Brit. J. Anaesth. (1968), 40, 666. REBREATHING CIRCUITS A Review

Size: px
Start display at page:

Download "Brit. J. Anaesth. (1968), 40, 666. REBREATHING CIRCUITS A Review"

Transcription

1 Brit. J. Anaesth. (1968), 40, 666 REBREATHING CIRCUITS A Review BY M, K. SYKES The effects of rebreathing depend on the composition of the gas which is re-inhaled and whether this gas passes to the alveoli (where it interferes with gas exchange) or only to the anatomical deadspace of the patient. It is therefore important to differentiate between the rebreathing of deadspace gas, alveolar gas and mixed expired gas and to know in what order these gases are reinhaled. The deadspace gas may be divided into apparatus and anatomical components. Gas from the apparatus deadspace will vary in composition depending on the arrangement of the circuit. Gas expired from the anatomical (conducting airway) deadspace has a composition similar to inspired gas but is saturated with water vapour and at a temperature of about 32 C. Alveolar gas is also saturated but contains 5-6 per cent carbon dioxide and per cent oxygen, whilst mixed expired gas usually contains 3-4 per cent carbon dioxide and per cent oxygen (when breathing 21 per cent oxygen). Unfortunately, although it is possible to forecast the general behaviour of a circuit on theoretical grounds, it is not possible to make detailed predictions of its behaviour under clinical conditions. There are a number of reasons for this. First, the effective volume of the apparatus deadspace may not equal that determined by volumetric displacement (Clarke, 1958; Harrison, Ozinsky and Jones, 1959). This difference arises from variations in the pattern of gas flow due to channelling, turbulence and the cone-front effect (Briscoe, Forster and Comroe, 1954). However, under certain circumstances this problem can be overcome by utilizing other methods to measure apparatus deadspace (e.g. Voss, 1963; Bethune and Collis, 1967). The second reason why predictions fail is that the size of the anatomical deadspace, and hence the volume of deadspace gas, may vary (Sykes, 1967). The third reason for failure is that the composition of alveolar gas depends not only on the overall relation of alveolar ventilation to oxygen and carbon dioxide exchange but also on the pattern of ventilation and blood flow within the lungs (fig. 1). The effect of rebreathing alveolar gas will therefore depend on which portion of this gas is re-inhaled. Finally, the patient may react to a change in alveolar gas composition by a change in alveolar ventilation (Clappison and Hamilton, 1956; Elam and Brown, 1956). Although these factors lead to difficulties in theoretical analysis there is now general agreement on the underlying principles of most of the common rebreathing circuits. to, EXPIRATION- -t^ INSPIRATION FIG. 1 Expired oxygen tension (Po,) and carbon dioxide tension (PcoJ. Patient with normal longs (solid lines) and patient with ventilation/perfusion inequality (dotted lines).

2 REBREATHING CIRCUITS 667 EFFECTS OF REBREATHING 1. Retention of water and heat. Since the specific heat of air is low (0.24 cal/g= cal/ml at 20 C) whilst the latent heat of vaporization of water is high (580 cal/g), most of the heat loss in the respiratory tract is due to the evaporation of water. Hence there is little difference in heat loss whether hot or cold air is inhaled, providing that it is dry. However, the inhalation of warm, moist, air markedly reduces heat loss. In this respect rebreathing may be beneficial. On the other hand, the inhalation of saturated air at a temperature above 37 C results in the addition of heat to the body. This may lead to hyperthermia, particularly if the normal heat-regulating mechanisms are depressed by the drugs used in anaesthesia. In most anaesthetic circuits heat is rapidly lost to the atmosphere and the gas which is re-inhaled has a lower temperature and water content than expired gas. However, in some circuits, such as the Waters to-and-fro absorption circuit, the tubes are short and heat loss is minimal; in addition, heat is added to the circuit by the chemical reaction taking place in the soda-lime. In this situation the inspired gas may be fully saturated and at a temperature of C (Adriani, 1960; Burton, 1962). Dangerous hyperthermia may result 2. Alteration of inspired and alveolar gas tensions. The general equation governing the alveolar concentration (FA,) of a gas (x) is: where ty, (the volume exchanged in unit time) is positive for uptake and negative for elimination, \^A is the alveolar ventilation in unit time and Fix is the inspired concentration (Nunn and Newman, 1964). From this equation it can be seen that alveolar concentration is dependent on inspired concentration and that the alveolar concentration only equals the inspired concentration when gas exchange is zero or alveolar ventilation is infinite. Normally the partial pressures of the individual components of fresh gas entering the lungs are lowered only by the tension of water vapour. When rebreathing of alveolar gas occurs there will be a further reduction in inspired oxygen tension and an increase in inspired carbon dioxide tension (Swartz, Adriani and Mih, 1953; Ruben, 1953; Firton, 1958, 1963). Similar changes will occur in the alveolar gas tensions unless the alveolar ventilation undergoes a compensatory increase. In addition to its effects on the respiratory gases, rebreathing can also cause an alteration in the inspired and alveolar tensions of an anaesthetic gas or vapour. During induction the alveolar tension is less than the inspired; rebreathing of alveolar gas therefore causes a reduction in inspired and alveolar tension and prolongs induction. During recovery alveolar tension exceeds inspired. In this situation rebreathing maintains a higher alveolar tension and so slows the elimination of an anaesthetic gas or vapour. In a similar fashion rebreathing slows down the change in alveolar gas composition in response to a change in inspired gas composition. This slowing is additional to the delay caused by replacing the gas in the circuit with gas of a new composition. MEASUREMENT OF REBREATHING As the rebreathing of deadspace gas is of little clinical significance most studies have been concerned with the rebreathing of alveolar gas. This has usually been detected by the analysis of carbon dioxide since physical methods for continuous analysis of this gas have been readily available. Furthermore, it is easier to add measured quantities of carbon dioxide to model lung systems than to arrange for a continuous uptake of oxygen. However, in clinical studies similar results may be obtained by continuous analysis of carbon dioxide or oxygen concentrations (Sykes, 1959a; Norman, Adams and Sykes, 1968). There are three main methods for the detection of rebreathing when using carbon dioxide as the indicator gas: (1) measurement of its concentration in the inspired gas; (2) measurement of the increase in alveolar carbon dioxide concentration at constant ventilation; (3) measurement of the increase in alveolar ventilation at constant arterial or alveolar Pco,. The measurement of the quantity of carbon dioxide in the inspired gas is most simply performed by deflecting the whole of one inspiration

3 668 BRITISH JOURNAL OF ANAESTHESIA into a reservoir so that the gas composition and volume can be measured. This method was used in model experiments by Woolmer and Lind (1954). A disadvantage is that the equilibrium of the system is temporarily upset Alternatively, the inspired carbon dioxide concentration can be continuously monitored and replotted against inspired volume (determined by integration of a pneumotachograph signal). The area under the curve is then indicative of the volume of carbon dioxide in the inspired gas (fig. 2). The method is time-consuming and requires impeccable instrumentation. Its main use is in the detailed analysis of the pattern of gas flow in a system. where FAooi=alveolar carbon dioxide concentration; Flcoa = inspired carbon dioxide concentration; Vco,=volume of carbon dioxide entering lungs (constant); VA=alveolar ventilation. Another method of quantifying the degree of rebreathing is to equate the increase in alveolar carbon dioxide or decrease in alveolar ventilation with the effect of an added deadspace. If VD represents the total deadspace and VT and f represent tidal volume and respiratory frequency respectively then V)f or VD=VTnow V*CO,=^AXFA 0O, or VA=CVC VD=VT-[Vco 1 /(FA 0Ol xf)] If VD is determined before and after adding the circuit under test the degree of rebreathing can be expressed as an increase in deadspace over that determined in the absence of rebreathing (Sykes, 1959a; Voss, 1963). ;:.. <} < > Volume Inspired FIG. 2 Mapleson "C circuit, spontaneous ventilation. Fresh gas inflow=15 L/tnin. Inspired carbon dioxide concentration replotted against volume inspired (ml). The initial dip in the carbon dioxide concentration is due to the fresh gas which accumulates under the valve mount during the expiratory period. The second and third methods are based on the assumption that carbon dioxide output is constant and that in a steady respiratory state the volume of carbon dioxide entering the lungs is equal to the volume removed by the alveolar ventilation. Hence any increase in alveolar carbon dioxide concentration or alveolar ventilation must be due to an increase in inspired carbon dioxide concentration. This is illustrated by the application of the general equation (1) to carbon dioxide: CIRCUITS IN WHICH REBREATHING MAY OCCUR These will be classified according to the presence or absence of valves and, if present, on their number and position. Circuits with no valves. The simplest example is the open-drop gauze mask. This becomes an extension of the anatomical deadspace. The degree of rebreathing is determined by the volume of the mask and the thickness of the material used to cover the frame. Rebreathing is increased (and dilution of anaesthetic vapour with fresh air minimized) by the application of a "chimney" (fig. 3). The inhalers of Clover and Ombredanne also act as an extension of the patient's deadspace, but the volume of rebreathing is more clearly defined. Complete rebreathing occurs in the Oxford and Goldman's vinesthene vaporizers since the circuit is completely dosed. Another circuit which has been widely used is the T-piece. This was originally described by Ayre (1937a, b) and further discussed by him in 1956 and The principle has been examined theoretically by Mapleson (1954, 1958),

4 REBREATHING CIRCUITS 669 providing a flow of fresh gas which exceeds the peak inspiratory flow rate. This normally approximates to three times the minute volume (fig. 4). -tlnspiratioi I In 30,- -Expiration- L/rren Schimmelbusch mask FIG. 3 Schimmelbusch mask with towel used to increase concentration of anaesthetic vapour. Onchi, Hayashi and Ueyama (1957), and Lewis and Spoerel (1961). Clinical and experimental investigations have been reported by Woolmer and Lind (1954), Inkster (1956), Voss (1963) and Harrison (1964a). The circuit was designed to eliminate the resistance of the expiratory valve and it was originally suggested that the volume of the expiratory limb should be less than the patient's tidal volume. A number of modifications have since been described (Brooks, Stuart and Gabel, 1958; Voss, 1963: Harrison 1964b) but the manner in which the T-piece functions depends primarily on the volume and shape of the expiratory limb. The expiratory limb may consist of a long tube with a volume greater than the patient's tidal volume, a tube of similar volume with a reservoir bag at the distal end (Rees, 1950), or the expiratory limb may be reduced in length so that the internal volume is less than the patient's tidal volume. The extreme example of this is the modification of Lewis and Spoerel (1961) and of Picken (1950) in which the volume of the expiratory limb is reduced to zero, the expired gases being eliminated through a hole in the wall of the catheter mount With the latter arrangement rebreathing cannot occur, since the patient breathes out directly to atmosphere. However, with this type of T-piece dilution of fresh gases with room air can only be prevented by FIG. 4 Pattern of inspiratory and expiratory flow rate during spontaneous ventilation. Minute volume 10 L/min. As the volume of the expiratory limb is increased so the risk of air dilution diminishes. When the volume of the expiratory limb exceeds the tidal volume dilution is unlikely to occur. The functional analysis of such a T-piece is shown in figure 5. Fresh gas flows continuously Patient I Dead space gas I Alveolar gas FIG. 5 T-piece (Mapleson "E" circuit). Disposition of gases at end of expiration. FGF=fresh gas flow. through the side arm. The peak expiratory flow rate occurs early in expiration and flow then diminishes in an approximately exponential fashion (fig. 4). The proportion of fresh gas which is added to the expired gas as it passes down the expiratory limb therefore increases as expiration progresses and during the expiratory pause fresh gas continues to accumulate at the patient end of the expiratory limb. Rebreathing cannot occur if the fresh gas flow rate exceeds the peak inspiratory flow rate, for all the inspired gas is then supplied from the fresh gas inlet. Under most circumstances a smaller fresh gas flow rate may be used without rebreathing. Favourable factors are a high inspiratory: expiratory time ratio, a slow

5 670 BRITISH JOURNAL OF ANAESTHESIA rise in inspiratory flow rate, a low flow rate during the last part of expiration and a long expiratory pause. All these factors provide time for the fresh gas flow to flush the expiratory limb; this minimizes the quantity of alveolar gas which is re-inhaled. Furthermore, it does not matter if alveolar gas is re-inhaled into the anatomical deadspace since no gas exchange takes place in this portion of the respiratory tract. Under the most favourable conditions during spontaneous ventilation the fresh gas flow rate may be reduced to 2-2^ times the minute volume without rebreathing occurring. Similar considerations apply when ventilation is controlled. It should be noted that the addition of a bag to the expiratory limb (Rees, 1950) will not affect the function of the system providing that the bag is separated from the patient by a tube with an internal volume which exceeds the patient's tidal volume. If the tube volume is too small re-inhalation of mixed expired gas from the bag may occur. The system then approximates to a Mapleson "C" circuit (Mapleson, 1954). One further valveless circuit deserves mention. This is the use of a circle absorption system in which the direction of gas flow is controlled, not by valves, but by a turbine-operated blower (Roffey, Revell and Morris, 1961). The advantage claimed is that resistance to breathing is minimized and that dangers arising from faulty action of the valves are eliminated. This system behaves as a T-piece at the mouth; circulating volume flow must therefore exceed three times the patient's minute volume to ensure that expired gas is not rebreathed until it has passed through the absorber. One of the dangers of this type of apparatus is that any increased resistance in the absorber or other part of the circuit may so reduce the flow of gas round th: circuit that rebreathing results. The principle of the T-piece is also utilized in many oxygen masks and tracheostomy humidifier boxes: since high oxygen flows are often undesirable, rebreathing is usually minimized by dilution with room air. T-piece circuits have the advantage that they are small, simple and have small apparatus deadspace. However, they require relatively large gas flows. They are, therefore, most suitable for paediatric use. Circuits with Expiratory Valves only. Spontaneous ventilation. The classification used by Mapleson (1954) is shown in figure 6. - Constant gas flow from anaesthetic machine Reservoir bag Corrugated tubing -i~i- Expiratory valve / \ Face mask FIG. 6 Classification of circuits (Mapleson, 1954). The Magill circuit (Mapleson "A") has been studied in most detail. The efficiency of this circuit during spontaneous ventilation depends on the separation of expired gas into deadspace and alveolar portions (Wynne, 1941; Molyneux and Pask, 1951; Domaingue, 1951). On expiration the deadspace portion travels up the tube towards the bag, followed by the alveolar gas portion (fig. 7). The deadspace gas meets the fresh gas entering the circuit and, as the bag fills and overflows during the expiratory pause, the deadspace gas is driven back towards the patient. This causes the expiratory valve to open and to discharge the alveolar gas lying at the patient's end of the corrugated tube. If the fresh gas Sow

6 REBREATHING CIRCUITS ?-^ [ used without the occurrence of rebreathing. Using this technique on patients who were breathing spontaneously under light halothane anaesthesia they found that the fresh gas flow could be reduced to two-thirds, and in some cases to half the minute volume, before rebreathing occurred. Norman, Adams and Sykes (1968) repeated this work in conscious volunteers and showed that rebreathing was not seen until the fresh gas flow was approximately 70 per cent of the minute volume. These authors also demonstrated that there was a fall in inspired and alveolar Po a when [~*1 Fresh gas rebreathing occurred. I Dead space gas ( The Mapleson "B" circuit was originally introduced in an attempt to ensure that changes in Alveolar gas fresh gas composition would be rapidly reflected in the alveoli. However, when the circuit diagram is examined it becomes apparent that there is a closed limb terminating in the reservoir bag which is never flushed with fresh gas. The analysis of this system is complicated, but it is obvious that rebreathing will occur if the flow of fresh gas is less than the peak inspiratory flow rate. Under these conditions it is a mixture of alveolar gas and fresh gas which is re-inhaled. A flow of fresh gas more than double the minute volume will, in most clinical circumstances, reduce the degree FIG. 7 Circuits with expiratory valves: spontaneous ventila- of rebreathing to acceptable levels. tion. Disposition of gases at end-expiration. From The Mapleson "C" system (Waters to-and-fro above down, Mapleson "A", "B", "C", "D" circuits. P=patient FGF=fresh gas flow. absorption circuit without a carbon dioxide absorber) may be regarded as similar to system "B" rate is great enough, the deadspace component except that there is no corrugated tubing to will also be washed out through the valve; how- maintain the separation of expired gas into deadever, as mentioned earlier, rebreathing of this space and alveolar components. When rebreathportion of the expired gas will not affect the ing occurs it is therefore a combination of mixed patient. Mapleson (1954) predicted that rebreath- expired gas and fresh gas which is inhaled (fig. 2). ing would not occur if the fresh gas flow rate For safety in clinical practice the fresh gas flow equalled or exceeded the alveolar ventilation. should be greater than twice the minute volume. Since this prediction was based on a number of The Mapleson ad" circuit is similar to the assumptions (the main one being that there would Mapleson "B" except that the expiratory valve be no mixing of alveolar, deadspace and fresh is situated close to the reservoir bag. If the tidal gas), he suggested that, in clinical practice, the volume is less than the volume of the corrugated fresh gas flow should always exceed the minute tube (500 ml) the system behaves as a T-piece. volume. Woolmer and Lind (1954) and Bracken If, however, the tidal volume exceeds this figure and Sanderson (1955) confirmed these predictions mixed expired gas from the bag is re-inhaled. experimentally and De d i v e Lowe (1956) and To prevent rebreathing at normal tidal volumes Davies, Verner and Bracken (1956) confirmed the fresh gas flow should exceed twice the them under clinical conditions. Kain and Nunn patient's minute volume. At fresh gas flow rates (1967) approached the problem by determining below the peak inspiratory flow and with a northe lowest fresh gas flow rate which could be mal pattern of breathing and tidal volume the

7 mt BRITISH JOURNAL OF ANAESTHESIA system is more efficient than the "B" or " C circuit. This is due to the fact that all the fresh gas passes into the patient whereas in the " B " and " C " circuits it is chiefly fresh gas which passes out through the valve during the expiratory pause. When the tidal volume is increased the difference between the circuits becomes less marked. Controlled t>entilation. The Magill attachment behaves very differently when ventilation is controlled (fig. 8). To force consists mainly of deadspace and fresh gas. Rebreathing under these conditions is therefore marked (Sykes, 1959a). The effects can be diminished by increasing the tidal volume or by increasing the fresh gas flow rate. Elimination of carbon dioxide can be further improved by applying a high pressure early in inspiration so that more alveolar gas is discharged to the atmosphere. PrtJsurt FGF-i Flow through valvt -HNSPIRATION* Fresh gas Dead space gas Alveolar gas Pro. 8 Circuits with expiratory valves: controlled ventilation. Disposition of gases at end-expiration. FGF=fresh gas flow. gas into the lungs the expiratory valve spring is tightened and pressure applied to the bag. As the pressure in the circuit is increased gas flows into the lungs, the alveolar gas being the first portion to return. Eventually the pressure under the valve causes it to open and the excess gas is discharged to the atmosphere (fig. 9). This gas -HNSPllWtON- FIG. 9 Flow of gas through expiratory valve with two differently-shaped inspiratory pressure curves. The Mapleson "B" circuit behaves in a manner similar to that described for spontaneous ventilation. However, rebreathing is slightly less since the expiratory valve is closed during the expiratory pause and the fresh gas therefore accumulates at the patient end of the corrugated tubing. When the bag is compressed most of the fresh gas passes into the lungs whilst the mixed expired gas tends to pass out through the valve. Similar remarks apply to the Mapleson " C " system although in this case there is a smaller volume available for the storage of fresh gas during the expiratory pause, most of the fresh gas mixing with the expired gas in the bag. In both these circuits a fresh gas flow of at least twice the minute volume is required to reduce rebreathing to clinically acceptable levels during controlled ventilation. The Mapleson " D " system tends to cause less rebreathing than systems "A", "B" or "C" during controlled ventilation because the valve is situated away from the patient. In the Mapleson "B" circuit some of the fresh gas passes out through the valve during inspiration whereas in

8 REBREATHING CIRCUITS 673 the "D" circuit practically all the fresh gas input is driven into the patient, whilst the gas which is eliminated from die system consists mostly of alveolar gas and deadspace. The order of merit of the four systems considered is therefore "D", "B", "C", "A" (Waters and Mapleson 1961). For this reason Waters (1961) has suggested a combined system using a Mapleson "A" circuit for spontaneous ventilation and a Mapleson "D" system for controlled ventilation. It must be remembered, however, that the tidal volume during controlled ventilation is usually greater than during spontaneous ventilation. The fresh gas flow required to ensure the complete absence of rebreathing may therefore approach 20 l./min. In clinical practice the rebreathing which occurs under conditions of controlled ventilation is usually compensated by the increased alveolar ventilation which is achieved. Smaller fresh gas flow rates than those mentioned may therefore prove acceptable (Nightingale, Richards and Glass, 1965; Marshall and Henderson, 1968). Circuits with Valves controlling the Direction of Gas Flow to and from the Patient. These may be divided into circuits with the valves close to the patient and those with the valves placed remotely from the patient. Valves placed close to the patient are usually termed non-rebreathing valves. Although there are many different designs (Forreger, 1959; Sykes, 1959b) these valves basically consist of an inlet port, which prevents reflux of fresh gas, and an outlet port, which prevents rebreathing of expired gas or dilution with room air. If the valve functions perfectly rebreathing is therefore related to the size of the added apparatus deadspace. Normally this is of the order of ml Unfortunately, in a number of these valves gas leaks past the valve flaps, either because the movement of the flaps is sluggish or because the valve fails to seat correctly (Kerr and Evers, 1958; Loehning, Davis and Safar, 1964). The reflux of patient's expired gas into the inspiratory limb may amount to per cent of the tidal volume if the inlet valve is sluggish and the pressure in the bag is lowered slowly. Fortunately it is usually only the deadspace portion of the expired gas which is re-inhaled; harmful rebreathing therefore only occurs when the valve sticks in the inspiratory position. During controlled respiration some inspired gas often slips past the expiratory valve during the initial part of inspiration. This reduces the tidal volume delivered. During spontaneous ventilation the expiratory valve may stick in the open position so that air may be inhaled. If the expiratory port is connected by a tube (e.g. to an absorber) the gas which is re-inhaled is alveolar gas; harmful rebreathing might therefore occur under these conditions. When the valves are situated at a distance from the patient two additional factors may cause rebreathing. During spontaneous ventilation some mixing of inspired and expired air streams occurs at the T-piece due to turbulence, delay in valve action and the pressure fluctuations in the system. During controlled ventilation this mixing may be accentuated by variations in the elasticity of the breathing tubes and by differences in the compressible volume on the two sides of the circuit Finally, the elimination of carbon dioxide from circle systems without an absorber must be considered. This depends largely on the fresh gas inflow and on the position of the expiratory valve. If the valve is situated close to the patient's mouth and the fresh gas flow equals the patient's minute volume almost complete elimination of alveolar carbon dioxide may occur. However, if the valve is placed at some distance from the patient considerable rebreathing may occur unless the fresh gas flow is increased (Brown, Seniff and Ham, 1964). CONCLUSION The elimination of carbon dioxide from semidosed circuits without absorbers depends on the fresh gas flow, tidal volume and to a lesser extent on the pattern of breathing. Some circuits arc surprisingly efficient; this is because the expired carbon dioxide is concentrated in the alveolar portion of the expired gas. If the design of the circuit is such that this portion is eliminated through the expiratory valve rebreathing will be minimal Circuits which do not maintain the separation between fresh gas, deadspace gas and alveolar gas tend to be inefficient. REFERENCES Adriani, J. (1960). Disposal of carbon dioxide from devices used for inhalations] anesthesia. Anetthesiology, 21, 742.

9 674 BRITISH JOURNAL OF ANAESTHESIA Ayre, P. (1937a). Anaesthesia for intracranial operations. Lancet, 1, 561. (1937b). Endotracheal anesthesia for babies: with special reference to hare-lip and cleft palate operations. Curr. Res. Anesth., 16, 330. (1956). The T-piece technique. Brit. J. Anaesth., 28, 520. (1967.) Theme and variations (on a T-piece). Anaesthesia, 22, 359. Bethune, D. W., and Collis, J. M. (1967). The evaluation of oxygen masks (a mechanical method). Anaesthesia, 22, 43. Bracken, A., and Sanderson, D. M. (1955). Carbon dioxide concentrations found in various anaesthetic circuits. Brit. J. Anaesth., 27, 428. Briscoe, W. A., Forster, R. E., and Comroe, J. H. (1954). Alveolar ventilation at very low tidal volumes. J. appl. Physiol., 7, 27. Brooks, W., Stuart, P., and Gabel, P. V. (1958). The T-piece technique in anesthesia. Curr. Res. Anesth., 37, 191. Brown, E. S., Seniff, A. M., and Elam, J. O. (1964). Carbon dioxide elimination in semiclosed systems. Anesthesiology, 25, 31. Burton, J. D. K. (1962). Effects of dry anaesthetic gases on the respiratory mucous membrane. Lancet, 1, 235. Oappison, G. B., and Hamilton, W. K. (1956). Respiratory adjustments to increases in external dead space. Anesthesiology, 17, 643. Clarke, A. D. (1958). Potential deadspace in an anaesthetic mask and connectors. Brit. J. Anaesth., 30, 176. Davies, R. M., Verner, I. R., and Bracken, A. (1956). Carbon dioxide elimination from semiclosed systems. Brit. J. Anaesth., 28, 196. De Clive Lowe, S. G. (1956). Carbon dioxide in anaesthesia. Proc. roy. Soc. Med., 49, 215. Domaingue, F. G. (1951). Rebreathing in a semiclosed system. Brit. J. Anaesth., 23, 249. Elam, J. O., and Brown, E. S. (1956). Carbon dioxide homeostasis during anesthesia. Ill: Ventilation and carbon dioxide elimination. Anesthesiology, 17, 116. Fitton, E. P. (1958). A theoretical investigation of orygen concentrations in gases inspired from various semiclosed anaesthetic systems. Brit. J. Anaesth., 30, 269. (1963). Theoretical investigation of oxygen concentrations: correction and additional observations. Brit. J. Anaesth., 35, 276. Forregcr, R. (1959). The classification and performance of respiratory valves. Anesthesiology, 20, 296. Harrison, G. A. (1964a). The effect of the respiratory flow pattern on rebreathing in a T-piece system. Brit. J. Anaesth., 36, 206. (1964b). Ayre's T-piece: a review of its modifications. Brit. J. Anaesth., 36, 115. Harrison, G. G., Ozinsky, J., and Jones, C. S. (1959). Choice of an anaesthetic facepiece. Brit. J. Anaesth., 31, 269. Inkster, J. S. (1956). The T-piece technique in anaesthesia. Brit. J. Anaesth., 28, 512. Kain, M. L., and Nunn, J. F. (1967). Fresh gas flow and rebreathing in the Magill circuit with spontaneous respiration. Proc. roy. Soc. Med., 60, 749. Kerr, J. H., and Evers, J. L. (1958). Carbon dioxide accumulation: valve leaks and inadequate absorption. Canad. Anaesth. Soc. J., 5, 154. Lewis, A., and Spoerel, A. (1961). A modification of Ayre's technique. Canad. Anaesth. Soc. J., 8, 501. Loshning, R. W., Davis, A., and Safar, P. (1964). Rebreathing with "non-rebreathing" valves. Anesthesiology, 25, 854. Mapkson, W. W. (1954). The elimination of rebreathing in various semiclosed anaesthetic systems. Brit. J. Anaesth., 26, 323. (1958). Theoretical considerations of the effect of rebreathing in two semi-closed anaesthetic systems. Brit. med. Bull., 14, 64. Marshall, M, and Henderson, G. A. (1968). Positive presure ventilation using a semiclosed system: a reassessment. Brit. J. Anaesth., 40, 265. Molyneux, L., and Pask, E. A. (1951). The flow of gases in a semiclosed anaesthetic system. Brit. J. Anaesth., 23, 81. Nightingale, D. A., Richards, C C, and Glass, A. (1965). An evaluation of rebreathing in a modified T-piece system during controlled ventilation of anaesthetized children. Brit. J. Anaesth., 37, 762. Norman, J., Adams, A. P., and Sykes, M. K. (1968). Rebreathing with the Magill attachment. Anaesthesia, 23, 75. Nunn, J. F., and Newman, H. C. (1964). Inspired gas, rebreathing and apparatus deadspace. Brit. J. Anaesth., 36, 5. Onchi, Y., Hayashi, T, and Ueyama, M. (1957). Studies on the Ayre T-piece technique. Far East J. Anaesth., 1, 8, 30. Picken, D. K. W. (1950). An endotracheal connector for infants. Brit. med. J., 1, 954. Rees, G. J. (1950). Anaesthesia in the newborn. Brit, med. J., 2, Roffey, P. J., Revell, D. G., and Morris, L. E. (1961). An assessment of the Revell circulator. Anesthesiology, 22, 583. Ruben, H. M. (1953). Concerning the concentration of inhaled gases in semiclosed anesthesia systems. Anesthesiology, 14, 459. Swam, C H., Adriani, J., and Mih, A. (1953). Semiclosed inhalers: studies of oxygen and carbon dioxide tensions during varying conditions of use. Anesthesiology, 14, 437. Sykes, M. K. (1959a). Rebreathing during controlled respiration with the Magill attachment Brit. J. Anaesth., 31, 247. (1959b). Non-rebreathing valves. Brit. J. Anaesth., 31, 450. (1967). Aspects of dead space in anaesthetic practice and artificial pulmonary ventilation. Chapter in Modem Trends in Anaesthesia (eds. Evans and Gray), vol. 3, p. 80. London: Butterworths. Voss, T. J. V. (1963). Deadspace in paediatric anaesthetic apparatus. Brit. J. Anaesth., 35, 454. Waters, D. J. (1961). A composite semidosed anaesthetic system suitable for controlled or spontaneous respiration. Brit. J. Anaesth., 33, 417. Mapleson, W. W. (1961). Rebreathing during controlled respiration with various semiclosed anaesthetic systems. Brit. J. Anaesth., 33, 374. Woolmer, R., and Land, B. (1954). Rebreathing with a semiclosed system. Brit. J. Anaesth., 26, 316. Wynne, R. L. (1941). Mechanism of partial rebreathing in anaesthesia. Brit. med. J., 1, 155.

CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS

CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS Brit. J. Anaesth. (1956), 28, 196 CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS BY RUSSELL M. DAVIES, I. R. VERNER Queen Victoria Hospital, East Grinstead AND A. BRACKEN Research and Development Centre,

More information

Breathing Systems. Professor Khalid Bashir

Breathing Systems. Professor Khalid Bashir Breathing Systems Professor Khalid Bashir Introduction The devices that connect and deliver anaesthetic gases from anaesthetic machine to patient Spontaneous respiration / IPPV Ports for gas sampling,

More information

Breathing systems. Non-rebreathing system using a unidirectional valve pair. Non-rebreathing systems

Breathing systems. Non-rebreathing system using a unidirectional valve pair. Non-rebreathing systems Breathing systems A Timothy Lovell By convention, rebreathing in an anaesthetic system refers to inhalation of some or all of the previously exhaled gases, including carbon dioxide and water. This is often

More information

Breathing Circuits. Product training

Breathing Circuits. Product training Breathing Circuits Product training Agenda Introduction to Breathing Circuits Anaesthesia Circuits Intensive Care Circuits Accessories 2 Covidien Introduction 3 Covidien What is a breathing circuit? In

More information

HALOTHANE AND REBREATHING. Research Department of Anaesthetics, Royal College of Surgeons of England

HALOTHANE AND REBREATHING. Research Department of Anaesthetics, Royal College of Surgeons of England Brit. J. Anaesth. (1958), 30, 555 HALOTHANE AND REBREATHING BY H. C. NEWMAN Research Department of Anaesthetics, Royal College of Surgeons of England THIS paper presents some elementary theoretical concepts

More information

Selecting and Connecting Breathing Systems

Selecting and Connecting Breathing Systems Selecting and Connecting Breathing Year Group: BVSc3 + Document number: CSL_A03 Equipment for this station: Equipment list: Pen Paper Calculator T-piece (in CSL a strip of white tape is around this system)

More information

Anesthesiology Research and Scholarship. Follow this and additional works at:

Anesthesiology Research and Scholarship. Follow this and additional works at: University of New Mexico UNM Digital Repository Anesthesiology Research and Scholarship Anesthesiology 9-1-2013 Breathing Systems Janet Brierley Follow this and additional works at: https://digitalrepository.unm.edu/anesthesiology_pubs

More information

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT IV - 2013 MECHANICAL VENTILATORS Prof. Yasser Mostafa Kadah Mechanical Ventilator A ventilator is a machine, a system of related elements designed to alter, transmit, and direct energy

More information

Lung Volumes and Capacities

Lung Volumes and Capacities Lung Volumes and Capacities Normally the volume of air entering the lungs during a single inspiration is approximately equal to the volume leaving on the subsequent expiration and is called the tidal volume.

More information

Chapter 4: Ventilation Test Bank MULTIPLE CHOICE

Chapter 4: Ventilation Test Bank MULTIPLE CHOICE Instant download and all chapters Test Bank Respiratory Care Anatomy and Physiology Foundations for Clinical Practice 3rd Edition Will Beachey https://testbanklab.com/download/test-bank-respiratory-care-anatomy-physiologyfoundations-clinical-practice-3rd-edition-will-beachey/

More information

Section Two Diffusion of gases

Section Two Diffusion of gases Section Two Diffusion of gases Lecture 5: Partial pressure and the composition of gasses in air. Factors affecting diffusion of gases. Ventilation perfusion ratio effect on alveolar gas concentration.

More information

Capnography in the Veterinary Technician Toolbox. Katie Pinner BS, LVT Bush Advanced Veterinary Imaging Richmond, VA

Capnography in the Veterinary Technician Toolbox. Katie Pinner BS, LVT Bush Advanced Veterinary Imaging Richmond, VA Capnography in the Veterinary Technician Toolbox Katie Pinner BS, LVT Bush Advanced Veterinary Imaging Richmond, VA What are Respiration and Ventilation? Respiration includes all those chemical and physical

More information

Flow meter. bellow vaporizer. APL valve. Scavenging system

Flow meter. bellow vaporizer. APL valve. Scavenging system Introductory Lecture Series: The Anesthesia Machine PORNSIRI WANNADILOK Objectives Anesthesia Machine Ventilators Scavenging Systems System Checkout 1 Flow meter ventilator bellow vaporizer Corrugated

More information

Let s talk about Capnography

Let s talk about Capnography Let s talk about Capnography This is one of a series of articles by Keith Simpson BVSc MRCVS MIET (Electronics) discussing the practical aspects of some common monitoring techniques. Capnometry is the

More information

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS 1. INTRODUCTION An anaesthesia delivery system includes any machine, equipment or apparatus which supplies gases, vapours, local anaesthesia and/or

More information

Anaesthesia breathing circuits

Anaesthesia breathing circuits 98 Anaesthesia breathing circuits John W.R. Mclntyre FFARCS (Eng) FRCPC During clinical anaesthesia gases pass through a complex conduit interposed between the anaesthesia machine and the patient. This

More information

Collin County Community College. Lung Physiology

Collin County Community College. Lung Physiology Collin County Community College BIOL. 2402 Anatomy & Physiology WEEK 9 Respiratory System 1 Lung Physiology Factors affecting Ventillation 1. Airway resistance Flow = Δ P / R Most resistance is encountered

More information

Recommendations on Checking Anaesthesia Delivery Systems

Recommendations on Checking Anaesthesia Delivery Systems Page 1 of 11 Recommendations on Checking Anaesthesia Delivery Version Effective Date 1 Oct 1992 (reviewed Feb 07, Feb 02) 2 2004 3 Nov 2011 4 Dec 2016 Document No. HKCA T1 v4 Prepared by College Guidelines

More information

Medical Instruments in the Developing World

Medical Instruments in the Developing World 2.2 Ventilators 2.2.1 Clinical Use and Principles of Operation Many patients in an intensive care and the operating room require the mechanical ventilation of their lungs. All thoracic surgery patients,

More information

J.A. BAIN AND W.E. SPOEREL~

J.A. BAIN AND W.E. SPOEREL~ PREDICTION OF ARTERIAL CARBON DIOXIDE TENSION DURING CONTROLLED VENTILATION WITH A MODIFIED MAPLESON D SYSTEM* J.A. BAIN AND W.E. SPOEREL~ ThE OBSERVATION that relatively low fresh gas inflows are adequate

More information

By: Aseel Jamil Al-twaijer. Lec : physical principles of gas exchange

By: Aseel Jamil Al-twaijer. Lec : physical principles of gas exchange By: Aseel Jamil Al-twaijer Lec : physical principles of gas exchange Date:30 /10/2017 this lecture is about the exchange of gases between the blood and the alveoli. I might add some external definitions

More information

Monitoring, Ventilation & Capnography

Monitoring, Ventilation & Capnography Why do we need to monitor? Monitoring, Ventilation & Capnography Keith Simpson BVSc MRCVS MIET(Electronics) Torquay, Devon. Under anaesthesia animals no longer have the ability to adequately control their

More information

PICU Resident Self-Study Tutorial The Basic Physics of Oxygen Transport. I was told that there would be no math!

PICU Resident Self-Study Tutorial The Basic Physics of Oxygen Transport. I was told that there would be no math! Physiology of Oxygen Transport PICU Resident Self-Study Tutorial I was told that there would be no math! INTRODUCTION Christopher Carroll, MD Although cells rely on oxygen for aerobic metabolism and viability,

More information

The Anesthesia Machine (con t) The primary cause of machine malfunction is failure to check Never start without the American Express items What is the

The Anesthesia Machine (con t) The primary cause of machine malfunction is failure to check Never start without the American Express items What is the Anesthesia Machine The Anesthesia Machine You are the master of the machine You are responsible for checking the machine prior to each case 1 The Anesthesia Machine (con t) The primary cause of machine

More information

have expressed widely varying opinions on this topic (Catterall and Snow, 1960; Catterall, 1960;

have expressed widely varying opinions on this topic (Catterall and Snow, 1960; Catterall, 1960; An evaluation Thorax (1967), 22, 221. of oxygen therapy equipment Experimental study of various devices on the human subject D. W. BETHUNE AND J. M. COLLIS From the Department of Anaesthesia, St. Bartholomew's

More information

RESPIRATORY REGULATION DURING EXERCISE

RESPIRATORY REGULATION DURING EXERCISE RESPIRATORY REGULATION DURING EXERCISE Respiration Respiration delivery of oxygen to and removal of carbon dioxide from the tissue External respiration ventilation and exchange of gases in the lung Internal

More information

Ventilators. Dr Simon Walton Consultant Anaesthetist Eastbourne DGH KSS Basic Science Course

Ventilators. Dr Simon Walton Consultant Anaesthetist Eastbourne DGH KSS Basic Science Course Ventilators Dr Simon Walton Consultant Anaesthetist Eastbourne DGH KSS Basic Science Course Objectives Discuss Classification/ terminology Look at Modes of ventilation How some specific ventilators work

More information

CLINICAL ANAESTHESIA AND THE MULTIPLE-GAUZE CONDENSER-HUMIDIFIER

CLINICAL ANAESTHESIA AND THE MULTIPLE-GAUZE CONDENSER-HUMIDIFIER Br. J. Anaesth. (1974), 46, 773 CLINICAL ANAESTHESIA AND THE MULTIPLE-GAUZE CONDENSER-HUMIDIFIER C. A. SHANKS SUMMARY Airway temperatures and humidities were examined during adult anaesthesia before and

More information

DPA 02 TM Diamedica Portable Anaesthesia System INSTRUCTIONS AND USER GUIDE

DPA 02 TM Diamedica Portable Anaesthesia System INSTRUCTIONS AND USER GUIDE DPA 02 TM Diamedica Portable Anaesthesia System INSTRUCTIONS AND USER GUIDE INSTRUCTIONS FOR PORTABLE ANAESTHETIC MACHINE DPA 02 TM The Diamedica Portable Anaesthetic machine DPA 2 TM has three principal

More information

Unit II Problem 4 Physiology: Diffusion of Gases and Pulmonary Circulation

Unit II Problem 4 Physiology: Diffusion of Gases and Pulmonary Circulation Unit II Problem 4 Physiology: Diffusion of Gases and Pulmonary Circulation - Physical principles of gases: Pressure of a gas is caused by the movement of its molecules against a surface (more concentration

More information

GASEOUS HOMEOSTASIS AND THE CIRCLE SYSTEM

GASEOUS HOMEOSTASIS AND THE CIRCLE SYSTEM Br. J. Anaesth. (1986), 58, 1167-1180 GASEOUS HOMEOSTASIS AND THE CIRCLE SYSTEM Factors Influencing Anaesthetic Gas Exchange C. M. CONWAY When an inhaled anaesthetic agent is administered at a constant

More information

Volumetric Gas Measurements Measurement Site Considerations

Volumetric Gas Measurements Measurement Site Considerations W H I T E P A P E R Volumetric Gas Measurements Measurement Site Considerations ABSTRACT For the measurement of gas exchange (such as carbon dioxide elimination), the choice of the gas and flow measurement

More information

UNIQUE CHARACTERISTICS OF THE PULMONARY CIRCULATION THE PULMONARY CIRCULATION MUST, AT ALL TIMES, ACCEPT THE ENTIRE CARDIAC OUTPUT

UNIQUE CHARACTERISTICS OF THE PULMONARY CIRCULATION THE PULMONARY CIRCULATION MUST, AT ALL TIMES, ACCEPT THE ENTIRE CARDIAC OUTPUT UNIQUE CHARACTERISTICS OF THE PULMONARY CIRCULATION THE PULMONARY CIRCULATION MUST, AT ALL TIMES, ACCEPT THE ENTIRE CARDIAC OUTPUT UNIQUE CHARACTERISTICS OF THE PULMONARY CIRCULATION THE PULMONARY CIRCULATION

More information

Unit 15 Manual Resuscitators

Unit 15 Manual Resuscitators 15-1 Unit 15 Manual Resuscitators GOAL On completion of this unit, the student should comprehend the proper operation of self-inflating resuscitation bags, flow-inflating resuscitation bags and gas-powered

More information

Respiration (revised 2006) Pulmonary Mechanics

Respiration (revised 2006) Pulmonary Mechanics Respiration (revised 2006) Pulmonary Mechanics PUL 1. Diagram how pleural pressure, alveolar pressure, airflow, and lung volume change during a normal quiet breathing cycle. Identify on the figure the

More information

Sign up to receive ATOTW weekly -

Sign up to receive ATOTW weekly - THE PHYSICS OF FLOW ANAESTHESIA TUTORIAL OF THE WEEK 84 9TH APRIL 2008 Paul Clements, SpR in Anaesthetics, Hope Hospital, Salford, UK. Carl Gwinnutt, Consultant Anaesthetist, Hope Hospital, Salford, UK.

More information

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting This is a quick reference for setting up, using, and troubleshooting the Hallowell EMC 2000 and 2002 Small Animal Ventilator.

More information

TRICHLOROETHYLENE CONCENTRATION FROM A " BOYLE " TYPE ANAESTHETIC APPARATUS. Department of Anaesthetics, Welsh National School of Medicine, Cardiff

TRICHLOROETHYLENE CONCENTRATION FROM A  BOYLE  TYPE ANAESTHETIC APPARATUS. Department of Anaesthetics, Welsh National School of Medicine, Cardiff Brit. J. Anaesth. (1957), 29, 3 CONCENTRATN FROM A " BOYLE " TYPE ANAESTHETIC APPARATUS BY W. W. MAPLESON Department of Anaesthetics, Welsh National School of Medicine, Cardiff INTRODUCTN GILCHRIST and

More information

Respiratory System. Part 2

Respiratory System. Part 2 Respiratory System Part 2 Respiration Exchange of gases between air and body cells Three steps 1. Ventilation 2. External respiration 3. Internal respiration Ventilation Pulmonary ventilation consists

More information

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE Surgical Care at the District Hospital 1 15 Anesthetic Infrastructure and Supplies Key Points 2 15.1 Equipment & Supplies for Different Level Hospitals However well trained you are as an anaesthetist,

More information

Mechanical Ventilation

Mechanical Ventilation Mechanical Ventilation Chapter 4 Mechanical Ventilation Equipment When providing mechanical ventilation for pediatric casualties, it is important to select the appropriately sized bag-valve mask or endotracheal

More information

(Received 16 January 1946)

(Received 16 January 1946) 186 J. Physiol. (I946) I05, I86-I90 6I2.2I5.9 THE ABSORPTION OF FLUIDS FROM THE LUNGS BY F. C. COURTICE AND P. J. PHIPPS From the Experimental Station, Porton and the Laboratory of Physiology, Oxford (Received

More information

CHAPTER 3: The respiratory system

CHAPTER 3: The respiratory system CHAPTER 3: The respiratory system Practice questions - text book pages 56-58 1) When the inspiratory muscles contract, which one of the following statements is true? a. the size of the thoracic cavity

More information

2. State the volume of air remaining in the lungs after a normal breathing.

2. State the volume of air remaining in the lungs after a normal breathing. CLASS XI BIOLOGY Breathing And Exchange of Gases 1. Define vital capacity. What is its significance? Answer: Vital Capacity (VC): The maximum volume of air a person can breathe in after a forced expiration.

More information

The physiological functions of respiration and circulation. Mechanics. exercise 7. Respiratory Volumes. Objectives

The physiological functions of respiration and circulation. Mechanics. exercise 7. Respiratory Volumes. Objectives exercise 7 Respiratory System Mechanics Objectives 1. To explain how the respiratory and circulatory systems work together to enable gas exchange among the lungs, blood, and body tissues 2. To define respiration,

More information

Initiation and Management of Airway Pressure Release Ventilation (APRV)

Initiation and Management of Airway Pressure Release Ventilation (APRV) Initiation and Management of Airway Pressure Release Ventilation (APRV) Eric Kriner RRT Pulmonary Critical Care Clinical Specialist Pulmonary Services Department Medstar Washington Hospital Center Disclosures

More information

Anaesthetic infrastructure and supplies

Anaesthetic infrastructure and supplies 15 Anaesthetic infrastructure and supplies Key Points 15.1 EQUIPMENT AND SUPPLIES FOR DIFFERENT LEVEL HOSPITALS However well trained you are as an anaesthetist, your ability to provide safe anaesthesia

More information

6 th Accredited Advanced Mechanical Ventilation Course for Anesthesiologists. Course Test Results for the accreditation of the acquired knowledge

6 th Accredited Advanced Mechanical Ventilation Course for Anesthesiologists. Course Test Results for the accreditation of the acquired knowledge 6 th Accredited Advanced Mechanical Ventilation Course for Anesthesiologists Course Test Results for the accreditation of the acquired knowledge Q. Concerning the mechanics of the newborn s respiratory

More information

Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo

Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo Link full download: http://testbankcollection.com/download/pilbeams-mechanicalventilation-physiological-and-clinical-applications-5th-edition-test-bank-cairo

More information

(a) (i) Describe how a large difference in oxygen concentration is maintained between a fish gill and the surrounding water.

(a) (i) Describe how a large difference in oxygen concentration is maintained between a fish gill and the surrounding water. 1. Answers should be written in continuous prose. Credit will be given for biological accuracy, the organisation and presentation of information and the way in which an answer is expressed. Fick s law

More information

Question 1: Define vital capacity. What is its significance? Vital capacity is the maximum volume of air that can be exhaled after a maximum inspiration. It is about 3.5 4.5 litres in the human body. It

More information

CHAPTER 3: The cardio-respiratory system

CHAPTER 3: The cardio-respiratory system : The cardio-respiratory system Exam style questions - text book pages 44-45 1) Describe the structures involved in gaseous exchange in the lungs and explain how gaseous exchange occurs within this tissue.

More information

Respiratory Physiology Gaseous Exchange

Respiratory Physiology Gaseous Exchange Respiratory Physiology Gaseous Exchange Session Objectives. What you will cover Basic anatomy of the lung including airways Breathing movements Lung volumes and capacities Compliance and Resistance in

More information

(Received 9 September 1940)

(Received 9 September 1940) 257 J. Physiol. (I 94I) 99, 257-264 6I2.2II A METHOD OF RECORDING THE RESPIRATION BY J. H. GADDUM From the College of the Pharmaceutical Society, 17 Bloomsbury Square, London, W.C. 2 (Received 9 September

More information

APPENDIX. working blood volume was also rather large; Evans, Grande, and. equilibrated to the new mixture is partially dependent upon the rate

APPENDIX. working blood volume was also rather large; Evans, Grande, and. equilibrated to the new mixture is partially dependent upon the rate 612.172-5 APPENDIX A SIMPLIFIED HEART OXYGENATOR CIRCUIT FOR BLOOD- FED HEARTS. By J. YULE BOG-UE and R. A. GREGORY.' SINCE 1934 studies on the carbohydrate metabolism of the blood-fed heart without lungs

More information

VENTILATION AND PERFUSION IN HEALTH AND DISEASE. Dr.HARIPRASAD VS

VENTILATION AND PERFUSION IN HEALTH AND DISEASE. Dr.HARIPRASAD VS VENTILATION AND PERFUSION IN HEALTH AND DISEASE Dr.HARIPRASAD VS Ventilation Total ventilation - total rate of air flow in and out of the lung during normal tidal breathing. Alveolar ventilation -represents

More information

VENTILATORS PURPOSE OBJECTIVES

VENTILATORS PURPOSE OBJECTIVES VENTILATORS PURPOSE To familiarize and acquaint the transfer Paramedic with the skills and knowledge necessary to adequately maintain a ventilator in the interfacility transfer environment. COGNITIVE OBJECTIVES

More information

Manual: Biphasic Positive Airway Pressure (BiPAP) Ventilation

Manual: Biphasic Positive Airway Pressure (BiPAP) Ventilation RCH@Home Manual: Biphasic Positive Airway Pressure (BiPAP) Ventilation 1. Commonly used terms... 2 1.1 Inspiration... 2 1.2 Expiration... 2 1.3 Breath rate (bpm)... 2 1.4 Ventilation... 2 1.5 Biphasic

More information

I Physical Principles of Gas Exchange

I Physical Principles of Gas Exchange Respiratory Gases Exchange Dr Badri Paudel, M.D. 2 I Physical Principles of Gas Exchange 3 Partial pressure The pressure exerted by each type of gas in a mixture Diffusion of gases through liquids Concentration

More information

EXCESSIVE WORK OF BREATHING DURING INTERMITTENT MANDATORY VENTILATION

EXCESSIVE WORK OF BREATHING DURING INTERMITTENT MANDATORY VENTILATION Br. J. Anaesth. (1986), 58, 1048-1054 EXCESSIVE WORK OF BREATHING DURING INTERMITTENT MANDATORY VENTILATION J. S. MECKLENBURGH, I. P. LATTO, T. A. A. AL-OBAIDI, E. A. SWAI AND W. W. MAPLESON Ventilators

More information

DURING the course of certain investigations it became

DURING the course of certain investigations it became VOLUMETRIC DETERMINATION OF ETHER OR CYCLOPROPANE, CARBON DIOXIDE, NITROUS OXIDE AND OXYGEN IN ANESTHETIC MIXTURES By F. J. PRIME DURING the course of certain investigations it became necessary to be able

More information

VOLUNTARY BREATHHOLDING. I. PULMONARY GAS

VOLUNTARY BREATHHOLDING. I. PULMONARY GAS VOLUNTARY BREATHHOLDING. I. PULMONARY GAS EXCHANGE DURING BREATHHOLDING'1 By CHARLES D. STEVENS, EUGENE B. FERRIS, JOSEPH P. WEBB, GEORGE L. ENGEL, AND MYRTLE LOGAN (From the Departments of Internal Medicine

More information

Figure 1. A schematic diagram of the human respiratory system.

Figure 1. A schematic diagram of the human respiratory system. Introduction to Respiration In this experiment, you will investigate various aspects of normal breathing, hyperventilation, rebreathing the effect of changing airway resistance and ways in which to measure

More information

FRESH GAS UTILIZATION OF EIGHT CIRCLE SYSTEMS

FRESH GAS UTILIZATION OF EIGHT CIRCLE SYSTEMS British Journal of Anaesthesia 1991; 67: 492^199 FRESH GAS UTILIZATION OF EIGHT CIRCLE SYSTEMS A. M. ZBINDEN, P. FEIGENWINTER AND M. HUTMACHER SUMMARY The fresh gas utilization (FGU) of a semi-closed breathing

More information

TERMINOLOGY AND SYMBOLS USED LN RESPIRATORY PHYSIOLOGY. Assistant, Medical Unit, Middlesex Hospital, London, W.i

TERMINOLOGY AND SYMBOLS USED LN RESPIRATORY PHYSIOLOGY. Assistant, Medical Unit, Middlesex Hospital, London, W.i Brit. J. Anaesth. (1957), 29, 534 TERMINOLOGY AND SYMBOLS USED LN RESPIRATORY PHYSIOLOGY BY E. J. MORAN CAMPBELL Assistant, Medical Unit, Middlesex Hospital, London, W.i INTRODUCTION MANY anaesthetists

More information

Homecare Pneumology NEONATOLOGY ANAESTHESIA INTENSIVE CARE VENTILATION Step beyond the future

Homecare Pneumology NEONATOLOGY ANAESTHESIA INTENSIVE CARE VENTILATION Step beyond the future Homecare Pneumology NEONATOLOGY ANAESTHESIA INTENSIVE CARE VENTILATION Step beyond the future Sleep Diagnostics Service Patient Support The future starts now! For over 30 years, Heinen + Löwenstein has

More information

I.CHEM.E. SYMPOSIUM SERIES NO. 97 BUOYANCY-DRIVEN NATURAL VENTILATION OP ENCLOSED SPACES

I.CHEM.E. SYMPOSIUM SERIES NO. 97 BUOYANCY-DRIVEN NATURAL VENTILATION OP ENCLOSED SPACES BUOYANCY-DRIVEN NATURAL VENTILATION OP ENCLOSED SPACES M. R. Marshall* and P. L. Stewart-Darling* A simple mathematical model for the buoyancy driven ventilation of an enclosed space, using a two-pipe

More information

BREATH-BY-BREATH METHOD

BREATH-BY-BREATH METHOD BREATH-BY-BREATH METHOD COR-MAN-0000-005-IN / EN Issue A, Rev. 2 2013-07 INNOISION ApS Skovvænge DK-5620 Glamsbjerg Denmark Tel.: +45 65 95 91 00 Fax: +45 65 95 78 00 info@innovision.dk www.innovision.dk

More information

THE PERFORMANCE OF A WALTON FIVE ANAESTHETIC MACHINE

THE PERFORMANCE OF A WALTON FIVE ANAESTHETIC MACHINE Brit. J. Anaesth. (1961), 33, 440 THE PERFORMANCE OF A WALTON FIVE ANAESTHETIC MACHINE BY W. D. A. SMITH Research Department of Anaesthetics, Royal College of Surgeons of England London, W.C.I THE Walton

More information

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products Saturn Evo SATURN EVO COLOR A N A E S T H E S I A S Y S T E M i n t e l l i g e n t s i m p l i c i t y Manufacturer of high quality anaesthesia products SATURN EVO COLOR PATIENT SAFETY AND USER PATIENT

More information

BASIC PHYSICS APPLIED TO ANAESTHESIOLOGY

BASIC PHYSICS APPLIED TO ANAESTHESIOLOGY BASIC PHYSICS APPLIED TO ANAESTHESIOLOGY Dr.R.Selvakumar.M.D.D.A.DNB Professor of Anaesthesiology, K.A.P.Viswanatham Govt medical college, Trichy. The current practice of Anaesthesiology demands knowledge

More information

Selecting the Ventilator and the Mode. Chapter 6

Selecting the Ventilator and the Mode. Chapter 6 Selecting the Ventilator and the Mode Chapter 6 Criteria for Ventilator Selection Why does the patient need ventilatory support? Does the ventilation problem require a special mode? What therapeutic goals

More information

RESPIRATORY MUSCLES IN HEALTH AND EMPHYSEMA *

RESPIRATORY MUSCLES IN HEALTH AND EMPHYSEMA * THE OXYGEN CONSUMPTION AND EFFICIENCY OF THE RESPIRATORY MUSCLES IN HEALTH AND EMPHYSEMA * BY REUBEN M. CHERNIACK t (From The Winnipeg General Hospital and the Departments of Medicine and Physiology and

More information

Mechanical Ventilation

Mechanical Ventilation PROCEDURE - Page 1 of 5 Purpose Scope Physician's Order Indications Procedure Mechanical Artificial Ventilation refers to any methods to deliver volumes of gas into a patient's lungs over an extended period

More information

RESPIRATORY PHYSIOLOGY. Anaesthesiology Block 18 (GNK 586) Prof Pierre Fourie

RESPIRATORY PHYSIOLOGY. Anaesthesiology Block 18 (GNK 586) Prof Pierre Fourie RESPIRATORY PHYSIOLOGY Anaesthesiology Block 18 (GNK 586) Prof Pierre Fourie Outline Ventilation Diffusion Perfusion Ventilation-Perfusion relationship Work of breathing Control of Ventilation 2 This image

More information

Concentration and second-gas effects in the water analogue

Concentration and second-gas effects in the water analogue British Journal of Anaesthesia 1998; 81: 837 843 CLINICAL INVESTIGATIONS Concentration and second-gas effects in the water analogue W. W. MAPLESON AND B. KORMAN Summary The water analogue provides a visual

More information

Lab 3. The Respiratory System (designed by Heather E. M. Liwanag with T.M. Williams)

Lab 3. The Respiratory System (designed by Heather E. M. Liwanag with T.M. Williams) Name Lab Partners Lab 3. The Respiratory System (designed by Heather E. M. Liwanag with T.M. Williams) Part 1. Lung Volumes and Capacities Objectives 1. Obtain graphical representation of lung capacities

More information

Chapter 37: Pulmonary Ventilation. Chad & Angela

Chapter 37: Pulmonary Ventilation. Chad & Angela Chapter 37: Pulmonary Ventilation Chad & Angela Respiratory Structures Basic Structures of Respiration Nasal/Oral Cavities Larynx Trachea Bronchi Secondary Bronchi Bronchioles Alveoli Mechanics of Ventilation

More information

Respiratory Medicine. A-A Gradient & Alveolar Gas Equation Laboratory Diagnostics. Alveolar Gas Equation. See online here

Respiratory Medicine. A-A Gradient & Alveolar Gas Equation Laboratory Diagnostics. Alveolar Gas Equation. See online here Respiratory Medicine A-A Gradient & Alveolar Gas Equation Laboratory Diagnostics See online here Alveolar gas equation helps to calculate the partial pressure of oxygen in alveoli and A-a gradient is the

More information

BREATHING AND EXCHANGE OF GASES

BREATHING AND EXCHANGE OF GASES 96 BIOLOGY, EXEMPLAR PROBLEMS CHAPTER 17 BREATHING AND EXCHANGE OF GASES MULTIPLE CHOICE QUESTIONS 1. Respiration in insects is called direct because a. The cell exchange O 2 directly with the air in the

More information

1) Kety and others have attempted to predict

1) Kety and others have attempted to predict BY J. W. SEVERINGHAUS 2 (From the Department of Anesthesia, Hospital of the University of Pennsylvania, and Harrison Department of Surgical Research, University of Pennsylvania, Philadelphia, Pa.) (Submitted

More information

Chapter 9 Airway Respirations Metabolism Oxygen Requirements Respiratory Anatomy Respiratory Anatomy Respiratory Anatomy Diaphragm

Chapter 9 Airway Respirations Metabolism Oxygen Requirements Respiratory Anatomy Respiratory Anatomy Respiratory Anatomy Diaphragm 1 Chapter 9 Airway 2 Respirations Every cell of the body requires to survive Oxygen must come in and carbon must go out 3 Metabolism Metabolism--Process where the body s cells convert food to Adequate

More information

ALVEOLAR - BLOOD GAS EXCHANGE 1

ALVEOLAR - BLOOD GAS EXCHANGE 1 ALVEOLAR - BLOOD GAS EXCHANGE 1 Summary: These notes examine the general means by which ventilation is regulated in terrestrial mammals. It then moves on to a discussion of what happens when someone over

More information

OXYGEN THERAPY. (Non-invasive O2 therapy in patient >8yrs)

OXYGEN THERAPY. (Non-invasive O2 therapy in patient >8yrs) OXYGEN THERAPY (Non-invasive O2 therapy in patient >8yrs) Learning aims Indications and precautions for O2 therapy Targets of therapy Standard notation O2 delivery devices Taps, tanks and tubing Notation

More information

Pulmonary Circulation Linda Costanzo Ph.D.

Pulmonary Circulation Linda Costanzo Ph.D. Pulmonary Circulation Linda Costanzo Ph.D. OBJECTIVES: After studying this lecture, the student should understand: 1. The differences between pressures in the pulmonary and systemic circulations. 2. How

More information

TV = Tidal volume (500ml) IRV = Inspiratory reserve volume (3,000 ml)

TV = Tidal volume (500ml) IRV = Inspiratory reserve volume (3,000 ml) By: Amin alajlouni Lec: 2nd record Date: 29/10/2017 First of all, this is my first sheet so excuse any mistakes I might make and let's start: As we said before in our last lecture about lung capacities

More information

Breathing Process: Inhalation

Breathing Process: Inhalation Airway Chapter 6 Breathing Process: Inhalation Active part of breathing Diaphragm and intercostal muscles contract, allowing the lungs to expand. The decrease in pressure allows lungs to fill with air.

More information

Department of Biology Work Sheet Respiratory system,9 class

Department of Biology Work Sheet Respiratory system,9 class I. Name the following : Department of Biology Work Sheet Respiratory system,9 class 1. A muscular sheet separating the thoracic and abdominal cavities. 2. A respiratory tube supported by cartilaginous

More information

RESPIRATORY GAS EXCHANGE

RESPIRATORY GAS EXCHANGE RESPIRATORY GAS EXCHANGE Alveolar PO 2 = 105 mmhg; Pulmonary artery PO 2 = 40 mmhg PO 2 gradient across respiratory membrane 65 mmhg (105 mmhg 40 mmhg) Results in pulmonary vein PO 2 ~100 mmhg Partial

More information

Respiratory Physiology. Adeyomoye O.I

Respiratory Physiology. Adeyomoye O.I Respiratory Physiology By Adeyomoye O.I Outline Introduction Hypoxia Dyspnea Control of breathing Ventilation/perfusion ratios Respiratory/barometric changes in exercise Intra-pulmonary & intra-pleural

More information

BIOLOGY 12: UNIT J - CHAPTER 15 - REVIEW WORKSHEET RESPIRATORY SYSTEM

BIOLOGY 12: UNIT J - CHAPTER 15 - REVIEW WORKSHEET RESPIRATORY SYSTEM BIOLOGY 12: UNIT J - CHAPTER 15 - REVIEW WORKSHEET RESPIRATORY SYSTEM A. CHAPTER REVIEW 1. Define the four components of respiration. 2. What happens to the air as it moves along the air passages? What

More information

Human gas exchange. Question Paper. Save My Exams! The Home of Revision. Cambridge International Examinations. 56 minutes. Time Allowed: Score: /46

Human gas exchange. Question Paper. Save My Exams! The Home of Revision. Cambridge International Examinations. 56 minutes. Time Allowed: Score: /46 Human gas exchange Question Paper Level Subject Exam oard Topic Sub Topic ooklet O Level iology ambridge International Examinations Respiration Human gas exchange Question Paper Time llowed: 56 minutes

More information

Anesthesia Delivery Systems What Do I Need to Know About My Machine? A Workshop James H. Philip, MEE, MD, CCE

Anesthesia Delivery Systems What Do I Need to Know About My Machine? A Workshop James H. Philip, MEE, MD, CCE Anesthesia Delivery Systems What Do I Need to Know About My Machine? A Workshop James H. Philip, MEE, MD, CCE Anesthesiologist and Director Bioengineering Department of Anesthesiology, Pain, and Peri-operative

More information

Keeping anaesthetic pollution to minimum in theatre environment

Keeping anaesthetic pollution to minimum in theatre environment Vet Times The website for the veterinary profession https://www.vettimes.co.uk Keeping anaesthetic pollution to minimum in theatre environment Author : Gianluca Bini Categories : General, Vets Date : September

More information

BREATHING AND EXCHANGE OF GASES

BREATHING AND EXCHANGE OF GASES 96 BIOLOGY, EXEMPLAR PROBLEMS CHAPTER 17 BREATHING AND EXCHANGE OF GASES MULTIPLE CHOICE QUESTIONS 1. Respiration in insects is called direct because a. The tissues exchange O 2 directly with the air in

More information

Chapter 17 The Respiratory System: Gas Exchange and Regulation of Breathing

Chapter 17 The Respiratory System: Gas Exchange and Regulation of Breathing Chapter 17 The Respiratory System: Gas Exchange and Regulation of Breathing Overview of Pulmonary Circulation o Diffusion of Gases o Exchange of Oxygen and Carbon Dioxide o Transport of Gases in the Blood

More information

throughout. The constant-flow respiration was administered through a intravenously at appropriate intervals (in addition to the general

throughout. The constant-flow respiration was administered through a intravenously at appropriate intervals (in addition to the general 414 6I2.22I:6I2.2I5.5 GASEOUS INTERCHANGES THROUGH THE VISCERAL PLEURA OF THE CAT. By M. KREMER, A. T. WILSON AND SAMSON WRIGHT. (Department of Physiology, Middlesex Hospital Medical School.) (Received

More information

2) an acute situation in which hypoxemia is suspected.

2) an acute situation in which hypoxemia is suspected. I. Subject: Oxygen Therapy II. Policy: Oxygen therapy shall be initiated upon a physician's order by health care professionals trained in the set-up and principles of safe oxygen administration. Oxygen

More information

LUNG CLEARANCE INDEX. COR-MAN IN Issue A, Rev INNOVISION ApS Skovvænget 2 DK-5620 Glamsbjerg Denmark

LUNG CLEARANCE INDEX. COR-MAN IN Issue A, Rev INNOVISION ApS Skovvænget 2 DK-5620 Glamsbjerg Denmark LUNG CLEARANCE INDEX METHOD COR-MAN-0000-008-IN Issue A, Rev. 3 2013-07-01 INNOVISION ApS Skovvænget 2 DK-5620 Glamsbjerg Denmark Tel.: +45 65 95 91 00 Fax: +45 65 95 78 00 info@innovision.dk www.innovision.dk

More information

ekom Compressor for Medical Air Sonata with Ekom

ekom Compressor for Medical Air Sonata with Ekom Compressor for Medical Air The compressor A40 is an ideal accessory for the intensive ventilators elisa and sonata. The A40 can be used either as the main or as a back-up gas supply. In the event of a

More information