase no matter how high the back pressure from the patient may rise. Another arrangement which operates as a constant-flow generator is shown in FG. 1. The reducing valve, on the top of the compressed- L 5-60 p.s.i cm RE S STANCE J. TO PATENT m S G R d cm H20 FG. 1 A constant-flow generator gas cylinder, delivers a steady pressure somewhere between 5 and 60 lb/in2 (between 300 and 4000cm H20). This is connected through a high resistance to the patient. The pressure in the patient s lungs rises during inflation from 0 to perhaps 30cm H2O - but this is a very small change compared to the 300 to 4000cm H20 at the reducing valve. Therefore the pressure difference across the high resistance is substantially constant throughout the phase and therefore the flow is also constant. t may be adjusted to any value up to say 100 litres/min by altering the magnitude of the high resistance, but once it is set the flow stays constant throughout every inspiratory phase. This system is found in the Pneumotron and Williams ventilators and also in some modes of operation of the Bird. *Descriptions of all the ventilators mentioned in this paper (apart from the Cyclator and Takaoka) will be found in Automatic Ventilation of the Lungs*

4 ANAESTHESA 303 Another mechanism which comes very near to being a constant-flow generator is a highly efficient injector such as is found in the Cyclator and also, though perhaps to a lesser extent, in the Aga and Drager ventilators. PRESSURE GENERATORS : MECHANSMS The discussion will be restricted to the case of a constant-pressure generator. Here it is the pressure at the mouth which is determined (and held constant) by the ventilator while the flowrate of gas and the volume and pressure in the alveoli result from the effect of this pressure on the patient s lung characteristics. (For other types of pressure generator and for an account of what happens when the pressure is held constant at some point separated from the mouth by a substantial resistance see Mushin et ap). One form of constant-pressure generator is shown in FG. 2. The weight provides a constant force acting on the constant crosssectional area of the bellows. Force per unit area is pressure so the pressure within the bellows is constant no matter whether the outlet is occluded and no flow occurs or whether the outlet is wide open and the air comes out in a rush. n use the bellows are connected to the patient s lungs FG. 2 A constant-pressure generator so that the lungs are inflated until the pressure $thin them equals the pressure in the bellows - unless the inspiratory phase is stopped beforehand. This system is found in the RPR and Radcliffe ventilators. Another form of constant-pressure generator is the second-stage reducing valve found in the Bennett ventilator. THE EFFECT OF THE VENTLATOR OUTPUT ON THE LUNGS Constant -30 w generators FG. 3 shows in graphical form the constant flow produced by a constant-flow generator and the resultant variation in the volume and pressures in the lungs. Numerical values have been indicated for the sake of illustration.

5 304 ANESTHESA At the start of the inspiratory phase the flow into the lungs rises suddenly to some value (30 litres/min in FG. 3) and stays there throughout the phase. Therefore (assuming there areno leaks in the system) the volume in the lungs increases steadily - at 30 litres/min=+ litre/sec and therefore reaches 500ml at the end of 1 sec. The compliance of the lungs may be assumed constant throughout the phase and therefore the steady increase in volume is accompanied by a steady increase in the pressure in the alveoli (PALV). f the total compliance of a normal anasthetised patient is taken to be 50ml/cm H20 then the steady increase in volume to 500ml entails a steady increase in alveolar pressure to locm H2O. The airway resistance is constant throughout the inspiratory phase and therefore the constant flow produces a steady pressure difference (PKPALv) across it. f the total airway resistance (including connectors) for a normal intubated patient is taken to be 6cm HzO per litre/sec the constant flow of 3 litre/sec will produce a constant pressure difference across it, i.e. between the mouth and the alveoli (PM- PaLv), of 3cm H2O. The pressure at the mouth is simply the sum of the pressure in the alveoli and the pressure difference between the mouth and the alveoli: it rises suddenly to 3cm H20 due to the pressure difference across the airway resistance and then rises steadily through a further locm H20 in 1 sec due to the steady rise in pressure in the alveoli. FLOW 30 ( /m in) 0 TME (sec) 0' 1 FG. 3 Flow, volume and pressure curves in the lungs for a constant-flow generator (30l/min) connected to a patient with a compliance of 50 ml/cm H2O and an airway resistance of 6 cm H2O per 1 /sec. PAL" =pressure in the alveoli, P,=pressure at the mouth. Constant-pressure generators A constant-pressure generator applies some pressure to the mouth at the start of the inspiratory phase and holds it constant throughout the phase. Suppose the generated pressure is 12cm H20 : PMoUTH= 12cm HzO

6 ANESTHESA 305 Suppose also that the total (intubated) airway resistance is as before: R = 6cm H20 per litre/sec At the start of inspiration, there is no pressure in the alveoli and therefore the whole of the 12cm H20 pressure at the mouth is dropped across the airway resistance. Therefore since the airway resistance is 6cm H2O for 1 litre/sec a pressure drop of 12cm H20 will produce a flow of 2 litres/sec: nitial flow into lungs = 2 litres/sec Therefore nitial volume increase in lungs = 200ml in 1 / 10 sec f the compliance is 50ml/cm H20, as before, a volume increase of 200ml will involve a pressure increase of 4cm H20. Therefore nitial pressure increase in alveoli = 4cm H20 in 1/10 sec. But as the pressure in the alveoli increases the pressure difference between the mouth and the alveoli decreases (because the mouth pressure is held constant by the ventilator). t is this pressure difference which provides the force driving gas into the lungs; therefore as it decreases, so the flow into the lungs decreases and the rate at which the volume in the lungs increases becomes less and the rate at which the pressure in the alveoli increases becomes less : however the alveolar pressure does go on increasing, but the closer it gets to the mouth pressure the slower it increases (FG. 4). From a mathematical point of view equilibrium will occur only after an infinite time; but from a clinical point of view there would be no advantage (and possibly some disadvantage) in prolonging the inspiratory phase to get the last few per cent of pressure or volume into the lungs. t is therefore important to know how long it takes to achieve 'near equilibrium' - say 90 to 99 % equilibrium. There are only three independent factors within the system (of ventilator plus patient) on which the time to reach a given degree of equilibrium could possibly depend : 1 Compliance 2 Resistance 3 Generated pressure VOLUME / rnl FG. 4 Pressure, flow and volume curves in the lungs for a constant-pressure generator (1 2 cm H2O) connected to a patient with a compliance of 50 ml/cm H20 and an airway resistance of 6 cm H20 per l/sec.

7 306 ANRSTHESA All the other variables within the system are derived from the effect of the generated pressure on the compliance and resistance of the patient's lungs. n fact the generated pressure has no influence on the time taken to reach a given degree of equilibrium so that only the compliance and resistance can be involved*. How can these parameters affect the time for equilibration? Consider the following calculation in which compliance and resistance are given the same values as before. C = 0.05 litre/cm H20, litre - - 0'05cm H20 ' R = 6cm H20 per L/sec, 6 cm H20 sec - litre litre cm H20 sec Therefore C x R = 0.05 x 6 X 9 cm H20 litre = 0.3 sec. The product of the compliance and airway resistance of a normal, ansesthetised, intubated patient is 0.3 sec. t seems odd that a lung should have a time, but the significance of this time can be seen from the following calculation. Assume that, as in the previous calculation for a constant-pressure generator: P,,,, = 12cm HzO nitial pressure increase in lungs = 4cm H20 in 1/10 sec. Now C x R = 0.3 = 3/10 sec. Therefore if initial rate of pressure increase in lungs were maintained for 3/10 sec the alveolar pressure would reach 3 x 4 = 12cm H20. Therefore this time of 3/10 sec, the time which is equal to the product of the compliance and resistance of the patient's lungs, is the time in which the inflation of the lungs wouzd be complete if the initial rate of inflation were maintained. n fact, as has already been shown, the initial rate of inflation is not maintained and therefore inflation is not complete at the end of this time. *n a more accurate analysis it would be necessary to take account of the way in which compliance wries with the volume in the lungs and resistance varies with flowrate so that the size of the generated pressure would have some indirect influence on the time taken to achieve a given degree of equilibrium. For the present discussion however the compliance and resistance may be regarded as independent of volume and flow, and hence of generated pressure, which then indeed has no influence on the time to achieve any given degree of equilibration.

8 ANESTHESA 307 This process of gradual approach towards equilibrium in which the rate of approach diminishes in proportion to the nearness to which the limit is approached is not peculiar to the process of inflating lungs but occurs in such diverse fields as the charging of an electrical condenser through a resistor or the decay of a radioactive substance. n these fields of knowledge the characteristic time derived above (the time in which the change would be complete if the initial rate of change were maintained) is given the name time constant and is usually represented by the Greek letter T (tau). Thus at the end of a period of time T (equal to the product of the compliance and resistance of the lungs) the lungs would be fully inflated if the initial rate of inflation were maintained. n fact the rate of inflation falls off as the alveolar pressure rises (FG. 5) and at the end PRESSURE MOUTH FG. 5 ncrease of alveolar pressure with time (in units of time constant, T) for a r - constant-pressure generator. The very thin broken line shows that the alveolar / pressure would reach the mouth pressure after one time constant if the initial rate of increase were maintained. CT T+ of one time constant the change is only about 2/3 complete - or 1/3 incomplete. n the period of the second time constant the same thing happens again (in a sense) : the pressure difference between mouth and alveoli is reduced to approximately 1/3 of what it was at the beginning of the period - and therefore to 1/3 of 1/3 = 1/9 of what it was at the start of the inspiratory phase. At the end of three time constants the remaining pressure difference is about 1/3 of 1/3 of 1/3 = 1/27 of what it was at the start of the inspiratory phase. n fact, at the end of a period of one time constant, the fraction of change which is incomplete is not exactly 1/3 but where e is the base 1 of natural logarithms and is equal to = 37 %. Therefore at the end of one time constant the change is 37 % incomplete and 63 % 1 1 complete. At the end of two time constants it is- x -= 13 % incomplete and 87 % complete. At the end of three time constants it is 5 % incomplete and 95 % complete. These results are summarised and extended in TABLE 1. They are true no matter what the lung character-

9 308 ANESTHESA Table 1 Exponential ncrease of Alveolar Pressure in a Lung Connected to a Constant-pressure Generator. Time is given in units of T where T is the time constant of the lungs (the product of compliance and airway resistance). Time T 2i 3T 4T istics may be, in the sense that, for instance, inflation will always be 95 complete in a time which is three times as great as that obtained by multiplying the compliance and airway resistance together. For the normal, anasthetised, intubated patient, for whom the time constant is 0.3 sec, inflation will be 95% complete in 3 x 0.3 = 0.9 sec. Thus in these circumstances there will be little point in prolonging the inspiratory phase beyond 1 sec*. The foregoing has established the patterns of flow, volume and pressure produced in normal lungs by constant-flow generators and constant-pressure generators. t is now possible to go on to consider how these patterns are modified by changes in lung characteristics. THE EFFECT OF CHANGES OF LUNG CHARACTERSTCS Constant-pressure generator FGURE 6 shows the alveolar pressure and volume curves, plotted against time, for a constant-pressure generator connected to a normal lung (continuous curves in all three pairs of graphs) and to lungs in which the compliance has been halved, the airway resistance doubled, and in which both changes have occurred simultaneously (broken curves). The blobs on the curves mark off intervals of one time constant. When the compliance is halved (perhaps due to the surgeon leaning on the chest) the most obvious effect is that the final steady volume is halved. But halving the compliance also halves the product of compliance and resistance - the time constant. Therefore it takes only half as long to reach any given degree of equilibration between the alveolar pressure and the mouth pressure. This can be seen by studying the blobs on the curve in FG. 6: at the end of one time constant (the first blob) both the continuous and the broken curves have reached 63 % of the applied pressure but the time constant is only half as long for *f apparatus resistance is in series with the airway resistance leading to an increased time constant, or if a constant-flow generator is used, then prolonging the inspiratory phase beyond 1 sec will indeed lead to further inflation of the lungs.

10 PALV tr ""if- ANESTHESA 309 Rx2 FG. 6 The pressure and volume in the alveoli plotted against time for a constant-pressure generator. The continuous curves are for a normal lung and the broken curves apply when compliance and airway resistance deviate from normal in the manner indicated. The blobs on the curves mark off intervals equal to one time constant. the broken line as for the continuous line because of the halved compliance. Although the final steady volume in the lungs is halved because of the halved compliance, this lower volume is approached more quickly because of the halved time constant. f instead of the compliance being halved the resistance is doubled (perhaps due to an obstruction or the accumulation of secretions) the ultimate pressure or volume in the alveoli is unaltered but because the time constant is doubled (by the doubling of the resistance) it takes twice as long to reach any given degree of equilibration. f the compliance is halved and the resistance is doubled at the same time (one main bronchus occluded perhaps) then the time constant (the product of the two) is unaltered. Therefore the alveolar pressure rises in exactly the same way as before. The final steady volume in the

11 3 10 ANESTHESA lungs is halved (by the halved compliance) but this lower volume is approached at the same rate as that at which the full volume is approached in the normal lung - because the time constant is the same. The foregoing explains the changes in the volume and pressure patterns produced by a constant-pressure generator when the lung characteristics change in certain ways; but the clinician is more interested in the resultant changes in tidal volume and ventilation. To determine these it is necessary to take account not only of what sort of generator the ventilator is working as, but also what sort of cycling mechanism is in operation. For instance if the constant-pressure generator is time-cycled (e.g. the Radcliffe) and the patient s compliance is halved, either with or without a doubling of airway resistance, (the top two sets of curves in FG. 6) then the tidal volume will certainly be reduced, probably down to half what it was before. f the change-over from expiration to inspiration is similarly time-cycled then the duration of the respiratory cycle, and hence the respiratory rate, will be fixed. Therefore the total ventilation will be reduced in the same proportion as the tidal volume. f the change in lung characteristics consists only of an increase of airway resistance (lower set of curves in FG. 6) then the effect on tidal volume and ventilation depends largely on the length of the inspiratory phase set by the time-cycling mechanism. f this is long (beyond the end of the lower graph in FG. 6) then the volume in the lungs will have near enough reached its equilibrium value, even with the doubled time constant consequent upon the doubled airway resistance. Therefore the change in airway resistance has little or no effect on the tidal volume. However, if the time-cycling mechanism is set to give rather a short inspiratory phase - say one-third of the way along the time axis of the lower set of graphs in FG. 6 (that is two normal time constants or one double time constant) - then doubling the airway resistance results in a substantial reduction in tidal volume. f instead of being time-cycled the constant-pressure generator is volume-cycled (as in the French RP R ventilator) then it is clear from FG. 6 that any decrease in compliance or increase in airway resistance will result in the ventilator taking a longer time to deliver the preset tidal volume. Therefore, unless there is a compensating decrease in expiratory time the rate and hence the total ventilation will be decreased. ndeed, if the cycling mechanism is set to operate at a volume which is greater than half the final equilibrium volume for a normal lung, and if the compliance is then halved (with or without an increase in airway resistance) the cycling volume will never be achieved and the rate, and therefore the total ventilation, will drop to zero.

12 FLOW - -- ANiESTHESA 311 palv! / E 0..<.. Rx2 0 p 1 / / / k /'..!..... Cx+ Rx2 FG. 7 Flow, volume and pressure curves in the lungs for a constant-flow generator when connected to a normal lung (continuous lines) or when connected to lungs with the indicated changes from normality (broken lines) of compliance and airway resistance. Constant-Jow generator FGURE 7 shows the flow, volume and pressure patterns produced by a constant-flow generator when connected to a normal lung (the continuous lines in all three sets of graphs - which are the same as shown in FG. 3) and also when connected to lungs in which the compliance has been halved, or the airway resistance doubled, or both changes have occurred simultaneously (the broken lines in the three successive sets of graphs). Changes of lung characteristics can have no effect on the fl ow from the ventilator, nor on the pattern of volume increase in the lungs

13 312 ANRSTHESA since, assuming there are no leaks, the volume is simply the integral of the flow into the lungs. However, when the compliance is halved (broken lines in the left-hand set of graphs in FG. 7) the resultant pressure rise in the alveoli is doubled. Therefore the pressure at the mouth, after a normal initial step (due to the pressure difference across the normal airway resistance), increases at twice the normal rate due to the halved compliance. When the airway resistance is doubled (centre set of graphs) there is no change in the alveolar pressure curve but the pressure difference between the mouth and alveoli (due to the constant flow through the doubled airway resistance) is doubled. Therefore the pressure at the mouth has a doubled initial step but then rises at a normal rate. When the compliance is halved and the airway resistance doubled at the same time (broken lines in the right-hand set of graphs in FG. 7) the two effects occur simultaneously. Therefore the pressure at the mouth shows both a doubled initial step and a doubled subsequent rate of rise. As with the constant-pressure generator it is necessary to take account of the cycling mechanisms to determine the effect of these changes in waveform, consequent upon changes in lung characteristics, on the tidal volume and ventilation. f a constant-flow generator is time-cycled (e.g. the Pneumotron) then the constant flow for a preset time will deliver a fixed tidal volume despite any changes in lung characteristics. Assuming there is no change in the expiratory time, the rate will stay the same and so the total ventilation wil be unaltered. The same will also be true of a constant-flow generator which is volume-cycled (e.g. the Greer and Donald) since the constant flow will take a fixed time to deliver the preset tidal volume and, providing the expiratory time is fixed, the rate and total ventilation will again be independent of changes in lung characteristics. f a constant-flow generator is pressure-cycled the situation is very different. Suppose that when connected to a normal lung the cycling pressure is set so that the inspiratory phase lasts just as long as the graphs in FG. 7 so that the tidal volume delivered is that indicated by the ends of the volume curves. n that case the cycling pressure must be that indicated by the dotted lines in FG. 7. Therefore when the compliance is reduced or the airway resistance increased or when both occur simultaneously the cycling pressure is reached much earlier and therefore the tidal volume is decreased. To take a specific instance : if the compliance is halved the inspiratory time is halved and the tidal volume is halved. f the expiratory time is fixed (as in the Aintree, Blease, Cyclator and Fazakerley) then the complete respiratory cycle will be only slightly shortened and the halving of the tidal volume will

14 ANBSTHESA 313 result in a large reduction of total ventilation. However, if when the compliance is halved, it is found that the halved tidal volume is not only inflated in half the time, but also expired in half the time (as it will be if the ventilator is a pressure-cycled constant-flow generator in expiration as well as in inspiration e.g. approximately, the Aga, Drager Poliomat and Takaoka ventilators) then, although the tidal volume is halved the rate is doubled and the total ventilation is unchanged. SUMMARY AND CONCLUSONS This paper does not give a comprehensive account of the functioning of all types of ventilator in all circumstances but illustrates a method of analysis by describing the most important aspects of the problem in detail. Thus only the inspiratory phase has been considered and even within that phase the many types of non-constant-flow and nonconstant-pressure generators have been omitted. Likewise no mention has been made of the effects of pressure and volume limits or of various other devices which modify the functioning of some ventilators. A more comprehensive, but much more concise, account will be found in Automatic Ventilation of the Lungsl. The present treatment is intended to give a fuller and more leisurely account of some of the fundamental aspects of the problem in the hope that it will lead to a greater understanding which can then be applied to other circumstances not covered here. CONSTANT-FLOW However, it is convenient to sum- GENERATOR marize the results that have been TME CYC. o o 0 TDAL NSF! TOTAL VOL. TME VENT, + + +or0 derived in this account and this is O O O done in FG. 8. An important con- CONST~-PRESSURE clusion which can be drawn from GENERATOR this Figure is that it is dangerous to. ; ; ;,. ; : 0 4 try to relate the responses of a t +or0 ventilator to changes in lung char- FG. 8 The effect of decreased compliacteristics to a single aspect of the ance or increased resistance on the performance of various types of ventilator s functioning. For inventilator (assuming no leaks from stance, a volume-cycled ventilator the system). is traditionally regarded as immune from changes in lung characteristics; it is true that the tidal volume is held constant by this means, but if the ventilator operates as a pressure-generator then the inspiratory time is increased and, unless there is a compensating decrease in expiratory time, the rate and therefore the total ventilation, are decreased. Similarly a pressure-cycled ventilator is traditionally regarded as particularly susceptible to

15 314 ANESTHESA changes in lung characteristics; but if it operates as a flow generator in both phases (and is pressure-cycled in both phases) any change in tidal volume resulting from a change in lung characteristics is precisely compensated for by a change in rate so that the total ventilation is held constant. t may be concluded that the only way to determine how a given type of ventilator will perform in the face of changes in lung characteristics is either to try it experimentally or else to work systematically through a complete functional analysis of its operation in the manner outlined above. Acknowledgements t is a pleasure to acknowledge the encouragement of Professor William W. Mushin firstly to develop this method of analysis of the functioning of automatic ventilators and secondly to develop what hope is a clear exposition of it. Reference 1MUSHN, W. W., RENDELL-BAKER, L., and THOMPSON, P. W. (1959). Automatic Ventilation of the Lungs (with chapters on the Physics of Automatic Ventilators by W. W. Mapleson). Oxford. Blackwell Scientific Publications.

RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE

RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE Course n : Course 3 Title: RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE Sub-category: Intensive Care for Respiratory Distress Topic: Pulmonary Function and

More information

Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo

Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo Link full download: http://testbankcollection.com/download/pilbeams-mechanicalventilation-physiological-and-clinical-applications-5th-edition-test-bank-cairo

More information

PROBLEM SET 8. SOLUTIONS April 15, 2004

PROBLEM SET 8. SOLUTIONS April 15, 2004 Harvard-MIT Division of Health Sciences and Technology HST.542J: Quantitative Physiology: Organ Transport Systems Instructors: Roger Mark and Jose Venegas MASSACHUSETTS INSTITUTE OF TECHNOLOGY Departments

More information

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT IV - 2013 MECHANICAL VENTILATORS Prof. Yasser Mostafa Kadah Mechanical Ventilator A ventilator is a machine, a system of related elements designed to alter, transmit, and direct energy

More information

The Time Constant of The Respiratory System

The Time Constant of The Respiratory System The Time Constant of The Respiratory System by Enrico Bulleri Today s post aims to facilitate the understanding of a difficult but fascinating concept of mechanical ventilation: the time constant. In the

More information

Lung Volumes and Capacities

Lung Volumes and Capacities Lung Volumes and Capacities Normally the volume of air entering the lungs during a single inspiration is approximately equal to the volume leaving on the subsequent expiration and is called the tidal volume.

More information

Guide to Understand Mechanical Ventilation Waveforms

Guide to Understand Mechanical Ventilation Waveforms Do No Harm Ventilate Gently Guide to Understand Mechanical Ventilation Waveforms Middle East Critical Care Assembly 1/30/2015 Mazen Kherallah, MD, FCCP http://www.mecriticalcare.net Email: info@mecriticalcare.net

More information

The Manley ventilator

The Manley ventilator Anaesthesia, 1995, Volume 50, pages 6471 CLASSIC PAPER The Manley ventilator Roger Manley was senior house officer (SHO) in the anaesthetic department of Westminster Hospital when he made and tested his

More information

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR: PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: 1. **Identify and name the filters on the 7200ae. 2. **Explain how each filter is sterilized. 3. **Trace the gas flow through the ventilator

More information

Medical Instruments in the Developing World

Medical Instruments in the Developing World 2.2 Ventilators 2.2.1 Clinical Use and Principles of Operation Many patients in an intensive care and the operating room require the mechanical ventilation of their lungs. All thoracic surgery patients,

More information

Principles of Mechanical Ventilation: A Graphics-Based Approach

Principles of Mechanical Ventilation: A Graphics-Based Approach Principles of Mechanical Ventilation: A Graphics-Based Approach Steven M. Donn, MD, FAAP Professor of Pediatrics Neonatal-Perinatal Medicine C.S. Mott Children s Hospital University of Michigan Health

More information

APRV: Moving beyond ARDSnet

APRV: Moving beyond ARDSnet APRV: Moving beyond ARDSnet Matthew Lissauer, MD Associate Professor of Surgery Medical Director, Surgical Critical Care Rutgers, The State University of New Jersey What is APRV? APRV is different from

More information

Initiation and Management of Airway Pressure Release Ventilation (APRV)

Initiation and Management of Airway Pressure Release Ventilation (APRV) Initiation and Management of Airway Pressure Release Ventilation (APRV) Eric Kriner RRT Pulmonary Critical Care Clinical Specialist Pulmonary Services Department Medstar Washington Hospital Center Disclosures

More information

ALVEOLAR - BLOOD GAS EXCHANGE 1

ALVEOLAR - BLOOD GAS EXCHANGE 1 ALVEOLAR - BLOOD GAS EXCHANGE 1 Summary: These notes examine the general means by which ventilation is regulated in terrestrial mammals. It then moves on to a discussion of what happens when someone over

More information

Classification of Mechanical Ventilators

Classification of Mechanical Ventilators Classification of Mechanical Ventilators Kacmarek s 12 Point Classification Positive/Negative Pressure Powering Mechanism Driving Mechanism Single or Double Circuited Modes of Ventilation Cycling Parameter

More information

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance Introduction Expired Gas Analysis or indirect calorimetry, can be used to measure ventilation and the fractions of oxygen and carbon dioxide in expired air. From these measurements, the body's oxygen consumption

More information

Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 6th Edition by Cairo

Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 6th Edition by Cairo Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 6th Edition by Cairo Link full download: http://testbankair.com/download/test-bank-for-pilbeams-mechanicalventilation-physiological-and-clinical-applications-6th-edition-by-cairo/

More information

Selecting the Ventilator and the Mode. Chapter 6

Selecting the Ventilator and the Mode. Chapter 6 Selecting the Ventilator and the Mode Chapter 6 Criteria for Ventilator Selection Why does the patient need ventilatory support? Does the ventilation problem require a special mode? What therapeutic goals

More information

Unit 15 Manual Resuscitators

Unit 15 Manual Resuscitators 15-1 Unit 15 Manual Resuscitators GOAL On completion of this unit, the student should comprehend the proper operation of self-inflating resuscitation bags, flow-inflating resuscitation bags and gas-powered

More information

The physiological functions of respiration and circulation. Mechanics. exercise 7. Respiratory Volumes. Objectives

The physiological functions of respiration and circulation. Mechanics. exercise 7. Respiratory Volumes. Objectives exercise 7 Respiratory System Mechanics Objectives 1. To explain how the respiratory and circulatory systems work together to enable gas exchange among the lungs, blood, and body tissues 2. To define respiration,

More information

Let s talk about Capnography

Let s talk about Capnography Let s talk about Capnography This is one of a series of articles by Keith Simpson BVSc MRCVS MIET (Electronics) discussing the practical aspects of some common monitoring techniques. Capnometry is the

More information

Mechanical Ventilation

Mechanical Ventilation Mechanical Ventilation Understanding Modes Rob Chatburn, RRT-NPS, FAARC Research Manager Respiratory Therapy Cleveland Clinic Associate Professor Case Western Reserve University 1 Overview Characteristics

More information

Using Common Ventilator Graphics to Provide Optimal Ventilation

Using Common Ventilator Graphics to Provide Optimal Ventilation Using Common Ventilator Graphics to Provide Optimal Ventilation David Vines, MHS, RRT, FAARC Associate Professor Chair / Program Director Department of Respiratory Care RUSH UNIVERSITY MEDICAL CENTER Disclosure

More information

4/18/12 MECHANISM OF RESPIRATION. Every Breath You Take. Fun Facts

4/18/12 MECHANISM OF RESPIRATION. Every Breath You Take. Fun Facts Objectives MECHANISM OF RESPIRATION Dr Badri Paudel Explain how the intrapulmonary and intrapleural pressures vary during ventilation and relate these pressure changes to Boyle s law. Define the terms

More information

Ventilators. Dr Simon Walton Consultant Anaesthetist Eastbourne DGH KSS Basic Science Course

Ventilators. Dr Simon Walton Consultant Anaesthetist Eastbourne DGH KSS Basic Science Course Ventilators Dr Simon Walton Consultant Anaesthetist Eastbourne DGH KSS Basic Science Course Objectives Discuss Classification/ terminology Look at Modes of ventilation How some specific ventilators work

More information

Advanced Ventilator Modes. Shekhar T. Venkataraman M.D. Professor Critical Care Medicine and Pediatrics University of Pittsburgh School of Medicine

Advanced Ventilator Modes. Shekhar T. Venkataraman M.D. Professor Critical Care Medicine and Pediatrics University of Pittsburgh School of Medicine Advanced Ventilator Modes Shekhar T. Venkataraman M.D. Shekhar T. Venkataraman M.D. Professor Critical Care Medicine and Pediatrics University of Pittsburgh School of Medicine Advanced modes Pressure-Regulated

More information

UNDERSTANDING NEONATAL WAVEFORM GRAPHICS. Brandon Kuehne, MBA, RRT-NPS, RPFT Director- Neonatal Respiratory Services

UNDERSTANDING NEONATAL WAVEFORM GRAPHICS. Brandon Kuehne, MBA, RRT-NPS, RPFT Director- Neonatal Respiratory Services UNDERSTANDING NEONATAL WAVEFORM GRAPHICS Brandon Kuehne, MBA, RRT-NPS, RPFT Director- Neonatal Respiratory Services Disclosures Purpose: To enhance bedside staff s knowledge of ventilation and oxygenation

More information

HAMILTON-C2 HAMILTON-C2. The universal ventilation solution

HAMILTON-C2 HAMILTON-C2. The universal ventilation solution HAMILTON-C2 HAMILTON-C2 The universal ventilation solution The universal ventilation solution HAMILTON-C2 - The compact ventilation solution The HAMILTON-C2 mechanical ventilator is a universal ventilation

More information

BASIC PHYSICS APPLIED TO ANAESTHESIOLOGY

BASIC PHYSICS APPLIED TO ANAESTHESIOLOGY BASIC PHYSICS APPLIED TO ANAESTHESIOLOGY Dr.R.Selvakumar.M.D.D.A.DNB Professor of Anaesthesiology, K.A.P.Viswanatham Govt medical college, Trichy. The current practice of Anaesthesiology demands knowledge

More information

Emergency Transport and Ventilation

Emergency Transport and Ventilation Emergency Transport and Ventilation When you get a call and the patient is not breathing, are you and your equipment ready? Can you make the difference? Pneupac portable gas powered ventilators (PGPVs)

More information

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS 1. INTRODUCTION An anaesthesia delivery system includes any machine, equipment or apparatus which supplies gases, vapours, local anaesthesia and/or

More information

A Hare-Lynx Simulation Model

A Hare-Lynx Simulation Model 1 A Hare- Simulation Model What happens to the numbers of hares and lynx when the core of the system is like this? Hares O Balance? S H_Births Hares H_Fertility Area KillsPerHead Fertility Births Figure

More information

mechanical ventilation Arjun Srinivasan

mechanical ventilation Arjun Srinivasan Respiratory mechanics in mechanical ventilation Arjun Srinivasan Introduction Mechanics during ventilation PV curves Application in health & disease Difficulties & pitfalls The future. Monitoring Mechanics

More information

Chapter 4: Ventilation Test Bank MULTIPLE CHOICE

Chapter 4: Ventilation Test Bank MULTIPLE CHOICE Instant download and all chapters Test Bank Respiratory Care Anatomy and Physiology Foundations for Clinical Practice 3rd Edition Will Beachey https://testbanklab.com/download/test-bank-respiratory-care-anatomy-physiologyfoundations-clinical-practice-3rd-edition-will-beachey/

More information

Operation and Maintenance of the EPV200 Portable Ventilator

Operation and Maintenance of the EPV200 Portable Ventilator Operation and Maintenance of the EPV200 Portable Ventilator 1 Applications of the EPV200 The EPV200 Portable Ventilator is a gas powered electronically controlled mechanical ventilator, designed to provide

More information

Basics of Mechanical Ventilation. Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity

Basics of Mechanical Ventilation. Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity Basics of Mechanical Ventilation Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity Overview of topics 1. Goals 2. Settings 3. Modes 4. Advantages and disadvantages

More information

EXCESSIVE WORK OF BREATHING DURING INTERMITTENT MANDATORY VENTILATION

EXCESSIVE WORK OF BREATHING DURING INTERMITTENT MANDATORY VENTILATION Br. J. Anaesth. (1986), 58, 1048-1054 EXCESSIVE WORK OF BREATHING DURING INTERMITTENT MANDATORY VENTILATION J. S. MECKLENBURGH, I. P. LATTO, T. A. A. AL-OBAIDI, E. A. SWAI AND W. W. MAPLESON Ventilators

More information

Hospital and Transport for Controlled Breathing

Hospital and Transport for Controlled Breathing Hospital and Transport for led Breathing When transporting a critically ill patient you need a ventilator that can go anywhere in any situation. Smiths Medical Pneupac small portable gas powered ventilators

More information

Mechanical Ventilation. Which of the following is true regarding ventilation? Basics of Ventilation

Mechanical Ventilation. Which of the following is true regarding ventilation? Basics of Ventilation Mechanical Ventilation Jeffrey L. Wilt, MD, FACP, FCCP Associate Professor of Medicine Michigan State University Associate Program Director MSU-Grand Rapids Internal Medicine Residency Which of the following

More information

Characterizers for control loops

Characterizers for control loops Characterizers for control loops By: F. G. Shinskey (May 1999) Introduction Commercial controllers such as the PID series (proportional, integral, derivative, and their combinations) are linear devices

More information

Blease900 Ventilator PATIENT CENTERED VENTILATION

Blease900 Ventilator PATIENT CENTERED VENTILATION Blease900 Ventilator PATIENT CENTERED VENTILATIN PATIENT CENTERED VENTILATIN From the first Blease/Manley ventilator sixty years ago, Spacelabs Healthcare has built a heritage of innovative and intuitive

More information

VENTILATION SERVO-s EASY AND RELIABLE PATIENT CARE

VENTILATION SERVO-s EASY AND RELIABLE PATIENT CARE VENTILATION SERVO-s EASY AND RELIABLE PATIENT CARE Critical Care SERVO-s 3 SERVO-s SIMPLY MAKES SENSE MAQUET THE GOLD STANDARD Leading the way: MAQUET is a premier international provider of medical products

More information

O-Two Self-Study Guide. e600 Transport Ventilator Ventilation Modes

O-Two Self-Study Guide. e600 Transport Ventilator Ventilation Modes O-Two Self-Study Guide e600 Transport Ventilator Ventilation Modes e600 VENTILATION MODES The e600 ventilator has 13 ventilation modes: Manual Ventilation, Controlled Mandatory Ventilation (CMV), Assist

More information

Pulmonary Circulation Linda Costanzo Ph.D.

Pulmonary Circulation Linda Costanzo Ph.D. Pulmonary Circulation Linda Costanzo Ph.D. OBJECTIVES: After studying this lecture, the student should understand: 1. The differences between pressures in the pulmonary and systemic circulations. 2. How

More information

VENTILATORS PURPOSE OBJECTIVES

VENTILATORS PURPOSE OBJECTIVES VENTILATORS PURPOSE To familiarize and acquaint the transfer Paramedic with the skills and knowledge necessary to adequately maintain a ventilator in the interfacility transfer environment. COGNITIVE OBJECTIVES

More information

12. School travel Introduction. Part III Chapter 12. School travel

12. School travel Introduction. Part III Chapter 12. School travel 12. School travel 12.1 Introduction This chapter presents the evidence on changes in travel patterns for the journey to school in the three towns over the period of the Sustainable Travel Town project.

More information

PICU Resident Self-Study Tutorial The Basic Physics of Oxygen Transport. I was told that there would be no math!

PICU Resident Self-Study Tutorial The Basic Physics of Oxygen Transport. I was told that there would be no math! Physiology of Oxygen Transport PICU Resident Self-Study Tutorial I was told that there would be no math! INTRODUCTION Christopher Carroll, MD Although cells rely on oxygen for aerobic metabolism and viability,

More information

Pressure Controlled Modes of Mechanical Ventilation

Pressure Controlled Modes of Mechanical Ventilation Pressure Controlled Modes of Mechanical Ventilation Christopher Junker Department of Anesthesiology & Critical Care Medicine George Washington University Saturday, August 20, 2011 Assist Control Hypoxemic

More information

Gas exchange measurement in module M- COVX

Gas exchange measurement in module M- COVX Gas exchange measurement in module M- COVX This document explains how the gas exchange measurement in the gas module M-COVX works. 1. Basic equations To obtain the oxygen consumption VO2 of a patient,

More information

http://www.priory.com/cmol/hfov.htm INTRODUCTION The vast majority of patients who are admitted to an Intensive Care Unit (ICU) will need artificial ventilation (Jones et al 1998). The usual means through

More information

Automatic Transport Ventilator

Automatic Transport Ventilator Automatic Transport Ventilator David M. Landsberg, MD, FACP, FCCP, EMT-P Luke J. Gasowski, RRT, NPS, ACCS, CCP-C, FP-C Christopher J. Fullagar, MD, FACEP, EMT-P Stan Goettel, MS, EMT-P Author credits /

More information

AUTOVENT 4000 VENTILATOR

AUTOVENT 4000 VENTILATOR OVERVIEW AUTOVENT 4000 Only properly trained and approved Escambia County Bureau of Public Safety Paramedics are to use the AutoVent 4000 ventilator manufactured by LSP to transport patients already on

More information

Mechanical Ventilation. Mechanical Ventilation is a Drug!!! is a drug. MV: Indications for use. MV as a Drug: Outline. MV: Indications for use

Mechanical Ventilation. Mechanical Ventilation is a Drug!!! is a drug. MV: Indications for use. MV as a Drug: Outline. MV: Indications for use Mechanical Ventilation is a Drug!!! Mechanical Ventilation is a drug I am an employee of Philips Healthcare Hospital Respiratory Care Group and they help me pay for my kids education Jim Laging, RRT, RCP

More information

6 th Accredited Advanced Mechanical Ventilation Course for Anesthesiologists. Course Test Results for the accreditation of the acquired knowledge

6 th Accredited Advanced Mechanical Ventilation Course for Anesthesiologists. Course Test Results for the accreditation of the acquired knowledge 6 th Accredited Advanced Mechanical Ventilation Course for Anesthesiologists Course Test Results for the accreditation of the acquired knowledge Q. Concerning the mechanics of the newborn s respiratory

More information

Virginia Beach EMS. Oxylator EMX. Debra H. Brennaman, RN, MPA, NREMT-P

Virginia Beach EMS. Oxylator EMX. Debra H. Brennaman, RN, MPA, NREMT-P Virginia Beach EMS Oxylator EMX Debra H. Brennaman, RN, MPA, NREMT-P Oxylator EMX Overview Patient responsive oxygen powered resuscitation / ventilation device intended to provide emergency ventilatory

More information

INTRODUCTION TO BI-VENT (APRV) INTRODUCTION TO BI-VENT (APRV) PROGRAM OBJECTIVES

INTRODUCTION TO BI-VENT (APRV) INTRODUCTION TO BI-VENT (APRV) PROGRAM OBJECTIVES INTRODUCTION TO BI-VENT (APRV) INTRODUCTION TO BI-VENT (APRV) PROGRAM OBJECTIVES PROVIDE THE DEFINITION FOR BI-VENT EXPLAIN THE BENEFITS OF BI-VENT EXPLAIN SET PARAMETERS IDENTIFY RECRUITMENT IN APRV USING

More information

The Merlin Small Animal Ventilator by Vetronic Services Ltd. Contents page. Introduction.2. Features of Merlin. 3

The Merlin Small Animal Ventilator by Vetronic Services Ltd. Contents page. Introduction.2. Features of Merlin. 3 The Merlin Small Animal Ventilator by Vetronic Services Ltd Contents page Introduction.2 Features of Merlin. 3 A brief over-view of how Merlin works 4 Merlin Front Panel and Controls. 6 Gas Port Connections..

More information

How Ventilators Work. Chapter 3

How Ventilators Work. Chapter 3 How Ventilators Work Chapter 3 To care for a ventilator patient, you need to know: The various functions of the ventilator used How the ventilator interacts with the patient How changes in lung condition

More information

Modelling and Simulation of Pressure Controlled Mechanical Ventilation System

Modelling and Simulation of Pressure Controlled Mechanical Ventilation System J. Biomedical Science and Engineering, 2015, 8, 707-716 Published Online October 2015 in SciRes. http://www.scirp.org/journal/jbise http://dx.doi.org/10.4236/jbise.2015.810068 Modelling and Simulation

More information

(Received 9 September 1940)

(Received 9 September 1940) 257 J. Physiol. (I 94I) 99, 257-264 6I2.2II A METHOD OF RECORDING THE RESPIRATION BY J. H. GADDUM From the College of the Pharmaceutical Society, 17 Bloomsbury Square, London, W.C. 2 (Received 9 September

More information

Recommendations on Checking Anaesthesia Delivery Systems

Recommendations on Checking Anaesthesia Delivery Systems Page 1 of 11 Recommendations on Checking Anaesthesia Delivery Version Effective Date 1 Oct 1992 (reviewed Feb 07, Feb 02) 2 2004 3 Nov 2011 4 Dec 2016 Document No. HKCA T1 v4 Prepared by College Guidelines

More information

DATA SHEET. VENTILATION SERVO-s

DATA SHEET. VENTILATION SERVO-s DATA SHEET VENTILATION The Gold Standard Critical Care 3 PROVEN SERVO SIMPLICITY, RELIABILITY AND SENSITIVITY MAQUET THE GOLD STANDARD Leading the way: MAQUET is a premier international provider of medical

More information

Design and development of microcontroller based human airway pressure measurement system with real time graphical display for anaesthesia ventilator

Design and development of microcontroller based human airway pressure measurement system with real time graphical display for anaesthesia ventilator Journal of Scientific & Industrial Research Vol. 65, December 2006, pp. 986-991 Design and development of microcontroller based human airway pressure measurement system with real time graphical display

More information

Principles of mechanical ventilation. Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands

Principles of mechanical ventilation. Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands Principles of mechanical ventilation Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands Disclosure Research grant Chiesi Pharmaceuticals Research grant CareFusion GA: 27 weeks,

More information

Calculation of Trail Usage from Counter Data

Calculation of Trail Usage from Counter Data 1. Introduction 1 Calculation of Trail Usage from Counter Data 1/17/17 Stephen Martin, Ph.D. Automatic counters are used on trails to measure how many people are using the trail. A fundamental question

More information

plethysmographic methods that when the subject was pinched on the upper

plethysmographic methods that when the subject was pinched on the upper 24 J. Physiol. (I95I) II2, 24-2I 6I2.I5.6II.976 THE DECREASE IN HAND BLOOD FLOW FOLLOWING INFLATION OF AN ARTERIAL OCCLUSION CUFF ON THE OPPOSITE ARM BY IAN C. RODDIE From the Department of Physiology,

More information

Presentation Overview. Monitoring Strategies for the Mechanically Ventilated Patient. Early Monitoring Strategies. Early Attempts To Monitor WOB

Presentation Overview. Monitoring Strategies for the Mechanically Ventilated Patient. Early Monitoring Strategies. Early Attempts To Monitor WOB Monitoring Strategies for the Mechanically entilated Patient Presentation Overview A look back into the future What works and what may work What s all the hype about the WOB? Are ventilator graphics really

More information

Technical Data and Specifications

Technical Data and Specifications Technical Data and Specifications INTENDED USE Ventilator designed to provide Invasive and Non-invasive ventilation for the critical care management of adult, pediatric and neonate-infant (including premature)

More information

Optimizing Compressed Air Storage for Energy Efficiency

Optimizing Compressed Air Storage for Energy Efficiency University of Dayton ecommons Mechanical and Aerospace Engineering Faculty Publications Department of Mechanical and Aerospace Engineering 4-2011 Optimizing Compressed Air Storage for Energy Efficiency

More information

Operating Instructions for Microprocessor Controlled Ventilators

Operating Instructions for Microprocessor Controlled Ventilators Page 1 of 5 Operating Instructions for Microprocessor Controlled Ventilators Purpose Audience Scope Physician's Order To provide guidelines for the procedure for the use of microprocessor controlled ventilators.

More information

Respiratory Pulmonary Ventilation

Respiratory Pulmonary Ventilation Respiratory Pulmonary Ventilation Pulmonary Ventilation Pulmonary ventilation is the act of breathing and the first step in the respiratory process. Pulmonary ventilation brings in air with a new supply

More information

Monitoring, Ventilation & Capnography

Monitoring, Ventilation & Capnography Why do we need to monitor? Monitoring, Ventilation & Capnography Keith Simpson BVSc MRCVS MIET(Electronics) Torquay, Devon. Under anaesthesia animals no longer have the ability to adequately control their

More information

Peter Kremeier, Christian Woll. 2nd. Understanding and comparing modes of ventilation. The Kronberg List of Ventilation Modes

Peter Kremeier, Christian Woll. 2nd. Understanding and comparing modes of ventilation. The Kronberg List of Ventilation Modes Extended Edition Revised and 2nd Peter Kremeier, Christian Woll Understanding and comparing modes of ventilation The Kronberg List of Ventilation Modes Contents Preface... 3 Comparison Table - Parameters...6

More information

Servo UNIVERSAL ADVANCED VENTILATION PLATFORM ALL PATIENT CATEGORIES

Servo UNIVERSAL ADVANCED VENTILATION PLATFORM ALL PATIENT CATEGORIES DATA ADVANCED VENTILATION PLATFORM ALL PATIENT CATEGORIES CRITICAL CARE HIGHLIGHTS! For all patient categories, from neonate and pediatric to adult! A platform for the future choose the level of system

More information

Mechanics of proportional-assist ventilation

Mechanics of proportional-assist ventilation CHAPTER 10 Mechanics of proportional-assist ventilation A.C. Lua & K.C. Shi School of Mechanical and Aerospace Engineering, Nanyang Technological University, 50, Nanyang Avenue, Singapore 639798. Abstract

More information

Minimum size for maximum performance

Minimum size for maximum performance HAMILTON-C1 HAMILTON-C1 Minimum size for maximum performance Minimum size for maximum performance HAMILTON-C1 - One for all The high-performance capabilities of the HAMILTON-C1 include advanced lung protective

More information

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator)

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator) VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator) VT PLUS HF provides a special mode for evaluating the performance of high frequency ventilators while connected

More information

PART SEVEN THE HISTORY AND APPLICATION OF HIGH FREQUENCY OSCILLATORY VENTILATION (HFOV)

PART SEVEN THE HISTORY AND APPLICATION OF HIGH FREQUENCY OSCILLATORY VENTILATION (HFOV) PART SEVEN THE HISTORY AND APPLICATION OF HIGH FREQUENCY OSCILLATORY VENTILATION (HFOV) Reciprocating pistons with an eccentric travel speed, moving to and fro within a cylinder (with a common inlet/outlet),

More information

Mitos Fluika Pressure and Vacuum Pumps Datasheet

Mitos Fluika Pressure and Vacuum Pumps Datasheet Unit 1, Anglian Business Park, Orchard Road, Royston, Hertfordshire, SG8 5TW, UK T: +44 (0)1763 242491 F: +44 (0)1763 246125 E: sales@dolomite-microfluidics.com W: www.dolomite-microfluidics.com Dolomite

More information

A bench study comparison of volume delivery in the presence of leak

A bench study comparison of volume delivery in the presence of leak A bench study comparison of volume delivery in the presence of leak Study objective To compare the volume delivery of the Trilogy1 ventilator to the volume delivery of the LTV 1 as a representative of

More information

Fluid Flow. Link. Flow» P 1 P 2 Figure 1. Flow Model

Fluid Flow. Link. Flow» P 1 P 2 Figure 1. Flow Model Fluid Flow Equipment: Water reservoir, output tubes of various dimensions (length, diameter), beaker, electronic scale for each table. Computer and Logger Pro software. Lots of ice.temperature probe on

More information

SLE5000 Infant Ventilator with HFO

SLE5000 Infant Ventilator with HFO SLE5000 Infant Ventilator with HFO When the smallest thing matters SLE5000 - The Total Solution for Infant Ventilation Shown on optional roll stand. Ventilator may be mounted either way round on stand.

More information

Collin County Community College. Lung Physiology

Collin County Community College. Lung Physiology Collin County Community College BIOL. 2402 Anatomy & Physiology WEEK 9 Respiratory System 1 Lung Physiology Factors affecting Ventillation 1. Airway resistance Flow = Δ P / R Most resistance is encountered

More information

VENTILATION STRATEGIES FOR THE CRITICALLY UNWELL

VENTILATION STRATEGIES FOR THE CRITICALLY UNWELL VENTILATION STRATEGIES FOR THE CRITICALLY UNWELL Dr Nick Taylor Visiting Emergency Specialist Teaching Hospital Karapitiya Senior Specialist and Director ED Training Clinical Lecturer, Australian National

More information

PERFORMANCE OF NEEDLE VALVES

PERFORMANCE OF NEEDLE VALVES Br. J. Anaesth. (1986), 58, 919-924 PERFORMANCE OF NEEDLE VALVES P. HUTTON AND R. W. BOADEN The use of needle valves to control theflowof gas in anaesthetic breathing systems is almost universal. We became

More information

HAMILTON-C3 HAMILTON-C3. The compact high-end ventilator

HAMILTON-C3 HAMILTON-C3. The compact high-end ventilator HAMILTON-C3 HAMILTON-C3 The compact high-end ventilator The compact high-end ventilator HAMILTON-C3 - The all-rounder for ICUs The HAMILTON-C3 ventilator is a modular high-end ventilation solution for

More information

CPR MEDICAL DEVICES INC. OPERATING MANUAL

CPR MEDICAL DEVICES INC. OPERATING MANUAL CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR OXYLATOR EMX RESUSCITATION AND INHALATION SYSTEM Symbols used in the Operating Manual for the OXYLATOR EMX and/or on the OXYLATOR EMX unit: WARNING! Do not

More information

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting This is a quick reference for setting up, using, and troubleshooting the Hallowell EMC 2000 and 2002 Small Animal Ventilator.

More information

Flight Medical presents the F60

Flight Medical presents the F60 Flight Medical presents the F60 Reliable Ventilation Across the Spectrum of Care Adult & Pediatric Pressure/Volume Control Basic/Advanced Modes Invasive/NIV High Pressure/Low Flow O2 Up to 12 hours batteries

More information

Advanced nasal CPAP system www.hamilton-medical.com/arabella Noninvasive, low-cost, effective and safe Neonatal nasal CPAP therapy Nasal CPAP therapy aims to support neonates, especially pre-term and

More information

Effect of Fluid Density and Temperature on Discharge Coefficient of Ogee Spillways Using Physical Models

Effect of Fluid Density and Temperature on Discharge Coefficient of Ogee Spillways Using Physical Models RESEARCH ARTICLE Effect of Fluid Density and Temperature on Discharge Coefficient of Ogee Spillways Using Physical Models M. SREENIVASULU REDDY 1 DR Y. RAMALINGA REDDY 2 Assistant Professor, School of

More information

PROBLEM SET 9. SOLUTIONS April 23, 2004

PROBLEM SET 9. SOLUTIONS April 23, 2004 Harvard-MIT Division of Health Sciences and Technology HST.542J: Quantitative Physiology: Organ Transport Systems Instructors: Roger Mark and Jose Venegas MASSACHUSETTS INSTITUTE OF TECHNOLOGY Departments

More information

LUNG CLEARANCE INDEX. COR-MAN IN Issue A, Rev INNOVISION ApS Skovvænget 2 DK-5620 Glamsbjerg Denmark

LUNG CLEARANCE INDEX. COR-MAN IN Issue A, Rev INNOVISION ApS Skovvænget 2 DK-5620 Glamsbjerg Denmark LUNG CLEARANCE INDEX METHOD COR-MAN-0000-008-IN Issue A, Rev. 3 2013-07-01 INNOVISION ApS Skovvænget 2 DK-5620 Glamsbjerg Denmark Tel.: +45 65 95 91 00 Fax: +45 65 95 78 00 info@innovision.dk www.innovision.dk

More information

Organis TestChest. Flight Simulator for Intensive Care Clinicians

Organis TestChest. Flight Simulator for Intensive Care Clinicians Organis TestChest Flight Simulator for Intensive Care Clinicians Organis TestChest Critical Care challenges Training on ventilation modes with simulation is crucial for patient safety The equipment and

More information

(AS AT 31 st MARCH, 2002)

(AS AT 31 st MARCH, 2002) ACACA PROTOCOL 2000 (AS AT 31 st MARCH, 2002) ACACA PROTOCOL 2000 INCLUDES (A) CODE OF PRACTICE FOR MANUFACTURERS AND/OR SUPPLIERS OF COMMERCIAL AIR COMPRESSORS AND METHOD FOR DETERMINING (B) RECIPROCATING

More information

Mechanical power and opening pressure. Fellowship training program Intensive Care Radboudumc, Nijmegen

Mechanical power and opening pressure. Fellowship training program Intensive Care Radboudumc, Nijmegen Mechanical power and opening pressure Fellowship training program Intensive Care Radboudumc, Nijmegen Mechanical power Energy applied to the lung: Ptp * V (Joule) Power = Energy per minute (J/min) Power

More information

CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS

CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS Brit. J. Anaesth. (1956), 28, 196 CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS BY RUSSELL M. DAVIES, I. R. VERNER Queen Victoria Hospital, East Grinstead AND A. BRACKEN Research and Development Centre,

More information

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12 LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/ This experiment will introduce you to the kinetic properties of low-pressure gases. You will make observations on the rates with which selected

More information

V8600 Ventilator. Integrated Invasive & Noninvasive Ventilation

V8600 Ventilator. Integrated Invasive & Noninvasive Ventilation V8600 Ventilator Integrated Invasive & Noninvasive Ventilation 0123 V8600 Ventilator Invasive Ventilation With state-of-the-art turbine technology, V8600 helps achieve sequential ventilation in various

More information

Narkomed MRI-2. A perfect fit for the MRI environment. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

Narkomed MRI-2. A perfect fit for the MRI environment. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care Narkomed MRI-2 A perfect fit for the MRI environment Emergency Care Perioperative Care Critical Care Perinatal Care Home Care Because you care The complete anesthesia solution for the MRI environment A

More information